W-55 (METHAFILCON A) AND HORIZON 55 SOFT EXTENDED WEAR CONTACT LENSES
P990072 · Westcon Contact Lens Co., Inc. · LPM · Aug 22, 2000 · Ophthalmic
Device Facts
Record ID
P990072
Device Name
W-55 (METHAFILCON A) AND HORIZON 55 SOFT EXTENDED WEAR CONTACT LENSES
Applicant
Westcon Contact Lens Co., Inc.
Product Code
LPM · Ophthalmic
Decision Date
Aug 22, 2000
Decision
APWD
Regulation
21 CFR 886.5925
Device Class
Class 3
Indications for Use
The Horizon 55 EW and Horizon 55 EW Westint (methafilcon A) Soft Hydrophilic Contact Lenses for Extended Wear are indicated for extended wear from 1 to 7 days between removal for cleaning, rinsing and disinfecting as recommended by the eye care practitioner. The lenses are indicated for the correction of visual acuity in not-aphakic persons with non-diseased eyes that are myopic or hyperopic. The lenses may be worn by persons who exhibit astigmatism of 1.50 diopters or less that does not interfere with visual acuity.
Device Story
Hydrophilic contact lens made of methafilcon A; used for vision correction in myopic or hyperopic patients. Lens acts as refractive medium to focus light on retina. Worn on extended basis (1-7 days) between removals for cleaning/disinfection. Prescribed by eye care practitioners; patient-operated. Requires chemical lens care system for disinfection. Benefits include visual acuity correction. Device is a hemispherical shell; 15mm diameter; 0.06mm center thickness; 45% water content.
Clinical Evidence
Clinical study of 152 patients (302 eyes) over 12 months. Primary endpoints: safety (slit lamp findings, adverse reactions) and effectiveness (visual acuity, wearing time). One corneal ulcer reported (resolved). Slit lamp findings (edema, staining, injection) were within expected limits. Visual acuity results demonstrated effectiveness in correcting myopia/hyperopia. Study concluded device is safe and effective for intended use.
Technological Characteristics
Material: methafilcon A (hydroxyethylmethacrylate polymer). Water content: 45%. Refractive index: 1.415. Oxygen permeability: 31.3 x 10^-9 (cm/sec) mL O2/mL x mmHg at 25°C. Dimensions: 15.00 mm diameter, 8.6 mm base curve, 0.06 mm center thickness. Shelf-life: 3 years. Chemical disinfection required; heat disinfection contraindicated.
Indications for Use
Indicated for myopic or hyperopic, not-aphakic persons with non-diseased eyes and refractive astigmatism ≤ 1.50 D. Contraindicated for acute/subacute anterior segment inflammation, corneal/conjunctival disease, lacrimal insufficiency, corneal hypoesthesia, systemic disease affecting the eye, active corneal infections, or allergy to lens care solution ingredients.
Regulatory Classification
Identification
A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.
{0}
Summary of Safety and Effectiveness
I. General Information
A. Device Generic Name: methafilcon A hydrophilic contact lenses
B. Device Trade Names: Horizon 55 EW and Horizon 55 EW Westint™ (methafilcon A) Soft Hydrophilic Contact Lenses for Extended Wear
C. Applicant's Name and Address: Westcon Contact Lens Co., Inc.
611 Eisenhauer Street
Grand Junction, CO 81505
D. Premarket Approval Application (PMA) Number: P990072
E. Date of Notice of Approval to Applicant: AUG 22 2000
II. Indications
The Horizon 55 EW and Horizon 55 EW Westint (methafilcon A) Soft Hydrophilic Contact Lenses for Extended Wear are indicated for extended wear from 1 to 7 days between removal for cleaning, rinsing and disinfecting as recommended by the eye care practitioner. The lenses are indicated for the correction of visual acuity in not-aphakic persons with non-diseased eyes that are myopic or hyperopic. The lenses may be worn by persons who exhibit astigmatism of 1.50 diopters or less that does not interfere with visual acuity.
III. Center for Devices and Radiological Health (CDRH) Decision
The application includes by reference the data in PMA P850078 and related supplements for the Kontur Soft EW (methafilcon A) Hydrophilic Contact Lens. Kontur Kontact Lens Co., Inc., in conjunction with two other companies, including Coast Vision, Inc., conducted a clinical study of the methafilcon A contact lens material for daily wear and extended wear. Kontur Kontact Lens Co., Inc., having joint ownership of the data, received approval of P850078/S1 for the extended wear lens under a licensing agreement from Coast Vision, Inc. (P850079/S1), on June 25, 1986. Kontur Kontact Lens Co., Inc. has authorized Westcon Contact Lens Co., Inc. to incorporate by reference the information contained in its approved PMA and related supplements to manufacture the lenses by the lathing technique.
CDRH approval of the Westcon Contact Lens Co., Inc. PMA is based on (1) the safety and effectiveness data contained in PMA P850078 and related supplements and (2) the results of the FDA inspection of Westcon Contact Lens Co., Inc. manufacturing facility. A copy of the approval order for P850078/S1 and the summary of the safety and effectiveness data for the Coast Vision, Inc. PMA supplement, P850079/S1 appears in Attachment A.
{1}
- 2 -
In accordance with the provisions of section 515(c)(2) of the act as amended by the Safe Medical Devices Act of 1990, this PMA was not referred to the Ophthalmic Devices Panel for review and recommendation because the information in the PMA substantially duplicates information previously reviewed by this panel. CERH issued an approval order on AUG 22 2000. The applicant's manufacturing facility was inspected on APR 28 2000, and was found to be in compliance with the device Good Manufacturing Practice regulations.
The device shelf-life has been established and approved at 3 years.
### III. Potential Adverse Effects of the Device on Health
Potential adverse effects on health resulting from the use of this device are listed in the package insert under "ADVERSE REACTIONS" (Attachment B).
### IV. Conditions of Approval
In addition to the standard "Conditions of Approval" enclosed with the approval order, the subject devices are limited to prescription use in accordance with 21 CFR 801.109. A draft copy of the approved package insert is attached (Attachment B). Copies of final printed labeling are available to interested persons for inspection at:
Food and Drug Administration
Center for Devices and Radiological Health
PMA Document Mail Center (HFZ-401)
9200 Corporate Blvd.
Rockville, MD 20850
Attachments A and B
9
{2}
DEPARTMENT OF HEALTH & HUMAN SERVICES
Attachment A
Public Health Service
Food and Drug Administration
8757 Georgia Avenue
Silver Spring MD 20910
JUN 25 1986
David G. Ewell, O.D.
President
Kontur Kontact Lens Co., Inc.
200 South Garrard Blvd.
Richmond, California 94804
Re: P850078/S1
Kontur Soft EW (methfilcon A)
Hydrophilic Contact Lens
Filed: October 2, 1985
Amended: June 5, 1986
Dear Dr. Ewell:
The Center for Devices and Radiological Health (CDRH) of the Food and Drug Administration has completed its review of your premarket approval application (PMA) supplement for the spherical Kontur Soft EW (methafilcon A) Hydrophilic Contact Lens. The lens is indicated for extended wear from 1 to 21 days between removals for cleaning and disinfecting as recommended by the eye care practitioner. The lens is indicated for the correction of visual acuity in not-aphakic persons with nondiseased eyes that are myopic or hyperopic. The lens may be worn by persons who exhibit refractive astigmatism of 1.50 diopters (D) or less that does not interfere with visual acuity. The lens ranges in powers from -10.00 D to +10.00 D. The lens is to be disinfected using a chemical lens care system. The PMA supplement is approved subject to the conditions described below and in the "Conditions of Approval" (enclosed). The shelf-life for the referenced device has been established and approved as 3 years. You may begin production and marketing of the device upon receipt of this letter.
CDRH will publish a notice of its decision to approve your PMA in the FEDERAL REGISTER. The notice will state that a summary of the safety and effectiveness data upon which the approval is based is available to the public upon written request. In addition, the notice will state that a copy of all approved labeling (which may be a draft of the final labeling) is available for public inspection at CDRH. Within 30 days of publication of the notice of approval in the FEDERAL REGISTER, any interested person may seek review of this decision by requesting an opportunity for administrative review, either through a hearing or review by an independent advisory committee, under section 515(g) of the Federal Food, Drug, and Cosmetic Act (act).
{3}
Page 2 - David G. Ewell, O.D.
CDRH approval is subject to full compliance with the conditions described in the enclosure. Additionally, the sale, distribution, and use of the device shall be restricted to prescription use in accordance with 21 CFR 801.109.
All stated requirements are subject to change upon publication of a final premarket approval procedural regulation. Failure to comply with the conditions of approval invalidates this approval order. Commercial distribution of a device that is not in compliance with these conditions is a violation of the act.
You shall submit all required documents in triplicate to the Food and Drug Administration, Center for Devices and Radiological Health, PMA Document Mail Center (HFZ-401), 8757 Georgia Avenue, Silver Spring, Maryland 20910. You shall refer to the above PMA number in all further correspondence to expedite processing.
If you have any questions concerning this approval order, please contact Mr. David M. Whipple at (301) 427-7940.
Sincerely yours,
Kshitij Mohan, Ph.D.
Director
Office of Device Evaluation
Center for Devices and Radiological Health
Enclosure
{4}
Summary of Safety and Effectiveness
## I. General Information
A. Device Generic Name: methafilcon A hydrophilic contact lens
B. Device Trade Name: Hydrasoft XW (methafilcon A) Hydrophilic Contact Lens
C. Applicant's Name and Address: Coast Vision, Inc.
18368 Enterprise Lane
Huntington Beach, California 92648
D. Premarket Approval Application (PMA) Supplement Number: P850079/S1
E. Date of Panel Recommendation: May 23, 1986
F. Date of Notice of Approval to Applicant: JUN 25 1986
## II. Indications
The spherical Hydrasoft (methafilcon A) Hydrophilic Contact Lens (hereafter referred to as the Hydrasoft XW Lens) is indicated for extended wear from 1 to 21 days between removals for cleaning and disinfection as recommended by the eye care practitioner. The lens is indicated for the correction of visual acuity in not-aphakic persons with nondiseased eyes that are myopic or hyperopic. The lens may be worn by persons who may exhibit refractive astigmatism of 1.50 diopters (D) or less that does not interfere with visual acuity. The lens ranges in powers from -10.00 D to +10.00 D. The lens is to be disinfected using a chemical lens care system only.
## III. Device Description
The Hydrasoft XW Lens is a hydrophilic lens made from methafilcon A material.
Physical parameters of the lens are:
Refractive index: 1.415
Light transmission: 98.8%
Surface character: hydrophilic
Water content: 45%
Oxygen permeability: $31.3 \times 10^{-9}$ (cm/sec) mL $O_2/mL$ X mmHg at $25^{\circ}C$ measured by Schema Versat ae model 920 connected to a polarographic cell
{5}
Dimensions of the lens are:
Base curve: 8.6 mm
Chord diameter: 15.00 mm
Center thickness: 0.06 mm
Lens powers: -10.00 D to +10.00 D
IV. Alternative Practices or Procedures
Alternative practices or procedures available to the patient are the use of other extended wear or daily wear contact lenses, or spectacles for the same indication.
V. Summary of Studies
A. Preclinical:
The PMA includes by reference, to the original PMA P820017 and all related supplements, the results of toxicological, microbiological, chemistry and manufacturing data and results. The results of these data provide reasonable assurance that the lens material is safe when manufactured into a lens for use in a clinical study and has a shelf-life of 3 years.
B. Clinical:
Purpose of the Study
The purpose of this clinical study was to demonstrate the safety and effectiveness of the device for its intended use. The clinical study was conducted in accordance with "Clinical Guidelines, Testing Guidelines for Class III Contact Lenses," an FDA guideline dated May 1983.
Patient Selection Criteria
Patients in this study were to meet the following criteria:
1. have need of an optical correction;
2. have spherical ametropia;
3. have reasonable expectation of improvement in visual acuity with contact lenses;
4. use no ocular medication; and
5. have nondiseased eyes which are normal or which have a pre-existing ocular condition which should not be expected to interfere with the patient's ability to wear the contact lens successfully. Such conditions should be adequately documented.
{6}
- 3 -
## Study Population
A total of 152 patients (302 eyes) was enrolled by 11 investigators into the clinical study. Of the 152 enrolled patients, 106 patients (210 eyes) completed the 12-month study, 19 patients (38 eyes) remained active (10 patients had not completed the 12-month study period, and 12-month visit forms had not been received by the applicant for 9 patients at the time the PMA supplement was submitted to the Center for Devices and Radiological Health (CDRH) for review). Twenty-seven patients (54 eyes) were discontinued from the study. Of the 106 completed patients, there were 48 males, 54 females, and 4 with sex not reported, with ages ranging from 15 to 55 years. The completed patients included 184 myopic eyes and 26 hyperopic eyes. All patients in the study were not-aphakic except one. For completed patients pre-existing pathologies were reported as abnormal pupil (1 eye), abnormal retina (1 eye), exotropia (3 eyes), corneal edema (1 eye), neovascularization (1 eye), staining (1 eye), infection (5 eyes), and pinguicula; puntate keratitis (3 eyes). Lenses used in this study ranged in powers from +20.00 D to -9.75 D. All patients in the study used chemical lens care systems. Previous lens wear experience was reported as successful daily wear for 179 eyes, successful extended wear for 4 eyes, unsuccessful daily wear for 27 eyes, no previous lens wear experience for 90 eyes, and unreported for 2 eyes.
## Study Period
The study was initiated on November 29, 1983. The cut-off date for clinical data in the PMA was July 1, 1985. One year of extended wear was required for completed patients.
## Findings
1. Safety:
### Adverse Reactions
In evaluating this device, an adverse reaction was considered to be a serious vision-threatening problem that was unanticipated, but which might have been attributed to the use of the study device.
There was one corneal ulcer reported during the course of this study. The corneal ulcer was reported at the 2-week visit along with grade 3 edema, grade 1 iritis, and grade 2 staining. This patient was discontinued from the study but continued to be seen at follow-up visits to monitor the conditions. At the next visit the patient had grade 1 edema with the ulcer clearing. Five days later all findings for this patient were normal. The investigator reported the ulcer followed a scratched cornea caused by the patient removing the lens.
{7}
- 4 -
Conclusion:
The Center for Devices and Radiological Health (CDRH) has determined that the causal reason and outcome of the adverse reaction does not raise a significant safety concern when the lens is used in accordance with the approved labeling.
## Slit Lamp Findings
A positive slit lamp finding is considered to be a routinely occurring complication that would be expected with or without the presence of contact lenses. The degree of severity can range from very slight to serious. At the least severe, the findings present no medical concern and are noticeable only by microscopic slit lamp examination. In a severe state, the findings require medical treatment.
Slit lamp examinations were performed initially (first visit with lenses) and periodically throughout the study. The applicant used the slit lamp quantification scheme "Classification of Slit Lamp Observations" as outlined in the FDA guideline.
The positive slit lamp findings for the 106 patients (210 eyes) completing the study were reported as follows:
| Slit Lamp Finding | Initial* Visit | Follow-up Visits | Final Visit |
| --- | --- | --- | --- |
| No. Eyes | 188 | 1,607 | 210 |
| Edema | | | |
| Grade 1 | 4 | 75 | 8 |
| Grade 2 | 0 | 5 | 0 |
| Grade 4 | 0 | 2 | 0 |
| Neovascularization | | | |
| Grade 1 | 0 | 2 | 2 |
| Staining | | | |
| Grade 1 | 0 | 34 | 2 |
| Grade 2 | 0 | 7 | 0 |
| Grade 3 | 0 | 1 | 0 |
| Injection | | | |
| Grade 1 | 6 | 82 | 7 |
| Grade 2 | 0 | 3 | 0 |
| Iritis | | | |
| Grade 1 | 0 | 2 | 0 |
| Other | | | |
| Grade 1 | 0 | 6 | 0 |
| Grade 2 | 0 | 2 | 0 |
*22 eyes missed the first visit (2 to 4 hours) with lens, but were available for the 1-week visit.
15
{8}
- 5 -
One patient with handling problems was reported to have grade 2 edema in both eyes at the 1-month visit and again at the 3-month visit in one eye. All findings were normal for this patient when seen 18 days after the 4-month visit. One patient had grade 2 edema in one eye at the 2-month visit. The lens was replaced, and the condition cleared by the next visit. One patient had grade 2 edema at the 2-week visit which cleared by the 1-month visit. One patient had grade 4 edema in both eyes at the 4-month visit. The condition cleared by the next visit.
One patient had grade 3 staining at the 5-month visit due to allergy to thimerosal. The lens was replaced and the patient instructed to use thimerosal free solutions. All findings were normal at the next visit. One patient accounted for 4 grade 2 stainings. The patient had grade 2 staining in both eyes at the 4-month visit which cleared by the 5-month visit and reoccurred at the 8-month visit. The patient was changed to daily wear lenses, and all findings were normal. One patient had grade 2 staining at the 3-month visit which cleared by the 5-month visit.
One patient had grade 2 injection at the 2-month visit along with burning, itching, and tearing. At the 5-month visit all findings were normal for this patient.
Grade 2 "Other" was reported as subepithelial infiltrates for one patient at the 5-month visit. The lens was replaced, and all findings were normal.
Positive slit lamp findings above grade 2 for discontinued patients showed only 1 grade 3 edema finding in 224 follow-up visits. Grade 1 and 2 slit lamp findings were only slightly more frequent than for completed patients.
Conclusion:
The incidence and severity of positive slit lamp findings for discontinued patients were comparable to those for the completed patients in this study. These findings do not raise any significant concerns about the safety of the device for its intended use.
Symptoms, Problems and Complaints
Patient symptoms, problems and complaints were reported by the investigators at each patient visit. A total of 2,638 patient eyes were examined during the course of the study, and multiple patient symptoms, problems and complaints were reported as follows:
16
{9}
- 6 -
| Symptoms, Problems and Complaints | No. Incidences |
| --- | --- |
| Blurred vision | 106 |
| Cleaning/handling | 65 |
| Tearing | 37 |
| Itching | 35 |
| Deposits | 29 |
| Burning | 28 |
| Dryness | 23 |
| Photophobia | 21 |
| Halo | 8 |
| Soiled lens | 7 |
| Redness | 6 |
| Comfort | 5 |
| Red eye | 5 |
| Excessive secretions | 4 |
| Flare | 3 |
| Cleaning | 2 |
| Headache | 2 |
| Fogged vision | 2 |
| Mild GPC* | 2 |
| Corneal ulcer | 1 |
| Scratchy | 1 |
| Lens loose | 1 |
| Discolor | 1 |
| Lump | 1 |
*Giant papillary conjunctivitis
Conclusion:
The patient symptoms, problems and complaints reported during this study are within expected limits and do not raise any significant concerns about the safety and effectiveness of the device.
2. Effectiveness:
**Visual Acuity**
Visual acuity for the 210 eyes completing the study was reported as follows:
| | Initial Best Corrected | Initial with Lens | Final with Lens |
| --- | --- | --- | --- |
| No. Eyes | 210 | 210 | 210 |
| VA | | | |
| 20/20 or better | 192 | 168 | 181 |
| 20/25 | 12 | 16 | 21 |
| 20/30 | 5 | 5 | 6 |
| 20/40 | 0 | 1 | 1 |
| 20/50 | 1 | 2 | 1 |
| Not available | 0 | 18 | 0 |
{10}
- 7 -
For the 210 eyes completing the study, visual acuity decreased 2 or more Snellen lines for 3 eyes for 2 patients. Visual acuity for one patient (2 eyes) changed from 20/20 to 20/30. The patient required a power change in both lenses. New lenses were dispensed, and visual acuity was 20/20 in both eyes. One patient was fit for monovision. The left eye was over corrected for good near vision.
Conclusion:
CDRH has concluded that a decrease of 1 Snellen line is not unusual in a contact lens study due to measuring techniques and normal fluctuation and is not significant in terms of visual acuity. The visual acuity results of this study demonstrate that the incidence of decreased vision for patients is within the expected limits for such patients. CDRH has determined that the decreases were not caused by ocular problems associated with contact lens wear. The visual acuity results in this study demonstrate the effectiveness of the device in correcting visual acuity in myopic and hyperopic not-aphakic eyes.
Lens Wearing Time
At the 6-month visit 136 of 178 eyes were wearing the lens from 7 to 14 days before removal. At the 12-month visit 142 of 194 eyes wore lenses from 7 to 14 days before removal and 30 eyes wore lenses for more than 30 days before removal. The major reason for lens removal was cleaning.
Conclusion:
The lens wearing time data provide reasonable assurance that the lens can be worn from 1 to 21 days as recommended by the eye care practitioner between removals without concerns regarding the safety or effectiveness of the device.
Discontinued Patients
A total of 27 patients (54 eyes) discontinued during the course of this study. Reasons for discontinuing the study were as follows:
{11}
- 8 -
| Reason | No. Eyes |
| --- | --- |
| Unsatisfactory fit, all reasons related to | 10 |
| Patient motivation | 8 |
| Deposits and soiled lenses | 8 |
| Patient lost-to-follow-up | 8 |
| Comfort, all reasons related to | 8 |
| Visual acuity, all reasons related to | 6 |
| Dryness and redness | 2 |
| Reaction to solution | 2 |
| Slit lamp finding (GPC) | 2 |
| Conclusion: | |
The reasons for and incidence of discontinuance in this clinical study are expected and do not raise any significant concerns regarding the safety or effectiveness of the device for its intended use.
## Lens Replacements
There were 107 lenses replaced during the course of this clinical study. Reasons for lens replacements were as follows:
| Reason | No. Replacements |
| --- | --- |
| Physiological | |
| Visual acuity | 17 |
| Visual acuity/deposits | 8 |
| Fit | 7 |
| Fit/vision. | 1 |
| Red eye infiltrates | 1 |
| Physical | |
| Deposits | 25 |
| Lost | 17 |
| Lost/damaged | 15 |
| Soiled lens | 6 |
| Damaged | 5 |
| GPC | 2 |
| Mucous | 1 |
| Handling thin lens | 1 |
| No reason given | 1 |
19
{12}
- 9 -
Conclusion:
The rate of lens replacements in this clinical study is expected when fitting soft (hydrophilic) contact lenses for extended wear.
VI. Potential Adverse Effects of the Device on Health
Potential adverse effects on health resulting from the use of this device are indicated in the package insert under "Adverse Reactions" (Attachment A).
VII. Conclusions Drawn From the Studies
The data provide reasonable assurance that the device is safe and effective for its intended use.
VIII. Panel Recommendation
On May 23, 1986, the Ophthalmic Devices Panel unanimously recommended approval of the PMA subject to the conditions that all administrative requirements be met and that the applicant be in compliance with the device Good Manufacturing Practice regulations.
IX. CDRH Decision
CDRH concluded that the applicant has met the above conditions. Based upon this conclusion, upon information in the PMA and upon review of the labeling, CDRH concurred with the Panel recommendation and approved the application and draft final labeling on
The device shelf-life has been established and approved as 3 years. Based on an on-site Inspection on October 6 and 12, 1982 and a review of the firm's regulatory history, the manufacturing facility is regarded as in compliance with the device Good Manufacturing Practice regulations.
X. Conditions of Approval
In addition to the standard "Conditions of Approval" (Attachment B), the subject device is limited to prescription use in accordance with 21 CFR 801.109 and is limited to use with the chemical lens care system specified in the approved labeling. A copy of the approved draft package insert is attached (Attachment A). Copies of approved draft labeling are available to interested persons for inspection at:
Food and Drug Administration
Center for Devices and
Radiological Health
PMA Document Mail Center (HFZ-401)
8757 Georgia Avenue
Silver Spring, Maryland 20910
Attachments A and B
{13}
Attachment A
# PACKAGE INSERT
Hydrasoft XW (methafilcon A) Hydrophilic Contact Lens
## DESCRIPTION
The Hydrasoft XW (methafilcon A) Hydrophilic Contact Lens is a hemispherical shell of approximately 15mm chord diameter, 0.06mm thickness, power range of -10.00 to +10.00, and base curve of 8.60mm. The lens material is a hydrophilic polymer of hydroxyethylmethacrylate. The lens consists of 45% methafilcon A and 55% water by weight when immersed in normal saline. The material has a refractive index of 1.415 and an oxygen permeability of 31.3 X 10⁻⁹ (cm/sec) ml θ₂/ml X mmHg at 25° C as measured by Schema Versatae model 920 connected to a polarographic cell. The surface character of the lens is hydrophilic. The light transmittance is 98.8%.
## ACTIONS
When placed on the human cornea in its hydrated state, the Hydrasoft XW lens acts as a refractive medium to focus light rays on the retina.
## INDICATIONS
Hydrasoft XW lens is indicated for extended wear from 1 to 21 days between removals for cleaning and disinfecting as recommended by the eye care practitioner. The lens is indicated for the correction of visual acuity in not-aphakic persons with non-diseased eyes that are myopic or hyperopic. The lens may be worn by persons who exhibit refractive astigmatism of 1.50 diopters or less that does not interfere
{14}
with visual acuity. The lens ranges in powers from -10.00 diopters to +10.00 diopters.
## CONTRAINDICATIONS
Hydrasoft XW contact lenses are contraindicated by the presence of any of the following conditions:
1. Acute and subacute inflammations of the anterior segment of the eye.
2. Any eye disease which affects the cornea or conjunctiva.
3. Insufficiency of lacrimal secretion.
4. Corneal hypoesthesia.
5. Any systemic disease which may affect the eye or be exaggerated by wearing contact lenses.
6. Any active corneal infections; bacterial, fungal, viral, or purulent.
7. Allergy to any ingredient such as mercury or thimerosal in a solution which must be used to care for the lens.
## ADVERSE REACTIONS
Serious corneal damage may result from wearing a lens which has been soaked in conventional hard lens solution. Eye irritation may occur within a short time after putting on a lens stored in a solution of improper pH or tonicity. Removal of the lens will relieve the irritation. Excessive tearing or redness, unusual eye secretions, reduced visual acuity, blurred vision, halos of light in the field of vision, and light sensitivity are not normal; if these symptoms occur, the patient should be examined by an eye-care practitioner to determine the cause.
## WARNINGS
{15}
## Medicaments and Eye Drops:
The Hydrasoft XW contact lens must be stored only in the recommended solutions. No ophthalmic solutions or medicaments should be used unless directed by your eye-care practitioner. Never use conventional hard contact lens solutions. Only the recommended disinfectant soaking solutions or cleaning solutions and rinsing solutions should be used. Lubricating solutions (see recommended list of solutions) may be used on the Hydrasoft XW contact lens.
## Abrasions and Infections:
If the lens becomes less comfortable than when it was first placed on the cornea, or the vision becomes less clear, this may indicate the presence of a foreign body, an improperly cleaned lens or the lens may be in the incorrect eye. The lens should be removed immediately and examined. If any corneal abrasion, ulceration, irritation or infection is present, an eye-care practitioner should be consulted immediately.
## Visual Blurring:
When visual blurring occurs, the lens must be removed and cleaned. If visual blurring continues, consult your eye-care practitioner immediately.
## Wearing Restrictions:
The Hydrasoft XW contact lens should not be worn when swimming or in the presence of noxious and irritating vapors.
## PRECAUTIONS
### Lens Handling:
Patients must wash and rinse hands thoroughly and dry with a lint-free towel before handling the lenses. Cosmetics, lotions, soaps, and creams must not come into contact with the
23
{16}
lenses since eye irritation or lens discoloration may result. If hair spray is used while the lenses are being worn, the eyes must be kept closed until the spray has settled.
Fluorescein:
Never use fluorescein while wearing the lens because the lens will become discolored. Whenever fluorescein is used, flush the eyes with normal saline solution and wait at least one hour before replacing the lens. Too early replacement may allow the lens to absorb residual fluorescein irreversibly.
LENS CARE DIRECTIONS
Lens Care and Handling:
Patients must be supplied with lens care instructions and supplies. Regular post-fitting visits are necessary to assure patient health and compliance with instructions.
Storage:
Hybrasoft XW contact lenses must be stored only in the recommended solutions. If left exposed to air, the lenses will dehydrate. If a lens dehydrates, it should be soaked in the recommended storage solution a minimum of four hours.
Cleaning:
Removal for cleaning every three weeks is recommended. However, at the discretion of the eye-care practitioner, the lenses may be removed at a frequency adjusted to the needs of each patient. Each time the lenses are removed from the wearer's eyes, both surfaces of the lenses must be cleaned using several drops of the recommended cleaner.
Chemical Disinfection:
CHEMICAL DISINFECTION of the contact lens is necessary to remove potentially harmful microorganisms from the lens before
24
{17}
placing the lens on the eye. This can be accomplished by using the recommended solutions. Hydrasoft XW contact lenses must be cleaned and rinsed after wearing, with the recommended cleaning solution and saline rinse. The lens case must be emptied and refilled with fresh recommended storage disinfection solution prior to disinfecting the lenses. Fresh storage and disinfection solution must be used for storage and disinfecting the lenses.
WARNING: HEAT DISINFECTION SHOULD NOT BE USED WITH THE HYDRASOFT XW CONTACT LENS.
DOSAGE AND ADMINISTRATION
Fitting Information:
The eye-care practitioner will use conventional methods to fit the Hydrasoft XW contact lens for vision correction. For a detailed description of the fitting technique, refer to the appropriate hydrophilic contact lens professional fitting guide, copies of which are available from CoastVision, Inc., 18368 Enterprise Lane, Huntington Beach, California 92648.
Wearing Schedule:
Not every patient is able to wear a lens on an extended basis, even if able to wear the same lens on a daily basis. The eye-care practitioner should determine the wearing schedule. Periodic checkups by the eye-care practitioner are extremely important, especially for patients on an extended wear regimen. With extended-wear there may be increased risks of eye problems such as irritation, infection, corneal thickening and corneal ulcers.
HOW SUPPLIED
Each lens is supplied sterile in a glass vial containing
25
{18}
sterile normal saline solution USP, buffered with sodium bicarbonate USP. The glass vial is marked with the base curve, dioptric power, diameter, manufacturing lot number, and expiration date of the lens.
To assure proper lens care and handling, each Hydrasoft XW patient MUST BE supplied with the Hydrasoft chemical disinfection patient care kit and the Hydrasoft XW Wearer's Guide. The recommended accessory products required for the chemical disinfection, cleaning and storage of Hydrasoft contact lenses consist of the following:
1. Cleaning - Preflex, Pliagel, Softmate Cleaner, or Allergan Cleaning and Disinfecting Solution.
2. Rinsing - Allergan Hydrocare Preserved Saline Solution, Unisol, or Prepared Unpreserved Normal Saline.
3. Disinfecting - Allergan Cleaning and Disinfecting Solution, Flexcare, or Softmate Disinfection and Storage Solution.
4. Lubricating Solutions - Adapettes or Clerz.
Caution: Federal Law prohibits dispensing without a prescription.
CoastVision, Inc.
18368 Enterprise Lane
Huntington Beach, California 92648
Printed 5/86
{19}
Attachment B REVISED MAY 1, 198
I
# CONDITIONS OF APPROVAL
**Approved Labeling.** As soon as possible, and before commercial distribution of your device, submit two copies of an amendment to this PMA submission with copies of all approved labeling in final printed form to the Food and Drug Administration (FDA), Center for Devices and Radiological Health, PMA Document Mail Center (HFZ-401), 8757 Georgia Avenue, Silver Spring, Maryland 20910.
**Premarket Approval Application (PMA) Supplement.** Before making any change that could affect the safety or effectiveness of the device, submit a PMA supplement for review and approval by Center for Devices and Radiological Health (CDRH). Such changes may include, but are not limited to:
1. new indications for use;
2. labeling changes;
3. changes in existing manufacturing facilities, methods or quality control procedures;
4. the use of a different facility or establishment to manufacture, process, or package the device;
5. changes in sterilization procedures;
6. changes in packaging;
7. changes in the performance or design specification, circuits, parts, components, accessories, ingredients, or physical layout of the device; and
8. extension of the expiration date of the device based on data obtained under a new or revised testing protocol that has not been approved by CDRH. If the protocol has been approved, the change shall be submitted along with the supporting data in the next periodic report required in the PMA approval order. An approved protocol is one included in FDA guidelines applicable to the device or in a PMA submission for the device for which the approval order granted the expiration dating requested by you. Otherwise, you must submit and obtain CDRH approval of a PMA supplement for an expiration dating protocol.
Changes described below that enhance the safety of the device or safety in the use of the device may be placed into effect before your receipt of a written FDA order approving the PMA supplement provided that:
1. the PMA supplement and its mailing cover are plainly marked "Special PMA Supplement - Changes Being Effected";
2. the PMA supplement provides a full explanation of the basis for the changes;
3. the applicant has received acknowledgement of FDA receipt of the PMA supplement;
4. the PMA supplement specifically identifies the date that such changes are being effected; and
{20}
(5) the changes are among the following:
(a) labeling changes that add or strengthen a contraindication, warning, precaution, or adverse reaction or effect;
(b) labeling changes that add or strengthen an instruction that is intended to enhance the safe use of the device;
(c) labeling changes that delete misleading, false, or unsupported indications; or
(d) changes in the manufacturing process or quality controls that add a new specification or test method, or otherwise provide additional assurance of purity, identity, strength or reliability of the device.
You need submit only three (3) copies of a PMA supplement and include only information relevant to the proposed or effected changes in the device. The submission shall include a separate section that identifies all changes for which approval is being requested. You shall submit additional copies and additional information if requested by CDRH.
FDA may, as experience permits, issue guidelines listing specific types of changes that do not require FDA approval before implementation.
Post-Approval Reports. Continued approval of your device is contingent upon the submission of 2 copies of post-approval reports to the Food and Drug Administration, Center for Devices and Radiological Health, PMA Document Mail Center (HFZ-401), 8757 Georgia Avenue, Silver Spring, Maryland 20910 at intervals of 1 year from the date of this letter. The required contents of these reports will be described in the final order for the premarket approval procedural regulation which will be published in the FEDERAL REGISTER in the future. Until this regulation is published in final form, each periodic report shall consist of information that previously has not been submitted as part of a PMA supplement and which you have obtained since the last post-approval report or since receipt of this letter, whichever is later;
(1) a bibliography and summary of reports in the scientific literature involving the device and unpublished reports of in vitro, animal and clinical experience studies, investigations, and tests conducted by, reported to, or reasonably available to you involving the device or a related device—if, after reviewing the bibliography and summary, CDRH concludes that it needs a copy of the published and unpublished reports, CDRH will notify you that copies of such reports shall be submitted;
(2) written promotional material; and
(3) a description of changes made in the device not previously submitted in a PMA supplement.
Adverse Reaction and Device Defect Reporting. You shall submit 3 copies of a written report to the Food and Drug Administraiton, Center for Devices and Radiological Health (CDRH), 8757 Georgia Avenue, Silver Spring, Maryland 20910 within 18 days after you receive or have knowledge of information about:
(1) a mixup of the device or its labeling with another article;
{21}
(2) any significant chemical, physical, or other change or deterioration in the device, or any failure of one or more of the devices to meet the specifications established in the application;
(3) any adverse reaction, side effect, injury, toxicity, or sensitivity reaction that is attributable to the device; and
(a) has not been addressed by the device's labeling or
(b) has been addressed by the device's labeling, but is occurring with unexpected severity or frequency.
Reporting under the Medical Device Reporting (MDR) Regulation. The Medical Device Reporting (MDR) Regulation became effective on December 13, 1984, and requires that all manufacturers and importers of medical devices, including in vitro diagnostic devices, report to FDA whenever they receive or otherwise become aware of information that reasonably suggests that one of its marketed devices (1) may have caused or contributed to a death or serious injury or (2) has malfunctioned and that the device or any other device marketed by the manufacturer, or importer would be likely to cause or contribute to a death or serious injury if the malfunction were to reoccur.
The conditions of approval accompanying PMA approval orders may require that the same events subject to reporting under the MDR Regulation must also be included in periodic reports to the PMA. FDA has determined that such duplicative reporting is unnecessary. Whenever an event involving a device is subject to reporting under both the MDR Regulation and the conditions of approval for this PMA, you shall submit the appropriate reports required by the MDR Regulation and identified with the PMA reference number to the following office:
Device Monitoring Branch (HFZ-343)
Center for Devices and Radiological Health
Food and Drug Administration
8757 Georgia Avenue
Silver Spring, Maryland 20910
Telephone (301) 427-7500
Copies of the MDR Regulation and a FDA publication entitled, "An Overview of the Medical Device Reporting Regulation", are available by written request to the above address or by telephoning (301) 427-8100.
Note: All conditions of approval are subject to change upon publication of the final order for the premarket approval procedural regulation.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.