AI/ML, Software as a Medical Device, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K221762 · Nov 16, 2022
BoneMRI v1.4
Mrlguidance B.V.
Retrospective clinical imaging data (MRI and CT scans)
Retrospective clinical data from 73 patients were used to perform a quantitative voxel-by-voxel validation of the BoneMRI software for the lumbar spine region, comparing bone morphology and radiodensity against co-registered CT scans.
73 patients with indications including sacroiliitis, degenerative spine and pelvic diseases, spondylolisthesis, radiculopathy, spondylosis, and spinal fractures.; Sample Size: 73; Number of Sites: Multiple (cross-site validation)
Standard CT scans (reference standard)
Accuracy of 3D bone morphology, radiodensity, and radiodensity contrast (voxel-by-voxel HUs).
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
Lumbar spine bone structure visualization
CNN with cascaded filter banks
p < 0.05
Mean absolute cortical delineation error < 1.0 mm; mean deviation < 25 HU (tissue) and < 55 HU (bone); mean HU correlation coefficient > 0.75 (bone).
Algorithm training sets included information from multiple clinical sites, multiple anatomies and multiple scanners.
—
Quantitative voxel-by-voxel validation on imaging data from 73 patients (sacroiliitis, degenerative spine and pelvic diseases, spondylolisthesis, radiculopathy, spondylosis and spinal fractures) from Europe and Asia.
—
Indications for Use
BoneMRI is an image processing software that can be used for image enhancement in MRI images. It can be used to visualize the bone structures in MRI images with enhanced contrast with respect to the surrounding soft tissue. It is to be used in the pelvic region, which includes the bony anatomy of the sacrum, hip bones and femoral heads; and the lumbar spine region, which includes the bony anatomy of the vertebrae from L3 to S1. BoneMRI is not to be used for diagnosis or monitoring of (primary or metastatic) tumors. Warning: BoneMRI images are not intended to replace CT images.
Device Story
BoneMRI is standalone image processing software; analyzes 3D gradient echo MRI scans; constructs 3D tomographic radiodensity contrast images (BoneMRI images). Used in clinic/hospital networks by radiologists or orthopedic surgeons; operates on DICOM files. Software uses convolutional neural network to assign Hounsfield Unit (HU) values to volume elements based on intensity and context. Output is DICOM images viewed on existing PACS. Enhances bone visualization relative to soft tissue; assists clinical assessment of bone morphology and radiodensity. Does not replace CT; not for tumor diagnosis/monitoring.
Clinical Evidence
Retrospective validation study using 73 patients (48% male, 52% female; mean age 50±15 years). Data included patients with sacroiliitis, degenerative spine/pelvic diseases, spondylolisthesis, radiculopathy, spondylosis, and spinal fractures. BoneMRI images compared to co-registered CT scans (reference standard). Results: mean absolute cortical delineation error <1.0 mm; mean radiodensity deviation <25 HU (overall) and <55 HU (bone); mean HU correlation coefficient >0.75 for bone. All endpoints met pre-specified acceptance criteria (p<0.05).
Technological Characteristics
Software package running on Linux. Inputs/outputs: DICOM MRI images. Core technology: Convolutional neural network (CNN) image enhancement. Processing: Cascade of filter banks with thresholding and scaling; neural network-based filters assign Hounsfield Unit (HU) values to volume elements. Connectivity: Server-based application within hospital/clinic networks. No hardware components.
Indications for Use
Indicated for adult patients requiring visualization of bone structures in the pelvic region (sacrum, hip bones, femoral heads) and lumbar spine (L3-S1) via MRI. Not for diagnosis or monitoring of primary or metastatic tumors.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue, followed by the words "U.S. FOOD & DRUG" in a larger, bold blue font. Below that is the word "ADMINISTRATION" in a smaller, regular blue font.
November 16, 2022
MRIguidance B.V. % Sujith Shetty Executive Vice President MAXIS Medical 3031 Tisch Way, Suite 1010 SAN JOSE CA 95128
Re: K221762
Trade/Device Name: BoneMRI v1.4 Regulation Number: 21 CFR 892.2050 Regulation Name: Medical image management and processing system Regulatory Class: Class II Product Code: OIH Dated: October 7, 2022 Received: October 11, 2022
Dear Sujith Shetty:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Daniel M. Krainak, Ph.D. Assistant Director Magnetic Resonance and Nuclear Medicine Team DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT 8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K221762
Device Name BoneMRI V1.4
#### Indications for Use (Describe)
BoneMRI is an image processing software that can be used for image enhancement in MRI images. It can be used to visualize the bone structures in MRI images with enhanced contrast with respect to the surrounding soft tissue. It is to be used in the pelvic region, which includes the bony anatomy of the sacrum, hip bones and femoral heads; and the lumbar spine region, which includes the bony anatomy of the vertebrae from L3 to S1. BoneMRI is not to be used for diagnosis or monitoring of (primary or metastatic) tumors.
Warning: BoneMRI images are not intended to replace CT images.
Type of Use (Select one or both, as applicable)
| <div> <span> <span style="font-size:16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> <span style="font-size:16px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
|-------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------|
|-------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------|
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for MRI Guidance. The logo consists of a circle with an arrow pointing to the upper left, followed by the text "MRI" in a larger font and "guidance" in a smaller font. The logo is set against a blue background.
# 510(K) SUMMARY
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CRF 807.92.
510 (k) number: K221762
# Applicant Information
MRIguidance B.V. Gildstraat 91a 3572 EL, Utrecht The Netherlands info@mriguidance.com www.mriguidance.com +31 854000810
### Contact Person
Marijn van Stralen Chief Technology Officer MRIquidance B.V. Email: marijn@mriguidance.com Tel.: +31 610 505 649 Date Prepared: December 22, 2021
# Official Correspondent
Dr. Sujith Shetty Executive Vice President MAXIS LLC Email: sjshetty@maxismedical.com
# Device Information
| Trade Name: | BoneMRI application |
|----------------------|-------------------------------------------------------------------|
| Common Name: | MRI image enhancement software |
| Classification name: | Medical image management and processing<br>system (21CRF892.2050) |
| Regulatory Class: | Class II |
| Product Code: | QIH |
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image is a logo for MRI guidance. The logo features a circle with an arrow pointing upwards and to the left. To the right of the circle, the text "MRI guidance" is written in a sans-serif font. The word "MRI" is in a darker shade of blue than the word "guidance."
### Predicate Device
| Name | Manufacturer | 510(k)# |
|--------------|------------------|---------|
| BoneMRI v1.2 | MRIguidance B.V. | K202404 |
odicato Dovico
This predicate has not been subject to a design-related recall. No reference devices were used in this submission.
# Device Description
The BoneMRI application is a standalone image processing software application that analyses 3D gradient echo MRI scans acquired with a dedicated MRI scan protocol. From the analysis, 3D tomographic radiodensity contrast images, called BoneMRI images, are constructed.
The BoneMRI images can be used to visualize the bone structures in MR images with enhanced contrast with respect to the surrounding soft tissue. The application is designed to be used by imaging experts, such as radiologists or orthopedic surgeons, typically in a physician's office.
The BoneMRI application is a server application running in the clinic or hospital networks. It returns the reconstructed BoneMRI images as DICOM images. The application uses an algorithm to detect bone images from MRIs obtained using a specific acquisition sequence. The algorithm training sets included information from multiple clinical sites, multiple anatomies and multiple scanners to ensure that the trained algorithm was robust with respect to the approved indications for use. None of the data used in the training dataset was used subsequently in the validation dataset.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for MRI guidance. The logo consists of a circle with an arrow pointing to the right inside of it. To the right of the circle, the text "MRI guidance" is written in a sans-serif font.
#### Indications for Use
BoneMRI is an image processing software that can be used for image enhancement in MRI images. It can be used to visualize the bone structures in MRI images with enhanced contrast with respect to the surrounding soft tissue. It is to be used in the pelvic region, which includes the bony anatomy of the sacrum, hip bones, and femoral heads; and the lumbar spine region, which includes the bony anatomy of the vertebrae from L3 to S1. BoneMRI is not to be used for diagnosis or monitoring of (primary or metastatic) tumors.
Warning: BoneMRI images are not intended to replace CT images.
### Comparison of Technological Characteristics with the Predicate Device:
A comparison of the intended use, indication for use, and technological characteristics of the BoneMRI v1.4 application to the predicate device, BoneMRI v1.2, is presented below. We have included the attributes suggested in FDA's website guidance for this comparison.
| | Predicate Device<br>BoneMRI v1.2 | Subject Device<br>BoneMRI v1.4 | Comment |
|---------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | BoneMRI is an<br>image processing<br>software that can be<br>used for image<br>enhancement in<br>MRI images. It can<br>be used to visualize<br>the bone structures<br>in MRI images with<br>enhanced contrast<br>with respect to the | BoneMRI is an<br>image processing<br>software that can be<br>used for image<br>enhancement in<br>MRI images. It can<br>be used to visualize<br>the bone structures<br>in MRI images with<br>enhanced contrast<br>with respect to the | Similar –<br>BoneMRI<br>v1.4 has an<br>expanded<br>indications for<br>use<br>including the<br>boney structures<br>of the lumbar<br>spine. |
#### A. Intended Use
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for MRI guidance. The logo consists of a circle with an arrow pointing to the right and the text "MRI guidance" to the right of the circle. The text is in a sans-serif font and is white. The background is blue.
| | Predicate Device<br>BoneMRI v1.2 | Subject Device<br>BoneMRI v1.4 | Comment |
|-------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------|
| | surrounding soft<br>tissue. It is to be<br>used in the pelvic<br>region, which<br>includes the boney<br>anatomy of the<br>sacrum, hip bones<br>and femoral heads.<br>Warning: BoneMRI<br>images are not<br>intended to replace<br>CT images and are<br>not to be used for<br>diagnosis or<br>monitoring of<br>(primary or<br>metastatic) tumors. | surrounding soft<br>tissue. It is to be<br>used in the pelvic<br>region, which<br>includes the bony<br>anatomy of the<br>sacrum, hip bones<br>and femoral heads;<br>and the lumbar<br>spine region, which<br>includes the bony<br>anatomy of the<br>vertebrae from L3 to<br>S1. BoneMRI is not<br>to be used for<br>diagnosis or<br>monitoring of<br>(primary or<br>metastatic) tumors.<br>Warning: BoneMRI<br>images are not<br>intended to replace<br>CT images | |
| 21CFR Section | 892.2050 | 892.2050 | The same |
| Product Code | QIH | QIH | The same |
| Target Population | Adults | Adults | The same |
# B. Technological Characteristics
| | Predicate Device BoneMRI v1.2 | Subject Device BoneMRI v1.4 | Comment |
|--------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------|
| Device Nature | Software package | Software package | The same |
| Operating System | Linux | Linux | The same |
| Data input | MRI images in DICOM format | MRI images in DICOM format | The same |
| | Predicate Device<br>BoneMRI v1.2 | Subject Device<br>BoneMRI v1.4 | Comment |
| Data output | MRI images in<br>DICOM format | MRI images in<br>DICOM format | The same |
| Processing<br>Algorithms | MRIguidance<br>software<br>implements an<br>image<br>enhancement<br>algorithm using<br>convolutional<br>neural network.<br>Original images are<br>enhanced by<br>running them<br>through a cascade<br>of filter banks,<br>where thresholding<br>and scaling<br>operations are<br>applied. Separate<br>neural network-<br>based filters are<br>obtained to assign<br>a Hounsfield Unit<br>(HU) value to a<br>single volume<br>element, based on<br>intensity and<br>contextual<br>information. The<br>parameters of the<br>model were<br>obtained through<br>an algorithm<br>development<br>pipeline. | MRIguidance<br>software<br>implements an<br>image<br>enhancement<br>algorithm using<br>convolutional<br>neural network.<br>Original images<br>are enhanced by<br>running them<br>through a cascade<br>of filter banks,<br>where thresholding<br>and scaling<br>operations are<br>applied. Separate<br>neural network-<br>based filters are<br>obtained to assign<br>a Hounsfield Unit<br>(HU) value to a<br>single volume<br>element, based on<br>intensity and<br>contextual<br>information. The<br>parameters of the<br>model were<br>obtained through<br>an algorithm<br>development<br>pipeline. | The same |
| User Interface | None - enhanced<br>images are viewed<br>on existing PACS | None - enhanced<br>images are viewed<br>on existing PACS | The same |
| | Predicate Device<br>BoneMRI v1.2 | Subject Device<br>BoneMRI v1.4 | Comment |
| Workflow | The software<br>operates on<br>DICOM files on the<br>file system,<br>enhances the<br>images, and stores<br>the enhanced<br>images on the file<br>system. The receipt<br>of original DICOM<br>image files and<br>delivery of<br>enhanced images<br>as DICOM files<br>depends on other<br>software systems.<br>Enhanced images<br>co-exist with the<br>original images. | The software<br>operates on<br>DICOM files on the<br>file system,<br>enhances the<br>images, and stores<br>the enhanced<br>images on the file<br>system. The<br>receipt of original<br>DICOM image files<br>and delivery of<br>enhanced images<br>as DICOM files<br>depends on other<br>software systems.<br>Enhanced images<br>co-exist with the<br>original images. | The same |
{7}------------------------------------------------
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image shows the logo for MRI guidance. The logo consists of a circle with an arrow pointing to the right and the text "MRI guidance" to the right of the circle. The text is in a sans-serif font and is white. The background is blue.
#### Performance Data:
Previous performance testing to verify and validate the software of BoneMRI v1.2 for the pelvic and hip region was submitted and approved under K202404.
BoneMRI conducted the following performance testing:
- Software verification and validation testing 1.
- Studies that utilized retrospective clinical data to demonstrate the software 2. enhanced imaging quality in MR images via an enhancement of bone.
### BoneMRI Lumbar spine region – Voxel-by-Voxel analysis
A quantitative voxel-by-voxel validation of BoneMRI was performed on imaging data from 73 patients. The demographics of the patient population is described in the table below.
{9}------------------------------------------------
Image /page/9/Picture/1 description: The image shows the logo for MRI guidance. The logo consists of a circle with an arrow pointing upwards and to the right, followed by the text "MRI" in a larger font and "guidance" in a smaller font. The logo is in shades of blue and teal.
| Validation data demographics | |
|------------------------------|--------------------------------------------------------------------------------------------------------------------------------|
| Number of patients | 73 |
| Indications | sacroiliitis, degenerative spine and pelvic<br>diseases, spondylolisthesis, radiculopathy,<br>spondylosis and spinal fractures |
| Gender | Male: 48%<br>Female: 52% |
| Age | $50 \pm 15$ years |
| Data origin/Ethnicity | Europe<br>Asia |
The imaging data consists of the BoneMRI and a standard CT of the same patient in the same anatomical region, acquired during previously conducted clinical investigations. The validations were conducted by MRIguidance based on an algorithm to detect bone images from MRIs obtained using a specific acquisition sequence.
Training and test datasets were selected and maintained to be appropriately independent of one another. All training and validation activities were recorded to ensure independence. In addition, validation was performed on data from independent sites (cross-site validation) to ensure that validation was performed on data from unseen centers.
The objective was to validate the quantitative accuracy of BoneMRI for the lumbar spine region using rigorous, objective, and unbiased statistical tests comparing bone morphology, radiodensity, and radiodensity contrast in BoneMRI and CT images. Therefore, the endpoints of this testing were the metrics that described the accuracy of 3D bone morphology, radiodensity, and radiodensity contrast versus co-registered CT scans (reference standard) in terms of voxel-by-voxel HUs and standard deviations around these HU values.
The results from the validation testing were compared to the accuracy acceptance criteria, specified below, and were found to fall within the pre-specified acceptance criteria (p<0.05).
{10}------------------------------------------------
The results demonstrate clinically acceptable accuracy on each of these endpoints.
The data provided demonstrate that BoneMRI application can
- o accurately reconstruct the 3D bone morphology with a mean absolute cortical delineation error below 1.0 mm on average;
- o accurately reconstructs the tissue radiodensity, with a mean deviation below 25 HU on average and a mean deviation below 55 HU specifically for bone;
- o accurately reconstructs the tissue radiodensity contrast, with a mean HU correlation coefficient above 0.75 specifically for bone.
CONCLUSION: BoneMRI demonstrates accurate bone morphology, radiodensity, and radiodensity contrast. Thus, BoneMRI is a useful tool to qualitatively and quantitatively assess the lumbar spine region.
#### Conclusions:
BoneMRI v1.4, based on the indications for use, product performance, and clinical information provided in this notification, has been shown to be substantially equivalent to the currently marketed predicate device, its predecessor, BoneMRI v1.2. The two devices have the same technological characteristics: both algorithms use the same image-based reconstruction, and both methods have optimized parameters to ensure the robustness of the algorithm. This 510(k) submission includes information on the BoneMRI v1.4 technological characteristics, as well as performance data and verification and validation activities demonstrating that BoneMRI is as safe and effective as the predicate. This evaluation did not raise any new issues pertaining to safety or effectiveness.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.