SCANSCOPE XT SYSTEM

K080564 · Aperio Technologies · NOT · Aug 14, 2009 · Hematology

Device Facts

Record IDK080564
Device NameSCANSCOPE XT SYSTEM
ApplicantAperio Technologies
Product CodeNOT · Hematology
Decision DateAug 14, 2009
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 864.1860
Device ClassClass 2

AI Performance

OutputAlgorithmAcceptanceObservedDev DSDev ReadersTest DSTest Readers
HER2/neu (c-erbB-2) protein expression levelDeterministic optimization algorithm using complete search of threshold combinations—Percent agreement between manual microscopy and image analysis ranged from 79.0% to 90.0%100 HER2 slides from clinical site 1, stratified into 0, 1+, 2+, and 3+ classes3 (pathologists)180 formalin-fixed, paraffin-embedded breast tissue specimens from two clinical sites3 (pathologists)

Indications for Use

The ScanScope® System is an automated digital slide creation, management, viewing and analysis system. It is intended for in vitro diagnostic use as an aid to the pathologist in the display, detection, counting and classification of tissues and cells of clinical interest based on particular color, intensity, size, pattern and shape. The IHC HER2 Tunable Image Analysis application is intended for use as an aid to the pathologist in the detection and semi-quantitative measurement of HER2/neu (c-erbB-2) in formalin-fixed, paraffin-embedded normal and neoplastic tissue. The IHC HER2 Tunable Image Analysis application is intended for use as an accessory to the DakoHercepTest™ to aid in the detection and semi-quantitative measurement of HER2/neu (c-erbB-2) in formalin-fixed, paraffin-embedded normal and neoplastic tissue. When used with the Dako HercepTest™, it is indicated for use as an aid in the assessment of breast cancer patients for whom HERCEPTIN® (Trastuzumab) treatment is being considered. Note: The IHC HER2 Image Analysis application is an adjunctive computer-assisted methodology to assist the reproducibility of a qualified pathologist in the acquisition and measurement of images from microscope slides of breast cancer specimens stained for the presence of HER-2 receptor protein. The accuracy of the test result depends upon the quality of the immunohistochemical staining. It is the responsibility of a qualified pathologist to employ appropriate morphological studies and controls as specified in the instructions for the Dako HercepTest™ to assure the validity of the IHC HER2 Image Analysis application assisted HER-2/neu score. The actual correlation of the Dako HercepTest™ to Herceptin® clinical outcome has not been established.

Device Story

System automates digital slide creation, management, viewing, and analysis of IHC-stained breast cancer specimens. Input: high-resolution digitized images of Dako HercepTest™ stained slides. Operation: pathologist outlines tumor regions; software algorithm quantifies membrane staining intensity and completeness. Algorithm uses deterministic optimization based on 20 training slides to set intensity thresholds (0, 1+, 2+, 3+). Output: HER2 score, percentages of 0-3+ cells, and color-coded markup image. Used in clinical pathology labs; operated by pathologists. Output provides quantitative data to supplement qualitative assessment; assists in reproducibility of HER2 scoring; aids clinical decision-making for Trastuzumab therapy. Benefits: improved consistency in HER2 scoring compared to manual microscopy.

Clinical Evidence

Multi-site study (n=180 specimens) compared image analysis to manual microscopy. Primary endpoints: HER2 score agreement between methods and inter-pathologist reproducibility. Results: Percent agreement between image analysis and manual microscopy ranged from 79.0% to 90.0% (95% CI: 69.7%–95.1%). Inter-pathologist agreement for image analysis (85.0%–94.0%) was comparable to or better than manual microscopy (76.3%–91.3%). Precision studies (intra-day, inter-day, inter-system) showed 100% agreement for HER2 scores. No clinical outcome data provided.

Technological Characteristics

System includes automated digital slide scanner, server (Digital Slide Repository), and workstation software. Scanner uses linear-array scanning with 20x/40x magnification. Electrical requirements: 100vAC-240vAC, 50/60Hz. Software utilizes colorimetric pattern recognition algorithms for image analysis. Connectivity includes support for remote access over wide-area networks. Data storage via RAID array and SQL database.

Indications for Use

Indicated for use as an aid to pathologists in the detection and semi-quantitative measurement of HER2/neu protein in formalin-fixed, paraffin-embedded normal and neoplastic breast tissue specimens, specifically for patients being considered for HERCEPTIN (Trastuzumab) treatment.

Regulatory Classification

Identification

Immunohistochemistry test systems (IHC's) are in vitro diagnostic devices consisting of polyclonal or monoclonal antibodies labeled with directions for use and performance claims, which may be packaged with ancillary reagents in kits. Their intended use is to identify, by immunological techniques, antigens in tissues or cytologic specimens. Similar devices intended for use with flow cytometry devices are not considered IHC's.

Special Controls

(2) Class II (special control, guidance document: “FDA Guidance for Submission of Immunohistochemistry Applications to the FDA,” Center for Devices and Radiologic Health, 1998). These IHC's are intended for the detection and/or measurement of certain target analytes in order to provide prognostic or predictive data that are not directly confirmed by routine histopathologic internal and external control specimens. These IHC's provide the pathologist with information that is ordinarily reported as independent diagnostic information to the ordering clinician, and the claims associated with these data are widely accepted and supported by valid scientific evidence. Examples of class II IHC's are those intended for semiquantitative measurement of an analyte, such as hormone receptors in breast cancer.

Predicate Devices

Submission Summary (Full Text)

{0} 1 # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY A. 510(k) Number: k080564 B. Purpose for Submission: Marketing product in the U.S. C. Manufacturer and Instrument Name: Aperio Technologies, Inc. ScanScope® XT System, IHC HER2 Breast Tissue Tunable Image Analysis D. Type of Test or Tests Performed: Computer-assisted image analyzer for immunohistochemistry HER2 slides E. System Descriptions: 1. Device Description: The ScanScope® XT System is an automated digital slide creation, management, viewing and analysis system which consists of an automated digital microscope, slide scanner, computer, color monitor, keyboard, digital pathology information management software and image analysis software. In this particular application, IHC HER2 Breast Tissue Tunable Image Analysis, the image analysis software assists the pathologist in quantitative assessment of immunohistochemistry stained histological specimens for human epidermal growth factor Receptor 2 (HER2) using a tuneable algorithm. This algorithm determines the cell classification thresholds based on a set of 20 training slides. The system software makes no independent interpretations of the data. 2. Principles of Operation: The ScanScope® XT System is intended to provide quantitative input to the pathologist to supplement the quantitative interpretation of HER2 immunohistochemistry stained breast cancer specimens. Formalin-fixed, paraffin embedded breast cancer specimens are stained with the Dako HercepTest™ according to the package insert. Slides are then scanned and digitized at high resolution using the ScanScope XT digital slide scanner. The pathologist then outlines tumor cell only regions and runs the image analysis algorithm. Between 15 and 20 regions and a minimum of 1000 tumor cells should be analyzed to maximize analysis results. Intensity thresholds for the IHC HER2 Breast Tissue Tuneable Image Analysis algorithm are established by a previously scanned training set. Twenty training slides are annotated with a representative set of tumor regions and the "right" scores the algorithm should provide for the slides. Only the three cell classification intensity thresholds (0 to 1+, 1+ to 2+, and 2+ to 3+) are automatically determined by the training algorithm. The training algorithm is a deterministic optimization algorithm that uses a complete search of all possible threshold combinations. The algorithm generates the scores for all possible threshold combinations and compares the algorithm scores to the "right" scores for all slides. The threshold combination that yields the best overall agreement {1} between algorithm scores and "right" scores is the optimum parameter set. The IHC HER2 Breast Tissue Tuneable Image Analysis algorithm then detects the membrane staining for the individual tumor cells in the selected regions and quantifies the intensity and completeness of the membrane staining. Tumor cells are individually classified as 0, 1+, 2+, and 3+ based on their membrane staining intensity and completeness. The HER2 score is then calculated based on the percentages of 0, 1+, 2+, and 3+ cells according to the HER2 scoring scheme. A markup image highlights the detected cell features (black = nuclei and membrane) and the membrane staining which is color-coded according to the cell classification (blue = 0, yellow = 1+, orange = 2+, red = 3+). The pathologist is then provided with HER2 score and the percentages of 0, 1+, 2+, and 3+ cells. The pathologist makes a final call based on both the qualitative and quantitative information and should follow all appropriate instructions in the Dako HercepTest™ product insert. 3. Modes of Operation: Computer-assisted interpretation 4. Specimen Identification: Specimens are identified by slide label (a digital image is taken of the slide label and stored with the digital slide) or by barcode, if provided by the user's laboratory information system. 5. Specimen Sampling and Handling: Immunohistochemical stained microslides can be loaded in the ScanScope XT manually (one at a time) or automatically. The ScanScope XT can automatically scan 120 slides contained in slide racks. 6. Calibration: Calibration of the ScanScope XT is an automated process which is re-verified as part of the scanning process for every scanned slide. If the calibration is not within predefined limits, then the user is prevented from scanning the slide and must take steps to assure that the scan is within acceptable limits. When the user scans a slide, the controller software automatically performs a "prescan". The prescan is a scan of a small region of the slide which contains clear glass or "white space". The brightness and color characteristics of the image are used to correct the resulting scanned image. The main functions of the prescan process are to automatically verify that no significant tissue is present, flatten the illumination field, correct the white balance, and measure bulb brightness. 7. Quality Control: The accuracy of the system depends on the laboratory Dako HercepTest™ kit. 8. Software: FDA has reviewed applicant's Hazard Analysis and Software Development processes for this line of product types: Yes ☐ X ☐ or No ☐ F. Regulatory Information: {2} 1. Regulation section: 21 CFR §864.1860 Immunohistochemistry reagents and kits 2. Classification: Class II 3. Product code: NOT (microscope, automated, image analysis, operator intervention) 4. Panel: Pathology 88 G. Intended Use: 1. Indication(s) for Use: The ScanScope® System is an automated digital slide creation, management, viewing and analysis system. It is intended for in vitro diagnostic use as an aid to the pathologist in the display, detection, counting and classification of tissues and cells of clinical interest based on particular color, intensity, size, pattern and shape. The IHC HER2 Tunable Image Analysis application is intended for use as an aid to the pathologist in the detection and semi-quantitative measurement of HER2/neu (c-erbB-2) in formalin-fixed, paraffin-embedded normal and neoplastic tissue. The IHC HER2 Tunable Image Analysis application is intended for use as an accessory to the DakoHercepTest™ to aid in the detection and semi-quantitative measurement of HER2/neu (c-erbB-2) in formalin-fixed, paraffin-embedded normal and neoplastic tissue. When used with the Dako HercepTest™, it is indicated for use as an aid in the assessment of breast cancer patients for whom HERCEPTIN® (Trastuzumab) treatment is being considered. Note: The IHC HER2 Image Analysis application is an adjunctive computer-assisted methodology to assist the reproducibility of a qualified pathologist in the acquisition and measurement of images from microscope slides of breast cancer specimens stained for the presence of HER-2 receptor protein. The accuracy of the test result depends upon the quality of the immunohistochemical staining. It is the responsibility of a qualified pathologist to employ appropriate morphological studies and controls as specified in the instructions for the Dako HercepTest™ to assure the validity of the IHC HER2 Image Analysis application assisted HER-2/neu score. The actual correlation of the Dako HercepTest™ to Herceptin® clinical outcome has not been established. 2. Special Conditions for Use Statement(s): For prescription use only. H. Substantial Equivalence Information: 1. Predicate Device Name(s) and 510(k) numbers: ChromaVision Medical Systems, Automated Cellular Imaging System (ACIS), k032113 Applied Imaging Ariol™, k031715 2. Comparison with Predicate Device: {3} | Similarities | | | | | --- | --- | --- | --- | | Item | Device | Predicate K032113 | Predicate K031715 | | Device type | Examines formalin-fixed paraffin-embedded breast cancer specimens stained by DakoCytomation HercepTest™ for Her2/neu receptor protein. | Same | Same | | Specimen Type | Formalin-fixed, paraffin-embedded stained by immunohistochemistry | Same | Same | | Device Components | Automated digital slide scanner, computer, color monitor, keyboard, image analysis software and digital pathology information management software | Same | Same | | Differences | | | | | --- | --- | --- | --- | | Item | Device | Predicate | Predicate | | Image algorithm training | Intensity thresholds are established by a previously scanned manually scored training set. The three cell classification intensity thresholds (0 to 1+, 1+ to 2+, and 2+ to 3+) are automatically determined by the training algorithm. | Image algorithm is permanently set and cannot be modified. No training is involved. | User trains the classifiers with breast cancer slides previously manually scored by a pathologist as 1+ and 3+. | I. Special Control/Guidance Document Referenced (if applicable): Guidance for Industry and FDA Staff: Format for Traditional and Abbreviated 510(k)s Guidance for Industry and FDA Staff: Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices J. Performance Characteristics: 1. Analytical Performance: a. Accuracy: A multi-site study was conducted at two clinical sites to compare the performance of Aperio’s IHC HER2 Breast Tissue Tunable Image Analysis system to manual microscopy. One hundred and eighty (180) formalin-fixed, paraffin-embedded breast tissue specimens from two clinical sites were used {4} for this study; 80 specimens with approximately equal HER2 score distribution from site 1 and 100 routine specimens from site 2. The specimens were immunohistochemically stained using Dako's HercepTest. At each site, the IHC HER2 Breast Tissue Tunable Image Analysis system was set up using the automatic algorithm training procedure on a training data set of 20 HER2 slides with approximately equal HER2 score distribution scored independently by three pathologists. At each site, three pathologists performed a blinded read of the glass slides using a microscope and reported the HER2 score for each of the slides. The glass slides were scanned at Aperio using a different ScanScope for each site, and after a wash-out period and randomization of the slides, the same three pathologists remotely viewed and outlined a representative set of tumor regions to be analyzed by the IHC HER2 Breast Tissue Tunable Image Analysis application. The algorithm itself was run in batch mode blinded from the pathologists to avoid any influence of the pathologists in their choice of the tumor regions. The algorithm reported the HER2 score for each of the three pathologists for each of the slides. The statistical analyses are presented across all slides for each of the methods: manual microscopy and image analysis, and comparatively between methods for manual microscopy against image analysis. 5 {5} The pair wise observations of the HER2 Score categories are summarized in 4x4 tables. | | Pathologist 1 | | | | | | | --- | --- | --- | --- | --- | --- | --- | | | | 0 | 1+ | 2+ | 3+ | Total | | Pathologist 2 | 0 | 24 | 1 | | | 25 | | | 1+ | 9 | 15 | 3 | | 27 | | | 2+ | 1 | 1 | 10 | 2 | 14 | | | 3+ | | | 0 | 14 | 14 | | | Total | 34 | 17 | 13 | 16 | 80 | | | Pathologist 1 | | | | | | | --- | --- | --- | --- | --- | --- | --- | | | | 0 | 1+ | 2+ | 3+ | Total | | Pathologist 3 | 0 | 19 | 1 | 1 | | 21 | | | 1+ | 12 | 2 | | | 14 | | | 2+ | 3 | 14 | 12 | | 29 | | | 3+ | | | | 16 | 16 | | | Total | 34 | 17 | 13 | 16 | 80 | | | Pathologist 2 | | | | | | | --- | --- | --- | --- | --- | --- | --- | | | | 0 | 1+ | 2+ | 3+ | Total | | Pathologist 3 | 0 | 17 | 4 | | | 21 | | | 1+ | 6 | 8 | | | 14 | | | 2+ | 2 | 15 | 12 | | 29 | | | 3+ | | | 2 | 14 | 16 | | | Total | 25 | 27 | 14 | 14 | 80 | Manual Microscopy - Clinical Site 1 - Inter-Pathologists | | Pathologist 1 | | | | | | | --- | --- | --- | --- | --- | --- | --- | | | | 0 | 1+ | 2+ | 3+ | Total | | Pathologist 2 | 0 | 15 | 3 | | | 18 | | | 1+ | 5 | 30 | 5 | | 40 | | | 2+ | | 3 | 19 | 6 | 28 | | | 3+ | | | 2 | 12 | 14 | | | Total | 20 | 36 | 26 | 18 | 100 | | | Pathologist 1 | | | | | | | --- | --- | --- | --- | --- | --- | --- | | | | 0 | 1+ | 2+ | 3+ | Total | | Pathologist 3 | 0 | 17 | 8 | 2 | | 27 | | | 1+ | 3 | 28 | 9 | | 40 | | | 2+ | | | 14 | 6 | 20 | | | 3+ | | | 1 | 12 | 13 | | | Total | 20 | 36 | 26 | 18 | 100 | | | Pathologist 2 | | | | | | | --- | --- | --- | --- | --- | --- | --- | | | | 0 | 1+ | 2+ | 3+ | Total | | Pathologist 3 | 0 | 16 | 10 | 1 | | 27 | | | 1+ | 2 | 30 | 8 | | 40 | | | 2+ | | | 19 | 1 | 20 | | | 3+ | | | | 13 | 13 | | | Total | 18 | 40 | 28 | 14 | 100 | Manual Microscopy - Clinical Site 2 - Inter-Pathologists {6} | | Pathologist 1 | | | | | | | --- | --- | --- | --- | --- | --- | --- | | | | 0 | 1+ | 2+ | 3+ | Total | | Pathologist 2 | 0 | 18 | 0 | 0 | 0 | 18 | | | 1+ | 7 | 23 | 2 | 0 | 32 | | | 2+ | 0 | 0 | 10 | 0 | 10 | | | 3+ | 0 | 0 | 5 | 15 | 20 | | | Total | 25 | 23 | 17 | 15 | 80 | | | Pathologist 1 | | | | | | | --- | --- | --- | --- | --- | --- | --- | | | | 0 | 1+ | 2+ | 3+ | Total | | Pathologist 3 | 0 | 18 | 1 | 0 | 0 | 19 | | | 1+ | 7 | 20 | 1 | 1 | 29 | | | 2+ | 0 | 2 | 14 | 0 | 16 | | | 3+ | 0 | 0 | 2 | 14 | 16 | | | Total | 25 | 23 | 17 | 15 | 80 | | | Pathologist 2 | | | | | | | --- | --- | --- | --- | --- | --- | --- | | | | 0 | 1+ | 2+ | 3+ | Total | | Pathologist 3 | 0 | 16 | 3 | 0 | 0 | 19 | | | 1+ | 2 | 26 | 0 | 1 | 29 | | | 2+ | 0 | 3 | 9 | 4 | 16 | | | 3+ | 0 | 0 | 1 | 15 | 16 | | | Total | 18 | 32 | 10 | 20 | 80 | Image Analysis - Clinical Site 1 - Inter-Pathologists | | Pathologist 1 | | | | | | | --- | --- | --- | --- | --- | --- | --- | | | | 0 | 1+ | 2+ | 3+ | Total | | Pathologist 2 | 0 | 15 | 2 | 0 | 0 | 17 | | | 1+ | 3 | 35 | 8 | 0 | 46 | | | 2+ | 0 | 5 | 16 | 2 | 23 | | | 3+ | 0 | 0 | 0 | 14 | 14 | | | Total | 18 | 42 | 24 | 16 | 100 | | | Pathologist 1 | | | | | | | --- | --- | --- | --- | --- | --- | --- | | | | 0 | 1+ | 2+ | 3+ | Total | | Pathologist 3 | 0 | 14 | 1 | 0 | 0 | 15 | | | 1+ | 4 | 39 | 2 | 0 | 45 | | | 2+ | 0 | 2 | 21 | 1 | 24 | | | 3+ | 0 | 0 | 1 | 15 | 16 | | | Total | 18 | 42 | 24 | 16 | 100 | | | Pathologist 2 | | | | | | | --- | --- | --- | --- | --- | --- | --- | | | | 0 | 1+ | 2+ | 3+ | Total | | Pathologist 3 | 0 | 13 | 2 | 0 | 0 | 15 | | | 1+ | 4 | 37 | 4 | 0 | 45 | | | 2+ | 0 | 7 | 17 | 0 | 24 | | | 3+ | 0 | 0 | 2 | 14 | 16 | | | Total | 17 | 46 | 23 | 14 | 100 | Image Analysis - Clinical Site 2 - Inter-Pathologists {7} | Pathologist 1 | Image Analysis | | | | | | | --- | --- | --- | --- | --- | --- | --- | | | | 0 | 1+ | 2+ | 3+ | Total | | Manual Microscopy | 0 | 22 | 11 | 1 | 0 | 34 | | | 1+ | 2 | 10 | 5 | 0 | 17 | | | 2+ | 1 | 2 | 10 | 0 | 13 | | | 3+ | 0 | 0 | 1 | 15 | 16 | | | Total | 25 | 23 | 17 | 15 | 80 | | Pathologist 2 | Image Analysis | | | | | | | --- | --- | --- | --- | --- | --- | --- | | | | 0 | 1+ | 2+ | 3+ | Total | | Manual Microscopy | 0 | 17 | 8 | 0 | 0 | 25 | | | 1+ | 1 | 23 | 3 | 0 | 27 | | | 2+ | 0 | 1 | 7 | 6 | 14 | | | 3+ | 0 | 0 | 0 | 14 | 14 | | | Total | 18 | 32 | 10 | 20 | 80 | | Pathologist 3 | Image Analysis | | | | | | | --- | --- | --- | --- | --- | --- | --- | | | | 0 | 1+ | 2+ | 3+ | Total | | Manual Microscopy | 0 | 17 | 4 | 0 | 0 | 21 | | | 1+ | 2 | 12 | 0 | 0 | 14 | | | 2+ | 0 | 12 | 15 | 2 | 29 | | | 3+ | 0 | 1 | 1 | 14 | 16 | | | Total | 19 | 29 | 16 | 16 | 80 | Manual Microscopy vs. Image Analysis - Clinical Site 1 - same Pathologist | Pathologist 1 | Image Analysis | | | | | | | --- | --- | --- | --- | --- | --- | --- | | | | 0 | 1+ | 2+ | 3+ | Total | | Manual Microscopy | 0 | 15 | 5 | 0 | 0 | 20 | | | 1+ | 3 | 32 | 1 | 0 | 36 | | | 2+ | 0 | 5 | 20 | 1 | 26 | | | 3+ | 0 | 0 | 3 | 15 | 18 | | | Total | 18 | 42 | 24 | 16 | 100 | | Pathologist 2 | Image Analysis | | | | | | | --- | --- | --- | --- | --- | --- | --- | | | | 0 | 1+ | 2+ | 3+ | Total | | Manual Microscopy | 0 | 14 | 4 | 0 | 0 | 18 | | | 1+ | 3 | 32 | 5 | 0 | 40 | | | 2+ | 0 | 10 | 15 | 3 | 28 | | | 3+ | 0 | 0 | 3 | 11 | 14 | | | Total | 17 | 46 | 23 | 14 | 100 | | Pathologist 3 | Image Analysis | | | | | | | --- | --- | --- | --- | --- | --- | --- | | | | 0 | 1+ | 2+ | 3+ | Total | | Manual Microscopy | 0 | 13 | 12 | 2 | 0 | 27 | | | 1+ | 2 | 33 | 5 | 0 | 40 | | | 2+ | 0 | 0 | 16 | 4 | 20 | | | 3+ | 0 | 0 | 1 | 12 | 13 | | | Total | 15 | 45 | 24 | 16 | 100 | Manual Microscopy vs. Image Analysis - Clinical Site 2 - same Pathologist {8} Statistical analyses are provided for a trichotomous categorization of the HER2 scores combining 0 and 1+ and leaving 2+ and 3+ uncombined. Percentage Agreement (PA) along with an exact 95% Confidence Interval (CI) are presented overall for all trichotomous HER2 score categories combined and for each of the trichotomous HER2 score categories separately using a dichotomous outcome of that category vs. the two other categories. | | Pathologist 1 v 2 | | Pathologist 1 v 3 | | Pathologist 2 v 3 | | | --- | --- | --- | --- | --- | --- | --- | | | PA | PA 95% CI | PA | PA 95% CI | PA | PA 95% CI | | Clinical Site 1 | 91.3% | (82.8, 96.4) | 77.5% | (66.8, 86.1) | 76.3% | (65.4, 85.1) | | Clinical Site 2 | 84.0% | (75.3, 90.6) | 82.0% | (73.1, 89.0) | 90.0% | (82.4, 95.1) | Manual Microscopy - Inter-Pathologists - Agreements. | | Pathologist 1 v 2 | | Pathologist 1 v 3 | | Pathologist 2 v 3 | | | --- | --- | --- | --- | --- | --- | --- | | | PA | PA 95% CI | PA | PA 95% CI | PA | PA 95% CI | | Clinical Site 1 | 91.3% | (82.8, 96.4) | 92.5% | (84.4, 97.2) | 88.8% | (79.7, 94.7) | | Clinical Site 2 | 85.0% | (76.5, 91.4) | 94.0% | (87.4, 97.8) | 87.0% | (78.8, 92.9) | Image Analysis - Inter-Pathologists - Agreements. | | Pathologist 1 | | Pathologist 2 | | Pathologist 3 | | | --- | --- | --- | --- | --- | --- | --- | | | PA | PA 95% CI | PA | PA 95% CI | PA | PA 95% CI | | Clinical Site 1 | 87.5% | (78.2, 93.8) | 87.5% | (78.2, 93.8) | 80.0% | (69.6, 88.1) | | Clinical Site 2 | 90.0% | (82.4, 95.1) | 79.0% | (69.7, 86.5) | 88.0% | (80.0, 93.6) | Manual Microscopy vs Image Analysis - same Pathologist - Agreements. The percent agreements between the pathologists' manual microscopy and Image Analysis ranged from 79.0% to 90.0% with confidence bounds from 69.7% to 95.1%; the inter-pathologists agreements for manual microscopy ranged from 76.3% to 91.3% with confidence bounds from 65.4% to 96.4% and the inter-pathologists agreements for Image Analysis ranged from 85.0% to 94.0% with confidence bounds from 76.5% to 97.8%. Note that these image analysis results were obtained by having the Pathologists choose and outline a representative set of tumor regions anywhere on the entire slide, completely blinded from each other, and blinded from the image analysis results (there was no influence on the Pathologists in their choice of the tumor regions). b. Precision: Eight HER2 slides with two slides per HER2 score 0, 1+, 2+ and 3+ were sampled from one of the clinical sites that used Dako's HercepTest to be used in a suite of precision studies. The slides were sampled in sequential order using the rounded average score of the manual microscopy scores provided by the three pathologists. {9} 10 | | Slide 1 | Slide 2 | Slide 3 | Slide 4 | Slide 5 | Slide 6 | Slide 7 | Slide 8 | | --- | --- | --- | --- | --- | --- | --- | --- | --- | | Pathologist 1 | 0 | 0 | 0 | 1 | 2 | 2 | 3 | 3 | | Pathologist 2 | 0 | 0 | 1 | 1 | 2 | 2 | 3 | 3 | | Pathologist 3 | 1 | 0 | 1 | 0 | 2 | 2 | 3 | 3 | | Average | 0 | 0 | 1 | 1 | 2 | 2 | 3 | 3 | HER2 scores provided by 3 Pathologists for the sampled slides. Separate studies were conducted to analyze the system variability separately from the variability introduced by the pathologists outlining the tumor regions for the analysis. System precision studies used the same tumor regions for analysis over all runs to eliminate the influence by the pathologists. Pathologist precision studies used the same digital slides to outline tumor regions and run the analysis to eliminate the influence of the system. The precision studies analyzed the changes in the system response by extending the analysis of the coarse HER2 score to the underlying cumulative percentages of 3+, 2+ and 1+ cells on which the HER2 score calculations are based; allowing detecting and quantifying smaller changes of the system. ## Intra-Day/Intra-System The eight HER2 slides were scanned 10 times on the same ScanScope system. The image analysis results show perfect agreement (100%) for the calculated HER2 scores and an overall average standard deviation of 0.70% (maximum 2.46%) and average range (maximum – minimum) of 1.30% (maximum 7.14%) for the cumulative percentages of 3+, 2+ and 1+ cells (range from 0.0 to 100.0%) across all runs. ## Inter-Day/Intra-System The eight HER2 slides were scanned on the same ScanScope system over 20 times on different days. The image analysis results show perfect agreement (100%) for the calculated HER2 scores and an overall average standard deviation of 0.69% (maximum 2.43%) and average range of 1.75% (maximum 12.07%) for the cumulative percentages of 3+, 2+ and 1+ cells across all runs. ## Inter-System The same eight HER2 slides were scanned 10 times on three different ScanScope systems. The image analysis results show perfect agreement (100%) for the calculated HER2 scores across all systems and all runs. The image analysis results on each of the three ScanScope systems show an overall average standard deviation of 0.70%, 0.56% and 0.57% (maximum 2.46%, 1.65% and 1.34%) and average range of 1.30%, 1,07% and 1,17% (maximum 7.14%, 5.09% and 4.70%) for the cumulative percentages of 3+, 2+ and 1+ cells respectively over all runs. The image analysis results of the three ScanScope systems combined show an overall average standard deviation of 0.80% (maximum 2.41%) and average range of 1.94% {10} (maximum 8.95%) for the cumulative percentages of 3+, 2+ and 1+ cells respectively over all runs. ## Intra-Pathologist One pathologist outlined the tumor regions for analysis on the same eight HER2 slides 5 times. The image analysis results show 4 cases out of 40 (10%) where the HER2 scores differed from the median HER2 score and an overall average standard deviation of 2.69% (maximum 8.08%) and average range of 3.90% (maximum 18.61%) for the cumulative percentages of 3+, 2+ and 1+ cells. ## Inter-Pathologists Three pathologists outlined the tumor regions for analysis on the same eight HER2 slides as part of the clinical comparison to manual microscopy study. The image analysis results show 3 cases out of 24 (12.5%) where the HER2 scores differed from the median HER2 score and an overall average standard deviation of 10.03% (maximum 27.09%) and average range of 11.74% (maximum 48.26%) for the cumulative percentages of 3+, 2+ and 1+ cells (range from 0.0 to 100.0%). A summary of the overall average and maximum Standard Deviation (SD) and range for the different precision studies is shown in the following table. | | | Average SD | Maximum SD | Average Range | Maximum Range | | --- | --- | --- | --- | --- | --- | | Intra-Run/Intra-System | | 0.70% | 2.46% | 1.30% | 7.14% | | Inter-Run/Intra-System | | 0.69% | 2.43% | 1.75% | 12.07% | | Inter-System | ScanScope #1 | 0.70% | 2.46% | 1.30% | 7.14% | | | ScanScope #2 | 0.56% | 1.65% | 1.07% | 5.09% | | | ScanScope #3 | 0.57% | 1.34% | 1.17% | 4.70% | | | Combined | 0.80% | 2.41% | 1.94% | 8.95% | | Intra-Pathologist | | 2.69% | 8.08% | 3.90% | 18.61% | | Inter-Pathologist | | 10.03% | 27.09% | 11.74% | 48.26% | ## Algorithm Training Set 100 HER2 slides from clinical site 1 were stratified into 0, 1+, 2+ and 3+ classes based on the average HER2 score provided by three pathologists using manual microscopy. Three different algorithm training and evaluation runs were conducted. Each time, the 100 slides were separated into a training data set and an evaluation data set. The training data set consisted of 5 slides for each 0, 1+, 2+ and 3+ HER2 class that were selected randomly from the available slides within the HER2 classes (stratified-random selection)—a total of 20 slides. The remaining 80 slides were used as the evaluation data set. The training data set was used to tune the IHC HER2 Breast Tissue Tunable Image Analysis {11} application according to the procedure outlined in the previous sections of this chapter. The tuned HER2 image analysis application was then run on the 80 slides of the test data set using the tumor region outlines provided by three pathologists during the digital read in the substantial equivalence study. The inter-pathologists variations for manual microscopy and image analysis as well as the inter-method variations are reported as previously in the substantial equivalence study. The agreements between the pathologists' manual microscopy and Image Analysis ranged from 75.0% to 88.8% with confidence bounds from 66.5% to 95.7%; the inter-pathologists agreements for manual microscopy ranged from 75% to 90% with confidence bounds from 66.5% to 96.6% and the inter-pathologists agreements for Image Analysis ranged from 86.3% to 92.5% with confidence bounds from 78.7% to 97.7%. c. Linearity: Not applicable. d. Carryover: Not applicable. e. Interfering Substances: Not applicable. 2. Other Supportive Instrument Performance Data Not Covered Above: Precision study to assess variability due to different training sets: The sponsor did not perform studies to assess how the use of different training sets affects the performance of the device. FDA requested a precision study to demonstrate that using different training sets does not affect performance. While these results did show different threshold estimations for each of the three training sets the difference was not reflected in the image analysis HER2 score determination. Disagreements in HER2 scoring more than 2 degrees, i.e. 0 to 2+, or 1+ to 3+, could be a cause for concern as the results determine patient treatment. There were no such differences for the image analysis system while there were 4 such disagreements for manual scoring (reference method). The different threshold estimations lack of disagreement for the image analysis slides could be an artifact of the sample set selected; still there is less disagreement compared to the reference method. The risk of a miscall due to variation in the training sets is mitigated by mandatory pathologist review of all results and labeling stressing the importance of test quality control and validation. K. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. L. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision. 12
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...