Browse hierarchy: [Neurology (NE)](/submissions/NE) → [Subpart F — Physical Medicine Therapeutic Devices](/submissions/NE/subpart-f%E2%80%94physical-medicine-therapeutic-devices) → [21 CFR 890.5420](/submissions/NE/subpart-f%E2%80%94physical-medicine-therapeutic-devices/890.5420) → QOL — Electroencephalography (Eeg)-Driven Upper Extremity Powered Exerciser

# QOL · Electroencephalography (Eeg)-Driven Upper Extremity Powered Exerciser

_Neurology · 21 CFR 890.5420 · Class 2_

**Canonical URL:** https://fda.innolitics.com/productcode/QOL

## Overview

- **Product Code:** QOL
- **Device Name:** Electroencephalography (Eeg)-Driven Upper Extremity Powered Exerciser
- **Regulation:** [21 CFR 890.5420](/submissions/NE/subpart-f%E2%80%94physical-medicine-therapeutic-devices/890.5420)
- **Device Class:** 2
- **Review Panel:** [Neurology](/submissions/NE)

## Identification

An EEG-driven upper extremity powered exerciser is a non-invasive prescription device intended for rehabilitation by driving movement or exercise of an impaired upper extremity in response to the detection of purpose oriented electrical activity produced by the patient's brain.

## Classification Rationale

Class II (special controls). The special controls for this device are:

## Special Controls

In combination with the general controls of the FD&C Act, the electroencephalography (EEG)driven powered exerciser is subject to the following special controls:

*Classification.* Class II (special controls). The special controls for this device are:(1) Clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. Testing must capture any adverse events observed during clinical use and must demonstrate that the EEG signal can be translated into intended motion.
(2) Software verification, validation, and hazard analysis must be performed.
(3) Performance data must demonstrate the electromagnetic compatibility, electrical safety, battery safety, and wireless compatibility of the device.
(4) The device components that contact the patient must be demonstrated to be biocompatible.
(5) Performance data must validate the reprocessing instructions for the reusable components of the device.
(6) Labeling must include:
(i) Instructions on fitting the device to the patient;
(ii) Information on how the device operates and the typical sensations experienced during treatment; and
(iii) Reprocessing instructions.

## Recent Cleared Devices (1 of 1)

| Record | Device Name | Applicant | Decision Date | Decision |
| --- | --- | --- | --- | --- |
| [DEN200046](/device/DEN200046.md) | Neurolutions Upper Extremity Rehabilitation System | Neurolutions, Inc. | Apr 23, 2021 | DENG |

## Top Applicants

- Neurolutions, Inc. — 1 clearance

---

**Source:** [https://fda.innolitics.com/productcode/QOL](https://fda.innolitics.com/productcode/QOL)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com