The DETOUR™ System is indicated for use for percutaneous revascularization in patients with symptomatic femoropopliteal lesions from 200mm to 460mm in length with chronic total occlusions (100mm to 425mm) or diffuse stenosis >70% who may be considered suboptimal candidates for surgical or alternative endovascular treatments. The DETOUR™ System, or any of its components, is not for use in the coronary and cerebral vasculature.
Device Story
The DETOUR System performs endovascular femoral-popliteal bypass using the femoral vein as a conduit. The system includes the TORUS Stent Graft (Nitinol wire frame with ePTFE/FEP film) and the ENDOCROSS Device (dual guidewire delivery tool with 0.025” Nitinol needle). Physicians use the ENDOCROSS Device to create an arterio-venous connection above the lesion and a veno-arterial connection below the lesion. TORUS Stent Grafts are then deployed sequentially from distal to proximal to create a bypass. The procedure is performed in a clinical setting (e.g., cath lab) by a physician. The device provides an alternative to surgical bypass for long-segment arterial disease, potentially improving blood flow and patient quality of life. The system is single-use and terminally sterilized.
Clinical Evidence
Evidence based on the prospective, multi-center, single-arm DETOUR2 clinical study (n=220). Primary safety endpoint: freedom from Major Adverse Events (MAE) at 30 days (93.0% success, 97.5% CI lower bound 88.5% vs 84% PG). Primary effectiveness endpoint: 12-month primary patency (68.1% success, 97.5% CI lower bound 60.9% vs 60.4% PG). Secondary endpoints included clinical success (RCC improvement), ABI, and 6MWT. Subgroup analysis indicated lower patency in females and US cohorts, potentially linked to anatomical factors. Supplemental data from the DETOUR1 trial (n=78) supported safety and effectiveness.
Technological Characteristics
TORUS Stent Graft: self-expanding Nitinol wire frame, ePTFE/FEP film. Delivery System: 8 Fr, 135 cm length, internal pulley mechanism, PtIr radiopaque markers. ENDOCROSS Device: 8 Fr, 133 cm length, 0.025” Nitinol needle, dual 0.014” guidewire ports. Sterilization: Ethylene Oxide (ISO 11135:2014). Biocompatibility: ISO 10993-1:2018. MRI: MR Conditional (ASTM 2503).
Indications for Use
Indicated for percutaneous revascularization in patients (age >18) with symptomatic femoropopliteal lesions (200-460mm length) including chronic total occlusions (100-425mm) or diffuse stenosis >70% who are suboptimal candidates for surgical/endovascular treatment. Contraindicated in patients with distal common femoral artery <7mm, high DVT risk, untreated flow-limiting aortoiliac disease, lack of patent tibial runoff, coagulopathy, or hypersensitivity to Nitinol, PTFE, or anticoagulants.
Regulatory Classification
Identification
For bypassing stenosis in the superficial femoral artery via the femoral vein
Related Devices
P220025 — LimFlow System · LimFlow, Inc. · Sep 11, 2023
P040037 — VIABAHN ENDOPROSTHESIS · W. L. Gore & Associates, Inc. · Jun 14, 2005
K260188 — LimFlow Vector · LimFlow, Inc. · Mar 18, 2026
K221902 — LimFlow Vector · LimFlow, Inc. · Dec 21, 2022
{0}
SUMMARY OF SAFETY AND EFFECTIVENESS DATA (SSED)
I. GENERAL INFORMATION
Device Generic Name: Stent Graft, Bypass, Superficial Femoral Artery
Device Trade Name: DETOUR™ System
Device Procode: QWM
Applicant’s Name and Address: Endologix LLC
2 Musick
Irvine, CA 92618
Date(s) of Panel Recommendation: None
Premarket Approval Application (PMA) Number: P220021
Date of FDA Notice of Approval: June 7, 2023
Breakthrough Device: Granted breakthrough device status on September 2, 2020 for percutaneous revascularization of symptomatic femoropopliteal lesions 200mm to 460mm with a chronic total occlusion 100mm to 425mm, and/or moderate-to-severe calcification, and/or in-stent-restenosis in patients with severe peripheral arterial disease.
II. INDICATIONS FOR USE
The DETOUR™ System is indicated for use for percutaneous revascularization in patients with symptomatic femoropopliteal lesions from 200mm to 460mm in length with chronic total occlusions (100mm to 425mm) or diffuse stenosis >70% who may be considered suboptimal candidates for surgical or alternative endovascular treatments. The DETOUR™ System, or any of its components, is not for use in the coronary and cerebral vasculature.
III. CONTRAINDICATIONS
The DETOUR™ System is contraindicated in patients with:
- A distal common femoral artery (CFA) <7 mm in diameter.
- Increased risk of deep vein thrombosis (DVT), such as patients with a recent history of DVT, thrombophilia, and disseminated malignancy.
- Untreated flow-limiting aortoiliac occlusive disease.
- Lack of patent single vessel tibial runoff to ankle.
- Known coagulopathy, bleeding diathesis, or thrombocytopenia that cannot be medically managed.
- Known hypersensitivities, allergies or contraindications to: Nitinol; PTFE; aspirin; heparin; antiplatelet; anticoagulant or thrombolytic therapy; or contrast media that cannot otherwise be medically managed.
PMA P220021: FDA Summary of Safety and Effectiveness Data
{1}
PMA P220021: FDA Summary of Safety and Effectiveness Data
Page 2 of 48
# IV. WARNINGS AND PRECAUTIONS
The warnings and precautions can be found in the DETOUR™ System labeling.
# V. DEVICE DESCRIPTION
The DETOUR™ System is comprised of two (2) main components:
- TORUS™ Stent Graft System; comprised of the:
- TORUS™ Stent Graft
- TORUS™ Stent Graft Delivery System
- ENDOCROSS™ Device
## TORUS Stent Graft System
The implantable TORUS Stent Graft (Figure 1) is a flexible, self-expanding composite structure made of a Nitinol (NiTi) wire frame encapsulated in an Expanded Polytetrafluoroethylene (ePTFE) film and Fluorinated Ethylene Propylene (FEP). The TORUS Stent Graft is pre-loaded onto the TORUS Stent Graft Delivery System.
Figure 1: TORUS Stent Graft
The TORUS Stent Graft is available in a variety of sizes outlined in Table 1. These can be deployed in a variety of configurations as outlined in Table 2.
Table 1: Vessel Diameter
| Labeled Device Diameter (mm) | Reference Vessel Diameter (mm) | Available Device Nominal Lengths (mm) | Recommended Balloon Diameter for Post-Dilation (mm) |
| --- | --- | --- | --- |
| 5.5 | 4.5-5.5 | 200 | 5.5 |
| 6.0 | 5.6-6.0 | 100, 150, 200 | 6.0 |
| 6.7 | 6.1-6.7 | 100, 150, 200 | 7.0 |
{2}
Table 2: Recommended Sizing Chart
| Lesion Length (mm) | Number of Stent Grafts* | Stent Graft 1 Nominal Length (mm) | Stent Graft 2 Nominal Length (mm) | Stent Graft 3 Nominal Length (mm) |
| --- | --- | --- | --- | --- |
| 200-220 | 2 | 150 | 150 | N/A |
| 230-270 | 2 | 200 | 150 | N/A |
| 280-320 | 2 | 200 | 200 | N/A |
| 330-410 | 3 | 200 | 200 | 150 |
| 420-460 | 3 | 200 | 200 | 200 |
*60mm of stent graft overlap is required
The TORUS Stent Graft Delivery System (Figure 2) is an 8 French (Fr) system. It is compatible with a 0.035” guidewire and has a 135 cm working length. The handle of the delivery system consists of an internal pulley mechanism activated through turning an external knob. The handle also features fluid flush ports for the inner lumen and guidewire lumen. The TORUS Stent Graft Delivery System uses an outer sheath to maintain the TORUS Stent Graft implant in a compressed state. Once at the target site, the user can slide the outer sheath proximally by turning the knob in the direction of the arrow to expose the self-expanding TORUS Stent Graft. The TORUS Stent Graft Delivery System has Platinum-Iridium (PtIr) radiopaque markers on both the proximal and distal ends of the TORUS Stent Graft landing zone (part of the inner shaft), and a marker band on the outer sheath to allow visualization of the position of the inner and outer sheath during deployment.

Figure 2: TORUS Stent Graft Delivery System
ENDOCROSS Device
The ENDOCROSS Device (Figure 3) is a dual guidewire delivery tool that uses a 0.025” Nitinol needle. The needle exits the delivery tool at an angle approximately 45° to the ENDOCROSS Device shaft. The ENDOCROSS Device is an 8 Fr compatible device with a
PMA P220021: FDA Summary of Safety and Effectiveness Data
{3}
133 cm working length, dual 0.014” guidewire ports, a rapid exchange (RX) guidewire port, and a needle guidewire port. The RX guidewire port is a back-loaded, rapid-exchange design used for initial device placement. The needle guidewire port is the central lumen, and the lumen exits through the needle and is used to deliver guidewire(s) to the desired location. The ENDOCROSS Device features are controlled using the outer handle and the button on the ENDOCROSS Device handle.
The outer handle controls spring loading, stabilizer deployment, and needle activation. The user moves the handle slider proximal to distal, then rotates the handle counterclockwise to load the spring, deploy the stabilizer, and activate the needle for deployment in a single motion. The ENDOCROSS Device shaft is keyed to ensure that the needle deploys in the same orientation as the marker band. Subsequent depression of the handle button deploys the needle in the direction indicated by the marker band.

Figure 3: ENDOCROSS Device
## Overview of the Procedure Using the DETOUR System
The procedure using the DETOUR System is an endovascular femoral-popliteal bypass procedure using the femoral vein as a conduit for the TORUS Stent Grafts which provides a bypass of the diseased arterial segment. Utilizing standard endovascular techniques, the ENDOCROSS Device is used to create an arterio-venous connection above the diseased arterial segment, and then a veno-arterial connection below the diseased arterial segment. TORUS Stent Grafts are then placed from distal to proximal sequentially to provide the bypass. Multiple TORUS Stent Grafts may be utilized during the procedure.
A graphical representation of the procedure using the DETOUR System is provided in (Figure 4).
PMA P220021: FDA Summary of Safety and Effectiveness Data
Page 4 of 48
{4}

Figure 4: Procedure Utilizing the DETOUR System
# VI. ALTERNATIVE PRACTICES AND PROCEDURES
There are several other alternatives for the correction of atherosclerotic disease of the superficial femoral and proximal popliteal arteries. Minimally-invasive approaches include endovascular intervention such as percutaneous transluminal angioplasty with a plain or drug-coated balloon, stent (e.g., bare metal, drug-eluting, or covered), and atherectomy. Surgical bypass procedures using autogenous or synthetic grafts are an option for long segment disease. Each alternative has its own advantages and disadvantages. A patient should fully discuss these alternatives with his/her physician to select the method that best meets expectations and lifestyle.
# VII. MARKETING HISTORY
The DETOUR System has been used in one (1) training case that was conducted outside of the clinical study. The training case was conducted in Germany while the device was CE Marked. The DETOUR System is no longer CE Marked and has not undergone full commercial launch in the United States or any foreign country.
# VIII. POTENTIAL ADVERSE EFFECTS OF THE DEVICE ON HEALTH
Below is a list of potential adverse effects (e.g., complications) associated with the use of the DETOUR System.
- Access vessel (arterial / venous) occlusion
- Amputation
- Aneurysm or pseudoaneurysm
PMA P220021: FDA Summary of Safety and Effectiveness Data
{5}
- Arteriovenous (AV) fistula
- Bleeding complications
- Death
- Device or deployment malfunction/failure
- Drug reactions to antiplatelet agents or contrast medium
- Edema
- Embolism (peripheral or pulmonary)
- Fever in absence of infection
- Hemorrhage or hematoma
- Hypotension / hypertension
- Infection (local or systemic including bacteremia or septicemia)
- Malposition
- Migration
- Myocardial infarction
- Pain (insertion site, leg and/or foot)
- Peripheral ischemia
- Renal insufficiency or failure secondary to contrast medium
- Shock
- Side branch vessel occlusion
- Stenosis or occlusion
- Stroke or transient ischemic attack
- Thrombosis
- Vessel wall trauma (dissection, perforation, or rupture)
- Vessel spasm
- Venous flow disruption (deep vein thrombosis, phlebitis, leg swelling and/or development of varicose veins)
- Worsening claudication
For the specific adverse events that occurred in the clinical study, please see Section D.1, Safety Results below.
## IX. SUMMARY OF NONCLINICAL STUDIES
### A. Laboratory Studies
The biocompatibility of the DETOUR System was evaluated per ISO 10993-1:2018 and FDA's Guidance for Industry and Food and Drug Administration Staff: Use of International Standard ISO 10993-1, "Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process" (September 2020).
Tests were conducted separately on products manufactured, packaged, and sterilized using materials and procedures intended for the marketed product for the TORUS Stent Graft, TORUS Stent Graft Delivery System, and ENDOCROSS Device.
PMA P220021: FDA Summary of Safety and Effectiveness Data
Page 6 of 48
{6}
The TORUS Stent Graft is categorized as a permanent implant device that comes into contact with circulating blood (>30 days). The TORUS Stent Graft Delivery System and the ENDOCROSS Device are classified as externally communicating devices that come into contact with circulating blood for a limited duration (<24 hours). A summary of the biocompatibility testing conducted can be found in Table 3 below.
Table 3: Summary of the DETOUR System Biocompatibility Testing
| Test Performed | Test Description | Stent Graft | Delivery System | Crossing Device | Results |
| --- | --- | --- | --- | --- | --- |
| Cytotoxicity | ISO MEM Elution Using L-929 Mouse Fibroblast Cells | X | X | X | Non-toxic |
| Sensitization | ISO Guinea Pig Maximization Sensitization Test | X | X | X | Non-sensitizing |
| Irritation | ISO Intracutaneous Study in Rabbits | X | X | X | Non-irritating |
| Pyrogenicity | ISO Materials Mediated Rabbit Pyrogen | X | X | X | Non-pyrogenic |
| Acute Systemic Toxicity | ISO Acute Systemic Toxicity Study in Mice | X^{a} | X | X | Non-toxic |
| Hemocompatibility | ASTM Hemolysis Assay (Direct and Indirect) | X | X | X | Non-hemolytic |
| | Complement Activation SC5b-9 Assay | X^{b} | X^{b} | X^{b} | Not a complement activator |
| | In Vivo Thrombogenicity | X^{b} | | | Non-thrombogenic |
| Genotoxicity | Chemical characterization testing and a toxicological risk assessment | X^{a} | | | Non-mutagenic, non-genotoxic |
| Subacute / Sub-chronic Toxicity | Chemical characterization testing and a toxicological risk assessment | X^{a} | | | Non-toxic |
| Chronic Toxicity | Chemical characterization testing and a | X^{a} | | | Non-toxic |
PMA P220021: FDA Summary of Safety and Effectiveness Data
Page 7 of 48
{7}
| Test Performed | Test Description | Stent Graft | Delivery System | Crossing Device | Results |
| --- | --- | --- | --- | --- | --- |
| | toxicological risk assessment | | | | |
| Carcinogenicity | Chemical characterization testing and a toxicological risk assessment | Xa | | | Non-carcinogenic |
${}^{a}$ Evaluated as part of the chemical characterization / toxicological risk assessment
${}^{b}$ Evaluated as part of the GLP chronic animal studies
The data demonstrated that the medical device materials and processing agents used in the manufacture of the DETOUR System have an acceptable biological safety profile.
# B. Animal Studies
Following completion of developmental animal testing, the DETOUR System was subjected to preclinical in-vivo evaluations compliant with the requirements of 21 CFR Part 58 – Good Laboratory Practices for Non-Clinical Laboratory Studies. These studies were conducted in an ovine animal model. Four chronic studies evaluating near final device design were conducted to support initiation of the DETOUR clinical trials. Two additional definitive ovine studies evaluating the final device were conducted to support approval of the PMA and the results support the safety and performance of the DETOUR System as summarized in Table 4.
Table 4: DETOUR System Animal Studies
| Study Name | Study Design | Findings |
| --- | --- | --- |
| GLP Study of the Performance, Chronic Safety, and Thrombogenicity of the PQ Bypass TORUS Stent Graft using the DETOUR Procedure in an Ovine Model (180 Days) | Ovine (n=8) at 191±1 days On day 90, two of the eight sheep were randomly chosen to be anesthetized and subjected to an interim angiogram procedure to evaluate the target arteries and distal vasculature. The In-Vivo Thrombogenicity biocompatibility endpoint was also evaluated. | All study objectives were met in their entirety with no complications at any phase. The test devices passed all evaluation determinants and acceptance criteria. There was no thrombus present on the ENDOCROSS Device or TORUS Stent Graft Delivery System during deployment, and all scores were recorded as “0”. |
| A GLP Study of the Performance, Efficacy, Chronic Safety and Thrombogenicity Evaluation of the PQ Bypass TORUS Stent Graft in the Peripheral Arteries of an Ovine Animal Model at 180 days with an Interim Angiographic | Ovine (n=6) at 180±7 days At the 90-day time point, two study sheep were arbitrarily chosen and subjected to a fluoroscopic evaluation to assess the patency, thrombus, flow or overt pathology within and around the test device. | All study objectives were met in their entirety with no problems encountered at any phase. The test device passed all evaluation determinants and acceptance criteria. None of the study sheep showed signs of thrombus, occlusion, stenosis, or other adverse signs. |
PMA P220021: FDA Summary of Safety and Effectiveness Data
{8}
| Study Name | Study Design | Findings |
| --- | --- | --- |
| Evaluation of a Sub-Cohort of Animals and Clinical Pathology at 90 days | The Implantation biocompatibility endpoint was also evaluated. | |
## C. Cadaver Study
A cadaver evaluation was conducted to confirm the usability of commercially available vascular snares with the DETOUR System during use in the procedure using the DETOUR System. The study considered the following design attributes to compare the usability and performance of snares:
- Venous Access Location
- Snare Outer Diameter
- Device Working Length
- Fluoroscopy Visibility
- Functional Guidewire Capture Area
- Difficulty of Retraction
The cadaver evaluation demonstrated that commercially available vascular snares can meet the usability and performance requirements necessary for the procedure using the DETOUR System.
## D. In Vitro Bench Testing
In vitro bench testing was performed to support the DETOUR System. Bench testing was performed per Endologix’s test protocols, which incorporated applicable international standards and FDA Guidance Non-Clinical Engineering Tests and Recommended Labeling for Intravascular Stents and Associated Delivery Systems – Guidance for Industry and FDA Staff (April 2010), as applicable.
Testing was conducted on all device configurations, a subset of device configurations, or worst-case for each test, as appropriate, to represent the entire DETOUR System ranges available. In vitro bench testing is summarized in Table 5, Table 6, and Table 7 according to testing of the TORUS Stent Graft, TORUS Stent Graft Delivery System, and ENDOCROSS Device.
Table 5: TORUS Stent Graft Testing
| Test | Test Purpose | Acceptance Criteria | Results |
| --- | --- | --- | --- |
| Dimensional verification | To verify dimensional specifications are met at t=0 and t=12 months accelerated aging. | Test samples must meet design specifications of the device. | PASS |
| Finite Element Analysis (FEA) | To evaluate the finite element analysis (FEA) results to the self-expanding Torus Stent Graft under simulated in vivo pulsatile | FEA is used to determine the worst-case stent graft sizing for durability fatigue testing. Therefore, there is no pre- | Worst case sizes for accelerated durability testing were identified |
PMA P220021: FDA Summary of Safety and Effectiveness Data
{9}
| Test | Test Purpose | Acceptance Criteria | Results |
| --- | --- | --- | --- |
| | and multi-mode physiological loading conditions. | defined acceptance criteria for FEA for bending fatigue. | |
| Accelerated Durability Testing | To evaluate the integrity of the TORUS Stent Graft in a silicone vessel model exposed to biomechanical loading conditions of pulsatile and axial / bending / torsional and crush fatigue for a minimum of 10 years post-implantation. | To stent must maintain structural integrity over a 10-year equivalent of pulsatile and multi-mode bending cycles in intravascular configurations. | PASS |
| Radial Resistive Force / Chronic Outward Force: TORUS Stent Graft COF/RRF | To assess chronic outward force / radial resistive force at time points t=0 and t=12 months, respectively. | Radial Resistive Force (RRF) ≥0.43 N/mm at 10% compression. Chronic Outward Force (COF) ≥0.14 N/mm and ≤0.60 N/mm at 10% compression. | PASS |
| TORUS Stent Graft Migration Resistance | To assess intravascular migration resistance at t=0 and t=12 months. | Migration resistance in intravascular placement >1.6N. | PASS |
| Simulated Use | To verify that the simulated use specifications are met at t=0 and t=12 months accelerated aging for the TORUS Stent Graft System. | The Stent Graft must successfully demonstrate the following functionality: • Shaft flushing • Guidewire compatibility • Catheter flexibility & trackability • Pushability: ≤20N • Outer shaft travel length: 100mm stent: ≥115mm; 200mm stent: ≥220mm • Deployment reliability and accuracy: tracks to location easily and deploys within ±3mm of target • Deployment force: ≤80N | PASS |
| Stent Integrity | Evaluated following Simulated Use testing. To verify the endoprosthesis has no clinically significant defects or flaws after deployment. | No failures visible under microscope for failures and stent graft portions conform to the mock artery wall. No anomalies (e.g., kinks, twists, component separation, nonuniform expansions, prosthesis damage). | PASS |
| Permeability: Porosity; Stent | To assess porosity of Torus Stent Grafts at t=0 and t=12 months; to evaluate the WEP of the Torus | Porosity: 5 microns ≤ internodal distance (IND) ≤ 30 microns. | Porosity testing inconclusive. As per ISO 25539- |
| | t=12 months, respectively. | t=12 months, respectively. | t=12 months, respectively. |
PMA P220021: FDA Summary of Safety and Effectiveness Data
{10}
PMA P220021: FDA Summary of Safety and Effectiveness Data
Page 11 of 48
| Test | Test Purpose | Acceptance Criteria | Results |
| --- | --- | --- | --- |
| Graft Water Entry Pressure (WEP) | Stent Graft at t=0 and t=12 months. | | 1:2017 guidance, refer to WEP. |
| | | Water Entry Pressure (WEP): 3 PSI, 155mmHg without gross continuous leaks, streams, or jets of fluid within the body of the stent graft. | PASS |
| Stent Graft De-Coupling | To evaluate stent graft decoupling force of two overlapped Torus Stent Grafts at time point t=0 and t=12 months. | Decoupling pull force ≥0.8N ≥6cm overlap). | PASS |
| Local Compression Testing | To evaluate local compression at t=0 and t=12 months. | ≥1.4N when compressed by 75% to 25% of nominal OD. | PASS |
| Flex/Kink Testing | To evaluate flex/kink resistance of the Torus Stent Graft at t=0 and t=12 months. | The stent must not kink at a 15mm bend radius without kinking (50% lumen loss). | PASS |
| Burst Testing | To evaluate burst testing of the Torus Stent Graft at t=0 and t=12 months. | Device can withstand ≥42kPa (6psi) of hydrostatic pressure delivered at a continuous rate of 10kPa/second. | PASS |
| Tensile Testing | To evaluate the longitudinal tensile strength of the Torus Stent Graft at t=0 and t=12 months. | Tensile strength must be ≥15N. | PASS |
| Corrosion Testing for Bare Stent Frame | To evaluate the crevice and pitting corrosion resistance of the bare Torus stent frame, i.e., with no ePTFE membrane. | The stent must have a minimum breakdown potential above 400 mV. | The endoprosthesis exhibits acceptable corrosion resistance based on testing per ASTM F2129-19a. |
| MRI Compatibility for Overlapping Torus Stent Grafts | To assess the safety and compatibility of the stent in the MRI environment. | The stent shall be MR conditional to 1.5 and 3 Tesla. | The stent does not pose additional risk to patients and may be labeled MR Conditional according to ASTM 2503. |
| Material Composition | Characterize the stent material composition to ensure it is acceptable for the intended use. | Chemical composition is within specification and complies with ASTM F2063-12. | The stent material conforms to implant material standards. |
| Mechanical Properties | Characterization testing performed to determine tensile and fatigue properties as inputs to support stress/strain and fatigue analysis. | | |
{11}
Table 6: TORUS Stent Graft Delivery System Testing
| Test | Test Purpose | Acceptance Criteria | Results |
| --- | --- | --- | --- |
| Dimensional Verification | To verify dimensional specifications are met at t=0 and t=12 months accelerated aging. | Test samples must meet design specifications of the device. | PASS |
| Simulated Use | See summary under TORUS Stent Graft Testing at t=0 and t=12 months accelerated aging. | See summary under TORUS Stent Graft Testing. | PASS |
| Bond/Joint Strength | To evaluate the strength of delivery system bonds at t=0 and t=12 months accelerated aging. | The delivery system bonds must maintain integrity above the specified load levels during stent deployment and delivery system retraction. | PASS |
| Kink Resistance | To verify that the Stent Graft Delivery System is resistant to kinking at t=0 and t=12 months accelerated aging. | Catheter will not kink when navigating a simulated anatomy with a radius of curvature 26.30mm. | PASS |
| Atraumatic Tip | To ensure that the force exerted by the Stent Graft Delivery System tip minimizes trauma to vessels during use at t=0 and t=12 months accelerated aging. | Force by the distal tip of the Delivery System is ≤20N. | PASS |
| Torque Strength | To assess the torque strength of the Stent Graft Delivery System at t=0 and t=12 months accelerated aging. | The stent graft delivery system must withstand handle rotations up to and including 360° clockwise and counterclockwise. | PASS |
Table 7: ENDOCROSS Device Testing
| Test | Test Purpose | Acceptance Criteria | Results |
| --- | --- | --- | --- |
| Dimensional Verification | To verify dimensional specifications are met at t=0 and t=12 months accelerated aging. | Test samples must meet design specifications of the device. | PASS |
| Simulated Use Testing | To verify that the ENDOCROSS Device simulated use specifications are met at t=0 and t=12 months accelerated aging. | The device must successfully demonstrate the following functionality: • Guidewire Rapid Exchange compatibility • Guidewire, needle compatibility | PASS |
PMA P220021: FDA Summary of Safety and Effectiveness Data
{12}
PMA P220021: FDA Summary of Safety and Effectiveness Data
Page 13 of 48
| Test | Test Purpose | Acceptance Criteria | Results |
| --- | --- | --- | --- |
| | | - Needle & stabilizer deployment control and multiple deployment and retractions
- Catheter flushing
- Catheter trackability
- Catheter pushability: ≤20N
- Catheter retraction force: ≤20N
- Needle guidewire retraction force: ≤3N | |
| Bond/Joint Strength | To verify that the ENDOCROSS Device bond strength specifications are met at t=0 and t=12 months accelerated aging. | The delivery system bonds must maintain integrity above the specified load levels during use. | PASS |
| Kink Resistance | To verify that the ENDOCROSS Device is resistant to kinking at t=0 and t=12 months accelerated aging. | Catheter will not kink when navigating simulated anatomy with a radius of curvature of 26.30mm (1.04 inches) at the inner surface. | PASS |
| Atraumatic Tip | To determine, at t=0 and t=12 months accelerated aging: (1) tensile force that will separate the distal tip on the ENDOCROSS Device, and (2) pressure exerted by the ENDOCROSS Device stabilizer to a simulated artery. | Atraumatic tip: ≤12N
Atraumatic stabilizer: ≤4psi in 4mm simulated artery | PASS |
| Needle Function | To verify, at t=0 and t=12 months accelerated aging: (1) distal end of the ENDOCROSS Device is non-coring, and (2) ability of the needle to puncture both arterial and venous walls in a single stroke during use. | Non-coring needle: No visible simulated tissue present after test.
Needle penetration: Needle penetrates through two 0.006” thick sheets of polyethylene with three needle firings. | PASS |
| Torque Response and Torque Strength | To assess the torque ability of the ENDOCROSS Device distal tip and the number of rotations until failure of the ENDOCROSS Device at t=0 and t=12 months accelerated aging. | Torque response: Amount of torque transmitted ≤300°
Torque strength: The catheter must withstand handle rotations up to and including 360° clockwise and counterclockwise. | PASS |
{13}
E. Sterilization
The DETOUR System (TORUS Stent Graft System and ENDOCROSS Device) is single-use and terminally sterilized using Ethylene Oxide (EO). The sterilization process is validated to demonstrate a sterility assurance level (SAP) of 10⁻⁶ in accordance with ISO 11135:2014: Sterilization of health care products – Ethylene oxide: Requirements for development, validation and routine control of a sterilization process for medical devices.
F. Packaging and Shelf Life
Packaging tests of the DETOUR System were conducted to verify that the integrity, as well as safety and performance, were not compromised due to the effects of storage and packaging stress. Baseline and aged packaging testing included a visual assessment, gross (bubble) leak testing, seal tensile strength, simulated shipping and distribution testing of the TORUS Stent Graft System and ENDOCROSS packaging. All tests passed requirements. A shelf life of one (1) year has been established for the DETOUR System based on product and package shelf-life testing.
X. SUMMARY OF PRIMARY CLINICAL STUDY
The applicant performed a clinical study (DETOUR2) to establish a reasonable assurance of safety and effectiveness of the procedure with the DETOUR System for percutaneous revascularization in patients with symptomatic femoropopliteal lesions from 200mm to 460mm in length in the US, Latvia, and Germany under IDE G170083. Data from this clinical study were the basis for the PMA approval decision. A summary of the clinical study is presented below.
A. Study Design
Patients were treated in the DETOUR2 study between December 13, 2017, and October 5, 2020. The database for the PMA reflected data collected through November 15, 2022, and included 220 patients. There were 36 investigational sites.
The study was a prospective, multi-center, single-arm, international, non-randomized premarket clinical study to evaluate the safety and effectiveness of the DETOUR System for the percutaneous treatment of patients with symptomatic femoropopliteal occlusive disease, as assessed in comparison to the prespecified safety and effectiveness performance goals (PGs).
The primary safety endpoint was freedom from major adverse events (MAEs) through 30 days post-procedure. Major Adverse Events include death, clinically driven target lesion revascularization (CD-TLR), amputation of the treated limb, occlusive-symptomatic deep vein thrombosis (DVT), pulmonary embolism (PE), or procedure-related bleeding requiring any transfusion of packed red blood cells or surgery. A performance goal for freedom from 30-day MAE of 84.0% was established for this endpoint. The safety performance goal was based on an aggregate of published trial data as described by VIVA Physicians Inc. (VPI) and adjusted to reflect the greater risk associated with the DETOUR2 study population. The study
PMA P220021: FDA Summary of Safety and Effectiveness Data
Page 14 of 48
{14}
device was considered to have achieved the safety objective if the lower limit of the one-sided lower 97.5% confidence limit based on the exact method is greater than 84%.
The primary effectiveness endpoint was patency at 12 months post procedure. Patency is defined as the absence of CD-TLR and absence of recurrent target lesion diameter stenosis >50% by imaging (e.g., duplex ultrasound peak systolic velocity ratio peak systolic velocity ratio [PSVR] of >2.5 within the stent or immediately 1cm above or below the treated segment). If both duplex ultrasound and angiography were available, angiography took precedence. The primary effectiveness endpoint of 12-month patency was evaluated by comparing the proportion of successful subjects to a literature-derived performance goal of 60.4%.
The final study sample size was based on the number required to test the primary effectiveness endpoint. To account for a 10% attrition rate, it was calculated that 202 subjects should be enrolled such that 181 subjects were expected to provide evaluable data for the 12-month patency endpoint.
An independent Clinical Events Committee (CEC) was used to review and adjudicate primary and secondary safety endpoints (defined in the study protocol), including major adverse events. A Data Safety Monitoring Board (DSMB) consisting of non-Investigator experts, reviewed safety data and established stopping rules for early termination of the trial. All angiograms, duplex ultrasound studies and x-rays obtained during this study per study requirements were submitted to the central Imaging Core Lab for analysis. The Core Lab assessed x-ray imaging for stent fracture at the applicable study time points. An independent Medical Monitor (MM) reviewed each adverse event to determine endpoints in accordance with guidelines provided by the CEC Charter.
1. Clinical Inclusion and Exclusion Criteria
Enrollment in the DETOUR2 Study was limited to patients who met the following inclusion criteria:
**General Inclusion Criteria**
- Age > 18 and ≤ 90 years.
- Willing and above to provide informed consent.
- Subject is willing to undergo all follow-up assessments according to the specified schedule over 36 months.
**Clinical Inclusion Criteria**
- Chronic, symptomatic lower limb ischemia defined as Rutherford Clinical Categories 3, 4, or 5.
- Venous Clinical Severity Score < 3.
- Subject is a suitable candidate for angiography and endovascular intervention and, if required, is eligible for standard surgical repair.
PMA P220021: FDA Summary of Safety and Effectiveness Data
Page 15 of 48
{15}
PMA P220021: FDA Summary of Safety and Effectiveness Data
Page 16 of 48
# Angiographic Inclusion Criteria
- Symptomatic femoropopliteal chronic total occlusions ≥20 cm (TASC D) that can include de novo, restenotic, or in-stent restenotic lesions; or
- Symptomatic femoropopliteal lesions ≥24 cm (total lesion length) that can include a chronic total occlusion or a ≥70% lesion that includes de novo, restenotic or in-stent restenosis (complex TASC C), by investigator visual assessment.
- Reference vessel diameter ≥4.5 and ≤6.7 mm, by investigator visual assessment
- Subject has a patent popliteal artery (<50% stenosis) distal to the landing zone.
- Able to successfully access the SFA origin for entry of the crossing device.
- At least one patent infrapopliteal vessel (<50% stenosis) with run-off to the ankle or foot.
- A significant stenosis (≥50%) or occlusion of an ipsilateral, inflow artery (e.g., aortoiliac, common femoral) must be successfully treated (use of investigational treatment prohibited) prior to treatment of the target lesion. Successful treatment is defined as no complications and less than 30% residual stenosis following intervention.
Patients were not permitted to enroll in the DETOUR2 study if they met any of the following exclusion criteria:
# General Exclusion Criteria
- Participating in another investigational study.
- Anticipated life expectancy less than 1 year or medical comorbid condition(s) that could limit the subject's ability to comply with the requirements of the trial.
- Subject has other medical, social or psychological problems that, in the opinion of the investigator, preclude them from receiving this treatment, and the procedures and evaluations pre- and post-treatment.
# Clinical Exclusion Criteria
- History of deep vein thrombosis on either limb.
- Thrombophlebitis, within the previous 30 days.
- Subject has a known coagulopathy or has bleeding diatheses, thrombocytopenia with platelet count less than 100,000/microliter or INR >1.8.
- Planned amputation of the target limb, including minor amputations.
- Prior distal amputation (above the trans metatarsal) of the target limb.
- Known or suspected active infection at the time of the procedure (e.g., WIfI foot infection grade 3: Severe infection. Local infection systemic inflammatory response syndrome [SIRS]).
- Rutherford Clinical Category 0, 1, 2 or 6.
- Has acute or chronic renal disease with GRF ≤30 ml/min per 1.73 m² and/or elevated serum creatinine >2.5 mg/dL (220 μmol/L) or on dialysis.
- Known hypersensitivity/allergy to the investigational devices and/or required pharmacotherapy that cannot be safely managed.
- Morbid obesity that does not allow for safe vascular access or imaging.
{16}
- Requires coronary or peripheral procedure within 30 days prior to or planned within 30 days post treatment of the target lesion.
- Has a known history of intracranial bleeding or aneurysm, myocardial infarction or stroke within the last 3 months.
- Subject is pregnant or breast-feeding.
## Angiographic Exclusion Criteria
- Stent within 3cm of SFA ostium.
- Previous bypass surgery on the target limb.
- Subject has significant disease or obstruction (≥50%) of the inflow tract that has not been successfully treated at the time of the index procedure (success measured as ≤30% residual stenosis, without complication).
- Presence of aneurysm or acute thrombus in the target limb.
- Thrombolysis of the target vessel within 72 hours prior to the index procedure, where complete resolution of the thrombus was not achieved.
## 2. Follow-up Schedule
All patients were scheduled to return for follow-up examinations at 1 month (±7 days), 6 months (±30 days), 12 months (±30 days), 24 months (±60 days), and 36 months (±60 days) post-procedure or when a subject withdrew prematurely from the study. The schedule of evaluations to be completed at each study interval is presented in Table 8.
Table 8: Evaluations to be Completed at Each Study Interval
| Activity | Pre-procedural Baseline Screening | Procedure | Discharge | Follow-Up Visits (M) | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | | 1 | 6 | 12 | 24 | 36 |
| Review of Inclusion / Exclusion Criteria | X | X | | | | | | |
| Medical History / Demographics | X | | | | | | | |
| Brief Physical Exam / Health Status | X | | X | X | X | X | X | X |
| Routine Laboratory Tests (CBC and Chem-7) | X | | | | | | | |
| Serum Creatinine to Calculate eGFR | X | | | | | | | |
| Ankle-Brachial Index (or Toe-Brachial Index¹) | X | | | X | X | X | X | X |
| Rutherford Assessment | X | | | X | X | X | X | X |
PMA P220021: FDA Summary of Safety and Effectiveness Data
Page 17 of 48
{17}
| Activity | Pre-procedural Baseline Screening | Procedure | Discharge | Follow-Up Visits (M) | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | | 1 | 6 | 12 | 24 | 36 |
| Venous Clinical Severity Score and Villalta Scale | X | | | X | X | X | X | X |
| Final Eligibility Angiograph / Venogram or Venous Ultrasound^{2} | | X | | | | | | |
| Venous Ultrasound^{3} and Venous Observation Scale | | | | X | X | X | X | X |
| Arterial Ultrasound^{3} | | | | X | X | X | X | X |
| Stent Graft X-Ray | | | | | | X | | |
| Adverse Event Assessment | | X | X | X | X | X | X | X |
| Dual Antiplatelet Therapy^{4} | X | X | X | X | X | X | X | X |
| VascuQOL | X | | | X | | X | | |
| EQ-5D-5L QOL | X | | | X | | X | | |
| SF-12 | X | | | X | | X | | |
| 6 Minute Walk Test^{5} | X | | | X | | X | | |
1 Perform Toe Brachial Index (TBI) only if unable to reliably assess ABI reading. Tests are performed in resting state.
2 Angiography performed throughout the index procedure. Angiography required to be submitted to the angiographic core lab for all target limb revascularization procedures.
3 Duplex Ultrasound is to be completed at each follow-up visit.
4 Effective anticoagulation therapy should be maintained throughout the procedure (minimum ACT >250 seconds is recommended). Dual anti-platelet or anti-coagulation therapy is recommended for at least three years.
5 The 6-Minute Walk Test is not routinely performed at all clinical sites; therefore, select sites shall complete a 6MWT on select subjects as agreed upon. This will therefore result in a sub-group of clinical sites and subjects and not be study wide.
## 3. Clinical Endpoints
The primary safety endpoint was freedom from major adverse events (MAEs) through 30 days post-procedure. The performance goal of freedom from primary safety event is 84.0%, which was set at 4% below the aggregate of published trial data as described by VIVA Physicians Inc. (VPI). The primary safety hypothesis was tested by calculating the lower one-sided 97.5% confidence limit, using the Exact method, for the rate of freedom from MAEs. The primary safety endpoint was evaluated on a per subject basis and was calculated as the percent of all Intent to Treat (ITT) subjects that are free from all elements of the primary safety endpoint composite at 30 days.
PMA P220021: FDA Summary of Safety and Effectiveness Data
Page 18 of 48
{18}
The primary effectiveness endpoint was primary patency at 12 months post procedure. Primary patency is defined as the absence of clinically driven target lesion revascularization (CD-TLR) and absence of recurrent target lesion diameter stenosis >50% by imaging (e.g., duplex ultrasound peak systolic velocity ratio peak systolic velocity ratio [PSVR] of >2.5 or as measured by invasive angiography) within the stent or immediately 1 cm above or below the treated segment.
The primary effectiveness endpoint was evaluated against a literature-derived performance goal (PG) of 60.4%, which was determined as the weighted mean of historical studies of endovascular treatment of similar lesions. The primary effectiveness hypothesis was tested by calculating the one-sided lower 97.5% confidence limit, using the Exact method. The primary effectiveness endpoint was evaluated on patients who received the device and with evaluable images as defined in Section X.B. Accountability of PMA Cohort.
## Secondary Endpoints
The following secondary safety endpoints were evaluated:
- Major Adverse Events (MAE) at 6, 12, 24 and 36 months.
- Major Adverse Limb Event (MALE) defined as above-ankle amputation or major reintervention including placement of a new bypass graft, interposition graft, thrombectomy, or thrombolysis.
- Major bleeding is defined as any bleeding event requiring transfusion of ≥2 units packed red blood cells (PRBC) or surgical repair through 30 days.
- Symptomatic Deep Vein Thrombosis (DVT) on ipsilateral limb at 30 days and 6, 12, 24 and 36 months.
- Pulmonary embolism at 30 days and 6, 12, 24 and 36 months.
- Perioperative myocardial infarctions through 30 days.
- Hematoma ≥8cm related to the device or procedure through 30 days.
- The combined rate of death, target lesion revascularization (TLR), index limb amputation, and an increase in Rutherford Clinical Category by 2 classes (comparing pre- to post-procedural assessments) at 30 days and 6, 12, 24 and 36 months.
- Stent graft separation and migration identified via ultrasound imaging at 30 days and 6, 12, 24 and 36 months.
- Stent graft separation and migration identified via x-ray at 12 months.
- Stent graft fracture identified via x-ray at 12 months.
The following secondary effectiveness endpoints were evaluated:
- Technical Success defined as successful delivery of the investigational devices to the identified area and removal of delivery system.
- Procedural Success defined as successful delivery of the investigational devices to the identified area and removal of delivery system in the absence of in-hospital MAEs.
PMA P220021: FDA Summary of Safety and Effectiveness Data
Page 19 of 48
{19}
- Clinical Success defined as limb ischemia improvement as assessed by Rutherford Clinical Category (improvement in scale by ≥1) at 30 days and 6, 12, 24 and 36 months.
- Limb ischemia by Rutherford Clinical Category through follow-up at 30 days and 6, 12, 24 and 36 months.
- Primary Patency at 30 days and 6, 24 and 36 months.
- Primary assisted patency defined as revascularization of nonocclusive (<99%) stenosis within the stent graft or immediately above or below the treated arterial segment with less than 50% residual stenosis at 30 days and 6, 12, 24 and 36 months.
- Secondary patency defined as revascularization of occlusion (100%) within the stent graft or immediately above or below the treated arterial segment with less than 50% residual stenosis at 30 days and 6, 12, 24 and 36 months.
- Target Vessel Revascularization defined as any repeat percutaneous intervention or surgical bypass of any segment of the target vessel at 30 days and 6, 12, 24 and 36 months.
- Ankle-Brachial Index (ABI) or Toe Brachial Index (TBI) of both the index limb and contralateral limb at 30 days and 6, 12, 24 and 36 months.
- Number of major index limb amputations at 30 days and 6, 12, 24 and 36 months.
- Major procedure-related infections within 30 days.
- Length of post-procedure hospital stay.
- Length of ICU stay.
- Hospital re-admissions through 30 days.
- Change from baseline in VascuQol QOL Score at 30 days and 12 months.
- Change from baseline in EQ-5D-5L QOL Score at 30 days and 12 months.
- Change from baseline in Duplex Venous Observation Scale at 30 days and 6, 12, 24 and 36 months.
- Change from baseline in Villalta and Venous Clinical Severity (VCSS) Scales at 30 days and 6, 12, 24 and 36 months.
- SF-12 Score at baseline, 30 days and 12 months.
## Subgroup Analyses
The following subgroups were analyzed for the primary endpoints: Female gender, diabetics, and US cohorts.
## B. Accountability of PMA Cohort
At the time of database lock, 220 patients enrolled in the DETOUR2 study at 36 sites. 85.5% (188) patients are available for analysis for the 12-month post-operative visit. Overall, 220 subjects were treated with the ENDOCROSS Device.
The analyzed patient cohorts are defined as:
Roll-ins: The first two subjects at each U.S. site may be considered roll-in subjects. Roll-in subjects were pre-identified. Not all sites enrolled roll-in subjects.
PMA P220021: FDA Summary of Safety and Effectiveness Data
Page 20 of 48
{20}
Intention to Treat (ITT): All subjects who received the intervention and were enrolled in the IDE Study (excluding roll-ins). The ITT Cohort was the primary analysis set to determine if the primary safety endpoint was met in the study.
Modified ITT (MITT): Only those subjects/lesions where a DETOUR System: TORUS Stent Graft was implanted. The MITT Cohort is a subset of the ITT Cohort. The MITT Cohort is the primary analysis set to determine if the primary effectiveness endpoint was met in the study.
Per Protocol (PP): Subset of the MITT group which excludes all subjects with major protocol deviations (e.g., violations of eligibility criteria) and missing the data required to evaluate the primary effectiveness endpoint.
Of the 220 subjects who received the ENDOCROSS Device, 18 were considered roll-in subjects and remaining 202 (91.8%) are considered part of the ITT Cohort. Two subjects from the ITT group did not receive the TORUS Stent Graft, thus 200 patients were part of the MITT Cohort. Of the MITT Cohort, 197 (89.5%) were considered part of the Per Protocol Cohort.
Figure 5 depicts the accountability of subjects at the study follow-up time points. Of the 220 subjects enrolled, 83.6% (184) patients are available for analysis for the 12-month post-operative visit. Eighteen (18) subjects in the ITT group exited before the 12-month window: 2 enrolled but did not receive the ENDOCROSS device, 7 subject withdrawals, 2 physician withdrawal, 5 deaths, and 2 lost-to-follow-up.
PMA P220021: FDA Summary of Safety and Effectiveness Data
Page 21 of 48
{21}

Figure 5: Subject Disposition
# C. Study Population Demographics and Baseline Parameters
Specific demographics and baseline characteristics for the DETOUR2 Study ITT population are presented in Table 9. Medical history and health status are provided in Table 10. Subgroup analysis of the ITT population is provided in Section D Safety and Effectiveness Results.
The mean age was $68.9 \pm 9.4$ years and $26.2\%$ were female. $86.6\%$ of the patients were white. Comorbidities included coronary artery disease $(87.6\%)$ , hypertension $(87.6\%)$ , diabetes $(34.7\%)$ , prior history of smoking $(91.1\%)$ , and renal insufficiency $(10.9\%)$ . The mean ABI was $0.61 \pm 0.22$ at baseline. Most patients $(77.7\%)$ were Rutherford Clinical Category (RCC) 3. The remainder of the patients $(22.3\%)$ were RCC 4 and RCC 5.
Table 9: Baseline Demographics - ITT Population
| Variable | All Subjects |
| --- | --- |
| Age at Consent (years) | 68.9 ± 9.38 (202) |
| | 69.0 (47, 88) |
| Body Mass Index (BMI) | 28.74 ± 5.028 (202) |
| | 28.50 (18.3, 45.0) |
PMA P220021: FDA Summary of Safety and Effectiveness Data
{22}
| Variable | All Subjects |
| --- | --- |
| Sex | -- |
| Female | 26.2% (53/202) |
| Male | 73.8% (149/202) |
| Race1 | -- |
| American Indian or Alaska Native | 0.5% (1/202) |
| Asian | 0.0% (0/202) |
| Black or African American | 8.9% (18/202) |
| Native Hawaiian or Other Pacific Islander | 0.0% (0/202) |
| White | 86.6% (175/202) |
| Other | 4.5% (9/202) |
| Ethnicity | -- |
| Hispanic or Latino | 4.6% (9/196) |
| Not Hispanic or Latino | 79.6% (156/196) |
| Not Reported | 15.8% (31/196) |
Categorical variables presented as $\%$ $(n / N)$ and continuous variables presented as mean $\pm SD$ (N) median (min, max) where $N$ is the number of subjects with available data.
One subject identified under two (2) categories (White, American Indian or Alaska Native).
Table 10: Medical History and Health Status - ITT Population
| Variable | All Subjects |
| --- | --- |
| Renal Insufficiency (with or without intervention) | 10.9% (22/202) |
| History of Smoking | -- |
| Current/Previous | 91.1% (184/202) |
| Never | 8.9% (18/202) |
| History of Peripheral Venous Disease (DVT, Thrombophlebitis, etc.) | 0.0% (0/202) |
| Peripheral Arterial Disease | 98.0% (198/202) |
| Previous Peripheral Intervention | 60.5% (121/200) |
| Previous Peripheral Vascular Surgery | 16.8% (34/202) |
| Hypertension | 87.6% (177/202) |
| Diabetes Mellitus | 34.7% (70/202) |
| Type 1 | 1.4% (1/70) |
| Type 2 | 98.6% (69/70) |
| Hypercholesterolemia | 73.2% (145/198) |
PMA P220021: FDA Summary of Safety and Effectiveness Data
{23}
| Variable | All Subjects |
| --- | --- |
| Hyperlipidemia | 73.3% (140/191) |
| Coronary Artery Disease | 46.0% (92/200) |
| Congestive Heart Failure | 12.4% (25/201) |
| NYHA I | 16.0% (4/25) |
| NYHA II | 48.0% (12/25) |
| NYHA N.A. | 36.0% (9/25) |
| History of Myocardial Infarction | 21.9% (44/201) |
| History of Cerebrovascular Disease | 13.9% (28/201) |
| Stroke | 60.7% (17/28) |
| TIA | 21.4% (6/28) |
| Other | 17.9% (5/28) |
| Target Limb Ankle-Brachial Index (ABI) | 0.61 ± 0.22 (193) 0.60 (0.00, 1.55) |
| Contralateral Limb Ankle-Brachial Index (ABI) | 0.82 ± 0.26 (189) 0.82 (0.00, 1.63) |
| Target Limb Toe-Brachial Index (TBI) | 0.45 ± 0.19 (58) 0.43 (0.00, 0.83) |
| Contralateral Limb Toe-Brachial Index (TBI) | 0.59 ± 0.23 (57) 0.60 (0.00, 1.10) |
| Rutherford Clinical Category | -- |
| Grade 3 | 77.7% (157/202) |
| Grade 4 | 17.8% (36/202) |
| Grade 5 | 4.5% (9/202) |
Categorical variables presented as $\%$ $(n / N)$ and continuous variables presented as mean $\pm SD$ (N) median (min, max) where $N$ is the number of subjects with available data.
Table 11 summarizes the baseline lesion characteristics determined by the Core Lab. Target lesions were relatively equally distributed between the right and left limb. Pre-procedure chronic occlusion (CTO; $100\%$ stenosis) was observed in $96.0\%$ (194/202) of subjects and diffuse stenosis ( $>70\%$ stenosis) was observed in $97.0\%$ (196/202) of subjects. The mean lesion length was $327.14 \pm 61.38\mathrm{mm}$ . The mean CTO length was $217.31 \pm 85.98\mathrm{mm}$ . The calcification grade for lesions was predominantly severe ( $70.4\%$ , 126/179) followed by none/mild ( $29.1\%$ , 52/179) and then moderate ( $0.6\%$ , 1/179).
Two vessel run-off to the foot was observed in $69.0\%$ (129/187) of subjects. There were no subjects who had zero run-off vessels to the foot. The popliteal artery was involved in approximately $10\%$ of lesions, and approximately $44\%$ of subjects had below-the-knee arterial disease as well.
PMA P220021: FDA Summary of Safety and Effectiveness Data
{24}
Table 11: Core Lab Baseline Lesion Characteristics - ITT Population
| Variable | All Subjects |
| --- | --- |
| Total Occlusion (100% stenosis) | 96.0% (194/202) |
| Diffuse Stenosis (>70% stenosis) | 97.0% (196/202) |
| In-Stent Restenosis | 17.3% (35/202) |
| Lesion Length (Normal to Normal, mm) | 327.14 ± 61.38 (196) 328.15 (194.6, 520.3) |
| Calcified Length (mm) | 64.12 ± 77.50 (178) 41.10 (0.0, 415.7) |
| CTO Length (mm) | 217.31 ± 85.98 (191) 232.50 (0.0, 436.1) |
| Calcification | -- |
| None/Mild | 29.1% (52/179) |
| Moderate | 0.6% (1/179) |
| Severe | 70.4% (126/179) |
Categorical variables presented as $\%$ $(n / N)$ and continuous variables presented as mean $\pm SD$ (N) median (min, max) where $N$ is the number of subjects with available data.
Of the 202 ITT subjects, the device was implanted in 200. Two (2) subjects did not receive the device: one (1) case was aborted because there was a complication with a percutaneous transluminal angioplasty (PTA) balloon which extended the procedure time, and one (1) case was aborted because venous access was difficult.
Table 12 summarizes site reported procedure characteristics. The majority of subjects were treated under conscious sedation (40.6%; 82/202) or local anesthesia (28.2%; 57/202). Mean procedure time was $181.4 \pm 90.55$ minutes, with a mean fluoroscopy time of $46.4 \pm 19.51$ minutes. An average of $208.1 \mathrm{ml} \pm 111.74$ of contrast was used per subject. Estimated blood loss was $50 \pm 57.50 \mathrm{ml}$ per subject.
Table 12: Site Reported Procedure Characteristics - ITT
| Variable | All Subjects |
| --- | --- |
| Type of Anesthesia Used | -- |
| General | 12.4% (25/202) |
| Local | 28.2% (57/202) |
| Epidural/Spinal | 5.9% (12/202) |
| Conscious Sedation | 40.6% (82/202) |
| Other | 12.9% (26/202) |
| Estimated Blood Loss (ml) | 50.0 ± 57.50 (201) 30.0 (0, 400) |
| Contrast Volume Used (ml) | 208.1 ± 111.74 (199) 180.0 (50, 900) |
PMA P220021: FDA Summary of Safety and Effectiveness Data
{25}
| Variable | All Subjects |
| --- | --- |
| Fluoroscopy Time (min) | 46.4 ± 19.51 (199) |
| | 42.0 (12, 122) |
| Total Procedure Time (min) | 181.4 ± 90.55 (202) |
| | 163.0 (55, 495) |
Categorical variables presented as % (n/N) and continuous variables presented as mean ± SD (N) median (min, max) where N is the number of subjects with available data.
Core laboratory reported post-procedure angiographic data are presented in Table 13.
Table 13: Post-Procedure Core Lab Angiography Data
| Post-Procedure Characteristics | All Subjects |
| --- | --- |
| Distance of Proximal TORUS Edge from Superficial Femoral Artery Ostium (mm) | 5.91 ± 5.672 (183)
4.40 (0.0, 50.8) |
| Distance from Distal TORUS Edge to Tibial Plateau (mm) | 42.55 ± 36.22 (183)
28.00 (2.4, 150.5) |
| Proximal Intra-Arterial TORUS Length | 48.10 ± 17.68 (187)
45.90 (13.2, 125.6) |
| Distal Intra-Arterial TORUS Length | 59.44 ± 20.82 (187)
54.40 (19.2, 129.4) |
| Overlap Lengths (mm) | -- |
| Proximal (or only) | 72.00 ± 20.54 (169)
69.00 (28.5, 139.1) |
| Middle | 43.60 ± 23.11 (3)
35.80 (25.4, 69.6) |
| Distal | 66.14 ± 16.54 (134)
63.15 (28.1, 179.9) |
| Run-Off Vessels to the Foot | -- |
| 0 | 0.0% (0/183) |
| 1 | 15.3% (28/183) |
| 2 | 67.2% (123/183) |
| 3 | 17.5% (32/183) |
Categorical variables presented as % (n/N) and continuous variables presented as mean ± SD (N) median (min, max) where N is the number of subjects with available data.
Technical success in the MITT population, defined as successful delivery of the investigational devices to the identified area and removal of delivery system was 100%, in the 200 subjects. Procedural success in the MITT population, defined as successful delivery of the investigational device to the identified area and removal of the delivery system in the absence of in-hospital MAEs, was 98.5% (197/200). Three (3) patients had MAEs prior to
PMA P220021: FDA Summary of Safety and Effectiveness Data
Page 26 of 48
{26}
discharge: Two (2) subjects were reported to have major bleeding, and one (1) subject was reported to have had a symptomatic DVT.
## D. Safety and Effectiveness Results
### 1. Safety Results
The primary safety endpoint was freedom from a composite of Major Adverse Events (MAE) at 30 days (Table 14) as adjudicated by the CEC. MAEs were defined as any of the following: all-cause mortality, CD-TLR, amputation of the treated limb, occlusive-symptomatic deep vein thrombosis, pulmonary embolism, or procedure-related bleeding. There were 15 total 30-day MAEs in 14 subjects: 3 CD-TLR, 5 DVT and 7 major bleeding events.
There were 199 subjects of the 202 subjects in the ITT group evaluated for the primary safety endpoint at 30 days. Freedom from MAE at 30 days was 93.0% (185/199), with a lower exact one-sided 97.5% confidence interval of 88.5% compared to the PG 84%. Thus, the primary safety endpoint was met. The composite primary safety endpoint is summarized in Table 14. There were 199 subjects of the 202 subjects in the ITT group evaluated for the primary safety endpoint at 30 days. The 30-day MAE rate for the roll-in cohort (n=18) was comparable to the pivotal [ITT] cohort (5.6% in roll-in versus 7.0% in pivotal ITT).
Major adverse events are reported in Table 15. Follow-up is on-going through 36 months.
Table 14: Freedom from Major Adverse Event (MAE) at 30 Days - ITT Population
| | All Subjects | Lower 97.5% Exact Confidence Limit |
| --- | --- | --- |
| Freedom from MAE^{1} | 93.0% (183/199)^{2}
15 total 30-day MAE events in 14 subjects:
• 3 CD-TLR
• 5 DVT
• 7 Major Bleeding Events | 88.5% > 84% P |
1 Freedom from a major adverse event (MAE) at 30-days post-procedure defined as any occurrence of the following events: Death, Clinically Driven Target Lesion Revascularization (CD-TLR), Amputation of the Treated Limb, Symptomatic Deep Vein Thrombosis (DVT), Pulmonary Embolism (PE) or procedure-related bleeding requiring any transfusion of packed red blood cells or surgery.
2 There were 199 subjects of the 202 subjects in the ITT group evaluated for the primary safety endpoint at 30 days.
Table 15: Major Adverse Events Over Time: Kaplan Meier Estimates - ITT Population
| Event | 30 Days | 6 Months | 12 Months |
| --- | --- | --- | --- |
| MAE | 7.0%
[3.5%, 10.5%] | 13.6%
[8.8%, 18.3%] | 19.8%
[14.2%, 25.4%] |
PMA P220021: FDA Summary of Safety and Effectiveness Data
Page 27 of 48
{27}
| Event | 30 Days | 6 Months | 12 Months |
| --- | --- | --- | --- |
| Death Events1 | 0.0% [0.0%, 0.0%] | 0.5% [0.0%, 1.5%] | 2.6% [0.4%, 4.9%] |
| CD-TLR | 1.5% [0.0%, 3.2%] | 6.6% [3.1%, 10.0%] | 12.3% [7.7%, 17.0%] |
| Amputation of Treated Limb | 0.0% [0.0%, 0.0%] | 1.0% [0.0%, 2.4%] | 2.1% [0.1%, 4.1%] |
| Deep Vein Thrombosis | 2.5% [0.3%, 4.7%] | 3.5% [1.0%, 6.1%] | 4/1% [1.3%, 6.8%] |
| Pulmonary Embolism | 0.0% [0.0%, 0.0%] | 0.0% [0.0%, 0.0%] | 0.0% [0.0%, 0.0%] |
| Procedure Related Bleeding | 3.5% [1.0%, 6.0%] | 3.5% [1.0%, 6.0%] | 3.5% [1.0%, 6.0%] |
Data presented as $\%$ (n/N) [m] where $n$ is the number of subjects experiencing at least 1 event, $N$ is the number of subjects, and $m$ is the total number of events.
There were 11 adverse events that resulted in death: none of these adverse events were related to the device or the procedure. Three (3) deaths were reported for unknown causes.
# Serious Adverse Events that Occurred in the PMA Clinical Study:
Overall, $66.3\%$ (134/202) of subjects had a serious adverse event (SAE) within 12 months; almost half were vascular disorders based on the System Organ Category (SOC) classification. Of all SAEs, $24.3\%$ (49/202) were device related SAEs, as adjudicated by the CEC. The most common SOC for device-related SAEs were Vascular Disorders, with 48/202 subjects $(23.8\%)$ experiencing at least one SAE in the SOC.
Overall, $17.3\%$ (35/202) procedure related SAEs were adjudicated by the CEC, were observed through 12-months. The most common SOC for procedure-related SAEs was Vascular Disorders, with 21/202 subjects $(10.4\%)$ experiencing at least one adverse event in this SOC.
Table 16: Serious Adverse Events - ITT Population
| Adjudicated Event | Body System or Organ Class / Preferred Level Term | Through 12 Months % (n/N) [m] |
| --- | --- | --- |
| All Serious Adverse Events | Any Event | 66.3% (134/202) [355] |
| | Vascular Disorders | 48.5% (98/202) [167] |
| Serious Device Related Adverse Event | Any Event | 24.3% (49/202) [67] |
| | Vascular Disorders | 23.8% (48/202) [59] |
PMA P220021: FDA Summary of Safety and Effectiveness Data
{28}
| Adjudicated Event | Body System or Organ Class / Preferred Level Term | Through 12 Months % (n/N) [m] |
| --- | --- | --- |
| Serious Procedure Related Adverse Event | Any Event | 17.3% (35/202) [57] |
| | Vascular Disorders | 10.4% (21/202) [26] |
Data presented as $\%$ $(n / N)$ [m] where $n$ is the number of subjects experiencing at least 1 event, $N$ is the number of subjects, and $m$ is the total number of events.
# 2. Secondary Safety Endpoints
Secondary safety endpoint Kaplan Meier Estimates were performed on the ITT analysis set (Table 17). Major Adverse Limb Events (MALE) were reported at 30 days, 6 months, and 12 months, with a rate of $4.0\%$ , $10.5\%$ , and $15.7\%$ , respectively.
Major Bleeding was reported in eight (8) subjects, with nine (9) events occurring within 30 days of the index procedure. Eight (8) of these events were related to the procedure. Five (5) of these eight (8) procedural related events were due to blood loss anemia, there were two (2) gastrointestinal bleeds and one retroperitoneal hematoma.
Symptomatic DVT occurred in eight (8) subjects through 12 months with no new symptomatic DVT in the subjects followed through 24 months and 36 months. Most DVT (7/8) occurred within 6 months of the index procedure. There has been no Pulmonary Embolism (PE) observed in the ITT Cohort at the time of this report. There was one (1) perioperative Myocardial Infarction (MI) in the ITT population, through 30 days $(0.5\%; 1/199)$ . There were three (3) MIs through 12 months in the ITT cohort and seven (7) MIs total in patients followed through 36 months.
The percentage of hematoma $(\geq 8\mathrm{cm})$ within 30 days of the index procedure that were related to the device or procedure is $0.5\%$ (1/199). At 30-days follow-up, the stent thrombosis Kaplan Meier estimate was $3.5\%$ .
Table 17: Secondary Safety Endpoints Over Time: Kaplan Meier Estimates - ITT Population
| Event | 30 Days | 6 Months | 12 Months |
| --- | --- | --- | --- |
| Major Adverse Limb Events (MALE) | 4.0% [1.3%, 6.7%] | 10.5% [6.3%, 14.8%] | 15.7% [10.7%, 20.9%] |
| Major Bleeding | 4.0% [1.3%, 6.7%] | 6.0% [2.7%, 9.3%] | 6.6% [3.1%, 10.0%] |
| Symptomatic Deep Vein Thrombosis (DVT) | 2.5% [0.3%, 4.7%] | 3.5% [1.0%, 6.1%] | 4.1% [1.3%, 6.8%] |
PMA P220021: FDA Summary of Safety and Effectiveness Data
{29}
| Event | 30 Days | 6 Months | 12 Months |
| --- | --- | --- | --- |
| Pulmonary Embolism (PE) | 0.0% [0.0%, 0.0%] | 0.0% [0.0%, 0.0%] | 0.0% [0.0%, 0.0%] |
| Myocardial Infarction (MI) | 0.5% [0.0%, 1.5%] | 0.5% [0.0%, 1.5%] | 1.6% [0.0%, 3.3%] |
| Hematoma | 0.5% [0.0%, 2.8%] | N/A | N/A |
| Stent Thrombosis | 3.5% [1.0%, 6.0%] | 9.0% [5.1%, 13.0%] | 15.8% [10.7%, 20.9%] |
Data presented as $(n / N)\%$ where $n$ is the number of subjects experiencing at least 1 event and $N$ is the number of subjects.
# Venous Outcomes
The definition of a DVT is any organized clot that is occluded within the deep venous system and results in a lack of flow and a lack of compressibility of the vein. This definition is distinct from catheter-related thrombosis or protein/fibrin deposition that have typically been reported with the use of indwelling central venous catheters and cardiac implantable electronic devices, which are largely asymptomatic, do not occlude the vein, and can be found in the vicinity of the device. Venous events are described below based on CEC adverse event definitions and clinical relevance. The methodologies of categorization of these events are not mutually exclusive. Table 18 categorizes venous events based upon the CEC definitions of serious and non-serious.
Serious DVTs are those that resulted in either death, life-threatening illness/injury, permanent impairment of a body structure or a body function, inpatient or prolonged hospitalization, or medical or surgical intervention to prevent life-threatening illness or injury or permanent impairment.
Non-serious DVTs are those that did not result in death, life-threatening illness/injury, permanent impairment of a body structure or a body function, inpatient or prolonged hospitalization, or medical or surgical intervention to prevent life-threatening illness or injury or permanent impairment. These may include fibrin sheaths which did not occlude the vein, are largely asymptomatic, and can be found in the vicinity of the device.
Table 18: Deep Vein Thrombosis (DVT) - ITT Population
| Event | 30 Days % (95% CI) {n/N} [m] | 6 Months % (95% CI) {n/N} [m] | 12 Months % (95% CI) {n/N} [m] |
| --- | --- | --- | --- |
| Non-serious DVT | 10.9% (7.0, 16.0) {22/202} [22] | 13.9% (9.4, 19.4) {28/202} [30] | 13.9% (9.4, 19.4) {28/202} [30] |
| Non-serious Device Related DVT | 6.9% (3.8, 11.4) {14/202} | 8.9% (5.4%, 13.7%) {18/202} | 8.9% (5.4%, 13.7%) {18/202} |
PMA P220021: FDA Summary of Safety and Effectiveness Data
{30}
| Event | 30 Days % (95% CI) {n/N} [m] | 6 Months % (95% CI) {n/N} [m] | 12 Months % (95% CI) {n/N} [m] |
| --- | --- | --- | --- |
| | [14] | [18] | [18] |
| Non-serious Procedure Related DVT | 10.9% (7.0, 16.0) {22/202} [22] | 10.9% (7.0, 16.0) {22/202} [22] | 10.9% (7.0, 16.0) {22/202} [22] |
| Serious DVT (or Endpoint DVT) | 2.5% (0.8, 5.7) {5/202} [5] | 4.0% (1.7, 7.7) {8/202} [8] | 4.0% (1.7, 7.7) {8/202} [8] |
| Serious Device Related DVT | 1.5% (0.3, 4.3) {3/202} [3] | 2.5% (0.8, 5.7) {5/202} [5] | 2.5% (0.8, 5.7) {5/202} [5] |
| Serious Procedure Related DVT | 2.5% (0.8, 5.7) {5/202} [5] | 3.0% (1.1, 6.4) {6/202} [6] | 3.0% (1.1, 6.4) {6/202} [6] |
| All DVT Events (serious and non-serious, regardless of relation to device or procedure) | 27 events | 38 events | 38 events |
Data presented as (95% exact CI) $\{n / N\}$ [m] where $n$ is the number of subjects experiencing at least 1 event, $N$ is the number of subjects, and $m$ is the total number of events.
Figure 6 below organizes the venous events by clinical relevance. The events were categorized as symptomatic DVT, which are venous events that are occlusive within a deep vein and associated with clinical symptoms, and those that are not. Symptomatic DVTs (or endpoint DVTs) were considered as part of the safety endpoint. Venous events that are not symptomatic are likely the result of fibrin deposition/sheath. Details on gender and treatment are also reported.
There were a total of 38 venous events reported in 36 patients from patients reporting of symptoms and venous duplex surveillance at follow-up visits. Most were found incidentally on follow-up imaging. All of these events were reviewed by an independent Clinical Events Committee and Medical Monitor. There were eight (8) symptomatic DVTs in eight (8) patients that were considered as part of the safety endpoint. However, one (1) was later assessed to be an arterial occlusion in the bypass graft. Symptomatic DVT occurred in $7.5\%$ (4/53) of females and $2.0\%$ (3/147) of males. All 7 symptomatic DVTs resolved.
There were a total of 30 venous events in 28 patients that were not considered symptomatic or occlusive, and most were found incidentally on follow up imaging. One (1) was later assessed to be an arterial occlusion of the SFA and popliteal artery, Venous events occurred in $15.1\%$ (8/53) of females and $12.9\%$ (19/147) of males; 14 patients started on anticoagulation and 13 patients had no additional treatment. All but one (1) of
PMA P220021: FDA Summary of Safety and Effectiveness Data
{31}
the venous events resolved within 6 months. The one venous event was resolved by year 2, without any further reintervention.
Overall, there were no significant interventions (e.g., lysis, thrombectomy) used in the treatment of any of the venous events. There was no progression in size, or progression to pulmonary embolism.

Figure 6: Deep Vein Thrombosis - MITT Accountability
$^{*}N = 8$ events were counted against the primary safety endpoint (i.e., freedom from MAEs at 30 days), however $N = 1$ event was later determined to be a bypass occlusion
$^{\dagger}N = 28$ events were reported as venous events but not counted against the primary safety endpoint, however $N = 1$ event was an arterial occlusion
# Duplex Venous Observation Scale (DVOS)
Duplex Venous Observation Scale was performed at 30 days, 6 months and 12 months, as shown in Table 19 in the ITT Cohort. The analysis demonstrates a DVOS of 0 for the majority of subjects indicating the vein was patent with no presence of fibrin or thrombus.
Table 19: Duplex Venous Observation Scale - ITT
| DVOS | 30 Days (N=1731) | 6 Months (N=175) | 12 Months (N=171) |
| --- | --- | --- | --- |
| 0 | 83.2% (144) | 94.3% (165) | 96.5% (165) |
| 1 | 0.6% (1) | 1.1% (2) | 0.0% (0) |
| 2 | 1.7% (3) | 0.0% (0) | 0.6% (1) |
| 3 | 0.6% (1) | 0.0% (0) | 0.0% (0) |
PMA P220021: FDA Summary of Safety and Effectiveness Data
{32}
| DVOS | 30 Days (N=173¹) | 6 Months (N=175) | 12 Months (N=171) |
| --- | --- | --- | --- |
| 4 | 11.0% (19) | 4.0% (7) | 2.3% (4) |
| 5 | 2.9% (5) | 0.6% (1) | 0.6% (1) |
¹Missing imaging assessment from the expected ITT Cohort group (202 at 30 days)
# Venous Clinical Severity Scale (VCSS) and Villalta Scale Results
Results for the VCSS scales at baseline and for each visit interval (30 days, 6 months, 12 months, 24 months and 36 months) are summarized in Table 20. The VCSS instrument provides a general assessment for patients with chronic venous disease. A negative change from baseline indicates improvement in the parameter assessed. At 30 days, 6 months and 12 months, patients reported less pain. In the patients that have reached the 24- and 36-months follow-up, the pain and varicose veins were improved. However, overall VCSS score at 1 year was $0.3 \pm 1.98$ , indicating on average the study subjects had a decline in their status with regard to their preexisting chronic venous disease.
Table 20: Venous Clinical Severity Score - ITT
| Visit | All Available | Change from Baseline (Follow-up – Baseline) |
| --- | --- | --- |
| Baseline | 0.9 ± 0.89 (202) 1.0 (0,2) | -- |
| 30 Day | 1.7 ± 2.23 (194) 1.0 (0.14) | 0.7 ± 2.11 (194) 0.0 (-2, 12) |
| 6 Months | 1.2 ± 2.08 (189) 1.0 (0, 19) | 0.3 ± 1.99 (189) 0.0 (-2, 17) |
| 12 Months | 1.3 ± 1.97 (173) 1.0 (0, 12) | 0.3 ± 1.98 (173) 0.0 (-2, 11) |
Data presented as mean $\pm$ SD (N) median (min, max) where $N$ is the number of subjects with available data.
Results for the Villalta scales scores for each visit interval (30 days, 6 months, 12 months, 24 months and 36 months) are summarized in Table 21. The Villalta score is used for post-thrombotic syndrome. It combines a patient self-assessment of symptoms and a healthcare assessment of clinical signs of post-thrombotic syndrome. A negative change from baseline indicates improvement in the parameter assessed. At 30 days, 6 months, 12 months, the overall Villalta score showed an improvement. The overall change in score at 1 year was $-0.6 \pm 3.23$ . In the patients who had reached 24- and 36-months follow-up, an improvement remains.
PMA P220021: FDA Summary of Safety and Effectiveness Data
{33}
Table 21: Villalta Scale - ITT
| Visit | All Available | Change from Baseline (Follow-up – Baseline) |
| --- | --- | --- |
| Baseline | 2.0 ± 3.08 (202) 1.0 (0.17) | -- |
| 30 Day | 1.7 ± 2.86 (194) 0.5 (0, 19) | -0.4 ± 3.29 (194) 0.0 (-12, 17) |
| 6 Month | 1.3 ± 2.37 (189) 0.0 (0, 16) | -0.8 ± 3.16 (189) 0.0 (-14, 13) |
| 12 Month | 1.5 ± 2.76 (173) 0.0 (0.17) | -0.6 ± 3.23 (173) 0.0 (-13, 11) |
Data presented as mean $\pm$ SD (N) median (min, max) where $N$ is the number of subjects with available data.
# Stent Graft Separation and Migration
Assessment of Stent Graft Separation and Migration was performed at 30 days, 6 months, 12 months, 24 months and 36 months, as shown in Table 22, through differing imaging modalities. One ultrasound detected stent graft separation and migration at the 12-month follow-up visit. One 12-month X-ray detected three (3) migrations (1.8%; 3/169) and one stent separation (0.6%; 1/169). There were no stent graft fractures at 12-months, and no additional separations, migrations, or fractures have been seen in patients who have reached follow-up through 36 months.
Table 22: Stent Graft Separation and Migration via Ultrasound and X-ray - ITT
| | | 30 Days | 6 Months | 12 Months |
| --- | --- | --- | --- | --- |
| Ultrasound | Stent Graft Separation | 0.0% (0/192) [0.0%, 1.9%] | 0.0% (0/188) [0.0%, 1.9%] | 0.6% (1/180) [0.0%, 3.1%] |
| | Stent Graft Migration | 0.0% (0/192) [0.0%, 1.9%] | 0.0% (0/188) [0.0%, 1.9%] | 0.6% (1/180) [0.0%, 3.1%] |
| | Stent Graft Separation or Migration | 0.0% (0/192) [0.0%, 1.9%] | 0.0% (0/188) [0.0%, 1.9%] | 0.6% (1/180) [0.0%, 3.1%] |
| X-Ray | Stent Graft Separation | N/A | N/A | 1.8% (3/169) [0.4%, 5.1%] |
| | Stent Graft Migration | N/A | N/A | 0.6% (1/169) [0.0%, 3.3%] |
| | Stent Graft Separation or Migration | N/A | N/A | 1.8% (3/169) [0.4%, 5.1%] |
Data presented as $\%$ (n/N) [95% CI] where $N$ is the number of subjects with available data.
PMA P220021: FDA Summary of Safety and Effectiveness Data
{34}
# 3. Effectiveness Results
The analysis of effectiveness was based on the 188 evaluable patients at the 12-month time point. The primary effectiveness endpoint was primary patency, defined as freedom from clinically driven target lesion revascularization (CD-TLR) and absence of recurrent target lesion restenosis of $>50\%$ diameter reduction by imaging (duplex ultrasound peak systolic velocity ratio of $>2.5$ or angiography) within the stent or $1\mathrm{cm}$ immediately above or below the treated segment at 12 months. When both imaging modalities are available, angiography takes precedence.
Primary patency at 12 months was $68.1\%$ (128/188). The lower bound of an exact one-sided $97.5\%$ confidence interval is $60.9\%$ . As this is above the prespecified $60.4\%$ threshold of the null hypothesis, the null hypothesis is rejected, and the evidence supports that the primary effectiveness is met. The primary effectiveness outcomes are summarized in Table 23. Information on the 12-month primary patency rate with the roll-in cohort is limited by a low number of patients available for analysis $(n = 18)$ . The primary patency Kaplan-Meier estimate is $57.1\%$ and is within the $95\%$ confidence band of the roll-in cohort.
Table 23: Primary Effectiveness Endpoint, Primary Patency at 12 Months - MITT
| | All Subjects | Lower 97.5% Exact Confidence Limit |
| --- | --- | --- |
| Absence of CD-TLR1 and absence of recurrent target lesion diameter stenosis >50% | 68.1% (128/188) Patency failures 30% (60/200) CD-TLR: 14% (28/200) [35] PSVR >2.5: 24% (48/200) | 60.9% > 60.4% |
Target lesion revascularization performed due to target lesion diameter stenosis $>50\%$ (e.g., duplex ultrasound peak systolic velocity ratio of $>2.5$ or invasive angiography) within the stent or immediately 1cm above or below the treated segment, AND either evidence of clinical or functional ischemia (e.g., recurrent/progressive intermittent claudication, critical limb ischemia) OR recurrence of the clinical syndrome for which the initial procedure was performed.
In order to meet the criteria for Clinically Driven (CD), the CEC members reviewed the reported TLR source documentation against the CD-TLR definition. To meet the definition of CD-TLR, at least two (2) reviewers must have determined that the event met at least two of the first three criteria. It is noted that the criteria did not need to be the same between the reviewers, only that at least two clinical criteria are met. All TLR are described in Figure 7 below.
PMA P220021: FDA Summary of Safety and Effectiveness Data
{35}

Figure 7: Arterial System at 1 Year - MITT Accountability
# 4. Secondary Effectiveness Endpoints
All secondary effectiveness analyses were performed on the primary analysis cohort, the MITT analysis set.
Secondary Effectiveness assessments of the MITT group were performed at 30 days, 6 months, 12 months, 24 months and 36 months as shown in Table 24 and Table 25.
Clinical success as defined by subjects showing a $\geq 1$ category improvement in RCC was $92.9\%$ (182/196), and the percentage of subjects with an improvement in RCC remained high through 36 months for those who have completed their 3-years visits. Limb ischemia, identified using the RCC, is reported as change from baseline. A negative value for change from baseline indicates an improvement in RCC.
At each time point in the study, the change from baseline showed RCC improved. At 30 days, 6 months, and 12 months, RCC improved by an average score of 2.7. At 24 months and 36 months, RCC improved by a score of 2.8 and 2.9.
Using Kaplan Meier methodology, the primary patency at 1 year was $72.3\%$ , and the secondary patency at 1 year was $81.8\%$ . Ankle-Brachial Index (ABI) change from baseline on the target limb averaged a 0.4 ( $\pm 0.25$ ) change at 30 days and a 0.4 ( $\pm 0.27$ ) change as 12 months. 6 Minute Walk Test (6MWT) change from baseline averaged a 46.8 ( $\pm 105.29$ ) change at 30 days and a 79.6 ( $\pm 144.71$ ) change at 12 months.
Table 24: Secondary Effectiveness Endpoints - MITT Population
| | 30 Days | 6 Months | 12 Months |
| --- | --- | --- | --- |
| Clinical Success | 92.9% (182/196) | 94.8% (182/192) | 97.2% (173/178) |
PMA P220021: FDA Summary of Safety and Effectiveness Data
{36}
Data presented as $\%$ (n/N) Kaplan Meier Estimates rate [95% CI] where $N$ is the number of subjects with available data.
Table 25: Secondary Effectiveness Baseline Comparison - MITT Population
| | | Baseline | 30 Days | 6 Months | 12 Months |
| --- | --- | --- | --- | --- | --- |
| Limb Ischemia: Rutherford Clinical Category (continuous) - MITT | | | | | |
| | All Available | 3.3 ± 0.53(200)3.0 (3,5) | 0.6 ± 1.04(196)0.0 (0, 5) | 0.5 ± 1.05(192)0.0 (0, 5) | 0.6 ± 0.96(178)0.0 (0, 5) |
| | Paired Data | Baseline | 3.2 ± 0.50(196)3.0 (3, 5) | 3.3 ± 0.53(192)3.0 (3, 5) | 3.3 ± 0.54(178)3.0 (3, 5) |
| | Paired Data | Follow-up | 0.6 ± 1.04(196)0.0(0, 5) | 0.5 ± 1.05(192)0.0(0, 5) | 0.6 ± 0.96(178)0.0(0, 5) |
| | Paired Data | Change from Baseline (Follow-up - Baseline) | -2.7 ± 1.12(196)-3.0 (-5, 1) | -2.7 ± 1.07(192)-3.0 (-5, -2) | -2.7 ± 1.05(178)0.0 (0, 5) |
| Ankle-Brachial Index (ABI) - MITT | | | | | |
| Target Limb | All Available | 0.61 ± 0.22(191)0.60(0.00, 1.55) | 0.99 ± 0.18(183)1.00(0.00, 1.71) | 0.99 ± 0.21(173)1.00(0.37, 1.98) | 0.95 ± 0.21(170)0.97(0.00, 1.97) |
| | Paired Data | Baseline | 0.61 ± 0.22(178)0.61(0.00, 1.55) | 0.61 ± 0.22(169)0.59(0.00, 1.55) | 0.60 ± 0.22(166)0.59(0.00, 1.55) |
| | Paired Data | Follow-up | 0.99 ± 0.18(178)1.00(0.00, 1.71) | 0.98 ± 0.21(169)1.00(0.37, 1.98) | 0.96 ± 0.21(166)0.97(0.00, 1.97) |
PMA P220021: FDA Summary of Safety and Effectiveness Data
{37}
| | | | Baseline | 30 Days | 6 Months | 12 Months |
| --- | --- | --- | --- | --- | --- | --- |
| | Paired Data | Change from Baseline (Follow-up - Baseline) | -- | 0.38 ± 0.25 (178) 0.38 (-0.66, 1.20) | 0.38 ± 0.24 (169) 0.38 (-0.49, 1.12) | 0.36 ± 0.27 (166) 0.37 (-0.74, 1.04) |
| Contralateral Limb | All Available | | 0.81 ± 0.26 (187) 0.81 (0.00, 1.63) | 0.87 ± 0.24 (181) 0.90 (0.21, 1.88) | 0.89 ± 0.24 (170) 0.93 (0.28, 1.97) | 0.88 ± 0.25 (163) 0.89 (0.02, 1.97) |
| | Paired Data | Baseline | -- | 0.82 ± 0.26 (172) 0.82 (0.00, 1.63) | 0.83 ± 0.26 (163) 0.82 (0.00, 1.63) | 0.82 ± 0.26 (157) 0.82 (0.00, 1.63) |
| | Paired Data | Follow-up | -- | 0.87 ± 0.24 (172) 0.90 (0.21, 1.88) | 0.89 ± 0.24 (163) 0.92 (0.28, 1.97) | 0.88 ± 0.25 (157) 0.89 (0.02, 1.97) |
| | Paired Data | Change from Baseline (Follow-up - Baseline) | -- | 0.05 ± 0.21 (172) 0.04 (-0.44, 1.16) | 0.06 ± 0.23 (163) 0.04 (-0.53, 1.25) | 0.06 ± 0.28 (157) 0.04 (-1.51, 1.25) |
| 6 Minute Walk Test (6MWT) - MITT | | | | | | |
| | All Available | | 208.8 ± 133.18 (142) 194.5 (18, 800) | 259.7 ± 147.84 (131) 229.8 (0, 864) | -- | 288.4 ± 141.51 (118) 291.5 (0, 1004) |
| | Paired Data | Baseline | -- | 213.4 ± 132.63 (130) 199.6 (23, 800) | -- | 209.8 ± 118.73 (117) 198.0 (23, 650) |
| | Paired Data | Follow-up | -- | 260.3 ± 148.26 (130) 230.7 (0, 864) | -- | 289.4 ± 141.76 (117) 292.0 (0, 1004) |
| |…
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.