ING · Elevator, Wheelchair, Portable

Physical Medicine · 21 CFR 890.3930 · Class 2

Overview

Product CodeING
Device NameElevator, Wheelchair, Portable
Regulation21 CFR 890.3930
Device ClassClass 2
Review PanelPhysical Medicine
3rd-Party ReviewableYes

Identification

A permanently mounted wheelchair platform lift is a motorized vertical or inclined platform lift device permanently installed in one location that is intended for use in mitigating mobility impairment caused by injury or other disease by providing a guided platform to move a person from one level to another, with or without a wheelchair. A portable wheelchair elevator is a motorized lift device that is not permanently mounted in one location and that is intended for use in mitigating mobility impairment caused by injury or other disease by providing a means to move a person, with or without a wheelchair, from one level to another (e.g., portable platform lifts, attendant-operated stair climbing devices for wheelchairs).

Classification Rationale

Class II. The permanently mounted wheelchair platform lift is exempt from premarket notification procedures in subpart E of part 807 of this chapter, subject to § 890.9 and the following conditions for exemption. Portable wheelchair elevators are Class II.

Special Controls

(i) Appropriate analysis and nonclinical testing (such as that outlined in the currently FDA-recognized edition of ASME A18.1 “Safety Standard for Platform Lifts and Stairway Chair Lifts”) must demonstrate that the safety controls are adequate to prevent a free fall of the platform in the event of a device failure; (ii) Appropriate analysis and nonclinical testing (such as that outlined in the currently FDA-recognized edition of ASME A18.1 “Safety Standard for Platform Lifts and Stairway Chair Lifts”) must demonstrate the ability of the device to withstand the rated load with an appropriate factor of safety; (iii) Appropriate analysis and nonclinical testing (such as that outlined in the currently FDA-recognized edition of ASME A18.1 “Safety Standard for Platform Lifts and Stairway Chair Lifts”) must demonstrate the ability of the enclosures to prevent the user from falling from the device; and (iv) Appropriate analysis and nonclinical testing (such as that outlined in the currently FDA-recognized editions of AAMI/ANSI/IEC 60601-1-2, “Medical Electrical Equipment—Part 1-2: General Requirements for Safety—Collateral Standard: Electromagnetic Compatibility—Requirements and Tests,” and ASME A18.1 “Safety Standard for Platform Lifts and Stairway Chair Lifts”) must validate electromagnetic compatibility and electrical safety.

Recent Cleared Devices (20 of 29)

Showing 20 most recent of 29 cleared devices.

RecordDevice NameApplicantDecision DateDecision
K113065RL, UL, SL, EL, AND EPL VERTICAL PLATFORM LIFTSBella Elevator, LLCFeb 22, 2012SESE
K110495LIFTKAR PTSano Transportgerate GmbHMay 20, 2011SESE
K102088JOURNEY INCLINED PLATFORM LIFT AND SUPRA CURVED INCLINE PLATFORM LIFTThyssenkrupp Ceteco SrlOct 20, 2010SESE
K091881RPL/CPL VERTICAL PLATFORM LIFTSHarmar Summit, LLCJul 1, 2009SESE
K090856JOURNEY (OR RPSP), JOURNEY CUBIC (OR SUPRA OR TP02)Thyssenkrupp Ceteco SrlMay 27, 2009SESE
K073513MERITS E700 PORCE LIFTMerits Health Products Co., Ltd.Jan 14, 2008SESE
K061514BRUNO VERTICAL PLATFORM LIFT, MODEL VPL-3100Bruno Independent Living Aids, Inc.Jul 13, 2006SESE
K033469GENESIS VERTICAL PLATFORM LIFTGaraventa AccessibilityJan 7, 2004SESE
K024346THE MULTILIFTServices Industriels Savaria, Inc.Mar 4, 2003SESE
K020933GARAVENTA SUPER-TRAC TRE-52Garaventa AccessibilityApr 4, 2002SESE
K983526VERTICAL LIFTBarnes,Richardson & ColburnDec 11, 1998SESE
K981486STAIR-LIFTGaraventa (Canada) , Ltd.Jul 1, 1998SESE
K964134TRUS<T>LIFT PORCH LIFTTrus T LiftDec 31, 1996SESE
K960739V-1504 VERTICAL PLATFORM LIFTBarnes,Richardson & ColburnApr 1, 1996SESE
K960276WHEELCHAIR ELEVATORServices Industriels Savaria, Inc.Apr 1, 1996SESE
K952946VANGATOR IICrow River Industries, Inc.Nov 29, 1995SESE
K941642ASCENDAAmerican Stair-Glide Corp.Jun 13, 1994SESE
K934553WHEEL-O-VATORThe National Wheel-O-Vator Co., Inc.Feb 17, 1994SESE
K920105SCALAMOBILAlber Technologies, Inc.Aug 6, 1992SESE
K922624ULTRAGATER WHEELCHAIR LIFTCrow River Industries, Inc.Jul 13, 1992SESE

Top Applicants

Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...