Retrospective clinical specimens from hospital patients; Retrospective specimens from volunteer whole blood donors; Retrospective specimens from individuals at increased risk (hemodialysis, hemophilia, injection drug users); Characterized HCV infection cohorts (acute, chronic, asymptomatic)
Retrospective clinical specimens were used to evaluate the prevalence of anti-HCV, assess assay reactivity/non-reactivity, and determine percent agreement positive/negative compared to reference assays.
Prevalence study in random populations; Retrospective analysis of clinical specimens
Hospital patients, first-time volunteer whole blood donors, and random volunteer whole blood donors; Sample Size: 1,003 hospital patients; 2,508 first-time donors; 1,006 random donors; Number of Sites: Multiple (2 sites for hospital patients, 4 for first-time donors, 2 for random donors)
Abbott HCV EIA 2.0 and supplemental strip immunoblot assay
Prevalence of anti-HCV
Reactivity and Agreement Study; Retrospective evaluation of characterized specimens
Individuals with characterized HCV infection, hospital patients, and individuals at increased risk (hemodialysis, hemophilia, injection drug users); Sample Size: 679 specimens; Number of Sites: 3 clinical sites
Abbott HCV EIA 2.0 and FDA-approved strip immunoblot assay
Percent Agreement Positive and Percent Agreement Negative
Indications for Use
AxSYM® Anti-HCV (hepatitis C virus) is a Microparticle Enzyme Immunoassay (MEIA) for the qualitative detection of anti-HCV IgG to HCV recombinant proteins in human serum or plasma (potassium EDTA, sodium EDTA, sodium heparin, lithium heparin, sodium citrate, and potassium oxalate). Assay results, in conjunction with other laboratory results and clinical information, may be used to provide presumptive evidence of infection with HCV virus (state of infection or associated disease not determined) in persons with signs or symptoms of hepatitis and in persons at risk for hepatitis C infection.
Device Story
AxSYM Anti-HCV is a Microparticle Enzyme Immunoassay (MEIA) used on the AxSYM System. Input: human serum or plasma samples. Process: sample diluted; incubated with recombinant HCV antigen-coated microparticles (HCr43, c200, c100-3) to form antibody-antigen complexes; complexes bind to glass fiber matrix; washed; reacted with Anti-Human IgG: Alkaline Phosphatase Conjugate; substrate (4-methylumbelliferyl phosphate) added; alkaline phosphatase catalyzes fluorescent product (4-methylumbelliferone). Output: fluorescence rate measured by optical assembly; compared to cut-off rate (0.12 x Index Calibrator mean rate) to determine reactivity (S/CO ratio). Used in clinical laboratories by trained personnel. Output aids physicians in presumptive diagnosis of HCV infection when combined with clinical assessment and supplemental testing (e.g., immunoblot, PCR). Benefits: identifies potential HCV infection, enabling appropriate clinical management and public health screening.
Clinical Evidence
Clinical evaluation across seven sites (n=5,623 specimens). Evaluated prevalence in hospital patients (2.9%), first-time blood donors (2.4%), and random donors (0%). Percent Agreement Positive was 100% (157/157) in characterized HCV infection populations. Compared to Abbott HCV EIA 2.0 and supplemental strip immunoblot assays. Sensitivity and specificity established; reproducibility tested via NCCLS EP5-A guidelines with total %CV ranging from 7.3% to 23.2%.
Technological Characteristics
MEIA technology; uses recombinant HCV antigens (HCr43, c200, c100-3) expressed in E. coli or S. cerevisiae. Energy source: AxSYM System optical assembly for fluorescence measurement. Form factor: Reagent pack for automated analyzer. Connectivity: Standalone system (AxSYM). Software: Automated data processing and assay control. Sterilization: Not applicable (reagents).
Indications for Use
Indicated for qualitative detection of anti-HCV IgG in human serum or plasma for individuals with signs/symptoms of hepatitis or those at risk for HCV infection. No specific age or gender contraindications.
Regulatory Classification
Identification
A hepatitis C virus (HCV) antibody test is identified as an in vitro diagnostic device intended for use with human serum, plasma, or other matrices as a prescription device that aids in the diagnosis of HCV infection in persons with signs and symptoms of hepatitis and in persons at risk for hepatitis C infection. The test is not intended for screening blood, plasma, cell, or tissue donors.
Special Controls
*Classification.* Class II (special controls). The special controls for this device are:(1) The labeling required under § 809.10(b) of this chapter must include:
(i) A prominent statement that the test is not intended for the screening of blood, plasma, and cell or tissue donors.
(ii) Limitations, which must be updated to reflect current clinical practice and disease presentation and management. The limitations must include, but are not limited to, statements that indicate:
(A) When appropriate, the performance characteristics of the test have not been established in populations of immunocompromised or immunosuppressed patients or, other special populations where test performance may be affected.
(B) The detection of HCV antibodies indicates a present or past infection with hepatitis C virus, but does not differentiate between acute, chronic, or resolved infection.
(C) The specimen types for which the device has been cleared, and that use of the test with specimen types other than those specifically cleared for this device may result in inaccurate test results.
(D) Test results are to be interpreted by qualified licensed healthcare professionals in conjunction with the individual's clinical presentation, history, and other laboratory results.
(E) A non-reactive test result may occur early during acute infection, prior to development of a host antibody response to infection, or when analyte levels are below the limit of detection of the test.
(iii) A detailed explanation of the principles of operation and procedures for performing the test.
(2) Design verification and validation must include the following:
(i) A detailed device description, including all parts that make up the device, ancillary reagents required but not provided, an explanation of the device methodology, and design of the antigen(s) and capture antibody(ies) sequences, rationale for the selected epitope(s), degree of amino acid sequence conservation of the target, and the design and nature of all primary, secondary, and subsequent standards used for calibration.
(ii) Documentation and characterization (
*e.g.,* supplier, determination of identity, and stability) of all critical reagents (including description of the antigen(s) and capture antibody(ies)), and protocols for maintaining product integrity throughout its labeled shelf life.(iii) Risk analysis and management strategies, such as Failure Modes Effects Analysis and/or Hazard Analysis and Critical Control Points summaries and their impact on test performance.
(iv) Final release criteria to be used for manufactured test lots with appropriate evidence that lots released at the extremes of the specifications will meet the claimed analytical and clinical performance characteristics as well as the stability claims.
(v) Stability studies for reagents must include documentation of an assessment of real-time stability for multiple reagent lots using the indicated specimen types and must use acceptance criteria that ensure that analytical and clinical performance characteristics are met when stability is assigned based on the extremes of the acceptance range.
(vi) All stability protocols, including acceptance criteria.
(vii) Final release test results for each lot used in clinical studies.
(viii) Multisite reproducibility study that includes the testing of three independent production lots.
(ix) Analytical performance studies and results for determining the limit of blank (LoB), limit of detection (LoD), cutoff, precision (reproducibility) including lot-to-lot and/or instrument-to-instrument precision, interference, cross reactivity, carryover, hook effect, seroconversion panel testing, matrix equivalency, specimen stability, reagent stability, and cross-genotype antibody detection sensitivity, when appropriate.
(x) Analytical sensitivity of the test is the same or better than that of other cleared or approved tests.
(xi) Detailed documentation of clinical performance testing from a multisite clinical study. Performance must be analyzed relative to an FDA cleared or approved HCV antibody test, or a comparator that FDA has determined is appropriate. This study must be conducted using appropriate patient samples, with an acceptable number of HCV positive and negative samples in applicable risk categories. Additional relevant patient groups must be validated as appropriate. The samples may be a combination of fresh and repository samples, sourced from geographically diverse areas. The study designs, including number of samples tested, must be sufficient to meet the following criteria:
(A) Clinical sensitivity of the test must have a lower bound of the 95 percent confidence interval of greater than or equal to 95 percent.
(B) Clinical specificity of the test must have a lower bound of the 95 percent confidence interval of greater than or equal to 96 percent.
(3) For any HCV antibody test intended for Point of Care (PoC) use, the following special controls, in addition to those listed in paragraphs (b)(1) and (2) of this section, apply:
(i) Clinical studies must be conducted at PoC sites.
(ii) Additional labeling must include a brief summary of the instructions for use that are appropriate for use in a PoC environment.
Reference Devices
Abbott HCV EIA 2.0
AxSYM System (k974651)
AxSYM HCV version 3.0 (List No. 3B44)
AxSYM Abbott HCV (List No. 1D30)
Submission Summary (Full Text)
{1}
# Summary of Safety and Effectiveness Data
## I. GENERAL INFORMATION
Device Generic Name: A Microparticle Enzyme Immunoassay for the qualitative detection of Antibody to Hepatitis C Virus (anti-HCV)
Device Trade Name: Abbott AxSYM® Anti-HCV
Applicant’s Name and Address: Abbott Laboratories
Abbott Diagnostics Division
100 Abbott Park Road
Abbott Park, Illinois 60064
Premarket Approval Application (PMA) Number: P970027
Date of Panel Recommendation: January 21, 2000
Date of Notice of Approval to the Applicant: February 5, 2004
## II. INDICATIONS FOR USE
AxSYM® Anti-HCV (hepatitis C virus) is a Microparticle Enzyme Immunoassay (MEIA) for the qualitative detection of anti-HCV IgG to HCV recombinant proteins in human serum or plasma (potassium EDTA, sodium EDTA, sodium heparin, lithium heparin, sodium citrate, and potassium oxalate).
Assay results, in conjunction with other laboratory results and clinical information, may be used to provide presumptive evidence of infection with HCV virus (state of infection or associated disease not determined) in persons with signs or symptoms of hepatitis and in persons at risk for hepatitis C infection.
## III. DEVICE DESCRIPTION
Abbott AxSYM® Anti-HCV (hereinafter referred to as AxSYM Anti-HCV) detects antibodies to putative structural and nonstructural proteins of the HCV genome. The relationship between the recombinant HCV proteins in AxSYM Anti-HCV and the putative structural and nonstructural proteins is depicted in the following figure.
{2}
Recombinant HCV Proteins in AxSYM Anti-HCV Polyprotein AA # 1 to 1931
| Core | E1 | E2/NS1 | NS2 | NS3 | NS4 |
| --- | --- | --- | --- | --- | --- |
| | | | | | |
| | | | | | |
| Core | | | | 33c | |
| HCr43 | AA # 1192 to 1457 + 1 to 150 | | | | |
| c200 | AA # 1192 to 1931 | | hSOD | c200 | |
| c100-3 | AA # 1569 to 1931 | | hSOD | c100-3 | |
* The complete HCV polyprotein contains the additional amino acids # 1932 to 3011 which are not represented in this figure.
AxSYM Anti-HCV utilizes three recombinant antigens (HCr43, c200, and c100-3) to detect anti-HCV. These antigens are used to prepare the AxSYM Anti-HCV Microparticles. All three antigens are purchased as purified proteins from Chiron® Corporation, Emeryville, California. A brief description of each is as follows:
- HCr43: The HCr43 antigen is a protein expressed in *Escherichia coli* (E. coli) and is composed of two noncontiguous coding regions of the Chiron HCV genome sequence. The first region represents amino acids 1192 to 1457 (33c) of the Chiron HCV sequence. The second of the two regions represent amino acids 1 to 150 (core) of the Chiron HCV sequence. Because of the similarity of the genomic organization of other flaviviruses, it is suggested that the first sequence be from the NS3 coding region and the second sequence is from the core-coding region of HCV.
- c200: The c200 antigen is a recombinant HCV protein expressed in the yeast *Saccharomyces cerevisiae*. It contains HCV amino acids 1192 to 1931 of the Chiron HCV sequence. Because of the similarity of the genomic organization of other flaviviruses, it is suggested that this sequence be from the NS3 and NS4 regions of the HCV genome. The c200 antigen is a chimeric fusion protein with 154 amino acids of human superoxide dismutase (hSOD).
- c100-3: The c100-3 antigen is a recombinant HCV protein expressed in the yeast *Saccharomyces cerevisiae*. The genomic organization of other flaviviruses suggests that the clone sequence is contained within the putative nonstructural (NS3 and NS4) regions of HCV. The c100-3 protein is a chimeric fusion protein with 154 amino acids of hSOD, five linker amino acids, amino acids number 1569 to 1931 of the HCV polyprotein, and an additional five linker amino acid chain at the carboxyl terminus.
8
{3}
The AxSYM Anti-HCV reagents require the use of the AxSYM System, which is cleared under 510(k) k974651. The analyzer performs all sample and reagent transfers, incubations, data processing, and completes the assay with a printed report.
The sample and all AxSYM Anti-HCV reagents required for one test are pipetted by the Sampling Probe into various wells of a reaction vessel (RV) in the Sampling Center. The RV is immediately transferred into the Processing Center. The Processing Probe performs further pipetting steps in the Processing Center.
The reactions occur in the following sequence:
- Sample is diluted with Specimen Diluent 1 in one RV well and then further diluted with Specimen Diluent 2 in a second RV well.
- Recombinant HCV Antigen Coated Microparticles are added to the diluted sample. During the incubation of this reaction mixture, the anti-HCV in the sample binds to the Recombinant HCV Antigen Coated Microparticles, forming an antibody-antigen complex.
- A portion of the reaction mixture is transferred to the matrix cell. The microparticles bind irreversibly to the glass fiber matrix.
- The matrix cell is washed to remove materials not bound to the microparticles.
- Anti-Human IgG (Goat): Alkaline Phosphatase Conjugate is dispensed onto the matrix cell. During the incubation of this reaction mixture, the Anti-Human IgG (Goat): Alkaline Phosphatase Conjugate binds to the antibody-antigen complex.
- The matrix cell is washed to remove materials not bound to the microparticles.
- The substrate, 4-methylumbelliferyl phosphate, is added to the matrix cell. The alkaline phosphatase-labeled conjugate catalyzes the removal of a phosphate group from the substrate, yielding the fluorescent product, 4-methylumbelliferone. The MEIA optical assembly measures this fluorescent product.
The presence or absence of anti-HCV in the sample is determined by comparing the rate of formation of fluorescent product to the cut-off rate, which is calculated from a previous AxSYM Anti-HCV Index Calibration. If the rate of formation of fluorescent product in the sample is greater than or equal to the cut-off rate, the sample is considered reactive for anti-HCV.
## IV. CONTRAINDICATIONS, WARNINGS AND PRECAUTIONS
There are no known contraindications for the AxSYM Anti-HCV assay.
Warnings and Precautions for the AxSYM Anti-HCV assay are stated in the product labeling.
{4}
4
10
# V. ALTERNATE PRACTICES AND PROCEDURES
Determining the presence of anti-HCV in patients may be achieved by using a variety of commercially available, FDA-approved serological tests. When these tests results are used in combination with a physician’s assessment and other laboratory test results, infection with HCV can be identified.
# VI. MARKETING HISTORY
This product (AxSYM Anti-HCV, List No. 5C36) has not been marketed in any other country. Two similar assays (AxSYM HCV version 3.0, List No. 3B44 and AxSYM Abbott HCV, List No. 1D30) are manufactured and marketed in the following countries outside the United States since 1997.
| Algeria | El Salvador | Malta | Slovakia |
| --- | --- | --- | --- |
| Argentina | Estonia | Mauritius | Slovenia |
| Australia | Finland | Mexico | South Africa |
| Austria | France | Morocco | South Korea |
| Bahrain | Germany | Namibia | Spain |
| Belarus | Greece | Netherlands | Sweden |
| Belgium | Guam / Saipan | New Zealand | Switzerland |
| Bosnia / Herzegovina | Guatemala | North Cypress | Syria |
| Brazil | Honduras | North Yemen | Taiwan |
| Bulgaria | Hungary | Norway | Thailand |
| Canada | India | Oman | Tunisia |
| Chile | Indonesia | Pakistan | Turkey |
| China | Ireland | Panama | U.A.B. |
| C.I.S. | Israel | Paraguay | U.A.E. (Dubai) |
| Columbia | Italy | Peru | Ukraine |
| Costa Rica | Ivory Coast | Philippines | United Kingdom |
| Croatia | Japan | Poland | Uruguay |
| Curacao | Jordan | Portugal | Venezuela |
| Cyprus | Kenya | Qatar | Vietnam |
| Czech Republic | Kuwait | Romania | Yugoslavia |
| Denmark | Latvia | Saudi Arabia | Zimbabwe |
| Dominican Republic | Lebanon | Senegal | |
| Ecuador | Lithuania | Serbia | |
| Egypt | Malaysia | Singapore | |
The devices have not been withdrawn from any of these countries for reasons related to safety and effectiveness of the device.
{5}
5
1
# VII. POTENTIAL ADVERSE EFFECTS OF THE DEVICE ON HEALTH
There are no known potential direct adverse effects of this device on the health of the user if the device is used according to the instructions in the package insert. However, failure of the test to perform as indicated or human error during performance of the test may lead to a false diagnosis and improper patient management.
A false positive result using an anti-HCV assay is not considered a patient or public health concern since a reactive EIA results in a clinical lab should be followed up with supplemental tests (e.g., strip immunoblot assay (SIA) and/or PCR for detection of HCV RNA) to determine inactive or resolved infection versus active HCV replication. Treatment of the patient with chronic HCV infection is initiated only after extensive clinical, laboratory and behavioral assessment of the patient (e.g., elevated ALT levels for six months, detectable serum HCV RNA, liver biopsy with portal fibrosis, patient compliance, and abstinence from drugs and alcohol).
A false negative anti-HCV result in a diagnostic setting may lead to a patient with HCV going unidentified. Under these circumstances, there is a safety concern for both the patient and the public, since such individuals may be capable of transmitting HCV infection. However, if a patient is known to be at high risk of HCV infection, or is symptomatic, and the physician’s suspicion of HCV infection is high, HCV RNA testing is often employed and is of diagnostic value, even after an initial negative anti-HCV test result.
# VIII. SUMMARY OF NON CLINICAL STUDIES
The applicant performed in-house studies in order to evaluate the following performance characteristics of the AxSYM Anti-HCV assay: Cut-off determination and verification, sample handling and collection, interference, sample carryover, and stability.
## Cut-off Determination and Verification
The cut-off value was selected in order to have an appropriate assay cut-off that would result in optimal specificity while maintaining acceptable sensitivity for AxSYM Anti-HCV. In order to select the preliminary cut-off, data obtained from the in-house studies (n = 1,262) and (n = 146) were evaluated. The preliminary cut-off was then challenged or verified during the clinical studies (n = 2,209) and (n = 679).
The cut-off for AxSYM Anti-HCV is described as a fraction of the Index Calibrator (IC) rate. The sponsor conducted studies to determine the optimal IC multiplication factor to be used for calculating the cut-off rate.
Receiver-operating characteristic (ROC) analysis was used to assist in the determination and verification of the AxSYM Anti-HCV assay cutoff. The cut-off was verified by ROC analysis after re-categorization of clinical specimens and expressed as follows:
Cut-off Rate (CO) = 0.12 Index Calibrator mean rate
{6}
Assay results are expressed as the ratio of the sample rate to the cut-off rate. The following AxSYM Anti-HCV result calculation is used:
$$
\mathrm{S/CO} = \frac{\text{Sample Rate (S)}}{\text{Cut-off Rate (CO)}}
$$
## Clinical Sample Types (Serum and Plasma)
The sponsor conducted a study using the AxSYM Anti-HCV assay to determine if there was any difference in results when using plasma specimens or serum specimens.
Data obtained demonstrated that there were no statistically significant differences and support the use of the AxSYM Anti-HCV assay with serum specimens collected in serum tubes or serum separator tubes, and plasma specimens collected in the following anticoagulants: potassium EDTA, sodium EDTA, sodium heparin, lithium heparin, sodium citrate, and potassium oxalate.
## Sample Freeze/Thaw
The sponsor conducted studies to assess the performance of AxSYM Anti-HCV with plasma specimens that have undergone multiple freeze/thaw cycles.
Data obtained support the use of the AxSYM Anti-HCV assay with specimens that have undergone up to six freeze/thaw cycles.
## Sample Heat Inactivation
The sponsor conducted a study to assess the performance of the AxSYM Anti-HCV assay with plasma specimens that have been heat-inactivated.
The data do not support the use of the AxSYM Anti-HCV assay with specimens that have been heat-inactivated at $56^{\circ}\mathrm{C}$. This information is provided in the Sample collection and Preparation for Analysis section of the labeling.
## "Off the Cells" Sample Shipping/Storage Conditions
A study was conducted to assess the performance of the AxSYM Anti-HCV assay with plasma specimens after removal from the collection tubes (i.e., "off the cells") subjected to various shipping/storage conditions.
The data support the use of the AxSYM Anti-HCV assay with specimens that have been stored off the cells at the following conditions: 2 to $8^{\circ}\mathrm{C}$ for 28 days; 15 to $30^{\circ}\mathrm{C}$ for 14 days; $37^{\circ}\mathrm{C}$ for 3 days.
The data do not support the use of the AxSYM Anti-HCV assay with specimens that have been stored off the cells at $45^{\circ}\mathrm{C}$. This information is provided in the Sample collection and Preparation for Analysis section of the labeling.
6
12
{7}
7
13
# "On the Cells" Sample Shipping/Storage Conditions
The sponsor conducted a study to assess the performance of the AxSYM Anti-HCV assay with plasma specimens stored in the original collection tubes (i.e., "on the cells") subjected to various shipping/storage conditions.
The data support the use of the AxSYM Anti-HCV assay with specimens that have been stored on the cells at the following conditions: 2 to 8°C for 28 days; 15 to 30°C for 14 days; and 37°C for 7 days.
The data do not support the use of the AxSYM Anti-HCV assay with specimens that have been stored on the cells at 45°C. This information is provided in the Sample collection and Preparation for Analysis section of the labeling.
# Potentially Interfering Substances (Triglycerides)
The sponsor conducted a study to assess the performance of the AxSYM Anti-HCV assay to determine the potential interference associated with use of specimens containing various levels (30 to 3,000 mg/dL) of triglycerides. Expected serum or plasma levels of triglycerides in an apparently healthy, fasting adult population can range from 30 to 220 mg/dL.
The data support the use of the AxSYM Anti-HCV assay with specimens that contain up to 3,000 mg/dL of triglycerides.
# Potentially Interfering Substances (Total Protein)
The sponsor conducted a study to assess the performance of the AxSYM Anti-HCV assay to determine the potential interference associated with use of specimens containing various levels (6.1 to 12.0 g/dL) of total protein. Expected serum or plasma levels of total protein in an apparently healthy population can range from 6.4 to 8.2 g/dL.
The data support the use of the AxSYM Anti-HCV assay with specimens that contain up to 12 g/dL of total protein.
# Potentially Interfering Substances (Total Bilirubin)
The sponsor conducted a study using the AxSYM Anti-HCV assay to assess the potential assay interference associated with use of specimens containing various levels (< 0.2 to 20 mg/dL) of total bilirubin. Expected levels of total bilirubin in an apparently healthy, adult population can range from 0.2 to 1.2 mg/dL.
The data support the use of the AxSYM Anti-HCV assay with specimens that contain up to 20 mg/dL of total bilirubin.
# Potentially Interfering Substances (Hemoglobin)
{8}
The sponsor conducted a study to assess the AxSYM Anti-HCV assay for potential interference associated with use of specimens containing various levels (0 to 500 mg/dL) of hemoglobin. Expected serum or plasma levels of hemoglobin in an apparently healthy, adult population can range from 0 to 2.5 mg/dL.
The data support the use of the AxSYM Anti-HCV assay with specimens that contain up to 500 mg/dL of hemoglobin.
## Potentially Interfering Substances (Red Blood Cells)
The sponsor conducted a study to assess the AxSYM Anti-HCV assay for potential assay interference associated with use of specimens containing various levels (0.2 to 0.4%v/v) of red blood cells.
The data support the use of the AxSYM Anti-HCV assay with specimens that contain up to 0.4%(v/v) of red blood cells.
## Microbial Contamination of Specimens
The sponsor conducted a study to assess the AxSYM Anti-HCV assay for potential interference associated with specimens contaminated with microorganisms.
The data do not support the use of the AxSYM Anti-HCV assay with specimens that may have been microbially contaminated with as much as 1×10⁶ CFU/mL of spores (Bacillus subtilis, Candida albicans), mold (Aspergillus niger), vegetative cells (Escherichia coli, Pseudomonas aeruginosa, Staphylococcus aureus), or an environmental isolate (Pseudomonas fluorescens). This is indicated in both the Warnings and Precautions section and Specimen Collection and Preparation for Analysis section of the labeling.
## Sample Carryover
The sponsor conducted a study to evaluate the effect of sample carryover within the AxSYM Anti-HCV assay or from other representative AxSYM assays when processing samples that contain high levels of anti-HCV.
The data indicate that no significant sample carryover was detected within AxSYM Anti-HCV. However, AxSYM Anti-HCV is susceptible to specimen carryover from AxSYM Total T3 when a high-titer anti-HCV specimen is tested by AxSYM Total T3 and is then followed by a specimen tested by AxSYM Anti-HCV. This carryover may result in a falsely reactive AxSYM Anti-HCV result. No significant sample carryover was detected from other AxSYM assays into AxSYM Anti-HCV. This effect is stated in the Warnings and Precautions section of the labeling.
## Onboard Reagent Stability
14
{9}
A study was conducted to determine how long the AxSYM Anti-HCV Reagent Pack can be stored onboard the AxSYM System without recalibration.
The data support the storage of the AxSYM Anti-HCV Reagent Pack onboard the AxSYM System for 336 hours without recalibration. However, to be conservative, the package insert indicates that the AxSYM Anti-HCV Reagent Pack may be on board the AxSYM System for a maximum of 112 hours.
## Reagent Stability
A real-time study was conducted to demonstrate the shelf-life integrity of the AxSYM Anti-HCV Reagent Pack and Controls at the recommended storage conditions (2 to 8°C) and to demonstrate the efficacy of these reagents and controls following warehouse/shipping simulation.
The data support a shelf-life of 12 months when stored at 2 to 8°C for the AxSYM Anti-HCV Reagent Pack and 14 months for the AxSYM Anti-HCV Controls. The data also support ambient shipment of AxSYM Anti-HCV Reagent Packs and Controls based on simulated warehouse shipping/storage for 72 hours followed by 2 to 8°C storage.
## Calibration Stability
A Calibration Check Panel was used to evaluate the calibration curve stability of the AxSYM Anti-HCV assay. The first run of each day was the six-member Calibration Check Panel consisting of Index Calibrator, Negative Control, Positive Control, and a supplemental panel consisting of high, medium, and low panel members. The Calibration Check Panel was tested using one reagent master lot at one clinical site for 15 working days following precision testing and then once each week until the end of the study. The AxSYM Anti-HCV calibration curve was demonstrated to be stable for up to 39 days after calibration.
## Analytical Specificity (Cross-Reactivity)
In order to evaluate analytical specificity of the AxSYM Anti-HCV, one site tested a total of 200 specimens containing substances that could potentially be cross-reactive. See Table V in the Specific Performance Characteristics section of the labeling for a list of the substances tested.
One hundred eighty-four specimens were nonreactive and 15 specimens were reactive by AxSYM Anti-HCV. Thirteen (86.67%) of the 15 AxSYM Anti-HCV reactive specimens were positive for anti-HCV by the supplemental assay.
The Analytical specificity of AxSYM Anti-HCV was estimated to be 98.9% (185/187) with a 95% confidence interval of 96.2% to 99.9% by the binomial distribution.
{10}
10
16
# Analytical Sensitivity
In order to evaluate analytical sensitivity, the sponsor conducted a qualitative study using a ten-member lot release panel (Antibody to Hepatitis C Virus [anti-HCV] Panel #10) obtained from the Center for Biologics Evaluation and Research (CBER).
Table I
| CBER
Anti-HCV Panel #10
Designation | AxSYM Anti-HCV
Mean S/CO
Results | AxSYM Anti-HCV
Qualitative Results | Expected
Qualitative Results |
| --- | --- | --- | --- |
| Member 1001 | 1.27 | R | R or NR |
| Member 1002 | 4.86 | R | R |
| Member 1003 | 0.62 | NR | R or NR |
| Member 1004 | 17.07 | R | R |
| Member 1005 | 2.74 | R | R |
| Member 1006 | 9.50 | R | R |
| Member 1007 | 6.33 | R | R |
| Member 1008 | 8.54 | R | R |
| Member 1009 | 10.37 | R | R |
| Member 1010 | 0.36 | NR | NR |
R = Reactive, NR = Nonreactive
The AxSYM Anti-HCV test results agreed with the expected qualitative results provided in the anti-HCV Panel #10 data sheet.
# Reproducibility
The reproducibility of AxSYM Anti-HCV was determined using the recommendations stated in the NCCLS Approved Guideline EP5-A. Three sites tested a coded 35-member panel consisting of seven unique members repeated five times within the panel. The panel was tested with one reagent master lot over a period of 20 days. Each run consisted of seven panel members tested in duplicate and two replicates each of the AxSYM Anti-HCV Negative Control (NC), Positive Control (PC), and Index Calibrator (IC). Two runs were performed daily with a minimum of 2 hours between each daily run. The total percent coefficient of variation (%CV) ranged from 7.3 to 12.2 for the anti-HCV positive panel members, from 9.4 to 10.8 for the borderline positive panel members, and from 9.1 to 14.6 for the negative panel members.
An additional study was performed to assess between-lot and between-site precision using a coded 24-member panel consisting of six unique members, four anti-HCV positive members (mean Specimen to cut-off (S/CO) ranged from 1.30 to 14.48) and two anti-HCV negative members, repeated four times within the panel. The panel was tested with three AxSYM Anti-
{11}
HCV reagent master lots twice daily for 5 days at three clinical sites. Each daily run included two replicates of the AxSYM Anti-HCV IC and the AxSYM Anti-HCV Controls. The PROC MIXED procedure in SAS® was used to estimate variance components for the mixed model. The between-lot %CV for the panel members was less than or equal to 20.0, the between-site %CV was less than or equal to 22.2, and the total %CV was less than or equal to 23.2.
Table 2
| Unique Member | Panel Members | Expected Range in S/CO |
| --- | --- | --- |
| A | 1, 8, 21, 22, 29 | 11.0 to 17.0 |
| B | 2, 12, 19, 24, 35 | 7.0 to 10.0 |
| C | 3, 13, 17, 23, 31 | 4.0 to 6.0 |
| D | 4, 10, 15, 28, 32 | 2.0 to 4.0 |
| E | 5, 11, 16, 25, 33 | 0.9 to 1.1 |
| F | 6, 14, 18, 26, 34 | 0.5 to 0.8 |
| G | 7, 9, 20, 27, 30 | 0.1 to 0.4 |
The reproducibility of AxSYM Anti-HCV was evaluated by testing the Proficiency Panel to determine within-run, between-run, between-day, and total variability. Precision data are presented in Tables 3 through 5. The PROC MIXED procedure in SAS was utilized to produce variance components for all models. These variance components were used to create estimates of variance (standard deviation) and percent coefficient of variation (%CV) for within-run, between-run, between-day, and total. All effects were considered as random for these analyses.
The following model was used to analyze the data by site:
Results = Day + Run (Day) + Error
Where: Day is the main effect
Run is nested within each Day
Error is the Replicate or within-run effect
Note: Results are the Rate values for the Index Calibrator and S/CO values for the controls and panels.
The estimate of variance for Total will be the sum of the variance components for Day, Run (Day), and Error. The estimate of variance for Between-Day will be the variance component for Day.
The estimate of variance for Between-Run will be the variance component for Run (Day).
The estimate of variance for Within-Run will be the variance component for Error.
{12}
Table 3 -AxSYM Anti-HCV Reproducibility
Site – Abbott Laboratories
| Panel Member | Total No. Reps | Grand Mean (S/CO) | Within-Run | | Between-Run | | Between-Day | | Total | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| A | 80 | 14.21 | 1.019 | 7.2 | 0.000 | 0.0 | 0.214 | 1.5 | 1.042 | 7.3 |
| B | 80 | 8.75 | 0.593 | 6.8 | 0.259 | 3.0 | 0.000 | 0.0 | 0.647 | 7.4 |
| C | 80 | 4.77 | 0.358 | 7.5 | 0.129 | 2.7 | 0.000 | 0.0 | 0.380 | 8.0 |
| D | 80 | 3.12 | 0.322 | 10.3 | 0.000 | 0.0 | 0.052 | 1.7 | 0.326 | 10.4 |
| E | 80 | 1.04 | 0.093 | 9.0 | 0.000 | 0.0 | 0.062 | 5.9 | 0.112 | 10.8 |
| F | 80 | 0.65 | 0.053 | 8.1 | 0.000 | 0.0 | 0.030 | 4.6 | 0.061 | 9.3 |
| G | 80 | 0.23 | 0.022 | 9.5 | 0.009 | 4.0 | 0.004 | 1.8 | 0.024 | 10.4 |
| NC | 80 | 0.23 | 0.034 | 14.3 | 0.000 | 0.0 | 0.016 | 6.7 | 0.037 | 15.8 |
| PC | 80 | 4.34 | 0.303 | 7.0 | 0.100 | 2.3 | 0.047 | 1.1 | 0.323 | 7.4 |
| Panel Member | Total No. Reps | Grand Mean (Rate) | Within-Run | | Between-Run | | Between-Day | | Total | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| IC | 80 | 97.31 | 8.033 | 8.3 | 2.007 | 2.1 | 0.630 | 0.6 | 8.303 | 8.5 |
Table 4 - AxSYM Anti-HCV Reproducibility
Site – J.T. Mather Memorial Hospital
| Panel Member | Total No | | | Grand Mean | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | Reps | (S/CO) | SD | %CV | SD | Within-Run %CV | Between-Run SD | Between-Day %CV | SD | Total %CV |
| A | 80 | 11.14 | 0.875 | 7.9 | 0.000 | 0.0 | 0.470 | 4.2 | 0.993 | 8.9 |
| B | 80 | 7.32 | 0.764 | 10.4 | 0.000 | 0.0 | 0.319 | 4.4 | 0.828 | 11.3 |
| C | 80 | 4.03 | 0.291 | 7.2 | 0.038 | 0.9 | 0.188 | 4.7 | 0.349 | 8.7 |
| D | 80 | 2.44 | 0.212 | 8.7 | 0.000 | 0.0 | 0.080 | 3.3 | 0.227 | 9.3 |
| E | 80 | 0.87 | 0.063 | 7.3 | 0.041 | 4.7 | 0.032 | 3.7 | 0.082 | 9.4 |
| F | 80 | 0.56 | 0.048 | 8.7 | 0.000 | 0.0 | 0.014 | 2.6 | 0.050 | 9.1 |
| G | 80 | 0.21 | 0.021 | 9.7 | 0.007 | 3.1 | 0.006 | 2.7 | 0.022 | 10.5 |
| NC | 80 | 0.21 | 0.026 | 12.5 | 0.000 | 0.0 | 0.011 | 5.2 | 0.028 | 13.5 |
| PC | 80 | 3.89 | 0.267 | 6.9 | 0.000 | 0.0 | 0.079 | 2.0 | 0.278 | 7.2 |
| Panel Member | Total No. Reps | Grand Mean (Rate) | Within-Run | | Between-Run | | Between-Day | | Total | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| IC | 80 | 90.65 | 8.362 | 9.2 | 0.000 | 0.0 | 1.396 | 1.5 | 8.477 | 9.4 |
{13}
Table 5
AxSYM Anti-HCV Reproducibility
Site – Sacramento Medical Foundation
| Panel Member | Total No. Reps | Grand Mean (S/CO) | Within-Run | | Between-Run | | Between-Day | | Total | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| A | 80 | 12.79 | 1.005 | 7.9 | 0.233 | 1.8 | 0.406 | 3.2 | 1.109 | 8.7 |
| B | 80 | 8.32 | 0.635 | 7.6 | 0.450 | 5.4 | 0.394 | 4.7 | 0.872 | 10.5 |
| C | 80 | 4.53 | 0.364 | 8.0 | 0.139 | 3.1 | 0.247 | 5.4 | 0.461 | 10.2 |
| D | 80 | 2.82 | 0.282 | 10.0 | 0.184 | 6.5 | 0.068 | 2.4 | 0.343 | 12.2 |
| E | 80 | 0.99 | 0.099 | 10.0 | 0.000 | 0.0 | 0.037 | 3.8 | 0.105 | 10.7 |
| F | 80 | 0.63 | 0.064 | 10.2 | 0.013 | 2.0 | 0.017 | 2.8 | 0.067 | 10.7 |
| G | 80 | 0.23 | 0.031 | 13.5 | 0.012 | 5.3 | 0.000 | 0.0 | 0.034 | 14.6 |
| NC | 80 | 0.21 | 0.034 | 16.1 | 0.000 | 0.0 | 0.012 | 5.6 | 0.035 | 17.0 |
| PC | 80 | 4.15 | 0.306 | 7.4 | 0.000 | 0.0 | 0.143 | 3.4 | 0.337 | 8.1 |
| Panel Member | Total No. Reps | Grand Mean (Rate) | Within-Run | | BetweenRun | | Between-Day | | Total | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| IC | 80 | 93.76 | 8.811 | 9.4 | 0.000 | 0.0 | 5.181 | 5.5 | 10.221 | 10.9 |
Reproducibility has not been established for specimens containing the following anticoagulants: EDTA, heparin, sodium citrate, and potassium oxalate. The sponsor states in the labeling that the user is responsible for establishing assay reproducibility when using these matrices.
## IX. Summary of Clinical Investigation
The objective of this clinical study was to assess the efficacy of AxSYM Anti-HCV for detecting anti-HCV antibodies in human serum or plasma as presumptive evidence of HCV infection. The following performance characteristics were studied:
Assay reactivity was evaluated by testing retrospective specimens which included previously characterized populations of HCV positive specimens, populations known to be at increased risk for HCV infection, HCV seroconversion panels, and HCV genotype panel.
Assay non-reactivity was evaluated by testing prospective specimens which included specimens from random volunteer whole blood donors, first-time volunteer whole blood donors, and random hospital patients, and by testing retrospective specimens which included patients with disease states other than HCV and specimens containing potentially interfering substances. Performance of AxSYM Anti-HCV was compared to the current Abbott HCV EIA 2.0 with regards to reactivity and non-reactivity based upon clinical specimen testing.
Clinical utility of the ASYM Anti-HCV assay was assessed by testing specimens from hospital patients with physicians' orders for hepatitis testing.
{14}
Clinical evaluation of AxSYM Anti-HCV was conducted at seven sites. However, not all sites conducted the same studies. The utility of AxSYM Anti-HCV for the qualitative detection of anti-HCV in human serum and plasma was determined by measuring precision, sensitivity, and specificity. One of the seven clinical sites tested specimens from patients with physicians' orders for hepatitis testing and four of the seven sites evaluated the prevalence of anti-HCV in first-time volunteer whole blood donors. The seven clinical sites included a hospital laboratory, a plasmapheresis center, two blood research laboratories, a university medical center laboratory, a reference laboratory, and Abbott Laboratories. A total of 5,623 clinical specimens were tested. Four AxSYM Anti-HCV reagent master lots were used in the evaluations, and the comparative assay was the FDA-approved Abbott HCV EIA 2.0. Testing was performed according to the AxSYM Anti-HCV assay clinical protocol/ brochure and the Abbott HCV EIA 2.0 package insert. All reactive specimens and AxSYM Anti-HCV gray zone specimens (S/CO = 0.80 to 1.20) were tested by a FDA-approved supplemental strip immunoblot assay (supplemental assay).
## Prevalence
In order to evaluate the prevalence of anti-HCV detected by AxSYM Anti-HCV, random populations of hospital patients, first-time volunteer whole blood donors, and volunteer whole blood donors were tested. Two sites tested a total of 1,003 specimens from random hospital patients. Thirty-seven (3.7%) of the 1,003 random hospital patient specimens were reactive by AxSYM Anti-HCV. Twenty-nine (78.4%) of the 37 specimens were positive by the supplemental assay and considered to be positive for the presence of anti-HCV. Four specimens were repeatedly in the gray zone by AxSYM Anti-HCV and were excluded from the calculations. The estimated prevalence of anti-HCV detected by AxSYM Anti-HCV in a population of random hospital patients was 2.9% (29/999) with a 95% confidence interval of 2.0 to 4.1% by the binomial distribution. This prevalence is notably higher than the normal population.
Four sites tested a total of 2,508 specimens from first-time volunteer whole blood donors. Seventy-one (2.8%) of the 2,508 first-time volunteer whole blood donor specimens were reactive by AxSYM Anti-HCV. Sixty (84.5%) of the 71 specimens were positive by the supplemental assay and considered to be positive for the presence of anti-HCV. Three specimens were repeatedly in the gray zone by AxSYM Anti-HCV and were excluded from the calculations. The estimated prevalence of anti-HCV detected by AxSYM Anti-HCV in a population of first-time volunteer whole blood donors was 2.4% (60/2,505) with a 95% confidence interval of 1.8% to 3.1% by the binomial distribution.
Two sites tested a total of 1,006 random volunteer whole blood donors; one (0.1%) of the 1,006 random volunteer whole blood donor specimens was reactive by AxSYM Anti-HCV. This specimen was nonreactive by the supplemental assay and considered to be negative for the presence of anti-HCV. One specimen was repeatedly in the gray zone by AxSYM Anti-HCV and was excluded from the calculations. The estimated prevalence of anti-HCV detected by AxSYM Anti-HCV in a population of random volunteer whole blood donors was 0% (0/1,005) with a 95% confidence interval of 00.0 to 00.4% by the binomial distribution.
## Reactivity (Percent Agreement Positive)
The data in Table 6 are not the basis for claiming performance of AxSYM Anti-HCV in any particular population. While the data for different categories of individuals (including types of characterized HCV infections) are descriptive of populations that were studied, it is not known if the type of HCV infection or
20
{15}
associated diseases affects performance of AxSYM Anti-HCV, particularly with regard to effects that might be different from those on performance of other current assays that presumptively detect anti-HCV.
Table 6 presents AxSYM Anti-HCV results for 679 specimens, each representing an individual pertinent to AxSYM Anti-HCV intended use. All of these specimens were characterized by testing for anti-HCV and reference assays. Those specimens that contain anti-HCV were defined by reactive results with Abbott HCV EIA 2.0, that uses the same antigens as AxSYM Anti-HCV: core, 33c, c200, and c100-3, and with an FDA-approved strip immunoblot assay that uses similar antigens: c22 representing core, 33c, c100-3, and 5.1.1 representing NS4. Specimens not shown to contain anti-HCV were defined by any of the following patterns of results: HCV EIA 2.0 nonreactive, HCV EIA 2.0 reactive and immunoblot-nonreactive, or HCV EIA 2.0 reactive and immunoblot-indeterminate. AxSYM Anti-HCV results were analyzed for reactivity (AxSYM Anti-HCV-non-reactive results among specimens not shown to contain anti-HCV).
Percent Agreement Positive and Percent Agreement Negative between AxSYM Anti-HCV and the Abbott HCV EIA 2.0 and the strip immunoblot assay were calculated and 95 percent confidence intervals (95% CI) were obtained. These results are exhibited in Table 6.
| Category | Number of Specimens Tested | % Agreement Positive (95% Confidence Interval) | % Agreement Negative (95% Confidence Interval) |
| --- | --- | --- | --- |
| Characterized HCV
Infection† | 157 | 157/157 (100.0%)
(97.7 - 100.0) | |
| Individuals with Anti-HCV‡ | 273 | 273/273 (100.0%)
(98.7 - 100.0) | |
| Hospital Patients with Physicians’ Orders for Hepatitis Testing
Individuals at Increased Risk: | 99 | 5/5 (100.0%)
(47.8 - 100.0) | 93/94 (98.9%)
(94.2 - 100.0) |
| - Hemodialysis Patients | 50 | 26/26 (100.0%)
(86.8 - 100.0) | 16/24§ (66.7%)
(44.7 - 84.4) |
| - Hemophilia Patients | 50 | 50/50 (100.0%)
(92.9 - 100.0) | |
| - Injection Drug Users | 50 | 15/15 (100.0%)
(78.2 - 100.0) | 35/35 (100.0%)
(90.0 - 100.0) |
# Information about age and gender for individual is not available
† Characterized infections include 33 acute (15 seroconversions and 18 with jaundice or high aminotransferase levels and with other likely causes excluded), 24 chronic (anti-HCV detected for ≥ 6 months and one or more among histopathologic evidence of chronic hepatitis, treatment with interferon, or detectable HCV RNA), and 100 asymptomatic (identified as HCV-infected during donor screening). Tested specimens were selected from frozen collections and did not represent random samples of the studied populations.
‡ Insufficient information to enable determination of the state of HCV infection.
§ AxSYM Anti-HCV results for these 24 specimens: 16 nonreactive, 6 reactive, and 2 equivocal.
21
{16}
# Characterized HCV Infection
In order to determine Percent Agreement Positive, three clinical sites evaluated a total of 157 specimens from individuals who were clinically classified as being asymptomatic, with acute or chronic HCV infection. One hundred specimens were from normal, healthy blood donors who had asymptomatic HCV infection at the time of donation. Thirty-three specimens were from individuals with acute HCV infection; these included 15 seroconversions and 18 with jaundice or high aminotransferase levels. The remaining 24 specimens were from individuals who were clinically and serologically diagnosed with chronic HCV infection. Percent Agreement Positive of AxSYM Anti-HCV was estimated to be 100.00% (157/157) with a 95% confidence interval of 97.7% to 100.0% by the binomial distribution.
HCV RNA was detected in 11 of the 15 seroconversion panels. There were no HCV RNA results for the remaining four panels. Detection of anti-HCV by AxSYM Anti-HCV was coincident with HCV EIA 2.0 in 5 of the 15 seroconversion panels. AxSYM Anti-HCV detected anti-HCV earlier than HCV EIA 2.0 in 9 of the 15 seroconversion panels. HCV EIA 2.0 detected anti-HCV one bleed earlier than AXSYM Anti-HCV in one seroconversion panel.
# HCV Genotypes
A study was done to determine if AxSYM Anti-HCV could yield reactive results during infections with different variants of HCV. One hundred twenty-seven specimens were characterized by testing for anti-HCV with reference assays (as described above in Table 6) and by research methods for determining genotype or subtype (restriction fragment length polymorphism, line probe assay, sequence analysis of 5' nontranslated or core regions, or primer-specific polymerase chain reaction). The geographic origins of these specimens were North America, Central America, Africa, and Asia. Anti-HCV was detected in 125 specimens. Genotype or subtype was determined for all 127 specimens as 1a, 1b, 2a, 2b, 3, 3a, 4, 5, or 6a; subtype and genotype were respectively 2b and 5 for the 2 specimens shown not to contain anti-HCV. AxSYM results were reactive for 100% of the 125 specimens with anti-HCV and nonreactive for the 2 specimens shown not to contain anti-HCV. These data are suggestive of AxSYM Anti-HCV reactivity for a broad range of HCV variants but, because the accuracy of the genotyping methods is not known, and because the assay detects antibodies to HCV, it is not possible to determine any anti-HCV assay's effectiveness among all HCV variants.
# X. CONCLUSIONS DRAWN FROM THE STUDIES
The data from the non-clinical studies demonstrated acceptable analytical sensitivity and specificity, reproducibility, and stability of the AxSYM Anti-HCV when used according to the instructions for use as stated in the labeling, the warnings and precautions, and the Specimen Collection and Preparation for Analysis and Limitations section of the labeling.
The clinical studies in this application indicate that the AxSYM Anti-HCV assay is safe and effective when used according to the directions for use in the labeling. The sponsor has provided scientific data to support the utility of this assay in conjunction with other laboratory results and clinical information as an aid in the diagnosis of HCV associated disease.
# Risk Benefit Analysis
22
{17}
The nonclinical and clinical studies contained in this application demonstrated that the device is safe and effective as indicated. It can therefore be concluded that the benefits of the AxSYM Anti-HCV, when used according to the provided directions and in conjunction with other serological and clinical information outweigh any risk that may be associated with its use.
## XI. PANEL RECOMMENDATION
On January 21, 2000, the Microbiology Devices Advisory Panel recommended that the PMA be approved with conditions.
The following approval conditions were stipulated:
1. Inclusion of a narrative below the intended use statement that repeat reactives should be retested by a supplemental assay.
2. Inclusion in the package insert that performance characteristics data are descriptive of the population studies but are not intended to be representative of any particular disease states. Test results are intended to provide an assessment of antibody status but individual diagnosis should be made only in the context of clinical information and other laboratory parameters.
3. Sub-division of "Individuals at Increased Risk" populations by individual groups, that is, dialysis patients, etc.
4. Under the Interpretation of results section, inclusion of the following sentence: "A non-reactive anti-HCV result does not exclude infection with HCV".
## XII. CDRH DECISION
CDRH concurred with the panel's recommendation. The sponsor made the labeling changes as recommended by the Panel. CDRH issued an approvable letter dated March 24, 2000, pending the satisfactory compliance with the Quality System Regulation (21 CFR Part 820) as documented in the November 2, 1999, consent decree. The applicant’s manufacturing and control facilities were inspected on October 28-November 13, 2003, and the facilities were found to be in compliance with the Quality System Regulations on December 18, 2003. Abbott's AxSYM® Anti-HCV was approved on February 5, 2004.
## XIII. APPROVAL SPECIFICATIONS
Directions for Use: See attached labeling.
Hazards to Health from use of Device: See Indications, Contraindications, Warnings and Precautions and Adverse Events in the labeling.
Conditions of Approval: CDRH approval of this PMA is subject to full compliance with the conditions described in the approval order.
Post-approval Requirements and Restrictions: See Approval Order
23
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.