cobas EGFR MUTATION TEST v2
P150044 · Roche Molecular Systems, Inc. · OWD · Sep 28, 2016 · Pathology
Device Facts
| Record ID | P150044 |
| Device Name | cobas EGFR MUTATION TEST v2 |
| Applicant | Roche Molecular Systems, Inc. |
| Product Code | OWD · Pathology |
| Decision Date | Sep 28, 2016 |
| Decision | APCB |
| Device Class | Class 3 |
| Attributes | Real-World Evidence, Expedited Review |
Real-World Evidence
| Submission | Device | Sponsor | RWD Sources | RWE Use Summary | Key Tags |
|---|
| P150044 · Sep 28, 2016 | cobas EGFR MUTATION TEST v2 | Roche Molecular Systems, Inc. | Banked clinical plasma samples from AURA1 Ext and AURA2 clinical studies | Retrospective analysis of banked plasma specimens from patients previously enrolled in clinical trials was used to bridge the performance of the cobas® EGFR Plasma Test v2 to the previously approved tissue-based test and to establish clinical effectiveness for selecting patients for osimertinib treatment. | Retrospective analysis; Banked clinical samples; Bridging study; Companion diagnostic |
Clinical Evidence
| Study Design | Population | Comparator | Key Endpoints |
|---|
| AURA2 (Protocol D5160C00002) Bridging Study; Retrospective analysis of banked clinical samples; Follow-up/Duration: Not applicable (retrospective) | Patients with advanced or metastatic NSCLC who progressed following prior EGFR-TKI therapy; Sample Size: 344 patients; Number of Sites: Approximately 50 sites worldwide | Validated quantitative next generation sequencing (NGS) method | Agreement between cobas® EGFR Plasma Test v2 and NGS method; ORR for T790M+ patients |
Indications for Use
The cobas® EGFR Mutation Test v2 is a real-time PCR test for the qualitative detection of defined mutations of the epidermal growth factor receptor (EGFR) gene in non-small cell lung cancer (NSCLC) patients. Defined EGFR mutations are detected using DNA isolated from formalin-fixed paraffin-embedded tumor tissue (FFPET) or circulating-free tumor DNA (cfDNA) from plasma derived from EDTA anti-coagulated peripheral whole blood. The test is indicated as a companion diagnostic to aid in selecting NSCLC patients for treatment with the targeted therapies listed in Table 1 below in accordance with the approved therapeutic product labeling: Table 1 Drug FFPET Plasma TARCEVA® (erlotinib) Exon 19 deletions and L858R Exon 19 deletions and L858R TAGRISSOTM (osimertinib) T790M T790M* Patients with positive cobas® EGFR Mutation Test v2 test results using plasma specimens for the presence of the EGFR mutations listed above are eligible for treatment with the corresponding drug as indicated in Table 1 (see Note* for T790M). Patients who are negative for these mutations by this test should be reflexed to routine biopsy and testing for EGFR mutations with the FFPET sample type. *The efficacy of TAGRISSOTM (osimertinib) has not been established in the EGFR T790M plasma-positive, tissue-negative or unknown population and clinical data for T790M plasma-positive patients are limited; therefore, testing using plasma specimens is most appropriate for consideration in patients from whom a tumor biopsy cannot be obtained.
Device Story
Real-time PCR test for qualitative detection of EGFR mutations in NSCLC patients; inputs include DNA from FFPET or cfDNA from plasma; utilizes allele-specific PCR (AS-PCR) chemistry with Z05-AS1 DNA polymerase and fluorescently labeled probes; automated amplification and detection performed on cobas z 480 analyzer; system software processes fluorescent signals to determine mutation status; results provided to clinicians to guide selection of targeted therapies (erlotinib or osimertinib); benefits patients by identifying eligibility for targeted treatment, particularly when tissue biopsy is unavailable.
Clinical Evidence
Retrospective analysis of 344 patients from AURA2 study. Primary endpoint: agreement between cobas® EGFR Plasma Test v2 and validated NGS method for T790M detection. PPA 91.5% (85.7-95.1%), NPA 91.1% (86.0-94.4%). Clinical efficacy (ORR) for osimertinib in T790M+ plasma patients was 64.9% (52.1-70.4%), comparable to tissue-positive cohort (64.1%).
Technological Characteristics
Real-time PCR; allele-specific PCR (AS-PCR) chemistry; Z05-AS1 DNA polymerase; fluorescent dye-labeled oligonucleotide probes; manual sample preparation (glass fiber filter-based); automated amplification/detection on cobas z 480 analyzer; internal control (EGFR exon 28 WT allele); connectivity to LIS.
Indications for Use
Indicated for NSCLC patients to detect defined EGFR mutations in FFPET or plasma cfDNA to aid in selecting patients for targeted therapies (erlotinib or osimertinib). Testing plasma is appropriate when tumor biopsy cannot be obtained.
Regulatory Classification
Identification
The somatic gene mutation detection system is an in vitro diagnostic nucleic acid amplification and real-time detection system for the identification of genetic mutations in DNA extracted from human tumor tissues.
Predicate Devices
- cobas® EGFR Mutation Test (v1) (P120019)
- cobas® EGFR Mutation Test v2 (P150047)
Submission Summary (Full Text)
{0}
# SUMMARY OF SAFETY AND EFFECTIVENESS DATA (SSED)
## I. GENERAL INFORMATION
Device Generic Name: Real-time PCR test
Device Trade Name: cobas® EGFR Mutation Test v2
Device Procode: OWD
Applicant's Name and Address: Roche Molecular Systems, Inc. (RMS)
4300 Hacienda Drive
Pleasanton, CA 94588-2722
Date(s) of Panel Recommendation: None
Premarket Approval Application (PMA) Number: P150044
Date of FDA Notice of Approval: September 28, 2016
Priority Review: Granted priority review status on December 15, 2015 because the device addresses an unmet medical need, as demonstrated by significant clinically meaningful advantage.
The cobas® EGFR Mutation Test v2 was originally approved as a prior version for the qualitative detection of defined mutations in the epidermal growth factor receptor (EGFR) gene in DNA isolated from formalin-fixed paraffin-embedded tumor tissue (FFPET) or circulating-free tumor DNA (cfDNA) from plasma derived from EDTA anti-coagulated peripheral whole blood and associated therapeutic indications. The indications for use of the device as previously approved were to aid in identifying patients with non-small cell lung cancer (NSCLC) whose tumors have defined EGFR mutations and for whom safety and efficacy of a drug have been established as shown in the table below. The SSEDs to support the previously approved FFPET and plasma indications are available on the CDRH website.
Table 1. Previously Approved cobas® EGFR Mutation Test Indications for FFPET and Plasma Specimens.
| Drug | FFPET Specimen Type | | Plasma Specimen Type | |
| --- | --- | --- | --- | --- |
| | Submission # Approval Date (Test version) | Specific EGFR Mutations | Submission # Approval Date (Test version) | Specific EGFR Mutations |
| TARCEVA® (erlotinib) | P120019
May 14, 2013
(v1) | Exon 19 deletions and Exon 20 (L858R) | P150047
June 1, 2016
(v2) | Exon 19 deletions and Exon 20 (L858R) |
PMA P150044: FDA Summary of Safety and Effectiveness Data
{1}
| Drug | FFPET Specimen Type | | Plasma Specimen Type | |
| --- | --- | --- | --- | --- |
| | Submission # Approval Date (Test version) | Specific EGFR Mutations | Submission # Approval Date (Test version) | Specific EGFR Mutations |
| | | substitution mutations | | substitution mutations |
| TAGRISSOTM (osimertinib) | P120019/S007 November 13, 2015 (v2) | Exon 20 T790M substitution mutation | Current submission | Exon 20 T790M substitution mutation |
The current PMA was submitted to add an additional indication for the cobas® EGFR Mutation Test v2 to identify NSCLC patients with the exon 20 T790M substitution mutation from circulating-free tumor DNA (cfDNA) isolated from plasma, for treatment with TAGRISSOTM (osimertinib).
## II. INDICATIONS FOR USE
The cobas® EGFR Mutation Test v2 is a real-time PCR test for the qualitative detection of defined mutations of the epidermal growth factor receptor (EGFR) gene in non-small cell lung cancer (NSCLC) patients. Defined EGFR mutations are detected using DNA isolated from formalin-fixed paraffin-embedded tumor tissue (FFPET) or circulating-free tumor DNA (cfDNA) from plasma derived from EDTA anti-coagulated peripheral whole blood.
The test is indicated as a companion diagnostic to aid in selecting NSCLC patients for treatment with the targeted therapies listed in Table 1 below in accordance with the approved therapeutic product labeling:
Table 1
| Drug | FFPET | Plasma |
| --- | --- | --- |
| TARCEVA® (erlotinib) | Exon 19 deletions and L858R | Exon 19 deletions and L858R |
| TAGRISSOTM (osimertinib) | T790M | T790M* |
Patients with positive cobas® EGFR Mutation Test v2 test results using plasma specimens for the presence of the EGFR mutations listed above are eligible for treatment with the corresponding drug as indicated in Table 1 (see Note* for T790M). Patients who are negative for these mutations by this test should be reflexed to routine biopsy and testing for EGFR mutations with the FFPET sample type.
*The efficacy of TAGRISSOTM (osimertinib) has not been established in the EGFR T790M plasma-positive, tissue-negative or unknown population and clinical data for T790M plasma-positive patients are limited; therefore, testing using plasma specimens is most appropriate for consideration in patients from whom a tumor biopsy cannot be obtained.
PMA P150044: FDA Summary of Safety and Effectiveness Data
{2}
Drug safety and efficacy have not been established for the EGFR mutations listed in Table 2 below that are also detected by the cobas® EGFR Mutation Test v2:
Table 2
| Drug | FFPET | Plasma |
| --- | --- | --- |
| TARCEVA®
(erlotinib) | G719X, exon 20 insertions,
T790M, S768I and L861Q | G719X, exon 20 insertions,
T790M, S768I and L861Q |
| TAGRISSO™
(osimertinib) | G719X, exon 19 deletions,
L858R, exon 20 insertions,
S768I, and L861Q | G719X, exon 19 deletions,
L858R, exon 20 insertions,
S768I, and L861Q |
For manual sample preparation, FFPET specimens are processed using the cobas® DNA Sample Preparation Kit and plasma specimens are processed using the cobas® cfDNA Sample Preparation Kit. The cobas z 480 analyzer is used for automated amplification and detection.
## III. CONTRAINDICATIONS
There are no known contraindications.
## IV. WARNINGS AND PRECAUTIONS
The warnings and precautions can be found in the cobas® EGFR Mutation Test v2 labeling.
## V. DEVICE DESCRIPTION
The cobas® EGFR Mutation Test v2 is based on two major processes: (1) manual sample preparation to obtain DNA from FFPET or plasma; and (2) PCR amplification and detection of target DNA using complementary primer pairs and oligonucleotide probes labeled with fluorescent dyes.
The cobas® EGFR Mutation Test v2 is comprised of the following:
1. The cobas® EGFR Mutation Test v2 kit provides reagents for automated real-time PCR amplification and detection of the EGFR mutations.
2. The cobas® DNA Sample Preparation Kit provides reagents for manual specimen preparation to obtain genomic DNA from formalin-fixed, paraffin-embedded tumor tissue (FFPET).
3. The cobas® cfDNA Sample Preparation Kit provides reagents for manual specimen preparation to obtain cfDNA from plasma.
Two external run controls are provided and the EGFR exon 28 wild-type (WT) allele serves as an internal, full process control.
PMA P150044: FDA Summary of Safety and Effectiveness Data
{3}
# A. Specimen Preparation - Plasma
Plasma specimens are processed and cfDNA is isolated using the cobas® cfDNA Sample Preparation Kit, a generic manual specimen preparation based on nucleic acid binding to glass fibers. The components of the cobas® cfDNA Sample Preparation Kit are identical to those of the cobas® DNA Sample Preparation Kit, with the exception of the High Pure Extension Assembly (HPEA) units.
Two milliliters (mL) of plasma are processed with a protease and chaotropic lysis/binding buffer that protects the cfDNA from DNases. Subsequently, isopropanol is added to the binding mixture that is then centrifuged through a column with a glass fiber filter insert. During centrifugation, the cfDNA is bound to the surface of the glass fiber filter. Unbound substances, such as salts, proteins and other impurities, are removed by centrifugation. The adsorbed nucleic acids are washed and then eluted with an aqueous solution.
# B. PCR Amplification and Detection
# Target Selection and Amplification
The cobas® EGFR Mutation Test v2 kit uses primers that define specific base-pair sequences for each of the targeted mutations. For the exon 19 deletion mutations, sequences ranging from 125 to 141 base pairs are targeted; for the L858R substitution mutation in exon 21, a 138 base pair sequence is targeted; for the T790M substitution mutation in exon 20, a 118 base pair sequence is targeted; for the G719X substitution mutation in exon 18, sequences ranging from 104 to 106 base pairs are targeted; for the S768I substitution mutation in exon 20, a 133 base pair sequence is targeted; for the exon 20 insertion mutations, sequences ranging from 125 to 143 base pairs are targeted; for the L861Q substitution mutation in exon 21, a 129 base pair sequence is targeted; for the internal control in exon 28, an 87 base pair sequence is targeted. Amplification occurs only in the regions of the EGFR gene between the primers; the entire EGFR gene is not amplified.
The cobas® EGFR Mutation Test v2 uses allele-specific PCR (AS-PCR) chemistry for amplification and detection. The selected AS-PCR primers specifically amplify the targeted mutant sequences over the WT sequences and/or other human DNA. The cobas® EGFR Mutation Test v2 is designed to use three master mix (MMx) reagents which are run in three separate wells. The number and types of primers and probes differ based on the particular target(s). The cobas® EGFR Mutation Test v2 detects the following EGFR mutations in exons 18, 19, 20, and 21:
Table 2. EGFR Mutations Detected by the cobas® EGFR Mutation Test v2
| Exon | EGFR Mutation | EGFR Nucleic Acid Sequence | COSMIC ID1 |
| --- | --- | --- | --- |
| Exon 18 | G719X | 2156G>C | 6239 |
| | | 2155G>A | 6252 |
PMA P150044: FDA Summary of Safety and Effectiveness Data
{4}
| Exon | EGFR Mutation | EGFR Nucleic Acid Sequence | COSMIC ID1 |
| --- | --- | --- | --- |
| | | 2155G>T | 6253 |
| Exon 19 | Ex. 19del | 2240_2251del12 | 6210 |
| | | 2239_2247del9 | 6218 |
| | | 2238_2255del18 | 6220 |
| | | 2235_2249del15 | 6223 |
| | | 2236_2250del15 | 6225 |
| | | 2239_2253del15 | 6254 |
| | | 2239_2256del18 | 6255 |
| | | 2237_2254del18 | 12367 |
| | | 2240_2254del15 | 12369 |
| | | 2240_2257del18 | 12370 |
| | | 2239_2248TTAAGAGAAG>C | 12382 |
| | | 2239_2251>C | 12383 |
| | | 2237_2255>T | 12384 |
| | | 2235_2255>AAT | 12385 |
| | | 2237_2252>T | 12386 |
| | | 2239_2258>CA | 12387 |
| | | 2239_2256>CAA | 12403 |
| | | 2237_2253>TTGCT | 12416 |
| | | 2238_2252>GCA | 12419 |
| | | 2238_2248>GC | 12422 |
| | | 2237_2251del15 | 12678 |
| | | 2236_2253del18 | 12728 |
| | | 2235_2248>AATTC | 13550 |
| | | 2235_2252>AAT | 13551 |
| | | 2235_2251>AATTC | 13552 |
| | | 2253_2276del24 | 13556 |
| | | 2237_2257>TCT | 18427 |
| | | 2238_2252del15 | 23571 |
| 2233_2247del15 | 26038 | | |
| Exon 20 | S768I | 2303G>T | 6241 |
| | T790M | 2369C>T | 6240 |
| | Ex. 20ins | 2307_2308ins9GCCAGCGTG | 12376 |
| | | 2319_2320insCAC | 12377 |
| | | 2310_2311insGGT | 12378 |
| | | 2311_2312ins9GCGTGGACA | 13428 |
| | | 2309_2310AC>CCAGCGTGGAT | 13558 |
| Exon 21 | L858R | 2573T>G | 6224 |
| | | 2573_2574TG>GT | 12429 |
| | L861Q | 2582T>A | 6213 |
$^{1}$ Catalogue of Somatic Mutations in Cancer (COSMIC), 2011, v.51. http://www.sanger.ac.uk/genetics/CGP/cosmic.
PMA P150044: FDA Summary of Safety and Effectiveness Data
{5}
MMx1 (first amplification reaction) contains:
- Fourteen AS-PCR primers, one common primer, and one common probe are used to detect the Exon 19 deletion and complex (defined as the combination of a deletion and an insertion) mutations.
- One AS-PCR primer, one common primer, and one common probe are used to detect the S768I mutation.
MMx2 (second amplification reaction) contains:
- One AS-PCR primer, one common primer, and one common probe are used to detect the L858R mutation.
- One AS-PCR primer, one common primer, and one common probe are used to detect the T790M mutation.
MMx3 v2 (third amplification reaction) contains:
- Three AS-PCR primers, one common primer, and one common probe are used to detect G719X mutations.
- Three AS-PCR primers, one common primer, and one common probe are used to detect Exon 20 insertion mutations.
- One AS-PCR primer, one common primer, and one common probe are used to detect the L861Q mutation.
A derivative of Thermus species Z05-AS1 DNA polymerase is utilized for target amplification. The PCR reaction mixture is heated to denature the DNA and expose the primer target sequences. As the mixture cools, the upstream and downstream primers anneal to the target DNA sequences. The Z05-AS1 DNA polymerase, in the presence of divalent metal ion and excess dNTPs, extends each annealed primer, thus synthesizing a second DNA strand. This completes the first cycle of PCR, yielding a double-stranded DNA copy, which includes the targeted base-pair regions of the EGFR gene. This process is repeated for a number of cycles, with each cycle effectively doubling the amount of amplicon DNA.
Selective amplification of target nucleic acid from the specimen is achieved in the cobas® EGFR Mutation Test v2 by the use of AmpErase® (uracil-N-glycosylase) enzyme and deoxyuridine triphosphate (dUTP), which are included in the Master Mix reagents. The AmpErase enzyme recognizes and catalyzes the destruction of DNA strands containing deoxyuridine, but not DNA containing thymidine. Deoxyuridine is always present in the amplicons due to the use of dUTP as one of the nucleotide triphosphates in the Reaction Mix reagent; therefore, only the amplicons contain deoxyuridine. The AmpErase® enzyme is inactive at temperatures above 55°C, i.e., throughout the thermal cycling steps, and therefore does not destroy the target amplicon.
## Automated Real-time Detection
The cobas® EGFR Mutation Test v2 utilizes real-time PCR technology. Each target-specific, oligonucleotide probe in the reaction is labeled with a fluorescent dye that
PMA P150044: FDA Summary of Safety and Effectiveness Data
Page 6
{6}
serves as a reporter, and with a quencher molecule that absorbs (quenches) fluorescent emissions from the reporter dye within an intact probe. During each cycle of amplification, a probe complementary to the single-stranded DNA sequence in the amplicon binds and is subsequently cleaved by the 5′ to 3′ nuclease activity of the Z05-AS1 DNA polymerase. Once the reporter dye is separated from the quencher by this nuclease activity, fluorescence of a characteristic wavelength can be measured when the reporter dye is excited by the appropriate spectrum of light. Two different reporter dyes are used to label the mutations targeted by the test. Amplification of the targeted EGFR sequences are detected independently across three reactions by measuring fluorescence at the two characteristic wavelengths in dedicated optical channels.
## Instrument and Software
The cobas® 4800 system is controlled by the cobas® 4800 system software, which provides the core software engines and user interfaces. This core system software was designed to allow multiple assays to be performed on the system using assay specific analysis package software (ASAP). The cobas® z480 analyzer component of the test system also has its own internal instrument control software, which is driven by the core software.
A dedicated Control Unit computer runs the cobas® 4800 system software and provides an interface to the cobas® z 480 and Laboratory Information System (LIS). The computer also processes the fluorescent signals with the assay specific analysis package (ASAP) and stores the test results in a controlled database. The complete system allows a user to create a test work order for each specimen either manually or automatically when connected to a LIS. A software wizard guides the user through the necessary steps to perform a run, which includes cobas z 480 maintenance handling, test selection, specimen ID entry, reagent and microwell plate barcode entry, microwell plate loading and run start.
The cobas® 4800 system tracks each specimen during processing and analysis on the cobas z 480 analyzer. Once the thermal run is complete the ASAP software processes the fluorescence data using data analysis algorithms, assesses the validity of the controls and determines the results using the assay specific result interpretation logic. The software then provides the results to the user in three formats: a printable PDF results report, a GUI based result viewer and a result export file that can be exported to the LIS.
The cobas® 4800 system software includes the assay-specific cobas® 4800 EGFR Analysis Package (AP) software, which contains an algorithm to determine sample results and run validity. The analysis package used to analyze the EGFR results from cfDNA specimens isolated from plasma differs from the one used to analyze EGFR results from DNA specimens isolated from FFPET specimens. The overall cobas® 4800 system components are shown in Figure 1 below.
PMA P150044: FDA Summary of Safety and Effectiveness Data
Page 7
{7}
The final version of the cobas® 4800 core system software is v2.1.0 or v2.2.0 and the ASAP software used to analyze all studies in this PMA was EGFR Plasma P2 AP v1.0.1.1555. The final commercial ASAP software is EGFR Plasma P2 Analysis Package v1.0.1.1567.

Fig 1. cobas® 4800 System
# Interpretation of Results
If the run is valid, then the cycle threshold (Ct) or CtR (relative cycle threshold) values for each sample will be evaluated against acceptable ranges for each channel. The CtR value is determined by calculating the difference between the mutations observed Ct and the corresponding Internal Control (IC) Ct value from the same Master Mix. Ct values are not available to the user. Tables 3 and 4 below summarize how the individual amplification Master Mix results are combined to provide an overall result.
Table 3. Individual Amplification Master Mix Results to Overall Reported Results.
| MMx1 Result | MMx2 Result | MMx3 v2 Result | Reported Result |
| --- | --- | --- | --- |
| Valid, No Mutation Detected | Valid, No Mutation Detected | Valid, No Mutation Detected | Valid, No Mutation Detected |
| Valid, No Mutation Detected | Valid, No Mutation Detected | Valid, Mutation Detected | Valid, Mutation Detected |
| Valid, No Mutation Detected | Valid, No Mutation Detected | Invalid | Invalid |
| Valid, No Mutation Detected | Valid, Mutation Detected | Valid, No Mutation Detected | Valid, Mutation Detected |
| Valid, No Mutation Detected | Valid, Mutation Detected | Valid, Mutation Detected | Valid, Mutation Detected |
| Valid, No Mutation Detected | Valid, Mutation Detected | Invalid | Invalid |
PMA P150044: FDA Summary of Safety and Effectiveness Data
{8}
Table 4. Result Interpretation of the cobas® EGFR Mutation Test v2
| Test Result | Mutation Result** | Interpretation |
| --- | --- | --- |
| Mutation Detected (MD) | Ex.19del L858R T790M | Mutation detected in specified targeted EGFR region. |
PMA P150044: FDA Summary of Safety and Effectiveness Data
{9}
| Test Result | Mutation Result** | Interpretation |
| --- | --- | --- |
| | S768I
G719X
Ex. 20ins
L861Q
(More than one mutation may be present) | |
| No Mutation Detected* (NMD) | N/A | No mutation detected in targeted EGFR regions |
| Invalid | N/A | Specimen result is invalid. |
| Failed | N/A | Failed run due to hardware or software failure. |
*A No Mutation Detected (NMD) result does not preclude the presence of a mutation in the targeted EGFR regions, because results depend on percent mutant sequences, adequate specimen integrity, absence of inhibitors, and sufficient DNA to be detected.
**Italicized mutation results consist of mutations that are intended for analytical detection only for both FFPET and plasma specimens and the T790M mutation is for analytical detection only when identified in plasma specimens.
## Test Controls
One EGFR mutant control and one EGFR negative control are provided. The EGFR WT allele located on exon 28 serves as an internal, full process control.
1. **EGFR Mutant Control**: The Mutant Control is a blend of seven DNA plasmids containing specified EGFR mutation sequences and cell line DNA that is WT for EGFR. The Mutant Control is composed of plasmids representing the most frequently observed mutation for each mutation class detected by the test. The Mutant Control will be included in every run and will serve as a process control for amplification and detection. The Mutant Control must yield Cycle threshold (Ct) values for all targeted mutations and the Internal Control (IC) within the respective acceptable ranges for the run to be considered valid.
2. **EGFR Negative Control**: The Negative Control is a full process contamination control for a given test batch of specimens. The Negative Control consists of a blank vial containing no specimen (specimen diluent only) which is processed through specimen preparation and the resulting eluate is subsequently used for amplification and detection. The Negative Control Ct values must be either not detected or greater than the pre-established Ct maximum value for all targeted mutation groups and the IC for the run to be considered valid.
3. **EGFR WT Internal Control (IC)**: The Internal Control in EGFR exon 28 from test specimens serves as a full process control. This control ensures that every step of the process from specimen preparation to amplification and detection has been completed successfully.
## VI. ALTERNATIVE PRACTICES AND PROCEDURES
PMA P150044: FDA Summary of Safety and Effectiveness Data
Page 10
{10}
There are no other FDA-cleared or -approved alternatives for the testing of plasma for EGFR mutation status in the selection of patients who are eligible for treatment with TAGRISSO™ (osimertinib).
## VII. MARKETING HISTORY
The cobas® EGFR Mutation Test (v1) was introduced into the United States starting on May 14, 2013. The cobas® EGFR Mutation Test (v1) is commercially available in the following countries: Argentina, Australia, Austria, Belgium, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Ecuador, Estonia, Finland, France, Germany, Greece, Guatemala, Hong Kong, Hungary, Iceland, India, Indonesia, Ireland, Italy, Japan, Korea, Latvia, Liechtenstein, Lithuania, Luxembourg, Malaysia, Malta, Mexico, Netherlands, New Zealand, Nicaragua, Norway, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Romania, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, United States, Uruguay, Venezuela, Vietnam.
The cobas® EGFR Mutation Test (v1) was replaced by the cobas® EGFR Mutation Test v2, which was introduced into the United States starting on November 13, 2015 for use with FFPET specimens. The cobas® EGFR Mutation Test v2 for use with FFPET and plasma specimens is commercially available in Australia, Austria, Belgium, Bolivia, Bulgaria, Colombia, Croatia, Cyprus, Czechia, Denmark, Ecuador, Estonia, Finland, France, Germany, Greece, Guatemala, Hong Kong, Hungary, Iceland, Ireland, Italy, Latvia, Liechtenstein, Lithuania, Luxembourg, Malta, Netherlands, New Zealand, Norway, Panama, Paraguay, Poland, Portugal, Romania, Singapore, Slovakia, Slovenia, Spain, Sweden, Switzerland, Thailand, Turkey, United Arab Emirates, United Kingdom, Uruguay, and Vietnam. The cobas® EGFR Mutation Test v2 for use with FFPET specimens is also commercially available in Japan. The cobas® EGFR Mutation Test v2 for use with plasma specimens to select patients for treatment with TARCEVA® (erlotinib) was approved on June 1, 2016 and is available in the US.
Neither version of the cobas® EGFR Mutation Test for either specimen type, has been withdrawn from the market for reasons related to safety and effectiveness.
## VIII. POTENTIAL ADVERSE EFFECTS OF THE DEVICE ON HEALTH
Failure of the device to perform as expected or failure to correctly interpret test results may lead to incorrect cobas® EGFR Mutation Test v2 results and subsequently improper patient management decisions in NSCLC treatment. No adverse events were reported in connection with the studies used to support this PMA as the studies were performed retrospectively using banked samples.
## IX. SUMMARY OF PRECLINICAL STUDIES
### A. Laboratory Studies
PMA P150044: FDA Summary of Safety and Effectiveness Data
Page 11
{11}
# 1. Correlation with Reference Method Using Clinical Samples
The analytical accuracy of the cobas® EGFR Mutation Test v2 using plasma specimens (cobas® EGFR Plasma Test v2) was assessed by comparing its results to a validated quantitative next generation sequencing (NGS) method using banked clinical samples from the AURA2 clinical study. The percent mutation present in specimens was determined for all specimens used to demonstrate performance. Plasma specimens from a second clinical study, AURA1 Extension (AURA1 Ext), were used to supplement the AURA2 study specimens to establish the detection of rare EGFR mutations.
# a. Specimen Processing, Disposition, and Testing
Plasma specimens were sent to one of the participating central laboratories after written informed consent was confirmed. cfDNA was isolated manually according to the Instructions for Use for the cobas® cfDNA Sample Preparation Kit. Testing was performed according to the Instructions for Use for the cobas® EGFR Plasma Test v2. Tissue specimens were processed according to the Instructions for Use for the cobas® DNA Sample Preparation Kit.
# b. Accountability of the AURA2 Cohort
Three hundred forty-four (344) of the 383 patients with matching tissue specimens had sufficient plasma samples available for testing by the cobas® EGFR Plasma Test v2. Of those, 322 (93.6%) samples had sufficient plasma for testing by the validated NGS reference method (see Table 6 below).
# c. Agreement to Reference Method
# i. Agreement for T790M Mutation:
Patient enrollment into the AURA2 study was based on FFPET biopsy test results using the cobas® EGFR Mutation Test v1 (cobas® EGFR Tissue Test v1), which were later bridged to test version 2 (v2). Based on the plasma samples tested from the AURA2 study, the agreements of the cobas® EGFR Plasma Test v2 and the NGS method for detection of the T790M mutation are presented in Table 5 below, stratified by the cobas® EGFR Tissue Test v1 results. Agreement between the three test methods, when stratified by the cobas® EGFR Tissue Test v2 also yielded similar results.
Table 5. Three-Way Agreement Among cobas® EGFR Tissue Test v1, cobas® EGFR Plasma Test v2, and NGS for Detection of T790M
| cobas® EGFR Tissue Test v1 | cobas® EGFR Plasma Test v2 | NGS | | | Total |
| --- | --- | --- | --- | --- | --- |
| | | T790M+ | T790M(-) | No Plasma Sample | |
| T790M+ | T790M+ | 109 | 13 | 9 | 131 |
PMA P150044: FDA Summary of Safety and Effectiveness Data
{12}
| cobas® EGFR Tissue Test v1 | cobas® EGFR Plasma Test v2 | NGS | | | Total |
| --- | --- | --- | --- | --- | --- |
| | | T790M+ | T790M(-) | No Plasma Sample | |
| | T790M(-) | 8 | 79 | 5 | 92 |
| | Invalid/No Plasma Sample | 3 | 0 | 7 | 10 |
| | Total | 120 | 92 | 21 | 233 |
| T790M(-) | T790M+ | 18 | 3 | 1 | 22 |
| | T790M(-) | 3 | 79 | 7 | 89 |
| | Invalid/No Plasma Sample | 0 | 0 | 29 | 29 |
| | Total | 21 | 82 | 37 | 140 |
| Invalid | T790M+ | 2 | 0 | 0 | 2 |
| | T790M(-) | 1 | 5 | 0 | 6 |
| | Invalid/No Plasma Sample | 0 | 0 | 2 | 2 |
| | Total | 3 | 5 | 2 | 10 |
T790M+ = T790M mutation detected (MD); T790M(-) = T790M mutation not detected (NMD)
Percent agreements and their 95% 2-sided confidence intervals (CIs) for the T790M mutation between the cobas® EGFR Plasma Test v2 and the NGS comparator method are shown in Table 6. A sensitivity analysis was conducted for the worst case scenario where invalid or missing results were treated as discordant results between the cobas® EGFR Plasma Test v2 and the NGS method. The estimates are shown in Table 6.
Table 6. Agreement between cobas® EGFR Plasma Test v2 and NGS for Detection of T790M Mutation
| | NGS | | | | Total | |
| --- | --- | --- | --- | --- | --- | --- |
| | | T790M+ | T790M(-) | Invalid | | No Plasma Sample |
| cobas® EGFR Plasma Test v2 | T790M(+) | 129 | 16 | 0 | 10 | 155 |
| | T790M(-) | 12 | 163 | 0 | 12 | 187 |
| | Invalid | 2 | 0 | 0 | 0 | 2 |
| | Total | 143 | 179 | 0 | 22 | 344 |
| Valid Results Only (Total N=320) | PPA (95% CI) | 91.5% (85.7%, 95.1%) | | | | |
| | NPA (95% CI) | 91.1% (86.0%, 94.4%) | | | | |
| | PPV (95% CI) | 89.0% (82.8%, 93.1%) | | | | |
| | NPV (95% CI) | 93.1% (88.4%, 96.0%) | | | | |
| Sensitivity Analysis (Worst Case Scenario) (Total N=344) | PPA (95% CI) | 83.2% (76.6%, 88.3%) | | | | |
| | NPA (95% CI) | 86.3% (80.6%, 90.4%) | | | | |
| | PPV (95% CI) | 83.2% (76.6%, 88.3%) | | | | |
| | NPV (95% CI) | 86.3% (80.6%, 90.4%) | | | | |
PMA P150044: FDA Summary of Safety and Effectiveness Data
{13}
ii. Agreement for Ex. 19del and L858R Mutations:
As part of the enrollment criteria into the AURA2 study, all patients were required to demonstrate the presence of an activating mutation in addition to T790M. Based on the plasma samples tested from the AURA2 study, the agreements of the cobas® EGFR Plasma Test v2 and the NGS method for detection of EGFR Ex. 19del and L858R mutations are presented in Table 7 below.
Table 7. Agreement between cobas® EGFR Plasma Test v2 and NGS for Detection of Individual Ex 19del and L858R Sensitizing Mutations
| cobas® EGFR Plasma Test v2 | NGS Method | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | Ex. 19del | | | | L858R | | | | |
| | MD | NMD | No Sample | Total | MD | NMD | No Sample | Invalid | Total |
| MD | 137 | 12 | 16 | 165 | 83 | 11 | 0 | 3 | 97 |
| MND | 11 | 160 | 6 | 177 | 3 | 222 | 1 | 19 | 245 |
| Invalid | 1 | 1 | 0 | 2 | 1 | 1 | 0 | 0 | 2 |
| Total | 149 | 173 | 22 | 344 | 87 | 234 | 1 | 22 | 344 |
| PPA (95% CI) | 92.6% (87.2%, 95.8%) | | | | 96.5% (90.2%, 98.8%) | | | | |
| NPA (95% CI) | 93.0% (88.2%, 96.0%) | | | | 95.3% (91.7%, 97.3%) | | | | |
The agreement between the cobas® EGFR Plasma Test v2 and the NGS method for the detection of mutations in exons 18, 19, 20 and 21 in aggregate in the EGFR gene are shown in Table 8 below.
Table 8. Agreement between cobas® EGFR Plasma Test v2 and NGS for Detection of All EGFR Mutations
| | NGS | | | | Total | |
| --- | --- | --- | --- | --- | --- | --- |
| | | EGFRm (+) | EGFRm (-) | Invalid | | No Plasma Sample |
| cobas® EGFR Plasma Test v2 | EGFRm(+) | 232 | 16 | 0 | 21 | 269 |
| | EGFRm (-) | 15 | 57 | 0 | 1 | 73 |
| | Invalid | 2 | 0 | 0 | 0 | 2 |
| | Total | 249 | 73 | 0 | 22 | 344 |
| Valid Results Only (Total N=320) | PPA (95% CI) | 93.9% (90.2%, 96.3%) | | | | |
| | NPA (95% CI) | 78.1% (67.3%, 86.0%) | | | | |
| | PPV (95% CI) | 93.5% (89.8%, 96.0%) | | | | |
| | NPV (95% CI) | 79.2% (68.4%, 86.9%) | | | | |
| Sensitivity Analysis (Worst Case Scenario) (Total N=344) | PPA (95% CI) | 92.8% (88.9%, 95.4%) | | | | |
| | NPA (95% CI) | 60.6% (50.5%, 69.9%) | | | | |
| | PPV (95% CI) | 86.2% (81.6%, 89.9%) | | | | |
| | NPV (95% CI) | 76.0% (65.2%, 84.2%) | | | | |
$\mathrm{EGFRm}(+) = \mathrm{EGFR}$ mutation detected (MD), $\mathrm{EGFRm}(-) = \mathrm{EGFR}$ mutation not detected (NMD)
PMA P150044: FDA Summary of Safety and Effectiveness Data
{14}
iii. Rare Mutations:
All four rare mutations that the cobas® EGFR Plasma Test v2 was designed to detect (i.e., Ex. 20ins, L861Q, S768I, and G719X) were detected and confirmed by the validated NGS method using plasma samples from the AURA1 Ext and AURA2 studies. The results are summarized in Table 9.
Table 9. Agreement of Rare Mutations between cobas® EGFR Plasma Test v2 and NGS.
| Study | Rare Mutations | MD by cobas® EGFR Plasma Test v2, NMD by NGS | MD by both cobas® EGFR Plasma Test v2 and NGS | NMD by cobas® EGFR Plasma Test v2, MD by NGS | PPA % |
| --- | --- | --- | --- | --- | --- |
| Total | G719X | 1 | 10 | 6 | 62.5 |
| | Ex 20ins | 0 | 0 | 1 | 0 |
| | S768I | 0 | 5 | 1 | 83.33 |
| | L861Q | 0 | 4 | 0 | 100 |
In four of five (4/5) of the missed calls by the cobas® EGFR Plasma Test v2 for G719X from AURA2, percent mutation in the specimens were above the NGS LoD (0.18%) but below the mutation cut-off for the cobas® EGFR Plasma Test v2. For the fifth G719X and the one S768I mutations, which were missed by the cobas® EGFR Plasma Test v2, the specimens' Ct cut-off values exceeded the maximum Ct cut-offs for the respective mutations by the cobas® EGFR Plasma Test v2. The Ex. 20ins mutation that was not detected by the cobas® EGFR Plasma Test v2 was not a variant that the test was designed to detect. No other Ex. 20ins mutations were detected by either the cobas® EGFR Plasma Test v2 or NGS.
2. Contrived Sample Comparison
See Summary of Safety and Effectiveness Data for P150047.
3. Analytical Sensitivity
See Summary of Safety and Effectiveness Data for P150047.
4. Analytical Specificity
Studies to assess both Inclusivity/Cross-Reactivity and Exclusivity were not performed for this PMA as there has been no change to the assay's primers, probes, or targeted mutations. (See Summary of Safety and Effectiveness Data for P120019/S007.)
PMA P150044: FDA Summary of Safety and Effectiveness Data
{15}
Studies to assess interference by potentially interfering endogenous and exogenous substances, and microorganism testing are summarized in the Summary of Safety and Effectiveness Data for P150047.
5. Reproducibility and Precision
See Summary of Safety and Effectiveness Data for P150047.
6. Lot-to-Lot Reproducibility
See Summary of Safety and Effectiveness Data for P150047.
7. Specimen Handling
See Summary of Safety and Effectiveness Data for P150047.
8. Guard banding
See Summary of Safety and Effectiveness Data for P150047.
9. Stability Studies
See Summary of Safety and Effectiveness Data for P150047.
B. Animal Studies
None.
C. Additional Studies
None.
X. SUMMARY OF PRIMARY CLINICAL STUDY
Roche Molecular Systems, Inc. conducted a bridging study to support the safety and effectiveness of the cobas® EGFR Plasma Test v2 to select patients for treatment with osimertinib by detecting the presence of the EGFR T790M substitution mutation in EDTA plasma specimens. The specimens used in the bridging study consisted of plasma specimens, collected at baseline, from patients with NSCLC who progressed following prior EGFR tyrosine kinase inhibitor (EGFR-TKI) therapy.
Patient specimen cohorts from two AstraZeneca clinical studies, AURA1 Ext (Protocol D5160C00001) and AURA2 (Protocol D5160C00002), were utilized to support this PMA submission. The AURA1 Ext Phase 1 study and the AURA2 Phase II study enrolled patients with advanced or metastatic NSCLC who had progressed following an EGFR-TKI for second line treatment with osimertinib along with patients who
PMA P150044: FDA Summary of Safety and Effectiveness Data
Page 16
{16}
progressed after treatment with an EGFR-TKI and at least one platinum-based doublet chemotherapy. The patients were enrolled based on an FFPET specimen positive EGFR sensitizing mutation and a T790M substitution mutation result using the cobas® EGFR Mutation Test v1. Enrollment for both studies was conducted at approximately 50 sites worldwide. These clinical studies were used to support accelerated approval of TAGRISSO (osimertinib) under NDA 208065. The AURA1 Ext study specimens were used only to supplement specimens with rare EGFR mutations (see Section IX.A.1.c.iii, above). The AURA2 study supported the approval of the cobas® EGFR Mutation Test v2 using FFPET specimens under P120019/S007.
## A. Study Design
The study design for the AURA2 study describing the enrollment of patients based on testing of FFPET specimens is available in the Summary of Safety and Effectiveness for P120019/S007.
### 1. Inclusion/Exclusion Criteria
Patients were enrolled into the AURA2 study and continued on treatment until the occurrence of disease progression (based on RESIST 1.1) or unacceptable toxicity. For the inclusion/exclusion criteria for the original AURA2 study and the bridging study between the first and second versions of the cobas® EGFR Tissue Test, see the Summary of Safety and Effectiveness for P120019/S007.
The inclusion/exclusion criteria for inclusion into the bridging study between the cobas® EGFR Tissue Test and the cobas® EGFR Plasma Test v2 are summarized below:
#### Inclusion Criteria:
Samples from subjects who were eligible for screening under the AURA1 Ext and AURA2 study protocols with a valid EGFR mutation result were eligible for this protocol. The criteria also indicated that:
- An FFPET NSCLC biopsy following progression of disease on the latest line of therapy must have been available for mutation analysis.
- A concurrently obtained plasma sample must have been available for analysis.
#### Exclusion criterion:
Samples were excluded if there was insufficient plasma volume for testing.
### 2. Follow-up Schedule
See the Summary of Safety and Effectiveness for P120019/S007.
### 3. Clinical Endpoints
See the Summary of Safety and Effectiveness for P120019/S007.
PMA P150044: FDA Summary of Safety and Effectiveness Data
Page 17
{17}
# 4. Bridging Study
In this bridging study, plasma specimens were collected at baseline screening from patients enrolled in the AURA2 clinical study. The available specimens from the study were tested by both the cobas® EGFR Plasma Test v2 and a validated NGS method. Agreements between the detection of EGFR T790M mutation status and EGFR mutations in exons 18, 19, 20 and 21, in aggregate, between the patients' plasma and tissue specimen results are shown in the Tables 10 and 12 below, respectively. Sensitivity analyses to account for missing data were also performed and the results are included in Tables 11 and 13 below.
Table 10. Agreement between cobas® EGFR Plasma Test v2 and cobas® EGFR Tissue Test v1 and v2 for Detection of T790M
| cobas® EGFR Plasma Test v2 | cobas® EGFR Tissue Test v1 | | | | cobas® EGFR Tissue Test v2 | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | T790M+ | T790M (-) | Invalid | Total | T790M+ | T790M (-) | Invalid | Total |
| T790M+ | 131 | 22 | 2 | 155 | 128 | 22 | 5 | 155 |
| T790M(-) | 92 | 89 | 6 | 187 | 91 | 90 | 6 | 187 |
| Invalid | 2 | 0 | 0 | 2 | 2 | 0 | 0 | 2 |
| No Plasma Sample | 8 | 29 | 2 | 39 | 10 | 27 | 2 | 39 |
| Total | 233 | 140 | 10 | 383 | 231 | 139 | 13 | 383 |
| PPA (95% CI) | 58.7% (52.2%, 65.0%) | | | | 58.4% (51.8%, 64.8%) | | | |
| NPA (95% CI) | 80.2% (71.8%, 86.5%) | | | | 80.4% (72.0%, 86.7%) | | | |
| PPV (95% CI) | 85.6% (79.2%, 90.3%) | | | | 85.3% (78.8%, 90.1%) | | | |
| NPV (95% CI) | 49.2% (42.0%, 56.4%) | | | | 49.7% (42.5%, 56.9%) | | | |
Table 11. Sensitivity Analysis of Agreement of cobas® EGFR Plasma Test v2 and cobas® EGFR Tissue Test v2 in Detecting T790M Mutation
| | PPA (95% CI) | NPA (95% CI) | PPV (95% CI) | NPV (95% CI) |
| --- | --- | --- | --- | --- |
| With invalid as discordant result | 56.4% (49.9%, 62.7%) | 76.9% (68.5%, 83.6%) | 82.6% (75.8%, 87.7%) | 47.6% (40.6%, 54.7%) |
| With invalid & no samples as discordant result | 54.0% (47.6%, 60.2%) | 62.5% (54.4%, 70.0%) | 70.3% (63.3%, 76.5%) | 45.2% (38.5%, 52.2%) |
| Verification bias adjusted for invalid and no sample w/ bootstrap method for CI | 58.4% (51.9%, 64.6%) | 80.7% (74.2%, 86.8%) | 83.1% (78.0%, 88.0%) | 54.3% (49.6%, 59.3%) |
Table 12. Agreement between cobas® EGFR Plasma Test v2 and Tissue Test v2 for Overall EGFR Mutation Detection
| cobas® EGFR Plasma Test v2 | cobas® EGFR Tissue Test v2 | | | Total |
| --- | --- | --- | --- | --- |
| | EGFRm+ | EGFRm(-) | Invalid | |
| EGFRm+ | 257 | 2 | 10 | 269 |
| EGFRm(-) | 62 | 10 | 1 | 73 |
PMA P150044: FDA Summary of Safety and Effectiveness Data
{18}
| cobas® EGFR Plasma Test v2 | cobas® EGFR Tissue Test v2 | | | Total |
| --- | --- | --- | --- | --- |
| | EGFRm+ | EGFRm(-) | Invalid | |
| Invalid | 2 | 0 | 0 | 2 |
| No Plasma Sample | 32 | 5 | 2 | 39 |
| Total | 353 | 17 | 13 | 383 |
| PPA (95% CI) | 80.6% (75.9%, 84.5%) | | | |
| NPA (95% CI) | 83.3% (83.3%, 95.3%) | | | |
EGFRm+ = EGFR mutation detected (MD); EGFRm(-) = EGFR mutation not detected (NMD)
Table 13. Sensitivity Analysis of Agreement of cobas® EGFR Plasma Test v2 and cobas® EGFR Tissue Test v2 for Overall EGFR Mutation
| Scenario | PPA (95% CI) | NPA (95% CI) |
| --- | --- | --- |
| With Invalid as discordant result | 79.8% (75.1%, 83.8%) | 45.5% (26.9%, 65.3%) |
| With Invalid and no samples as discordant result | 72.6% (67.7%, 77.0%) | 37.0% (21.5%, 55.8%) |
| Verification Bias Adjusted for invalid and no samples with Bootstrap method for confidence intervals | 80.6% (76.4%, 84.9%) | 83.3% (65.2%, 96.3%) |
# B. Accountability of PMA Cohort
A total of 472 patients were screened for AURA2; 383 patients had tissue samples tested by the cobas® EGFR Tissue Test v1 and were eligible for the study. Of those eligible patients, 344 (89.9%) had a plasma sample tested by the cobas® EGFR Plasma Test v2. Of the eligible patients with a valid tissue test (v1) result (n = 373), 233 were T790M+ and 140 were T790M(-) as shown in the Table 14 below. Of 233 patients who were T790M+ (by v1 of the tissue test), 210 were enrolled in the AURA2 study and received at least one dose of osimertinib and were considered as the tissue full analysis set (tFAS). Of the 210 tFAS patients enrolled and treated in the AURA2 study, 207 (98.6%) had a plasma sample and were tested by cobas® EGFR Plasma Test v2. Sample availability and accountability are described further in Figure 2 and Table 14 below.
PMA P150044: FDA Summary of Safety and Effectiveness Data
{19}

Fig. 2. Patient Accountability for Plasma Test from AURA2
Table 14. Plasma Sample Availability from AURA2 by Category
| Category | Total | Plasma Sample n(%) |
| --- | --- | --- |
| Total screened patients | 472 | 382 (80.9%) |
| Tested by cobas® Tissue Test v1 (Eligible) | 383 | 344 (89.8%) |
| T790M+ | 233 | 225 (96.6%) |
| Enrolled and Treated with Tagrisso™ | 210 | 207 (98.6%) |
PMA P150044: FDA Summary of Safety and Effectiveness Data
{20}
| Category | Total | Plasma Sample n(%) |
| --- | --- | --- |
| T790M(-) | 140 | 111 (79.3%) |
| Invalid | 10 | 8 (80.0%) |
| Not Tested by cobas® EGFR Tissue Test v1 | 89 | 38 (42.7%) |
| Pathology Unsuccessful* | 34 | 25 (73.5%) |
| Without Tumor Tissue | 55 | 13 (23.6%) |
* Tissue samples from 4 patients with tissue pathology that were ‘unsuccessful’ were mistakenly tested by the cobas® EGFR Tissue Test v1. The tissue results were treated as ‘not tested’ by the cobas® EGFR Tissue Test v1.
## C. Study Population Demographics and Baseline Parameters
The average age for the 383 eligible patients is 62.6 years, ranging from 33 to 88 years old, approximately 68% were female, 96.6% were not Hispanic or Latino, 65% were Asian, and 73% never smoked. Approximately 78% of patients were classified as metastatic/stage IV cancer and approximately 95% had a histologic diagnosis of adenocarcinoma (NOS). The demographics and baseline clinical characteristics are shown in Table 15 below.
Table 15. Demographics and Baseline Clinical Characteristics of All Screened Patients by Plasma Specimen Availability
| Characteristics | Overall (n=383) | Available for Plasma Test | | | |
| --- | --- | --- | --- | --- | --- |
| | | Yes (n=344) | No (n=39) | p-value | |
| Age (Years) | | | | | |
| | Mean ± SD | 62.6±10.7 | 62.6±10.8 | 62.1±9.9 | 0.7852 |
| | Median (Range) | 63.0 (33, 88) | 64.0 (33, 88) | 63.0 (37, 85) | |
| Sex, n (%) | | | | | |
| | Male | 122 (31.9%) | 111 (32.3%) | 11 (28.2%) | 0.6058 |
| | Female | 261 (68.1%) | 233 (67.7%) | 28 (71.8%) | |
| Race, n (%) | | | | | |
| | White | 125 (32.6%) | 103 (29.9%) | 22 (56.4%) | 0.0002 |
| | Asian | 248 (64.8%) | 234 (68.0%) | 14 (35.9%) | |
| | Other | 10 (2.6%) | 7 (2.0%) | 3 (7.7%) | |
| Ethnic Origin, n (%) | | | | | |
| | Hispanic Or Latino | 11 (2.9%) | 10 (2.9%) | 1 (2.6%) | 0.0001 |
| | Not Hispanic Or Latino | 370 (96.6%) | 334 (97.1%) | 36 (92.3%) | |
| | Missing | 2 (0.5%) | 0 (0.0%) | 2 (5.1%) | |
| Region, n (%) | | | | | |
| | Asia | 215 (56.1%) | 207 (60.2%) | 8 (20.5%) | <0.0001 |
| | Europe | 67 (17.5%) | 60 (17.4%) | 7 (17.9%) | |
| | North America | 101 (26.4%) | 77 (22.4%) | 24 (61.5%) | |
| Smoking Status, n (%) | | | | | |
| | Never | 279 (72.8%) | 246 (71.5%) | 33 (84.6%) | 0.0812 |
| | Current/Former | 104 (27.2%) | 98 (28.5%) | 6 (15.4%) | |
| AJCC Disease Stage, n (%) | | | | | |
| | I/II/III | 76 (19.8%) | 70 (20.3%) | 6 (15.4%) | <0.0001 |
| | IV | 298 (77.8%) | 270 (78.5%) | 28 (71.8%) | |
PMA P150044: FDA Summary of Safety and Effectiveness Data
{21}
| Characteristics | Overall (n=383) | Available for Plasma Test | | | |
| --- | --- | --- | --- | --- | --- |
| | | Yes (n=344) | No (n=39) | p-value | |
| | Missing | 9 (2.3%) | 4 (1.2%) | 5 (12.8%) | |
| Overall Disease Classification | | | | | |
| | Primary | 216 (56.4%) | 187 (54.4%) | 29 (74.4%) | 0.0568 |
| | Metastatic | 166 (43.3%) | 156 (45.3%) | 10 (25.6%) | |
| | Tumor (not specified) | 1 (0.3%) | 1 (0.3%) | 0 (0.0%) | |
| Histology Type, n (%) | | | | | |
| | Adenocarcinoma (NOS) | 363 (94.8%) | 328 (95.3%) | 35 (89.7%) | 0.0001 |
| | Other | 18 (4.7%) | 16 (4.7%) | 2 (5.1%) | |
| | Missing | 2 (0.5%) | 0 (0.0%) | 2 (5.1%) | |
| cobas® EGFR Tissue Test v1 T790M Result | | | | | |
| | T790M+ | 233 (60.8%) | 225 (65.4%) | 8 (20.5%) | <0.0001 |
| | T790M(-) | 140 (36.6%) | 111 (32.3%) | 29 (74.4%) | |
| | Invalid | 10 (2.6%) | 8 (2.3%) | 2 (5.1%) | |
Note: In Race, category 'Other' combines Black or African American, Native Hawaiian or Other Pacific Islander and Other.
All the variables other than age and sex are significantly different for the patients with and without plasma samples while smoking status and overall disease classification were considered of borderline significance. A multiple logistic regression model with stepwise selection was used to compare the patient demographics and clinical characteristics within each T790M status [T790M+, T790M(-), and T790M status unknown by the cobas® EGFR Tissue Test v1]. Study participants' region (i.e., location), smoking status, and tissue T790M status, were related to the availability of plasma samples. The impact of variables (region of study participants and smoking) were further investigated on the overall agreement (concordant or discordant) as well as the 3-group agreement (positive concordant, negative concordant and discordant) of T790M status between tissue and plasma for those with valid results from both tests by a propensity score method. The equal distribution across the T790M subgroups indicates that those variables did not affect the agreement between the plasma and tissue tests.
To further ensure that the impact of the identified covariates to the agreement estimates were minor, the distributions of the covariates conditioned on each assay agreement status were assessed. Chi-square tests reported for each subgroup indicated that neither the smoking status nor the region of the study participants has significant connection with the cobas® EGFR Plasma Test v2 results provided the cobas® EGFR Tissue Test v1 result (smoking status p-values = 0.9825 - 0.9868 and region of the study participants p-values = 0.9795 - 09803).
## D. Safety and Effectiveness Results
### 1. Safety Results
The safety with respect to treatment with osimertinib was addressed in the original drug approval and was summarized as part of the original approval of the
PMA P150044: FDA Summary of Safety and Effectiveness Data
{22}
cobas® EGFR Tissue Test v2. See Summary of Safety and Effectiveness Data for P120019/S007. No adverse events were reported in connection with the studies used to support this PMA as the studies were performed retrospectively using banked samples.
## 2. Effectiveness Results
The objective response rate (ORR) from the AURA2 study based on enrollment with a T790M+ tissue result using the cobas® EGFR Tissue Test v1 was 61.0% (128/210; 95% CI: 57.0%, 70.8%) in the tFAS population. There were 198 patients in the tFAS who also had measurable disease at baseline according to blinded independent central review (BICR) and these were considered as the cobas® EGFR Tissue Test v1 tissue. Evaluable for Response Analysis Set (tERAS). The ORR was 64.1% (127/198; 95% CI: 54.0%, 67.6%) in the tERAS.
Only three (1.4%) of the 210 tFAS patients did not have a plasma sample. Of the 207 tFAS patients with a plasma sample, 117 were T790M+ (i.e., T790M+ by both the cobas® EGFR Tissue Test v1 and cobas® EGFR Plasma Test v2) and 89 were T790M(-) (i.e., T790M+ by the cobas® EGFR Tissue Test v1 and T790M(-) by the cobas® EGFR Plasma Test v2), and one had an invalid result by the cobas® EGFR Plasma Test v2.
Table 16 summarizes the number of patients with a T790M+ plasma positive result from the original AURA2 tFAS population.
Table 16. Patient Enrollment Status by cobas® EGFR Plasma Test v2 and cobas® EGFR Tissue Test v1
| | Enrollment Based on cobas® EGFR Tissue Test v1 Result | | | |
| --- | --- | --- | --- | --- |
| | | Enrolled (tFAS) | Not Enrolled T790M(-) | Total |
| cobas® EGFR Plasma Test v2 | T790M+ | 117 | 22 | 139 |
| | T790M(-) | 89 | 89 | 178 |
| | Invalid | 1 | 0 | 1 |
| | No Plasma Sample | 3 | 29 | 32 |
| | Total | 210 | 140 | 350 |
To estimate the drug efficacy for patients who were T790M+ by the cobas® EGFR Plasma Test v2 was calculated by assuming the ORR to be 48.8% (80% of ORR in tFAS) to 0% (no drug efficacy) among those patients who were T790M+ by the cobas® EGFR Plasma Test v2 and T790M(-) by the cobas® EGFR Tissue Test v1. The estimates are shown in Table 17, below. To compare efficacy estimates, ORR data (based on 3 months after the enrollment) for patients who were T790M(-) in tissue from a Phase I study group (cohort 3) from the AURA study was provided. In this study group, both EGFR T790M+ and T790M(-) NSCLC patients were enrolled based on the cobas® EGFR Tissue Test v1 and received the 80mg dose of TAGRISSO™ (osimertinib). Based on the 20.7% ORR observed in the cohort 3 patients who were T790M(-) by tissue, the estimated
PMA P150044: FDA Summary of Safety and Effectiveness Data
Page 23
{23}
ORR for those patients who were T790M+ by the cobas® EGFR Plasma Test v2 was 54.6%. These results indicated that the drug efficacy was robust in patients who were T790M+ by the cobas® EGFR Plasma Test v2.
Table 17. ORR (BICR) for T790M+ by the cobas® EGFR Plasma Test v2
| Patients Not in FAS
[cobas® EGFR Tissue Test v1
T790M(-) and cobas® EGFR Plasma
Test v2 T790M+] Assumed Response Rate (%) | All Patients Plasma
T790M+
ORR (95 % CI) |
| --- | --- |
| 48.8% | 59.0% (53.3%, 64.7%) |
| 36.6% | 57.1% (51.2%, 62.9%) |
| 30.5% | 56.1% (50.2%, 62.1%) |
| 18.3% | 54.2% (48.0%, 60.4%) |
| 6.1% | 52.3% (45.7%, 58.8%) |
| 0% | 51.3% (44.5%, 58.1%) |
| 20.7% (AURA cohort 3) | 54.6% (48.1%, 61.1%) |
A total of 111 tERAS patients were T790M+ by the cobas® EGFR Plasma Test v2. The ORR for this subset was 64.9% (95% CI: 52.1%, 70.4%), which is similar to the 64.1% observed ORR in the tERAS population. The ORRs for patients who were T790M(-) by the cobas® EGFR Plasma Test v2 and T790M+ by the cobas® EGFR Tissue Test v1 were also similar to the observed ORRs in FAS and ERAS population. These data are summarized in Table 18, below.
Table 18. ORR Rate by Plasma Result Status among Enrolled Patients with Confirmed Responses
| Analysis Population | Plasma Test Result | N Enrolled | # of Patients with ORR | ORR (95% CI) |
| --- | --- | --- | --- | --- |
| tFAS | T790M+ | 117 | 72 | 61.5% (55.2%, 73.7%) |
| | T790M(-) | 89 | 53 | 59.6% (51.3%, 73.0%) |
| | Overall | 210 | 128 | 61.0% (57.0%, 70.8%) |
| tERAS | T790M+ | 111 | 72 | 64.9% (52.1%, 70.4%) |
| | T790M(-) | 83 | 52 | 62.7% (48.6%, 69.8%) |
| | Overall | 198 | 127 | 64.1% (54.0%, 67.6%) |
Drug efficacy for T790M+ patients based on the cobas® EGFR Plasma Test v2 was established by bridging the efficacy results from the cobas® EGFR Tissue Test v1. Some of the T790M(-) patients by the cobas® EGFR Tissue Test v1 were T790M+ by the cobas® EGFR Plasma Test v2. These patients were not enrolled/dosed (by design) in the clinical trial and thus no drug efficacy data is available for these patients. To estimate the drug efficacy for all patients T790M+ by the cobas® EGFR Plasma Test v2, the ORR for this subset of patients was assumed to be: 1) a proportion (c = 0.8, 0.6, 0.5, 0.3, 0.1, 0) of the drug efficacy (ORR = 61%) in the tFAS population, and 2) equal to the ORR reported for the tissue T790M(-) patients with unknown plasma results from cohort 3 of the
PMA P150044: FDA Summary of Safety and Effectiveness Data
Page 24
{24}
AURA study, i.e., 20.7% with 95% CI of 8.0% to 39.7%. Furthermore, the different proportions of plasma testing between the tFAS population and the T790M(-) (by the cobas® EGFR Tissue Test v1) screen fail population was also adjusted in drug efficacy for T790M+ by the cobas® EGFR Plasma Test v2. The weighted average was used for drug efficacy estimation in this scenario.
The 210 patients in tFAS population along with the estimated 28 patients who would be T790M+ by cobas® EGFR Plasma Test v2 and T790M(-) by cobas® EGFR Tissue Test v1 represented the screening population of patients with T790M mutation by either the cobas® EGFR Tissue Test v1 or cobas® EGFR Plasma Test v2. A sensitivity analysis based on a different assumed drug efficacy on these 28 patients was performed. An assumed number of responders and response rates among the 28 patients who would be T790M+ by cobas® EGFR Plasma Test v2 and T790M(-) by cobas® EGFR Tissue Test v1 ranged from 16 (57.1%) to 0 (0.0%). The ORRs for patients who are T790M+ by either the cobas® EGFR Plasma Test v2 or cobas® EGFR Tissue Test v1 (Triage FAS) are summarized in Table 20 below.
A total of 209 patients would be enrolled as tFAS if the enrollment were based on cobas® EGFR Tissue Test v2. Of those, 128 were categorized as a responder (ORR = 61.2%). The assumed number of responders and response rates [17 (56.7%) to 0 (0.0%)] amongst the 30 patients with T790M mutation by the cobas® EGFR Plasma Test v2 and who were T790M(-) or invalid by the cobas® EGFR Tissue Test v2 are summarized in Table 19.
Table 19. Comparison of ORRs for T790M+ Patients Based on Different Versions of Tissue Test for Triage Algorithm (FAS Population)
| T790M+ Based on | Enrolled tFAS (Responders, ORR) | Not Enrolled (projected T790M- by tissue and T790M+ by plasma) | | Total Triage FAS | |
| --- | --- | --- | --- | --- | --- |
| | | N | Assumed ORR | N | ORR |
| cobas® EGFR Tissue Test v2 or Plasma Test v2 | 209 (128, 61.2%) | 30 | 56.7% to 0% | 239 | 60.7% (54.1%, 67.3%) to 53.6% (46.8%, 60.4%) |
| cobas® EGFR Tissue Test v1 or Plasma Test v2 | 210 (128, 61%) | 28 | 57.1% to 0% | 238 | 60.5% (53.9%, 67.1%) to 53.8% (47.1%, 60.5%) |
As shown in Table 19, the ORR results for the Triage FAS patients by the cobas® EGFR Tissue Test v2 or the cobas® EGFR Plasma Test v2 (T790M+ based on the cobas® EGFR Tissue Test v2 or cobas® EGFR Plasma Test v2) are similar to those for Triage FAS patients by cobas® EGFR Tissue Test v1 or cobas® EGFR Plasma Test v2 (T790M+ based on cobas® EGFR Tissue Test v1 or cobas® EGFR Plasma Test v2). This demonstrates that in patients with a known T790M substitution mutation in a FFPET specimen, which also had a T790M positive
PMA P150044: FDA Summary of Safety and Effectiveness Data
Page 25
{25}
plasma result had clinical benefit similar to those treated with TAGRISSO™ (osimertinib) based on a tissue result.
3. Pediatric Extrapolation
In this premarket application, existing clinical data was not leveraged to support approval of a pediatric patient population.
E. Financial Disclosure
The Financial Disclosure by Clinical Investigators regulation (21 CFR 54) requires applicants who submit a marketing application to include certain information concerning the compensation to, and financial interests and arrangement of, any clinical investigator conducting clinical studies covered by the regulation. The pivotal clinical study included three investigators. None of the clinical investigators had disclosable financial interests/arrangements as defined in sections 54.2(a), (b), (c), and (f). The information provided does not raise any questions about the reliability of the data.
XI. SUMMARY OF SUPPLEMENTAL CLINICAL INFORMATION
Additional data from paired tissue and EDTA plasma results from an independent cohort were provided to support specimen handling and storage conditions based on similar protocols. See Summary of Safety and Effectiveness Data for P150047.
Very preliminary and limited data were also provided from a small cohort of patients from the AURA study and from an ongoing independent external study demonstrating the efficacy of TAGRISSO™ (osimertinib) in EGFR T790M mutation positive patients selected using the cobas® EGFR Plasma Test v2. In this study patients may also be enrolled based on a T790M mutation positive result in a FFPET specimen; however for those enrolled based on a plasma specimen result, the presence of the mutation in tissue may be unknown.
XII. PANEL MEETING RECOMMENDATION AND FDA'S POST-PANEL ACTION
In accordance with the provisions of section 515(c)(3) of the act as amended by the Safe Medical Devices Act of 1990, this PMA was not referred to the Molecular and Clinical Genetics Panel of Medical Devices, an FDA advisory committee, for review and recommendation.
XIII. CONCLUSIONS DRAWN FROM PRECLINICAL AND CLINICAL STUDIES
A. Effectiveness Conclusions
The clinical benefit of the cobas® EGFR Plasma Test v2 in the detection of the T790M substitution mutation in cfDNA extracted from plasma, where the mutation was known to be present in NSCLC tissue specimens, was demonstrated in
PMA P150044: FDA Summary of Safety and Effectiveness Data
Page 26
{26}
retrospective analyses of patients enrolled in the Phase II AURA2 study as summarized in Tables 16 - 19. Analytical performance studies with the cobas® EGFR Plasma Test v2, when used according to the Instructions for Use, demonstrate the ability to detect the T790M substitution mutation with an analytical sensitivity of 100 cp/mL of cfDNA extracted from NSCLC patient plasma.
Very preliminary and limited data were provided demonstrating the efficacy of TAGRISSO™ (osimertinib) in EGFR T790M mutation positive patients selected using the cobas® EGFR Plasma Test v2 but in whom the patient's tumor tissue may be T790M mutation negative or unknown. The data provided however was not sufficient to draw conclusions regarding the effectiveness of the cobas® EGFR Plasma Test v2 in the plasma-positive,
The safety and effectiveness of TAGRISSO™ (osimertinib) has not been established in patients whose plasma specimens have Ex. 19del, L858R, G719X, Ex. 20ins, S768I or L861Q mutations, which are also detected by the cobas® EGFR Plasma Test v2.
## B. Safety Conclusions
As a diagnostic test, the cobas® EGFR Plasma Test v2 involves testing on plasma isolated from EDTA anti-coagulated peripheral blood collected from NSCLC patients. The risks of the cobas® EGFR Plasma Test are associated with the potential mismanagement of patients resulting from false results of the test. Failure of the device to perform as expected or failure to correctly interpret test results may lead to incorrect EGFR test results, and consequently improper patient management decisions in NSCLC treatment. Since a patient with a negative result (including a false negative result) from the cobas® EGFR Plasma Test v2 will be reflexed to having their EGFR status determined from an FFPET specimen, the risks of the cobas® EGFR Plasma Test v2 are largely associated with a false positive result in a patient, who may then undergo treatment with TAGRISSO™ (osimertinib) with inappropriate expectation of therapeutic benefit and experience side effects.
The patient population included in the study used to support the safety and effectiveness of the cobas® EGFR Plasma Test v2 for selecting patients for treatment with TAGRISSO™ (osimertinib) was originally selected using FFPET specimens. Clinical data for the EGFR T790M plasma-positive, tissue-negative (or unknown) population is limited, and therefore additional clinical data is needed to establish the clinical performance of the test in this population (see Section XIV. below). Given the lack of data, testing with the cobas® EGFR Plasma Test is considered most appropriate for patients from whom a tumor biopsy cannot be obtained
## C. Benefit-Risk Conclusions
The probable benefits of the device are based on data collected in the AURA2 study, which were used to support PMA approval as described in Tables 16 - 19 above. The
PMA P150044: FDA Summary of Safety and Effectiveness Data
Page 27
{27}
clinical benefit of the cobas® EGFR Plasma Test v2 was demonstrated in a retrospective analysis of efficacy and safety data obtained from a global Phase II, open-label, single-arm study, in patients with locally advanced or metastatic NSCLC (Stage IIIB-IV) and a positive T790M result in tumor specimens, who had progressed following prior therapy with an approved EGFR TKI agent. TAGRISSO™ (osimertinib) demonstrated an ORR of 61.5% (72/117; 95% CI: 55.2%, 73.7%) in all patients with positive T790M results in tumor specimens and 64.9% (72/111; 95% CI: 52.1%, 70.4%) in a subset of these patients whose plasma specimens also demonstrated a T790M positive result. This demonstrated that in patients with a known T790M substitution mutation in a FFPET specimen in which also had a T790M positive plasma result had clinical benefit similar to those treated with TAGRISSO™ (osimertinib) based on a tissue result.
The risks associated with use of the cobas® EGFR Plasma Test v2 are mitigated in the description of the performance observed in the Instructions for Use which also supports the accompanying update to the TAGRISSO™ (osimertinib) prescribing information which was reviewed by CDER. In particular, the efficacy of TAGRISSO™ (osimertinib) has not been established in the EGFR T790M plasma-positive, tissue-negative or unknown population and for which clinical data for T790M plasma-positive patients are limited. Prospective clinical data for patients with EGFR positive metastatic NSCLC who are T790M plasma-positive (but tissue-negative or unknown) to further characterize outcomes with TAGRISSO™ (osimertinib) therapy from real world patients who have been selected for treatment on the basis of an EGFR T790M mutation positive result from plasma (cfDNA) using the cobas® EGFR Mutation Test v2 is planned as a condition of approval.
The risks of the cobas® EGFR Plasma Test v2 are associated with the potential mismanagement of patients resulting from erroneous test results. The device is a key part of diagnostic evaluation for NSCLC in decisions regarding treatment with TAGRISSO™ (osimertinib). There is currently no FDA approved test for the selection of candidate metastatic NSCLC patients for treatment with TAGRISSO™ (osimertinib) based on identification of a T790M substitution mutation from patient's plasma specimens.
In conclusion, given the available information above, the data supports the use of the cobas® EGFR Plasma Test v2 as an aid in selecting NSCLC patients for TAGRISSO™ (osimertinib) treatment based on a cobas® EGFR Plasma Test v2 "Mutation Detected" result for EGFR T790M mutations when detected in plasma, that the probable benefits outweigh the probable risks. However because the outcome of patients with a plasma T790M+ result, but who are T790M(-) or unknown by tissue, who are treated with TAGRISSO™ (osimertinib) is uncertain, testing using the cobas® EGFR Plasma Test v2 is most appropriate for patients from whom a tumor biopsy cannot be obtained.
Patient Perspective Information
PMA P150044: FDA Summary of Safety and Effectiveness Data
Page 28
{28}
This submission did not include specific information on patient perspectives for this device.
## D. Overall Conclusions
The data in this application support the reasonable assurance of safety and effectiveness of this device when used in accordance with the indications for use. Data from the AURA2 Phase II clinical study support the utility of the cobas® EGFR Plasma Test v2 as an aid in selecting patients with advanced NSCLC and T790M mutation for whom TAGRISSO™ (osimertinib), an EGFR TKI, is indicated. Osimertinib demonstrated clinical benefit in terms of ORR that appears to be robust and of a magnitude to reasonably predict clinical benefit for osimertinib in patients with T790M mutation identified with the cobas® EGFR Plasma Test v2.
## XIV. CDRH DECISION
CDRH issued an approval order on September 28, 2016. The final conditions of approval cited in the approval order are described below.
1. The data provided in the PMA application demonstrates that the cobas® EGFR Mutation Test v2 using plasma can identify patients who will benefit from osimertinib therapy; however the trial was conducted in patients previously demonstrated to have the EGFR T790M mutation in tumor samples. Limited data was provided for patients who are EGFR T790M mutation positive in plasma (cfDNA) but whose FFPET specimens were EGFR T790M mutation negative or unknown. Therefore, as part of the agreed upon post-market commitment (PMC #3119-1) to NDA 208065, additional data regarding overall response rate is needed for from NSCLC patients treated with osimertinib from "real-world" cohorts of patients who have been selected for treatment on the basis of an EGFR T790M mutation positive result from plasma using the cobas® EGFR Mutation Test v2. Please also provide duration of response data and, if available, the tissue EGFR T790M status on these patients.
2. Limited specimen handling and specimen storage stability information was included in the PMA application. Additional testing and data are required to establish robustness in regard to the pre-analytical specimen handling instructions to users to ensure consistent performance of the cobas® EGFR Plasma Test v2 with clinical plasma specimens.
The applicant’s manufacturing facilities have been inspected and found to be in compliance with the device Quality System (QS) regulation (21 CFR 820).
## XV. APPROVAL SPECIFICATIONS
Directions for use: See device labeling.
PMA P150044: FDA Summary of Safety and Effectiveness Data
Page 29
{29}
Hazards to Health from Use of the Device: See Indications, Warnings, Precautions, and Limitations in the device labeling. Refer to the drug label for TAGRISSO™ (osimertinib) for additional information related to use of the drug.
Post-approval Requirements and Restrictions: See approval order.
PMA P150044: FDA Summary of Safety and Effectiveness Data
Page 30