The enVista® One-Piece Hydrophobic Acrylic Toric IOL (Model MX60T) is indicated for primary implantation in the capsular bag of the eye in adult patients for visual correction of aphakia and corneal astigmatism following removal of a cataractous lens for improved uncorrected distance vision.
Device Story
The enVista® Model MX60T is a posterior chamber, hydrophobic acrylic, one-piece toric intraocular lens (IOL) designed for implantation in the capsular bag following cataract surgery. It features an aspheric, aberration-neutral optic and a 360-degree posterior squared step edge (SureEdge™) to inhibit posterior capsule opacification. The lens includes Accuset™ haptics for stability and posterior axis marks to guide alignment with the steep corneal meridian to correct astigmatism. Surgeons use a proprietary web-based 'enVista Toric calculator' to determine IOL selection and axis parameters. The device is implanted by ophthalmologists during standard phacoemulsification. By correcting both aphakia and corneal astigmatism, the lens improves uncorrected distance visual acuity, potentially reducing patient reliance on glasses or contact lenses post-surgery.
Clinical Evidence
Prospective, multicenter, parallel, partially randomized, double-masked clinical trial (IDE #G120193) with 191 subjects (112 toric, 79 control). Primary endpoints included axial stability, dioptric cylinder reduction, and BCDVA. At 6 months, 100% of toric lenses showed ≤ 5° rotation. Mean cylinder reduction was significantly higher in the 1.25 D toric group (0.865 D) vs. control (0.479 D) (p < 0.001). 94.5% of toric subjects achieved UCDVA of 20/40 or better. No secondary surgical interventions for rotational instability occurred. Safety profile was favorable with no increased incidence of adverse events compared to the control group.
Technological Characteristics
Hydrophobic acrylic (HEMA-poly(EG)PEA-styrene copolymer) one-piece lens; 12.5 mm overall diameter; 6.0 mm optic diameter; 0° haptic angulation; aspheric, aberration-neutral optic; 360° square posterior edge; refractive index 1.54 @ 35°C; UV-absorbing. Sterilized via radiation. Software: web-based Toric Calculator (Major Level of Concern).
Indications for Use
Indicated for adult patients undergoing cataract extraction for visual correction of aphakia and corneal astigmatism via primary implantation in the capsular bag.
Regulatory Classification
Identification
An intraocular lens is a device made of materials such as glass or plastic intended to be implanted to replace the natural lens of an eye.
Predicate Devices
enVista® One-Piece Hydrophobic Acrylic Intraocular Lens, Model MX60 (P910056/S010)
Submission Summary (Full Text)
{1}
# SUMMARY OF SAFETY AND EFFECTIVENESS DATA (SSED)
## I. GENERAL INFORMATION
Device Generic Name: Toric Posterior Chamber Intraocular Lens (IOL)Device Trade Name: enVista® One-Piece Hydrophobic Acrylic Toric Intraocular Lens; Model MX60TDevice Procode: MJPApplicant's Name and Address: Bausch + Lomb, Inc.
400 Somerset Corporate Blvd.
Bridgewater, NJ 08807Date(s) of Panel Recommendation: NonePremarket Approval Application (PMA) Number: P910056/S027Date of FDA Notice of Approval: June 8, 2018
The original PMA P910056 was approved on January 17, 1992 and incorporated by reference here. The enVista® One-Piece Hydrophobic Acrylic Toric Intraocular Lens; Model MX60T is a modified version of the enVista® One-Piece Acrylic Intraocular Lens, Model MX60 approved under PMA P910056/S010 on April 30, 2012 and incorporated by reference here. . The current supplement was submitted to expand the indication for the enVista® Posterior Chamber Intraocular Lens to include correction of aphakia and astigmatism.
## II. INDICATIONS FOR USE
The enVista® One-Piece Hydrophobic Acrylic Toric IOL (Model MX60T) is indicated for primary implantation in the capsular bag of the eye in adult patients for visual correction of aphakia and corneal astigmatism following removal of a cataractous lens for improved uncorrected distance vision.
## III. CONTRAINDICATIONS
There are no known contraindications.
## IV. WARNINGS AND PRECAUTIONS
PMA P910056/S027: FDA Summary of Safety and Effectiveness Data
Page 1
{2}
The warnings and precautions can be found in the enVista® Toric IOL labeling packaged with the device.
# V. DEVICE DESCRIPTION
The enVista® One-Piece Hydrophobic Acrylic Toric Intraocular Lens, Model MX60T (see Figure 1, below), is an ultraviolet-absorbing, posterior chamber, hydrophobic acrylic one-piece astigmatism-correcting intraocular lens. It has an overall diameter of 12.5 mm, and an optic body diameter of 6.0 mm. The enVista® Toric IOL has an aspheric optic that is designed to be free of spherical aberration through the incorporation of a proprietary spherical aberration-neutral optic design that does not influence the pre-surgical corneal spherical aberration profile. The optic is designed with the SureEdge™, posterior squared step edge to provide a 360 degree posterior capsule opacification (PCO) barrier. The enVista® Toric IOL employs an Accuset™ haptic with a broad, modified C-loop design and optic-haptic offset to facilitate improved contact and stability within the capsular bag. The posterior located cylinder axis marks denote the meridian with the lowest power. The enVista® material that makes up the TruSight™ optic has been assessed for glistening-free capacity and scratch resistance. The enVista® One-Piece Hydrophobic Acrylic Toric Intraocular Lens, Model MX60T, is a modification to the B+L enVista® One-Piece Hydrophobic Acrylic Intraocular Lens, Model MX60. The only difference between the enVista® Toric IOL and the currently approved enVista® IOL is the incorporation of a toric optic surface and axis marks on the posterior optic surface of the lens. The lens toricity results in a periodic change in refractive power between the flat and steep lens meridians. When the flat meridian of the toric IOL is correctly aligned with the steep corneal meridian, the toric IOL corrects the corneal astigmatism. The enVista® Toric Intraocular Lens is available in seven different cylinder powers (+1.25 D, +2.00 D, +2.75 D, +3.50 D, +4.25 D, +5.00 D and +5.75 D) and in spherical equivalent powers of +6.0 D to +30.0 D, in 0.5 D increments. To facilitate IOL selection and axis parameters, B+L provides a web-based proprietary tool, the enVista Toric calculator.

Figure 1: Schematic of MX60T IOL
PMA P910056/S027: FDA Summary of Safety and Effectiveness Data
Page 2
{3}
The physical characteristics of the enVista® Toric Intraocular Lens are shown in Table 1.
Table 1: PHYSICAL CHARACTERISTICS OF ENVISTA TORIC INTRAOCULAR LENS
| Characteristic | Model MX60T |
| --- | --- |
| Optic Body Diameter | 6.0 mm |
| Overall Diameter | 12.5 mm |
| Haptic Angle | 0° |
| Lens Configuration | One-Piece Modified C-loop Haptics; Zero Angulation; Step-Vaulted 360° square posterior edge |
| Anterior Surface | Aspheric Optic |
| Posterior Surface | Toric Optic with Cylinder Axis Marks 360° Square Edge |
| Lens / Haptic Material | Hydrophobic Acrylic (hydroxyethyl methacrylate (HEMA)-polyethylene glycol phenyl ether acrylate (poly(EG)PEA)-styrene copolymer, cross-linked with ethylene glycol dimethacrylate) |
| Spherical Equivalent Power | +6.0 D to +30.0 D, in 0.5 D Increments. |
| Cylinder Power (Diopter, D) | (+1.25 D, +2.00 D, +2.75 D, +3.50 D, +4.25 D, +5.00 D and +5.75 D) |
| Refractive Index | 1.54 @ 35°C |
| UV Transmittance | 363 nm |
| Suggested A-Constant | 119.1 (Optical Biometry) |
# VI. ALTERNATIVE PRACTICES AND PROCEDURES
There are several other alternatives for the correction of restoring functional vision of the aphakic eye. Non-surgical options include special cataract glasses or contact lenses. Surgical options such as monofocal, multifocal, simultaneous vision or accommodative IOLs are also available. Each alternative has its own advantages and disadvantages. A patient should fully discuss these alternatives with his/her physician to select the method that best meets expectations and lifestyle.
PMA P910056/S027: FDA Summary of Safety and Effectiveness Data
Page 3
{4}
## **VII. MARKETING HISTORY**
The enVista® One-Piece Hydrophobic Acrylic Toric IOL has not been marketed in the United States or any foreign country although it has been available in the European Union since receiving the CE Mark in April, 2012. Subsequently, the product has been marketed in Canada, Latin America, and in the Asia Pacific region. The enVista® One-Piece Hydrophobic Acrylic Toric IOL has not been withdrawn from marketing in any country for any reason.
## **VIII. PROBABLE ADVERSE EFFECTS OF THE DEVICE ON HEALTH**
Below is a list of the probable adverse effects (e.g., complications) associated with the use of the device. Potential adverse events and complications accompanying cataract or implant surgery may include, but are not limited to the following: corneal endothelial damage, infection (endophthalmitis), retinal detachment, vitritis, cystoid macular edema, corneal edema, pupillary block, cyclitic membrane, iris prolapsed, hypopyon, transient or persistent glaucoma, acute corneal decompensation, toxic anterior segment syndrome (TASS), and secondary surgical intervention.
Secondary surgical interventions include, but are not limited to, lens repositioning, lens replacement, vitreous aspiration or iridectomy for pupillary block, wound leak repair, and retinal detachment repair.
For the specific adverse events that occurred in the clinical studies, please see Section 10.
## **IX. SUMMARY OF NONCLINICAL STUDIES**
The enVista® Toric IOL is composed of identical materials to the currently approved enVista® One-Piece Hydrophobic Acrylic IOL, MX60 (material parent, P910056/S010). Preclinical studies were performed on the material parent device and are incorporated by referenced. These tests are as follows: physicochemical and biocompatibility testing (with the exception of inorganic insolubility testing), sterilization, packaging, shelf-life, and transport stability testing. Insoluble inorganics testing and results are shown in **Table 2**.
### **Table 2: PHYSICAL-CHEMICAL AND BIOLOGICAL TESTING**
PMA P910056/S027: FDA Summary of Safety and Effectiveness Data
Page 4
{5}
| ISO 11979-5 | Insoluble Inorganics | Aluminum Assay Extraction | Passed |
| --- | --- | --- | --- |
| | | Using ICP-MS or ICP-OES on digested material to detect material on and in the lens | Passed |
### *Optical and Mechanical Testing*
The optical properties of the Model MX60T were tested per “ISO 11979-2: Ophthalmic implants – Intraocular lenses: Optical properties and test methods and ANSI Z80.30: Toric intraocular lenses.” The mechanical properties were tested per “ISO 11979-3: Ophthalmic implants – Intraocular lenses: mechanical properties and test methods.”
The lens testing was performed using fully processed and sterilized Model MX60T lenses supplied from product manufactured at the B+L Clearwater, Florida manufacturing facility as part of the process validation of the Model MX60T.
Testing was performed on lenses representing the low, medium and the high diopter ranges: (spherical equivalent power (SE) = +6.0 D with cylinder = 1.25 D; SE = +20.0 D with cylinder = 2.0 D; SE = +30.0 D with cylinder = 2.75 D; SE=+6.0 D with cylinder = 3.50; SE = + 21.0 with cylinder 4.25 and SE = + 30.0 D with cylinder = 5.75D.). The results of the testing demonstrate that the Model MX60T meets the ISO and ANSI standards for optical and mechanical properties and qualifies as a Level A modification to the approved enVista® Model MX60 per ISO/TR 22979. The results of the testing demonstrate that the Model MX60T meets all of the ISO and ANSI standards for optical and mechanical properties, as summarized in **Table 3** and **Table 4**.
**Table 3: SUMMARY OF OPTICAL TESTING FOR MODEL MX60T - ISO 11979-2 & ANSI Z80.30**
| SUMMARY OF OPTICAL TESTING FOR MODEL MX60T - ISO 11979-2 & ANSI Z80.30 | | |
| --- | --- | --- |
| Optical Property (Design criteria) | Requirement (Design Input) | Results (Design Output) |
| Dioptric Power (SE) | +6.0 ≤ SE ≤ + 15.0D: ± 0.3D + 15.0 < SE ≤ + 25.0D: ± 0.4D + 25.0 < SE ≤ + 30.0D: ± 0.5D | Passed |
PMA P910056/S027: FDA Summary of Safety and Effectiveness Data
Page 5
{6}
# **SUMMARY OF OPTICAL TESTING FOR MODEL MX60T - ISO 11979-2 & ANSI Z80.30**
| Optical Property (Design criteria) | Requirement (Design Input) | Results (Design Output) |
| --- | --- | --- |
| Cylinder Power | Cylinder SKU range (D) | Cylinder Acceptance Criteria |
| | +1.25 and +2.00 Cylinders | SE +6.0 to < +25.0 ± 0.3D SE ≥+25.0 to ≤ +30.0 ± 0.4D Passed |
| | +2.75, +3.50, and +4.25 Cylinders | SE +6.0 to < +25.0 ± 0.4D SE +≥25.0 to ≤ +30.0 ± 0.4D Passed |
| | +5.00 and +5.75 Cylinders | SE +6.0 to < +25.0 ± 0.5D SE ≥ +25.0 to ≤ +30.0 ± 0.5D Passed |
| Axis Orientation Marks | ± 5.0 Degrees from Toric Meridian | Passed |
| Imaging Quality | MTF ≥ 0.08 @ 70.0% cut-off freq. MTF ≥ 0.36 @ 36.9% cut-off freq. (Equivalent to 0.43 @ 100 lp/mm per ISO 11979-2) | Passed |
| Spherical Aberration | 0 ± 0.05 µm | Passed |
PMA P910056/S027: FDA Summary of Safety and Effectiveness Data
Page 6
{7}
**Table 4: SUMMARY OF MECHANICAL TESTING FOR MODEL MX60T - ISO 11979-3**
| SUMMARY OF MECHANICAL TESTING FOR MODEL MX60T - ISO 11979-3 | | |
| --- | --- | --- |
| Mechanical Property (Design Criteria) | Requirement (Design Input) | Results (Design Output) |
| Clear Optic Diameter | ≥ 4.25 mm | Passed |
| Dimensions: • Overall Diameter • Optic Diameter • Vault Height • Sagitta | • 12.5 mm ± 0.2 mm • 6.0 mm ± 0.1 mm • Satisfy dimension criteria • Satisfy dimension criteria | Passed |
| Surface and Bulk Homogeneity | Free of surface and bulk defects at 10x magnification | Passed |
| Optic Decentration | Sum of mean and two standard deviations must not exceed 10% of clear optic | Passed |
| Optic Tilt | Sum of mean and two standard deviations must not exceed 5° | Passed |
| Axial Displacement in Compression | Measured and reported at same diameters used for compression force | Complete |
| Haptic Pull | Withstand a force of ≥0.25 N | Passed |
| Contact Angle | Informational only | 56° |
| | | 56° (post decay) |
| Compression Force | Informational only | 0.125 g |
| | | 0.119 g (post decay) |
| Mechanical Characterization | Contact Angle within 40% of parent at 10 mm compressed diameter | Passed |
| | UFB/AC and LFB/AC overlaps with parent model | Passed |
| Dynamic Fatigue Durability Testing | Withstand 250,000 cycles (± 0.25 mm; 1 Hz and 10 Hz) | Passed |
| Fold & Recovery | After release from deformed state, return to specifications within 24 hrs. | Passed |
| Spectral Transmittance | Record Range 300 nm to 1100 nm | Complete |
PMA P910056/S027: FDA Summary of Safety and Effectiveness Data
Page 7
{8}
*Standards*
Standards used to conduct the testing presented in this section were conducted in accordance with the standards cited in **Table 3** as well as:
- EN ISO 11137-1: 2012-Sterilization of health care products – Radiation – Part 1: Requirements for development, validation, and routine control of a sterilization process for medical devices
- EN ISO 11137-2: 2015-Sterilization of health care products – Radiation – Part 2: Establishing a sterilization dose
- EN11607-1: 2009+A1:2014-Packaging for terminally sterilized medical devices – Part 1: Requirements for materials, sterile barrier systems and packaging systems
- EN 11607-2:2006+A1:2014-Packaging for terminally sterilized medical devices – Part 2: Validation requirements for forming, sealing and assembly processes
- DIN EN 11737-1: 2016-Sterilization of medical devices – Microbiological methods – Part 1: Determination of a population of microorganisms on products
- EN 11737-2:2015-Sterilization of medical devices – Microbiological methods – Part 2: Tests of sterility performed in the definition, validation and maintenance of a sterilization process
**A. Additional Studies**
*Software Validation*
A software validation was performed for the B+L Toric Calculator per the procedures described in FDA's Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices, Software Level of Concern: MAJOR. The software for this device was developed under an appropriate software development program. A hazard analysis was performed from both the patient's and user's standpoint and all identified
PMA P910056/S027: FDA Summary of Safety and Effectiveness Data
Page 8
{9}
hazards were addressed. These procedures provide the foundation for assuring, to the extent possible, that the software will operate in a manner described in the specifications, and in no other way. In addition, software standard, ISO 63204:2006 - Medical device software – Software Lifecycle Processes was followed.
## **X. SUMMARY OF PRIMARY CLINICAL STUDY**
The applicant performed a clinical study to establish a reasonable assurance of safety and effectiveness of subjects undergoing cataract extraction and IOL implantation with the enVista® Toric IOL, Model MX60T for primary implantation in the capsular bag of the eye in adult subjects for visual correction of aphakia and corneal astigmatism following removal of a cataractous lens for improved uncorrected distance vision in the US under IDE # G120193. Data from this clinical study were the basis for the PMA approval decision. A summary of the clinical study is presented below.
### **A. Study Design**
Subjects were treated between March 6, 2013 and June 2, 2016. The database for this Panel Track Supplement (PTS) reflected data collected through February 3, 2017 and included 191 subjects. There were 9 investigational sites.
The study was a prospective, multicenter, parallel, partially randomized, partially controlled, double-masked, monocular clinical trial. The study was designed to evaluate the safety and effectiveness of the enVista® Toric IOL (Model MX60T) in reducing postoperative refractive astigmatism in subjects undergoing cataract extraction. IOL implantations were conducted in the US with study duration of 6 months for subjects. Subjects in the lowest astigmatic power (1.25 D) cohort were randomized to undergo implantation of either the toric test lens or the non-toric spherical control lens (enVista® One-Piece Hydrophobic IOL Model MX60) in one eye. Subjects in the non-randomized higher astigmatic power cohorts (2.00 D, 2.75 D) were implanted with the test lens in one eye.
The control group were fellow aphakic subjects that only had spherical corrective lenses implanted (one-piece monofocal IOL; Model MX60).
PMA P910056/S027: FDA Summary of Safety and Effectiveness Data
Page 9
{10}
# 1. Clinical Inclusion and Exclusion Criteria
Enrollment in the enVista® Toric IOL study was limited to subjects who met the following inclusion criteria:
- Clinically documented diagnosis of age-related cataract (cortical, nuclear, subcapsular, or a combination) considered amenable to treatment with standard phacoemulsification / extracapsular cataract extraction,
- 18 years of age or older and met any applicable local minimum age requirements for IOL implantation following cataract surgery,
- Clear intraocular media other than cataract in study eye,
- BCVA equal to or worse than 20/40, with or without a glare source,
- Ability to return for scheduled follow-up examinations and had mental capacity to cooperate when undergoing a detailed postoperative exam,
- Contact lens wearers demonstrate stable refraction (within ± 0.50 D) in operative eye, determined by manifest refraction spherical equivalent (MRSE) on 2 consecutive exam dates, under the following conditions:
- Lenses not worn for at least 2 weeks (rigid or toric contact lenses) or 3 days (soft contact lenses),
- The 2 refractions were performed at least 7 days apart,
- Preoperative dilated pupil size greater than or equal to 5.0 mm,
- Underwent primary IOL implantation for correction of aphakia following continuous curvilinear anterior capsulotomy and phacoemulsification cataract extraction,
- Required spherical lens power of 16.00 D to 27.00 D,
- Minimum preoperative corneal astigmatism of 0.90 D in operated eye,
- Predicted postoperative corneal astigmatism between 0.90 D and 2.40 D, as determined by the toric calculator, and
- Potential for BCDVA of 20/30 or better in the operated eye (determined by medical judgment or measured by retinal acuity meter / potential acuity meter testing).
Subjects were not permitted to enroll in the enVista® Toric IOL study if they met any of the following exclusion criteria:
PMA P910056/S027: FDA Summary of Safety and Effectiveness Data
Page 10
{11}
- Anterior segment pathology for which extracapsular phacoemulsification cataract surgery would be contraindicated (e.g., chronic uveitis, iritis, iridocyclitis, rubeosis, iridis, clinically significant corneal dystrophy, clinically significant Fuch's dystrophy, clinically significant anterior membrane dystrophy)
- Corneal pathology potentially affecting topography
- Diagnoses of degenerative visual disorders (e.g., macular degeneration or other retinal disorders) predicted to cause future acuity losses to a level of 20/30 or worse
- Conditions associated with increased risk of zonular rupture (zonule rupture during cataract extraction procedure that may affect postoperative rotation, centration, or tilt of the lens)
- Inability to achieve pupil dilation of 5.0 mm
- Any inflammation or edema (swelling) of the cornea, including but not limited to keratitis, keratoconjunctivitis and keratouveitis
- Any ocular condition (e.g., amblyopia) that could prevent the possibility of a visual outcome better than 20/30 in the operative eye
- Presence of rubella, bilateral congenital, traumatic or complicated cataract
- Previous intraocular or corneal surgery
- Uncontrolled glaucoma
- Prior retinal detachment
- Irregular corneal astigmatism
- Diabetic retinopathy (proliferative or nonproliferative)
- Marked microphthalmos or aniridia
- Clinically significant retinal pigment or epithelium/macular changes
- Iris or chorioretinal neovascularization
- Presence of optic atrophy
- Chronic use of systemic steroids or immunosuppressive medications
PMA P910056/S027: FDA Summary of Safety and Effectiveness Data
Page 11
{12}
- Concurrent participation in another clinical trial or participation in another clinical trial within 30 days prior to enrollment in this study
- Difference in corneal astigmatism measured with the IOL Master and the topographer greater than 0.50 D using vector analysis
# 2. Follow-up Schedule
All subjects were scheduled to return for follow-up examinations at:
- Postoperative Visit 1 (Form 1) Day 1 to 2
- Postoperative Visit 2 (Form 2) Day 7 to 14
- Postoperative Visit 3 (Form 3) Day 30 to 60
- Postoperative Visit 4 (Form 4) Day 120 to 180
Preoperatively, subjects were evaluated to obtain a medical history and to establish a baseline for their ocular condition. Subjects in the lowest astigmatic power group were randomized to undergo implantation of either the test or control lens. No corneal or refractive procedures (such as limbal relaxing incisions or astigmatic keratotomies) were permitted at any time during the course of the study. Postoperatively, the objective parameters measured during the study included the following:
- Analysis of IOL axial stability included continuous summary statistics, 95% confidence intervals (CIs) around the mean, and the proportion of lenses with axial rotation ≤ to 5° for cylindrical powers of 1.25 D, 2.00 D and 2.75 D.
- Analysis of the dioptric reduction in cylinder included continuous summary statistics. Additionally, a statistical model including the effects of treatment and site was constructed to test the null hypothesis that the dioptric reduction in cylinder of the eyes implanted with IOL cylinder power 1.25 D was less than or equal to the dioptric reduction in cylinder of the eyes implanted with the control IOL.
- Analysis of the percent reduction in absolute cylinder included continuous summary statistics and 95% confidence intervals (CIs) around the mean. For the endpoint of the percent of eyes with reduction of cylinder within 0.50 D and 1.00 D of target, categorical summary statistics were provided, including exact 95% binomial CIs around the percent of eyes within 0.50
PMA P910056/S027: FDA Summary of Safety and Effectiveness Data
Page 12
{13}
D and 1.00 D.
- Analysis of lens axis misalignment included continuous summaries of the absolute value of the misalignment, the signed value of the misalignment (including a 2-sided 95% tolerance interval, which contains at least 90% of the population), and the proportion of lenses with axis misalignment in the following categories: ≤ 5, ≥ 10, ≤ 20, ≤ 30, and > 30 degrees.
- Analysis of the best corrected distance visual acuity (BCDVA) includes a comparison to historical controls to test the null hypothesis that the true rate of cases with BCDVA 0.3 logMAR or better is greater than or equal to the historical control rate.
- Analysis of uncorrected distance visual acuity (UCDVA), expressed as the logarithm of the minimum angle of resolution, logMAR, includes a comparison of the Toric 1.25 D arm to the spherical control group to test the null hypothesis that the true UCDVA of the Toric 1.25 D eyes is equivalent to the mean logMAR UCDVA of the spherical control group.
Adverse events and complications were recorded at all visits.
The key timepoints are shown below in the tables summarizing safety and effectiveness.
### 3. Clinical Endpoints
With regards to safety, the study endpoints were the incidence of cumulative and persistent adverse events for eyes implanted with toric IOLs.
With regards to effectiveness, primary endpoints were:
- IOL axial stability, expressed as the degree of rotation between consecutive visits,
- Dioptric reduction in cylinder,
- Percent reduction in absolute cylinder, expressed as a percentage of the intended reduction in cylinder, calculated as:
$$\frac{(\text{Postoperative Manifest Refractive Cylinder}) - \text{Preoperative Keratometric Cylinder}}{(\text{Intended Postoperative Manifest Refractive Cylinder}) - \text{Preoperative Keratometric Cylinder}} * 100$$
- Percent of eyes with “reduction of cylinder” within 0.50 D and within 1.00 D of intended,
PMA P910056/S027: FDA Summary of Safety and Effectiveness Data
Page 13
{14}
- Lens axis misalignment as determined by a direct measurement method; and
- Best corrected distance visual acuity (BCDVA).
A secondary effectiveness endpoint was:
- Uncorrected distance visual acuity (UCDVA).
### B. Accountability of PMA Cohort
At the time of database lock, of 191 subjects enrolled in the PMA study, 98.4% (188/191) subjects are available for analysis at the completion of the study, the 6 month post-operative visit. Refer to Table 5 below for subject accountability by each scheduled visit.
Table 5: ACCOUNTABILITY OF SUBJECTS AT EACH FORM VISIT (ALL ENROLLED)
| Subject Status | Op Visit n (%) | Form 1 n (%) | Form 2 n (%) | Form 3 n (%) | Form 4 n (%) |
| --- | --- | --- | --- | --- | --- |
| Available for Analysis | 191 (100.0) | 190 (99.5) | 189 (99.0) | 188 (98.4) | 188 (98.4) |
| Discontinued | 0 | 0 | 0 | 0 | 1 (0.5) |
| Missing at scheduled visit but seen later | 0 | 1 (0.5) | 2 (1.0) | 3 (1.6) | 1 (0.5) |
| Lost to follow-up | 0 | 0 | 0 | 0 | 1 (0.5) |
| Active | 0 | 0 | 0 | 0 | 0 |
| Percent Accountability | - | 190/191 (99.5) | 189/191 (99.0) | 188/191 (98.4) | 188/190 (98.9) |
Abbreviations: Op = Operative; n = number of subjects at a study visit in a stated category
Figure 2: summarizes the analysis of the various populations used in this study: All Enrolled Cohort, Safety Cohort, Repositioned Lens Cohort, Effectiveness ITT Cohort, Best Case Cohort; Not Best Case Cohort; and Per Protocol Cohort.
PMA P910056/S027: FDA Summary of Safety and Effectiveness Data
Page 14
{15}
**Figure 2: STUDY POPULATIONS ANALYSIS FLOW CHART**

*A total of 4 subject eyes (1 toric, 3 control) reporting 4 ocular pathologies were included in the Not Best Case Cohort. Note: n=Control/Toric 1.25 D/Toric 2.00 D/Toric 2.75 D IOL
### C. Study Population Demographics and Baseline Parameters
The demographics (Table 6) of the study population are typical for an IOL study performed in the US. In total, 89 subjects were male, and 102 subjects were female. The mean age for subjects in the Control Cohort was 69.5 years and the mean age for subjects in the All Toric Cohort was 70.5 years. The age range for all subjects was 40 to 89 years. With regard to gender, the majority for the Control Cohort were male at 54.4% (43/79) while the majority for the All Toric Cohort was female at 58.9% (66/112).
PMA P910056/S027: FDA Summary of Safety and Effectiveness Data
Page 15
{16}
**Table 6: DEMOGRAPHICS (ALL ENROLLED)**
| | Control IOL (N=79) | Toric IOL 1.25 D (N=80) | Toric IOL 2.00 D (N=20) | Toric IOL 2.75 D (N=12) | All Toric IOL (N=112) |
| --- | --- | --- | --- | --- | --- |
| **Age (years)** | | | | | |
| Total Non-Missing | 79 | 80 | 20 | 12 | 112 |
| Mean (SD) | 69.5 (8.6) | 71.1 (8.8) | 68.7 (9.1) | 69.6 (8.8) | 70.5 (8.9) |
| Median | 69.0 | 71.0 | 67.0 | 67.5 | 70.0 |
| Min, Max | 52, 87 | 40, 89 | 58, 86 | 56, 85 | 40, 89 |
| **Age Category, n (%)** | | | | | |
| < 60 | 11 (13.9%) | 7 (8.8%) | 3 (15.0%) | 1 (8.3%) | 11 (9.8%) |
| 60 to 69 | 29 (36.7%) | 24 (30.0%) | 10 (50.0%) | 6 (50.0%) | 40 (35.7%) |
| 70 to 79 | 28 (35.4%) | 35 (43.8%) | 4 (20.0%) | 3 (25.0%) | 42 (37.5%) |
| ≥ 80 | 11 (13.9%) | 14 (17.5%) | 3 (15.0%) | 2 (16.7%) | 19 (17.0%) |
| **Gender, n (%)** | | | | | |
| Total Non-Missing | 79 | 80 | 20 | 12 | 112 |
| Male | 43 (54.4%) | 31 (38.8%) | 12 (60.0%) | 3 (25.0%) | 46 (41.1%) |
| Female | 36 (45.6%) | 49 (61.3%) | 8 (40.0%) | 9 (75.0%) | 66 (58.9%) |
| **Operative Eye, n (%)** | | | | | |
| Total Non-Missing | 79 | 80 | 20 | 12 | 112 |
| Oculus Dexter (OD) | 40 (50.6%) | 48 (60.0%) | 12 (60.0%) | 8 (66.7%) | 68 (60.7%) |
| Oculus Sinister (OS) | 39 (49.4%) | 32 (40.0%) | 8 (40.0%) | 4 (33.3%) | 44 (39.3%) |
#### **D. Safety and Effectiveness Results**
##### **1. Safety Results**
The analysis of safety was based on the safety cohort of 188 implanted subject eyes available for the 6-month evaluation. The key safety outcomes for this study are presented below in **Table 7** and **Table 8**. Adverse effects are reported in Tables 7 to 9.
PMA P910056/S027: FDA Summary of Safety and Effectiveness Data
Page 16
{17}
# **Adverse effects that occurred in the PMA clinical study:**
**Table 7: ISO 11979-7 SAFETY AND PERFORMANCE ENDPOINTS (SPE) ADVERSE EVENTS REPORTED AT EACH POSTOPERATIVE VISIT, IMPLANTED SUBJECTS (SAFETY, ALL TORIC IOL)**
| Adverse Event | Form 1 n/N (%)^{1} | Form 2 n/N (%)^{1} | Form 3 n/N (%)^{1} | Form 4 n/N (%)^{1} | Cumulative n/N (%)^{1} | ISO 11979-7 SPE rate (%) | p-value |
| --- | --- | --- | --- | --- | --- | --- | --- |
| Endophthalmitis | 0/112 | 0/112 | 0/112 | 0/112 | 0/112 | 0.1 | >0.999 |
| Hypopyon | 0/112 | 0/112 | 0/112 | 0/112 | 0/112 | 0.3 | >0.999 |
| Lens Dislocated from Posterior Chamber | 0/112 | 0/112 | 0/112 | 0/112 | 0/112 | 0.1 | >0.999 |
| Cystoid Macular Edema | 0/112 | 1/112 (0.9) | 1/112 (0.9) | 0/112 | 2/112 (1.8) | 3.0 | 0.853 |
| Pupillary Block | 0/112 | 0/112 | 0/112 | 0/112 | 0/112 | 0.1 | >0.999 |
| Retinal Detachment | 0/112 | 0/112 | 0/112 | 0/112 | 0/112 | 0.3 | >0.999 |
| Secondary Surgical Intervention | 0/112 | 0/112 | 0/112 | 0/112 | 0/112 | 0.8 | >0.999 |
| **Persistent Adverse Events Noted at Form 4^{2}** | | | | | | | |
| Corneal Stromal Edema | | | | 0/112 | | 0.3 | >0.999 |
| Iritis | | | | 0/112 | | 0.3 | >0.999 |
| Cystoid Macular Edema | | | | 0/112 | | 0.5 | >0.999 |
| Raised IOP Requiring Treatment | | | | 0/112 | | 0.4 | >0.999 |
$^{1}$Only non-zero percent values are shown.
$^{2}$Cumulative versus persistent AEs are defined by the FDA SPE Grid and ISO 11979-7 as those occurring at Form 4 in this clinical study.
PMA P910056/S027: FDA Summary of Safety and Effectiveness Data
Page 17
{18}
**Table 8: ISO 11979-7 SPE ADVERSE EVENTS REPORTED AT EACH POSTOPERATIVE VISIT, IMPLANTED SUBJECTS (SAFETY, CONTROL IOL)**
| Adverse Event | Form 1 n/N (%) | Form 2 n/N (%) | Form 3 n/N (%) | Form 4 n/N (%) | ISO SPE % | p-Value |
| --- | --- | --- | --- | --- | --- | --- |
| Endophthalmitis | 0/79 | 0/79 | 0/79 | 0/79 | 0.1 | >0.999 |
| Hypopyon | 0/79 | 0/79 | 0/79 | 0/79 | 0.3 | >0.999 |
| Lens Dislocated from Posterior | 0/79 | 0/79 | 0/79 | 0/79 | 0.1 | >0.999 |
| Cystoid Macular Edema | 0/79 | 0/79 | 0/79 | 2/79 (2.5) | 3.0 | 0.853 |
| Pupillary Block | 0/79 | 0/79 | 0/79 | 0/79 | 0.1 | >0.999 |
| Retinal Detachment | 0/79 | 0/79 | 0/79 | 0/79 | 0.3 | >0.999 |
| Secondary Surgical Intervention | 0/79 | 0/79 | 0/79 | 0/79 | 0.8 | >0.999 |
| **Persistent Adverse Events Noted at Form 4^{2}** | | | | | | |
| Corneal Stromal Edema | | | | 0/79 | 0.3 | >0.999 |
| Iritis | | | | 0/79 | 0.3 | >0.999 |
| Cystoid Macular Edema | | | | 1/79 (1.3) | 0.5 | >0.999 |
| Raised IOP Requiring Treatment | | | | 0/79 | 0.4 | >0.999 |
**Note:** Only non-zero percent values are shown
PMA P910056/S027: FDA Summary of Safety and Effectiveness Data
Page 18
{19}
**Table 9: MEAN TORIC LENS AXIS MISALIGNMENT FROM SURGICAL MARKINGS AT FORM 4 (ITT POPULATION)**
| Axis Misalignment | Toric IOL | | | All Toric IOL (N=112) |
| --- | --- | --- | --- | --- |
| | 1.25 D (N=80) | 2.00 D (N=20) | 2.75 D (N=12) | |
| **Form 4 signed axis misalignment, degrees** | | | | |
| n | 77 | 16 | 11 | 104 |
| Mean ±SD | 1.11±8.69 | 3.08±10.51 | 2.52±3.03 | 1.56±8.57 |
| 95% tolerance interval for 90% of the population | -15.51 to 17.73 | -22.54 to 28.70 | -5.78 to 10.82 | -14.45 to 17.57 |
| **Form 4 absolute misalignment, degrees** | | | | |
| n | 77 | 16 | 11 | 104 |
| Mean ±SD | 4.77±7.33 | 5.15±9.61 | 3.32±2.01 | 4.68±7.33 |
| 95% tolerance interval for 90% of the population | -9.25 to 18.79 | -18.28 to 28.58 | -2.19 to 8.83 | -9.02 to 18.38 |
| **Form 4 signed axis misalignment category, n (%)** | | | | |
| ≤ 5 degrees | 56 (72.7%) | 13 (81.3%) | 9 (81.8%) | 78 (75.0%) |
| ≤ 10 degrees | 71 (92.2%) | 15 (93.8%) | 11 (100.0%) | 97 (93.3%) |
| **Form 4 signed axis misalignment category, n (%)** | | | | |
| -10.00 to -5.01 degrees | 4 (5.2%) | 2 (12.5%) | 0 (0.0%) | 6 (5.8%) |
| -5.00 to -0.01 degrees | 31 (40.3%) | 3 (18.8%) | 2 (18.2%) | 36 (34.6%) |
| 0.00 degrees | 0 (0.0%) | 0 (0.0%) | 0 (0.0%) | 0 (0.0%) |
| +0.001 to +5.00 degrees | 25 (32.5%) | 10 (62.5%) | 7 (63.6%) | 42 (40.4%) |
| +5.01 to 10.00 degrees | 11 (14.3%) | 0 (0.0%) | 2 (18.2%) | 13 (12.5%) |
## 2. Effectiveness Results
The analysis of effectiveness was based on the 188 evaluable subjects at the 6-month time point. Key effectiveness outcomes are presented in Tables 11 to 14.
The primary analysis of effectiveness was based on the 191 intent-to-treat (ITT) subjects who were allocated to a randomly assigned treatment IOL or the non-randomized cohort, and for whom phacoemulsification was successfully completed. Effectiveness analyses were also conducted using the 188 subjects in the per protocol (PP) population who were implanted with a study lens and who had no major protocol deviations. The effective corneal powers for each of the test lens plane cylindrical powers of the test IOLs are shown in **Table 10**. Key effectiveness outcomes are presented in **Table 11**, **Table 12**, **Table 13**, **Table 14**, and **Table 15**. The results achieved for 191 subject eyes followed
PMA P910056/S027: FDA Summary of Safety and Effectiveness Data
Page 19
{20}
postoperatively for six months provide data to support the conclusion that eyes implanted with an enVista® Toric IOL following cataract extraction achieve adequate visual correction of aphakia and astigmatism.
The data for subject eyes specifically support a significant dioptric reduction in cylinder and reduction in absolute cylinder, rotational stability of the lens, and improvement of both best corrected and uncorrected visual acuity at distance following implantation of an enVista® Toric IOL.
A primary effectiveness endpoint was mean Toric IOL axial stability from Form 3 to Form 4. Table 12 shows the absolute axial stability from Form 3 to Form 4 for subjects who received an enVista® Toric IOL in the ITT population, with 95% confidence intervals and notation of the percent of subjects in each Toric IOL treatment group who demonstrated ≤ 5 degrees rotation during this interval. All subjects in the Toric IOL treatment groups demonstrated ≤ 5 degrees rotation from Form 3, including those subjects with missing data for which 5 rounds of multiple data imputation were used to simulate the entire Toric IOL ITT population.
The percent of eyes within 0.50 D and 1.00 D of intended correction for the enVista® Toric Posterior Chamber Intraocular Lens (All Toric Cohort) at Form 4 was 57.3% (63/112) and 90.9% (100/112), respectively. The minimum preoperative keratometric cylinder treated with a toric IOL was 1.24 D. The results for percent of non-missing eyes in the ITT population with reduction in cylinder within 0.50 D and 1.00 D of intended cylinder are shown in Table 11.
Table 10: ENVISTA TORIC IOL CYLINDER POWER
| Cylinder Power at IOL Plane (D) | Cylinder Power at Corneal Plane (D) | Range of Predicted Postoperative Corneal Cylinder^{1} (D) |
| --- | --- | --- |
| 1.25 | 0.90 | 0.90-1.39 |
| 2.00 | 1.40 | 1.40-1.92 |
| 2.75 | 1.93 | 1.93-2.40 |
| ^{1} Each Surgeon's individual surgically-induced astigmatism (SIA) was added to the recommended preoperative correction range to determine eligibility based on preoperative corneal cylinder. Once the SIA was estimated, this value stayed constant during the study for each investigator. | | |
PMA P910056/S027: FDA Summary of Safety and Effectiveness Data
Page 20
{21}
Table 11: MEAN PERCENT OF EYES WITH REDUCTION IN CYLINDER WITHIN 0.50 D AND 1.00 D OF INTENDED – FORM 4 (ITT POPULATION)
| | Control IOL (N=79) n (%) | Toric IOL 1.25 D (N=80) n (%) | Toric IOL 2.00 D (N=20) n (%) | Toric IOL 2.75 D (N=12) n (%) | All Toric IOL (N=112) n (%) |
| --- | --- | --- | --- | --- | --- |
| Total Non-Missing, n | 78 | 80 | 18 | 12 | 110 |
| Within 0.50 D of Intended, n(%) | 27 (34.6%) | 43 (53.8%) | 12 (66.7%) | 8 (66.7%) | 63 (57.3%) |
| Within 1.00 D of Intended, n (%) | 45 (57.7%) | 71 (88.8%) | 17 (94.4%) | 12 (100.0%) | 100 (90.9%) |
Table 12: MEAN TORIC IOL AXIAL STABILITY FROM FORM 3 TO FORM 4 (ITT POPULATION)
| Absolute rotation (degrees) | Toric IOL | | | All Toric IOL (N=112) |
| --- | --- | --- | --- | --- |
| | 1.25 D (N=80) | 2.00 D (N=20) | 2.75 D (N=12) | |
| Absolute rotation from Form 3 to Form 4 | | | | |
| n | 74 | 15 | 12 | 101 |
| Mean ± SD | 1.15 ± 1.08 | 0.92 ± 1.09 | 1.08 ± 0.73 | 1.11 ± 1.04 |
| ≤ 5 degrees rotation | 74 (100.0%) | 15 (100.0%) | 12 (100.0%) | 101 (100.0%) |
| 95% exact CI | 95.1% to 100.0% | 78.2% to 100.0% | 73.5% to 100.0% | 96.4% to 100.0% |
| Five multiple imputations of absolute rotation from Form 3 to Form 4 for missing data | | | | |
| ≤ 5 degrees rotation | | | | 560 (100.0%) |
| Abbreviations: CI = confidence interval; D = diopter; IOL = intraocular lens; Max = maximum; Min = minimum; SD = standard deviation. | | | | |
Another primary effectiveness endpoint was the mean (dioptric) cylinder reduction achieved at Form 4. Dioptric cylinder reduction (preoperative keratometric cylinder magnitude – postoperative manifest refraction cylinder magnitude referenced to the corneal plane) was determined at Form 4. At Form 4, mean cylinder reduction from preoperative keratometric cylinder measurements in the randomized ITT population was 0.479 ± 0.665 D among those subjects with control IOLs and 0.865 ± 0.487 D among those subjects with 1.25 D toric IOLs (Table 13), showing a statistically significant improvement favoring the 1.25 D toric IOLs (P < 0.001). Similar results were seen for the PP population. The dioptric cylinder reduction treatment effect (1.25 D IOL minus control IOL) in the ITT population was 0.39 D (95% CI: 0.228 to 0.545 D). Covariate
PMA P910056/S027: FDA Summary of Safety and Effectiveness Data
Page 21
{22}
analyses of dioptric cylinder reduction across age, gender, race, and site showed there were no statistically significant inconsistent results between the randomized treatment groups. Additional sensitivity analyses with best case and worst case imputation methods showed a significant benefit in the 1.25 D toric IOL arm (P < 0.001 for both imputation methods).
PMA P910056/S027: FDA Summary of Safety and Effectiveness Data
Page 22
{23}
Table 13: MEAN DIOPTRIC CYLINDER REDUCTION FROM PREOPERATIVE MEASUREMENTS, (ITT POPULATION)
| Cylinder reduction (D) | Control IOL (N=79) | Toric IOL | | | All Toric IOL (N=112) |
| --- | --- | --- | --- | --- | --- |
| | | 1.25 D (N=80) | 2.00 D (N=20) | 2.75 D (N=12) | |
| Form 2 | | | | | |
| n | 77 | 80 | 20 | 12 | 112 |
| Mean reduction ± SD | 0.640 ± 0.591 | 0.966 ± 0.466 | 1.486 ± 0.498 | 2.115 ± 0.328 | 1.182 ± 0.594 |
| Form 3 | | | | | |
| n | 79 | 79 | 18 | 12 | 109 |
| Mean reduction ± SD | 0.532 ± 0.627 | 0.864 ± 0.455 | 1.446 ± 0.519 | 1.926 ± 0.364 | 1.077 ± 0.584 |
| Form 4 | | | | | |
| n | 78 | 80 | 18 | 12 | 110 |
| Mean reduction ± SD | 0.479 ± 0.665 | 0.865 ± 0.487 | 1.413 ± 0.532 | 1.944 ± 0.327 | 1.072 ± 0.601 |
| Treatment effect at Form 4 | | 0.39 | | | |
| 95% CI of effect | | 0.228 to 0.545 | | | |
| Multiple imputation p-value^{a} | | < 0.001 | | | |
| Abbreviations: CI = confidence interval; D = diopter; IOL = intraocular lens; Max = maximum; Min = minimum; SD = standard deviation. ^{a}P-values and treatment effects are from a linear model Type II analysis, which include an effect for investigator and compare the control and 1.25 D toric IOLs at Form 4. Note: Dioptric change in cylinder = preoperative keratometric cylinder- postoperative manifest cylinder | | | | | |
At Form 4, no eyes in any treatment arm lost ≥ 10 letters from baseline. However, as required by the study statistical analysis plan, an analysis of eyes that lost ≥ 10 letters between an evaluation and a later postoperative evaluation (Form 2 through Form 4) identified 3 such eyes (2.7%; 3/112) in the overall population. One subject (1.2%; 1/112) in the control IOL group lost 14 letters from Form 3 to Form 4, and two subjects in the 1.25 D toric IOL group transiently lost 10 and 11 letters from Form 2 to Form 3, respectively. Preservation of best-corrected distance visual acuity showed 99.1% (111/112) of eyes in the ITT population implanted with a Toric IOL reported a VA of 20/40 or better at Form 4. Best-corrected distance visual acuity (BCDVA) results for the all Toric IOL treatment group are presented in Table 14 and Table 15. Uncorrected distance visual acuity (UCDVA) results for all Toric IOL treatment groups are presented in Table 16. The data provided in these tables are summaries for all subject eyes enrolled in the study, including 79 eyes with the control lens MX60 and 112 eyes with the test lens (MX60T).
PMA P910056/S027: FDA Summary of Safety and Effectiveness Data
Page 23
{24}
**Table 14: PRESERVATION OF BCDVA AT EACH EXAMINATION (ALL TORIC IOLS, ITT POPULATION)**
| | Preoperative N=112 | Form 2 N=112 | Form 3 N=112 | Form 4 N=112 |
| --- | --- | --- | --- | --- |
| 20/40 or Better, n(%) | 27 (24.1%) | 108 (96.4%) | 109 (100%) | 109 (99.1%) |
| Worse than 20/40, n(%) | 85 (75.9%) | 4 (3.6%) | 0 (0.0%) | 1 (0.9%) |
**Table 15: BCDVA WITHOUT GLARE AT FORM 4 (ITT POPULATION)**
| | Control IOL (N=79) | Toric IOL 1.25 D (N=80) | Toric IOL 2.00 D (N=20) | Toric IOL 2.75 D (N=12) | All Toric IOL (N=112) |
| --- | --- | --- | --- | --- | --- |
| **BCDVA (logMAR)** | | | | | |
| Total Non-Missing, (n) | 78 | 80 | 18 | 12 | 110 |
| Mean (±SD) | 0.01 (±0.09) | 0.00 (±0.09) | 0.05 (±0.10) | -0.01 (±0.09) | 0.01 (±0.09) |
| **BCDVA (Snellen)** | | | | | |
| 20/40 or Better, n (%) | 78 (100.0%) | 79 (98.8%) | 18 (100.0%) | 12 (100%) | 109 (99.1%) |
| Worse than 20/40, n (%) | 0 | 1 (1.3%) | 0 | 0 | 1 (0.9%) |
PMA P910056/S027: FDA Summary of Safety and Effectiveness Data
Page 24
{25}
Table 16: UCDVA AT FORM 4 (ITT POPULATION)
| | Control IOL (N=79) | Toric IOL 1.25 D (N=80) | Toric IOL 2.00 D (N=20) | Toric IOL 2.75 D (N=12) | All Toric IOL (N=112) |
| --- | --- | --- | --- | --- | --- |
| UCDA (logMAR) | | | | | |
| Total Non-Missing, n | 78 | 80 | 18 | 12 | 110 |
| Mean (±SD) | 0.19 (±0.16) | 0.11 (±0.14) | 0.12 (±0.11) | 0.13 (±0.18) | 0.11 (±0.14) |
| UCDVA (Snellen) | | | | | |
| 20/40 or Better, n (%) | 65 (83.3%) | 76 (95.0%) | 18 (100.0%) | 10 (83.3%) | 104 (94.5%) |
| Worse than 20/40, n (%) | 13 (16.7%) | 4 (5.0%) | 0 | 2 (16.7%) | 6 (5.5%) |
### 3. Subgroup Analyses
The following preoperative characteristics were evaluated for potential association with outcomes: The reduction in cylinder and UCDVA (in LogMAR units) following preoperative characteristics were evaluated for potential association with device effectiveness in subgroups prospectively defined across the range of preoperative corneal astigmatism values (Table 18) and (Table 19). Table 17 shows the dioptric change in cylinder (D) in the ITT population at Form 4 by intended preoperative IOL cylinder power for the randomized control and 1.25 D toric IOL treatment groups, where change in cylinder is preoperative keratometric cylinder minus postoperative manifest refraction cylinder.
PMA P910056/S027: FDA Summary of Safety and Effectiveness Data
Page 25
{26}
Table 17: MEAN REDUCTION IN CYLINDER AT FORM 4 STRATIFIED BY PREOPERATIVE KERATOMETRIC CYLINDER
| Preoperative Keratometric Cylinder (D) | Magnitude of Refractive Cylinder Reduction (D) | |
| --- | --- | --- |
| | Toric IOL 1.25 D (n) mean | Sphere IOL (n) mean |
| 1.00 to 1.25 | (1) 1.240 | (1) 0.550 |
| 1.26 to 1.50 | (35) 0.890 | (25) 0.290 |
| 1.51 to 1.75 | (33) 0.823 | (42) 0.519 |
| 1.76 to 2.00 | (11) 0.881 | (9) 0.708 |
| 2.01 to 2.25 | (0) --- | (1) 1.510 |
Table 18: MEAN LOGMAR UNCORRECTED DISTANCE VISUAL ACUTITY (UCDVA) AT FORM 4 STRATIFIED BY PREOPERATIVE KERATOMETRIC CYLINDER
| | logMAR UCDVA | |
| --- | --- | --- |
| Preoperative Keratometric Cylinder (D) | Toric IOL 1.25 D (n) mean | Sphere IOL (n) mean |
| 1.00 to 1.25 | (1) 0.020 | (1) 0.200 |
| 1.26 to 1.50 | (35) 0.125 | (25) 0.193 |
| 1.51 to 1.75 | (33) 0.103 | (42) 0.200 |
| 1.76 to 2.00 | (11) 0.096 | (9) 0.189 |
A related effectiveness endpoint was the percent reduction in absolute cylinder, defined as the absolute value of the difference between the postoperative magnitude of the subjective Manifest Refractive cylinder (converted to the corneal plane) and the preoperative magnitude of the keratometric cylinder, divided by the intended reduction in absolute cylinder, with the ratio expressed as a percentage. Table 19 summarizes the mean percent reduction in absolute cylinder at Form 4.
PMA P910056/S027: FDA Summary of Safety and Effectiveness Data
Page 26
{27}
**Table 19: MEAN PERCENT REDUCTION IN ABSOLUTE CYLINDER AT FORM 4**
| | Control IOL Mean ± SD (N=79) | Toric IOL 1.25 D Mean ± SD (N=80) | Toric IOL 2.00 D Mean ± SD (N=20) | Toric IOL 2.75 D Mean ± SD (N=12) | All Toric IOL Mean ± SD (N=112) |
| --- | --- | --- | --- | --- | --- |
| Number of Subjects at Form 4, n | 78 | 80 | 18 | 12 | 110 |
| Mean^{1} % Reduction in Absolute Cylinder (± SD) | 36.8% ±50.49% | 64.8% ±36.80% | 81.0% ±31.3% | 82.8% ±13.0% | 69.4% ±34.8% |
$^{1}$Mean of all subject (N) results at Form 4 - n (%)
#### 4. Pediatric Extrapolation
In this premarket application, existing clinical data was not leveraged to support approval of a pediatric patient population.
#### E. Financial Disclosure
The Financial Disclosure by Clinical Investigators regulation (21 CFR 54) requires applicants who submit a marketing application to include certain information concerning the compensation to, and financial interests and arrangement of, any clinical investigator conducting clinical studies covered by the regulation. The pivotal clinical study included 9 investigators. None of the clinical investigators had disclosable financial interests/arrangements as defined in sections 54.2(a), (b), (c), and (f). The information provided does not raise any questions about the reliability of the data.
### XI. PANEL MEETING RECOMMENDATION AND FDA'S POST-PANEL ACTION
In accordance with the provisions of section 515(c)(3) of the act as amended by the Safe Medical Devices Act of 1990, this PMA was not referred to the Ophthalmic Devices Panel, an FDA advisory committee, for review and recommendation because the information in the PMA substantially duplicates information previously reviewed by this panel.
PMA P910056/S027: FDA Summary of Safety and Effectiveness Data
Page 27
{28}
# XII. CONCLUSIONS DRAWN FROM PRECLINICAL AND CLINICAL STUDIES
# A. Effectiveness Conclusions
Significant reduction of astigmatism was achieved for all models of the enVista® Toric IOL. All of the primary effectiveness endpoints of this study were met, including all Toric IOL axial stability, 1.25 D toric IOL dioptric reduction in cylinder compared with control IOL (P< 0.001). The proportion of subjects in the ITT population at Form 4 with BCDVA of 20/40 or better was not statistically inferior (P>0.999) compared to the historical control rate (92.5%). In addition, the secondary effectiveness endpoint of UCDVA was met, also showing a significant advantage for the 1.25 D toric IOL arm compared to the control IOL arm (P < 0.001).
Forty-five percent (50/110) of toric IOL subjects had UCDVA of 20/20 or better, while only 16.7% (13/78) of the non-toric subjects achieved UCDVA of 20/20 or better at Form 4; this difference is highly statistically significant (P<0.01), and supports the expectation astigmatic subjects receiving an enVista® Toric IOL could rely less on glasses for distance vision.
# B. Safety Conclusions
The risks of the device are based on nonclinical laboratory and animal studies as well as data collected in a clinical study conducted to support PMA approval as described above. The safety of the enVista® One-Piece Hydrophobic Acrylic Toric IOL is highly favorable. All clinical safety endpoints of the clinical trial (G120193) are met. FDA historical controls provided in ISO 11979-7 SPE grids for both cumulative and persistent AEs are also not exceeded in the toric IOL arms of G120193. No secondary surgical interventions (SSIs) occurred during the trial for rotational instability or for any other condition.
Overall, no safety signals were associated with the toric IOLs during G120193. Neither the incidence nor the severity of AEs in the toric IOL arms was increased compared with the control IOL arm, and no SAEs were reported in the study eye. No AEs were related to either the device or the study protocol and no AEs resulted in study discontinuation. Mean IOPs in the toric IOL arms are similar to those in the control IOL arm across time points and no subject in a toric IOL arm had IOP ≥ 22 mmHg at Form 4. It is also true that no subject in the control IOL arm or the All Toric IOL arm have a BCDVA loss > 10 letters (2 lines) from baseline at Form 4. Slit lamp examinations, fundus examinations,
PMA P910056/S027: FDA Summary of Safety and Effectiveness Data
Page 28
{29}
and central posterior capsule assessments did not reveal any clinically meaningful increase in abnormal findings at Form 4.
### C. Benefit-Risk Determination
The probable benefits of the device are also based on data collected in a clinical study conducted to support PMA approval as described above. The clinical benefits of the device are based on data collected in G120193, which demonstrate that the enVista® Toric IOL is a safe and effective option for cataract subjects with astigmatism who, with the counsel of their surgeon, select the enVista® Toric IOL. This can produce excellent vision outcomes for subjects undergoing cataract surgery and may reduce the need for subjects to use glasses, contact lens, or other refractive correction for distance vision. These findings support a favorable risk-benefit ratio for visual correction of aphakia and astigmatism with the enVista® Toric IOL.
The probable risks of the device are also based on data collected in a clinical study conducted to support PMA approval as described above. The clinical benefits of the device are based on data collected in the G120193, which demonstrate that the enVista® Toric IOL is a safe and effective option for cataract patients with astigmatism who, with the counsel of their surgeon, select the enVista® Toric IOL.
1. Patient Perspectives
This submission did not include specific information on patient perspectives for this device.
In conclusion, given the available information above, the data support that for primary implantation in the capsular bag of the eye in adult subjects for visual correction of aphakia and corneal astigmatism following removal of a cataractous lens for improved uncorrected distance vision the probable benefits outweigh the probable risks.
### D. Overall Conclusions
The data in this application support the reasonable assurance of safety and effectiveness of this device when used in accordance with the indications for use.
PMA P910056/S027: FDA Summary of Safety and Effectiveness Data
Page 29
{30}
### XIII. CDRH DECISION
CDRH issued an approval order on June 8, 2018.
The applicant's manufacturing facilities have been inspected and found to be in compliance with the device Quality System (QS) regulation (21 CFR 820).
### XIV. APPROVAL SPECIFICATIONS
Directions for use: See device labeling.
Hazards to Health from Use of the Device: See Indications, Contraindications, Warnings, Precautions, and Adverse Events in the device labeling.
Post-approval Requirements and Restrictions: See approval order.
PMA P910056/S027: FDA Summary of Safety and Effectiveness Data
Page 30
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.