DUROLANE

P170007 · Bioventus, LLC · MOZ · Aug 29, 2017 · Orthopedic

Device Facts

Record IDP170007
Device NameDUROLANE
ApplicantBioventus, LLC
Product CodeMOZ · Orthopedic
Decision DateAug 29, 2017
DecisionAPPR
Device ClassClass 3
AttributesTherapeutic

Indications for Use

DUROLANE® is indicated for the treatment of pain in osteoarthritis of the knee in patients who have failed to respond adequately to conservative non-pharmacological therapy or simple analgesics, e.g., acetaminophen.

Device Story

DUROLANE is a high molecular weight, non-animal, stabilized sodium hyaluronate gel (NASHA technology) derived from bacterial fermentation. It is administered as a single intra-articular injection into the knee joint by a physician. The gel acts as a viscoelastic supplement to synovial fluid, providing lubrication and shock absorption to reduce pain associated with osteoarthritis. It is designed to be more resistant to degradation than unmodified or cross-linked HA preparations. The device provides a therapeutic alternative to multi-injection HA regimens, corticosteroids, or analgesics. Patients benefit from a single-injection treatment protocol, which may improve compliance and reduce the burden of repeated clinical visits. Clinical output is measured by patient-reported pain reduction (WOMAC scores).

Clinical Evidence

Pivotal prospective, randomized, multicenter, non-inferiority trial (TG1018DLN) compared single-injection DUROLANE to a 5-injection HA product in 319 evaluable patients (Chinese population). Primary endpoint: WOMAC pain subscale score change from baseline over 18 weeks. Results: DUROLANE LSM -5.97 vs. 5-injection HA -5.87 (difference -0.09; 95% CI: -0.58, 0.39), meeting non-inferiority margin of +1.6. Secondary endpoints (physical function, stiffness, global assessment) also met non-inferiority. Safety profile comparable to comparator; most common AEs were transient arthralgia and injection site pain. No device-related SAEs reported.

Technological Characteristics

Stabilized sodium hyaluronate (20 mg/mL) in phosphate buffered saline. Derived from Streptococcus equi fermentation. NASHA proprietary stabilization technology. Supplied in 3 mL pre-filled glass syringes. Viscoelastic properties; resistant to free radical degradation. Biocompatibility per ISO 10993 standards.

Indications for Use

Indicated for patients aged 40-80 with mild to moderate osteoarthritis of the knee (Kellgren-Lawrence grade 2 or 3) who failed conservative therapy. Contraindicated in patients with knee joint infections, skin disease at injection site, or known hypersensitivity to HA preparations.

Submission Summary (Full Text)

{0} SUMMARY OF SAFETY AND EFFECTIVENESS DATA (SSED) I. GENERAL INFORMATION Device Generic Name: Hyaluronic acid, intra-articular Device Trade Name: DUROLANE® Device Procode: MOZ Applicant’s Name and Address: Bioventus LLC 4721 Emperor Boulevard, Suite 100 Durham, NC 27703 Date(s) of Panel Recommendation: None Premarket Approval Application (PMA) Number: P170007 Date of FDA Notice of Approval: August 29, 2017 II. INDICATIONS FOR USE DUROLANE® is indicated for the treatment of pain in osteoarthritis of the knee in patients who have failed to respond adequately to conservative non-pharmacological therapy or simple analgesics, e.g., acetaminophen. III. CONTRAINDICATIONS - Do not inject DUROLANE with knee joint infections, infections, or skin disease in the area of the injection site. - Do not administer to patients with known hypersensitivity (allergy) to HA preparations. IV. WARNINGS AND PRECAUTIONS The warning and precautions can be found in the labeling for DUROLANE. V. DEVICE DESCRIPTION DUROLANE is a high molecular weight, non-animal, sodium hyaluronate, stabilized gel produced using a unique, proprietary technology, NASHA®. The device is administered as a single injection. PMA P170007: FDA Summary of Safety and Effectiveness Data Page 1 {1} Each 3 mL glass syringe of DUROLANE contains 20 mg/mL of sodium hyaluronate, dissolved in phosphate buffered saline. The sodium hyaluronate is derived from bacterial fermentation (Streptococcus equi). Each pre-filled syringe contains the following: | Component | Each mL contains | | --- | --- | | Stabilized Sodium Hyaluronate | 20 mg | | Sodium Chloride | 9 mg | | Potassium Dihydrogen Phosphate | 0.03 mg | | Disodium Hydrogen Phosphate Dihydrate | 0.14 mg | | Water for Injection | q.s. 1mL | VI. ALTERNATIVE PRACTICES AND PROCEDURES There are several other alternatives for the treatment of pain in osteoarthritis (OA) of the knee. For patients who have failed to respond to conservative, non-pharmacologic therapy, alternative practices and procedures include simple analgesics, such as acetaminophen, non-steroidal anti-inflammatory drugs (NSAIDs), intraarticular injection of corticosteroids, avoidance of activities that cause joint pain, exercise, physical therapy, and removal of excess fluid from the knee. For patients who fail to respond to these treatments, surgical interventions including arthroscopic surgery and total knee replacement are also alternatives. Each alternative has its own advantages and disadvantages. A patient should fully discuss these alternatives with his/her physician to select the method that best meets expectations and lifestyle. VII. MARKETING HISTORY DUROLANE has been CE marked within the European Union since 2001. DUROLANE is marketed globally. DUROLANE has not been required to be withdrawn from any market for any reason related to safety or effectiveness of the device. VIII. POTENTIAL ADVERSE EFFECTS OF THE DEVICE ON HEALTH Below is a list of the potential adverse effects (e.g., complications) associated with the use of the device and, in general, associated with intra-articular injection devices for the treatment of pain in osteoarthritis of the knee: - Aggravated osteoarthritis - Arthralgia (knee pain) PMA P170007: FDA Summary of Safety and Effectiveness Data {2} - Arthropathy - Arthrosis - Baker’s cyst - Bursitis - Immune Response - Infection - Injection site erythema - Injection site edema - Injection site pain - Injection site reaction - Localized osteoarthritis - Joint (knee) disorder - Joint (knee) swelling - Joint (knee) effusion - Joint (knee) stiffness - Pain in limb - Paraesthesia - Phlebitis - Puritis - Tendonitis For the specific adverse events that occurred in the clinical studies, please see Section X below. ## IX. SUMMARY OF NONCLINICAL STUDIES DUROLANE has undergone extensive pre-clinical testing in a variety of in-vitro and in-vivo studies. See Table 1. The majority of the biocompatibility evaluations were performed on product formulations which are chemically equivalent to DUROLANE and which are manufactured using the same NASHA gel technology. These gels are used for different indications and fill volume. However, given the chemical equivalence, these tests are considered applicable to DUROLANE. Where there was a parameter which could impact results of the test (e.g., gel particle size), testing was done either on DUROLANE or a formulation with the same specifications. Biocompatibility tests were performed in accordance with the applicable subsection of ISO-10993, where appropriate. PMA P170007: FDA Summary of Safety and Effectiveness Data Page 3 {3} Table 1: Key Pre-clinical Evaluations of DUROLANE | Study | Product Evaluated | Applicable Standard | Results | | --- | --- | --- | --- | | USP Rabbit Pyrogen Test | Chemically equivalent NASHA gel | USP | No pyrogens | | Intra-articular injection in rabbits - acute and sub-chronic toxicity | DUROLANE | N/A – internally-developed test | No toxicity reaction after 1 and 3 weeks intra-articular injection | | Repeated intra-articular injections in rabbit - chronic toxicity | DUROLANE | N/A – internally-developed test | No toxicity reaction after 1 and 13 weeks intra-articular injection | | Intradermal injection in rabbits - acute, sub-chronic and chronic toxicity | Chemically equivalent NASHA gel | N/A – internally-developed test | No toxicity reaction after 1, 2, 3 or 52 weeks intradermal injection | | Systemic toxicity in rats | Chemically equivalent NASHA gel | ISO 10993-11 | No evidence of systemic toxicity 13 weeks post subcutaneous injection | | Cytotoxicity study, ISO elution Method | Chemically equivalent NASHA gel | ISO 10993-5 | Non-cytotoxic (cytotoxicity grade < 2) | | Cytotoxicity test in V79 cells, colony formation test | Chemically equivalent NASHA gel | Japanese Pharmaceutical Affairs Bureau MHLW notification #99: 1995 | Non-cytotoxic | | Ames Test | Chemically equivalent NASHA gel | EU guidelines B14: 1992 | Non-mutagenic | | In vitro chromosomal aberration study in mammalian cells | Chemically equivalent NASHA gel | ISO 10993-3 | Non-genotoxic | | Mouse bone marrow micronucleus study | Chemically equivalent NASHA gel | ISO 10993-3 | Non-genotoxic | | Cancer Risk Assessment | Chemically equivalent NASHA gel | N/A – toxicological evaluation of chemical stabilizer used in NASHA process | Cancer risk from exposure minimal | | ISO Maximization Sensitization study | Chemically equivalent NASHA gel | ISO 10993-10 | No sensitization at injection site | | ISO Intracutaneous study (solution) | Chemically equivalent NASHA gel | ISO 10993-10 | Slight irritation at the injection site | | Muscle implantation test in the rabbit (4 weeks) | Chemically equivalent NASHA gel | ISO 10993-6 | Well tolerated by local muscle tissue | | Muscle implantation test in the rabbit (90 days) | Chemically equivalent NASHA gel | ISO 10993-6 | Slight general foreign body reaction | | Residence time in rabbits | 14C-labeled DUROLANE | N/A – internally developed test | 28-day half-life | PMA P170007: FDA Summary of Safety and Effectiveness Data {4} | Residence time in humans | 131I-labeled DUROLANE | N/A – internally developed test | 4-week half-life | | --- | --- | --- | --- | | Viscoelastic properties following free radical degradation | DUROLANE | N/A – internally developed test | More resistant to degradation effects compared to an unmodified and a cross- linked HA preparation | | Viscoelastic properties under simulated osteoarthritic joint conditions | DUROLANE | N/A – internally developed test | More resistant to degradation effects when diluted as compared to a cross-linked HA preparation | | Anti-nociceptive effects in rats | DUROLANE | N/A – internally developed test | High mechanical threshold required for pain induction; approximate equal weight distribution between affected and unaffected limbs | | Sequestration of pain-inducing agents | DUROLANE | N/A – internally developed test | Significant absorbance of free bradykinin and PGE2 in solution. | # X. SUMMARY OF PRIMARY CLINICAL STUDIES The applicant performed a clinical study to establish a reasonable assurance of safety and effectiveness of a single intra-articular injection of DUROLANE for the treatment of pain in osteoarthritis of the knee in patients who have failed to respond adequately to conservative non-pharmacological therapy or simple analgesics in the People's Republic of China. Data from this clinical study were the basis for the PMA approval decision. Summaries of this clinical study, TG1018DLN, as well as three preceding clinical development studies, 35GA0001, 35GA0301, and 35GA0608, and a repeated injection safety study, 35GA09901E, are presented below. # A. Original Clinical Development Studies 35GA0001, 35GA0301, and 35GA0608 The original clinical development of DUROLANE was founded upon three randomized, controlled trials: 35GA0001, 35GA0301, and 35GA0608. The initial two trials were superiority studies versus saline; the third was a non-inferiority trial versus the commonly used corticosteroid, methylprednisolone acetate (MPA). All trials evaluated the outcome measure associated with a pain responder rate, defined as a minimum $40\%$ reduction from baseline in WOMAC pain scores and an absolute reduction of at least 5 points in that score. The saline controlled trials are summarized in Table 2: PMA P170007: FDA Summary of Safety and Effectiveness Data {5} Table 2: Original Clinical Development Studies-DUROLANE vs. Saline | Clinical Study No. | n | # of Centers | Pain Responder Rate % | | | | | --- | --- | --- | --- | --- | --- | --- | | | | | Visits | DUROLANE | SALINE | p-value | | 35GA0001 | DUROLANE: 172 Saline: 174 | 7 US 6 Canada 5 Sweden | 2 weeks | 29.1 | 36.2 | 0.16 | | | | | 6 weeks | 36.6 | 29.9 | 0.18 | | | | | 3 months | 32.0 | 35.1 | 0.54 | | | | | 6 months | 29.1 | 32.2 | 0.53 | | 35GA0301 | DUROLANE: 108 Saline: 110 | 6 Canada 4 United Kingdom 2 Germany | 2 weeks | 19.4 | 25.5 | 0.29 | | | | | 4 weeks | 26.9 | 26.4 | 0.94 | | | | | 6 weeks | 30.6 | 26.4 | 0.49 | Neither of these saline-controlled studies demonstrated superiority over saline. Confounding factors were identified following subgroup analyses after completion of each trial. Study 35GA001 included patients with poly-articular pain, making it difficult to discriminate if pain reported was from the signal knee joint versus other joints. Study 35GA0301 identified an additional confounding factor of joint effusion. The presence of effusion may signify the presence of an active inflammatory process which can lead to degradation of HA through pro-inflammatory cytokine activity. These two studies informed the design of the third study, 35GA0608, outside of the United States comparing DUROLANE versus MPA. Results from this study are summarized in Table 3. Table 3: Original Clinical Development Studies-DUROLANE vs. MPA | Clinical Study No. | # of Centers | Pain Responder Rate % | | | | | --- | --- | --- | --- | --- | --- | | | | Visits | DUROLANE n=221 | MPA n=221 | Difference 95% Confidence Interval* | | 35GA0608 | 15 Canada 5 Sweden 4 UK | 6 weeks | 47.7 | 50.2 | (-11.9%; +6.9%) | | | | 12 weeks | 44.6 | 46.2 | (-11.2%; +7.9%) | | | | 18 weeks | 43.0 | 45.2 | (-11.9%; +7.4%) | | | | 26 weeks | 43.9 | 36.9 | (-2.5%; +16.6%) | *Hypothesis testing required the lower bound to be greater than -0.15 Adverse events (AEs) from these three trials are collectively presented in Table 4. Table 4: Summary of Adverse Events Reported in Original Clinical Development Studies | Preferred Class* | DUROLANE All three studies (n=502) | MPA 35GA0608 (n=221) | Saline 35GA0001 and 35GA0301 (n=284) | | --- | --- | --- | --- | | Related to product or injection procedure or both | | | | | Nausea | - | 2 (0.9%) | - | PMA P170007: FDA Summary of Safety and Effectiveness Data {6} | Pyrexia | 2 (0.4%) | - | - | | --- | --- | --- | --- | | Injection site haematoma | - | - | 2 (0.7%) | | Injection site haemorrhage | - | - | 1 (0.4%) | | Injection site pain | 15 (3.0%) | 1 (0.5%) | 2 (0.7%) | | Injection site swelling | 2 (0.4%) | - | - | | Blood glucose increased | 1 (0.2%) | - | - | | Arthralgia | 54 (10.8%) | 7 (3.2%) | 8 (2.8%) | | Arthropathy | 9 (1.8%) | - | 5 (1.8%) | | Joint crepitation | 1 (0.2%) | - | - | | Joint effusion | 1 (0.2%) | 1 (0.5%) | - | | Joint lock | 1 (0.2%) | - | - | | Joint stiffness | 4 (0.8%) | - | - | | Joint swelling | 5 (1.0%) | 1 (0.5%) | - | | Joint warmth | 1 (0.2%) | - | - | | Muscle spasms | 1 (0.2%) | - | - | | Pain in extremity | 1 (0.2%) | 1 (0.5%) | - | | Sensation of heaviness | | 1 (0.5%) | | | Synovitis | 1 (0.2%) | | 1 (0.4%) | | Anxiety | 1 (0.2%) | - | - | | Depression | 1 (0.2%) | - | - | | Dermatitis | - | - | 1 (0.4%) | | Headache | 2 (0.4%) | - | - | | Haemarthrosis | 1 (0.2%) | - | - | | Myalgia | 1 (0.2%) | - | - | | Oedema peripheral | - | - | 1 (0.4%) | | Osteoarthritis | 2 (0.4%) | 1 (0.5%) | - | | Nervousness | 1 (0.2%) | - | - | * AEs in the 35GA001 and 35GA0301 studies were classified using World Health Organisation, Adverse Reaction Terminology (WHO ART). In 35GA0608, AEs were classified using Medical Dictionary for Regulatory Activities (MedDRA). ## B. Repeated Injection Safety Study, 35GA09901E, and Extension Phase of 35GA0608 The safety of repeated use of DUROLANE is supported by one open label study conducted at two centers in Sweden, 35GO9901E, and an extension phase to the non-inferiority trial versus MPA, 35GA0608. In 35GA09901E, patients were offered a second injection of DUROLANE 3 months following the initial injection at approximately 7 months from the first injection. Safety assessments were performed at 2 and 4 weeks post-injection. In 35GA0608, patients were offered an injection of DUROLANE following completion of the 26-week blinded phase of the trial. PMA P170007: FDA Summary of Safety and Effectiveness Data {7} Safety assessments were performed at 2 weeks following the second injection and at 9 and 12 months post-initial injection. The adverse event rates of the three groups were comparable, and the adverse events from these studies are summarized in Table 5. Table 5: Summary of Related Adverse Events Reported for Repeat Injections of Durolane | Preferred term* | 35GA0608 1st injection: DUROLANE (n=163) | 35GA0608 1st injection: MPA (n=179) | 35GO9901E DUROLANE (n=53) | | --- | --- | --- | --- | | Related to product or injection procedure or both | | | | | Arthralgia | 30 (18.4%) | 31 (17.3%) | 9 (17.0%) | | Arthropathy | - | - | 2 (3.8%) | | Joint dislocation | 1 (0.6%) | - | - | | Joint effusion | 1 (0.6%) | - | - | | Joint stiffness | 1 (0.6%) | 3 (1.7%) | - | | Joint swelling | 2 (1.2%) | 1 (0.6%) | - | | Joint warmth | - | 1 (0.6%) | - | | Musculoskeletal discomfort | 3 (1.8%) | - | - | | Urticaria | - | 1 (0.6%) | - | * AEs in the 35GO9901E study were classified using World Health Organisation, Adverse Reaction Terminology (WHO ART) and Medical Dictionary for Regulatory Activities (MedDRA) for 35GA0308. ## C. Study Design of Pivotal Trial, TG1018DLN The safety and effectiveness of DUROLANE was studied in pivotal trial, TG1018DLN – “Comparative Study of Safety and Efficacy of Two Hyaluronic Acids for the Treatment of Knee Osteoarthritis.” The database for this PMA reflected data collected between January 10, 2011, and February 2, 2012, and 349 patients. There were 7 investigational sites in the People’s Republic of China. Study TG1018DLN was a prospective, randomized, controlled, multicenter clinical study intended to demonstrate that DUROLANE was non-inferior to a commercially-available, 5-injection regimen hyaluronic acid (HA) product in the treatment of pain associated with knee OA over 18 weeks. The primary outcome measure was based on the Western Ontario and McMaster Universities Osteoarthritis Index (WOMAC) 20-point Likert-scale pain score. The non-inferiority margin was established as 8% (i.e., +1.6 units of the Likert-scale). Other outcome measures collected included WOMAC subscale domains of stiffness and physical function, along with subject global assessment. ## 1. Clinical Inclusion and Exclusion Criteria Enrollment in the TG1018DLN study was limited to patients who met the following inclusion criteria: - Subject (female or male) with 40-80 years of age, inclusive. PMA P170007: FDA Summary of Safety and Effectiveness Data {8} - Documented diagnosis of mild to moderate OA of the studied knee and fulfilled the American College of Rheumatology (ACR) criteria*. *ACR criteria for diagnosis of knee OA: Knee pain plus osteophytes on radiographs and at least one of the following conditions: age &gt; 50 years, morning stiffness &lt; 30 minutes in duration and crepitus on motion. - Radiographic evidence of OA in the studied knee (Kellgren-Lawrence [KL] radiographic score = 2 or 3), obtained from postero-anterior view of standing weight-bearing semi-flexed radiographs of the studied knee assessed by an assigned x-ray film reader in a study center. - WOMAC pain score from 7 to 17 (with WOMAC Likert 3.1 A) in the studied knee at both the screening and baseline visit. - Pain in the studied knee demonstrating a score of 2 or 3 (i.e. moderate or severe) with WOMAC Likert 3.1 A1 pain scoring. Patients were not permitted to enroll in the TG1018DLN study if they met any of the following exclusion criteria: - Clinically apparently tense effusion of the studied knee on examination determined by either a positive bulge sign or positive ballottement of the patella (patellar tap). - Symptomatic OA of the contralateral knee or of either hip that was not responsive to acetaminophen (paracetamol) and/or required any therapies prohibited in the protocol. - Low back pain requiring chronic ongoing analgesic therapy that confounded measurement of pain in the studied knee. - Systemic or intra-articular injection of corticosteroids in any joint within 6 months prior to screening. - Intra-articular HA injection in any joint including the studied knee within 9 months prior to screening. - Treatment with glucosamine-chondroitin sulfate initiated within the past 3 months, or dosage not stable for the past 3 months. - Serious injuries to the studied knee in the past resulting in significant restriction in ambulation. - Arthroscopy and/or other surgical procedure in the studied knee within the past 12 months. - Pregnant or breastfeeding women or women with potential childbearing without adequate contraception measure. 2. Follow-up Schedule Patients were followed for 26 weeks. Effectiveness was assessed at Weeks 6, 10, 14, 18, and 26. Safety was assessed at screening and at Weeks 0, 1, 2, 3, 4, 6, PMA P170007: FDA Summary of Safety and Effectiveness Data {9} 10, 14, 18, and 26. To address patient blinding due to the different injection regimens for the products, the DUROLANE group was given 3 mL at Week 0 (baseline) and received subcutaneous skin punctures (i.e., the needle did not enter the joint space) with an empty syringe at Weeks 1, 2, 3, and 4. The 5-injection HA group was administered using 2.5 mL injections of product at the same five time points. Before the baseline visit, the current use of analgesics was required to have elapsed by at least 5 half-lives; within 48 hours before each visit, patients were not allowed to take any acetaminophen (paracetamol) or any other analgesic. Acetaminophen was permitted as rescue analgesia during the course of the study. ## 3. Clinical Endpoints With regards to safety, safety assessments included collection of adverse events (AEs), physical examinations (including the knee), vital sign assessments, and blood and urine laboratory assessments. With regards to effectiveness, the primary effectiveness variable was the WOMAC Likert pain subscale score Change from Baseline (CFB) over Weeks 6, 10, 14, and 18. Secondary effectiveness variables were analyzed in a step-wise, hierarchical fashion; these variables included: - WOMAC physical function subscale score CFB - Subject global assessment CFB - WOMAC stiffness subscale score CFB Variables were tested for non-inferiority in the order as listed above. The same 8% non-inferiority margin utilized for the primary effectiveness variable was used for all secondary variables. Testing was first conducted over Weeks 6, 10, 14, and 18 for a given variable; contingent upon a determination of non-inferiority, testing was extended out through Week 26. Each variable tested over these two timeframes required a conclusion of non-inferiority in order to continue hierarchical non-inferiority testing. With regards to success/failure criteria, the primary effectiveness analysis involved a non-inferiority test of the WOMAC pain subscale variable using a Mixed Model for Repeated Measures (MMRM) over 18 weeks. The non-inferiority margin was established as +1.6 units of the Likert 20-point scale, corresponding to 8 units on a 100-point scale (i.e., 8%). PMA P170007: FDA Summary of Safety and Effectiveness Data Page 10 {10} D. Accountability of PMA Cohort At the time of database lock, a total of 404 subjects were screened at seven (7) sites. There were 349 (86.4%) subjects (175 DUROLANE and 174 5-injection HA) who received the study treatment after randomization. Among subjects treated, there were about 5% of subjects who did not complete the 18-week primary analysis period. For the 17 subjects who did not complete Visit 10 (Week 18), five refused to participate; five were due to protocol violations; three were due to adverse events (none of which were device-related); three were due to other reasons; and one subject was lost to follow-up. A total of 331 (94.8%) subjects (168 DUROLANE and 163 5-injection HA) received all five HA injections or a single injection of DUROLANE and four needle-sticks; had at least one effectiveness assessment; and met all inclusion/exclusion criteria, therein forming the Full Analysis Set (FAS) population. A total of 319 (91.4%) subjects (161 DUROLANE and 158 5-injection HA) were included in the Per Protocol Set (PPS) population. Twelve (12) subjects from the FAS were excluded from the PPS due to various protocol deviations. E. Study Population Demographics and Baseline Parameters The study was conducted in the People's Republic of China, and all patients were of Chinese ethnicity. Demographic and baseline characteristics were balanced between the two groups; see Table 6. Table 6: Demographic Data and Baseline Characteristics | Variable | Measure | DUROLANE | 5-injection HA | TOTAL | | --- | --- | --- | --- | --- | | | | (N=161) | (N=158) | (N=319) | | Age (years) | Mean (SD) | 60.2 (8.1) | 60.4 (7.8) | 60.3 (7.9) | | | Median | 60 | 59 | 59 | | | Min; Max | 40; 78 | 42; 78 | 40; 78 | | Sex [n (%)] | Female | 119 (73.9) | 127 (80.4) | 246 (77.1) | | | Male | 42 (26.1) | 31 (19.6) | 73 (22.9) | | | Total | 161 (100.0) | 158 (100.0) | 319 (100.0) | | Nationality [n (%)] | Han | 155 (96.3) | 157 (99.4) | 312 (97.8) | | | Other | 6 (3.7) | 1 (0.6) | 7 (2.2) | | | Total | 161 (100.0) | 158 (100.0) | 319 (100.0) | | Weight (kg) | n | 161 | 158 | 319 | | | Mean (SD) | 66.5 (10.2) | 66.8 (10.8) | 66.6 (10.5) | | | Median | 65.0 | 67.0 | 65.0 | PMA P170007: FDA Summary of Safety and Effectiveness Data {11} | | Min; Max | 44.0; 100.0 | 44.5; 106.0 | 44.0; 106.0 | | --- | --- | --- | --- | --- | | Height (cm) | n | 161 | 158 | 319 | | | Mean (SD) | 162.5 (6.7) | 162.4 (7.7) | 162.4 (7.2) | | | Median | 162 | 160 | 162.0 | | | Min; Max | 147; 183 | 145; 190 | 145; 190 | | BMI (kg/m2) | n | 161 | 158 | 319 | | | Mean (SD) | 25.1 (3.2) | 25.3 (3.2) | 25.2 (3.2) | | | Median | 24.8 | 25.1 | 25.0 | | | Min; Max | 18.4; 33.9 | 19.0; 35.0 | 18.4; 35.0 | | BMI classification | Underweight | 1 (0.6) | 0 (0.0) | 1 (0.3) | | | Normal range | 81 (50.3) | 76 (48.1) | 157 (49.2) | | | Overweight | 65 (40.4) | 69 (43.7) | 134 (42.0) | | | Obese | 14 (8.7) | 13 (8.2) | 27 (8.5) | | | Total | 161 (100.0) | 158 (100.0) | 319 (100.0) | The basic demographics and baseline characteristics of the Chinese population studied in TG1018DLN were also found to be reasonably similar to other non-inferiority HA studies conducted in both China and the United States. ## F. Safety and Effectiveness Results ### 1. Safety Results The analysis of safety was based on the prospective and randomized cohort of 349 patients available for the 18-week evaluation. The key safety outcomes for this study are presented below in Tables 7 and 8. There were three subjects (1 DUROLANE; 2 5-injection HA) who discontinued the study due to an AE; none of these were considered device-related. **Adverse Effects that occurred in the PMA clinical study:** Subjects with device-related AEs were 13.1% and 9.8% in the DUROLANE and 5-injection HA groups, respectively. The most common device-related AE was arthralgia for both groups (DUROLANE: 8.6%; 5-injection HA: 7.5%). The severities of device-related AEs in both groups were mainly mild and moderate; only one injection site pain in the DUROLANE group and two cases of arthralgia and one case of joint swelling in the 5- INJECTION HA group were classified as severe. The proportion of subjects with Serious Adverse Events (SAEs) was 1.7% (3/175) and 3.4% (6/174) in the DUROLANE and 5-injection HA groups, PMA P170007: FDA Summary of Safety and Effectiveness Data {12} respectively. No SAEs were considered related to the investigational products. No deaths occurred in this study. The adverse events of the two groups were comparable. Table 7: Summary of Adverse Events – safety set (SS) | | DUROLANE | 5-injection HA | | --- | --- | --- | | | (N=175) | (N=174) | | | n (%) | n (%) | | Subjects with at least one treatment emergent AE | 83 (47.4) | 74 (42.5) | | Subjects with device-related AEs | 23 (13.1) | 17 (9.8) | | Subjects with a severe AE | 6 (3.4) | 8 (4.6) | | Subjects with a SAEs | 3 (1.7) | 6 (3.4) | | Subjects with a device-related SAE | - | - | | Death | - | - | Table 8: Subjects with Device Related AEs | Primary System Organ Classification (SOC) Preferred term | DUROLANE | 5-injection HA | | --- | --- | --- | | | (N=175) | (N=174) | | | n (%) | n (%) | | General disorders and administration site conditions | 4 (2.3) | 2 (1.1) | | Injection site pain | 4 (2.3) | 2 (1.1) | | Musculoskeletal and connective tissue disorders | 18 (10.3) | 16 (9.2) | | Arthralgia | 15 (8.6) | 13 (7.5) | | Joint swelling | 3 (1.7) | 3 (1.7) | | Arthropathy | 1 (0.6) | - | | Epicondylitis | - | 1 (0.6) | | Joint effusion | 1 (0.6) | - | | Limb discomfort | - | 1 (0.6) | | Muscular weakness | 1 (0.6) | - | | Musculoskeletal discomfort | 1 (0.6) | - | | Myalgia | - | 1 (0.6) | | Pain in extremity | 1 (0.6) | - | | Skin and subcutaneous tissue disorders | 1 (0.6) | - | | Erythema | 1 (0.6) | - | PMA P170007: FDA Summary of Safety and Effectiveness Data Page 13 {13} # 2. Effectiveness Results The analysis of effectiveness was based on the 319 evaluable patients at the 26-week time point. Key effectiveness outcomes are presented in Table 9. # Primary Effectiveness Endpoint The analysis of the effectiveness of DUROLANE was based on the PPS $(n = 319)$ . The results demonstrated that DUROLANE was non-inferior to the 5-injection HA. The Least Square Mean (LSM) WOMAC pain subscale score change from baseline (CFB) over 18 weeks was -5.97 for DUROLANE and -5.87 for the 5-injection HA, with a difference (DUROLANE - 5-injection HA) of -0.09 (95% CI:-0.58, 0.39). See Table 9 for a tabular summary of the results of the primary endpoint by visit. Table 9: Mixed Model Repeated Measures (MMRM) Analysis of WOMAC Pain Subscale Score CFB by Visit -PPS | Visit (week) | Measure | Actual Result | | Change from Baseline | | Baseline Change Difference (95%CI)a | | --- | --- | --- | --- | --- | --- | --- | | | | DUROLANE (N=161) | 5-injection HA (N=158) | DUROLANE (N=161) | 5-injection HA (N=158) | DUROLANE - 5-injection HA | | Baseline (Week 0) | Mean(SD)b | 9.4 (1.98) | 9.5 (1.80) | | | | | Visit 7 (Week 6) | Mean(SD) | 4.6 (3.32) | 4.6 (2.93) | -4.9 (3.16) | -5.0 (2.68) | | | | LSMc(95% CI) | | | -5.02 (-5.46; -4.58) | -5.06 (-5.50; -4.61) | 0.04 (-0.58; 0.65) | | | p-value | | | <0.0001 | <0.0001 | 0.91 | | Visit 8 (Week 10) | Mean(SD) | 3.7 (3.19) | 3.7 (2.81) | -5.7 (3.03) | -5.8 (2.65) | | | | LSM (95% CI) | | | -5.45 (-5.85; -5.05) | -5.49 (-5.89; -5.09) | 0.04 (-0.51; 0.59) | | | p-value | | | <0.0001 | <0.0001 | 0.89 | | Visit 9 (Week 14) | Mean(SD) | 3.2 (2.90) | 3.4 (2.69) | -6.2 (2.87) | -6.1 (2.59) | | | | LSM (95% CI) | | | -5.76 (-6.14; -5.39) | -5.73 (-6.11; -5.36) | -0.03 (-0.54; 0.48) | | | p-value | | | <0.0001 | <0.0001 | 0.91 | | Visit 10 (Week 18) | Mean(SD) | 3.0 (2.88) | 3.3 (2.75) | -6.5 (2.79) | -6.2 (2.75) | | | | LSM (95% CI) | | | -5.97 (-6.32; -5.61) | -5.87 (-6.23; -5.52) | -0.09 (-0.58; 0.39) | | | p-value | | | <0.0001 | <0.0001 | 0.70 | | Visit 11 (Week 26) | Mean(SD) | 2.8 (2.73) | 2.9 (2.68) | -6.6 (2.67) | -6.6 (2.58) | | | | LSM (95% CI) | | | -6.15 (-6.49; -5.81) | -6.05 (-6.39; -5.71) | -0.10 (-0.56; 0.37) | | | p-value | | | <0.0001 | <0.0001 | 0.68 | PMA P170007: FDA Summary of Safety and Effectiveness Data {14} a $\mathrm{CI} =$ confidence interval b $\mathrm{SD} =$ standard deviation [Note: SD are descriptive statistics using at visit observations.] c LSM = Least Square Mean [Note: LSM (95% CI) and p-values are repeated measure statistics using at visit and all preceding visit observations.] As the upper bound of the confidence interval did not exceed the pre-specified non-inferiority margin of $+1.6$ , non-inferiority was established. See Figure 1. Figure 1: MMRM $95\%$ CI Plot ![img-0.jpeg](img-0.jpeg) Note: confidence interval information is cumulative first week 6, then week 6 and 10, etceteras to week 26. # Secondary Effectiveness Endpoints The same $8\%$ non-inferiority margin utilized for the primary effectiveness variable was used for all secondary variables. Results through 26 weeks for the WOMAC pain subscale score CFB are included in Table 9 above. The remaining secondary variables were tested for non-inferiority in a stepwise order as outlined in Table 10 below. PMA P170007: FDA Summary of Safety and Effectiveness Data {15} Table 10: Results of Stepwise Non-inferiority Analyses of Other Secondary Effectiveness Variables | Secondary Variable (order of importance) | Baseline Change Mean (SD) | | (DUROLANE-5-injection HA) LSMa (95%CIb) | Non-Inferiority 8% Margins | Conclusion | | --- | --- | --- | --- | --- | --- | | | DUROLANE | 5-injection HA | | | | | WOMAC Physical Function CFB (over 18 weeks) | -12.75 (-13.60; -11.91) | -12.10 (-12.95; -11.26) | -0.65 (-1.81,0.51) | +5.44 | Non-inferior | | WOMAC Physical Function CFB (over 26 weeks) | -12.58 (-13.39; -11.77) | -13.16 (-13.97; -12.35) | -0.58 (-1.69,0.53) | +5.44 | Non-inferior | | Subject Global Assessment CFB (over 18 weeks) | 2.70 (2.48; 2.92) | 2.55 (2.33; 2.77) | 0.15 (-0.15,0.45) | -0.8 | Non-inferior | | Subject Global Assessment CFB (over 26 weeks) | 2.81 (2.59; 3.02) | 2.67 (2.45; 2.88) | 0.14 (-0.16,0.43) | -0.8 | Non-inferior | | WOMAC Knee Stiffness CFB (over 18 weeks) | -1.87 (-2.00; -1.73) | -1.73 (-1.87; -1.59) | -0.14 (-0.33,0.05) | +0.64 | Non-inferior | | WOMAC Knee Stiffness CFB (over 26 weeks) | -1.95 (-2.08; -1.82) | -1.80 (-1.93; -1.67) | -0.15 (-0.33,0.03) | +0.64 | Non-inferior | $^{\mathrm{a}}$ LSM = Least Square Mean ${}^{\mathrm{b}}$ CI = confidence interval ## 3. Subgroup Analyses Subgroup analyses were conducted on the PPS and included subjects with different age, gender, and osteoarthritis condition (e.g. joint fluid removal and KL grade). Results showed that at Weeks 18 and 26, the DUROLANE WOMAC pain subscale scores in all subgroup populations were non-inferiority to the 5-injection HA group. The only exception involved subjects who had joint fluid aspirated at baseline. However, the results were not superior for the 5-injection HA group and the number of subjects in this subgroup were too small to draw any conclusions based on these findings. Additionally, there was no statistically significant difference between DUROLANE and the 5-injection HA group in the corresponding WOMAC pain responder rate at Weeks 18 and 26 in all subgroup populations, inclusive of the joint aspirated subgroup. ## 4. Pediatric Extrapolation In this premarket application, existing clinical data were not leveraged to support approval of a pediatric patient population. ## G. Financial Disclosure PMA P170007: FDA Summary of Safety and Effectiveness Data {16} The Financial Disclosure by Clinical Investigators regulation (21 CFR 54) requires applicants who submit a marketing application to include certain information concerning the compensation to, and financial interests and arrangement of, any clinical investigator conducting clinical studies covered by the regulation. The pivotal clinical study was conducted in the People's Republic of China and included 7 investigators. Financial disclosure was not conducted before or during the study, and there were no responses from the investigators during attempts to collect these disclosures retrospectively. However, the information provided does not raise any questions about the reliability of the data. ## XI. PANEL MEETING RECOMMENDATION AND FDA'S POST-PANEL ACTION In accordance with the provisions of section 515(c)(3) of the act as amended by the Safe Medical Devices Act of 1990, this PMA was not referred to the Orthopaedic and Rehabilitation Devices Panel, an FDA advisory committee, for review and recommendation because the information in the PMA substantially duplicates information previously reviewed by this panel. ## XII. CONCLUSIONS DRAWN FROM PRECLINICAL AND CLINICAL STUDIES ### A. Effectiveness Conclusions The primary clinical evidence for reasonable assurance of the effectiveness of DUROLANE was the results from a 26-week randomized controlled clinical non-inferiority trial comparing DUROLANE to a commercially available HA with 319 evaluable patients. The results demonstrated that DUROLANE was non-inferior to the 5-injection HA. The LSM WOMAC pain subscale score CFB over 18 weeks was -5.97 for DUROLANE and -5.87 for the 5-injection HA, with a difference (DUROLANE - 5-injection HA) of -0.09 (95% CI: -0.58, 0.39). See Table 9 for a tabular summary of the results of the primary endpoint by visit. ### B. Safety Conclusions The risks of the device are based on the pre-clinical laboratory studies, animal studies, and clinical studies conducted to support PMA approval as have been described in the preceding sections. Pre-clinical studies have shown that DUROLANE is biocompatible. Results from the pivotal clinical trial showed comparable adverse events rates between DUROLANE and a multi-injection commercially available HA product. The most common side effects include transient pain and swelling, which are common types of reactions for any type of intra-articular therapy. Additional clinical studies have not shown any safety concerns related to its use, including its use in patients receiving a repeated course of therapy. PMA P170007: FDA Summary of Safety and Effectiveness Data Page 17 {17} C. Benefit-Risk Determination The probable benefits of the device are also based on data collected in clinical studies conducted to support PMA approval as described above. A single injection of DUROLANE provides probable benefit for pain reduction in patients with osteoarthritis in the knee. An additional benefit of DUROLANE is the ability for patients to be treated with a single injection versus a series of injections required of multi-injection HA formulations. The data support the conclusion that the benefits outweigh the risks of transitory adverse events such as pain and swelling in treatment of pain associated with osteoarthritis of the knee. 1. Patient Perspectives Patient perspectives considered during the review included patient-reported global assessments of pain that were collected in the pivotal trial as a secondary endpoint. Mean reductions from baseline in these reported pain assessments are summarized in Table 10. In conclusion, given the available information above, the data support that for the treatment of pain in osteoarthritis of the knee in patients who have failed to respond adequately to conservative non-pharmacological therapy or simple analgesics the probable benefits outweigh the probable risks. D. Overall Conclusions The data contained within the PMA support the reasonable assurance of safety and effectiveness of DUROLANE when used in accordance with its labeled indications. Pre-clinical and clinical data support the safety and effectiveness of a single injection of DUROLANE. Results from a randomized, controlled non-inferiority pivotal study demonstrated that DUROLANE was non-inferior to a commercially available 5-injection HA for the treatment of pain in OA of the knee. Results from this pivotal study also showed the adverse event rates and safety profiles of DUROLANE and the 5-injection HA to be comparable. Clinical data in the PMA also supported the safe, repeated use of the product. XIII. CDRH DECISION CDRH issued an approval order on August 29, 2017. The applicant’s manufacturing facilities have been inspected and found to be in compliance with the device Quality System (QS) regulation (21 CFR 820). XIV. APPROVAL SPECIFICATIONS Directions for use: See device labeling. PMA P170007: FDA Summary of Safety and Effectiveness Data Page 18 {18} Hazards to Health from Use of the Device: See Indications, Contraindications, Warnings, Precautions, and Adverse Events in the device labeling. Post-approval Requirements and Restrictions: See approval order. PMA P170007: FDA Summary of Safety and Effectiveness Data Page 19
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...