# FKR · Subsystem, Proportioning

_Gastroenterology, Urology · 21 CFR 876.5820 · Class 2_

**Canonical URL:** https://fda.innolitics.com/productcode/FKR

## Overview

- **Product Code:** FKR
- **Device Name:** Subsystem, Proportioning
- **Regulation:** [21 CFR 876.5820](/submissions/GU/subpart-f%E2%80%94therapeutic-devices/876.5820)
- **Device Class:** 2
- **Review Panel:** [Gastroenterology, Urology](/submissions/GU)
- **Life-sustaining:** yes

## Identification

A hemodialysis system and accessories is a device that is used as an artificial kidney system for the treatment of patients with renal failure or toxemic conditions and that consists of an extracorporeal blood system, a conventional dialyzer, a dialysate delivery system, and accessories. Blood from a patient flows through the tubing of the extracorporeal blood system and accessories to the blood compartment of the dialyzer, then returns through further tubing of the extracorporeal blood system to the patient. The dialyzer has two compartments that are separated by a semipermeable membrane. While the blood is in the blood compartment, undesirable substances in the blood pass through the semipermeable membrane into the dialysate in the dialysate compartment. The dialysate delivery system controls and monitors the dialysate circulating through the dialysate compartment of the dialyzer.(1) The extracorporeal blood system and accessories consists of tubing, pumps, pressure monitors, air foam or bubble detectors, and alarms to keep blood moving safely from the blood access device and accessories for hemodialysis (§ 876.5540) to the blood compartment of the dialyzer and back to the patient. (2) The conventional dialyzer allows a transfer of water and solutes between the blood and the dialysate through the semipermeable membrane. The semipermeable membrane of the conventional dialyzer has a sufficiently low permeability to water that an ultrafiltration controller is not required to prevent excessive loss of water from the patient's blood. This conventional dialyzer does not include hemodialyzers with the disposable inserts (Kiil type) (§ 876.5830) or dialyzers of high permeability (§ 876.5860). (3) The dialysate delivery system consists of mechanisms that monitor and control the temperature, conductivity, flow rate, and pressure of the dialysate and circulates dialysate through the dialysate compartment of the dialyzer. The dialysate delivery system includes the dialysate concentrate for hemodialysis (liquid or powder) and alarms to indicate abnormal dialysate conditions. This dialysate delivery system does not include the sorbent regenerated dialysate delivery system for hemodialysis (§ 876.5600), the dialysate delivery system of the peritoneal dialysis system and accessories (§ 876.5630), or the controlled dialysate delivery system of the high permeability hemodialysis system § 876.5860). (4) Remote accessories to the hemodialysis system include the unpowered dialysis chair without a scale, the powered dialysis chair without a scale, the dialyzer holder set, dialysis tie gun and ties, and hemodialysis start/stop tray.

## Classification Rationale

(1) Class II (performance standards) for hemodialysis systems and all accessories directly associated with the extracorporeal blood system and the dialysate delivery system.(2) Class I for other accessories of the hemodialysis system remote from the extracorporeal blood system and the dialysate delivery system, such as the unpowered dialysis chair, hemodialysis start/stop tray, dialyzer holder set, and dialysis tie gun and ties. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.

## Recent Cleared Devices (12 of 12)

| Record | Device Name | Applicant | Decision Date | Decision |
| --- | --- | --- | --- | --- |
| [K140571](/device/K140571.md) | NXSTAGE PUREFLOW SL | Nxstage Medical, Inc. | May 15, 2014 | SESE |
| [K111174](/device/K111174.md) | NXSTAGE PUREFLOW SL | Nxstage Medical, Inc. | Sep 19, 2011 | SESE |
| [K080919](/device/K080919.md) | NXSTAGE PUREFLOW SL | Nxstage Medical, Inc. | Oct 15, 2008 | SESE |
| [K060296](/device/K060296.md) | NXSTAGE DIALYSATE PREPARATION MODULE | Nxstage Medical, Inc. | Mar 31, 2006 | SESE |
| [K043436](/device/K043436.md) | NXSTAGE DIALYSATE PREPARATION MODULE, MODEL NX0269 | Nxstage Medical, Inc. | Mar 17, 2005 | SESE |
| [K850435](/device/K850435.md) | SARTORIUS AQUASART SM 40035 - DIALYSATE SYS | Sartorius Filters, Inc. | Jun 4, 1985 | SESE |
| [K850756](/device/K850756.md) | AUTO-D AUTOMATIC DISINFECTION ACCESS/ATTACHMENTS | Travenol Laboratories, S.A. | Mar 20, 1985 | SESE |
| [K810605](/device/K810605.md) | CHEMATIC BICARBONATE DIALYSATE SYSTEM | Trimedyne, Inc. | Jun 16, 1981 | SESE |
| [K810127](/device/K810127.md) | GAMBRO BICARBONATE MODULE, BCM-10 | Gambro, Inc. | Feb 25, 1981 | SESE |
| [K802465](/device/K802465.md) | BICARB-O-MATE HEMODIALYSIS SOD. BICARB. | Renal Systems, Inc. | Nov 12, 1980 | SESE |
| [K800772](/device/K800772.md) | MOD. DM-357 BICARBONATE RETROFIT KIT | Extracorporeal Medical Specialities, Inc. | May 14, 1980 | SESE |
| [K781303](/device/K781303.md) | BIO-PRO UNIT | Bd Becton Dickinson Vacutainer Systems Preanalytic | Oct 2, 1978 | SESE |

## Top Applicants

- Nxstage Medical, Inc. — 5 clearances
- Bd Becton Dickinson Vacutainer Systems Preanalytic — 1 clearance
- Extracorporeal Medical Specialities, Inc. — 1 clearance
- Gambro, Inc. — 1 clearance
- Renal Systems, Inc. — 1 clearance

---

**Source:** [https://fda.innolitics.com/productcode/FKR](https://fda.innolitics.com/productcode/FKR)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com