Browse hierarchy: [Ear, Nose, Throat (EN)](/submissions/EN) → [Subpart F — Therapeutic Devices](/submissions/EN/subpart-f%E2%80%94therapeutic-devices) → [21 CFR 874.5900](/submissions/EN/subpart-f%E2%80%94therapeutic-devices/874.5900) → PKA — External Upper Esophageal Sphincter (Ues) Compression Device

# PKA · External Upper Esophageal Sphincter (Ues) Compression Device

_Ear, Nose, Throat · 21 CFR 874.5900 · Class 2_

**Canonical URL:** https://fda.innolitics.com/productcode/PKA

## Overview

- **Product Code:** PKA
- **Device Name:** External Upper Esophageal Sphincter (Ues) Compression Device
- **Regulation:** [21 CFR 874.5900](/submissions/EN/subpart-f%E2%80%94therapeutic-devices/874.5900)
- **Device Class:** 2
- **Review Panel:** [Ear, Nose, Throat](/submissions/EN)

## Identification

An external upper esophageal sphincter compression device is a prescription device used to apply external pressure on the cricoid cartilage for the purpose of reducing the symptoms of laryngopharyngeal reflux disease.

## Classification Rationale

Class II (special controls). The special controls for this device are:

## Special Controls

In combination with the general controls of the FD&C Act, the External UES Compression Device is subject to the following special controls:

*Classification.* Class II (special controls). The special controls for this device are:(1) The patient contacting components must be demonstrated to be biocompatible.
(2) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be demonstrated:
(i) Mechanical integrity testing (
*e.g.,* tensile strength testing, fatigue testing) and(ii) Shelf life testing.
(3) The technical specifications must include pressure measurement accuracy to characterize device performance.
(4) Clinical performance testing must document any adverse events observed during clinical use, and demonstrate that the device performs as intended under anticipated conditions of use.
(5) Labeling must include the following:
(i) Appropriate warnings and precautions,
(ii) A detailed summary of the clinical testing pertinent to use of the device including a detailed summary of the device-related complications or adverse events,
(iii) Detailed instructions on how to fit the device to the patient, and
(iv) Instructions for reprocessing of any reusable components.
(6) Patient labeling must be provided and must include:
(i) Relevant warnings, precautions, and adverse effects/complications,
(ii) Information on how to correctly wear the device,
(iii) The potential risks and benefits associated with the use of the device,
(iv) Alternative treatments, and
(v) Reprocessing instructions.

## Recent Cleared Devices (2 of 2)

| Record | Device Name | Applicant | Decision Date | Decision |
| --- | --- | --- | --- | --- |
| [K173934](/device/K173934.md) | Reza Band, Reflux Band | Somna Therapeutics, LLC | Apr 13, 2018 | SESE |
| [DEN130046](/device/DEN130046.md) | REZA BAND(TM) UPPER ESOPHAGEAL (UES) SPHINCTER ASSIST DEVICE | Somna Therapeutics, LLC | Mar 6, 2015 | DENG |

## Top Applicants

- Somna Therapeutics, LLC — 2 clearances

---

**Source:** [https://fda.innolitics.com/productcode/PKA](https://fda.innolitics.com/productcode/PKA)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
