Browse hierarchy: [Pathology (PA)](/submissions/PA) → [Subpart G — Tumor Associated Antigen Immunological Test Systems](/submissions/PA/subpart-g%E2%80%94tumor-associated-antigen-immunological-test-systems) → [21 CFR 866.6100](/submissions/PA/subpart-g%E2%80%94tumor-associated-antigen-immunological-test-systems/866.6100) → QDC — Dna-Based Test For Minimal Residual Disease For Hematologic Malignancies

# QDC · Dna-Based Test For Minimal Residual Disease For Hematologic Malignancies

_Pathology · 21 CFR 866.6100 · Class 2_

**Canonical URL:** https://fda.innolitics.com/productcode/QDC

## Overview

- **Product Code:** QDC
- **Device Name:** Dna-Based Test For Minimal Residual Disease For Hematologic Malignancies
- **Regulation:** [21 CFR 866.6100](/submissions/PA/subpart-g%E2%80%94tumor-associated-antigen-immunological-test-systems/866.6100)
- **Device Class:** 2
- **Review Panel:** [Pathology](/submissions/PA)

## Identification

A DNA-based test to measure minimal residual disease in hematological malignancies is a prescription in vitro diagnostic device that identifies and quantifies specific nucleic acid sequences within human tissues to estimate the percentage of cells that harbor the specific sequence(s). The test is intended to be used as an aid to measure minimal residual disease to assess the change in burden of disease during monitoring of treatment. The test is indicated for use by qualified healthcare professionals in accordance with professional guidelines for clinical decision-making, in conjunction with other clinicopathological features.

## Classification Rationale

Class II (special controls). The special controls for this device are:

## Special Controls

A DNA-based test to measure minimal residual disease in hematological malignancies must comply with the following special controls:

*Classification.* Class II (special controls). The special controls for this device are:(1) Design verification and validation must include:
(i) A detailed description of the device, including:
(A) A detailed description of all test components, reagents, instrumentation, and software, including software applications and any hardware-based devices that incorporate software.
(B) A detailed description of all genomic regions that are detected and quantified by the assay.
(C) A detailed description of the methodology and protocols for each step of the test, including description of the quality metrics, thresholds, and filters at each step of the test that are implemented for final result reporting and a description of the metrics for run-failures, specimen-failures, and invalids, as appropriate.
(D) Detailed specifications and procedures for sample collection, processing, and storage.
(E) A description of the internal and external controls that are recommended or provided. The description must identify those control elements that are incorporated into the testing procedure. If appropriate, this description must include a description of the controls and control procedures used during the sequencing and data analysis.
(ii) Identification of risk mitigation elements used by the device, including a detailed description of all additional procedures, methods, and practices incorporated into the instructions for use that mitigate risks associated with use of the device.
(iii) As part of the risk management activities, an appropriate end user device training program must be offered as an effort to mitigate the risk of failure from user error, as appropriate.
(iv) Description of analytical and clinical studies, including:
(A) Device performance data that demonstrates the ability to measure minimal residual disease in the claimed specimen type(s) from patients that are representative of the intended use population. Data can be obtained via:
(
*1* ) A method comparison study comparing the device to a predicate device with clinical data for the specified hematological neoplastic indication using the specified specimen type(s); or(
*2* ) A clinical study demonstrating clinical validity using well characterized clinical specimens from patients with known clinical outcomes using a study design deemed acceptable by FDA.(B) Device precision (repeatability and reproducibility) data using clinical samples covering the range of minimal residual disease frequencies reported by the test and covering the stated range of DNA inputs that are indicated as allowable for use with the test. Results shall be reported as the standard deviation and/or percentage coefficient of variation with the 95 percent confidence interval for each level tested. The study must evaluate all sources of variability, including, as appropriate, between-site and between operator (minimum of three sites of which two must be external with a minimum of two operators per site), between-day (minimum of 3 days), between-run, within-run, between-lot (minimum of three lots), between instrument (minimum of three instruments), and total variation.
(C) Device linearity data generated from samples covering the device measuring range using a dilution panel created from clinical samples.
(D) Device accuracy by comparison to flow cytometry across the measuring interval or to the predicate method across the measuring interval.
(E) Device analytic sensitivity data, including limit of blank, limit of detection, and limit of quantitation, using a dilution panel created from clinical samples.
(F) Analytical specificity data, including interference and cross-contamination, and index cross-contamination, as appropriate.
(G) Validation of pre-analytical methods, including DNA extraction methods and cell enrichment methods, as appropriate.
(H) Device stability data, including real-time stability of reagents under various storage times and temperatures.
(I) Specimen and prepared sample stability data established for each specimen matrix in the anticoagulant combinations and storage/use conditions that will be indicated, including specimen transport, as appropriate.
(2) The intended use for the labeling required under § 809.10(a)(4) of this chapter and for the labeling required under § 809.10(b)(5)(ii) of this chapter, as applicable, must include:
(i) The clinical hematopoietic malignancy for which the assay was designed and validated (
*e.g.,* multiple myeloma or B-cell acute lymphoblastic leukemia);(ii) Specimen type (
*e.g.,* bone marrow);(iii) The specific DNA regions that are being identified and quantified (
*e.g.,* rearranged IgH (VDJ), IgH (DJ), IgK, and IgL receptor gene sequences); and(iv) A statement that the results are indicated to be interpreted by qualified healthcare professionals in accordance with professional guidelines for clinical decision-making in conjunction with other clinicopathological features.
(3) The labeling required under § 809.10(b) of this chapter must include information that demonstrates the performance characteristics of the test, including a detailed summary of the performance studies conducted and their results, as described in paragraphs (b)(1)(iv)(A) through (I) of this section.
(4) The device output, including any test report, must include the estimated minimal residual disease (MRD) frequency and an appropriate range of the uncertainty of that frequency based on the amount of DNA that was evaluated by the test and the number of specific nucleic acid sequences that were detected (
*e.g.,* “MRD = 1.2 × 10−5 [Range = 0.8 × 10−6 to 2.0 × 10−5 ]”).

## Recent Cleared Devices (2 of 2)

| Record | Device Name | Applicant | Decision Date | Decision |
| --- | --- | --- | --- | --- |
| [K200009](/device/K200009.md) | Adaptive Biotechnologies clonoSEQ Assay | Adaptive Biotechnologies Corporation | Aug 5, 2020 | SESE |
| [DEN170080](/device/DEN170080.md) | Adaptive Biotechnologies clonoSEQ Assay | Adaptive Biotechnologies Corporation | Sep 28, 2018 | DENG |

## Top Applicants

- Adaptive Biotechnologies Corporation — 2 clearances

---

**Source:** [https://fda.innolitics.com/productcode/QDC](https://fda.innolitics.com/productcode/QDC)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
