VISTAKON (LENEFILCON A) SOFT (HYDROPHILIC) CONTACT LENS, CLEAR AND VISIBILITY TINTED WITH UV BLOCKER
P990085 · Johnson & Johnson Vision Care, Inc. · LPM · Feb 16, 2001 · Ophthalmic
Device Facts
Record ID
P990085
Device Name
VISTAKON (LENEFILCON A) SOFT (HYDROPHILIC) CONTACT LENS, CLEAR AND VISIBILITY TINTED WITH UV BLOCKER
Applicant
Johnson & Johnson Vision Care, Inc.
Product Code
LPM · Ophthalmic
Decision Date
Feb 16, 2001
Decision
APWD
Regulation
21 CFR 886.5925
Device Class
Class 3
Indications for Use
The VISTAKON Spherical (lenefilcon A) Contact Lens is indicated for daily and extended wear for the correction of refractive ametropia (myopia and hyperopia) in not-aphakic persons with non-diseased eyes. The lens is indicated for daily wear in aphakic persons. The lens may be worn by persons who have astigmatism of 1.00 diopter or less which does not interfere with visual acuity. The VISTAKON BIFOCAL (lenefilcon A) Contact Lens is indicated for daily and extended wear for the correction of distance and near vision in presbyopic, not-aphakic persons with non-diseased eyes. The lens is indicated for daily wear in aphakic persons. The lens may be worn by persons who have astigmatism of 1.00 diopter or less which does not interfere with visual acuity. The VISTAKON TORIC (lenefilcon A) Contact Lens is indicated for daily and extended wear for the correction of visual acuity in not-aphakic persons with non-diseased eyes that are myopic or hyperopic and have astigmatism of 10.00 diopters or less. The lens is indicated for daily wear in aphakic persons. The VISTAKON TORIC BIFOCAL (lenefilcon A) Contact Lens is indicated for daily and extended wear for the correction of distance and near vision in presbyopic not-aphakic persons with non-diseased eyes that have astigmatism of 10.00 diopters or less. The lens is indicated for daily wear in aphakic persons. VISTAKON (lenefilcon A) UV Blocking Contact Lenses help protect against transmission of harmful UV radiation to the cornea and into the eye. The lenses may be prescribed for either daily wear or extended wear from 1 – 7 days between removals for cleaning and disinfection or disposal, as recommended by the eye care practitioner. Eye care practitioners may prescribe the lens either for single-use disposable wear or frequent/planned replacement wear, with cleaning, disinfection and scheduled replacement. When prescribed for frequent/planned replacement wear, the lens may be disinfected using a chemical disinfection system recommended for soft (hydrophilic) contact lenses.
Device Story
VISTAKON (lenefilcon A) soft contact lenses; spherical, bifocal, toric, and toric bifocal designs; 55% water content; 45% lenefilcon A copolymer. Intended for daily or extended wear (up to 7 days). Prescribed by eye care practitioners for refractive error correction. Lenses provide UV protection. Clinical performance evaluated via prospective, randomized, multi-center study (n=457) comparing against group IV control lens. Primary endpoints: safety (adverse events, slit lamp findings) and effectiveness (wearing time, visual acuity). Results showed comparable performance to control; no severe adverse reactions in trial group. Benefits include vision correction and extended wear convenience. Packaging modifications addressed early adhesion issues. Sterilized lenses with 4-year shelf life.
Clinical Evidence
Prospective, randomized, controlled, multi-center clinical trial (n=457; 305 test, 152 control). 1-year duration. Primary endpoints: safety (adverse events, slit lamp findings) and effectiveness (wearing time, visual acuity). 85.8% of test subjects maintained 7-day extended wear. No statistically significant differences in wearing time or refractive changes compared to control. No severe adverse reactions reported in trial group. Bench testing included biocompatibility (ocular irritation, cytotoxicity, systemic toxicity) and physical/chemical characterization.
Technological Characteristics
Material: 55% water, 45% lenefilcon A (copolymer of 2-hydroxyethyl 2-methyl-2-propenoate and glycerol monomethacrylate cross-linked with poly (oxy-1,2-ethanediyl) (4) bis (2-methyl-2-propenoate)). Group II (high water, non-ionic). Oxygen permeability (Dk): 16.4 (cm²/sec)(ml.mmHG) x 10⁻¹¹. Refractive index: 1.43. Specific gravity: 1.05–1.19. UV blocking capability. Sterilized.
Indications for Use
Indicated for not-aphakic persons (myopia, hyperopia, presbyopia, astigmatism up to 10.00D) and aphakic persons (daily wear only) with non-diseased eyes. Contraindicated for those with anterior chamber inflammation/infection, eye disease/injury, severe dry eyes, corneal hypoesthesia, systemic eye-affecting diseases, ocular allergies, or active corneal infections.
Regulatory Classification
Identification
A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.
Submission Summary (Full Text)
{0}
# SUMMARY OF SAFETY AND EFFECTIVENESS DATA
## I. GENERAL INFORMATION
Device Generic Name: lenefilcon A hydrophilic contact lens
Device Trade Name: VISTAKON (lenefilcon A) Soft (hydrophilic) Contact Lenses (Clear and Visibility Tinted with UV Blocker) for Extended Wear
Applicant’s Name and Address: VISTAKON
Division of Johnson & Johnson
Vision Care, Inc.
P.O. Box 10157
Jacksonville, FL 32247-0157
Date(s) of Panel Recommendation: None
Premarket Approval Application (PMA) Number: P990085
Date of Quality System Good Manufacturing Inspection: February 2, 2001
Date of Notice of Approval to Applicant: February 16, 2001
## II. INDICATIONS FOR USE
The VISTAKON Spherical (lenefilcon A) Contact Lens is indicated for daily and extended wear for the correction of refractive ametropia (myopia and hyperopia) in not-aphakic persons with non-diseased eyes. The lens is indicated for daily wear in aphakic persons. The lens may be worn by persons who have astigmatism of 1.00 diopter or less which does not interfere with visual acuity.
The VISTAKON BIFOCAL (lenefilcon A) Contact Lens is indicated for daily and extended wear for the correction of distance and near vision in presbyopic, not-aphakic persons with non-diseased eyes. The lens is indicated for daily wear in aphakic persons. The lens may be worn by persons who have astigmatism of 1.00 diopter or less which does not interfere with visual acuity.
The VISTAKON TORIC (lenefilcon A) Contact Lens is indicated for daily and extended wear for the correction of visual acuity in not-aphakic persons with non-diseased eyes that are myopic or hyperopic and have astigmatism of 10.00 diopters or less. The lens is indicated for daily wear in aphakic persons.
{1}
The VISTAKON TORIC BIFOCAL (lenefilcon A) Contact Lens is indicated for daily and extended wear for the correction of distance and near vision in presbyopic not-aphakic persons with non-diseased eyes that have astigmatism of 10.00 diopters or less. The lens is indicated for daily wear in aphakic persons.
VISTAKON (lenefilcon A) UV Blocking Contact Lenses help protect against transmission of harmful UV radiation to the cornea and into the eye.
The lenses may be prescribed for either daily wear or extended wear from 1 – 7 days between removals for cleaning and disinfection or disposal, as recommended by the eye care practitioner. Eye care practitioners may prescribe the lens either for single-use disposable wear or frequent/planned replacement wear, with cleaning, disinfection and scheduled replacement. When prescribed for frequent/planned replacement wear, the lens may be disinfected using a chemical disinfection system recommended for soft (hydrophilic) contact lenses.
The subject lenses were cleared for daily wear under K983912 on January 26, 1999.
### III. CONTRAINDICATIONS
Do not use the VISTAKON (lenefilcon A) Contact Lenses when any of the following conditions exists:
- Acute or subacute inflammation or infection of the anterior chamber of the eye
- Any eye disease, injury or abnormality that affects the cornea, conjunctiva, or eyelids
- Severe insufficiency of lacrimal secretion (dry eyes)
- Corneal hypoesthesia (reduced corneal sensitivity), if not-aphakic
- Any systemic disease that may affect the eye or be exaggerated by wearing contact lenses
- Allergic reactions of ocular surfaces or adnexa that may be induced or exaggerated by wearing contact lenses or use of contact lens solutions
- Allergy to any ingredient, such as mercury or Thimerosal, in a solution which is to be used to care for the VISTAKON Contact Lens
- Any active corneal infection (bacterial, fungal, protozoal or viral)
- If eyes become red or irritated
### IV. WARNINGS AND PRECAUTIONS
The warnings and precautions can be found in the VISTAKON (lenefilcon A) Soft (hydrophilic) Contact Lens labeling (Attachment 1)
### V. DEVICE DESCRIPTION
The VISTAKON (lenefilcon A) Soft (hydrophilic) Contact Lenses are available in spherical, bifocal, toric, and toric bifocal designs. The lens material is 55% water and
Page 2
{2}
45% lenefilcon A, a copolymer of 2-hydroxyethyl 2-methyl-2-propenoate and glycerol monomethacrylate cross-linked with poly (oxy-1,2-ethanediyl) (4) bis (2-methyl-2-propenoate).
The lens may be prescribed for extended wear in spherical lens powers ranging from +8.00 D to -13.00 D, and toric lens powers from +8.00 D to -10.00 D sphere with cylinder powers from -0.25 D to -10.00 D. Bifocal lens add powers range from +0.25 D to +5.00 D
## VI. ALTERNATIVE PRACTICES AND PROCEDURES
The alternative practices and procedures to correcting vision include other daily wear and extended wear contact lenses, rigid gas permeable daily wear and extended wear contact lenses, spectacles, and corrective surgeries such as radial keratotomy, photorefractive keratectomy and LASIK.
## VII. MARKETING HISTORY
The subject device has not been marketed in the United States or any other country for extended wear.
## VIII. POTENTIAL ADVERSE EFFECTS OF THE DEVICE ON HEALTH
Potential adverse effects on health associated with extended wear contact lenses are included in the device labeling. Please refer to page 6 of this summary for a list of adverse reactions which were observed during the clinical study.
## IX. SUMMARY OF PRECLINICAL STUDIES
The objective of the preclinical studies were to provide reasonable assurance of the safety of the VISTAKON (lenefilcon A) Soft (hydrophilic) Contact Lenses prior to clinical testing.
### Biocompatibility Studies
The following toxicology tests were performed: Primary Ocular Irritation Test, Cytotoxicity Test, Acute Systemic Toxicity and Leachables Testing. The test results raise no acute toxicological concerns and support the safety of the study lens for its intended use.
//
{3}
# Physical and Chemical Characterization Studies:
The physical/optical properties of the lens are:
| % Water Content | 55 |
| --- | --- |
| Oxygen Permeability (Dk): | 16.4 |
| (cm²/sec)(ml.mmHG) x 10⁻¹¹ | |
| % Light Transmittance | 95% Clear, 85% Tinted |
Mechanical Properties:
| Refractive Index: | 1.43 |
| --- | --- |
| Specific Gravity | 1.05 – 1.19 |
# Solution Compatibility Studies:
The subject device is a group II (high water content, non-ionic material) contact lens. Currently marketed lens care solutions have been tested with group II lenses. The applicant provided information from 13 brief (2 to 4 week) small scale studies. These studies enrolled only 14 to 37 subjects each, and were conducted to evaluate the role of biguanide preserved solutions on the presence of superficial punctate keratitis (SPK). Although the prevalence of SPK observed in some of these studies was higher than anticipated, the results were inconclusive.
# Shelf-life Stability Studies
An expiration date of 4 years has been established for sterilized lenses.
# X. SUMMARY OF CLINICAL STUDIES
## Objective
The objective of this clinical trial was to evaluate the safety and effectiveness of the subject device for not-aphakic extended wear from 1 to 7 days.
## Study Design
This clinical trial was a prospective, randomized, controlled, multi-center clinical trial lasting one year. A total of 457 subjects (914 eyes) (305 test subjects and 152 control subjects) were enrolled at 18 investigative sites throughout the United States. There were 10 subjects enrolled but not dispensed lenses because they did not meet the eligibility criteria. Subjects were randomized to wear either the test or control lens bilaterally for one year. The control lens used in this study was a group IV, 55% water content lens currently marketed for extended wear.
Page 4
{4}
For the purposes of this study, extended wear meant that lenses were to be applied and worn around the clock, including during sleep for no more than 7 days/6 nights. Weekly extended wear meant at the end of any six (6) nights of extended wear, the patient had to remove the lens for one night prior to beginning a new cycle of lens wear. All subjects used the Opti-Free Multi-Purpose Solution as the lens care regimen when lenses were removed.
Follow-up visits were to be scheduled at day 1, week 1, month 1, 3, 6, 9 and 12. Primary safety data collected during the study included:
- Adverse events
- Slit lamp findings
- Subjective symptoms and problems
- Keratometric and refractive changes
- Visual acuity
The primary efficacy data collected during the study included:
- Lens wearing time
- Discontinuations
- Lens Replacements
Subjects were eligible for study participation if they were at least 18 years old, signed informed consent, and complied with inclusion and exclusion criteria specified in the protocol.
## Demographic Data
The study population consisted of subjects representative of the general population attending offices for contact lens care. Of the subjects enrolled, 305 test and 152 control subjects were dispensed lenses. Lens power ranged from $-2.00$ D to $-5.00$ D.
Gender distribution for the enrolled subjects consisted of 301 females and 166 males enrolled into the clinical study and ranged in age from 18 to 60 years. Although the potential exists for minor differences in physiological response by gender for the target population, minimal number of clinically significant findings does not indicate that gender differences are of clinical importance for this device.
{5}
# Data Analysis and Results
## Adverse Reactions
Adverse events were defined as including, but not limited to a hazardous, sight threatening condition such as: corneal ulcer, iritis, other ocular infections, inflammations, corneal scarring, or permanent loss of vision.
There were 125 adverse reactions (50 in the control group and 75 in the trial group) reported for 40 control eyes and 64 trial eyes during the clinical study. In some cases, multiple adverse reactions were reported for eyes in both groups. There were no adverse reactions reported as serious/severe for the trial group. Adverse reactions were reported as follows:
| Finding | No. Events | |
| --- | --- | --- |
| | Control Group | Trial Group |
| Infiltrates | 8 | 3 |
| Abrasion | 3 | 2 |
| Corneal ulcer | 6 | 0 |
| Conjunctivitis | 21 | 41 |
| CLARE | 5 | 3 |
| Iritis | 5 | 1 |
| Hypoxia | 2 | 0 |
| GPC | 6 | 13 |
| Foreign body | 1 | 0 |
| Keratitis | 4 | 9 |
| Uvitis | 0 | 2 |
| SPK | 0 | 2 |
| Bacterial infection | 0 | 1 |
There were no adverse reactions reported as severe in the Trial group. All adverse reactions in this group were reported as mild or moderate.
## Slit Lamp Findings
Slit lamp exam results were graded 0 to 4. There were 10 non-serious events (3 control, 7 trial) requiring treatment. There was no apparent trend over time for either group. Findings were not statistically different between the 2 groups. For eyes that completed the study positive slit lamp findings greater than grade 2 were:
Page 6
{6}
corneal edema – 0.11% had grade 3+ (control 0.28%).
staining – 0.03% had grade 3+ (control 0.39%).
vascularization – 0% had grade 3+ (control 0%).
injection – 0.19% had grade 3+ (control 0.39%)
infiltrate – 0% had grade 3+ (control 0.06%). No corneal ulcers in trial eyes that completed the study.
tarsal abnormalities – 0.03% had grade 3+ (control 0.11%).
“other” – most prevalent grade 3 findings were conjunctivitis or lid edema, grade 2 were compression rings. 0.27% had grade 3+ (control 0.62%).
In each case for both groups the conditions resolved with no damage to the eyes with the exception of one eye with a grade 3 infiltrate having a scar after treatment. All visual acuities were restored to acuities prior to the events occurring.
## Symptoms/Problems/Complaints
There were 29 non-serious events (10 control, 19 trial) that required treatment; none resulted in damage to eye. Among completed eyes, dryness was most prevalent (15.5% trial, 10.5% control), followed by blurred vision (5% trial, 5.4% control) and discomfort (4.5% trial, 5.1% control). Among discontinued eyes, dryness was most prevalent (10.5% trial, 10.5% control), followed by discomfort (5.6% trial, 8.7% control) and other (6.3% trial, 5.3% control). There were no statistically significant differences in tearing, photophobia, halos, itch/burning, spectacle blur, and other
The trial lenses and control lenses performed similarly with respect to patient reported symptoms, problems and complaints. The most frequently reported symptom was dryness for both groups.
## Keratometric/Refractive Changes
**Keratometry Changes:** There were 2 reports in the trial group of mire clarity change from baseline to final exam, but no associated change in BCVA. K measurement changes were substantially clinically equivalent between trial and control groups. Mean K change at the final exam was 0.31/0.34D for completed controls, and 0.34/0.36D for completed trial eyes.
**Refractive Changes:** Refractive changes were substantially clinically equivalent between trial and control groups. Of completed eyes at 12 months, mean change was 0.22D control and 0.23D trial eyes. Discontinued eyes also had similar mean refractive changes, 0.13D control and 0.12D trial eyes.
## Visual Acuity
Visual acuity was reported initially and at each follow-up visits for subjects in both the Control and Trial groups of the study. At the final visit 60.6% of eyes in the Control
15
{7}
group had visual acuity of 20/20 or better and 87.4% had visual acuity better than 20/40. In the Trial group 72.8% had 20/20 or better and 93.4% had better than 20/40 final visual acuity. In the Trial group there were no eyes with final visual acuity worse than 20/40
## Wearing Time
The primary effectiveness endpoint was the percentage of subjects in each group able to successfully maintain the extended wear schedule. For subjects in the Trial group, 85.8% were able to wear lenses for 7 days. For subjects in the Control group, 83.8% were able to wear lenses for 7 days. There are no statistically significant differences for wearing time between the two lens groups with both having an average lens wearing time of 6.87 days.
## Discontinued Patients
There were 82 eyes (27%) discontinued from the Control group and 146 eyes (23.9%) discontinued from the Trial group. Multiple reasons were cited for discontinuance. The main reasons for discontinuing the study were as follows:
| Reason | Control | Trial |
| --- | --- | --- |
| Protocol violation | 6.6% | 6.9% |
| Lost-to-follow-up | 5.9% | 4.3% |
| Discomfort | 3.3% | 3.3% |
Reasons for discontinuance were comparable in both groups.
## Lens Replacements
Of the 304 lenses dispensed in the Control group and 610 lenses dispensed in the Trial group, there were 175 lenses (57.6%) replaced in the Control group and 531 lenses (87.0%) replacements in the Trial group. The most prevalent reasons for replacement were as follows:
| Reason | Control | Trial |
| --- | --- | --- |
| Torn | 17.4% | 56.4% |
| Comfort | 15.8% | 11.6% |
| Visual acuity | 9.2% | 4.8% |
There was a higher rate of unscheduled replacements for the trial group early in the study due to lenses adhering to the blister packages, but this was remedied (trial and control similar rates) once the packaging design was modified.
/
{8}
Page 9
//
# XI. CONCLUSIONS DRAWN FROM STUDY
The results of the preclinical and clinical studies provide reasonable assurance of the safety and effectiveness of the VISTAKON (lenefilcon A) soft contact lens for the patient population, refractive conditions and specified duration of wear.
# XII. PANEL RECOMMENDATION
In accordance with the provisions of section 515(c)(2) of the act as amended by the Safe Medical Devices Act of 1990, this PMA was not referred to the Ophthalmic Devices Panel, an FDA advisory committee, for review and recommendation because the information in the PMA substantially duplicates information previously reviewed by this panel.
# XIII. CDRH DECISION
FDA issued an approval order on February 16, 2001. The applicant’s manufacturing facility was inspected on February 2, 2001 and was found to be in compliance with the device Quality System Regulations.
# XIV. APPROVAL SPECIFICATIONS
Directions for use: See the labeling
Hazards to Health from Use of the Device: See the Indications, Contraindications, Warnings, Precautions and Adverse Events in the labeling.
Postapproval Requirements and Restrictions: See approval order.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.