Retrospective standalone performance study: 600 anonymized head CT cases from 5 sites in US and Europe
3 (US board certified neuroradiologists)
Indications for Use
syngo.CT Brain Hemorrhage is designed to assist the radiologist in prioritizing cases of suspected intracranial hemorrhage on non-contrast CT examinations of the head. It makes case-level output available to a CT scanner or other PACS system for worklist prioritization. The output is intended for informational purposes only and is not intended for diagnostic use. The device does not alter the original medical image and is not intended to be used as a stand-alone diagnostic device.
Device Story
Image processing software using AI algorithms to analyze non-contrast head CT DICOM images; identifies findings suspicious for acute intracranial hemorrhage (ICH). Operates as a triage and notification tool in acute care settings; integrates with CT scanners or PACS to prioritize worklists. Provides case-level output (suspected hemorrhage vs. no suspected hemorrhage) to clinicians. Does not alter original images; intended for informational purposes to assist radiologist workflow; does not remove cases from standard-of-care worklists. Benefits include faster identification and prioritization of urgent ICH cases.
Clinical Evidence
Retrospective standalone performance study using 600 anonymized head CT cases (balanced positive/negative for ICH). Ground truth established by majority read of 3 US board-certified neuroradiologists (>10 years experience). Sensitivity: 92.8% (95% CI: 89.3%-95.2%); Specificity: 94.5% (95% CI: 91.3%-96.5%). Average processing time: 13.67 seconds.
Technological Characteristics
Image processing software; AI/ML-based algorithm. DICOM-compliant (PS 3.1-3.20). Standards: IEC 62304 (Software Life Cycle), ISO 14971 (Risk Management), IEC 62366-1 (Usability). Operates as a notification-only, parallel workflow tool.
Indications for Use
Indicated for radiologists to prioritize non-contrast head CT examinations for suspected intracranial hemorrhage (ICH). Not for diagnostic use; not a stand-alone diagnostic device.
Regulatory Classification
Identification
Radiological computer aided triage and notification software is an image processing prescription device intended to aid in prioritization and triage of radiological medical images. The device notifies a designated list of clinicians of the availability of time sensitive radiological medical images for review based on computer aided image analysis of those images performed by the device. The device does not mark, highlight, or direct users' attention to a specific location in the original image. The device does not remove cases from a reading queue. The device operates in parallel with the standard of care, which remains the default option for all cases.
Special Controls
Radiological computer aided triage and notification software must comply with the following special controls: 1. Design verification and validation must include: i. A detailed description of the notification and triage algorithms and all underlying image analysis algorithms including, but not limited to, a detailed description of the algorithm inputs and outputs, each major component or block, how the algorithm affects or relates to clinical practice or patient care, and any algorithm limitations. ii. A detailed description of pre-specified performance testing protocols and dataset(s) used to assess whether the device will provide effective triage (e.g., improved time to review of prioritized images for pre-specified clinicians). iii. Results from performance testing that demonstrate that the device will provide effective triage. The performance assessment must be based on an appropriate measure to estimate the clinical effectiveness. The test dataset must contain sufficient numbers of cases from important cohorts (e.g., subsets defined by clinically relevant confounders, effect modifiers, associated diseases, and subsets defined by image acquisition characteristics) such that the performance estimates and confidence intervals for these individual subsets can be characterized with the device for the intended use population and imaging equipment. iv. Standalone performance testing protocols and results of the device. v. Appropriate software documentation (e.g., device hazard analysis; software requirements specification document; software design specification document; traceability analysis; description of verification and validation activities including system level test protocol, pass/fail criteria, and results). 2. Labeling must include the following: i. A detailed description of the patient population for which the device is indicated for use. ii. A detailed description of the intended user and user training that addresses appropriate use protocols for the device. iii. Discussion of warnings, precautions, and limitations must include situations in which the device may fail or may not operate at its expected performance level (e.g., poor image quality for certain subpopulations), as applicable. iv. A detailed description of compatible imaging hardware, imaging protocols, and requirements for input images. v. Device operating instructions. vi. A detailed summary of the performance testing, including: test methods, dataset characteristics, triage effectiveness (e.g., improved time to review of prioritized images for pre-specified clinicians), diagnostic accuracy of algorithms informing triage decision, and results with associated statistical uncertainty (e.g., confidence intervals), including a summary of subanalyses on case distributions stratified by relevant confounders, such as lesion and organ characteristics, disease stages, and imaging equipment.
*Classification.* Class II (special controls). The special controls for this device are:(1) Design verification and validation must include:
(i) A detailed description of the notification and triage algorithms and all underlying image analysis algorithms including, but not limited to, a detailed description of the algorithm inputs and outputs, each major component or block, how the algorithm affects or relates to clinical practice or patient care, and any algorithm limitations.
(ii) A detailed description of pre-specified performance testing protocols and dataset(s) used to assess whether the device will provide effective triage (
*e.g.,* improved time to review of prioritized images for pre-specified clinicians).(iii) Results from performance testing that demonstrate that the device will provide effective triage. The performance assessment must be based on an appropriate measure to estimate the clinical effectiveness. The test dataset must contain sufficient numbers of cases from important cohorts (
*e.g.,* subsets defined by clinically relevant confounders, effect modifiers, associated diseases, and subsets defined by image acquisition characteristics) such that the performance estimates and confidence intervals for these individual subsets can be characterized with the device for the intended use population and imaging equipment.(iv) Stand-alone performance testing protocols and results of the device.
(v) Appropriate software documentation (
*e.g.,* device hazard analysis; software requirements specification document; software design specification document; traceability analysis; description of verification and validation activities including system level test protocol, pass/fail criteria, and results).(2) Labeling must include the following:
(i) A detailed description of the patient population for which the device is indicated for use;
(ii) A detailed description of the intended user and user training that addresses appropriate use protocols for the device;
(iii) Discussion of warnings, precautions, and limitations must include situations in which the device may fail or may not operate at its expected performance level (
*e.g.,* poor image quality for certain subpopulations), as applicable;(iv) A detailed description of compatible imaging hardware, imaging protocols, and requirements for input images;
(v) Device operating instructions; and
(vi) A detailed summary of the performance testing, including: test methods, dataset characteristics, triage effectiveness (
*e.g.,* improved time to review of prioritized images for pre-specified clinicians), diagnostic accuracy of algorithms informing triage decision, and results with associated statistical uncertainty (*e.g.,* confidence intervals), including a summary of subanalyses on case distributions stratified by relevant confounders, such as lesion and organ characteristics, disease stages, and imaging equipment.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services seal on the left and the FDA acronym along with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name written out to the right of the square.
Siemens Medical Solutions USA, Inc. % Tabitha Estes Regulatory Affairs Specialist 810 Innovation Drive KNOXVILLE TN 37932
Re: K203260
January 28, 2022
Trade/Device Name: syngo.CT Brain Hemorrhage Regulation Number: 21 CFR 892.2080 Regulation Name: Radiological computer aided triage and notification software Regulatory Class: Class II Product Code: QAS Dated: December 15, 2021 Received: December 17, 2021
Dear Tabitha Estes:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
{1}------------------------------------------------
https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K203260
Device Name syngo.CT Brain Hemorrhage
### Indications for Use (Describe)
syngo.CT Brain Hemorrhage is designed to assist the radiologist in prioritizing cases of suspected intracranial hemorrhage on non-contrast CT examinations of the head. It makes case-level output available to a CT scanner or other PACS system for worklist prioritization. The output is intended for informational purposes only and is not intended for diagnostic use. The device does not alter the original medical image and is not intended to be used as a stand-alone diagnostic device.
Type of Use (Select one or both, as applicable)
| <span></span> | <input checked="True" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) |
|---------------|--------------------------------------------------------------------------------------|
| <span></span> | <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(K) SUMMARY FOR SYNGO.CT BRAIN HEMORRHAGE
# K203260
Date Prepared: January 26th, 2022
## Identification of the Submitter
Importer/Distributor Siemens Medical Solutions USA, Inc. 40 Liberty Boulevard Malvern, PA 19355 Establishment Registration Number 2240869
Manufacturing Site Siemens Healthcare GmbH Siemensstr 1 D-91301 Forchheim, Germany
Establishment Registration Number 3004977335
### Submitter Contact Person:
Tabitha Estes Regulatory Affairs Specialist Siemens Medical Solutions, Inc. USA 810 Innovation Drive Knoxville, TN 37932 Phone: (865) 804-4553 Email: tabitha.estes(@siemens-healthineers.com
Alternate Contact: Alaine Medio
#### I. Device Name and Classification
Product Name: syngo.CT Brain Hemorrhage Propriety Trade Name: syngo.CT Brain Hemorrhage Classification Name: Radiological Computer-Assisted Triage and Notification Software Classification Panel: Radiology 21 CFR §892.2080 CFR Section: Device Class: Class II Product Code: QAS
{4}------------------------------------------------
#### II. Predicate Device
| Trade Name: | Accipiolx |
|-----------------------|-----------------------------------------------------------------|
| 510(k) Number: | K182177 |
| Clearance Date: | 07/26/2019 |
| Classification Name: | Radiological Computer-Assisted Triage and Notification Software |
| Classification Panel: | Radiology |
| CFR Section: | 21 CFR §892.2080 |
| Device Class: | Class II |
| Product Code: | QAS |
# III. Device Description
The subject device syngo.CT Brain Hemorrhage is an image processing software that utilizes artificial intelligence learning algorithms to support qualified clinicians in analysis and prioritization of noncontrast head CT DICOM images by algorithmically identifying findings suspicious of acute intracranial hemorrhage. The subject device provides a pipeline for the analysis and identification of potential ICH as well as a finding notification mechanism.
#### IV. Indications for Use
syngo.CT Brain Hemorrhage is designed to assist the radiologist in prioritizing cases of suspected intracranial hemorrhage on non-contrast CT examinations of the head. It makes case-level output available to a CT scanner or other PACS system for worklist prioritization. The output is intended for informational purposes only and is not intended for diagnostic use. The device does not alter the original medical image and is not intended to be used as a stand-alone diagnostic device.
#### V. Comparison of Technological Characteristics with the Predicate Device
The differences and similarities between the above referenced predicate device are listed at a high-level in the following table:
| Feature | Subject Device | Predicate Device | Comparison<br>Table |
|----------------------------------------------------------------|-----------------------------------------------|-----------------------------------------------|---------------------|
| | syngo.CT Brain Hemorrhage | MaxQ-AI Accipiolx | |
| Notification-only, parallel<br>workflow tool | Yes | Yes | Same |
| Intended User | Hospital networks and qualified<br>clinicians | Hospital networks and qualified<br>clinicians | Same |
| Setting | Acute Care | Acute Care | Same |
| Identify patients with a<br>prespecified clinical<br>condition | Yes | Yes | Same |
| Clinical condition | Intracranial hemorrhage | Intracranial hemorrhage | Same |
| Alert to finding | Yes; flagged for review | Yes; flagged for review | Same |
| Independent of standard<br>of care workflow | Yes; No cases are removed from<br>worklist | Yes; No cases are removed from<br>worklist | Same |
| Modality | Non-contrast CT | Non-contrast CT | Same |
| Body Part | Head | Head | Same |
| Artificial Intelligence<br>algorithm | Yes | Yes | Same |
| Limited to analysis of<br>imaging data | Yes | Yes | Same |
{5}------------------------------------------------
| Feature | Subject Device | Predicate Device | Comparison Table |
|---------|------------------------------------------------|------------------------------------------------|------------------|
| Output | syngo.CT Brain Hemorrhage | MaxQ-AI Accipiolx | Same |
| | Suspected hemorrhage / No suspected hemorrhage | Suspected hemorrhage / No suspected hemorrhage | |
The subject device syngo.CT Brain Hemorrhage does not have changes in fundamental scientific technology compared to the predicate device. The post-processing software functionality remains unchanged from the subject device and the predicate device. The operating principle and the scientific technology are the same; therefore, Siemens believes that syngo.CT Brain Hemorrhage application is substantially equivalent to the predicate device.
#### VI. Performance Data
The following performance data were provided in support of the substantial equivalence determination.
### Non-Clinical Testing
Non-clinical tests (integration and functional) were conducted for syngo.CT Brain Hemorrhage during product development. These tests have been performed to test the ability of the included features of the subject device. The results of these tests demonstrate that the subject device performs as intended. The results of all conducted testing were found acceptable to support the claim of substantial equivalence.
### Performance Evaluation of the Algorithm
The performance of the syngo.CT Brain Hemorrhage device has been validated in a retrospective standalone performance study. Sensitivity and specificity of syngo.CT Brain Hemorrhage in processing of non-contrast head CT have been analyzed by comparison to a ground truth established by majority read of 3 US board certified neuroradiologists with more than 10 years of experience. The data cohort consisted of 600 anonymized head CT cases from 5 sites in US and Europe with approximately equal distribution of positive (case with ICH) and negative (case without ICH) cases.
Sensitivity was observed to be 92.8% (95% CI: 89.3%-95.2%). Specificity was observed to be 94.5% (95% CI: 91.3%-96.5%). Thus, sensitivity and specificity exceeded the 80% performance goal.
The average per-case processing time was 13.67 seconds (95% CI: 7.48-19.86 seconds). This is comparable to the predicate device.
To conclude, syngo.CT Brain Hemorrhage is a safe and effective triage and notification device for intracranial hemorrhages which is substantially equivalent to the predicate device.
Siemens hereby certifies that syngo.CT Brain Hemorrhage will meet the following voluntary standards covering electrical and mechanical safety listed below, prior to introduction into interstate commerce:
| Recognition<br>Number | Product<br>Area | Title of Standard | Date of<br>Recognition | Standards<br>Development<br>Organization |
|-----------------------|--------------------------|------------------------------------------------------------------------------------------------------------|------------------------|------------------------------------------|
| 12-300 | Radiology | Digital Imaging and Communications in<br>Medicine (DICOM) Set; PS 3.1 - 3.20 | 06/27/2016 | NEMA |
| 13-79 | Software | Medical Device Software -Software Life Cycle<br>Processes; 62304:2006 (1st Edition) | 01/14/2019 | AAMI, ANSI,<br>IEC |
| 5-40 | Software/<br>Informatics | Medical devices – Application of risk<br>management to medical devices; 14971 Second<br>Edition 2007-03-01 | 06/27/2016 | ISO |
| 5-114 | General I<br>(QS/RM) | Medical devices - Part 1: Application of<br>usability engineering to medical devices<br>IEC 62366-1:2015 | 12/23/2016 | IEC |
{6}------------------------------------------------
### VII. Conclusion
syngo.CT Brain Hemorrhage has the same intended use and similar indication for use as the predicate device. The result of all testing conducted was found acceptable to support the claim of substantial equivalence. The comparison of technological characteristics, clinical and non-clinical performance data, and software validation demonstrates that the subject device is as safe and effective when compared to the predicate device that is currently marketed for the same intended use. Siemens considers syngo.CT Brain Hemorrhage to be as safe, as effective and with performance substantially equivalent to the commercially available predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.