MMH · Blood Establishment Computer Software And Accessories

Hematology · 21 CFR 864.9165 · Class 2

Overview

Product CodeMMH
Device NameBlood Establishment Computer Software And Accessories
Regulation21 CFR 864.9165
Device ClassClass 2
Review PanelHematology

Identification

Blood establishment computer software (BECS) is a device used in the manufacture of blood and blood components to assist in the prevention of disease in humans by identifying ineligible donors, by preventing the release of unsuitable blood and blood components for transfusion or for further manufacturing into products for human treatment or diagnosis, by performing compatibility testing between donor and recipient, or by performing positive identification of patients and blood components at the point of transfusion to prevent transfusion reactions. This generic type of device may include a BECS accessory, a device intended for use with BECS to augment the performance of the BECS or to expand or modify its indications for use.

Classification Rationale

Class II (special controls). The special controls for these devices are:

Special Controls

*Classification.* Class II (special controls). The special controls for these devices are:(1) Software performance and functional requirements including detailed design specifications ( *e.g.,* algorithms or control characteristics, alarms, device limitations, and safety requirements).(2) Verification and validation testing and hazard analysis must be performed. (3) Labeling must include: (i) Software limitations; (ii) Unresolved anomalies, annotated with an explanation of the impact on safety or effectiveness; (iii) Revision history; and (iv) Hardware and peripheral specifications. (4) Traceability matrix must be performed. (5) Performance testing to ensure the safety and effectiveness of the system must be performed, including when adding new functional requirements ( *e.g.,* electrical safety, electromagnetic compatibility, or wireless coexistence).

Recent Cleared Devices (20 of 20)

RecordDevice NameApplicantDecision DateDecision
BK261333Blood Product Questionnaire Module (v2.0.0)Buoy Software, LLC Austin, TX 78704Jun 8, 2026SESE
BK251256ImmuLINK (v3.3)Immucor, Inc. Norcross, Georgia 30071Oct 24, 2025SESE
BK251241SafeTrace Tx® Software 5.0.0Haemonetics Corporation Boston, MA 02110Sep 10, 2025SESE
BK251188Patient Health Software (P.H.S) v11.0.0.0MAK-SYSTEM Group Ltd. London United KingdomJun 11, 2025SESE
BK241136Lookback Notification System version 2.0BioLife Plasma Services LP Bannockburn, ILFeb 20, 2025SESE
BK241061bexWISE API v1.0IT Synergistics, LLC Flowood, MSOct 25, 2024SESE
BK241081Blood Product Administration Module (BPAM 4-2023)Epic Systems Corporation Verona, WIAug 21, 2024SESE
BK241080ImmuLINK (v3.2)Immucor, Inc. Norcross, GAAug 6, 2024SESE
BK241069Blood Bank Information Management System (BBIMS) v3.01Carter BloodCare Bedford, TXJul 9, 2024SESE
BK241062BloodTrack Software 4.16.0Haemonetics Corporation Boston, MAJun 21, 2024SESE
BK230958HemDonER 1.0.1eDonorSoft, LLC Sugarland, TXMay 31, 2024SESE
BK230814SmartDMSLife Plasma, Inc. Reading, PAApr 16, 2024SESE
BK231036NexLynk DMS® Donor Management System 4.10.0Haemonetics Corporation Boston, MAMar 15, 2024SESE
BK231004Blood Product Questionnaire ModuleBuoy Software, LLC Westport, CTNov 16, 2023SESE
BK230986ePROGESA v5.0.3MAK-SYSTEM Group Ltd. London United KingdomNov 3, 2023SESE
BK230811Blood Typing Manager, version 3.0Diagnostic Grifols, S. A. Barcelona, SpainJul 24, 2023SESE
BK230826WellSky® Transfusion 2022 R3WellSky Corporation Overland Park, KSJul 19, 2023SESE
BK220791e-Delphyn DonorGPI USA, Inc. Houston, TXJun 26, 2023SESE
BK230806ImmuLINK (v3.0)Immucor, Inc. Norcross, GAApr 25, 2023SESE
BK220790Blood Bank Control System, Version 6.0Blood Bank Computer Systems, Inc Bonney Lake, WAFeb 17, 2023SESE

Top Applicants

Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...