# Medtronic CoreValve™ Evolut™ R System and Medtronic CoreValve™ Evolut™ Pro System (P130021S058)

_Medtronic, Inc. · NPT · Aug 16, 2019 · Cardiovascular · APPR_

**Canonical URL:** https://fda.innolitics.com/device/P130021S058

## Device Facts

- **Applicant:** Medtronic, Inc.
- **Product Code:** [NPT](/productcode/NPT.md)
- **Decision Date:** Aug 16, 2019
- **Decision:** APPR
- **Device Class:** Class 3
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic, Real-World Evidence

## Real-World Evidence

| Submission | Device | Sponsor | RWD Sources | RWE Use Summary | Key Tags |
| --- | --- | --- | --- | --- | --- |
| P130021S058 · Aug 16, 2019 | Medtronic CoreValve™ Evolut™ R System and Medtronic CoreValve™ Evolut™ Pro System | Medtronic, Inc. | Society of Thoracic Surgeons/American College of Cardiology Transcatheter Valve Therapy (TVT) Registry; Centers for Medicare and Medicaid Services (CMS) claims database | The FDA mandated a post-approval surveillance study using the TVT Registry and CMS claims data to characterize clinical outcomes and assess real-world use of the device in the 'low risk' patient population over a 10-year period. | Post-market surveillance; Registry; Claims data; Real-world use |

### Clinical Evidence

| Study Design | Population | Comparator | Key Endpoints |
| --- | --- | --- | --- |
| Medtronic CoreValve Evolut R System and CoreValve Evolut PRO System Registry-Based Continued Access Protocol (CAP) Cohort and “Low Risk” Indication Real-World Use Surveillance; Registry-based surveillance with linkage to claims database; Follow-up/Duration: 10 years post implantation; Study Period: First 2 years of commercial use (enrollment period) | Patients treated with the Medtronic CoreValve Evolut R System and CoreValve Evolut PRO System for the 'low risk' indication; Sample Size: All consecutive patients treated within the first 2 years entered into the TVT Registry | Not applicable for this study | Device success, all-cause mortality, all stroke, life-threatening/major bleeding, new requirement for dialysis, peri-procedural myocardial infarction, repeat procedure for valve-related dysfunction, vascular complications, and quality of life (KCCQ) |

## Indications for Use

The Medtronic CoreValve Evolut R System and Medtronic CoreValve Evolut PRO System are indicated for relief of aortic stenosis in patients with symptomatic heart disease due to severe native calcific aortic stenosis who are judged by a heart team, including a cardiac surgeon, to be appropriate for the transcatheter heart valve replacement therapy.

## Device Story

The Medtronic CoreValve Evolut R and Evolut PRO systems are self-expanding transcatheter heart valves (TAV) designed for percutaneous replacement of native aortic valves. The system comprises the TAV, a delivery catheter system (DCS), and a loading system. The Evolut R allows for resheathing and recapture during deployment; the Evolut PRO includes a porcine pericardial tissue wrap to reduce paravalvular regurgitation. Used in a catheterization lab or OR by a heart team, the device is delivered via an over-the-wire approach (typically transfemoral). The valve frame is made of Nitinol. The device provides a less invasive alternative to surgical aortic valve replacement (SAVR), aiming to improve valve hemodynamics, functional status (NYHA class), and quality of life in patients with severe aortic stenosis. Clinical outcomes are monitored via echocardiography and patient-reported quality of life metrics.

## Clinical Evidence

Prospective, randomized, multi-center trial (Low Risk Trial) of 1468 patients comparing TAVR (Evolut R/PRO) to SAVR. Primary endpoint: all-cause mortality or disabling stroke at 24 months. TAVR was non-inferior to SAVR (posterior probability >0.999). TAVR showed superior mean gradient and EOA at 12 months. Adverse events included conduction disturbances requiring permanent pacemakers (22.7% TAVR vs 7.6% SAVR at 24 months). CT substudy evaluated leaflet thickening and mobility; clinical significance of findings remains unknown.

## Technological Characteristics

Self-expanding transcatheter heart valve frame constructed of Nitinol (titanium/nickel alloy). Evolut PRO includes a porcine pericardial tissue wrap. Delivery systems are intravascular, over-the-wire, compatible with 0.035" guidewires. Sterilization and specific ASTM standards are referenced in the original PMA (P130021).

## Regulatory Identification

To replace a patient's aortic heart valve.  They are different from the classified device (heart valves) in that they are placed percutaneously and do not require open chest surgery or a cardiotomy for placement.

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a PMA](https://innolitics.com/services/regulatory/), [a 510(k)](https://innolitics.com/services/510ks/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{1}

# SUMMARY OF SAFETY AND EFFECTIVENESS DATA

## I. GENERAL INFORMATION

|  Device Generic Name: | Aortic valve, prosthesis, percutaneously delivered  |
| --- | --- |
|  Device Trade Name: | Medtronic CoreValve Evolut R System Medtronic CoreValve Evolut PRO System  |
|  Device Procode: | NPT  |
|  Applicant Name and Address: | Medtronic CoreValve LLC 3576 Unocal Place Santa Rosa, CA 95403  |
|  Date of Panel Recommendation: | None  |
|  Premarket Approval Application (PMA) Number: | P130021/S058  |
|  Date of FDA Notice of Approval: | August 16, 2019  |

The original PMA of the Medtronic CoreValve System, P130021, was first approved on January 17, 2014. The device has since undergone two new design iterations: the Evolut R System was approved under P130021/S014 (for sizes 23, 26, and 29 mm) and P130021/S025 (for size 34 mm) on June 22, 2015, and October 26, 2016, respectively; and the Evolut PRO System was approved under P130021/S029 on March 20, 2017. The indication has also since been expanded in Panel Track PMA Supplements P130021/S002, P130021/S010, and P130021/S033 on June 12, 2014, March 30, 2015, and July 10, 2017, respectively, to include: (1) patients with symptomatic heart disease due to severe native calcific aortic stenosis who are judged by a heart team, including a cardiac surgeon, to be at intermediate or greater risk for open surgical therapy (i.e., predicted risk of surgical mortality $\geq 3\%$ at 30 days, based on the Society of Thoracic Surgeons (STS) risk score and other clinical comorbidities unmeasured by the STS risk calculator); and (2) patients with symptomatic heart disease due to failure (stenosed, insufficient, or combined) of a surgical bioprosthetic aortic valve who are judged by a heart team, including a cardiac surgeon, to be at high or greater risk for open surgical therapy (i.e., STS predicted risk of operative mortality score $\geq 8\%$ or at a $\geq 15\%$ risk of mortality at 30 days).

The SSEDs to support the indication are available on the following FDA websites and are incorporated by reference herein:

- P130021: http://www.accessdata.fda.gov/cdrh_docs/pdf13/P130021b.pdf

PMA P130021/S058: FDA Summary of Safety and Effectiveness Data

Page 1

{2}

- P130021/S002:
http://www.accessdata.fda.gov/cdrh_docs/pdf13/P130021S002b.pdf
- P130021/S010:
http://www.accessdata.fda.gov/cdrh_docs/pdf13/P130021S010B.pdf
- P130021/S033:
https://www.accessdata.fda.gov/cdrh_docs/pdf13/p130021s033b.pdf

The current supplement was submitted to expand the indications for use of the Evolut R System and Evolut PRO System to include patients with severe symptomatic native calcific aortic stenosis who are deemed to be at low risk for surgical aortic valve replacement (SAVR).

## II. INDICATIONS FOR USE

The Medtronic CoreValve Evolut R System and Medtronic CoreValve Evolut PRO System are indicated for relief of aortic stenosis in patients with symptomatic heart disease due to severe native calcific aortic stenosis who are judged by a heart team, including a cardiac surgeon, to be appropriate for the transcatheter heart valve replacement therapy.

### III. CONTRAINDICATIONS

The CoreValve Evolut R system and the CoreValve Evolut PRO system are contraindicated in patients who cannot tolerate Nitinol (Titanium or Nickel), an anticoagulation/antiplatelet regimen, or who have active bacterial endocarditis or other active infections.

### IV. WARNINGS AND PRECAUTIONS

The warnings and precautions can be found in the Medtronic Evolut R System and Evolut PRO System labeling.

### V. DEVICE DESCRIPTION

The Medtronic Evolut R and Evolut PRO Systems each consists of 3 components: the Transcatheter Aortic Valve (TAV), the Delivery Catheter System (DCS), and the Loading System (LS).

- Medtronic Evolut R System

The Evolut R TAV (models EVOLUTR-23-US, EVOLUTR-26-US, EVOLUTR-29-US, and EVOLUTR-34-US), as shown in Figure 1, is a design iteration of the CoreValve TAV. It provides the optional capability of allowing for resheathing and/or complete recapture and redeployment during valve deployment. The Evolut R TAV is fully functional at approximately 2/3 partial deployment from the DCS. Once the TAV is fully deployed, it is not retrievable from the site of implantation.

PMA P130021/S058: FDA Summary of Safety and Effectiveness Data

Page 2

{3}

**Figure 1: Evolut R Transcatheter Aortic Valve**

![img-0.jpeg](img-0.jpeg)

The Evolut R TAV can be delivered interchangeably using the EnVeo R DCS (models ENVEOR-US, ENVEOR-N-US, ENVPRO-14-US, and ENVPRO-16-US) or EnVeo PRO DCS (models ENVEOR-N-US and ENVPRO-16-US), which is a single use, intravascular, over-the-wire delivery catheter, as shown in Figure 2 and Figure 3, respectively. Both systems are designed to be compatible with commercially available 0.035' intravascular wires and incorporate a protective deployment sheath that houses and deploys the prosthesis.

**Figure 2: EnVeo R Delivery Catheter System**

![img-1.jpeg](img-1.jpeg)

**Figure 3: EnVeo PRO Delivery Catheter System**

![img-2.jpeg](img-2.jpeg)

The Evolut R TAV can be loaded onto the delivery system using the EnVeo R LS (models LS-ENVEOR-23US, LS-ENVEOR-2629US, LS-ENVEOR-34US, LS-ENVPRO-14-US, and L-ENVPRO-16-US) or EnVeo PRO LS (models LS-MDT2-23-US, and LS-MDT2-2629-US, L-ENVPRO1623-US, and L-ENVPRO-16-US), as shown in Figure 4 and Figure 5, respectively.

PMA P130021/S058: FDA Summary of Safety and Effectiveness Data

Page 3

{4}

Figure 4: EnVeo R Loading System

![img-3.jpeg](img-3.jpeg)

Figure 5: EnVeo PRO Loading System

![img-4.jpeg](img-4.jpeg)

1. Inflow ring
2. Inflow cone
3. Outflow cone
4. Back plate
5. Capsule Guide Tube
6. Tip guide tube

# • Medtronic Evolut PRO System

The Evolut PRO TAV, as shown in Figure 6, is a design iteration of the Evolut R TAV, with the addition of a porcine pericardial tissue wrap on the outside of the frame (outer wrap) that covers the inflow portion of the TAV to reduce paravalvular regurgitation.

Figure 6: Evolut PRO Transcatheter Aortic Valve

![img-5.jpeg](img-5.jpeg)

All three sizes of the Evolut PRO TAVs are deployed using the 20 Fr EnVeo R DCS or 16 Fr equivalent EnVeo PRO DCS.

The EnVeo R LS used with the Evolut PRO TAV is similar to that used with the Evolut R TAV, with minor design modifications to the inflow cone, the inflow ring, and the outflow cone. The Evolut PRO TAVs can also be used with the EnVeo PRO LS, similar to that used with the Evolut R TAVs.

PMA P130021/S058: FDA Summary of Safety and Effectiveness Data

Page 4

{5}

## **VI. ALTERNATIVE PRACTICES AND PROCEDURES**

There are several other alternatives for the correction of severe native calcific aortic stenosis in patients deemed to be at low risk for open surgical therapy, including surgical aortic valve replacement (SAVR), temporary relief using balloon aortic valvuloplasty (BAV), or medical therapy (no obstruction-relieving intervention). Each alternative has its own advantages and disadvantages. A patient should fully discuss these alternatives with his/her physician to select the method that best meets expectations and lifestyle.

## **VII. MARKETING HISTORY**

The Medtronic CoreValve Evolut R System and CoreValve Evolut PRO System have not been marketed in the United States or any foreign country for the “low risk” transcatheter aortic valve replacement (TAVR) indication.

## **VIII. POTENTIAL ADVERSE EFFECTS OF THE DEVICE ON HEALTH**

Below is a list of the potential adverse effects (e.g., complications) associated with the use of the Medtronic CoreValve Evolut R System and CoreValve Evolut PRO System:

- – Death
- – Myocardial infarction, cardiac arrest, cardiogenic shock, cardiac tamponade
- – Coronary occlusion, obstruction, or vessel spasm (including acute coronary closure)
- – Cardiovascular injury (including rupture, perforation, tissue erosion, or dissection of vessels, ascending aorta trauma, ventricle, myocardium, or valvular structures that may require intervention)
- – Emergent surgical or transcatheter intervention (for example, coronary artery bypass, heart valve replacement, valve explant, percutaneous coronary intervention [PCI], balloon valvuloplasty)
- – Prosthetic valve dysfunction (regurgitation or stenosis) due to fracture; bending (out-of-round configuration) of the valve frame; underexpansion of the valve frame; calcification; pannus; leaflet wear, tear, prolapse, or retraction; poor valve coaptation; suture breaks or disruption; leaks; mal-sizing (prosthesis-patient mismatch); malposition (either too high or too low)/malplacement
- – Prosthetic valve migration/embolization
- – Prosthetic valve endocarditis
- – Prosthetic valve thrombosis
- – Delivery catheter system malfunction resulting in the need for additional re-crossing of the aortic valve and prolonged procedural time
- – Delivery catheter system component migration/embolization
- – Stroke (ischemic or hemorrhagic), transient ischemic attack (TIA), or other neurological deficits
- – Individual organ (for example, cardiac, respiratory, renal [including acute kidney failure]) or multi-organ insufficiency or failure

PMA P130021/S058: FDA Summary of Safety and Effectiveness Data

Page 5

{6}

- Major or minor bleeding that may require transfusion or intervention (including life-threatening or disabling bleeding)
- Vascular access-related complications (for example, dissection, perforation, pain, bleeding, hematoma, pseudoaneurysm, irreversible nerve injury, compartment syndrome, arteriovenous fistula, stenosis)
- Mitral valve regurgitation or injury
- Conduction system disturbances (for example, atrioventricular node block, left-bundle branch block, asystole), which may require a permanent pacemaker
- Infection (including septicemia)
- Hypotension or hypertension
- Hemolysis
- Peripheral ischemia
- Bowel ischemia
- Abnormal lab values (including electrolyte imbalance)
- Allergic reaction to antiplatelet agents, contrast medium, or anesthesia
- Exposure to radiation through fluoroscopy and angiography
- Permanent disability

For the specific adverse events that occurred in the clinical study, please see Section X.

# IX. SUMMARY OF PRECLINICAL STUDIES

A summary of previously reported preclinical studies can be found in the SSED for the original PMA. No additional preclinical study was performed for the current application.

# X. SUMMARY OF PRIMARY CLINICAL STUDY

The applicant performed a clinical study to establish a reasonable assurance of safety and effectiveness of transcatheter aortic valve replacement (TAVR) with the CoreValve Evolut R System and CoreValve Evolut PRO System for patients with severe, native, calcific, aortic stenosis deemed by a heart team to be at low risk for open surgical therapy under IDE #G160022 (entitled the “Low Risk Trial”). Data from this clinical study were the basis for the PMA approval decision. A summary of the clinical study is presented below.

# A. Study Design

Patients were enrolled between March 25, 2016 and November 26, 2018. The database for this Panel Track PMA Supplement reflected data collected through November 30, 2018 and included 1468 randomized patients. There were 86 investigational sites in the US, Canada, Australia, New Zealand, Europe, and Japan.

The Low Risk Trial was a prospective, randomized (1:1), multi-center investigational study intended to determine whether TAVR is non-inferior to SAVR (with an absolute margin, δ, of 0.06) with respect to the primary endpoint. The randomization was stratified by investigational site and the need for revascularization. The sample size of the trial was

PMA P130021/S058: FDA Summary of Safety and Effectiveness Data

Page 6

{7}

1200 patients. The trial employed Bayesian adaptive statistical methods to allow for a first “early win” analysis to be performed when 850 patients would have been followed for 12 months. At the first “early win” analysis, if the posterior probability, $Prob(H_{A,\delta=0.06}|data)$ with $H_A$ being defined as the alternative hypothesis, was to be greater than 0.972, non-inferiority would be declared at this time; otherwise, a second “early win” analysis would occur when 1200 patients would have reached 12 months follow-up. If non-inferiority was not to be reached, all 1200 patients would be followed to 24 months when a final analysis would occur. At the final analysis, the standard for trial success would again be $Prob(H_{A,\delta=0.06}|data) > 0.972$.

A subset of patients were enrolled in a computed tomography (CT) substudy to investigate the prevalence of Hypoattenuated Leaflet Thickening (HALT) and reduced leaflet mobility.

Additional patients were enrolled in the trial after 1200 patients had been enrolled to complete the CT substudy and a cohort in Japan, resulting in a combined sample size of 1468 randomized patients at the time of the database lock.

Independent designees were utilized for interpretation and analysis of data for several aspects of the study, including: an independent Data Safety Monitoring Board (DSMB) with an independent statistician, a Clinical Events Committee (CEC) that was responsible for adjudicating adverse events, an echocardiography core laboratory, and a contract research organization that participated in source data verification. A computer tomography (CT) core laboratory was used for assessment of CT images acquired in the CT substudy.

### 1. Clinical Inclusion and Exclusion Criteria

Enrollment in the Low Risk Trial was limited to patients who met the following inclusion criteria:

- Severe aortic stenosis, defined as follows:

- For symptomatic patients:

  o Aortic valve area ≤1.0 cm² (or aortic valve area index of ≤0.6 cm²/m²), OR mean gradient ≥40 mmHg, OR maximal aortic valve velocity ≥4.0 m/sec by transthoracic echocardiography at rest.

- For asymptomatic patients:

  o Very severe aortic stenosis with an aortic valve area of ≤1.0 cm² (or aortic valve area index of ≤0.6 cm²/m²), AND maximal aortic velocity ≥5.0 m/sec, or mean gradient ≥60 mmHg by transthoracic echocardiography at rest, OR
  o Aortic valve area of ≤1.0 cm² (or aortic valve area index of ≤0.6 cm²/m²), AND a mean gradient ≥40 mmHg or maximal aortic valve velocity ≥4.0 m/sec by transthoracic echocardiography at rest, AND an exercise tolerance test that demonstrates a limited exercise capacity, abnormal BP response, or arrhythmia, OR

PMA P130021/S058: FDA Summary of Safety and Effectiveness Data

Page 7

{8}

○ Aortic valve area of ≤1.0 cm² (or aortic valve area index of ≤0.6 cm²/m²), AND mean gradient ≥40 mmHg, or maximal aortic valve velocity ≥4.0 m/sec by transthoracic echocardiography at rest, AND a left ventricular ejection fraction <50%.
- Patient is considered low risk for SAVR, where low risk is defined as predicted risk of mortality for SAVR <3% at 30 days per multidisciplinary local heart team assessment.
- The subject and the treating physician agree that the subject will return for all required post-procedure follow-up visits.

Patients were not permitted to enroll in the Low Risk Trial if they met any of the following clinical or anatomical exclusion criteria:

- Any condition considered a contraindication for placement of a bioprosthetic valve (eg, subject is indicated for mechanical prosthetic valve).
- A known hypersensitivity or contraindication to any of the following that cannot be adequately pre-medicated:
  • aspirin or heparin (HIT/HITTS) and bivalirudin
  • ticlopidine and clopidogrel
  • Nitinol (titanium or nickel)
  • contrast media
- Blood dyscrasias as defined: leukopenia (WBC <1000 mm³), thrombocytopenia (platelet count <50,000 cells/mm³), history of bleeding diathesis or coagulopathy, or hypercoagulable states.
- Ongoing sepsis, including active endocarditis.
- Any percutaneous coronary or peripheral interventional procedure with a bare metal stent within 30 days prior to randomization, or drug eluting stent performed within 180 days prior to randomization.
- Multivessel coronary artery disease with a Syntax score >22 and/or unprotected left main coronary artery.
- Symptomatic carotid or vertebral artery disease or successful treatment of carotid stenosis within 10 weeks of Heart Team assessment.
- Cardiogenic shock manifested by low cardiac output, vasopressor dependence, or mechanical hemodynamic support.
- Recent (within 2 months of Heart Team assessment) cerebrovascular accident (CVA) or transient ischemic attack (TIA).
- Gastrointestinal (GI) bleeding that would preclude anticoagulation.
- Subject refuses a blood transfusion.
- Severe dementia (resulting in either inability to provide informed consent for the trial/procedure, prevents independent lifestyle outside of a chronic care facility, or will fundamentally complicate rehabilitation from the procedure or compliance with follow-up visits).
- Estimated life expectancy of less than 24 months due to associated non-cardiac co-morbid conditions.
- Other medical, social, or psychological conditions that in the opinion of the

PMA P130021/S058: FDA Summary of Safety and Effectiveness Data

Page 8

{9}

investigator precludes the subject from appropriate consent or adherence to the protocol required follow-up exams.

- Currently participating in an investigational drug or another device trial (excluding registries).
- Evidence of an acute myocardial infarction ≤30 days before the trial procedure due to unstable coronary artery disease (WHO criteria).
- Need for emergency surgery for any reason.
- Subject is pregnant or breastfeeding.
- Subject is less than legal age of consent, legally incompetent, or otherwise vulnerable.
- Pre-existing prosthetic heart valve in any position.
- Severe mitral regurgitation amenable to surgical replacement or repair.
- Severe tricuspid regurgitation amenable to surgical replacement or repair.
- Moderate or severe mitral stenosis amenable to surgical replacement or repair.
- Hypertrophic obstructive cardiomyopathy with left ventricular outflow gradient.
- Bicuspid aortic valve verified by echocardiography, multiple detector computed tomography (MDCT), or magnetic resonance imaging (MRI).
- Prohibitive left ventricular outflow tract calcification.
- Sinus of Valsalva diameter unsuitable for placement of the self-expanding bioprosthesis.
- Aortic annulus diameter of <18 or >30 mm.
- Significant aortopathy requiring ascending aortic replacement.
- For transfemoral or transaxillary (subclavian) access:

Access vessel mean diameter <5.0 mm for Evolut 23R, 26R, or 29R mm TAV, or access vessel mean diameter <5.5 mm for Evolut 34R mm or Evolut PRO TAV. However, for transaxillary (subclavian) access in patients with a patent LIMA, access vessel mean diameter <5.5mm for Evolut 23R, 26R, or 29R mm TAV, or access vessel mean diameter <6.0 mm for the Evolut 34R or Evolut PRO TAV.

## 2. Follow-Up Schedule

All patients were scheduled for follow-up examinations at discharge, 30 days, 6 months, 1 year, 18 months, 2 years, and annually thereafter to a minimum of 10 years post-procedure, with the clinical assessments at 6, 8, and 9 years being conducted via telephone. Preoperative and post-operative assessments included physical assessment and patient interview, laboratory measurements, imaging tests, and health status/quality of life (QoL) questionnaire. Adverse events and complications were recorded at all visits.

PMA P130021/S058: FDA Summary of Safety and Effectiveness Data

Page 9

{10}

### 3. Clinical Endpoints

# *Primary Endpoint:*

The primary endpoint was all-cause mortality or disabling stroke rate at 24 months, with the following alternative hypothesis:

$$H_{\mathrm{A}}: \pi_{\mathrm{TAVR}} < \pi_{\mathrm{SAVR}} + 0.06$$

where $\pi_{\mathrm{TAVR}}$ and $\pi_{\mathrm{SAVR}}$ denote binary rates of all-cause mortality or disabling stroke at 24 months for the TAVR (treatment) and SAVR (control) cohorts, respectively.

# *Secondary Endpoints:*

The following ordered list of secondary endpoints, as shown in Table 1, were evaluated in a hierarchical testing scheme:

Table 1: Ordered List of Secondary Endpoints for Hierarchical Testing

|  Order | Secondary Endpoint | Alternative Hypothesis  |
| --- | --- | --- |
|  #1 | Transvalvular mean gradient at 12 months (non-inferiority) | $H_{\mathrm{A}}: \mu_{\mathrm{TAVR}} < \mu_{\mathrm{SAVR}} + 5$  |
|  #2 | Effective orifice area (EOA) at 12 months (non-inferiority) | $H_{\mathrm{A}}: \mu_{\mathrm{TAVR}} > \mu_{\mathrm{SAVR}} - 0.1$  |
|  #3 | Change in Kansas City Cardiomyopathy Questionnaire (KCCQ) overall score from baseline to 12 months (non-inferiority) | $H_{\mathrm{A}}: \mu_{\mathrm{TAVR}} > \mu_{\mathrm{SAVR}} - 5$  |
|  #4 | Change in New York Heart Association (NYHA) classification from baseline to 12 months (non-inferiority) | $H_{\mathrm{A}}: \mu_{\mathrm{TAVR}} > \mu_{\mathrm{SAVR}} - 0.375$  |
|  #5 | Transvalvular mean gradient at 12 months (superiority) | $H_{\mathrm{A}}: \mu_{\mathrm{TAVR}} < \mu_{\mathrm{SAVR}}$  |
|  #6 | EOA at 12 months (superiority) | $H_{\mathrm{A}}: \mu_{\mathrm{TAVR}} > \mu_{\mathrm{SAVR}}$  |
|  #7 | Change in KCCQ overall score from baseline to 30 days (superiority) | $H_{\mathrm{A}}: \mu_{\mathrm{TAVR}} > \mu_{\mathrm{SAVR}}$  |

### B. Accountability of PMA Cohort

At the time of database lock, a total of 1468 patients were randomized in this study, including 734 TAVR patients and 734 SAVR patients.

There were four different analysis populations defined in the statistical analysis plan of the study: intention-to-treat (ITT), as treated (AT), implanted, and per protocol

PMA P130021/S058: FDA Summary of Safety and Effectiveness Data

Page 10

{11}

(PP), as summarized in Table 2 and Figure 7. The primary analysis was the AT analysis.

**Table 2: Analysis Populations**

|  Analysis Population | Definition | Number of Patients  |   |
| --- | --- | --- | --- |
|   |   |  SAVR | TAVR  |
|  Intention-to-treat (ITT) | All randomized patients | 734 | 734  |
|  As treated (AT) | All ITT patients with an attempted implant procedure* | 725 | 678  |
|  Implanted | All AT patients who were actually implanted with a valve | 722 | 680  |
|  Per protocol (PP) | Based on the International Council for Harmonisation (ICH) E9 Statistical Principals: – All implanted patients who were implanted according to their randomization; and – Patients without early exit (e.g., lost to follow-up) before 24 months (730 days), except those experiencing the primary endpoint (death or disabling stroke) prior to the early exit; and – Patients without crossover to a different type of procedure from their first attempted procedure type before their 24-month visits; and – Patients must satisfy all inclusion/exclusion criteria. | 702 | 647  |

*Attempted implant procedure was defined as when the subject was brought into the procedure room and any of the following had occurred: anesthesia administered, vascular line placed, transesophageal echocardiography probe placed, or any monitoring line placed. Patients were analyzed according to their first attempted procedure (TAVR or SAVR).

PMA P130021/S058: FDA Summary of Safety and Effectiveness Data

Page 11

{12}

Figure 7: Population Flowchart

![img-6.jpeg](img-6.jpeg)

Of the 722 patients in the implanted TAVR cohort, 534 patients were implanted with the Evolut R TAV, 162 patients with the Evolut PRO TAV, and 26 patients with the CoreValve 31 mm TAV.

The overall follow-up compliance of the trial is summarized in Table 3. The compliance rates were similar for TAVR and SAVR patients at each visit through 24 months.

PMA P130021/S058: FDA Summary of Safety and Effectiveness Data

Page 12

{13}

**Table 3: Overall Study Compliance (ITT Population)**

|  Visit Interval | Number Expected* | Visit Completed | Study Exits |   |   |   |   | Pending Next Visit  |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
|   |   |   |  Not Eligible | Died | Withdrew† | Lost to Follow Up | Other  |   |
|  **TAVR**  |   |   |   |   |   |   |   |   |
|  Randomized | 734 | 100.0% | 1 | 0 | 9 | 0 | 0 | 2  |
|  Procedure | 722 | 100.0% | 0 | 3 | 0 | 0 | 0 | 0  |
|  Discharge | 719 | 100.0% | 0 | 0 | 0 | 0 | 0 | 0  |
|  1-Month | 719 | 99.9% | 0 | 6 | 2 | 1 | 0 | 2  |
|  6-Month | 708 | 98.7% | 0 | 6 | 3 | 0 | 0 | 81  |
|  12-Month | 618 | 98.5% | 0 | 2 | 5 | 0 | 0 | 222  |
|  18-Month | 389 | 96.4% | 0 | 4 | 0 | 1 | 0 | 190  |
|  24-Month | 194 | 96.4% | 0 | 0 | 0 | 0 | 0 | 190  |
|  **SAVR**  |   |   |   |   |   |   |   |   |
|  Randomized | 734 | 100.0% | 4 | 1 | 45 | 2 | 0 | 1  |
|  Procedure | 681 | 100.0% | 0 | 6 | 2 | 0 | 0 | 0  |
|  Discharge | 673 | 100.0% | 0 | 2 | 3 | 0 | 0 | 0  |
|  1-Month | 668 | 99.1% | 0 | 5 | 10 | 0 | 0 | 2  |
|  6-Month | 651 | 96.3% | 0 | 6 | 9 | 1 | 0 | 95  |
|  12-Month | 540 | 96.9% | 0 | 2 | 5 | 2 | 0 | 208  |
|  18-Month | 323 | 94.1% | 0 | 0 | 1 | 0 | 0 | 160  |
|  24-Month | 162 | 96.3% | 0 | 1 | 2 | 0 | 0 | 156  |

\*Number of expected visits in an interval = (# of expected visits in the previous interval - # not eligible - # died - # withdrew - # lost to follow up - # other - # pending).

†Withdrew includes subjects who withdrew consent and who were withdrawn from study by physician.

### **C. Study Population Demographics and Baseline Parameters**

The demographics and baseline characteristics of the study population are typical for a TAVR study performed in the U.S., as summarized in Table 4. The treatment cohorts were generally well balanced with respect to age, gender, baseline NYHA classification, and STS risk score.

PMA P130021/S058: FDA Summary of Safety and Effectiveness Data

Page 13

{14}

**Table 4: Patient Demographics and Baseline Characteristics (AT Population)**

|  Demographics and Baseline Characteristics | Summary Statistics*  |   |   |
| --- | --- | --- | --- |
|   |  TAVR | SAVR | Difference (TAVR – SAVR) (95% BCI)  |
|  Age (years) | 74.1 ± 5.8 (725) | 73.6 ± 5.9 (678) | (-0.17, 1.07)  |
|  Gender female (%) | 36.0% (261/725) | 33.8% (229/678) | (-2.77%, 7.18%)  |
|  NYHA class  |   |   |   |
|  I | 10.5% (76/725) | 9.3% (63/678) | (-1.95%, 4.30%)  |
|  II | 64.4% (467/725) | 62.2% (422/678) | (-2.85%, 7.21%)  |
|  III | 25.0% (181/725) | 28.0% (190/678) | (-7.64%, 1.57%)  |
|  IV | 0.1% (1/725) | 0.4% (3/678) | (-1.07%, 0.34%)  |
|  STS score, % | 1.9 ± 0.7 (725) | 1.9 ± 0.7 (678) | (-0.03, 0.11)  |
|  Peripheral arterial disease | 7.5% (54/718) | 8.3% (56/677) | (-3.62%, 2.09%)  |
|  Previous myocardial infarction | 6.6% (48/725) | 4.9% (33/678) | (-0.70%, 4.20%)  |
|  Previous reintervention  |   |   |   |
|  Coronary artery bypass Surgery | 2.5% (18/725) | 2.1% (14/678) | (-1.20%, 2.02%)  |
|  Percutaneous coronary Intervention (PCI) | 14.2% (103/725) | 12.8% (87/678) | (-2.21%, 4.94%)  |
|  Cerebrovascular disease | 10.2% (74/725) | 11.8% (80/678) | (-4.90%, 1.67%)  |
|  Immunosuppressive therapy | 2.1% (15/725) | 0.9% (6/678) | (-0.11%, 2.53%)  |
|  Chronic lung disease/COPD | 15.0% (104/695) | 18.0% (117/649) | (-7.04%, 0.90%)  |
|  Diabetes | 31.4% (228/725) | 30.5% (207/678) | (-3.91%, 5.73%)  |
|  Creatinine level > 2 mg/dl | 0.4% (3/725) | 0.1% (1/678) | (-0.41%, 0.98%)  |
|  Atrial fibrillation/atrial flutter | 15.4% (111/722) | 14.5% (98/676) | (-2.86%, 4.60%)  |
|  Pre-existing permanent pacemaker or defibrillator | 3.2% (23/725) | 3.8% (26/677) | (-2.66%, 1.28%)  |
|  Hypertension | 84.8% (614/724) | 82.6% (559/677) | (-1.63%, 6.11%)  |
|  Dialysis | 0.0% (0/725) | 0.1% (1/678) | (-0.72%, 0.31%)  |
|  Echocardiographic findings - Implanted Population  |   |   |   |
|  Aortic valve area (cm^{2}) | 0.8 ± 0.2 (716) | 0.8 ± 0.2 (673) | (-0.02, 0.02)  |
|  Mean gradient (mmHg) | 47.0 ± 12.1 (724) | 46.6 ± 12.2 (678) | (-0.87, 1.69)  |

\*Continuous measures - Mean ± SD (Total no.); categorical measures - % (no./Total no.)

#### **D. Safety and Effectiveness Results**

At the time of the first “early win” analysis, 168 patients had been followed for 24 months in the original dataset. Subsequently, a supplemental analysis was performed on an expanded dataset during the review of the PMA application, which included

PMA P130021/S058: FDA Summary of Safety and Effectiveness Data

Page 14

{15}

additional follow-up data collected through May 3, 2019 when 410 patients had been followed for 24 months. The data presented in this section reflect the results of the supplemental analysis unless noted otherwise. Specifically, all hypothesis testing was conducted on the original dataset.

### 1. Primary Endpoint:

The first “early win” assessment of the primary endpoint of all-cause mortality or disabling stroke rate at 24 months included all patients in the AT population (N=1403). The median of the posterior distribution for the primary endpoint event rate was 5.3% for the TAVR cohort and 6.7% for the SAVR cohort, with a median of the posterior distribution of the difference in the primary endpoint event rate of -1.4% (TAVR-SAVR) and a 95% Bayesian credible interval (BCI) of (-4.9%, 2.1%), as summarized in Table 5. The posterior probability of non-inferiority with a margin of 6% was >0.999, which is greater than the pre-specified threshold of 0.972, thus the primary endpoint non-inferiority could be concluded.

Similarly, the supplemental analysis showed that the median of the posterior distribution for the primary endpoint event rate was 4.4% for the TAVR cohort and 6.2% for the SAVR cohort, with a median of the posterior distribution of the difference in the primary event rate of -1.8% (TAVR – SAVR) and a 95% BCI of (-4.6%, 1.0%), as summarized in Table 5. Hypothesis testing was not repeated on the expanded dataset because it was not prespecified; the supplemental analysis for the posterior probability of non-inferiority with a margin of 6% is shown for context.

**Table 5: All-Cause Mortality or Disabling Stroke at 24 Months - AT Population**

|   | “Early Win” Analysis* |   | Supplemental Analysis†  |   |
| --- | --- | --- | --- | --- |
|   |  TAVR (N=725) | SAVR (N=678) | TAVR (N=725) | SAVR (N=678)  |
|  Posterior median (95% BCI) | 5.3% (3.3%, 8.0%) | 6.7% (4.4%, 9.6%) | 4.4% (2.9%, 6.4%) | 6.2% (4.3%, 8.6%)  |
|  Difference (TAVR- SAVR) posterior median (95% BCI) | -1.4% (-4.9%, 2.1%) |   | -1.8% (-4.6%, 1.0%)  |   |
|  Primary objective – Non-inferiority  |   |   |   |   |
|  Posterior probability $$P(H_{A,S=0.06}|data)$$ | > 0.999 |   | > 0.999  |   |
|  Posterior threshold for non-inferiority | 0.972 |   |   |   |
|  Non-inferiority test | Passed |   |   |   |

\*Conducted on the original dataset

†Conducted on the expanded dataset

PMA P130021/S058: FDA Summary of Safety and Effectiveness Data

Page 15

{16}

The Kaplan-Meier (K-M) curve of all-cause mortality or disabling stroke is shown in Figure 8.

**Figure 8: All-Cause Mortality or Disabling Stroke through 24 Months (AT Population)**

![img-7.jpeg](img-7.jpeg)

**Note:** The confidence intervals were calculated without multiplicity adjustment. The adjusted confidence intervals could be wider than presented here. As such, confidence intervals are provided to illustrate the variability only and should not be used to draw any statistical conclusion.

## 2. Secondary Endpoints

### Hypothesis testing

Hypothesis testing was performed hierarchically on pre-specified secondary endpoints based on the original dataset, as shown in Table 6. TAVR was found to be non-inferior to SAVR within the pre-specified non-inferiority margins in terms of mean gradient and effective orifice area (EOA) at 12 months, the NYHA functional classification change from baseline to 12 months, and the Kansas City Cardiomyopathy Questionnaire (KCCQ) overall score change from baseline to 12 months. TAVR was found to be superior to SAVR with respect to mean gradient and EOA at 12 months and the KCCQ score change from baseline to 30 days (posterior probability > 0.999 for all).

PMA P130021/S058: FDA Summary of Safety and Effectiveness Data

Page 16

{17}

**Table 6: Secondary Endpoints Hierarchical Testing**

| Secondary Endpoint | TAVR Mean±SD (N) | SAVR Mean±SD (N) | Difference (TAVR – SAVR) (90% BCI) | Posterior Probability Prob (H_{A} | data) | Threshold | Test Result |
| --- | --- | --- | --- | --- | --- | --- | --- |
| **Non-inferiority testing** |
| #1 Mean gradient at 12 months | 8.6 ± 3.7 (409) | 11.2 ± 4.9 (339) | -2.6 (-3.1, -2.1) | >0.999 | 0.95 | Passed |
| #2 EOA at 12 months | 2.3 ± 0.7 (341) | 2.0 ± 0.6 (293) | 0.3 (0.2, 0.4) | >0.999 | 0.95 | Passed |
| #3 NYHA change (baseline – 12 months) | 0.9 ± 0.7 (428) | 1.0 ± 0.7 (342) | -0.1 (-0.2, 0.0) | >0.999 | 0.95 | Passed |
| #3 KCCQ overall score change (12 months – baseline) | 22.2 ± 20.3 (428) | 20.9 ± 21.0 (347) | 1.3 (-1.2, 3.8) | >0.999 | 0.95 | Passed |
| Secondary Endpoint | TAVR Mean±SD (N) | SAVR Mean±SD (N) | Difference (TAVR – SAVR) (95% BCI) | Posterior Probability Prob (H | data) | Threshold | Test Result |
| **Superiority testing** |
| #4 Mean gradient at 12 months | 8.6 ± 3.7 (409) | 11.2 ± 4.9 (339) | -2.6 (-3.2, -2.0) | >0.999 | 0.975 | Passed |
| #5 EOA at 12 months | 2.3 ± 0.7 (341) | 2.0 ± 0.6 (293) | 0.3 (0.2, 0.4) | >0.999 | 0.975 | Passed |
| #6 KCCQ overall score change (30 day – baseline) | 20.0 ± 21.1 (713) | 9.1 ± 22.3 (636) | 10.9 (8.6, 13.2) | >0.999 | 0.975 | Passed |

Note: The Implanted population was used for the mean gradient and EOA, and the AT population was used for the rest. All testing was conducted on the original dataset.

### Valve Performance

The EOA, mean aortic gradient, total aortic regurgitation (AR), and paravalvular regurgitation values obtained over time for the TAVR and SAVR patients are shown in Figure 9 through Figure 12, respectively. The increase in EOA and decrease in gradient were sustained through 24 months in both cohorts. In the TAVR cohort, the proportion of patients with total AR ≥ moderate was 4.6% at 12 months and 5.6% at 24 months, while in the SAVR cohort, the corresponding proportion was 1.4% at 12 months and 2.1% at 24 months. The proportion of patients with paravalvular regurgitation ≥ moderate was 4.0% at 12 months and 4.1% at 24 months in the TAVR cohort, as compared to 0.8% at 12 months and 0.7% at 24 months in the SAVR cohort.

PMA P130021/S058: FDA Summary of Safety and Effectiveness Data

Page 17

{18}

**Figure 9: Effective Orifice Area through 24 Months (Implanted Population)**

![img-8.jpeg](img-8.jpeg)

Note: Line plot with mean and standard deviation.

**Figure 10: Mean Aortic Gradient through 24 Months (Implanted Population)**

![img-9.jpeg](img-9.jpeg)

Note: Line plot with mean and standard deviation.

PMA P130021/S058: FDA Summary of Safety and Effectiveness Data

Page 18

{19}

**Figure 11: Total Aortic Regurgitation (Implanted Population)**

![img-10.jpeg](img-10.jpeg)

Note: Values < 1.0% are not labeled.

**Figure 12: Paravalvular Aortic Regurgitation by Visit (Implanted Population)**

![img-11.jpeg](img-11.jpeg)

Note: Values < 1.0% are not labeled.

PMA P130021/S058: FDA Summary of Safety and Effectiveness Data

Page 19

{20}

# NYHA Functional Class

The NYHA classifications by visit are presented in Figure 13. At baseline, 25.1% of TAVR patients and 28.4% of SAVR patients were in NYHA III/IV. At 24 months, this percentage decreased to 3.3% in TAVR patients and 1.3% in SAVR patients.

Figure 13: NYHA Classification by Visit (AT Population)

![img-12.jpeg](img-12.jpeg)

Note: Values < 1.0% are not labeled.

# QoL

# KCCQ

The KCCQ overall and clinical summary scores for the two treatment cohorts are shown in Figure 14 and Figure 15, respectively. In TAVR patients, the mean KCCQ overall summary score increased from 68.7 at baseline to 90.5 at 12 months and 90.1 at 24 months, and the mean KCCQ clinical summary score increased from 74.4 at baseline to 89.8 at 12 months and 89.5 at 24 months. Similar trends were observed in SAVR patients.

PMA P130021/S058: FDA Summary of Safety and Effectiveness Data

Page 20

{21}

**Figure 14: KCCQ Overall Summary Score (AT Population)**

![img-13.jpeg](img-13.jpeg)

Note: Line plot with mean and standard deviation.

**Figure 15: KCCQ Clinical Summary Score (AT Population)**

![img-14.jpeg](img-14.jpeg)

Note: Line plot with mean and standard deviation.

PMA P130021/S058: FDA Summary of Safety and Effectiveness Data

Page 21

{22}

### *EuroQoL (EQ-5D)*

The EQ-5D index scores for the two treatment cohorts are shown in Figure 16. The mean score was 0.7 at baseline, 0.8 at 30 days, and 0.8 at 12 months in TAVR patients, as compared to 0.7 at baseline, 0.7 at 30 days, and 0.8 at 12 months in SAVR patients.

**Figure 16: EQ-5D Index Score (AT Population)**

![img-15.jpeg](img-15.jpeg)

**Note:** Line plot with mean and standard deviation.

### 3. Adverse Events

The key adverse events that occurred in the trial through 24 months are presented in Table 7.

**Table 7: Adverse Events through 24 Months (AT Population)**

|  Events | Kaplan-Meier Rate *  |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- | --- |
|   |  0-30 Days |   | 0-12 Months |   | 0-24 Months  |   |
|   |  TAVR | SAVR | TAVR | SAVR | TAVR | SAVR  |
|  All-cause mortality or disabling stroke | 0.7% (5, 6) | 2.5% (17, 20) | 2.6% (18, 21) | 4.5% (29, 34) | 4.6% (24, 28) | 5.9% (33, 39)  |
|  All-cause mortality | 0.4% (3, 3) | 1.2% (8, 8) | 2.2% (15, 15) | 2.8% (18, 18) | 4.0% (20, 20) | 3.6% (21, 21)  |
|  Cardiovascular | 0.4% (3, 3) | 1.2% (8, 8) | 1.6% (11, 11) | 2.5% (16, 16) | 2.7% (14, 14) | 2.8% (17, 17)  |

PMA P130021/S058: FDA Summary of Safety and Effectiveness Data

Page 22

{23}

|  Events | Kaplan-Meier Rate *  |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- | --- |
|   |  0-30 Days |   | 0-12 Months |   | 0-24 Months  |   |
|   |  TAVR | SAVR | TAVR | SAVR | TAVR | SAVR  |
|  Non-cardiovascular | 0.0% (0, 0) | 0.0% (0, 0) | 0.6% (4, 4) | 0.3% (2, 2) | 1.3% (6, 6) | 0.8% (4, 4)  |
|  Reintervention | 0.3% (2, 2) | 0.3% (2, 2) | 0.6% (4, 4) | 0.5% (3, 3) | 0.8% (5, 5) | 1.3% (5, 5)  |
|  All stroke | 3.5% (25, 25) | 3.3% (22, 23) | 4.3% (31, 33) | 4.4% (29, 31) | 6.4% (37, 39) | 6.4% (33, 35)  |
|  Disabling stroke | 0.4% (3, 3) | 1.6% (11, 12) | 0.8% (6, 6) | 2.3% (15, 16) | 1.5% (8, 8) | 3.1% (17, 18)  |
|  Non-disabling stroke | 3.0% (22, 22) | 1.6% (11, 11) | 3.5% (25, 27) | 2.2% (15, 15) | 4.9% (29, 31) | 3.4% (17, 17)  |
|  Life threatening/disabling bleeding | 2.3% (17, 17) | 7.5% (51, 51) | 3.5% (25, 25) | 8.7% (58, 59) | 4.1% (28, 28) | 8.7% (58, 59)  |
|  Major vascular complication | 3.7% (27, 27) | 3.1% (21, 21) | 3.7% (27, 27) | 3.4% (23, 23) | 4.2% (28, 28) | 3.7% (24, 24)  |
|  Acute kidney injury - Stage 3 | 0.4% (3, 3) | 1.8% (12, 12) | 0.4% (3, 3) | 1.8% (12, 12) | 0.4% (3, 3) | 1.8% (12, 12)  |
|  Myocardial infarction | 0.8% (6, 6) | 1.3% (9, 9) | 1.8% (13, 15) | 1.6% (11, 12) | 2.0% (14, 16) | 1.6% (11, 12)  |
|  Aortic valve hospitalization^{†} | 1.1% (8, 8) | 2.4% (16, 17) | 3.3% (23, 29) | 6.2% (40, 44) | 5.0% (30, 39) | 7.5% (44, 53)  |
|  New permanent pacemaker implantation^{‡} | 17.3% (125, 125) | 6.1% (41, 41) | 19.1% (138, 138) | 6.7% (45, 45) | 22.7% (150, 150) | 7.6% (48, 48)  |

*Kaplan-Meier rate (# patients, # events).

†Not adjudicated by CEC.

‡Patients with pacemaker or ICD at baseline were not counted as new events. Not adjudicated by CEC.

The patient prosthesis mismatch adjudicated by the core laboratory is summarized in Table 8.

**Table 8: Patient Prosthesis Mismatch (Implanted Population)**

|  Severity^{†} | Summary Statistics*  |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- | --- |
|   |  30 Days |   | 12 Months |   | 24 Months  |   |
|   |  TAVR | SAVR | TAVR | SAVR | TAVR | SAVR  |
|  Severe | 1.1% (7/610) | 4.4% (24/545) | 1.8% (9/489) | 6.8% (30/438) | 1.3% (2/154) | 2.5% (3/120)  |
|  Moderate | 10.0% (61/610) | 16.0% (87/545) | 5.5% (27/489) | 16.7% (73/438) | 7.1% (11/154) | 14.2% (17/120)  |

PMA P130021/S058: FDA Summary of Safety and Effectiveness Data

Page 23

{24}

|  Severity† | Summary Statistics*  |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- | --- |
|   |  30 Days |   | 12 Months |   | 24 Months  |   |
|   |  TAVR | SAVR | TAVR | SAVR | TAVR | SAVR  |
|  None | 88.9% (542/610) | 79.6% (434/545) | 92.6% (453/489) | 76.5% (335/438) | 91.6% (141/154) | 83.3% (100/120)  |

*Observed rate - % (no./total no.)

†Severe: (Body mass index [BMI] < 30 and effective orifice area index [EOAI] < 0.65) OR (BMI ≥ 30 and EOAI < 0.60); moderate: (BMI < 30 and 0.65 ≤ EOAI ≤ 0.85) OR (BMI ≥ 30 and 0.60 ≤ EOAI ≤ 0.70); none: (BMI < 30 and EOAI > 0.85) OR (BMI ≥ 30 and EOAI > 0.70)

### 4. Subgroup Analyses

The protocol specified subgroup analyses of the primary endpoint of all-cause mortality or disabling stroke at 24 months by randomization designation (TAVR vs. SAVR) for patients with and without revascularization and for patients of different genders.

All-Cause Mortality or Disabling Stroke Stratified by Need for Revascularization:

The K-M curves of all-cause mortality or disabling stroke are shown in Figures 17 and 18 for patients with and without the need for concomitant revascularization, respectively.

PMA P130021/S058: FDA Summary of Safety and Effectiveness Data

Page 24

{25}

**Figure 17: All-Cause Mortality or Disabling Stroke for Patients with Need for Revascularization – AT Population**

![img-16.jpeg](img-16.jpeg)

**Note:** The confidence intervals were calculated without multiplicity adjustment. The adjusted confidence intervals could be wider than presented here. As such, confidence intervals are provided to illustrate the variability only and should not be used to draw any statistical conclusion. The trial was not powered to assess the difference between the two subgroups.

PMA P130021/S058: FDA Summary of Safety and Effectiveness Data

Page 25

{26}

**Figure 18: All-Cause Mortality or Disabling Stroke for Patients without Need for Revascularization – AT Population**

![img-17.jpeg](img-17.jpeg)

**Note:** The confidence intervals were calculated without multiplicity adjustment. The adjusted confidence intervals could be wider than presented here. As such, confidence intervals are provided to illustrate the variability only and should not be used to draw any statistical conclusion. The trial was not powered to assess the difference between the two subgroups.

PMA P130021/S058: FDA Summary of Safety and Effectiveness Data

Page 26

{27}

# All-Cause Mortality or Disabling Stroke Stratified by Gender:

The K-M curves of all-cause mortality or disabling stroke are shown in Figures 19 and 20, for the male and female patients, respectively.

**Figure 19: All-Cause Mortality or Disabling Stroke for Male Patients - AT Population**

![img-18.jpeg](img-18.jpeg)

**Note:** The confidence intervals were calculated without multiplicity adjustment. The adjusted confidence intervals could be wider than presented here. As such, confidence intervals are provided to illustrate the variability only and should not be used to draw any statistical conclusion. The trial was not powered to assess the difference between the two subgroups.

PMA P130021/S058: FDA Summary of Safety and Effectiveness Data

Page 27

{28}

**Figure 20: All-Cause Mortality or Disabling Stroke for Female Patients - AT Population**

![img-19.jpeg](img-19.jpeg)

**Note:** The confidence intervals were calculated without multiplicity adjustment. The adjusted confidence intervals could be wider than presented here. As such, confidence intervals are provided to illustrate the variability only and should not be used to draw any statistical conclusion. The trial was not powered to assess the difference between the two subgroups.

## 5. Other Study Observations

### Procedure Data

The procedure data of the TAVR and SAVR cohorts are summarized in Table 9 and Table 10, respectively.

**Table 9: TAVR Procedure Data (AT Population)**

|  Procedure Data | Summary Statistics* (N=725)  |
| --- | --- |
|  Number of index procedures | 724  |
|  Total delivery catheter in the body time (min) | 17.4 ± 19.4  |
|  Type of anesthesia |   |
|  General | 56.9% (412/724)  |
|  Local | 43.1% (312/724)  |
|  Access site |   |
|  Iliofemoral | 99.0% (717/724)  |

PMA P130021/S058: FDA Summary of Safety and Effectiveness Data

Page 28

{29}

|  Procedure Data | Summary Statistics* (N=725)  |
| --- | --- |
|  Non-iliofemoral | 1.0% (7/724)  |
|  Valve size |   |
|  23 mm | 1.2% (9/721)  |
|  26 mm | 19.6% (141/721)  |
|  29 mm | 42.7% (308/721)  |
|  31 mm | 3.6% (26/721)  |
|  34 mm | 32.9% (237/721)  |
|  Total time in catheterization laboratory or operating room (min) | 148.2 ± 55.1  |
|  Embolic protection device used | 1.2% (9/722)  |
|  Pre-TAVR balloon valvuloplasty performed | 34.9% (253/724)  |
|  Post-TAVR balloon valvuloplasty performed | 31.3% (226/723)  |
|  Concomitant procedure (percutaneous coronary intervention; PCI) | 6.9% (50/724)  |
|  Length of index hospitalization (days) | 2.6 ± 2.1  |

*Continuous measures - Mean ± SD; categorical measures - % (no./total no.). Data included subjects with the index procedure defined as the first procedure in which the delivery catheter was introduced. If a patient had two implant procedures, the index procedure was used.

**Table 10: SAVR Procedure Data (AT Population)**

|  Procedure Data | Summary Statistics*  |
| --- | --- |
|   |  SAVR (N=678)  |
|  Procedure aborted† | 0.4% (3/678)  |
|  Valve size |   |
|  19 mm | 3.6% (24/675)  |
|  21 mm | 18.4% (124/675)  |
|  23 mm | 31.3% (211/675)  |
|  25 mm | 28.0% (189/675)  |
|  27 mm | 7.3% (49/675)  |
|  29 mm | 0.4% (3/675)  |
|  Other‡ | 11.1% (75/675)  |
|  Total aortic cross clamp time (min) | 68.6 ± 28.9  |
|  Total time in catheterization laboratory or operating room (min) | 276.6 ± 79.5  |
|  SAVR approach |   |

PMA P130021/S058: FDA Summary of Safety and Effectiveness Data

Page 29

{30}

|  Procedure Data | Summary Statistics*  |
| --- | --- |
|   |  SAVR (N=678)  |
|  Full sternotomy | 65.9% (446/677)  |
|  Mini sternotomy | 14.5% (98/677)  |
|  Right anterior thoracotomy | 19.4% (131/677)  |
|  Other | 0.3% (2/677)  |
|  Concomitant procedures^{§}  |   |
|  Aortic root enlargement | 1.6% (11/678)  |
|  Coronary artery bypass grafting (CABG) | 13.6% (92/678)  |
|  Permanent pacemaker implantation | 0.0% (0/678)  |
|  Surgical treatment of atrial fibrillation | 3.5% (24/678)  |
|  Automatic implantable cardioverter-defibrillator (AICD) implantation | 0.0% (0/678)  |
|  Left atrial appendage (LAA) closure | 6.2% (42/678)  |
|  Patent foramen ovale (PFO) closure | 0.7% (5/678)  |
|  Mitral valve repair | 0.6% (4/678)  |
|  Mitral valve replacement | 0.0% (0/678)  |
|  Other | 5.0% (34/678)  |
|  Length of index hospitalization (days) | 6.2 ± 3.3  |

*Continuous measures - mean ± SD (n); categorical measures - % (no./Total no.).

†Adjudicated by CEC: Aborted procedure or SAVR conversion to alternate procedure.

‡Others included sutureless valves categorized as “S,” “M,” or “L” for valve size.

§Subjects might have more than one concomitant procedure.

## CT Substudy

There were 197 TAVR and 177 SAVR patients at 30 days and 112 and 94 patients at 12 months, respectively, who had an adequate CT for leaflet assessments at both time points. The HALT and leaflet mobility imaging findings are summarized in Table 11, along with the associated mean aortic pressure gradients. The mean aortic pressure gradients at 12 months stratified by HALT and leaflet mobility at 30 days are summarized in Table 12 and Table 13, respectively. The rate of death, stroke or TIA at 1 year stratified by HALT and leaflet mobility at 30 days are summarized in Table 14 and Table 15, respectively. The CT substudy was not powered to compare the relative incidence or the severity of HALT or reduced leaflet mobility between the TAVR and SAVR cohorts, or to determine whether late clinical outcomes were affected by the presence of HALT or reduced leaflet mobility.

PMA P130021/S058: FDA Summary of Safety and Effectiveness Data

Page 30

{31}

**Table 11: HALT and Leaflet Mobility Findings and Associated Mean Gradients**

|  Findings | Summary Statistics*  |   |   |   |
| --- | --- | --- | --- | --- |
|   |  At 30 Days |   | At 12 Months  |   |
|   |  TAVR (N=197) | SAVR (N=177) | TAVR (N=112) | SAVR (N=94)  |
|  Proportion of patients on oral anticoagulants at time of scan^{†} | 9.1% (18/197) | 22.0% (39/177) | 12.5% (14/112) | 10.6% (10/94)  |
|  **HALT^{‡}**  |   |   |   |   |
|  No HALT (no thickening) | 82.2% (162/197) | 87.6% (155/177) | 70.5% (79/112) | 75.5% (71/94)  |
|  Mean gradient (mmHg) | 8.6 ± 3.6 (160) | 10.5 ± 3.6 (153) | 8.2 ± 3.2 (77) | 11.4 ± 4.6 (69)  |
|  Presence of HALT | 17.8% (35/197) | 12.4% (22/177) | 29.5% (33/112) | 24.5% (23/94)  |
|  <25% leaflet length thickened | 10.7% (21/197) | 2.3% (4/177) | 17.0% (19/112) | 6.4% (6/94)  |
|  Mean gradient (mmHg) | 7.2 ± 3.0 (21) | 9.2 ± 4.6 (4) | 8.4 ± 2.5 (19) | 8.6 ± 2.5 (6)  |
|  25%-50% leaflet length thickened | 3.0% (6/197) | 4.5% (8/177) | 7.1% (8/112) | 8.5% (8/94)  |
|  Mean gradient (mmHg) | 8.1 ± 1.6 (6) | 11.1 ± 3.8 (8) | 7.2 ± 2.4 (7) | 10.4 ± 4.1 (8)  |
|  50%-75% leaflet length thickened | 2.0% (4/197) | 3.4% (6/177) | 3.6% (4/112) | 6.4% (6/94)  |
|  Mean gradient (mmHg) | 6.8 ± 3.0 (2) | 12.2 ± 5.6 (6) | 7.9 ± 4.9 (4) | 11.9 ± 3.9 (6)  |
|  >75% leaflet length thickened | 2.0% (4/197) | 1.7% (3/177) | 1.8% (2/112) | 3.2% (3/94)  |
|  Mean gradient (mmHg) | 5.9 ± 1.4 (4) | 6.9 ± 3.5 (3) | 11.6 ± NA (1) | 10.0 ± 2.3 (3)  |
|  **Number of leaflets with HALT**  |   |   |   |   |
|  0 leaflet | 82.2% (162/197) | 87.6% (155/177) | 70.5% (79/112) | 75.5% (71/94)  |
|  1 leaflet thickening | 11.7% (23/197) | 5.1% (9/177) | 13.4% (15/112) | 13.8% (13/94)  |
|  2 leaflets thickening | 5.1% (10/197) | 5.1% (9/177) | 12.5% (14/112) | 8.5% (8/94)  |

PMA P130021/S058: FDA Summary of Safety and Effectiveness Data

Page 31

{32}

|  Findings | Summary Statistics*  |   |   |   |
| --- | --- | --- | --- | --- |
|   |  At 30 Days |   | At 12 Months  |   |
|   |  TAVR (N=197) | SAVR (N=177) | TAVR (N=112) | SAVR (N=94)  |
|  3 leaflets thickening | 1.0% (2/197) | 2.3% (4/177) | 3.6% (4/112) | 2.1% (2/94)  |
|  **Leaflet mobility^{§}**  |   |   |   |   |
|  Unrestricted | 84.6% (148/175) | 89.0% (153/172) | 70.6% (77/109) | 77.5% (69/89)  |
|  Mean gradient (mmHg) | 8.6 ± 3.7 (146) | 10.5 ± 3.6 (151) | 8.2 ± 3.2 (76) | 11.3 ± 4.6 (67)  |
|  Partially restricted (<25%) | 9.7% (17/175) | 5.2% (9/172) | 20.2% (22/109) | 7.9% (7/89)  |
|  Mean gradient (mmHg) | 7.6 ± 3.2 (17) | 9.6 ± 3.4 (9) | 8.3 ± 2.6 (22) | 7.7 ± 2.7 (7)  |
|  Partially restricted (25%-50%) | 3.4% (6/175) | 4.1% (7/172) | 6.4% (7/109) | 10.1% (9/89)  |
|  Mean gradient (mmHg) | 7.0 ± 2.1 (5) | 12.8 ± 5.5 (7) | 8.0 ± 2.0 (6) | 11.8 ± 3.7 (9)  |
|  Partially restricted (50%-75%) | 1.7% (3/175) | 1.2% (2/172) | 1.8% (2/109) | 3.4% (3/89)  |
|  Mean gradient (mmHg) | 7.8 ± 1.6 (2) | 10.6 ± 6.3 (2) | 9.8 ± 6.8 (2) | 12.4 ± 3.4 (3)  |
|  Largely immobile | 0.6% (1/175) | 0.6% (1/172) | 0.9% (1/109) | 1.1% (1/89)  |
|  Mean gradient (mmHg) | 5.9 ± NA (1) | 9.7 ± NA (1) | NA (0) | 11.0 ± NA (1)  |
|  **Number of leaflets partially restricted or largely immobile**  |   |   |   |   |
|  0 leaflet | 84.6% (148/175) | 89.0% (153/172) | 70.6% (77/109) | 77.5% (69/89)  |
|  1 leaflet | 10.3% (18/175) | 4.1% (7/172) | 13.8% (15/109) | 12.4% (11/89)  |
|  2 leaflets | 4.0% (7/175) | 4.7% (8/172) | 11.9% (13/109) | 7.9% (7/89)  |

PMA P130021/S058: FDA Summary of Safety and Effectiveness Data

Page 32

{33}

|  Findings | Summary Statistics*  |   |   |   |
| --- | --- | --- | --- | --- |
|   |  At 30 Days |   | At 12 Months  |   |
|   |  TAVR (N=197) | SAVR (N=177) | TAVR (N=112) | SAVR (N=94)  |
|  3 leaflets | 1.1% (2/175) | 2.3% (4/172) | 3.7% (4/109) | 2.2% (2/89)  |

*Continuous measures - mean ± SD (n); categorical measures - % (no./total no.). The analysis population for the 30-day analysis included all the patients enrolled in the CT substudy and had an adequate CT for leaflet assessments at 30 days; the analysis population for the 12-month analysis had an adequate CT for leaflet assessments at both time points.

†During the course of the substudy enrollment, a protocol amendment removed the requirement for discontinuation of anticoagulation therapy prior to the CT scan at 30 days.

‡HALT was defined as: the presence of any hypoattenuated leaflet thickening in any singular leaflet as identified by an independent CT core laboratory. The extent of the hypoattenuated leaflet thickening was graded with regards to the entire leaflet as: none, <25%, 25-50%, 50-75%, >75%. If more than one leaflet had the appearance of HALT, the thickening measure of the most impacted leaflet was used. One SAVR subject was identified as having one thickened leaflet; however, the extent of thickening was not recorded, and the percentages do not sum to 100%.

§Leaflet mobility was determined by an independent CT core laboratory and included: unrestricted, partially restricted mobility limited to the base of a leaflet, partially restricted mobility involving more than the base of the leaflet but less than 50% of the leaflet, partially restricted mobility involving more than 50% of the leaflet but less than 75% of the leaflet, and/or a largely immobile leaflet. Presence of immobility any degree of restriction or immobility on any one leaflet rendered a finding.

**Table 12: Mean Aortic Gradient at 1 Year Stratified by HALT at 30 Days**

|   | Summary Statistics*  |   |   |   |
| --- | --- | --- | --- | --- |
|   |  No HALT at 30 Days |   | HALT at 30 Days  |   |
|   |  TAVR (N=162) | SAVR (N=155) | TAVR (N=35) | SAVR (N=22)  |
|  Mean gradient | 8.1 ± 2.9 (112) | 11.5 ± 4.4 (93) | 6.8 ± 3.4 (18) | 10.1 ± 3.8 (17)  |

*Mean ± SD (n). The analysis population included all the patients enrolled in the CT substudy and had an adequate CT for leaflet assessments at 30 days.

PMA P130021/S058: FDA Summary of Safety and Effectiveness Data

Page 33

{34}

**Table 13: Mean Aortic Gradient at 1 Year Stratified by Leaflet Mobility at 30 Days**

|   | Summary Statistics*  |   |   |   |
| --- | --- | --- | --- | --- |
|   |  Unrestricted at 30 Days |   | Reduced Leaflet Mobility at 30 Days  |   |
|   |  TAVR (N=148) | SAVR (N=153) | TAVR (N=27) | SAVR (N=19)  |
|  Mean gradient | 7.9 ± 2.7 (98) | 11.5 ± 4.5 (91) | 6.5 ± 3.6 (14) | 10.5 ± 3.8 (15)  |

*Mean ± SD (n). The analysis population included all the patients enrolled in the CT substudy and had an adequate CT for leaflet assessments at 30 days.

**Table 14: All-Cause Mortality, All Stroke or TIA at 1 Year Stratified by HALT at 30 Days**

|  1-Year Endpoint | Kaplan-Meier Rate*  |   |   |   |
| --- | --- | --- | --- | --- |
|   |  No HALT at 30 Days |   | HALT at 30 Days  |   |
|   |  TAVR (N=162) | SAVR (N=155) | TAVR (N=35) | SAVR (N=22)  |
|  All-cause mortality | 0.0% (0, 0) | 0.9% (1, 1) | 0.0% (0, 0) | 4.5% (1, 1)  |
|  All stroke | 2.5% (4, 4) | 1.9% (3, 3) | 2.9% (1, 2) | 0.0% (0, 0)  |
|  TIA | 1.9% (3, 3) | 0.0% (0, 0) | 5.7% (2, 2) | 0.0% (0, 0)  |
|  All-cause mortality or all stroke or TIA | 4.3% (7, 7) | 2.8% (4, 4) | 8.6% (3, 4) | 4.5% (1, 1)  |

*Kaplan-Meier rate (# patients, # events). The analysis population included all the patients enrolled in the CT substudy and had an adequate CT for leaflet assessments at 30 days. The Kaplan-Meier analysis used the procedure date as the start date in determining time to event. Presence of any degree of HALT on any one leaflet rendered a finding and inclusion in the HALT cohort.

PMA P130021/S058: FDA Summary of Safety and Effectiveness Data

Page 34

{35}

**Table 15: All-Cause Mortality, All Stroke or TIA at 1 Year Stratified by Leaflet Mobility at 30 Days**

|  1-Year Endpoint | Kaplan-Meier Rate*  |   |   |   |
| --- | --- | --- | --- | --- |
|   |  Unrestricted at 30 Days |   | Reduced Leaflet Mobility at 30 Days  |   |
|   |  TAVR (N=148) | SAVR (N=153) | TAVR (N=27) | SAVR (N=19)  |
|  All-cause mortality | 0.0% (0, 0) | 0.9% (1, 1) | 0.0% (0, 0) | 5.3% (1, 1)  |
|  All stroke | 2.7% (4, 4) | 2.0% (3, 3) | 3.7% (1, 2) | 0.0% (0, 0)  |
|  TIA | 1.4% (2, 2) | 0.0% (0, 0) | 7.4% (2, 2) | 0.0% (0, 0)  |
|  All-cause mortality or all stroke or TIA | 4.1% (6, 6) | 2.8% (4, 4) | 11.1% (3, 4) | 5.3% (1, 1)  |

*Kaplan-Meier rate (# patients, # events). The analysis population included all the patients enrolled in the CT substudy and had an adequate CT for leaflet assessments at 30 days. The Kaplan-Meier analysis used the procedure date as the start date in determining time to event. The presence of any degree of restriction or immobility on any one leaflet rendered a finding and inclusion in the reduced leaflet mobility cohort.

6. Pediatric Extrapolation

In this premarket application, existing clinical data was not leveraged to support approval of a pediatric patient population.

**E. Financial Disclosure**

The Financial Disclosure by Clinical Investigators regulation (21 CFR 54) requires applicants who submit a marketing application to include certain information concerning the compensation to, and financial interests and arrangement of, any clinical investigator conducting clinical studies covered by the regulation. The pivotal clinical study included 1032 investigators, of which none were full-time or part-time employees of the sponsor and 46 had disclosable financial interests/arrangements related to the Low Risk study as defined in 21 CFR 54.2(a), (b), (c) and (f) and described below:

- Compensation to the investigator for conducting the study where the value could be influenced by the outcome of the study: 12
- Significant payment of other sorts: 34
- Proprietary interest in the product tested held by the investigator: 0
- Significant equity interest held by investigator in sponsor of covered study: 2

The applicant has adequately disclosed the financial interest/arrangements with clinical investigators. Statistical analyses were conducted by FDA to determine whether the financial interests/arrangements had any impact on the clinical study outcome. The

PMA P130021/S058: FDA Summary of Safety and Effectiveness Data

Page 35

{36}

information provided does not raise any questions about the reliability of the data.

## XI. PANEL MEETING RECOMMENDATION AND FDA'S POST-PANEL ACTION

In accordance with the provisions of section 515(c)(2) of the Act as amended by the Safe Medical Devices Act of 1990, this PMA was not referred to the Circulatory System Devices panel, an FDA advisory committee, for review and recommendation because the information in the PMA substantially duplicates information previously reviewed by this panel.

## XII. CONCLUSIONS DRAWN FROM THE PRECLINICAL AND CLINICAL STUDIES

### A. Effectiveness Conclusions

In the clinical study, patients overall demonstrated clinically significant improvement in valve hemodynamics from baseline to 12 months. On average, the EOA increased from 0.8 cm² at baseline to 2.2 cm² at 12 months, and the mean pressure gradient decreased from 44.8 mmHg at baseline to 8.7 mmHg at 12 months in the TAVR patients. These trends were consistent with those observed in the SAVR patients. In the TAVR cohort, the proportion of patients with total AR ≥ moderate was 4.6% at 12 months, while in the SAVR cohort, the proportion was 1.4% at 12 months. The proportion of patients with paravalvular AR ≥ moderate was 4.0% at 12 months, as compared to 0.8% at 12 months in the SAVR cohort.

The improvement in valve hemodynamics in the TAVR patients was further demonstrated through improvements in NYHA classification and QoL. In the TAVR cohort, about 1.7% of the patients were in NYHA Class III or IV at 12 months as compared to 25.1% at baseline; similar results were seen in the SAVR cohort. In addition, clinically significant improvement in the KCCQ overall summary score was observed in the TAVR patients, which increased from 68.7 at baseline to 88.7 and 90.5 at 30 days and 12 months, respectively. Furthermore, the mean total time in the catheterization laboratory or operating room and index procedure hospital stay were 148.2 minutes and 2.6 days, respectively, for TAVR, which were significantly longer for SAVR (276.6 minutes and 6.2 days, respectively).

### B. Safety Conclusions

The risks of the device are based on nonclinical laboratory and animal studies as well as data collected in the clinical study conducted to support PMA approval as described above. The results from the nonclinical laboratory (e.g., biocompatibility, hydrodynamic performance, durability, and structural integrity) and animal studies demonstrated that the device is suitable for long-term implant.

The posterior median estimate of all-cause mortality or disabling stroke at 24 months (i.e., the primary endpoint) was 5.3% for TAVR and 6.7% for SAVR. TAVR with the

PMA P130021/S058: FDA Summary of Safety and Effectiveness Data

Page 36

{37}

Evolut R or Evolut PRO TAV was found to be non-inferior to SAVR in the primary endpoint within a non-inferiority margin of 6% with a posterior probability of >0.999. The K-M rate of all stroke at 30 days was 3.5% for TAVR and 3.3% for SAVR, while the rates of disabling stroke were 0.4% and 1.6%, respectively. The K-M rate of all-cause mortality for TAVR was 0.4% at 30 days, 2.2% at 12 months, and 4.0% at 24 months, as compared to 1.2%, 2.8%, and 3.6%, respectively, for SAVR.

The CT substudy revealed that 17.8% and 29.5% of TAVR patients had various degree of leaflet thickening at 30 days and 12 months, respectively, as compared to 12.4% and 24.5% of SAVR patients. In addition, various degrees of restricted leaflet mobility were observed in 15.4% of patients at 30 days and 29.4% of patients at 12 months in the TAVR cohort, which was 11.0% and 22.5%, respectively, in the SAVR cohort. The long-term clinical sequelae of these imaging findings are presently unknown.

### C. Benefit-Risk Determination

The probable benefits of TAVR with the Evolut R or Evolut PRO TAV include improved valve hemodynamic performance, improved functional status as measured by the NYHA classification, and improved QoL at 1 year post-procedure.

The probable risks of TAVR with the Evolut R or Evolut PRO TAV include procedure-related complications such as death, stroke, myocardial infarction, major vascular complications, bleeding, conduction disturbance, and acute kidney injury.

#### **1. Patient Perspectives**

This application did not include specific information on patient perspectives for TAVR with the Evolut R or Evolut PRO TAV. However, since TAVR provides a less invasive alternative to SAVR, FDA believes that many patients would prefer the TAVR therapy. However, the long-term durability of the Evolut R or Evolut PRO TAV compared to surgically implanted valves have not been established. Patients, especially younger ones, should discuss available treatment options with their heart care team to select the appropriate therapy.

In conclusion, given the available information above, the data support that for patients with severe native aortic stenosis who are at low risk for open aortic valve replacement surgery, the probable benefits of TAVR with the Evolut R or Evolut PRO TAV outweigh the probable risks.

### D. Overall Conclusions

The data in this application support the reasonable assurance of safety and effectiveness of the Evolut R System and Evolut PRO System for the replacement of native aortic valves in symptomatic severe aortic stenosis patients who are deemed to be at low surgical risk.

PMA P130021/S058: FDA Summary of Safety and Effectiveness Data

Page 37

{38}

### **XIII. CDRH DECISION**

CDRH issued an approval order on August 16, 2019. The final conditions of approval cited in the approval order are described below:

The applicant must conduct one post-approval study as well as participate in and support continued surveillance:

1. **Post-Approval Study - Continued Follow-up of the Medtronic CoreValve Evolut R System and CoreValve Evolut PRO System “Low Risk” Indication Premarket Pivotal Cohort:** The study will consist of all living patients who were enrolled in the pivotal cohort under the IDE. The objective of this study is to characterize the clinical outcomes annually through 10 years post-procedure. The safety and effectiveness endpoints include all-cause mortality, all stroke (disabling and non-disabling), life-threatening bleeding, acute kidney injury at stage 2 or 3, coronary artery obstruction requiring intervention, major vascular complication, valve-related dysfunction requiring repeat procedure, new permanent pacemaker implantation, prosthetic valve endocarditis, prosthetic valve thrombosis, NYHA classification, KCCQ score, and hemodynamic performance metrics by Doppler echocardiography.

2. **Medtronic CoreValve Evolut R System and CoreValve Evolut PRO System Registry-Based Continued Access Protocol (CAP) Cohort and “Low Risk” Indication Real-World Use Surveillance:** The applicant has agreed to work with the Society of Thoracic Surgeons/American College of Cardiology Transcatheter Valve Therapy (TVT) Registry to ensure that FDA surveillance occurs for the registry-based CAP cohort per approved protocol and for commercial uses of the Medtronic CoreValve Evolut R System and CoreValve Evolut PRO System for the “low risk” indication. The surveillances will be carried out to characterize the clinical outcomes of the CAP cohort annually through 10 years post implantation and to assess the real-world use of the commercial Medtronic CoreValve Evolut R System and CoreValve Evolut PRO System to ensure that the device is used in appropriate circumstances, respectively. The surveillance of the CAP cohort will consist of all living CAP patients who were enrolled at participating institutions, and the surveillance of the real-world use will involve all consecutive patients treated within the first 2 years that are entered into the TVT Registry (enrollment period). The applicant has also agreed to link the data to the Centers for Medicare and Medicaid Services (CMS) claims database for long-term surveillance of these patients through 10 years post implantation (follow-up duration). This surveillance will monitor the following: (1) device success (intra-procedure); (2) all-cause mortality, all stroke, life-threatening/major bleeding, new requirement for dialysis, peri-procedural myocardial infarction, and repeat procedure for valve-related dysfunction (surgical or interventional therapy) at 30 days and 12 months; (3) neurological (non-stroke), vascular complications, and quality of life (KCCQ) outcomes at 30 days and 12 months; and (4) all-cause mortality, all stroke, and repeat procedure for valve-related dysfunction (surgical or interventional therapy) at 2-10 year post implantation.

PMA P130021/S058: FDA Summary of Safety and Effectiveness Data

Page 38

{39}

The applicant’s manufacturing facilities have been inspected and found to be in compliance with the device Quality System (QS) regulation (21 CFR 820).

# **XIV. APPROVAL SPECIFICATIONS**

Directions for use: See final approved labeling (Instructions for Use).

Hazards to health from use of the device: See indications, contraindications, warnings, precautions, and adverse events in the final labeling (Instructions for Use).

Post-approval requirements and restrictions: See approval order.

PMA P130021/S058: FDA Summary of Safety and Effectiveness Data

Page 39

---

**Source:** [https://fda.innolitics.com/device/P130021S058](https://fda.innolitics.com/device/P130021S058)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a PMA](https://innolitics.com/services/regulatory/), [a 510(k)](https://innolitics.com/services/510ks/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).