Edwards SAPIEN 3 Transcatheter Pulmonary Valve System with Alterra Adaptive Prestent

P200015S011 · Edwards Lifesciences, LLC · NPV · Dec 16, 2021 · Cardiovascular

Device Facts

Record IDP200015S011
Device NameEdwards SAPIEN 3 Transcatheter Pulmonary Valve System with Alterra Adaptive Prestent
ApplicantEdwards Lifesciences, LLC
Product CodeNPV · Cardiovascular
Decision DateDec 16, 2021
DecisionAPPR
Device ClassClass f
AttributesTherapeutic, Pediatric

Indications for Use

The Edwards SAPIEN 3 Transcatheter Pulmonary Valve System with Alterra Adaptive Prestent is indicated for use in the management of pediatric and adult patients with severe pulmonary regurgitation as measured by echocardiography who have a native or surgically-repaired right ventricular outflow tract and are clinically indicated for pulmonary valve replacement.

Device Story

Transcatheter pulmonary valve replacement system; utilizes Alterra Adaptive Prestent as docking adaptor for 29-mm SAPIEN 3 THV. Input: patient RVOT anatomy. Operation: Alterra prestent (nitinol frame/PET fabric) deployed via delivery system to RVOT; SAPIEN 3 THV (cobalt-chromium frame/bovine pericardial tissue) deployed within prestent. Used in cath lab/hybrid suite by interventional cardiologists/surgeons. Output: restored pulmonary valve function; reduced regurgitation. Clinical benefit: improved hemodynamics; reduced RVEDV; potential delay of surgical reoperation. Healthcare providers use echocardiography/imaging to monitor valve function and RV remodeling.

Clinical Evidence

Prospective, single-arm, multicenter study (Alterra Clinical Study, N=85). Main cohort (N=60) evaluated THV dysfunction at 6 months; 0% rate (95% CI: 0.0-6.1%), meeting 25% performance goal. PDS registry cohort (N=25) achieved 88% acute success. Secondary outcomes: 96.7% technical success; significant reduction in RVEDV (295.9 to 220.4 ml) and pulmonary regurgitant fraction (47.4% to 4.1%) at 1 year. Adverse events included arrhythmia (33.3%) and bleeding (21.7%).

Technological Characteristics

System includes 29-mm SAPIEN 3 THV (cobalt-chromium MP35N frame, bovine pericardial tissue, PET skirts) and Alterra Adaptive Prestent (self-expanding nitinol frame, PET covering). Delivery systems include Commander and Pulmonic models. Sterilization: E-beam (Alterra) and Ethylene Oxide (Pulmonic Delivery System). Standards: ISO 5840-1/3, ISO 10993-1, ISO 11137, ISO 11135.

Indications for Use

Indicated for pediatric and adult patients with severe pulmonary regurgitation (measured by echocardiography) and dysfunctional native or surgically-repaired right ventricular outflow tract (RVOT) requiring pulmonary valve replacement. Contraindicated in patients unable to tolerate anticoagulation/antiplatelet therapy or with active infections/endocarditis.

Regulatory Classification

Identification

These devices are intended to replace a patient's pulmonary heart valve. They are different from the classified device (heart valves) in that they are placed percutaneously and do not require open chest surgery or a cardiotomy for placement.

Reference Devices

Submission Summary (Full Text)

{0} # SUMMARY OF SAFETY AND EFFECTIVENESS DATA ## I. GENERAL INFORMATION | Device Generic Name: | Pulmonary valve, prosthesis, percutaneously delivered | | --- | --- | | Device Trade Name: | Edwards SAPIEN 3 Transcatheter Pulmonary Valve System with Alterra Adaptive Prestent | | Device Procode: | NPV | | Applicant Name and Address: | Edwards Lifesciences LLC One Edwards Way Irvine, CA 92614 | | Date of Panel Recommendation: | None | | Premarket Approval Application (PMA) Number: | P200015/S011 | | Date of FDA Notice of Approval: | December 16, 2021 | The original PMA (P200015) of the Edwards SAPIEN 3 Transcatheter Pulmonary Valve System, which included the Edwards SAPIEN 3 Transcatheter Heart Valve System and Edwards Commander Delivery System was approved on August 31, 2020. The device was indicated for use in the management of pediatric and adult patients who have a clinical indication for intervention on a dysfunctional right ventricular outflow tract (RVOT) conduit or surgical bioprosthetic valve in the pulmonic position with ≥ moderate regurgitation and/or a mean RVOT gradient of ≥ 35 mmHg. The SSED to support this indication is available on the following FDA website and is incorporated by reference herein: https://www.accessdata.fda.gov/cdrh_docs/pdf20/P200015B.pdf The current supplement was submitted to introduce the Edwards Alterra Adaptive Prestent System and Edwards Pulmonic Delivery System and to expand the indications for use to include pediatric and adult patients with severe pulmonary regurgitation (PR) who have a dysfunctional native or surgically-repaired RVOT and are clinically indicated for pulmonary valve replacement. ## II. INDICATIONS FOR USE The Edwards SAPIEN 3 Transcatheter Pulmonary Valve System with Alterra Adaptive Prestent is indicated for use in the management of pediatric and adult patients with severe pulmonary regurgitation as measured by echocardiography who have a native or surgically- PMA P200015/S011: FDA Summary of Safety and Effectiveness Data Page 1 {1} repaired right ventricular outflow tract and are clinically indicated for pulmonary valve replacement. ### III. CONTRAINDICATIONS The Edwards SAPIEN 3 Transcatheter Pulmonary Valve System with Alterra Adaptive Prestent is contraindicated in patients who cannot tolerate an anticoagulation/antiplatelet regimen or who have active bacterial endocarditis or other active infections. ### IV. WARNINGS AND PRECAUTIONS The warnings and precautions can be found in the Edwards SAPIEN 3 Transcatheter Pulmonary Valve System with Alterra Adaptive Prestent labeling. ### V. DEVICE DESCRIPTION The Edwards SAPIEN 3 Transcatheter Pulmonary Valve System with Alterra Adaptive Prestent consists of the 29-mm Edwards SAPIEN 3 Transcatheter Heart Valve (THV), the Edwards Alterra Adaptive Prestent System, the Edwards Commander Delivery System, the Edwards Pulmonic Delivery System, the Edwards Crimper, and the Qualcrimp Crimping Accessory. During use, the Edwards Alterra Adaptive Prestent is first positioned within the patient's RVOT, serving as a docking adaptor for the Edwards SAPIEN 3 THV, which is subsequently deployed within the Edwards Alterra Adaptive Prestent. #### Edwards SAPIEN 3 THV The Edwards SAPIEN 3 THV, as shown in Figure 1, comprises a balloon-expandable, radiopaque, cobalt-chromium (MP35N) frame, a trileaflet bovine pericardial tissue valve, a polyethylene terephthalate (PET) internal fabric skirt, and a PET external sealing skirt for reduction of paravalvular leakage (PVL). The leaflets are treated according to the Edwards ThermaFix process. Although the Edwards SAPIEN 3 THV is available in multiple sizes, only the 29- mm valve is included in the scope of this PMA application. The Edwards SAPIEN 3 THV can be deployed using either the Edwards Commander Delivery System or the Edwards Pulmonic Delivery System. Figure 1: Edwards SAPIEN 3 Transcatheter Heart Valve ![img-0.jpeg](img-0.jpeg) #### Edwards Commander Delivery System The Edwards Commander Delivery System, as shown in Figure 2, includes a handle that PMA P200015/S011: FDA Summary of Safety and Effectiveness Data Page 2 {2} provides a flex wheel for articulation of the flex catheter, a tapered tip at the distal end, a balloon catheter for deployment of the THV, and radiopaque markers. A Loader component is used to aid insertion of the delivery system into the sheath. Figure 2: Edwards Commander Delivery System ![img-1.jpeg](img-1.jpeg) # Edwards Pulmonic Delivery System The Edwards Pulmonic Delivery System, as shown in Figure 3, consists of an inline sheath, a balloon catheter for deployment of the Edwards SAPIEN 3 THV, an outer shaft to cover the THV during insertion and tracking to the intended deployment location, and a tapered tip at the distal end to facilitate crossing of right heart structures. A 28F Dilator is used to predilate the vessel prior to insertion of the delivery system, if necessary. Figure 3: Edwards Pulmonic Delivery System ![img-2.jpeg](img-2.jpeg) # Qualcrimp Crimping Accessory The Qualcrimp Crimping Accessory, as shown in Figure 4, is a non-patient contacting device that is placed around the THV to protect the leaflets during the crimping process. It is manufactured of tubular polyester polyurethane foam and laminated cylindrically on both the inner and outer surfaces with a polyether urethane material. Figure 4: Qualcrimp Crimping Accessory # Edwards Crimper The Edwards Crimper, as shown in Figure 5, is composed of various molded plastic components and is used to compress the SAPIEN 3 THV to a controlled aperture. The aperture is created by rotating the handle until it abuts the Crimp Stopper, a component used to ensure the THV is crimped correctly. PMA P200015/S011: FDA Summary of Safety and Effectiveness Data Page 3 {3} **Figure 5: Edwards Crimper** ![img-3.jpeg](img-3.jpeg) ### **Edwards Alterra Adaptive Prestent System** The Edwards Alterra Adaptive Prestent System includes the Edwards Alterra Adaptive Prestent and Edwards Alterra Delivery System, as shown in Figure 6. The Edwards Alterra Adaptive Prestent is used as a docking adaptor for the 29-mm SAPIEN 3 THV. It comprises a self-expanding, radiopaque, nitinol frame assembly, and PET fabric covering. The Edwards Alterra Delivery System includes a handle which consists of a wheel that allows for deployment, two primary shafts with a flush port to flush the delivery system, and a long compliant tapered tip at the distal end to facilitate tracking through the vasculature. **Figure 6: Edwards Alterra Adaptive Prestent System** ![img-4.jpeg](img-4.jpeg) (a) Edwards Alterra Adaptive Prestent ![img-5.jpeg](img-5.jpeg) (b) Edwards Alterra Delivery System ## **VI. ALTERNATIVE PRACTICES AND PROCEDURES** There are several other alternatives for the correction of severe PR in patients with a native or surgically-repaired RVOT, including surgical placement of a right ventricle-pulmonary artery (RV-PA) conduit or a prosthetic valve and implantation of another transcatheter pulmonary valve. Each alternative has its own advantages and disadvantages. A patient should fully discuss these alternatives with his/her physician to select the method that best meets expectations and lifestyle. ## **VII. MARKETING HISTORY** The Edwards SAPIEN 3 Transcatheter Pulmonary Valve System with Alterra Adaptive Prestent has not been marketed in the United States or any foreign country. ## **VIII. POTENTIAL ADVERSE EFFECTS OF THE DEVICE ON HEALTH** Below is a list of the potential adverse effects (e.g., complications) associated with the use of PMA P200015/S011: FDA Summary of Safety and Effectiveness Data Page 4 {4} the device: - Death - Allergic reaction to anesthesia, contrast media, antithrombotic therapy, device materials - Anemia - Angina - Aortic root distortion - Atelectasis - Arrhythmia - Arteriovenous (AV) fistula - Blood loss requiring transfusion - Cardiovascular or vascular injury, such as perforation or damage (dissection) of vessels, myocardium or valvular structures including rupture of the RVOT that may require intervention - Cardiac arrest - Cardiac failure - Cardiogenic shock - Chest pain/discomfort - Conduction system injury - Coronary flow obstruction/transvalvular flow disturbance - Deep vein thrombosis - Device acute migration or malposition - Device dysfunction (regurgitation and/or stenosis) - Device embolization - Device thrombosis - Dislodgement of previously implanted devices (i.e., pacing lead) - Dyspnea - Electrolyte imbalance - Embolic event: air, calcific material, thrombus, device fragments - Embolic event: device fragments - Endocarditis - Exercise intolerance or weakness - Fever - Hematoma or ecchymosis - Hemolysis/hemolytic anemia - Hypertension or hypotension - Infection including incisional site infection, septicemia and endocarditis - Inflammation - Injury to tricuspid valve - Mechanical failure of delivery system, and/or accessories - Myocardial infarction - Pain - Pericardial effusion/cardiac tamponade PMA P200015/S011: FDA Summary of Safety and Effectiveness Data Page 5 {5} - Pleural effusion - Pneumothorax - Pulmonary edema - Radiation injury - Renal insufficiency or renal failure - Respiratory insufficiency or respiratory failure - Stroke/transient ischemic attack - Syncope - Systemic or peripheral ischemia - Systemic or peripheral nerve injury For the specific adverse events that occurred in the clinical study, please see Section X. # IX. SUMMARY OF PRECLINICAL STUDIES A summary of previously reported preclinical studies on the Edwards SAPIEN 3 THV System with the Edwards Commander Delivery System can be found in the SSED for PMA P140031 (https://www.accessdata.fda.gov/cdrh_docs/pdf14/P140031B.pdf) for the aortic valve replacement indication and in the SSED for PMA P200015 (https://fda.report/PMA/P200015/20/P200015B.pdf) for the pulmonic valve replacement indication. The following additional preclinical studies were performed to support the Edwards Alterra Adaptive Prestent System and Edwards Pulmonic Delivery System. # A. Laboratory Studies Nonclinical laboratory studies on the Edwards Alterra Adaptive Prestent System and Edwards Pulmonic Delivery System were performed in accordance with ISO 5840-1:2015, Cardiovascular implants – Cardiac valve prostheses – Part 1: General requirements and ISO 5840-3:2013, Cardiovascular implants – Cardiac valve prostheses – Part 3: Heart valve substitutes implanted by transcatheter techniques. # 1. Biocompatibility Biocompatibility assessments were completed on the Edwards Alterra Adaptive Prestent System and Edwards Pulmonic Delivery System in accordance with ISO 10993-1:2018, Biological Evaluation of Medical Devices Part 1: Evaluation and testing within a risk management process, and the FDA Guidance for Industry and Food and Drug Administration Staff, Use of International Standard ISO 10993-1, “Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process.” The required testing for each component was determined based on the nature and duration of body contact per ISO 10993-1:2018. Test articles consisted of the patient-contacting device components after exposure to all manufacturing processes, including sterilization. The biocompatibility test results for the Edwards Alterra Adaptive Prestent System and Edwards Pulmonic Delivery System are summarized in Table 1 and Table 2, respectively. PMA P200015/S011: FDA Summary of Safety and Effectiveness Data Page 6 {6} Table 1: Summary of Alterra Adaptive Prestent System Biocompatibility Assessments | Biological Effect per ISO 10993-1 | Test Method | Results | | --- | --- | --- | | Alterra Adaptive Prestent | | | | Cytotoxicity | Medium eluate method using human fibroblast cells | Pass | | Sensitization | Guinea pig maximization test | Pass | | Irritation/intracutaneous reactivity | Rabbit intracutaneous reactivity test | Pass | | Acute systemic toxicity | Mouse systemic injection test | Pass | | Pyrogenicity | Materials mediated rabbit pyrogenicity | Pass | | Genotoxicity | Ames test – plate incorporation | Pass | | | Chromosomal aberration assay | Pass | | Implantation | 4-week rabbit intramuscular implantation test | Pass | | | 90-day systemic toxicity in rabbits via intramuscular implantation | Pass | | Hemocompatibility | *In vitro* hemolysis (indirect) | Pass | | | *In vitro* hemolysis (direct method) | Pass | | | Complement activation | Pass | | | *In vivo* thrombogenicity | Pass | | Physicochemical | Chemical characterization and toxicological risk assessment | Pass | | Alterra Delivery System | | | | Cytotoxicity | Medium eluate method using human fibroblast cells | Pass | | Sensitization | Guinea pig maximization test | Pass | | Irritation/intracutaneous reactivity | Rabbit intracutaneous reactivity test | Pass | | Acute systemic toxicity | Mouse systemic injection test | Pass | | Pyrogenicity | Materials mediated rabbit pyrogenicity | Pass | | Hemocompatibility | *In vitro* hemolysis (indirect) | Pass | | | *In vitro* hemolysis (direct method) | Pass | | | Complement activation | Pass | PMA P200015/S011: FDA Summary of Safety and Effectiveness Data Page 7 {7} Table 2: Summary of Pulmonic Delivery System Biocompatibility Assessments | Biological Effect per ISO 10993-1 | Test Method | Results | | --- | --- | --- | | Cytotoxicity | Medium eluate method using L-929 mouse fibroblast cells | Pass | | Sensitization | Guinea pig maximization test | Pass | | Irritation/intracutaneous reactivity | Rabbit intracutaneous reactivity test | Pass | | Acute systemic toxicity | Mouse systemic injection test | Pass | | Pyrogenicity | Materials mediated rabbit pyrogenicity | Pass | | Hemocompatibility | In vitro hemolysis (indirect) | Pass | | | In vitro hemolysis (direct method) | Pass | | | Complement activation | Pass | | | Thrombogenicity (Yorkshire Cross pig model) | Pass | ## 2. Bench testing Summaries of the bench testing results for the Edwards Alterra Adaptive Prestent System and Edwards Pulmonic Delivery System are provided in Table 3 and Table 4, respectively. Table 3: Summary of Edwards Alterra Adaptive Prestent System Bench Testing | Test | Purpose | Results | | --- | --- | --- | | Alterra Adaptive Prestent | | | | Frame characterization | To characterize properties of the material and finished frame. | For characterization only | | Frame corrosion and nickel ion release | To assess the potential for pitting corrosion and for the leaching of nickel ions from the frame. | Pass | | Durability and migration | To assess the durability, fretting resistance, and migration resistance of the Alterra Adaptive Prestent and 29-mm SAPIEN 3 THV when deployed together. | Pass | | Marker visualization and skirt integrity | To assess the radiopaque markers sewn within the Alterra component. | Pass | | Finite element analysis | To investigate the strain distribution in the Alterra Adaptive Prestent. | Pass | | Radial pulsatile fatigue testing | To assess the fatigue resistance of the Alterra Adaptive Prestent under cyclic radial loading for up to 600 million cycles. | Pass | PMA P200015/S011: FDA Summary of Safety and Effectiveness Data Page 8 {8} | Test | Purpose | Results | | --- | --- | --- | | Magnetic resonance imaging (MRI) compatibility | To evaluate the MRI compatibility of the Alterra Adaptive Prestent. | Pass | | Alterra Delivery System | | | | Visual inspection | To verify the external surface of the effective length of the Alterra Delivery System is free from surface defects and extraneous matter. | Pass | | Dimensional inspection | To verify the working or effective dimensions of the Alterra Delivery System are met. | Pass | | Hemostasis | To ensure the Alterra Delivery System maintains hemostasis. | Pass | | Tensile testing | To ensure the prespecified tensile strengths of the Alterra Delivery System are met. | Pass | | Corrosion testing | To ensure the Alterra Delivery System is corrosion resistant per ISO 10555-1. | Pass | | Simulated use | To simulate the use of the Alterra Delivery System with the Alterra Adaptive Prestent, including loading, tracking, and deployment in a clinical setting. | Pass | | Edwards Alterra Adaptive Prestent System | | | | Design validation study | To confirm that the user needs are met for the Alterra Adaptive Prestent System. | Pass | **Table 4: Summary of Edwards Pulmonic Delivery System Bench Testing** | Test | Purpose | Results | | --- | --- | --- | | Visual inspection | To verify the Pulmonic Delivery System is free from surface defects and extraneous matter. | Pass | | Dimensional inspection | To ensure the working or effective dimensions of the Pulmonic Delivery System are met. | Pass | | Hemostasis | To ensure the Pulmonic Delivery System maintains hemostasis. | Pass | | Tensile testing | To ensure the prespecified tensile strengths of the Pulmonic Delivery System are met. | Pass | | Simulated use | To simulate the use of the Pulmonic Delivery System with the Edwards | Pass | PMA P200015/S011: FDA Summary of Safety and Effectiveness Data Page 9 {9} | Test | Purpose | Results | | --- | --- | --- | | | SAPIEN 3 THV, including loading, tracking, and deployment in a clinical setting. | | | Design validation study | To confirm that the user needs are met for the Edwards Pulmonic Delivery System. | Pass | ### **B. Animal Studies** The Edwards Alterra Adaptive Prestent System was evaluated in a Good Laboratory Practice (GLP)-compliant chronic study in a porcine model, as summarized in Table 5. The results of this study demonstrated that the Edwards Alterra Delivery System delivered and deployed the Edwards Alterra Adaptive Prestent as intended and the implant showed appropriate healing. An earlier chronic study evaluated the implantation of a SAPIEN 3 THV inside an Alterra Adaptive Prestent in a porcine model, which showed that the valves were well-seated and functional with no signs of migration, embolization, thrombus, or valve dysfunction after a minimum 90-day duration. **Table 5: Summary of Alterra Adaptive Prestent System Chronic Animal Study** | Description | Details | | --- | --- | | Animal model/sample size | Ten (10) pigs. | | Test Article | Edwards Alterra Adaptive Prestent System. | | Technique | Transcatheter implantation of the Edwards Alterra Adaptive Prestent through the femoral vein to the pulmonary position. Native pulmonic leaflets were blocked open and valve function was completely disabled. | | Results | Four (4) early deaths occurred prior to the minimum 90-day implant duration due to complications unrelated to the device. Six (6) animals survived to a minimum of 90 days. | | Conclusion | All devices were delivered and implanted as intended. No clinically significant injuries to the right heart structures including the tricuspid valve and chordae were observed. A single device apex embolized and was found embedded in a branch of the pulmonary artery at explant, although no clinically significant injury to this site was discovered. | ### **C. Sterilization** The Edwards Alterra Adaptive Prestent System is sterilized using electron beam (e-beam) radiation in accordance with ISO 11137-1:2006, *Sterilization of health care products - Radiation - Part 1: Requirements for development, validation and routine control of a sterilization process for medical devices*, and ISO 11137-2:2013, *Sterilization of health* PMA P200015/S011: FDA Summary of Safety and Effectiveness Data Page 10 {10} care products - Radiation - Part 2: Establishing the sterilization dose, which demonstrated a sterility assurance level (SAL) of $10^{-6}$ under a minimum dose of 25 kGy. The Edwards SAPIEN 3 Pulmonic Delivery System is sterilized using ethylene oxide (EO) gas per ISO 11135:2014, *Sterilization of health-care products - Ethylene oxide - Requirements for the development, validation and routine control of a sterilization process for medical devices*, ISO 11737-1:2018, *Sterilization of health care products - Microbiological methods - Part 1: Determination of a population of microorganisms on products*, and ISO 10993-7:2008, *Biological evaluation of medical devices - Part 7: Ethylene oxide sterilization residuals*, which demonstrated an SAL of $10^{-6}$ under 100% EO sterilization. #### **D. Packaging and Shelf Life** The Edwards Alterra Adaptive Prestent System components are packaged together. The Edwards Alterra Adaptive Prestent is supplied preloaded within the Edwards Alterra Delivery System, which is secured in a tray/retainer. The secured device is sealed within a Tyvek/Nylon pouch and placed in a solid bleach sulfate (SBS) shelf box. The shelf box is housed in a corrugated shipper for transport. The shelf life of all components of the Edwards Alterra Adaptive Prestent System is 2 years as demonstrated by packaging integrity and product functional testing on aged samples. The Edwards Pulmonic Delivery System is secured in a polyethylene terephthalate glycol (PETG) tray, along with a Dilator, a 2-piece Crimp Stopper, and a Qualcrimp Crimping Accessory. A PETG retainer is then placed over the tray, and the tray is placed within a Tyvek/Nylon sealed pouch. The pouched system is placed in an SBS shelf carton, which is then placed in a corrugated shipper box for transport. The shelf life of all components of the Edwards Pulmonic Delivery System is 1 year as demonstrated by packaging integrity and product functional testing on aged samples. ### **X. SUMMARY OF CLINICAL STUDY** The applicant performed a clinical study to establish a reasonable assurance of the safety and effectiveness of transcatheter pulmonary valve replacement (TPVR) with the Edwards SAPIEN 3 Transcatheter Pulmonary Valve System with Alterra Adaptive Prestent under IDE G170053 (entitled the 'Alterra Clinical Study'). The data from this study were the basis for the PMA approval decision. A summary of the clinical study is presented below. #### **A. Study Design** The Alterra Clinical Study was a prospective, single-arm, multicenter study. It consisted of two cohorts, namely, the main cohort and the pulmonic delivery system (PDS) registry cohort. The main cohort comprised patients implanted with the Edwards Commander Delivery System, the enrollment of which took place between August 2017 and September 2019 at 11 investigational sites in the U.S. The main purpose of the PDS registry cohort was to evaluate the use of the Edwards Pulmonic Delivery System, which was developed later in the clinical study. The enrollment of the PDS registry cohort occurred between January 2020 PMA P200015/S011: FDA Summary of Safety and Effectiveness Data Page 11 {11} and August 2020 at 13 investigational sites. The database for this PMA application reflected data collected through June 2, 2021. The Alterra Clinical Study used an independent Data Safety Monitoring Board (DSMB) that was instructed to notify the applicant of any safety or compliance issues and a Clinical Events Committee (CEC) that was responsible for adjudicating endpoint-related events reported during the study. An independent echocardiographic core laboratory was used for standardized assessment of echocardiograms. 1. Clinical Inclusion and Exclusion Criteria Enrollment in the Alterra Clinical Study was limited to patients who met the following inclusion criteria: - The candidate/candidate's legal guardian has been informed of the nature of the study, agrees to its provisions and has provided written informed consent. - Weight is ≥ 20 kg (44 lbs). - RVOT/pulmonary valve (PV) with moderate or greater PR by transthoracic echocardiogram (TTE). - RVOT/PV proximal and distal landing zone diameter ≥ 27 mm and ≤ 38 mm, and minimum of 35 mm from contractile tissue to lowest pulmonary artery takeoff. Patients were not permitted to enroll in the study if they met any of the following exclusion criteria: - Active infection requiring current antibiotic therapy (if temporary illness, patient may be a candidate 2 weeks after discontinuation of antibiotics). - History of or active endocarditis (active treatment with antibiotics) within the past 180 days. - Leukopenia (white blood cell < 2000 cells/μL), anemia (hemoglobin (Hgb) < 7 g/dL), thrombocytopenia (platelets < 50,000 cells/μL) or any known blood clotting disorder. - Inappropriate anatomy for introduction and delivery of Alterra or the SAPIEN 3. - Need for concomitant atrial septal defect or ventricular septal defect closure or other concomitant interventional procedures other than pulmonary artery or branch pulmonary artery stenting or angioplasty. - Interventional/surgical procedures within 30 days prior to the Alterra or valve implant procedure. - Any planned surgical, percutaneous coronary or peripheral procedure to be performed within the 30-day follow-up from the Alterra or valve implant procedure. - History of or current intravenous drug use. - Major or progressive non-cardiac disease resulting in a life expectancy of less than one year. - Known hypersensitivity to aspirin or heparin and cannot be treated with other antiplatelet and/or antithrombotic medications. PMA P200015/S011: FDA Summary of Safety and Effectiveness Data Page 12 {12} - Known hypersensitivity to nitinol, cobalt-chromium, nickel or contrast media that cannot be adequately pre-medicated. - Currently participating in an investigational drug or another device study. [Note: Trials requiring extended follow-up for products that were investigational, but have since become commercially available, are not considered investigational devices.] - Positive urine or serum pregnancy test in female patients of child-bearing potential. - Renal insufficiency (creatinine > 3.0 mg/dL) and/or renal replacement therapy. ## 2. Follow-up Schedule Follow-up time points included discharge, 30 days, 6 months, and annually through 5 years post procedure. Preoperative and post-operative assessments included physical assessment and medical history, laboratory measurements, imaging tests, and health surveys. Adverse events and complications were recorded at all visits. ## 3. Clinical Endpoints The primary endpoint for the main cohort was THV dysfunction at 6 months defined as a non-hierarchical composite of: - RVOT/PV reintervention - Moderate or greater total PR via TTE - Mean RVOT gradient ≥ 35 mmHg via TTE A performance goal of 25% was pre-specified for the primary endpoint of the main cohort. The hypothesis for the primary endpoint was as follows: $$H_0: \pi_{\text{Alterra}} \ge 25\%$$ $$H_A: \pi_{\text{Alterra}} < 25\%$$ where $\pi_{\text{Alterra}}$ represented the composite event rate at 6 months. If the upper limit of the two-sided 95% confidence interval for the composite event was less than 25%, the performance goal would be met. The hypothesis was tested at a one-sided significance level of 0.025. The primary endpoint for the PDS registry cohort was acute PDS success, defined as a non-hierarchical composite of: - Single THV implanted in the desired location. - Right ventricular (RV)-pulmonary artery (PA) peak-to-peak gradient < 35 mmHg post-THV implantation. - Less than moderate total PR by discharge TTE (or earliest evaluable TTE). - Free of SAPIEN 3/Alterra explant at 24 hours post-implantation. PMA P200015/S011: FDA Summary of Safety and Effectiveness Data Page 13 {13} The primary endpoint for the PDS registry cohort was analyzed descriptively. Other outcome measures for the main cohort included: - Technical success at exit from the procedure room, defined as a composite of: o Alive, and o Successful access, delivery, and retrieval of the Alterra Delivery System, and o Successful access, delivery, and retrieval of the SAPIEN 3 delivery system, and o Single Alterra deployed in the desired location, and o Single THV implanted in the desired location within the Alterra (staged procedures allowed), and o No need for additional unplanned or emergency surgery or intervention related to the device or access procedure. - Device success at 30 days and 6 months, defined as a composite of: o Alive, and o Original intended device(s) in place, and o No additional surgical or interventional procedures related to access or the device since completion of the original procedure (i.e., exit from the procedure room), and o Intended performance of the device: ▪ Structural performance: no migration, embolization, fracture, detachment, hemolysis, thrombosis (including reduced leaflet mobility if detected) or endocarditis, and ▪ Hemodynamic performance: net pulmonary valve peak gradient < 35 mmHg and transvalvular PR ≤ mild, and ▪ Absence of para-device complications (paravalvular regurgitation > mild, need for a permanent pacemaker, erosion, new pulmonary embolism). - Procedure success at 30 days, defined as a composite of: o Device success, and o No device or procedure related serious adverse events (SAEs) from the following list: ▪ Life threatening bleed; major vascular or access site complications requiring unplanned reintervention or surgery; stage 2 or 3 acute kidney injury (AKI) (including new dialysis); severe heart failure or hypotension requiring intravenous inotrope, ultrafiltration or mechanical circulatory support; prolonged intubation > 48 hours. - THV hemodynamic function. - New York Heart Association (NYHA) Functional Class. - Characterization of right ventricular remodeling. ### B. Accountability of the PMA Cohorts At the time of database lock, a total of 85 patients were enrolled in the study, including 60 in PMA P200015/S011: FDA Summary of Safety and Effectiveness Data Page 14 {14} the main cohort and 25 in the PDS registry cohort. There were 3 different analysis populations defined in the protocol: All Treated (AT), Attempted Implant (AI), and Valve Implant (VI), as summarized in Table 6. Table 6: Analysis Populations | Analysis Population | Definition | Number of Patients | | | --- | --- | --- | --- | | | | Main Cohort | PDS Registry | | All Treated (AT) | All patients who signed informed consent, passed screening and for whom the procedure was begun (defined as the time of vascular access – incision or puncture). | 60 | 25 | | Attempted Implant (AI) | All AT patients who had an attempted implant (the introducer sheath for vascular delivery of the Alterra Adaptive Prestent was inserted). | 60 | 25 | | Valve Implant (VI) | All AI patients who received and retained a SAPIEN 3 THV upon leaving the catheterization laboratory/hybrid suite. | 60 | 25 | Study visit compliance is summarized in Table 7. One patient in the main cohort had not completed the 6-month visit at the time of the database lock. Table 7: Study Visit Compliance | | Main Cohort (N = 60) | | PDS Registry (N = 25) | | | --- | --- | --- | --- | --- | | | 30 Days | 6 Months | 30 Days | 6 Months | | Ineligible* | 0 | 0 | 1 | 1 | | Eligible | 60 | 60 | 24 | 24 | | Visit completed | 60 (100.0%) | 59 (98.3%) | 24 (100%) | 23 (95.8%) | *Ineligible patients included those who had died, withdrawn, lost to follow-up, or not reached the visit window. ### C. Study Population Demographics and Baseline Characteristics The demographics and baseline characteristics of the study population are typical for a transcatheter pulmonary valve study performed in the U.S., as shown in Table 8. PMA P200015/S011: FDA Summary of Safety and Effectiveness Data Page 15 {15} **Table 8: Demographics and Baseline Characteristics (AT Population)** | Demographics and Baseline Characteristics | Summary Statistics* | | | --- | --- | --- | | | Main Cohort (N = 60) | PDS Registry (N = 25) | | Age - years | 29.5 ± 2.14 (60) | 29.1 ± 3.45 (25) | | <12 years (child) | 1.7% (1/60) | 0.0% (0/25) | | 12-21 years (adolescent) | 48.3% (29/60) | 44.0% (11/25) | | >21 years (adult) | 50.0% (30/60) | 56.0% (14/25) | | Male sex | 56.7% (34/60) | 56.0% (14/25) | | Weight (kg) | 73.1 ± 3.05 (60) | 73.7 ± 4.44 (25) | | New York Heart Association (NYHA) Class | | | | Class I | 53.3% (32/60) | 37.5% (9/24) | | Class II | 38.3% (23/60) | 45.8% (11/24) | | Class III | 8.3% (5/60) | 16.7% (4/24) | | Class IV | 0.0% (0/60) | 0.0% (0/24) | | NYHA Class grouped | | | | Class I/II | 91.7% (55/60) | 83.3% (20/24) | | Class III/IV | 8.3% (5/60) | 16.7% (4/24) | | Primary diagnosis | | | | Pulmonary atresia | 5.0% (3/60) | 4.0% (1/25) | | Pulmonary valve stenosis | 23.3% (14/60) | 20.0% (5/25) | | Tetralogy of Fallot | 70.0% (42/60) | 68.0% (17/25) | | Other | 1.7% (1/59) | 8.0% (2/25) | | Secondary diagnosis | | | | Atrial septal defect | 18.3% (11/60) | 12.0% (3/25) | | Coarctation of the aorta | 3.3% (2/60) | 4.0% (1/25) | | Ventricular septal defect | 30.0% (18/60) | 12.0% (3/25) | | Other | 21.7% (13/60) | 20.0% (5/25) | | Most recent RVOT/PV intervention or surgery | | | | Pericardial/transannular patch | 65.0% (39/60) | 60.0% (15/25) | | Valvuloplasty alone | 25.0% (15/60) | 20.0% (5/25) | | Other^{†} | 10.0% (6/60) | 20.0% (5/25) | *Continuous measures - mean ± SD (total no.); categorical measures - % (no./total no.) †Tetralogy of Fallot repairs, unspecified RVOT/PV procedures, or none. The breakdowns of prior cardiac interventions by patient age in the AT population are shown in Table 9 for the main cohort and Table 10 for the PDS registry cohort. PMA P200015/S011: FDA Summary of Safety and Effectiveness Data Page 16 {16} **Table 9: Prior Cardiac Interventions by Age Group (AT Population) - Main Cohort** | Most Recent RVOT/PV Intervention or Surgery | Summary Statistics* | | | | --- | --- | --- | --- | | | <12 Years (N = 1) | 12-21 Years (N = 29) | ≥22 Years (N = 30) | | Pericardial/transannular patch | 100.0% (1/1) | 82.8% (24/29) | 46.7% (14/30) | | Valvuloplasty alone | 0.0% (0/1) | 17.2% (5/29) | 33.3% (10/30) | | Other | 0.0% (0/1) | 0.0% (0/29) | 20.0% (6/30) | | Tetralogy of Fallot repairs | 0.0% (0/0) | 0.0% (0/0) | 66.7% (4/6) | | Unspecified RVOT/PV procedure | 0.0% (0/0) | 0.0% (0/0) | 16.7% (1/6) | | None | 0.0% (0/0) | 0.0% (0/0) | 16.7% (1/6) | *Event rate (no./total no.) **Table 10: Prior Cardiac Interventions by Age Group (AT Population) - PDS Registry** | Most Recent RVOT/PV Intervention or Surgery | Summary Statistics* | | | --- | --- | --- | | | 12-21 Years (N = 11) | ≥ 22 Years (N = 14) | | Pericardial/transannular patch | 72.7% (8/11) | 50.0% (7/14) | | Valvuloplasty alone | 18.2% (2/11) | 21.4% (3/14) | | Other | 9.1% (1/11) | 28.6% (4/14) | | Tetralogy of Fallot repairs | 100.0% (1/1) | 25.0% (1/4) | | Unspecified RVOT/PV procedure | 0.0% (0/1) | 50.0% (2/4) | | None | 0.0% (0/1) | 25.0% (1/4) | *Event rate (no./total no.) #### **D. Safety and Effectiveness Results** ##### **1. Primary Endpoint** The primary endpoint result for the main cohort is presented in Table 11. The rate of THV dysfunction at 6 months was 0% (two-sided 95% confidence interval: 0.0% to 6.1%). Since the upper limit of the two sided 95% confidence interval for the primary endpoint event rate was < 25%, the endpoint was met. PMA P200015/S011: FDA Summary of Safety and Effectiveness Data Page 17 {17} **Table 11: Primary Endpoint Result - Main Cohort** | Endpoint | Summary Statistics* (N = 60) | 95% Confidence Interval† | Less Than Prespecified Performance Goal (25%)? | | --- | --- | --- | --- | | THV dysfunction‡ | 0.0% (0/59) | [0.0%, 6.1%] | Yes | | RVOT/PV reintervention | 0.0% (0/60) | | | | Moderate or greater pulmonary regurgitation | 0.0% (0/59) | | | | Mean RVOT gradient ≥35 mmHg | 0.0% (0/59) | | | *Event rate (no./total no.) †One patient did not have evaluable echocardiogram data. ‡Two-sided 95% confidence interval. The primary endpoint result for the PDS registry cohort is presented in Table 12. Acute PDS success was achieved in 88.0% of the patients. **Table 12: Primary Endpoint Result - PDS Registry** | Endpoint | Summary Statistics* (N = 25) | | --- | --- | | Acute PDS success | 88.0% (22/25) | | Single THV implanted in the desired location | 100.0% (25/25) | | RV-PA peak-to-peak gradient < 35 mmHg post implantation | 100.0% (25/25) | | Less than moderate PR by discharge TTE (or earliest evaluable TTE) | 88.0% (22/25) | | Free of explant at 24 hours post implantation | 100.0% (25/25) | *Event rate (no./total no.) ## 2. Additional Outcome Measures ### Technical Success Technical success at exit from the procedure room was achieved in 96.7% of the patients in the main cohort, as summarized in Table 13. **Table 13: Technical Success Result - Main Cohort** | Endpoint | Summary Statistics* (N = 60) | | --- | --- | | Technical Success | 96.7% (58/60) | | Alive | 100.0% (60/60) | | Successful access delivery and retrieval of the Alterra delivery system | 100.0% (60/60) | PMA P200015/S011: FDA Summary of Safety and Effectiveness Data Page 18 {18} | Endpoint | Summary Statistics* (N = 60) | | --- | --- | | Successful access delivery and retrieval of the SAPIEN 3 delivery system | 100.0% (60/60) | | Single Alterra deployed in the desired location | 100.0% (60/60) | | Single SAPIEN 3 valve deployed to the desired location | 96.7% (58/60) | | No need for unplanned or emergency surgery or intervention related to the device or access procedure | 100.0% (60/60) | *Event rate (no./total no.) ### Device Success The device success rate in the main cohort was 88.3% at 30 days and 76.3% at 6 months, as summarized in Table 14. **Table 14: Device Success Result - Main Cohort** | Endpoint | Summary Statistics* | | | --- | --- | --- | | | 30 Days (N = 60) | 6 Months (N = 59) | | Device Success | 88.3% (53/60) | 76.3% (45/59) | | Alive | 100.0% (60/60) | 100% (59/59) | | No device explant | 100.0% (60/60) | 100% (59/59) | | No additional surgical or interventional procedures related to access or the device since completion of the original procedure | 100.0% (60/60) | 100% (59/59) | | Intended performance of the device† | 88.3% (53/60) | 76.3% (45/59) | | Structural performance‡ | 90.0% (54/60) | 76.3% (45/59) | | Hemodynamic performance | 100.0% (60/60) | 100% (59/59) | | Absence of para-device complications | 98.3% (59/60) | 100% (59/59) | *Event rate (no./total no.) †Intended performance of the device is defined as a composite of: (1) structural performance - no migration, embolization, fracture, thrombosis (including reduced leaflet mobility if detected) or endocarditis; (2) hemodynamic performance - net pulmonary valve peak gradient ≤ 35mmHg and transvalvular PR ≤ mild; and (3) Absence of para-device complications – pulmonary valve regurgitation > mild, need for a permanent pacemaker, or new pulmonary embolism. ‡The structural performance events included 4 Alterra Prestent fracture events, 1 SAPIEN 3 THV thrombosis event, 1 Alterra Prestent thrombosis event, and 1 Alterra Prestent migration event at 30 days and an additional 8 Alterra Prestent fracture events at 6 months. Multiple events occurred on one patient are counted as one. PMA P200015/S011: FDA Summary of Safety and Effectiveness Data Page 19 {19} ### Procedural Success Procedural success was met in 88.3% of patients in the main cohort at 30 days, as summarized in Table 15. **Table 15: Procedural Success Result - Main Cohort** | Endpoint | Summary Statistics* (N = 60) | | --- | --- | | Procedure Success | 88.3% (53/60) | | Device success | 88.3% (53/60) | | No device or procedure related SAEs^{†} | 100.0% (60/60) | *Event rate (no./total no.) $^{†}$SAEs: life threatening bleed; major vascular or access site complications requiring unplanned reintervention or surgery; stage 2 or 3 acute kidney injury (including new dialysis); severe heart failure or hypotension requiring intravenous inotrope, ultrafiltration or mechanical circulatory support; prolonged intubation > 48 hours. ### THV Hemodynamic Function The mean RVOT gradient, peak RVOT gradient, total PR, and paravalvular regurgitation results through 1 year for the main cohort are shown in Figure 7 through Figure 10, respectively. The mean and peak RVOT gradients remained largely stable through 1 year post implant (9.7 mmHg and 17.9 mmHg, respectively, at 1 year). The proportion of patients with total PR ≥ moderate decreased from 100% to 9.6% at 1 year. The proportion of patients with paravalvular regurgitation ≥ moderate was 1.7% at 30 days and 0% at both 6 months and 1 year. **Figure 7: Mean RVOT Gradient (VI Population) - Main Cohort** ![img-6.jpeg](img-6.jpeg) **Note:** Line plot with mean and standard error. The total number of patients at each visit time point only counted the patients with evaluable values. PMA P200015/S011: FDA Summary of Safety and Effectiveness Data Page 20 {20} Figure 8: Peak RVOT Gradient (VI Population) - Main Cohort ![img-7.jpeg](img-7.jpeg) Note: Line plot with mean and standard error. The total number of patients at each visit time point only counted the patients with evaluable values. Figure 9: Total Pulmonary Regurgitation (VI Population) - Main Cohort ![img-8.jpeg](img-8.jpeg) Note: The total number of patients at each visit time point only counted the patients with evaluable values. PMA P200015/S011: FDA Summary of Safety and Effectiveness Data Page 21 {21} **Figure 10: Paravalvular Regurgitation (VI Population) - Main Cohort** ![img-9.jpeg](img-9.jpeg) **Note:** The total number of patients at each visit time point only counted the patients with evaluable values. #### NYHA Functional Class The NYHA classifications by visit are presented for the main cohort in Figure 11. At baseline, 91.7% of patients in the main cohort were in NYHA Class I/II, which increased to 100% at 30 days and 6 months and 98.1% at 1 year. **Figure 11: NYHA Class by Visit (VI Population) - Main Cohort** ![img-10.jpeg](img-10.jpeg) PMA P200015/S011: FDA Summary of Safety and Effectiveness Data Page 22 {22} # Characterization of Right Ventricular Remodeling Right ventricular remodeling was characterized via magnetic resonance imaging (MRI) and computed tomography (CT) in the main cohort. Right ventricular end diastolic volume (RVEDV), RVEDV index, and main pulmonary artery regurgitant fraction at baseline and 1 year are presented in Figure 12 through Figure 14, respectively. The RVEDV decreased from 295.9 ml to 220.4 ml, with the corresponding RVEDV index decreasing from 162.7 ml/m² to 118.9 ml/m². The main pulmonary artery regurgitant fraction decreased from 47.4% to 4.1%. Figure 12: Right Ventricular End Diastolic Volume (VI Population) - Main Cohort ![img-11.jpeg](img-11.jpeg) Note: The total number of patients at each visit time point only counted the patients with evaluable paired values. PMA P200015/S011: FDA Summary of Safety and Effectiveness Data Page 23 {23} Figure 13: Right Ventricular End Diastolic Volume Index (VI Population) - Main Cohort ![img-12.jpeg](img-12.jpeg) Note: The total number of patients at each visit time point only counted the patients with evaluable paired values. Figure 14: Main Pulmonary Artery Regurgitant Fraction (VI Population) - Main Cohort ![img-13.jpeg](img-13.jpeg) Note: The total number of patients at each visit time point only counted the patients with evaluable paired values. PMA P200015/S011: FDA Summary of Safety and Effectiveness Data Page 24 {24} ### 3. Adverse Events Kaplan-Meier estimates of the CEC-adjudicated adverse events through 1 year for the main cohort and through 6 months for the PDS registry cohort are presented in Table 16 and Table 17, respectively. **Table 16: CEC-Adjudicated Adverse Events Through 1 Year (AT Population) | Event | Summary Statistics* | | | | --- | --- | --- | --- | | | 30 Days (N = 60) | 6 Months (N = 59) | 1 Year (N = 49) | | All-cause death | 0.0% (0, 0) | 0.0% (0, 0) | 0.0% (0, 0) | | Reintervention | 0.0% (0, 0) | 0.0% (0, 0) | 0.0% (0, 0) | | Arrhythmia | 33.3% (21, 20) | 33.3% (21, 20) | 33.3% (21, 20) | | Permanent pacemaker | 0.0% (0, 0) | 0.0% (0, 0) | 0.0% (0, 0) | | Acute kidney injury | 0.0% (0, 0) | 0.0% (0, 0) | 0.0% (0, 0) | | Bleeding | 18.3% (12, 11) | 21.7% (14, 13) | 23.5% (16, 14) | | Life threatening or disabling | 0.0% (0, 0) | 1.7% (1, 1) | 3.5% (2, 2) | | Major | 1.7% (1, 1) | 1.7% (1, 1) | 1.7% (1, 1) | | Minor | 16.7% (11, 10) | 18.4% (12, 11) | 20.1% (13, 12) | | Coronary artery compression | 0.0% (0, 0) | 0.0% (0, 0) | 0.0% (0, 0) | | Endocarditis | 0.0% (0, 0) | 0.0% (0, 0) | 0.0% (0, 0) | | Myocardial infarction | 0.0% (0, 0) | 0.0% (0, 0) | 0.0% (0, 0) | | Pulmonary embolism | 0.0% (0, 0) | 0.0% (0, 0) | 0.0% (0, 0) | | Stroke | 0.0% (0, 0) | 0.0% (0, 0) | 0.0% (0, 0) | | Transient ischemic attack | 0.0% (0, 0) | 0.0% (0, 0) | 0.0% (0, 0) | | Vascular injury or access site complication | 5.0% (3, 3) | 5.0% (3, 3) | 6.8% (4, 4) | | Major | 0.0% (0, 0) | 0.0% (0, 0) | 1.8% (1, 1) | | Minor | 5.0% (3, 3) | 5.0% (3, 3) | 5.0% (3, 3) | *Kaplan-Meier estimate (no. events, no. patients with event) **Table 17: CEC-Adjudicated Adverse Events Through 6 Months (AT Population) | Event | Summary Statistics* | | | --- | --- | --- | | | 30 Days (N = 24) | 6 Months (N = 24) | | All-cause death | 4.0% (1, 1) | 4.0% (1, 1) | | Cardiovascular | 0.0% (0, 0) | 0.0% (0, 0) | | Non-Cardiovascular | 4.0% (1, 1) | 4.0% (1, 1) | | Reintervention | 0.0% (0, 0) | 0.0% (0, 0) | | Arrhythmia | 48.0% (15, 12) | 48.0% (15, 12) | PMA P200015/S011: FDA Summary of Safety and Effectiveness Data Page 25 {25} | Event | Summary Statistics* | | | --- | --- | --- | | | 30 Days (N = 24) | 6 Months (N = 24) | | Permanent pacemaker | 0.0% (0, 0) | 0.0% (0, 0) | | Acute kidney injury | 0.0% (0, 0) | 0.0% (0, 0) | | Bleeding | 20.0% (6, 5) | 20.0% (6, 5) | | Life threatening or disabling | 0.0% (0, 0) | 0.0% (0, 0) | | Major | 0.0% (0, 0) | 0.0% (0, 0) | | Minor | 20.0% (6, 5) | 20.0% (6, 5) | | Coronary artery compression | 0.0% (0, 0) | 0.0% (0, 0) | | Endocarditis | 0.0% (0, 0) | 0.0% (0, 0) | | Myocardial infarction | 0.0% (0, 0) | 0.0% (0, 0) | | Pulmonary embolism | 0.0% (0, 0) | 0.0% (0, 0) | | Stroke | 0.0% (0, 0) | 0.0% (0, 0) | | Transient ischemic attack | 0.0% (0, 0) | 0.0% (0, 0) | | Vascular injury or access site complication | 8.0% (4, 2) | 8.0% (4, 2) | | Major | 0.0% (0, 0) | 0.0% (0, 0) | | Minor | 8.0% (4, 2) | 8.0% (4, 2) | *Kaplan-Meier estimate (no. events, no. patients with event) #### 4. Other Study Observations ##### Procedural Information Procedural data for the main cohort and PDS registry cohort are summarized in Table 18. Concomitant Alterra Adaptive Prestent and SAPIEN 3 THV procedures were performed on the majority of patients (98.3% for the main cohort and 100% for PDS registry cohort). **Table 18: Procedural Data (AT Population)** | Variable | Summary Statistics* | | | --- | --- | --- | | | Main Cohort (N = 60) | PDS Registry (N = 25) | | Procedure type | | | | Staged procedure | 1.7% (1/60) | 0.0% (0/25) | | Prestent procedure sheath time (min) | 61.0 | N/A | | SAPIEN 3 procedure sheath time (min) | 23.0 | N/A | | Concomitant Alterra Adaptive Prestent + SAPIEN 3 THV procedure | 98.3% (59/60) | 100.0% (25/25) | | Procedure sheath time (min) | 66.8 ± 4.73 (59) | 58.2 ± 8.42 (25) | | Alterra Adaptive Prestent procedure complications | 3.3% (2/60) | 0.0% (0/25) | PMA P200015/S011: FDA Summary of Safety and Effectiveness Data Page 26 {26} | Variable | Summary Statistics* | | | --- | --- | --- | | | Main Cohort (N = 60) | PDS Registry (N = 25) | | SAPIEN 3 procedure complications | 10.0% (6/60) | 8.0% (2/25) | | Total fluoroscopy time | 37.2 ± 3.04 (60) | 38.0 ± 4.88 (25) | | Alterra Adaptive Prestent procedure post-dilatation performed | 0.0% (0/60) | 0.0% (0/25) | | SAPIEN 3 procedure post-dilatation performed | 11.7% (7/60) | 0.0% (0/25) | | Alterra Prestent implanted in the intended location at the time the patient left the procedure room | 100.0% (60/60) | 100.0% (25/25) | | SAPIEN 3 THV implanted in the intended location at the time the patient left the procedure room | 100.0% (60/60) | 100.0% (25/25) | | Single SAPIEN 3 THV implanted | 96.7% (58/60) | 100.0% (25/25) | *Continuous measures - Mean ± SD (total no.); categorical measures - % (no./total no.) #### CT Core Laboratory Findings The results of an imaging assessment by the CT Core Laboratory through 1 year for the main cohort and through 30 days for the PDS registry cohort are presented in Table 19 and Table 20, respectively. There were no device reinterventions associated with these findings. **Table 19: CT Core Laboratory Findings (VI Population) – Main Cohort** | Finding | Summary Statistics* | | | --- | --- | --- | | | 30 Days (N = 60) | 1 Year (N = 55) | | Alterra Adaptive Prestent penetration† | | | | Minor penetration - into adjacent vasculature or cardiac structure without extravasation, pseudoaneurysm, or erosion | 2.6% (1/39) | 5.2% (1/19) | | Major penetration - protrusion into surrounding tissue with blood pooling, or erosion, or pseudoaneurysm | 0.0% (0/39) | 0.0% (0/19) | | Pedunculated mobile mass† | 2.6% (1/39) | 0.0% (0/19) | | Alterra Adaptive Prestent fracture‡ | | | | Fracture not requiring intervention | 6.7% (4/60) | 21.8% (12/55) | | Fracture requiring intervention | 0.0% (0/60) | 0.0% (0/55) | | SAPIEN 3 valve frame fracture‡ | 0.0% (0/60) | 0.0% (0/55) | *Rate (no./total no.) †Only included patients with evaluable images at visit time point. ‡Fracture was determined by either chest x-ray or fluoroscopy. PMA P200015/S011: FDA Summary of Safety and Effectiveness Data Page 27 {27} **Table 20: CT Core Laboratory Findings (VI Population) – PDS Registry** | Finding | Summary Statistics* | | --- | --- | | | 30 Days (N = 25) | | Alterra Adaptive Prestent penetration† | | | Minor penetration - into adjacent vasculature or cardiac structure without extravasation, pseudoaneurysm, or erosion | 8.7% (2/23) | | Major penetration - protrusion into surrounding tissue with blood pooling, or erosion, or pseudoaneurysm | 0.0% (0/23) | | Pedunculated mobile mass† | 4.3% (1/23) | | Alterra Adaptive Prestent fracture‡ | | | Fracture not requiring intervention | 20.0% (5/25) | | Fracture requiring intervention | 0.0% (0/25) | | SAPIEN 3 valve frame fracture‡ | 0.0% (0/25) | *Rate (no./total no.) †Only included patients with evaluable images at visit time point. ‡Fracture was determined by either chest x-ray or fluoroscopy. 5. Pediatric Extrapolation In this premarket application, existing clinical data was not leveraged to support approval of a pediatric patient population. **E. Financial Disclosure** The Financial Disclosure by Clinical Investigators regulation (21 CFR 54) requires applicants who submit a marketing application to include certain information concerning the compensation to, and financial interests and arrangement of, any clinical investigator conduction clinical studies covered by the regulation. The Alterra Clinical Study involved 43 investigators of which none were full-time or part-time employees of the sponsor and 15 investigators had disclosable financial interests/arrangements as defined in 21 CFR 54.2(a), (b), (c) and (f), as described below: - Compensation to the investigator for conducting the study where the value could be influenced by the outcome of the study: None - Significant payment of other sorts: 15 - Proprietary interest in the product tested held by the investigator: None - Significant equity interest held by investigator in sponsor of covered study: None The applicant has adequately disclosed the financial interest/arrangements with clinical investigators. Statistical analyses were conducted by FDA to determine whether the financial interests/arrangements had any impact on the clinical study outcome. The information provided does not raise any questions about the reliability of the data. PMA P200015/S011: FDA Summary of Safety and Effectiveness Data Page 28 {28} ## **XI. PANEL MEETING RECOMMENDATION AND FDA'S POST-PANEL ACTION** In accordance with the provisions of section 515(c)(3) of the act as amended by the Safe Medical Devices Act of 1990, this PMA was not referred to the Circulatory Systems Devices panel, an FDA advisory committee, for review and recommendation because the information in the PMA substantially duplicates information previously reviewed by this panel. ## **XII. CONCLUSIONS DRAWN FROM THE PRECLINICAL AND CLINICAL STUDIES** ### **A. Effectiveness Conclusions** In the clinical study, 100% of the patients in the main cohort were free from THV dysfunction at 6 months post implant, with no RVOT/PV reintervention, moderate or greater total PR, or mean RVOT gradient $\geq 35$ mmHg. The proportion of patients with severe PR decreased from 93.3% at baseline to 0% at 6 months. Positive changes in right ventricular function were seen post implant, as evidenced by the reduction in the mean right ventricular end diastolic volume (RVEDV: 295.9 to 220.4 ml; RVEDV index: 162.7 to 118.9 ml/m$^2$) and decrease in main pulmonary artery regurgitant fraction (47.4% to 4.1%) from baseline to 6 months. In the main cohort, technical success was achieved in 96.7% of the patients at exit from the procedure room; device success was achieved in 88.3% of the patients at 30 days and 76.3% of the patients at 6 months; and procedural success was achieved in 88.3% of patients at 30 days. In the PDS registry cohort, acute device success was met in 88.0% of the patients. All patients in the main cohort were NYHA functional class I/II (I: 89.8%; II: 10.2%) at 6 months compared to 91.7% (I: 53.3%; II: 38.3%) of patients at baseline, with more patients in class I post implant. ### **B. Safety Conclusions** The risks of the device are based on nonclinical laboratory and animal studies as well as data collected in a clinical study conducted to support PMA approval as described above. The results from the nonclinical laboratory (e.g., biocompatibility, hydrodynamic performance, durability, and structural integrity) and animal studies demonstrated that the Edwards Alterra Adaptive Prestent and Edwards SAPIEN 3 THV are suitable for long-term implant. In the Alterra Clinical Study, there were no deaths at 6 months in the main cohort. The most frequent adverse event observed at 6 months in the main cohort was arrhythmia (33.3%), followed by bleeding (21.7%, including 18.4% minor bleeding), and minor vascular injury or access site complication (5.0%). ### **C. Benefit-Risk Determination** The probable benefits of TPVR with the Edwards SAPIEN 3 Transcatheter Pulmonary Valve System with Alterra Adaptive Prestent demonstrated in the Alterra Clinical Study include PMA P200015/S011: FDA Summary of Safety and Effectiveness Data Page 29 {29} improved RVOT/PV hemodynamics and patient functional status as measured by the NYHA classification at 6 months post implant. These improvements may delay the need for surgical replacement of the patient’s dysfunctional RVOT/PV. For patients that require multiple reoperations over a lifetime, with incremental increases in risk of surgical morbidity and mortality with each reoperation, TPVR with the Edwards SAPIEN 3 Transcatheter Pulmonary Valve System with Edwards Alterra Adaptive Prestent may reduce the cumulative risks associated with such operations. The probable risks of TPVR with the Edwards SAPIEN 3 Transcatheter Pulmonary Valve System with Alterra Adaptive Prestent include arrhythmia, bleeding, vascular injury or access site complications, and THV dysfunction. #### 1. Patient Perspectives This application did not include specific information on patient perspectives for TPVR with the Edwards SAPIEN 3 Transcatheter Pulmonary Valve System with Alterra Adaptive Prestent. In conclusion, given the available information above, the data support that for patients with severe PR who have a native or surgically-repaired RVOT, the probable benefits of the Edwards SAPIEN 3 Transcatheter Pulmonary Valve System with Alterra Adaptive Prestent outweigh the probable risks. ### D. Overall Conclusions The data in this application support the reasonable assurance of safety and effectiveness of the Edwards SAPIEN 3 Transcatheter Pulmonary Valve System with Alterra Adaptive Prestent for the management of pediatric and adult patients with severe PR who have a native or surgically-repaired right ventricular outflow tract and are clinically indicated for pulmonary valve replacement. ### XIII. CDRH DECISION CDRH issued an approval order on December 16, 2021. The final conditions of approval cited in the approval order are described below. The applicant must conduct two post-approval studies: 1. **Continued Follow-up of the Alterra IDE Cohort:** This study will be conducted in accordance with the protocol, entitled, “Post-Approval Data Analysis Protocol: Continued Follow-up of IDE Cohorts”, dated December 3, 2021. The study will consist of 119 patients enrolled in the IDE study (including the Continued Access Protocol investigation). The objective of the study is to characterize the clinical outcomes annually through 10 years post implant. The safety and effectiveness endpoints include RVOT/PV reintervention, THV hemodynamic function, and adverse events (including device frame fracture). PMA P200015/S011: FDA Summary of Safety and Effectiveness Data Page 30 {30} 2. **Alterra New Enrollment Study:** This study will be conducted in accordance with the protocol, entitled, “Multicenter Post-Approval Study of Congenital Pulmonic Valve Dysfunction Studying the Edwards SAPIEN 3 Transcatheter Pulmonary Valve System with Alterra Adaptive Prestent”, dated November 30, 2021. The study will enroll 150 patients at up to 25 sites in the U.S. The objective of the study is to characterize the real-world performance of the Edwards SAPIEN 3 Transcatheter Pulmonary Valve System with Edwards Alterra Adaptive Prestent through 10 years post implant. The safety and effectiveness endpoints include acute device success, THV dysfunction, echocardiographic assessments, NYHA functional class, and adverse events device (including device frame fracture). The applicant’s manufacturing facilities have been inspected and found to be in compliance with the device Quality System (QS) regulation (21 CFR 820). #### **XIV. APPROVAL SPECIFICATIONS** Directions for use: See final approved labeling (Instructions for Use). Hazards to Health from Use of the Device: See Indications, Contraindications, Warnings, Precautions, and Adverse Events in the final labeling (Instructions for Use). Post-approval Requirements and Restrictions: See Approval Order. PMA P200015/S011: FDA Summary of Safety and Effectiveness Data Page 31
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...