← Product Code [MGR](/productcode/MGR) · P950032

# APLIGRAF (GRAFTSKIN) (P950032)

_Organogenesis, Inc. · MGR · May 22, 1998 · General, Plastic Surgery · APCV_

**Canonical URL:** https://fda.innolitics.com/device/P950032

## Device Facts

- **Applicant:** Organogenesis, Inc.
- **Product Code:** [MGR](/productcode/MGR.md)
- **Decision Date:** May 22, 1998
- **Decision:** APCV
- **Device Class:** Class 3
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic, Expedited Review

## Indications for Use

Apligraf is indicated for use with standard therapeutic compression for the treatment of non-infected partial and full-thickness skin ulcers due to venous insufficiency of greater than 1 month duration and which have not adequately responded to conventional ulcer therapy.

## Device Story

Apligraf is a viable, bi-layered skin construct consisting of living human keratinocytes (epidermal-like layer) and human dermal fibroblasts (dermal-like layer) in a Type I bovine collagen matrix. Supplied as a 75 mm diameter circular disk in an agarose gel nutrient medium. Used in clinical settings by healthcare providers; applied to venous ulcers alongside standard therapeutic compression. Acts as a biological skin substitute; provides barrier function; secretes cytokines/matrix proteins to promote healing. Benefits patients by accelerating wound closure rates compared to compression therapy alone. Single-use; non-terminally sterilized; requires aseptic handling.

## Clinical Evidence

Prospective, randomized, multi-center trial (n=297). Primary endpoints: incidence of 100% wound closure per unit time and by 6 months. Apligraf + compression showed statistically significant improvement in closure per unit time (Cox’s Proportional Hazards, p=0.0223) vs. control + compression. 6-month closure rates: 55.4% (Apligraf) vs 49.1% (Control). Safety: Suspected wound infection higher in Apligraf group (29.2%) vs Control (14.0%). No evidence of immune rejection or antibody response to bovine collagen/human cellular components.

## Technological Characteristics

Bi-layered construct; Type I bovine collagen matrix; allogeneic human infant foreskin-derived fibroblasts and keratinocytes. Dimensions: 75 mm diameter, 0.75 mm thickness. Supplied in agarose gel medium (includes L-glutamine, insulin, transferrin, etc.). Aseptically manufactured; not terminally sterilized. Storage: 20-31°C in 10% CO2/air atmosphere.

## Reference Devices

- Neutral Allografts-Lack of Allogeneic Stimulation by Cultured Human Cells Expressing MHC Class I and Class II Antigens (Transplantation 1993;55:128-33)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a PMA](https://innolitics.com/services/regulatory/), [a 510(k)](https://innolitics.com/services/510ks/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

P950032

# SUMMARY OF SAFETY AND EFFECTIVENESS DATA

## I. GENERAL INFORMATION

DEVICE GENERIC NAME: Graftskin

DEVICE TRADE NAME: Apligraf™ (Graftskin)

APPLICANT: Organogenesis Inc.
150 Dan Road
Canton, MA 02021

PREMARKET APPROVAL APPLICATION (PMA): P950032

DATE OF PANEL RECOMMENDATION: January 29, 1998

DATE OF GMP INSPECTION: April 8, 1996

DATE OF NOTICE OF APPROVAL OF APPLICATION: May 22, 1998

EXPEDITED REVIEW: Expedited processing was authorized on May 30, 1995, based on the potential of Apligraf™ to provide a clinically important advance over existing alternatives in the treatment of chronic venous insufficiency ulcers.

## II. INTENDED USE / INDICATIONS

Apligraf is indicated for use with standard therapeutic compression for the treatment of non-infected partial and full-thickness skin ulcers due to venous insufficiency of greater than 1 month duration and which have not adequately responded to conventional ulcer therapy.

## III. DEVICE DESCRIPTION

Apligraf is a viable, bi-layered, skin construct, which contains Type I bovine collagen, extracted and purified from bovine tendons and viable allogeneic human fibroblast and keratinocyte cells isolated from human infant foreskin. Apligraf™ consists of two primary layers. The upper "epidermal-like" layer, formed of living human keratinocytes, has a well

{1}

differentiated stratum corneum which has been shown in *in vitro* experiments to provide a natural barrier to topical infection and wound desiccation. In the supporting “dermis-like” layer of Apligraf™, the major cell type is the fibroblast. Apligraf™ fibroblasts produce many of the matrix proteins found in human dermis, such as collagen type IV, tenascin, decorin, hyaluronate, and fibronectin. In addition, collagen type IV, laminin, laminin 5, heparin sulfate, proteoglycan, and β₄ integrin are present at the dermal-epidermal junction. Apligraf also expresses many of the cytokines found in human skin including PDGF-A, PDGF-B, TGFα, TGFβ₁, TGFβ₃, ECGF, FGF-1, FGF-2, FGF-7, IGF-1, IGF-2, CSF, IL-1α, IL-6, IL-8 and IL-11. Other cells found in human skin, Langerhans cells, melanocytes, macrophages and lymphocytes as well as secondary structures such as blood vessels and hair follicles are not present in Apligraf.

Apligraf is supplied ready to use, in a plastic container/carrier and is intended for single use only. This container protects and supports the product and provides a supply of agarose gel nutrient medium to maintain cell viability until use. The carrier is sealed in a heavy gauge polyethylene bag containing a 10% CO₂/air atmosphere. Apligraf is kept in the sealed bag at 20-31°C until use. Apligraf is supplied as a circular disk 75 mm in diameter. The thickness of the product is 0.75 mm. The agarose shipping medium contains agarose, L-glutamine, hydrocortisone/bovine serum albumin, bovine insulin, human transferrin, triiodothyronine, ethanolamine, O-phosphorylethanolamine, adenine, selenious acid, DMEM powder, HAM’s F-12 powder, sodium bicarbonate, calcium chloride and water for injection.

To maintain cell viability, the product is aseptically manufactured, but not terminally sterilized. Apligraf is shipped following a preliminary sterility test with a 48 hour incubation to determine the absence of microbial growth. Final (14 day incubation) sterility tests results are not available at the time of application.

Information concerning the following sections of this Summary of Safety and Effectiveness Data is included in the product labeling at the end of this document:

## IV. CONTRAINDICATIONS

- Apligraf is contraindicated for use on clinically infected wounds.
- Apligraf is contraindicated in patients with known allergies to bovine collagen.
- Apligraf is contraindicated in patients with a known hypersensitivity to the components of the Apligraf agarose shipping medium.

The warnings and precautions can be found in the Apligraf labeling.

Page 2

{2}

Page 3
# V. ALTERNATIVE PRACTICES AND PROCEDURES

Compression therapy is the standard of treatment for ulcers caused by venous disease. Surgical alternatives for venous ulcers include vein stripping, vein ligation and skin grafting.

# VI. POTENTIAL ADVERSE EFFECTS

A total of 297 patients (161 Apligraf, 136 active control) were evaluated for safety in a clinical trial for the treatment of venous ulcers. Adverse events were recorded as mild, moderate, severe or life-threatening.

There were 1 life-threatening and 3 severe infections reported in the Apligraf™ group and none in the control arm. Of these, two severe infections were considered related to treatment: however one occurred one month after the last application of Apligraf and the other occurred following application on a pre-existing *Pseudomonas* infection.

All reported adverse events which occurred in the Apligraf cohort in the pivotal clinical study at an incidence of 1% or greater are listed in Table 1. The adverse events are listed in descending order according to frequency.

{3}

Table 1
Adverse Events Reported in Greater than 1.0% of Apligraf Patients

|   | Apligraf
(n = 161) | Control
(n= 136)  |
| --- | --- | --- |
|   |  Total | Total  |
|  Suspected Wound Infection^{1} (study site) | 47 (29.2%) | 19 (14.0%)  |
|  Suspected Wound Infection^{1} (non-study site) | 16 (9.9%) | 15 (11.0%)  |
|  Cellulitis^{2} (study site) | 13 (8.1%) | 11 (8.1%)  |
|  Cellulitis^{2} (non-study site) | 12 (7.5%) | 7 (5.1%)  |
|  Dermatitis (non-study site) | 10 (6.2%) | 10 (7.4%)  |
|  Exudate (study site) | 9 (5.6%) | 0 (0.0%)  |
|  Peripheral Edema | 8 (5.0%) | 7 (5.1%)  |
|  Pain (study site) | 7 (4.3%) | 7 (5.1%)  |
|  Death | 6 (3.7%) | 6 (4.4%)  |
|  Skin Ulcer (non-study site) | 6 (3.7%) | 5 (3.7%)  |
|  Pain (non-study site) | 5 (3.1%) | 4 (2.9%)  |
|  Pruritus (non-study site) | 5 (3.1%) | 2 (1.5%)  |
|  Skin Ulcer (study site) | 5 (3.1%) | 3 (2.2%)  |
|  Infection (non-wound) | 4 (2.5%) | 1 (0.7%)  |
|  Positive Wound Culture^{3} (study site) | 4 (2.5%) | 3 (2.2%)  |
|  Rhinitis | 4 (2.5%) | 1 (0.7%)  |
|  Dermatitis (study site) | 4 (2.5%) | 2 (1.5%)  |
|  Pain (overall body) | 3 (1.8%) | 2 (1.5%)  |
|  Congestive Heart Failure | 3 (1.8%) | 0 (0.0%)  |
|  Accidental Injury (musculoskeletal) | 3 (1.8%) | 0 (0.0%)  |
|  Dyspnea | 3 (1.8%) | 1 (0.7%)  |
|  Pharyngitis | 3 (1.8%) | 0 (0.0%)  |
|  Rash (study site) | 3 (1.8%) | 2 (1.5%)  |
|  Accidental Injury (overall body) | 2 (1.3%) | 1 (0.7%)  |
|  Asthenia | 2 (1.3%) | 0 (0.0%)  |
|  Arrhythmia | 2 (1.3%) | 0 (0.0%)  |
|  Abscess (non-study site) | 2 (1.3%) | 0 (0.0%)  |
|  Arthralgia | 2 (1.3%) | 2 (1.5%)  |
|  Cough Increased | 2 (1.3%) | 0 (0.0%)  |
|  Rash (non-study site) | 2 (1.3%) | 5 (3.7%)  |
|  Erythema (study site) | 2 (1.3%) | 1 (0.7%)  |
|  Kidney Failure | 2 (1.3%) | 0 (0.0%)  |
|  Urinary Tract Infection | 2 (1.3%) | 5 (3.7%)  |

Page 4
12

{4}

In the clinical trial the following definitions were used:

¹Suspected Wound infection: a wound with at least some clinical signs and symptoms of infection such as increased exudate, odor, redness, swelling, heat, pain, tenderness to the touch and purulent discharge; quantitative culture was not required.

²Cellulitis: a non-suppurative inflammation of the subcutaneous tissues extending along connective tissue planes and across intercellular spaces; widespread swelling, redness and pain without definite localization.

³Positive wound culture: reported as an adverse event, but not reported as a wound infection.

## VII. MARKETING HISTORY

Apligraf™ is approved for marketing in Canada and has been commercially available since August 12, 1997.

## VIII. SUMMARY OF PRE-CLINICAL STUDIES

This section provides brief summaries of important preclinical tests performed on Apligraf followed by Table 2 which describes a number of non-clinical laboratory studies performed in the development and evaluation of Apligraf. Table 2 has been divided into studies chosen to evaluate the following categories: Development and Characterization, Immunology, Microbial and Toxicology studies. The studies reported here include a range of topics assessing safety, device attributes, practical aspects of device delivery, and potential clinical use issues.

Presence of Blood Group Antigens on Apligraf - DNA coding the Rh Factor was identified by PCR analysis of the cells used to make Apligraf for pivotal study. Weak, patchy staining of the B Blood Group antigen in the epidermal layer of this Apligraf was detected by immunohistochemical (IHC) analysis. No expression of the Rh antigen by Apligraf was observed in flow cytometry measurements.

Apligraf Compatibility with Antimicrobial Agents - In in vitro and in vivo histology studies, exposure to Dakin's solution, Mafenide Acetate, Scarlet Red Dressing, Tincoban, Zinc Sulfate, Povidone-iodine solution, or Chlorhexidine degraded the overall histology of Apligraf. Device exposure to Mafenide acetate, Polymixin/Nystatin or Dakin’s Solution also reduced Apligraf cell viability.

Karyology analyses of keratinocyte cells used in device manufacture revealed a limited number of chromosomal abnormalities. These same cells were not neoplastic in in vitro and in vivo assays.

The fibroblast and keratinocyte cells, from which Apligraf is manufactured, are from human infant foreskin tissue. Products made from human tissue may contain infectious

Page 5
13

{5}

agents. The risk that Apligraf will transmit a pathogenic agent is reduced by extensive testing. A comprehensive medical history of the mother was taken and blood of the mother was screened for HIV-1, HIV-2, HIV-p24, HTLV-1, HTLV-2, Hepatitis A, Hepatitis B surface, Hepatitis B core, Hepatitis B, Hepatitis C, Cytomegalovirus and Epstein Barr viruses. Additionally, human fibroblasts and keratinocytes used to form Apligraf are derived from cell banks which were tested for HIV-1, HIV-2, HIV-p24, HTLV-1, HTLV-2, Hepatitis B surface, Hepatitis C, Cytomegalovirus, Epstein-Barr virus, bacteria, fungi, yeast, mycoplasma, in vitro virus, in vivo virus, karyology, isoenzymes, virus by EM, Retrovirus by RT, Herpesvirus 6 and tumorigenicity. Product manufacture also includes reagents derived from animal materials including bovine pituitary extract. All animal-derived reagents are tested for viruses, retroviruses, bacteria, fungi, yeast, and mycoplasma before use and all bovine material is obtained from countries free of Bovine Spongiform Encephalopathy (BSE). To maintain cell viability, the product is aseptically manufactured, but not terminally sterilized. Apligraf is shipped following a preliminary sterility test with a 48 hour incubation to determine the absence of microbial growth. Final (14 day incubation) sterility tests results are not available at the time of application. The final product is also tested for morphology, cell viability, epidermal coverage, mycoplasma, and physical container integrity.

Page 6

{6}

Table 2
Apligraf™ Pre-Clinical Studies

|  Development & Characterization Studies  |   |
| --- | --- |
|  Study | Results/Conclusions  |
|  Cytokine and receptor analysis of Apligraf by RT-PCR | The cytokine and cytokine receptor mRNA profile of Apligraf and component cells were similar to human skin. However, hematopoietic cell derived cytokines (IL-2 and IL-4) were not detected. This is consistent with the use of pure HEP and HDF populations.  |
|  Apligraf: Determination of cell purity in HEP and HDF cell banks by flow cytometry | Results demonstrated that each HEP and HDF cell strain contained no detectable levels of professional APC (endothelial cells and Langerhan cells).  |
|  Determination of residual bovine serum proteins in Apligraf | NBCS in Apligraf G100 = 2.6 ± 0.07% total dry weight (4.5 mg per G100 unit).  |
|  Morphological development and maturation of Apligraf | Apligraf epidermis underwent a sequence of morphologic changes during development and maturation in vitro resulting in an organotypic skin culture with a morphology very similar to that of normal human skin.
Changes in morphology parallel biochemical and functional events. Established morphological characteristics serve as device specifications.  |
|  Effect of Apligraf development on graft performance in vivo | 100% graft take in mice; Apligraf epidermis remained throughout study; Basement membrane formed by day 15; TEM analysis confirmed the presence of the ultrastructural features of a differentiated epidermis.  |
|  Effect of Apligraf development on graft performance and barrier function formation in vivo | In vitro barrier function developed rapidly in mice between 14 and 20 days of culture. Apligraf (14d old) failed to integrate and persist on mice, while 16d old Apligraf persisted when grafted onto mice. The barrier function of the 16d old Apligraf grafted onto mice was slightly less than human skin. The barrier function of 20d old Apligraf grafted onto mice was comparable to human skin.  |
|  Immunology Studies  |   |
|  Neutral allograft study^{1} | HEPs and HDFs of Apligraf did not, but HUVECs did, stimulate T cell proliferation in a mixed lymphocyte reaction (MLR) assay.  |
|  Hu-SCID mouse study: Part I survival of Apligraf on Hu-SCID mice | Graft survival of Apligraf was significantly higher than human skin on hu-SCID mice (p < 0.05). After 28 days, 88% (n=7/8) of Apligraf grafts integrated and persisted on hu-SCID. In contrast, after 14 days, only 28% (n=2/7) of the human skin grafts persisted on hu-SCID mice. The survival of Apligraf and human skin on control SCID mice was not significantly different.  |
|  Hu-SCID mouse study: Part II survival of MHC class-II^{+} Apligraf on Hu-SCID mice. | The persistence of IFN-γ treated Apligraf on hu-SCID mice (100% survival; n=9/9) was equivalent to the percent survival of untreated Apligraf on hu-SCID mice (100% survival; n=10/10).  |
|  Regulation of T cell proliferation by keratinocyte derived soluble factors: part I | HEPs produce soluble factors that significantly inhibit the proliferation of anti-CD3 activated T cells.  |
|  Regulation of T cell proliferation by keratinocyte derived soluble factors: part II | HEPs produce soluble factors that significantly inhibit the proliferation of allogeneic T cells.  |

Page 7

{7}

Table 2 (cont.)
Apligraf™ Pre-Clinical Studies

|  Immunology Studies (cont.)  |   |
| --- | --- |
|  Study | Results/Conclusions  |
|  Identification of keratinocyte-derived T cell inhibitory factor | HEP inhibition of T cell proliferation did not require cell contact, was inducible in the presence of FBS, and could be partially blocked by addition of indomethecin or anti-TGF-β Mab. These results suggest that HEPs can regulate the response to antigen presented by other APC through the production of soluble factors.  |
|  Microbial Studies  |   |
|  Can Apligraf act as a barrier to topical infection? | No evidence of bacterial penetration through the Apligraf was seen in a system where bacteria were seeded on the device supported on a membrane permeable to bacteria above sterile bacterial growth medium.  |
|  Toxicology Studies  |   |
|  General Safety Test | Apligraf is non-toxic.  |
|  Primary Skin Irritation Study | No reactivity. Apligraf was scored as a non-irritant.  |
|  Kligman Maximization Study (sensitization assays) | No reactivity. Apligraf showed no primary irritancy response.  |
|  Tissue Culture-Agar Diffusion Test (cytotoxicity) | No reactivity. Apligraf met the requirements of the Agar Diffusion Test, USPXXII.  |
|  Systemic Injection Test | No reactivity. Apligraf is non-toxic.  |
|  Intracutaneous Test | No reactivity. Apligraf is non-toxic.  |
|  Hemolysis Test | No reactivity. Apligraf is non-hemolytic.  |
|  Subcutaneous Injection Test-Subchronic Toxicity | Apligraf caused a significant response when tested in albino rabbits. The protocol was designed for plastics or relatively inert materials. The validity of the test was compromised by the nature of Apligraf. Therefore, this test is not considered a valid measurement of toxicity for Apligraf.  |

APC: antigen presenting cell; HDF: human dermal fibroblast; HEP: human epidermal keratinocyte; RT-PCR: reverse transcriptase-polymerase chain reaction; NBCS: new born calf serum; TEM: transmission electron microscopy; IFN-γ: interferon-gamma; SCID: severe combined immunodeficient; MHC: major histocompatibility complex; HUVEC: human vascular endothelial cells; FBS: fetal bovine serum.

{8}

Page 9
(7)

# IX. SUMMARY OF THE RESULTS OF THE CLINICAL INVESTIGATION

The following is a summary of the large scale study designed to support approval, “Protocol 92-VSU-001, “Multi-Center Parallel Group Controlled Clinical Trial to Determine the Efficacy and Safety of Apligraf in the Treatment of Chronic Venous Insufficiency Leg Ulcers”.

## Study Design:

A prospective, randomized, controlled, multi-center, multi-specialty, unmasked study was conducted to evaluate the safety and effectiveness of Apligraf and compression therapy in comparison to an active treatment concurrent control of zinc paste gauze and compression therapy. The study population included consenting patients who were 18-89 years old, available for one year follow-up, with venous insufficiency confirmed by plethysmography (venous reflux &lt; 20 sec.); associated with non-infected partial and / or full thickness skin loss ulcer (IAET Stage 2 or 3) of greater than one month duration and which had not adequately responded to conventional ulcer therapy. Patients were excluded for ankle brachial index &lt; 0.65, severe rheumatoid arthritis, collagen vascular disease, pregnancy/lactation, cellulitis, osteomyelitis, ulcer with necrotic, avascular or bone/tendon/fascia exposed-bed, clinically significant wound healing impairment due to uncontrolled diabetes, or renal, hepatic, hematologic, neurologic or immune insufficiency or due to immunosuppressive agents such as corticosteroids (&gt; 15 mg/day), radiation therapy or chemotherapy; or enrollment in studies within the past 30 days for investigational devices or within the past three months for investigational drugs related to wound healing.

Extremities with multiple ulcers were enrolled; however, only one ulcer per extremity was studied. Non-study ulcer care was not specifically defined. Study ulcer care was defined for the treatment (Apligraf and compression therapy) and control (zinc paste gauze and compression therapy), treatment groups in two phases:

1) Active Phase (0-8 weeks): All patients received: i) a non-adherent, ii) a non-occlusive and iii) a therapeutic compression dressing on day 0, mid-week during the first week (day 3-5), and at weeks 1-8. Control treated patients also received zinc impregnated gauze at each visit. All Apligraf patients received Apligraf on day 0. At the day 3-5 and weeks 1, 2 and 3 visits, if less than 50% Apligraf take was observed, then patients received an additional application of Apligraf. Patients were not allowed to receive more than 5 Apligraf applications total.

2) Maintenance Phase (8-52 weeks): Closed-ulcer extremities received non-specified elastic compression stockings. Open-ulcer extremities continued with dressing changes.

{9}

# Study Endpoints:

The primary study endpoints were: 1) the incidence of 100% wound closure per unit time and 2) the overall incidence of 100% wound closure by 6 months. “Complete Wound Closure” was defined as full epithelialization of the wound with the absence of drainage. “Epithelialization” was defined as a thin layer of epithelium visible on the open wound surface. Secondary endpoint measurements included: the incidence of ulcer recurrence, duration of wound closure, immune responses against the human cellular and bovine device components and analyses of changes in ulcer: depth (IAET staging), erythema, edema, wound pain, fibrin, exudate, granulation tissue and overall assessment from baseline visit to the 6 month visit.

# Listing of Study Centers and Patient Treatment Group Assignment:

The study enrollment is displayed below in Table 3.

|  Investigator & Center | Total # of Patients at a site  |
| --- | --- |
|  1. Gerit Mulder, Denver, CO | 55  |
|  2. Oscar Alvarez, New Brunswick, NJ | 50  |
|  3. Frank Maggiacomo, Providence, RI | 41  |
|  4. Morton Altman, San Francisco, CA | 24  |
|  5. Duyen Faria, Detroit, MI | 22  |
|  6. Vincent Falanga, Miami, FL | 19  |
|  7. James Snyder, Las Vegas, NV | 18  |
|  8. Thomas Garland, Lawrenceville, NJ | 17  |
|  9. George Mueller, San Diego, CA | 16  |
|  10. Steven Bowman, Clearwater, FL | 11  |
|  11. David Margolis, Philadelphia, PA | 7  |
|  12. Thomas Schnitzer, Chicago, IL | 6  |
|  13. Arnold Luterman, Mobile, AL | 6  |
|  14. Marketa Limova, San Francisco, CA | 4  |
|  15. John Hansbrough, San Diego, CA | 1  |
|  TOTAL | 297  |

# Notes

*The product effectiveness dataset excluded the results from all patients treated at one clinical site, because FDA audit raised sufficient concerns about the reliability of the clinical records at this site. Consequently, the clinical outcome of these patients was excluded from study effectiveness analyses, but was included in all safety analyses. The dataset for product effectiveness included a total of 240 patients, i.e., 130 Apligraf and 110 Control patients.

{10}

# Results:

## Baseline Demographics:

The baseline demographics in both the Apligraf and Control arms were comparable for gender, race, age and ulcer area. Ulcer size was a little larger and longer in duration, (but not significantly) in the Apligraf treatment arm as displayed in Table 4.

## Study Drop-outs:

The discontinuation rate for all patients prior to the 6 month evaluation was 76/291 (26%) and 105/291 (36%) at 12 months. Within the safety cohort, 59 Apligraf and 50 Control patients discontinued prior to 12 month visit.

## Intent-to-treat Analyses of Ulcer Healing:

Apligraf use with standard therapeutic compression provided a statistically significant improvement in the incidence of ulcer closure per unit time for all patients enrolled in the effectiveness cohort when compared to control therapy in a Cox’s Proportional Hazards Regression Analysis. The incidence of wound closure by 6 months was numerically superior, but not statistically significantly improved, in Apligraf-treated patients.

![img-0.jpeg](img-0.jpeg)
Chart 1

![img-1.jpeg](img-1.jpeg)
Chart 2

The incidence of wound closure at set visits up to 6 months presented as the raw data results (Chart 1) and the results after adjustment for pooled center, baseline ulcer duration and baseline area (Chart 2).

Page 11

{11}

Page 12
20

## Incidence of Closure per Unit Time

In a Kaplan-Meier life table analysis, median times of 140 and 181 days were calculated for when 50% of the Apligraf and Control patients achieved wound closure, respectively, (p=0.3916). A Cox’s Proportional Hazards Regression Analysis of these data determined that the covariables of pooled center, duration of ulcer and ulcer area had significant effects on the time to 100% wound closure for all patients. Adjusted median times to closure from this analysis were 99 and 184 days for Apligraf and Control patients, respectively.

## Incidence of 100% Wound Closure

The overall closure rate was 55.4% (72/130) for Apligraf and 49.1% (54/110) for Control patients by 6 months (p= 0.365 by a Fisher’s Exact 2-tailed test, Chart 1). In a logistic regression analysis of these data, the covariables of pooled center, baseline ulcer duration and baseline ulcer area were found to impact 100% wound closure for all patients. A logistic regression analysis, which adjusted for these factors, predicted that 58.8% of Apligraf and 44.0% of Control patients would achieve ulcer closure by six months (p = 0.0530). In a Cox’s Regression Analysis, which accounted for the healing pattern over the six month timeline, closure rates of 56.8% and 39.8% by 24 weeks were predicted for Apligraf and Control patients, respectively, (p=0.0223, Chart 2).

## Duration of Wound Closure

In this analysis once a patient achieved wound closure, the patient was judged a treatment success even if the duration of wound closure was short. The durability of complete wound closure was calculated from the first and last study days in which a Wound Closure case report form (CRF) was checked closed. In this analysis, the mean number of days for wound closure for patients who attained complete closure by 6 months and completed the study was 233 days for Apligraf patients and 219 days for Control patients. Similarly, the mean number of days of ulcer closure for all patients showed no significant differences for Apligraf (190 days) and Control (182 days) patients.

## Correlation between photographs and Case Report Form records of wound closure

was evaluated in a masked review with two evaluators of 437 study photographs composed of: 1) photographs of the baseline, time of first report of healing and the 6 month study visit for all 126 patients whose wound closure CRF was checked closed and 2) photographs at baseline, study week 8 and study month 6 of 20 Apligraf and 20 Control patients randomly selected from the 114 non-healing patients in the effectiveness cohort. These photographs, which represent 166 of the 240 patients in the effectiveness cohort, were randomly ordered to reduce bias or unmasking that might have resulted from a sequential ordering of the photographs. Results of this analysis revealed a good correlation between the data in Case Report Forms and the two reviewers with Kappa statistics ranging from 0.711 to 0.781.

## Revised Effectiveness Cohort

Analysis of a revised dataset that selected only patients who met the precise study inclusion and exclusion criteria was performed. In this subset, the results of 32 patients were excluded from the intent-to-treat population because either: 1) they were over 85 years old, 2) their ulcers were not believed to be of non-venous etiology or 3) their ulcers were not of an appropriate size. The results of two additional Apligraf patients’ ulcers were also switched from closed to open at 6 months after an FDA review of

{12}

clinical photographs. The improvement observed in Apligraf-treated over control treated patients in the incidence of ulcer closure per unit time remained statistically significant for this cohort (n=208).

Ulcer recurrence - At six months, the incidence of ulcer recurrence was 8.3% (6/72) for Apligraf- and 7.4% (4/54) for control-treated patients. The incidence of ulcer recurrence by 12 months was 18.1% (13/72) in the Apligraf group and 22.2% (12/54) in the control group.

Baseline status impact on wound closure - The impact of patient baseline status on wound closure was evaluated for the patients above and below the median values for ulcer duration and ulcer size as well as for baseline IAET Ulcer Stage, the presence of diabetes and a patient's Ankle Brachial Index. In these analyses, Apligraf use with standard compression provided statistically significant improvements in both: 1) the incidence of ulcer closure per unit time and 2) the incidence of ulcer closure by 6 months for patients with baseline ulcer durations greater than one year at baseline. The impact of baseline status impact on wound closure for different subgroups is displayed in Table 4.

Table 4
Pre-Treatment Status and Wound Closure
Effectiveness Cohort (n=240 patients)

|   | Pre-Treatment Status |   | Number and Percent of Wound Closure by 6 months  |   |
| --- | --- | --- | --- | --- |
|   |  No. and (%) Apligraf Pts. | No. and (%) Control Pts. | Apligraf | Control  |
|  |   |   |   |   |
|  Total | 130 Patients | 110 Patients | 72/130 (55.4%) | 54/110 (49.1%)  |
|  |   |   |   |   |
|  Ulcer Duration  |   |   |   |   |
|  ≤ 1 year | 58 (44.6%) | 62 (56.3%) | 38/58 (65.5%) | 45/62 (72.6%)  |
|  > 1 year | 72 (55.4%) | 48 (43.6%) | 34/72 (47.2%) | 9/48 (18.8%)  |
|  *Ulcer Area  |   |   |   |   |
|  < 500 mm² | 65 (50.0%) | 60 (54.5%) | 45/65 (69.2%) | 35/60 (58.3%)  |
|  > 500 mm² | 63 (48.5%) | 50 (45.5%) | 26/63 (41.3%) | 19/50 (38.0%)  |
|  IAET Staging  |   |   |   |   |
|  Stage II | 63 (48.5%) | 56 (50.9%) | 34/63 (54.0%) | 32/56 (57.1%)  |
|  Stage III | 67 (51.5%) | 54 (49.1%) | 38/67 (56.7%) | 22/54 (40.7%)  |
|  Diabetes  |   |   |   |   |
|  Yes¹ | 25 (19.2%) | 11 (10.0%) | 12/25 (48.0%) | 4/11 (36.4%)  |
|  No | 105 (80.8%) | 99 (90.0%) | 60/105 (57.1%) | 50/99 (50.5%)  |
|  **Ankle Brachial Index data (ABI)  |   |   |   |   |
|  > 0.65 - < 0.8 | 9 (6.9%) | 10 (9.1%) | 4/9 (44.4%) | 4/10 (40.0%)  |
|  >0.8 - <1.0 | 43 (33.1%) | 50 (45.5%) | 26/43 (60.5%) | 27/50 (54.0%)  |
|  >1.0 | 75 (57.7%) | 49 (44.5%) | 40/75 (53.3%) | 22/49 (44.9%)  |

{13}

*Baseline ulcer area missing for two patients in the Apligraf group
**ABI data is missing for 3 Apligraf and 1 control patient
¹ This category includes both insulin-dependent and non-insulin dependent diabetes patients, because the insulin-dependence of patients was not determined in this clinical trial

Secondary effectiveness endpoints - Changes in ulcer depth (IAET staging), erythema, edema, wound pain, fibrin, exudate, granulation tissue and overall assessment were assessed. Statistically significant differences between treatment groups were found for wound exudate at day 3-5 and week 2. The Apligraf group experienced earlier improvement in fibrin, while the Control group showed earlier improvement in wound exudate. While both treatment arms showed statistically significant improvement in all clinical parameters and patient overall assessments when comparing the baseline and 6 month visits, no statistically significant differences between treatment groups were observed at the 6 month visit.

Gender and Wound Closure - 36/70 (51%) of the men and 36/60 (60%) of the women in the Apligraf treatment group achieved wound closure by six months. In the Control group 19/52 (37%) of the men and 35/58 (60.8%) of the women in the Control treatment arm achieved wound closure by six months. The distribution of men (36.5%) and women (60.3%) attaining 100% wound closure in the Control arm was statistically significant by a Fisher's Exact 2-tail test (p=0.014).

## Device Safety

### Study Withdrawals:

13 patients withdrew from Study 92-VSU-001 due to adverse events or intercurrent illness. Per treatment arm the division was 5 Apligraf patients (1 male and 4 females) and 8 Control patients (4 males and 4 females).

### Adverse events: Are displayed in section VI.

Suspected wound infection at the study ulcer: - 47/161 Apligraf (29.2%) and 19/136 (14.0%) Control patients had reports of localized suspected wound infections at the study site as defined by a wound with at least some clinical signs and symptoms of infection such as redness, swelling, heat, pain, tenderness to the touch and purulent discharge. The difference between treatment arms was significant (p=0.002) for all wound infections and non-significant (p=0.190) for wound infections judged to be device related. Overall there were 12/46 (26.1%) and 6/18 (33.3%) suspected wound infections judged as related to Apligraf and Control treatments, respectively. There were 1 life-threatening and 3 severe infections in the Apligraf group and none in the control arm. While the life threatening infection was judged as unrelated to device application, two of three severe infections were judged as Apligraf treatment-related. One Control (and no Apligraf) patient was hospitalized for infection at the study ulcer.

Page 14

{14}

Because quantitative wound culture was not performed routinely in the study, the true incidence of wound infection associated with Apligraf use remains unknown. Diagnosis of wound infection may be complicated by the white or yellow appearance of Apligraf after it becomes hydrated with wound fluid.

## Immune response:

In tests of patients’ sera there were no observations of antibody responses against bovine type I collagen, bovine serum proteins or the Class I HLA antigens on human dermal fibroblasts and human epidermal cells. T-cell specific responses were not observed against bovine type I collagen, human fibroblasts or human keratinocytes. There was also no clinical evidence of Apligraf rejection by any patient.

## X. CONCLUSIONS DRAWN FROM THE STUDY

This study provides reasonable assurance of the safety and effectiveness of Apligraf with standard therapeutic compression for the treatment of non-infected partial and full-thickness skin ulcers due to venous insufficiency of greater than 1 month duration and which have not adequately responded to conventional ulcer therapy. This study demonstrated that:

- Apligraf provides a statistically significant advantage in the incidence of wound closure per unit time when used with standard therapeutic compression for the treatment of non-infected partial and full-thickness skin ulcers due to venous insufficiency of greater than 1 month duration and which have not adequately responded to conventional ulcer therapy. The incidence of wound closure by 6 months was numerically superior, but not statistically significantly improved in patients treated with Apligraf.

- In the controlled clinical study conducted in patients with ulcers due to venous insufficiency of greater than one month in duration, suspected infection was reported more frequently in Apligraf-treated (29.2%) than control patients (14.0%). There were 1 life-threatening and 3 severe infections in the Apligraf group and none in the control arm.

- There were no observations of antibody responses against bovine type I collagen, bovine serum proteins or the Class I HLA antigens on human dermal fibroblasts, and human epidermal cells. T-cell specific responses were also not observed against bovine type I collagen, human fibroblasts or human keratinocytes.

## XI. PANEL RECOMMENDATION

On January 29, 1998, the General and Plastic Surgery Devices Panel recommended approval without conditions of Organogenesis’ PMA for Apligraf. In these discussions the Panel agreed that the definition of wound healing used in the pivotal study, (i.e., full epithelialization of the wound with the absence of drainage, where epithelialization was

Page 15
23

{15}

defined as a thin layer of epithelium visible on the open wound surface) was consistent with the definition of a “healed” ulcer.

## XII. CDRH ACTION

Expedited processing was authorized on May 30, 1995, based on the potential of Apligraf™ to provide a clinically important advance over existing alternatives in the treatment of chronic venous insufficiency ulcers.

Inspection of the sponsor’s manufacturing facilities was completed on April 8, 1996 and was found to be in compliance with the device Good Manufacturing Practice regulations.

After the Panel meeting, FDA completed review of preclinical testing and product manufacturing issues. These issues involved assessing the purity and composition of device components and manufacturing reagents. In specific:

1) It was determined that the use of bovine pituitary extract (obtained from a BSE-free country) in the keratinocyte growth media should be identified in the device description.

2) The keratinocyte cells in this device were found to weakly express the B Blood Group antigen, but not the Rh antigen. Fibroblasts did not express either antigen. These results are consistent with the scientific literature on cultured skin products. Based on the results of studies with Apligraf, the extensive clinical use of cadaver skin and the scientific literature on cultured skin products, the weak expression of Blood Group antigens on the device was not believed to be clinically significant.

3) The purity of transferrin used in device manufacture was reviewed and determined to be safe. Submission of formal documentation about the methods of inactivating blood-borne viruses during preparation this plasma-derived reagent was requested as a condition of product approval.

4) Karyology analyses of keratinocyte cells used in device manufacture revealed a limited number of chromosomal abnormalities. These same cells were not neoplastic in *in vitro* and *in vivo* assays. As a condition, of approval the sponsor was requested to evaluate these findings with respect to the longevity of Apligraf cells on venous ulcer patients and the karyology, morphology and neoplastic potential of all keratinocyte and fibroblast cell lines used in the manufacture of future commercial products.

FDA issued an approval order on May 22, 1998.

Page 16
24

{16}

Page 17
25

# APPROVAL SPECIFICATIONS

Directions for Use: See product labeling.

Postapproval Requirement and Restrictions:

1. Regarding the purity of the transferrin used in device manufacture, data documenting the viral inactivation properties of the processing procedures used by your supplier will be submitted to FDA immediately after receipt of these data by Organogenesis, Inc. from the supplier.

2. The significance of the karyology data observed on the keratinocyte and fibroblast cells used in product manufacture needs to be further evaluated. Please submit within one month of an approval order for this product the following protocols:

a. A protocol designed to determine the longevity of Apligraf cells on patients with venous ulcers.

b. Protocols for evaluating the karyology, morphology and neoplastic potential of all keratinocyte and fibroblast cell lines that will be used in future commercial products. Such data should include evaluations at both the MWCB stage and a cell stage that is as close to cellular senescence as possible. These evaluations should not only quantitate the extent of chromosomal changes, but also look for specific markers known to predict neoplastic transformation of keratinocyte cells. All such analyses should be performed in a manner consistent with the methods published in “Report of Ad Hoc Committee on Karyological Control of Human Cell Substrates,” *J. of Biol. Standard.* 1979, 7, 397-404 or a justification should be supplied.

{17}

Page 18
26

# REFERENCE

1. Theobald VA, Lauer JD, Kaplan FA, Baker KB, Rosenberg M. Neutral Allografts-Lack of Allogeneic Stimulation by Cultured Human Cells Expressing MHC Class I and Class II Antigens. Transplantation 1993;55:128-33.

---

**Source:** [https://fda.innolitics.com/device/P950032](https://fda.innolitics.com/device/P950032)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a PMA](https://innolitics.com/services/regulatory/), [a 510(k)](https://innolitics.com/services/510ks/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
