← Product Code [OBF](/productcode/OBF) · P160016

# VERSANT HCV GENOTYPE 2.0 ASSAY (LiPA) (P160016)

_Fujirebio Europe N.V. · OBF · Mar 14, 2017 · Microbiology · APRL_

**Canonical URL:** https://fda.innolitics.com/device/P160016

## Device Facts

- **Applicant:** Fujirebio Europe N.V.
- **Product Code:** [OBF](/productcode/OBF.md)
- **Decision Date:** Mar 14, 2017
- **Decision:** APRL
- **Regulation:** 21 CFR 866.3170
- **Device Class:** Class 2
- **Review Panel:** Microbiology
- **Attributes:** Real-World Evidence

## Real-World Evidence

| Submission | Device | Sponsor | RWD Sources | RWE Use Summary | Key Tags |
| --- | --- | --- | --- | --- | --- |
| P160016 · Mar 14, 2017 | VERSANT HCV GENOTYPE 2.0 ASSAY (LiPA) | Fujirebio Europe N.V. | Retrospective banked clinical trial samples | Retrospective analysis of banked patient samples with known HCV treatment outcomes was used to demonstrate the clinical validity of the assay by correlating genotype results with sustained virologic response (SVR12). | Retrospective study; Banked samples; Clinical validity; Treatment outcome |

### Clinical Evidence

| Study Design | Population | Comparator | Key Endpoints |
| --- | --- | --- | --- |
| Clinical Validity Study; Retrospective study using banked samples; Follow-up/Duration: Not applicable (retrospective) | 217 individuals with chronic HCV infection (genotypes 1a, 1b, 2, 3, 4, 5, 6) with known treatment outcomes; Sample Size: 217 evaluable subjects; Number of Sites: 3 | NS5b HCV genotyping reference method | Association between HCV genotype and probability of achieving sustained virologic response (SVR12) |

## Indications for Use

The VERSANT® HCV Genotype 2.0 Assay (LiPA) is a line probe assay, for in vitro diagnostic use, which identifies Hepatitis C virus (HCV) genotypes 1 to 6 and subtypes 1a and 1b in human serum or plasma (K₂-EDTA, ACD-A CPD, and CPDA) samples. The VERSANT® HCV Genotype 2.0 Assay (LiPA) is intended to be used as an aid in the management of patients with chronic HCV infection to guide the selection of antiviral treatment. The VERSANT® HCV Genotype 2.0 Assay (LiPA) is not approved for use as a donor screening test for HCV or as a diagnostic test to confirm the presence of HCV. The VERSANT® Sample Preparation 1.0 Reagents kit is intended to be used, along with the VERSANT® kPCR Molecular System Sample Preparation module (SP Module), for the isolation and purification of nucleic acids for in vitro diagnostic applications. The product is intended for use by laboratory professionals trained in molecular biology techniques. The VERSANT® HCV Amplification 2.0 Kit (LiPA) is designed for use with the VERSANT® HCV Genotype 2.0 Assay (LiPA) and the VERSANT® HCV Control 2.0 Kit (LiPA) products. The HCV Amplification 2.0 kit is for in vitro diagnostic use for reverse transcription and amplification of the 5' untranslated region (5'UTR) and core region of the Hepatitis C Virus (HCV) genome. The VERSANT® HCV Control 2.0 Kit (LiPA) is designed for use with the VERSANT® HCV Amplification 2.0 Kit (LiPA) and the VERSANT® HCV Genotype 2.0 Assay (LiPA) products. The VERSANT® HCV Control 2.0 Kit (LiPA) is for in vitro diagnostic use to monitor the performance of all steps of the VERSANT® HCV Genotype 2.0 assay.

## Device Story

System uses serum/plasma samples; automated nucleic acid extraction via magnetic silica beads on VERSANT kPCR Molecular System SP. Viral RNA reverse-transcribed to cDNA; biotinylated primers amplify 5' UTR and core regions via PCR. Amplified products undergo reverse hybridization to sequence-specific oligonucleotide probes immobilized on nitrocellulose membrane strips (LiPA). Streptavidin-alkaline phosphatase conjugate and BCIP/NBT chromogen substrate produce visible purple-brown bands. Results interpreted manually via reading card/chart or semi-automated via LiPA-Scan software. Used by trained laboratory professionals in clinical settings. Output identifies HCV genotype/subtype to guide antiviral therapy selection; incorrect results may delay optimal treatment.

## Clinical Evidence

Retrospective clinical validity study (n=217 evaluable) compared LiPA results to NS5b reference sequencing and treatment outcomes (SVR12). Genotyping accuracy ranged from 98.6% to 100% across genotypes. SVR12 rates ranged from 72.5% to 100% depending on genotype. Analytical studies established LoD (400 IU/mL plasma, 650 IU/mL serum), specificity, and reproducibility. No interference observed from endogenous substances, microorganisms, or antiviral drugs.

## Technological Characteristics

Line probe assay (LiPA) using reverse hybridization. Components: nitrocellulose strips with immobilized oligonucleotide probes, magnetic silica bead extraction, RT-PCR amplification (5' UTR and core regions). Detection via streptavidin-alkaline phosphatase conjugate and BCIP/NBT substrate. Manual or semi-automated (LiPA-Scan software) interpretation. Storage: 2-8°C for strips/reagents, -20°C for amplification kit.

## Regulatory Identification

A nucleic acid-based hepatitis C virus (HCV) ribonucleic acid (RNA) test is identified as an in vitro diagnostic device intended for prescription use as an aid in the diagnosis of HCV infection in specified populations, and/or as an aid in the management of HCV-infected patients including guiding the selection of genotype-specific treatment in individuals with chronic HCV infection. The test is intended for use with human serum or plasma. The test is not intended for use as a donor screening test for the presence of HCV antibodies in blood, blood products, or tissue donors.

## Special Controls

*Classification.* Class II (special controls). The special controls for this device are:(1) For all nucleic acid-based HCV RNA tests, the labeling required under § 809.10(b) of this chapter must include:
(i) A prominent statement that the test is not intended for use as a donor screening test for the presence of HCV RNA from human cells, tissues, and cellular and tissue-based products.
(ii) A detailed explanation of the principles of operation and procedures for performing the assay.
(iii) A detailed explanation of the interpretation of results.
(iv) Limitations, which must be updated to reflect current clinical practice and disease presentation and management. These limitations must include, but are not limited to, statements that indicate:
(A) The specimen types for which the device has been cleared and that use of this test kit with specimen types other than those specifically cleared for this device may result in inaccurate test results.
(B) When applicable, that assay performance characteristics have not been established in populations of immunocompromised or immunosuppressed patients or, other populations where test performance may be affected.
(C) Test results are to be interpreted by qualified licensed healthcare professionals in conjunction with the individual's clinical presentation, history, and other laboratory results.
(2) For all nucleic acid-based HCV RNA tests, the design verification and validation must include:
(i) Detailed device description, including the device components, ancillary reagents required but not provided, and an explanation of the device methodology. Additional information appropriate to the technology must be included such as design of primers and probes, rationale for the selected gene targets, specifications for amplicon size, and degree of nucleic acid sequence conservation.
(ii) For devices with assay calibrators, the design and nature of all primary, secondary, and subsequent quantitation standards used for calibration as well as their traceability to a standardized reference material that FDA has determined is appropriate (
*e.g.,* a recognized consensus standard). In addition, analytical testing must be performed following the release of a new lot of the standard material that was used for device clearance or approval, or when there is a transition to a new calibration standard.(iii) Documentation and characterization (
*e.g.,* determination of the identity, supplier, purity, and stability) of all critical reagents (including nucleic acid sequences for primers and probes) and protocols for maintaining product integrity.(iv) Detailed documentation of analytical performance studies conducted as appropriate to the technology, specimen types tested, and intended use of the device, including, but not limited to, limit of detection (LoD), upper and lower limits of quantitation (ULoQ and LLoQ, respectively), linearity, precision, endogenous and exogenous interferences, cross reactivity, carryover, matrix equivalency, and sample and reagent stability. Samples selected for use in analytical studies or used to prepare samples for use in analytical studies must be from subjects with clinically relevant circulating genotypes in the United States. Cross-reactivity studies must include samples from HCV RNA negative subjects with other causes of liver disease, including autoimmune hepatitis, alcoholic liver disease, chronic hepatitis B virus, primary biliary cirrhosis, and nonalcoholic steatohepatitis, when applicable. The effect of each claimed nucleic-acid isolation and purification procedure on detection must be evaluated.
(v) Risk analysis and management strategies, such as Failure Modes Effects Analysis and/or Hazard Analysis and Critical Control Points summaries and their impact on test performance.
(vi) Final release criteria to be used for manufactured test lots with appropriate evidence that lots released at the extremes of the specifications will meet the claimed analytical and clinical performance characteristics as well as the stability claims.
(vii) Multisite reproducibility study that includes the testing of three independent production lots.
(viii) All stability protocols, including acceptance criteria.
(ix) Final release test results for each lot used in clinical studies.
(x) Analytical sensitivity and specificity of the test must be the same or better than that of other cleared or approved tests.
(xi) Lot-to-lot precision studies, as appropriate.
(3) For devices intended for the qualitative detection of HCV RNA, in addition to the special controls listed in paragraphs (b)(1) and (2) of this section, the design verification and validation must include detailed documentation of performance from a multisite clinical study. Performance must be analyzed relative to an FDA cleared or approved qualitative HCV RNA test, or a comparator that FDA has determined is appropriate. This study must be conducted using appropriate patient samples, with appropriate numbers of HCV positive and negative samples in applicable risk categories. Additional genotypes must be validated using appropriate numbers and types of samples. The samples may be a combination of fresh and repository samples, sourced from within and outside the United States, as appropriate. The study designs, including number of samples tested, must be sufficient to meet the following criteria:
(i) Clinical sensitivity of the test must have a lower bound of the 95 percent confidence interval of greater than or equal to 95 percent.
(ii) Clinical specificity of the test must have a lower bound of the 95 percent confidence interval of greater than or equal to 96 percent.
(4) For devices intended for the quantitative detection of HCV RNA, the following special controls, in addition to those listed in paragraphs (b)(1) and (2) of this section, apply:
(i) Labeling required under § 809.10(b) of this chapter must include a prominent statement that the test is not intended as a diagnostic test to confirm the presence of active HCV infection, when applicable.
(ii) Design verification and validation must include the following:
(A) Detailed documentation of the following analytical performance studies conducted as appropriate to the technology, specimen types tested, and intended use of the device, including but not limited to: LoD, ULoQ and LLoQ. LoD, LLoQ, and linearity studies must demonstrate acceptable device performance with all HCV genotypes detected by the device.
(B) Detailed documentation of clinical performance testing from either:
(
*1* ) A multisite clinical study with an appropriate number of clinical samples from chronically HCV infected patients in which the results are compared to an FDA-cleared or approved quantitative HCV RNA test, or a comparator that FDA has determined is appropriate. This study must include a sufficient number of HCV positive samples containing an analyte concentration near the LLoQ to describe performance at this level. Clinical samples must cover the full range of the device output and must be consistent with the distribution of these genotypes in the U.S. population. Clinical samples may be supplemented with diluted clinical samples for those viral load concentrations that are not sufficiently covered by natural clinical specimens, or(
*2* ) A clinical study with prospectively collected samples demonstrating clinical validity of the device.(C) Detailed documentation of a qualitative analysis near the lower end of the measuring range demonstrating acceptable performance when used as an aid in diagnosis.
(5) For devices intended for HCV RNA genotyping, in addition to the special controls listed in paragraphs (b)(1) and (2) of this section, design verification and validation must include the following:
(i) Detailed documentation of an analytical performance study demonstrating the LoD for all HCV genotypes detected by the device.
(ii) Detailed documentation, including results, of a multisite clinical study that assesses genotyping accuracy (
*i.e.,* the proportion of interpretable results that match with the reference method result) and the genotyping rate (*i.e.,* the proportion of results that were interpretable).(6) For any nucleic acid-based HCV RNA test intended for Point of Care (PoC) use, the following special controls, in addition to those listed in paragraphs (b)(1) and (2) of this section, apply:
(i) Clinical studies must be conducted at PoC sites.
(ii) Additional labeling must include a brief summary of the instructions for use that are appropriate for use in a PoC environment.

## Reference Devices

- Abbott RealTime HCV Genotype II assay

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a PMA](https://innolitics.com/services/regulatory/), [a 510(k)](https://innolitics.com/services/510ks/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

SUMMARY OF SAFETY AND EFFECTIVENESS DATA (SSED)

I. GENERAL INFORMATION

Device Generic Name: Genotyping assay for hepatitis C virus

Device Trade Name: VERSANT® Sample Preparation 1.0 Reagents
VERSANT® HCV Amplification 2.0 Kit (LiPA)
VERSANT® HCV Genotype 2.0 Assay (LiPA)
VERSANT® HCV Control 2.0 Kit (LiPA)

Device Procode: OBF

Applicant's Name and Address: Siemens Healthcare Diagnostics
725 Potter Street
Berkeley, CA 94710

Date(s) of Panel Recommendation: None

Premarket Approval Application (PMA) Number: P160016

Date of FDA Notice of Approval: March 14, 2017

II. INDICATIONS FOR USE

The genotyping device for hepatitis C virus (LiPA) consists of the following 4 separate kits:

- VERSANT® HCV Genotype 2.0 Assay (LiPA)
- VERSANT® Sample Preparation 1.0 Reagents
- VERSANT® HCV Amplification 2.0 Kit (LiPA)
- VERSANT® HCV Control 2.0 Kit (LiPA)

VERSANT® HCV Genotype 2.0 Assay (LiPA)

The VERSANT® HCV Genotype 2.0 Assay (LiPA) is a line probe assay, for in vitro diagnostic use, which identifies Hepatitis C virus (HCV) genotypes 1 to 6 and subtypes 1a and 1b in human serum or plasma (K₂-EDTA, ACD-A CPD, and CPDA) samples.

The VERSANT® HCV Genotype 2.0 Assay (LiPA) is intended to be used as an aid in the management of patients with chronic HCV infection to guide the selection of antiviral treatment.

The VERSANT® HCV Genotype 2.0 Assay (LiPA) is not approved for use as a donor screening test for HCV or as a diagnostic test to confirm the presence of HCV.

VERSANT® Sample Preparation 1.0 Reagents

The VERSANT® Sample Preparation 1.0 Reagents kit is intended to be used, along with the VERSANT® kPCR Molecular System Sample Preparation module (SP Module), for the isolation and purification of nucleic acids for in vitro diagnostic

PMA P160016: FDA Summary of Safety and Effectiveness Data

{1}

applications. The product is intended for use by laboratory professionals trained in molecular biology techniques.

## VERSANT® HCV Amplification 2.0 Kit (LiPA)

The VERSANT® HCV Amplification 2.0 Kit (LiPA) is designed for use with the VERSANT® HCV Genotype 2.0 Assay (LiPA) and the VERSANT® HCV Control 2.0 Kit (LiPA) products. The HCV Amplification 2.0 kit is for in vitro diagnostic use for reverse transcription and amplification of the 5' untranslated region (5'UTR) and core region of the Hepatitis C Virus (HCV) genome.

## VERSANT® Control 2.0 Kit (LiPA)

The VERSANT® HCV Control 2.0 Kit (LiPA) is designed for use with the VERSANT® HCV Amplification 2.0 Kit (LiPA) and the VERSANT® HCV Genotype 2.0 Assay (LiPA) products.

The VERSANT® HCV Control 2.0 Kit (LiPA) is for in vitro diagnostic use to monitor the performance of all steps of the VERSANT® HCV Genotype 2.0 assay.

All four kits are required to extract nucleic acids from patient samples, serve as external controls, amplify specific regions of the HCV sequence, hybridize the amplified material to probes and identify the HCV genotype.

## III. CONTRAINDICATIONS

None

## IV. WARNINGS AND PRECAUTIONS

The warnings and precautions can be found in the labeling for the VERSANT® Sample Preparation 1.0 Reagents, VERSANT® HCV Amplification 2.0 Kit (LiPA), VERSANT® HCV Genotype 2.0 Assay (LiPA) and VERSANT® Control 2.0 Kit (LiPA).

## V. DEVICE DESCRIPTION

Four VERSANT® reagent kits are required to obtain an HCV genotype from a serum or plasma sample from a patient infected with HCV using the VERSANT® HCV Genotype 2.0 Assay (LiPA):

1. Extraction and Purification of Viral RNA

The VERSANT® kPCR Molecular System SP is a fully automated sample preparation instrument that utilizes VERSANT® Sample Preparation 1.0 Reagent for the isolation and purification of nucleic acids using magnetic bead extraction technology. The VERSANT® Sample Preparation 1.0 Reagent consists of Box 1 that contains buffers, beads and wash solutions and Box 2 (SMN 10286027) that contains the enzyme to lyse cells.

PMA P160016: FDA Summary of Safety and Effectiveness Data
Page 2

{2}

2. Reverse Transcription – Polymerase Chain Reaction

Amplification of the HCV RNA is performed using the VERSANT® HCV Amplification 2.0 Kit (LiPA). The kit contains 4 or 10 tubes of enzyme mix for the 40 test and 100 test kits, respectively, as well as either 1 or 2 tubes of amplification mix for the 40 test and 100 test kits, respectively. The enzyme mix consists of reverse transcriptases, a DNA polymerase and Uracil-N-Glycosylase. The amplification mix contains synthetic oligonucleotides in buffer with dNTP/dUTP and MgCl₂.

3. Genotyping with the VERSANT® HCV Genotype 2.0 Assay (LiPA)

The detection and identification of HCV genotypes using the VERSANT® HCV Genotype 2.0 Assay (LiPA) (SMN 10492004/40 tests and 10492005/100 tests) is based upon reverse hybridization of the RT-PCR (Reverse Transcriptase Polymerase chain reaction) amplified nucleic acid to HCV sequence specific oligonucleotides (probes) bound to a membrane strip. The VERSANT® HCV Genotype 2.0 Assay (LiPA) Kit contains the reagents listed in the table below.

PMA P160016: FDA Summary of Safety and Effectiveness Data
Page 3

{3}

Table 1: Components in the VERSANT® HCV Genotype 2.0 Assay (LiPA)

|  Component | Standard Kit Quantity | Large Kit Quantity | Description | Storage  |
| --- | --- | --- | --- | --- |
|  Strips | 2 x 20 | 5 x 20 | Nitrocellulose membranes coated with oligonucleotide probes. Identified by a green marker line | 2° to 8°C  |
|  100X Conjugate | 1 x 1.5 mL | 2 x 1.5 mL | Concentrated conjugate containing streptavidin- labeled with alkaline phosphatase with protein stabilizers and preservatives | 2° to 8°C  |
|  Conjugate Dilution Buffer | 1 x 150 mL | 2 x 150 mL | Conjugate diluent containing phosphate buffer with 0.1% 2- Chloroacetamide, detergents, protein stabilizers and other preservatives | 2° to 8°C  |
|  Denaturation Solution | 1 x 1 mL | 2 x 1 mL | Denaturation solution containing 1.7% sodium hydroxide | 2° to 8°C  |
|  Hybridization and Stringent Wash Solution | 2 x 220 mL | 5 x 220 mL | Hybridization and stringent wash solution containing sodium chloride, sodium citrate buffer with detergent and preservatives | 2° to 8°C  |
|  5X Rinse Solution | 1 x 150 mL | 2 x 150 mL | Concentrated rinse solution containing phosphate buffer with 0.5% 2- Chloroacetamide, NaCl, detergent and other preservatives | 2° to 8°C  |
|  Substrate Buffer | 2 x 180 mL | 3 x 180 mL | Substrate buffer containing TRIS buffer with 0.1% 2- Chloroacetamide, MgCl₂, NaCl and other preservatives. | 2° to 8°C  |
|  100X BCIP/NBT Substrate | 1 x 1.5 mL | 2 x 1.5 mL | Substrate containing 1.6% 5-Bromo-4-chloro-3- indolyl-phosphate and 4-Nitroblue tetrazolium in 83% dimethylformamide | 2° to 8°C  |
|  Reading Card | 1 | 1 | Used to identify positive bands on a strip |   |
|  Data Reporting Sheets | 4 | 10 | Pages used for reporting and storing the results of an assay |   |
|  Interpretation Chart | 1 | 1 | Provides banding patterns for identification of genotypes |   |

PMA P160016: FDA Summary of Safety and Effectiveness Data

{4}

PMA P160016: FDA Summary of Safety and Effectiveness Data
Page 5

4. Controls

The VERSANT® HCV Control 2.0 Kit (LiPA) (SMN 10719668) contains 2 tubes of negative control (human plasma, nonreactive for HCV) and 2 tubes of positive control (Armored RNA construct containing HCV 5' UTR (untranslated region) and Core sequences for Genotype 3 in human plasma that is nonreactive for HCV). The positive and negative controls require the same isolation and purification steps to extract the nucleic acids from the plasma matrix, as the test samples.

Sample Preparation

The VERSANT® kPCR Molecular System SP is a fully automated liquid handling workstation used in conjunction with VERSANT® Sample Preparation 1.0 Reagents to isolate nucleic acids. The VERSANT® kPCR Molecular System consists of the bar code reader, instrument and software. Assay parameters specific to the VERSANT® HCV Genotype 2.0 Assay (LiPA) assay are contained in an assay specific test definition file. The magnetic silica technology is a nonspecific capture method and is target independent; it captures any DNA or RNA present in the sample. Due to their extremely small size and homogeneous shape, the magnetic beads can be fully dispersed in solution, allowing thorough nucleic acid binding, washing and elution.

The VERSANT® Sample Preparation technology employs chaotropic high salt conditions to disrupt cells and release nucleic acids, as well as protect them from cellular nucleases. The nucleic acids are captured on silica coated beads. The beads are separated by a magnetic field and washed to remove proteins, nucleases, and other cellular impurities. The nucleic acids are eluted in a small volume of elution buffer, ready for subsequent analysis.

Two hundred and fifty microliters (250 μL) of patient sample, or Positive or Negative controls are processed using the automated VERSANT® kPCR Molecular System SP. The Positive and Negative controls from the VERSANT® Control HCV 2.0 Kit (LiPA) are processed along with the patient samples and can be used for run validation at all steps of the assay.

The VERSANT® kPCR Molecular System SP produces 50 μL of eluate and a volume of 20 μL is needed for the amplification step. The remainder of the eluate can be stored at -80 °C for further testing. Additionally, depending on time constraints, the user can continue with the next step or store the eluate at -80 °C until the next day.

Reverse Transcription and Amplification

Twenty microliters (20 μL) of extracted RNA is added to the RT-PCR mastermix prepared from reagents in the VERSANT® HCV Amplification 2.0 Kit (LiPA). All reagents needed for both reactions are added prior to initiation of the reverse transcription step so that the tube does not need to be manipulated or opened until the PCR amplification reactions are complete.

The genomic HCV RNA is reversed transcribed into cDNA using a mixture of 2 reverse transcriptases enzymes, an HCV specific primer, and an excess of a mixture of four deoxynucleoside triphosphates (dNTP) and deoxyuridine triphosphate (dUTP). Reverse

{5}

transcriptase extends the primer along the target RNA, thus copying the RNA into its exact DNA complement (cDNA).

The mixture is then heated to inactivate the reverse transcriptases and activate the Taq polymerase. The reaction uses two pairs of biotinylated primers, and a thermostable Taq DNA polymerase to co-amplify two distinct biotinylated DNA fragments specific for the 5' UTR and core regions of HCV. These primers are designed to permit amplification of two regions of the genome which are used to determine the different genotypes. These primers are labeled with biotin to facilitate visualization of the biotinylated DNA end product in the subsequent LiPA genotyping procedure.

## PCR Product Contamination Control

PCR amplification is performed in the presence of dUTP to generate amplification products which can be selectively degraded by treatment with uracil-N- glycosylase (UNG) and heat. UNG cleaves the glycosidic bond between the uracil base and the deoxyribose ring, creating a basic site that is susceptible to heat. The ENZ MIX component of the VERSANT® HCV Amplification 2.0 Kit (LiPA) contains a heat-labile UNG enzyme. Following the addition of the purified viral RNA to the master mix and prior to the initiation of the RT step, the reaction mixture is allowed to incubate at room temperature for 10 minutes. This step removes the uracil base from any contaminating amplification product made with dUTP. The UNG is inactivated when the temperature is increased to 50 °C to initiate the RT reaction, preventing the removal of uracil from the newly synthesized cDNA. As the temperature is further increased during the first PCR step, any abasic amplification products are completely degraded.

## Genotyping with the VERSANT® HCV Genotype Assay (LiPA)

The detection and identification of HCV genotypes using the VERSANT® HCV Genotype 2.0 Assay (LiPA) is based upon reverse hybridization of the RT-PCR amplified nucleic acid to HCV sequence specific oligonucleotides (probes) bound to a membrane strip.

The genotyping regions chosen for use in the assay are the 5'UTR and core regions of the Hepatitis C virus. The 5'UTR region is optimal for genotyping due to the high conservation within—but increased variability between—genotypes. Probes specific for the core region have been added to allow improved discrimination between subtypes a and b of genotype 1 and to enable the differentiation of the rare genotype 6, subtypes c-1, from genotype 1.

The amplified, biotinylated DNA derived from a patient specimen (the target nucleic acid) is hybridized to a set of sequence-specific oligonucleotide probes that are immobilized on a nitrocellulose membrane. The probes are attached to the nitrocellulose support by a poly(T) tail and are specific for the 5'UTR or core region of the different HCV genotypes. Each type of probe is applied as a different line, and thus the test is referred to as a line probe assay or LiPA. Under well-defined hybridization conditions, including pH and temperature, such probes allow the specific capture of matching DNA fragments from the patient specimen in a single reaction and subsequent visual detection as described below.

PMA P160016: FDA Summary of Safety and Effectiveness Data
Page 6

{6}

PMA P160016: FDA Summary of Safety and Effectiveness Data
Page 7

# Line Probe Assay Test Principle
## Line Probe Assay (LiPA)

Reverse hybridization principle

![img-0.jpeg](img-0.jpeg)

Once the hybridization procedure is completed, a wash step is performed to remove any unbound or weakly bound material. An alkaline phosphatase-labeled streptavidin conjugate is then added, which binds to the biotinylated hybrid. Upon addition of the chromogen substrate, 5-bromo-4-chloro-3-indolyl-phosphate and 4-nitro blue tetrazolium (BCIP/NBT), a purple brown precipitate is formed that reacts with the streptavidin-alkaline phosphatase complex. This results in a pattern of positive bands on the strip that are indicative of the patient's genotype. Due to small differences in probe affinity, reaction conditions, or amount of amplified material, the color intensity of the different lines may vary. Because the assay is a qualitative test, lines are read as positive or negative based on both visual detection and calibrated optical density cut-off values. The HCV genotype of the specimen is interpreted from the resulting pattern of probe lines on the strip.

# Genotype Interpretation

The LiPA strips contain three control lines that assure specific steps of the assay are successfully performed:

The conjugate control line (line 1) is designed to ensure that the color development reaction has taken place. The absence of conjugate control line indicates that an error occurred during the detection step of the assay, such as inappropriate amount of a reagent, the wrong reagent added during strip processing or a reagent was improperly prepared. Strips that do not develop a conjugate control line are considered invalid and the assay will need to be repeated.

Two amplification control lines have been included on the strip, one specific for the 5'UTR region (line 2) and the other for the core region of the HCV genome (line 3). These lines are designed to confirm the presence of the resulting amplicon. If the 5'UTR control line on the LiPA strip is negative, it may indicate that the RT-PCR reaction was inhibited, a low

{7}

concentration of HCV RNA was present in the sample, the sample was from an HCV negative patient, or an operator error occurred. If the core amplification control line is negative, it is not possible to differentiate genotype 6 subtypes c-1 from genotype 1, and it is not possible to differentiate subtypes a and b of genotype 1. If the amplification control lines give the appropriate positive results, the positive bands for each specimen strip can be used to interpret the genotype.

The pattern of bands determines the genotype identity. HCV genotypes and subtypes react in distinct patterns, each of which is genotype and/or subtype- specific.

There are two methods to identify the genotype of a patient:

1. manually using the reading card and interpretation chart
2. semi-automated using the LiPA-Scan Software

## Manual HCV Genotype Interpretation

In order to identify the genotype of a patient, a plastic reading card is placed over the strip. HCV genotype is determined by aligning the assay strips with the reading card and recording the line numbers corresponding to the positively colored bands that have appeared. The genotype corresponding to the banding patterns can then be derived using the reading card and the interpretation chart.

![img-1.jpeg](img-1.jpeg)
LiPA Strip Design

## Semi-automated HCV Genotype Interpretation

The developed strips are scanned and analyzed with LiPA-Scan software. The software interprets the band pattern and reports a genotype. The operator confirms that the results are accurate.

## VI. ALTERNATIVE PRACTICES AND PROCEDURES

Currently, there is one other FDA approved medical device for determining the HCV genotype(s) in plasma or serum, the Abbott RealTime HCV Genotype II assay. The only currently available conventional alternative methodology for this indication is

PMA P160016: FDA Summary of Safety and Effectiveness Data

{8}

nucleic acid sequencing. Each alternative has its own advantages and disadvantages. A patient should fully discuss the alternatives with his/her physician to select the method that best meets expectations and lifestyle.

## VII. MARKETING HISTORY

The VERSANT® Sample Preparation Reagent 1.0 Kit (Box 1 (SMN 10286026 and Box 2 SMN 10286027) has been on sale outside the US since November 2008.

The VERSANT® Sample Preparation Reagent 1.0 Kit is sold in the following countries: Algeria, Australia, Austria, Belgium, Canada, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Egypt, El Salvador, Finland, France, Germany, Greece, Honduras, Hungary, Iceland, Ireland, Italy, Jordan, Korea, Liechtenstein, Lithuania, Luxembourg, Malta, Mexico, Netherlands, New Zealand, Nicaragua, Norway, Peru, Philippines, Portugal, Serbia, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Syria, Tunisia, Turkey, United Kingdom, United States, Vietnam.

The VERSANT® HCV Genotype 2.0 Assay composed of the following:

- VERSANT® HCV Amplification 2.0 Kit (LiPA)
- VERSANT® HCV Genotype 2.0 Assay (LiPA)
- VERSANT® Control 2.0 Kit (LiPA)
- VERSANT® Sample Preparation 1.0 Reagents

has been sold in the following countries: Algeria, Argentina, Australia, Austria, Belgium, Brazil, Canada, Canary Islands, Chile, China, Croatia, Czech Republic, Denmark, France, Georgia, Germany, Great Britain, Greece, Hong Kong, Hungary, Indonesia, Ireland, Israel, Italy, Kuwait, Lithuania, Luxembourg, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Norway, Oman, Poland, Portugal, Qatar, Republic of Korea, Saudi Arabia, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand.

This product has not been withdrawn from the market from any country related to safety or effectiveness, or for any other reasons.

## VIII. POTENTIAL ADVERSE EFFECTS OF THE DEVICE ON HEALTH

When used according to the instructions in the package insert, there are no known direct adverse effects of this device on the health of the user.

The VERSANT® HCV Genotype 2.0 Assay (LiPA) is intended to be used as an aid in the management of patients with chronic HCV infection to guide the selection of individuals being considered for antiviral treatment. Therefore, patients have already been diagnosed as HCV positive by another method eliminating the possibility of false negative samples.

Direct acting antivirals (DAA’s) can act on more than one HCV genotype. For example, Sofosbuvir can be used in the treatment of HCV genotypes 1, 2, 3 and 4. Nevertheless, it is still recommended to determine the genotype prior to initiating treatment. If a patient is assigned an incorrect genotype, the patient would not respond optimally to the treatment provided. This would result in a second test to determine the correct genotype. Hence, incorrect genotypes result in a delay of optimal treatment.

PMA P160016: FDA Summary of Safety and Effectiveness Data
Page 9

{9}

IX. SUMMARY OF NONCLINICAL STUDIES

Nonclinical studies were performed at Siemens Healthcare Diagnostics, Berkeley, CA to evaluate the performance characteristics of the VERSANT® HCV Genotype 2.0 assay (LiPA). The studies are described below. All of the nonclinical studies were performed using the four reagent kits that comprise this assay, multiple VERSANT® Molecular Systems SP, three different thermal cyclers and two different auto strip processors.

A. Laboratory Studies

1. Limit of Detection (LoD)

The purpose of this study was to establish the Limit of Detection (LoD) of the VERSANT® HCV Genotype 2.0 Assay (LiPA) in serum and plasma for each genotype 1a, 1b, 2, 3, 4, 5, and 6. The experimental design followed the CLSI guideline EP17-A2. Four (4) lots of the VERSANT® Sample Preparation 1.0 Reagent Kit, 4 lots of the VERSANT® HCV Control 2.0 Kit (LiPA), 7 lots of the VERSANT® HCV Amplification 2.0 Kit (LiPA) and 7 lots of the VERSANT® HCV Genotype 2.0 Assay (LiPA) Kit were used for this study.

The study was conducted in two phases. The goal of Phase 1 was to establish the Limit of Detection of the VERSANT® HCV Genotype 2.0 Assay (LiPA) for each genotype/subtype separately in serum and plasma. At least twenty replicates at target concentrations ranging from 25 IU/mL to 1000 IU/mL of HCV genotypes 1-6 in serum and plasma were tested. Each target concentration was tested with a minimum of 2 replicates per run, with 2 runs per day for at least 4 days. Using probit analysis, the LoD for each genotype was estimated separately for serum and plasma and separately for the semi-automated results and for the manually interpreted results.

The goal of Phase 2 was to verify the Limit of Detection of the VERSANT® HCV Genotype 2.0 Assay (LiPA) for each genotype separately in serum and plasma. The LoD for plasma was verified separately in K₂EDTA plasma, ACD plasma, and CPD plasma. A single concentration at the LoD claim for each genotype/subtype obtained in Phase 1 was tested in 20 replicates each for serum and plasma (K₂EDTA, ACD and CPD). The acceptance criteria stated that each genotype/subtype provides at least 95% detection rate at the concentration tested.

The LoD claim for genotype 1-6 in plasma and serum are shown in the table below. Detection rate at the LoD claim in serum and plasma was ≥ 95% for each genotype.

PMA P160016: FDA Summary of Safety and Effectiveness Data
Page 10

{10}

Table 2: Limit of Detection in Plasma and Serum

|  Limit of Detection for HCV genotype/subtype in plasma and serum  |   |   |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- | --- | --- |
|   | 1a | 1b | 2 | 3 | 4 | 5 | 6  |
|  Plasma (IU/mL) | 400 | 300 | 200 | 150 | 300 | 300 | 350  |
|  Serum (IU/mL) | 350 | 350 | 300 | 200 | 300 | 650 | 350  |

The LoD claim for the VERSANT® HCV Genotype 2.0 Assay (LiPA) is 400 IU/mL in plasma and 650 IU/mL in serum.

## 2. Analytical Specificity

### Potentially Interfering Substances

Two independent studies were conducted to address potentially interfering substances:

- Interference by antiviral drugs
- Interference by endogenous substances

Both studies followed CLSI guideline EP7-A2, Interference Testing in Clinical Chemistry; Approved Guideline – Second Edition. Studies were performed with HCV genotypes 1a, 1b, 2, and 3 at the concentrations shown below.

Table 3: Target HCV Concentrations Used for Testing Interference

|  HCV Genotype | Endogenous Substance Interference Target HCV Concentration (IU/mL) | Antiviral Drug Interference Target HCV Concentration (IU/mL)  |
| --- | --- | --- |
|  1a | 1300 | 714  |
|  1b | 5000 | 6360  |
|  2 | 600 | 738  |
|  3 | 600 | 378  |

These genotypes were chosen because they resemble the current genotype distribution in the US. All studies passed the acceptance criteria.

PMA P160016: FDA Summary of Safety and Effectiveness Data

{11}

# Endogenous Substances

Panels were prepared by adding one specimen each of HCV genotypes 1a, 1b, 2 and 3 to pooled HCV negative human plasma. Potentially interfering endogenous substances were then added into HCV samples at the concentrations listed in the table below. Potential effects on genotyping rate and accuracy were tested using these samples.

Table 4 Endogenous Substances Tested

|  Endogenous Substance | Concentrations Tested  |
| --- | --- |
|  Hemoglobin | 500 mg/dL  |
|  Triglycerides (Intralipid) | 3000 mg/dL  |
|  Conjugated bilirubin | 20 mg/dL  |
|  Unconjugated bilirubin | 20 mg/dL  |
|  Protein (Albumin solution) | 9 g/dL  |

Three (3) replicates were prepared for each combination of endogenous substance and HCV genotype to yield a total sixty (60) test samples. These samples were tested alongside two (2) replicates of each HCV genotype sample (no endogenous substance) as positive controls and two (2) replicates of each endogenous substance (no HCV) as specificity controls.

The results met the acceptance criteria for genotyping rate $(\geq 90\%)$ and genotyping accuracy $(\geq 99\%)$ for each HCV genotype tested.

# Anti-viral Drugs

The drugs evaluated included eight (8) different types of medications: nucleoside HIV RT inhibitors, nucleotide DNA polymerase inhibitors, HIV and HCV protease inhibitors, non-nucleoside RT inhibitors, fusion inhibitors, HIV integrase inhibitors, HCV and HBV antivirals, and immune modulators (see table below). The drugs were pooled based on their solubility properties into 5 groups and spiked into negative human serum at three times their $C_{\mathrm{max}}$ concentrations. Aliquots were then individually spiked with HCV genotypes 1a, 1b, 2 and 3. For each drug pool tested, there were a total of 12 replicates, 3 replicates each from each genotype.

Table 5: List of Anti-Viral Drugs Tested

|  Drug | Drug Name  |
| --- | --- |
|  Nucleoside HIV RT Inhibitors | Abacavir sulfate (ABC)
Didanosine (ddI)
Emtricitabine (FTC) Lamivudine (3TC)
Stavudine (d4T) Zidovudine (AZT)  |
|  Nucleotide DNA Polymerase Inhibitors | Tenofovir
Adefovir dipivoxil  |

PMA P160016: FDA Summary of Safety and Effectiveness Data

{12}

|  Drug | Drug Name  |
| --- | --- |
|  HIV and HCV Protease Inhibitors | Atazanavir, Boceprievir, Darunavir, Fosamprenavir, Indinavir, Lopinavir, Nelfinavir, Ritonavir, Saquinavir, Telaprevir, Tipranavir  |
|  Non-Nucleoside RT Inhibitors | Efavirenz, Nevirapine  |
|  Fusion Inhibitor | Enfuvirtide, Maraviroc  |
|  HIV Integrase | Raltegrivir  |
|  HCV and HBV antivirals | Acyclovir, Entecavir, Ganciclovir, Ribavirin, Telbivudie Valganciclovir  |
|  Immune Moderators | Interferon alfacon-1, Interferon alpha-2b, Peginterferon alpha-2a, Peginterferon alpha-2b  |

Results:

No interference in the performance of the VERSANT® HCV Genotype 2.0 Assay (LiPA) was observed in the presence of anti-viral drugs for all HCV positive and negative samples tested.

# 3. Cross-reactivity with non-HCV Disease States and Other Blood-borne Microorganisms

The purpose of this study was to evaluate potential cross-reactivity with three non-HCV disease states and seventeen common microorganisms on the genotyping rate and genotyping accuracy of HCV samples using the VERSANT® HCV Genotype 2.0 Assay (LiPA). The table below lists the non-HCV diseases and microorganisms that were tested.

Table 6: List of Non-HCV Disease States/Microorganisms Tested

|   | Disease State/ Microorganism  |
| --- | --- |
|  non-HCV Disease | Anti-nuclear antigen  |
|   |  Rheumatoid factor  |
|   |  Systemic Lupus Erythematosus  |
|   | Cytomegalovirus  |
|   |  Epstein-Barr Virus  |
|   |  Hepatitis A Virus  |
|   |  Hepatitis B Virus  |
|   |  Human Immunodeficiency Virus  |
|   |  Candida albicans  |

PMA P160016: FDA Summary of Safety and Effectiveness Data

{13}

|   | Disease State/ Microorganism  |
| --- | --- |
|  Microorganism | Enterobacter cloacae  |
|   |  Escherichia coli  |
|   |  Haemophilus influenzae  |
|   |  Klebsiella pneumoniae  |
|   |  Pseudomonas aeruginosa  |
|   |  Pseudomonas fluorescens  |
|   |  Serratia marcescens  |
|   |  Streptococcus agalactiae  |
|   |  Staphylococcus aureus  |
|   |  Staphylococcus epidermidis  |
|   |  Streptococcus pneumonia  |

Protocol for Non-HCV Disease:
Patient serum samples from 3 different non-HCV disease conditions were spiked with HCV positive plasma and/or serum samples of HCV genotype 1a, 1b, 2, or 3. Each HCV genotype was spiked into two patient serum samples from each disease state (i.e., two patient samples with Rheumatoid factor were spiked with HCV genotype 1a while another two patient samples with Rheumatoid factor were spiked with HCV genotype 1b, etc.).

Protocol for Microorganisms:
Potentially interfering microorganisms were tested for
- Cross reactivity – the microorganism does not produce a false positive result for an HCV genotype
- Interference – the microorganism does not prevent proper genotyping of HCV genotypes tested at 1.5 X LoD

Each microorganism was spiked into serum sample tubes except for Hepatitis A virus (HAV). In this case, HAV positive patient sample tubes were spiked with a HCV genotype/subtype.

Results:
Two hundred and ninety (290) samples were analyzed in this study. All cross reactivity samples that contained potentially interfering microorganisms were negative. At least 9 of the 10 replicates of each of the non-HCV disease states and Hepatitis A virus samples and at least 11 of the 12 replicates of each of the potentially interfering organism samples gave the correct genotype result. Thus, the samples from three different non-HCV disease states or samples containing 17 different potentially interfering microorganisms yielded results that met the requirements for ≥90% genotyping rate and ≥99% genotyping accuracy with the VERSANT® HCV Genotype 2.0 Assay (LiPA).

4. Analytical Carryover

This study was designed to demonstrate that strips processed from HCV RNA-negative samples are not contaminated by samples with high HCV viral concentrations when tested simultaneously with the VERSANT® HCV Genotype 2.0 Assay (LiPA).

PMA P160016: FDA Summary of Safety and Effectiveness Data
Page 14

{14}

A high titer HCV genotype 1a stock (1.2 x 10⁶ IU/mL) was used as the positive sample. Pooled human K₂EDTA plasma that was pre-qualified to be HCV RNA negative using an FDA approved Transcript Mediated Amplification (TMA) Assay (Hologic, Marlborough, MA) was used as the HCV negative test sample. Seventy two (72) replicates of each of the positive and negative samples were arranged in a checkerboard pattern to maximize cross contamination during nucleic acid extraction and PCR amplification.

Samples were processed using three thermocyclers: the GeneAmp PCR System 9700, the Agilent Mx3005P or the Biorad PTC-200, and with the Auto-LiPA 48 or the AutoBlot 3000H in order to test for potential contamination across all equipment used in the assay.

Results demonstrate that analyte carryover specificity was 100%. These results demonstrate that the VERSANT® HCV Genotype 2.0 Assay (LiPA) met acceptance criteria and is able to process HCV negative and high HCV viral load samples simultaneously without cross-contamination.

# 5. Specimen Stability

The purpose of this study was to evaluate HCV specimen stability in serum and plasma. Specimen stability was evaluated by diluting one specimen each from HCV Genotypes 1a, 1b, 2 and 3 in K₂EDTA plasma, plasma from PPT tubes, serum and serum from SST tubes at the target concentrations shown in Table 7.

Table 7: Target HCV Concentrations Used to Test Specimen Stability in K₂EDTA Plasma, PPT tubes, Serum and SST Tubes

|  HCV Genotype | Target HCV Concentration (IU/mL)  |
| --- | --- |
|  1a | 1300  |
|  1b | 450  |
|  2 | 600  |
|  3 | 600  |

The study demonstrated that specimens in K₂EDTA plasma, plasma from PPT tubes, serum and serum from SST tubes may be stored at 2 to 8°C for 48h, -20 ± 5°C for 72 h or at -60 to -80°C for long term storage for use with the VERSANT® HCV Genotype 2.0 Assay (LiPA). Specimens should not exceed four freeze-thaw cycles prior to use with the VERSANT® HCV Genotype 2.0 Assay (LiPA).

# 6. Reagent Stability

The purpose of this study was to determine the stability of the three VERSANT® HCV Genotype 2.0 Assay (LiPA) reagent kits and the Sample Preparation 1.0 Reagent Kit together. The four kits were tested as a unit for their ability to genotype clinical samples of HCV genotypes 1a, 1b, 2, 3, 4, 5 and 6 diluted to 3 X LoD in pooled K₂EDTA plasma. At each stability time point, 76 replicates were tested; 11 replicates of each genotype except for genotype 6 which was tested 10 times.

PMA P160016: FDA Summary of Safety and Effectiveness Data
Page 15

{15}

Baseline, 13-month and end of shelf life real-time stability testing of unopened kits was assessed for three lots of each of the four reagent kits. Genotyping rate and genotyping accuracy at all test points passed all acceptance criteria. Data from the baseline real-time stability test point verify that reagents kits sequestered for stability testing of the VERSANT® HCV Genotype 2.0 Assay (LiPA) had acceptable product performance at the start of the study. Data from 13-month real-time stability test point verifies that reagents kits still had acceptable performance. Data from the end of shelf life testing verifies the expiration dating for all four reagent kits included in the product labeling.

Based on the stability data, the shelf life for each of the four VERSANT® HCV Genotype 2.0 Assay (LiPA) reagent kits are as follows.

Table 8: Shelf Life for the Four VERSANT® HCV Genotype 2.0 Assay (LiPA) Reagent Kits

|  Kit Description | Storage Temp | Shelf Life  |
| --- | --- | --- |
|  VERSANT® Sample Preparation 1.0 Reagent Kit | 15-30°C (Box1)
2-8°C (Box2) | 24 months  |
|  VERSANT® HCV | 2-8°C | 24 months  |
|  VERSANT® HCV Amplification 2.0 Kit | -20 ± 5°C | 29 months  |
|  VERSANT® HCV Genotype 2.0 Assay (LiPA) Kit | 2-8°C | 18 months  |

Shipping stability, which was assessed concurrently with real-time stability, also passed all acceptance criteria. This verifies that product performance is acceptable when shipping conditions are met.

PMA P160016: FDA Summary of Safety and Effectiveness Data
Page 16

{16}

# 7. Reproducibility

Reproducibility of the VERSANT® HCV Genotype 2.0 Assay (LiPA) was evaluated using a 14 member reproducibility panel. Single specimens containing HCV genotypes 1a, 1b, 2, 3, 4, 5 and 6 were each diluted in HCV negative human plasma to a low target and a high target HCV concentration as shown in Table 9.

Table 9: Target HCV Concentrations Used for Evaluating Reproducibility

|  HCV Genotype | Low Target HCV Concentration (IU/mL) | High Target HCV Concentration (IU/mL)  |
| --- | --- | --- |
|  1a | 714 | 100,000  |
|  1b | 6,099 | 100,000  |
|  2 | 738 | 100,000  |
|  3 | 378 | 100,000  |
|  4 | 3,531 | 100,000  |
|  5 | 579 | 100,000  |
|  6 | 378 | 100,000  |

A total of 3 lots each of the VERSANT® Sample Preparation 1.0 Reagent Kit, the VERSANT® HCV Control 2.0 Kit (LiPA), the VERSANT® HCV Amplification 2.0 Kit (LiPA) and the VERSANT® HCV Genotype 2.0 Assay (LiPA) Kit were used to process 54 replicates of each panel member at each of 3 clinical test sites.

No significant difference in performance of the VERSANT® HCV Genotype 2.0 Assay (LiPA) was observed with different reagent kit lots using either the manual or the semi-automated method of interpretation.

PMA P160016: FDA Summary of Safety and Effectiveness Data

{17}

Table 10: Effect of Reagent Kit Lot on Assay Performance

|  Kit Lot | Manual Method of Interpretation |   |   |   | Semi-Automated Method of Interpretation  |   |   |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
|   |  Genotyping Ratea[Fraction] | Genotyping Rate [Percent] | Genotyping Accuracyb[Fraction] | Genotyping Accuracy [Percent] | Genotyping Rate [Fraction] | Genotyping Rate [Percent] | Genotyping Accuracy [Fraction] | Genotyping Accuracy [Percent]  |
|  Lot 1 | 740/753 | 98.27 | 735/740 | 99.32 | 736/756 | 97.74 | 732/736 | 99.46  |
|  Lot 2 | 740/756 | 97.88 | 738/740 | 99.73 | 735/756 | 97.22 | 733/735 | 99.73  |
|  Lot 3 | 743/759 | 97.89 | 742/743 | 99.87 | 739/759 | 97.36 | 738/739 | 99.86  |

a Genotyping rate is the proportion of valid genotype results that were interpretable.
b Genotyping accuracy is the proportion of interpretable results that match with the reference method results.

Assay performance, measured by genotyping rate and genotyping accuracy, at Site 3 was significantly different from the other two sites (p-values ≤ 0.05) as shown in Table 11. No significant difference in assay performance was observed between Site 1 and Site 2. The difference at Site 3 was attributable to one operator (Table 12). After further training the site corrected this problem, as shown by an inter-site comparison of a subset of the specimens used to evaluate assay accuracy as shown in Table 13.

PMA P160016: FDA Summary of Safety and Effectiveness Data

{18}

Table 11: Effect of Clinical Trial Site on Assay Performance

|  Clinical Site | Manual Method of Interpretation |   |   |   | Semi-Automated Method of Interpretation  |   |   |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
|   |  Genotyping | Genotyping | Genotyping | Genotyping | Genotyping | Genotyping | Genotyping | Genotyping  |
|   |  Ratea | Rate | Accuracyb | Accuracy | Ratea | Rate | Accuracyb | Accuracy  |
|  Site 1 | 748/756 | 98.94 | 747/748 | 99.87 | 749/756 | 99.07 | 748/749 | 99.87  |
|  Site 2 | 750/756 | 99.21 | 749/750 | 99.87 | 745/756 | 98.54 | 744/745 | 99.87  |
|  Site 3 | 725/756 | 95.90 | 719/725 | 99.17 | 716/756 | 94.71 | 711/716 | 99.30  |
|  |   |   |   |   |   |   |   |   |

a Genotyping rate is the proportion of valid genotype results that were interpretable.
b Genotyping accuracy is the proportion of interpretable results that match with the reference method results.

Table 12: Assay Performance at Site 3 Stratified by Operators

|  Operator at Site 3a | Manual Method of Interpretation |   |   |   | Semi-Automated Method of Interpretation  |   |   |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
|   |  Genotyping Rateb | Genotyping Rate | Genotyping Accuracyc | Genotyping Accuracy | Genotyping Rate [Fraction] | Genotyping Rate [Percent] | Genotyping Accuracy [Fraction] | Genotyping Accuracy [Percent]  |
|  Operator 1 | 352/378 | 93.1 | 348/352 | 98.9 | 369/378 | 97.6 | 368/369 | 99.7  |
|  Operator 2 | 373/378 | 98.7 | 371/373 | 99.5 | 347/378 | 91.8 | 343/347 | 98.8  |

a For a given run, the semi-automated method of interpretation was conducted by one operator, while a second operator at the site interpreted the processed assay strips for each run using the manual method of interpretation. Hence, the results from the manual method of interpretation for operator 1 correspond to the results from the semi-automated method of interpretation for operator 2. These samples were processed by the same operator.
b Genotyping rate is the proportion of valid genotype results that were interpretable.
c Genotyping accuracy is the proportion of interpretable results that match with the reference method results.

PMA P160016: FDA Summary of Safety and Effectiveness Data

{19}

Table 13: Effect of Clinical Trial Site on Assay Performance with a Subset of Specimens used to Evaluate Assay Accuracy

|  Clinical Site | Manual Method of Interpretation |   |   |   | Semi-Automated Method of Interpretation  |   |   |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
|   |  Genotyping Rate [Fraction] | Genotyping Rate [Percent] | Genotyping Accuracy [Fraction] | Genotyping Accuracy [Percent] | Genotyping Rate [Fraction] | Genotyping Rate [Percent] | Genotyping Accuracy [Fraction] | Genotyping Accuracy [Percent]  |
|  Site 1 | 189/193 | 97.93 | 189/189 | 100 | 187/193 | 96.89 | 187/187 | 100  |
|  Site 2 | 191/194 | 98.45 | 189/191 | 98.95 | 188/194 | 96.91 | 186/188 | 98.94  |
|  Site 3 | 186/193 | 96.37 | 185/186 | 99.46 | 186/193 | 96.37 | 185/186 | 99.46  |
|  |   |   |   |   |   |   |   |   |

PMA P160016: FDA Summary of Safety and Effectiveness Data

{20}

Results of Reproducibility Studies:

There were no significant differences in assay performance between kit lots for both manual and semi-automatic interpretation method.

There were no significant differences in assay performance between the manual method of interpretation and the semi-automated method of interpretation.

Assay performance at one clinical trial site was shown to be significantly different compared to the other two sites. Further investigation determined that operator laboratory skills were the cause of this discrepancy. Following further training, assay performance was the same at all three sites.

## Conclusions Drawn from Non Clinical Studies

The VERSANT® HCV Genotype 2.0 Assay (LiPA) was evaluated to demonstrate performance claims for specificity, accuracy, endogenous substances interference, carryover, reproducibility, mixed infections, limit of detection and stability. These performance characteristics were measured using genotyping accuracy and genotyping rate. Together with the clinical trial results, these data support the intended use claims of the VERSANT® HCV Genotype 2.0 Assay (LiPA).

## X. SUMMARY OF PRIMARY CLINICAL STUDIES

## A. Study Design

### Clinical Validity

The overall objective of the Clinical Validity Study was to show a relationship between the HCV genotype obtained by the VERSANT® HCV Genotype 2.0 Assay (LiPA) System and the probability of achieving sustained virologic response (SVR₁₂) for individuals receiving HCV treatment. HCV genotypes 1a, 1b, 2, 3, 4, 5, and 6 collected from individuals with chronic HCV infection who had known HCV treatment outcome were tested.

Specimens from 217 individuals infected with HCV genotype 1a, 1b, 2, 3, 4, 5 or 6 as determined by the NS5b HCV genotyping reference method were tested at three clinical trial sites. The specimens were obtained from individuals prior to initiating treatment with Direct-Acting Antivirals (DAAs). All specimens from the particular genotype were treated with the identical DAA regime. Samples were collected retrospectively for the clinical trial. Inclusion and exclusion criteria are listed below:

### 1. Clinical Inclusion and Exclusion Criteria

Inclusion Criteria:

- At least 18 years of age
- Specimen available from a time point prior to or during HCV treatment
- Specimen collected from subjects infected with the same HCV genotype who have also received the same HCV treatment.
- HCV RNA viral load &gt; 2000 IU/mL
- Documented HCV treatment outcome available

PMA P160016: FDA Summary of Safety and Effectiveness Data
Page 21

{21}

Exclusion Criteria:

- Subject quit HCV treatment and viral load was not documented upon exiting the study

Specimens of genotype 4 and subtype 1a and 1b were collected from patients treated with Sofosbuvir (SOF), Pegylated Interferon (peg-IFN) and Ribavirin (RBV). Genotype 2 and 3 specimens were collected from patients treated with SOF and RBV, and genotype 5 and 6 specimens were collected from patients treated with SOF and Ledipasvir (LDV). Tables 16a and 16b show the demography of the patients, the genotype distribution and the treatment received. Individuals were enrolled in clinical trials¹ and were treated according to the standard of care or until treatment was discontinued. Individuals enrolled in the study attained SVR₁₂ or NSVR (Non-Sustained Virologic Response) as documented at a follow-up visit after 12 weeks of treatment.

2. Follow-up Schedule:

All studies were retrospective using banked samples. Follow-up did not apply.

3. Endpoints:

Results were compared to sequencing for the rate and accuracy study and to treatment outcome for the clinical validity study to determine accuracy, as noted above.

Table 14: Distribution of HCV Genotype Specimens in the Clinical Validity Study

|  HCV Genotype (NS5b) | Quantity (% of total specimens) | Treatment Received  |
| --- | --- | --- |
|  1a | 40 (18.2%) | SOF+peg-IFN+RBV  |
|  1b | 40 (18.2%) | SOF+peg-IFN+RBV  |
|  2 | 40 (18.2%) | SOF+RBV  |
|  3 | 40 (18.2%) | SOF+RBV  |
|  4 | 20 (9.1%) | SOF+peg-IFN+RBV  |
|  5 | 20 (9.1%) | SOF+LDV  |
|  6 | 20 (9.1%) | SOF+LDV  |
|  Total | 220 |   |
|  SOF: Sofosbuvir; peg-IFN: Pegylated Interferon; RBV: Ribavirin; LDV: Ledipasvir  |   |   |

PMA P160016: FDA Summary of Safety and Effectiveness Data
Page 22

{22}

PMA P160016: FDA Summary of Safety and Effectiveness Data
Page 23

B. Accountability of PMA Cohort

Two hundred and twenty samples from patients with known treatment outcomes were used in the clinical validity study. Of these, 217 gave evaluable results and three samples were invalid and therefore not included in the calculations.

C. Study Population Demographics and Baseline Parameters

Table 15: Demography of HCV Infected Subjects Enrolled in the Clinical Validity Study

|  Characteristics | Category | Number of Subjects | Percentage of Total  |
| --- | --- | --- | --- |
|  Age | <40 Years | 21 | 9.5  |
|   |  ≥40 Years | 199 | 90.5  |
|  Gender | Female | 85 | 38.6  |
|   |  Male | 135 | 61.4  |
|  Race/Ethnicity | White | 173 | 79  |
|   |  American Indian or Alaska | 7 | 3.2  |
|   |  Asian | 18 | 8.2  |
|   |  Black | 20 | 9.1  |
|   |  Native Hawaiian or Other Pacific Islander | 1 | 0.5  |
|   |  Other | 1 | 0.5  |
|  Baseline HCV RNA (Genotype 1a & b) | < 3E6 IU/mL | 29 | 36.25  |
|   |  ≥ 3E6 IU/mL | 51 | 63.75  |
|  Baseline HCV RNA (Genotype Non 1a & b) | < 3E6 IU/mL | 91 | 65  |
|   |  ≥ 3E6 IU/mL | 49 | 35  |

{23}

D. Safety and Effectiveness Results

1. Safety Results

Since the study utilized retrospective patient samples there were no adverse effects, and safety analysis of the study is not applicable.

2. Effectiveness Results

The results from the Clinical Validity Study for genotyping rate and genotyping accuracy are summarized in Tables 16 and 17 below.

Table 16: Genotyping Rate and Genotyping Accuracy using the VERSANT® HCV Genotype 2.0 Assay (LiPA) Manual Method of Interpretation

|  Genotype | Total Tested | Genotyping Rate^{a} [Fraction] | Genotyping Rate [Percent (95% CI^{c})] |   | Genotyping Accuracy^{b} [Fraction] | Genotyping Accuracy [Percent (95% CI)]  |   |
| --- | --- | --- | --- | --- | --- | --- | --- |
|  1a | 140 | 138/140 | 98.6 | (94.6, 99.82) | 137/138 | 99.3 | (96.01, 99.87)  |
|  1b | 140 | 140/140 | 100 | (96.30, 100) | 138/140^{d} | 98.6 | (94.94, 99.61)  |
|  2 | 140 | 140/140 | 100 | (96.30, 100) | 140/140 | 100 | (97.33, 100)  |
|  3 | 140 | 140/140 | 100 | (96.30, 100) | 140/140 | 100 | (97.33, 100)  |
|  4 | 80 | 78/80 | 97.5 | (88.64, 99.08) | 78/78 | 100 | (95.31, 100)  |
|  5 | 80 | 75/80 | 93.8 | (86.30, 98.29) | 75/75 | 100 | (95.13, 100)  |
|  6 | 80 | 72/80 | 90.0 | (75.83, 93.09) | 72/72 | 100 | (94.93, 100)  |

a Genotyping rate is the proportion of valid genotype results that were interpretable.
b Genotyping accuracy is the proportion of interpretable results that match with the reference method result.
c The 95% upper and lower confidence limits.
d The two discordant genotype 1b samples were recombinant genotype 2/1b specimens.

PMA P160016: FDA Summary of Safety and Effectiveness Data
Page 24

{24}

Table 17: Genotyping Rate and Genotyping Accuracy using the VERSANT® HCV Genotype 2.0 Assay (LiPA) Semi-Automated Method of Interpretation

|  Genotype | Total Tested | Genotyping Ratea [Fraction] | Genotyping Rate [Percent (95% CIc)] |   | Genotyping Accuracyb [Fraction] | Genotyping Accuracy [Percent (95% CI)]  |   |
| --- | --- | --- | --- | --- | --- | --- | --- |
|  1a | 140 | 136/140 | 97.1 | (92.88, 98.88) | 135/136 | 99.26 | (95.95, 99.87)  |
|  1b | 140 | 138/140d | 98.6 | (94.94, 99.61) | 136/138d | 98.55 | (94.87, 99.6)  |
|  2 | 140 | 140/140 | 100 | (97.33, 100) | 140/140 | 100 | (97.33, 100)  |
|  3 | 140 | 140/140 | 100 | (97.33, 100) | 140/140 | 100 | (97.33, 100)  |
|  4 | 80 | 78/80 | 97.5 | (91.34, 99.31) | 78/78 | 100 | (95.31, 100)  |
|  5 | 80 | 76/80 | 95 | (87.84, 98.04) | 76/76 | 100 | (95.19, 100)  |
|  6 | 80 | 70/80 | 87.5 | (78.5, 93.07) | 70/70 | 100 | (94.8, 100)  |

a Genotyping rate is the proportion of valid genotype results that were interpretable.
b Genotyping accuracy is the proportion of interpretable results that match with the reference method result.
c The  $95\%$  upper and lower confidence limits.
d The two discordant genotype 1b samples were recombinant genotype 2/1b specimens.

PMA P160016: FDA Summary of Safety and Effectiveness Data

{25}

# Rate of  $\mathrm{SVR}_{12}$  in Response to HCV Treatment for each Genotype

Viral load was measured 12 weeks after completion of treatment for patients enrolled in clinical trials. Refer to Tables 18a, 18b, and 18c for SVR/non-SVR distributions based on genotype results obtained with the NS5b, and the VERSANT® HCV Genotype 2.0 Assay (LiPA) System manual, and semi- automated methods of interpretation.

Table 18a: Summary of the SVR12 Results Using the Reference Method NS5b to Determine Genotype

|  Genotype (NS5b) | # SVR12 | # Non-SVR | Total  |
| --- | --- | --- | --- |
|  1a | 35 | 5 | 40  |
|  1b | 29 | 11 | 40  |
|  2 | 40 | 0 | 40  |
|  3 | 32 | 8 | 40  |
|  4 | 19 | 1 | 20  |
|  5 | 20 | 0 | 20  |
|  6 | 19 | 1 | 20  |
|  Total | 194 | 26 | 220  |

Table 18b: Summary of the SVR12 Results Using the VERSANT® HCV Genotype 2.0 Assay (LiPA) System Manual Method of Interpretation to Determine Genotype

|  Genotype (Manual) | # SVR12 | # Non-SVR | Total  |
| --- | --- | --- | --- |
|  1a | 35 | 4 | 39  |
|  1b | 29 | 11 | 40  |
|  2 | 40 | 0 | 40  |
|  3 | 32 | 8 | 40  |
|  4 | 19 | 1 | 20  |
|  5 | 18 | 0 | 18  |
|  6 | 19 | 1 | 20  |
|  Total | 192 | 25 | 217  |

PMA P160016: FDA Summary of Safety and Effectiveness Data

{26}

Table 18c: Summary of the SVR Results Using the VERSANT® HCV Genotype 2.0 Assay (LiPA) System Semi-Automated Method of Interpretation to Determine Genotype

|  Genotype (Semi-Auto) | # SVR12 | # Non-SVR | Total  |
| --- | --- | --- | --- |
|  1a | 34 | 4 | 38  |
|  1b | 29 | 11 | 40  |
|  2 | 40 | 0 | 40  |
|  3 | 32 | 8 | 40  |
|  4 | 19 | 1 | 20  |
|  5 | 19 | 0 | 19  |
|  6 | 19 | 1 | 20  |
|  Total | 192 | 25 | 217  |

The clinical utility of the VERSANT® Genotype 2.0 Assay (LiPA) was assessed by evaluating the association between HCV genotype and the probability of achieving SVR₁₂. For individual genotypes/subtypes, the observed SVR₁₂ rates ranged from 72.5% to 100%. Patients with subtype 1b had the lowest SVR₁₂ rate at 72.5% (29/40), while genotypes 2 and 5 had the highest rate, with 100% SVR₁₂.

Analysis of the SVR₁₂ rate was performed to compare SVR₁₂ rates between genotype 1 (pooled 1a and 1b) and the other non-1 genotypes by calculating the SVR₁₂ rate ratios. The odds ratios for SVR₁₂ between genotype 1 (pooled 1a and 1b) and the other non-1 genotypes were calculated to assess the relative odds for SVR₁₂ between the genotypes. The results for the analysis are summarized in Tables 19a and 19b for the VERSANT® HCV Genotype 2.0 Assay (LiPA) System manual and semi-automated methods of interpretation, respectively.

Tables 19a and 19b show that the odds for SVR₁₂ for genotype 1 were generally lower relative to the other genotypes, except for genotype 1 relative to genotype 3 for which the odds ratio was about 1. In particular, the odds for SVR₁₂ is statistically significantly lower for genotype 1 relative to genotype 2 (odds ratio of 0.051 for both manual and semi-automated methods of interpretation). Furthermore, as a group, genotype 1 (pooled 1a and 1b) resulted in statistically significantly lower odds for SVR₁₂ compared to the group of all non-1 genotypes (pooled genotypes 2, 3, 4, 5, 6). All other odds ratios were not statistically significant as the confidence limits contain 1.

PMA P160016: FDA Summary of Safety and Effectiveness Data

{27}

Table 19a: Ratios of SVR12 Rates and Odds Ratios for SVR12 by Genotype Based on the Manual Method of Interpretation

|  Genotype Comparison | SVR Rate Comparison | SVR Rate Ratio | Odds Ratio | 95% CI for OR  |
| --- | --- | --- | --- | --- |
|  1 vs 2 | 81.01% vs. 100% | 0.810 | 0.051 | (0.003, 0.882)  |
|  1 vs 3 | 81.01% vs. 80% | 1.013 | 1.067 | (0.410, 2.778)  |
|  1 vs 4 | 81.01% vs. 95% | 0.853 | 0.225 | (0.028, 1.812)  |
|  1 vs 5 | 81.01% vs. 100% | 0.810 | 0.112 | (0.006, 1.970)  |
|  1 vs 6 | 81.01% vs. 95% | 0.853 | 0.225 | (0.028, 1.812)  |
|  1 vs all non-1 | 81.01% vs. 92.75% | 0.873 | 0.333 | (0.142, 0.783)  |

CI= 95% confidence interval (lower and upper limit, respectively)

Table 19b: Ratios of SVR12 Rates and Odds Ratios for SVR12 by Genotype Based on the Semi-Automated Method of Interpretation

|  Genotype Comparison | SVR Rate Comparison | SVR Rate Ratio | Odds Ratio | 95% CI for OR  |
| --- | --- | --- | --- | --- |
|  1 vs 2 | 80.77% vs. 100% | 0.808 | 0.051 | (0.003, 0.869)  |
|  1 vs 3 | 80.77% vs. 80% | 1.010 | 1.050 | (0.403, 2.736)  |
|  1 vs 4 | 80.77% vs. 95% | 0.850 | 0.221 | (0.027, 1.784)  |
|  1 vs 5 | 80.77% vs. 100% | 0.808 | 0.105 | (0.006, 1.837)  |
|  1 vs 6 | 80.77% vs. 95% | 0.850 | 0.221 | (0.027, 1.784)  |
|  1 vs all non-1 | 80.77% vs. 92.81% | 0.870 | 0.326 | (0.138, 0.766)  |

CI= 95% confidence interval (lower and upper limits, respectively)

The LiPA results matched 100% with the NS5b reference method. Patients with HCV genotype 1 are less likely to achieve SVR12 than patients with genotype 2 or genotype 3. Current HCV therapy with DAA’s resulted in shorter duration of therapy and high rates of SVR12 (89%-90%). The observed SVR12 rate by genotype/subtype is consistent with published results using similar treatments with DAA drugs.

3. **Subgroup Analyses**

No subgroup analyses were performed.

4. **Pediatric Extrapolation**

In this premarket application, existing clinical data was not leveraged to support approval of a pediatric patient population.

E. **Financial Disclosure**

PMA P160016: FDA Summary of Safety and Effectiveness Data
Page 28

{28}

The Financial Disclosure by Clinical Investigators regulation (21 CFR 54) requires applicants who submit a marketing application to include certain information concerning the compensation to, and financial interests and arrangement of, any clinical investigator conducting clinical studies covered by the regulation. The pivotal clinical study included three investigators. None of the clinical investigators had disclosable financial interests/arrangements as defined in sections 54.2(a), (b), (c), and (f). The information provided does not raise any questions about the reliability of the data.

## XI. PANEL MEETING RECOMMENDATION AND FDA'S POST-PANEL ACTION

In accordance with the provisions of section 515(c)(3) of the act as amended by the Safe Medical Devices Act of 1990, this PMA was not referred to the FDA Microbiology Devices Advisory Panel, an FDA advisory committee, for review and recommendation.

## XII. CONCLUSIONS DRAWN FROM PRECLINICAL AND CLINICAL STUDIES

### A. Effectiveness Conclusions

The limit of detection was determined for each genotype in both serum and plasma. The LoD claim for the VERSANT® HCV Genotype 2.0 Assay (LiPA) is 400 IU/mL in plasma and 650 IU/mL in serum.

The VERSANT® HCV Genotype Assay 2.0 (LiPA) met requirements for reproducibility for different lots of reagents over multiple days of testing.

The genotyping rate and accuracy study used 800 unique samples. The percentage of specimens that yielded a genotype result ranged from 90% to 100% for the individual genotypes using the manual method of interpretation as shown in Table 17. The percentage of correct genotypes ranged from 98.6% to 100% for the individual genotypes.

Potentially interfering substances studies including endogenous substances, microorganisms and current anti-viral drugs, showed no interference with genotyping rate or accuracy.

Specimen stability was evaluated in K₂EDTA Plasma, Plasma from PPT tubes, Serum and Serum from SST tubes at the target concentrations. There was no impact to genotyping accuracy or rate.

Specimens may be stored in blood collection tubes at 2 to 8°C for 48h, -20 ± 5°C for 72 h or at -60 to -80°C for long term storage for use with the VERSANT® HCV Genotype 2.0 Assay (LiPA). Specimens should not exceed four freeze-thaw cycles prior to use with the VERSANT® HCV Genotype 2.0 Assay (LiPA).

The SVR₁₂ rate for all patients (all genotypes and subtypes pooled) was 88.48% (192/217) for both the manual and semi-automated methods of interpretation. For individual genotypes/subtypes, the observed SVR₁₂ rates ranged from 72.5% to 100%. Patients with subtype 1b had the lowest SVR₁₂ rate at 72.5% (29/40), while genotypes 2 and 5 were the highest, with 100%

PMA P160016: FDA Summary of Safety and Effectiveness Data
Page 29

{29}

$\mathrm{SVR}_{12}$

The VERSANT® HCV Genotype 2.0 Assay (LiPA) provides an accurate diagnosis of HCV genotypes 1a, 1b, 2, 3, 4, 5, and 6 as shown by a genotyping accuracy rate of $\geq 99\%$ as compared to the reference method. The SVR12 rate by genotype is consistent with published results using similar treatments with direct acting antiviral drugs as demonstrated in the clinical validity study. their entirety and were found to be genotype 2/1b recombinants with breakpoints in the NS2 region. These recombinant samples were reported as genotype 1b by NS5b sequencing and as genotype 2 by the VERSANT® HCV Genotype 2.0 Assay which utilizes the 5'UTR region.

## B. Safety Conclusions

The risks of the device (i.e. incorrect genotype result) are based on nonclinical laboratory as well as data collected in a clinical studies conducted to support PMA approval as described above.

The genotyping rate and accuracy study included 800 unique samples. The percentage of specimens that yielded a genotype result was $97.9\%$ for all genotypes and ranged from $90\%$ to $100\%$ for the individual genotypes using the manual method of interpretation as shown in Table 17. The percentage of correct genotypes was $99.6\%$ for all genotypes and ranged from $98.6\%$ to $100\%$ for the individual genotypes.

The VERSANT® HCV Genotype Assay 2.0 (LiPA) has the same inherent risk to patients as any other in vitro diagnostic test that requires that blood be drawn. Therefore, common safety practices must be observed when drawing the sample. There were no significant adverse effects of the VERSANT® HCV Genotype Assay 2.0 (LiPA) assay reported during the studies.

## C. Benefit-Risk Determinaton

The probable benefits of the device are based on data collected in the analytical and clinical studies conducted to support the pre-market approval as described above.

The benefit of using the VERSANT® HCV Genotype Assay 2.0 (LiPA) is that clinicians will be able to accurately determine the HCV genotype for each patient. Thus the patient will receive the correct treatment and achieve SVR without timely delays.

Additional factors to be considered in determining probable risks and benefits for the VERSANT® HCV Genotyping Assay 2.0 (LiPA) device included: inability to decipher between genotype 6 and genotypes 1a or 1b because the core region information is unavailable, inability to identify mixed infections (although this is a rare event in the USA) and recombinant HCV. In all cases, the risk to patient is a delay in treatment or incorrect DAA. Since the course of treatment for the current DAAs is short, the delay is not substantial.

PMA P160016: FDA Summary of Safety and Effectiveness Data
Page 30

{30}

1. Patient Perspectives: This submission did not include specific information on patient perspectives for this device.

In conclusion, given the available information above, the data support that, for detection of HCV genotype for a patient, the probable benefits outweigh the probable risks.

## D. Overall Conclusions

The data in this application support the reasonable assurance of safety and effectiveness of this device when used in accordance with the instructions for use. Data from pre- clinical and clinical data show that the VERSANT® HCV Genotype Assay 2.0 (LiPA) is an accurate method to determine the genotype of HCV present in the patient.

## XIII. CDRH DECISION

CDRH issued an approval order on March 14, 2017. The final conditions of approval cited in the approval order are described below.

## XIV. APPROVAL SPECIFICATIONS

Directions for use: See device labeling.

Hazards to Health from Use of the Device: See Indications, Contraindications, Warnings, Precautions, and Adverse Events in the device labeling.

Post-approval Requirements and Restrictions: See approval order.

PMA P160016: FDA Summary of Safety and Effectiveness Data
Page 31

---

**Source:** [https://fda.innolitics.com/device/P160016](https://fda.innolitics.com/device/P160016)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a PMA](https://innolitics.com/services/regulatory/), [a 510(k)](https://innolitics.com/services/510ks/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
