OVP · Vibratory Counter-Stimulation

Physical Medicine · 21 CFR 882.5895 · Class 2

Overview

Product CodeOVP
Device NameVibratory Counter-Stimulation
Regulation21 CFR 882.5895
Device ClassClass 2
Review PanelPhysical Medicine

Identification

A vibratory counter-stimulation device is a prescription device that provides electrically powered mechanical vibration to improve the quality of sleep in patients with primary Restless Legs Syndrome.

Classification Rationale

Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 882.9. The special controls for this device are:

Special Controls

*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 882.9. The special controls for this device are:(1) Appropriate analysis/testing must demonstrate electromagnetic compatibility (EMC), electrical safety, and thermal safety. (2) If the device contains software or firmware, appropriate verification, validation, and hazard analysis must be performed. (3) The elements of the device that contact the patient must be assessed to be biocompatible. (4) Non-clinical testing data (including vibration frequency, amplitude, and acceleration) must demonstrate that the device performs as intended under anticipated conditions of use. (5) Labeling must include: (i) Specific information pertinent to use of the device by the intended patient population and the treatment regimen; (ii) Warning to only use the device on normal, intact, clean, healthy skin; (iii) Warning to not use the device if the user has leg skin disorders, such as eczema, psoriasis, cellulitis, non-healing wounds; (iv) Warning to discontinue use if Restless Leg Syndrome symptoms worsen; and (v) Instructions for end users to contact the device manufacturer and MedWatch in case they experience any adverse events when using this device.

In combination with the general controls of the FD&C Act, the Vibratory Counter-Stimulation device is subject to the following special controls: - 1. Appropriate analysis/testing must demonstrate electromagnetic compatibility (EMC), electrical safety, and thermal safety. - 2. If the device contains software or firmware, appropriate verification, validation, and hazard analysis must be performed. - 3. The elements of the device that contact the patient must be assessed to be biocompatible. - 4. Non-clinical testing data (including vibration frequency, amplitude and acceleration) must demonstrate that the device performs as intended under anticipated conditions of use. - 5. Labeling must include: - a. Specific information pertinent to use of the device by the intended patient population and the treatment regimen. - Warning to only use the device on normal, intact, clean, healthy skin. b. - Warning to not use the device if the user has leg skin disorders, such as eczema, C. psoriasis, cellulitis, non-healing wounds. - d. Warning to discontinue use if restless leg syndrome symptoms worsen. - Instructions for end users to contact the device manufacturer and MedWatch in e. case they experience any adverse events when using this device.

Recent Cleared Devices (1 of 1)

RecordDevice NameApplicantDecision DateDecision
DEN110011SYMPHONY DEVICESensory Medical, Inc.Dec 18, 2013DENG

Top Applicants

Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...