ATS OPEN PIVOT BILEAFLET HEART VALVE

P990046 · Medtronic Ats Medical, Inc. · LWQ · Oct 13, 2000 · Cardiovascular

Device Facts

Record IDP990046
Device NameATS OPEN PIVOT BILEAFLET HEART VALVE
ApplicantMedtronic Ats Medical, Inc.
Product CodeLWQ · Cardiovascular
Decision DateOct 13, 2000
DecisionAPPR
Device ClassClass 3
AttributesTherapeutic

Indications for Use

The ATS Open Pivot® Bileaflet Heart Valve is indicated for the replacement of diseased, damaged, or malfunctioning native or prosthetic aortic or mitral valves.

Device Story

Mechanical bileaflet heart valve; pyrolytic carbon orifice ring and two mirror-image leaflets; low-profile design with pivot areas contained within orifice ring. Leaflets rotate around opposing spheres; inflow/outflow stops limit motion. Orifice ring houses leaflets; sewing cuff (double velour polyester) allows in situ rotation during surgery. Mitral sizes 29-33mm include PTFE liner in cuff. Implanted by cardiac surgeons in OR. Device replaces native/prosthetic valves; restores unidirectional blood flow. Benefits include hemodynamic performance and structural durability. MRI safe up to 1.5 Tesla.

Clinical Evidence

Prospective, multi-center, international study of 965 patients (685 AVR, 280 MVR) with 1323 patient-years follow-up. Primary endpoints: safety (complications) and effectiveness (NYHA classification, echocardiographic assessment). Results: 1-year actuarial freedom from valve-related death 98.17% (AVR) and 98.31% (MVR). NYHA class I/II improvement observed post-operatively. Hemodynamic data (mean pressure gradients, effective orifice area) showed stable valve function. No structural dysfunction reported.

Technological Characteristics

Materials: pyrolytic carbon coated graphite substrate (20% tungsten impregnated for radiopacity), pyrolytic carbon orifice, double velour polyester sewing cuff, titanium stiffening/lock rings. Design: bileaflet, low-profile. Sterilization: steam (SAL 10^-6). MRI compatibility: safe at 1.5T or less. Standards: ISO 10993-1 (biocompatibility), ISO 5840 (hydrodynamic/structural performance).

Indications for Use

Indicated for patients requiring replacement of diseased, damaged, or malfunctioning native or prosthetic aortic or mitral valves. Contraindicated in patients unable to tolerate anticoagulation therapy.

Submission Summary (Full Text)

{0} p990046 # Summary of Safety and Effectiveness Data ## ATS Open Pivot® Bileaflet Heart Valve ## Table of Contents 1. GENERAL INFORMATION ... 2 2. INDICATIONS FOR USE ... 2 3. DEVICE DESCRIPTION ... 2 4. CONTRAINDICATIONS ... 3 5. WARNINGS AND PRECAUTIONS ... 3 5.1. WARNINGS ... 3 6. PRECAUTIONS ... 3 6.1. PRECAUTIONS PRIOR TO USE ... 3 6.2. PRECAUTIONS DURING USE ... 4 7. ALTERNATIVE PRACTICES AND PROCEDURES ... 4 8. MARKETING HISTORY ... 4 9. ADVERSE EVENTS ... 4 9.1. OBSERVED ADVERSE EVENTS ... 5 9.2. POTENTIAL ADVERSE EVENTS ... 6 10. SUMMARY OF NONCLINICAL STUDIES ... 7 10.1. BENCH TESTING ... 7 10.1.1. Biocompatibility Studies ... 7 10.1.2. Hydrodynamic Performance ... 8 10.1.3. Structural Performance ... 9 10.2. ANIMAL TESTING ... 10 10.3. STERILIZATION AND SHELF LIFE ... 11 10.4. MAGNETIC RESONANCE IMAGING (MRI) TESTING ... 11 11. SUMMARY OF CLINICAL STUDIES ... 11 11.1. DESCRIPTION OF PATIENTS AND ANALYSIS FOR GENDER BIAS ... 14 12. CONCLUSIONS DRAWN FROM THE STUDIES ... 15 13. PANEL RECOMMENDATIONS ... 15 14. FDA DECISION ... 15 15. APPROVAL SPECIFICATIONS ... 15 10 {1} SUMMARY of SAFETY and EFFECTIVENESS DATA ATS Open Pivot® Bileaflet Heart Valve 1. GENERAL INFORMATION Device Generic Name: Replacement Heart Valve Device Trade Name: ATS Open Pivot® Bileaflet Heart Valve Aortic Models 500FA and 501DA, and Mitral Models 500DM and 501DM Applicant’s Name and Address: ATS Medical, Inc. Suite 105 3905 Annapolis Lane Minneapolis, MN 55447 PMA Application Number: P990046 Date of Notice of Approval to the Applicant: OCT 13 2000 2. INDICATIONS FOR USE The ATS Open Pivot® Bileaflet Heart Valve is indicated for the replacement of diseased, damaged, or malfunctioning native or prosthetic aortic or mitral valves. 3. DEVICE DESCRIPTION The ATS Open Pivot® Bileaflet Heart Valve is a low profile bileaflet valve consisting of pyrolytic carbon orifice ring and leaflets. The prosthesis consists of an orifice housing two mirror image leaflets. The low profile of the prosthesis results from the bileaflet design where the pivot areas are located entirely within the orifice ring, which minimizes the overall height of the valve. Pivot guides located on the inner circumference of the orifice ring control the range of leaflet motion. The pivot geometry consists of arc-shaped notches at either end of each leaflet and spherical protrusions at four places on the orifice. Each leaflet rotates around two opposing spheres. The inflow and outflow stops are adjacent to each sphere on the orifice, which limit the rotation of the leaflets. There are no recesses or cavities in the pivot area. In the closed position, the plane of each leaflet forms a nominal angle of $25^{\circ}$ relative to the plane of the orifice ring. In the fully open position, the plane of each leaflet forms a nominal angle of $85^{\circ}$ relative to the plane of the orifice ring. The leaflets consist of pyrolytic carbon coated over a graphite substrate. The graphite substrate is impregnated with $20\%$ tungsten for radiopacity. The orifice consists entirely of pyrolytic carbon. - 2 - {2} The valve sewing cuff is constructed of double velour polyester fabric and is mounted on the orifice using a titanium stiffening ring and secured with two titanium lock rings and a lock wire. This method of sewing cuff attachment to the orifice allows for rotation of the sewing cuff *in situ*, during surgical placement. The sewing cuff of mitral sizes 29, 31, and 33 mm contains a polytetrafluoroethylene (PTFE) liner inside the double velour polyester fabric. The ATS Open Pivot® Bileaflet Heart Valve is available in the aortic and mitral configurations in two sewing cuff styles, the Standard and Advanced Performance (AP) styles. The AP model, with a reduced cuff, is a supra-annular configuration of the Standard model. The ATS Open Pivot® Bileaflet Heart Valve Standard model is available in sizes 21 through 29 mm in the aortic position (Model 500FA), and sizes 29 through 33 mm in the mitral position (Model 500DM). The ATS Open Pivot® Bileaflet Heart Valve AP model is available in sizes 18 through 26 mm in the aortic position (Model 501DA), and sizes 26 and 28 mm in the mitral position (Model 501DM). ## 4. CONTRAINDICATIONS The ATS Open Pivot® Bileaflet Heart Valve is contraindicated in patients unable to tolerate anticoagulation therapy. ## 5. WARNINGS AND PRECAUTIONS ### 5.1. Warnings - FOR SINGLE USE ONLY. - Avoid damaging the prosthesis. Only handle the prosthesis with the accessories provided by ATS Medical. Touching of the valve with gloved fingers or any surgical instrument may cause damage to the valve surface not seen with the unaided eye that may lead to accelerated valve structural deterioration or leaflet escape, or serve as a nidus for thrombus formation. - Do NOT pass a catheter through the prosthesis as this may cause valvular insufficiency, disc dislodgment, or catheter entrapment. - Do NOT apply force to the leaflets, attempt to change the position of the leaflets, or remove a leaflet. ## 6. PRECAUTIONS ### 6.1. Precautions Prior to Use - Do NOT use the ATS Open Pivot® Bileaflet Heart Valve if the prosthesis has been dropped, damaged, or mishandled in any way. Should the valve be damaged during implantation or removal from the package, do not use for implantation. - 3 - 12 {3} - Do NOT use the ATS Open Pivot® Bileaflet Heart Valve if the tamper evident seal is broken, or if the expiration date has elapsed. - Do NOT resterilize any ATS Open Pivot® Bileaflet Heart Valve. ## 6.2. Precautions During Use - Use only the ATS Medical Valve Sizer to select the proper valve size as other sizers may result in improper valve selection. - When seating the valve, ensure that no suture material or anatomic structures interfere with leaflet motion. The valve’s rotation capability may be helpful in avoiding abnormal residual pathology that could interfere with leaflet motion. ## 7. ALTERNATIVE PRACTICES AND PROCEDURES Alternative forms of treatment include medical therapy with drugs or surgical treatments such as annuloplasty or valvuloplasty with or without the use of implantable materials (i.e., annuloplasty rings, sutures). When the patient requires replacement of his/her native or previously placed prosthetic valve, the option of choosing a mechanical or biological valve exists. The choice of replacement valve depends upon factors that include the patient’s age, preoperative conditions, cardiac anatomy, and ability to tolerate long term anticoagulation therapy. ## 8. MARKETING HISTORY The ATS Open Pivot® Bileaflet Heart Valve is distributed in Argentina, Australia, Austria, Belgium, Brazil, Canada, China, Croatia, France, Germany, Greece, India, Israel, Italy, Japan, South Korea, Lebanon, Mauritius, Mexico, Morocco, Netherlands, Norway, Portugal, Russia, South Africa, Spain, Switzerland, Tunisia, Turkey, United Kingdom, and Yugoslavia. The ATS Open Pivot® Bileaflet Heart Valve has not been withdrawn from the market in any country for any reason. ## 9. ADVERSE EVENTS A multi-center, non-randomized, prospective, international clinical study was conducted of patients implanted with the ATS Open Pivot® Bileaflet Heart Valve. In this study, 685 patients had isolated aortic valve replacement (AVR), 280 patients had isolated mitral valve replacement (MVR), and 35 patients had double valve replacement (DVR) where both the aortic and the mitral valves were replaced with an ATS Open Pivot® Bileaflet Heart Valve. The study was conducted between 1994 and 1999. Patients were evaluated pre-operatively, intra-operatively, and post-operatively at discharge, at 3-6 months, at 1 year, and annually thereafter. Adverse events were captured throughout the post-operative period. The adverse event rates were based on 965 patients at 20 centers. The cumulative follow-up was 1323 patient-years with a mean follow-up of 1.4 years (range 0 to 5 years). - 4 - 13 {4} A total of 56 deaths occurred during the study. Twenty (20) of these deaths were characterized as valve-related. The causes of valve-related deaths were endocarditis (2 patients), paravalvular leak (1 patient), thromboembolism (3 patients), anticoagulant-related hemorrhage (10 patients), and unknown (4 patients). ## 9.1. Observed Adverse Events Table 1 shows the observed adverse events for early events (≤ 30 days), the linearized rates for late events (>30 days post-operatively), and the actuarial adverse event rates at one and five years post-operatively. | | Early Events % of pts. (N) | Late Events %/pt-yr. (N) | Actuarial Freedom by Kaplan-Meier | | | --- | --- | --- | --- | --- | | | | | 1 Year [95% CI] | 5 years [95% CI] | | Aortic Valve Replacement, All patients implanted: N= 685, Cumulative Follow-up= 866.4 patient years | | | | | | Deaths (all causes) | 2.04% (14) | 2.77% ± 1.06% (24) | .9735 [± .01] | .9031 [± .05] | | Death (valve-related/unexplained) | 0.58% (4) | 1.15% ± 0.74% (10) | .9817 [± .01] | .9539 [± .03] | | Anticoagulant-Related Hemorrhage (All) | 4.67% (32) | 1.96% ± 0.91% (17) | .9781 [± .01] | .9340 [± .04] | | Anticoagulant-Related Hemorrhage (Major) | 3.21% (22) | 1.27% ± 0.76% (11) | .9878 [± .01] | .9473 [± .04] | | Thromboembolism (All) | 1.75% (12) | 2.08% ± 0.93% (18) | .9733 [± .01] | .9283 [± .05] | | Permanent Neurological Events | 0.88% (6) | 0.69% ± 0.60% (6) | .9920 [± .01] | .9706 [± .04] | | Transient Neurological Events | 0.88% (6) | 1.39% ± 0.78% (12) | .9812 [± .01] | .9564 [± .03] | | Valve Thrombosis | 0.00% (0) | 0.00% ± 0.00% (0) | 1.000 [± .00] | 1.000 [± .00] | | Perivalvular Leak (All) | 0.15% (1) | 0.46% ± 0.52% (4) | .9966 [± .00] | .9898 [± .01] | | Perivalvular Leak (Major) | 0.15% (1) | 0.12% ± 0.33% (1) | .9983 [± .00] | .9983 [± .00] | | Endocarditis | 0.00% (0) | 0.35% ± 0.46% (3) | .9960 [± .01] | .9908 [± .01] | | Hemolysis | 0.00% (0) | 0.00% ± 0.00% (0) | 1.000 [± .00] | 1.000 [± .00] | | Structural Dysfunction | 0.00% (0) | 0.00% ± 0.00% (0) | 1.000 [± .00] | 1.000 [± .00] | | Nonstructural Dysfunction | 0.00% (0) | 0.00% ± 0.00% (0) | 1.000 [± .00] | 1.000 [± .00] | | Reoperation | 0.15% (1) | 0.35% ± 0.46% (3) | .9961 [± .01] | .9914 [± .01] | | Explant | 0.00% (0) | 0.23% ± 0.41% (2) | .9978 [± .00] | .9931 [± .01] | Notes: 1. Cumulative probability of freedom from event estimate at the end of the interval (Pc) is based on the Kaplan-Meier method. 2. The 95% confidence interval bound for the cumulative freedom rate at the end of the interval = 1.96 X SE, where SE is the standard error estimate of the cumulative probability of freedom from study heart valve related or unexplained event estimate calculated using Greenwood's formula. 3. The actuarial hazard rate estimates are calculated at the midpoint of each interval. 4. The 95% confidence interval bound for the hazard rate at the midpoint of each interval = 1.96 X SE, where SE is the standard error estimate of the hazard rate estimate at the midpoint of the interval. {5} Table 1: Observed Adverse Events for AVR and MVR - continued | | Early Events % of pts. (N) | Late Events %/pt-yr. (N) | Actuarial Freedom by Kaplan-Meier 1 Year [95% CI] | 5 years [95% CI] | | --- | --- | --- | --- | --- | | Mitral Valve Replacement, All patients implanted: N=280, Cumulative Follow-up= 374.7 patient years | | | | | | Deaths (all causes) | 1.79% (5) | 3.47% ± 1.88% (13) | .9814 [± .02] | .8099 [± .11] | | Death (valve-related/unexplained) | 0.71% (2) | 1.07% ± 1.19% (4) | .9831 [± .02] | .9342 [± .07] | | Anticoagulant-Related Hemorrhage (All) | 3.21% (9) | 0.53% ± 0.95% (2) | .9958 [± .01] | .9673 [± .06] | | Anticoagulant-Related Hemorrhage (Major) | 3.21% (9) | 0.53% ± 0.95% (2) | .9958 [± .01] | .9673 [± .06] | | Thromboembolism (All) | 3.21% (9) | 4.00% ± 2.00% (15) | .9534 [± .03] | .8589 [± .09] | | Permanent Neurological Events | 1.79% (5) | 0.80% ± 1.08% (3) | .9910 [± .01] | .9807 [± .02] | | Transient Neurological Events | 1.43% (4) | 3.20% ± 1.82% (12) | .9621 [± .03] | .8758 [± .09] | | Valve Thrombosis | 0.00% (0) | 0.53% ± 0.95% (2) | .9947 [± .01] | .9866 [± .02] | | Perivalvular Leak (All) | 0.71% (2) | 1.07% ± 1.19% (4) | .9819 [± .02] | .9819 [± .02] | | Perivalvular Leak (Major) | 0.36% (1) | 0.53% ± 0.95% (2) | .9915 [± .01] | .9915 [± .01] | | Endocarditis | 0.36% (1) | 0.53% ± 0.95% (2) | .9957 [± .01] | .9861 [± .02] | | Hemolysis | 0.00% (0) | 0.53% ± 0.95% (2) | .9952 [± .01] | .9814 [± .03] | | Structural Dysfunction | 0.00% (0) | 0.00% ± 0.00% (0) | 1.000 [± .00] | 1.000 [± .00] | | Nonstructural Dysfunction | 0.00% (0) | 0.00% ± 0.00% (0) | 1.000 [± .00] | 1.000 [± .00] | | Reoperation | 0.71% (2) | 1.07% ± 1.19% (4) | .9874 [± .01] | .9697 [± .03] | | Explant | 0.36% (1) | 0.53% ± 0.95% (2) | .9959 [± .01] | .9783 [± .03] | Notes: 1. Cumulative probability of freedom from event estimate at the end of the interval (Pc) is based on the Kaplan-Meier method. 2. The 95% confidence interval bound for the cumulative freedom rate at the end of the interval = 1.96 X SE, where SE is the standard error estimate of the cumulative probability of freedom from study heart valve related or unexplained event estimate calculated using Greenwood's formula. 3. The actuarial hazard rate estimates are calculated at the midpoint of each interval. 4. The 95% confidence interval bound for the hazard rate at the midpoint of each interval = 1.96 X SE, where SE is the standard error estimate of the hazard rate estimate at the midpoint of the interval. ## 9.2. Potential Adverse Events Adverse events potentially associated with the use of prosthetic heart valves (in alphabetical order) include: - cardiac arrhythmias - death - leaflet entrapment (impingement) - endocarditis - hemolysis - hemorrhage, anticoagulation-related - leak, transvalvular or perivalvular - prosthesis thrombosis - structural deterioration - valve thromboembolism 15 {6} # 10. SUMMARY OF NONCLINICAL STUDIES ## 10.1. Bench testing In vitro studies were performed for the ATS Open Pivot® Bileaflet Heart Valve as recommended in the FDA’s Draft Replacement Heart Valve Guidance (1994). ### 10.1.1. Biocompatibility Studies Selected short-term tests recommended in the Tripartite Biocompatibility Guidance for Medical Devices document have been conducted on material used in the ATS Open Pivot® Bileaflet Heart Valve and accessories. Biocompatibility tests were performed according to the requirements of ISO 10993-1, with the exception of carcinogenicity and hemocompatibility testing. Carcinogenicity testing was determined to be unnecessary since the test devices demonstrated no mutagenic potential at the levels at or above those intended for clinical use. Hemocompatibility evaluation was determined in animal implant studies. See Section 10.2. All studies were performed in accordance with FDA Good Laboratory Practices Regulation (21 CFR Part 58). A matrix of tests performed and the corresponding results is provided in Table 2. Table 2: Biocompatibility Studies Testing Results | Test Performed | Test Objective | Samples: Control | Samples: Test article | Results | | --- | --- | --- | --- | --- | | In Vitro Inhibition of Cell Growth | Assess the effect of an extract of the material on the normal growth of cells in culture. | Negative control: Extract only Positive control: Known toxic material | ATS Open Pivot Bileaflet Heart Valve | Non-inhibitory to cell growth | | USP Systemic Toxicity | Assess the systemic effect of a material extract in mice. | Negative control: Normal Saline and cottonseed oil | ATS Open Pivot Bileaflet Heart Valve | All mice normal. Non-toxic | | Hemolysis | Assess the hemocompatibility of an extract of the device. | Negative control: Normal Saline Positive control: Known hemolytic material. | ATS Open Pivot Bileaflet Heart Valve | Mean hemolysis = 0% | | Genotoxicity-Gene Mutation | Assess the mutagenicity of carbon components with extracts of Saline and DMSO. | Negative controls: Normal Saline and DMSO. | Carbon Components | Extracts were non-mutagenic. | | Rabbit Implantation 7 days | Assess implantation effects. | 3 USP control strips and 3 sections of test article per rabbit | Carbon Components | Macroscopic reaction was not significant as compared to the negative control implant material. | - 7 - 16 {7} Table 2: Biocompatibility Studies Testing Results - continued | Rabbit Implantation 90 days | Assess implantation effects. | 3 USP control strips and 3 sections of test article per rabbit | Carbon Components | Macroscopic reaction was not significant as compared to the control. | | --- | --- | --- | --- | --- | | Cytotoxicity | Assess the effect of an extract on the normal growth of cells in culture. | Negative Control: MEM, Positive Control Known material | Accessories | Non-cytotoxic | | Sensitization-Guinea Pig Maximization | Assess the potential of a material to produce sensitization when material saline and cottonseed oil extracts are exposed to the test animal. | Negative control: Normal Saline and cottonseed oil | Accessories | No evidence of causing delayed dermal contact sensitization in the guinea pig. | | Acute Systemic Toxicity - USP 22 Mouse Injection | Assess the systemic effect of material extract in mice. | Negative control: Normal Saline and cottonseed oil. | Accessories | All mice normal. Non-toxic | | Hemocompatibility | Assess the effect of an extract on blood. | Normal Saline extract | Accessories | Non-hemolytic | | Pyrogenicity USP 22 | Assess the pyrogenicity of an extract | Non-pyrogenic Saline | Accessories | Non-pyrogenic | ## 10.1.2. Hydrodynamic Performance *In vitro* hydrodynamic performance studies of the ATS Open Pivot® Bileaflet Heart Valve were completed according to test recommendations outlined in FDA’s *Draft Replacement Heart Valve Guidance* (1994) and ISO 5840 *Cardiovascular Implants-Cardiac Valve Prostheses*. Commercially available mechanical valves were used as controls. All test and control valves were final production samples. A matrix of the hydrodynamic tests is provided in Table 3. Table 3: Hydrodynamic Performance Testing and Results | Test performed | Sample Size: Control | Sample Size: ATS Open Pivot® Bileaflet Heart Valve | Results | | --- | --- | --- | --- | | Steady Forward Flow Pressure Drop | 3 valves of each size (19, 23, and 31 mm) | 3 valves of each size (16/19, 18/2, 20/23, 22/25, 24/27, 26/29, and 28/31/33 mm) | The test results revealed that steady flow pressure drop is approximately the same in both the aortic and mitral positions and consistent with results reported in the literature. | | Steady Backflow leakage Testing | 1 each 23 mm | 3 valves of each size (16/19, 18/2, 20/23, 22/25, 24/27, 26/29, and 28/31/33 mm) | The leakage rate of the test valves was less than the control valve. | | Pulsatile Flow Pressure Drop | 3 each 19, 23, and 31 mm | 3 valves of each size (16/19, 18/2, 20/23, 22/25, 24/27, 26/29, and 28/31/33 mm) | Results were comparable to the control valve. | - 8 - {8} Table 3: Hydrodynamic Performance Testing and Results - continued | Pulsatile Flow Regurgitation | 3 each 19, 23 and 31 mm | 3 valves of each size (16/19, 18/2, 20/23, 22/25, 24/27, 26/29, and 28/31/33 mm) | Results were comparable to the control valve. | | --- | --- | --- | --- | | Laser Doppler Anemometry | 1 each size 27 mm | 1 each size 27 mm | Similar flow patterns were observed with the test and control valves. The highest turbulent shear stress for the ATS valve was measured to be 25.2 Pa with the corresponding values for the control 45 Pa. | | Flow Visualization | N/A | 1 each size 19 and 27 mm | There were no adverse flow characteristics observed with test valves. | | Verification of the Bernoulli Relationship | N/A | 1 each size 19, 23, 27, and 31 mm | Transvalvular pressure drops determined by Doppler ultrasound showed good correlation. | ## 10.1.3. Structural Performance *In vitro* structural performance studies of the ATS Open Pivot® Bileaflet Heart Valve were performed in accordance with testing recommendations outlined in FDA’s *Draft Replacement Heart Valve Guidance (1994)* and ISO 5840 *Cardiovascular Implants- Cardiac Valve Prostheses*. Commercially available valves were used as controls in those studies that require concurrent testing. All control and reference valves were final production samples. A matrix of the structural performance tests performed is provided in Table 4. Table 4: Structural Performance Testing and Results | Test performed | Sample Size: Control | Sample Size: ATS Open Pivot® Bileaflet Heart Valve | Results | | --- | --- | --- | --- | | Material characterization testing - carbon coated graphite | N/A | Carbon-coated graphite specimen | Crack initiation fracture toughness (K_{i}) values between 1.00 to 2.59 MPa/m; worst case Paris law coefficients for fatigue crack growth rates were da/dN=56.5 (delta K)^{63} in units of m/cycle; worst case crack growth rate of da/dt=2.7 x 10^{-9}K^{75} | | Material characterization testing - pyrolytic carbon | N/A | Pyrolytic carbon specimen | Crack initiation fracture toughness (K_{i}) value of 1.0MPa/m; worst case fatigue crack growth rates were da/dN=4141 (delta K)^{88,9} in units of m/cycle; worst case crack growth rate of da/dt=21.36 x 10^{-7}K^{13,8} | - 9 - 18 {9} Table 4: Structural Performance Testing and Results - continued | Accelerated Wear Testing | 2 each 31 mm | 10 each 31 mm | The magnitude of leaflet depth wear varied on the tester used. For the two sets of ATS valves, the average inflow leaflet wear depths were 34.8 microns and 59.1 microns. For the corresponding control valves, average inflow wear depths were 32 and 47.2 microns. | | --- | --- | --- | --- | | Stress Analysis | N/A | 19 and 31 mm sizes | Worse case stresses were used in fracture mechanics formulae to calculate the critical crack size at fracture. Paris law fatigue crack growth computations were used to calculate the initial length of cracks that would grow to critical size in 100 years or less at 72 bpm. | | Fatigue Lifetime Determination | N/A | 19 and 31 mm sizes | A proof test was designed by fracture mechanics methods to ensure valves placed in service at 200 mmHg systolic pressure will survive at least 100 years at 72 bpm | | Dynamic Failure Mode | 1 each 31 mm | 6 each 31 mm | The test suggested the ultimate failure mode is fracturing of either the orifice or the leaflet at the point of contact between the major radius of the leaflet and orifice. Failures did not occur until loading conditions were greater than 780 mmHg. No failures occurred in the pivot regions. | | Sewing Ring Integrity Testing | N/A | 1 each standard and AP valves | Cuff push-off force was >100 lbs for all sizes except size 16. Size 16 valve tested in excess of 50 lbs before fixturing difficulties ended the test. | | Leaflet Escape Testing | 1 each 19, 21, 23, 29, and 31 | 3 each size standard valves and 3 each size AP valves | Minimum load to leaflet escape is 29.6 lbs and comparable to control valves. | | Leaflet Impingement Testing | Literature | Minimum 3 each size Standard and AP valves | Minimum load to impingement 5 lbs compared to 3.5 lbs for literature control valve. | ## 10.2. Animal testing Animal studies were conducted in compliance with Good Laboratory Practices. Ten (10) juvenile sheep underwent mitral valve replacement with either a size 23 mm ATS Open Pivot® Bileaflet Heart Valve (n=8) or a 23 mm commercially available valve (n=2). Parameters evaluated included ease of surgical implantation, assessment of valve-related deaths, hemodynamic performance, and pathological impact following 150 days implantation. There were no surgical complications and no technical failures. All animals underwent angiography and sacrifice after 150 days post-operatively or greater. Blood abnormalities were not detected. Angiography analysis demonstrated normally functioning valves with minimal regurgitation and small transvalvular pressure gradients. Pathological evaluation of the 10 surviving animals revealed all implant devices functioning fully at autopsy. Paravalvular leaks detected in animals implanted with the ATS Open Pivot® Bileaflet Heart Valve or the control valves were attributed to surgical technique. - 10 - 19 {10} # 10.3. Sterilization and Shelf life The ATS Open Pivot® Bileaflet Heart Valve is terminally sterilized using steam to a sterility assurance level (SAL) of at least $10^{-6}$. Product is quarantined until sterility is verified. Revalidation of this process is performed annually. Resterilization and cleaning of accessory components (sizers) was validated using an artificial soil inoculated with *B. stearothermophilus*. The cleaning method achieved a > 4 log reduction of spores. Terminal sterilization of the accessories using steam yielded a SAL of at least $10^{-6}$. Both packaging and product integrity studies were conducted to ensure that the shelf life for the package and product is maintained for a minimum of three years. These studies consisted of real-time and accelerated aging. # 10.4. Magnetic Resonance Imaging (MRI) Testing The ATS Open Pivot® Bileaflet Heart Valve has been shown to be safe when tested using MR systems operating with shielded static magnetic field strengths of 1.5 Tesla or less. However, the testing may cause significant MRI image artifacts or distortion. This phenomenon produced no harmful effects to the patient. # 11. SUMMARY OF CLINICAL STUDIES The safety endpoints captured in the prospective studies were complications. The safety results are provided in Table 1 above. Effectiveness endpoints were New York Heart Association (NYHA) functional classification and echocardiographic assessments. Preoperative and operative patient demographics are presented below, followed by the effectiveness results. Table 5: Patient Characteristics All patients implanted, N = 965, 1323 patient-years | Description of Patients | Aortic Valve N = 685 (70.98%) | Mitral Valve N = 280 (29.02%) | | --- | --- | --- | | Age at implant in years | | | | 0-9 | 5 (0.7%) | 6 (2.1%) | | 10-19 | 4 (0.6%) | 0 (0.0%) | | 20-29 | 11 (1.6%) | 2 (0.7%) | | 30-39 | 33 (4.8%) | 14 (5.0%) | | 40-49 | 72 (10.5%) | 31 (11.1%) | | 50-59 | 142 (20.7%) | 52 (18.6%) | | 60-69 | 238 (34.7%) | 107 (38.2%) | | 70-79 | 155 (22.6%) | 62 (22.1%) | | 80 & over | 25 (3.6%) | 6 (2.1%) | | Gender | | | | Male | 460 (67.2%) | 120 (42.9%) | | Female | 225 (32.8%) | 160 (57.1%) | | Etiology of valve disease | | | | Stenosis | 541 (79.0%) | 118 (42.1%) | | Insufficiency | 330 (48.2) | 210 (75.0%) | | Mixed | 205 (29.9%) | 63 (22.5%) | | Other | 12 (1.8%) | 12 (4.3%) | {11} Figure 1: Number of Patients by Implant Location over Time All patients implanted, N=965 ![img-0.jpeg](img-0.jpeg) | Year | 0 | 1 | 2 | 3 | 4 | 5 | | --- | --- | --- | --- | --- | --- | --- | | Total Patients | 965 | 624 | 331 | 135 | 92 | 2 | | Aortic Patients | 685 | 436 | 231 | 93 | 58 | 2 | | Mitral Patients | 280 | 188 | 100 | 42 | 34 | 0 | Table 6: Number of Patients Implanted and Maximum Number of Patients with Hemodynamic Data at > 6 Months Follow-Up By implant location and valve size, n¹ = 493, N=965 | Implant Location | Valve size (mm) | | | | | | | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | | | 16/19 | 18/21 | 20/23 | 22/25 | 24/27 | 26/29 | 28/31/33 | Total | | Aortic (N/n) | 23/9 | 100/61 | 202/108 | 206/113 | 111/70 | 39/32 | 4/4 | 685/397 | | Mitral (N/n) | 0/0 | 3/1 | 5/4 | 3/2 | 23/8 | 71/39 | 175/106 | 280/160 | | Total (N/n) | 23/9 | 103/62 | 207/112 | 209/115 | 134/78 | 110/71 | 179/110 | 965/557 | Note: 1. n = number of patients with hemodynamic data; N = number of patients implanted Table 7: Number of Patient-Years by Implant Location and Valve Size By implant location and valve size, all patients implanted, N = 965 | Implant Location | Valve size (mm) | | | | | | | | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | | 16/19 | 18/21 | 20/23 | 22/25 | 24/27 | 26/29 | 28/31 | 33 | Total | | Aortic | 15.7 | 159.3 | 275.0 | 255.6 | 150.1 | 61.1 | 7.4 | 0 | 924.2 | | Mitral | 0 | 2.2 | 4.8 | 2.4 | 17.6 | 99.8 | 97.7 | 174.0 | 398.5 | | Total | 15.7 | 161.5 | 279.8 | 258.0 | 167.7 | 160.9 | 105.1 | 174.0 | 1322.7 | {12} Table 8: Effectiveness Outcomes, Functional New York Heart (NYHA) Classification | NYHA Class | Pre-op | | 1 Year (11-14 Months) | | 2 Year (23-26 Months) | | 3 Year (35-38 Months) | | 4 Year (47-50 Months) | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | | n/N¹ | % | n/N | % | n/N | % | n/N | % | n/N | % | | Aortic Valve Replacement, N = 685 | | | | | | | | | | | | I | 4/685 | 0.6% | 399/431 | 92.5% | 205/231 | 88.7% | 85/94 | 90.4% | 53/58 | 91.4% | | II | 267/685 | 38.9% | 25/431 | 5.8% | 20/231 | 8.7% | 7/94 | 7.4% | 5/58 | 8.6% | | III | 340/685 | 49.6% | 1/431 | 0.2% | 1/231 | 0.4% | 0/94 | 0% | 0/58 | 0% | | IV | 68/685 | 9.9% | 1/431 | 0.2% | 1/231 | 0.4% | 0/94 | 0% | 0/58 | 0% | | Missing | 6/685 | 0.9% | 5/431 | 1.2% | 4/231 | 1.7% | 2/94 | 2.1% | 0/58 | 0% | | Mitral Valve Replacement, N = 280 | | | | | | | | | | | | I | 0/280 | 0% | 158/182 | 86.8% | 78/100 | 78.0% | 33/42 | 78.5% | 28/34 | 82.4% | | II | 74/280 | 26.4% | 19/182 | 10.4% | 19/100 | 19.0% | 8/42 | 19.0% | 5/34 | 14.7% | | III | 155/280 | 55.4% | 3/182 | 1.6% | 2/100 | 2.0% | 1/42 | 2.4% | 1/34 | 2.9% | | IV | 45/280 | 16.1% | 0/182 | 0% | 0/100 | 0% | 0/42 | 0% | 0/34 | 0% | | Missing | 6/280 | 2.1% | 2/182 | 1.1% | 1/100 | 1.0% | 0/42 | 0% | 0/34 | 0% | Note: 1. N = all values reported; n = number in subgroup Table 9: Effectiveness Outcomes – Hemodynamics, Valvular Regurgitation | Valvular Regurgitation | Early¹ % patients, n/N³ | Late² % patients, n/N³ | | --- | --- | --- | | Aortic Valve Replacement, N = 685 | | | | No Regurgitation | 41.4%, 244/590 | 34.8%, 119/342 | | Trivial Regurgitation | 57.6%, 340/590 | 64.6%, 221/342 | | Mild Regurgitation | 1.0%, 6/590 | 0.6%, 2/342 | | Moderate Regurgitation | 0%, 0/590 | 0%, 0/342 | | Severe Regurgitation | 0%, 0/590 | 0%, 0/342 | | Mitral Valve Replacement, N = 280 | | | | No Regurgitation | 60.7%, 128/211 | 60.5%, 78/129 | | Trivial Regurgitation | 37.4%, 79/211 | 38.0%, 49/129 | | Mild Regurgitation | 1.9%, 4/211 | 1.6%, 2/129 | | Moderate Regurgitation | 0%, 0/211 | 0%, 0/129 | | Severe Regurgitation | 0%, 0/211 | 0%, 0/129 | Notes: 1. Early post-operative evaluation conducted at 30-days post-implantation or hospital discharge. 2. Late post-operative evaluation = 11-14 months post-implantation. 3. N = all values reported; n = number in subgroup - 13 - 22 {13} Table 10: Effectiveness Outcomes – Hemodynamics, Mean Pressure Gradient and Effective Orifice Area | Endpoint | Early^{1} n/N^{3}, mean ± SD (min, max) | Late^{1} n/N, mean ± SD (min, max) | | --- | --- | --- | | Aortic Valve Replacement, N = 685 | | | | Mean Gradient (mm Hg) | | | | 16/19mm | 15/23, 25.8 ± 5.1 (10.5, 49.0) | 9/23, 20.2 ± 2.8 (15.0, 26.6) | | 18/21mm | 87/100, 18.7 ± 1.7 (2.4, 42.0) | 61/100, 18.0 ± 1.6 (7.0, 36.0) | | 20/23mm | 181/202, 14.3 ± 0.8 (2.6, 37.0) | 107/202, 13.1 ± 0.8 (5.1, 30.1) | | 22/25mm | 187/206, 12.8 ± 0.8 (2.6, 29.5) | 112/206, 11.1 ± 0.8 (3.2, 26.0) | | 24/27mm | 102/111, 10.0 ± 0.7 (1.8, 22.0) | 70/111, 8.0 ± 0.8 (1.3, 16.7) | | 26/29mm | 38/39, 9.2 ± 1.1 (3.5, 18.0) | 32/39, 7.8 ± 1.1 (2.0, 13.0) | | 28/31mm | 3/4, 3.0 ± 0.7 (2.6, 3.8) | 4/4, 5.1 ± 3.3 (1.4, 9.3) | | Effective Orifice Area (cm^{2}) | | | | 16/19mm | 11/23, 1.1 ± 0.2 (0.7, 1.8) | 8/23, 1.2 ± 0.3 (0.8, 1.9) | | 18/21mm | 81/100, 1.5 ± 0.1 (0.8, 3.7) | 60/100, 1.5 ± 0.1 (0.7, 3.4) | | 20/23mm | 165/202, 1.7 ± 0.1 (0.8, 6.6) | 102/202, 1.7 ± 0.1 (0.9, 3.7) | | 22/25mm | 173/206, 2.0 ± 0.1 (1.1, 4.0) | 103/206, 2.1 ± 0.1 (1.0, 4.9) | | 24/27mm | 97/111, 2.4 ± 0.2 (1.1, 4.8) | 65/111, 2.5 ± 0.2 (1.5, 4.9) | | 26/29mm | 34/39, 3.0 ± 0.3 (1.4, 4.7) | 28/39, 3.1 ± 0.4 (1.4, 5.4) | | 28/31mm | 3/4, 2.8 ± 0.8 (2.0, 3.4) | 3/4, 3.1 ± 1.6 (1.6, 4.5) | | Mitral Valve Replacement, N = 280 | | | | Mean Gradient (mmHg) | | | | 16/19mm | 0 | 0 | | 18/21mm | 0/3 | 1/3, 6.0 (6.0, 6.0) | | 20/23mm | 3/5, 5.3 ± 3.6 (3, 9) | 4/5, 4.6 ± 0.9 (4.0, 6.0) | | 22/25mm | 1/3, 10.0 (10, 10) | 2/3, 5.4 ± 4.7 (3.0, 7.8) | | 24/27mm | 15/23, 4.4 ± 0.9 (2.3, 7.9) | 8/23, 4.5 ± 0.9 (2.4, 6.2) | | 26/29mm | 66/71, 3.7 ± 0.4 (1.5, 10.0) | 39/71, 3.7 ± 0.7 (1.3, 9.9) | | 28/31/33mm | 154/175, 3.5 ± 0.3 (0.7, 9.0) | 106/175, 3.1 ± 0.2 (0.3, 7.3) | | Effective Orifice Area (cm^{2}) | | | | 16/19mm | 0 | 0 | | 18/21mm | 0/3 | 1/3^{4} | | 20/23mm | 3/5, 2.9 ± 0.6 (2.3, 3.3) | 3/5, 1.6 ± 0.3 (1.3, 1.8) | | 22/25mm | 1/3, 1.6 (1.6, 1.6) | 2/3, 1.8 ± 0.5 (1.5, 2.0) | | 24/27mm | 15/23, 3.3 ± 0.5 (1.8, 5.2) | 8/23, 2.9 ± 0.9 (1.6, 5.7) | | 26/29mm | 63/71, 3.3 ± 0.2 (1.6, 5.0) | 38/71, 2.8 ± 0.3 (1.0, 4.6) | | 28/31/33mm | 140/175, 3.0 ± 0.2 (1.3, 7.6) | 95/175, 2.9 ± 0.2 (1.5, 7.4) | Notes: 1. Early post-operative evaluation conducted at 30-days post implantation or hospital discharge. 2. Late post-operative evaluation = 11-14 months post implantation 3. N= all values reported; n = number in subgroup 4. Echo confirmed valve function. ## 11.1. Description of Patients and Analysis for Gender Bias A gender bias was not found in the ATS Open Pivot® Bileaflet Heart Valve clinical study. Of the 965 patients, 60% (580/965) were male and 40% (385/965) were female. This gender distribution is consistent with incidence of patients presenting for valve replacement. The results of analysis for morbidity/mortality by gender were not significant following valve replacement. {14} - 15 - 24 ## 12. CONCLUSIONS DRAWN FROM THE STUDIES The results from pre-clinical laboratory studies performed on the ATS Open Pivot® Bileaflet Heart Valve for biocompatibility testing, hydrodynamic performance testing, and structural performance testing demonstrate that this device is suitable for long-term implant. The animal studies show that the ATS Open Pivot® Bileaflet Heart Valve is safe for valve replacement. The clinical studies submitted in the PMA provide scientific evidence that the ATS Open Pivot® Bileaflet Heart Valve is safe and effective for the replacement of native or prosthetic aortic or mitral valves. ## 13. PANEL RECOMMENDATIONS In accordance with the provisions of section 510(c)(2) of the act as amended by the Safe Medical Devices Act of 1990, this PMA was not referred to the Circulatory Systems Device Panel, a FDA advisory committee, for review and recommendation because the information in the PMA substantially duplicates information previously reviewed by this panel. ## 14. FDA DECISION FDA issued an approval order on OCT 13 2000. The applicant’s manufacturing and control facilities were inspected on September 13, 1999, and the facilities were found to be in compliance with the Good Manufacturing Practices (GMP) regulation. ## 15. APPROVAL SPECIFICATIONS Direction for Use: See Final Draft Labeling (Instructions for Use). Hazards to Health from Use of the Device: See Indications, Contraindications, Warnings, Precautions, and Adverse Events in the Final Draft Labeling (Instructions for Use). Post-approval Requirements and Restrictions: See Approval Order.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...