← Product Code [LMH](/productcode/LMH) · P140029

# Restylane Refyne, Restylane Defyne (P140029)

_Q-Med AB · LMH · Dec 9, 2016 · General, Plastic Surgery · APPR_

**Canonical URL:** https://fda.innolitics.com/device/P140029

## Device Facts

- **Applicant:** Q-Med AB
- **Product Code:** [LMH](/productcode/LMH.md)
- **Decision Date:** Dec 9, 2016
- **Decision:** APPR
- **Device Class:** Class 3
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

Restylane Refyne is indicated for injection into the mid-to-deep dermis for the correction of moderate to severe facial wrinkles and folds (such as nasolabial folds) in patients over the age of 21. Restylane Defyne is indicated for injection into the mid-to-deep dermis for the correction of moderate to severe deep facial wrinkles and folds (such as nasolabial folds) in patients over the age of 21.

## Device Story

Restylane Refyne and Defyne are sterile, cross-linked sodium hyaluronate (HA) gels (20 mg/mL) containing 3 mg/mL lidocaine hydrochloride for pain relief. The device is injected into the mid-to-deep dermis by a physician to provide structural support and correct facial wrinkles/folds. The gel acts as a physical filler to replace lost volume or support underlying structures. The device is supplied in a pre-filled syringe with specific needle gauges (30G for Refyne, 27G for Defyne). Clinical outcomes are assessed by clinicians using the Wrinkle Severity Rating Scale (WSRS) and by patients using the Subject Self-Assessment (SSA) scale. The inclusion of lidocaine provides immediate pain relief during the injection procedure. The device benefits patients by improving facial appearance and reducing wrinkle severity, with effects lasting through 48 weeks in clinical studies.

## Clinical Evidence

Two prospective, randomized, evaluator-blinded, active-controlled, split-face US pivotal studies (N=170 for Refyne, N=162 for Defyne) compared the devices to non-lidocaine HA fillers. Primary endpoint: mean change in WSRS score from baseline at 24 weeks. Both devices demonstrated non-inferiority to control (lower limit of 95% CI > -0.4). Safety profile showed common injection-site reactions (erythema, swelling, bruising, pain) consistent with HA fillers. No serious device-related adverse events reported. Supplemental European studies confirmed non-inferiority.

## Technological Characteristics

Sterile, cross-linked sodium hyaluronate gel (20 mg/mL) with 3 mg/mL lidocaine hydrochloride in phosphate buffered saline (pH 7). Differentiated by degree of cross-linking and extrusion grid size. Supplied in 1 mL pre-filled syringes. Sterilized via moist heat (autoclave) per ISO 17665-1 (SAL 10^-6). Biocompatibility per ISO 10993.

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a PMA](https://innolitics.com/services/regulatory/), [a 510(k)](https://innolitics.com/services/510ks/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

SUMMARY OF SAFETY AND EFFECTIVENESS DATA (SSED)

I. GENERAL INFORMATION

Device Generic Name: Injectable Dermal Filler

Device Trade Name: Restylane® Defyne, Restylane® Refyne

Device Procode: LMH

Applicant’s Name and Address: Q-Med AB
Seminariegatan 21
Uppsala S-75228 Sweden

Date(s) of Panel Recommendation: None

Premarket Approval Application (PMA) Number: P140029

Date of FDA Notice of Approval: December 9, 2016

II. INDICATIONS FOR USE

- Restylane Refyne is indicated for injection into the mid-to-deep dermis for the correction of moderate to severe facial wrinkles and folds (such as nasolabial folds) in patients over the age of 21.
- Restylane Defyne is indicated for injection into the mid-to-deep dermis for the correction of moderate to severe deep facial wrinkles and folds (such as nasolabial folds) in patients over the age of 21.

III. CONTRAINDICATIONS

- Contraindicated for patients with severe allergies manifested by a history of anaphylaxis or history or presence of multiple severe allergies.
- Contains trace amounts of gram-positive bacterial proteins, and is contraindicated for patients with a history of allergies to such material.
- Contains lidocaine and is contraindicated for patients with a history of allergies to such material.

IV. WARNINGS AND PRECAUTIONS

The warnings and precautions can be found in the Restylane Refyne and Restylane Defyne physician labeling.

V. DEVICE DESCRIPTION

PMA P140029: FDA Summary of Safety and Effectiveness Data
Page 1

{1}

Restylane Refyne and Restylane Defyne are sterile transparent gels composed of cross-linked sodium hyaluronate of bacterial origin. The products have a hyaluronic acid (HA) concentration of 20 mg/mL in phosphate buffered saline at pH 7 and contain 3 mg/mL lidocaine hydrochloride added to the gel to achieve a pain-relieving effect. The action of the drug substance is ancillary to that of the device. Restylane Refyne and Restylane Defyne are differentiated by their degree of cross linking and the calibration size of the grid used to extrude the gel to a defined size. The products are supplied in a sealed Tyvek blister package containing a syringe with 1 mL sterile gel and two needles (30 G½ for Restylane Refyne and 27 G½ for Restylane Defyne).

## VI. ALTERNATIVE PRACTICES AND PROCEDURES

There are several other alternatives for the correction of moderate to severe facial wrinkles and folds (such as nasolabial folds). Patients frequently seek correction of facial contour deformities that are: (1) age-related loss of facial fat or weakening of underlying supportive structures; (2) sun damaged skin; or, (3) related to specific diseases or their treatments that may cause facial wasting, scarring, or structural damage (e.g., prior surgery, acne vulgaris). Treatment of photo-damaged skin, with its associated wrinkling and changes in texture and pigmentation, is often accomplished by use of topical creams (e.g., retinoids), chemical peeling procedures, or laser resurfacing. Deeper wrinkles, folds, scars, and other depressed lesions are often treated with surgery (e.g., rhytidectomy), botulinum toxin Type A injections, or by implantation of dermal filler substances (e.g., injection of collagen, other FDA approved soft tissue fillers, or autologous fat). Each alternative has its own advantages and disadvantages. A patient should fully discuss these alternatives with his/her physician to select the method that best meets expectations and lifestyle.

## VII. MARKETING HISTORY

Restylane Refyne and Restylane Defyne were approved for marketing and sale in Europe in January 2010. The products are approved in multiple countries globally; Argentina, Australia, Brazil, Canada, Chile, Colombia, Hong Kong, India, Israel, Korea, Kuwait, Mexico, New Zealand, Philippines, Serbia, Turkey and United Arab Emirates. The devices have not been withdrawn from marketing in any country for any reason related to the safety or effectiveness of the devices.

## VIII. POTENTIAL ADVERSE EFFECTS OF THE DEVICE ON HEALTH

Below is a list of the potential adverse effects (e.g., complications) associated with the use of the device.

As of September 2014, the following post market adverse events have been reported for the Restylane Refyne or Restylane Defyne since release to markets outside the U.S. in 2011; atrophy/scaring, bruising, capillary disorders such as telangiectasia, dermatitis, discoloration, erythema, eye pain, eye swelling, granuloma, hypersensitivity, induration, infection, inflammation, ischemia, mass, neurological symptoms such as hypoesthesia,

PMA P140029: FDA Summary of Safety and Effectiveness Data
Page 2

{2}

paraesthesia, pain/tenderness, papules/nodules, pruritus, reactivation of herpes infection, short duration of effect, swelling and urticaria.

Other potential adverse events that have been reported following injection of HA gels in general and may occur when using the product include the following: abscess, acne, angioedema, blisters, device dislocation, fistula, necrosis, rash, and visual disturbance.

Vascular compromise may occur due to an inadvertent intravascular injection or as a result of vascular compression associated with implantation of any injectable product. This may manifest as ischemia or necrosis at the implant site or in the area supplied by the blood vessels affected; or rarely as ischemic events in other organs due to embolization.

Vision abnormalities including blindness have been reported following injection of dermal fillers including HA, with and without lidocaine, into the nose, glabella, periorbital areas, and/or cheek, with a time to onset ranging from immediate to a few days following injection. Reported treatments include anticoagulant, epinephrine, aspirin, hyaluronidase, corticosteroid treatment and hyperbaric oxygen. Outcomes ranged from resolved to ongoing at the time of last contact. Events requiring medical intervention, and events where resolution information is not available were reported after injection of HA with or without lidocaine.

There has been an isolated case report and the scientific literature also refers to rare events of cerebral infarction after treatment with dermal fillers.

For the specific adverse events that occurred in the clinical studies, please see Section X below.

## IX. SUMMARY OF NONCLINICAL STUDIES

### A. Laboratory Studies

Restylane Refyne and Restylane Defyne have been extensively tested and characterized through physical and chemical analyses. To ensure that Restylane Refyne and Restylane Defyne degrade naturally during their clinical lifespan, degradation assays have also been performed.

Table 1: Summary of Key Bench Testing on Restylane Refyne and Restylane Defyne

|  Test | Purpose | Results  |
| --- | --- | --- |
|  Extrusion force (N) | Ensures extrusion force meets specification | Passed  |
|  pH | Ensures pH meets specification | Passed  |
|  Rheology (tan δ) | Ensures rheological properties meet specification | Passed  |
|  HA concentration (mg/mL) | Ensures HA concentration meets specification | Passed  |
|  Gel content (%) | Ensures gel content meets specification | Passed  |
|  Lidocaine concentration (mg/mL) | Ensures lidocaine concentration meets specification | Passed  |
|  Residual crosslinker (ppm) | Ensures residual crosslinker meets specification | Passed  |

PMA P140029: FDA Summary of Safety and Effectiveness Data

{3}

|  Endotoxin (EU/mL) | Ensures endotoxin meets specification | Passed  |
| --- | --- | --- |
|  Sterility | Ensures device is sterile | Passed  |

## B. Biocompatibility Studies

According to ISO 10993-1, Biological evaluation of medical devices, Restylane Refyne and Restylane Defyne are categorized as implant devices in contact with tissue where contact duration is more than 30 days. In order to evaluate the biological safety of the products, a biocompatibility program was defined and performed taking similarities and differences between the products into consideration. Testing has been conducted on Restylane Refyne and/or Restylane Defyne as outlined in Table 2 below, including multiple long-term implantation studies evaluating long-term tolerability and device degradation.

Table 2: Summary of Biocompatibility Testing

|  Test | Method | ISO standard | Results  |
| --- | --- | --- | --- |
|  Cytotoxicity | Agar overlay | 10993-5 | Not cytotoxic  |
|  Sensitization | Guinea pig maximization test | 10993-10 | Not sensitizing  |
|  Irritation/intracutaneous reactivity | Evaluated as part of implantation study |  | Well tolerated  |
|  Subchronic toxicity | Evaluated as part of implantation study |  | Well tolerated  |
|  Acute systemic toxicity | Intraperitoneal injection in mice | ISO 10993-11 | Not systemically toxic  |
|  Pyrogenicity | Rabbit pyrogen study |  | Non-pyrogenic  |
|  Genotoxicity | Bacterial reverse mutation, Micronucleus, and Mouse lymphoma | ISO 10993-3 | Not mutagenic and not genotoxic  |
|  Intradermal implantation | Multiple rabbit studies (3 years, 2.5 years, 1.5 years) | ISO 10993-6 | Well tolerated  |

## C. Additional Studies

Filled syringes are sterilized using a validated moist heat process in a pressurized autoclave. The sterilization cycle has been validated according to ISO 17665-1. The validated sterilization cycle provides a minimum Sterility Assurance Level (SAL) of $10^{-6}$.

A shelf life of 18 months for Restylane Refyne and 24 months for Restylane Defyne has been established. For Resylane Refyne, the shelf life is based on stability data collected through 18-24 months at $25^{\circ}\mathrm{C} / 40\%$ and $30^{\circ}\mathrm{C} / 35\%$ relative humidity, and through 6 months at $40^{\circ}\mathrm{C} / 25\%$ relative humidity. For Restylane Defyne, the shelf life is based on stability data collected through 24 months at $25^{\circ}\mathrm{C} / 40\%$ and $30^{\circ}\mathrm{C} / 35\%$ relative humidity, and through 6 months at $40^{\circ}\mathrm{C} / 25\%$ relative humidity. At each time point, product was characterized via microbiological, physical, chemical, lidocaine hydrochloride content, and lidocaine-related degradant parameters. Conformance of real-time aged product with all specifications was confirmed.

PMA P140029: FDA Summary of Safety and Effectiveness Data

{4}

PMA P140029: FDA Summary of Safety and Effectiveness Data
Page 5

# X. SUMMARY OF PRIMARY CLINICAL STUDIES

The applicant performed two clinical studies to establish a reasonable assurance of safety and effectiveness of injection into the mid-to-deep dermis with Restylane Refyne or Restylane Defyne for the correction of moderate to severe facial wrinkles and folds (such as nasolabial folds, or NLFs) in patients over the age of 21 in the US under IDE # G090150. Data from these clinical studies were the basis for the PMA approval decision. A summary of the clinical studies are presented below.

## A. Study Design

Patients were treated between September 23, 2010 and January 19, 2012 for the Refyne study, and between September 27, 2010 and January 12, 2012 for the Defyne study. The databases for this PMA reflected data collected through January 19, 2012 and included 170 patients for the Refyne study, and 162 patients for the Defyne study. There were 12 investigational sites for the Refyne study and 11 investigational sites in the Defyne study.

Both studies were nearly identical in design. The two studies were both randomized, evaluator-blinded, active-controlled, multi-center, split-face clinical studies. For each study, the Control treatment was an FDA approved HA soft tissue filler, with similar characteristics to either Restylane Refyne or Restylane Defyne, respectively. Both control products did not contain lidocaine and are legally marketed alternatives with similar indications for use.

### 1. Key Clinical Inclusion and Exclusion Criteria

Enrollment in both pivotal studies were limited to patients who met the following key inclusion criteria:

- The subject was a male or female 18 years of age or older. (Note: all patients who enrolled were over the age of 21.)
- The subject had bilateral NLFs that, in the opinion of both the Blinded Evaluating Investigator and the Unblinded Injecting Investigator, could be corrected with an injectable dermal implant.
- The subject had the same WSRS score of 3 or 4 (moderate or severe) for each NLF.

Patients were not permitted to enroll in either study if they met any of the following key exclusion criteria:

- The subject had active skin disease or inflammation on or near an NLF that, in the Principal Investigator’s opinion, would interfere with the study device injections and/or study assessments.
- The subject had a history of sensitivity to HA.

{5}

- The subject had a history of sensitivity to lidocaine or other amide-type anesthetics.
- The subject was, in the Investigator’s opinion, at undue risk based on the precautions, warnings, and contraindications for local lidocaine anesthetics.
- The subject was pregnant, nursing, or planning to conceive a child in the next year.
- The subject had undergone soft tissue augmentation in an NLF within a certain time frame (6 months: bovine collagen; 12 months: porcine or human collagen; 18 months: HA or hydroxylapatite).
- The subject had undergone soft tissue augmentation with any of the following at any time: autologous fat, polymethylmethacrylate or other acrylates, polyacrylamide, polyethylene oxide, polylactic acid, liquid silicone, or other similar implant materials.
- The subject had in the previous 6 months undergone, or planned to undergo, certain aesthetic treatment procedures that would interfere with the treatment area (such as facelift, laser resurfacing, chemical peel, botulinum toxin A injections, etc.).

2. Follow-up Schedule

In both pivotal studies, subjects underwent initial treatment to the NLFs and one touch-up treatment as appropriate after 3 weeks. Another injectable HA gel was included as control treatment as previously specified. The follow-up period consisted of safety and effectiveness follow-up visits at 1 week after the initial injection, 3 weeks after the initial or touch-up injections, and 12, 24, 36, and 48 weeks after the final initial injection (defined as the latter of the initial injection or initial touch-up injection).

At 48 weeks after the final initial injection, in an open label phase of both studies, an optional retreatment injection of one or both NLFs was performed with either Restylane Refyne or Restylane Defyne, as appropriate. A retreatment touch-up injection was administered as appropriate 3 weeks after the retreatment injection. The subjects were then followed-up for 4 weeks after the final retreatment injection (defined as the latter of the retreatment injection or retreatment touch-up injection).

For each treatment, the Unblinded Injecting Investigator determined the appropriate product injection technique and injection volume. Subjects received a maximum of two injections per side of the face at each treatment, and completed a safety diary for 3 weeks after each treatment.

3. Clinical Endpoints

With regards to safety, in both studies adverse outcomes were evaluated by comparing the incidence and intensity of clinical events reported in patient diaries during the 21 days after treatment, and the adverse events assessed during study visits by the investigator.

PMA P140029: FDA Summary of Safety and Effectiveness Data
Page 6

{6}

With regards to effectiveness, in both studies the primary effectiveness endpoint was the change from baseline in the Wrinkle Severity Rating Scale (WSRS) score at 24 weeks after the final initial injection. The WSRS is a 5-point scale (Table 3) and each side was evaluated by a Blinded Evaluating Investigator.

Table 3: 5-Point Wrinkle Severity Rating Scale (WSRS)

|  WSRS Score | Description  |
| --- | --- |
|  1 | Absent: No visible fold; continuous skin line.  |
|  2 | Mild: Shallow but visible fold with a slight indentation; minor facial feature; implant is expected to produce a slight improvement in appearance.  |
|  3 | Moderate: Moderately deep folds; clear facial feature visible at normal appearance but not when stretched; excellent correction is expected from injectable implant.  |
|  4 | Severe: Very long and deep folds; prominent facial feature; less than 2 mm visible fold when stretched; significant improvement is expected from injectable implant.  |
|  5 | Extreme: Extremely deep and long folds, detrimental to facial appearance; 2 to 4 mm visible V-shaped fold when stretched; unlikely to have satisfactory correction with injectable implant alone.  |

Secondary effectiveness endpoints, for both studies, included:

- Change from baseline at each visit after the final initial injection in WSRS score;
- Change from baseline at each visit after the final initial injection in Subject Self Assessment (SSA) score (see Table 4 below);
- WSRS Response rate, defined as the percentage of subjects with an at least 1-grade improvement in WSRS score at each visit after the final initial injection;
- SSA Response rate, defined as the percentage of subjects with an at least 1-grade improvement at each visit after the final initial injection in SSA score;
- Relative WSRS Response and Relative SSA Response (defined by 3 response categories: test better than control, test same as control, or test worse than control), and
- Subject's pain assessment using 11-point Numeric Pain Intensity Scale (NPIS) at 0, 15, 30, 45, 60 minutes, and 24 hours after the injection (Initial Injection and if appropriate, the Initial Touch-up Injection).

Table 4: Subject Self-Assessment (SSA) of Wrinkle Severity Scale

|  SSA Score | Description  |
| --- | --- |
|  1 | Absent: No visible fold; continuous skin line.  |
|  2 | Mild: Shallow but visible fold with a slight indentation.  |
|  3 | Moderate: Moderately deep folds.  |
|  4 | Severe: Very long and deep fold.  |
|  5 | Extreme: Extremely deep and long folds.  |

With regard to success/failure criteria, an individual subject was defined as a responder if he/she had at least a 1-grade improvement in WSRS score at each visit after the final initial injection. With regard to the success of each study, a change in

PMA P140029: FDA Summary of Safety and Effectiveness Data

{7}

WSRS score of 0.4 was considered in the statistical test between groups (i.e., a 95% confidence interval approach with a non-inferiority margin of -0.4 for difference in WSRS score between groups).

## B. Accountability of PMA Cohort

Restylane Refyne: A total of 170 subjects were treated. At study completion, 21 randomized subjects had prematurely discontinued the study; 14 were due to subject request, 1 was due to a study protocol violation, 4 subjects were lost to follow-up, 1 subject was due to noncompliance to study protocol procedures, and 1 subject took unpermitted medication and was therefore ineligible for additional treatment. No subject discontinued due to an adverse event.

Restylane Defyne: A total of 162 subjects were treated. At study completion, 26 subjects had prematurely discontinued the study. 4 subjects discontinued the study due to adverse events; 1 subject with a lung cancer, 1 subject with a right hip injury, both unrelated to treatment, 1 subject with granulomatous dermatitis, 1 subject with allergic contact dermatitis, both related to treatment. Of the remaining discontinued subjects, 11 were due to subject's request, 9 subjects were lost to follow-up, 1 subject due to a study protocol violation, and 1 subject due to noncompliance to study protocol procedures.

Table 5: Restylane Refyne Study Summary of Subject Enrollment

|  Population | Restylane Refyne | Control | All Subjects  |
| --- | --- | --- | --- |
|  Screen Failure (Not Randomized) | - | - | 39  |
|  - Inclusion/Exclusion criteria not fulfilled | - | - | 24  |
|  - Subject’s request | - | - | 1  |
|  - Other | - | - | 14  |
|  Randomization | 84 | 87 | 171^{a}  |
|  Intent-to-Treat (ITT) | 84 | 86 | 170  |
|  Per-Protocol (PP) | 75 | 79 | 154  |
|  Safety | 84 | 86 | 170  |

a One subject discontinued just after randomization and did not receive any injection (excluded from ITT, PP, and Safety Populations).

Table 6: Restylane Defyne Study Summary of Subject Enrollment

|  Population | Restylane Defyne | Control | All Subjects  |
| --- | --- | --- | --- |
|  Screen Failure (Not Randomized) | - | - | 22  |
|  - Inclusion/Exclusion criteria not fulfilled | - | - | 20  |
|  - Subject’s request | - | - | 1  |
|  - Other | - | - | 1  |
|  Randomization | 80 | 82 | 162  |
|  Intent-to-Treat (ITT) | 80 | 82 | 162  |
|  Per-Protocol (PP) | 68 | 74 | 142  |
|  Safety | 80 | 82 | 162  |

PMA P140029: FDA Summary of Safety and Effectiveness Data

{8}

# C. Study Population Demographics and Baseline Parameters

The demographics of the study population are typical for a pivotal study performed in the US. Subject demographics for the participants in both studies are presented in Table 7 below.

Table 7: Subject Demographics for Both Studies

|   | Restylane Refyne study (N=170) | Restylane Defyne study (N=162)  |
| --- | --- | --- |
|  Gender, n (%)  |   |   |
|  Male | 10 (5.9) | 6 (3.7)  |
|  Female | 160 (94.1) | 156 (96.3)  |
|  Age (Years)  |   |   |
|  Mean | 53.9 | 53.7  |
|  Standard Deviation | 8.22 | 8.68  |
|  Median | 54.5 | 53.0  |
|  Minimum, Maximum | 26, 77 | 34, 75  |
|  Age range, n (%)  |   |   |
|  18 to 64 | 159 (93.5) | 146 (90.1)  |
|  65 and above | 11 (6.5) | 16 (9.9)  |
|  Ethnicity, n (%)  |   |   |
|  Caucasian | 116 (68.2) | 129 (79.6)  |
|  Black | 15 (8.8) | 20 (12.3)  |
|  Hispanic | 38 (22.4) | 10 (6.2)  |
|  Other | 1 (0.6) | 3 (1.9)  |
|  Fitzpatrick Skin Type, n (%)  |   |   |
|  I | 6 (3.5) | 6 (3.7)  |
|  II | 39 (22.9) | 28 (17.3)  |
|  III | 52 (30.6) | 67 (41.4)  |
|  IV | 55 (32.4) | 34 (21.0)  |
|  V | 9 (5.3) | 19 (11.7)  |
|  VI | 9 (5.3) | 8 (4.9)  |
|  Baseline Wrinkle Severity Rating Score (WSRS)*, n (%)  |   |   |
|  Moderate (3) | 93 (54.7) | 118 (72.8)  |
|  Severe (4) | 77 (45.3) | 44 (27.2)  |
|  Baseline Subject Self-Assessment Score (SSA)*, n (%)  |   |   |
|  Mild (2) | 3 (1.8) | 5 (3.1)  |
|  Moderate (3) | 76 (44.7) | 101 (62.3)  |
|  Severe (4) | 76 (44.7) | 45 (27.8)  |
|  Extreme (5) | 15 (8.8) | 11 (6.8)  |

* Data for nasolabial folds treated with Restylane Refyne or Restylane Defyne only

PMA P140029: FDA Summary of Safety and Effectiveness Data

{9}

Although the inclusion criteria specified a minimum of 18 years of age, all patients enrolled across both studies were over the age of 21 (that is, 22 years of age and older).

**Ethnic Representation**: The majority of patients enrolled in the studies were Caucasian (68.2% in the Refyne study and 79.6% in the Defyne study). Forty-three (43) percent of patients had a Fitzpatrick Skin Type of IV, V or VI in the Refyne study, and 38 percent in the Defyne study.

**Treatment Material Administered**: In the Restylane Refyne study, the mean total volume injected per NLF for the treatment sessions (initial and touch-up) was 1.52 mL (range 0.5 to 4.0 mL) for the Restylane Refyne side and 1.34 mL (range 0.5 to 3.2 mL) for the Control side. The corresponding mean total volumes for the retreatment sessions (initial and touch-up) was 0.95 mL (range 0.1 to 2.5 mL) for the Restylane Refyne side and 0.89 mL (range 0.1 to 2.2 mL) for the Control side.

In the Restylane Defyne study, the mean total volume injected per NLF for the treatment sessions (initial and touch-up) was 1.45 mL (range 0.4 to 3.3 mL) for the Restylane Defyne side and 1.39 mL (range 0.3 to 3.7 mL) for the Control side. The corresponding mean total volumes for the retreatment sessions (initial and touch-up) was 0.74 mL (range 0.1 to 2.1 mL) for the Restylane Defyne side and 0.70 mL (range 0.2 to 1.7 mL) for the Control side.

## D. Safety and Effectiveness Results

### 1. Safety Results

The analysis of safety was based on the cohort of 170 (Refyne study) and 162 (Defyne study) patients available for the up to 48 week evaluation. The key safety outcomes for this study are presented below in Tables 8 to 11. Adverse effects are reported in Tables 12 and 13.

### Summary of Observations Recorded in Subject Diaries

In a randomized order, subjects received in one side of the face either Restylane Refyne (Refyne study) or Restylane Defyne (Defyne study), and in the other side of the face a control HA gel. If necessary, a touch-up injection was administered 3 weeks after the initial injection. Subjects recorded their observations of treatment responses for each side of the face in pre-printed diaries once each day during the first 21 days following each treatment. The diaries included check boxes for commonly expected treatment responses, e.g., redness, swelling, bruising, lump/bump formation, pain/tenderness, or itching, at the injection/application sites. An "Other" category was included in the diary for the subject to report any sign/symptom not covered by the previously mentioned responses, and this category was followed-up as an adverse event. A diary was completed for each initial and subsequent touch-up treatment. It should be noted that the study subjects were required to record the presence and level of intensity for each observed treatment response as "Mild," "Moderate," "Severe," or "None." A

PMA P140029: FDA Summary of Safety and Effectiveness Data
Page 10

{10}

summary of the maximum intensity and duration of the subject observations is presented in Table 8 through Table 11 on the following pages.

Table 8: Restylane Refyne versus Control Injection Site Responses By Maximum Intensity (Number / % of Subject NLFs)

|   | Restylane Refyne n$ (%) |   |   |   | Control** n$ (%)  |   |   |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
|   | Mild | Moderate | Severe | Total | Mild | Moderate | Severe | Total  |
|  Post-Initial Injection* (N= 168 NLFs for Restylane Refyne and N= 168 NLFs for Control**)  |   |   |   |   |   |   |   |   |
|  Redness | 60 (35.7) | 24 (14.3) | 9 (5.4) | 93 (55.4) | 62 (36.9) | 22 (13.1) | 14 (8.3) | 98 (58.3)  |
|  Swelling | 66 (39.3) | 30 (17.9) | 5 (3.0) | 101 (60.1) | 52 (31.0) | 44 (26.2) | 6 (3.6) | 102 (60.7)  |
|  Bruising | 49 (29.2) | 29 (17.3) | 18 (10.7) | 96 (57.1) | 51 (30.4) | 31 (18.5) | 14 (8.3) | 96 (57.1)  |
|  Lump/Bump Formation | 39 (23.2) | 26 (15.5) | 10 (6.0) | 75 (44.6) | 45 (26.8) | 27 (16.1) | 11 (6.5) | 83 (49.4)  |
|  Pain/Tenderness | 58 (34.5) | 11 (6.5) | 3 (1.8) | 72 (42.9) | 65 (38.7) | 19 (11.3) | 2 (1.2) | 86 (51.2)  |
|  Itching | 21 (12.5) | 3 (1.8) | 1 (0.6) | 25 (14.9) | 28 (16.7) | 1 (0.6) | 0 | 29 (17.3)  |
|  Post-Initial Touch-up Injection* (N= 89 NLFs for Restylane Refyne and N= 85 NLFs for Control**)  |   |   |   |   |   |   |   |   |
|  Redness | 23 (25.8) | 9 (10.1) | 3 (3.4) | 35 (39.3) | 23 (27.1) | 9 (10.6) | 4 (4.7) | 36 (42.4)  |
|  Swelling | 23 (25.8) | 12 (13.5) | 3 (3.4) | 38 (42.7) | 22 (25.9) | 12 (14.1) | 4 (4.7) | 38 (44.7)  |
|  Bruising | 23 (25.8) | 9 (10.1) | 2 (2.2) | 34 (38.2) | 21 (24.7) | 6 (7.1) | 3 (3.5) | 30 (35.3)  |
|  Lump/Bump Formation | 15 (16.9) | 10 (11.2) | 3 (3.4) | 28 (31.5) | 18 (21.2) | 11 (12.9) | 5 (5.9) | 34 (40.0)  |
|  Pain/Tenderness | 16 (18.0) | 3 (3.4) | 0 | 19 (21.3) | 23 (27.1) | 7 (8.2) | 0 | 30 (35.3)  |
|  Itching | 12 (13.5) | 1 (1.1) | 0 | 13 (14.6) | 9 (10.6) | 1 (1.2) | 1 (1.2) | 11 (12.9)  |
|  * Number of subjects who completed subject diaries. ** A legally marketed injectable HA gel without lidocaine $ Number of subject NLFs with each specific injection site response  |   |   |   |   |   |   |   |   |

PMA P140029: FDA Summary of Safety and Effectiveness Data

{11}

Table 9: Restylane Defyne versus Control Injection Site Responses by Maximum Intensity (Number / % of Subject NLFs)

|   | Restylane Defyne n$ (%) |   |   |   | Control** n$ (%)  |   |   |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
|   |  Mild | Moderate | Severe | Total | Mild | Moderate | Severe | Total  |
|  Post-Initial Injection* (N= 160 NLFs for Restylane Defyne and N= 160 NLFs for Control**)  |   |   |   |   |   |   |   |   |
|  Redness | 66 (41.3) | 27 (16.9) | 8 (5.0) | 101 (63.1) | 65 (40.6) | 27 (16.9) | 9 (5.6) | 101 (63.1)  |
|  Swelling | 60 (37.5) | 45 (28.1) | 14 (8.8) | 119 (74.4) | 61 (38.1) | 42 (26.3) | 15 (9.4) | 118 (73.8)  |
|  Bruising | 48 (30.0) | 36 (22.5) | 11 (6.9) | 95 (59.4) | 47 (29.4) | 37 (23.1) | 13 (8.1) | 97 (60.6)  |
|  Lump/Bump Formation | 51 (31.9) | 39 (24.4) | 15 (9.4) | 105 (65.6) | 43 (26.9) | 37 (23.1) | 14 (8.8) | 94 (58.8)  |
|  Pain/Tenderness | 64 (40.0) | 28 (17.5) | 8 (5.0) | 100 (62.5) | 60 (37.5) | 23 (14.4) | 11 (6.9) | 94 (58.8)  |
|  Itching | 28 (17.5) | 9 (5.6) | 3 (1.9) | 40 (25.0) | 24 (15.0) | 8 (5.0) | 1 (0.6) | 33 (20.6)  |
|  Post-Initial Touch-up Injection* (N= 84 NLFs for Restylane Defyne and N= 79 NLFs for Control**)  |   |   |   |   |   |   |   |   |
|  Redness | 28 (33.3) | 6 (7.1) | 1 (1.2) | 35 (41.7) | 30 (38.0) | 5 (6.3) | 2 (2.5) | 37 (46.8)  |
|  Swelling | 32 (38.1) | 9 (10.7) | 3 (3.6) | 44 (52.4) | 22 (27.8) | 12 (15.2) | 7 (8.9) | 41 (51.9)  |
|  Bruising | 21 (25.0) | 8 (9.5) | 1 (1.2) | 30 (35.7) | 19 (24.1) | 8 (10.1) | 3 (3.8) | 30 (38.0)  |
|  Lump/Bump Formation | 24 (28.6) | 11 (13.1) | 3 (3.6) | 38 (45.2) | 23 (29.1) | 10 (12.7) | 6 (7.6) | 39 (49.4)  |
|  Pain/Tenderness | 30 (35.7) | 6 (7.1) | 1 (1.2) | 37 (44.0) | 27 (34.2) | 5 (6.3) | 2 (2.5) | 34 (43.0)  |
|  Itching | 7 (8.3) | 0 | 0 | 7 (8.3) | 2 (2.5) | 0 | 1 (1.3) | 3 (3.8)  |
|  * Number of subjects who completed subject diaries. ** A legally marketed injectable HA gel without lidocaine $ Number of subject NLFs with each specific injection site response  |   |   |   |   |   |   |   |   |

PMA P140029: FDA Summary of Safety and Effectiveness Data

{12}

Table 10: Restylane Refyne versus Control Duration of Injection Site Responses Post-Initial Injection (Number / % of Subject NLFs)

|  Injection site response | Restylane Refyne n$ (%) |   |   |   | Control** n$ (%)  |   |   |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  Duration£ | 1-3 Days | 4-7 Days | 8-14 Days | >14 Days | 1-3 Days | 4-7 Days | 8-14 Days | >14 Days  |
|  Post-Initial Injection* (N= 168 NLFs for Restylane Refyne and N= 168 NLFs for Control**)  |   |   |   |   |   |   |   |   |
|  Redness | 71 (42.3) | 13 (7.7) | 5 (3.0) | 4 (2.4) | 73 (43.5) | 18 (10.7) | 5 (3.0) | 2 (1.2)  |
|  Swelling | 58 (34.5) | 23 (13.7) | 14 (8.3) | 6 (3.6) | 55 (32.7) | 24 (14.3) | 17 (10.1) | 6 (3.6)  |
|  Bruising | 31 (18.5) | 30 (17.9) | 27 (16.1) | 8 (4.8) | 28 (16.7) | 35 (20.8) | 29 (17.3) | 4 (2.4)  |
|  Lump/bump Formation | 33 (19.6) | 19 (11.3) | 9 (5.4) | 14 (8.3) | 34 (20.2) | 21 (12.5) | 16 (9.5) | 12 (7.1)  |
|  Pain/tenderness | 53 (31.5) | 11 (6.5) | 8 (4.8) | 0 | 63 (37.5) | 19 (11.3) | 4 (2.4) | 0  |
|  Itching | 22 (13.1) | 2 (1.2) | 1 (0.6) | 0 | 26 (15.5) | 3 (1.8) | 0 | 0  |
|  Post-Initial Touch-up Injection* (N= 89 NLFs for Restylane Refyne and N= 85 NLFs for Control**)  |   |   |   |   |   |   |   |   |
|  Redness | 22 (24.7) | 10 (11.2) | 2 (2.2) | 1 (1.1) | 23 (27.1) | 6 (7.1) | 5 (5.9) | 2 (2.4)  |
|  Swelling | 26 (29.2) | 9 (10.1) | 2 (2.2) | 1 (1.1) | 26 (30.6) | 9 (10.6) | 3 (3.5) | 0  |
|  Bruising | 16 (18.0) | 13 (14.6) | 4 (4.5) | 1 (1.1) | 17 (20.0) | 10 (11.8) | 3 (3.5) | 0  |
|  Lump/bump Formation | 15 (16.9) | 6 (6.7) | 2 (2.2) | 5 (5.6) | 14 (16.5) | 11 (12.9) | 5 (5.9) | 4 (4.7)  |
|  Pain/tenderness | 16 (18.0) | 3 (3.4) | 0 | 0 | 27 (31.8) | 3 (3.5) | 0 | 0  |
|  Itching | 11 (12.4) | 1 (1.1) | 0 | 1 (1.1) | 7 (8.2) | 4 (4.7) | 0 | 0  |
|  * Number of subjects who completed subject diaries. ** A legally marketed injectable HA gel without lidocaine $ Number of subject NLFs with each specific injection site response £ Duration refers to number of days from symptom onset until resolution, irrespective of date of implantation.  |   |   |   |   |   |   |   |   |

PMA P140029: FDA Summary of Safety and Effectiveness Data

{13}

Table 11: Restylane Defyne vs. Control Duration of Injection Site Responses Post-Initial Injection (Number / % of Subject NLFs)

|  Injection site response | Restylane Defyne n$ (%) |   |   |   | Control** n$ (%)  |   |   |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  Duration£ | 1-3 Days | 4-7 Days | 8-14 Days | >14 Days | 1-3 Days | 4-7 Days | 8-14 Days | >14 Days  |
|  Post-Initial Injection* (N= 160 NLFs for Restylane Defyne and N= 160 NLFs for Control**)  |   |   |   |   |   |   |   |   |
|  Redness | 74 (46.3) | 19 (11.9) | 6 (3.8) | 2 (1.3) | 72 (45.0) | 22 (13.8) | 6 (3.8) | 1 (0.6)  |
|  Swelling | 61 (38.1) | 31 (19.4) | 22 (13.8) | 5 (3.1) | 57 (35.6) | 32 (20.0) | 20 (12.5) | 9 (5.6)  |
|  Brusing | 25 (15.6) | 42 (26.3) | 23 (14.4) | 5 (3.1) | 29 (18.1) | 32 (20.0) | 35 (21.9) | 1 (0.6)  |
|  Lump/bump Formation | 34 (21.3) | 35 (21.9) | 12 (7.5) | 24 (15.0) | 40 (25.0) | 15 (9.4) | 15 (9.4) | 24 (15.0)  |
|  Pain/tenderness | 63 (39.4) | 26 (16.3) | 7 (4.4) | 4 (2.5) | 58 (36.3) | 22 (13.8) | 11 (6.9) | 3 (1.9)  |
|  Itching | 29 (18.1) | 8 (5.0) | 2 (1.3) | 1 (0.6) | 24 (15.0) | 7 (4.4) | 2 (1.3) | 0  |
|  Post-Initial Touch-up Injection* (N= 84 NLFs for Restylane Defyne and N= 79 NLFs for Control**)  |   |   |   |   |   |   |   |   |
|  Redness | 25 (29.8) | 8 (9.5) | 0 | 2 (2.4) | 26 (32.9) | 8 (10.1) | 1 (1.3) | 2 (2.5)  |
|  Swelling | 26 (31.0) | 12 (14.3) | 4 (4.8) | 2 (2.4) | 20 (25.3) | 13 (16.5) | 6 (7.6) | 2 (2.5)  |
|  Brusing | 13 (15.5) | 11 (13.1) | 5 (6.0) | 1 (1.2) | 12 (15.2) | 9 (11.4) | 7 (8.9) | 2 (2.5)  |
|  Lump/bump Formation | 18 (21.4) | 4 (4.8) | 8 (9.5) | 8 (9.5) | 14 (17.7) | 9 (11.4) | 7 (8.9) | 9 (11.4)  |
|  Pain/tenderness | 30 (35.7) | 6 (7.1) | 1 (1.2) | 0 | 29 (36.7) | 4 (5.1) | 1 (1.3) | 0  |
|  Itching | 7 (8.3) | 0 | 0 | 0 | 3 (3.8) | 0 | 0 | 0  |
|  * Number of subjects who completed subject diaries.
** A legally marketed injectable HA gel without lidocaine
$ Number of subject NLFs with each specific injection site response
£ Duration refers to number of days from symptom onset until resolution, irrespective of date of implantation.  |   |   |   |   |   |   |   |   |

## Device and Injection Related Adverse Events

In the Restylane Refyne study, among the 170 treated subjects, 7.1% (12/170) experienced device- and injection-related AEs following initial and touch-up treatment, as well as retreatment and touch-up treatment. These subjects reported a total of 9 events related to control treatment and 10 events related to Restylane Refyne. The most common of treatment site responses was injection site erythema, which was reported for 4 subjects. Other treatment site responses (swelling, mass, hematoma, pruritus, and anesthesia) were reported for 1 subject each. Other related adverse events included presyncope (2 subjects), headache (1 subject), syncope (1 subject), and ecchymosis (1 subject). No related AEs led to discontinuation for any subject and no severe related AEs were reported during the study.

In the Restylane Defyne study, among the 162 treated subjects, 11.7% (19/162) experienced device- and injection-related AEs following initial and touch-up

PMA P140029: FDA Summary of Safety and Effectiveness Data

{14}

treatment, as well as retreatment and touch-up treatment. These subjects reported a total of 19 events related to control treatment and 20 events related to Restylane Defyne. The most common of treatment site responses was injection site swelling, which was reported for 5 subjects. Injection site erythema and injection site pain were reported for 4 and 2 subjects, respectively. Other treatment site responses (inflammation, hematoma, discomfort, and mass) were reported for 1 subject each. Other related adverse events included skin tightness, dermatitis allergic, interstitial granulomatous dermatitis, salivary hypersecretion, sensitivity of teeth, presyncope, and arterial stenosis which were all reported for 1 subject each. Two related AEs led to the discontinuation of 2 subjects (1 subject with interstitial granulomatous dermatitis, and 1 subject with allergic contact dermatitis). No severe related AEs were reported during the study.

Tabular data for related AEs is provided in Tables 12 and 13 below.

## Serious Adverse Events

A total of 2 subjects reported serious adverse events during the Restylane Refyne study. None was on the treated area and none was assessed as related to the study product.

A total of 7 subjects reported 8 serious adverse events during the Restylane Defyne study. None of the SAEs were in the treated area and none were assessed as related to the study products.

## Ethnicity

Subject Fitzpatrick skin type did not appear to have an effect on the distribution of AEs in either study, with similar incidence rates observed in subjects with both Skin Type subgroups I to III and IV to VI.

PMA P140029: FDA Summary of Safety and Effectiveness Data
Page 15

{15}

Table 12: Restylane Refyne Adverse Events Related to Study Product, Safety Population

|  System Organ Class/Preferred Terma | Restylane Refyne (N=170) n (%) | Control (N=170) n (%) | All Subjects (N=170) n (%)  |
| --- | --- | --- | --- |
|  Total Number of AE(s) | 10 | 9 | 16  |
|  Total Number (%) of Subjects with AE(s)b | 10 (5.9) | 8 (4.7) | 12 (7.1)  |
|  General Disorders And Administration Site Conditions | 6 (3.5) | 5 (2.9) | 8 (4.7)  |
|  Injection Site Erythema | 2 (1.2) | 4 (2.4) | 4 (2.4)  |
|  Injection Site Swelling | 1 (0.6) | 0 | 1 (0.6)  |
|  Injection Site Mass | 1 (0.6) | 0 | 1 (0.6)  |
|  Injection Site Hematoma | 1 (0.6) | 0 | 1 (0.6)  |
|  Injection Site Pruritus | 1 (0.6) | 0 | 1 (0.6)  |
|  Injection Site Anaesthesia | 0 | 1 (0.6) | 1 (0.6)  |
|  Nervous System Disorders | 3 (1.8) | 4 (2.4) | 4 (2.4)  |
|  Presyncope | 2 (1.2) | 2 (1.2) | 2 (1.2)  |
|  Headache | 0 | 1 (0.6) | 1 (0.6)  |
|  Syncope | 1 (0.6) | 1 (0.6) | 1 (0.6)  |
|  Skin And Subcutaneous Tissue Disorders | 1 (0.6) | 0 | 1 (0.6)  |
|  Ecchymosis | 1 (0.6) | 0 | 1 (0.6)  |

a Multiple occurrences within a System Organ Class by a subject were counted once per System Organ Class. Multiple occurrences of a Preferred Term by a subject were counted once per Preferred Term.
b A subject was counted once even if the subject experienced more than one AE during the study.
If the treated side where an AE occurred was not applicable, this AE was summarized under each treatment group.

PMA P140029: FDA Summary of Safety and Effectiveness Data

{16}

Table 13: Restylane Defyne Adverse Events Related to Study Product, Safety Population

|  System Organ Class/ Preferred Terma | Restylane Defyne (N=162) n, (%) | Control (N=162) n, (%) | All Subjects (N=162) n, (%)  |
| --- | --- | --- | --- |
|  Total Number of AE(s) | 20 | 19 | 37  |
|  Total Number (%) of Subjects with AE(s)b | 16 (9.9) | 15 (9.3) | 19 (11.7)  |
|  General Disorders and Administration Site Conditions | 9 (5.6) | 10 (6.2) | 13 (8.0)  |
|  Injection Site Swelling | 4 (2.5) | 3 (1.9) | 5 (3.1)  |
|  Injection Site Erythema | 3 (1.9) | 3 (1.9) | 4 (2.5)  |
|  Injection Site Pain | 2 (1.2) | 2 (1.2) | 2 (1.2)  |
|  Injection Site Inflammation | 1 (0.6) | 1 (0.6) | 1 (0.6)  |
|  Injection Site Hematoma | 0 | 1 (0.6) | 1 (0.6)  |
|  Injection Site Discomfort | 1 (0.6%) | 0 | 1 (0.6)  |
|  Injection Site Mass | 0 | 1 (0.6) | 1 (0.6)  |
|  Skin and Subcutaneous Tissue Disorders | 3 (1.9) | 2 (1.2) | 3 (1.9)  |
|  Skin Tightness | 1 (0.6) | 0 | 1 (0.6)  |
|  Dermatitis Allergic | 1 (0.6) | 1 (0.6) | 1 (0.6)  |
|  Interstitial Granulomatous Dermatitis | 1 (0.6) | 1 (0.6) | 1 (0.6)  |
|  Gastrointestinal Disorders | 2 (1.2) | 2 (1.2) | 2 (1.2)  |
|  Salivary Hypersecretion | 1 (0.6) | 1 (0.6) | 1 (0.6)  |
|  Sensitivity of Teeth | 1 (0.6) | 1 (0.6) | 1 (0.6)  |
|  Injury, Poisoning and Procedural Complications | 1 (0.6) | 1 (0.6) | 1 (0.6)  |
|  Underdose | 1 (0.6) | 1 (0.6) | 1 (0.6)  |
|  Nervous System Disorders | 1 (0.6) | 1 (0.6) | 1 (0.6)  |
|  Presyncope | 1 (0.6) | 1 (0.6) | 1 (0.6)  |
|  Vascular Disorders | 1 (0.6) | 0 | 1 (0.6)  |
|  Arterial Stenosis | 1 (0.6) | 0 | 1 (0.6)  |

a Multiple occurrences within a System Organ Class by a subject were counted once per System Organ Class. Multiple occurrences of a Preferred Term by a subject were counted once per Preferred Term.
b A subject was counted once even if the subject experienced more than one AE during the study.
If the treated side where an AE occurred was Not Applicable, this AE was summarized both treatment sides.

## Retreatment Phase

Of the enrolled patients, 112/170 (65.9%) and 128/162 (79.0%) in the Restylane Refyne study and Restylane Defyne study, respectively, took part in an open label retreatment extension of the study 48 weeks after their final initial treatment. The retreatment option for both sides of the face was Restylane Refyne (Refyne study) or Restylane Defyne (Defyne study). An optional touch-up injection was administered 3 weeks after the retreatment injection, based on the judgment of the investigator. The safety profile observed after retreatment was similar to that described above during the initial treatment in the two studies.

PMA P140029: FDA Summary of Safety and Effectiveness Data

{17}

PMA P140029: FDA Summary of Safety and Effectiveness Data
Page 18

2. Effectiveness Results

The analysis of effectiveness was based on the 170 evaluable patients (Refyne study) and the 162 evaluable patients (Defyne study) at the 24-week time point.

Non-inferiority testing for Restylane Refyne (Refyne study) and Restylane Defyne (Defyne study) relative to Control was based on the 95% confidence interval approach with a non-inferiority margin of 0.4 units in WSRS. The confidence interval was constructed for the mean paired difference (Control minus Restylane Refyne or Restylane Defyne, respectively) in change from baseline in WSRS score at 24 weeks after the final initial injection. Non-inferiority was declared if the 95% confidence interval was fully above -0.4 (i.e., the lower limit was greater than -0.4).

In the Refyne study at 24 weeks after the final initial injection, the mean change in WSRS score from baseline was -1.1 for Restylane Refyne, as compared to -1.2 for Control. The lower limit of the 95% confidence interval in the comparison between treatments was -0.15, which was greater than the non-inferiority criterion of -0.4. Hence, the WSRS scores demonstrated non-inferiority of Restylane Refyne to Control.

In the Defyne study at 24 weeks after the final initial injection, the mean change in WSRS score from baseline was -1.1 for both Restylane Defyne and Control. The lower limit of the 95% confidence interval in the comparison between treatments was -0.18, which was greater than the non-inferiority criterion of -0.4. Hence, the WSRS scores demonstrated non-inferiority of Restylane Defyne to Control also in this study.

Secondary Endpoints

Pain Assessments: The benefit of pre-incorporated lidocaine in Restylane Refyne and Restylane Defyne was observed immediately at the time of injection and up to 1 hour after injection in both the Refyne and Defyne studies. The pain score at the Control (without lidocaine) treated side was statistically higher (worse) than the Restylane Refyne and Restylane Defyne treated side up to 60 minutes after injection for both the initial injection and the initial touch-up injection. However, after the initial injection, just as many subjects reported pain or tenderness over the next 21 days.

Subject Self-Assessment: The SSA scale used a similar 1-5 point scale as the WSRS and evaluated the subjects' evaluation of their own responses to treatment. The data in Table 14 below demonstrated consistency between clinician scores and patient reported outcomes.

{18}

Table 14: Restylane Refyne and Restylane Defyne Subject Self-Assessment

|   | Subject Self-Assessment |   | Responder Rate  |   |
| --- | --- | --- | --- | --- |
|  Refyne Study | Restylane Refyne | Control | Restylane Refyne | Control  |
|  Week 24 | -1.3 | -1.4 | 78.8% | 83.5%  |
|  Week 36 | -1.2 | -1.3 | 78.0% | 80.5%  |
|  Week 48 | -0.9 | -1.0 | 66.7% | 72.8%  |
|  |   |   |   |   |
|  Defyne Study | Restylane Defyne | Control | Restylane Defyne | Control  |
|  Week 24 | -1.2 | -1.3 | 77.8% | 85.2%  |
|  Week 36 | -1.1 | -1.2 | 73.6% | 81.8%  |
|  Week 48 | -0.9 | -1.0 | 64.4% | 75.2%  |

# 3. Subgroup Analyses

The following characteristics were evaluated for potential association with effectiveness outcomes:

- Skin Type (I to VI) and Skin Type by group (I to III, and IV to VI)
race (Caucasian versus non-Caucasian)
- gender (male/female)
- injection technique at the first injection (e.g., serial puncture, linear threading, fanning, crosshatching, or other)
- total injection volume (&lt;1 mL, 1 to 2 mL, or &gt;2 mL)
- previous use of dermal implants in treatment of NLFs (Implant versus non-Implant)
- investigational site

Age was significant in the models for both studies, but in the Refyne study older subjects had better outcomes, while in the Defyne study younger subjects had better outcomes. In both studies, the investigational and control results were similar in each age group.

There were 12 sites in the Refyne study and 11 sites in the Defyne study. In both studies, site was highly statistically significant. While there were differences between sites, at no site did the investigational device perform significantly worse than the control. There was one site in the Refyne study in which neither group showed meaningful improvement. A majority of the subjects in both groups showed no improvement (and one got worse). However, aesthetic studies have often had site to site differences.

There do not appear to be meaningful differences in either study based on skin type or race, or other subgroups considered. The treatment device (Restylane Refyne or Restylane Defyne) and their Control products both performed equally well or equally poorly in all subgroups.

PMA P140029: FDA Summary of Safety and Effectiveness Data

{19}

PMA P140029: FDA Summary of Safety and Effectiveness Data
Page 20

## E. Financial Disclosure

The Financial Disclosure by Clinical Investigators regulation (21 CFR 54) requires applicants who submit a marketing application to include certain information concerning the compensation to, and financial interests and arrangement of, any clinical investigator conducting clinical studies covered by the regulation. The two pivotal clinical studies included a total of 71 investigators of which none were full-time or part-time employees of the sponsor and 1 had disclosable financial interests/arrangements as defined in 21 CFR 54.2(a), (b), (c) and (f) and described below:

- Compensation to the investigator for conducting the study where the value could be influenced by the outcome of the study: none
- Significant payment of other sorts: 1
- Proprietary interest in the product tested held by the investigator: none
- Significant equity interest held by investigator in sponsor of covered study: none

The applicant has adequately disclosed the financial interest/arrangements with clinical investigators. Statistical analyses were conducted by FDA to determine whether the financial interests/arrangements had any impact on the clinical study outcome. The information provided does not raise any questions about the reliability of the data.

## XI. SUMMARY OF SUPPLEMENTAL CLINICAL INFORMATION

### Antibody Testing

Possible antibody formation to HA was tested in two European multi-center studies in which two phases were considered; an open label and a blind phase (the blind phase of the studies are reviewed below in Section Other Clinical Studies). In one study on Restylane Refyne (without lidocaine, but otherwise with same composition as Restylane Refyne with lidocaine) a total of 70 blood samples were tested from 25 subjects for IgG antibodies. One subject had an antibody value greater than the assay threshold before administration and at 24 and 48 weeks after administration. No clinical symptoms were reported for the subject and no reaction against Restylane Refyne was seen. In a second study on Restylane Defyne (without lidocaine, but otherwise with same composition as Restylane Defyne with lidocaine) a total of 53 blood samples were tested from 18 subjects for IgG antibodies. One subject had one sample with an antibody value greater than the assay threshold at 24 weeks after administration. An additional blood sample was obtained at 48 weeks that confirmed the positive result at 24 weeks. The subject had an adverse event of scarlet fever diagnosed within 2 weeks after the retreatment injection (Week 36). The adverse event was of moderate intensity and was completely resolved after treatment with antibiotics. The subject completed the study as planned.

{20}

PMA P140029: FDA Summary of Safety and Effectiveness Data
Page 21

# Other Clinical Studies

A randomized, evaluator-blinded, multi-center, active-controlled, intra-individual split-face comparison of Restylane Refyne (without lidocaine but otherwise with same composition as Restylane Refyne with lidocaine) with Control HA gel (without lidocaine) in the treatment of moderate facial wrinkles (WSRS score of 3 at inclusion) was conducted at 9 European centers. The randomized phase included 81 Caucasian subjects where the major Fitzpatrick skin type was III (64% of subjects) but with otherwise similar subject demographics as in the US studies. Up to two injections for the initial treatment were allowed, and the total injection volume was in average 0.9 mL for Restylane Refyne and 1.0 mL for Control.

The primary effectiveness endpoint of the study was non-inferiority testing of Restylane Refyne versus Control at 24 weeks after administration. The 95% confidence interval of the mean difference between treatments (Control minus Restylane Refyne) for the change from baseline in WSRS score was calculated. Non-inferiority was declared if the 95% confidence interval was fully above -0.5 (i.e., the lower limit was greater than -0.5). At 24 weeks following treatment, a mean change from baseline in WSRS of -0.86 units was observed for Restylane Refyne and a mean change of -0.95 units for Control. The lower limit of the 95% confidence interval for the mean difference between the treatments was -0.21 and hence above the non-inferiority criterion threshold set at -0.5. Therefore, non-inferiority in efficacy of Restylane Refyne to Control was demonstrated.

There were no discontinuations due to adverse events. One serious adverse event was reported: a case of appendicitis in the open label phase of the study, which was considered unrelated to Restylane Refyne.

In another randomized, evaluator-blinded, multi-center, active-controlled, intra-individual split-face comparison of Restylane Defyne (without lidocaine but otherwise with same composition as Restylane Defyne with lidocaine) with Control HA gel (without lidocaine) in the treatment of severe facial wrinkles (WSRS score of 4 at inclusion) was conducted at 6 European centers. The randomized phase included 68 Caucasian subjects where the major Fitzpatrick skin type was III (75.0% of subjects) but with otherwise similar subject demographics as in the US studies. Up to two injections for the initial treatment were allowed, and the total injection volume was in average 1.0 mL for Restylane Defyne and 1.0 mL for Control.

The primary effectiveness endpoint of the study and non-inferiority testing with reference to Control was the same as in the previous reviewed European study. At 24 weeks after treatment, a mean change from baseline in WSRS of -1.65 units was observed for Restylane Defyne and a mean change of -1.41 units for Control. The lower limit of the 95% confidence interval for the mean difference between the treatments was 0.1 units and hence above the non-inferiority criterion threshold set at -0.5. Therefore, non-inferiority in efficacy of Restylane Defyne versus Control was demonstrated.

There were no discontinuations due to adverse events. Four (4) serious adverse events were reported, assessed as not related to treatment; intestinal obstruction, bursitis subacromial, accidental trauma (impalement of left orbit) and breast cancer.

{21}

For both European studies, local responses and duration as well as investigator's reported adverse events showed a safety profile, which was similar to that described above in the pivotal studies (Section X).

## XII. PANEL MEETING RECOMMENDATION AND FDA'S POST-PANEL ACTION

In accordance with the provisions of section 515(c)(3) of the act as amended by the Safe Medical Devices Act of 1990, this PMA was not referred to the General and Plastic Surgery Devices Advisory Panel, an FDA advisory committee, for review and recommendation because the information in the PMA substantially duplicates information previously reviewed by this panel.

## XIII. CONCLUSIONS DRAWN FROM PRECLINICAL AND CLINICAL STUDIES

### A. Effectiveness Conclusions

Assessment of product effectiveness is based on the results of the two U.S. pivotal studies. These submitted data provided a reasonable assurance that Restylane Refyne and Restylane Defyne are effective for injection into the mid-to-deep dermis for the correction of moderate to severe facial wrinkles and folds (such as nasolabial folds) in patients over the age of 21.

Non-inferiority testing for Restylane Refyne and Restylane Defyne relative to Control was based on the 95% confidence interval approach with a non-inferiority margin of 0.4 units in WSRS. The confidence interval was constructed for the mean paired difference (Control minus Restylane Refyne or Restylane Defyne, respectively) in change from baseline in WSRS score at 24 weeks after the final initial injection. Non-inferiority was declared if the 95% confidence interval was fully above -0.4 (i.e., the lower limit was greater than -0.4).

In the Refyne study at 24 weeks after the final initial injection, the mean change in WSRS score from baseline was -1.1 for Restylane Refyne, as compared to -1.2 for Control. The lower limit of the 95% confidence interval in the comparison between treatments was -0.15, which was greater than the non-inferiority criterion of -0.4. Hence, the WSRS scores demonstrated non-inferiority of Restylane Refyne to Control.

In the Defyne study at 24 weeks after the final initial injection, the mean change in WSRS score from baseline was -1.1 for both Restylane Defyne and Control. The lower limit of the 95% confidence interval in the comparison between treatments was -0.18, which was greater than the non-inferiority criterion of -0.4. Hence, the WSRS scores demonstrated non-inferiority of Restylane Defyne to Control also in this study.

PMA P140029: FDA Summary of Safety and Effectiveness Data
Page 22

{22}

B. Safety Conclusions

The risks of the device are based on nonclinical laboratory and animal studies as well as data collected in clinical studies conducted to support PMA approval as described above. These submitted data provided a reasonable assurance of safety of injection of Restylane Refyne and Restylane Defyne for injection into the mid-to-deep dermis for the correction of moderate to severe facial wrinkles and folds (such as nasolabial folds) in patients over the age of 21. The specific conclusions are:

In the Refyne study, among the 170 treated subjects, 7.1% (12/170) experienced device- and injection-related AEs following initial and touch-up treatment, as well as retreatment and touch-up treatment. These subjects reported a total of 9 events related to control treatment and 10 events related to Restylane Refyne. The most common of treatment site responses was injection site erythema, which was reported for 4 subjects. Other treatment site responses (swelling, mass, hematoma, pruritus, and anesthesia) were reported for 1 subject each. Other related adverse events included presyncope (2 subjects), headache (1 subject), syncope (1 subject), and ecchymosis (1 subject). No related AEs led to discontinuation for any subject and no severe related AEs were reported during the study.

In the Defyne study, among the 162 treated subjects, 11.7% (19/162) experienced device- and injection-related AEs following initial and touch-up treatment, as well as retreatment and touch-up treatment. These subjects reported a total of 19 events related to control treatment and 20 events related to Restylane Defyne. The most common of treatment site responses was injection site swelling, which was reported for 5 subjects. Injection site erythema and injection site pain were reported for 4 and 2 subjects, respectively. Other treatment site responses (inflammation, hematoma, discomfort, and mass) were reported for 1 subject each. Other related adverse events included skin tightness, dermatitis allergic, interstitial granulomatous dermatitis, salivary hypersecretion, sensitivity of teeth, presyncope, and arterial stenosis which were all reported for 1 subject each. Two related AEs led to the discontinuation of 2 subjects (1 subject with interstitial granulomatous dermatitis, and 1 subject with allergic contact dermatitis). No severe related AEs were reported during the study.

In both studies, serious adverse events were reported (2 subjects in the Refyne study, 7 subjects in the Defyne study). However, no adverse events were in the treated area and none were assessed as related to the study product.

Subject Fitzpatrick skin type did not appear to have an effect on the distribution of AEs in either study, with similar incidence rates observed in subjects with both Skin Type subgroups I to III and IV to VI.

C. Benefit-Risk Determination

The probable benefits of the device are also based on data collected in clinical studies conducted to support PMA approval as described above. Both studies were well-

PMA P140029: FDA Summary of Safety and Effectiveness Data
Page 23

{23}

designed prospective, comparator controlled, non-inferiority studies using a validated scale and blinded, live evaluations. The data are considered to be as robust as possible for an aesthetic endpoint. For the primary effectiveness variable of mean change in WSRS from Baseline at Week 24 post-Final Initial Injection, both study products caused reduction in WSRS score, with the Control products showing similar reduction in mean WSRS compared to Restylane Refyne or Restylane Defyne. A summary of the effectiveness conclusions is provided above.

Additional factors to be considered in determining probable risks and benefits for the Restylane Refyne and Restylane Defyne devices included: 7.1% (12/170) and 11.7% (19/162) patients in the Restylane Refyne and Restylane Defyne studies, respectively, experienced device- and injection-related AEs across initial and retreatment injections. and touch-up treatment.

The probable benefits outweigh the probable risks, as determined by the robustness of the comparative effectiveness results, the lack of long term sequelae, and the improvement in subject self-assessment. The risks of short term adverse outcomes seen after injection and rare adverse events are sufficiently well understood for patients to make informed decisions about device use.

Patient Perspectives: Patient perspectives considered during the review included patient reported outcome information provided above, demonstrating that Subject Self-Assessment scores consistent with clinician scores. Patients are willing to accept the probable risk of reported adverse events, as seen with the willingness of the majority of patients to undergo retreatment at 48 weeks (112/170 (65.9%) and 128/162 (79.0%) in the Refyne study and Defyne study, respectively).

In conclusion, given the available information above, the data support the use of Restylane Refyne or Restylane Defyne for injection into the mid-to-deep dermis for the correction of moderate to severe facial wrinkles and folds (such as nasolabial folds) in patients over the age of 21, and the probable benefits outweigh the probable risks.

D. Overall Conclusions

The data in this application support the reasonable assurance of safety and effectiveness of this device when used in accordance with the indications for use.

XIV. CDRH DECISION

CDRH issued an approval order on December 9, 2016.

The applicant's manufacturing facilities have been inspected and found to be in compliance with the device Quality System (QS) regulation (21 CFR 820).

PMA P140029: FDA Summary of Safety and Effectiveness Data
Page 24

{24}

PMA P140029: FDA Summary of Safety and Effectiveness Data
Page 25

# XV. APPROVAL SPECIFICATIONS

Directions for use: See device labeling.

Hazards to Health from Use of the Device: See Indications, Contraindications, Warnings, Precautions, and Adverse Events in the device labeling.

Post-approval Requirements and Restrictions: None.

---

**Source:** [https://fda.innolitics.com/device/P140029](https://fda.innolitics.com/device/P140029)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a PMA](https://innolitics.com/services/regulatory/), [a 510(k)](https://innolitics.com/services/510ks/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
