SYNVISC ONE

P940015 · Sanofi Genzyme Corp. · MOZ · Aug 8, 1997 · Orthopedic

Device Facts

Record IDP940015
Device NameSYNVISC ONE
ApplicantSanofi Genzyme Corp.
Product CodeMOZ · Orthopedic
Decision DateAug 8, 1997
DecisionAPPR
Device ClassClass 3
AttributesTherapeutic, Real-World Evidence

Real-World Evidence

SubmissionDeviceSponsorRWD SourcesRWE Use SummaryKey Tags
P940015 · Aug 8, 1997SYNVISC ONESanofi Genzyme Corp.Postmarket clinical use reports (Canada and Sweden)Postmarket surveillance data from Canada and Sweden were used to characterize the safety profile of the device, specifically identifying adverse events such as pain, swelling, effusion, and rare systemic reactions.postmarket surveillance; adverse events; clinical use

Clinical Evidence

Study DesignPopulationComparatorKey Endpoints
Clinical use in Canada and Sweden; Postmarket surveillance / observational clinical use; Follow-up/Duration: Since 1992 (Canada) and 1995 (Sweden); Study Period: 1992-1997Patients treated with Synvisc in clinical practice; Number of Sites: Not applicableNot applicable for this studySafety and adverse event reporting

Indications for Use

Synvisc® is indicated for the treatment of pain in osteoarthritis (OA) of the knee in patients who have failed to respond adequately to conservative non-pharmacologic therapy and simple analgesics (e.g., acetaminophen).

Device Story

Synvisc (hylan G-F 20) is an elastoviscous fluid derived from chicken combs; contains hylan A and hylan B gel in buffered saline. Administered via intra-articular knee injection by a physician. Acts as a lubricant and shock absorber in the joint. Supplied in 2.25 mL glass syringes containing 2.0 mL of sterile, nonpyrogenic fluid. Used in clinical settings for patients with chronic idiopathic OA of the knee. Output is the physical presence of the elastoviscous fluid within the synovial space, which provides pain relief and improved joint function. Clinical benefit includes reduction of knee pain for up to 12 weeks, potentially serving as an alternative to NSAIDs or surgical intervention.

Clinical Evidence

Seven clinical trials (738 subjects) evaluated safety and effectiveness. Studies #1, #2, #3, #5, #6, and #7 were double-blind. Primary endpoints were pain reduction on 0-100 mm VAS scales. Studies #2 and #3 demonstrated Synvisc superiority over saline control (p < 0.05) for up to 12 weeks. Study #5 showed Synvisc was not more effective than arthrocentesis alone in the general population, but showed significant improvement in a subpopulation of 'flare' subjects. Study #6 showed Synvisc was at least as effective as NSAID therapy. Adverse events included knee pain/swelling (2.3% of injections).

Technological Characteristics

Elastoviscous fluid; hylan polymers (hylan A and hylan B) derived from chicken combs; buffered physiological sodium chloride solution (pH 7.2). Elasticity (G') 111 ± 13 Pa; viscosity (G") 25 ± 2 Pa at 2.5 Hz. Supplied in 2.25 mL glass syringe (2.0 mL fill). Sterile and nonpyrogenic.

Indications for Use

Indicated for treatment of pain in osteoarthritis (OA) of the knee in patients who have failed to respond adequately to conservative non-pharmacologic therapy and simple analgesics.

Reference Devices

Submission Summary (Full Text)

{0} SUMMARY OF SAFETY AND EFFECTIVENESS DATA # I. GENERAL INFORMATION DEVICE GENERIC NAME: Hylan G-F 20 DEVICE TRADE NAME: Synvisc® APPLICANT'S NAME: Biomatrix, Incorporated 65 Railroad Avenue Ridgefield, NJ 07657 PREMARKET APPROVAL (PMA) APPLICATION NUMBER: P940015 DATE OF PANEL RECOMMENDATION: November 20, 1996 DATE OF NOTICE OF APPROVAL TO THE APPLICANT: August 8, 1997 # II. INDICATIONS FOR USE Synvisc® is indicated for the treatment of pain in osteoarthritis (OA) of the knee in patients who have failed to respond adequately to conservative non-pharmacologic therapy and simple analgesics (e.g., acetaminophen). # III. DEVICE DESCRIPTION Synvisc® (hylan G-F 20) is an elastoviscous fluid containing hylan polymers produced from chicken combs. Hylans are derivatives of hyaluronan (sodium hyaluronate), a natural complex sugar of the glycosaminoglycan family. Hyaluronan is a long-chain polymer containing repeating disaccharide units of Na-glucuronate-N-acetylglucosamine. Synvisc® contains hylan A (average molecular weight 6,000,000) and hylan B hydrated gel in a buffered physiological sodium chloride solution, pH 7.2. Synvisc® has an elasticity (storage modulus G') at 2.5 Hz of 111 ± 13 Pascals (Pa) and a viscosity (loss modulus G") of 25 ± 2 Pa (elasticity and viscosity of knee synovial fluid of 18-27 year old humans measured with a comparable method at 2.5 Hz: G' = 117 ± 13 Pa; G" = 45 ± 8.2 Pa). Synvisc® is supplied in a 2.25 mL glass syringe containing 2.0 mL Synvisc®. The contents of the syringe are sterile and nonpyrogenic. Each syringe of Synvisc® contains: | Hylan polymers (hylan A + hylan B) | 16 mg | | --- | --- | | Sodium chloride | 17 mg | | Disodium hydrogen phosphate | 0.32 mg | | Sodium dihydrogen phosphate monohydrate | 0.08 mg | | Water for injection | q.s. to 2.0 mL | {1} Page 2 Information concerning the following sections of this Summary of Safety and Effectiveness Data is included in the product labeling at the end of this document: ## IV. CONTRAINDICATIONS - Do not administer to patients with known hypersensitivity (allergy) to hyaluronan (sodium hyaluronate) preparations. - Do not inject Synvisc® in the knees of patients having knee joint infections or skin diseases or infections in the area of the injection site. Precautions and warnings can be found in the device labeling. ## V. ALTERNATIVE PRACTICES AND PROCEDURES For patients who have failed to respond adequately to conservative nonpharmacologic therapy and simple analgesics, e.g., acetaminophen, alternative therapies to Synvisc® include nonsteroidal anti-inflammatory drugs (NSAIDs); intraarticular injections of corticosteroids or injections of unmodified hyaluronan (sodium hyaluronate). For patients who have failed the above treatments, surgical interventions such as arthroscopic surgery and total knee replacement surgery are also alternative treatments. ## VI. POTENTIAL ADVERSE EFFECTS A total of 511 subjects (559 knees) received 1771 injections in seven clinical trials of Synvisc®. There were 37 reports in 35 subjects (2.3% of injections, 7% of subjects) of knee pain and/or swelling after these injections. Ten subjects (10 knees) were treated with arthrocentesis and removal of joint effusion. Two additional subjects (two knees) received treatment with intra-articular steroids. Two subjects (two knees) received NSAIDs. One of these subjects also received arthrocentesis. One subject was treated with arthroscopy. The remaining subjects with adverse events localized to the knee received no treatment or only analgesics. Systemic adverse events occurred in 10 (2.0%) of the Synvisc® treated subjects. There was one case each of rash (thorax and back) and itching of the skin following Synvisc® injections in these studies. These symptoms did not recur when these subjects received additional Synvisc® injections. The remaining generalized adverse events reported were calf cramps, hemorrhoid problems, ankle edema, muscle pain, tonsillitis with nausea, tachyrhythmia, phlebitis with varicosities and low back sprain. In three concurrently controlled clinical trials with a total of 112 subjects who received Synvisc® and 110 subjects who received either saline or arthrocentesis, there were no statistically significant differences in the {2} Page 3 numbers or types of adverse events between the group of subjects that received Synvisc® and the group that received control treatments. In clinical use in Canada (since 1992) and Sweden (since 1995), the most common adverse events reported have been pain, swelling, and/or effusion in the injected knees. Other adverse events reported were one case each of: generalized urticaria; recurring small hives; pain on one side of the body with nausea, anxiety and listlessness; facial flush with swelling of lips; nausea with dizziness; shivering with headache, nausea, respiratory difficulties; and prickling in body which did not recur after subsequent Synvisc® injections. No cases of anaphylaxis or anaphylactoid reactions have been reported. No deaths have been associated with the use of Synvisc®. Intra-articular infections did not occur in any of the clinical trials, but have occurred in clinical use following Synvisc® injections. ## VII. MARKETING HISTORY Synvisc® has been manufactured and marketed in Canada since November, 1992, and in Sweden since June, 1995. CE Mark Approval was granted in the European Community in November, 1995. Synvisc® has not been withdrawn from marketing in any countries. ## VIII. SUMMARY OF PRECLINICAL STUDIES Synvisc® (hylan G-F 20) and each of its two components (hylan A and hylan B) were studied in nonclinical laboratory studies to characterize their biological properties and to ensure safety. **Irritation tests:** Hylan A, hylan B and Synvisc® (hylan G-F 20) were each evaluated in intracutaneous toxicity studies in NZW rabbits using the USP method. A volume of 0.2 ml was injected into each of five intradermal sites. Injection sites were examined 24, 48, and 72 hours post-injection for erythema and edema. No evidence of significant irritation or of other signs of local toxicity was observed in any of the studies. It was concluded that neither Synvisc® nor its two components induced irritation. **Sensitization and immunogenicity assays:** Dermal sensitization potential was evaluated for hylan A, hylan B and Synvisc® using the test described by Magnusson and Kligman, 1970. Following two induction phases, each of 10 treated guinea pigs were challenged with undiluted test material (hylan A, hylan B or Synvisc®) using a Hill Top chamber. At 24, 48, 72 and 96 hours, the dermal reaction was evaluated according to Draize criteria. Synvisc®, hylan A and hylan B did not induce detectable dermal sensitization under the test conditions. {3} Page 4 The immunogenicity of Synvisc® was also evaluated in primates. Twenty owl monkeys received weekly bilateral intra-articular injections of Synvisc® for six months. Monthly dermal challenge testing for up to 31 weeks produced negative skin test results indicative of an absence of cell-mediated immunity. Serum analysis by ELISA, carried out monthly through six months and at one year, gave no indication of humoral immunity. **Cytotoxicity:** Hylan A, hylan B and Synvisc® were assessed for cytotoxicity using *in vitro* methods (MEM elution method, current USP procedure) in L929 mouse fibroblast cell line. None of the hylans produced cell lysis or cytotoxicity under the conditions tested. **Acute systemic toxicity:** Hylan A, hylan B and Synvisc® were evaluated for systemic toxicity in mice using the USP procedure. A dose of 50 ml per kg of test article was administered intraperitoneally to each animal. Mice were observed for adverse reactions at four, 24, 48 and 72 hours. None of the test articles appeared to be systemically toxic in mice. Eight studies of acute toxicity were conducted in NZW rabbits following intra-arterial, intra-articular, subcutaneous and subconjunctival administration of hylan preparations. Mild to moderate focal or multifocal inflammations of the synovial membranes were observed seven days after intra-articular injections of the hylans in one study. These were not found in joints studied four weeks after such injections. **Hemocompatibility:** The hemocompatibility of hylan A, hylan B and Synvisc® was evaluated in three *in vitro* hemolysis studies in which the USP direct contact method was used. Hylan A, hylan B and Synvisc® did not show any hemolytic activity in these studies. Eleven studies were carried out to evaluate the effect of hylan A, hylan B and Synvisc® on platelet function and blood clotting. None of the studies produced evidence that Synvisc® and its components (hylan A and hylan B) interfere with platelet function or with the clotting mechanism. **Pyrogenicity:** In the rabbit pyrogen test (USP test for material-mediated pyrogenicity), Synvisc® (reduced elastoviscosity) was injected intravenously into NZW rabbits in a single dose of 10 ml/kg of diluted test article. The body temperature of the rabbits was then measured during a three-hour observation period. No elevations in body temperature were observed in rabbits injected with the Synvisc® preparation or with samples taken from 14 batches of hylan A and from five batches of hylan B. **Tissue Implantation:** Seven-day and 30-day intramuscular implantation studies were conducted in rabbits using hylan A, hylan B or Synvisc®. All hylan test articles produced macroscopic responses that were comparable to a negative control implant material. {4} Page 5 Ten studies using rats, guinea pigs and owl monkeys were conducted to characterize any tissue reactions to implanted hylans. Hylan B was injected subcutaneously (5 ml) into each of 28 female rats. Saline was injected into eight rats, which served as controls. A small amount of fibrosis was noted in 75% of the injection sites three months after hylan B implantation. Hylan B remained present in the injection site 12 months after implantation, but was not associated with fibrosis or with any other tissue reaction. Sixteen guinea pigs were injected at multiple sites with either hylan B, injectable collagen (Zyplast® and Zyderm®, Collagen Corp.) or physiological saline. Tissue samples were taken from two animals at each of the following time points: three days and one, two, four, nine, 13, 26 and 52 weeks. Hylan B proved to be biocompatible and remained present 52 weeks after injection. The collagen implants could not be detected at Week 52. The liquid vitreous of owl monkeys was replaced with hylan A, hylan B, and Synvisc® in the monkey vitreous test (Balazs and Denlinger, 1980). The inflammatory reaction was quantified by measuring the level of infiltration of leukocytes into the anterior chamber 48 hours after insertion of the hylan test article. Ocular changes were evaluated using slit lamps, ophthalmoscopy, and measurements of intraocular pressure periods for up to 74 months. Hylan A, hylan B and Synvisc® produced no inflammation or any other evidence of toxicity in any of the tested eyes. **Mutagenicity**: Hylan A, hylan B and Synvisc® were evaluated for mutagenic and clastogenic potential in a series of test procedures, which included the Ames assay, a chromosome aberration assay in Chinese hamster ovary (CHO) cells (Hsie et al.), a CHO/hypoxanthine guanine phosphoribosyl transferase (HGPRT) mutation assay, and an unscheduled DNA synthesis (UDS) assay. The hylan materials were not found to be mutagenic or clastogenic in any of these tests. Synvisc®, hylan A, and hylan B (degraded and intact) were evaluated in the plate incorporation assay (gene mutation assay, Ames et al.). All assays included activated (S-9 fraction of rat liver homogenate from Aroclor 1254-induced rats) and non-activated systems. None of the hylan preparations produced evidence of mutagenicity. The ability of Synvisc® to induce chromosome aberrations was tested using cultured CHO cells with and without metabolic activity. Synvisc® did not induce any chromosomal abnormalities in any these tests. Synvisc® was tested for its ability to induce gene mutation at the HGPRT locus in cultured CHO cells, with and without an exogenous S-9 activation system at a range of various concentrations. Synvisc® showed no indication of mutagenic activity at any dose tested. {5} Page 6 Synvisc$^{\text{R}}$ was tested for its potential to cause unscheduled DNA synthesis (UDS) in rat primary hepatocytes (Mirsalis and Butterworth). Hepatocytes were obtained from Sprague-Dawley rats that had been treated intraperitoneally with three doses of Synvisc$^{\text{R}}$. UDS was evaluated by measuring the incorporation of [³H]-thymidine into the extracted hepatocytes. None of the three doses of Synvisc$^{\text{R}}$ appeared to produce any increase in UDS or any primary DNA damage. **Subchronic toxicity:** Five parenteral repeat-dose toxicity studies of two-week to five-week duration were conducted in guinea pigs and rabbits to determine potential toxicity of hylan A and hylan B, the two components of Synvisc$^{\text{R}}$. Hylan A and hylan B did not produce any remarkable systemic or local toxicity. Individual summaries of the studies are given below: Hylan A, when administered in two weekly intraperitoneal injections of 10 mg/ml (2.5 ml/kg to each group of male and female test animals), produced no changes in body weight or organ weights. Hematology and clinical chemistry comparisons were similar to that of control animals. No treatment-related gross or microscopic changes were observed. Four weekly intra-arterial injections of hylan A (3 mg/kg) produced no changes in body weights, clinical pathology parameters or blood hyaluronan levels in 14 NZW rabbits. No gross or histological treatment-related changes were observed. Knee joints of 10 male rabbits were injected bilaterally with 0.5 ml of hylan B (2.6 mg/ml), or physiological saline solution twice at an interval of one week. Animals were sacrificed 3 weeks after the second injection. Body weight and clinical pathology values were not affected by the treatment. Minimal focal or multifocal inflammation in synovial membranes was associated with the treatment. Four consecutive weekly intra-articular injections of hylan A (0.3 ml/joint) into the knee joints (bilateral) of five NZW rabbits produced no changes in normal weight gains. At the end of the fourth week, no changes were found in synovial fluid volume and its hyaluronan content and hexuronic acid or hexosamine content of joint cartilage. No immunogenic skin responses (erythema and edema) or gross joint abnormalities were observed in any animals in this study. **Chronic toxicity in primates:** Synvisc$^{\text{R}}$ was tested in owl monkey joints for its potential to induce local and systemic effects following repeated weekly bilateral intra-articular (knee) injections for 31 weeks (average of 24 injections per knee) with a one-year follow-up period. The study consisted of two groups of animals: a treatment group of 20 (eight male and 12 female) monkeys and a control group of 14 (six male and eight female) monkeys. The animals in the treatment group received an amount of Synvisc$^{\text{R}}$ that in toto represented a dose that was greater than 20 times the total human clinical regimen when adjusted for body mass. {6} Page 7 The following parameters were evaluated in this study: clinical signs (specifically related to the functions and health of the joint) and behavioral changes (daily); body weight gains (monthly); intradermal skin sensitization testing (monthly); blood sampling for immune reactivity by ELISA methodology (monthly); and determination of hematology and clinical chemistry parameters and urinalysis (at 16 or 21 months after the first injection). Synvisc® was not associated with any detectable local or systemic adverse effects during the observation period. No detectable changes related to overall health, behavior or physical activity were observed that appeared related to Synvisc® treatment. There were no clinical signs of any local reactions (swelling, heat, redness) to Synvisc® in the knees or the legs. Analysis of hematology, clinical chemistry and urinalysis data did not reveal any treatment-related abnormalities. Monthly intradermal skin sensitization test results did not show any cell-mediated or delayed hypersensitivity reactions. Two animals developed low titers of antibodies to Synvisc® (possibly due to contamination by gram-negative bacterial lipopolysaccharide), while five animals developed antibodies to chicken proteins. Four animals treated with Synvisc® died during this study. The death rate in the Synvisc® treated group did not exceed that for the remainder of the monkey colony during this time period. **Clearance:** The kinetics of the clearance of hylan A, hylan B, and Synvisc® were studied by injection of radiolabeled hylan A and/or hylan B into the knees of rabbits at therapeutic concentrations. The half-lives of the elimination of these materials were calculated from measurements of the material remaining in the synovial fluid at various times after injection; hylans adhering to tissue surfaces were considered part of an elimination compartment and were not included in the calculation of the half-lives. The clearance of Synvisc® was measured using Synvisc® made from a combination of radiolabeled hylan A and hylan B. There was no significant inhibition of the clearance of one component of Synvisc® by the other (Table 1). The half-life of hylan A in the blood was measured in a competitive-binding radioimmunoassay after the injection of unlabeled hylan A into the auricular artery of rabbits at levels about 500 times that of the normal blood level. By this method, hylan A had a half-life in the blood of 16-17 minutes (Table 2). This is longer than the 3-5 minute physiological half-life in blood reported in the scientific literature for native hyaluronan (hyaluronic acid). The longer half-life of hylan A may reflect the higher molecular weight of the cross-linked hylan as compared to native hyaluronan. The clearance of intravascular hylan B from the blood cannot be directly determined by intravascular administration because hylan B is an insoluble particulate gel. Therefore, unlabeled hylan B was degraded and solubilized by breaking glycosidic bonds with acid hydrolysis to produce hylan B polysaccharides, which can be injected intravascularly. These polysaccharides are similar to the solubilized form of hylan B which exists in the joint and which enters the blood stream after hylan B is injected into the joint and {7} Page 8 degraded. Following intravascular injection, the half life in the blood of unlabeled soluble degraded hylan B was found to be 22 minutes (Table 2). ## IX. SUMMARY OF CLINICAL INVESTIGATIONS The safety and/or effectiveness of Synvisc® for the treatment of osteoarthritis (OA) of the knee were evaluated in seven clinical investigations conducted in three countries. The investigations had a total of 738 human subjects (Table 3). All of the investigations were performed in accordance with the Declaration of Helsinki and/or with Good Clinical Practice guidelines. Six double-blind studies (Studies #1, #2, #3, #5, #6 and #7) compared Synvisc® to a control treatment (intra-articular saline, hyaluronan or nonelastoviscous hylan fluid, or arthrocentesis). Control subjects in Study #6 received NSAID therapy as well as arthrocentesis. Study #5, conducted in the U.S.A. under IDE #G890108, evaluated the safety and effectiveness of a single course three Synvisc® injections as well as the safety of a second course of Synvisc® treatment. The study contained an alternative treatment (arthrocentesis) for the first course of Synvisc®, but did not contain a concurrent control group for the second course of Synvisc®. Study #4 was intended to evaluate the safety of Synvisc® without a comparison to an alternative treatment. All subjects in this study received Synvisc®. All but two of the trials (Studies #1 and #2) were conducted at more than one investigational site. ## Clinical Investigation Methodology **Subject Population:** All subjects participating in the seven clinical studies had chronic idiopathic osteoarthritis of the knee with moderate to severe pain. Most of the studies limited enrollment to subjects with Grade I to Grade III radiological changes of the study knee, but three studies, #4, #5 and #7, included subjects with Grade IV X-rays. Subjects were excluded from all studies if they had been diagnosed with rheumatoid arthritis, arthritis of metabolic origins or chondromalacia or had been treated with arthroscopy within two months or with intra-articular steroid within three months of the beginning the study. Subjects were also excluded if they were limited in their daily activities because of disorders other than OA of the knee, if they were taking steroid therapy, or if they were pregnant. In five studies, subjects were excluded if they had effusions in the joints of their treated knees. A total of 738 subjects were enrolled in these studies. Ninety-five percent of the Synvisc®-treated subjects and 98% of controls completed the studies. A total of 511 subjects were treated with 1,771 Synvisc® injections. {8} Page 9 Table 4 summarizes the demographic and disease characteristics of the subjects in these seven clinical trials. The majority of subjects had mid-stage X-ray grades (76% of subjects had Grades II and III on the Larsen or Kellgren-Lawrence scales) and had a disease duration of between one and five years (48%). ## Effusion Two studies (#5 and #6) contained subjects who presented with an effusion (34% of the subjects in Study #5 and 15% of the subjects in Study #6). If an effusion was present before any treatment or control injection, it was removed as completely as possible before the injection. Studies #1, #2, #3, #4 and #7 excluded subjects who had an effusion in the knee joint prior to the first intra-articular treatment in that knee. ## Alternative Treatments (Controls) All subjects participating in these studies received arthrocentesis with removal of effusion, if present. All subjects received either Synvisc® or one of three types of alternative (control) treatments. In Studies #1, #2, and #3, the alternative treatment consisted of arthrocentesis followed by intra-articular injection of 2 ml phosphate-buffered saline solution, the solvent for Synvisc®. In Studies #5 and #6, nothing was injected into the control knees following arthrocentesis. In Study #6, a group of control subjects received continued oral administration of an NSAID which had been well-tolerated by the subject for at least one month prior to the beginning of the study, but which had not provided adequate pain relief. Another group received Synvisc® as well as oral NSAIDs. In Study #7, alternative-treatment subjects received one of three different nonelastoviscous hylan or hyaluronan preparations. Each of these preparations had lower elastoviscous properties than did Synvisc®. ## Blinding In studies #1, #2, #3, #5, #6 and #7, patients, injectors, and evaluators were blinded as to treatment. In Studies #5 and #6, where arthrocentesis alone was the only intra-articular treatment, a screen was set up so that the subject could not observe the procedure being performed. Studies #1, #2, and #3 compared the results of injecting Synvisc® and saline intra-articularly in humans. To determine whether investigators could distinguish between these materials during the injection procedure, three separate studies were performed in which experienced and inexperienced blinded investigators injected one or both of the materials into the joints of rabbit knees. In all three animal studies, neither experienced nor inexperienced {9} Page 10 injectors were able to differentiate between these two treatments while performing injections. ## Pretreatment Medication With the exception of Study #6, each of the five controlled clinical studies contained a "washout" period during which subjects withdrew from all anti-inflammatory medications before beginning intra-articular treatments. In Studies #1, #2, #3, #4 and #7, all subjects were required to discontinue all medication, including analgesics, during the two-week period immediately preceding the first intra-articular treatment. In Study #5, subjects discontinued all anti-inflammatory medications for a four-week washout period preceding the first intra-articular treatment, but were permitted to take acetaminophen as an analgesic. Subjects in Study #6 did not discontinue their arthritis medications prior to Synvisc® treatment because the study was intended to evaluate Synvisc® therapy as a direct replacement for NSAID therapy. ## Concomitant Medication and Rescue Therapy In studies #1, #2, #3, #4, and #7 subjects were permitted to take any concomitant medications that they desired to treat their arthritis after treatment with the study medications had begun. All such medications were documented on the subjects' case report forms. Studies #5 and #6 permitted only the concomitant use of acetaminophen for analgesia. In all studies, any concomitant medications taken by the subjects during the course of the trial were recorded, whether or not the medications were related to the subjects' arthritic pain. In studies #1, #2, #3, #4, and #7, concurrent therapies other than medications were not recorded. All concurrent therapies, including medications, physical therapy, and exercise, were recorded in Studies #5 and #6. ## Subjects with Osteoarthritis in Both Knees Subjects with OA in both knees were included in all seven studies. In Studies #4, #5 and #7, both knees of bilaterally affected subjects could be entered, treated and evaluated. In Study #6, only the most painful knee of bilaterally affected subjects was followed for the analysis of both safety and effectiveness. If the contralateral knee was painful, it could receive the same treatment as the target knee, but was followed only for the analysis of safety. In Studies #1, #2, and #3, bilaterally affected subjects were permitted to be treated only in the most painful knee. {10} Page 11 # Outcome Measures The primary outcome measures of effectiveness in all of the trials were the subjects' evaluations of pain on 0-100 mm horizontal visual analog scales (VAS). All of the studies measured movement-related pain, requesting evaluator and/or subject assessment of "pain with motion", "weight-bearing pain", and "pain while walking". All of the studies also measured subjects' overall evaluations, night pain, and restriction of activity. Rest pain was measured in Studies #5 and #6. Except in Study #5, night pain and rest pain were not predominant symptoms at baseline. Blinded evaluators also questioned the subjects about the subjects' pain and activity levels and marked the VAS accordingly. Based on this information and on the evaluators' overall assessments of the subjects' conditions, the evaluators also recorded on the VAS their impressions of treatment success. # Statistical Methods For all statistical analyses, two-tailed tests were used with alpha = 0.05. Improvements from baseline were calculated for individual subjects. Least squares means were then calculated from the individual subject improvements and were used for comparisons between treatment groups. The statistical significance of improvements from baseline was calculated by paired t-test. Between-group comparisons of improvement from baseline were calculated by one-way ANOVA. Repeated measures analyses were conducted, when appropriate, using PROC MIXED. These analyses used all evaluation visits occurring after completion of treatment. Covariates to be used for both the repeated measures analyses and ANOVA were selected by stepwise regression. # Pivotal Studies Three clinical studies (Studies #2, #3 and #5) compared three weekly injections of Synvisc with concurrent placebo control treatments (saline or arthrocentesis). FDA considered these three investigations to be the most important for the determination of device effectiveness. # Clinical Studies of Effectiveness ## Studies #1 and #2: Single-Center Treatment Regimen Studies The first two clinical studies compared a treatment regimen of two injections applied two weeks apart (Study #1) to a regimen of three injections applied weekly (Study #2) at a single center. Each study compared Synvisc treatment to the corresponding regimen of saline injections. {11} Page 12 Synvisc® treatment showed statistically significant superiority (p &lt; 0.05) over saline treatment for the two biweekly and the three weekly treatment regimens from Week 3 or 4 through the final study visit at Week 12 for most outcome measures. Three weekly injections of Synvisc® were statistically superior to two biweekly injections of Synvisc® from Week 8 onward. Table 5 illustrates this superiority for the subject evaluation of weight-bearing pain. ## Study #3: Synvisc® vs. Saline Control Multicenter Study Study #3 was designed to provide a multicenter, randomized, controlled double-blind comparison of a single course of three weekly Synvisc® injections vs. the same regimen of saline injections. Tables #6 and #7 compare the Synvisc® and saline groups throughout the study with respect to their improvement (in mm) for all subject-evaluated VAS outcome measures. Subjects in both the Synvisc® and saline groups significantly improved after the first injection. The superiority of the Synvisc® group over the saline group was statistically significant in a combined analysis of data from all outcome measures at Weeks 3, 8, and 12 (1, 6, and 10 weeks after the last injection) (Table 6) and for each of the individual outcome measures at Weeks 8 and 12 (Table 7). ## Study #4: Open-Label Study Study #4 was a multicenter open-label study designed primarily to assess product safety in over 200 subjects treated with three weekly injections of Synvisc®. Measurements of outcome measures for effectiveness were also performed on VAS to document changes in these measures as compared to baseline and to compare such changes among different subpopulations of patients. There were no comparable control groups or alternative treatments in this study. All of the evaluated outcome measures showed improvements as compared to baseline measurements that had been made before treatment had begun. Subjects with X-ray Grade IV OA experienced improvements in their own assessments of weight-bearing pain as compared to baseline that were comparable to those improvements observed in subjects with less severe disease (Table 8). ## Study #5: Evaluation Three Synvisc® Injections vs. Three Arthrocenteses Following a Four Week Washout Period and Evaluation of a Second Synvisc® Treatment Study #5 was a multicenter, randomized, controlled double-blind trial designed to establish safety and effectiveness of Synvisc® treatment as compared to arthrocentesis alone following a four-week washout period during which subjects received no treatments for their disease. The four-week washout period was intended to determine the natural history of OA after potential investigational subjects had discontinued all therapies for OA. {12} Page 13 At the end of the four week no-treatment period, subjects electing to enter the first treatment phase of this study received either three weekly SynviscR injections or three weekly arthrocenteses as a control. All subjects were permitted to take acetaminophen as an escape medication for pain after treatment began if they so desired. Subjects were evaluated to determine their responses to treatment four weeks after their first intra-articular injections (two weeks after their third and last injections). This was the last evaluation point before any subjects were permitted to enter a second treatment phase of this investigation. Upon their request and fulfillment of entry criteria, subjects entered the second phase of this investigation and received a course of three SynviscR injections four or more weeks after they had completed their first course of treatment. Subjects were allowed to enter the second phase whether their first course of treatment had consisted of SynviscR or of arthrocentesis alone. Adverse events were evaluated during both treatment phases. The second phase of this study did not contain a concurrent control group that received an alternative treatment to SynviscR. The effectiveness of SynviscR effectiveness was therefore not evaluated after the second course of treatment had begun. Both the SynviscR and the arthrocentesis-treated subjects improved significantly in all outcome measures (p &lt; 0.05) as compared to baseline. Two weeks after the first treatment, subjects injected with saline showed a significantly greater improvement in night pain than did patients injected with SynviscR (p &lt; 0.004) (Table 10). However, no other statistically significant difference (p &gt; 0.05) in improvement from baseline was found between the SynviscR and arthrocentesis groups in any outcome measure at any time throughout the four week evaluation period of the first treatment phase of this study in the intent-to-treat population, whether the analyses were corrected for covariates (Table 9) or not (Table 10). Thus, SynviscR was no more effective than arthrocentesis alone in this population. In a retrospective analysis, a subpopulation of 31 "flare" subjects were identified who had appeared to have responded better to SynviscR than to arthrocentesis alone during the first treatment phase. In these 31 subjects, motion pain or rest pain had increased more than 20 mm on the VAS during the four-week "no-treatment" period. The 15 SynviscR-treated "flare" subjects exhibited significantly greater improvements in most outcome measures than did the 16 comparable subjects that had been treated only with arthrocentesis (Tables 11 and 12). There were no unanticipated adverse events in the 35 subjects who received two courses of SynviscR treatment. The incidence and types of adverse events were similar to those in 43 subjects whose SynviscR injections followed treatments with arthrocentesis alone. {13} Page 14 # Study #6: NSAID Alone vs. Synvisc® + NSAID Study #6 was a multicenter, randomized, double-blind controlled study designed to evaluate the effectiveness of Synvisc® treatment either as a replacement or as an addition to NSAID therapy in subjects whose NSAID therapy had not provided adequate pain relief during the 30 days prior to Synvisc® treatment. Upon entering the trial, subjects were randomized into one of three treatment groups: - continuation of NSAID therapy plus three weekly arthrocentesis (NSAID-only group) - discontinuation of NSAID therapy and replacement with three weekly Synvisc® injections (Synvisc®-only group) - continuation of NSAID therapy plus three weekly Synvisc® injections (Synvisc®-plus-NSAID group) There was no "washout period" in this study, nor were any subjects treated with saline or arthrocentesis alone. All subjects received arthrocentesis with removal of any effusions at each treatment visit. Subjects were evaluated at follow-up visits for up to twelve weeks after receiving the first treatment. The seven and twelve week evaluations occurred after the completion of treatment. Subjects in all three treatment groups experienced significant decreases of pain from baseline. There were no statistically significant differences (p &lt; 0.05) in the outcome measures between the three treatment groups after the completion of treatment, except that the subjects taking only Synvisc® experienced a greater improvement in night pain than did the subjects treated only with NSAIDs (Table 13). The data showed that a course of three Synvisc® injections is as least as effective as a continuation of NSAID therapy plus arthrocentesis in subjects who had not obtained adequate pain relief with NSAIDs prior to entering the study. They also showed that the pain-relieving effect of three Synvisc® injections can last for 12 weeks and is at least as effective as an NSAID administered continuously over the same 12-week period. A combination of the two treatments was no more or less effective than either one alone. All subjects received arthrocentesis, but none received arthrocentesis alone. Therefore, the study was unable to determine whether the observed improvements were due to arthrocentesis, escape medication (acetaminophen) and the natural course of OA, or whether they were due to treatment with Synvisc® and/or NSAIDs. {14} Page 15 # Study #7: Relationship of Effectiveness and Elastoviscosity (Clinical Study #7) Clinical Study #7, a double-blind, multicenter study, compared four different preparations composed of hyaluronans or hylans to determine whether a correlation could be established between the elastoviscosity of these products and their clinical effectiveness. The following preparations were tested: - Synvisc® (elastoviscous hylan) - Degraded Synvisc® (nonelastoviscous hylan) - 1% hyaluronan, average molecular weight 750,000 - 1% hyaluronan, average molecular weight 2,000,000 All preparations had similar polysaccharide structures and concentrations (1%), and differed only in their average molecular weights and therefore in their elastoviscosities. Synvisc® had the highest elasticity and viscosity of those substances tested. All subjects received three weekly injections of a hylan or hyaluronan preparation following a two week period during which they received no arthritis treatments. No subjects were treated with either saline or with arthrocentesis alone. Subjects treated with Synvisc® often showed greater improvements in pain scores than did subjects receiving the other three treatments. However, these were differences were often not statistically significant, and in some instances were minimal. Synvisc® showed significantly greater improvements in most outcome measures than did three injections of the 750,000 molecular weight hyaluronan preparation. For subject evaluations of weight-bearing pain at Week 12, Synvisc®-treated subjects showed a mean improvement of 38 ± 4 mm on the VAS. This was a significantly greater improvement (p = 0.03) than that experienced by subjects treated with the 750,000 molecular weight hyaluronan. However, subjects treated with nonelastoviscous hylan showed an improvement in subject-evaluated weight-bearing pain measure of 36 ± 4 mm. This improvement was not significantly different from that experienced by Synvisc®-treated subjects. Moreover, there were no significant differences between the improvements in the VAS for any outcome measures between the group treated with Synvisc® and the group treated with the 2,000,000 molecular weight hyaluronan preparation. This preparation has an elastoviscosity that is lower than that of Synvisc® but is higher than that of the 750,000 molecular weight hyaluronan. Comparison of the Synvisc® group with less elastoviscous hyaluronan and hylan preparations suggested that the effectiveness of these materials may be related to their viscosities and elasticities. However, the study did not conclusively demonstrate this relationship. {15} Page 16 ## Gender Analysis: The gender of subjects were reported in all studies. There appear to be no effects of gender on the safety and effectiveness of Synvisc®. ## X. CONCLUSIONS DRAWN FROM THE STUDIES These studies provide reasonable assurance of the safety and effectiveness of Synvisc® for the treatment of pain in OA of the knee in patients who have failed to respond adequately to conservative nonpharmacologic therapy and simple analgesics (e.g., acetaminophen). These studies demonstrated that: - in subjects with OA of the knee who lack intra-articular effusions, who have not responded adequately to previous arthritis therapies, and who have not received any therapies during the previous two weeks, three weekly intra-articular knee injections of Synvisc® reduce knee pain to a greater extent than do comparable injections of saline for up to twelve weeks following the beginning of treatment (Studies #2 and #3). - three weekly injections of Synvisc® are more effective than two weekly injections of the product (Studies #1 and #2). - under certain investigational conditions, treatment with Synvisc® is no more effective than arthrocentesis alone (Study #5). - injections of Synvisc® do not interfere with the pain-relieving effect of arthrocentesis when prior NSAID treatment is continued in subjects who have not responded adequately to NSAIDs during the month before they receive Synvisc® or arthrocentesis (Study #6). ## XI. PANEL RECOMMENDATIONS On November 20, 1996, the Orthopedics and Rehabilitation Devices Advisory Panel recommended approval of Biomatrix Corp.'s PMA for Synvisc® subject to the following: 1. The device is for use in subjects who are resistant to treatment with analgesics and NSAIDs, for up to three injections given in one course of treatment, 2. Conduct a post-approval study or studies which will include parameters which were not available at that time. These parameters include: synovial fluid analysis, quality of life, walking time, height, weight, outcome measures, safety and effectiveness in comparison to NSAIDs, and long term effects of treatment on joints. 3. Conduct a post-approval study to include multiple courses of injection and studies relating to the development of hypersensitivity reactions and pathology of the joints and lymph nodes as available. {16} Page 17 4. Perform pathological studies in the monkeys previously treated with Synvisc after each monkey expires. 5. The data required further analysis and reanalysis. Reanalysis would include analysis of covariates. 6. Address the possible influence of residual formaldehyde used in manufacturing Synvisc on the safety and effectiveness of the final product. 7. Address the adequacy of using of immunodeficient nude mice in preclinical toxicity studies. ## XII. CDRH DECISION On February 22, 1997, CDRH informed Biomatrix, Inc., that further information on the clinical studies was needed, including a covariate analyses of data, on Synvisc's formaldehyde content and the effects of the formaldehyde on Synvisc's safety and effectiveness, and on the adequacy of using immunodeficient mice in its toxicity studies. Biomatrix, Inc., responded satisfactorily to these concerns. Biomatrix, Inc., additionally agreed to perform pathological studies on each of the monkeys previously treated with Synvisc after each animal expires. CDRH did not agree with the advisory panel's recommendation that post-approval studies were needed to determine the safety and effectiveness of Synvisc for the above indication, as CDRH considered that Biomatrix, Inc., had already provided sufficient data to support the indication. CDRH did not agree with the advisory panel's recommendation that OA subjects must have failed to respond to NSAID therapy before receiving Synvisc, since treatment with Synvisc does not appear to be associated with the adverse gastrointestinal reactions (bleeding, nausea, etc.) that are known to occur with NSAIDs. On August 8, 1997, CDRH approved a single course of three intra-articular Synvisc injections for the treatment of pain in OA of the knee in subjects who have not responded adequately to conservative non-pharmacologic therapy and simple analgesics (e.g., acetaminophen). ## XIII. APPROVAL SPECIFICATIONS Directions for Use: See product labeling. Postapproval Requirement and Restrictions: See approval order. {17} Page 18 # XIV. REFERENCES Ames, B.N., McCann, J. and Jamasaki, E. (1975). Methods for detecting carcinogens and mutagens with the salmonella/mammalian-microsome mutagenicity test. Mutation Res. 31, 347-367. Balazs, E.A. and Denlinger, J.L. (1980). Replacement of the liquid vitreus with sodium hyaluronate in monkeys. I. Short term evaluation. Exp. Eye Res. 31, 81-99. Denlinger, J.L. (1982). Metabolism of sodium hyaluronate in articular and ocular tissues. Ph.D. thesis, Université des Sciences et Techniques de Lille, Lille, France, 1982. Engström-Laurent, A., Laurent, U.B.G., Lilja, K. and Laurent, T.C. (1985). Concentration of sodium hyaluronate in serum. Scand. J. Clin. Lab. Invest. 45, 497-504. Hsie, A.W., Casciano, D.A., Couch, D.B., Krahn, D.F., O'Neill, J.P. and Whitfield, B.L. (1981). The use of Chinese hamster ovary cells to quantify specific locus mutations and to determine mutagenicity of chemicals. Mutation Res. 86, 193-214. Magnusson, B. and Kligman, A.M. (1970). Identifications of contact allergens. In Allergic Contact Dermatitis in the Guinea Pig, Charles C. Thomas, Springfield, 102-123. Mirsalis, J.C. and Butterworth, B.E. (1982). Induction of unscheduled DNA synthesis in rat hepatocytes following in vivo treatment with dinitrotoluene. Carcinogenesis 3, 241-245. Mirsalis, J.C., Tyson, C.K. and Butterworth, B.E. (1982). Detection of genotoxic carcinogens in the in vivo-in vitro hepatocytes DNA repair assay. Environ. Mutagen. 4, 553-562. {18} TABLE 1 Intra-Articular Clearance Studies in Rabbits | Test Article | Half-Life | | --- | --- | | [³H]-hylan A fluid (1%) (avg. MW: 6 million) | 1.2 ± 0.1 days | | [³H]-hylan B gel (0.4%) | 7.7 ± 1.0 days | | [³H]/[¹⁴C]-Synvisc® ¹⁴C-hylan A fluid | 1.5 ± 0.2 days | | ³H-hylan B gel | 8.8 ± 0.9 days | | Hyaluronan 1% (avg. MW: 1.7-2.6 million)^{1} | 11 hours | ¹ Reference: Denlinger, 1982 TABLE 2 Clearance of Hylan A and Hylan B From Blood | Test Article | Dose (mg/kg) | Blood Half-Life (minutes) | % Cleared in 60 Minutes | Clearance Rate^{1} (mg/kg/day) | | --- | --- | --- | --- | --- | | hylan A | 0.35^{2} | 16 | 96% | 15.9 | | hylan A | 0.67^{2} | 17 | 81% | 28.5 | | hylan B^{4} | 23.7^{3} | 22 | 96% | 77.4 | 1. daily clearance rates were calculated from the dose injected and the measured clearance rates (half-lives) 2. test species is New Zealand rabbit, assumes 3 kg. total body weight 3. test species is Sprague-Dawley rat, assumes 200g body weight 4. soluble degraded hylan B {19} # TABLE 3 Description of the Clinical Trials Evaluating the Safety and Effectiveness of Synvisc® | Study No. | Description | Country | Number of Subjects | | --- | --- | --- | --- | | 1 | Double-blind comparison of the safety and effectiveness of two Synvisc® injections 14 days apart vs. the corresponding regimen of saline injections | Germany | 48 | | 2 | Double-blind comparison of the safety and effectiveness of three weekly Synvisc® injections vs. the corresponding regimen of saline injections | Germany | 30 | | 3 | Double-blind comparison of the safety and effectiveness of three weekly Synvisc® injections vs. the corresponding regimen of saline injections | Germany | 110 | | 4 | Evaluation of the safety of three weekly Synvisc® injections and comparison of effectiveness in subpopulations; no control group | Germany | 222 | | 5 | Double-blind comparison of the safety and effectiveness of three weekly Synvisc® injections vs. the corresponding regimen of arthrocenteses, with a four-week pretreatment medication washout period and an evaluation of the safety of a second course of Synvisc® treatment | U.S.A. | 94 | | 6 | Double-blind comparison of the safety and effectiveness of three weekly Synvisc® injections with and without concomitant NSAID therapy vs. NSAID therapy with the corresponding regimen of arthrocenteses | Canada | 102 | | 7 | Double-blind comparison of the effectiveness of three weekly injections of Synvisc® and of three other hyaluronan and hylan preparations vs. the relative elastoviscosities of the preparations | Germany | 132 | {20} # TABLE 4 ## Demographic Analysis of the Study Population for the Seven Clinical Studies | Male | 296 | (40%) | | --- | --- | --- | | Female | 442 | (60%) | | Age (yrs.) | 61.3 ± 0.5¹ | | | X-Ray Grade² | | | | I | 96 | (12%) | | II | 302 | (38%) | | III | 305 | (38%) | | IV | 95 | (12%) | | Disease Duration | | | | Mean (yrs) | 6.0 ± 0.2 | | | <1 yr | 122 | (15%) | | 1-5 yrs | 384 | (48%) | | >5 years | 288 | (36%) | ¹ Mean ± Standard Error of Mean ² X-ray grade on Larsen or Kellgren-Lawrence scales {21} # TABLE 5 ## Studies #1 &amp; #2 ### Treatment Regimen Comparison of Weight-Bearing Pain #### One-Way Analysis of Variance | Time (Weeks) | Saline Mean ± SEM^{1} n = 39 | Two Injections Mean ± SEM n = 23 | Three Injections Mean ± SEM n = 15 | p-value | | | | --- | --- | --- | --- | --- | --- | --- | | | | | | Control vs. Two Injections | Control vs. Three Injections | Two Injections vs. Three Injections | | 0^{*} | 68 ± 3 | 61 ± 4 | 65 ± 5 | 0.1 | 0.5 | 0.5 | | 1^{*} | 64 ± 4 | NA^{2} | 54 ± 4 | NA | 0.06 | NA | | 2^{*} | 47 ± 4 | 41 ± 5 | 34 ± 6 | 0.4 | 0.06 | 0.3 | | 3 | 44 ± 5 | NA | 22 ± 5 | NA | 0.004 | NA | | 4 | 47 ± 5 | 32 ± 5 | NA | 0.02 | NA | NA | | 8 | 48 ± 4 | 32 ± 5 | 13 ± 6 | 0.006 | 0.0001 | 0.01 | | 12 | 49 ± 4 | 27 ± 5 | 11 ± 6 | 0.0003 | 0.0001 | 0.03 | 1. Mean ± SEM = Mean (mm) ± Standard error of mean on 100 mm VAS (raw scores) * Injection Date 2. NA = Not Applicable {22} # TABLE 6 ## Study #3 Improvements from Baseline (Weeks 3, 8 and 12) Patients ≥ 40 Years of Age Repeated Measures Analysis (PROC MIXED) | Outcome Measure | Covariates^{1} | Synvisc^{®} Mean^{2} ± SEM^{3} N = 52 | Saline Mean ± SEM N = 57 | p-value between treatments | | --- | --- | --- | --- | --- | | Patient evaluated weight-bearing pain | Center, baseline, age, x-ray grade and any concurrent therapy | 39 ± 3 | 15 ± 3 | 0.0001 | | | p-value from baseline | 0.0001 | 0.0001 | | | Patient evaluated night pain | Baseline, gender and any concurrent therapy | 24 ± 2 | 14 ± 2 | 0.001 | | | p-value from baseline | 0.0001 | 0.0001 | | | Patient assessment most painful knee movement | Center and concurrent therapy-NSAIDs | 61 ± 4 | 33 ± 4 | 0.0001 | | | p-value from baseline | 0.0001 | 0.0001 | | | Patient assessment success of treatment | Center, age, any concurrent therapy, x-ray grade and concurrent therapy -NSAIDs | 69 ± 4 | 39 ± 4 | 0.0001 | | | p-value from baseline | 0.0001 | 0.0001 | | | Evaluator assessment decrease of activity during daily chores | Center, baseline, age, x-ray grade and any concurrent therapy | 33 ± 3 | 14 ± 2 | 0.0001 | | | p-value from baseline | 0.0001 | 0.0001 | | 1. The analyses were corrected for all covariates found to be statistically significant by stepwise regression. Each significant covariate is listed in the table with the individual outcome measure. 2. Mean of assessments on VAS of 0 to 100 mm 3. SEM = Standard Error of the Mean {23} # TABLE 7 ## Study #3 ### Improvements from Baseline (Weeks 1 – 12) #### Patients ≥ 40 Years of Age ##### Analysis of Variance¹ | | Baseline Mean² ± SEM³ | Improvement (Change from Baseline) Mean ± SEM | | | | | | --- | --- | --- | --- | --- | --- | --- | | Week | 0 | 1 | 2 | 3 | 8 | 12 | | Patient evaluated weight-bearing pain | | | | | | | | Synvisc®-treated | 69.7 ± 2.3 | 12.0 ± 1.8 | 26.5 ± 2.5 | 37.9 ± 3.2 | 45.9 ± 3.2 | 46.5 ± 3.5 | | p-value from baseline | NA⁴ | 0.0001 | 0.0001 | 0.0001 | 0.0001 | 0.0001 | | Saline-treated | 75.1 ± 2.3 | 9.0 ± 2.2 | 17.0 ± 2.3 | 23.0 ± 3.0 | 16.8 ± 4.0 | 16.4 ± 4.1 | | p-value from baseline | NA | 0.0001 | 0.0001 | 0.0001 | 0.0001 | 0.0002 | | p-value between treatments | 0.1 | 0.3 | 0.01 | 0.0008 | <0.0001 | <0.0001 | | Patient evaluated night pain | | | | | | | | Synvisc®-treated | 41.6 ± 4.0 | 9.2 ± 1.8 | 20.0 ± 2.7 | 26.4 ± 3.5 | 28.3 ± 3.4 | 29.8 ± 3.7 | | p-value from baseline | NA | 0.0001 | 0.0001 | 0.0001 | 0.0001 | 0.0001 | | Saline-treated | 45.7 ± 4.1 | 9.5 ± 2.3 | 15.2 ± 2.6 | 21.2 ± 3.1 | 18.4 ± 3.6 | 17.3 ± 3.8 | | p-value from baseline | NA | 0.0001 | 0.0001 | 0.0001 | 0.0001 | 0.0001 | | p-value between treatments | 0.5 | 0.9 | 0.2 | 0.3 | 0.05 | 0.02 | | Patient assessment most painful knee movement | | | | | | | | Synvisc®-treated | NA | 36.8 ± 4.0 | 44.7 ± 4.0 | 57.9 ± 3.8 | 71.5 ± 3.5 | 72.6 ± 4.0 | | p-value from baseline | NA | NA | NA | NA | NA | NA | | Saline-treated | NA | 30.2 ± 3.8 | 31.4 ± 3.4 | 38.9 ± 3.6 | 38.1 ± 4.2 | 38.8 ± 4.3 | | p-value from baseline | NA | NA | NA | NA | NA | NA | | p-value between treatments | NA | 0.2 | 0.01 | 0.0004 | <0.0001 | <0.0001 | ¹ No correction for covariates ² Mean of assessments on VAS of 0 to 100 mm ³ SEM = Standard Error of the Mean ⁴ NA = Not Applicable {24} # TABLE 7 (continued) ## Study #3 Improvements from Baseline (Weeks 1 – 12) Patients ≥ 40 Years of Age Analysis of Variance¹ | | Baseline Mean² ± SEM³ | Improvement (Change from Baseline) Mean ± SEM | | | | | | --- | --- | --- | --- | --- | --- | --- | | Week | 0 | 1 | 2 | 3 | 8 | 12 | | Patient assessment success of treatment | | | | | | | | Synvisc®-treated | NA⁴ | 44.6 ± 4.8 | 59.6 ± 4.3 | 69.8 ± 3.7 | 78.9 ± 3.2 | 81.2 ± 3.3 | | p-value from baseline | NA | NA | NA | NA | NA | NA | | Saline-treated | NA | 39.4 ± 4.2 | 41.8 ± 4.3 | 46.7 ± 4.2 | 43.6 ± 4.4 | 42.3± 4.5 | | p-value from baseline | NA | NA | NA | NA | NA | NA | | p-value between treatments | NA | 0.4 | 0.005 | 0.0001 | <0.0001 | <0.0001 | | Evaluator assessment decrease of activity during daily chores | | | | | | | | Synvisc®-treated | 60.2 ± 3.1 | 11.2 ± 1.7 | 22.3 ± 2.4 | 31.1± 2.9 | 37.3 ± 3.1 | 37.6 ± 3.4 | | p-value from baseline | NA | 0.0001 | 0.0001 | 0.0001 | 0.0001 | 0.0001 | | Saline-treated | 68.0 ± 3.2 | 8.0 ± 1.6 | 14.5 ± 2.0 | 17.4 ± 2.3 | 17.2 ± 3.2 | 15.0 ± 3.2 | | p-value from baseline | NA | 0.0001 | 0.0001 | 0.0001 | 0.0001 | 0.0001 | | p-value between treatments | 0.08 | 0.2 | 0.01 | 0.0003 | <0.0001 | <0.0001 | ¹ No correction for covariates ² Mean of assessments on VAS of 0 to 100 mm ³ SEM = Standard Error of the Mean ⁴ NA = Not Applicable {25} # TABLE 8 ## Study #4 ### Improvements in Subjects' Assessments of Weight-Bearing Pain Following Synvisc® Treatment Comparison of Improvements in Subjects With X-Ray Grades I-III Osteoarthritis and in Subjects With X-Ray Grade IV Osteoarthritis One-Way Analysis of Variance¹ | Timepoint (Week) | Improvement Mean² ± SEM³ | | | | --- | --- | --- | --- | | | Grades I - III n = 210 | Grade IV n = 42 | p-value | | 1 | 11 ± 1 | 12 ± 3 | 0.6 | | 2 | 21 ± 1 | 23 ± 3 | 0.6 | | 3 | 29 ± 2 | 30 ± 4 | 0.9 | | 8 | 34 ± 2 | 33 ± 4 | 0.9 | | 12 | 35 ± 2 | 36 ± 4 | 0.9 | ¹ No correction for covariates ² Mean of assessments (in mm) on VAS of 0 to 100 mm at specified time points subtracted from the mean of assessments on VAS at Week 0. ³ SEM = Standard Error of the Mean {26} # TABLE 9 ## Study #5 ### Improvement from Baseline Four Weeks After First Injection #### (Phase I Evaluation Point) ##### Intent-to-Treat Population ###### Patients ≥ 40 Years of Age **Analysis of Variance** | Outcome Measure | Covariates^{1} | Synvisc^{®} Mean^{2} ± SEM^{3} N = 47 | Arthrocentesis Mean ± SEM N = 46 | p-value between treatments | | --- | --- | --- | --- | --- | | Patient assessment walking pain | Baseline, body mass index, x-ray grade and duration of disease | 19 ± 4 | 15 ± 4 | 0.5 | | | p-value from baseline | 0.0001 | 0.001 | | | Patient assessment motion pain | Baseline, gender, and x-ray grade | 20 ± 5 | 16 ± 5 | 0.5 | | | p-value from baseline | 0.0001 | 0.002 | | | Patient assessment night pain | Center, baseline and x-ray grade | 25 ± 4 | 23 ± 5 | 0.7 | | | p-value from baseline | 0.0001 | 0.0001 | | | Patient assessment restriction of activity | Baseline and gender | 14 ± 4 | 13 ± 4 | 0.8 | | | p-value from baseline | 0.001 | 0.005 | | | Patient overall assessment arthritic pain | Center, baseline, age, x-ray grade and contralateral knee | 20 ± 4 | 16 ± 4 | 0.4 | | | p-value from baseline | 0.0001 | 0.0003 | | 1. The analyses were corrected for all covariates found to be statistically significant by stepwise regression. Each significant covariate is listed in the table with the individual outcome measure. 2. Mean of assessments on VAS of 0 to 100 mm 3. SEM = Standard Error of the Mean {27} # TABLE 10 ## Study #5 ### Improvement from Baseline During Phase I #### Intent-to-Treat Population ##### Patients ≥ 40 Years of Age ###### Analysis of Variance | | Baseline Mean² ±SEM³ | Improvement (Change from Baseline) Mean ± SEM | | | | --- | --- | --- | --- | --- | | Week | 0 | 1 | 2 | 4 | | Patient assessment walking pain | | | | | | Synvisc®-treated | 78.1 ± 2.3 | 12.1 ± 3.0 | 18.5 ± 3.6 | 21.0 ± 4.3 | | p-value from baseline | NA¹ | 0.0002 | 0.0001 | 0.0001 | | Saline-treated | 80.1 ± 2.3 | 10.1 ± 3.4 | 23.7 ± 4.6 | 19.5 ± 4.6 | | p-value from baseline | NA | 0.004 | 0.0001 | 0.0001 | | p-value between treatments | 0.5 | 0.7 | 0.4 | 0.8 | | Patient assessment motion pain | | | | | | Synvisc®-treated | 67.3 ± 2.4 | 12.9 ± 3.2 | 18.9 ± 3.8 | 21.3 ± 3.9 | | p-value from baseline | NA | 0.0002 | 0.0001 | 0.0001 | | Saline-treated | 69.4 ± 3.2 | 9.4 ± 3.6 | 21.2 ± 4.5 | 19.1 ± 4.7 | | p-value from baseline | NA | 0.01 | 0.0001 | 0.0002 | | p-value between treatments | 0.6 | 0.5 | 0.7 | 0.7 | | Patient assessment night pain | | | | | | Synvisc®-treated | 61.0 ± 3.7 | 19.0 ± 3.9 | 17.9 ± 3.5 | 22.8 ± 4.5 | | p-value from baseline | NA | 0.0001 | 0.0001 | 0.0001 | | Saline-treated | 76.0 ± 3.0 | 23.3 ± 4.6 | 36.3 ± 5.2 | 29.8± 5.3 | | p-value from baseline | NA | 0.0001 | 0.0001 | 0.0001 | | p-value between treatments | 0.002 | 0.5 | 0.004 | 0.3 | ¹ No correction for covariates ² Mean of assessments on VAS of 0 to 100 mm ³ SEM = Standard Error of the Mean ⁴ NA = Not Applicable {28} TABLE 10 (continued) Study #5 Improvement from Baseline During Phase I Intent-to-Treat Population Patients ≥ 40 Years of Age Analysis of Variance¹ | | Baseline Mean² ± SEM³ | Improvement (Change from Baseline) Mean ± SEM | | | | --- | --- | --- | --- | --- | | Week | 0 | 1 | 2 | 4 | | Patient assessment restriction of activity | | | | | | Synvisc®-treated | 69.2 ± 3.0 | 9.3 ± 3.1 | 10.7± 3.0 | 15.4 ± 3.5 | | p-value from baseline | NA⁴ | 0.004 | 0.0009 | 0.0001 | | Saline-treated | 63.1 ± 4.0 | 9.5 ± 3.1 | 14.7 ± 4.8 | 15.6 ± 3.7 | | p-value from baseline | NA | 0.003 | 0.003 | 0.0001 | | p-value between treatments | 0.2 | 1.0 | 0.5 | 1.0 | | Patient overall assessment arthritic pain | | | | | | Synvisc®-treated | 76.5 ± 2.2 | 12.8 ± 2.8 | 19.2 ± 3.4 | 21.4± 4.0 | | p-value from baseline | NA | 0.0001 | 0.0001 | 0.0001 | | Saline-treated | 78.0 ± 2.7 | 11.8 ± 3.4 | 23.8 ± 4.3 | 20.7 ± 4.6 | | p-value from baseline | NA | 0.0009 | 0.0001 | 0.0001 | | p-value between treatments | 0.7 | 0.8 | 0.4 | 0.9 | ¹ No correction for covariates ² Mean of assessments on VAS of 0 to 100 mm ³ SEM = Standard Error of the Mean ⁴ NA = Not Applicable {29} # TABLE 11 ## Study #5 ### Improvement from Baseline Four Weeks After First Injection #### (Phase I Evaluation Point) ##### Flare Population Patients ≥ 40 Years of Age ##### Analysis of Variance | Outcome Measure | Covariates^{1} | Synvisc^{®} Mean^{2} ± SEM^{3} N = 14 | Arthrocentesis Mean ± SEM N = 16 | p-value between treatments | | --- | --- | --- | --- | --- | | Patient assessment walking pain | Baseline, body mass index, x-ray grade and duration of disease | 41 ± 5 | 13 ± 4 | 0.0003 | | | p-value from baseline | 0.0001 | 0.008 | | | Patient assessment motion pain | Baseline, gender, and x-ray grade | 34 ± 5 | 11 ± 5 | 0.003 | | | p-value from baseline | 0.0001 | 0.04 | | | Patient assessment night pain | Center, baseline and x-ray grade | 59 ± 8 | 26 ± 8 | 0.005 | | | p-value from baseline | 0.0001 | 0.003 | | | Patient assessment rest pain | Center, baseline and x-ray grade | 46 ± 5 | 19 ± 5 | 0.0007 | | | p-value from baseline | 0.0001 | 0.002 | | | Patient assessment restriction of activity | Baseline and gender | 23 ± 5 | 16 ± 4 | 0.2 | | | p-value from baseline | 0.0001 | 0.001 | | | Patient overall assessment arthritic pain | Center, baseline, age, x-ray grade and contralateral knee | 50 ± 7 | 18 ± 6 | 0.0001 | | | p-value from baseline | 0.0001 | 0.007 | | 1. The analyses were corrected for all covariates found to be statistically significant by stepwise regression. Each significant covariate is listed in the table with the individual outcome measure. 2. Mean of assessments on VAS of 0 to 100 mm 3. SEM = Standard Error of the Mean {30} # TABLE 12 ## Study #5 ### Improvement from Baseline During Phase I #### Flare Population ##### All Patients¹ ##### Analysis of Variance² | | Baseline Mean² ± SEM⁴ | Improvement (Change from Baseline) Mean ± SEM | | | | --- | --- | --- | --- | --- | | Week | 0 | 1 | 2 | 4 | | Patient assessment walking pain | | | | | | Synvisc®-treated | 80.5 ± 3.7 | 23.7 ± 4.2 | 25.7 ± 5.8 | 34.4 ± 5.8 | | p-value from baseline | NA⁵ | 0.0001 | 0.0001 | 0.0001 | | Saline-treated | 81.0 ± 3.5 | 7.1 ± 4.1 | 23.6 ± 5.6 | 18.3 ± 5.4 | | p-value from baseline | NA | 0.09 | 0.0002 | 0.002 | | p-value between treatments | 0.9 | 0.009 | 0.8 | 0.05 | | Patient assessment motion pain | | | | | | Synvisc®-treated | 70.8 ± 4.7 | 23.5 ± 5.6 | 25.9 ± 6.3 | 37.5 ± 6.4 | | p-value from baseline | NA | 0.0002 | 0.0003 | 0.0001 | | Saline-treated | 66.9 ± 4.5 | 0.3 ± 5.4 | 18.1± 6.1 | 16.3 ± 6.0 | | p-value from baseline | NA | 1.0 | 0.006 | 0.01 | | p-value between treatments | 0.6 | 0.006 | 0.4 | 0.02 | | Patient assessment night pain | | | | | | Synvisc®-treated | 67.2 ± 4.6 | 24.9 ± 7.3 | 26.7 ± 7.9 | 42.8 ± 8.0 | | p-value from baseline | NA | 0.002 | 0.002 | 0.0001 | | Saline-treated | 83.1 ± 4.5 | 19.4 ± 7.1 | 37.0± 7.7 | 21.7± 7.5 | | p-value from baseline | NA | 0.01 | 0.0001 | 0.007 | | p-value between treatments | 0.02 | 0.6 | 0.4 | 0.06 | ¹ One Synvisc® patient &lt; 40 years old ² No correction for covariates ³ Mean of assessments on VAS of 0 to 100 mm ⁴ SEM = Standard Error of the Mean ⁵ NA = Not Applicable {31} TABLE 12 (continued) Study #5 Improvement from Baseline During Phase I Flare Population All Patients¹ Analysis of Variance² | | Baseline Mean³ ± SEM⁴ | Improvement (Change from Baseline) Mean ± SEM | | | | --- | --- | --- | --- | --- | | Week | 0 | 1 | 2 | 4 | | Patient assessment rest pain | | | | | | Synvisc®-treated | 67.8 ± 5.7 | 19.6 ± 7.0 | 30.3 ± 7.6 | 41.1 ± 7.1 | | p-value from baseline | NA⁵ | 0.009 | 0.0004 | 0.0001 | | Saline-treated | 66.0 ± 5.5 | 5.5 ± 6.8 | 21.7 ± 7.4 | 18.4 ± 6.7 | | p-value from baseline | NA | 0.4 | 0.006 | 0.01 | | p-value between treatments | 0.8 | 0.2 | 0.4 | 0.03 | | Patient assessment restriction of activity | | | | | | Synvisc®-treated | 70.8 ± 6.5 | 13.2 ± 3.6 | 22.1± 6.2 | 25.6 ± 4.7 | | p-value from baseline | NA | 0.001 | 0.001 | 0.0001 | | Saline-treated | 68.6 ± 6.3 | 9.8 ± 3.5 | 21.9 ± 6.0 | 18.8 ± 4.4 | | p-value from baseline | NA | 0.009 | 0.001 | 0.0002 | | p-value between treatments | 0.8 | 0.5 | 1.0 | 0.3 | | Patient overall assessment arthritic pain | | | | | | Synvisc®-treated | 77.3 ± 4.4 | 19.7 ± 4.4 | 30.6 ± 6.3 | 36.5± 6.6 | | p-value from baseline | NA | 0.0001 | 0.0001 | 0.0001 | | Saline-treated | 79.7 ± 4.3 | 8.5 ± 4.3 | 23.1 ± 6.1 | 17.9 ± 6.2 | | p-value from baseline | NA | 0.06 | 0.0007 | 0.007 | | p-value between treatments | 0.7 | 0.08 | 0.4 | 0.05 | ¹ One Synvisc® patient &lt; 40 years old ² No correction for covariates ³ Mean of assessments on VAS of 0 to 100 mm ⁴ SEM = Standard Error of the Mean ⁵ NA = Not Applicable {32} # TABLE 13 ## Study #6 ### Improvements from Baseline (Weeks 7 and 12) ### Patients ≥ 40 Years of Age ### Repeated Measures Analysis (PROC MIXED) | Outcome Measure: | Covariates^{1} | Synvisc® Mean^{2} ± SEM^{3} (n) | NSAIDs Mean ± SEM (n) | Synvisc® & NSAIDs Mean ± SEM (n) | p-value between groups | | | | --- | --- | --- | --- | --- | --- | --- | --- | | | | | | | Synvisc® vs. NSAIDs | Synvisc® & NSAIDs vs. NSAIDs | Synvisc® & NSAIDs vs. Synvisc® | | Motion Pain | Center, baseline, age, x-ray grade and duration of disease | 27 ± 5 (26) | 20 ± 4 (32) | 23 ± 4 (33) | 0.2 | 0.6 | 0.4 | | | p-value from baseline | 0.0001 | 0.0001 | 0.0001 | | | | | Night Pain | Center, baseline and duration of disease | 18 ± 3 (26) | 11 ± 3 (32) | 15 ± 3 (34) | 0.05 | 0.2 | 0.5 | | | p-value from baseline | 0.0001 | 0.0001 | 0.001 | | | | | Rest Pain | Baseline, duration of disease, contralateral knee and any concurrent therapy | 17 ± 3 (26) | 13 ± 3 (32) | 16 ± 3 (34) | 0.2 | 0.4 | 0.6 | | | p-value from baseline | 0.0001 | 0.001 | 0.0001 | | | | | Restriction of Activity | Baseline, age, x-ray grade and duration of disease | 16 ± 4 (26) | 15 ± 4 (32) | 13 ± 4 (33) | 0.9 | 0.7 | 0.6 | | | p-value from baseline | 0.0004 | 0.0003 | 0.001 | | | | | Pt. Evaluated Overall Pain | Baseline, age and duration of disease | 27 ± 4 (26) | 20 ± 4 (32) | 21 ± 4 (34) | 0.2 | 0.8 | 0.3 | | | p-value from baseline | 0.0001 | 0.0001 | 0.0001 | | | | 1. The analyses were corrected for all covariates found to be statistically significant by stepwise regression. Each significant covariate is listed in the table with the individual outcome measure. 2. VAS in mm 3. SEM = Standard Error of the Mean
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...