K232501 · Exo, Inc. · QIH · Nov 17, 2023 · Radiology
Device Facts
Record ID
K232501
Device Name
AI Platform (AIP001)
Applicant
Exo, Inc.
Product Code
QIH · Radiology
Decision Date
Nov 17, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
AI/ML, Software as a Medical Device, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K232501 · Nov 17, 2023
AI Platform (AIP001)
Exo, Inc.
Routine clinical ultrasound images from multiple clinical sites
Retrospective clinical ultrasound images were used to validate the performance of the AI Platform's cardiac (LVEF) and lung (A-line/B-line) analysis tools against expert-adjudicated reference standards.
The AI Platform is intended for noninvasive processing of ultrasound images to detect, measure, and calculate relevant medical parameters of structures and function of adult patients with suspected disease. The device is intended to be used on images from adult patients.
Device Story
Software as a medical device (SaMD) for ultrasound image analysis; processes DICOM images from ultrasound scanners. Features LVEF AI for cardiac ejection fraction quantification and Lung AI for lung structure/artifact suggestion. Operates via deep convolutional neural networks for segmentation and landmark detection. Used in standard clinical settings by trained healthcare providers to assist in evaluation, quantification, and reporting. Provides tools for viewing, landmark placement, and report generation. Does not replace manual assessment; results intended for review by clinicians to support decision-making. Benefits include workflow simplification and standardized measurement of cardiac/lung parameters.
Clinical Evidence
Performance evaluated on independent test data (n=125 for lung, n=151 for cardiac) from diverse demographics (age 20-96, BMI 15.3-52.8). Cardiac performance (LVEF) measured via ICC and RMSD against expert consensus; overall ICC 0.93 (95% CI 0.91-0.95). Lung performance measured via Cohen's kappa for A-lines (0.84) and ICC for B-lines (0.97). Consistent performance across clinical confounders including heart failure, COPD, and pneumonia.
Technological Characteristics
Software-only device; DICOM-compliant. Implements non-adaptive machine learning (Deep Convolutional Neural Networks) for segmentation and landmark detection. Supports single and multi-frame ultrasound images. Complies with IEC 62304:2006/AC:2015 for software life cycle processes.
Indications for Use
Indicated for noninvasive processing of ultrasound images to detect, measure, and calculate medical parameters of cardiac and lung structures and function in adult patients with suspected disease.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food & Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services seal on the left and the FDA acronym followed by the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a solid blue square and the agency name in a lighter blue. The text reads "FDA U.S. FOOD & DRUG ADMINISTRATION".
Exo Inc Jacqueline Murray Senior Regulatory Affairs Specialist 4201 Burton Drive Santa Clara, CA 95054
November 17, 2023
Re: K232501
Trade/Device Name: AI Platform (AIP001) Regulation Number: 21 CFR 892.2050 Regulation Name: Medical Image Management And Processing System Regulatory Class: Class II Product Code: QIH Dated: October 25, 2023 Received: October 26, 2023
Dear Jacqueline Murray:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrb/cfdocs/cfpmn/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jessica Lamb
Jessica Lamb, Ph.D. Assistant Director Imaging Software Team DHT 8B: Division of Radiological Imaging Devices and Electronic Products OHT 8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K232501
Device Name AI Platform (AIP001)
Indications for Use (Describe)
The AI Platform is intended for noninvasive processing of ultrasound images to detect, measure, and calculate relevant medical parameters of structures and function of adult patients with suspected disease. The device is intended to be used on images from adult patients.
Type of Use (Select one or both, as applicable)
| <div> <span>☑</span> <span>Prescription Use (Part 21 CFR 801 Subpart D)</span> </div> |
|---------------------------------------------------------------------------------------|
| <div> <span>☐</span> <span>Over-The-Counter Use (21 CFR 801 Subpart C)</span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for EXO. The logo consists of a pattern of blue and green dots on the left, followed by the word "EXO" in a dark gray sans-serif font on the right. The dots are arranged in a grid-like pattern, with the color transitioning from blue to green.
# 510(k) Summary
K232501
General Information
| 510(k) Sponsor | Exo Imaging |
|-----------------------|--------------------------------------------|
| Address | 4201 Burton Drive<br>Santa Clara, CA 95054 |
| Correspondence Person | Jacqueline Murray |
| Contact Information | jmurray@exo.inc<br>Cell: +1 236 838-5056 |
| Date Prepared | September 21st, 2023 |
Proposed Device
| Proprietary Name | AI Platform (AIP001) |
|---------------------|--------------------------------------------------|
| Common Name | AI Platform |
| Classification Name | Automated Radiological Image Processing Software |
| Regulation Number | 21 CFR 892.2050 |
| Product Code | QIH |
| Regulatory Class | II |
Predicate Device
| Proprietary Name | LVivo Software Application |
|------------------------|--------------------------------------------------|
| Premarket Notification | K210053 |
| Classification Name | Automated Radiological Image Processing Software |
| Regulation Number | 21 CFR 892.2050 |
| Product Code | QIH |
| Regulatory Class | II |
Reference Device
| Proprietary Name | Lumify Diagnostic Ultrasound System |
|------------------------|------------------------------------------|
| Premarket Notification | K223771 |
| Classification Name | Ultrasonic pulsed doppler imaging system |
| Regulation Number | 21 CFR 892.1550 |
| Product Code | IYN, IYO, ITX, QIH |
| Regulatory Class | II |
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image contains the logo for EXO. On the left side of the logo is a pattern of blue circles arranged in a grid-like fashion. The circles are arranged in a way that they form a larger X shape. To the right of the circles is the word "EXO" in a sans-serif font. The letters are dark gray.
#### Device Description
Exo Al Platform is a software as a medical device (SaMD) that helps qualified users with image-based assessment of ultrasound examinations in adult patients. It is designed to simplify workflow by helping trained healthcare providers evaluate, quantify, and generate reports for ultrasound images. The device is intended to generate images and a report that can be reviewed in a typical standard of care setting.
Al Platform takes as an input imported Digital Imaging and Communications in Medicine (DICOM) images from ultrasound scanners of a specific range and allows users to detect, measure, and calculate relevant medical parameters of structures and function of patients with suspected disease. It provides users with a specific toolset for viewing ultrasound images of the lung and heart, placing landmarks, and creating reports.
Key features of the software are
- LVEF AI: an Al-assisted tool for quantification of ejection on cardiac ultrasound images.
- . Lung Al: an Al-assisted tool to suggest presence of lung structures and artifacts on ultrasound images.
Exo Al Platform does not perform any function that could not be accomplished by a trained user manually. It's important to note that patient management decisions should not be made solely on the results of the Al Platform analysis.
#### Indications for Use
The Al Platform is intended for noninvasive processing of ultrasound images to detect, measure, and calculate relevant medical parameters of structures and function of adult patients with suspected disease. The device is intended to be used on images from adult patients.
#### Comparison of Technological Characteristics with the Predicate Device
| Feature/<br>Function | Subject Device<br>Exo Al Platform | Predicate Device<br>LVivo Software Application<br>(K210053) | Reference Device<br>Lumify Diagnostic<br>Ultrasound System<br>(K223771) |
|--------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------|-------------------------------------------------------------------------|
| Image input | Complies with DICOM<br>Standard | Same as subject device | Same as subject device |
| Scan type | Single and Multi-frame<br>ultrasound images | Same as subject device | Same as subject device |
| Feature/<br>Function | Subject Device<br>Exo Al Platform | Predicate Device<br>LVivo Software Application<br>(K210053) | Reference Device<br>Lumify Diagnostic<br>Ultrasound System<br>(K223771) |
| Image display<br>mode | Static | Same as subject device | Same as subject device |
| Image navigation<br>and manipulation<br>tools | Slice-scroll, pane layout,<br>reset | Same as subject device | Same as subject device |
| Image review | Yes, capable of reviewing<br>all frames of multi-frame<br>(multi-slice) images | Same as subject device | Same as subject device |
| Principle of<br>Operation and<br>Technology | Ultrasound image<br>processing software<br>implementing artificial<br>intelligence including<br>non-adaptive machine<br>learning algorithms trained<br>with clinical data intended<br>for non-invasive analysis of<br>ultrasound data | Same as subject device | Same as subject device |
| Al Algorithm | Deep Convolutional Neural<br>Networks for<br>Segmentation or<br>Landmark Detection | Same as subject device | Same as subject device |
| Manual<br>Adjustments or<br>Editing by User<br>Allowed | Yes | Same as subject device | Same as subject device |
| Anatomical Sites | Heart, Lungs | Heart, Bladder | Lungs |
| Ejection Fraction<br>Measurement<br>Views | AP4, AP2, Bi-plane, PLAX | AP4, AP2, Bi-plane | No |
| Non Cardiac<br>functions | A-lines, B-lines | Bladder Volume | B-lines |
| Report creation | Yes | Same as subject device | Same as subject device |
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image contains a logo with two distinct parts. On the left, there is a cluster of blue and cyan gradient circles arranged in a pattern. To the right of the circles, the letters 'EXO' are written in a bold, dark gray sans-serif font. The overall design is clean and modern.
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows a logo with two distinct parts. On the left, there is a cluster of blue and light blue gradient circles arranged in a pattern resembling a stylized network or constellation. To the right of the circles, the letters 'EXO' are displayed in a bold, sans-serif font, with a dark gray color. The overall design is clean and modern, suggesting a technology-oriented or innovative company.
#### Performance Data
Safety and performance of the AI Platform has been evaluated and verified in accordance with software specifications and applicable performance standards through software verification and validation testing. Additionally, the software validation activities were performed in accordance with IEC 62304:2006/AC:2015 - Medical device software - Software life cycle processes, FDA's 'Content of Premarket Submissions for Device Software Functions'' Guidance for Industry and Food and Drug Administration Staff Document issued on June 14, 2023 and FDA Guidance (June 2022) "Technical performance assessment of quantitative imaging in radiological device premarket submissions".
#### Validation Performance Testing
The clinical performance of the Al platform was successfully evaluated on a test data encompassing diverse demographic variables, including gender, age (ranging from 20 to 96), BMI (ranging from 15.3 to 52.8), and ethnicity from multiple clinical sites in metropolitan cities with diverse racial patient populations. The Lung function was evaluated with 125 subjects, on images acquired during a routine clinical practice from cart-based and portable ultrasound devices (with frequency ranging from 1.5 to 7 MHz). The LVEF function was evaluated with 151 subjects, on images acquired from cart-based and portable ultrasound devices (with frequency ranging from 1.2 to 4 MHz).
The test data was entirely separated from the training/validation datasets and was not used for any part of the training. To ensure data separation and generalizability, the data sources used in the test set are chosen to be different from the data sources used in the training set. We also established auditability measures, by assigning a unique identification number to each study and its corresponding images.
The ground truth for ejection fraction (reference data) was obtained as the average ejection fraction measurement of three experts. Performance was assessed by calculating the intraclass correlation coefficient (ICC) and ejection fraction root mean square difference (RMSD).
The measurement accuracy of Al Platform for cardiac ultrasound images compared with reference data is summarized in Table 1 below:
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the logo for EXO. The logo consists of a pattern of blue-green gradient dots arranged in a triangular shape on the left, followed by the word "EXO" in a dark gray sans-serif font on the right. The dots are arranged in a way that suggests a network or constellation, while the word "EXO" is simple and modern.
| Subgroup (View) | ICC (95% CI) | RMSD (95% CI) |
|------------------------------------------------|--------------------|--------------------|
| Ejection Fraction Parasternal<br>Long-axis | 0.93 (0.89 - 0.96) | 6.12 (5.30 - 8.36) |
| Ejection Fraction Apical Biplane | 0.95 (0.90 - 0.98) | 4.81 (3.99 - 7.25) |
| Ejection Fraction Apical (AP4)<br>Single Plane | 0.92 (0.88 - 0.95) | 6.06 (5.27 - 8.20) |
| Ejection Fraction Apical (AP2)<br>Single Plane | 0.92 (0.87 - 0.95) | 6.25 (5.33 - 8.82) |
| All | 0.93 (0.91 - 0.95) | 5.90 (5.35 - 7.23) |
Table 1: Summary of Al Platform accuracy and reliability for cardiac ultrasound images
The ground truth of the presence of A-line was determined by consensus of two or more experts. Performance was assessed by measuring the agreement using Cohen's kappa coefficient (k), The ground truth of B-line counts was determined as the average of B-line counts from three experts. Performance was assessed by calculating the intraclass correlation coefficient (ICC). The reliability of Al platform for lung ultrasound images compared with reference data is summarized in Table 2 below:
Table 2: Summary of Al Platform reliability for lung ultrasound images
| | Reliability |
|---------|--------------|
| A-lines | Kappa = 0.84 |
| B-lines | ICC = 0.97 |
The device performance was also assessed across a wide range of Ultrasound manufacturer, demographic subgroups, (including gender and BMI) and clinical confounders present including heart failure with reduced ejection fraction, Covid-19, Chronic obstructive pulmonary disease (COPD), Pneumonia, Pulmonary Edema, Coronary artery disease (CAD) and Cardiomyopathy. The evaluation concluded that the device performance was consistent among clinically meaningful subgroups.
### Conclusions
Exo's Al Platform is substantially equivalent in intended use, design, principles of operation, technological characteristics, and safety features to the predicate device. There are no different questions of safety and/or effectiveness introduced by the Al Platform when used as intended.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.