K233253 · Agilemd, Inc. · QNL · Jun 21, 2024 · Cardiovascular
Device Facts
Record ID
K233253
Device Name
eCARTv5 Clinical Deterioration Suite (“eCART”)
Applicant
Agilemd, Inc.
Product Code
QNL · Cardiovascular
Decision Date
Jun 21, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2210
Device Class
Class 2
Attributes
AI/ML, Software as a Medical Device, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K233253 · Jun 21, 2024
eCARTv5 Clinical Deterioration Suite (“eCART”)
Agilemd, Inc.
Electronic Health Records (EHR); Routine clinical vital signs; Laboratory data; Patient demographics
Retrospective and prospective validation of the eCARTv5 algorithm performance (AUROC, sensitivity, specificity, PPV, NPV) for predicting clinical deterioration and mortality in adult ward patients across three health systems.
eCARTv5 Retrospective and Prospective Validation; Retrospective and prospective observational cohort study; Follow-up/Duration: Deterioration/Mortality within 24 hours of score; Study Period: Retrospective: 2009-2023; Prospective: 2023-2024
Adult ward patients (18+ years) across three geographically distinct health systems; Sample Size: Retrospective: 1,769,461 encounters (934,454 patients); Prospective: 205,946 encounters (151,233 patients); Number of Sites: 3
Not applicable for this study
Composite outcome of death or ICU transfer (clinical deterioration); Mortality
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
Clinical Deterioration
Gradient Boosted Machine
—
AUROC 0.828 (Prospective)
Retrospective analysis of admissions between 2009 and 2023 in three health systems (N=1,769,461)
—
Prospective validation in three healthcare systems including non-overlapping admissions between 2023 and 2024 (N=205,946)
—
Mortality
Gradient Boosted Machine
—
AUROC 0.913 (Prospective)
Retrospective analysis of admissions between 2009 and 2023 in three health systems (N=1,769,461)
—
Prospective validation in three healthcare systems including non-overlapping admissions between 2023 and 2024 (N=205,946)
—
Indications for Use
eCART is a software product that provides automated risk stratification and early warning for impending patient deterioration, signified as the composite outcome of death or ICU transfer. It is intended to be used on hospitalized ward patients 18 years of age or older by trained medical professionals. As a clinical decision support device, eCART's risk score and trend analysis is intended to aid clinical teams in identifying which patients are most likely to clinically deteriorate. eCART provides additional information and does not replace the standard of care or clinical judgment. eCART scoring is initiated by the documentation of any vital sign on an adult ward patient. The device calculates risk only from validated EHR data, such as vitals that have been confirmed by a registered nurse (RN); unvalidated data streaming from monitors/devices will not be used until confirmed by a healthcare professional. The product predictions are for reference only and no therapeutic decisions should be made based solely on the eCART scores.
Device Story
eCARTv5 is a cloud-based clinical decision support software integrated into hospital EHRs; it processes routine vital signs, laboratory data, and patient demographics to predict clinical deterioration (death or ICU transfer) in adult ward patients. The system uses a Gradient Boosted Machine (GBM) learning algorithm to generate a real-time risk score (0-100) and trend analysis. Outputs are displayed in an intuitive UI for med-surg nursing staff and physicians; the score indicates the odds of deterioration within 24 hours. The device flags at-risk patients to guide clinical teams in following institutional workflows for monitoring and evaluation. It does not replace clinical judgment or standard of care; therapeutic decisions must not be based solely on eCART scores. The system requires validated EHR data confirmed by healthcare professionals; unvalidated streaming data is excluded. Benefits include early identification of patients at risk of deterioration, potentially enabling timely intervention.
Clinical Evidence
Clinical evidence includes retrospective (2009-2023, N=1,769,461) and prospective (2023-2024, N=205,946) validation studies across three health systems. Primary endpoints were death or ICU transfer within 24 hours. Retrospective AUROC for deterioration was 0.835; prospective AUROC was 0.828. Mortality AUROC was 0.923 (retrospective) and 0.913 (prospective). Subgroup analyses for comorbidities (CHF, COVID-19, COPD, Sepsis) and race demonstrated consistent performance. Sensitivity and specificity metrics were provided for moderate (≥93) and high (≥97) risk thresholds.
Technological Characteristics
Cloud-based software integrated with hospital EHR. Uses Gradient Boosted Machine (GBM) learning algorithm. Inputs: vital signs, lab data, demographics. Output: real-time risk score (0-100) and color-coded risk categorization. Connectivity: EHR interface. No physical materials or energy sources.
Indications for Use
Indicated for hospitalized ward patients 18 years of age or older to provide automated risk stratification and early warning for impending patient deterioration (death or ICU transfer). Used by trained medical professionals as clinical decision support.
Regulatory Classification
Identification
The adjunctive predictive cardiovascular indicator is a prescription device that uses software algorithms to analyze cardiovascular vital signs and predict future cardiovascular status or events. This device is intended for adjunctive use with other physical vital sign parameters and patient information and is not intended to independently direct therapy.
Special Controls
*Classification.* Class II (special controls). The special controls for this device are:(1) A software description and the results of verification and validation testing based on a comprehensive hazard analysis and risk assessment must be provided, including:
(i) A full characterization of the software technical parameters, including algorithms;
(ii) A description of the expected impact of all applicable sensor acquisition hardware characteristics and associated hardware specifications;
(iii) A description of sensor data quality control measures;
(iv) A description of all mitigations for user error or failure of any subsystem components (including signal detection, signal analysis, data display, and storage) on output accuracy;
(v) A description of the expected time to patient status or clinical event for all expected outputs, accounting for differences in patient condition and environment; and
(vi) The sensitivity, specificity, positive predictive value, and negative predictive value in both percentage and number form.
(2) A scientific justification for the validity of the predictive cardiovascular indicator algorithm(s) must be provided. This justification must include verification of the algorithm calculations and validation using an independent data set.
(3) A human factors and usability engineering assessment must be provided that evaluates the risk of misinterpretation of device output.
(4) A clinical data assessment must be provided. This assessment must fulfill the following:
(i) The assessment must include a summary of the clinical data used, including source, patient demographics, and any techniques used for annotating and separating the data.
(ii) The clinical data must be representative of the intended use population for the device. Any selection criteria or sample limitations must be fully described and justified.
(iii) The assessment must demonstrate output consistency using the expected range of data sources and data quality encountered in the intended use population and environment.
(iv) The assessment must evaluate how the device output correlates with the predicted event or status.
(5) Labeling must include:
(i) A description of what the device measures and outputs to the user;
(ii) Warnings identifying sensor acquisition factors that may impact measurement results;
(iii) Guidance for interpretation of the measurements, including a statement that the output is adjunctive to other physical vital sign parameters and patient information;
(iv) A specific time or a range of times before the predicted patient status or clinical event occurs, accounting for differences in patient condition and environment;
(v) Key assumptions made during calculation of the output;
(vi) The type(s) of sensor data used, including specification of compatible sensors for data acquisition;
(vii) The expected performance of the device for all intended use populations and environments; and
(viii) Relevant characteristics of the patients studied in the clinical validation (including age, gender, race or ethnicity, and patient condition) and a summary of validation results.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 21, 2024
AgileMD, Inc. % Kelliann Payne Partner Hogan Lovells US LLP 1735 Market Street 23rd Floor Philadelphia, Pennsylvania 19103
Re: K233253
Trade/Device Name: eCARTy5 Clinical Deterioration Suite ("eCART") Regulation Number: 21 CFR 870.2210 Regulation Name: Adjunctive Predictive Cardiovascular Indicator Regulatory Class: Class II Product Code: ONL Dated: May 17, 2024 Received: May 17, 2024
## Dear Kelliann Payne:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Robert T. Kazmierski -S
for
1. d. F
Page 2
LCDR Stephen Browning Assistant Director
{2}------------------------------------------------
Division of Cardiac Electrophysiology, Diagnostics, and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
510(k) Number (if known) K233253
Device Name
eCART
Indications for Use (Describe)
eCART is a software product that provides automated risk stratification and early warning for impending patient deterioration, signified as the composite outcome of death or ICU transfer. It is intended to be used on hospitalized ward patients 18 years of age or older by trained medical professionals.
As a clinical decision support device, eCART's risk score and trend analysis is intended to aid clinical teams in identifying which patients are most likely to clinically deteriorate. eCART provides additional information and does not replace the standard of care or clinical judgment.
eCART scoring is initiated by the documentation of any vital sign on an adult ward patient. The device calculates risk only from validated EHR data, such as vitals that have been confirmed by a registered nurse (RN); unvalidated data streaming from monitors/devices will not be used until confirmed by a healthcare professional. The product predictions are for reference only and no therapeutic decisions should be made based solely on the eCART scores.
区 Prescription Use (Part 21 CFR 801 Subpart D) □ Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number.
{4}------------------------------------------------
#### 510(k) SUMMARY
#### eCARTv5 Clinical Deterioration Suite
#### Submitter
AgileMD, Inc. 2261 Market Street #4378 San Francisco CA, 94114 Phone: 415-650-0522 Contact Person: Borna Safabakhsh
Date Prepared: June 20, 2024
Name of Device: eCARTv5 Clinical Deterioration Suite ("eCART")
Common or Usual Name: Clinical Monitor
Classification Regulation: 21 CFR 870.2210 (Adjunctive Predictive Cardiovascular Indicator)
Regulatory Class: II
Product Code: QNL
Predicate Device: CLEWICU (K200717)
Reference Device: PeraServer and PeraTrend (K172959)
#### Device Description
The AgileMD eCARTv5 Clinical Deterioration Suite ("eCART") is a cloud-based software device that is integrated into the electronic health record ("EHR") in order to anticipate clinical deterioration in adult ward patients, which is signified as either of the following two predicted outcomes: (1) death or (2) ICU transfer. The tool synthesizes routine vital signs, laboratory data, and patient demographics into a single value that can be used to flag patients at-risk of the composite outcome of clinical deterioration for additional evaluation and monitoring. eCARTv5 requires the healthcare system within which it will be used, to provide an EHR connection and data interfaces through which the patient data necessary to run the software will be transmitted.
The primary functions of the system are imparted by the Gradient Boosted Machine ("GBM") learning algorithm that takes input directly from the EHR, in real time, to provide an assessment of patients and displays its outputs in an intuitive user interface which drives providers to follow standardized clinical workflows (established by their institutions) for elevated-risk patients.
eCARTv5's end users include med-surg nursing staff, physicians and other providers caring for these patients. The eCARTv5 composite score is determined from the model output (predicted probability of deterioration) scaled from 0-100, based on the specificity (true negative rate). The observed rate of deterioration at each eCART score threshold, displayed as the odds of deterioration in the next 24 hours, is presented to the user along with the scaled score. Default thresholds are set to an eCART of 93 and 97, respectively, for moderate and high risk categorization.
{5}------------------------------------------------
### Intended Use / Indications for Use
eCART is a software product that provides automated risk stratification and early warning for impending patient deterioration, signified as the composite outcome of death or ICU transfer. It is intended to be used on hospitalized ward patients 18 years of age or older by trained medical professionals.
As a clinical decision support device, eCART's risk score and trend analysis is intended to aid clinical teams in identifying which patients are most likely to clinically deteriorate. eCART provides additional information and does not replace the standard of care or clinical judgment.
eCART scoring is initiated by the documentation of any vital sign on an adult ward patient. The device calculates risk only from validated EHR data, such as vitals that have been confirmed by a registered nurse (RN); unvalidated data streaming from monitors/devices will not be used until confirmed by a healthcare professional. The product predictions are for reference only and no therapeutic decisions should be made based solely on the eCART scores.
The differences in indications for use from the predicate device are not critical to the intended use of eCARTv5, nor do they raise different questions of safety or effectiveness when the subject device is used as labeled. The minor differences are supported by adequate performance testing that show the subject device is substantially equivalent for the proposed indications for use. Both devices are intended to provide trained healthcare providers with a patient status score that reflects the underlying patient condition to supplement standard of care and informed decision making.
## Summary of Technological Characteristics
At a high level, the subject and predicate devices are based on the following same technological elements:
- · A risk-predictive output generated using a machine-learning algorithm.
- Inputs to the software device include vital signs, assessments, and laboratory data collected . from the hospital EHR system.
- . The output is a color-defined (red-yellow), real-time risk categorization to provide users information regarding potential risk of patient deterioration.
- Embedded workflow features enable users to acknowledge the risk prediction and record next . steps for medical management, based on their independent medical judgment.
The primary technological differences between the devices are that eCART's algorithm was trained on ward patients to predict the probability of ICU transfer or death, whereas the predicate is trained in ICU patients to predict respiratory failure and/or hemodynamic instability. Additionally, the predicate device does not provide a risk score (but rather, a risk category). The minor technological differences between the devices raise no different questions of safety or effectiveness. With respect to the output, the reference device (PeraServer and PeraTrend (K172959) provides a risk score (the Rothman Index), which was validated on the outcome of mortality within 24 hours and included training and testing on data from adult ward patients (similar to eCART). Performance data for eCART, including retrospective and prospective clinical testing as well as human factors validation, demonstrates that the subject device is substantially equivalent to the predicate device.
## Performance Data
eCART was assessed in retrospective and prospective validation studies with adult ward patients from three geographically distinct health systems. Test characteristics for both cohorts are shown in
{6}------------------------------------------------
Table A below. The retrospective analysis included admissions between 2009 and 2023 in three health systems. Deterioration is defined as death or ward to ICU transfer within 24 hours following a score. The mortality outcome is defined as death within 24 hours following a score. The prospective validation was performed in the same three healthcare systems where the retrospective analysis was undertaken. This analysis included non-overlapping admissions between 2023 and 2024.
Encounters (N) are defined as unique hospitalizations for a single patients may have more than one encounter in the data set if they were admitted and discharged from a study hospital more than once during the study period. Each encounter includes all eCART scores (observations, n) generated during that hospitalization.
| | | DETERIORATION | | MORTALITY | |
|------------------------------------|--------------------|-------------------------|-----------------------|-------------------------|-----------------------|
| | | Retrospective | Prospective | Retrospective | Prospective |
| | | N=1,769,461 | N=205,946 | N=1,769,461 | N=205,946 |
| AUROC | | 0.835 (0.834, 0.835) | 0.828 (0.827, 0.829) | 0.923 (0.923, 0.924) | 0.913 (0.911, 0.914) |
| | | n=132,873,833 | n=21,516,964 | n=132,873,833 | n=21,516,964 |
| Outcome Prevalence | | 1.3% | 1.3% | 0.2% | 0.2% |
| | | 1,744,044/132,873,833 | 284,678/21,516,964 | 249,636/132,873,833 | 39,609/21,516,964 |
| Moderate-<br>risk<br>threshold | Positivity<br>Rate | 7.5% | 7.3% | 7.5% | 7.3% |
| | | 9,982,577/132,873,833 | 1,563,648/21,516,964 | 9,982,577/132,873,833 | 1,563,648/21,516,964 |
| | Sensitivity | 51.8% (51.7%, 51.8%) | 48.8% (48.7%, 49.0%) | 75.8% (75.6%, 75.9%) | 71.7% (71.3%, 72.1%) |
| | | 902,951/1,744,044 | 139,043/284,678 | 189,171/249,636 | 28,399/39,609 |
| | Specificity | 93.1% (93.1%, 93.1%) | 93.3% (93.3%, 93.3%) | 92.6% (92.6%, 92.6%) | 92.9% (92.8%, 92.9%) |
| | | 122,050,163/131,129,789 | 19,807,681/21,232,286 | 122,830,791/132,624,197 | 19,942,106/21,477,355 |
| eCART≥93 | PPV | 9.0% (9.0%, 9.1%) | 8.9% (8.8%, 8.9%) | 1.9% (1.9%, 1.9%) | 1.8% (1.8%, 1.8%) |
| | | 902,951/9,982,577 | 139,043/1,563,648 | 189,171/9,982,577 | 28,399/1,563,648 |
| | NPV | 99.3% (99.3%, 99.3%) | 99.3% (99.3%, 99.3%) | 100.0% (100.0%, 100.0%) | 99.9% (99.9%, 99.9%) |
| | | 122,050,163/122,891,256 | 19,807,681/19,953,316 | 122,830,791/122,891,256 | 19,942,106/19,953,316 |
| | Risk Diff | 8.4% (8.3%, 8.4%) | 8.2% (8.1%, 8.2%) | 1.8% (1.8%, 1.9%) | 1.8% (1.7%, 1.8%) |
| High-risk<br>threshold<br>eCART≥97 | Positivity<br>Rate | 3.6% | 3.1% | 3.6% | 3.1% |
| | | 4,753,332/132,873,833 | 675,279/21,516,964 | 4,753,332/132,873,833 | 675,279/21,516,964 |
| | Sensitivity | 38.6% (38.5%, 38.7%) | 33.7% (33.6%, 33.9%) | 64.7% (64.5%, 64.9%) | 58.1% (57.6%, 58.6%) |
| | | 673,539/1,744,044 | 96,051/284,678 | 161,531/249,636 | 23,025/39,609 |
| | Specificity | 96.9% (96.9%, 96.9%) | 97.3% (97.3%, 97.3%) | 96.5% (96.5%, 96.5%) | 97.0% (97.0%, 97.0%) |
| | | 127,049,996/131,129,789 | 20,653,058/21,232,286 | 128,032,396/132,624,197 | 20,825,101/21,477,355 |
| | PPV | 14.2% (14.1%, 14.2%) | 14.2% (14.1%, 14.3%) | 3.4% (3.4%, 3.4%) | 3.4% (3.4%, 3.5%) |
| | | 673,539/4,753,332 | 96,051/675,279 | 161,531/4,753,332 | 23,025/675,279 |
| | NPV | 99.2% (99.2%, 99.2%) | 99.1% (99.1%, 99.1%) | 99.2% (99.2%, 99.2%) | 99.9% (99.9%, 99.9%) |
| | | 127,049,996/128,120,501 | 20,653,058/20,841,685 | 127,049,996/128,120,501 | 20,825,101/20,841,685 |
| | Risk Diff | 13.3% (13.3%, 13.4%) | 13.3% (13.2%, 13.4%) | 3.3% (3.3%, 3.3%) | 3.3% (3.3%, 3.4%) |
## Table A. eCART Prediction of Deterioration and Mortality in the Full External Retrospective and Prospective Cohorts
Note: The encounter and observation data above represents 934,454 and 151,233 unique patients in the retrospective and prospective cohorts, respectively. 95% confidence intervals were calculated using the Clopper-Pearson method.
Subgroup analyses for select comorbidities are presented in Tables B-C below for deterioration and mortality, respectively, and show comparable performance across conditions. Sequencing data for the SARS-CoV-2 variants included in the COVID subgroup analyses was not available, but the sample likely included the common variants circulating in the US between 2020 and the early part of 2024 (see https://cov-lineages.org/lineage list.html). The comparative effectiveness of the model between existing SARS-CoV-2 strains or to future strains is unknown.
{7}------------------------------------------------
| | | Comorbidity | | | |
|--------------------|--------------------|--------------------------------------------|------------------------|---------------------------------------------|-----------------------|
| | DETERIORATION | Congestive Heart<br>Failure<br>N = 306,140 | COVID-19<br>N = 49,834 | Chronic Pulmonary<br>Disease<br>N = 443,263 | Sepsis<br>N = 639,802 |
| | AUROC | 0.810 (0.809, 0.810) | 0.858 (0.857, 0.859) | 0.824 (0.823, 0.824) | 0.836 (0.836, 0.836) |
| | | n = 31,445,409 | n = 5,961,084 | n = 38,638,180 | n = 72,055,421 |
| | Outcome Prevalence | 2.2% | 2.3% | 1.7% | 1.8% |
| | | 695,781/31,445,409 | 136,077/5,961,084 | 646,095/38,638,180 | 1,295,361/72,055,421 |
| | Sensitivity | 53.9% (53.8%, 54.1%) | 69.6% (69.4%, 69.9%) | 53.9% (53.8%, 54.0%) | 56.4% (56.3%, 56.5%) |
| | | 375,328/695,781 | 94,726/136,077 | 348,271/646,095 | 730,846/1,295,361 |
| Moderate- | Specificity | 89.8% (89.7%, 89.8%) | 84.6% (84.6%, 84.7%) | 91.1% (91.0%, 91.1%) | 90.8% (90.8%, 90.8%) |
| risk | | 27,600,037/30,749,628 | 4,929,294/5,825,007 | 34,593,348/37,992,085 | 64,243,126/70,760,060 |
| threshold | PPV | 10.6% (10.6%, 10.7%) | 9.6% (9.5%, 9.6%) | 9.3% (9.3%, 9.3%) | 10.1% (10.1%, 10.1%) |
| (≥93) | | 375,328/3,524,919 | 94,726/990,439 | 348,271/3,747,008 | 730,846/7,247,780 |
| | NPV | 98.9% (98.8%, 98.9%) | 99.2% (99.2%, 99.2%) | 99.1% (99.1%, 99.1%) | 99.1% (99.1%, 99.1%) |
| | | 27,600,037/27,920,490 | 4,929,294/4,970,645 | 34,593,348/34,891,172 | 64,243,126/64,807,641 |
| | Sensitivity | 40.2% (40.1%, 40.3%) | 55.9% (55.6%, 56.2%) | 40.1% (40.0%, 40.3%) | 42.6% (42.6%, 42.7%) |
| | | 279,748/695,781 | 76,086/136,077 | 259,354/646,095 | 552,417/1,295,361 |
| | Specificity | 95.2% (95.1%, 95.2%) | 90.7% (90.6%, 90.7%) | 95.8% (95.8%, 95.9%) | 95.6% (95.6%, 95.6%) |
| High-risk | | 29,259,307/30,749,628 | 5,280,458/5,825,007 | 36,413,561/37,992,085 | 67,648,075/70,760,060 |
| threshold<br>(≥97) | PPV | 15.8% (15.8%, 15.9%) | 12.3% (12.2%, 12.3%) | 14.1% (14.1%, 14.2%) | 15.1% (15.1%, 15.1%) |
| | | 279,748/1,770,069 | 76,086/620,635 | 259,354/1,837,878 | 552,417/3,664,402 |
| | NPV | 98.6% (98.6%, 98.6%) | 98.9% (98.9%, 98.9%) | 98.9% (98.9%, 99.0%) | 98.9% (98.9%, 98.9%) |
| | | 29,259,307/29,675,340 | 5,280,458/5,340,449 | 36,413,561/36,800,302 | 67,648,075/68,391,019 |
### Table B. eCART Prediction of Clinical Deterioration in Select Comorbidities in the Full External Retrospective Cohort
Note: The encounter and observation data above represents 136,197 unique patients with congestive heart failure, 44,231 with chronic obstructive pulmonary disease, and 385,997 with sepsis were calculated using the Cloper-Pearson method.
#### Table C. eCART Prediction of Mortality in Select Comorbidities in the Full External Retrospective Cohort
| MORTALITY | Comorbidity | | | |
|------------------------------------------------------|--------------------------------------------------------------|------------------------------------------------------------|--------------------------------------------------------------|--------------------------------------------------------------|
| | Congestive Heart<br>Failure | COVID-19<br>N = 49,834 | Chronic Pulmonary<br>Disease<br>N = 443,263 | Sepsis<br>N = 639,802 |
| | N = 306,140 | | | |
| | | | | |
| AUROC | 0.893 (0.892, 0.894) | 0.924 (0.923, 0.926) | 0.912 (0.911, 0.913) | 0.914 (0.914, 0.915) |
| | n = 31,445,409 | n = 5,961,084 | n = 38,638,180 | n = 72,055,421 |
| Outcome Prevalence | 0.3%<br>103,194/31,445,09 | 0.5%<br>30,714/5,961,084 | 0.2%<br>85,582/38,638,180 | 0.2%<br>160,348/72,055,421 |
| Moderate-<br>risk<br>threshold<br>(≥93) | Sensitivity<br>73.6% (73.3%, 73.9%)<br>75,961/103,194 | Sensitivity<br>86.7% (86.3%, 87.1%)<br>26,639/30,714 | Sensitivity<br>76.1% (75.8%, 76.4%)<br>65,111/85,582 | Sensitivity<br>77.% (76.8%, 77.2%)<br>123,467/160,348 |
| | Specificity<br>89.0% (89.0%, 89.0%)<br>27,893,257/31,342,215 | Specificity<br>83.7% (83.7%, 83.8%)<br>4,966,570/5,930,370 | Specificity<br>90.4% (90.4%, 90.5%)<br>34,870,701/38,552,598 | Specificity<br>90.1% (90.1%, 90.1%)<br>64,770,760/71,895,073 |
| PPV<br>2.2% (2.1%, 2.2%)<br>75,961/3,524,919 | PPV<br>2.7% (2.7%, 2.7%)<br>26,639/990,439 | PPV<br>1.7% (1.7%, 1.8%)<br>65,111/3,747,008 | PPV<br>1.7% (1.7%, 1.7%)<br>123,467/7,247,780 | |
| NPV<br>99.9% (99.9%, 99.9%)<br>27,893,257/27,920,490 | NPV<br>99.9% (99.9%, 99.9%)<br>4,966,570/4,970,645 | NPV<br>99.9% (99.9%, 99.9%)<br>34,870,701/34,891,172 | NPV<br>99.9% (99.9%, 99.9%)<br>64,770,760/64,807,641 | |
| High-risk<br>threshold<br>(≥97) | Sensitivity<br>61.7% (61.4%, 62.0%)<br>63,697/103,194 | Sensitivity<br>79.2% (78.7%, 79.6%)<br>24,315/30,714 | Sensitivity<br>64.9% (64.5%, 65.2%)<br>55,507/85,582 | Sensitivity<br>66.1% (65.8%, 66.3%)<br>105,915/160,348 |
| | Specificity<br>94.6% (94.5%, 94.6%)<br>29,635,843/31,342,215 | Specificity<br>89.9% (89.9%, 90.0%)<br>5,334,050/5,930,370 | Specificity<br>95.4% (95.4%, 95.4%)<br>36,770,227/38,552,598 | Specificity<br>95.1% (95.%, 95.1%)<br>68,336,586/71,895,073 |
| PPV<br>3.6% (3.6%, 3.6%)<br>63,697/1,770,069 | PPV<br>3.9% (3.9%, 4.0%)<br>24,315/620,635 | PPV<br>3.0% (3.0%, 3.0%)<br>55,507/1,837,878 | PPV<br>2.9% (2.9%, 2.9%)<br>105,915/3,664,402 | |
| NPV<br>99.9% (99.9%, 99.9%)<br>29,635,843/29,675,340 | NPV<br>99.9% (99.9%, 99.9%)<br>5,334,050/5,340,449 | NPV<br>99.9% (99.9%, 99.9%)<br>36,770,227/36,800,302 | NPV<br>99.9% (99.9%, 99.9%)<br>68,336,586/68,391,019 | |
Note: The encounter and observation data above represents 136,197 unique patients with congestive heart failure, 44,8231 with chronic obstructive pulmonary disease, and 385,997 with sepss. 95% confidence intervals were calculated using the Cloper-Pearson method.
{8}------------------------------------------------
Subgroup analyses for race are presented in Table D, below, for deterioration. All five groups met the performance thresholds for AUROC and sensitivity.
| | | | Table D. eCART Retrospective Validation Test Characteristics for the Outcomes of Clinical | | | |
|-----------------------------------------|-----------------------|--------------------------------------------------|-------------------------------------------------------------------------------------------|-----------------------------------------------|-----------------------------------------------------------|------------------------------------------------|
| | Deterioration by Race | | | | | |
| DETERIORATION | Race | American Indian or<br>Alaska Native<br>N = 6,468 | Asian/Mideast Indian<br>N = 26,681 | Black/African-<br>American<br>N = 252,982 | Native<br>Hawaiian/Other<br>Pacific Islander<br>N = 2,496 | White/Caucasian<br>N = 1,384,075 |
| AUROC | | 0.814 (0.808, 0.820)<br>486,615 | 0.847 (0.844, 0.850)<br>1,694,681 | 0.831 (0.830, 0.832)<br>20,057,696 | 0.862 (0.854, 0.870)<br>177,513 | 0.834 (0.833, 0.834)<br>104,522,118 |
| Outcome Prevalence | | 1.2%<br>6,026/486,615 | 1.2%<br>20,736/1,694,681 | 1.2%<br>234,979/20,057,696 | 1.3%<br>2,395/177,513 | 1.3%<br>1,403,867/104,522,118 |
| Moderate-<br>risk<br>threshold<br>(≥93) | Pos<br>Rate | 6.7%<br>32,455/486,615 | 7.2%<br>121,532/1,694,681 | 6.5%<br>1,301,626/20,057,696 | 7.2%<br>12,738/177,513 | 7.7%<br>8,085,868/104,522,118 |
| | Sensitivity | 46.6% (45.3%, 47.9%)<br>2,809/6,026 | 52.9% (52.2%, 53.6%)<br>10,968/20,736 | 48.1% (47.9%, 48.3%)<br>112,939/234,979 | 56.5% (54.5%, 58.5%)<br>1,353/2,395 | 52.3% (52.2%, 52.3%)<br>733,566/1,403,867 |
| | Specificity | 93.8% (93.8%, 93.9%)<br>450,943/480,589 | 93.4% (93.4%, 93.4%)<br>1,563,381/1,673,945 | 94.0% (94.0%, 94.0%)<br>18,634,030/19,822,717 | 93.5% (93.4%, 93.6%)<br>163,733/175,118 | 92.9% (92.9%, 92.9%)<br>95,765,949/103,118,251 |
| | PPV | 8.7% (8.4%, 9.0%)<br>2,809/32,455 | 9.0% (8.9%, 9.2%)<br>10,968/121,532 | 8.7% (8.6%, 8.7%)<br>112,939/1,301,626 | 10.6% (10.1%, 11.2%)<br>1,353/12,738 | 9.1% (9.1%, 9.1%)<br>733,566/8,085,868 |
| | NPV | 99.3% (99.3%, 99.3%)<br>450,943/454,160 | 99.4% (99.4%, 99.4%)<br>1,563,381/1,573,149 | 99.3% (99.3%, 99.4%)<br>18,634,030/18,756,070 | 99.4% (99.3%, 99.4%)<br>163,733/164,775 | 99.3% (99.3%, 99.3%)<br>95,765,949/96,436,250 |
| | Risk Diff | 7.9% (7.6%, 8.3%) | 8.4% (8.2%, 8.6%) | 8.0% (8.0%, 8.1%) | 10.0% (9.5%, 10.5%) | 8.4% (8.4%, 8.4%) |
| High-risk<br>threshold<br>(≥97) | Pos<br>Rate | 3.1%<br>14,933/486,615 | 3.6%<br>60,178/1,694,681 | 3.1%<br>611,210/20,057,696 | 3.5%<br>6,174/177,513 | 3.7%<br>3,831,413/104,522,118 |
| | Sensitivity | 35.2% (34.0%, 36.4%)<br>2,122/6,026 | 40.9% (40.3%, 41.6%)<br>8,491/20,736 | 34.9% (34.7%, 35.1%)<br>81,982/234,979 | 40.0% (38.0%, 42.0%)<br>957/2,395 | 39.1% (39.0%, 39.1%)<br>548,404/1,403,867 |
| | Specificity | 97.3% (97.3%, 97.4%)<br>467,778/480,589 | 96.9% (96.9%, 96.9%)<br>1,622,258/1,673,945 | 97.3% (97.3%, 97.3%)<br>19,293,489/19,822,717 | 97.0% (96.9%, 97.1%)<br>169,901/175,118 | 96.8% (96.8%, 96.8%)<br>99,835,242/103,118,251 |
| | PPV | 14.2% (13.7%, 14.8%)<br>2,122/14,933 | 14.1% (13.8%, 14.4%)<br>8,491/60,178 | 13.4% (13.3%, 13.5%)<br>81,982/611,210 | 15.5% (14.6%, 16.4%)<br>957/6,174 | 14.3% (14.3%, 14.3%)<br>548,404/3,831,413 |
| | NPV | 99.2% (99.1%, 99.2%)<br>467,778/471,682 | 99.3% (99.2%, 99.3%)<br>1,622,258/1,634,503 | 99.2% (99.2%, 99.2%)<br>19,293,489/19,446,486 | 99.2% (99.1%, 99.2%)<br>169,901/171,339 | 99.2% (99.1%, 99.2%)<br>99,835,242/100,690,705 |
| | Risk Diff | 13.4% (12.8%, 13.9%) | 13.4% (13.1%, 13.6%) | 12.6% (12.5%, 12.7%) | 14.7% (13.8%, 15.6%) | 13.5% (13.4%, 13.5%) |
Note: The encounter and observation data above represents 3,282 unique American Indian or Alaska Native patients, 17,657 Asian/Mideast Asian patients, 116,642 Black/African-American patients, 1,427 Native Hawaiian/Other Pacific Islander patients, 95% confidence intervals were calculated using the Clopper-Pearson method.
# Conclusion
eCARTv5 has the same intended use and similar indications for use, technological characteristics, and principles of operation as its predicate device, CLEWICU. The minor differences in indications for use do not alter the intended fundamental clinical purpose of the device, nor affect its safety and effectiveness when used as labeled. In addition, the minor technological differences between eCARTv5 and its predicate device raise no different questions of safety or effectiveness and are further supported by performance data. Thus, the device is substantially equivalent.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.