Browse hierarchy: [General Hospital (HO)](/submissions/HO) → [Subpart F — General Hospital and Personal Use Therapeutic Devices](/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices) → [21 CFR 880.5130](/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/880.5130) → ONM — Radiant Infant Warmer Cabinet

# ONM · Radiant Infant Warmer Cabinet

_General Hospital · 21 CFR 880.5130 · Class 2_

**Canonical URL:** https://fda.innolitics.com/productcode/ONM

## Overview

- **Product Code:** ONM
- **Device Name:** Radiant Infant Warmer Cabinet
- **Regulation:** [21 CFR 880.5130](/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/880.5130)
- **Device Class:** 2
- **Review Panel:** [General Hospital](/submissions/HO)
- **3rd-party reviewable:** yes

## Identification

The infant radiant warmer is a device consisting of an infrared heating element intended to be placed over an infant to maintain the infant's body temperature by means of radiant heat. The device may also contain a temperature monitoring sensor, a heat output control mechanism, and an alarm system (infant temperature, manual mode if present, and failure alarms) to alert operators of a temperature condition over or under the set temperature, manual mode time limits, and device component failure, respectively. The device may be placed over a pediatric hospital bed or it may be built into the bed as a complete unit.

## Classification Rationale

Class II (Special Controls)

## Special Controls

*Classification.* Class II (Special Controls):(1) The Association for the Advancement of Medical Instrumentation (AAMI) Voluntary Standard for the Infant Radiant Warmer;
(2) A prescription statement in accordance with § 801.109 of this chapter (restricted to use by or upon the order of qualified practitioners as determined by the States); and
(3) Labeling for use only in health care facilities and only by persons with specific training and experience in the use of the device.

## Recent Cleared Devices (0 of 0)

_No cleared devices on record._

---

**Source:** [https://fda.innolitics.com/productcode/ONM](https://fda.innolitics.com/productcode/ONM)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
