The Vibrant System is an orally administered Capsule that is indicated for the treatment of adults with chronic idiopathic constipation who have not experienced relief of their bowel symptoms by using laxative therapies at the recommended dosage for at least one month.
Device Story
Vibrant System is an orally ingested, electric, swallowable capsule designed for treatment of chronic idiopathic constipation. Capsule travels through gastrointestinal tract; provides mechanical stimulation to colon to promote bowel movements. Device is prescription-only; intended for adult patients. System includes capsule and associated components; operates via internal electrical mechanism. Clinical benefit derived from mechanical stimulation of colon wall, aiding motility in patients refractory to laxative therapy. Safety profile includes monitoring for potential adverse events like abdominal discomfort, nausea, or failure to excrete capsule. Healthcare providers use clinical data and patient history to determine suitability for therapy.
Clinical Evidence
Prospective, randomized, multi-center, double-blind, placebo-controlled pivotal study (V270) of 349 subjects. Primary endpoints: CSBM1 (≥1 increase) and CSBM2 (≥2 increase) in weekly CSBMs for 6 of 8 weeks. mITT results: CSBM1 success 40.51% (treatment) vs 22.92% (control), p=0.0011; CSBM2 success 23.42% (treatment) vs 11.81% (control), p=0.0085. Secondary endpoints showed improved stool consistency; no difference in bloating. Safety: 14.5% related AEs in treatment vs 3.36% in control; most common AE was vibration sensation.
Technological Characteristics
Electric swallowable capsule; transient gastrointestinal use. Components must be biocompatible. Performance requirements include dimensional verification, leak testing, bite force resistance, and pH resistance. Electrical components require safety and electromagnetic compatibility (EMC) testing. Software validation, verification, and hazard analysis are required. Shelf life must be supported by package integrity and functionality testing. Bioburden testing required for infection risk mitigation.
Indications for Use
Indicated for adults with chronic idiopathic constipation who have failed to achieve relief with laxative therapies at recommended dosages for at least one month.
Regulatory Classification
Identification
An orally ingested transient device for constipation is an electric swallowable capsule that naturally passes through the gastrointestinal tract for the treatment of constipation. The Vibrant System is indicated for the treatment of adults with chronic idiopathic constipation who have not experienced relief of their bowel symptoms by using laxative therapies at the recommended dosage for at least one month.
Special Controls
In combination with the general controls of the FD&C Act, an orally ingested transient device for constipation is subject to the following special controls:
Submission Summary (Full Text)
{0}------------------------------------------------
### DE NOVO CLASSIFICATION REQUEST FOR VIBRANT SYSTEM
#### REGULATORY INFORMATION
FDA identifies this generic type of device as:
Orally ingested transient device for constipation. An orally ingested transient device for constipation is an electric swallowable capsule that naturally passes through the gastrointestinal tract for the treatment of constipation.
NEW REGULATION NUMBER: 21 CFR 876.5940
CLASSIFICATION: Class II
PRODUCT CODE: QTN
#### BACKGROUND
DEVICE NAME: Vibrant System
SUBMISSION NUMBER: DEN210052
DATE DE NOVO RECEIVED: December 1, 2021
SPONSOR INFORMATION: Vibrant Ltd. Hakochav P.O. Box 516 Yokneam, Israel, 2069206
#### INDICATIONS FOR USE
The Vibrant System is an orally administered Capsule that is indicated for the treatment of adults with chronic idiopathic constipation who have not experienced relief of their bowel symptoms by using laxative therapies at the recommended dosage for at least one month.
#### LIMITATIONS
The sale, distribution, and use of the Vibrant System is restricted to prescription use in accordance with 21 CFR 801.109.
The Vibrant capsules are contraindicated for use under the following conditions:
- History of complicated/obstructive diverticular disease .
- History of intestinal or colonic obstruction, or suspected intestinal obstruction .
- Clinical evidence of current and significant gastroparesis .
{1}------------------------------------------------
- History of significant gastrointestinal disorder, including any form of inflammatory . bowel disease or gastrointestinal malignancy (celiac disease is accepted if the subject has been treated and is in remission), and/or anal fissures and fistulas
- History of Zenker's diverticulum, dysphagia, esophageal stricture, eosinophilic . esophagitis, or achalasia
- . Women who are pregnant or lactating
# PLEASE REFER TO THE LABELING FOR A COMPLETE LIST OF WARNINGS, PRECAUTIONS AND CONTRAINDICATIONS.
### DEVICE DESCRIPTION
The Vibrant System is comprised of two components; a Pod and a single use Capsule. The Pod activates the Capsule using radio frequency (RF) communication. After activation, the Capsules are taken orally. After a pre-programmed activation delay, the Capsules begin to vibrate. The marketed device will have an activation delay of 14 hours. (The 14 hours activation delay corresponds to Active Mode 1 of the pivotal study described below: the pivotal study also included an Active Mode 2 with a 9 hour activation delay.) As they pass through the gastrointestinal tract, the Capsules vibrate for four pre-programmed periods of 100 min, 120 min, 100 min. 180 min. It takes about three days for the Capsules to pass through the gastrointestinal tract, at which point thev are excreted in a bowel movement.
The Capsule is comprised of a motor that vibrates (2.5 Vdc. 70 ma), electronic card, batteries (2x 1.5V 60 mAh, alkaline/0% Hg coin battery) and connector. The capsule is a two-piece shell made from Polycarbonate Makrolon 2458. The Capsule dimensions are as follows: length: 24.5 ± 0.1 mm; diameter: 11.20 ± 0.05 mm; weight 4.0 ± 0.1 g. The Capsules have a vibration force between 200 and 550 gf, with vibration cycles of 3 seconds on and16 seconds off (190 cycles per hour).
Figure 1: Vibrant System: Capsule
{2}------------------------------------------------
Figure 2: Vibrant System: Pod
Image /page/2/Picture/1 description: The image shows a white and blue device with a clear plastic lid that is open. The device has two circular indentations on the top, and a small, round, light-colored object is in the center of the indentations. There is a green light on the lower left corner of the device, and a black rectangular button is located near the bottom center.
### SUMMARY OF NONCLINICAL/BENCH STUDIES
#### BIOBURDEN AND SHELF LIFE
The Vibrant System is not provided sterile. Vibrant Capsules are routinely monitored during manufacturing for bioburden levels to ensure the total bioburden is no greater than forming units per capsule. Packaging and shelf-life testing confirmed the Vibrant System packaging maintained integrity and device performance over the 1-year labeled shelf life. See Table 1 for list of bench testing conducted to support shelf-life.
#### SOFTWARE
The Vibrant System contains software to control the operation and vibration of the Capsule and for activation and time calibration of the Capsule by the Pod. Software verification and validation testing was performed and the results demonstrated the software functions as intended and all requirements specifications were met. The software/firmware was reviewed according to the "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices," dated May 11, 2005. The software has a moderate level of concern.
{3}------------------------------------------------
# ELECTROMAGNETIC COMPATIBILITY AND ELECTRICAL SAFETY
The Vibrant System conforms to the requirements for electrical safety and electromagnetic compatibility in ANSI AAMI ES60601-1:2005/@2012, A1:2012, C1:2009/(R)2012 and A2:2010/(R)2012; IEC 60601-1-11:2015; IEC 60601-1-2:2014.
# BIOCOMPATIBILITY
The Vibrant capsules are classified as mucosal membrane contacting with a cumulative duration of exposure of >30 days. Capsule biocompatibility was evaluated in accordance with ISO 10993-1, Biological evaluation of medical devices and FDA Guidance: Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process". The following biocompatibility endpoints were evaluated with testing and chemical characterization/toxicological risk assessment:
- . cytotoxicity
- sensitization .
- irritation .
- acute systemic toxicity .
- materials mediated pyrogenicity .
- . subchronic toxicity
- genotoxicity .
- . chronic toxicity
The results of these evaluations support the biocompatibility of the Vibrant Capsules.
The Vibrant System Pod is non-patient contacting and biocompatibility assessment was not necessary.
# PERFORMANCE TESTING - BENCH
The performance of the Vibrant Capsule was evaluated with the nonclinical bench testing summarized in Table 1.
| Test | Test Methods | Acceptance Criteria | Results |
|----------------------|----------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| Mechanical Integrity | Capsules were placed under loads ranging from (b)(4) to (b)(4). The capsule was subjected to (b)(4) for (b)(4) | No cracks, ruptures, loose particles, or debris are observable following (b)(4) application of (b)(4) | Pass |
| Seal test | Capsule ultrasonic welds were evaluated for seal integrity using leak tester (b)(4) | No major leak, and the pressure of the test capsule must be less than (b)(4) the control capsule pressure. | Pass |
| pH Resistance | Capsules were incubated for (b)(4) in the worse-case pH conditions of pH (b)(4) | Capsules must not show defects in the seal, no ruptures, no shell cracks, or any other damage to the capsule. The Capsules must also pass the sealing | Pass |
# Table 1: Summary of Performance Testing - Bench Studies
{4}------------------------------------------------
| Test | Test Methods | Acceptance Criteria | Results |
|-----------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| Mode of Operation | Following Capsule activation, the<br>following parameters were recorded:<br>vibration start time, vibration<br>duration, vibration force, vibration<br>sequence and rest time. These<br>parameters were recorded for each of<br>the four vibration rounds. | Capsules must meet the predefined<br>specifications for start time, duration,<br>vibrational force, vibrational<br>sequence. | Pass |
| Dimensional<br>Verification | A digital caliper was used to measure<br>the length and diameter of the<br>capsules. | Capsule length must be<br><span style="background-color: #ffaaaa">(b)(4)</span><br>Capsule diameter must be<br><span style="background-color: #ffaaaa">(b)(4)</span> | Pass |
| Visual Inspection | The Capsules were visually inspected<br>with a magnifying glass to check the<br>Capsule for correct assembly of the<br>shell components, check for any<br>damage to the shell (e.g., scratches,<br>kinks), and to check for the presence<br>of extraneous matter on the shell. | External surface of Capsule is free of<br>surface defects and extraneous matter. | Pass |
### PERFORMANCE TESTING - ANIMAL
An earlier version of the Vibrant Capsule was evaluated in proof-of-concept [ by ]studies. The testing evaluated preliminary safety and performance of the capsules. The studies did not identify any significant safety signals.
# SUMMARY OF CLINICAL INFORMATION
A pivotal study (V270) was conducted to evaluate the safety and effectiveness of the Vibrant Capsules. The study was conducted in compliance with 21 CFR parts 50. 56, and 812. V270 was a prospective, randomized, multi-center, double-blinded, placebo-controlled pivotal study. The study included two active treatment arms that used different vibration patterns (i.e., Active Mode I and Active Mode 2) and a control arm. For the first part of the study, subjects were randomized equally to each arm in a 1:1:1 ratio, for a treatment period of 8 weeks. A "drop the loser" design was pre-specified, whereby the lower performing active treatment was dropped from the study following an interim analysis. After the lower performing active arm was dropped, future subjects were randomized to each of the remaining arms on a 1:1 ratio. The alpha value was adjusted to two-sided 0.025 to account for this design.
Vibrant Capsules were administered 5 times per week, and the treatment arm was compared to a control arm. The sham used in the control arm was a placebo consisting of a biodegradable softgel capsule filled with soybean oil, beeswax, and calcium carbonate. The study population was comprised of subjects with chronic idiopathic constipation (CIC) who were refractory to existing OTC or Rx treatments or could not tolerate with the side effects. Patient eligibility was assessed during the initial screening visit and was re-assessed and confirmed during the Baseline visit following a 2-3-week run-in period. Randomization occurred following the 2-3-week run-in
{5}------------------------------------------------
period and verification that the patient satisfied all inclusion criteria and did not meet any exclusion criteria.
# Endpoints
### Primary Effectiveness endpoint
- Complete Spontaneous Bowel Movements (CSBM1) success rate, defined as an increase . from the run-in period of at least one weekly CSBM during at least 6 of the 8 weeks of treatment.
- · Complete Spontaneous Bowel Movements (CSBM2) success rate, defined as an increase from the run-in period of at least two weekly CSBM during at least 6 of the 8 weeks of treatment.
Study success was defined as either the CSBM1 or the CSBM2 success rate being statistically significantly higher in the active arm that was continued after interim analysis, than in the control arm, as evaluated in the modified Intent to Treat (mITT) population, which is described in the "Subject Accountability" section, below.
### Primary Safety endpoint
All adverse events (AEs) related and unrelated to the study treatment, in the Intent to Treat (ITT) population, as described in the "Subject Accountability" section, below.
### Secondary endpoints
- Change from baseline in average straining, using a scale from 0 (no straining) to 10 . (unbearable straining).
- Change from baseline in average stool consistency, using the Bristol Stool Scale. .
- . Change from baseline in average bloating, using a scale from 0 (no bloating) to 10 (unbearable bloating).
# Selected additional secondary endpoints
- Treatment satisfaction score using the Treatment Satisfaction Questionnaire for Medication . (TSQM)
- . Change from baseline in quality of life using the Patient Assessment of Constipation Quality of Life Ouestionnaire (PAC-OOL) questionnaire
# Subject accountability
A total of 904 subjects were consented and screened. 555 subjects were not randomized: 89% (493/555) were considered screen failures, 5% (28/555) withdrew consent prior to randomization, 2% (10/555) were lost to follow up before randomization, and the other 4% (24/555) for other reasons. A total of 349 consenting subjects were enrolled in the study and randomized. It was subsequently determined that one of the 349 randomized subjects did not meet the eligibility criteria, and 12/349 subjects had less than 2 weeks of diary entries with at least 5 days/week (requirement for randomization); these subjects were thus included in the Intention to Treat (ITT) analysis set but excluded from the modified ITT (mITT) analysis set. Three subjects in the mITT analysis set had major protocol deviations and were thus excluded from the Per Protocol (PP) analysis set. Subjects with less than 6 weeks of diary data were
{6}------------------------------------------------
considered as failures for the CSBM1, and CSBM2 success rates. The final treatment allocation and analysis sets are shown in Table 2.
| | Active Mode 1 | Active Mode 2 | Control | Total |
|-------------------|---------------|---------------|---------|-------|
| ITT analysis set | 163 | 37 | 149 | 349 |
| mITT analysis set | 158 | 34 | 144 | 336 |
| PP analysis set | 155 | 34 | 144 | 333 |
Table 2: Treatment Allocation
Patient demographics are shown in Table 3.
| | | | Active Mode 1 | Active Mode 2 | Control | Total | p-value (Active Mode<br>1 vs Control) |
|----------------|---------------------------------|-------------------|---------------------|---------------------|---------------------|---------------------|---------------------------------------|
| Age, | | N | 163 | 37 | 149 | 349 | |
| years | | Mean<br>(SD) | 47.1<br>(13.33) | 45.0<br>(12.25) | 45.9<br>(13.47) | 46.4<br>(13.26) | 0.4391(*) |
| | | Median<br>[Range] | 47.4<br>[20.6;77.8] | 44.0<br>[22.2:76.0] | 44.8<br>[22.1;78.1] | 45.0<br>[20.6:78.1] | |
| Gender | Male | %<br>(n/N) | 12.3%<br>(20/163) | 24.3%<br>(9/37) | 15.4%<br>(23/149) | 14.9%<br>(52/349) | |
| | Female | %<br>(n/N) | 87.7%<br>(143/163) | 75.7%<br>(28/37) | 84.6%<br>(126/149) | 85.1%<br>(297/349) | 0.4177(#) |
| Race/Ethnicity | Caucasian | %<br>(n/N) | 47.2%<br>(77/163) | 59.5%<br>(22/37) | 40.3%<br>(60/149) | 45.6%<br>(159/349) | 0.7624(#) |
| | Hispanic<br>or Latino | %<br>(n/N) | 19.0%<br>(31/163) | 8.1%<br>(3/37) | 22.8%<br>(34/149) | 19.5%<br>(68/349) | |
| | Black or<br>African<br>American | %<br>(n/N) | 25.2%<br>(41/163) | 29.7%<br>(11/37) | 26.2%<br>(39/149) | 26.1%<br>(91/349) | |
| | Asian/<br>Pacific<br>Islander | %<br>(n/N) | 6.1%<br>(10/163) | 2.7%<br>(1/37) | 8.1%<br>(12/149) | 6.6%<br>(23/349) | |
| | Other | %<br>(n/N) | 2.5%<br>(4/163) | - | 2.7%<br>(4/149) | 2.3%<br>(8/349) | |
Table 3: Demographic Characteristics - ITT
(*) t-test
(#) x2 test
# Effectiveness results
The pre-specified interim analysis was conducted on the results available from 116 enrolled subjects. At the interim analysis timepoint, the Active Mode 2 arm was dropped, and the study continued with only 2 arms: the most promising effectiveness treatment arm (Active Mode 1) and the control arm. The effectiveness results presented below are for the Active Mode 1 treatment arm and the control arm.
Table 4 presents the CSBM1 and CSBM2 success rates by study arm and analysis set. In the primary analysis population, the mITT set, the CSBM1 success rate was 40.51% in the treatment arm, compared to 22.92% in the control arm, a difference of 17.6% (chi-square p-value = 0.0011); and the CSBM2 success rate was 23.42% in the treatment arm, compared to 11.81% in the control arm, a difference of 10.0% (chi-square p-value = 0.0085). The primary effectiveness results met the study success criteria with both the CSBM1 and the CSBM2 success rates
{7}------------------------------------------------
statistically significantly higher in the active arm compared to the control. Similar results were also found in the PP and ITT analysis sets.
| | | Treatment | | Control | | |
|------|------------------------------------------|--------------------|----------------------------|--------------------|----------------------------|-------------------------|
| | | % (n/N) | 97.5% CI<br>(Wilson Score) | % (n/N) | 97.5% CI<br>(Wilson Score) | p-value<br>(chi-square) |
| mITT | Improve by ≥ 1 CSBM,<br>6 out of 8 weeks | 40.51%<br>(64/158) | [32.18%;49.42%] | 22.92%<br>(33/144) | [16.06%;31.60%] | 0.0011 |
| | Improve by ≥ 2 CSBM,<br>6 out of 8 weeks | 23.42%<br>(37/158) | [16.76%;31.72%] | 11.81%<br>(17/144) | [7.03%;19.16%] | 0.0085 |
| PP | Improve by ≥ 1 CSBM,<br>6 out of 8 weeks | 40.65%<br>(63/155) | [32.23%;49.65%] | 22.92%<br>(33/144) | [16.06%;31.60%] | 0.0010 |
| | Improve by ≥ 2 CSBM,<br>6 out of 8 weeks | 23.23%<br>(36/155) | [16.54%;31.60%] | 11.81%<br>(17/144) | [7.03%;19.16%] | 0.0098 |
| ITT | Improve by ≥ 1 CSBM,<br>6 out of 8 weeks | 39.26%<br>(64/163) | [31.13%;48.04%] | 22.15%<br>(33/149) | [15.50%;30.61%] | 0.0011 |
| | Improve by ≥ 2 CSBM,<br>6 out of 8 weeks | 22.70%<br>(37/163) | [16.23%;30.80%] | 11.41%<br>(17/149) | [6.79%;18.55%] | 0.0085 |
Table 4: CSBM1 and CSBM2 Success Rates in mITT, ITT, and PP Analysis Sets
### Secondary endpoints
There was an observed reduction in straining in the active arm compared to the control arm (Table 5). Staining was measured using a subject reported scale from 0 (no straining) to 10 (unbearable straining).
| Table 5: Change from Baseline in Straining (Adjusted Baseline) | |
|----------------------------------------------------------------|--|
|----------------------------------------------------------------|--|
| | | Adj. Mean | SE | Two-sided 97.5% CI |
|------|---------------------------|-----------|------|--------------------|
| mITT | Treatment | -1.78 | 0.15 | [-2.12;-1.44] |
| | Control | -1.32 | 0.16 | [-1.68;-0.97] |
| | Diff. (Treatment-Control) | -0.46 | 0.22 | [-0.95;0.03] |
| PP | Treatment | -1.80 | 0.15 | [-2.15;-1.46] |
| | Control | -1.33 | 0.16 | [-1.69;-0.98] |
| | Diff. (Treatment-Control) | -0.47 | 0.22 | [-0.96;0.02] |
| ITT | Treatment | -1.71 | 0.15 | [-2.04;-1.37] |
| | Control | -1.29 | 0.16 | [-1.65;-0.94] |
| | Diff. (Treatment-Control) | -0.41 | 0.22 | [-0.90;0.08] |
There was an observed improvement in stool consistency (rated using the Bristol Stool scale) in the active arm compared to the control arm (Table 6).
| Table 6: Change from Baseline in Stool Consistency (Adjusted for Baseline) | |
|----------------------------------------------------------------------------|--|
|----------------------------------------------------------------------------|--|
| | | Adj. Mean | SE | Two-sided 97.5% CI |
|------|---------------------------|-----------|------|--------------------|
| mITT | Treatment | 1.03 | 0.08 | [0.85;1.21] |
| mITT | Control | 0.58 | 0.08 | [0.39;0.76] |
| mITT | Diff. (Treatment-Control) | 0.45 | 0.11 | [0.19;0.71] |
| PP | Treatment | 1.02 | 0.08 | [0.84;1.20] |
| PP | Control | 0.58 | 0.08 | [0.40;0.77] |
| PP | Diff. (Treatment-Control) | 0.44 | 0.12 | [0.18;0.70] |
| ITT | Treatment | 0.99 | 0.08 | [0.81;1.16] |
| ITT | Control | 0.57 | 0.08 | [0.38;0.75] |
| ITT | Diff. (Treatment-Control) | 0.42 | 0.11 | [0.16;0.68] |
{8}------------------------------------------------
No difference was observed in the bloating sensation endpoint (Table 7). Bloating was measured using a subject reported scale from 0 (no bloating) to 10 (unbearable bloating).
| | | Adj. Mean | SE | Two-sided 97.5% CI |
|------|---------------------------|-----------|------|--------------------|
| mITT | Treatment | -0.42 | 0.11 | [-0.67;-0.17] |
| | Control | -0.37 | 0.12 | [-0.63;-0.10] |
| | Diff. (Treatment-Control) | -0.06 | 0.16 | [-0.42;0.30] |
| PP | Treatment | -0.44 | 0.11 | [-0.70;-0.19] |
| | Control | -0.37 | 0.12 | [-0.63;-0.11] |
| | Diff. (Treatment-Control) | -0.08 | 0.16 | [-0.44;0.29] |
| ITT | Treatment | -0.40 | 0.11 | [-0.64;-0.15] |
| | Control | -0.36 | 0.11 | [-0.62;-0.10] |
| | Diff. (Treatment-Control) | -0.04 | 0.16 | [-0.39;0.32] |
Table 7: Change from Baseline in Bloating Sensation (Adjusted for Baseline)
# Safety results
The ITT data analysis set served as the principal set for the safety analysis: 200 treatment subjects (37 from Active Mode 2 arm and 163 from Active Mode 1 arm) and 149 control subjects. A total of 114 AEs were reported during the study: 74 AEs experienced by 62 treatment subjects (31.0%), and 40 AEs experienced by 26 subjects (17.45%) in the control arm. Two AEs, both in the control arm, were classified as serious adverse events (SAEs), but were not considered related to the study treatment. There were no SAEs reported in the treatment arms.
49 AEs were considered possibly, probably, or definitely related to the study treatment or procedure (41 in the treatment arm, and 8 in the control arm) by study investigators (Table 8), Device and procedure related AEs reported in the study included abdominal pain, abdominal distension, abdominal discomfort, vomiting, nausea, diarrhea, flatulence, and proctalgia (rectal spasm or cramp). Medical device discomfort events characterized as sensation of vibration were the most commonly reported device related AE, with 19 subjects (9.5%) in the Active Mode 1 and 2 treatment arms reporting 21 events (none in the control).
Table 8: Related adverse events that are possibly, or definitely related to the study treatment or procedure, by system organ class, preferred term and relationship (ITT)
| System Organ<br>Class | Preferred Term | Relation to Device | Treatment<br>(Active Modes 1 and 2) | | | Control | | |
|------------------------------------------------------|------------------------------|-------------------------|-------------------------------------|------------------|-----------|-----------------|------------------|-----------|
| | | | # of<br>reports | # of<br>subjects | Incidence | # of<br>reports | # of<br>subjects | Incidence |
| Total | Total | Total | 41 | 29 | 14.50% | 8 | 5 | 3.36% |
| | | Possibly related | 12 | 7 | 3.50% | 7 | 4 | 2.68% |
| | | Probably related | 8 | 6 | 3.00% | 1 | 1 | 0.67% |
| | | Related | 18 | 15 | 7.50% | - | - | - |
| | | Related to<br>procedure | 3 | 2 | 1.00% | - | - | - |
| | | | | | | | | |
| Gastrointestinal | Total | Possibly related | 9 | 5 | 2.50% | 7 | 4 | 2.68% |
| disorders | | Probably related | 5 | 5 | 2.50% | 1 | 1 | 0.67% |
| | Abdominal pain | Possibly related | 1 | 1 | 0.50% | 2 | 2 | 1.34% |
| | Vomiting | Possibly related | 3 | 3 | 1.50% | | | |
| | Nausea | Possibly related | 1 | 1 | 0.50% | | | |
| | | Probably related | 1 | 1 | 0.50% | | | |
| | Diarrhoea | Possibly related | 1 | 1 | 0.50% | | | |
| | | Probably related | 1 | 1 | 0.50% | | | |
| | Abdominal<br>distension | Possibly related | 1 | 1 | 0.50% | 1 | 1 | 0.67% |
| | Flatulence | Possibly related | 1 | 1 | 0.50% | | | |
| | Constipation | Possibly related | 1 | 1 | 0.50% | | | |
| | Proctalgia | Possibly related | | | | 1 | 1 | 0.67% |
| | | Probably related | 1 | 1 | 0.50% | | | |
| | Abdominal<br>discomfort | Probably related | 2 | 2 | 1.00% | | | |
| | Anorectal<br>discomfort | Possibly related | | | | 2 | 1 | 0.67% |
| | Abdominal<br>lower | pain Possibly related | | | | 1 | 1 | 0.67% |
| | Abdominal<br>upper | pain Probably related | | | | 1 | 1 | 0.67% |
| General disorders | Total | Possibly related | 1 | 1 | 0.50% | | | |
| and administration | | Probably related | 3 | 3 | 1.50% | | | |
| site conditions | | Related | 16 | 14 | 7.00% | | | |
| | | Related to<br>procedure | 2 | 1 | 0.50% | | | |
| | Medical | device Possibly related | | 1 | 0.50% | | | |
| | discomfort | Probably related | 3 | 3 | 1.50% | | | |
| | | Related | 16 | 14 | 7.00% | | | |
| | | Related to<br>procedure | 1 | 1 | 0.50% | | | |
| | Discomfort | Related to<br>procedure | 1 | 1 | 0.50% | | | |
| Reproductive | Total | Possibly related | 1 | 1 | 0.50% | | | |
| system and breast<br>disorders | Amenorrhoea | Possibly related | 1 | 1 | 0.50% | | | |
| Nervous system | Total | Possibly related | 1 | 1 | 0.50% | | | |
| disorders | Headache | Possibly related | 1 | 1 | 0.50% | | | |
| Psychiatric | Total | Related | 1 | 1 | 0.50% | | | |
| disorders | Sleep disorder | Related | 1 | 1 | 0.50% | | | |
| | Total | Related | 1 | 1 | 0.50% | | | |
| Injury, poisoning<br>and procedural<br>complications | Intentional device<br>misuse | Related | 1 | 1 | 0.50% | . | . | . |
| Investigations | Total | Related to<br>procedure | 1 | 1 | 0.50% | . | . | . |
| | Weight decreased | Related to<br>procedure | 1 | 1 | 0.50% | . | . | . |
{9}------------------------------------------------
De Novo Summary (DEN210052)
Page 10 of 15
{10}------------------------------------------------
### ADDITIONAL CLINICAL INFORMATION
Five smaller clinical trials were conducted prior to the V270 pivotal study. The studies included a total of 499 subjects and varied from the V270 study treatment protocol in the dosing regimen, number of capsule vibration sets while passing through the large intestine, and/or the number of capsule administered per week. The prior studies do not support effectiveness of the Vibrant System. However, FDA considered the cumulative safety data in these studies to support safety of the Vibrant System. Safety data from these prior studies are consistent in the type, severity and frequency of adverse events reported in the V270 pivotal study.
# Pediatric Extrapolation
In this De Novo request, existing clinical data were not leveraged to support the use of the device in a pediatric patient population.
# LABELING
Physician labeling includes the device indications for use, a description of the device, warnings and precautions, clinical data on the device, and instructions for the safe and effective use of the device. The labeling satisfies the requirements of 21 CFR 801.109 Prescription devices. Per the Special Controls for this generic type of device, labeling includes a summary of device effectiveness and device related adverse events.
#### RISKS TO HEALTH
Table 9 identifies the risks to health that may be associated with use of an orally ingested transient device for constipation, and the measures necessary to mitigate these risks.
| Identified Risks to Health | Mitigation Measures |
|--------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------|
| Infection | Bioburden testing<br>Labeling<br>Shelf life testing |
| Adverse tissue reaction | Biocompatibility evaluation |
| Device malfunction leading to injury | Electrical safety testing<br>Software validation, verification, and hazard analysis<br>Non-clinical performance testing<br>Labeling<br>Shelf life testing |
| Interference with other devices | Electromagnetic compatibility testing |
| Failure to excrete capsule | Clinical data |
# Table 9 Identified Risks to Health and Mitigation Measures
{11}------------------------------------------------
| | Labeling |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------|
| Device related adverse events including:<br>• Choking<br>• Abdominal pain<br>• Abdominal distension<br>• Abdominal discomfort<br>• Vomiting<br>• Nausea<br>• Proctalgia<br>• Diarrhea | Clinical data<br>Non-clinical performance testing<br>Labeling |
| Device ineffective leading to constipation and effects of delayed treatment | Clinical data<br>Labeling |
#### SPECIAL CONTROLS
In combination with the general controls of the FD&C Act, an orally ingested transient device for constipation is subject to the following special controls:
- Clinical data must demonstrate the device performs as intended and evaluate the following: (1)
- Treatment of constipation; and (i)
- (ii) All adverse events.
- (2) Non-clinical performance data must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested:
- Dimensional testing must verify device dimensions; (i)
- (ii) Performance bench testing must verify functional aspects of the device design;
- (iii) Leak testing must verify device integrity under worst case clinical conditions;
- Bite testing must demonstrate that the device can withstand bite forces; (iv)
- (v) pH resistance testing must evaluate integrity of the capsule when exposed to a physiological relevant range of pH values; and
- Bioburden testing must demonstrate the device does not pose an infection risk (vi) throughout the labeled shelf life.
- The patient-contacting components of the device must be demonstrated to be (3) biocompatible.
- Performance data must support the shelf life of the device by demonstrating continued (4) package integrity and device functionality over the labeled shelf life.
- (5) Software validation, verification, and hazard analysis must be performed.
- Electrical safety and electromagnetic compatibility (EMC) testing must be performed for (6) any electrical components of the device.
{12}------------------------------------------------
- (7) Labeling for the device must include:
- (i) A summary of clinical data for the device, including a discussion of adverse events and clinical benefit: and
- (ii) A shelf life.
### BENEFIT-RISK DETERMINATION
Nonclinical laboratory studies as well as clinical data from a pivotal study (V270) and prior studies were used to evaluate the safety and effectiveness of the Vibrant System.
### Summary of Benefits
The V270 pivotal study demonstrated a statistically significant and clinically meaningful benefit over the control. The study met both components of the primary endpoint with higher success rates in the treatment arm compared to the control arm. In 6 out of the 8 weeks of the study, 40.51% in the treatment arm had at least one more weekly CSBM, compared to 22.92% in the control arm, a difference of 17.6% (chi-square p-value = 0.0011, mITT, CSBM1). In 6 out of the 8 weeks of the study, 23.42% in the treatment arm had at least two more weekly CSBM, compared to 11.81% in the control arm, a difference of 10% (chi-square p-value = 0.0085, mITT, CSBM2). The improvements in the number of complete spontaneous bowel movements was statistically significant and clinically meaningful when considering an improvement of 1 and 2 bowel movements during at least 6 of the 8 weeks of treatment in a patient population with chronic idiopathic constipation who have not experienced relief of bowel symptoms by using laxative therapies at the recommended dosage for at least one month.
Factors that increase uncertainty in determining probable benefits for the Vibrant Systems include:
- Inconsistent or conflicting results between studies. Five clinical trials were conducted prior . to the V270 pivotal studies included a total of 499 subjects and varied from the V270 study treatment protocol in the dosing regimen, number of capsule vibration sets while passing through the large intestine, and/or the number of capsule administered per week. The prior studies did not support effectiveness of the Vibrant System.
- . The effectiveness endpoint required an increase of at least one weekly complete spontaneous bowel movement (CSBM) during at least 6 of the 8 weeks of treatment. The study met the primary effectiveness endpoint; however, effectiveness in the CSBM1 treatment arm was 40.5% indicating the response to the Vibrant capsule treatment was not experienced by 59.5% of the treated subjects.
- · The long-term effectiveness of the device has not been studied beyond 8 weeks.
#### Summary of Risks
A total of 114 AEs were reported during the study. Two AEs were classified as serious adverse events (SAEs) in the control arm that were not considered to the study treatment. There were no SAEs reported in the treatment arms. 49 AEs were considered possibly, or definitely related to the study treatment or procedure (41 in the treatment arm, and 8 in the control arm). Device and procedure related adverse events reported in the study included abdominal pain, abdominal distension, abdominal discomfort, vomiting, nausea, diarrhea,
De Novo Summary (DEN210052)
{13}------------------------------------------------
flatulence, and proctalgia. Medical device discomfort events characterized as sensation of vibration were the most commonly reported device related AE, with 19 subjects (9.5%) in the Active Mode 1 and 2 treatment arms reporting 21 events (none in the control).
Factors that increase uncertainty in determining probable risks for the Vibrant System include:
- . Insufficient patient numbers to detect less common serious events.
- . Insufficient duration of follow-up to detect delayed/late events. The long-term risks associated with continued repeat use of the device beyond 8 weeks are unknown. The pivotal study collected safety data for 8 weeks.
- . The V270 pivotal study did not include evaluation for capsule retrievals once they passed through the GI tract. An earlier safety study (V11) evaluated capsule retrieval directly in a 28 subject safety study and reported a capsule retrieval rate of 85%; however, subjects in the V111 study had difficulty identifying the capsules in bowel movements.
# Patient Perspectives
Patient perspectives were measured in the V270 trial as an additional endpoint assessment. Treatment Satisfaction Questionnaire for Medication (TSQM) scores were evaluated in the V270 study. The TSOM score for effectiveness was reported to be higher in the treatment arm relative to the control arm. The mean scores for side-effects, convenience, and global satisfaction were similar in both groups.
Change from baseline in quality of life using the Patient Assessment of Constipation Quality of Life Questionnaire (PAC-QOL) questionnaire2 was evaluated in the V270 study. More subjects reported an improvement in quality of life from baseline on the PAC-OOL in the active arm 1 than in the control arm (77.9% of active arm 1 subjects versus 66.2% of control arm subjects).
# Benefit/Risk Conclusion
The V270 pivotal study reports a statistically significant and clinically meaningful clinical benefit for the Vibrant System treatment of chronic idiopathic constipation with an improvement in the number of complete spontaneous bowel movements per week over an 8-week treatment course. Considering the types, numbers, and mild-to-moderate severity of adverse events in the treatment arm compared to the control arm, there is an overall favorable benefit-risk profile that supports granting this De Novo request.
<sup>1</sup> Atkinson MJ, Sinha A, Hass SL, Colman SS, Kumar RN, Brod M, Rowland CR. Validation of a general measure of treatment satisfaction, the Treatment Satisfaction (TSOM), using a national panel study of chronic disease. Health Qual Life Outcomes. 2004 Feb 26;2:12. doi: 10.1186/1477-7525-2-12. PMID: 14987333; PMCID: PMC398419.
<sup>2</sup> Marquis P. De La Loge C. Dubois D. McDermott A. Chassany O. Development and validation of the Patient Assessment of Constipation Quality of Life questionnaire. Scand J Gastroenterol. 2005 May:40(5):540-51. doi: 10.1080/00365520510012208. PMID: 16036506.
{14}------------------------------------------------
In conclusion, given the available information above, the probable benefits outweigh the probable risks for the Vibrant System when used as indicated. The device provides benefits, and the risks can be mitigated using general controls and the identified special controls.
#### CONCLUSION
The De Novo request for the Vibrant System is granted and the device is classified as follows:
Product Code: QTN Device Type: Orally ingested transient device for constipation Regulation Number: 21 CFR 876.5940 Class: II
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.