Low-Profile Visualized Intraluminal Support (LVIS) and LVIS Jr.

P170013 · MicroVention, Inc. · QCA · May 30, 2018 · Neurology

Device Facts

Record IDP170013
Device NameLow-Profile Visualized Intraluminal Support (LVIS) and LVIS Jr.
ApplicantMicroVention, Inc.
Product CodeQCA · Neurology
Decision DateMay 30, 2018
DecisionAPPR
Device ClassClass 3
AttributesTherapeutic, Real-World Evidence

Real-World Evidence

SubmissionDeviceSponsorRWD SourcesRWE Use SummaryKey Tags
P170013 · May 30, 2018Low-Profile Visualized Intraluminal Support (LVIS) and LVIS Jr.MicroVention, Inc.Published clinical literature (Bondi et al. 2007; Fiorella et al.; Maldonado et al.; Yahia et al.; Sedat et al. 2009)The sponsor used retrospective data from five published studies of HDE-approved neurovascular stents to derive performance goals (PGs) for safety and effectiveness, which served as the benchmarks for the pivotal clinical study's Bayesian statistical analysis.Performance Goals; Bayesian Analysis; Literature Review; Historical Control

Clinical Evidence

Study DesignPopulationComparatorKey Endpoints
Performance Goal Derivation Studies; Retrospective literature review of five published studies514 patients treated with HDE-approved Neuroform Stent Systems; Sample Size: 514Not applicable for this studyMajor stroke or neurological death rate; aneurysm occlusion rates

Indications for Use

The LVIS and LVIS Jr. are indicated for use with neurovascular embolization coils in patients ≥ 18 years of age for the treatment of wide-neck (neck width ≥ 4 mm or dome to neck ratio < 2) saccular intracranial aneurysms arising from a parent vessel with a diameter ≥ 2.0 mm and ≤ 4.5 mm.

Device Story

Self-expanding nitinol stent with radiopaque tantalum markers; delivered via microcatheter using detachable push wire. Used in neurovascular procedures to support embolization coils in wide-neck saccular intracranial aneurysms. Physician navigates delivery system through tortuous intracranial anatomy under fluoroscopy; stent deploys across aneurysm neck to provide structural support for coils. Stent promotes aneurysm occlusion by diverting flow and aiding coil stability. Benefits include reduced aneurysm rupture risk and improved occlusion rates in complex wide-neck anatomy. Clinical decision-making relies on angiographic assessment of aneurysm morphology and parent vessel diameter.

Clinical Evidence

Prospective, multi-center, single-arm, open-label pivotal study (IDE #G110188) of 171 patients (153 ITT). Primary safety endpoint: disabling stroke or neurological death within 12 months (observed 5.9%). Primary effectiveness endpoint: 100% aneurysm occlusion without significant in-stent stenosis or retreatment at 12 months (70.6% success; 81.0% composite success including stable Raymond-Roy II). Study used Bayesian statistical methods and compared results to performance goals derived from prior HDE-approved stent data.

Technological Characteristics

Self-expanding nickel-titanium (nitinol) braid; closed-cell tubular design with flared ends. Includes tantalum radiopaque wires and markers. LVIS compatible with 0.021" ID microcatheter; LVIS Jr. with 0.017" ID. MRI Conditional per ASTM2503. Sterilized via electron beam irradiation. Single-use, non-pyrogenic.

Indications for Use

Indicated for patients ≥ 18 years with wide-neck (neck ≥ 4 mm or dome-to-neck ratio < 2) saccular intracranial aneurysms arising from parent vessels 2.0–4.5 mm in diameter. Contraindicated in patients with contraindications to anti-platelet/anticoagulant therapy, hypersensitivity to nickel-titanium, active infection, or pre-existing stents at the target site.

Regulatory Classification

Identification

An intracranial coil-assist stent is a tubular device placed in an artery in the brain to assist in the placement and retention of embolic coils for the treatment of wide-neck intracranial aneurysms.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0} SUMMARY OF SAFETY AND EFFECTIVENESS DATA (SSED) I. GENERAL INFORMATION Device Generic Name: Intracranial Neurovascular Stent Device Trade Name: Low-Profile Visualized Intraluminal Support (LVIS) and LVIS Jr. Device Procode: QCA Applicant's Name and Address: MicroVention, Inc. 1311 Valencia Avenue Tustin, CA 92780 Date(s) of Panel Recommendation: None Premarket Approval Application (PMA) Number: P170013 Date of FDA Notice of Approval: May 30, 2018 II. INDICATIONS FOR USE The LVIS and LVIS Jr. are indicated for use with neurovascular embolization coils in patients ≥ 18 years of age for the treatment of wide-neck (neck width ≥ 4 mm or dome to neck ratio &lt; 2) saccular intracranial aneurysms arising from a parent vessel with a diameter ≥ 2.0 mm and ≤ 4.5 mm. III. CONTRAINDICATIONS The use of the LVIS and LVIS Jr. device are contraindicated under these circumstances: - Patients in whom anticoagulant, anti-platelet therapy or thrombolytic drugs are contraindicated. - Patients with known hypersensitivity to metal, such as nickel-titanium and metal jewelry. - Patients with anatomy that does not permit passage or deployment of the LVIS/LVIS Jr. device. - Patients with an active bacterial infection. - Patients with a pre-existing stent in place at the target aneurysm. IV. WARNINGS AND PRECAUTIONS The warnings and precautions can be found in the LVIS and LVIS Jr. labeling. PMA P170013: FDA Summary of Safety and Effectiveness Data Page 1 {1} # V. DEVICE DESCRIPTION The LVIS and LVIS Jr. consists of: 1) A self-expanding nickel titanium (nitinol) stent with radiopaque markers; and 2) A delivery system consisting of an introducer sheath and a detachable push wire. The self-expanding nitinol stent consists of a single wire braid, compliant, closed-cell tubular design with flared ends. The nitinol stent contains radiopaque wires (tantalum) that are woven into the implant in a helical configuration and radiopaque tantalum coil markers at each end to aid in stent placement under fluoroscopy. The delivery system contains an introducer sheath that is used to protect the stent in the package and facilitates the introduction of the device into the microcatheter hub. The delivery push wire is attached to the stent and is designed to facilitate delivery of the stent to the site of implantation. On the proximal end of the delivery push wire, there are three radiopaque markers that indicate when the stent should exit the distal end of the microcatheter and aids in stent placement (see Figures 1 and 2). The LVIS are deliverable through a 0.021" inner diameter (ID) microcatheter whereas the LVIS Jr. are deliverable through a 0.017" ID microcatheter. The devices are provided sterile for single use only and are non-pyrogenic. ![img-0.jpeg](img-0.jpeg) Figure 1: Overall Device Diagram ![img-1.jpeg](img-1.jpeg) Figure 2: Device Implant and Delivery System PMA P170013: FDA Summary of Safety and Effectiveness Data {2} Table 1 shows the device sizes available for the LVIS and LVIS Jr. Table 1: Sizes of LVIS and LVIS Jr. | Device | Outer Diameter (mm) | Lengths (mm) | | --- | --- | --- | | LVIS | 3.5 | 17 and 22 | | | 4.0 | 12, 17, 22, 28, and 31 | | | 4.5 | 18, 23, and 32 | | | 5.5 | 30 and 33 | | LVIS Jr. | 2.5 | 13, 17, 23, and 34 | | | 3.5 | 18, 23, 28, and 33 | # VI. ALTERNATIVE PRACTICES AND PROCEDURES There are several other alternatives for the treatment of wide-neck intracranial aneurysms including open surgical clipping, endovascular treatment using embolization coils supported with other Humanitarian Device Exemption (HDE) approved neurovascular stents or balloon catheter assisted coiling of the intracranial aneurysm, and flow diverters. The other neurovascular stents approved through the HDE regulatory pathway for stent-assisted coiling include the Stryker Neurovascular Neuroform EZ, 3, and Atlas Stent Systems (H020002) and the Codman &amp; Shurtleff, Inc. Enterprise Vascular Reconstruction Device and Delivery System (H060001). A similar HDE approved device that is indicated to support neurovascular embolization coils specifically for the treatment of unruptured wide-necked intracranial aneurysms originating on or near a vessel bifurcation of the basilar tip or carotid terminus is the Pulsar Vascular, Inc. PulseRider Aneurysm Neck Reconstruction Device (H160002). The Micro Therapeutics, Inc. d/b/a ev3 Neurovascular Pipeline Embolization Device (PED) (P100018) is the only approved flow diverter in the United States (US) and was approved with the intended use of endovascular treatment of large or giant wide-necked intracranial aneurysms in the internal carotid artery (ICA) from the petrous to the superior hypophyseal segment. The flow diverter is implanted in the parent vessel and is placed across the neck of the intracranial aneurysm. Its mechanism of action is to divert the blood flow from entering the intracranial aneurysm sac and endothelialization will occur on the implant over time to further promote complete intracranial aneurysm occlusion. The flow diverter is intended to be used by itself as a stand-alone device. In addition to these alternative treatments, certain intracranial aneurysms may be managed medically or by observation only with no treatment but with regular imaging follow-up examinations to ensure there are no morphological changes in the intracranial aneurysm(s) over time. Each alternative has its own advantages and disadvantages. A patient should fully discuss these alternatives with his/her physician to select the method that best meets expectations and lifestyle. PMA P170013: FDA Summary of Safety and Effectiveness Data {3} PMA P170013: FDA Summary of Safety and Effectiveness Data Page 4 # VII. MARKETING HISTORY The LVIS and LVIS Jr. are approved for marketing in the following countries: Argentina, Australia, Austria, Belarus, Belgium, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Cuba, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, Finland, France, Georgia, Germany, Honduras, Hong Kong, Hungary, Iceland, India, Iran, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Korea, Latvia, Lebanon, Libya, Lithuania, Malaysia, Morocco, Mexico, Mongolia, Netherlands, New Zealand, Norway, Pakistan, Panama, Paraguay, Peru, Poland, Portugal, Puerto Rico, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovenia, Spain, Sweden, Switzerland, Thailand, Tobago, Turkey, Ukraine, United Arab Emirates, United Kingdom, United States (under HDE H130005), Uruguay, Venezuela, Vietnam. The LVIS and LVIS Jr. have not been withdrawn from the market for safety or effectiveness reasons. # VIII. PROBABLE ADVERSE EFFECTS OF THE DEVICE ON HEALTH Below is a list of the probable adverse effects (e.g., complications) associated with the use of the device. - Allergic reaction, including but not limited to: contrast dye, nitinol metal, and any other medications used during the procedure - Aphasia - Blindness - Cardiac arrhythmia - Coil prolapsed or migration into normal vessel adjacent to intracranial aneurysm - Complications of arterial puncture including pain, local bleeding, local infection and injury to the artery, vein or adjacent nerves - Cranial neuropathy - Death - Device fracture, migration or misplacement - Dissection or perforation of the parent artery - Headache - Hemorrhage (i.e., intracranial hemorrhage (ICH), subarachnoid hemorrhage (SAH), retroperitoneal (or in other locations)) - Hemiplegia - Hydrocephalus - Infection - Injury to normal vessel or tissue - Ischemia - Mass effect - Myocardial infarction {4} - Neurological deficits - Occlusion of non-target side branches - Pseudo aneurysm formation - Reactions to anti-platelet/anti-coagulant agents - Reactions due to radiation exposure (i.e., alopecia, burns ranging in severity from skin reddening to ulcers, cataracts, delayed neoplasia) - Reactions to anesthesia and related procedures - Reactions to contrast agents - Renal failure - Aneurysm rupture - Stenosis of stented segment - Seizure - Stent thrombosis - Stroke or TIA (transient ischemic attack) - Thromboembolic event - Vasospasm - Visual impairment For the specific adverse events that occurred in the clinical study, please see Section X below. ## IX. SUMMARY OF NONCLINICAL STUDIES The LVIS and LVIS Jr. underwent non-clinical mechanical, functional, biocompatibility, sterilization validation, bacterial endotoxin, and animal testing to support the proposed intended use. ## A. Laboratory Studies ### Design Verification and Validation Testing Table 2 shows the design verification and validation testing conducted on the LVIS and LVIS Jr. Much of the testing was conducted based on the recommendations in the Guidance for Industry and FDA Staff, "Non-Clinical Engineering Tests and Recommended Labeling for Intravascular Stents and Associated Delivery Systems," issued on April 18, 2010. PMA P170013: FDA Summary of Safety and Effectiveness Data Page 5 {5} Table 2: Design Verification and Validation Testing | Test | Purpose | Results | | --- | --- | --- | | Material Composition | Verify if materials are suitable for implant. | Materials met pre-specified requirements. | | Austenite Finish Transition Temperature (At) | The implant will achieve its predetermined size and shape when exposed to normal body temperature. | The At for the LVIS and LVIS Jr. implant was confirmed to comply with the specification of < 30 °C. | | Pitting, Crevice, and Fretting Corrosion Potential | Determine the corrosion susceptibility of the stent. | No breakdown potential was reached. | | Galvanic Corrosion | Assess the galvanic corrosion susceptibility of two dissimilar metals (nitinol and tantalum). | No corrosion susceptibility was observed. | | Percent Surface Area | To determine the surface area of the LVIS/LVIS Jr. device. | Met pre-specified criteria. | | Foreshortening | To determine if the stent length shortens once deployed. | Met pre-specified criteria. | | Stent Integrity | Determine if there are any abnormalities on the LVIS/LVIS Jr. device. | All samples demonstrated the LVIS/LVIS Jr. device is able to deploy over the device shelf life of 3 years. | | Mechanical Properties | To specify the mechanical properties of incoming and post processing stent material. | The post processed nitinol wire is equivalent to the raw nitinol material. | | Stress/Strain and Fatigue Analysis | Fatigue resistance of the LVIS/LVIS Jr. stents were analyzed based on a 0.4% endurance limit. | No excessive localized stresses were detected. The finite element analysis (FEA) predicted the LVIS/LVIS Jr. fatigue lifetime has adequate safety built into the design with a safety factor of > 2.39. | | Accelerated Durability Testing | Following 380,000,000 cycles (10-yrs equivalent), the LVIS/LVIS Jr. device was tested for percent strain. | The LVIS/LVIS Jr. device met accelerated durability test specification with no loss in structural integrity or fragmentation. | | Particulate Evaluation | Quantify the particulate matter in injections for the LVIS/LVIS Jr. device under simulated use conditions in a neurovascular tortuosity model. | The particulate test was validated by demonstrating > 90% recovery for > 10 μm and > 25 μm particle size ranges. | PMA P170013: FDA Summary of Safety and Effectiveness Data {6} | Test | Purpose | Results | | --- | --- | --- | | Magnetic Resonance Imaging (MRI) Safety | To assess the MRI safety and compatibility of the LVIS/LVIS Jr. device. | The test results demonstrated the LVIS/LVIS Jr. device does not pose additional unacceptable risk to the patient. The LVIS/LVIS Jr. device was determined to be MRI Conditional per ASTM2503. The MRI Conditional scanning parameters are specified in the labeling. | | Radiopacity | Determine if the LVIS/LVIS Jr. device and delivery system are visible under fluoroscopy (imaging used during implantation). | The LVIS/LVIS Jr. device was verified to meet the acceptance criteria for radiopacity. | | Tensile Bond Strength | The tensile strength of the LVIS/LVIS Jr. implant is characterized at the: - Implant wire ends; - Proximal implant marker coil; and - Distal implant marker coil. | The LVIS/LVIS Jr. device met the tensile bond strength acceptance criteria at the bond locations. | | Dimensional Attributes – Delivery System and Stent | To confirm the delivery system and stent is within dimensional specifications. | The delivery systems and stents for the LVIS and LVIS Jr. met dimensional specifications. | | Radial Outward Force | The radial force of the LVIS/LVIS Jr. implant is characterized. | The LVIS/LVIS Jr. implant met the radial outward force acceptance criteria. | | Simulated Use Testing | Verify performance attributes in an in-vitro model simulating tortuous intracranial anatomy. - Compatibility with delivery catheter - Trackability through delivery catheter while advancing stent - Trackability through delivery catheter while retracting stent - Ease of stent deployment - Ease of stent detachment - Accuracy of stent placement - Wall apposition/conformance to vessel - Stent stability - Ability to cross/re-cross deployed stent with microcatheter | All devices tested passed the simulated use testing. | PMA P170013: FDA Summary of Safety and Effectiveness Data {7} PMA P170013: FDA Summary of Safety and Effectiveness Data Page 8 | Test | Purpose | Results | | --- | --- | --- | | | - Compatibility with ancillary microcatheter used for deploying embolic coils - Compatibility with embolic coils - Overall performance | | | Flexibility and Kink Test | The flexibility and kink test for the LVIS/LVIS Jr. were assessed during the “simulated use” testing of the device. During simulated use testing, the trackability of the devices is evaluated during advancement/retraction of the device in the microcatheter. The device including the delivery system needs to be flexible and kink resistant to be able to advance the device through tortuous anatomy. If not, the trackability of the device would be affected. | The trackability of the LVIS and LVIS Jr. were rated favorably during the simulated use testing and did not result in any kinks observed. | ## Biocompatibility Testing Biocompatibility testing for all materials used to manufacture the LVIS and LVIS Jr. were performed in accordance with ISO 10993-1:2009/(R)2013, Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing within a Risk Management Process. Tables 3 and 4 shows the biocompatibility tests conducted for the LVIS/LVIS Jr. implant and delivery system, respectively. Table 3: Implant - Biocompatibility | Test | Purpose | Results | | --- | --- | --- | | Cytotoxicity - L929 MEM Elution Test | Test for cell lysis. | No evidence of biological reactivity. | | Sensitization - Kligman Maximization Test | Test for allergenic potential or sensitization capacity of test article. | The test article extracts showed no reaction at the challenge. | | Irritation - Intracutaneous Injection Test in Rabbits | Test for irritation potential. | The test article sites did not show a significantly greater biological reaction than the sites injected with the control article. | | Systemic Toxicity - Systemic Injection Test Study in Mice | Test for systemic acute toxicity in mice following intravenous and intraperitoneal injections. | The test article did not induce a significantly greater biological reaction than the control extracts, when tested in mice. | | Systemic Toxicity - Rabbit Pyrogen Test (Material Mediated) | Test for pyrogenic response from the material. | The increases in temperature did not exceed the test limit for the maximum individual temperature rise. | | Hemocompatibility - | Test for red blood cell | The test article exhibited 0.29% hemolysis | {8} Table 4: Delivery System - Biocompatibility | Test | Purpose | Results | | --- | --- | --- | | Cytotoxicity - L929 MEM Elution Test | Test for cell lysis. | No evidence of biological reactivity. | | Sensitization - Kligman Maximization Test | Test for allergenic potential or sensitization capacity of test article. | The test article extracts showed no reaction at the challenge. | | Irritation - Intracutaneous Injection Test in Rabbits | Test for irritation potential. | The test article sites did not show a | | | | infection. | | Cell Death - 100000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000 | Test for cell lysis. | No evidence of biological reactivity. | PMA P170013: FDA Summary of Safety and Effectiveness Data {9} | Test | Purpose | Results | | --- | --- | --- | | | | significantly greater biological reaction than the sites injected with the control article. | | Systemic Toxicity - Systemic Injection Test Study in Mice | Test for systemic acute toxicity in mice following intravenous and intraperitoneal injections. | The test article did not induce a significantly greater biological reaction than the control extracts, when tested in mice. | | Systemic Toxicity - Rabbit Pyrogen Test (Material Mediated) | Test for pyrogenic response from the material. | The increases in temperature did not exceed the test limit for the maximum individual temperature rise. | | Hemocompatibility - ASTM Hemolysis (Indirect Contact) | Test for red blood cell hemolysis. | The test article exhibited 0.00% hemolysis above the level of hemolysis exhibited by the negative control via the indirect method. | | Hemocompatibility - Prothrombin Time Assay (Indirect Contact) | Test for coagulation response from the test material. | The mean value for the test article and each individual replicate value did not differ from the mean of the negative and untreated controls by more than one second. | | Hemocompatibility - Unactivated Partial Thromboplastin Time (UPTT) Assay (Indirect Contact) | Test to measure the ability to form blood clots. | The UPTT of the plasma exposed to the test article extract was not significantly decreased as compared to the untreated plasma and the plasma exposed to the negative control article. | | Hemocompatibility - Complement Activation Assay (C3a, SC5b-9) (Indirect Contact) | Test the potential for activation of the complement system. | The test article did not induce complement activation of C3 or C5 proteins in human plasma. | | Hemocompatibility - Thrombogenicity Study | Test to determine comparative thromboresistance of medical devices that are intended for blood contact. | No significant thrombosis was observed in the test sites and control sites. | ## Sterilization Validation The LVIS and LVIS Jr. are sterilized using electron beam irradiation. The sterilization method was validated to a sterility assurance level of $10^{-6}$ per ISO 11137-1:2006/(R)2010, Sterilization of Health Care Products – Radiation – Part 1: Requirements for Development, Validation, and Routine Control of a Sterilization Process for Medical Devices. The device was tested and met specifications for sterilization. ## Shelf Life The LVIS and LVIS Jr. were tested and determined to have a 3-year shelf-life. The 3-year shelf life was verified on accelerated aged devices. The samples were pre-conditioned for PMA P170013: FDA Summary of Safety and Effectiveness Data {10} simulated shipping and sterilization. The dimensional and functional attributes were tested and met acceptance criteria. In addition, packaging integrity testing (pouch) was verified and met acceptance criteria to support the 3-year shelf-life. ## B. Animal Studies A Good Laboratory Practice (GLP) animal study was conducted to evaluate acute and chronic safety and performance of the device at 0, 30, 90, and 180 days. The device was implanted in 20 rabbits in which elastase-induced aneurysms were created. The device performance characteristics during implantation were evaluated based upon a scaled scoring system used by the neurointerventionalists conducting the procedure. Prior to sacrifice, the animals were angiographically assessed for stent performance such as stability of the implant in the artery (absence of migration), parent vessel patency, embolization coil stability, blood flow or vessel irregularities, and aneurysm occlusion. Excised vessels were evaluated for histology – histopathology and vessel patency. At 180 days, minimal inflammation was observed. The results at all time points in the in vivo study demonstrated safety of the device for use in humans in the pivotal clinical study. ## X. SUMMARY OF PRIMARY CLINICAL STUDY The applicant performed a clinical study to establish a reasonable assurance of safety and effectiveness of stent assisted coiling (SAC) with the LVIS and LVIS Jr. for use with neurovascular embolization coils in the treatment of wide-necked saccular intracranial aneurysms in the US under IDE # G110188. Data from this clinical study were the basis for the PMA approval decision. A summary of the clinical study is presented below. ## A. Study Design Patients were treated between July 30, 2013 and October 28, 2014. The database for this PMA reflected data collected through February 24, 2017 and included 171 patients. There were 22 investigational sites, all in the US. The study, titled "Pivotal Study of the MicroVention, Inc. Neurovascular Self-Expanding Retrievable Stent System LVIS in the Treatment of Wide-Necked Intracranial Artery Aneurysms," was a multi-center, prospective, single-arm, open label clinical study. The pivotal study included follow-up at discharge, 30 days, 6 months, and 12 months. The pre-specified primary endpoints in the clinical study protocol were: - Safety: Any major stroke or death within 30 days, or major ipsilateral stroke or neurological death within 12 months. - Effectiveness: Successful aneurysm treatment with LVIS/LVIS Jr. as defined by complete (100%) aneurysm angiographic occlusion at 12 months without retreatment and no significant (≥ 50%) stenosis of the treated artery at 12 months. PMA P170013: FDA Summary of Safety and Effectiveness Data Page 11 {11} The primary endpoint results were compared to performance goals (PGs) developed using prior published data from the other HDE approved neurovascular stents including the Neuroform Stent Systems (H020002) and the Enterprise Vascular Reconstruction Device and Delivery System (H060001). Study analyses were conducted using Bayesian statistical methods described in the Guidance for Industry and FDA Staff, "Guidance for the Use of Bayesian Statistics in Medical Device Clinical Trials," issued on February 5, 2010. The Bayesian approach allows for incorporation of prior (external to the proposed study) information in decision-making. Bayesian analyses also allows the computation of one-sided 97.5% credible intervals and multivariate regression analyses. Bayesian analyses were performed using the program OpenBUGS, Version 3.2.3 (Lunn et al., The BUGS Project: Evolution, Critique and Future Directions. Statistics in Medicine 2009; 28(25): 3049-67). Posterior inferences were carried out by sampling from the posterior distribution of the parameters. Other analyses, including summaries, were conducted using SAS (version 9.2 or greater). Descriptive statistics were presented for continuous variables included mean, standard deviation, median, quartiles, minimum, maximum, and sample size for each treatment group. Categorical variables were summarized using counts and percentages. Proportions were calculated using known non-missing values. This study included an independent blinded Clinical Events Committee (CEC), an independent blinded Data Safety and Monitoring Board (DSMB), an angiographic Core Laboratory ("Core Lab"), and study monitors who confirmed neurological assessments, adverse events and study data with source documentation. 1. Clinical Inclusion and Exclusion Criteria Enrollment in the "Pivotal Study of the MicroVention, Inc. Neurovascular Self-Expanding Retrievable Stent System LVIS in the Treatment of Wide-Necked Intracranial Artery Aneurysms" was limited to patients who met the following inclusion criteria. - Subject whose age is ≥ 18 and ≤ 75 years. - Subject with an unruptured or ruptured (&gt; 30 days since occurrence), wide-necked (neck ≥ 4 mm or dome to neck ratio &lt; 2) intracranial, saccular aneurysms (≥ 4 mm and &lt; 20 mm maximum diameter in any plane) arising from a parent vessel with a diameter ≥ 2.0 mm and ≤ 4.5 mm who are candidates for endovascular coil embolization. - Subject or his/her Legally Authorized Representative understands the nature of the procedure, consents to participation in the study and provides a signed informed consent form. - Subject (woman of child-bearing potential) with a current negative pregnancy test who has agreed to an appropriate method of contraception throughout the trial. - Subject lives at a permanent address within commuting range of the investigational site and will be residing at that address during their 12 months of study participation. PMA P170013: FDA Summary of Safety and Effectiveness Data Page 12 {12} - Subject is willing to return to the investigational site for the 30-day, 6-month and 12-month follow-up evaluations. Patients were not permitted to enroll in the “Pivotal Study of the MicroVention, Inc. Neurovascular Self-Expanding Retrievable Stent System LVIS in the Treatment of Wide-Necked Intracranial Artery Aneurysms” if they met any of the following exclusion criteria. - Subject who presents with ruptured aneurysm, unless rupture occurred 30 days or more prior to screening. - Subject who presents with an intracranial mass (other than a meningioma) or currently undergoing radiation therapy for carcinoma or sarcoma of the head or neck region. - Subject with significant extracranial or intracranial stenosis of the parent artery (&gt; 50%) proximal to the target aneurysm. - Subject with an irreversible bleeding disorder, a platelet count of less than 100,000/ml (&lt; 100×10³ cells/mm³) or known platelet dysfunction or a contraindication to or inability to tolerate anticoagulants and/or antiplatelet agents. - Subject with serum creatinine level &gt; 3.0 mg/dL at time of enrollment (this will restrict the use of contrast) and not on dialysis. - Subject with known allergies to nickel-titanium metal, jewelries. - Subject with known allergies or contraindications to required anti-platelet and/or heparin medications required for treatment. - Subject with a life-threatening allergy to radiographic contrast (unless treatment for allergy is tolerated or can be managed medically). - Subject with a contraindication to CT (Computed Tomography) and MRI (Magnetic Resonant Imaging). - Subject who has a known cardiac disorder, likely to be associated with cardioembolic symptoms such as AFIB (atrial fibrillation). - Subject with any condition which in the opinion of the treating physician would place the subject at a high risk of embolic stroke. - Subject who is currently participating in another clinical research study with a conflicting protocol. - Subject who has had a previous intracranial stenting procedure associated with the target aneurysm. - Subject who is unable to complete the required follow-up. - Subject who is pregnant or breastfeeding; - Subject who has participated in a drug study within the last 30 days. Angiographic Exclusion Criteria - Subject has a cerebral diagnostic angiogram that demonstrates an aneurysm that is not appropriate for endovascular treatment. - Subject has a fusiform or dissecting aneurysm. PMA P170013: FDA Summary of Safety and Effectiveness Data Page 13 {13} - Subject is harboring more than one aneurysm with each aneurysm requiring treatment within 30 days. - Subject has an arteriovenous malformation (AVM) in the territory of the target aneurysm. ## 2. Follow-up Schedule All patients were scheduled to return for follow-up examinations at discharge, 30 days (± 7 days), 6 months (± 4 weeks), and 12 months (± 4 weeks) postoperatively. Preoperatively, the patients underwent a review of their concomitant medications, medical history, physical, clinical, neurological, laboratory, and angiographic evaluations. Postoperatively, the objective parameters measured during the study included a review of the concomitant medications, physical, clinical, neurological, and angiographic evaluations (see Table 5). Postoperatively, the angiographic examination was only conducted at the 12 month visit to evaluate intracranial aneurysm occlusion and parent artery patency. Adverse events and complications were recorded at all visits. The key timepoints are shown below in tables summarizing safety and effectiveness. Table 5: Study Timeline | | Pre-Procedure | Procedure | Post-Procedure | Discharge | 30 days ± 7 days | 6 months ± 4 weeks | 12 months ± 4 weeks | | --- | --- | --- | --- | --- | --- | --- | --- | | Medical History | X | | | | | | | | Physical Examination | X | | | X | X | X | X | | Informed Consent (1) | X | | | | | | | | Clinical Evaluation | X | | | X | X | X | X | | Neurological Evaluation | X | | | X | X | X | X | | Concomitant Medications | X | | | X | X | X | X | | Laboratory Assessment (2) | X | | | | | | | | Intracranial Stent Procedure | | X | | | | | | | Procedural Medications | | X | X | | | | | | Angiographic Evaluation | X | X | X | | | | X | | Adverse Event | X | X | X | X | X | X | X | | Serious Adverse Event (3) | X | X | X | X | X | X | X | | Unanticipated Adverse Event (3) | X | X | X | X | X | X | X | | Protocol Deviation (3) | X | X | X | X | X | X | X | | Death, or Device Related Adverse Event (3) | X | X | X | X | X | X | X | (1) Informed consent was signed before the patient was enrolled in the study. PMA P170013: FDA Summary of Safety and Effectiveness Data Page 14 {14} (2) Complete Blood Count (CBC) with differential, blood chemistry, PT/PTT, and blood sugar evaluation within 14 days from procedure and a pregnancy test for women of childbearing age at the time of enrollment. (3) A serious adverse event (e.g., death), protocol deviation, unanticipated adverse device effect or device related adverse event were reported to the Sponsor as soon as possible (i.e., within 24 hours) and no more than 10 working days from the date of becoming aware of the event or effect. ## 3. Clinical Endpoints With regards to safety, the number of patients who had a disabling stroke (defined as modified Rankin Scale (mRS) score ≥ 3 assessed at a minimum of 90 days post-stroke event) or neurological death within 12 months post-procedure was used to analyze the clinical study results. With regards to effectiveness, the number of patients who had complete (100%) occlusion (equivalent to Raymond-Roy I intracranial aneurysm occlusion classification) of the target intracranial aneurysm without clinically significant in-stent stenosis (≥ 50%) or target aneurysm re-treatment within 12 months post-procedure was used to analyze the clinical study results. In addition, the number of patients who had stable intracranial aneurysm occlusion assessed via two (2) imaging scans taken at a minimum of 6 months apart equivalent to a Raymond-Roy II intracranial aneurysm occlusion classification (~90-99% occlusion) without clinically significant in-stent stenosis or re-treatment of the target intracranial aneurysm were also considered a successful treatment with respect to effectiveness. These primary safety and effectiveness endpoints were determined to be most clinically meaningful for evaluating the safety and performance of the LVIS and LVIS Jr., and are consistent with the recommendations from a March 1, 2018 general issues meeting of the Neurological Devices Panel of the Medical Devices Advisory Committee to discuss the evaluation of benefits vs. risks of new endovascular medical devices intended to treat intracranial aneurysms. The pre-specified primary endpoints for the "Pivotal Study of the MicroVention, Inc. Neurovascular Self-Expanding Retrievable Stent System LVIS in the Treatment of Wide-Necked Intracranial Artery Aneurysms" are described in Section X (A. Study Design). With regard to success/failure criteria, the primary endpoints were compared to PGs. The clinical study would be considered a success for effectiveness if the primary effectiveness endpoint based on the rate of patients who had complete (100%) intracranial aneurysm occlusion without clinically significant in-stent stenosis or retreatment of the target aneurysm was greater than 46%. This PG for effectiveness was based on evaluable patient level data obtained from 514 patients from five (5) published studies using the HDE approved Neuroform Stent Systems (H020002) (Bondi et al. 2007; Fiorella et al.; Maldonado et al.; Yahia et al.; Sedat et al. 2009). The clinical study would be considered a success for safety if the primary safety endpoint was less than 20%. This PG for safety was calculated based on the rate of major stroke or PMA P170013: FDA Summary of Safety and Effectiveness Data Page 15 {15} neurological death observed in the prior clinical studies used to support HDE approval of the Neuroform Stent Systems (H020002) and the Enterprise Vascular Reconstruction Device and Delivery System (H060001), which was approximately 10% with an added non-inferiority margin of 10% to equal 20%. As part of the decision-making process for the subject PMA, the FDA did not consider the safety PG of 20% to be acceptable and utilized the observed 10% rate for primary safety events to be the PG when determining the safety profile of the LVIS and LVIS Jr. for study success. ## B. Accountability of PMA Cohort At the time of database lock, of 171 patients enrolled in the PMA study, 82% (141) patients are available for analysis at the completion of the study, the 12 month post-operative visit (see Figure 3). PMA P170013: FDA Summary of Safety and Effectiveness Data Page 16 {16} ![img-2.jpeg](img-2.jpeg) Figure 3: Subject Accountability Flow Chart [1] Subject 01-04 died on day 2 [2] Subject 22-05 had a missed 30 day visit [3] Subject 01-05 died on day 63 [4] Subjects 06-25, 09-02, 11-03, 12-04, 12-06, and 19-23 missed the 6 month visit [5] Subject 03-03 died on day 393; however, this subject completed the 12 month follow-up visit and is included. [6] Subjects 06-25 and 05-05 had a 12 month visit. 06-25 did not have an angiogram, 05-05 had an angiogram that was not evaluable. # C. Study Population Demographics and Baseline Parameters The demographics of the study population are typical for an intracranial aneurysm treatment study performed in the US. This disease predominantly affects more women than men, and most patients are White, similar to the demographic and baseline characteristics of the patient population in the "Pivotal Study of the MicroVention, Inc. Neurovascular Self-Expanding Retrievable Stent System LVIS in the Treatment of Wide PMA P170013: FDA Summary of Safety and Effectiveness Data {17} Necked Intracranial Artery Aneurysms" (see Table 6). The study population used to analyze the clinical study results are based on the Intent-to-Treat (ITT) population of 153 patients (N) who signed the informed consent form, met inclusion/exclusion criteria, and in which the device was attempted. Table 6: Demographics and Baseline Characteristics | Characteristic | Summary Statistic | | --- | --- | | Age (years) | | | Mean ± Standard Deviation (std) (N) | 58.3 ± 10.49 (153) | | Median (min, max) | 59.0 (18, 75) | | Gender, % (n/N) | | | Male | 28.1% (43/153) | | Female | 71.9% (110/153) | | Ethnicity, % (n/N) | | | Hispanic or Latino | 5.2 % (8/153) | | Not Hispanic or Latino | 94.1% (144/153) | | No willing to provide | 0.0 % (0/153) | | Unknown | 0.7 % (1/153) | | Missing1 | 0.0 % (0/153) | | Race, % (n/N) | | | American Indian or Alaska Native | 0.7 % (1/153) | | Asian | 1.3 % (2/153) | | Black or African American | 15.7% (24/153) | | Native Hawaiian or other Pacific Islander | 0.0 % (0/153) | | White | 80.4% (123/153) | | Other | 2.0 % (3/153) | | Missing1 | 0.7 % (1/153) | | Systolic Blood Pressure | | | Mean ± std (N) | 130.9 ± 17.80 (153) | | Median (min, max) | 131.0 (93, 179) | | Diastolic Blood Pressure | | | Mean ± std (N) | 74.8 ± 11.35 (153) | | Median (min, max) | 75.0 (40, 102) | | Temperature (°F) | | | Mean ± std (N) | 97.8 ± 0.90 (144) | | Median (min, max) | 97.9 (94.5, 99.5) | | Heart Rate (Beats per Minute (BPM)) | | | Mean ± std (N) | 73.0 ± 15.04 (153) | | Median (min, max) | 71.0 (45, 124) | Pre-procedure data is considered for this table. One subject is considered in both "Black or African American" and "Other" Missing information are considered under the category Missing. Baseline intracranial aneurysm characteristics are reported per the site evaluation. The target intracranial aneurysm locations in the neurovasculature are presented in Table 7, and Table 8 presents the intracranial aneurysm sizes and parent artery dimensions. The PMA P170013: FDA Summary of Safety and Effectiveness Data {18} mean intracranial aneurysm dome height was $6.0 \pm 2.15 \mathrm{~mm}$ and mean dome width was $5.5 \pm 2.33 \mathrm{~mm}$ . All intracranial aneurysms were wide-neck as defined by a neck width $\geq 4 \mathrm{~mm}$ or a dome to neck ratio $&lt; 2$ . There were only 12 intracranial aneurysms that were classified as large or giant since they had at least one aneurysm dimension (dome height or width) $\geq 10 \mathrm{~mm}$ . Table 7: Target Aneurysm Location | Location and Sublocation (site reported) | % of subjects (n/N) | | --- | --- | | Internal Carotid Artery (ICA) | 28.1% (43/153) | | Carotid Cavernous | 2.0% (3/153) | | Carotid Ophthalmic | 5.9% (9/153) | | Superior Hypophyseal | 6.5% (10/153) | | Posterior Communication Artery | 5.9% (9/153) | | Anterior Choroidal Artery | 1.3% (2/153) | | Internal Carotid Artery (Supraclinoid) | 4.6% (7/153) | | Carotid Bifurcation | 2.0% (3/153) | | Anterior Cerebral Artery (ACA) | 37.3% (57/153) | | Anterior Communicating Artery (AComm) | 33.3% (51/153) | | Pericallosal | 3.9% (6/153) | | Middle Cerebral Artery (MCA) | 11.1% (17/153) | | Posterior Cerebral Artery | 3.9% (6/153) | | Basilar Artery | 17.6% (27/153) | | Basilar Tip | 17.0% (26/153) | | Anterior Inferior Cerebellar Artery | 0.0% (0/153) | | Basilar Trunk | 0.7% (1/153) | | Superior Cerebellar Artery | 0.7% (1/153) | | Vertebral Artery | 1.3% (2/153) | | Posterior Inferior Cerebellar Artery (PICA) | 0.7% (1/153) | | Vertebrobasilar (VB) Junction | 0.7% (1/153) | Table 8: Target Aneurysm Characteristics | Characteristic | Mean ± std (N) | Median (min, max) | | --- | --- | --- | | Dome Height | 6.0 ± 2.15 (153) | 5.8 (2.0, 14.0) | | Dome Width (perpendicular to height) | 5.5 ± 2.33 (153) | 5.0 (1.4, 17.0) | | Neck Width | 4.2 ± 1.41 (153) | 4.0 (1.8, 10.0) | | Dome to Neck Ratio | 1.3 ± 0.38 (153) | 1.3 (0.5, 3.3) | | Distal Parent Artery Diameter (landing zone) | 2.5 ± 0.64 (153) | 2.2 (1.6, 4.8) | | Proximal Parent Artery Diameter (landing zone) | 2.8 ± 0.70 (153) | 2.5 (2.0, 4.5) | | Mean Parent Artery Diameter | 2.6 ± 0.64 (153) | 2.4 (2.0, 4.6) | PMA P170013: FDA Summary of Safety and Effectiveness Data {19} D. Safety and Effectiveness Results 1. Safety Results The analysis of safety was based on the ITT cohort of 153 patients available for the 12-month evaluation. The key safety outcomes for this study are presented below in Table 9. Adverse effects are reported in Table 10. Table 9: Primary Safety Events through 12 Months - ITT Population | Event Type | % of Subjects with Observations (n/N) | Posterior Mean, 95% Confidence Interval (CI) 1 | Posterior Probability 2 | | --- | --- | --- | --- | | Primary Safety Composite Rate (disabling stroke with mRS score ≥ 3 or neurological death within 12 months) | 5.9% (9/153) | 6.2% (3.0% - 10.5%) | 1 | | Primary Safety Failure Reasons 3 | | | | | Disabling stroke with mRS score ≥ 3 through 12 months* | 3.9% (6/153) | 4.2% (1.7% - 7.9%) | N/A | | Neurological death through 12 months | 2.0% (3/153) | 2.3% (0.6% - 5.1%) | N/A | 1 Posterior mean and 95% CI. Confidence intervals are calculated without multiplicity adjustment. As such, the confidence intervals are provided to show the variability only and should not be used to draw any statistical conclusions. 2 Posterior probability that the primary safety endpoint event rate is &lt; 20% (PG proposed by applicant). 3 Subjects may have more than one failed safety component. Three (3) subjects with stroke expired from neurological deaths. * mRS score ≥ 3 at any time point between 90 days and last available follow-up. There were nine (9) subjects who met the primary safety endpoint definition of disabling stroke or neurological death within 12 months post-procedure for a rate of 5.9% (9/153). Three (3) of the 9 primary safety events were deaths caused by strokes from which two (2) subjects expired from major ipsilateral strokes at 2 and 63 days post-procedure and the third subject sustained a fatal contralateral stroke 393 days post-procedure. Of the remaining six (6) disabling strokes out of the total of nine (9) primary safety events, two (2) strokes were determined by the CEC to have occurred in vessel territories that are distinct from those territories treated with LVIS/LVIS Jr. and may be caused by pre-existing conditions unrelated to the device or procedure. These two (2) disabling stroke events are analyzed within the primary safety endpoint as a worst-case analysis and it cannot be fully confirmed that the strokes were unrelated to the device and treatment. The posterior mean is 6.2% with a one-sided 97.5% CI upper bound of 10.5%. PMA P170013: FDA Summary of Safety and Effectiveness Data Page 20 {20} # Adverse effects that occurred in the PMA clinical study: Tables 10-12 present all adverse events, device-related serious adverse events, and procedure-related serious adverse events that were observed through 12 months in the pivotal study for the ITT population as adjudicated by the CEC, respectively. Tables 10-12 are also presented based on time of adverse event occurrence, whether periprocedural or post-procedure. In Tables 10-12, percentage of subjects with events may not match the number of events if more than one event occurred in the same subject. Table 10: All Adverse Events Observed through 12 Months - ITT Population | Adverse Event | Peri Procedure | | Post Procedure | | | --- | --- | --- | --- | --- | | | # of Events | % of subjects with event (n/N) | # of Events | % of subjects with event (n/N) | | Any Adverse Event | 145 | 39.9% (61/153) | 159 | 30.1% (46/153) | | Cardiac | 8 | 4.6% (7/153) | 10 | 5.2% (8/153) | | Cardiac arrhythmias | 4 | 2.6% (4/153) | 2 | 1.3% (2/153) | | Death | 0 | 0.0% (0/153) | 1 | 0.7% (1/153) | | Myocardial Infarction | 0 | 0.0% (0/153) | 1 | 0.7% (1/153) | | Other | 4 | 2.6% (4/153) | 6 | 3.9% (6/153) | | Gastrointestinal | 2 | 1.3% (2/153) | 13 | 5.9% (9/153) | | Bleeding | 0 | 0.0% (0/153) | 2 | 0.7% (1/153) | | Hemorrhage | 1 | 0.7% (1/153) | 0 | 0.0% (0/153) | | Infection | 0 | 0.0% (0/153) | 1 | 0.7% (1/153) | | Ischemia | 0 | 0.0% (0/153) | 1 | 0.7% (1/153) | | Other | 1 | 0.7% (1/153) | 9 | 4.6% (7/153) | | Infectious / Inflammatory | 0 | 0.0% (0/153) | 8 | 1.3% (2/153) | | Infection | 0 | 0.0% (0/153) | 8 | 1.3% (2/153) | | Musculoskeletal | 9 | 5.2% (8/153) | 17 | 9.8% (15/153) | | Ischemia | 0 | 0.0% (0/153) | 1 | 0.7% (1/153) | | Other | 9 | 5.2% (8/153) | 16 | 9.8% (15/153) | | Neurological / Neurovascular | 74 | 26.8% (41/153) | 50 | 19.0% (29/153) | | Aneurysm rupture | 4 | 2.6% (4/153) | 0 | 0.0% (0/153) | | Aphasia | 1 | 0.7% (1/153) | 0 | 0.0% (0/153) | | Device Failure | 10 | 6.5% (10/153) | 0 | 0.0% (0/153) | | Dissection or perforation of the parent artery | 2 | 1.3% (2/153) | 0 | 0.0% (0/153) | | Headache | 9 | 5.2% (8/153) | 8 | 4.6% (7/153) | | Hydrocephalus | 1 | 0.7% (1/153) | 1 | 0.7% (1/153) | | Intra-Parenchymal Hemorrhage | 2 | 1.3% (2/153) | 2 | 1.3% (2/153) | | Neurological deficits | 6 | 3.3% (5/153) | 3 | 1.3% (2/153) | | Other | 5 | 3.3% (5/153) | 12 | 7.2% (11/153) | | Seizure | 4 | 2.0% (3/153) | 1 | 0.0% (0/153) | | Stent Thrombosis | 3 | 2.0% (3/153) | 2 | 1.3% (2/153) | | Stroke | 6 | 3.3% (5/153) | 6 | 3.3% (5/153) | | Sub-Arachnoid Hemorrhage (SAH) | 2 | 1.3% (2/153) | 1 | 0.7% (1/153) | | Sub-Dural Hematoma (SDH) | 1 | 0.7% (1/153) | 2 | 1.3% (2/153) | | TIA (Transient Ischemic Attack) | 3 | 2.0% (3/153) | 3 | 2.0% (3/153) | | Target aneurysm retreatment | 0 | 0.0% (0/153) | 6 | 3.9% (6/153) | PMA P170013: FDA Summary of Safety and Effectiveness Data {21} Table 11: Device-Related Serious Adverse Events through 12 Months – ITT Population | Adverse Event | Peri Procedure | | Post Procedure | | | --- | --- | --- | --- | --- | | | # of Events | % of subjects with event (n/N) | # of Events | % of subjects with event (n/N) | | Any Serious Device Related Adverse Events | 22 | 11.1% (17/153) | 5 | 3.3% (5/153) | | Neurological / Neurovascular | 22 | 11.1% (17/153) | 5 | 3.3% (5/153) | | Device Failure* | 6 | 3.9% (6/153) | 0 | 0.0% (0/153) | | Dissection or perforation of the parent artery | 1 | 0.7% (1/153) | 0 | 0.0% (0/153) | | Other | 3 | 2.0% (3/153) | 0 | 0.0% (0/153) | | Stent Thrombosis† | 3 | 2.0% (3/153) | 2 | 1.3% (2/153) | | Stroke | 4 | 2.6% (4/153) | 2 | 1.3% (2/153) | | TIA (Transient Ischemic Attack) | 1 | 0.7% (1/153) | 1 | 0.7% (1/153) | | Vasospasm | 1 | 0.7% (1/153) | 0 | 0.0% (0/153) | | Visual impairment | 3 | 1.3% (2/153) | 0 | 0.0% (0/153) | PMA P170013: FDA Summary of Safety and Effectiveness Data {22} Table 12: Procedure-Related Serious Adverse Events through 12 Months – ITT Population | Adverse Event | Peri procedure | | Post Procedure | | | --- | --- | --- | --- | --- | | | # of Events | % of subjects with event (n/N) | # of Events | % of subjects with event (n/N) | | Any Serious Procedure Related Adverse Events | 61 | 28.8% (44/153) | 8 | 3.9% (6/153) | | Cardiac | 1 | 0.7% (1/153) | 0 | 0.0% (0/153) | | Cardiac arrhythmias | 1 | 0.7% (1/153) | 0 | 0.0% (0/153) | | Gastrointestinal | 1 | 0.7% (1/153) | 1 | 0.7% (1/153) | | Bleeding | 0 | 0.0% (0/153) | 1 | 0.7% (1/153) | | Hemorrhage | 1 | 0.7% (1/153) | 0 | 0.0% (0/153) | | Neurological / Neurovascular | 43 | 20.9% (32/153) | 6 | 2.6% (4/153) | | Aneurysm rupture | 4 | 2.6% (4/153) | 0 | 0.0% (0/153) | | Aphasia | 1 | 0.7% (1/153) | 0 | 0.0% (0/153) | | Device Failure | 9 | 5.9% (9/153) | 0 | 0.0% (0/153) | | Dissection or perforation of the parent artery | 1 | 0.7% (1/153) | 0 | 0.0% (0/153) | | Hydrocephalus | 1 | 0.7% (1/153) | 0 | 0.0% (0/153) | | Intra-Parenchymal Hemorrhage | 2 | 1.3% (2/153) | 0 | 0.0% (0/153) | | Neurological deficits | 2 | 1.3% (2/153) | 0 | 0.0% (0/153) | | Other | 1 | 0.7% (1/153) | 0 | 0.0% (0/153) | | Seizure | 1 | 0.7% (1/153) | 0 | 0.0% (0/153) | | Stent Thrombosis | 3 | 2.0% (3/153) | 0 | 0.0% (0/153) | | Stroke | 5 | 3.3% (5/153) | 1 | 0.0% (0/153) | | Sub-Arachnoid Hemorrhage (SAH) | 2 | 1.3% (2/153) | 0 | 0.0% (0/153) | | TIA (Transient Ischemic Attack) | 1 | 0.7% (1/153) | 1 | 0.7% (1/153) | | Target aneurysm retreatment | 0 | 0.0% (0/153) | 3 | 2.0% (3/153) | | Thromboembolic event | 1 | 0.7% (1/153) | 0 | 0.0% (0/153) | | Vasospasm | 7 | 4.6% (7/153) | 0 | 0.0% (0/153) | | Visual impairment | 2 | 1.3% (2/153) | 1 | 0.7% (1/153) | | Other | 4 | 2.6% (4/153) | 1 | 0.7% (1/153) | | Other | 4 | 2.6% (4/153) | 1 | 0.7% (1/153) | | Renal / Genitourinary | 2 | 1.3% (2/153) | 0 | 0.0% (0/153) | | Infection | 2 | 1.3% (2/153) | 0 | 0.0% (0/153) | | Respiratory / Pulmonary | 3 | 2.0% (3/153) | 0 | 0.0% (0/153) | | Infection | 1 | 0.7% (1/153) | 0 | 0.0% (0/153) | | Other | 2 | 1.3% (2/153) | 0 | 0.0% (0/153) | | Vascular | 7 | 4.6% (7/153) | 0 | 0.0% (0/153) | | Bleeding | 2 | 1.3% (2/153) | 0 | 0.0% (0/153) | | Complications of arterial puncture | 4 | 2.6% (4/153) | 0 | 0.0% (0/153) | | Hemorrhage | 1 | 0.7% (1/153) | 0 | 0.0% (0/153) | Table 13 below presents all the stroke events observed in the “Pivotal Study of the MicroVention, Inc. Neurovascular Self-Expanding Retrievable Stent System LVIS in the Treatment of Wide-Necked Intracranial Artery Aneurysms,” which was 16 events total that occurred in 14 subjects (9.2%, 14/153). PMA P170013: FDA Summary of Safety and Effectiveness Data {23} Table 13: All Stroke Events through 12 Months (ITT Population) | # | Subject ID* | mRS pre-procedure baseline | mRS at discharge | mRS at 30 days | mRS at 180 days | mRS at 12 months | | --- | --- | --- | --- | --- | --- | --- | | Strokes resulting in neurological death | | | | | | | | 1 | 01-04 | 0 | mRS = 6, neurological death | n/a | n/a | n/a | | 2 | 01-05 | 0 | 0 | 0 | mRS = 6, neurological death | n/a | | 3 | 03-03 € | 0 | 0 | Not done | 0 | mRS = 0 at 12 months. Subject expired from a neurological cause 393 days post-procedure | | Disabling strokes (mRS ≥ 3 at a minimum of 90 days) | | | | | | | | 4 | 03-02 ¥ | 0 | 4 | 1 | 3 | 1 | | 5 | 09-09 € | 0 | 3 | 2 | 2 | 3 | | 6 | 12-04 | 3 | 3 | 4 | Not done | 4 | | 7 | 14-18 | 1 | 1 | 0 | 5 | 5 | | Disabling strokes caused by pre-existing conditions | | | | | | | | 8 | 03-18 ¥ | 1 | 1 | 0 | 3 | 1 | | 9 | 04-09 | 3 | 3 | 3 | 3 | 3 | | Non-Disabling strokes | | | | | | | | 10 | 03-03 € | 0 | 0 | Not done | 0 | 0 | | 11 | 03-19 | 1 | 1 | 1 | 1 | 0 | | 12 | 06-07 | 0 | 0 | 1 | 1 | 0 | | 13 | 09-05 | 1 | 1 | 1 | 1 | 1 | | 14 | 09-09 € | 0 | 3 | 2 | 2 | 3 | | 15 | 09-13 | 0 | 0 | 0 | 0 | 0 | | 16 | 16-01 | 0 | 0 | 1 | 0 | 0 | *16 strokes occurred in 14 subjects, ID=Patient Identification. €Subjects 03-03 &amp; 09-09 each sustained both a minor stroke peri-operatively and a major stroke post-operatively. ¥Subjects 03-02 &amp; 03-18 sustained strokes that did not result in permanent neurological disability. One of the key secondary safety endpoints was assessment of the number of subjects who had a worsening in the mRS (measurement of patient disability) at 12 months post-procedure compared to their baseline mRS prior to the treatment. Table 14 shows all of the subjects in the "Pivotal Study of the MicroVention, Inc. Neurovascular Self-Expanding Retrievable Stent System LVIS in the Treatment of Wide-Necked Intracranial Artery Aneurysms" who had a worsening of the mRS at 12 months post-procedure compared to their baseline mRS, 16% (25/153). Fourteen PMA P170013: FDA Summary of Safety and Effectiveness Data {24} (14) out of the 25 patients with a deterioration in the mRS at 12 months post-procedure compared to baseline had a bad clinical outcome (mRS ≥ 3) resulting in a rate of 9.2% (14/153). This rate in bad clinical outcome based on the mRS score is less than the observed safety PG estimated for the pivotal study of 10%. Table 14: Subjects (n=25) with Worsening mRS Scores at 12 Months Compared to Baseline | Subject # | Baseline mRS | Discharge mRS | mRS at 30 days | mRS at 6 months | mRS at 12 months | Reason for Change | | --- | --- | --- | --- | --- | --- | --- | | Neurological Death | | | | | | | | 1* | 0 | 6 | | | | | | 2* | 0 | 0 | 0 | | | | | 3* | 0 | 0 | | 0 | 0 | | | Non-Neurological Death | | | | | | | | 4* | 1 | 1 | 1 | | | Cardiac arrest | | 5* | 0 | 0 | 0 | | | Drug overdose | | 6* | 2 | 2 | 2 | | | Suicide | | Neurological Decline from New Neurological Deficits | | | | | | | | 7 | 0 | 3 | 2 | 2 | 3 | Ataxia, Major Stroke | | 8 | 1 | 1 | 0 | 5 | 5 | Major stroke | | 9 | 0 | 0 | 0 | | 4 | Subdural Hematoma | | 10 | 0 | 0 | 0 | 2 | 3 | New diagnosis of multiple sclerosis | | 11 | 2 | 2 | 1 | 0 | 3 | Depression | | 12 | 3 | 4 | 4 | 3 | 4 | General Debilitation from visual impairment, anxiety, depression | | Neurological Decline from Pre-Existing Neurological Deficits | | | | | | | | 13 | 1 | 1 | 1 | 3 | 3 | Preexisting neuropathy | | 14 | 1 | 1 | 1 | 1 | 2 | General debilitation from right upper intrinsic weakness, right leg weakness, mild motor aphasia | | 15 | 3 | 3 | 4 | | 4 | Preexisting bilateral leg paresthesia & weakness | | Other (No New Neurological Deficit) | | | | | | | | 16 | 0 | 0 | 0 | 1 | 1 | Lightheaded episode | | 17 | 0 | 4 | 1 | 3 | 1 | Left foot weakness | PMA P170013: FDA Summary of Safety and Effectiveness Data {25} | 18 | 0 | 1 | 0 | 0 | 1 | Persisting headaches | | --- | --- | --- | --- | --- | --- | --- | | 19 | 0 | 0 | 0 | 1 | 1 | Left-sided weakness | | 20 | 0 | 0 | 0 | 0 | 1 | Headaches & Fatigue | | 21 | 0 | 0 | 0 | 1 | 1 | Arthritis | | 22 | 0 | 0 | 0 | 1 | 1 | Dizziness | | 23 | 0 | 0 | 1 | 1 | 1 | Exacerbation of pre-existing low back pain | | 24 | 0 | 0 | 1 | 1 | 1 | Headaches & dizziness | | 25 | 0 | 0 | 0 | 0 | 2 | Fatigue, CPAP issues | Six (6) subjects footnoted above died throughout the course of the LVIS study. *Subject (1) died 2 days post procedure, subject (2) died 63 days post procedure, subject (3) died 393 days post procedure, subject (4) died 310 days post procedure, subject (5) died 92 days post procedure, subject (6) died 202 days post procedure. ## 2. Effectiveness Results The analysis of effectiveness was based on the 153 evaluable patients at the 12-month time point. Key effectiveness outcomes are presented in Table 15. Table 15: Primary Effectiveness Endpoint Analysis through 12 Months – ITT Population | Endpoint [1] | % of Subjects with Observations (n/N) | Posterior Mean, 95% CI [2] | Posterior Probability [3] | | --- | --- | --- | --- | | Imputed Analysis per the Prespecified Primary Effectiveness Endpoint | | | | | Primary Effectiveness Composite Success (100% Aneurysm Occlusion without Clinically Significant In Stent Stenosis or Target Aneurysm Retreatment) | 70.6% (108/153) | 70.5% (63.0% – 77.4%) | 1 | | Imputed Analysis per the Modified Primary Effectiveness Endpoint | | | | | Modified Primary Effectiveness Composite Success (90% - 99% Aneurysm Occlusion without Clinically Significant In Stent Stenosis or Target Aneurysm Retreatment)Σ | 10.4% (16/153) | | | | Modified Primary Effectiveness Composite Success (90% - 100% Aneurysm Occlusion without Clinically Significant In Stent Stenosis or Target Aneurysm Retreatment) | 81.0% (124/153) | 80.8% (74.3% - 86.6%) | 1 | | Primary Effectiveness Endpoint Subcomponents | | | | | 90%-100% Aneurysm Occlusion* | 83.7% (128/153) | 83.4% (77.2% - 88.9%) | n/a | Σ: Standard Deviation; N: Number of patients; n: Number of patients with a significant difference; n: Number of patients with a significant difference. PMA P170013: FDA Summary of Safety and Effectiveness Data {26} | Endpoint [1] | % of Subjects with Observations (n/N) | Posterior Mean, 95% CI [2] | Posterior Probability [3] | | --- | --- | --- | --- | | Imputed Analysis per the Prespecified Primary Effectiveness Endpoint | | | | | Without Clinically Significant In Stent Stenosis (≥ 50%) of Parent Artery | 90.8% (139/153) | 90.6% (85.5% - 94.7%) | n/a | | No Target Aneurysm Retreatment | 96.1% (147/153) | 95.8% (92.1% - 98.4%) | n/a | | Evaluable Only Analysis per the Prespecified Primary Effectiveness Endpoint | % of Subjects with Observations (n/N) | | | | Primary Effectiveness Composite Success (100% Aneurysm Occlusion without Clinically Significant In Stent Stenosis or Target Aneurysm Retreatment) | 77.7% (108/139) | 77.5% (70.3% - 84%) | 1 | | Evaluable Only Analysis per the Modified Primary Effectiveness Endpoint | | | | | Modified Primary Effectiveness Composite Success (90% - 99% Aneurysm Occlusion without Clinically Significant In Stent Stenosis or Target Aneurysm Retreatment)Σ | 11.5% (16/139) | | | | Modified Primary Effectiveness Composite Success (90% - 100% Aneurysm Occlusion without Clinically Significant In Stent Stenosis or Target Aneurysm Retreatment) | 89.2% (124/139) | 82.5% (75.3% - 88.6%) | 1 | | Modified Primary Effectiveness Composite Success (90% - 100% Aneurysm Occlusion without Clinically Significant In Stent Stenosis or Target Aneurysm Retreatment) using only DSA at 12 months¥ | 88.6% (117/132) [124-7 / 139-7] | 88.3% (82.4%- 93.2%) | 1 | | Primary Effectiveness Endpoint Subcomponents | | | | | 90%-100% Aneurysm Occlusion* | 91.4% (127/139) | 91.2% (84.4% - 97.4%) | n/a | | Without Clinically Significant In Stent Stenosis (≥ 50%) of Parent Artery | 100.0% (139/139) | 99.6% (98.2% - 100%) | n/a | | No Target Aneurysm Retreatment | 95.7% (133/139) | 95.4% (91.3% - 98.2%) | n/a | Σ Only subjects with stable or positively progressing Raymond-Roy II occlusion between baseline and 12 months are included. ¥ Seven subjects who were assessed using magnetic resonance angiography (MRA) in lieu of digital subtraction angiography (DSA) are excluded. [1] Includes primary effectiveness data collected on all ITT subjects at or through 12-month follow-up visit. PMA P170013: FDA Summary of Safety and Effectiveness Data {27} [2] Posterior mean and two-sided 95% CI. The confidence intervals are calculated without multiplicity adjustment. As such, the confidence intervals are provided to show the variability only and should not be used to draw any statistical conclusions. [3] Posterior probability that the primary effectiveness endpoint success rate exceeds the prespecified PG at 12 months. *Subjects having negative progression from post-procedure to 12 months are considered failures. Missing data imputed as failures. The confidence intervals are calculated without multiplicity adjustment. As such, the confidence intervals are provided to show the variability only and should not be used to draw any statistical conclusions. Table 15 presents the primary effectiveness endpoint analyses individually for the number of patients in the ITT population who had complete (100%) occlusion (equivalent to Raymond-Roy I intracranial aneurysm occlusion classification) of the target intracranial aneurysm without clinically significant in-stent stenosis (≥ 50%) or target aneurysm re-treatment within 12 months post-procedure, and the number of patients who had stable intracranial aneurysm occlusion assessed via two imaging scans taken at a minimum of 6 months apart equivalent to a Raymond-Roy II intracranial aneurysm occlusion classification (~90-99% occlusion) without clinically significant in-stent stenosis or re-treatment of the target intracranial aneurysm, and a combined composite primary effectiveness rate with these two (2) results combined. The additional analysis of the primary effectiveness endpoint to include intracranial aneurysms with Raymond-Roy II classifications as a satisfactory outcome was based on the recommendations at the March 1, 2018 general issues meeting of the Neurological Devices Panel of the Medical Devices Advisory Committee. The additional analysis of the primary effectiveness endpoint to include "stable" Raymond-Roy II classification of treated intracranial aneurysms identified seventeen (17) patients from the "Pivotal Study of the MicroVention, Inc. Neurovascular Self-Expanding Retrievable Stent System LVIS in the Treatment of Wide-Necked Intracranial Artery Aneurysms" with stable or improved Raymond-Roy II intracranial aneurysm occlusions. Four (4) of the 17 subjects had "stable" intracranial aneurysm occlusion between post-procedure and 12 months and 13 had improved intracranial aneurysm occlusion at 12 months as compared to the immediate post-procedure angiogram. Of the 17 subjects, 16 subjects are included in the primary effectiveness endpoint composite analysis which includes freedom from clinically significant in-stent stenosis and target aneurysm retreatment (since one subject was retreated). As with the original ITT analysis, missing data were imputed as failures (not a primary endpoint success). Thus, 81.0% (124/153) of the subjects met the success criteria of the composite primary effectiveness endpoint analysis as shown in Table 15. The posterior mean for the composite primary effectiveness endpoint success proportion at 12 months post-treatment was 80.8% with an equitailed 95% CI of 74.3-86.6%, with the lower bound exceeding the 46% effectiveness PG. If using the primary effectiveness endpoint analysis of only those subjects who had complete (100%) intracranial aneurysm occlusion without clinically significant in-stent stenosis PMA P170013: FDA Summary of Safety and Effectiveness Data Page 28 {28} or retreatment of the target aneurysm, the posterior mean for success was 70.5% with an equitailed 95% CI of 63-77.4%, which still exceeded the effectiveness PG of 46%. The posterior probability of the alternative hypothesis was &gt; 0.9999; therefore, exceeding the pre-defined one-sided threshold of 0.975 (0.95 equitailed), and confirms the primary effectiveness analysis in the ITT population. Table 15 also presents an “Evaluable Only” analysis that excludes 14 subjects with missing angiographic data at 12 months. It is also noted that seven (7) subjects evaluated for this modified effectiveness endpoint were evaluated using magnetic resonance angiography (MRA) instead of digital subtraction angiography (DSA) at the 12 month follow-up visit, which is a protocol violation. With the MRA-evaluated subjects excluded, 88.6% (117/132) of the evaluable subjects with imaging available at the 12 month follow-up visit met the success criteria of the primary effectiveness endpoint as shown in Table 15. These results further support that the primary effectiveness endpoint was met and exceeds the effectiveness PG of 46%. Because the subject devices are intended to support neurovascular embolization coils in the treatment of saccular wide-neck intracranial aneurysms, FDA also evaluated two (2) key secondary effectiveness endpoints of successful delivery of the LVIS and LVIS Jr. and stent migration within 12 months post-procedure (see Table 16). There were no events of stent migration within 12 months post-procedure. There were 10 cases of technical success failure associated with delivery and implantation of the LVIS/LVIS Jr., and failures were primarily associated with inadequate sizing of the device resulting in poor or lack of apposition to the parent vessel walls and/or coil prolapse. Table 16. Secondary Endpoints of LVIS/LVIS Jr. Delivery Success and Stent Migration | Secondary Endpoint [1] | % Subjects with Observations (n/N) | Posterior Mean, 95% CI [2] | | --- | --- | --- | | LVIS/LVIS Jr. Delivery and Implant Success [3] | 93.5% (143/153) | 0.932 (0.888-0.966) | | Stent Migration | 0% (0/153) | 0.003 (0.000-0.016) | [1] Includes secondary endpoint data collected on all ITT patients at or through the 12-month follow-up visit. [2] Posterior mean and 95% CI. The confidence intervals are calculated without multiplicity adjustment. As such, the confidence intervals are provided to show the variability only and should not be used to draw any statistical conclusions. [3] Successful delivery of LVIS/LVIS Jr. is evaluated at the completion of the procedure. ## 3. Subgroup Analyses The following preoperative characteristics were evaluated for potential association with outcomes: intracranial aneurysm rupture status, intracranial aneurysm size, location in the neurovasculature (anterior vs. posterior circulation), sex, and use of device with hydrocoils or bare metal embolic coils. PMA P170013: FDA Summary of Safety and Effectiveness Data Page 29 {29} # Primary Safety Endpoint Sub-Group Analyses Subgroup analyses of the primary safety endpoint was performed based on sex, intracranial aneurysm rupture status, location in the neurovasculature, size, and use with hydrocoils or bare metal neurovascular embolic coils, and the subgroup analyses are presented in Tables 17-21. For sex (male vs. female), it appears there is a higher rate of probability of primary safety event occurrence in female patients as shown in Table 17. However, this result has limitations in that the pivotal study sample size was primarily female (71.9% (110/153)) and the primary safety event rates were small to make any definitive conclusions from this subgroup analysis that there may be a difference in clinical safety outcomes based on sex. Table 17: Subgroup Analysis of Primary Safety Endpoint through 12 Months Based on Sex - ITT Population | Event Type | Gender | % of Subjects with Observations (n/N) | Posterior Mean, 95% CI [1] | | --- | --- | --- | --- | | Modified Primary Safety Composite Rate* (Disabling stroke with mRS score ≥ 3 or neurological death within 12 months) | Male | 7.0% (3/43) | 0.070 (0.015 - 0.162) | | | Female | 5.5% (6/110) | 0.059 (0.023 - 0.109) | | Modified Primary Safety Failure Reasons [2] | | | | | Disabling stroke with mRS score ≥ 3 through 12 months * | Male | 7.0% (3/43) | 0.070 (0.015 - 0.163) | | | Female | 2.7% (3/110) | 0.032 (0.008 - 0.072) | | Neurological death through 12 months | Male | 0.0% (0/43) | 0.000 (0.000 - 0.001) | | | Female | 2.7% (3/110) | 0.031 (0.008 - 0.071) | [1] Posterior mean and 95% CI. The confidence intervals are calculated without multiplicity adjustment. As such, the confidence intervals are provided to show the variability only and should not be used to draw any statistical conclusions. [2] Subjects may have more than one failed safety component. [*] mRS score ≥ 3 at any time point between 90 days and last available follow-up. Three (3) subjects with stroke resulted in neurological deaths. Subgroup analyses were conducted for the primary safety endpoint based on intracranial aneurysm rupture status (unruptured vs. previously ruptured aneurysms that were treated with device &gt; 30 days after rupture), intracranial aneurysm location in the neurovasculature (anterior vs. posterior circulation), and intracranial aneurysm PMA P170013: FDA Summary of Safety and Effectiveness Data {30} size ( $\leq 5$ mm vs. $&gt;5$ mm) for the ITT population (see Tables 18-20). The analysis supports comparability of these subgroups of intracranial aneurysm type, location, and size for the primary safety endpoint events although there is limited data available for previously ruptured intracranial aneurysms that were treated with LVIS/LVIS Jr. $&gt;30$ days after rupture, posterior circulation aneurysms, and small aneurysms. Also, the "Pivotal Study of the MicroVention, Inc. Neurovascular Self-Expanding Retrievable Stent System LVIS in the Treatment of Wide-Necked Intracranial Artery Aneurysms" only enrolled and treated 12 $(12/153 = 7.8\%)$ intracranial aneurysms $\geq 10$ mm and there were $0\%$ (0/12) primary safety events in this subgroup and all nine (9) primary safety events occurred in patients with intracranial aneurysms $&lt; 10$ mm $(9/141 = 6.4\%)$ . This information is descriptive only and should not be used to make any definitive statistical or clinical conclusions on safety outcomes. Table 18: Subgroup Analysis of Primary Safety Endpoint through 12 Months Follow-Up by Intracranial Aneurysm Type - ITT Population | Event Type | Aneurysm Type | % of Subjects with Observations (n/N) | Posterior Mean, 95% CI [1] | | --- | --- | --- | --- | | Modified Primary Safety Composite Rate* (Disabling stroke with mRS score ≥ 3 or neurological death within 12 months) | Ruptured [3] | 0.0% (0/22) | 0.022 (0.000 - 0.103) | | | Unruptured | 6.9% (9/131) | 0.069 (0.032 - 0.118) | | Modified Primary Safety Failure Reasons [2] | | | | | Disabling stroke with mRS score ≥ 3 through 12 months * | Ruptured [3] | 0.0% (0/22) | 0.021 (0.000 - 0.105) | | | Unruptured | 4.6% (6/131) | 0.046 (0.017 - 0.088) | | Neurological death through 12 months | Ruptured [3] | 0.0% (0/22) | 0.021 (0.000 - 0.101) | | | Unruptured | 2.3% (3/131) | 0.023 (0.005 - 0.055) | [1] Posterior mean and $95\%$ CI. The confidence intervals are calculated without multiplicity adjustment. As such, the confidence intervals are provided to show the variability only and should not be used to draw any statistical conclusions. [2] Subjects may have more than one failed safety component. [3] Device treatment was initiated $&gt;30$ days since occurrence of intracranial aneurysm rupture. [*] mRS score $\geq 3$ at any time point between 90 days and last available follow-up. Three (3) subjects with stroke resulted in neurological deaths. PMA P170013: FDA Summary of Safety and Effectiveness Data {31} Table 19: Subgroup Analysis of Primary Safety Endpoint through 12 Months Follow-Up by Intracranial Aneurysm Location – ITT Population | Event Type | Aneurysm Location | % of Subjects with Observations (n/N)# | Posterior Mean, 95% CI [1] | | --- | --- | --- | --- | | Modified Primary Safety Composite Rate* (Disabling stroke with mRS score ≥ 3 or neurological death within 12 months) | Anterior circulation | 6.8% (8/118) | 0.071 (0.032 - 0.125) | | | Posterior circulation | 2.9% (1/34) | 0.029 (0.001 - 0.106) | | Modified Primary Safety Failure Reasons [2] | | | | | Disabling stroke with mRS score ≥ 3 through 12 months* | Anterior circulation | 5.1% (6/118) | 0.055 (0.021 - 0.103) | | | Posterior circulation | 0.0% (0/34) | 0.000 (0.000 - 0.001) | | Neurological death through 12 months | Anterior circulation | 1.7% (2/118) | 0.021 (0.004 - 0.053) | | | Posterior circulation | 2.9% (1/34) | 0.029 (0.001 - 0.105) | [1] Posterior mean and 95% CI. The confidence intervals are calculated without multiplicity adjustment. As such, the confidence intervals are provided to show the variability only and should not be used to draw any statistical conclusions. [2] Subjects may have more than one failed safety component. [*] mRS score ≥ 3 at any time point between 90 days and last available follow-up. Three (3) subjects with stroke resulted in neurological deaths. #One subject 16-03 has missing aneurysm location. Table 20: Subgroup Analysis of Primary Safety Endpoint through 12 Months Follow-Up by Intracranial Aneurysm Size – ITT Population | Event Type | % of Subjects with Observations | Posterior Mean, 95% CI [1] | Posterior Probability [2] | | --- | --- | --- | --- | | Primary Safety Composite Rate* (Disabling stroke with mRS score ≥ 3 or neurological death within 12 months) | Size ≤ 5 mm 6.8% (3/44) | 7.8% (1.9% - 17.1%) | 0.992 | | | Size > 5 mm 5.5% (6/109) | 5.9% (2.3% - 11.0%) | 1 | [1] Posterior Mean and 95% Credible Interval (CI). The confidence intervals are calculated without multiplicity adjustment. As such, the confidence intervals are provided to show the variability only and should not be used to draw any statistical conclusions. PMA P170013: FDA Summary of Safety and Effectiveness Data {32} [2] Posterior probability that the primary safety endpoint event rate is $&lt; 20\%$ (PG proposed by applicant). [*] mRS score $\geq 3$ at any time point between 90 days and last available follow-up. An additional subgroup analysis was conducted for the primary safety endpoint based on neurovascular embolization coil type used for treatment with the LVIS and LVIS Jr. for the ITT population (see Table 21). The analysis supports comparable safety outcomes for both coils types (hydrogel and non-hydrogel coils). Table 21: Subgroup Analysis of Primary Safety Endpoint through 12 Months Follow-Up by Type of Neurovascular Embolization Coil used with LVIS/LVIS Jr. - ITT Population | Event Type | Coil Type | % of Subjects with Observations (n/N)# | Posterior Mean, 95% CI [1] | | --- | --- | --- | --- | | Modified Primary Safety Composite Rate* (Disabling stroke with mRS score ≥ 3 or neurological death within 12 months) | Hydrogel Coils | 6.8% (5/74) | 0.073 (0.026 - 0.142) | | | Non-Hydrogel Coils | 5.2% (4/77) | 0.052 (0.014 - 0.111) | | Modified Primary Safety Failure Reasons [2] | | | | | Disabling stroke with mRS score ≥ 3 through 12 months* | Hydrogel coils | 4.1% (3/74) | 0.047 (0.012 - 0.104) | | | Non-Hydrogel coils | 3.9% (3/77) | 0.039 (0.008 - 0.092) | | Neurological death through 12 months | Hydrogel coils | 2.7% (2/74) | 0.033 (0.006 - 0.084) | | | Non-Hydrogel coils | 1.3% (1/77) | 0.013 (0.000 - 0.048) | [1] Posterior mean and $95\%$ CI. The confidence intervals are calculated without multiplicity adjustment. As such, the confidence intervals are provided to show the variability only and should not be used to draw any statistical conclusions. [2] Subjects may have more than one failed safety component. [*] mRS score $\geq 3$ at any time point between 90 days and last available follow-up. Three (3) subjects with stroke resulted in neurological deaths. Two subjects 06-06 and 06-19 have missing coil type. Primary Effectiveness Endpoint Sub-Group Analyses Subgroup analyses were conducted for the primary effectiveness endpoint based on sex, intracranial aneurysm type, location in the neurovasculature, size, and use with PMA P170013: FDA Summary of Safety and Effectiveness Data {33} type of neurovascular embolic coil (see Tables 22-26). For the subgroup analysis based on sex (male vs. female), the primary effectiveness endpoint results appear comparable between both groups with female patients exhibiting slightly less effectiveness. For the subgroup analysis of the primary effectiveness endpoint based on intracranial aneurysm type (unruptured vs. previously ruptured intracranial aneurysm with device implanted $&gt;30$ days after rupture), the results show that previously ruptured intracranial aneurysms with the device implanted $&gt;30$ days after rupture tended to have a higher primary effectiveness endpoint rate of success. Based on the limited sample size of male subjects and the number of previously ruptured intracranial aneurysms treated with the device $&gt;30$ days after rupture in the pivotal study, it is difficult to draw any statistical or clinical conclusions from these subgroup analyses. Table 22: Subgroup Analysis of Primary Effectiveness Endpoint through 12 Months by Sex - ITT Population | Primary Effectiveness Endpoint [1] | Sex | % of Subjects with Observations (n/N) | Posterior Mean, 95% CI [2] | | --- | --- | --- | --- | | Primary Effectiveness Composite Success* [100% Aneurysm Occlusion (Equivalent to Raymond-Roy I Classification) without Clinically Significant In Stent Stenosis or Target Aneurysm Retreatment] | Male | 74.4% (32/43) | 0.744 (0.606-0.862) | | | Female | 69.1% (76/110) | 0.689 (0.600-0.772) | [1] Includes primary effectiveness data collected on all ITT subjects at or through 12 month follow-up visit. [2] Posterior mean and two-sided $95\%$ CI. The confidence intervals are calculated without multiplicity adjustment. As such, the confidence intervals are provided to show the variability only and should not be used to draw any statistical conclusions. *Missing data imputed as failures. PMA P170013: FDA Summary of Safety and Effectiveness Data {34} Table 23: Subgroup Analysis of Primary Effectiveness Endpoint through 12 Months by Intracranial Aneurysm Type - ITT Population | Primary Effectiveness Endpoint [1] | Intracranial Aneurysm Type | % of Subjects with Observations (n/N) | Posterior Mean, 95% CI [2] | | --- | --- | --- | --- | | Primary Effectiveness Composite Success* [100% Aneurysm Occlusion (Equivalent to Raymond-Roy I Classification) without Clinically Significant In Stent Stenosis or Target Aneurysm Retreatment] | Ruptured [3] | 81.8% (18/22) | 0.803 (0.619-0.934) | | | Unruptured | 68.7% (90/131) | 0.687 (0.606-0.764) | [1] Includes primary effectiveness data collected on all ITT subjects at or through 12 month follow-up visit. [2] Posterior mean and two-sided $95\%$ CI. The confidence intervals are calculated without multiplicity adjustment. As such, the confidence intervals are provided to show the variability only and should not be used to draw any statistical conclusions. [3] Device treatment initiated $&gt;30$ days from intracranial aneurysm rupture. *Missing data imputed as failures. Table 24: Subgroup Analysis of Primary Effectiveness Endpoint through 12 Months by Intracranial Aneurysm Location - ITT Population | Primary Effectiveness Endpoint [1] | Location | % of Subjects with Observations (n/N)# | Posterior Mean, 95% CI [2] | | --- | --- | --- | --- | | Primary Effectiveness Composite Success* [100% Aneurysm Occlusion (Equivalent to Raymond-Roy I Classification) without Clinically Significant In Stent Stenosis or Target Aneurysm Retreatment] | Anterior Circulation | 74.6% (88/118) | 0.744 (0.662-0.817) | | | Posterior Circulation | 58.8% (20/34) | 0.588 (0.421-0.746) | [1] Includes primary effectiveness data collected on all ITT subjects at or through 12 month follow-up visit. [2] Posterior mean and two-sided $95\%$ CI. The confidence intervals are calculated without multiplicity adjustment. As such, the confidence intervals are provided to show the variability only and should not be used to draw any statistical conclusions. *Missing data imputed as failures. One subject 16-03 has missing intracranial aneurysm location by Core Lab assessment. Therefore, the total $N = 152$ . PMA P170013: FDA Summary of Safety and Effectiveness Data {35} As an additional subgroup analysis, the primary effectiveness endpoint was evaluated by intracranial aneurysm location (anterior vs. posterior circulation) per Core Lab assessment (see Table 24). The distribution of anterior versus posterior intracranial aneurysms in the LVIS and LVIS Jr. pivotal trial is representative of the prevalence of intracranial aneurysms located in these regions. In prior published studies, the initial International Study of Unruptured Intracranial Aneurysms (ISUIA) was a retrospective study that evaluated 1449 angiographically confirmed intracranial aneurysms, which showed only 207 (14%) were in the posterior circulation (Wiebers 1998). In a 2003 follow-up study by Wiebers, the results showed that 12% of the 4060 participating patients had intracranial aneurysms located in the posterior circulation (Wiebers 2003). The subgroup analysis results show that there is decreased device and treatment effectiveness for posterior circulation intracranial aneurysms. This result is not unexpected because there may be a greater quantity of sources of blood supply and small arteries in the posterior circulation and it may be more procedurally challenging to access and implant the device in the posterior circulation resulting in a decreased effectiveness to fully occlude the intracranial aneurysm to achieve a Raymond-Roy I occlusion classification. Table 25: Subgroup Analysis of Primary Effectiveness Endpoint through 12 Months by Intracranial Aneurysm Size – ITT Population | Primary Effectiveness Endpoint [1] | Size | % of Subjects with Observations (n/N) | Posterior Mean, 95% CI [2] | Posterior Probability [3] | | --- | --- | --- | --- | --- | | Primary Effectiveness Composite Success* [90% - 100% Aneurysm Occlusion (Equivalent to Raymond-Roy I and “Stable” Raymond-Roy II Classifications) without Clinically Significant In Stent Stenosis or Target Aneurysm Retreatment] | ≤ 5 mm | 86.4% (38/44) | 85.6% (74% - 94.1%) | 1 | | | > 5 mm | 78.9% (86/109) | 78.6% (70.5% - 85.7%) | 1 | [1] Includes primary effectiveness data collected on all ITT subjects at or through 12-month follow-up visit. [2] Posterior mean and two-sided 95% CI. The…
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...