K210187 · Overjet, Inc. · LLZ · May 19, 2021 · Radiology
Device Facts
Record ID
K210187
Device Name
Overjet Dental Assist
Applicant
Overjet, Inc.
Product Code
LLZ · Radiology
Decision Date
May 19, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
AI/ML, Software as a Medical Device, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K210187 · May 19, 2021
Overjet Dental Assist
Overjet, Inc.
Retrospective clinical dental radiographs (bitewing and periapical)
Retrospective clinical radiographs were used to evaluate the performance of the Overjet Dental Assist software in measuring interproximal bone levels. The software's measurements were compared against adjudicated ground truth measurements derived from these retrospective clinical images.
Overjet Dental Assist is a radiological semi-automated image processing software device intended to aid dental professionals in the measurements of mesial and distal bone levels associated with each tooth from bitewing and periapical radiographs. It should not be used in-lieu of full patient evaluation or solely relied upon to make or confirm a diagnosis. The system is to be used by trained professionals including, but not limited to, dentists and dental hygienists.
Device Story
Cloud-native software; processes bitewing and periapical dental radiographs. Inputs: digital radiographs imported from radiographic devices or practice management systems. Operation: semi-automated image processing; identifies key anatomical points to measure interproximal bone levels. Output: annotated radiographs with bone level measurements. Usage: clinical setting; operated by dentists or dental hygienists. Workflow: clinician reviews automated annotations, modifies as needed, and integrates findings into diagnostic process. Benefit: provides objective measurements to assist in periodontal assessment and treatment planning; does not replace clinical judgment.
Clinical Evidence
Retrospective study of 161 radiographs (65 bitewing, 96 periapical) from 63 subjects. Ground truth established by three dentists and adjudicated by two dental radiologists. Bitewing performance: 98.7% sensitivity, 95.0% specificity, 0.307mm mean absolute difference. Periapical bone level performance: 88.94% sensitivity, 95.96% specificity, 0.353mm mean absolute difference. Periapical root length performance: 90.9% sensitivity, 97.47% specificity, 0.567mm mean absolute difference. Bench testing on 8,777 radiographs showed 82.3-83.0% average precision and 89.5-90.0% average recall.
Technological Characteristics
Cloud-native software as a medical device (SaMD). Web-based interface (Edge, Chrome, Firefox). Compatible with SQL databases. Supports jpg, png, jfif, eop, etp, jif formats. Performs automated linear distance measurements via keypoint detection. Software level of concern: moderate.
Indications for Use
Indicated for dental professionals to aid in measuring mesial and distal bone levels from bitewing and periapical radiographs. Intended for patients 22 years or older with no remaining primary teeth. Not for use on primary dentition.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{1}------------------------------------------------
May 19, 2021
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Overjet, Inc. % Mr. Adam Heroux Head of Regulatory Affairs 560 Harrison Ave, Unit 403 BOSTON MA 02118
Re: K210187
Trade/Device Name: Overjet Dental Assist Regulation Number: 21 CFR 892.2050 Regulation Name: Medical image management and processing system Regulatory Class: Class II Product Code: LLZ Dated: April 12, 2021 Received: April 21, 2021
Dear Mr. Heroux:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
{2}------------------------------------------------
https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K210187
Device Name
Overjet Dental Assist
Indications for Use (Describe)
Overjet Dental Assist is a radiological semi-automated image processing software device intended to aid dental professionals in the measurements of mesial and distal bone levels associated with each tooth from bitewing and periapical radiographs.
It should not be used in-lieu of full patient evaluation or solely relied upon to make or confirm a diagnosis. The system is to be used by trained professionals including, but not limited to, dentists and dental hygienists.
Intended Patient Population:
The intended patient population of the device is patients living in the United States, who are 22 years old or older, and that do not have any remaining primary teeth. Overjet has not evaluated the performance of the device on primary dentition.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| <span style="display:inline-block; margin-right: 5px;"> <svg fill="none" height="16px" viewbox="0 0 16 16" width="16px" xmlns="http://www.w3.org/2000/svg"> <path d="M13.25 3.34L5.92 10.67L2.75 7.5L1.92 8.33L5.92 12.33L14.08 4.17L13.25 3.34Z" fill="black"></path> <rect height="15" rx="3.5" stroke="black" width="15" x="0.5" y="0.5"></rect> </svg> </span> Prescription Use (Part 21 CFR 801 Subpart D) | |
| <span style="display:inline-block; margin-right: 5px;"> <svg fill="none" height="16px" viewbox="0 0 16 16" width="16px" xmlns="http://www.w3.org/2000/svg"> <rect height="15" rx="3.5" stroke="black" width="15" x="0.5" y="0.5"></rect> </svg> </span> Over-The-Counter Use (21 CFR 801 Subpart C) | |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary
(K210187)
This summary of 510(k) information is being submitted in accordance with the requirements of 21CFR Part 807.92
- 1. Date 4/12//2021
- 2. Applicant
Overjet, Inc. 560 Harrison Ave Unit 403 Boston, MA 02118 Contact Person: Adam Heroux Email: adam@overjet.ai
- 3. Trade Name Overjet Dental Assist
- 4. Common Name Dental Imaging Software
- 5. Classification System, Imaging processing, radiological (21CFR 892.2050, Product code LLZ, Class 2, Radiology)
- 6. Device Description
Overjet Dental Assist developed by Overjet Inc, is a radiological semi automated image processing software device intended to aid dental professionals in the measurements of mesial and distal bone levels associated with each tooth from bitewing and periapical radiographs.
Overjet Dental Assist is a cloud native Software as a Medical Device that allows users to automate the measurement of interproximal bone levels for bitewing and periapical radiographs, review associated radiographs, view annotations, modify annotations.
- 7. Indications for Use
Overjet Dental Assist is a radiological semi-automated image processing software device intended to aid dental professionals in the measurements of mesial and distal bone levels associated with each tooth from bitewing and periapical radiographs.
{5}------------------------------------------------
lt should not be used in-lieu of full patient evaluation or solely relied upon to make or confirm a diagnosis. The system is to be used by trained professionals including, but not limited to, dentists and dental hygienists.
Intended Patient Population:
The intended patient population of the device is patients living in the United States, who are 22 years old or older, and that do not have any remaining primary teeth. Overjet has not evaluated the performance of the device on primary dentition.
- 8. Predicate Device
Device - EZOrtho Manufacturer - Ewoosoft Co., Ltd. 510k - K192888
- 9. Substantial Equivalence
| Device | Ewoosoft<br>EZOrtho v1.0<br>(predicate) | Overjet Dental Assist<br>(proposed) |
|----------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510k | K192888 | K210187 |
| Indications | EzOrtho is a software<br>indicated for use by dentists<br>who provide orthodontic<br>treatment for image analysis,<br>simulation, profilogram,<br>VTO/STO and patient<br>consultation. Results<br>produced by the software's<br>diagnostic, treatment planning<br>and simulation tools are<br>dependent on the<br>interpretation of trained and<br>licensed practitioners or<br>dentists | Overjet Dental Assist is a<br>radiological semi-automated<br>image processing software device<br>intended to aid dental<br>professionals in the<br>measurements of mesial and<br>distal bone levels associated with<br>each tooth. It should not be used<br>in-lieu of full patient evaluation or<br>solely relied upon to make or<br>confirm a diagnosis. The system<br>is to be used by trained<br>professionals including, but not<br>limited to, dentists and dental<br>hygienists |
| Patient<br>Population | Patients receiving dental care | Patients receiving dental care |
| Platform | IBM-compatible PC or PC<br>network | Web - Edge, Chrome, Firefox |
| OS | Microsoft Window 7, 8, 10 | any |
| User Interface | Mouse, Keyboard | Mouse, Keyboard, Trackpad |
| Image Input<br>Sources | Images can be scanned,<br>loaded from digital cameras<br>or card readers, or imported<br>from a radiographic imaging<br>device | Images imported from the<br>radiographic device, or from the<br>practice management system |
| Image format | DICOM, BMP, JPG, PNG, TIF | jpg, png, jfif, eop, etp, jif |
| Patient Database<br>Compatibility | SQL | SQL |
| Includes Image<br>Measurement<br>tools | Linear distance, angle | Linear distance |
| Image viewing | Full, side by side, thumbnail | Full, Thumbnail |
| Image<br>manipulation | Grayscale, invert, emboss,<br>brightness, contrast, gamma,<br>sharpen, median, despeckle,<br>hue, saturation, equalize flip,<br>mirror, masking, rotate,<br>annotation, cephalometric<br>tracing, implant simulations | annotation |
| Cephalometric<br>tracing | In addition to the<br>user-configured analysis,<br>standard orthodontic tracing<br>analysis include: Downs<br>Jarabek McNamara Ricketts<br>Jefferson | n/a |
| Implant module | Generic | n/a |
| 3D imaging<br>capability | None | none |
| Image annotation | Text, paint, ellipse, pointer,<br>select, draw, magnify, line,<br>rectangle, ruler, protractor,<br>brush, select region, copy /<br>paste | Line |
{6}------------------------------------------------
Overjet Dental Assist is determined to be substantially equivalent to the Ewoosoft EZ Ortho cleared as K192888. Both systems are software intended to support dental professionals in their diagnosis and treatment planning for their dental patients.
{7}------------------------------------------------
Both software systems automatically annotate dental radiographs for the dentist. EZ Ortho places calibration points within a radiograph for orthodontic planning, while the Overjet Dental Assist utilizes key points to measure interproximal bone levels. Both systems allow users to visualize the radiograph with the annotations, add their own annotations, and use the information as part of their diagnostic process.
Other similarities include both systems have no direct contact with the patient, both systems evaluate oral cavity radiographs, both systems utilize standard image types, and both systems connect to practice management systems.
Some differences between the systems include the location of the software, the user interface, and the availability of additional features. A primary difference is the EZ Otho is a local software while Overjet Dental Assist is a cloud native application. While EZ Ortho and Overjet Dental Assist have different user interfaces, both are accessed by computer and are intended for dental professionals to review annotations on dental radiographs. The EZ Ortho software contains additional annotation features that are not required for the Overjet Dental Assist use case. Overjet does not feel that the differences raise a concern of substantial equivalence and these differences do not interfere with the ability of the Overjet software to achieve its intended use.
- 10. Performance Testing
## Software Verification and Validation Testing
Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "moderate" level of concern, since a failure or latent flaw could indirectly result in minor injury to the patient or operator through incorrect or delayed information or through the action of a care provider.
## Bench Testing
Bench testing included evaluation of performance based upon a ground truth data set utilizing Object Keypoint Similarity assessment, Overiet Dental Assist was evaluated with two datasets of 2234 bitewing radiographs, and 6543 periapical radiographs and evaluated the precision and recall against the labeled keypoints within those radiographs. The Overjet met acceptable performance criteria with the following results:
| Metric | Bitewing | Periapical |
|-------------------|----------|------------|
| Average Precision | 82.3% | 83.0% |
| Average Recall | 89.5% | 90.0% |
{8}------------------------------------------------
#### Clinical Testing
Clinical testing of the Overjet Dental Assist software was performed utilizing retrospective clinical subject data from 65 bitewing and 96 periapical radiographs from 63 subjects. The demographics of the data set was as follows:
| Subject Age | Count (%) |
|----------------------------|-----------|
| under 30 | 15 (24%) |
| 30-40 | 9 (14%) |
| 40-50 | 8 (13%) |
| 50-60 | 12 (19%) |
| 60-70 | 6 (10%) |
| 70 + | 12(19%) |
| no information<br>provided | 1 (1%) |
| Subject Gender | Count (%) |
|----------------|-----------|
| F | 46 (73%) |
| M | 17 (27%) |
No information about ethnicity was available from the dental radiographs.
Overjet utilized three US licensed dentists to label interproximal bone levels using the measurement tool in Dexis Webview software for all of the radiographs. These measurements were then adjudicated by two US Dental Radiologists, The final adjudicated measurements were compared against the Overjet Dental Assist predicted measurements.
The Overjet Dental Assist software met the prespecified acceptance criteria of >85% sensitivity and specificity when compared to the ground truth, and accuracy of <1.5mm mean absolute difference versus ground truth.
The Overjet Dental Assist software demonstrated a modeled 98.7% sensitivity value and 95.0% specificity value, and a mean absolute difference from the ground truth of 0.307mm for measurement of interproximal bone levels in bitewing radiographs. In interproximal bone levels in periapical radiographs, Overjet Dental Assist scored 88.94% sensitivity, 95,96% specificity, and 0,353mm mean absolute difference, Performance of measurement of Periapical Root Length was measured at 90.9% sensitivity, 97.47% specificity, and 0.567mm mean absolute difference.
The interproximal bone level value in periapical images while meeting the performance acceptance criteria, did not meet the alpha value of <0.05. The alpha for the hypothesis test was 0.083. Analysis of this data indicated that 33% of the false negatives were
{9}------------------------------------------------
related to anatomical overlap that obstructed the CEJ point. In this case human readers were more prone to estimate a bone level point that was obscured. The overall sensitivity of 88% was consistent with the performance of the three ground truth dentists.
No adverse effects or complications were recorded associated with this study as all radiographs were retrospective and no treatment decisions were made.
All testing demonstrated that the Overjet Dental Assist software met prespecified performance requirements when evaluated against human labeling.
#### 11. Conclusion
Overjet Dental Assist is substantially equivalent to the predicate device, EZ Ortho. Differences do not raise any concerns about the safety or efficacy of the device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.