← Product Code [LOM](/productcode/LOM) · P230015

# Access HBsAg, Access HBsAg Confirmatory, Access HBsAg Calibrator, Access HBsAg QC (P230015)

_Beckman Coulter, Inc. · LOM · Oct 17, 2025 · Microbiology · APRL_

**Canonical URL:** https://fda.innolitics.com/device/P230015

## Device Facts

- **Applicant:** Beckman Coulter, Inc.
- **Product Code:** [LOM](/productcode/LOM.md)
- **Decision Date:** Oct 17, 2025
- **Decision:** APRL
- **Regulation:** 21 CFR 866.3172
- **Device Class:** Class 2
- **Review Panel:** Microbiology
- **Attributes:** Real-World Evidence, Pediatric

## Real-World Evidence

| Submission | Device | Sponsor | RWD Sources | RWE Use Summary | Key Tags |
| --- | --- | --- | --- | --- | --- |
| P230015 · Oct 17, 2025 | Access HBsAg, Access HBsAg Confirmatory, Access HBsAg Calibrator, Access HBsAg QC | Beckman Coulter, Inc. | Retrospective clinical specimens from patients with acute or chronic HBV infection | Retrospective clinical samples were used to supplement the prospective clinical study to evaluate the device's diagnostic performance (PPA) in known positive populations. | Retrospective study; Clinical performance; HBV infection |

### Clinical Evidence

| Study Design | Population | Comparator | Key Endpoints |
| --- | --- | --- | --- |
| PMA Clinical Study (Retrospective Cohort); Retrospective analysis of pre-characterized clinical specimens; Follow-up/Duration: Not applicable; Study Period: 2019-2025 | Adult subjects with acute or chronic HBV infection; Sample Size: 464; Number of Sites: 4 vendors | Comparator HBsAg assay | Positive Percent Agreement (PPA) |

## Indications for Use

Access HBsAg The Access HBsAg assay is a paramagnetic particle, chemiluminescent immunoassay for the in vitro qualitative detection of hepatitis B surface antigen (HBsAg) in human pediatric (7 through 21 years) and adult serum and serum separator tubes or plasma [lithium heparin, lithium heparin separator tubes, dipotassium (K₂) EDTA, tripotassium (K₃) EDTA, sodium citrate, acid citrate dextrose (ACD), and citrate phosphate dextrose (CDP)] using the DxI 9000 Access Immunoassay Analyzer. The Access HBsAg assay is used for the laboratory diagnosis of acute or chronic hepatitis B virus (HBV) infection in individuals with signs and symptoms of hepatitis or at risk for hepatitis B virus infection, when used in conjunction with other serological and clinical information. The assay may also be used to screen for HBV infection in pregnant women to identify neonates who are at risk of acquiring hepatitis B during the perinatal period. Access HBsAg Confirmatory The Access HBsAg Confirmatory assay is a paramagnetic particle, chemiluminescent immunoassay for the in vitro qualitative confirmation of the presence of hepatitis B surface antigen (HBsAg) in human pediatric (7 through 21 years) and adult serum and serum separator tubes or plasma [lithium heparin, lithium heparin separator tubes, dipotassium (K₂) EDTA, tripotassium (K₃) EDTA, sodium citrate, acid citrate dextrose (ACD) and citrate phosphate dextrose (CDP)] that have been found to be repeatedly reactive in the Access HBsAg assay using the Access DxI 9000 Immunoassay Analyzer. Access HBsAg Calibrator The Access HBsAg Calibrator is intended to calibrate the Access HBsAg and HBsAg Confirmatory assays for the in vitro qualitative detection and confirmation of hepatitis B surface antigen (HBsAg) in human serum and plasma using the DxI 9000 Access Immunoassay Analyzer. Access HBsAg QC The Access HBsAg QC is intended for monitoring system performance of the Access HBsAg and Access HBsAg Confirmatory assays. The Access HBsAg QC is for use on the DxI 9000 Access Immunoassay Analyzer.

## Device Story

Access HBsAg is a paramagnetic particle, chemiluminescent immunoassay performed on the DxI 9000 Access Immunoassay Analyzer. Input: human serum or plasma. Principle: one-step enzyme immunoassay where HBsAg binds to monoclonal antibody-coated paramagnetic particles and alkaline phosphatase-conjugated polyclonal antibodies. Unbound materials are washed away; chemiluminescent substrate is added; light output is measured by a luminometer. Confirmatory assay uses neutralization with HBsAg-specific antibodies to confirm reactive samples. Output: qualitative reactive/nonreactive result based on Signal/Cut-off (S/CO) ratio. Used in clinical laboratories by technicians. Results aid clinicians in diagnosing HBV infection, monitoring, and identifying pregnant women requiring neonatal prophylaxis. Benefits include accurate diagnosis, enabling timely antiviral treatment, and reducing disease transmission.

## Clinical Evidence

Clinical study evaluated 3,614 specimens (3,150 prospective, 464 retrospective) across three US sites. Prospective cohort included adults (at-risk/symptomatic), pregnant women, and pediatric subjects (7-21 years). Primary endpoints: PPA and NPA compared to reference HBsAg assays. Overall PPA was 99.5% (415/417) and NPA was 99.7% (3,189/3,197). Pregnant population NPA was 100.0% (494/494). Pediatric NPA was 100.0% (181/181). Confirmatory assay PPA on repeat reactive samples was 100.0% (415/415).

## Technological Characteristics

Paramagnetic particle, chemiluminescent immunoassay. Reagents include monoclonal (mouse) and polyclonal (caprine/equine) antibodies, alkaline phosphatase conjugate, and TRIS/MES buffers. Form factor: reagent packs for DxI 9000 analyzer. Connectivity: integrated with DxI 9000 system. Software: automated signal processing and S/CO calculation. Sterilization: not specified (reagents).

## Regulatory Identification

A qualitative hepatitis B virus (HBV) antigen assay is identified as an in vitro diagnostic device intended for prescription use for qualitative use with human serum, plasma, or other matrices that aids in the diagnosis of chronic or acute HBV infection. HBV surface antigen (HbsAg) is also used for screening of HBV infection in pregnant women to identify neonates who are at risk of acquiring hepatitis B during perinatal period. The assay is not intended for screening of blood, plasma, cells, or tissue donors.

## Special Controls

*Classification.* Class II (special controls). The special controls for this device are:(1) The labeling required under § 809.10(b) of this chapter must include:
(i) A prominent statement that the assay is not intended for the screening of blood, plasma, cells, or tissue donors.
(ii) A detailed explanation of the principles of operation and procedures for performing the assay.
(iii) A detailed explanation of the interpretation of results.
(iv) Limitations, which must be updated to reflect current clinical practice and disease presentation and management. The limitations must include statements that indicate:
(A) The specimen types for which the device has been cleared, and that use of this assay with specimen types other than those specifically cleared for this device may result in inaccurate assay results.
(B) When appropriate, performance characteristics of the assay have not been established in populations of immunocompromised or immunosuppressed patients or other populations where assay performance may be affected.
(C) Diagnosis of hepatitis B infection should not be established on the basis of a single assay result but should be determined by a licensed healthcare professional in conjunction with the clinical presentation, history, and other diagnostic procedures.
(D) Detection of HBV antigens indicates a current infection with hepatitis B virus but does not differentiate between acute or chronic infection. False reactive HbsAg result may occur for up to 2 weeks after vaccination with HbsAg containing vaccine.
(E) Current methods for the detection of hepatitis B antigens may not detect all potentially infected individuals. A non-reactive assay result does not exclude the possibility of exposure to or infection with hepatitis B virus. A non-reactive assay result in individuals with prior exposure to hepatitis B may be due to but not limited to antigen levels below the detection limit of this assay or lack of antigen reactivity to the antibodies in this assay. HBV mutants lacking the ability to produce antigens have been reported. These may occur as “escape” mutants in the presence of anti-HBV antibodies and such patients may be infectious.
(F) Results obtained with this assay may not be used interchangeably with results obtained with a different manufacturer's assay.
(2) Design verification and validation must include the following:
(i) A detailed device description, including all parts that make up the device, ancillary reagents required but not provided, an explanation of the device methodology, design of the capture antibody(ies), external controls, and computational path from collected raw data to reported result (
*e.g.,* how collected raw signals are converted into a reported signal and result), as applicable to the detection method and device design.(ii) For devices with assay calibrators, the design and composition of all primary, secondary, and subsequent quantitation standards used for calibration as well as their traceability to a standardized reference material that FDA has determined is appropriate (
*e.g.,* a recognized consensus standard). In addition, analytical testing must be performed following the release of a new lot of the standard material that was used for device clearance or approval, or when there is a transition to a new calibration standard.(iii) Documentation and characterization (
*e.g.,* supplier, determination of identity, purity, and stability) of all critical reagents (including description of the capture antibody(ies)), and protocols for maintaining product integrity throughout its labeled shelf life.(iv) Risk analysis and management strategies, such as Failure Modes Effects Analysis and/or Hazard Analysis and Critical Control Points summaries and their impact on assay performance.
(v) Final release criteria to be used for manufactured assay lots with appropriate evidence that lots released at the extremes of the specifications will meet the identified analytical and clinical performance characteristics as well as stability.
(vi) Stability studies for reagents must include documentation of an assessment of real-time stability for multiple reagent lots using the indicated specimen types and must use acceptance criteria that ensure that analytical and clinical performance characteristics are met when stability is assigned based on the extremes of the acceptance range.
(vii) All stability protocols, including acceptance criteria.
(viii) Final release assay results for each lot used in clinical studies.
(ix) Reproducibility study data that includes the testing of three independent production lots.
(x) Detailed documentation of analytical performance studies conducted, as appropriate to the technology, specimen types tested, and intended use of the device, including, the limit of blank (LoB), limit of detection (LoD), cutoff, precision (reproducibility) including lot-to-lot and/or instrument-to-instrument precision, interference, cross reactivity, carryover, hook effect, seroconversion panel testing, matrix equivalency, prominent mutants/variants detection (
*e.g.,* for HbsAg), specimen stability, reagent stability, and cross-genotype antigen detection sensitivity, when appropriate.(xi) Analytical sensitivity of the assay that is the same or better than that of other cleared or approved assays.
(xii) For devices with associated software or instrumentation, documentation must include a detailed description of device software, including software applications and hardware-based devices that incorporate software. The detailed description must include documentation of verification, validation, and hazard analysis and risk assessment activities, including an assessment of the impact of threats and vulnerabilities on device functionality and end users/patients as part of cybersecurity review.
(xiii) Detailed documentation and results from a clinical study. Performance must be analyzed relative to an FDA cleared or approved HBV antigen assay or a comparator that FDA has determined is appropriate. This study must be conducted using appropriate patient samples, with an appropriate number of HBV reactive and non-reactive samples in applicable risk and disease categories, and any applicable confirmatory testing. Additional relevant patient groups must be validated as appropriate. The samples must include prospective (sequential) samples for each identified specimen type and, as appropriate, additional characterized clinical samples. Samples must be sourced from geographically diverse areas. This study must be conducted in the appropriate settings by the intended users to demonstrate clinical performance.

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a PMA](https://innolitics.com/services/regulatory/), [a 510(k)](https://innolitics.com/services/510ks/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

SUMMARY OF SAFETY AND EFFECTIVENESS DATA (SSED)

I. GENERAL INFORMATION

Device Generic Name: Hepatitis B Surface Antigen (HBsAg)
Hepatitis B Surface Antigen Confirmatory

Device Trade Name: Access HBsAg
Access HBsAg Confirmatory
Access HBsAg Calibrator
Access HBsAg QC

Device ProCode: LOM

Applicant’s Name and Address: Beckman Coulter, Inc.
1000 Lake Hazeltine Drive
Chaska, MN 55318

Date(s) of Panel Recommendation: None

Premarket Approval Application (PMA) Number: P230015

Date of FDA Notice of Approval: October 17, 2025

II. INDICATIONS FOR USE

Access HBsAg
The Access HBsAg assay is a paramagnetic particle, chemiluminescent immunoassay for the in vitro qualitative detection of hepatitis B surface antigen (HBsAg) in human pediatric (7 through 21 years) and adult serum and serum separator tubes or plasma [lithium heparin, lithium heparin separator tubes, dipotassium (K₂) EDTA, tripotassium (K₃) EDTA, sodium citrate, acid citrate dextrose (ACD), and citrate phosphate dextrose (CDP)] using the DxI 9000 Access Immunoassay Analyzer.

The Access HBsAg assay is used for the laboratory diagnosis of acute or chronic hepatitis B virus (HBV) infection in individuals with signs and symptoms of hepatitis or at risk for hepatitis B virus infection, when used in conjunction with other serological and clinical information. The assay may also be used to screen for HBV infection in pregnant women to identify neonates who are at risk of acquiring hepatitis B during the perinatal period.

PMA P230015: FDA Summary of Safety and Effectiveness Data

{1}

The Access HBsAg assay is for use on the DxI 9000 Access Immunoassay Analyzer only.

This assay has not been approved for use in the screening of blood, plasma, or tissue donors or cadaveric specimens.

## Access HBsAg Confirmatory

The Access HBsAg Confirmatory assay is a paramagnetic particle, chemiluminescent immunoassay for the in vitro qualitative confirmation of the presence of hepatitis B surface antigen (HBsAg) in human pediatric (7 through 21 years) and adult serum and serum separator tubes or plasma [lithium heparin, lithium heparin separator tubes, dipotassium (K₂) EDTA, tripotassium (K₃) EDTA, sodium citrate, acid citrate dextrose (ACD) and citrate phosphate dextrose (CDP)] that have been found to be repeatedly reactive in the Access HBsAg assay using the Access DxI 9000 Immunoassay Analyzer.

The Access HBsAg Confirmatory assay is for use on the DxI 9000 Access Immunoassay System only.

## Access HBsAg Calibrator

The Access HBsAg Calibrator is intended to calibrate the Access HBsAg and HBsAg Confirmatory assays for the in vitro qualitative detection and confirmation of hepatitis B surface antigen (HBsAg) in human serum and plasma using the DxI 9000 Access Immunoassay Analyzer.

## Access HBsAg QC

The Access HBsAg QC is intended for monitoring system performance of the Access HBsAg and Access HBsAg Confirmatory assays. The Access HBsAg QC is for use on the DxI 9000 Access Immunoassay Analyzer.

## III. CONTRAINDICATIONS

There are no known contraindications.

## IV. WARNINGS AND PRECAUTIONS

The warnings and precautions can be found in the Access HBsAg, Access HBsAg Confirmatory, Access HBsAg Calibrator, and Access HBsAg QC labeling.

PMA P230015: FDA Summary of Safety and Effectiveness Data
Page 2

{2}

# V. DEVICE DESCRIPTION

# Kit Configurations and Components

The Access HBsAg assay (2 packs, 100 tests/pack) is for the qualitative determination of HBsAg, and is comprised of the following:

Table 1: Access HBsAg Reagent Pack Composition

|  Well | Contents | Ingredients  |
| --- | --- | --- |
|  R1a | 3.30 mL | Paramagnetic particles coated with streptavidin coupled to biotinylated monoclonal (mouse) HBsAg antibodies in TRIS buffer, surfactant, protein (bovine), < 0.1% sodium azide and 0.1% ProClin* 300.  |
|  R1b | 3.00 mL | MES buffer, protein (bovine), < 0.1% sodium azide, and 0.1% ProClin 300.  |
|  R1c | 3.20 mL | MES buffer, polyclonal (caprine) HBsAg antibody alkaline phosphatase conjugate, surfactant, protein (bovine, caprine and mouse), < 0.1% sodium azide and 0.1% ProClin 300.  |

*ProClin™ is a trademark of LANXESS Corp.

The Access HBsAg Confirmatory assay (2 packs, 50 determinations/pack) is for the confirmation of the presence of HBsAg, and is comprised of the following:

Table 2: Access HBsAg Confirmatory Reagent Pack Composition

|  Well | Contents | Ingredients  |
| --- | --- | --- |
|  R1a | 3.30 mL | Paramagnetic particles coated with streptavidin coupled to biotinylated monoclonal (mouse) HBsAg antibodies in TRIS buffer, surfactant, protein (bovine), < 0.1% sodium azide and 0.1% ProClin* 300.  |
|  R1b | 2.10 mL | MES buffer, protein (bovine), < 0.1% sodium azide, and 0.1% ProClin 300.  |
|  R1c | 3.20 mL | MES buffer, polyclonal (caprine) HBsAg antibody alkaline phosphatase conjugate, surfactant, protein (bovine, caprine and mouse), < 0.1% sodium azide and 0.1% ProClin 300.  |
|  R1d | 1.20 mL | MES buffer, polyclonal (equine) HBsAg antibody, protein (bovine), < 0.1% sodium azide and 0.1% ProClin 300.  |

*ProClin™ is a trademark of LANXESS Corp.

PMA P230015: FDA Summary of Safety and Effectiveness Data

{3}

The Access HBsAg Calibrator contains one level (positive for HBsAg) calibrator that is ready to use (C1): 2.0 mL/vial.

Table 3: Access HBsAg Calibrator Composition

|  Description | Contents | Ingredients  |
| --- | --- | --- |
|  C1 | 1 vial | TRIS buffer with HBs antigen (human), protein (bovine), < 0.1% sodium azide, and 0.5% ProClin* 300.  |
|  Calibration Card | 2 cards | Card 1 is for use with the Access HBsAg assay (REF C39422)
Card 2 is for use with the Access HBsAg Confirmatory assay (REF C39425)  |

*ProClin™ is a trademark of LANXESS Corp.

The Access HBsAg QC set contains two controls (negative and positive) that are ready to use (QC1-QC2): 4.0 mL/vial, 3 vials each level.

Table 4: Access HBsAg QC Composition

|  Description | Contents | Ingredients  |
| --- | --- | --- |
|  QC1 | 3 vials | Human serum negative for HBsAg, < 0.1% sodium azide, and 0.5% ProClin* 300.  |
|  QC2 | 3 vials | TRIS Buffer, human serum, HBs antigen, protein (bovine), < 0.1% sodium azide, and 0.5% ProClin 300.  |
|  QC Value Card | 2 cards | Card 1 is for use with the Access HBsAg assay (REF C39422)
Card 2 is for use with the Access HBsAg Confirmatory assay (REF C39425)  |

*ProClin™ is a trademark of LANXESS Corp.

In addition, the following components are required.

1. UniCel DxI Wash Buffer II is used to dilute patient samples for testing with the Access HBsAg Confirmatory assay.
2. The Access HBsAg and Access HBsAg Confirmatory assay protocol files (APF) must be installed on the DxI Access 9000 Immunoassay Analyzer prior to performing the assays.
3. The DxI Access 9000 Immunoassay Analyzer is designed to perform fully automated immunoassay tests based on the use of chemiluminescent detection technology, cleared under K221225.

PMA P230015: FDA Summary of Safety and Effectiveness Data
Page 4

{4}

Assay Principle and Format

Access HBsAg

The Access HBsAg assay is a one-step enzyme immunoassay. Paramagnetic particles coated with monoclonal HBsAg antibodies, alkaline phosphatase coupled to polyclonal HBsAg antibodies and sample are added to a reaction vessel. HBsAg present in the patient sample is bound by the anti-HBsAg coated on the paramagnetic particles and the anti-HBsAg-alkaline phosphatase conjugate. After incubation in a reaction vessel, materials bound to the solid phase are held in a magnetic field while unbound materials are washed away. A chemiluminescent substrate is added to the vessel and light generated by the reaction is measured with a luminometer. The light production is compared to the cut-off value defined during calibration on the instrument. Qualitative assessment is automatically determined from a stored calibration.

Access HBsAg Confirmatory

The Access HBsAg Confirmatory assay uses the principle of neutralization by an excess of HBsAg-specific antibodies (neutralization reagent) to confirm the presence of HBsAg in samples found repeatedly reactive in the Access HBsAg assay. The confirmatory assay reports percent neutralization (suppression) of a control assay with a neutralization reagent.

Paramagnetic particles coated with monoclonal HBsAg antibodies, alkaline phosphatase coupled to polyclonal HBsAg antibodies and sample are added to two separate reaction vessels, one containing neutralization reagent, and the other containing diluent (control). If HBsAg is present in the sample, the HBsAg specific antibodies of the neutralization reagent bind to the antigenic determinants which can no longer bind to the antibodies of the solid phase and conjugate. After incubation in a reaction vessel, materials bound to the solid phase are held in a magnetic field while unbound materials are washed away. A chemiluminescent substrate is added to each vessel and light generated by the reaction is measured with a luminometer. Both neutralization and diluent (control) tests are automatically run on the system. Percent neutralization is determined by comparison of the light generated with the neutralization reagent to that of the diluent (control).

Access HBsAg Calibrator

The Access HBsAg Calibrator is used to establish calibration (determine the cut-off value) for the Access HBsAg and Access HBsAg Confirmatory assays. By comparing the light intensity generated by a sample to the cut-off value, the presence or absence of hepatitis B virus surface antigen in the sample is determined.

PMA P230015: FDA Summary of Safety and Effectiveness Data
Page 5

{5}

Access HBsAg QC

Quality control (QC) materials simulate the characteristics of patient samples and are essential for monitoring the system performance of the Access HBsAg and Access HBsAg Confirmatory assays. In addition, they are an integral part of good laboratory practices. When performing assays with Access reagents for HBsAg, include quality control materials to validate the integrity of the assays. The assayed values should fall within the acceptable range if the test system is working properly.

Interpretation of Results

Access HBsAg

Test results are determined automatically by the system software. Results (Signal/Cut-off (S/CO)) are reported to be "reactive" or "nonreactive" as a function of their relationship with the "cut-off" (signal "greater than or equal to" or "less than" the cut-off value).

The high positive cut-off value (≥ 100.00 S/CO of HBsAg) for not performing repeat testing on Access HBsAg or confirmation testing on the Access HBsAg Confirmatory was determined using a population of reactive and nonreactive clinical samples. The cut-off was validated by comparing results from samples with initial S/CO values ≥ 100.00 with the final result using the Access HBsAg Confirmatory assay, as well as the final interpretation using commercially available reference HBsAg Qualitative and reference HBsAg Confirmatory assays. In this study, 399/423 (94.3%) samples initially reactive with Access HBsAg assay fell into the ≥ 100.00 S/CO range and 100.0% (399/399) were confirmed positive using the Access HBsAg Confirmatory assay. Positive percent agreement (PPA) of the Access HBsAg assay with the HBsAg reference assay final interpretation was also 100.0% with a lower bound 2-sided 95% score confidence interval of 99.0%.

Table 5: Access HBsAg Assay Results Interpretation

|  Initial Results | Result (S/CO) | Interpretation | Reporting Instructions | Retest Procedure  |
| --- | --- | --- | --- | --- |
|   |  < 1.00 | Nonreactive | Report result as nonreactive for HBsAg | No retest required  |
|   |  ≥ 1.00 to < 100.00 | Reactive | Report as initial reactive for HBsAg | Retest in duplicate  |
|   |  ≥ 100.00 | Reactive | Report as positive (reactive) for HBsAg | No retest or additional confirmatory testing required  |
|  Duplicate Retest Results | Both results < 1.00 | Nonreactive | Report result as nonreactive for HBsAg | No retest required  |
|   |  One or both results ≥ 1.00 | Reactive | Report result as repeat reactive for HBsAg | Confirm repeat reactive result in Access HBsAg Confirmatory assay (REF C39425). Samples that are confirmed using the neutralization reagent can be considered positive for the presence of HBsAg.  |

PMA P230015: FDA Summary of Safety and Effectiveness Data

{6}

# Access HBsAg Confirmatory

Access HBsAg Confirmatory test results are determined automatically by the system software. The presence of HBsAg in a reactive sample is confirmed if the calculations performed by the DxI 9000 Access Immunoassay Analyzer show a Signal/Cut-off (S/CO) ratio that is  $\geq 1.00$ , and a percent neutralization (%NT) that is  $\geq 50\%$ .

Samples with a S/CO ratio that is  $\geq 1.00$  and  $&lt; 50\%$  neutralization may be diluted up to 1:62,500, as indicated in the following table.

Table 6: Access HBsAg Confirmatory Assay Results Interpretation

|  Sample | Result (S/CO) | Percent Neutralization (%NT) | Reporting Instructions | Retest Procedure  |
| --- | --- | --- | --- | --- |
|  Neat sample | < 1.00 | Not applicable | Nonreactive for the presence of HBsAg | No retest required  |
|   |  ≥ 1.00 | ≥ 50% | Confirmed positive for the presence of HBsAg | No retest required  |
|   |  ≥ 1.00 | < 50% | Not Confirmed | Dilute sample 1:250 and run in Access HBsAg Confirmatory assay  |
|  1: 250 diluted sample | < 1.00 | Not applicable | Nonreactive for the presence of HBsAg | No retest required  |
|   |  ≥ 1.00 | ≥ 50% | Confirmed positive for the presence of HBsAg | No retest required  |
|   |  ≥ 1.00 | < 50% | Not Confirmed | Dilute sample 1:62,500 and run in Access HBsAg Confirmatory assay  |
|  1:62,500 diluted sample | < 1.00 | Not applicable | Nonreactive for the presence of HBsAg | No retest required  |
|   |  ≥ 1.00 | ≥ 50% | Confirmed positive for the presence of HBsAg | No retest required  |

# VI. ALTERNATIVE PRACTICES AND PROCEDURES

There are several other alternatives for the detection of hepatitis B virus (HBV) surface antigen (HBsAg). There are currently several FDA approved in vitro diagnostic tests commercially available for serological markers of hepatitis B virus (HBV) which, when used in conjunction with a patient's medical history, clinical examination, and other laboratory findings, may be used as an aid in the diagnosis of HBV infection in patients with symptoms of hepatitis or who may be at risk for HBV infection. The assay may be used to screen for hepatitis B infection in pregnant women to identify neonates at high risk of acquiring HBV during perinatal period. Each alternative has its own advantages and disadvantages. A patient should fully discuss these alternatives with his/her physician to select the method that best meets their expectations and lifestyle.

PMA P230015: FDA Summary of Safety and Effectiveness Data

{7}

# VII. MARKETING HISTORY

The Access HBsAg, Access HBsAg Confirmatory, Access HBsAg Calibrator, and Access HBsAg QC have not been marketed in the United States but is legally marketed in the European Union. The Access HBsAg, Access HBsAg Confirmatory, Access HBsAg Calibrator, and Access HBsAg QC have not been withdrawn from marketing for any reason related to its safety or effectiveness.

# VIII. POTENTIAL ADVERSE EFFECTS OF THE DEVICE ON HEALTH

Below is a list of the potential adverse effects (e.g., complications) associated with the use of the device. The risks associated with the device, when used as intended, are those related to the risk of false test results, failure to correctly interpret the test results and failure to correctly operate the instrument.

Risks of a false positive test include improper patient management, including further investigation of hepatitis B infection with other laboratory tests to determine if a patient is acutely or chronically infected. It is possible that a clinician would decide to treat hepatitis B infection with antiviral medications in a patient without hepatitis B infection. Antiviral medication has risks including toxicity and more rarely allergic reactions. Over time, viral resistance in patients who are co-infected but undiagnosed with other viruses using the same antiviral medication, such as HIV, can lead to viral resistance, however the likelihood of an undiagnosed co-infection in a patient tested for hepatitis B is exceedingly unlikely. These risks are likely mitigated by the fact that this test would then be part of a panel, and incongruous test results in a hepatitis panel would lead a clinician to retest the patient before starting treatment.

Risks of a false negative test include improper patient management, including missing the opportunity to treat chronic hepatitis B infection. A clinician may falsely believe that a patient is not acutely or chronically infected, but rather is currently susceptible or immune to the infection. False negative results may lead a clinician to vaccinate an infected patient. This risk is likely mitigated by the fact that this test is usually ordered as part of a panel of hepatitis B tests, and incongruous test results in a hepatitis panel would lead a clinician to retest the patient. A false negative result may alternatively result in a clinician missing the opportunity to further investigate and initiate treatment in a patient in whom treatment is otherwise be recommended, as HBsAg is often the first test sent as part of the evaluation of hepatitis B infection.

For the specific adverse events that occurred in the clinical study, please see Section X.D.1 below.

PMA P230015: FDA Summary of Safety and Effectiveness Data
Page 8

{8}

PMA P230015: FDA Summary of Safety and Effectiveness Data
Page 9

# IX. SUMMARY OF NON-CLINICAL STUDIES

## A. Laboratory Studies

1. Cut-Off Determination

The Access HBsAg and the Access HBsAg Confirmatory assays utilize a signal to cut-off ratio to define sample HBsAg status. The ratio of relative light units (RLUs) to an established cut-off determined during the assay calibration on the instrument indicates the presence or absence of HBsAg. The cut-offs of the Access HBsAg and the Access HBsAg Confirmatory assays were established by testing a total of 3,750 samples including characterized HBsAg positive samples, HBsAg negative samples, and samples. The evaluation was based on CLSI EP24-A2, Assessment of the Diagnostic Accuracy of Laboratory Tests Using Receiver Operating Characteristic Curves; Approved Guideline – 2nd Edition. Receiver Operating Characteristics (ROC) analyses were performed on the results of the specimens tested. The assays' cut-offs were evaluated with the observed results to demonstrate that their selection represent the best compromise between diagnostic specificity and diagnostic sensitivity. ROC curve analysis provided sensitivity (true positive proportion) and specificity (true negative proportion) values based on the manufacturer's working calibrator (WC-1) S/CO. The cut-off for the Access HBsAg assay is 1.0 S/CO. The cutoff for the Access HBsAg confirmatory assay is 1.0 S/CO with a percent neutralization ration (NR%) of ≥ 50%.

2. Analytical Sensitivity at the Cut-off (1.0 S/CO)

The ability of the Access HBsAg and Access HBsAg Confirmatory assays to detect low concentrations of the Third International Standard (IS) material for HBsAg, NIBSC code: 12/226 was evaluated.

The IS material was reconstituted according to manufacturer's instructions with 1 mL water to obtain a stock solution (47.30 IU/mL). A dilution series of six serum and six plasma levels were prepared with spiked amounts of the IS stock solution into each negative matrix to target HBsAg sample concentrations ranging from 0.01 – 0.1 IU/mL. Each dilution was tested in triplicate using three lots each of Access HBsAg assay, Access HBsAg Confirmatory assay, and Access HBsAg Calibrator. The analytical sensitivity for each sample type and reagent/calibrator lot combination was determined by a linear fit regression of sample S/CO values versus the IS stated concentrations (IU/mL). The point estimate of the IS concentration corresponding to 1.00 S/CO was used in the analytical sensitivity estimation. For each assay, the final analytical sensitivity is the maximum analytical sensitivity of all the reagent pack lot/calibration combinations tested.

{9}

The Access HBsAg and Access HBsAg Confirmatory assays were designed to have an analytical sensitivity of less than or equal to 0.03 IU/mL using the same IS material for HBsAg (HBV genotype B4, HBsAg subtypes ayw1/adw2) NIBSC code: 12/226. The overall results obtained are summarized below.

Access HBsAg assay: 0.025 IU/mL (95% CI: 0.024 – 0.026 IU/mL)
Access HBsAg Confirmatory assay: 0.019 IU/mL (95% CI: 0.018 – 0.020 IU/mL).

3. Limit of Detection

A limit of detection (LoD) study was conducted on the DxI 9000 Access Immunoassay Analyzer following methods described in CLSI guideline EP17-A2 using dilutions of the 3rd International standard material, NIBSC code: 12/226. Five (5) dilutions in serum and five (5) dilutions in plasma were prepared ranging from 0.020 IU/mL to 0.024 IU/mL, and each dilution level was tested across three reagent lots.

The LoD was determined for serum and plasma for each reagent lot by modeling the relationship between the percent of replicates detected (S/CO ≥ 1.00) and sample concentration using the Probit approach to calculate the concentration corresponding to 95% probability of detection. The maximum observed LoD value across reagent lots was 0.024 IU/mL for both plasma and serum sample types.

4. Genotype and Subtype Detection

Genotype detection was evaluated using two commercially available genotype panels (22 samples) and two patient samples containing genotypes A through H. A total of 24 panel members (A (4), B (3), C (4), D (4), E (2), F (3), G (1) and H (3)) were tested using the Access HBsAg and Access HBsAg Confirmatory assays. Fifteen of these samples were from the 1st International Reference Panel for Hepatitis B Virus (HBV) Genotypes for Hepatitis B Virus Surface Antigen (HBsAg) assays, PEI Code 6100/09.

A total of 9 commercially available subtypes, (adw2 (1), adw4 (1), adr (1), ayw1 (1), ayw2 (1), ayw3 (2), ayw4 (1), and ayr (1), were tested using the Access HBsAg and Access HBsAg Confirmatory assays.

Each sample was tested in a single replicate using one DxI 9000 Access Immunoassay Analyzer, one lot of Access HBsAg, one lot of Access HBsAg Confirmatory, and three lots of Access HBsAg Calibrator run in duplicate. All genotypes and subtypes were found to be reactive using the Access HBsAg assay, verified reactive using the reference HBsAg assay, and confirmed reactive for the presence of HBsAg using the Access HBsAg Confirmatory assay.

PMA P230015: FDA Summary of Safety and Effectiveness Data
Page 10

{10}

PMA P230015: FDA Summary of Safety and Effectiveness Data
Page 11

# 5. HBsAg Mutant Detection

A total of thirty HBsAg mutant samples (10 native and 20 recombinants internally created and verified by DNA sequencing), containing modifications between HBsAg amino acids 100 through 170 and representing mutants commonly reported in literature, were evaluated in singlicate with one lot each of the Access

All mutant HBsAg samples were recognized (100.0%) using the Access HBsAg assay, verified reactive with the reference assay, and confirmed reactive with HBsAg assay, Access HBsAg Confirmatory assay and HBsAg Calibrator on one DxI 9000 Immunoassay Analyzer. These same thirty samples were also tested with the reference assays (Qualitative and Confirmatory).

Access HBsAg Confirmatory and reference confirmatory assays. The list of all mutants tested and detected as reactive are listed in Table 7.

Table 7: HBsAg Mutations Evaluated (Native and Recombinant)

|  Mutant Type | Mutation  |
| --- | --- |
|  Native | Q129R+G130N, T126N, G130R+S132Y, P120S+F134I, T126A, Q129H+D144A, T118A (2), Y137F, M133L  |
|  Recombinant | P142S, S143L, G145R, C137W, P142S+G145R, 122NT, C124R, D144A, F134H, G130N, K122T, L109I, M133T, P127T, Y161F, Q129H, T123N, T126S, T131A, Y100C  |

# 6. Within Assay Sample Carryover (Intra-Assay)

The study was performed to evaluate the susceptibility of the Access HBsAg assay to within-assay sample carryover. High sample carryover testing was completed by running a sequence of alternating low (negative) and high HBsAg serum samples. Carryover testing was performed using one lot each of Access HBsAg and Access HBsAg Calibrator on one DxI 9000 Access Immunoassay Analyzer.

The acceptance criteria were met. Therefore, the Access HBsAg assay is not susceptible to within-assay sample carryover.

# 7. Interfering Substances (Endogenous/Exogenous)

The study was performed using a paired differences approach, as described in CLSI EP07, Interference Testing in Clinical Chemistry; Approved Guideline - 3rd Edition. Each substance was tested with matched samples (Control and Test for each interferent) run on the same day at three analyte levels (negative, low positive, and moderate positive) in seven replicates using one DxI 9000 Access Immunoassay Analyzer, one lot each of Access HBsAg, Access HBsAg Confirmatory, and Access HBsAg Calibrator. No significant interference was observed at the following concentrations. Refer to Table 8.

{11}

Table 8: Interfering Substances

|  Substance | Highest Concentration Added  |
| --- | --- |
|  Hemoglobin | 1,000 mg/dL  |
|  Total protein | 15 g/dL  |
|  Bilirubin Conjugated | 43 mg/dL  |
|  Bilirubin Unconjugated | 43 mg/dL  |
|  Intralipid | 38.54 mg/mL (37 mmol/L)  |
|  Biotin | 3,510 ng/mL  |
|  Aspirin (acetylsalicylic acid) | 167 μmol/L  |
|  Salicylic acid | 207 μmol/L  |
|  Acetaminophen (paracetamol) | 1,030 μmol/L  |
|  Ibuprofen | 1,060 μmol/L  |
|  Atorvastatin | 1.34 μmol/L  |
|  Lisinopril | 0.607 μmol/L  |
|  Levothyroxine | 0.552 μmol/L  |
|  Metformin | 92.9 μmol/L  |
|  Amlodipine | 0.183 μmol/L  |
|  Omeprazole | 862.26 μmol/L  |
|  Sertraline | 3.03 μmol/L  |

8. Cross reactivity (Analytical Specificity) and Microbial Interference
The study was conducted per CLSI EP07, Interference Testing in Clinical Chemistry-Approved Guideline, 3rd Edition to evaluate the Access HBsAg and Access HBsAg Confirmatory assays for potential cross-reactivity in specimens from individuals with various medical conditions. Confirmation of cross-reactivity status of serum or plasma samples was obtained from the certificate of analysis (CoA) of each stock sample.

A total of 401 HBsAg negative samples were tested using the Access HBsAg assay, and a total of 80 HBsAg positive samples were tested using the Access HBsAg Confirmatory assay. Sample status was verified using a reference HBsAg assay to confirm an initial negative result. For the Access HBsAg assay, viral antigen samples were diluted in normal human serum, and bacterial samples were prepared by spiking HBsAg negative samples with bacterial cultures before testing. For the Access HBsAg Confirmatory assay, viral antigen samples and bacterial samples were diluted in normal human serum spiked with HBsAg positive samples before testing.

All negative samples remained unimpacted in the presence of cross reactants. All HBsAg-spiked samples were found reactive and were confirmed with the Access HBsAg Confirmatory assay. Results are presented in Table 9.

PMA P230015: FDA Summary of Safety and Effectiveness Data

{12}

Table 9: Cross Reactivity (Analytical Specificity)

|  Category | Access HBsAg and Reference HBsAg Assay Final Interpretation |   |   | Access HBsAg Confirmatory Final Interpretation  |   |   |
| --- | --- | --- | --- | --- | --- | --- |
|   |  Samples tested | Number of Reactive samples | Number of Non-reactive samples | Number of spiked samples tested | Number of Reactive and Confirmed | Number of Non-reactive and Non-confirmed  |
|  Epstein-Barr Virus (EBV IgG) | 10 | 0 | 10 | 2 | 2 | 0  |
|  Cytomegalovirus (CMV IgG and IgM) | 10 | 0 | 10 | 2 | 2 | 0  |
|  Herpes Simplex Virus (HSV 1/2 IgG) | 10 | 0 | 10 | 2 | 2 | 0  |
|  Human Immunodeficiency Virus (anti-HIV-1/2) | 10 | 0 | 10 | 2 | 2 | 0  |
|  Hepatitis A Virus (HAV) | 10 | 0 | 10 | 2 | 2 | 0  |
|  Hepatitis C Virus (anti-HCV) | 10 | 0 | 10 | 2 | 2 | 0  |
|  Hepatitis E Virus (HEV) | 10 | 0 | 10 | 2 | 2 | 0  |
|  Alcoholic Liver Disease | 10 | 0 | 10 | 2 | 2 | 0  |
|  Primary Biliary Cirrhosis | 10 | 0 | 10 | 2 | 2 | 0  |
|  Flavivirus (Zika) | 11 | 0 | 11 | 2 | 2 | 0  |
|  Flavivirus (Dengue) | 8 | 0 | 8 | 2 | 2 | 0  |
|  Flavivirus (Zika + Dengue) | 2 | 0 | 2 | N/A | N/A | N/A  |
|  Flavivirus (West Nile) | 10 | 0 | 10 | 2 | 2 | 0  |
|  Influenza Post Vaccination | 10 | 0 | 10 | 2 | 2 | 0  |
|  HAMA | 10 | 0 | 10 | 2 | 2 | 0  |
|  Anti-Nuclear Antibody (ANA) | 10 | 0 | 10 | 2 | 2 | 0  |
|  Rheumatoid Factor | 10 | 0 | 10 | 2 | 2 | 0  |
|  Systemic Lupus Erythematosus (SLE) | 10 | 0 | 10 | 2 | 2 | 0  |
|  Pregnancy Multipara | 10 | 0 | 10 | 2 | 2 | 0  |
|  Pregnancy First Trimester | 10 | 0 | 10 | 2 | 2 | 0  |
|  Pregnancy Second Trimester | 10 | 0 | 10 | 2 | 2 | 0  |
|  Pregnancy Third Trimester | 10 | 0 | 10 | 2 | 2 | 0  |
|  Syphilis | 10 | 0 | 10 | 2 | 2 | 0  |
|  Toxoplasmosis (Anti-Toxo IgG) | 10 | 0 | 10 | 2 | 2 | 0  |
|  Transplant / Transplant Recipient | 10 | 0 | 10 | 2 | 2 | 0  |
|  Dialysis patients | 10 | 0 | 10 | 2 | 2 | 0  |
|  Hemophiliac / Clotting Factor | 10 | 0 | 10 | 2 | 2 | 0  |
|  S.aureus | 10 | 0 | 10 | 2 | 2 | 0  |
|  P.aeruginosa | 10 | 0 | 10 | 2 | 2 | 0  |
|  E.coli | 10 | 0 | 10 | 2 | 2 | 0  |

PMA P230015: FDA Summary of Safety and Effectiveness Data
Page 13

{13}

In addition, Access HBsAg and Access HBsAg Confirmatory assays were evaluated for 11 other disease categories by spiking 10 negative samples with antigen for each of the following categories: Cytomegalovirus (CMV), Epstein-Bar virus (EBV), Hepatitis A Virus (HAV), Hepatitis C Virus (HCV), Hepatitis E Virus (HEV), Human Immunodeficiency Virus (HIV), Herpes Simplex virus (HSV1/2), Rubella, Varicella Zoster Virus (VZV), Toxoplasmosis, and Syphilis. No cross reactivity was observed.

## 9. Sample Type Equivalency

The study was conducted using a protocol based on CLSI EP09c, Measurement Procedure Comparison and Bias Estimation Using Patient Samples Approved Guideline – 3rd Edition, to establish the equivalence of serum and plasma specimens. Matched sample sets from 80 individuals were used to create test samples for use in this study. A total of 40 of the matched sets of HBsAg negative samples were tested as true negative samples in the study, and the remaining 40 sets were used to prepare positive samples. Serum (without gel) served as the reference sample type. The samples from each sample set were tested in two replicates using one lot each of Access HBsAg, Access HBsAg Confirmatory, and Access HBsAg Calibrator on one DxI 9000 Access Immunoassay Analyzer. The results from reactive samples tested with the Access HBsAg assay were verified with the Access HBsAg Confirmatory assay to confirm the "reactive" status of the positive samples.

The results from the study demonstrate the equivalency between the reference sample type and the eight serum/plasma matrices evaluated. The results support the use of human serum (serum without gel and serum separator tubes) or plasma (lithium heparin, lithium heparin separator tube, dipotassium (K₂) EDTA, tripotassium (K₃) EDTA, sodium citrate, Acid Citrate Dextrose (ACD), and Citrate Phosphate Dextrose (CPD)) to detect hepatitis B virus surface antigen using Access HBsAg and Access HBsAg Confirmatory assays. Results are presented in Table 10.

PMA P230015: FDA Summary of Safety and Effectiveness Data
Page 14

{14}

Table 10: Sample Type Equivalence

|  Sample Type | Slope (95% Confidence Interval: Lower 95% CI – Upper 95% CI) | Correlation Coefficient (r)  |
| --- | --- | --- |
|  Serum | N/A | N/A  |
|  Serum separator tube | 1.02 (1.00 – 1.04) | 1.00  |
|  Plasma Lithium Heparin | 0.99 (0.97 – 1.01) | 1.00  |
|  Plasma Lithium Heparin separator tube | 0.99 (0.98 – 1.01) | 1.00  |
|  Plasma K2 EDTA | 0.99 (0.97 - 1.00) | 1.00  |
|  Plasma K3 EDTA | 0.99 (0.97 - 1.03) | 1.00  |
|  Plasma Sodium Citrate | 1.00 (0.98 - 1.01) | 1.00  |
|  Plasma ACD | 0.96 (0.95 – 0.98) | 1.00  |
|  Plasma CPD | 1.00 (0.98 - 1.02) | 1.00  |

# 10. Hook Effect

The purpose of this study was to determine whether high levels of HBsAg in patient specimens result in a hook effect impacting the results reported with the Access HBsAg and Access HBsAg Confirmatory assays on the DxI 9000 Immunoassay Analyzer. The highest available HBsAg positive human specimens at concentrations of  $3.01\mathrm{mg / mL}$  HBsAg Ad and  $2.52\mathrm{mg / mL}$  HBsAg Ay were used as positive stocks for this study. A series of ten-fold dilutions of the high positive samples were prepared using negative serum. The samples were tested in three replicates using one DxI 9000 Access Immunoassay Analyzer and one lot each of Access HBsAg, Access HBsAg Confirmatory, and Access HBsAg Calibrator. The Access HBsAg and Access HBsAg Confirmatory assays do not demonstrate any "hook" effect up to  $2.5\mathrm{mg / mL}$ .

# 11. Within Laboratory Precision (20 days)

A precision study was performed to evaluate the precision performance of the Access HBsAg assay based on guidance from the CLSI EP05-A3, Evaluation of Precision of Quantitative Measurement Procedures; Approved Guideline - 3rd Edition. The study design included two test runs per day over 20 test days. A ten-member panel of serum (S1-S5), plasma (P1-P5) patient samples and one lot of QC (QC1-QC2) were assayed in each run with patient samples tested in triplicate. Three lots of Access HBsAg reagents, two lots of calibrators and two lots of QC were tested on one DxI 9000 Access Immunoassay Analyzer for the study. The results presented in the following table are overall results.

The assay was designed to have within-laboratory imprecision of  $\leq 0.100$  S/CO SD at concentrations of  $&lt; 1.00$  S/CO and  $\leq 10.0\%$  CV at concentrations  $\geq 1.00$  S/CO.

PMA P230015: FDA Summary of Safety and Effectiveness Data

{15}

Table 11: Precision

|  Sample | Mean (S/CO) | N | Repeatability |   | Between-Run | Between-Day | Within Laboratory (Total)  |   |   |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|   |   |   |  (Within-Run)  |   |   |   |   |   |   |   |
|   |   |   |  SD | %CV | SD | %CV | SD | %CV | SD | %CV  |
|  QC1 | 0.12 | 1,440 | 0.00 | N/A | 0.00 | N/A | 0.00 | N/A | 0.01 | N/A  |
|  QC2 | 2.90 | 1,440 | 0.05 | 1.7 | 0.03 | 1.1 | 0.06 | 1.9 | 0.15 | 5.3  |
|  P1 | 0.12 | 1,440 | 0.00 | N/A | 0.00 | N/A | 0.00 | N/A | 0.01 | N/A  |
|  P2 | 1.28 | 1,440 | 0.02 | 1.5 | 0.01 | 1.1 | 0.02 | 1.7 | 0.07 | 5.3  |
|  P3 | 3.18 | 1,440 | 0.05 | 1.5 | 0.05 | 1.7 | 0.05 | 1.4 | 0.17 | 5.2  |
|  P4 | 6.23 | 1,440 | 0.10 | 1.7 | 0.08 | 1.4 | 0.11 | 1.8 | 0.34 | 5.4  |
|  P5 | 156.97 | 1,440 | 2.63 | 1.7 | 1.97 | 1.3 | 2.42 | 1.5 | 10.88 | 6.9  |
|  S1 | 0.11 | 1,440 | 0.00 | N/A | 0.00 | N/A | 0.00 | N/A | 0.01 | N/A  |
|  S2 | 1.23 | 1,440 | 0.02 | 1.5 | 0.01 | 1.1 | 0.02 | 1.7 | 0.06 | 5.1  |
|  S3 | 2.91 | 1,440 | 0.04 | 1.5 | 0.03 | 1.2 | 0.05 | 1.6 | 0.15 | 5.1  |
|  S4 | 4.51 | 1,440 | 0.07 | 1.6 | 0.05 | 1.1 | 0.08 | 1.7 | 0.23 | 5.1  |
|  S5 | 152.09 | 1,440 | 2.53 | 1.7 | 1.59 | 1.0 | 2.56 | 1.7 | 10.33 | 6.8  |

Note: %CV are not meaningful when S/CO approaches zero. Results are noted as N/A.

# 12. Analyte Detection in Serum (Seroconversion)

The study was conducted to determine the seroconversion sensitivity of the Access HBsAg and Access HBsAg Confirmatory assays. Thirty HBV seroconversion panels obtained from commercial vendors were tested on the DxI 9000 Access Immunoassay Analyzer using the Access HBsAg and Access HBsAg Confirmatory assays. The same panels were tested with the qualitative reference assay and samples found reactive were also tested using the confirmatory reference assay. The Access HBsAg assay shows detection of analyte equal to the reference assay for all panels except three. The Access HBsAg assay detected analyte and exhibited a reactive status one draw later than the reference assay in panels HBV6278 and PHM939, while it detected two draws later than the reference assay in panel HBV6290. All bleeds detected by Access HBsAg assay as reactive were confirmed reactive by the Access HBsAg Confirmatory assay. Results are summarized in the following table.

PMA P230015: FDA Summary of Safety and Effectiveness Data

{16}

Table 12: Seroconversion

|  Panel ID | HBsAg Positive Result from Initial Draw Date |   | Access HBsAg vs. Reference Assay  |
| --- | --- | --- | --- |
|   |  Access HBsAg (days) | Reference Assay (days) | Difference in Bleed Number*  |
|  SCP-HBV-002 | 16 | 16 | 0  |
|  SCP-HBV-004 | 18 | 18 | 0  |
|  SCP-HBV-006 | 13 | 13 | 0  |
|  HBV6273 | 14 | 14 | 0  |
|  HBV6275 | 7 | 7 | 0  |
|  HBV6277 | 33 | 33 | 0  |
|  HBV6278 | 12 | 8 | +1  |
|  HBV6279 | 26 | 26 | 0  |
|  HBV6280 | 13 | 13 | 0  |
|  HBV6282 | 19 | 19 | 0  |
|  HBV6283 | 26 | 26 | 0  |
|  HBV6284 | 50 | 50 | 0  |
|  HBV6285 | 40 | 40 | 0  |
|  HBV6286 | 33 | 33 | 0  |
|  HBV6290 | 21 | 14 | +2  |
|  HBV6291 | 27 | 27 | 0  |
|  HBV9099 | 16 | 16 | 0  |
|  HBV11000 | 21 | 21 | 0  |
|  HBV11001 | 44 | 44 | 0  |
|  HBV11003 | 142 | 142 | 0  |
|  HBV11007 | 34 | 34 | 0  |
|  HBV11011 | 103 | 103 | 0  |
|  HBV11012 | 18 | 18 | 0  |
|  HBV11017 | 40 | 40 | 0  |
|  HBV11029 | 35 | 35 | 0  |
|  HBV11052 | 37 | 37 | 0  |
|  HBV11056 | 33 | 33 | 0  |
|  PHM937 | 2 | 2 | 0  |
|  PHM939 | 11 | 3 | +1  |
|  PHM941 | 7 | 7 | 0  |

* The difference in bleed number is compared to the reference assay. For example, +1 indicates that the reference assay required 1 less bleed before reactivity was determined compared to the Access HBsAg assay.

PMA P230015: FDA Summary of Safety and Effectiveness Data

{17}

# 13. Reproducibility

A 5-day reproducibility study was performed based on CLSI EP05-A3, Evaluation of Precision of Quantitative Measurement Procedures; Approved Guideline - 3rd Edition. A ten-member panel of patient samples, including serum (S1-S5) and plasma (P1-P5) samples were assayed at three external clinical sites using one lot of Access HBsAg reagent kits, on three instruments. Each panel member was assayed in replicates of three with two runs per day. The results are summarized in Table 13.

Table 13: Combined Site Reproducibility

|  Sample | N | Mean (S/CO) | Between Site |   | Between Day |   | Between Run |   | Repeatability (Within Run) |   | Reproducibility  |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|   |   |   |  SD (S/CO) | %CV | SD (S/CO) | %CV | SD (S/CO) | %CV | SD (S/CO) | %CV | SD (S/CO) | %CV  |
|  S1 | 90 | 0.11 | 0.02 | N/A | 0.00 | N/A | 0.00 | N/A | 0.01 | N/A | 0.02 | N/A  |
|  S2 | 90 | 1.23 | 0.12 | 9.9 | 0.01 | 0.8 | 0.04 | 2.9 | 0.06 | 4.7 | 0.14 | 11.4  |
|  S3 | 90 | 2.94 | 0.32 | 10.8 | 0.05 | 1.8 | 0.07 | 2.2 | 0.04 | 1.5 | 0.33 | 11.2  |
|  S4 | 90 | 4.56 | 0.44 | 9.6 | 0.00 | 0.0 | 0.16 | 3.6 | 0.20 | 4.4 | 0.51 | 11.2  |
|  S5 | 90 | 155.93 | 19.45 | 12.5 | 5.48 | 3.5 | 4.29 | 2.8 | 2.29 | 1.5 | 20.79 | 13.3  |
|  P1 | 90 | 0.12 | 0.02 | N/A | 0.00 | N/A | 0.00 | N/A | 0.01 | N/A | 0.02 | N/A  |
|  P2 | 90 | 1.28 | 0.13 | 10.3 | 0.00 | 0.0 | 0.02 | 1.7 | 0.03 | 2.1 | 0.14 | 10.7  |
|  P3 | 90 | 3.20 | 0.34 | 10.5 | 0.04 | 1.1 | 0.03 | 0.9 | 0.06 | 1.7 | 0.34 | 10.7  |
|  P4 | 90 | 6.34 | 0.57 | 9.0 | 0.09 | 1.4 | 0.07 | 1.1 | 0.11 | 1.8 | 0.59 | 9.3  |
|  P5 | 90 | 159.12 | 17.86 | 11.2 | 5.79 | 3.6 | 3.46 | 2.2 | 2.86 | 1.8 | 19.3 | 12.1  |

Note: %CV are not meaningful when S/CO approaches zero. Results are noted as N/A.

# 14. Sample Stability

A study was conducted to evaluate the effect of sample handling and storage conditions at room temperature  $(20 - 25^{\circ}\mathrm{C})$ , refrigerated  $(2 - 8^{\circ}\mathrm{C})$ , and frozen  $(\leq -18^{\circ}\mathrm{C})$ . It also evaluated sample stability after multiple freeze/thaw cycles. The data demonstrated that samples may be used with the Access HBsAg and Access HBsAg Confirmatory assays when:

1. Stored at  $2 - 8^{\circ}\mathrm{C}$  for up to 6 days
2. Stored at  $20 - 25^{\circ}\mathrm{C}$  for up to 72 hours
3. Stored at  $-20^{\circ}\mathrm{C}$  or colder for up to 30 days. Do not thaw more than 5 times.

PMA P230015: FDA Summary of Safety and Effectiveness Data

{18}

15. Reagent Stability Studies

Real Time Stability: A stability study was performed to establish the shelf-life of Access HBsAg, Access HBsAg Confirmatory, Access HBsAg Calibrator, and Access HBsAg QC with three lots, each stored at the recommended storage temperature of 2-10°C throughout the study duration. The study demonstrated that reagents are stable when stored at 2-10°C as follows:

1. Access HBsAg reagent pack – 120 days
2. Access HBsAg Confirmatory reagent pack – 120 days
3. Access HBsAg Calibrator – 90 days
4. Access HBsAg QC – 318 days

In-Use Stability: A stability study was performed to verify the performance of Access HBsAg, Access HBsAg Confirmatory, Access HBsAg Calibrator, and Access HBsAg QC at the recommended open storage conditions (2-10°C). The results showed that Access HBsAg and Access HBsAg Confirmatory are stable for 56 days after initial use, Access HBsAg Calibrator is stable for 180 days after initial use, and Access HBsAg QC are stable for 90 days after initial use.

Stored Curve Stability: The stability study was performed to establish the stored curve stability of the Access HBsAg and Access HBsAg Confirmatory assays on the DxI 9000 Access Immunoassay Analyzer. The results showed that, for the Access HBsAg and the Access HBsAg Confirmatory assays, calibration is required every 56 days.

B. Animal Studies

Not Applicable

C. Additional Studies

Not Applicable

PMA P230015: FDA Summary of Safety and Effectiveness Data
Page 19

{19}

PMA P230015: FDA Summary of Safety and Effectiveness Data
Page 20

# X. SUMMARY OF PRIMARY CLINICAL STUDY

The applicant performed a clinical study to establish a reasonable assurance of safety and effectiveness of the Access HBsAg and Access HBsAg Confirmatory for the detection of hepatitis B surface antigen using samples that would routinely be tested for hepatitis in the US. Data from this clinical study were the basis for the PMA approval decision. A summary of the clinical study is presented below.

## A. Study Design

A multi-center study was conducted between 2019 and 2025 at three clinical testing sites in the United States. Specimens were collected from a total of forty-two (42) sites and 4 vendors. A total of 3,614 specimens were tested and analyzed with the Access HBsAg assay and a reference HBsAg assay from the following categories:

- 3,150 prospective samples:
- 2,474 adult samples (non-pregnant)
- 181 pediatric samples
- 495 pregnant women samples

- 464 retrospective specimens from known positive, acute or chronically ill adult subject specimens were tested and analyzed with the Access HBsAg assay and a Comparator HBsAg assay.

The prospective study population was as follows:

- The majority of patients were female (65% female and 35% male).
- 62.6% White, 26.7% Black or African American, 1.4% Asian, 1.4% American Indian or Alaska Native, 0.2% Native Hawaiian or other Pacific Islander, and 7.7% from unknown/other or unwilling to answer. 28.9% of the prospective study population was of Hispanic ethnicity.
- Patients in the prospective population were from the following states: Arizona (586, 18.6%), California (1, 0.0%), Connecticut (148, 4.7%), Florida (457, 14.5%), Georgia (118, 3.7%), Indiana (90, 2.9%), North Carolina (5, 0.2%), New Jersey (90, 2.9%), New York (518, 16.4%), Ohio (54, 1.7%), Pennsylvania (201, 6.4%), South Carolina (95, 3.0%), Tennessee, (225, 7.1%), Texas (392, 12.4%), and Virginia (170, 5.4%).

1. Clinical Inclusion and Exclusion Criteria

Enrollment in the Access HBsAg and Access HBsAg Confirmatory study was limited to patients who met the following inclusion:

- Subjects 2 years of age or older
- Subjects willing and able to donate required blood sample
- Subjects or legal guardian has signed the Informed Consent Form (a minor may need to sign an Assent Form if required by IRB)
- Subjects belong to one or more of the below categories:
- a. Person with an increased risk for hepatitis and/or with signs or symptoms of hepatitis

{20}

b. Pregnant person of any age for typical pre-natal HBV screening (low risk)

- Purchased HBV positive samples from acute and chronically ill subjects

Patients were not permitted to enroll in the Access HBsAg and Access HBsAg Confirmatory study if they met any of the following exclusion criteria:

- Subject who previously participated in the study.
- Subject who has received any experimental or investigational drugs or treatments pertaining to hepatitis B within 4 weeks of phlebotomy.

Subject samples that did not meet the required volume of sample were excluded from testing.

2. Follow-up Schedule

A follow-up schedule was not required for the study.

3. Clinical Endpoints

With regards to safety and effectiveness, the clinical endpoints are described in Section X.D.2.

B. Accountability of PMA Cohort

At the time of database lock, 3,593 subjects were prospectively enrolled and 472 retrospective samples were collected in the PMA study, 88% patients are available for analysis at the completion of the study.

HBV Classification

The HBV classification was determined for each specimen based on the reactivity patterns of the 4 HBV serological marker results (HBsAg, anti-HBc IgM, total anti-HBc, and anti-HBs), shown in Table 14 below.

Table 14: HBV Disease Staging by Serological Marker Results – 4 Markers

|  HBV Disease Staging | HBsAg^{a} | Anti-HBc IgM | Anti-HBc Total | Anti-HBs  |
| --- | --- | --- | --- | --- |
|  Early Acute | + | - | - | -  |
|  Acute | + | + | + | -  |
|   |  + | I | + | -  |
|  Recovering Acute/Undetectable | - | + | + | -  |
|  |   |   |   |   |
|  Recovering Acute | - | + | + | +  |
|  |   |   |   |   |
|  Chronic | + | - | + | +  |
|   |  + | - | + | -  |
|  |   |   |   |   |
|  Immune due to Natural Infection | - | - | + | +  |

PMA P230015: FDA Summary of Safety and Effectiveness Data

{21}

|  HBV Disease Staging | HBsAg^{a} | Anti-HBc IgM | Anti-HBc Total | Anti-HBs  |
| --- | --- | --- | --- | --- |
|  Immune due to HBV Vaccination | - | - | - | +  |
|  |   |   |   |   |
|  Susceptible | - | - | - | -  |
|  |   |   |   |   |
|  Uninterpretable | - | - | + | -  |
|   |  - | - | + | I  |
|   |  - | - | - | I  |
|   |  - | I | + | +  |
|   |  - | + | - | -  |

+ = Positive/Reactive; – = Negative; I = Indeterminate
a HBsAg results indicated as + were repeatedly reactive and confirmed by neutralization as required by the manufacturer’s Instructions for Use.

## C. Study Population Demographics and Baseline Parameters

The demographics of the study population are typical for an HBsAg detection study performed in the US. Demographics for the different study cohorts are shown in the following tables.

The following table summarizes the samples analyzed by gender and age for both the prospective and retrospective (purchased positives from acute or chronically ill patients) enrollment populations.

Table 15: Analyzed Samples by Gender and Age (Prospective and Purchased Positive Subjects)

|  Age Range (years) | Prospective |   | Retrospective |   | Total  |
| --- | --- | --- | --- | --- | --- |
|   |  Female | Male | Female | Male  |   |
|  7-12* | 14 | 19 | 0 | 0 | 33  |
|  13-18 | 43 | 24 | 0 | 0 | 67  |
|  19-21 | 109 | 32 | 0 | 0 | 141  |
|  22-29 | 520 | 126 | 40 | 103 | 789  |
|  30-39 | 480 | 167 | 44 | 106 | 797  |
|  40-49 | 260 | 174 | 27 | 64 | 525  |
|  50-59 | 340 | 270 | 18 | 28 | 656  |
|  60-69 | 207 | 203 | 9 | 16 | 435  |
|  70-79 | 62 | 64 | 1 | 4 | 131  |
|  80-89 | 13 | 17 | 1 | 0 | 31  |
|  90+ | 1 | 5 | 0 | 0 | 6  |
|  Unknown | 0 | 0 | 1 | 2 | 3  |
|  Total | 2,049 | 1,101 | 141 | 323 | 3,614  |

*Four (4) subjects outside of the pediatric intended use population of 7-21 are included in this analysis.

PMA P230015: FDA Summary of Safety and Effectiveness Data

{22}

Analyzed samples by race and enrollment cohort (prospective subjects only) are presented in the following table.

Table 16: Analyzed Samples by Race and Cohort (Prospective Subjects Only)

|  Race | Adults (Non-Pregnant) |   | Pediatrics (Non-Pregnant) |   | Pregnant |   | All Subjects  |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
|   |  N | % | N | % | N | % | N | %  |
|  American Indian or Alaska Native | 28 | 0.9% | 5 | 0.2% | 11 | 0.3% | 44 | 1.4%  |
|  Asian | 34 | 1.1% | 0 | 0.0% | 9 | 0.3% | 43 | 1.4%  |
|  Black or African American | 700 | 22.2% | 38 | 1.2% | 105 | 3.3% | 843 | 26.7%  |
|  Native Hawaiian or other Pacific Islander | 2 | 0.1% | 0 | 0.0% | 3 | 0.1% | 5 | 0.2%  |
|  Other | 158 | 5.0% | 7 | 0.2% | 27 | 0.9% | 192 | 6.1%  |
|  Unknown | 17 | 0.5% | 4 | 0.1% | 9 | 0.3% | 30 | 1.0%  |
|  Unwilling to specify | 17 | 0.5% | 2 | 0.1% | 0 | 0.0% | 19 | 0.6%  |
|  White | 1,518 | 48.2% | 125 | 4.0% | 331 | 10.5% | 1,974 | 62.6%  |
|  Total | 2,474 | 78.5% | 181 | 5.7% | 495 | 15.7% | 3,150 | 100.0%  |

Analyzed prospective samples for Hispanic and Non-Hispanic subjects by enrollment cohort are presented in the following table.

Table 17: Analyzed Hispanic and Non-Hispanic Subject Samples by Cohort (Prospective Subjects Only)

|  Ethnicity | Adults (Non-Pregnant) |   | Pediatrics (Non-Pregnant) |   | Pregnant |   | All Subjects  |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
|   |  N | % | N | % | N | % | N | %  |
|  Not Hispanic or Latino | 1,760 | 55.9% | 70 | 2.2% | 389 | 12.3% | 2,219 | 70.4%  |
|  Unknown or unwilling to specify | 13 | 0.4% | 0 | 0.0% | 7 | 0.2% | 20 | 0.6%  |
|  Hispanic or Latino | 701 | 22.3% | 111 | 3.5% | 99 | 3.1% | 911 | 28.9%  |
|  Total | 2,474 | 78.5% | 181 | 5.7% | 495 | 15.7% | 3,150 | 100.0%  |

PMA P230015: FDA Summary of Safety and Effectiveness Data
Page 23

{23}

The overall study population included 3,614 specimens, consisting of 3,150 serum samples that were prospectively collected, and an additional 464 pre-characterized retrospective samples (serum and plasma) collected from patients with acute or chronic HBV infection. Of the 3,150 specimens prospectively collected, 2,474 were from non-pregnant adults classified as increased risk for hepatitis due to lifestyle, behavior, occupation, or known exposure events, or individuals with signs and symptoms of hepatitis. 495 prospective specimens were from the pregnant screening population, of which 183 also were increased risk and/or showed signs and symptoms of hepatitis. Specimens from the pregnant population included 202 from the first trimester, 180 from the second trimester, and 113 from the third trimester. In addition, 181 prospective specimens were collected from the pediatric increased risk and/or signs &amp; symptoms population. The table below summarizes the number of specimens in each population.

Table 18: Summary of Specimens Used for the Access HBsAg Study

|   | Adult (non-pregnant)
(n = 2,938) |   | Pregnant
(n = 495) |   | Pediatric
(non-pregnant)  |
| --- | --- | --- | --- | --- | --- |
|   |  IR/S&S* | Pre-characterized acute/chronic HBV infection | IR/S&S | Low risk | IR/S&S  |
|  Prospective
(n=3,150) | 2,474 | N/A | 183 | 312 | 181  |
|  Retrospective
(n=464) | N/A | 464 | N/A | N/A | N/A  |
|  Total | 3,614  |   |   |   |   |

*IR/S&amp;S = Increased Risk and/or Signs &amp; Symptoms

The Access HBsAg results for the prospective population for all clinical trial sites combined by age group and gender are summarized in the table below. Samples were considered nonreactive if S/CO was &lt; 1.00 upon initial testing and reactive if S/CO was ≥ 1.00 during initial testing and repeat reactive if 2 of 3 replicates were ≥ 1.00 S/CO after duplicate testing.

PMA P230015: FDA Summary of Safety and Effectiveness Data
Page 24

{24}

Table 19: Distribution of Access HBsAg Assay Repeat Reactive and Nonreactive Results Among the Prospective Cohort by Age Group and Gender

|  Age Range (years) | Gender | Access HBsAg |   |   |   | Total  |
| --- | --- | --- | --- | --- | --- | --- |
|   |   |  Reactive |   | Nonreactive  |   |   |
|   |   |  N | % | N | %  |   |
|  7-12* | Female | 0 | 0.00 | 14 | 0.44 | 14  |
|   |  Male | 0 | 0.00 | 19 | 0.60 | 19  |
|  13-18 | Female | 0 | 0.00 | 43 | 1.37 | 43  |
|   |  Male | 0 | 0.00 | 24 | 0.76 | 24  |
|  19-21 | Female | 0 | 0.00 | 109 | 3.46 | 109  |
|   |  Male | 0 | 0.00 | 32 | 1.02 | 32  |
|  22-29 | Female | 1 | 0.03 | 519 | 16.48 | 520  |
|   |  Male | 0 | 0.00 | 126 | 4.00 | 126  |
|  30-39 | Female | 2 | 0.06 | 478 | 15.17 | 480  |
|   |  Male | 2 | 0.06 | 165 | 5.24 | 167  |
|  40-49 | Female | 3 | 0.10 | 257 | 8.16 | 260  |
|   |  Male | 3 | 0.10 | 171 | 5.43 | 174  |
|  50-59 | Female | 4 | 0.13 | 336 | 10.67 | 340  |
|   |  Male | 5 | 0.16 | 265 | 8.41 | 270  |
|  60-69 | Female | 0 | 0.00 | 207 | 6.57 | 207  |
|   |  Male | 5 | 0.16 | 198 | 6.29 | 203  |
|  70-79 | Female | 0 | 0.00 | 62 | 1.97 | 62  |
|   |  Male | 1 | 0.03 | 63 | 2.00 | 64  |
|  80-89 | Female | 0 | 0.00 | 13 | 0.41 | 13  |
|   |  Male | 0 | 0.00 | 17 | 0.54 | 17  |
|  90+ | Female | 0 | 0.00 | 1 | 0.03 | 1  |
|   |  Male | 0 | 0.00 | 5 | 0.16 | 5  |
|  Total |   | 26 | 0.83 | 3,124 | 99.17 | 3,150  |

*Four (4) subjects outside of the pediatric intended use population of 7-21 are included in this analysis.

PMA P230015: FDA Summary of Safety and Effectiveness Data

{25}

PMA P230015: FDA Summary of Safety and Effectiveness Data
Page 26

# Pregnant Females

495 serum samples were prospectively collected from a U.S. pregnant screening population, including 183 with at-risk factors for hepatitis and/or signs and symptoms of hepatitis. Samples were tested using the Access HBsAg assay and the reference HBsAg qualitative and reference HBsAg confirmatory assay to provide a final interpretation. Access HBsAg samples were considered nonreactive if S/CO was &lt; 1.00 upon initial testing and repeat reactive if S/CO was ≥ 1.00 during initial testing and if at least 2 of 3 replicates were ≥ 1.00 S/CO after duplicate testing. The results of HBsAg testing at all sites combined is presented in the following table.

Table 20: Pregnant Subjects Reference HBsAg Assay with Confirmation by Trimester

|  Pregnancy Cohort | Trimester | Reference HBsAg with Reference HBsAg Confirmatory assay result |   |   |   | Total (N)  |
| --- | --- | --- | --- | --- | --- | --- |
|   |   |  Reactive |   | Nonreactive  |   |   |
|   |   |  Access HBsAg  |   |   |   |   |
|   |   |  Repeat Reactive (N) | Nonreactive (N) | Repeat Reactive (N) | Nonreactive (N)  |   |
|  Pregnancy Screening (IR/S&S*) n = 183 | First | 1 | 0 | 0 | 74 | 75  |
|   |  Second | 0 | 0 | 0 | 71 | 71  |
|   |  Third | 0 | 0 | 0 | 37 | 37  |
|  Pregnancy Screening (Low Risk) n = 312 | First | 0 | 0 | 0 | 127 | 127  |
|   |  Second | 0 | 0 | 0 | 109 | 109  |
|   |  Third | 0 | 0 | 0 | 76 | 76  |
|  Total | Total | 1 | 0 | 0 | 494 | 495  |

*IR/S&amp;S = Increased Risk and/or Signs &amp; Symptoms

# Pediatrics

181 serum samples were prospectively collected from an at-risk and signs and symptoms U.S. pediatric non-pregnant population. Samples were tested using the Access HBsAg assay and the reference HBsAg qualitative and reference HBsAg confirmatory assay to provide a final interpretation. No reactive samples were found with Access HBsAg; therefore, no Access HBsAg Confirmatory testing was required. Negative percent agreement between the Access HBsAg assay and the reference HBsAg assay in pediatric samples was calculated. The results are presented in the following table.

Table 21: Pediatric Performance

|  PPA |   | NPA  |   |
| --- | --- | --- | --- |
|  % (n/N) | 95% CI | % (n/N) | 95% CI  |
|  N/A | N/A | 100.0 (181/181) | 97.9-100.0  |

{26}

Access HBsAg results for each classification were compared with the final interpretation from the same reference HBsAg assay used for classification in the table above. The results are described in the table below.

Table 22: Comparison of Access HBsAg assay versus HBsAg Reference Assay - All Populations by HBV Classification

|   | Reference HBsAg with Reference HBsAg Confirmatory Assay Result |   |   |   | Total  |   |   |
| --- | --- | --- | --- | --- | --- | --- | --- |
|   |  Reactive |   | Nonreactive  |   |   |   |   |
|   |  Access HBsAg  |   |   |   |   |   |   |
|  Cohort | Nonreactive (N) | Repeat Reactive (N) | Nonreactive (N) | Repeat Reactive (N) | Prospective (N) | Retrospective (N) | Total (N)  |
|  Early Acute | 0 | 3 | 0 | 0 | 3 | 0 | 3  |
|  Acute | 0 | 17 | 0 | 0 | 2 | 15 | 17  |
|  Recovering Acute | 0 | 0 | 4 | 0 | 4 | 0 | 4  |
|  Chronic | 1a | 393 | 0 | 0 | 18 | 376 | 394  |
|  Immune due to Natural Infection | 0 | 0 | 264 | 2 | 226 | 40 | 266  |
|  Immune due to HBV Vaccination | 0 | 0 | 1,164 | 1b | 1,163 | 2 | 1,165  |
|  Susceptible | 0 | 0 | 1,552 | 0 | 1,549 | 3 | 1,552  |
|  Uninterpretable | 0 | 0 | 203 | 5c | 183 | 25 | 208  |
|  Missing Classification | 1d | 2 | 2 | 0 | 2 | 3 | 5  |
|  Total | 2 | 415 | 3,189 | 8 | 3,150 | 464 | 3,614  |

a Sample tested PCR negative by certificate of analysis.
b Sample tested nonreactive using the Access HBsAg Confirmatory assay.
c Three of five samples were confirmed reactive using the Access HBsAg Confirmatory assay: available seroprofile information for these samples may indicate a low level of HBsAg that was not detected by the reference assay. The remaining two samples tested nonreactive with Access HBsAg Confirmatory.
d Sample tested PCR negative by supplementary testing and by certificate of analysis.

PMA P230015: FDA Summary of Safety and Effectiveness Data

{27}

D. Safety and Effectiveness Results

1) Safety Results

With regard to safety, as an *in vitro* diagnostic test, the Access HBsAg assay involves taking a sample of plasma or serum from a patient. The test therefore presents no more safety hazard to an individual being tested than other tests where blood samples were drawn.

There were no adverse events or device deficiencies reported during the conduct of the study.

2) Effectiveness Results

Effectiveness of the Access HBsAg and Access HBsAg Confirmatory assays were evaluated by comparing the final interpretation of Access HBsAg assay and Access HBsAg Confirmatory assay to the subject’s final HBsAg status as determined by the Comparator assay(s) on the same blood samples. Pre-specified endpoints were overall diagnostic effectiveness measured as positive percent agreement (PPA) and negative percent agreement (NPA) of the Access HBsAg assay compared to final HBsAg sample status as determined by the comparator assay(s).

Primary Endpoint

The primary endpoint evaluated for the Access HBsAg Confirmatory assay is as follows:

1) PPA for positive subjects from the overall population (HBsAg repeat reactive samples)

- The overall population includes all prospectively enrolled adult and pregnant subjects and purchased positive samples from acute and chronically infected adult patients.

The following pass criteria will be used for pass criteria for Primary Endpoints of the Access HBsAg assay:

1) PPA ≥ 98.0% on overall population with lower bound of a 95% 2-sided score confidence interval of PPA &gt; 95% for positive samples from HBV infected subjects
2) NPA ≥ 98.0% on overall population (excluding purchased positive samples from acute and chronically infected adult patients) with lower bound of a 95% 2-sided score confidence interval of NPA &gt; 95% for negative samples
3) Lower bound of a 95% 2-sided score confidence interval of NPA &gt; 98% in all pregnant subjects

The Access HBsAg Confirmatory assay will have the following pass criteria for primary endpoint:

- PPA ≥ 99.5% on overall population with lower bound of a 95% 2-sided score confidence interval of PPA &gt; 95%

PMA P230015: FDA Summary of Safety and Effectiveness Data
Page 28

{28}

PMA P230015: FDA Summary of Safety and Effectiveness Data
Page 29

# Primary Endpoint - PPA &amp; NPA in Overall Population

PPA (calculated on Access HBsAg assay only and Comparator HBsAg and confirmatory assays) and NPA (calculated on Access HBsAg assay only and Comparator HBsAg and confirmatory assays) for the overall population are shown in Table 23. The PPA overall population includes all subjects at-risk and/or with signs and symptoms, including the adult and pregnant population and purchased positive samples from acute and chronically ill adult patients that were HBsAg reactive by the Comparator during this study. No pediatrics are included as there were no HBsAg positive pediatric samples in this study. The NPA overall population includes All adult (at-risk and/or with signs and symptoms), pediatric (at-risk and/or with signs and symptoms) and pregnant (with or without documented risk factors and/or signs and symptoms) subjects and purchased HBV positive samples from acute and chronically infected adult patients that were HBsAg negative by the Comparator assay.

Table 23: Overall Population PPA &amp; NPA

|  Cohort | PPA |   | NPA  |   |
| --- | --- | --- | --- | --- |
|   |  % (n/N) | 95% CI | % (n/N) | 95% CI  |
|  Overall Population | 99.5% (415/417)a | 98.3-99.9% | 99.7% (3,189/3,197) | 99.5-99.9%  |

a The two samples that were found nonreactive using the Access HBsAg assay tested PCR negative by certificate of analysis or supplementary testing.

# Primary Endpoint - NPA in Pregnant Population

Pregnant subjects NPA (calculated with Access HBsAg assay only and compared the final HBsAg sample status) is shown in Table 24. Pregnant subject population includes all adult and pediatric Pregnant subjects with or without documented risk factors and/or signs and symptoms and vaccinated Pregnant subjects.

Table 24: Pregnant Subjects NPA

|  Cohort | NPA  |   |
| --- | --- | --- |
|   |  % (n/N) | 95% CI  |
|  Pregnant | 100.0% (494/494) | 99.2-100.0%  |

{29}

Primary Endpoint – HBsAg Repeat Reactive Samples with Confirmatory Test PPA

Primary endpoint PPA on 415 confirmed HBsAg positive samples from the overall population (including purchased positive samples from acute and chronically ill adult patients), with the Access HBsAg assay and the Access HBsAg Confirmatory assay compared to the Comparator HBsAg confirmatory assay, is shown in Table 25.

Table 25: Repeat Reactive HBsAg Samples Requiring Confirmatory Assay PPA
|  Cohort | PPA  |   |
| --- | --- | --- |
|   |  % (n/N) | 95% CI  |
|  Overall Population | 100.0% (415/415) | 99.1-100.0%  |

The following table provides a subanalysis of samples where the Access HBsAg initial results were reactive (1.00 ≤ S/CO &lt; 100.00) and required repeat testing and confirmation.

Table 26: Access HBsAg Initial Reactive Results (S/CO ≥ 1.00 and &lt; 100.00) and Required Repeat Testing
|  HBV Classification | Reference HBsAg with Reference HBsAg Confirmatory Assay Result |   |   |   | Total  |
| --- | --- | --- | --- | --- | --- |
|   |  Reactive |   | Nonreactive  |   |   |
|   |  Access HBsAg with Confirmatory  |   |   |   |   |
|   |  Initial Reactive (N) | Nonreactive (N) | Initial Reactive (N) | Nonreactive (N)  |   |
|  Early Acute | 1 | 0 | 0 | 0 | 1  |
|  Acute | 1 | 0 | 0 | 0 | 1  |
|  Chronic | 13 | 0 | 0 | 0 | 13  |
|  Immune due to Natural Infection | 0 | 0 | 2 | 0 | 2  |
|  Immune due to HBV Vaccination | 0 | 0 | 0 | 1 | 1  |
|  Uninterpretable | 0 | 0 | 3^{a} | 2 | 5  |
|  Missing Classification | 1 | 0 | 0 | 0 | 1  |
|  Total | 16 | 0 | 5 | 3 | 24  |

a Samples were confirmed reactive using Access HBsAg Confirmatory: available seroprofile information for these samples may indicate a low level of HBsAg that was not detected by the reference assay.

PMA P230015: FDA Summary of Safety and Effectiveness Data
Page 30

{30}

The following table provides a sub analysis of results from positive samples where no repeat or confirmatory testing was required using the high positive cut-off (S/CO ≥ 100.00).

Table 27: Access HBsAg Initial Reactive Results (S/CO ≥ 100.00) using the High Positive Cutoff

|  HBV Classification | Reference HBsAg with Reference HBsAg Confirmatory assay result |   |   |   | Total  |
| --- | --- | --- | --- | --- | --- |
|   |  Reactive |   | Nonreactive  |   |   |
|   |  Access HBsAg  |   |   |   |   |
|   |  Reactive (N) | Nonreactive (N) | Reactive (N) | Nonreactive (N)  |   |
|  Early Acute | 2 | 0 | 0 | 0 | 2  |
|  Acute | 16 | 0 | 0 | 0 | 16  |
|  Chronic | 380 | 0 | 0 | 0 | 380  |
|  Missing Classification | 1 | 0 | 0 | 0 | 1  |
|  Total | 399 | 0 | 0 | 0 | 399  |

Secondary Endpoint

Secondary endpoint NPA on both pediatric at-risk and/or signs and symptoms subjects (n=181) and pediatric pregnant subjects at-risk and/or with signs and symptoms (n=22) for a total of 203 subjects with Access HBsAg Assay only compared to the final HBsAg sample status, is shown in Table 28.

Table 28: Pediatric Subjects NPA

|  Cohort | NPA  |   |
| --- | --- | --- |
|   |  % (n/N) | 95% CI  |
|  Pediatric At-Risk or S&S | 100.0% (203/203) | 98.1-100.0%  |

3) Subgroup Analyses

The analysis of effectiveness was based on the 3,614 evaluable patients. Key effectiveness outcomes are presented in Section X.D.2 above. Subgroup analyses for each cohort discussed above are presented below.

The following tables compare the Access HBsAg assay results with the results obtained on an FDA-approved HBsAg reference assay by HBV disease classification for the different cohorts tested.

The distribution of Access HBsAg assay repeat reactive and non-reactive results by age and gender of the overall prospective population are presented below in Table 29.

PMA P230015: FDA Summary of Safety and Effectiveness Data

{31}

Table 29: Distribution of Access HBsAg Assay Repeat Reactive and Nonreactive Results Among the Prospective Cohort by Age Group and Gender

|  Age Range (years) | Gender | Access HBsAg Assay |   |   |   | Total  |
| --- | --- | --- | --- | --- | --- | --- |
|   |   |  Repeat Reactive |   | Nonreactive  |   |   |
|   |   |  N | % | N | %  |   |
|  7-12* | Female | 0 | 0.00% | 14 | 0.44% | 14  |
|   |  Male | 0 | 0.00% | 19 | 0.60% | 19  |
|  13-18 | Female | 0 | 0.00% | 43 | 1.37% | 43  |
|   |  Male | 0 | 0.00% | 24 | 0.76% | 24  |
|  19-21 | Female | 0 | 0.00% | 109 | 3.46% | 109  |
|   |  Male | 0 | 0.00% | 32 | 1.02% | 32  |
|  22-29 | Female | 1 | 0.03% | 519 | 16.48% | 520  |
|   |  Male | 0 | 0.00% | 126 | 4.00% | 126  |
|  30-39 | Female | 2 | 0.06% | 478 | 15.17% | 480  |
|   |  Male | 2 | 0.06% | 165 | 5.24% | 167  |
|  40-49 | Female | 3 | 0.10% | 257 | 8.16% | 260  |
|   |  Male | 3 | 0.10% | 171 | 5.43% | 174  |
|  50-59 | Female | 4 | 0.13% | 336 | 10.67% | 340  |
|   |  Male | 5 | 0.16% | 265 | 8.41% | 270  |
|  60-69 | Female | 0 | 0.00% | 207 | 6.57% | 207  |
|   |  Male | 5 | 0.16% | 198 | 6.29% | 203  |
|  70-79 | Female | 0 | 0.00% | 62 | 1.97% | 62  |
|   |  Male | 1 | 0.03% | 63 | 2.00% | 64  |
|  80-89 | Female | 0 | 0.00% | 13 | 0.41% | 13  |
|   |  Male | 0 | 0.00% | 17 | 0.54% | 17  |
|  90+ | Female | 0 | 0.00% | 1 | 0.03% | 1  |
|   |  Male | 0 | 0.00% | 5 | 0.16% | 5  |
|  Total |   | 26 | 0.83% | 3,124 | 99.17% | 3,150  |

*Four (4) subjects outside of the pediatric intended use population of 7-21 years are included in this analysis

The results are summarized in Table 30, categorized by HBV classification for the entire population.

PMA P230015: FDA Summary of Safety and Effectiveness Data

{32}

Table 30: PPA and NPA between the Access HBsAg Assay Results and the Reference HBsAg Assay

|  Cohort | PPA |   | NPA  |   |
| --- | --- | --- | --- | --- |
|   |  % (n/N) | 95% CI | % (n/N) | 95% CI  |
|  Early Acute | 100.0% (3/3) | 43.9-100.0% | N/A | N/A  |
|  Acute | 100.0% (17/17) | 81.6-100.0% | N/A | N/A  |
|  Recovering Acute | N/A | N/A | 100.0% (4/4) | 51.0-100.0%  |
|  Chronic | 99.7% (393/394)a | 98.6-100.0% | N/A | N/A  |
|  Immune due to Natural Infection | N/A | N/A | 99.2% (264/266) | 97.3-99.8%  |
|  Immune due to HBV Vaccination | N/A | N/A | 99.9% (1,164/1,165)b | 99.5-100.0%  |
|  Susceptible | N/A | N/A | 100.0% (1,552/1,552) | 99.8-100.0%  |
|  Uninterpretable | N/A | N/A | 97.6% (203/208)c | 94.5-99.0%  |
|  Missing Classification | 66.7% (2/3)d | 20.8-93.9% | 100.0% (2/2) | 34.2-100.0%  |
|  Total | 99.5% (415/417) | 98.3-99.9% | 99.7% (3,189/3,197) | 99.5-99.9%  |

a The one sample found nonreactive using Access HBsAg also tested PCR negative by certificate of analysis.
b The one sample found repeat reactive using Access HBsAg was found nonreactive using Access HBsAg Confirmatory.
c Of five samples found repeat reactive with Access HBsAg, three were confirmed reactive using Access HBsAg Confirmatory: available seroprofile information for these samples may indicate a low level of HBsAg that was not detected by the reference assay. The remaining two samples tested nonreactive with Access HBsAg Confirmatory.
d The one sample found nonreactive using Access HBsAg was found PCR negative by supplementary testing and by certificate of analysis.

The positive percent agreement between the Access HBsAg assay repeat reactive results and the reference assay final interpretation for the prospective, retrospective and combined populations is summarized in the following table.

Table 31: Comparison of Prospective and Retrospective Cohorts

|   | PPA |   | NPA  |   |
| --- | --- | --- | --- | --- |
|  Cohort | % (n/N) | 95% CI | % (n/N) | 95% CI  |
|  Prospective | 100.0 (23/23) | 85.7-100.0 | 99.7 (3,189/3,197) | 99.5-99.9  |
|  Retrospective | 99.5 (392/394)a | 98.2-99.9 | N/A | N/A  |
|  Combined | 99.5 (415/417) | 98.3-99.9 | N/A | N/A  |

a The two samples found nonreactive using Access HBsAg tested PCR negative by certificate of analysis or supplementary testing.

PMA P230015: FDA Summary of Safety and Effectiveness Data

{33}

PMA P230015: FDA Summary of Safety and Effectiveness Data
Page 34

# Pregnant Female Population

Four hundred and ninety-five (495) specimens were from pregnant women. Positive percent agreement (PPA) and negative percent agreement (NPA) between the Access HBsAg assay and the reference HBsAg assay in pregnant samples was calculated. The results are summarized in Table 32, categorized by trimester.

Table 32: Pregnant Population - PPA and NPA between the Access HBsAg Assay Results and the Reference HBsAg Assay

|  Pregnancy Cohort | PPA |   | NPA  |   |
| --- | --- | --- | --- | --- |
|  Trimester | % (n/N) | 95% CI | % (n/N) | 95% CI  |
|  First | 100.0% (1/1) | 20.7-100.0% | 100.0% (201/201) | 98.1-100.0%  |
|  Second | 0.0% (0/0) | N/A | 100.0% (180/180) | 97.9-100.0%  |
|  Third | 0.0% (0/0) | N/A | 100.0% (113/113) | 96.7-100.0%  |
|  Total | 100.0% (1/1) | 20.7-100.0% | 100.0% (494/494) | 99.2-100.0%  |

# Pediatric Population

One hundred and eighty-one (181) serum samples were prospectively collected from an at-risk and signs and symptoms U.S. pediatric non-pregnant population. Samples were tested using the Access HBsAg assay and the reference HBsAg qualitative and reference HBsAg confirmatory assay to provide a final interpretation. No positive samples were obtained and negative percent agreement between the Access HBsAg assay and the reference HBsAg assay in pediatric samples was calculated. The results are summarized in Table 33.

Table 33: Pediatric Population - PPA and NPA between the Access HBsAg Assay Results and the Reference HBsAg assay

|  PPA |   | NPA  |   |
| --- | --- | --- | --- |
|  % (n/N) | 95% CI | % (n/N) | 95% CI  |
|  N/A | N/A | 100.0% (181/181) | 97.9-100.0%  |

# 4) Pediatric Extrapolation

In this premarket application, existing clinical data was not leveraged to support approval of a pediatric patient population.

{34}

XI. Financial Disclosure

The Financial Disclosure by Clinical Investigators regulation (21 CFR 54) requires applicants who submit a marketing application to include certain information concerning the compensation to, and financial interests and arrangement of, any clinical investigator conducting clinical studies covered by the regulation. The pivotal clinical study included 10 investigators. None of the clinical investigators had disclosable financial interests/arrangements as defined in sections 54.2(a), (b), (c), and (f). The information provided does not raise any questions about the reliability of the data.

XII. SUMMARY OF SUPPLEMENTAL CLINICAL INFORMATION

Analytical comparison between pregnant and adult non-pregnant samples spiked with HBsAg

Thirty (30) samples from HBsAg negative pregnant patients were tested to determine if they provided equivalent results to HBsAg negative adult human serum pool samples (controls) when samples were spiked with the same individual HBsAg positive sample. These 30 samples included two specimens from women in the 1st trimester, 16 from the 2nd trimester, and 12 from the 3rd trimester of pregnancy.

Samples were spiked at target HBsAg values between 2.00 and 4.00 S/CO and tested with the Access HBsAg and Access HBsAg Confirmatory assays in replicates of five. The S/CO results for each pregnant sample were compared to the result obtained on each control sample and ranged from -6% to 13%. All samples were confirmed reactive with the Access HBsAg Confirmatory assay.

Analytical comparison between pediatrics and adult samples spiked with HBsAg

Thirty (30) samples from HBsAg negative pediatric patients aged 7 to 20 were tested to determine if they provided equivalent results to HBsAg negative adult human serum pool samples (controls) when samples were spiked with the same individual HBsAg positive sample.

Samples were spiked at a target HBsAg value between 2.00 and 4.00 S/CO and tested with the Access HBsAg and Access HBsAg Confirmatory assays in replicates of five. The S/CO results for each pediatric sample were compared to the result obtained on each control sample and ranged from -13% to 6%. All samples were confirmed reactive with Access HBsAg Confirmatory assay.

XIII. PANEL MEETING RECOMMENDATION AND FDA'S POST-PANEL ACTION

In accordance with the provisions of section 515(c)(3) of the act as amended by the Safe Medical Devices Act of 1990, this PMA was not referred to the Microbiology Panel, an FDA advisory committee, for review and recommendation because the information in the PMA substantially duplicates information previously reviewed by this panel.

PMA P230015: FDA Summary of Safety and Effectiveness Data
Page 35

{35}

XIV. CONCLUSIONS DRAWN FROM PRECLINICAL AND CLINICAL STUDIES

## A. Effectiveness Conclusions

The effectiveness of the Access HBsAg assay for the qualitative detection of hepatitis B surface antigen in human adult and pediatric (7 years to 21 years of age) serum and plasma is supported by the clinical study results. The results of this test may be used as an aid in the diagnosis of HBV infection in patients at risk or with signs and symptoms of hepatitis. The assay may also be used to screen for hepatitis B virus (HBV) infection in pregnant women to identify neonates who are at risk for acquiring hepatitis B during the perinatal period. See Section X.D.2 for Effectiveness Results.

## B. Safety Conclusions

The risks of the device are based on nonclinical laboratory studies as well as data collected in a clinical study conducted to support PMA approval as described above. Based on the results of these studies, the Access HBsAg assay, when used according to the manufacturer's instructions, can aid the physician in the diagnosis of HBV infection and screen for HBV infection in pregnant women to identify neonates who are at risk for acquiring hepatitis B during the perinatal period.

## C. Benefit-Risk Determination

The probable benefits of the device are also based on data collected in the clinical study conducted to support PMA approval as described above. The benefits of the assay are the determination, in conjunction with other serological and clinical information, the appropriate diagnosis and treatment of hepatitis B infection including initiation of appropriate monitoring, antiviral medications, and improved patient knowledge regarding the condition. Treatment for appropriate patients can mitigate the sequelae of hepatitis B infection and may result in reduced morbidity and mortality in these patients. Additionally, diagnosis and appropriate treatment can potentially decrease transmission and disease burden in the general population as well as in populations at high risk for hepatitis B infection. Accurate diagnosis of HBV infection also leads clinicians to evaluate and subsequently treat patients for human immunodeficiency virus (HIV) and hepatitis C virus (HCV) if indicated as these viruses share common risk factors and modes of transmission with HBV, and patients are often co-infected.

The risks associated with the device, when used as intended, are those related to the risk of false test results, failure to correctly interpret the test results and failure to correctly operate the instrument.

Risks of a false positive test include improper patient management, including further investigation of hepatitis B infection with other laboratory tests to determine if a patient is acutely or chronically infected. It is possible that a clinician would decide to treat hepatitis B infection with antiviral medications in a patient without hepatitis B infection. Antiviral medication has risks including toxicity and more rarely allergic

PMA P230015: FDA Summary of Safety and Effectiveness Data

{36}

reactions. Over time, viral resistance in patients who are co-infected but undiagnosed with other viruses using the same antiviral medication, such as HIV, can lead to viral resistance, however the likelihood of an undiagnosed co-infection in a patient tested for hepatitis B is exceedingly unlikely. These risks are likely mitigated by the fact that this test would then be part of a panel, and incongruous test results in a hepatitis panel would lead a clinician to retest the patient before starting treatment.

Risks of a false negative test include improper patient management, including missing the opportunity to treat chronic Hepatitis B infection. A clinician may falsely believe that a patient is not acutely or chronically infected, but rather is currently susceptible or immune to the infection. False negative results may lead a clinician to vaccinate an infected patient. This risk is likely mitigated by the fact that this test is usually ordered as part of a panel of hepatitis B tests, and incongruous test results in a hepatitis panel would lead a clinician to retest the patient. A false negative result may alternatively result in a clinician missing the opportunity to further investigate and initiate treatment in a patient in whom treatment is otherwise be recommended, as HBsAg is often the first test sent as part of the evaluation of hepatitis B infection.

## D. Patient Perspective

This submission either did not include specific information on patient perspectives or the information did not serve as part of the basis of the decision to approve or deny the PMA for this device.

In conclusion, given the available information above, the data support that for the claimed intended use the probable benefits outweigh the probable risks.

## E. Overall Conclusions

The data in this application support the reasonable assurance of safety and effectiveness of this device when used in accordance with the indications for use. The probable clinical benefits outweigh the potential risks for the proposed assay considering the performance of the device in the clinical study and the low risk and associated risk mitigations in clinical practice. The proposed assay labeling will facilitate accurate assay implementation and interpretation of results.…

---

**Source:** [https://fda.innolitics.com/device/P230015](https://fda.innolitics.com/device/P230015)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a PMA](https://innolitics.com/services/regulatory/), [a 510(k)](https://innolitics.com/services/510ks/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
