The AIMI-Triage CXR PTX Application is a notification-only triage workflow tool for use by hospital networks and clinics to identify and help prioritize chest X-rays acquired in the acute setting for review by hospital radiologists. The device operates in parallel to and independent of standard of care image interpretation workflow. Specifically, the device uses an artificial intelligence algorithm to analyze images for features suggestive of moderate to large sized pneumothorax; it makes case-level output available to a PACS/workstation for worklist prioritization or triage. Identification of suspected cases of moderate to large sized pneumothorax is not for diagnostic use beyond notification. The AIMI-Triage CXR PTX Application is limited to analysis of imaging data as a guide to possible urgency of adult chest X-ray image review, and should not be used in lieu of full patient evaluation or relied upon to make or confirm diagnoses. Notified radiologists are responsible for engaging in appropriate patient evaluation as per local hospital procedure before making care-related decisions or requests. The device does not replace review and diagnosis of the X-rays by radiologists. The device is not intended to be used with plain film X-rays.
Device Story
AIMI-Triage CXR PTX is a software-based triage tool for hospital radiologists. It inputs chest X-ray images from a PACS/workstation; uses an AI algorithm trained via pattern recognition to detect features suggestive of moderate to large pneumothorax; outputs a case-level notification to the PACS worklist to prioritize urgent review. It operates in parallel to standard clinical workflows; does not alter images; does not provide diagnoses; does not highlight specific image regions. Radiologists use the notification to adjust review priority. Benefits include faster identification of potentially urgent cases, improving clinical workflow efficiency.
Clinical Evidence
Blinded, retrospective, multi-center study of 300 frontal chest X-rays (168 male, 132 female; mean age 51.6). Ground truth established by 3 independent US-board certified radiologists. Primary endpoint: AUC of 96.7% (95% CI: 95.0%, 98.4%). Sensitivity 92% (95% CI: 86%, 96%). Average analysis time 20.3 seconds.
Technological Characteristics
Software-only device; AI algorithm trained via pattern recognition. Compatible with standard PACS/workstation. No hardware components. No image alteration or segmentation. Operates in parallel to standard of care.
Indications for Use
Indicated for use in hospital networks and clinics to identify and prioritize adult chest X-rays acquired in acute settings for features suggestive of moderate to large sized pneumothorax. Not for diagnostic use; not for use with plain film X-rays.
Regulatory Classification
Identification
Radiological computer aided triage and notification software is an image processing prescription device intended to aid in prioritization and triage of radiological medical images. The device notifies a designated list of clinicians of the availability of time sensitive radiological medical images for review based on computer aided image analysis of those images performed by the device. The device does not mark, highlight, or direct users' attention to a specific location in the original image. The device does not remove cases from a reading queue. The device operates in parallel with the standard of care, which remains the default option for all cases.
Special Controls
Radiological computer aided triage and notification software must comply with the following special controls: 1. Design verification and validation must include: i. A detailed description of the notification and triage algorithms and all underlying image analysis algorithms including, but not limited to, a detailed description of the algorithm inputs and outputs, each major component or block, how the algorithm affects or relates to clinical practice or patient care, and any algorithm limitations. ii. A detailed description of pre-specified performance testing protocols and dataset(s) used to assess whether the device will provide effective triage (e.g., improved time to review of prioritized images for pre-specified clinicians). iii. Results from performance testing that demonstrate that the device will provide effective triage. The performance assessment must be based on an appropriate measure to estimate the clinical effectiveness. The test dataset must contain sufficient numbers of cases from important cohorts (e.g., subsets defined by clinically relevant confounders, effect modifiers, associated diseases, and subsets defined by image acquisition characteristics) such that the performance estimates and confidence intervals for these individual subsets can be characterized with the device for the intended use population and imaging equipment. iv. Standalone performance testing protocols and results of the device. v. Appropriate software documentation (e.g., device hazard analysis; software requirements specification document; software design specification document; traceability analysis; description of verification and validation activities including system level test protocol, pass/fail criteria, and results). 2. Labeling must include the following: i. A detailed description of the patient population for which the device is indicated for use. ii. A detailed description of the intended user and user training that addresses appropriate use protocols for the device. iii. Discussion of warnings, precautions, and limitations must include situations in which the device may fail or may not operate at its expected performance level (e.g., poor image quality for certain subpopulations), as applicable. iv. A detailed description of compatible imaging hardware, imaging protocols, and requirements for input images. v. Device operating instructions. vi. A detailed summary of the performance testing, including: test methods, dataset characteristics, triage effectiveness (e.g., improved time to review of prioritized images for pre-specified clinicians), diagnostic accuracy of algorithms informing triage decision, and results with associated statistical uncertainty (e.g., confidence intervals), including a summary of subanalyses on case distributions stratified by relevant confounders, such as lesion and organ characteristics, disease stages, and imaging equipment.
*Classification.* Class II (special controls). The special controls for this device are:(1) Design verification and validation must include:
(i) A detailed description of the notification and triage algorithms and all underlying image analysis algorithms including, but not limited to, a detailed description of the algorithm inputs and outputs, each major component or block, how the algorithm affects or relates to clinical practice or patient care, and any algorithm limitations.
(ii) A detailed description of pre-specified performance testing protocols and dataset(s) used to assess whether the device will provide effective triage (
*e.g.,* improved time to review of prioritized images for pre-specified clinicians).(iii) Results from performance testing that demonstrate that the device will provide effective triage. The performance assessment must be based on an appropriate measure to estimate the clinical effectiveness. The test dataset must contain sufficient numbers of cases from important cohorts (
*e.g.,* subsets defined by clinically relevant confounders, effect modifiers, associated diseases, and subsets defined by image acquisition characteristics) such that the performance estimates and confidence intervals for these individual subsets can be characterized with the device for the intended use population and imaging equipment.(iv) Stand-alone performance testing protocols and results of the device.
(v) Appropriate software documentation (
*e.g.,* device hazard analysis; software requirements specification document; software design specification document; traceability analysis; description of verification and validation activities including system level test protocol, pass/fail criteria, and results).(2) Labeling must include the following:
(i) A detailed description of the patient population for which the device is indicated for use;
(ii) A detailed description of the intended user and user training that addresses appropriate use protocols for the device;
(iii) Discussion of warnings, precautions, and limitations must include situations in which the device may fail or may not operate at its expected performance level (
*e.g.,* poor image quality for certain subpopulations), as applicable;(iv) A detailed description of compatible imaging hardware, imaging protocols, and requirements for input images;
(v) Device operating instructions; and
(vi) A detailed summary of the performance testing, including: test methods, dataset characteristics, triage effectiveness (
*e.g.,* improved time to review of prioritized images for pre-specified clinicians), diagnostic accuracy of algorithms informing triage decision, and results with associated statistical uncertainty (*e.g.,* confidence intervals), including a summary of subanalyses on case distributions stratified by relevant confounders, such as lesion and organ characteristics, disease stages, and imaging equipment.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
RADLogics, Inc. % John J. Smith, M.D., J.D. Partner Hogan Lovells US LPP 555 13th Street NW WASHINGTON DC 20004
Re: K193300
Trade/Device Name: AIMI-Triage CXR PTX Regulation Number: 21 CFR 892.2080 Regulation Name: Radiological computer aided triage and notification software Regulatory Class: Class II Product Code: OFM Dated: March 9, 2020 Received: March 9, 2020
Dear Dr. Smith:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
April 8, 2020
{2}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
510(k) Number (if known)
K193300
Device Name
AIMI-Triage CXR PTX
Indications for Use (Describe)
The AIMI-Triage CXR PTX Application-only triage workflow tool for use by hospital networks and clinics to identify and help prioritize chest X-rays acquired in the acute setting for review by hospital radiologists. The device operates in parallel to and independent of care image interpretation workflow. Specifically, the device uses an artificial intelligence algorithm to analyze images for features suggestive of moderate to large sized pneumothorax; it makes case-level output available to a PACS/workstation for triage. ldentification of suspected cases of moderate to large sized pneumothorax is not for diagnostic use beyond notification.
The AIMI-Triage CXR PTX Application is limited to analysis of imaging data as a guide to possible urgency of adult chest X-ray image review, and should not be used in lieu of full patient evaluation or relied upon to make or confirm diagnoses. Notified radiologists are responsible for engaging in appropriate patient evaluation as per local hospital procedure before making care-related decisions or requests. The device does not replace review and diagnosis of the X-rays by radiologists. The device is not intended to be used with plain film X-rays.
Type of Use (Select one or both, as applicable)
- X Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB num
{4}------------------------------------------------
## 510(k) SUMMARY K193300 RADLogics Inc.'s AIMI-Triage CXR PTX
### Submitter
RADLogics Inc. 24 Westland Ave. #28 Boston, MA 02115
Phone: (408) 905-8054
Contact Person: Moshe Becker
Date Prepared: April 2, 2020
Name of Device: AIMI-Triage CXR PTX
Common or Usual Name: CADt software
Classification Name: Radiological computer-aided triage and notification software
Regulatory Class: II
Product Code: QFM
Predicate Device: Zebra Medical Vision Ltd.'s HealthPNX (K190362)
### Device Description
The AIMI-Triage CXR PTX provides a chest X-ray prioritization service for use by radiologists to identify features suggestive of moderate to large sized pneumothorax. The artificial intelligence algorithm, trained via pattern recognition, processes each chest X-ray and flags those that appear to contain a moderate to large sized pneumothorax for urgent radiologist review. X-rays without an identified anomaly are placed in the worklist for routine review, which is the current standard of care. The user interface is minimal, consisting of the radiologist's existing picture archiving and communication system (PACS) viewer and worklist in which positively identified images are flagged by the software to notify of the suspected anomaly. Images are not marked or otherwise altered, and no diagnoses are provided.
The device does not have any direct accessories. However, it interacts with hospital communication and database systems in order to read and analyze cases in the worklist of the hospital's PACS system in order to identify suspected abnormal findings and transmit corresponding notifications to reflect its recommended prioritization of patient examinations for radiologist review. The software output is compatible with any PACS viewer and worklist.
### Intended Use / Indications for Use
The AIMI-Triage CXR PTX Application is a notification-only triage workflow tool for use by hospital networks and clinics to identify and help prioritize chest X-rays acquired in the acute setting for
{5}------------------------------------------------
review by hospital radiologists. The device operates in parallel to and independent of standard of care image interpretation workflow. Specifically, the device uses an artificial intelligence algorithm to analyze images for features suggestive of moderate to large sized pneumothorax; it makes caselevel output available to a PACS/workstation for worklist prioritization or triage. Identification of suspected cases of moderate to large sized pneumothorax is not for diagnostic use beyond notification.
The AIMI-Triage CXR PTX Application is limited to analysis of imaging data as a guide to possible urgency of adult chest X-ray image review, and should not be used in lieu of full patient evaluation or relied upon to make or confirm diagnoses. Notified radiologists are responsible for engaging in appropriate patient evaluation as per local hospital procedure before making care-related decisions or requests. The device does not replace review and diagnosis of the X-rays by radiologists. The device is not intended to be used with plain film X-rays.
# Summary of Technological Characteristics
AIMI-Triage CXR PTX is technologically similar to the predicate device. Both are based on algorithms which have been trained to analyze images suggestive of pre-defined lung abnormalities and notify an appropriate clinician of these findings in parallel to standard of care image interpretation. The algorithms function similarly and with the same purpose of signaling potential pneumothorax without providing diagnosis or altering the original datasets. There are no notable technological differences between the subject and predicate devices. A table comparing the key features of the subject and predicate devices is provided in the substantial equivalence table below.
## Performance Data
The AIMI-Triage CXR PTX was evaluated and verified in a blinded, retrospective and multi-center study for detecting pneumothorax in chest X-rays. A total of 300 frontal chest X-rays (PA/AP) were collected from US and OUS patients representative of the intended population. 168 (56%) were male with mean age 51.6 years and SD=18.6 (range 18-91), and 132 (44%) were female with mean age 51.8 and SD=16.2 (range 23-86).
The AIMI-Triage CXR PTX output was compared to the ground truth established by 3 independent US-board certified radiologists (Truther involved in the ground truthing process was blinded to any other Truther's results, to any existing report, and to the results obtained by the AlMI-Triage CXR PTX software. Overall, the device was able to demonstrate an area under the curve (AUC) of 96.7% (95.0%, 98.4%)), which is substantially equivalent to the predicate device, and meets the required performance goal. The sensitivity of the device was 92% (95% CI: [86%, 96%]) and 90% (95% CI: [84%, 95%]), respectively.
In addition, the average time it takes the AIMI-Triage CXR PTX to analyze the study and send notification to the PACS worklist was measured to be 20.3 seconds, a timing performance that is substantially equivalent to the predicate (22.1 seconds). Lastly, device performance was evaluated by dataset/region and scanner spatial resolution as summarized in the tables below.
{6}------------------------------------------------
| Category | N | Sensitivity | Specificity | AUROC |
|----------------|-----|----------------------|----------------------|------------------------|
| NIH (US) | 147 | 97.6%<br>(93.2,99.2) | 90.8%<br>(84.5,94.7) | 0.987<br>(0.973,0.999) |
| PADCHEST (OUS) | 153 | 85.3%<br>(79.0,90.8) | 89.7%<br>(83.6,93.9) | 0.949<br>(0.918,0.979) |
## Device Performance by Dataset and Region
## Device Performance by Scanner Spatial Resolution
| Spatial Resolution<br>Category | N | Sensitivity | Specificity | AUROC |
|-----------------------------------------------|-----|----------------------|----------------------|------------------------|
| High range<br><0.145<br>(above 3.45 lp/mm) | 133 | 89.5%<br>(83.0,94.1) | 85.1%<br>(77.7,90.6) | 0.976<br>(0.944,0.999) |
| Mid range<br>0.145-0.170<br>(2.95-3.45 lp/mm) | 88 | 92.6%<br>(84.3,96.7) | 93.4%<br>(85.7,97.4) | 0.983<br>(0.961,0.999) |
| Low range<br>>0.170<br>(below 2.95 lp/mm) | 64 | 93.1%<br>(84.8,98.3) | 91.4%<br>(80.7,96.5) | 0.946<br>(0.911,0.980) |
Thus, the device also performs comparably to the predicate in terms of time savings. The AIMI-Triage CXR PTX has a safety and effectiveness profile that is similar to the predicate device.
# Conclusions
The AIMI-Triage CXR PTX has the same intended use and very similar indications, technological characteristics, and principles of operation as its predicate device. The minor differences in indications do not alter the device's intended use and do not affect its safety and effectiveness when used as labeled. In addition, the minor technological differences between the device and its predicate raise no new issues of safety or effectiveness. Performance data further demonstrate that the AIMI-Triage CXR PTX is as safe and effective as the HealthPNX (K190362). Thus, the subject device is substantially equivalent to the identified predicate.
| Substantial Equivalence Table | | |
|---------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | AIMI-Triage CXR PTX | HealthPNX (K190362) |
| Intended Use / Indications for Use | The AIMI-Triage CXR PTX Application is a notification-only triage workflow tool for use by hospital networks and clinics to identify and help prioritize chest X-rays acquired in the acute setting for review by hospital radiologists. The device operates in parallel to and independent of standard of care image interpretation workflow. Specifically, the device uses an artificial intelligence algorithm to analyze images for features suggestive of moderate to large sized pneumothorax; it makes case-level | The Zebra Pneumothorax device is a software workflow tool designed to aid the clinical assessment of adult Chest X-Ray cases with features suggestive of Pneumothorax in the medical care environment. HealthPNX analyzes cases using an artificial intelligence algorithm to identify suspected |
| | AIMI-Triage CXR PTX | HealthPNX (K190362) |
| | output available to a PACS/workstation for worklist<br>prioritization or triage. Identification of suspected<br>cases of moderate to large sized pneumothorax is<br>not for diagnostic use beyond notification. | findings. It makes case-level<br>output available to a<br>PACS/workstation for worklist<br>prioritization or triage.<br>HealthPNX is not intended to |
| | The AIMI-Triage CXR PTX Application is limited to<br>analysis of imaging data as a guide to possible<br>urgency of adult chest X-ray image review, and<br>should not be used in lieu of full patient evaluation<br>or relied upon to make or confirm diagnoses.<br>Notified radiologists are responsible for engaging in<br>appropriate patient evaluation as per local hospital<br>procedure before making care-related decisions or<br>requests. The device does not replace review and<br>diagnosis of the X-rays by radiologists. The device<br>is not intended to be used with plain film X-rays. | direct attention to specific<br>portions or anomalies of an<br>image. Its results are not<br>intended to be used on a<br>stand-alone basis for clinical<br>decision-making nor is it<br>intended to rule out<br>Pneumothorax or otherwise<br>preclude clinical assessment of<br>X-Ray cases. |
| Classification | 21 C.F.R. § 892,2080, Product Code QFM | |
| User | Radiologist | |
| Anatomical<br>Region | Lung | |
| Clinical<br>Condition | Pneumothorax | |
| Modality | Chest X-ray | |
| Segmentation<br>of region of<br>interest | No; device does not mark, highlight, or direct users' attention to a specific location in<br>the original image | |
| Alteration of<br>original<br>image | No | |
| Relation to<br>standard of<br>care workflow | Independent/parallel; no cases are removed from worklist queue | |
| Algorithm | Artificial intelligence algorithm with database of images | |
| Notification /<br>Prioritization | Yes | |
| Alert to<br>Finding | Passive notification - flagged for review | |
| Where<br>Results are<br>Received | PACS / Workstation | |
### Substantial Equivalence Table
{7}------------------------------------------------
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.