Browse hierarchy: [Gastroenterology, Urology (GU)](/submissions/GU) → [Subpart F — Therapeutic Devices](/submissions/GU/subpart-f%E2%80%94therapeutic-devices) → [21 CFR 876.5026](/submissions/GU/subpart-f%E2%80%94therapeutic-devices/876.5026) → QRC — Non-Implanted Electrical Stimulation Device For Management Of Premature Ejaculation

# QRC · Non-Implanted Electrical Stimulation Device For Management Of Premature Ejaculation

_Gastroenterology, Urology · 21 CFR 876.5026 · Class 2_

**Canonical URL:** https://fda.innolitics.com/productcode/QRC

## Overview

- **Product Code:** QRC
- **Device Name:** Non-Implanted Electrical Stimulation Device For Management Of Premature Ejaculation
- **Regulation:** [21 CFR 876.5026](/submissions/GU/subpart-f%E2%80%94therapeutic-devices/876.5026)
- **Device Class:** 2
- **Review Panel:** [Gastroenterology, Urology](/submissions/GU)

## Identification

A non-implanted electrical stimulation device for management of premature ejaculation is intended to be used in patients with premature ejaculation by delivery of electrical stimulation to the perineal muscles and nerves.

## Classification Rationale

Class II (special controls). The special controls for this device are:

## Special Controls

In combination with general controls of the Food Drug & Cosmetic Act, the non-implanted electrical stimulation device for management of premature ejaculation is subject to the following special controls:

*Classification.* Class II (special controls). The special controls for this device are:(1) The device must be demonstrated to be biocompatible.
(2) Performance testing must demonstrate the electromagnetic compatibility, electrical safety, and thermal safety of the device.
(3) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested:
(i) Mechanical performance;
(ii) Electrical stimulation parameters; and
(iii) Battery performance.
(4) Performance testing must support shelf life by demonstrating continued device functionality over the identified shelf life.
(5) Software verification, validation, and hazard analysis must be performed.
(6) Labeling must include:
(i) Specific instructions regarding safe placement and correct use of the device;
(ii) Warning(s) against use by patients with active implanted medical devices; and
(iii) A shelf life.

## Recent Cleared Devices (4 of 4)

| Record | Device Name | Applicant | Decision Date | Decision |
| --- | --- | --- | --- | --- |
| [K252809](/device/K252809.md) | in2 Smart (in2S) | Virility Medical , Ltd. | Apr 6, 2026 | SESE |
| [K241897](/device/K241897.md) | MOR | Morari, Inc. | Feb 11, 2025 | SESE |
| [K223595](/device/K223595.md) | vPATCH | Virility Medical , Ltd. | May 1, 2023 | SESE |
| [DEN210012](/device/DEN210012.md) | vPatch | Virility Medical | Nov 23, 2021 | DENG |

## Top Applicants

- Virility Medical , Ltd. — 2 clearances
- Morari, Inc. — 1 clearance
- Virility Medical — 1 clearance

---

**Source:** [https://fda.innolitics.com/productcode/QRC](https://fda.innolitics.com/productcode/QRC)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
