Browse hierarchy: [Cardiovascular (CV)](/submissions/CV) → [Subpart D — Cardiovascular Prosthetic Devices](/submissions/CV/subpart-d%E2%80%94cardiovascular-prosthetic-devices) → [21 CFR 870.3535](/submissions/CV/subpart-d%E2%80%94cardiovascular-prosthetic-devices/870.3535) → NKO — System, Balloon, Intra-Aortic And Control, Reprocessed

# NKO · System, Balloon, Intra-Aortic And Control, Reprocessed

_Cardiovascular · 21 CFR 870.3535 · Class 3_

**Canonical URL:** https://fda.innolitics.com/productcode/NKO

## Overview

- **Product Code:** NKO
- **Device Name:** System, Balloon, Intra-Aortic And Control, Reprocessed
- **Regulation:** [21 CFR 870.3535](/submissions/CV/subpart-d%E2%80%94cardiovascular-prosthetic-devices/870.3535)
- **Device Class:** 3
- **Review Panel:** [Cardiovascular](/submissions/CV)

## Identification

An intra-aortic balloon and control system is a prescription device that consists of an inflatable balloon, which is placed in the aorta to improve cardiovascular functioning during certain life-threatening emergencies, and a control system for regulating the inflation and deflation of the balloon. The control system, which monitors and is synchronized with the electrocardiogram, provides a means for setting the inflation and deflation of the balloon with the cardiac cycle.

## Classification Rationale

(1) Class II (special controls) when the device is indicated for acute coronary syndrome, cardiac and non-cardiac surgery, or complications of heart failure. (2) Class III (premarket approval) when the device is indicated for septic shock and pulsatile flow generation.

## Special Controls

*Classification.* (1) Class II (special controls) when the device is indicated for acute coronary syndrome, cardiac and non-cardiac surgery, or complications of heart failure. The special controls for this device are:(i) Appropriate analysis and non-clinical testing must be conducted to validate electromagnetic compatibility and electrical safety of the device;
(ii) Software verification, validation, and hazard analysis must be performed;
(iii) The device must be demonstrated to be biocompatible;
(iv) Sterility and shelf-life testing must demonstrate the sterility of patient-contacting components and the shelf life of these components;
(v) Non-clinical performance evaluation of the device must demonstrate mechanical integrity, durability, and reliability to support its intended purpose; and
(vi) Labeling must include a detailed summary of the device- and procedure-related complications pertinent to use of the device.
(2) Class III (premarket approval) when the device is indicated for septic shock and pulsatile flow generation.
(c)
*Date premarket approval application (PMA) or notice of completion of product development protocol (PDP) is required.* A PMA or notice of completion of a PDP is required to be filed with the Food and Drug Administration on or before March 31, 2014, for any intra-aortic balloon and control system indicated for septic shock or pulsatile flow generation that was in commercial distribution before May 28, 1976, or that has, on or before March 31, 2014, been found to be substantially equivalent to any intra-aortic balloon and control system indicated for septic shock or pulsatile flow generation that was in commercial distribution before May 28, 1976. Any other intra-aortic balloon and control system indicated for septic shock or pulsatile flow generation shall have an approved PMA or declared completed PDP in effect before being placed in commercial distribution.

## Recent Cleared Devices (0 of 0)

_No cleared devices on record._

---

**Source:** [https://fda.innolitics.com/productcode/NKO](https://fda.innolitics.com/productcode/NKO)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).