QNM · Mountable Electromechanical Surgical System For Transluminal Approaches

General, Plastic Surgery · 21 CFR 878.4961 · Class 2

Overview

Product CodeQNM
Device NameMountable Electromechanical Surgical System For Transluminal Approaches
Regulation21 CFR 878.4961
Device ClassClass 2
Review PanelGeneral, Plastic Surgery

Identification

A mountable electromechanical surgical system for transluminal approaches is a software-controlled, patient bed- and/or operating table-mounted electromechanical surgical system with human/device interfaces that allows a qualified user to perform transluminal endoscopic or laparoscopic surgical procedures using surgical instruments attached to an electromechanical arm.

Classification Rationale

Class II (special controls). The special controls for this device are:

Special Controls

*Classification.* Class II (special controls). The special controls for this device are:(1) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error. (2) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program. (3) The device manufacturer must conduct and complete post-market surveillance, including an impact of the training program on user learning, behavior, and performance, in accordance with an FDA-agreed-upon protocol. The device manufacturer must submit post-market surveillance reports that contain current data and findings in accordance with the FDA-agreed-upon protocol. (4) The device manufacturer must submit a report to FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information: (i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and (ii) Identification and rationale for changes made to the device, labeling or device-specific use training program, which did not require submission of a premarket notification during the reporting period. (5) Labeling must include: (i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified; (ii) A statement in the labeling that the safety and effectiveness of the device has not been evaluated for outcomes related to the treatment or prevention of cancer, including but not limited to risk reduction, overall survival, disease-free survival and local recurrence, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA; (iii) Identification of compatible devices; (iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification; (v) Reprocessing instructions for reusable components; (vi) A shelf life for any sterile components; (vii) A description of the device-specific use training program; (viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and (ix) A detailed summary of the post-market surveillance data collected under paragraph (b)(3) of this section and any necessary modifications to the labeling to accurately reflect outcomes based upon the post-market surveillance data collected under paragraph (b)(3) of this section. (6) Clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. (7) Human factors validation testing must be performed and must demonstrate that the user interfaces of the system support safe use in an operating room environment. (8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include: (i) Device motion accuracy and precision; (ii) System testing; (iii) Instrument reliability; (iv) Thermal effects on tissue; (v) Human-device interface; (vi) Mounting hardware testing; (vii) Workspace access testing; and (viii) Performance testing with compatible devices. (9) Software verification, validation, and hazard analysis must be performed. Software documentation must include an assessment of the impact of threats and vulnerabilities on device functionality and end users/patients as part of cybersecurity review. (10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed. (11) Performance data must demonstrate the sterility of all patient-contacting device components. (12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life. (13) Performance data must validate the reprocessing instructions for the reusable components of the device. (14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible. (15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.

Recent Cleared Devices (8 of 8)

RecordDevice NameApplicantDecision DateDecision
K251761Anovo Surgical System (Model 6Ne); Anovo Instrument ARM Curved Scissors ; Anovo Instrument ARM Hook ElectrodeMomentis Surgical , Ltd.Jul 8, 2025SESE
K250591Anovo Surgical System (6Ne)Momentis Surgical , Ltd.Jun 6, 2025SESE
K251056Anovo Instrument ARM Curved Scissors ; Anovo Instrument ARM Hook ElectrodeMomentis Surgical , Ltd.May 2, 2025SESE
K243182Anovo Instrument ARM Curved Scissors; Anovo Instrument ARM Hook ElectrodeMomentis Surgical , Ltd.Feb 3, 2025SESE
K242157Anovo Surgical System (Model 6Ne)Momentis Surgical , Ltd.Oct 21, 2024SESE
K241907Anovo Surgical System (model 6N)Momentis Surgical , Ltd.Oct 2, 2024SESE
K232146Anovo PedestalMomentis Surgical , Ltd.Sep 14, 2023SESE
DEN190022Hominis Surgical SystemMomentis Surgical , Ltd.Feb 26, 2021DENG

Top Applicants

Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...