The CP Stent and Mounted CP Stent are indicated for use in the treatment of native and/or recurrent coarctation of the aorta involving a compliant aortic isthmus or first segment of the descending aorta where there is adequate size and patency of at least one femoral artery and the balloon angioplasty is contraindicated or predicted to be ineffective. The Covered CP Stent and Covered Mounted CP Stent are indicated for use in the treatment of native and/or recurrent coarctation of the aorta involving the aortic isthmus or first segment of the descending aorta where there is adequate size and patency of at least one femoral artery associated with one or more of the following: - acute or chronic aortic wall injury - nearly atretic descending aorta of 3 mm or less in diameter - a non-compliant stenotic aortic segment found on pre-stent balloon dilation - a genetic or congenital syndrome associated with aortic wall weakening or ascending aortic aneurysm
Device Story
The Cheatham Platinum (CP) Stent System consists of a balloon-expandable platinum/iridium wire stent (bare or ePTFE-covered) mounted on a triaxial BIB (balloon-in-balloon) catheter. Used in cath labs by interventional cardiologists to treat aortic coarctation. The BIB catheter allows incremental balloon inflation for precise stent expansion. The stent provides permanent radial support to the aorta; the covered version creates a barrier to treat aortic wall injuries or aneurysms. The device is tracked over a guidewire under fluoroscopic guidance. By relieving the aortic obstruction, the device reduces systolic blood pressure gradients and improves clinical status. Benefits include minimally invasive treatment compared to surgery, reduced hospital stay, and effective repair of aortic wall defects.
Clinical Evidence
Evidence based on two prospective, multi-center, single-arm clinical studies (COAST and COAST II) and Continued Access Protocols (CAP). COAST (n=112) evaluated bare stents; COAST II (n=82) evaluated covered stents. Primary safety endpoints (serious/somewhat serious adverse events <18% at 30 days) were met. Effectiveness endpoints included reduction in arm-leg systolic blood pressure gradient and improvement in Severity of Illness Scale. Results showed sustained reduction in blood pressure gradients and effective repair of aortic wall injuries. Stent fractures were observed in bare stents (12% at 24 months, 36% at 60 months) but without loss of structural integrity or clinical sequelae.
Indicated for patients with native or recurrent aortic coarctation involving the aortic isthmus or first segment of the descending aorta. Requires adequate femoral artery size/patency. Covered stents indicated for patients with aortic wall injury, nearly atretic descending aorta (≤3mm), non-compliant stenotic segments, or genetic syndromes associated with aortic wall weakening. Contraindicated in patients too small for safe delivery, unfavorable anatomy, curved vasculature, systemic artery occlusion, active infection/endocarditis, allergy to antiplatelet agents/heparin, or pregnancy.
Regulatory Classification
Identification
Coarctation of the Aorta
Reference Devices
Melody Valve (PMA)
Related Devices
K160889 — BIB Stent Placement Catheter · NuMED, Inc. · May 20, 2016
P240003 — Minima Stent System · Renata Medical, Inc. · Aug 28, 2024
K213915 — BIB Stent Placement Catheter · NuMED, Inc. · Jan 12, 2022
P160021 — GORE VIABAHN VBX BALLOON EXPANDABLE ENDOPROSTHESIS · W. L. Gore & Associates, Inc. · Jan 27, 2017
DEN080015 — NUCLEUS-X PTV CATHETER · NuMED, Inc. · Jun 11, 2012
Submission Summary (Full Text)
{0}
SUMMARY OF SAFETY AND EFFECTIVENESS DATA (SSED)
I. GENERAL INFORMATION
Device Generic Name: Aortic Stent
Device Trade Name: Cheatham Platinum (CP) Stent System
(CPStent – Model 425
Covered CP Stent – Model 427
Mounted CP Stent – Model 426
Covered Mounted CP Stent – Model 428
BiB Stent Placement Catheter – Model 420/420.1)
Device Procode: PNF
Applicant’s Name and Address: NuMED, Inc.
2880 Main Street
Hopkinton, NY 12965
Date(s) of Panel Recommendation: None
Premarket Approval Application (PMA) Number: P150028
Date of FDA Notice of Approval: March 25, 2016
II. INDICATIONS FOR USE
The CP Stent and Mounted CP Stent are indicated for use in the treatment of native and/or recurrent coarctation of the aorta involving a compliant aortic isthmus or first segment of the descending aorta where there is adequate size and patency of at least one femoral artery and the balloon angioplasty is contraindicated or predicted to be ineffective.
The Covered CP Stent and Covered Mounted CP Stent are indicated for use in the treatment of native and/or recurrent coarctation of the aorta involving the aortic isthmus or first segment of the descending aorta where there is adequate size and patency of at least one femoral artery associated with one or more of the following:
- acute or chronic aortic wall injury
- nearly atretic descending aorta of 3 mm or less in diameter
- a non-compliant stenotic aortic segment found on pre-stent balloon dilation
- a genetic or congenital syndrome associated with aortic wall weakening or ascending aortic aneurysm
PMA P150028: FDA Summary of Safety and Effectiveness Data
{1}
PMA P150028: FDA Summary of Safety and Effectiveness Data
Page 2
# III. CONTRAINDICATIONS
1. Patients too small to allow safe delivery of the stent without compromise to the systemic artery used for delivery;
2. Unfavorable aortic anatomy that does not dilate with high pressure balloon angioplasty;
3. Curved vasculature;
4. Occlusion or obstruction of systemic artery precluding delivery of the stent;
5. Clinical or biological signs of infection;
6. Active endocarditis;
7. Known allergy to aspirin, other antiplatelet agents, or heparin;
8. Pregnancy
# IV. WARNINGS AND PRECAUTIONS
The warnings and precautions can be found in the CP Stent, Mounted CP Stent, Covered CP Stent, Covered Mounted CP Stent, and BIB Stent Placement Catheter labeling.
# V. DEVICE DESCRIPTION
The NuMED Cheatham Platinum (CP) Stent System includes a bare or covered CP Stent and a delivery catheter (BIB). Each stent is balloon expandable and intended for permanent implant. The device is available in the following configurations: The bare stent un-mounted (Cheatham Platinum (CP) Stent), bare stent mounted on delivery catheter (Mounted Cheatham Platinum (CP) Stent), covered stent un-mounted (Covered Cheatham Platinum (CP) Stent), and covered stent mounted on delivery catheter (Covered Mounted Cheatham Platinum (CP) Stent). Each configuration is available in the sizes listed in Table 1.
Table 1. Device Size Matrix
| Configuration (number of zigs) | Platinum Wire Diameter (inch) | Diameter (mm) | Labeled Stent Length (mm) | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | 16 | 22 | 28 | 34 | 39 | 45 |
| 8 | 0.013 | 12 | ✓ | ✓ | ✓ | ✓ | ✓ | ✓ |
| | | 14 | ✓ | ✓ | ✓ | ✓ | ✓ | ✓ |
| | | 15 | ✓ | ✓ | ✓ | ✓ | ✓ | ✓ |
| | | 16 | ✓ | ✓ | ✓ | ✓ | ✓ | ✓ |
| | | 18 | ✓ | ✓ | ✓ | ✓ | ✓ | ✓ |
| | | 20 | ✓ | ✓ | ✓ | ✓ | ✓ | ✓ |
| | | 22 | ✓ | ✓ | ✓ | ✓ | ✓ | ✓ |
| | | 24 | ✓ | ✓ | ✓ | ✓ | ✓ | ✓ |
The CP stent (Figure 1) is composed of a platinum/iridium wire that is arranged in a "zig" pattern, laser welded at each joint and over brazed with 24K gold. Each row of zigs is laser-welded to the next identical row in a repeating manner to accommodate the desired length of the stent. The number of rows determines the unexpanded length of the stent. The CP Stent is balloon expandable and intended for permanent implant.
{2}

Figure 1. Expanded Bare NuMED CP Stent
The Mounted CP Stent (Figure 2) is the CP Stent mounted on the NuMED BIB balloon expandable catheter.

Figure 2. Left: Mounted NuMED CP stent crimped Right: Mounted NuMED CP stent expanded.

The Covered CP Stent (Figure 3) is comprised of the CP Stent that is covered with an expandable sleeve of ePTFE. The sleeve covers the entire length of the stent. The sleeve is attached to each end of the stent with a cyanoacrylate adhesive on a physically etched section of the sleeve. Upon balloon expansion of the stent, the covering remains intact and expanded with the stent to create a barrier around the stent.
PMA P150028: FDA Summary of Safety and Effectiveness Data
{3}

Figure 3. Left: Expanded Covered NuMED CP stents Right: Hand crimped Covered NuMED CP stents
The Covered Mounted CP Stent (Figure 4) is the Covered CP Stent mounted on the NuMED BIB balloon expandable catheter.

Figure 4. Left: Covered Mounted NuMED CP stent crimped Right: Covered Mounted NuMED CP stent expanded.

The catheter is triaxial in construction with two lumens being used to inflate the balloons while one lumen is used for tracking over a guidewire. The double balloon catheter allows for incremental inflation for the purpose of dilating a stent. Radiopaque platinum marker bands are located under the balloon shoulders for placement using fluoroscopy. The catheter is composed of PES2, Pebax, Platinum/Iridium, and PES2 with colorants. The delivery catheter is compatible with 0.035" guidewires and 8-11 Fr introducers.
# VI. ALTERNATIVE PRACTICES AND PROCEDURES
There are several alternatives for the correction of coarctation of the aorta, including: balloon catheter dilatation and surgical intervention. Each alternative has its own advantages and disadvantages. A patient should fully discuss these alternatives with his/her physician to select the method that best meets expectations and lifestyle.
PMA P150028: FDA Summary of Safety and Effectiveness Data
{4}
# VII. MARKETING HISTORY
The CP andCovered CP Stent Systems and BiB Stent Placement Catheter are currently marketed in the following countries:
Table 2. Device Marketing Locations
| Product | Countries |
| --- | --- |
| CP Stent | Algeria, Argentina, Australia, Bahamas, Brazil, Brunei, Canada, Chile, Columbia, Costa Rica, Cuba, Dominican Republic, Ecuador, El Salvador, European Union, Guatemala, Honduras, Hong Kong, India, Israel, Jordan, Kenya, Kuwait, Malaysia, Mauritius Island, Mexico, Mongolia, New Zealand, Norway, Pakistan, Peru, Russia, Saudi Arabia, South Africa, Sultanate of Oman, Switzerland, Trinidad & Tobago, Turkey, Uganda, Uruguay, Vietnam. |
| Covered CP Stent | Algeria, Argentina, Australia, Bahamas, Brazil, Brunei, Canada, Chile, Columbia, Costa Rica, Cuba, Dominican Republic, Ecuador, El Salvador, European Union, Guatemala, Honduras, Hong Kong, India, Israel, Jordan, Kenya, Kuwait, Malaysia, Mauritius Island, Mexico, Mongolia, New Zealand, Norway, Pakistan, Peru, Russia, Saudi Arabia, South Africa, Sultanate of Oman, Switzerland, Trinidad & Tobago, Turkey, Uganda, Uruguay, Vietnam. |
| Mounted CP Stent | Algeria, Argentina, Australia, Bahamas, Brazil, Brunei, Canada, Chile, Columbia, Costa Rica, Cuba, Dominican Republic, Ecuador, El Salvador, European Union, Guatemala, Honduras, Hong Kong, India, Israel, Jordan, Kenya, Kuwait, Malaysia, Mauritius Island, Mongolia, Norway, Pakistan, Peru, Saudi Arabia, South Africa, Sultanate of Oman, Switzerland, Trinidad & Tobago, Turkey, Uganda, Uruguay, Vietnam. |
| Covered Mounted CP Stent | Algeria, Argentina, Australia, Bahamas, Brazil, Brunei, Canada, Chile, Columbia, Costa Rica, Cuba, Dominican Republic, Ecuador, El Salvador, European Union, Guatemala, Honduras, Hong Kong, India, Israel, Jordan, Kenya, Kuwait, Malaysia, Mauritius Island, Mongolia, Norway, Pakistan, Peru, Saudi Arabia, South Africa, Sultanate of Oman, Switzerland, Trinidad & Tobago, Turkey, Uganda, Uruguay, Vietnam. |
PMA P150028: FDA Summary of Safety and Effectiveness Data
{5}
| BIB Stent | Algeria, Argentina, Australia, Bahamas, Brazil, Brunei, Canada, Chile, Columbia, Costa Rica, Cuba, Dominican Republic, Ecuador, |
| --- | --- |
| Placement | El Salvador, European Union, Guatemala, Honduras, Hong Kong, India, Israel, Jordan, Kenya, Kuwait, Malaysia, Mauritius Island, Mongolia, New Zealand, Norway, Pakistan, Peru, Russia, Saudi Arabia, South Africa, Sultanate of Oman, Switzerland, Taiwan, Trinidad & Tobago, Turkey, Uganda, Uruguay, Vietnam. |
| Catheter | |
The device has not been withdrawn from marketing for any reason related to its safety and effectiveness.
## VIII. POTENTIAL ADVERSE EFFECTS OF THE DEVICE ON HEALTH
Below is a list of the potential adverse effects (e.g., complications) associated with the use of the device.
- Femoral Artery injury, thrombosis or psuedoaneurysm
- Stent Migration
- Stent Stenosis
- Stent Fracture
- Aortic Aneurysm/Pseudoaneurysm
- Aortic Rupture/Tear
- Stent Malposition
- Hematoma
- Sepsis/infection
- Thrombosis/Thromboembolism
- AV fistula formation
- Death
- Transitory arrhythmia
- Endocarditis
- Bleeding
- Cell necrosis at the site of implant
- Cerebrovascular Incident
For the specific adverse events that occurred in the clinical study, please see Section X below.
PMA P150028: FDA Summary of Safety and Effectiveness Data
{6}
PMA P150028: FDA Summary of Safety and Effectiveness Data
Page 7
# IX. SUMMARY OF PRECLINICAL STUDIES
## A. Laboratory Studies
### 1. In vitro Product Testing
Bench testing was performed on the CP Stent and Covered CP Stent as described below. The samples were exposed to 2X Ethylene Oxide sterilization cycle prior to testing. All applicable testing for each stent configuration and the BIB catheter was conducted with accessories representative of clinical use. Testing was conducted according to four corners of the available device unless otherwise noted. A matrix of tests performed and corresponding results are provided in Tables 3 to 5.
Table 3. Summary of in vitro Product Testing for CP Stent and Covered CP Stent
| Test | Purpose/Objective | Test/Reference Articles | Results |
| --- | --- | --- | --- |
| Material Characterization | | | |
| Material Composition | To identify the materials of construction of the stent and delivery system | Test: Two wire bundle samples | All data demonstrated that materials were free from contamination, discoloration and any form of damage that could impact the material. The major elements found were platinum and iridium. |
| Mechanical Properties | To confirm the ultimate tensile strength (UTS), yield strength (YS) and elongation of the raw material | Test: Raw material wire samples | All data demonstrated that the UTS and wire diameter of the raw material met the acceptance criteria, as confirmed in the Certificate of Conformance from the material manufacturer. |
{7}
PMA P150028: FDA Summary of Safety and Effectiveness Data
Page 8
| Stent Structural Performance Evaluation | | | |
| --- | --- | --- | --- |
| Corrosion Resistance | To address the corrosion resistance of the stent including pitting, fretting, crevice and galvanic, as appropriate | Test: 6 units of CP 8 Zig 4.5cm | All data demonstrated that the device will not be susceptible to pitting or corrosion. |
| Stress Analysis (FEA) | To demonstrate the range of stresses at critical locations on the stent during physiological loading | Modeling based on the stent material properties and geometries | All data demonstrated that the Fatigue Analysis and Goodman Diagram predicts the fatigue safety factor to be 1.2 throughout the area of the stent with tensile stress. |
| Stent Fatigue | To determine the fatigue stresses of the inflated stent design when subjected to fluctuating external pressures | Modeling based on the stent material properties and geometries | All data demonstrated that the Fatigue Analysis and Goodman Diagram predicts the fatigue safety factor to be 1.2 throughout the area of the stent with tensile stress. |
| Accelerated Durability Testing | To determine the long term durability of the stent | Test: 16 units of CP Stent 8 Zig 4.5cm | All stents survived the durability testing of 400 million cycles in the fatigue testers that applied cyclic pulsatile stresses, simulating the radial strain of |
{8}
PMA P150028: FDA Summary of Safety and Effectiveness Data
Page 9
| | | | an artery. |
| --- | --- | --- | --- |
| Stent Dimensional And Functional Attributes | | | |
| Dimensional Verification | To ensure that all dimensional specifications do not deviate from the design specifications | Test: 10 units of CP Stent 8 Zig in each of the following sizes: 1.6cm, 2.2cm, 2.8cm, 3.4cm, 3.9cm, 4.5cm | All stents met the acceptance criteria and the data showed no deviation from the design specifications. |
| Percent Stent Area | To determine the contact area of the stent structure, as a percentage of the conceptual solid circumferential area | Test: 1unit of 8 Zig 1.6 cm stent, 1unit of 8 Zig 4.5cm | There is no established criteria for this test, values are calculated and reported. Using ASTM F2081 to define the full cylindrical side surface area for the stents, the percent stent area for CP8Z16 at 12mm is 49% and CP8Z45 at 24mm is 35%. |
| Stent Foreshortening | To demonstrate the decrease in length of the stent between the catheter loaded condition and deployment to the maximum diameter per the IFU, determining the maximum nominal diameter for which the device is designed | Test: 10 units of the CP Stent mounted on BIB 8 Zig 1.6 cm, 10 units of the CP Stent mounted on BIB 8 Zig 4.5cm, 5 units covered CP 8 Zig 4.5cm | There is no established criteria for this test, values are calculated and reported. The average stent foreshortening of CP8Z16 was 36.4% and of CP8Z45 was 34.9%. |
| Stent Recoil | To determine the decrease in diameter of the stent, from the maximum balloon expanded condition per IFU | Test: 10 units of CP 8 Zig 1.6 cm stent, 10 units of CP 8 Zig 4.5cm, 5 random lengths of each Covered CP Stent having a diameter of 12, 14, 15, 16, 18, 20, 22, and 24mm | All stents met the acceptance criteria, namely that the stent recoil did not exceed 3.5%. |
{9}
Table 4. Design Specific Testing for Covered/Covered Mounted CP Stent
| Test | Purpose | Test/Reference Articles | Results |
| --- | --- | --- | --- |
| ePTFE Permeability/Leakage | To determine the physical properties of the covering material | Test: Raw ePTFE material | The physical properties of ePTFE, including porosity, water |
| | | Test: Raw ePTFE material | the physical properties of ePTFE, including porosity, water |
| pH | To determine the surface area of the covering material | Test: Raw pH | the surface area of the covering material |
| pH2 | To determine the surface area of the covering material | Test: Raw pH2 | the surface area of the covering material |
| pH3 | To determine the surface area of the covering material | Test: Raw pH3 | the surface area of the covering material |
PMA P150028: FDA Summary of Safety and Effectiveness Data
{10}
Table 5. BIB Delivery Catheter Compatibility Testing
| Test | Purpose | Test/Reference Articles | Results |
| --- | --- | --- | --- |
| Balloon and CP Stent Burst Pressure | To demonstrate the burst strength of the catheter | Test: 20 units of CP Stent 8 Zig 1.6cm with 12 x 2.5 BIB, 20 units of CP Stent 8 Zig 4.5cm with 12 x 5 BIB, 20 units of CP Stent 8 Zig 4.5cm with 14 x 5 BIB, 20 units of CP Stent 8 Zig 4.5cm with 15 x 5 BIB, 20 units CP Stent 8 Zig 4.5cm with 16 x 5 BIB, 20 units of CP Stent 8 Zig 4.5cm with 18 x 5 BIB, 20 units of CP Stent 8 Zig 4.5cm with 20 x 5 BIB, 20 units of CP Stent 8 Zig 4.5cm with 22 x 5 BIB, 20 units of CP Stent 8 Zig 1.6cm with 24 x 3 BIB, 20 units of CP Stent 8 Zig 4.5cm with 24 x 5 BIB | All data supported that statistically the balloons will not burst at or below the maximum recommended rated burst pressure. |
| Balloon Compliance | To demonstrate the stent ID versus inflation pressure characteristics | Test: 20 units of CP Stent 8 Zig 1.6cm with 12 x 2.5 BIB, 20 units of CP Stent 8 Zig 4.5cm with 12 x 5 BIB, 20 units of CP Stent 8 Zig 4.5cm with 14 x 5 BIB, 20 units of CP Stent 8 Zig 4.5cm with 15 x 5 BIB, 20 units CP Stent 8 Zig 4.5cm with 16 x 5 BIB, 20 units of CP Stent 8 Zig 4.5cm with 18 x 5 BIB, 20 units of CP Stent 8 Zig 4.5cm with 20 x 5 BIB, 20 units of CP | All data met the acceptance criteria that the inside diameter of the stent shall be +/- 10% of the rated balloon diameter at rated pressure. |
PMA P150028: FDA Summary of Safety and Effectiveness Data
{11}
| | | Stent 8 Zig 4.5cm with 22 x 5 BIB, 20 units of CP Stent 8 Zig 1.6cm with 24 x 3 BIB, 20 units of CP Stent 8 Zig 4.5cm with 24 x 5 BIB | |
| --- | --- | --- | --- |
| Balloon Fatigue | To determine the repeatability of successful balloon inflations to the RBP | Test: 30 units of CP Stent 8 Zig 1.6cm with 12x2.5 BIB, 30 units of CP Stent 8 Zig 4.5cm with 12x5cm BIB, 30 units of CP Stent 8 Zig 1.6cm with 24x3 BIB, CP Stent 8 Zig 4.5cm with 24x5 BIB | All catheters passed the acceptance criteria, with no failures including loss of pressure or bust at rated burst pressure. |
| Balloon Inflation/Deflation | To ensure that the catheter inflates and deflated within a specified time | Test: 10 units of CP Stent 8 Zig 1.6cm with 12x2.5BIB, 10 units of CP Stent 8 Zig 4.5 cm with 12x5 BIB, 10 units of CP Stent 8 Zig 1.6cm with 24x3BIB, 10 units of CP Stent 8 Zig 4.5cm with 24x5BIB | All BIBs met the acceptance criteria of a 15second inflation time and 25 second deflation time. |
| Balloon Deflatability | To ensure that the catheter deflates without interference | Test: 10 units of CP Stent 8 Zig 1.6cm with 12x2.5BIB, 10 units of CP Stent 8 Zig 4.5 cm with 12x5 BIB, 10 units of CP Stent 8 Zig 1.6cm with 24x3BIB, 10 units of CP Stent 8 Zig 4.5cm with 24x5BIB | All BIBs met the acceptance criteria with no interference with balloon deflation. |
| Catheter Bond Strength | To demonstrate the pull strength of the following: distal hub to extension, extension to “Y” connector, “Y” connector to shaft, proximal balloon bond, tip to balloon | Test: 10 8F, 10 9Fr | All samples exceeded the minimum pull strength of 8.9 Newtons. |
| Crossing Profile | To measure the crossing profile as the maximum diameter over the length from the proximal end of the mounted stent to | Test: 10 catheters of each balloon diameter were tested with a mounted stent of random length, 3 catheters of each balloon diameter was tested with a covered mounted stent of | All catheters passed through the appropriate Mullins sheath. |
PMA P150028: FDA Summary of Safety and Effectiveness Data
{12}
| | the distal tip of the delivery system | random length | |
| --- | --- | --- | --- |
| CP Stent Securement on BIB Delivery Catheter | To ensure that the stent remains intact and is not dislodged while being passed through the tortuous pathway | Test: 10 units of CP Stent 8 Zig 1.6cm with 12x2.5BIB, 10 units of CP Stent 8 Zig 4.5 cm with 12x5 BIB, 10 units of CP Stent 8 Zig 1.6cm with 24x3BIB, 10 units of CP Stent 8 Zig 4.5cm with 24x5BIB | No stents dislodged while passing through passageway. |
## 2. MRI Compatibility
Nonclinical testing and modeling of this device in magnetic fields of 1.5 and 3.0 Tesla showed that the device is MR Conditional. The Bare CP Stent, Bare Mounted CP stent, Covered CP stent, and Mounted Covered CP stent can be scanned safely under the following conditions:
- Static magnetic field of 1.5 T and 3 T
- Maximum spatial gradient magnetic field of 2500 gauss/cm (25 T/m)
- Maximum MR system reported, whole body averaged specific absorption rate (SAR) of 2.0 W/kg for 15 minutes of scanning (Normal Operating Mode)
## 3. Biocompatibility
The biological safety assessment of the CP Stent, Mounted CP Stent, Covered CP Stent, Covered Mounted CP Stent and the BIB catheter were conducted in accordance with ISO 10993 standard series “Biological Evaluation of Medical Devices.” Each stent in the CP Stent family is classified, per ISO 10993-1, as blood-contacting, permanent (> 30 days) implant devices. Based on the results of the biocompatibility testing performed and leveraged, along with consideration of the extensive clinical use of the stent in the field, the CP Stent family and BIB catheter were determined to be biocompatible. The mounted and un-mounted stents are identical given that the only difference is that the mounted stents are crimped onto the BIB catheter. The Covered CP Stent is considered the worst case in comparison to the bare CP Stent. Thus, the biocompatibility of the Bare CP Stent, Mounted CP Stent, Covered CP Stent, and Covered Mounted CP Stent was assessed with the covered stent. Carcinogenicity testing for the CP Stent family was leveraged from the data contained within the Melody Valve PMA in conjunction with a risk assessment and evaluation of clinical experience. A summary of the testing conducted on the Covered CP Stent and the BIB catheter is provided in Table 6 and 7, respectively.
PMA P150028: FDA Summary of Safety and Effectiveness Data
{13}
Table 6. Summary of Biocompatibility Testing for the Covered CP Stent
| Test | Objectives | Results |
| --- | --- | --- |
| Cytotoxicity (L929) | Assessment of biological reactivity of mammalian cell cultures following incubation with test device extracts | Non-cytotoxic |
| Sensitization (ISO Guinea Pig Maximization Test) | Determine the potential for the test device extract to elicit contact dermal allergenicity | Non-sensitizing |
| Irritation (ISO Rabbit Intracutaneous Reactivity) | Assess potential of the device to produce irritation following a single intradermal injection of specific extracts prepared from a test device | Non-irritant |
| Systemic Toxicity (ISO Mouse Systemic Injection) | Evaluate the adverse effects after a single injection of test device extract | Not systemically toxic |
| Material-Mediated Pyrogenicity (USP Rabbit Pyrogenicity) | Evaluate the test device extract for leachates that have the potential to induce material-mediated pyrogenicity following a single dose injection | Not pyrogenic |
| Genotoxicity (AMES) | Evaluate the mutagenic potential of the device test article by measuring its ability to induce DNA reverse mutations in S. typhimurium and E. coli in the presence and absence of microsomal enzymes | Not-mutagenic |
| Genotoxicity (Mouse Lymphoma Assay) | Determine the ability of the device test article to induce forward mutations at the thymidine kinase (TK) locus as assayed by colony growth of L5178Y mouse lymphoma cells in the presence of trifluorothymidine (TFT) | Not-mutagenic |
| Genotoxicity (Mouse Peripheral Micronucleus Study) | Evaluate the potential of the device test article to induce micronuclei formation in immature polychromatic erythrocytes (PCE) present in bone marrow of adult CD-1 mice | Not-mutagenic |
| Hemocompatibility (Hemolysis) | Determine the percent hemolysis of whole blood following direct contact exposure to the test article | Non-hemolytic |
| Hemocompatibility (Complement Activation) | In vitro evaluation to measure complement activation in normal human serum when serum is exposed to a test article | Not a Sc5b-9 or C3a complement activator |
| Sub-Chronic Toxicity (ISO Rabbit Subcutaneous Implantation) | Evaluate subchronic systemic toxicity and the local effects of the implant material on living tissue following subcutaneous implantation of the test sample | No toxicity, no irritation |
| Chronic Toxicity (ISO Rabbit Subcutaneous Implantation) | Evaluate both chronic systemic toxicity and local effects of an implant material on living tissue following subcutaneous implantation of | No toxicity, no irritation |
PMA P150028: FDA Summary of Safety and Effectiveness Data
{14}
| | the test sample | |
| --- | --- | --- |
| In Vivo Swine Thrombogenicity | Assess the comparative thromboresistance of the Mounted CP Stent and the Covered Mounted CP Stent by implanting the test articles in the aorta of the swine and comparing thromboresistance with an appropriate control device through gross pathology and a subjective thrombus scoring scale | Non-thrombogenic |
A risk assessment of carcinogenic potential of the materials and manufacturing agents of the Covered CP Stent was provided in lieu of carcinogenicity testing. In addition there have been no reported incidences of cancer development associated with the Covered CP Stent that has been marketed for 12 years.
Table 7. Summary of Biocompatibility Testing for the BIB Catheter
| Test | Objectives | Results |
| --- | --- | --- |
| Cytotoxicity (L929) | Assessment of biological reactivity of mammalian cell cultures following incubation with test device extracts | Non-cytotoxic |
| Sensitization (ISO Guinea Pig Maximization Test) | Determine the potential for the test device extract to elicit contact dermal allergenicity | Non-sensitizing |
| Irritation (ISO Rabbit Intracutaneous Reactivity) | Assess potential of the device to produce irritation following a single intradermal injection of specific extracts prepared from a test device | Non-irritant |
| Systemic Toxicity (ISO Mouse Systemic Injection) | Evaluate the acute adverse effects occurring after a single injection of test device extract | Not systemically toxic |
| Material-Mediated Pyrogenicity (USP Rabbit Pyrogenicity) | Evaluate the test device extract for leachates that have the potential to induce material-mediated pyrogenicity following a single dose injection | Not pyrogenic |
| Hemocompatibility (Hemolysis) | Determine the percent hemolysis of whole blood following direct contact exposure to the test article | Non-hemolytic |
| In Vivo (Canine) Thrombogenicity | Evaluate relative thromboresistance of the materials in vivo following an approximate 2 hour implant period | Non-thrombogenic |
# B. Animal Studies
A non-GLP porcine study was conducted in 2004 to evaluate the safety of the NuMED Cheatham Platinum (CP) Balloon Expandable Stent for use in the treatment of coarctation of the aorta. A summary of this study is provided in Table 8.
PMA P150028: FDA Summary of Safety and Effectiveness Data
{15}
Table 8. Summary of Non-GLP Porcine Study
| Non-GLP Porcine Study | |
| --- | --- |
| Sample Size/Animal Model | Ten 2-3 month old (23-33kg) healthy pigs |
| Test Articles | 11 Bare CP Stents were implanted in ten animals |
| Technique | Animals were anesthetized. Cardiac catheterization and IVUS imaging was performed to establish appropriate sizing of the CP Stent. Balloons 10% larger than the diameter of the aorta distal to the left subclavian artery were used. Under fluoroscopic guidance a delivery sheath was advanced using a guide wire and the balloon-stent assembly was deployed in the area of interest. Post-deployment imaging was performed and the animals were recovered. |
| Results | All animals underwent cardiac catheterization and descending aorta angiogram at three months. Stented vessels were dilated if needed using a balloon similar in size to the largest vessel diameter adjacent to the stent. One animal was euthanized at this stage for histopathology of the stented segmentAt six months cardiac catheterization was again performed for final angiography and IVUS. All animals were terminated at the end of the procedure for necropsy and histopathology of the stented segments.There were no acute complications during implantation. The stent did not inflate symmetrically in one pig, but angiography showed good flow proximal, inside and distal to the stent. At three months all stents were patent; however, two stents were not fully expanded. Both stents were re-expanded. At six months all stents were patent and no significant complications were documented.Histopathology showed all stents to be expanded and vessels to be patent, with significant overexpansion in two of the animals. 90% of the stent struts were endothelialized, with 100% endothelialization in 6 of 11 stents. Approximately 12.7% of all struts and 7/11 stents had at least one strut defined as mal-apposed. Although the overall mean arterial injury score was low, three stents demonstrated deep medial injury or healed medial rupture and 2-3+ inflammation scores at the stent sites. Of these three cases, two had radiographic evidence of significant oversizing. These sites were also associated with significant neointimal thickening. Stent associated inflammation ranged from 0-3+, with 55% of stent sections showing 1-2+ and 45% showing 2-3+. Six of 11 stents had small calcific deposits associated with struts. There was no luminal thrombus deposition noted.Proximal and distal aortic segments were unremarkable. There was an extensive healed medial rupture in one animal which also demonstrated significant stent oversizing. |
PMA P150028: FDA Summary of Safety and Effectiveness Data
{16}
C. Additional Studies
Sterilization
The CP Stent family is sterilized using Ethylene oxide. The sterilization process has been validated to a sterility assurance level (SAL) of 10⁻⁶.
Packaging and Shelf Life
The non-mounted Bare CP Stent and Covered CP Stent are packaged in a small bottle, which is then placed in an inner and outer Tyvek pouch and heat sealed.
The Mounted CP Stent and Covered Mounted CP Stent are packaged in two Tyvek pouches and heat sealed. Over the distal end of the catheter where the stent is mounted, an additional small pouch is placed over the stent to provide extra protection.
The BiB catheter is coiled and placed in two Tyvek pouches and heat sealed.
The shelf life of the CP Stent System has been established at 5 years. Shelf life and package integrity testing was performed on 2x EO sterilized devices that were aged to 5 years using real time aging. Testing demonstrated that the sterility barrier was maintained after 5 years.
X. SUMMARY OF PRIMARY CLINICAL STUDY
The applicant performed a clinical study to establish a reasonable assurance of safety and effectiveness of implantation of the Bare CP Stent and Covered CP Stent in the native and/or recurrent coarctation of the aorta in the US under IDE G060057. Data from these clinical studies were the basis for the PMA approval decision. A summary of the clinical studies are presented below.
A. Study Design
i. COAST
Patients were treated between February 8, 2008 and November 9, 2010. The database for this PMA reflected data collected through February 1, 2015 and included 112 patients in the safety cohort. Seven patients did not receive a bare metal CP stent, leaving 105 patients in the effectiveness cohort. There were 19 investigational sites.
The study was a prospective, multi-center, single-arm clinical study comparing stent treatment of native or recurrent aortic coarctation to a performance goal (PG) derived from surgical treatment. Surgical PGs are derived from retrospective data collection at selected participating centers and from the literature.
PMA P150028: FDA Summary of Safety and Effectiveness Data
Page 17
{17}
The study used a Data Coordinating Center (DCC) that was responsible for database development, data management, monitoring data quality, monitoring adherence to the protocol by each site, monitoring device accountability, coordinating flow of information to and from the angiographic core laboratory, coordinating activities of the Data and Safety Monitoring Board (DSMB), directing data analysis and complying with FDA regulatory reporting requirements. Core labs were used to independently evaluate angiograms, MR images and angiograms, and fluoroscopic images of the coarctation stent.
## 1. Clinical Inclusion and Exclusion Criteria
Enrollment in the COAST study was limited to patients who met the following inclusion criteria:
### Pre-catheterization Inclusion Criteria:
a. Native or recurrent aortic coarctation
b. Weight 35 kg
c. Noninvasive, arm-leg cuff systolic blood pressure* difference or catheter measured systolic coarctation gradient 20 mmHg
*Patients receiving antihypertensive therapy can be included in the study. The type and dose of the medication will be recorded and used for comparisons with follow up evaluations.
### Catheterization Inclusion Criteria:
a. Coarctation of the aorta, either native or recurrent that is demonstrated, angiographically to involve the aortic isthmus or first segment of the descending aorta
b. Coarctation of the aorta found to be compliant on pre-stent balloon dilation
c. Patency of at least one femoral artery
Patients were not permitted to enroll in the COAST study if they met any of the following exclusion criteria:
### Pre-catheterization Exclusion Criteria:
a. Age > 60 years
b. Connective tissue disorders, including Marfan syndrome and other genetic syndromes such as Turner syndrome and Noonan syndrome
c. Inflammatory aortitis
d. Bloodstream infection, including endocarditis
e. Pregnancy
f. Aortic aneurysm
g. Prior stent placement
h. Adults lacking capacity to consent
i. Foster children and/or wards of the court
PMA P150028: FDA Summary of Safety and Effectiveness Data
Page 18
{18}
# Catheterization Exclusion Criteria
a. Angiography that demonstrates aortic coarctation involving a "curved" region of the aorta, the transverse aortic arch, carotid arterial branches or obstruction extending into or beyond the mid-thoracic descending aorta
b. Complete aortic atresia demonstrated angiographically
c. Anatomic location of coarctation judged by operator to preclude safe placement of a stent
d. Coarctation of the aorta found to be non-compliant on pre-stent balloon dilation
# 2. Follow-up Schedule
All patients were scheduled to return for follow-up examinations at 1 month, 6 months, 12 months, 24 months, and annually to 5 years. Adverse events and complications were recorded at all visits. The preoperative and postoperative assessments are listed in Table 9.
Table 9. Assessment at Follow-up Assessments for COAST
| | Pre-Implant Within 3 months prior to implant | In Cat h Lab | Pr e-D/ C | 1 month F/U 2 - 6 wks | 6 month F/U (4-8 months) | 12 month F/U (10-14 months) | 24 month F/U (22-26 months) | Annual to 5 years (+/- 3 months) |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Screen | X | | | | | | | |
| Informed Consent | X | | | | | | | |
| Cardiac Hx/PE | X | | X | X | X | X | X | X |
| Medication History | X | | | X | X | X | X | X |
| Paradoxical hypertension | | X | X | | | | | |
| Arm-leg systolic BP gradient1 | X | | X | X | X | X | X | X |
| Echo/Doppler2 | X | | X | | X | X | | |
| Fluoroscopy3 | | | | | | X | X | X |
| 3D MRI or CT4 | | | | | | X | X | ** |
| Adverse Events | | X | X | X | X | X | X | X |
1. Arm-leg systolic BP gradient at pre-implant must be obtained no more than 4 weeks prior to implantation. For Continued Access patients BPs can be measured as per routine institutional practice with at least one set of 4 quadrant BPs obtained at baseline and at one and two year follow up.
2. Echocardiography is not required for continued access patients.
3. Fluoroscopy not required for 3 year follow up.
4. 3D MRI scan will be performed to evaluate patient for presence of late aneurysm formation at 1 and 2 years. Additional imaging may be performed on patients with stent fractures, if clinically indicated.
# F.2. Data Collection
Data collection for the use of stents will start at time of pre-implant and implant, and will continue at follow-up periods of 1 month, 6 months, 12 months, 24 months, 36 months, 48 months, and 60 months after the procedure. Physical exam, Echocardiogram, 3D imaging by MRI or CT scanning or angiography if cardiac catheterization is performed, and adverse events will be collected for
PMA P150028: FDA Summary of Safety and Effectiveness Data
{19}
PMA P150028: FDA Summary of Safety and Effectiveness Data
Page 20
# 3. Clinical Endpoints
With regards to safety, the following criteria were evaluated:
**Primary Safety Endpoint #1**: Occurrence of any serious or somewhat serious adverse event attributed to the stent or implantation procedure within 30 days of the catheterization procedure.
The following hypothesis was tested using a one-sample, one-sided test of proportions conducted at the 0.05 level of significance:
$$
\mathrm{H}_0: \mathrm{p} \geq 0.18 \text{ vs. } \mathrm{H}_A: \mathrm{p} < 0.18
$$
**Primary Safety Endpoint #2**: Occurrence of post-procedure paradoxical hypertension.
The following hypothesis was tested using a two-sided, one-sample test of proportions conducted at the 0.05 level of significance:
$$
\mathrm{H}_0: \mathrm{p} = 0.84 \text{ vs. } \mathrm{H}_A: \mathrm{p} \neq 0.84
$$
With regards to effectiveness, the following criteria were evaluated.
**Primary Effectiveness Endpoint #1**: Reduction in arm-leg systolic blood pressure gradient from pre-dilation to the 12-month post-dilation follow-up.
Assuming that $\mu$ represents the true mean gradient reduction among stent patients, the following hypothesis was tested using a one-sample, one-sided t test conducted at the 0.05 level of significance:
$$
\mathrm{H}_0: \mu \leq 31 \text{ vs. } \mathrm{H}_A: \mu > 31
$$
**Primary Effectiveness Endpoint #2**: Length of stay in the hospital, measured in days
The following hypothesis was evaluated using a two-sided, one-sample t test conducted at the 0.05 level of significance:
$$
\mathrm{H}_0: \mu = 3.5 \text{ vs. } \mathrm{HA}: \mu \neq 3.5
$$
# ii. COAST II
Patients were treated between May 8, 2008 and December 14, 2011. The database for this PMA reflected data collected through February 1, 2015 and included 82 patients. There were 19 investigational sites.
The study was a prospective, multi-center, single-arm clinical study that evaluates the Covered CP Stent for treatment of coarctation of the aorta. For effectiveness, each patient serves as his or her own control. For safety, a performance goal was derived from surgical literature.
{20}
The study used a Data Coordinating Center (DCC) that was responsible for database development, data management, monitoring data quality, monitoring adherence to the protocol by each site, monitoring device accountability, coordinating flow of information to and from the angiographic core laboratory, coordinating activities of the Data and Safety Monitoring Board (DSMB), directing data analysis and complying with FDA regulatory reporting requirements. Core labs were used to independently evaluate angiograms, MR images and angiograms, and fluoroscopic images of the coarctation stent.
## 1. Clinical Inclusion and Exclusion Criteria
Enrollment in the COAST II study was limited to patients who met the following inclusion criteria:
Native or recurrent aortic coarctation*associated with one or more of the following:
1. Acute or chronic aortic wall injury**
2. Nearly atretic descending aorta to 3 mm or less in diameter.
3. Genetic Syndromes associated with aortic wall weakening. Individuals with genetic syndromes such as Marfan Syndrome, Turner's Syndrome or familial bicuspid aortic valve and ascending aortic aneurysm.
4. Advanced age. Men and woman aged 60 years or older.
* The significance of aortic obstruction is left to the judgment of the participating investigator. Indications might include mild resting aortic obstruction associated with:
- Exercise related upper extremity hypertension;
- Severe coarctation with multiple and/or large arterial collaterals;
- Single ventricle physiology
- Left ventricular dysfunction
- Ascending aortic aneurysm
**Aortic wall injury might include:
- Descending aortic aneurysm
- Descending aortic pseudo-aneurysm
- Contained aortic wall rupture
- Non-contained rupture of the aortic wall
Patients were not permitted to enroll in the COAST II study if they met any of the following exclusion criteria:
a. Patient size too small for safe delivery of the device. The absolute lower limit for inclusion under this protocol is 20 kg. However, serious
PMA P150028: FDA Summary of Safety and Effectiveness Data
Page 21
{21}
femoral artery injury can occur in small patients, particularly those in the 20-30 kg range and this risk must be reviewed in detail with parents or guardians of children in this weight range.
b. Planned deployment diameter less than 10 mm or greater than 22 mm
c. Location requiring covered stent placement across a carotid artery*
d. Adults lacking capacity to consent
e. Pregnancy
*crossing or covering of a subclavian artery is acceptable in certain situations, but only after alternative treatments have been considered.
## 2. Follow-up Schedule
All patients were scheduled to return for follow-up examinations at 1 month, 6 months, 12 months, 24 months, and annually thereafter to 5 years. Adverse events and complications were recorded at all visits. The preoperative and postoperative assessments are listed in the Table 10.
Table 10. Summary of Follow-up Assessments for COAST II
| | Pre-Implant
Within 12 wks prior to implant | In Cath Lab | Pre-D/C | 1 month F/U
(2 - 6 wks) | 6 month F/U
(4-8 months) | 12 month F/U
(18-14 months) | 24 month F/U
(20-28 months) | Annual to 5 years
(+/- 4 months) |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Screening | X | | | | | | | |
| Informed Consent | X | | | | | | | |
| Cardiac Hx/PE | X | | X | X | X | X | X | X |
| Medication History | X | | | X | X | X | X | X |
| SIS Assessment | X | | | | | X | X | |
| Arm-leg systolic BP gradient* | X | | X | X | X | X | X | X |
| Fluoroscopy** | | | | | | X | X | X |
| 3D MRI scan* | | | | | | X | X | |
| Adverse Events | | X | X | X | X | X | X | X |
+ Replicate, four quadrant BPs at baseline, 1 and 2 yrs, with clinical, arm and leg pressure measurements recorded at interim time periods
* If prior stainless steel stent(s) are in place, then CT scanning will be required instead of MRI
**Fluoroscopy not necessary for the 3 year follow up visit, but is added at 4 and 5 years.
## 3. Clinical Endpoints
With regards to safety, the following criteria were evaluated.
**Primary Safety Endpoint:** Occurrence of any serious or somewhat serious adverse event attributed to the stent or implantation procedure within 30 days of the catheterization procedure.
PMA P150028: FDA Summary of Safety and Effectiveness Data
Page 22
{22}
The following hypothesis was tested using a one-sample, one-sided test of proportions conducted at the 0.05 level of significance:
$$
\mathrm{H}_0: \mathrm{p} \geq 0.18 \text{ vs. } \mathrm{H}_A: \mathrm{p} < 0.18
$$
**Secondary Safety Endpoint:** Proportion of patients experiencing any of the following adverse events related to the device or implant procedure post 1 year
- Underlying cardiac or non-cardiac disease, aortic wall injury, new aortic aneurysm formation within region of device, stent misplacement, malposition, stent fracture, aortic wall aneurysms, or restenosis requiring reintervention.
With regards to effectiveness, the following criteria were evaluated.
**Primary Effectiveness Endpoint #1:** Improvement of aortic wall injury and/or aortic arch obstruction by a median increase of at least one grade from pre-implantation baseline to 12-month follow-up using the Severity of Illness Scale (based on upper extremity (UE) systolic BP, UE to lower extremity (LE) systolic BP, and aortic wall injury).
The following hypothesis was tested using a one-sided Wilcoxon signed-rank test conducted at the 0.025 level of significance:
$$
\mathrm{H}_0: \text{median change in grade} \leq 0 \text{ vs. } \mathrm{H}_A: \text{median change in grade} > 0
$$
**Primary Effectiveness Endpoint #2:** Aortic wall injury and aortic arch obstruction at Grade 4 or above at the 12-month follow-up, based on the Severity of Illness Scale, with no clinical worsening.
The following hypothesis was tested using a one-sample, one-sided test of proportions conducted at the 0.05 level of significance:
$$
\mathrm{H}_0: \mathrm{p} \leq 0.70 \text{ vs. } \mathrm{H}_A: \mathrm{p} > 0.70
$$
**Secondary Effectiveness Endpoints:**
- Reduction of arm-leg systolic blood pressure gradients to less than $20\,\mathrm{mmHg}$ and less than $15\,\mathrm{mmHg}$.
- Reduction of upper extremity blood pressure at 1 year compared to baseline
- Repair of wall defect with $<10\%$ residual endoleak on MRI or CT in patients with aortic wall injury
- Hospital length of stay compared to length of stay for surgical repair of aortic coarctation.
PMA P150028: FDA Summary of Safety and Effectiveness Data
Page 23
{23}
B. Accountability of PMA Cohort
COAST
Of the 167 enrolled patients, 102 met the study eligibility criteria and were treated with the CP Stent. 112 patients were evaluated for safety, of which 5 patients crossed-over to the Covered CP Stent therapy. Approximately 107 stents were implanted in 105 patients and these patients were included in the evaluation of effectiveness.
At the time of database lock, of 112 safety cohort patients and 105 effectiveness cohort patients, 87 patients were available for analysis at the completion of the study, the 24 month post-operative visit. Study accountability is detailed in Table 11.
Table 11. COAST Accountability
| | Possible N (100%) | 1 Month Visit n (%) | 12 Month Visit n (%) | 24 Month Visit n (%) | 3 years n (%) | 4 years n (%) | 5 years n (%) |
| --- | --- | --- | --- | --- | --- | --- | --- |
| Safety Cohort | 112 | 102 (91%) | 94 (84%) | 90 (80%) | 80 (71%) | 76 (68%) | 73 (65%) |
| Effectiveness Cohort | 105* | 100 (95%) | 92 (88%) | 87 (83%) | 77 (73%) | 69 (66%) | 56 (53%) |
*112 patients underwent catheterization and pre-stenting balloon angioplasty, five then received Covered CP Stents and were entered into the COAST II trial where they have been followed since. Two patients did not receive study stents and are followed for safety outcomes only.
COAST II
At the time of database lock, of 82 patients enrolled in the PMA study, are available for analysis at the completion of the study endpoints (i.e. the 24 month post-implant visit). Study accountability is detailed in Table 12.
Table 12. COAST II Accountability
| COAST II Patients | Possible N (100%) | 1 Month Visit n (%) | 12 Month Visit n (%) | 24 Month Visit n (%) | 3 years n (%) | 4 years n (%) | 5 years n (%) |
| --- | --- | --- | --- | --- | --- | --- | --- |
| Safety Cohort | 82 | 82 (100%) | 69 (84%) | 67 (81.7%) | 55 (67.1%) | 38 (46.3%) | 22 (26.8%) |
| Effectiveness Cohort | 82 | 82 (100%) | 68 (83%) | 66 (80.5%) | 54 (65.8%) | 37 (45.1%) | 21 (25.6%) |
PMA P150028: FDA Summary of Safety and Effectiveness Data
{24}
# C. Study Population Demographics and Baseline Parameters
The demographics of the study population are typical for a coarctation of the aorta study performed in the US. The COAST and COAST II demographics are shown in Table 13.
Table 13. COAST and COAST II Patient Characteristics
| COAST | | |
| --- | --- | --- |
| Assessment | Number (Percent) or Median (Range) | |
| | Safety Cohort (n=112) | Efficacy Cohort (n=105) |
| Gender | | |
| Male | 77 (69%) | 73 (70%) |
| Female | 35 (31%) | 32 (30%) |
| Age, years | 16 (8 to 52) | 16 (8 to 52) |
| NYHA Classification | | |
| I | 88 (79%) | 82 (78%) |
| II | 22 (20%) | 21 (20%) |
| III | 1 (1%) | 1 (1%) |
| IV | 1 (1%) | 1 (1%) |
| Primary Indication | | |
| Native Coarctation | 65 (58%) | 60 (57%) |
| Recurrent Coarctation | 47 (42%) | 45 (43%) |
| COAST II | | | |
| --- | --- | --- | --- |
| Assessment | Number (Percent) or Median (Range) | | |
| | Prospective (n=29) | Legacy (n=53) | Total (n=82) |
| Gender | | | |
| Male | 21 (72%) | 31 (58%) | 52 (63%) |
| Female | 8 (28%) | 22 (42%) | 30 (37%) |
| Age, years | 20 (6 to 67) | 17 (6 to 66) | 18 (6 to 67) |
| Primary Indication | | | |
| Repair of aortic wall injury | 15 (52%) | 34 (64%) | 49 (60%) |
| Prevention of aortic wall injury1 | 14 (48%) | 19 (36%) | 33 (40%) |
| 1Includes 1 patient classified as not having pre-existing aortic wall injury, who was noted to have a small, localized intimal tear with a diameter of < 1/4 the aortic diameter (study number 013-501). | | | |
PMA P150028: FDA Summary of Safety and Effectiveness Data
{25}
D. Safety and Effectiveness Results
1. Safety Results
The analysis of safety was based on the implanted cohort of 112 COAST and 82 COAST II patients completing their implant procedures. The primary safety outcomes are presented in Table 14.
Table 14. Summary of COAST and COAST II Outcomes and Pre-Specified Safety Endpoints
| | Safety Endpoint | Event Rate | P Value (CI) |
| --- | --- | --- | --- |
| COAST Primary | Serious or Somewhat Serious Adverse event attributed to the Stent or Implantation procedure within 30 days of the procedure | 8.9% | 0.006 (4.9%, 14.7%)* |
| | Post-procedure paradoxical hypertension | 7.5% | <0.001 (3.3%, 14.2%)* |
| COAST II Primary | Serious or Somewhat Serious Adverse Events Attributed to the Stent, Implantation or Catheterization within 30 days of the procedure (includes data from COAST combined with COAST II) | 8.2% | <0.001 (5.2%, 12.3%)* |
| Secondary | Proportion of patients experiencing any AEs related to the device or implant procedure post 1 year (among 74 patients followed for at least 1 year) | 6.8% | N/A (2.2%, 15.1%)*# |
*90% Confidence interval
*95% Confidence Interval
# confidence interval provided to illustrate the variability only and should not be used to draw any statistical conclusion.
The COAST and COAST II primary safety endpoints were met with the occurrence of any serious or somewhat serious adverse event within 30 days post procedure being less than the predefined 18%. Post procedural paradoxical hypertension was observed in 7.5% of patients in the COAST Trial. The COAST primary safety endpoint for the incidence of paradoxical hypertension (< 84%) was met.
**Adverse effects that occurred in the PMA clinical study:**
The overall incidence and types of adverse events were within expected ranges. Aortic wall injuries were rare and treated appropriately without the need for emergency surgery. The results are durable out to 60 months for each study and re-coarctation was treated by transcatheter means when it occurred. There were no late complications related to device fracture noted, though incidence of stent fracture increased with time for COAST patients. Table 15 provides a summary of the adverse events reported under COAST and COAST II.
PMA P150028: FDA Summary of Safety and Effectiveness Data
Page 26
{26}
Table 15. Summary of Adverse Events (AEs) for COAST
| | Stent Related Events1(Rates) | Stent, Implantation, or Catheterization Related Events2(Rates) | All Events (Rates) |
| --- | --- | --- | --- |
| Patients with adverse events at 30 days | 1(0.9%) | 37(33.0%) | 50(44.6%) |
| Serious or somewhat serious events at 30 days | 1(0.9%) | 10(8.9%) | 11(9.8%) |
| Serious or somewhat serious events at 30 days, excluding stent fracture | 1(0.9%) | 10(8.9%) | 11(9.8%) |
| Serious event at 30 days | 0(0.0%) | 0(0.0%) | 0(0.0%) |
| Patients with adverse events at 12 Months | 4(3.6%) | 40(35.7%) | 69(61.6%) |
| Serious or somewhat serious events at 12 months | 4(3.6%) | 13(11.6%) | 16(14.3%) |
| Serious or somewhat serious events at 12 months, excluding stent fracture | 2(1.8%) | 11(9.8%) | 14(12.5%) |
| Serious event at 12 months | 0(0.0%) | 0(0.0%) | 0(0.0%) |
| Patients with adverse event at 24 Months | 13(11.6%) | 47(42.0%) | 73(65.2%) |
1Includes events that are due to or possible due to stent, and stent fractures.
2Includes events that are due to or possible due to stent, implantation, or catheterization, and stent fractures
PMA P150028: FDA Summary of Safety and Effectiveness Data
{27}
Table 16. Summary of Adverse Events (AEs) for COAST II
| | Stent Related Events1 (Rates) | Stent, Implantation, or Catheterization Related Events2 (Rates) | All Events (Rates) |
| --- | --- | --- | --- |
| Patients with adverse events at 30 days | 2 (2.4%) | 27 (32.9%) | 42 (51.2%) |
| Serious or somewhat serious events at 30 days | 1 (1.2%) | 6 (7.3%) | 7 (8.5%) |
| Serious or somewhat serious events at 30 days, excluding stent fracture | 1 (1.2%) | 6 (7.3%) | 7 (8.5%) |
| Serious event at 30 days | 0 (0.0%) | 1 (1.2%) | 1 (1.2%) |
| Patients with adverse events at 12 Months | 4 (4.9%) | 30 (36.6%) | 58 (70.7%) |
| Serious or somewhat serious events at 12 months | 3 (3.7%) | 7 (8.5%) | 14 (17.1%) |
| Serious or somewhat serious events at 12 months, excluding stent fracture | 3 (3.7%) | 7 (8.5%) | 14 (17.1%) |
| Serious event at 12 months | 1 (1.2%) | 1 (1.2%) | 4 (4.9%) |
| Patients with adverse event at 24 Months | 5 (6.1%) | 31 (37.8%) | 60 (73.2%) |
1Includes events that are due to or possible due to stent, and stent fractures.
2Includes events that are due to or possible due to stent, implantation, or catheterization, and stent fractures
Tables 17 through 19 document the stent related, implantation related and catheterization procedure related adverse events in the COAST and COAST II trials.
Table 17. Stent Related Adverse Events for COAST and COAST II
| | Event | n (Event Rate) |
| --- | --- | --- |
| COAST (n=112) | Aortic Aneurysm | 1 (0.9 %) |
| | Increased cardiac output, tachycardia/light-headedness | 1 (0.9 %) |
| COAST II (n =82) | Aortic Aneurysm | 2 (2.4 %) |
| | Asymmetric Stent Shortening | 1 (1.2 %) |
| | Left arm numbness and weakness | 1 (1.2 %) |
PMA P150028: FDA Summary of Safety and Effectiveness Data
{28}
Table 18. Implantation Related Adverse Events for COAST and COAST II
| | Event | n (Event Rate) |
| --- | --- | --- |
| COAST (n=112) | Aortic Aneurysm | 2 (1.8 %) |
| | Back Pain | 1 (0.9 %) |
| | Chest Pain | 5 (4.5 %) |
| | Confined Vascular Tear | 3 (2.7 %) |
| | Groin Pain | 4 (3.6 %) |
| | Intimal tear with vessel irregularity | 1 (0.9 %) |
| | Jailed left subclavian | 1 (0.9 %) |
| | Local hematoma groin | 1 (0.9 %) |
| | Pain | 1 (0.9 %) |
| | Right leg pain | 1 (0.9 %) |
| | Stent malposition | 1 (0.9 %) |
| COAST II (n =82) | Aneurysmal formation | 1 (1.2 %) |
| | Chest pain | 4 (4.9 %) |
| | Chest and back pain | 1 (1.2 %) |
| | Easy bruising on aspirin | 1 (1.2 %) |
| | Increased bruising | 1 (1.2 %) |
| | Right groin pain | 1 (1.2 %) |
| | Stent malposition | 2 (2.4 %) |
| | Wound bleeding | 1 (1.2 %) |
Table 19. Catheterization Related Adverse Events for COAST and COAST II
| | Event | n (Event Rate) |
| --- | --- | --- |
| COAST (n=112) | AV fistula | 1 (0.9 %) |
| | Bleeding | 1 (0.9 %) |
| | Corneal abrasion | 1 (0.9 %) |
| | Decreased pulse | 1 (0.9 %) |
| | Fever | 1 (0.9 %) |
| | Groin pain | 1 (0.9 %) |
| | Local hematoma groin | 4 (3.6 %) |
| | Right inguinal hematoma | 1 (0.9 %) |
| COAST II (n =82) | Aneurysm | 1 (1.2 %) |
| | Atrial arrhythmia | 1 (1.2 %) |
| | Brachial plexus injury | 1 (1.2 %) |
| | Contact skin rash | 1 (1.2 %) |
| | Corneal abrasion | 1 (1.2 %) |
| | Discomfort right eye | 1 (1.2 %) |
| | Dissection of iliac artery | 1 (1.2 %) |
| | Ecchymosis/groin tenderness | 1 (1.2 %) |
| | Femoral artery occlusion | 1 (1.2 %) |
| | Local hematoma groin | 2 (2.4 %) |
| | Localized groin bruising | 1 (1.2 %) |
| | Minimal bleeding/cough | 1 (1.2 %) |
| | Neck swelling | 1 (1.2 %) |
| | Pulsatile bleeding | 1 (1.2 %) |
| | Right iliac dissection/pulse loss | 1 (1.2 %) |
| | Superficial infection of groin | 1 (1.2 %) |
| | Wide complex non-sustained tachycardia | 1 (1.2 %) |
PMA P150028: FDA Summary of Safety and Effectiveness Data
{29}
There were five patients that crossed over from COAST to COAST II. One patient crossed over due to a small aneurysm after dilation, two patients due to a near atretic aorta, one patient due to localized intimal tear after dilation and one patient due to an acute, rapidly expanding aneurysm after dilation. In the COAST trial, one patient experienced stent malposition, representing 0.9% of patients, and this event was resolved in the catheterization lab with no permanent damage. In the COAST II trial, two patients experienced two events, representing 2.4% of patients, and both events were resolved using a second Covered CP stent to fully occlude the aneurysm developed with no permanent damage.
In the COAST trial, five patients experienced aortic wall injury, four that occurred prior to hospital discharge and one that occurred by 24-month follow-up. These five events are detailed in Table 20, below.
Table 20. COAST Aortic Wall Injuries by 24 Month Follow-up
| Patient | Bare Metal CP Stent Implanted | Type of Injury | Outcome of Event |
| --- | --- | --- | --- |
| 1 | No | Small aneurysm after dilation | Cross-over to Covered CP Stent |
| 2 | Yes | Therapeutic and localized tear | Resolved after implantation of Bare Metal CP Stent; tear no longer visible and not noted on further imaging |
| 3 | Yes | Contained rupture | Possible minimal aneurysm with no progression during admission; not noted on further imaging |
| 4 | Yes | Aneurysm | Implantation of Covered CP Stent |
| 5 | No | Acute, rapidly expanding aneurysm after dilation | Cross-over to Covered CP Stent |
In the COAST II trial, three patients experienced aortic wall injuries by the 24-month follow-up. These injuries are detailed in Table 21, below.
Table 21. COAST II Aortic Wall Injuries by 24 Month Follow-up
| Patient | Injury Detected | Intervention |
| --- | --- | --- |
| 1 | Neo-intimal proliferation | Therapy for new aortic wall injury – implantation of Covered CP Stent |
| 2 | Small aneurysm at 12 m visit | Therapy for new aortic wall injury – implantation of Covered CP Stent |
PMA P150028: FDA Summary of Safety and Effectiveness Data
{30}
In COAST, four patients with events underwent catheter reinterventions, representing 3.7% of patients with an event. No surgical interventions were performed. Table 22 provides a summary of these four interventions.
Table 22. COAST Coarctation-Related Reintervention by 24 Months
| Approximate Time to Intervention Post-procedure (Months) | Indication for Reintervention | Procedure Performed |
| --- | --- | --- |
| 11 | Planned reintervention to fully expand stent | Redilation of stent, implantation of non-study stent |
| 20 | Planned re-expansion of stent | Redilation of stent |
| 14 | Aneurysm detected at 12m visit | Therapy for new aortic wall injury - implantation of Covered CP Stent |
| 14 | Planned re-expansion of stent, signs of restenosis at coarctation site | Redilation of stent |
In COAST II, six patients experienced coarctation-related events, representing 7.3% of patients with events. These patients underwent catheter reinterventions. No surgical interventions were completed. Table 23 provides a summary of these interventions.
Table 23. COAST II Coarctation-Related Reintervention by 24 months
| Approximate Time to Intervention Post-procedure (Months) | Indication for Reintervention | Procedure Performed |
| --- | --- | --- |
| 7 | Persistent hypertension and gradient across aortic arch | Radiation of stent |
| 7 | Planned re-expansion of stent | Radiation of stent |
| 9 | Planned re-expansion of stent | Radiation of stent |
| 25 | Increased gradient with somatic growth; neo-intimal proliferation detected in cath lab | Therapy for new aortic wall injury - implantation of Covered CP Stent |
| 14 | Aneurysm detected by MRI at 12 m visit¹ | Therapy for new aortic wall injury - implantation of Covered CP Stent |
| 13 | Aneurysm detected by MRI at 12 m visit¹ | Therapy for new aortic wall injury - implantation of Covered CP Stent |
¹ Presence of aneurysm in this patient was not confirmed by core laboratory review of the MRI
In COAST, two patients experienced non-coarctation related reinterventions that were documented by the 24-month follow-up. These patients underwent surgical interventions. The time to intervention for one patient was 20 months, at which time the patient had an aortic valve replacement to address progressive and severe left ventricular enlargement and progression of exercise intolerance. The time to intervention for the second patient was approximately two months, when the patient received a mitral valve replacement to address mitral regurgitation. In COAST II, two patients experienced non-coarctation
PMA P150028: FDA Summary of Safety and Effectiveness Data
Page 31
{31}
related reinterventions that were documented by the 24 month follow-up, representing 2.4% of the patients with events. These patients underwent catheter reinterventions. The time to intervention for one patient was four months, when the patient received a coronary angiogram and graft angiogram to address symptoms of angina. The time to intervention for the second patient was 26 months when the Melody valve was implanted to address a high right ventricle to pulmonary artery conduit gradient.
Stent fracture was increasingly common during follow-up ranging from 12% at 24 months to 36% at 60 months in patients treated with bare metal stents (COAST trial). However, no loss of structural integrity and no complications resulting in patient injury were observed. The incidence of stent fracture for covered CP stents (COAST II) was much lower and was not observed to substantially increase over time. Also, relief from blood pressure gradient was maintained through 60 month follow-up and re-intervention was rare. When needed, this was accomplished using transcatheter interventions.
Table 24 shows the stent fracture events in the COAST pivotal cohort.
Table 24: COAST Pivotal Cohort –Stent Fracture
| | Completed 12 Month Fluoroscopy (n=91)^{1} | Completed 24 Month Fluoroscopy (n=87)^{3} |
| --- | --- | --- |
| Percentage of Eligible Subjects Undergoing Fluoroscopy | 91/104^{2} (88%) | 87/103^{4} (84%) |
| Stent Fracture | 2 (2.2%) | 11 (12.6%) |
| No loss of structural integrity | 2 | 11 |
| Loss of structural integrity | 0 | 0 |
1 Among 104 eligible subjects, excludes: 1 patient lost to follow-up at 1 month, 2 patients lost to follow-up at 6 months, and 3 patients lost to follow-up at 12 months. An additional 7 patients did not undergo fluoroscopy at 12 months.
2 Cross-over patients treated with Covered CP Stent (5) were followed only through implantation of Bare Metal CP Stent; intent to treat patients (2) were followed only through hospital discharge. One patient withdrew consent prior to the 12 month visit.
3 Among 103 eligible subjects, excludes: 6 patients previously lost to follow-up, and 5 patients lost to follow-up at 24 months. An additional 5 patients did not undergo fluoroscopy at 24 months.
4 In addition to cross-over patients treated with Covered CP Stent, intent to treat patients, and 1 patient who withdrew consent prior to the 12 month follow-up, 1 patient was noted to have an aneurysm at the 12 month visit and was treated using a Covered CP Stent; this patient is no longer followed for COAST and is currently enrolled in COAST II.
A summary of the number of stents implanted in COAST and COAST II enrolled patients who underwent cardiac catheterization for the purpose of Coarctation of the Aorta is provided in Tables 25 and 26.
PMA P150028: FDA Summary of Safety and Effectiveness Data
Page 32
{32}
Table 25: COAST Pivotal Cohort – Procedural Data
| | Number (Percent) |
| --- | --- |
| | Treated with Bare Metal CP Stent, or Meeting Study Eligibility Criteria but Not Treated with Bare Metal CP Stent (n=107)^{1} |
| Bare Metal CP Stent Implanted | 105 (98%) |
| Second Bare Metal CP Stent Implanted (n=105) | 2 (2%) |
1 Includes: Patients treated with Bare Metal CP Stent meeting study eligibility criteria (102), patients treated with Bare Metal CP Stent not meeting study eligibility criteria (3), patients meeting study eligibility but not treated with Bare Metal CP Stent (2).
Table 26: COAST II Pivotal Cohort – Procedural Data
| | Number (Percent) |
| --- | --- |
| | Total
(n=82) |
| Covered CP Stent Implanted | 82 (100%) |
| Second Covered CP Stent Implanted | 9 (11%) |
| Third Covered CP Stent Implanted | 2 (2%) |
| Patient Free of Explant 24 hours after Procedure | 82 (100%) |
2. Effectiveness Results
The analysis of effectiveness was based on the 105 COAST patients receiving bare metal CP stents and 82 patients receiving Covered CP Stents in the COAST II study. The key effectiveness outcomes are presented in Tables 27 through 29.
Although the COAST primary effectiveness endpoint for blood pressure gradient reduction was not met, the observed blood pressure gradients were clinically meaningful. Failure to meet the endpoint was a function of the endpoint chosen rather than a failure to achieve reduction of the underlying gradient. From a clinical perspective, relief of systolic blood pressure gradient was complete and sustained.
PMA P150028: FDA Summary of Safety and Effectiveness Data
Page 33
{33}
Table 27. Summary of Late Outcomes and Major Pre-Specified Primary Effectiveness Endpoints
| COAST Primary | Effectiveness Endpoint | Event Rate | P Value (CI) |
| --- | --- | --- | --- |
| | Mean Reduction in Systolic Blood Pressure Difference, Pre-Dilation to the 12 Month Post-Dilation Follow-Up | 30 ±22mmHg | 0.64 (26mmHg, 34mmHg)* |
| | Length of Stay in Hospital | 1.1±0.3 days | <0.001 (1.0 days, 1.1 days)+ |
| COAST II Primary | Severity of Illness Scale Grade 4 or 5 with No Clinical Worsening at 12 Month Follow-up | 80% | 0.048 (70.1%, 87%)* |
| COAST II Secondary | Proportion of patients with arm-leg systolic blood pressure differences less than 20mmHg and less than 15 mmHg at 12 month follow-up, compared to baseline | 87% (up from 46% at baseline) 79% (up from 38% at baseline) | p<0.001 (76%, 94%)+ p<0.001 (68%, 88%)+ |
| | Reduction of upper extremity blood pressure at 1 year compared to baseline | 12 ±20mmHg | N/A (7mmHg, 17mmHg)*# |
| | Complete repair of aortic wall defect with first Covered CP Stent (no residual endoleak during the catheterization procedure | 47 of 49 (96%) of patients treated for an aortic wall injury | N/A |
| | Proportion of patients with effective treatment of AWI with no residual aneurysm seen on MRI scanning | 37 of 39 (95%) patients treated for an aortic wall injury 1/39 (2.5%) with a small aneurysm and one patient's MRI could not be evaluated by core lab | N/A |
| | Hospital length of stay compared to length of stay for surgical repair of aortic coarctation | 1.2 ± 0.9 days | <0.001 (1.0 days, 1.4 days)+ |
*90% Confidence interval
+95% Confidence Interval
Confidence interval provided to illustrate the variability only and should not be used to draw any statistical conclusion.
PMA P150028: FDA Summary of Safety and Effectiveness Data
{34}
Table 28. COAST Pivotal Cohort – Systolic Blood Pressure
| | Number (Percent) or Median (Range) And Mean ± Standard Deviation | | |
| --- | --- | --- | --- |
| | Completed 1 Month Follow-up (n=100)¹ | Completed 12 Month Follow-up (n=92)² | Completed 24 Month Follow-up (n=87)³ |
| Upper Extremity Systolic Blood Pressure (mmHg) | 118 (83 to 148) | 122 (82 to 148) | 121 (87 to 175) |
| Median (range) | 120 ± 12 | 123 ± 12 | 122 ± 14 |
| Mean ± standard deviation | | | |
| Lower Extremity Systolic Blood Pressure (mmHg) | 119 (91 to 163) | 122 (89 to 180) | 123 (84 to 172) |
| Median (range) | 122 ± 15 | 123 ± 15 | 125 ± 16 |
| Mean ± standard deviation | | | |
| Systolic Blood Pressure Difference (mmHg) | -1 (-45 to 32) | -1 (-37 to 40) | -4 (-46 to 43) |
| Median (range) | -2 ± 13 | -1 ± 15 | -3 ± 15 |
| Mean ± standard deviation | | | |
| Reduction in Systolic Blood Pressure Difference Pre-Dilation (n=99⁴, 91⁵) | 31 ± 18 | 30 ± 22⁶ | 33 ± 20 |
¹ Among 104 eligible subjects, excludes: 1 patient lost to follow-up at 1 month. An additional 3 patients missed the 1 month visit.
² Among 104 eligible subjects, excludes: 1 patient previously lost to follow-up, 2 patients lost to follow-up at 6 months, and 3 patients lost to follow-up at 12 months. An additional 6 patients missed the 12 month visit.
³ Among 101 eligible subjects, excludes: 6 patients previously lost to follow-up, and 5 patients lost to follow-up at 24 months. An additional 3 patients missed the 24 month visit.
⁴ One patient does not have 1 month systolic blood pressure difference due to missing lower extremity pressure.
⁵ One patient does not have 12 month systolic blood pressure difference due to missing lower extremity pressure.
⁶ Primary effectiveness outcome.
PMA P150028: FDA Summary of Safety and Effectiveness Data
{35}
Table 29. COAST II Pivotal Cohort – Systolic Blood Pressure
| | Number (Percent) or Median (Range) And Mean ± Standard Deviation | |
| --- | --- | --- |
| | Completed 12 Month Follow-up (n=68)¹ | Completed 24 Month Follow-up (n=66)² |
| Upper Extremity Systolic Blood Pressure (mmHg) | 123 (98 to 166) | 126 (96 to 158) |
| Median (range) | 125 ± 14 | 126 ± 13 |
| Mean ± standard deviation | | |
| Lower Extremity Systolic Blood Pressure (mmHg) | 121 (90 to 199) | 124 (86 to 156) |
| Median (range) | 124 ± 21 | 125 ± 16 |
| Mean ± standard deviation | | |
| Systolic Blood Pressure Difference (mmHg) | 2 (-48 to 38) | 0 (-35 to 62) |
| Median (range) | 1 ± 16 | 1 ± 17 |
| Mean ± standard deviation | | |
| Systolic Blood Pressure Difference | | |
| < 10 mmHg | 46 (68%) | 52 (79%) |
| < 15 mmHg | 54 (79%) | 56 (85%) |
| < 20 mmHg | 59 (87%) | 60 (91%) |
¹ Among 81 eligible subjects, excludes: 3 patients lost to follow-up at 6 and 4 patients lost to follow-up at 12 months. An additional 6 patients missed the 12 month visit.
² Among 80 eligible subjects, excludes: 7 patients previously lost to follow-up, and 2 patients lost to follow-up at 24 months. An additional 5 patients missed the 24 month visit.
## 3. Subgroup Analyses
The following preoperative characteristics were evaluated for potential association with outcomes: age, gender, and condition. While the early adverse event rate was higher in females, likely due to peripheral artery size differences, the residual gradient not different by gender. These subgroup analyses did not reveal significant differences in treatment outcomes.
In the COAST trial, no subgroup analysis was performed by age. Under COAST II, in a comparison of outcome by age, there was no significant difference in the proportions of patients with a serious or somewhat serious adverse event attributed to the stent or procedure for patients younger or older than 16 years old. A subgroup analysis for gender was completed for COAST and COAST II. The outcome defined as the proportion of patients with a serious or somewhat serious adverse event attributed to the stent or procedure, exhibited a worse trend for females than males, although not statistically different. The remaining outcomes in each study did not present significant differences between genders. A comparison of outcome by
PMA P150028: FDA Summary of Safety and Effectiveness Data
Page 36
{36}
primary indication was performed for the COAST safety cohort. There was no significant difference in outcomes of the safety cohort with native coarctation or recurrent coarctation. There was also no difference in the outcomes of the safety cohort with repair of the aortic wall injury or prevention of aortic wall injury. Also, using the COAST II data, a comparison of outcomes by coarctation minimum diameter was performed and did not demonstrate any difference between the effectiveness cohort with a minimum diameter of ≤ 3.0mm and ≥ 3.1mm.
## E. Financial Disclosure
The Financial Disclosure by Clinical Investigators regulation (21 CFR 54) requires applicants who submit a marketing application to include certain information concerning the compensation to, and financial interests and arrangement of, any clinical investigator conducting clinical studies covered by the regulation. The pivotal clinical study included 2 investigators of which none were full-time or part-time employees of the sponsor and 2 had disclosable financial interests/arrangements as defined in 21 CFR 54.2(a), (b), (c) and (f) and described below:
- Compensation to the investigator for conducting the study where the value could be influenced by the outcome of the study: None
- Significant payment of other sorts: 1
- Proprietary interest in the product tested held by the investigator: 2
- Significant equity interest held by investigator in sponsor of covered study: None
The applicant has adequately disclosed the financial interest/arrangements with clinical investigators. Statistical analyses were conducted by FDA to determine whether the financial interests/arrangements had any impact on the clinical study outcome. The information provided does not raise any questions about the reliability of the data.
## XI. SUMMARY OF SUPPLEMENTAL CLINICAL INFORMATION
The supplemental clinical data to the PMA application came from the Continued Access studies for COAST and COAST II conducted in the United States. The objective was to collect additional safety and effectiveness data on the CP Stent Family in subjects with aortic coarctation.
## A. Study Design
### COAST CAP
The COAST Continued Access Protocol was a prospective, multi-center, single-arm clinical study allowing continued access to the CP Stent during regulatory review of the pre-market application for the CP Stent. Up to 19 investigational sites with prior CP Stent experience in the COAST Study were allowed to participate.
PMA P150028: FDA Summary of Safety and Effectiveness Data
Page 37
{37}
The endpoints of the COAST CAP registry were similar to those in the COAST Study but there were no pre-specified statistical hypotheses. The primary safety endpoints were the occurrence of device or implant related serious adverse events and the occurrence of post-procedure paradoxical hypertension. The primary effectiveness endpoints were reduction in arm-leg systolic blood pressure gradient and length of stay in the hospital. All patients were scheduled to return for follow-up examinations at 1 month, 6 months, 12 months, 24 months, and annually to 5 years.
## COAST II CAP
The COAST II Continued Access Protocol was a prospective, multi-center, single-arm clinical study allowing continued access to the Covered CP Stent during regulatory review of the pre-market application for the Covered CP Stent. Up to 19 investigational sites with prior Covered CP Stent experience in the COAST II Study were allowed to participate.
The endpoints of the COAST II CAP registry were similar to those in the COAST II Study but there were no pre-specified statistical hypotheses. The primary safety endpoint was the occurrence of device- or procedure- related serious adverse events. The primary effectiveness endpoint was improvement of aortic wall injury and/or aortic arch obstruction. All patients were scheduled to return for follow-up examinations at 1 month, 6 months, 12 months, 24 months, and annually to 5 years.
## B. Accountability of Non-Pivotal Trials
## COAST CAP
Sixteen patients underwent catheterization between May 28, 2011 and January 14, 2013 and were enrolled at eight sites. All sixteen patients met the study eligibility criteria and were treated with the Bare CP Stent. Table 30 documents the patient accountability for the COAST CAP.
Table 30. Patient Accountability (COAST CAP)
| | Possible N (100%) | 1 Month Visit n (%) | 12 Month Visit n (%) | 24 Month Visit n (%) |
| --- | --- | --- | --- | --- |
| Safety Cohort | 16 | 15 | 13 | 8 |
| Effectiveness Cohort | 16 | 15 | 13 | 8 |
The small number of patients at the 24-month time point mostly represents those who have yet to go through their 2-year window. Data collection is ongoing for the follow-up visits beyond 24 months, namely at 3, 4 and 5 years.
PMA P150028: FDA Summary of Safety and Effectiveness Data
Page 38
{38}
COAST II CAP
Forty-five patients underwent catheterizations between April 11, 2012 and August 16, 2013. Table 31 documents the patient accountability for the COAST II CAP.
Table 31. Patient Accountability (COAST II CAP)
| | Possible N (100%) | 1 Month Visit n (%) | 12 Month Visit n (%) | 24 Month Visit n (%) |
| --- | --- | --- | --- | --- |
| Safety Cohort | 45 | 45 | 40 | 16 |
| Effectiveness Cohort | 45 | 45 | 40 | 16 |
As noted for the COAST CAP, the small number of patients at the 24 month time point mostly represents those who have yet to go through their 2 year window. Data collection is ongoing for the follow-up visits beyond 24 months, namely at 3, 4 and 5 years.
## C. Study Population Demographics and Baseline Parameters
COAST CAP
The demographics of the COAST CAP population are shown in Table 32.
Table 32. Patient Demographics and Baseline Characteristics (COAST CAP)
| | Number (Percent) or Median (Range) |
| --- | --- |
| | Continued Access (n=16) |
| Gender | |
| Male | 12 (75%) |
| Female | 4 (25%) |
| Age, years | 16 (9 to 47) |
| Age Group, years | |
| 7 to 13 | 3 (19%) |
| 14 to 17 | 7 (44%) |
| 18 to 29 | 2 (13%) |
| 30 to 60 | 4 (25%) |
| Weight, kg | 68 (36 to 80) |
| Height, cm | 168 (141 to 185) |
| NYHA Classification | |
| I | 12 (75%) |
| II | 3 (19%) |
| III | 1 (6%) |
| Primary Indication | |
| Native coarctation | 9 (56%) |
| Recurrent coarctation | 7 (44%) |
PMA P150028: FDA Summary of Safety and Effectiveness Data
{39}
COAST II CAP
The demographics of the COAST II CAP population are shown in Table 33.
Table 33. Patient Demographics and Baseline Characteristics (COAST II CAP)
| | Number (Percent) or Median (Range) |
| --- | --- |
| | Continued Access (n=45) |
| Gender | |
| Male | 29 (64%) |
| Female | 16 (36%) |
| Age, years | 21 (7 to 65) |
| Age Group, years | |
| <10 | 3 (7%) |
| 10 to 13 | 5 (11%) |
| 14 to 17 | 9 (20%) |
| 18 to 29 | 10 (22%) |
| 30 to 60 | 17 (38%) |
| >60 | 1 (2%) |
| Weight, kg | 69 (23 to 142) |
| Height, cm | 164 (127 to 192) |
| Primary Indication | |
| Repair of aortic wall injury | 19 (42%) |
| Prevention of aortic wall injury | 26 (58%) |
## D. Safety and Effectiveness Results
### 1. Safety Endpoints
Table 34 details the summary of data pertaining to the safety endpoints for the COAST CAP and COAST II CAP.
PMA P150028: FDA Summary of Safety and Effectiveness Data
{40}
Table 34. COAST CAP and COAST II CAP Safety Endpoints
| | Safety Endpoint | N (Event Rate) |
| --- | --- | --- |
| COAST Primary | Serious or Somewhat Serious Adverse event attributed to the Stent or Implantation procedure within 30 days of the procedure | 1 (6.3%) |
| Primary | Post-procedure paradoxical hypertension | 0 (0%) |
| COAST II Primary | Serious or Somewhat Serious Adverse Events Attributed to the Stent, Implantation or Catheterization within 30 days of the procedure | 3 (6.7%) |
| Secondary | Proportion of patients experiencing any AEs related to the device or implant procedure post 1 year (among 42 patients followed for at least 1 year) | 0 (0%) |
An overview of the adverse events observed in the COAST CAP and COAST II CAP is presented in Table 35…
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.