P190005 · Roche Diagnostics · SEI · Feb 3, 2021 · Microbiology
Device Facts
Record ID
P190005
Device Name
Elecsys Anti-HBe, PreciControl Anti-HBe
Applicant
Roche Diagnostics
Product Code
SEI · Microbiology
Decision Date
Feb 3, 2021
Decision
APRL
Regulation
21 CFR 866.3173
Device Class
Class 2
Intended Use
Elecsys Anti-HBe Immunoassay for the in vitro qualitative detection of total antibodies to hepatitis B e antigen (anti-HBe) in human adult serum or plasma (potassium EDTA, lithium heparin, sodium citrate, sodium heparin) from individuals with symptoms of hepatitis or at risk for hepatitis B virus (HBV) infection. Assay results, in conjunction with other laboratory results and clinical information may be used as an aid in the diagnosis of hepatitis B virus (HBV) infection in patients with symptoms of hepatitis or who may be at risk for HBV infection. A reactive test is presumptive laboratory evidence of HBV seroconversion. Further HBV serological marker testing is required to define the specific disease state. The electrochemiluminescence immunoassay "ECLIA" is intended for use on the cobas e 602 immunoassay analyzer. PreciControl Anti-HBe PreciControl Anti-HBe is used for quality control of the Elecsys Anti-HBe immunoassay on the cobas e 602 immunoassay analyzer. The performance of PreciControl Anti-HBe has not been established with any other anti-HBe assay.
Device Story
Elecsys Anti-HBe is an in vitro diagnostic immunoassay using electrochemiluminescence (ECLIA) technology on the cobas e 602 analyzer. Input: 35 µL human serum or plasma. Principle: Two-incubation competition assay; sample anti-HBe binds to HBeAg; remaining binding sites occupied by biotinylated and ruthenium-labeled anti-HBe antibodies; complex captured on streptavidin-coated microparticles via magnetic electrode. Output: Chemiluminescent emission measured by photomultiplier; results calculated automatically by software comparing sample signal to cutoff. Used in clinical laboratories by technicians. Results aid physicians in diagnosing HBV infection and determining disease chronicity/seroconversion status when used as part of a multi-marker panel. Benefits include improved disease management and treatment decisions for HBV patients.
Clinical Evidence
Clinical study evaluated 1800 samples (1500 prospective adult at-risk, 300 supplemental). Performance compared to FDA-approved reference assay and an internal confirmatory assay. Overall PPA 87.5% (95% CI: 84.1–90.2%) and NPA 98.7% (95% CI: 97.9–99.2%) against confirmatory method. Study included analytical sensitivity (0.127 IU/mL), cross-reactivity (190 samples), endogenous interference (hemoglobin, bilirubin, intralipid, biotin, albumin), and reproducibility across three sites.
Technological Characteristics
ECLIA technology; competition principle; two-incubation format. Components: Streptavidin-coated microparticles, HBeAg (E. coli, rDNA), biotinylated monoclonal anti-HBe (mouse), ruthenium-labeled monoclonal anti-HBe (mouse). Connectivity: cobas e 602 analyzer. Software: Automated signal processing and result calculation. Sterilization: Not specified. Materials: Streptavidin-coated microparticles.
Indications for Use
Indicated for adult patients with symptoms of hepatitis or at risk for HBV infection to aid in the diagnosis of HBV infection and as presumptive evidence of HBV seroconversion.
Regulatory Classification
Identification
A hepatitis B virus (HBV) antibody assay is identified as an in vitro diagnostic device intended for prescription use in the detection of antibodies to HBV in human serum, plasma, or other matrices, and as a device that aids in the diagnosis of HBV infection in persons with signs and symptoms of hepatitis and in persons at risk for hepatitis B infection. Results from assays may be qualitative or quantitative, such as quantitative anti-HBs. In addition, results from an anti-HBc IgM (IgM antibodies to core antigen) assay indicating the presence of anti-HBc IgM are indicative of recent HBV infection. Anti-HBs (antibodies to surface antigen) assay results may be used as an aid in the determination of susceptibility to HBV infection in individuals prior to or following HBV vaccination or when vaccination status is unknown. The assay is not intended for screening of blood, plasma, cells, or tissue donors. The assay is intended as an aid in diagnosis in conjunction with clinical findings and other diagnostic procedures.
Special Controls
*Classification.* Class II (special controls). The special controls for this device are:(1) The labeling required under § 809.10(b) of this chapter must include:
(i) A prominent statement that the assay is not intended for the screening of blood, plasma, cells, or tissue donors.
(ii) A detailed explanation of the principles of operation and procedures for performing the assay.
(iii) A detailed explanation of the interpretation of results.
(iv) Limitations, which must be updated to reflect current clinical practice and disease presentation and management. The limitations must include statements that indicate:
(A) When appropriate, performance characteristics of the assay have not been established in populations of immunocompromised or immunosuppressed patients or other special populations where assay performance may be affected.
(B) Detection of HBV antibodies to a single viral antigen indicates a present or past infection with hepatitis B virus, but does not differentiate between acute, chronic, or resolved infection.
(C) The specimen types for which the device has been cleared, and that use of the assay with specimen types other than those specifically cleared for this device may result in inaccurate assay results.
(D) Diagnosis of hepatitis B infection should not be established on the basis of a single assay result but should be determined by a licensed healthcare professional in conjunction with the clinical presentation, history, and other diagnostic procedures.
(E) A non-reactive assay result may occur early during acute infection, prior to development of a host antibody response to infection, or when analyte levels are below the limit of detection of the assay.
(F) Results obtained with this assay may not be used interchangeably with results obtained with a different manufacturer's assay.
(v) For devices intended for the quantitative detection of HBV antibodies (anti-HBs), in addition to the special controls listed in paragraphs (b)(1) and (2) of this section, labeling required under § 809.10(b) of this chapter must include:
(A) The assay calibrators' traceability to a standardized reference material that FDA has determined is appropriate (
*e.g.,* a recognized consensus standard) and the limit of blank (LoB), limit of detection (LoD), limit of quantitation (LoQ), linearity, and precision to define the analytical measuring interval.(B) Performance results of the analytical sensitivity study testing a standardized reference material that FDA has determined is appropriate (
*e.g.,* a recognized consensus standard).(2) Design verification and validation must include the following:
(i) Detailed device description, including all parts that make up the device, ancillary reagents required but not provided, an explanation of the device methodology, and design of the antigen(s) and capture antibody(ies) sequences, rationale for the selected epitope(s), degree of amino acid sequence conservation of the target, and the design and composition of all primary, secondary and subsequent standards used for calibration.
(ii) Documentation and characterization (
*e.g.,* supplier, determination of identity, and stability) of all critical reagents (including description of the antigen(s) and capture antibody(ies)), and protocols for maintaining product integrity throughout its labeled shelf life.(iii) Risk analysis and management strategies, such as Failure Modes Effects Analysis and/or Hazard Analysis and Critical Control Points summaries and their impact on assay performance.
(iv) Final release criteria to be used for manufactured assay lots with appropriate evidence that lots released at the extremes of the specifications will meet the identified analytical and clinical performance characteristics as well as stability.
(v) Stability studies for reagents must include documentation of an assessment of real-time stability for multiple reagent lots using the indicated specimen types and must use acceptance criteria that ensure that analytical and clinical performance characteristics are met when stability is assigned based on the extremes of the acceptance range.
(vi) All stability protocols, including acceptance criteria.
(vii) When applicable, analytical sensitivity of the assay that is the same or better than that of other cleared or approved assays.
(viii) Analytical performance studies and results for determining the limit of blank (LoB), limit of detection (LoD), cutoff, precision (reproducibility), including lot-to-lot and/or instrument-to-instrument precision, interference, cross reactivity, carryover, hook effect, seroconversion panel testing, matrix equivalency, specimen stability, reagent stability, and cross-genotype antibody detection sensitivity, when appropriate.
(ix) For devices intended for the detection of antibodies for which a standardized reference material (that FDA has determined is appropriate) is available, the analytical sensitivity study and results testing the standardized reference material. Detailed documentation of that study and its results must be provided, including the study protocol, study report, testing results, and all statistical analyses.
(x) For devices with associated software or instrumentation, documentation must include a detailed description of device software, including software applications and hardware-based devices that incorporate software. The detailed description must include documentation of verification, validation, and hazard analysis and risk assessment activities, including an assessment of the impact of threats and vulnerabilities on device functionality and end users/patients as part of cybersecurity review.
(xi) Detailed documentation of clinical performance testing from a clinical study with an appropriate number of HBV reactive and non-reactive samples in applicable risk categories and conducted in the appropriate settings by the intended users. Performance must be analyzed relative to an FDA cleared or approved HBV antibody assay or a comparator that FDA has determined is appropriate. Additional relevant patient groups must be validated as appropriate. The samples must include prospective (sequential) samples for each identified specimen type and, as appropriate, additional characterized clinical samples. Samples must be sourced from geographically diverse areas.
(3) For any HBV antibody assay intended for quantitative detection of anti-HBV antibodies, the following special controls, in addition to those special controls listed in paragraphs (b)(1) and (2) of this section, also apply:
(i) Detailed documentation of the metrological calibration traceability hierarchy to a standardized reference material that FDA has determined is appropriate.
(ii) Detailed documentation of the following analytical performance studies conducted, as appropriate to the technology, specimen types tested, and intended use of the device, including upper and lower limits of quantitation (UloQ and LloQ, respectively), linearity using clinical samples, and an accuracy study using the recognized international standard material.
P090028 — VITROS IMMUNODIAGNOSTIC PRODUCTS HBEAG REAGENT PACK/PRODUCTS HBEAG CALIBRATOR/PRODUCTS HBE CONTROLS · Ortho-Clinical Diagnostics, Inc. · May 11, 2011
Submission Summary (Full Text)
{0}
SUMMARY OF SAFETY AND EFFECTIVENESS DATA (SSED)
I. GENERAL INFORMATION
Device Generic Name: Antibodies to Hepatitis B e antigen (Anti-HBe)
Device Trade Name: Elecsys Anti-HBe, PreciControl Anti-HBe
Device Procode: LOM
Applicant's Name and Address: Roche Diagnostics
9115 Hague Road
Indianapolis, IN 46250
Date(s) of Panel Recommendation: None
Premarket Approval Application (PMA) Number: P190005
Date of FDA Notice of Approval: February 3, 2021
II. INDICATIONS FOR USE
Elecsys Anti-HBe
Immunoassay for the in vitro qualitative detection of total antibodies to hepatitis B e antigen (anti-HBe) in human adult serum or plasma (potassium EDTA, lithium heparin, sodium citrate, sodium heparin) from individuals with symptoms of hepatitis or at risk for hepatitis B virus (HBV) infection. Assay results, in conjunction with other laboratory results and clinical information may be used as an aid in the diagnosis of hepatitis B virus (HBV) infection in patients with symptoms of hepatitis or who may be at risk for HBV infection. A reactive test is presumptive laboratory evidence of HBV seroconversion. Further HBV serological marker testing is required to define the specific disease state.
The electrochemiluminescence immunoassay "ECLIA" is intended for use on the cobas e 602 immunoassay analyzer.
PreciControl Anti-HBe
PreciControl Anti-HBe is used for quality control of the Elecsys Anti-HBe immunoassay on the cobas e 602 immunoassay analyzer. The performance of PreciControl Anti-HBe has not been established with any other anti-HBe assay.
PMA P190005: FDA Summary of Safety and Effectiveness Data
{1}
PMA P190005: FDA Summary of Safety and Effectiveness Data
# III. CONTRAINDICATIONS
There are no known contraindications.
# IV. WARNINGS AND PRECAUTIONS
The warnings and precautions can be found in the Elecsys Anti-HBe labeling.
# V. DEVICE DESCRIPTION
The Elecsys Anti-HBe immunoassay is a qualitative test that employs the electrochemiluminescence "ECLIA" technology. The assay employs a two-incubation step assay using the competition principle test format and a total assay time of 18 minutes.
- 1st incubation: Anti-HBe in the sample (35 µL) binds to the added HBeAg.
- 2nd incubation: After addition of biotinylated antibodies and ruthenium complex-labeled antibodies specific for HBeAg, together with streptavidin-coated microparticles, the still-free binding sites on the HBe antigens become occupied. The entire complex is then bound to the solid phase via interaction of biotin and streptavidin.
- The reaction mixture is aspirated into the measuring cell where the microparticles are magnetically captured onto the surface of the electrode. Unbound substances are then removed with ProCell M. Application of a voltage to the electrode then induces chemiluminescent emission which is measured by a photomultiplier.
- Results are determined automatically by the software by comparing the electrochemiluminescence signal obtained from the reaction product of the sample with the signal of the cutoff value previously obtained by calibration.
The Elecsys Anti-HBe consists of five components summarized below:
Table 1: Components of the Elecsys Anti-HBe
| | Name | Description |
| --- | --- | --- |
| Rackpack (bundle of three reagent bottles) labeled as A-HBE | M | Streptavidin-coated microparticles (transparent cap), 1 bottle 6.5 mL consists of streptavidin-coated microparticles 0.72 mg/mL; preservative. |
| | R1 | HBeAg (gray cap), 1 bottle, 12 mL consists of HBeAg (E. coli, rDNA) > 7 ng/mL; HEPES) buffer 36 mmol/L, pH 7.4; preservative. |
| | R2 | Anti-HBeAg-Ab~biotin; anti-HBeAg-Ab~Ru(bpy)3 (black cap), 1 bottle, 12 mL: Biotinylated monoclonal anti-HBe antibody (mouse) > 0.8 mg/L; monoclonal anti-HBe antibody (mouse) labeled with ruthenium |
2015
{2}
The following results are reported:
Table 2: Interpretation of Results
| Output Elecsys Anti-HBe (S/CO) | Test Result | Interpretation |
| --- | --- | --- |
| > 1.00 S/CO* | Non-Reactive | Indicates sample is non-reactive for anti-HBe. A negative test result does not exclude the possibility of infection with hepatitis B virus. |
| ≤ 1.00 S/CO | Reactive | Indicates sample is reactive for anti-HBe |
s/co=signal/cutoff
## VI. ALTERNATIVE PRACTICES AND PROCEDURES
There are several other alternatives for the detection of total antibodies to hepatitis B antigen (anti-HBe). There are currently several FDA approved in vitro diagnostic tests commercially available for serological markers of hepatitis B virus (HBV) infection which, when used in conjunction with a patient's medical history, clinical examination and other laboratory findings, may be used as an aid in the diagnosis of HBV infection in patients with symptoms of hepatitis or who may be at risk of HBV infection. The assay may be used as aid in determining acute infection. Each alternative has its own advantages and disadvantages. A patient should fully discuss these alternatives with his/her physician to select the method that best meets expectations and lifestyle.
## VII. MARKETING HISTORY
The device has been marketed in the following countries (Table 3) and has not been withdrawn from any country for reasons relating to safety and effectiveness.
Table 3: Marketing History
| Argentina | Ecuador | Mexico | Slovakia |
| --- | --- | --- | --- |
| Australia | Egypt | Middle East | Slovenia |
| Austria | Finland | Myanmar | South Africa |
| Baltics | France | Netherlands | Spain |
| Belgium | Germany | New Zealand | Sweden |
| Brazil | Greece | Norway | Switzerland |
| Canada | Hong Kong | Pakistan | Taiwan |
PMA P190005: FDA Summary of Safety and Effectiveness Data
{3}
| Central America | Hungary | Peru | Thailand |
| --- | --- | --- | --- |
| Chile | India | Philippines | Turkey |
| China | Indonesia | Poland | United Kingdom |
| Colombia | Italy | Portugal | Uruguay |
| Croatia | Japan | Romania | Venezuela |
| Czech Republic | Korean Republic | Russian Federation | Vietnam |
| Denmark | Malaysia | Singapore | Slovakia |
## VIII. POTENTIAL ADVERSE EFFECTS OF THE DEVICE ON HEALTH
Below is a list of the potential adverse effects (e.g., complications) associated with the use of the device. The Elecsys Anti-HBe is intended for *in vitro* diagnostic use, and as a result, there is no direct adverse effect on the patient. Standard good laboratory practices are considered sufficient to minimize risks to the end user.
Failure of the product to perform as intended or human error in the use of the test may lead to a false result. Appropriate Warnings and Precautions for identified risks are contained in the labeling and assay Instructions for Use.
The risks associated with the device, when used as intended, are those related to the risk of false test results, failure to correct interpret the test results and failure to correctly operate the instrument.
Risks of a false positive test includes improper patient management, including premature discontinuation of antiviral treatment should a clinician be falsely led to determine a patient has seroconverted. This risk is mitigated by the fact that this assay is usually repeated and is used as part of a panel. Repeatedly false positive results have the potential to lead to inappropriate treatment decisions, however anti-HBe is not used in isolation to determine seroconversion status. Because anti-HBe is sometimes included as part of a panel in clinical practice to diagnose hepatitis B but is more commonly utilized as part of a panel to determine chronicity of disease and immune-active versus chronic infection, the risk of a false positive will likely be somewhat mitigated as incongruous test results would lead a clinician to either retest the patient or further investigate the etiology of hepatitis.
Risk of a false negative test includes improper patient management, including continued treatment for hepatitis B with antiviral medication. Antiviral medication has risks including toxicity and more rarely allergic reactions. Over time, viral resistance in patients who are co-infected but undiagnosed with other viruses using the same antiviral medication, such as HIV, can lead to viral resistance, however the chance of an undiagnosed co-infection in a patient tested for hepatitis B is exceedingly unlikely. Anti-HBe is not used in this same manner as HBeAg to guide treatment decisions, and thus confers less clinical risk to this subpopulation than a false negative or false positive HBeAg test. Anti-HBe is not used in isolation to determine seroconversion status. Because anti-HBe is sometimes included as part of a panel in clinical practice to diagnose hepatitis B but is more commonly utilized as
PMA P190005: FDA Summary of Safety and Effectiveness Data
{4}
part of a panel to determine chronicity of disease and immune-active v. chronic infection, the risk of a false negative will likely be mitigated as incongruous test results would lead a clinician to either retest the patient or further investigate the etiology of hepatitis
# IX. SUMMARY OF NONCLINICAL STUDIES
# A. Laboratory Studies
# 1. Cut-Off Determination
The cut-off was established internally at Roche and verified by testing a total of 181 samples (84 positives and 97 negatives). A Receiver Operating Characteristics (ROC) analysis was performed on the results of the samples tested. The assay's cut-off was evaluated with the observed results to demonstrate that its selection represents the best level of specificity, without compromising sensitivity.
The cut-off value of 1.0 is within the optimal range determined by the ROC curve to discriminate between negative and positive results.
# 2. Analytical Sensitivity/Dilution Study with Standard
The WHO International Standard Anti-Hepatitis B virus e antigen (anti-HBe), code 129095/12, was diluted into HBV negative human serum by nine dilution steps to produce ten different concentrations (0, 0.02, 0.039, 0.078, 0.156, 0.313, 0.625, 1.25, 2.5, and $5\mathrm{IU / mL}$ ). The samples were measured in duplicate with one reagent lot of Elecsys Anti-HBe to establish a standard curve. The sensitivity at the cut-off for the Elecsys Anti-HBe on the cobas e 602 analyzer is $0.127\mathrm{IU / mL}$ .
# 3. Sensitivity/Seroconversion Panels
The seroconversion sensitivity of the Elecsys Anti-HBe has been demonstrated by testing 8 commercially seroconversion panels in comparison to a reference anti-HBe immunoassay in terms of number of days from initial draw to first positive sample, as well as the difference between the last negative results and the first positive results. The following table shows that the Elecsys Anti-HBe detected a positive result sooner by one or more blood draws than the comparator assay in 4 out of 8 panels.
Table 4: Seroconversion Performance of the Elecsys Anti-HBe
| Panel ID | Bleed day of the last NRa) test | | Bleed day of the first RXb) test | | Difference in daysc) to anti-HBe reactive |
| --- | --- | --- | --- | --- | --- |
| | Reference assay | Elecsys Anti-HBe assay | Reference assay | Elecsys Anti-HBe assay | |
| 6281 | 54 | 41 | N/A | 43 | >11 |
| 6510 | 56 | 14 | 70 | 28 | 42 |
| 9092 | 198 | 134 | N/A | 141 | >57 |
| 9093 | 173 | 173 | 182 | 182 | 0 |
| 11024 | 54 | 54 | N/A | N/A | N/A |
PMA P190005: FDA Summary of Safety and Effectiveness Data
{5}
| Panel ID | Bleed day of the last NRa) test | | Bleed day of the first RXb) test | | Difference in daysc) to anti-HBe reactive |
| --- | --- | --- | --- | --- | --- |
| | Reference assay | Elecsys Anti-HBe assay | Reference assay | Elecsys Anti-HBe assay | |
| BMX11071 | 81 | 69 | 88 | 81 | 7 |
| BMX11072 | 60 | 60 | 74 | 74 | 0 |
| BMX11073 | 78 | 78 | 140 | 140 | 0 |
a) $\mathrm{NR} =$ non-reactive
b) $\mathrm{RX} =$ reactive
c) The dates of the first reactive test results were compared between the Elecsys Anti-HBe assay and the Reference assay
# 4. Analytical Specificity (Cross-Reactivity)
A study was conducted to evaluate the Elecsys Anti-HBe for potential cross-reactivity in specimens from individuals with various medical conditions. The specificity of 190 samples with 23 sub-categories of potentially interfering diseases or medical conditions was evaluated with the Elecsys Anti-HBe assay on the cobas e 602 immunoassay analyzer and the reference assay.
Of the 190 samples, there were 6 samples that showed cross-reactivity. The results of each potential cross reactant are shown in table below.
Table 5: Cross-Reactivity Results
| Reactivity of the Elecsys Anti-HBe assay | | | | |
| --- | --- | --- | --- | --- |
| Category | Sub-category | \( RX^h) \) | \( NR^i) \) | Total |
| Immune disorders | ANA | 0 | 12 | 12 |
| | RF | 0 | 12 | 12 |
| Infections / disorders | T. pallidum | 0 | 11 | 11 |
| | Toxoplasmosis | 0 | 11 | 11 |
| Infectious Viral agents | CMV | 0 | 10 | 10 |
| | EBV | 1 | 11 | 12 |
| | HAV | 0 | 12 | 12 |
| | HCV | 2 | 10 | 12 |
| | HIV 1/2 | 0 | 10 | 10 |
| | HSV | 0 | 12 | 12 |
| | Parvo B19 | 1 | 11 | 12 |
| | Rubella | 0 | 12 | 12 |
| | VZV | 0 | 12 | 12 |
| Non-viral | Liver cancer | 2 | 8 | 10 |
PMA P190005: FDA Summary of Safety and Effectiveness Data
{6}
PMA P190005: FDA Summary of Safety and Effectiveness Data
7 of 27
| Reactivity of the Elecsys Anti-HBe assay | | | | |
| --- | --- | --- | --- | --- |
| Category | Sub-category | RX^{h)} | NR^{i)} | Total |
| liver disease | Chronic alcoholic liver disease | 0 | 7 | 7 |
| | Various cirrhosis | 0 | 5 | 5 |
| | Alcoholic fatty liver | 0 | 4 | 4 |
| | Alcoholic liver disease | 0 | 4 | 4 |
| | Abdominal pain / pelvic mass | 0 | 4 | 4 |
| | Unspecified jaundice | 0 | 2 | 2 |
| | Liver abscess or lesion | 0 | 2 | 2 |
| | Fatty infiltrate of liver | 0 | 1 | 1 |
| | Chronic passive congestion of liver | 0 | 1 | 1 |
| Total | | 6 | 184 | 190 |
5. Endogenous Interference
To evaluate the effect of elevated levels of hemoglobin, bilirubin, intralipid, biotin, and total protein on the Elecsys Anti-HBe assay, one negative, one high negative, one low positive, and one positive anti-HBe serum samples were spiked with potential interferents. Each interferent was evaluated at 11 concentrations. All samples were tested in duplicate. Interferences were tested up to the listed concentration and no impact on results was observed.
Table 6: Interfering Substances
| Compound | Concentration tested |
| --- | --- |
| Bilirubin | ≤ 66 mg/dL |
| Hemoglobin | ≤ 2000 mg/dL |
| Intralipid | ≤ 2000 mg/dL |
| Albumin | ≤ 7 g/dL |
6. Endogenous Interference-Biotin
Because this test employs strept-avidin technology and may be subject to potential interference with biotin, more extensive interference testing with biotin was performed. The following tables show the level of interference that was observed.
Table 7: Bias for Samples Containing Various Concentrations of Biotin
| Sample (COI)^{a} | Biotin concentration (ng/mL) | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | | | 96 | 112 | 128 | 144 | 160 |
| negative | 1.56 | | -1.8 | -5.2 | -8.2 | -12.5 | -17.1 |
{7}
| Sample (COI)a | Biotin concentration (ng/mL) | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | | | 96 | 112 | 128 | 144 | 160 |
| high negative | 1.14 | relative deviation (%) | -6.2 | -6.9 | -12.2 | -14.3* | -19.8* |
| low positive | 0.892 | absolute deviation | -0.025 | -0.052 | -0.069 | -0.108 | -0.154 |
| positive | 0.371 | | -0.019 | -0.032 | -0.043 | -0.058 | -0.075 |
| Sample (COI) | Biotin concentration (ng/mL) | | | | | |
| --- | --- | --- | --- | --- | --- | --- |
| | | | 200 | 300 | 600 | 1200 |
| negative | 1.56 | relative deviation (%) | -46.2* | -74.1* | -89.3* | -95.8* |
| high negative | 1.14 | | -45.5* | -73.7* | -88.4* | -93.1* |
| low positive | 0.892 | absolute deviation | -0.361 | -0.604 | -0.730 | -0.765 |
| positive | 0.371 | | -0.151 | -0.226 | -0.263 | -0.273 |
$^{\mathrm{a}}$ COI=cutoff index
*false positive
Specimens with biotin concentrations up to 112 ng/mL demonstrated ≤ 10 % bias in results. Biotin concentrations greater than 112 ng/mL lead to higher negative bias and in consequence can lead to false negative Elecsys Anti-HBe results in samples near the medical decision value. Pharmacokinetic studies have shown that serum concentrations of biotin can reach up to 355 ng/mL within the first hour after biotin ingestion for subjects consuming supplements of 20 mg biotin per day and up to 1160 ng/mL for subjects after a single dose of 300 mg biotin.
7. Drug Interference
A drug interference study was performed with 16 common therapeutic drugs. Each drug was spiked into one negative, one high negative, one low positive, and one positive sample. Each sample was tested in triplicate. Each drug was found to be non-interfering at the following claimed concentrations.
Table 8: Drug Interference
| Compound | Concentration (mg/L) |
| --- | --- |
| Acetyl cysteine | 150 |
| Ampicillin-Na | 1000 |
| Ascorbic acid | 300 |
| Cefoxitin | 2500 |
| Heparin | 5000 U/L |
| Levodopa | 20 |
PMA P190005: FDA Summary of Safety and Effectiveness Data
{8}
| Compound | Concentration (mg/L) |
| --- | --- |
| Methyldopa+ 1.5 | 20 |
| Metronidazole | 200 |
| Doxycycline | 50 |
| Acetylsalicylic acid | 1000 |
| Rifampicin | 60 |
| Cyclosporine | 5 |
| Phenylbutazone | 400 |
| Acetaminophen | 200 |
| Ibuprofen | 500 |
| Theophylline | 100 |
6. Human Anti-Mouse Antibody Effect (HAMA) Study
Two samples (one anti-HBe negative sample, one native anti-HBe negative samples with positive anti-HBe sample) were spiked with Human anti-mouse antibodies in ten concentration levels with a maximum concentration of 60 ug/mL. The samples were measured in duplicate. The percent recovery for the negative samples ranged from 98-102% while the standard deviation for the positive samples ranged from 0-0.019 s/co. Results met the acceptance criteria. No interference for Elecsys Anti-HBe was observed with 60 ug/mL HAMA titer.
7. Sample Equivalence/Matrix Equivalency
Studies were conducted to evaluate the suitability of the following seven sample types: serum/gel separation tubes, sodium heparin plasma, sodium citrate plasma, lithium heparin plasma, K₂-EDTA plasma, K₃-EDTA plasma and plasma tubes containing separation gel, to be used with the Elecsys Anti-HBe assay.
Samples were collected into matched serum and plasma collection tubes and assayed in single determinations. The study was conducted using negative, high negative, low positive, and positive samples for anti-HBe. The studies support the use of serum/gel separation tubes, and the following plasma types: lithium heparin, K₂-EDTA, K₃-EDTA, sodium heparin, sodium citrate, and plasma tubes containing separating gel.
8. Carry-Over Study
The use of disposable tips for sample pipetting on the cobas e 602 immunoassay analyzer should eliminate the risk of sample carryover. A study was performed by testing six anti-HBe positive samples with Elecsys Anti-HBe in triplicate and a high signal generating sample in an Elecsys assay (Toxo IgG immunoassay) in alternating patterns. The percent recovery ranged from 103% to 110%.
PMA P190005: FDA Summary of Safety and Effectiveness Data
9 of 27
{9}
PMA P190005: FDA Summary of Safety and Effectiveness Data
10 of 27
# 9. Stability Studies
## Sample Stability
Studies were performed to determine the storage stability of patient serum and plasma samples at storage temperatures of 2-8°C, room temperature (RT), and -20°C. A multiple freeze/thaw study was also performed.
Serum and plasma samples tested contained anti-HBs analyte levels of negative, high negative, low positive, and positive.
- 2-8°C study-samples were tested unstressed (T=0) and again after 1, 5, 7, 11, and 14 days at 2-8°C in triplicate.
- Room temperature study (RT)- samples were tested immediately after preparation and again after 1, 2, 3, and 7 days of storage at RT.
- -20°C Study-samples were tested unstressed (t=0) and stored at -20°C 1, 2, and 3 months.
- Freeze/thaw study-samples were tested unstressed (T=0) and after 6 freeze thaw cycles in triplicate.
Table 9: Sample Stability Claims in Serum and Plasma
| Sample Matrix | Number of Freeze and Thaw Cycles | Storage at 2-8°C | Storage at -20°C | Storage at Room Temperature |
| --- | --- | --- | --- | --- |
| Serum and Plasma | 6 | 14 days | 3 months | 14 days |
## Reagent Stability-Real-Time (Shelf-Life)
Studies were performed to establish the shelf-life for the Elecsys Anti-HBe. A negative, high negative, low positive, and positive sample were tested with three lots of Elecsys Anti-HBe were stored at the recommended storage temperature of 2-8°C throughout the study. Performance was assessed against clinically relevant acceptance criteria. Study demonstrated reagents are stable and continue to meet acceptance criteria for 24 months after date of manufacture.
## Reagent Stability- after First Opening
This study was performed to determine the time period over which the Elecsys Anti-HBe can be kept at 2-8°C once opened. One test kit was opened and the cobas e 602 analyzer calibrated. A negative, high negative, low positive, and positive serum sample and two Elecsys PreciControl Anti-HBe were tested with the opened reagent unstressed (T=0) and after 8 and 9 weeks at 2-8°C. Study demonstrated that reagents are stable for 8 weeks when stored at 2-8°C after first opening.
## Reagent Stability-Reagent On-Board
This study was performed to determine the time period for which the Elecsys Anti-HBe reagents can be stored on the analyzer once opened. The reagent packs were stored onboard for 8 and 9 weeks at 20-25°C. Unstressed reagent packs were opened and the analyzer was calibrated. A negative, high negative, low positive, and positive serum sample and two Elecsys PreciControl Anti-HBe were tested with the unstressed reagent
{10}
packs (stored at 2-8°C) and with the reagent packs which were stressed on-board for 8 and 9 weeks. For each time point, the calibration occurred seven days prior. Study demonstrated that reagents are stable for 8 weeks when stored on-board the cobas e 602 analyzer.
## Reagent Stability-Temperature Stress
This study was conducted to determine the effect of elevated temperature stress on the Elecsys Anti-HBe reagents during transportation. The reagent kit was stressed for one week at 35°C. A negative, high negative, low positive, and positive serum samples and both PreciControls were measured in duplicate with the stressed reagent kits and compared to the results from the unstressed kit. Study demonstrated that reagents are stable for one week at 35°C.
## Lot Calibration Stability
An unstressed rackpack of Elecsys Anti-HBe reagent was calibrated on the cobas e 602 analyzer. A negative, high negative, low positive, and positive serum sample and PreciControls were tested in duplicate (T=0). After 28 and 35 days, reagent of the same lot was run again using the initial calibration. Study supports a claim for lot calibration of 28 days.
## On-Board Calibration Stability
An Elecsys Anti-HBe reagent pack was tested unstressed (stored at 2-8°C) and after 7 and 14 days of storage on-board the cobas e 602 at 20-25°C, using the initial calibration of day 0 to demonstrate the stability of the initial calibration and the stability of the control measurements. Study supports a claim for on-board calibration stability of 7 days.
## Calibrator Stability-After First Opening
This study was performed to determine the time period in which the Elecsys Anti-HBe calibrators can be kept at 2-8°C once opened. A new reagent pack was opened and the cobas e 602 calibrated. The opened calibrators were then tested again in duplicate after 13 weeks stored at 2-8°C. Study supports calibrator stability for 8 weeks at 2-8°C after first opening.
## Calibrator Stability-On-Board
According to product specification, calibrators 1 and 2 may be used only once. Unless the entire volume is necessary for calibration on the analyzer, aliquots of the ready-for-use calibrators may be transferred into empty snap-cap bottles (CalSet vials), should be left on the analyzer only during calibration, and should be discarded after use on the cobas e 602 analyzer. This study was performed to assess stability of the calibrators on-board the cobas e 602 analyzer at 20-25°C. A pair of Elecsys Anti-HBe calibrators were opened and stored at 20-25°C to simulate on-board stress. After 7 hours of incubation at 20-25°C, the calibrators were tested in duplicate with a pair of unstressed calibrators. Study supports calibrator on-board stability of 6 hours.
## PreciControl Stability-Real Time Shelf-Life
PMA P190005: FDA Summary of Safety and Effectiveness Data
11 of 27
{11}
Shelf life was determined by testing one production lot in duplicate of PreciControl Anti-HBe stored at 2-8°C. Study supports PreciControl shelf life stability of 22 months at 2-8°C.
## PreciControl Stability-After First Opening
Stability studies were performed to determine the time period over which the PreciControl Anti-HBe can be kept at 2-8°C once opened. A new PreciControl kit pack was opened, tested in duplicate on day 0 (unstressed reference) on the cobas e 602 analyzer, and then tested again in duplicate after 8 and 9 weeks of storage at 2-8°C. Study supports PreciControl Anti-HBe stored for 8 weeks at 2-8°C after first opening.
## PreciControl Stability-On Board
Stability studies were performed to determine the time period over which PreciControl Anti-HBe can be kept on-board at 20-25°C once opened. A new PreciControl kit was opened and measured on the cobas e 602 analyzer in duplicate (unstressed reference). After stored open at 32°C for 7 hours, the control kit was run again on the cobas e 602 analyzer. Opened PreciControl Anti-HBe can be stored up to 6 hours on-board the cobas e 602 analyzer.
## PreciControl Stability-Temperature Stress
This study was conducted to determine the effect of elevated temperature stress on the PreciControl Anti-HBe during transport. One kit was stored at the recommended storage of 2-8°C and a second kit was stressed for one week at 35°C. PreciControl Anti-HBe are stable for 1 week at 35°C.
## 10. Within Laboratory Precision
A six member precision panel consists of four human serum (HS) pools (one negative, one high negative, one low positive, and one positive) and two PreciControls (PC1 and PC2) were tested in duplicate, on one cobas e 602 analyzer, at one site, with one reagent lot, in two runs per day for 12 days. Calibration was performed on day 1 and day 7 spanning at least two calibration cycles. Repeatability (within-run) and within-laboratory precision were calculated according to EP5-A3. Results are shown in the following table.
| | | Repeatability^{a)} | | Intermediate precision^{b)} | |
| --- | --- | --- | --- | --- | --- |
| Sample | Mean COI | SD COI | CV % | SD COI | CV % |
| HS^{c)}, negative | 1.50 | 0.016 | 1.0 | 0.030 | 2.0 |
| HS, high negative | 1.10 | 0.012 | 1.1 | 0.025 | 2.3 |
| HS, low positive | 0.799 | 0.008 | 1.0 | 0.020 | 2.5 |
| HS, positive | 0.023 | 0.001 | 2.8 | 0.002 | 7.0 |
| PC^{d)} A-HBe 1 | 1.49 | 0.027 | 1.8 | 0.034 | 2.3 |
PMA P190005: FDA Summary of Safety and Effectiveness Data
12 of 27
{12}
| | | Repeatabilitya) | | Intermediate precisionb) | |
| --- | --- | --- | --- | --- | --- |
| Sample | Mean COI | SD COI | CV % | SD COI | CV % |
| PC A-HBe 2 | 0.660 | 0.008 | 1.2 | 0.021 | 3.1 |
a) Repeatability = within-run precision
b) Intermediate precision = within-laboratory precision
c) HS = Human serum
d) PC = PreciControl
## 11. Reproducibility
The study was performed at three sites with three reagent lots with PreciControl Anti-HBe and sample concentrations shown in the table below in three replicates per run, two runs per day, for five days, on three cobas e 602 analyzers in a modified protocol CLSI EP05-A3. The results are shown in the following table.
Table 11: Reproducibility
| | | Repeatability | | Between-run | | Between-day | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| Sample | Mean COI | SD COI | CV % | SD COI | CV % | SD COI | CV % |
| HSPa) 01 | 0.576 | 0.009 | 1.5 | 0.012 | 2.2 | 0.015 | 2.6 |
| HSP 02 | 0.783 | 0.012 | 1.6 | 0.013 | 1.6 | 0.021 | 2.7 |
| HSP 03 | 0.885 | 0.013 | 1.5 | 0.011 | 1.3 | 0.024 | 2.7 |
| HSP 04 | 1.05 | 0.017 | 1.6 | 0.014 | 1.3 | 0.030 | 2.9 |
| HSP 05 | 1.08 | 0.019 | 1.8 | 0.012 | 1.1 | 0.029 | 2.7 |
| HSP 06 | 1.22 | 0.020 | 1.7 | 0.013 | 1.1 | 0.033 | 2.8 |
| PC A-HBe 1 | 1.53 | 0.029 | 1.9 | 0.018 | 1.2 | 0.031 | 2.1 |
| PC A-HBe 2 | 0.643 | 0.012 | 1.8 | 0.012 | 1.9 | 0.013 | 2.1 |
a) HSP = human serum pool
| | Between-lot | | Between-site | | Reproducibility | |
| --- | --- | --- | --- | --- | --- | --- |
| Sample | SD COI | CV % | SD COI | CV % | SD COI | CV % |
| HSP 01 | 0.019 | 3.2 | 0.024 | 4.1 | 0.037 | 6.4 |
| HSP 02 | 0.025 | 3.2 | 0.025 | 3.2 | 0.045 | 5.7 |
| HSP 03 | 0.025 | 2.9 | 0.023 | 2.6 | 0.045 | 5.1 |
| HSP 04 | 0.027 | 2.6 | 0.034 | 3.2 | 0.057 | 5.5 |
| HSP 05 | 0.034 | 3.2 | 0.022 | 2.0 | 0.055 | 5.1 |
| HSP 06 | 0.046 | 3.8 | 0.026 | 2.1 | 0.067 | 5.5 |
| PC A-HBe 1 | 0.025 | 1.6 | 0.013 | 0.83 | 0.054 | 3.5 |
| PC A-HBe 2 | 0.012 | 1.8 | 0.015 | 2.4 | 0.029 | 4.5 |
PMA P190005: FDA Summary of Safety and Effectiveness Data
{13}
12. Antimicrobial Effectiveness Testing
The purpose of this study was to prove the effectiveness of the added preservatives in the reagents. One lot each of reagent was tested with a panel of microorganisms. All reagents were plated on appropriate media prior to inoculation. Non-inoculated reagent (reference) was incubated in parallel with inoculated reagents and palted at each time. After inoculation, the reagents were incubated for 14 and 28. To pass United States Pharmacopoeia) USP criteria, the bacterial concentration is to be reduced to < 0.1 % of the original inoculum by day 14 and remain at or below this level until day 28. Yeast and molds are to remain at or below the original inoculum during the 28-day period. USP criteria suggest that a suitable inoculum should be between 1 x 10⁵ and 1 x 10⁶ organisms per mL. All reagents met the USP requirements for antimicrobial effectiveness testing.
B. Animal Studies
Not Applicable
C. Additional Studies
Not Applicable
X. SUMMARY OF PRIMARY CLINICAL STUDY
The applicant performed a clinical study to establish a reasonable assurance of safety and effectiveness for the detection of antibodies to hepatitis B e antigen with the Elecsys Anti-HBe using samples that would routinely be tested for hepatitis in the US. Data from this clinical study were the basis for the PMA approval decision. A summary of the clinical study is presented below.
A. Study Design
A multi-site clinical study was conducted to evaluate the clinical performance of the Elecsys Anti-HBe on samples that would routinely be tested for hepatitis and samples that were selected from individuals that were diagnosed with acute or chronic Hepatitis B infection.
The clinical agreement study involved the testing of 1800 samples (1500 adult specimens were prospectively collected from individuals at increased risk and 300 supplemental cohort specimens). The 1500 specimens were collected from seven collection sites located in Los Angeles, CA (32.0%), San Antonio, TX (24.9%), Baltimore, MD (17.9%), Miami, FL (7.9%), City of Industry, CA (6.9%), Darby, PA (6.8%), and Minneapolis, MN (3.6%). The 300 supplemental specimens were obtained from three external vendors. Testing of the specimens was performed at three clinical sites located in South Bend, IN, Iowa City, IA, and San Diego, CA.
B. Accountability of PMA Cohort
The clinical agreement study involved the testing of 1800 samples (1500 prospectively collected and 300 supplemental) on six (6) FDA approved reference assays, each detecting a unique serological marker (HBsAg, HBeAg, Anti-HBs, Anti-HBc, Anti-HBc
PMA P190005: FDA Summary of Safety and Effectiveness Data
14 of 27
{14}
IgM, and Anti-HBe) in order to determine the HBV classification for each of the samples tested.
The following table shows the different HBV specimen classifications.
Table 12: Serological Classification by FDA-Approved HBV Panel
| | HBsAg | HBeAg | Anti-HBc IgM | Anti-HBc | Anti-HBe | Anti-HBs |
| --- | --- | --- | --- | --- | --- | --- |
| Acute | (+) | (+) | (+) | (+) | (-), (+) | (-) |
| Acute | (+) | (+) | (-), (+) | (-) | (-) | (-) |
| Acute | (+) | (-) | (-) | (-) | (-) | (-) |
| Acute | (+) | (+) | (eq) | (+) | (-), (+) | (-) |
| Acute | (+) | (-) | (+) | (+) | (-) | (-) |
| Acute | (+) | (-) | (eq) | (+) | (+) | (-) |
| Acute (late) | (+) | (-) | (+) | (+) | (+) | (-), (+) |
| Chronic | (+) | (+) | (+) | (+) | (+) | (+) |
| Chronic | (+) | (-) | (-) | (+) | (+) | (-), (+) |
| Chronic | (+) | (-) | (-) | (+) | (eq) | (-) |
| Chronic | (+) | (-) | (-) | (+) | (-) | (-), (+) |
| Chronic | (+) | (+) | (+) | (+) | (-) | (+) |
| Chronic | (+) | (+) | (-) | (+) | (-) | (-), (+) |
| Chronic | (+) | (+) | (-) | (+) | (+) | (-) |
| Early recovery | (-) | (-) | (-) | (+) | (-), (+) | (-) |
| Early recovery | (-) | (-) | (+) | (+) | (-) | (-), (+) |
| Early recovery | (-) | (-) | (+) | (+) | (+) | (-), (+) |
| Recovery | (-) | (-) | (-) | (-), (+) | (+) | (+) |
| Recovery | (-) | (-) | (-) | (+) | (+) | (eq) |
| Recovered or immune due to natural infection | (-) | (-) | (-) | (+) | (-) | (+), (eq) |
| HBV vaccine response | (-) | (-) | (-) | (-) | (-) | (+) |
| HBV vaccination response | (-) | (-) | (-) | (-) | (-) | (eq) |
| Not previously infected | (-) | (-) | (-) | (-) | (-) | (-) |
| Not interpretable | (-) | (+) | (-) | (+) | (-) | (+) |
| Not interpretable | (-) | (-) | (-) | (-) | (+) | (-) |
PMA P190005: FDA Summary of Safety and Effectiveness Data
15 of 27
{15}
PMA P190005: FDA Summary of Safety and Effectiveness Data
16 of 27
| | HBsAg | HBeAg | Anti-HBc IgM | Anti-HBc | Anti-HBe | Anti-HBs |
| --- | --- | --- | --- | --- | --- | --- |
| Not interpretable | (-) | (+) | (-) | (+) | (+) | (-) |
| Not interpretable | (-) | (+) | (-) | (-) | (-) | (-), (eq), (+) |
## C. Study Population Demographics and Baseline Parameters
The demographics of the study population are typical for an Anti-HBe detection study performed in the US. The following tables show the demographics for the adult at increased risk (AIR) and supplemental study cohorts.
Table 13: Demographics of Clinical Population by Gender
| Cohort | Gender | N | % of cohort total |
| --- | --- | --- | --- |
| Adult AIR | Female | 686 | 45.7 |
| | Male | 814 | 54.3 |
| | Unknown | 0 | 0.0 |
| | Subtotal | 1500 | 100 |
| Supplemental | Female | 117 | 39.0 |
| | Male | 182 | 60.7 |
| | Unknown | 1 | 0.3 |
| | Subtotal | 300 | 100 |
| Total | | 1800 | |
Table 14: Demographics of Clinical Population by Ethnicity
| Cohort | Ethnicity | N | % of cohort total |
| --- | --- | --- | --- |
| Adult AIR | Hispanic or Latino | 421 | 28.1 |
| | Not Hispanic or Latino | 1073 | 71.5 |
| | Unknown | 6 | 0.4 |
| | Subtotal | 1500 | 100 |
| Supplemental | Hispanic or Latino | 2 | 0.7 |
| | Not Hispanic or Latino | 0 | 0.0 |
| | Unknown | 298 | 99.3 |
| | Subtotal | 300 | 100 |
| Total | | 1800 | |
{16}
Table 15: Demographics of Clinical Population by Race
| Cohort | Race | N | % of cohort total |
| --- | --- | --- | --- |
| Adult AIR | AIAN* | 17 | 1.1 |
| | Asian | 9 | 0.6 |
| | Black or African American | 735 | 49.0 |
| | Caucasian / White | 709 | 47.3 |
| | More than one race | 21 | 1.4 |
| | Nhopi | 4 | 0.3 |
| | Other | 1 | 0.1 |
| | Unknown | 4 | 0.3 |
| | Subtotal | 1500 | 100 |
| Supplemental | AIAN | 0 | 0.0 |
| | Asian | 166 | 55.3 |
| | Black or African American | 114 | 38.0 |
| | Caucasian / White | 15 | 5.0 |
| | More than one race | 0 | 0.0 |
| | Nhopi | 0 | 0.0 |
| | Other | 0 | 0.0 |
| | Unknown | 5 | 1.7 |
| | Subtotal | 300 | 100 |
| Total | | 1800 | |
*AIAN-American Indian and Alaskan Native
Table 16: Demographics of Clinical Population by Age
| Cohort | Age group (years) | N | % of cohort total |
| --- | --- | --- | --- |
| Adult AIR | 2-11 | 0 | 0.0 |
| | 12-21 | 0 | 0.0 |
| | 22-29 | 176 | 11.7 |
| | 30-39 | 253 | 16.9 |
| | 40-49 | 435 | 29.0 |
| | 50-59 | 481 | 32.1 |
| | 60-69 | 140 | 9.3 |
| | 70-79 | 12 | 0.8 |
PMA P190005: FDA Summary of Safety and Effectiveness Data
{17}
| Cohort | Age group (years) | N | % of cohort total |
| --- | --- | --- | --- |
| | ≥ 80 | 3 | 0.2 |
| | Unknown | 0 | 0.0 |
| | Subtotal | 1500 | 100 |
| Supplemental | 2-11 | 0 | 0.0 |
| | 12-21 | 0 | 0.0 |
| | 22-29 | 73 | 24.3 |
| | 30-39 | 80 | 26.7 |
| | 40-49 | 70 | 23.3 |
| | 50-59 | 50 | 16.7 |
| | 60-69 | 26 | 8.7 |
| | 70-79 | 1 | 0.3 |
| | ≥ 80 | 0 | 0.0 |
| | Unknown | 0 | 0.0 |
| | Subtotal | 300 | 100 |
| Total | | 1800 | |
## D. Safety and Effectiveness Results
### 1. Safety Results
With regard to safety, as an in vitro diagnostic test, the Elecsys Anti-HBe test involves taking a sample of plasma or serum from a patient. The test therefore presents no more safety hazard to an individual being tested than other tests where blood samples are drawn.
There were no adverse effects that occurred in the PMA clinical study.
### 2. Effectiveness Results
The analysis of effectiveness was based on 1800 evaluable patients. Key effectiveness outcomes are presented in the tables below.
Specimens were tested using the Elecsys Anti-HBe assay on the cobas e 602 immunoassay analyzer and a FDA-approved reference assay to establish the clinical performance characteristics. The following tables compare the Elecsys Anti-HBe results and performance with the results and performance obtained on
PMA P190005: FDA Summary of Safety and Effectiveness Data
{18}
an FDA-approved anti-HBe reference assay by HBV disease classification for the adult at increased risk cohort.
Table 17: Results comparing Elecsys Anti-HBe to the FDA-Approved Reference Assay by HBV Classification for Adults at Increased Risk Cohort
| HBV classification | FDA-approved reference assay | | | | Total |
| --- | --- | --- | --- | --- | --- |
| | Reactive | | Non-reactive | | |
| | Elecsys Anti-HBe | | Elecsys Anti-HBe | | |
| | RX^{1)} | NR^{2)} | RX | NR | |
| | n | n | n | n | |
| Acute | 0 | 0 | 0 | 7 | 7 |
| Chronic | 25 | 0 | 0 | 7 | 32 |
| Early recovery | 40 | 2 | 16 | 139 | 197 |
| Not previously infected | 0 | 0 | 1 | 554 | 555 |
| Recovered | 0 | 0 | 47 | 195 | 242 |
| Recovery | 129 | 3 | 0 | 0 | 132 |
| Vaccination | 0 | 0 | 0 | 335 | 335 |
| Total | 194 | 5 | 64 | 1237 | 1500 |
1) RX = reactive
2) NR = non-reactive
Table 18: Percent Agreement between Elecsys Anti-HBe and FDA-Approved Reference Assay by HBV Classification for Adults at Increased Risk Cohort
| | PPA^{3)} | | NPA^{4)} | |
| --- | --- | --- | --- | --- |
| HBV classification | % (n/N) | 95 % Exact CI^{5)} | % (n/N) | 95 % Exact CI |
| Acute | N/A (0/0) | N/A | 100 (7/7) | 59.0 to 100 |
| Chronic | 100 (25/25) | 86.3 to 100 | 100 (7/7) | 59.0 to 100 |
| Early recovery | 95.2 (40/42) | 83.8 to 99.4 | 89.7 (139/155) | 83.8 to 94.0 |
| Not previously infected | N/A (0/0) | N/A | 99.8 (554/555) | 99.0 to 100 |
| Recovered | N/A (0/0) | N/A | 80.6 (195/242) | 75.0 to 85.4 |
| Recovery | 97.7 (129/132) | 93.5 to 99.5 | N/A (0/0) | N/A |
| Vaccination | N/A (0/0) | N/A | 100 (335/335) | 98.9 to 100 |
PMA P190005: FDA Summary of Safety and Effectiveness Data
19 of 27
{19}
| | PPA3) | | NPA4) | |
| --- | --- | --- | --- | --- |
| HBV classification | % (n/N) | 95 % Exact CI5) | % (n/N) | 95 % Exact CI |
| Total | 97.5 (194/199) | 94.2 to 99.2 | 95.1 (1237/1301) | 93.8 to 96.2 |
3) PPA = positive percent agreement
4) NPA = negative percent agreement
5) CI = confidence interval
Table 19: Results comparing Elecsys Anti-HBe to the FDA-Approved Reference Assay by HBV Classification for Supplemental Cohort
| HBV classification | FDA-approved reference assay | | | | Total |
| --- | --- | --- | --- | --- | --- |
| | Reactive | | Non-reactive | | |
| | Elecsys Anti-HBe | | Elecsys Anti-HBe | | |
| | RX | NR | RX | NR | |
| | n | n | n | n | |
| Acute | 7 | 0 | 97 | 24 | 128 |
| Chronic | 45 | 0 | 10 | 117 | 172 |
| Total | 52 | 0 | 107 | 141 | 300 |
Table 20: Percent Agreement between Elecsys Anti-HBe and FDA-Approved Reference Assay by HBV Classification for Supplemental Cohort
| | PPA | | NPA | |
| --- | --- | --- | --- | --- |
| HBV classification | % (n/N) | 95 % Exact CI | % (n/N) | 95 % Exact CI |
| Acute | 100 (7/7) | 59.0 to 100 | 19.8 (24/121) | 13.1 to 28.1 |
| Chronic | 100 (45/45) | 92.1 to 100 | 92.1 (117/127) | 86.0 to 96.2 |
| Total | 100 (52/52) | 93.2 to 100 | 56.9 (141/248) | 50.4 to 63.1 |
In order to demonstrate the presence or absence of anti-HBe antibody activity, an internal confirmatory assay was developed. The confirmatory principle used HBeAg covalently coupled to sepharose to bind to anti-HBe antibodies in the sample. Depletion of Anti-HBe antibodies in the sample distinguished a confirmed reactive or positive result from a false-positive or unconfirmed reactive result. After testing was completed at the three external sites, samples were transferred to Roche for storage. An internal study was then conducted testing all Elecsys Anti-HBe reactive samples (N=421) and approximately 1/3 of the concordant non-reactive samples (N=476). The concordant non-reactive samples were randomly selected for testing from the remaining samples with enough sample volume to complete testing.
PMA P190005: FDA Summary of Safety and Effectiveness Data
{20}
The following algorithm was employed to obtain a composite method result for the samples evaluated with the confirmatory test.
Table 21: Composite Method Algorithm
| Confirmatory assay result | FDA-approved Reference assay result | Composite result interpretation |
| --- | --- | --- |
| Confirmed non-reactive | Non-reactive | Non-reactive |
| Confirmed non-reactive | Reactive | Reactive |
| Confirmed positive | Non-reactive | Reactive |
| Confirmed positive | Reactive | Reactive |
False positive confirmatory assay results were considered non-reactive for the purposes of this analysis. Because only a subset of the clinical samples were tested by the confirmatory assay, a verification bias analysis was employed to impute the results for those samples that were not tested by confirmatory assay. The following results were obtained for the adult cohort and supplemental cohort.
Table 22: Results Comparing the Elecsys Anti-HBe with Composite Method Algorithm For the Adult at Increased Risk Cohort
| HBV Classification | Composite Comparator | | | | Total | PPA (95% CI, Wilson Score) | NPA (95% CI, Wilson Score) |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | Reactive | | Non-reactive | | | | |
| | Elecsys Anti-HBe | | Elecsys Anti-HBe | | | | |
| | Reactive | Non-reactive | Reactive | Non-reactive | | | |
| Acute | | | | 7 | 7 | | 100% (7/7) |
| Chronic | 25 | | | 7 | 32 | 100% (25/25) | 100% (7/7) |
| Early Recovery | 56 | 10 | 0 | 131 | 197 | 84.8% (56/66) | 100% (131/131) |
| Not previously infected | | | | 554 | 554 | | 100% (554/554) |
| Recovered | 1 | | 1 | 155 | 157 | 100% (1/1) | 99.4% (155/156) |
| Recovery | 174 | 43 | | | 217 | 80.2% (174/217) | |
| Vaccination | | | | 335 | 335 | | 100% (335/335) |
| Not Interpretable | 1 | | | | 1 | | |
| Total | 257 | 53 | 1 | 1189 | 1500 | 82.9% (257/310) | 99.9% (1189/1190) |
PMA P190005: FDA Summary of Safety and Effectiveness Data
21 of 27
{21}
Table 23: Results Comparing the Elecsys Anti-HBe with Composite Method Algorithm For the Supplemental Cohort
| HBV Classification | Composite Comparator | | | | Total | PPA (95% CI, Wilson Score) | NPA (95% CI, Wilson Score) |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | Reactive | | Non-reactive | | | | |
| | Elecsys Anti-HBe | | Elecsys Anti-HBe | | | | |
| | Reactive | Non-reactive | Reactive | Non- reactive | | | |
| Acute | 103 | 2 | 2 | 21 | 128 | 98.1% (103/105) | 91.3% (21/23) |
| Chronic | 54 | 4 | 1 | 113 | 172 | 93.1% (54/58) | 99.1% (113/114) |
| Total | 157 | 6 | 3 | 134 | 300 | 96.3% (157/163) | 97.8% (134/137) |
The following section provides further analysis comparing the results of the FDA approved reference assay and the Elecsys Anti-HBe to the Roche Confirmatory assay.
Table 24: Results Comparing the FDA Approved Reference Assay Versus Confirmatory Assay
| FDA-approved reference assay | Confirmatory Reactive | Confirmatory Non-reactive | Total |
| --- | --- | --- | --- |
| Positive | 197 | 2 | 199 |
| Negative | 211 | 1090 | 1301 |
| Total | 40 | 1092 | 1500 |
Table 25: Results Comparing the Elecsys Anti-HBe Versus Confirmatory Assay
| Elecsys Anti-HBe | Confirmatory Reactive | Confirmatory Non-reactive | Total |
| --- | --- | --- | --- |
| Positive | 255 | 3 | 258 |
| Negative | 50 | 1192 | 1242 |
| Total | 305 | 1195 | 1500 |
## Adults at Increased Risk
A summary of performance, including percent agreements, of the Elecsys Ant-HBe and the FDA-approved reference assay compared to Confirmatory approach for the adult at increased risk cohort are presented in the table below.
PMA P190005: FDA Summary of Safety and Effectiveness Data
{22}
Table 26: Summary of Performance of Elecsys Anti-HBe and the FDA-Approved Assay Compared to the Confirmatory Assay for Adults at Increased Risk
| | Positive Percent Agreement | 95% CI | Negative Percent Agreement | 95% CI |
| --- | --- | --- | --- | --- |
| Elecsys Anti-HBe | 83.6 % | 79.0 – 87.3 % | 99.7 % | 99.3 – 99.9 % |
| FDA-approved reference assay | 48.3 % | 43.5 – 53.1 % | 99.8 % | 99.3 – 99.9 % |
## Supplemental Cohort
A summary of performance, including percent agreements, of the Elecsys Ant-HBe and the FDA-approved reference assay compared to Confirmatory approach for the supplemental cohort are presented in the table below.
Table 27: Summary of Results of Elecsys Anti-HBe and the FDA-Approved assay Compared to the Confirmatory Assay for Supplemental Cohort
| FDA-approved reference assay | Confirmatory Confirmed Reactive | Confirmatory Confirmed Non-reactive | Total |
| --- | --- | --- | --- |
| Positive | 50 | 2 | 52 |
| Negative | 143 | 105 | 248 |
| Total | 193 | 107 | 300 |
Table 28: Summary of Performance of Elecsys Anti-HBe and the FDA-Approved assay Compared to the Confirmatory Assay for Supplemental Cohort
| Elecsys Anti-HBe | Confirmatory Confirmed Reactive | Confirmatory Confirmed Non-reactive | Total |
| --- | --- | --- | --- |
| Positive | 144 | 15 | 159 |
| Negative | 7 | 134 | 141 |
| Total | 151 | 149 | 300 |
Table 29: Summary of Performance of Elecsys Anti-HBe and the FDA-Approved assay Compared to the Confirmatory Assay for Supplemental Cohort
| | Positive Percent Agreement | 95% CI | Negative Percent Agreement | 95% CI |
| --- | --- | --- | --- | --- |
| Elecsys Anti-HBe | 95.4 % | 90.7 – 97.7 % | 89.9 % | 84.1 – 93.8 % |
| FDA-approved reference assay | 25.9 % | 20.2 – 32.5 % | 98.1 % | 93.4 – 99.5 % |
## Overall (Combined AIR and Supplemental Cohort) Results
The overall analysis of the study results completed between the FDA approved assay and the Confirmatory approach are summarized in the following table.
PMA P190005: FDA Summary of Safety and Effectiveness Data
23 of 27
{23}
Table 30: Summary of Results of Elecsys Anti-HBe and the FDA-Approved assay Compared to the Confirmatory Assay Overall
| FDA-approved reference assay | Confirmatory Confirmed Reactive | Confirmatory Confirmed Non-reactive | Total |
| --- | --- | --- | --- |
| Positive | 247 | 4 | 251 |
| Negative | 415 | 1134 | 1549 |
| Total | 662 | 1138 | 1800 |
Table 31: Summary of Performance of Elecsys Anti-HBe and the FDA-Approved assay Compared to the Confirmatory Assay Overall
| Elecsys Anti-HBe | Confirmatory Confirmed Reactive | Confirmatory Confirmed Non-reactive | Total |
| --- | --- | --- | --- |
| Positive | 399 | 18 | 417 |
| Negative | 57 | 1326 | 1383 |
| Total | 456 | 1344 | 1800 |
Table 32: Summary of Performance of Elecsys Anti-HBe and the FDA-Approved assay Compared to the Confirmatory Assay Overall
| | Positive Percent Agreement | 95% CI | Negative Percent Agreement | 95% CI |
| --- | --- | --- | --- | --- |
| Elecsys Anti-HBe | 87.5 % | 84.1 – 90.2 % | 98.7 % | 97.9 – 99.2 % |
| FDA-approved reference assay | 37.3 % | 33.7 – 41.1 % | 99.6 % | 99.1 – 99.9 % |
3. Subgroup Analyses
The study design enabled an assessment of assay performance by subgroup as depicted in the tables in Section X.D.2 above which show subjects stratified by HBV classification.
4. Pediatric Extrapolation
In this premarket application, existing clinical data was not leveraged to support approval of a pediatric patient population.
E. Financial Disclosure
The Financial Disclosure by Clinical Investigators regulation (21 CFR 54) requires applicants who submit a marketing application to include certain information concerning the compensation to, and financial interests and arrangement of, any clinical investigator conducting clinical studies covered by the regulation. The pivotal clinical study included 11 investigators. None of the clinical investigators had disclosable financial interests/arrangements as defined in sections 54.2(a), (b), (c),
PMA P190005: FDA Summary of Safety and Effectiveness Data
24 of 27
{24}
and (f). The information provided does not raise any questions about the reliability of the data.
## XI. PANEL MEETING RECOMMENDATION AND FDA'S POST-PANEL ACTION
In accordance with the provisions of section 515(c)(3) of the act as amended by the Safe Medical Devices Act of 1990, this PMA was not referred to the Microbiology Panel, an FDA advisory committee, for review and recommendation because the information in the PMA substantially duplicates information previously reviewed by this panel.
## XII. CONCLUSIONS DRAWN FROM PRECLINICAL AND CLINICAL STUDIES
### A. Effectiveness Conclusions
The effectiveness of the Elecsys Anti-HBe for the qualitative detection of antibodies to hepatitis B e antigen in human serum and plasma (potassium EDTA, lithium heparin, sodium citrate, sodium heparin) samples is supported by the clinical study results. The results of this test may be used as an aid in the diagnosis of HBV infection in patients with symptoms of hepatitis. See Section X.D.2 for Effectiveness Results.
### B. Safety Conclusions
The risks of the device are based on nonclinical laboratory studies as well as data collected in a clinical study conducted to support PMA approval as described above. Based on the results of these studies, the Elecsys Anti-HBe when used according to the manufacturer's instructions can aid the physician in the diagnosis of HBV infection.
### C. Benefit-Risk Determination
The probable benefits of the device are also based on data collected in the clinical study conducted to support PMA approval as described above. The benefits of the assay are as part of a hepatitis B panel, the appropriate determination of HBV seroconversion as part of disease management and treatment. Treatment for appropriate patients can mitigate the sequelae of hepatitis B infection and may result in improved morbidity and mortality in these patients. Known sequelae of hepatitis B infection include continued symptoms, increases in all-cause mortality, liver disease-related complications and death, hepato-cellular carcinoma rates, and need for liver transplantation. Additionally, management and appropriate treatment for hepatitis B infection can potentially decrease transmission and disease burden in the general population and particularly in populations at high risk for hepatitis B infection. While the performance of the device in the clinical study suggests that patients will benefit from the assay, low prevalence of certain HBV classifications is a source of potential uncertainty when analyzing the samples. The wide confidence intervals for those subgroups is expected due to the biology of hepatitis B infection and is acceptable
PMA P190005: FDA Summary of Safety and Effectiveness Data
25 of 27
{25}
The risks associated with the device, when used as intended, are those related to the risk of false test results, failure to correctly interpret the test results and failure to correctly operate the instrument.
Risks of a false positive test includes improper patient management, including premature discontinuation of antiviral treatment should a clinician be falsely led to determine a patient has seroconverted. This risk is mitigated by the fact that this assay is usually repeated and is used as part of a panel. Repeatedly false positive results have the potential to lead to inappropriate treatment decisions, however anti-HBe is not used in isolation to determine seroconversion status. Because anti-HBe is sometimes included as part of a panel in clinical practice to diagnose hepatitis B but is more commonly utilized as part of a panel to determine chronicity of disease and immune-active v. chronic infection, the risk of a false positive will likely be somewhat mitigated as incongruous test results would lead a clinician to either retest the patient or further investigate the etiology of hepatitis.
Risk of a false negative test includes improper patient management, including continued treatment for hepatitis B with antiviral medication. Antiviral medication has risks including toxicity and more rarely allergic reactions. Over time, viral resistance in patients who are co-infected but undiagnosed with other viruses using the same antiviral medication, such as HIV, can lead to viral resistance, however the chance of an undiagnosed co-infection in a patient tested for hepatitis B is exceedingly unlikely. Anti-HBe is not used in this same manner as HBeAg to guide treatment decisions, and thus confers less clinical risk to this subpopulation than a false negative or false positive HBeAg test. Anti-HBe is not used in isolation to determine seroconversion status. Because anti-HBe is sometimes included as part of a panel in clinical practice to diagnose hepatitis B but is more commonly utilized as part of a panel to determine chronicity of disease and immune-active v. chronic infection, the risk of a false negative will likely be mitigated as incongruous test results would lead a clinician to either retest the patient or further investigate the etiology of hepatitis.
1. Patient Perspective
This submission either did not include specific information on patient perspectives or the information did not serve as part of the basis of the decision to approve or deny the PMA for this device.
In conclusion, given the available information above, the data support that for the claimed intended use and the probable benefits outweigh the probable risks.
D. Overall Conclusions
The data in this application support the reasonable assurance of safety and effectiveness of this device when used in accordance with the indications for use. The probable clinical benefits outweigh the potential risks for the proposed assay
PMA P190005: FDA Summary of Safety and Effectiveness Data
26 of 27
{26}
considering the performance of the device in the clinical study and the low risk and associated risk mitigations in clinical practice. The proposed assay labeling will facilitate accurate assay implementation and interpretation of results. The clinical performance observed suggests that errors will be uncommon and that the assay may provide substantial benefits to patients as an accurate and sensitive aid in the diagnosis of HBV infection when used in conjunction with other laboratory results and clinical information.
## XIII. CDRH DECISION
CDRH issued an approval order on February 3, 2021.
The applicant’s manufacturing facilities have been inspected and found to be in compliance with the device Quality System (QS) regulation (21 CFR 820).
## XIV. APPROVAL SPECIFICATIONS
Directions for use: See device labeling.
Hazards to Health from Use of the Device: See Indications, Contraindications, Warnings, Precautions, and Adverse Events in the device labeling.
Post-approval Requirements and Restrictions: See approval order.
PMA P190005: FDA Summary of Safety and Effectiveness Data
27 of 27
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.