MAESTRO RECHARGEABLE SYSTEM

P130019 · Reshape Lifesciences, Inc. · PIM · Jan 14, 2015 · Gastroenterology, Urology

Device Facts

Record IDP130019
Device NameMAESTRO RECHARGEABLE SYSTEM
ApplicantReshape Lifesciences, Inc.
Product CodePIM · Gastroenterology, Urology
Decision DateJan 14, 2015
DecisionAPWD
Device ClassClass 3
AttributesTherapeutic

Indications for Use

The MAESTRO® Rechargeable System is indicated for use in weight reduction in patients aged 18 years through adulthood who have a Body Mass Index (BMI) of 40 to 45 kg/m², or a BMI of 35 to 39.9 kg/m² with one or more obesity related co-morbid conditions, and have failed at least one supervised weight management program within the past five years.

Device Story

The MAESTRO® Rechargeable System provides vagal blocking (VBLOC®) therapy to treat obesity. The system consists of an implantable rechargeable neuroregulator (RNR) and two leads (anterior and posterior) placed on the abdominal vagus nerve trunks. The RNR delivers intermittent, square, biphasic, constant current electrical pulses (up to 8 mA, 5000 Hz) to block vagal nerve signals, which reduces gastric accommodation, delays gastric emptying, and promotes satiety. The system is implanted subcutaneously on the thoracic sidewall. Patients use an external mobile charger and transmit coil to transcutaneously recharge the RNR battery. Clinicians use a programmer to adjust therapy parameters and schedules. The device is intended to assist in weight loss by reducing food intake. It is not MRI-safe.

Clinical Evidence

PMA approval based on the ReCharge pivotal study (n=239, randomized 2:1 VBLOC vs. Sham). Primary effectiveness endpoints (10% super-superiority margin in %EWL at 12 months; responder rates) were not met. However, VBLOC group showed statistically significant weight loss compared to Sham (24.4% vs 15.9% EWL at 12 months). Safety endpoint met: 12-month SAE rate related to device/procedure was 3.7% (VBLOC), meeting the <15% performance goal. Adverse events included neuroregulator site pain, heartburn, and abdominal pain. No deaths or unanticipated adverse events reported.

Technological Characteristics

Implantable neuroregulator (Model 2002) with Li-ion battery; leads (Model 2200-47E) with platinum/iridium electrodes and 316L stainless steel support. Delivers constant current (0-8 mA), 5000 Hz, 90 μs pulse width. Connectivity via inductive telemetry (6.78 MHz). Materials: silicone, platinum/iridium, 316L stainless steel. Sterilization: ethylene oxide. Software: Major level of concern. MRI unsafe.

Indications for Use

Indicated for weight reduction in adults (18+) with BMI 40-45 kg/m² or BMI 35-39.9 kg/m² with obesity-related comorbidities who failed at least one supervised weight management program in the past 5 years. Contraindicated in patients with cirrhosis, portal hypertension, esophageal varices, significant hiatal hernia, or those requiring MRI, diathermy, or having other implanted electrical medical devices.

Regulatory Classification

Identification

Weight reduction by peripheral neuromodulation.

Reference Devices

Submission Summary (Full Text)

{0} # Summary of Safety and Effectiveness Data (SSED) ## I. GENERAL INFORMATION Device Generic Name: Rechargeable vagal blocking system Device Trade Name: MAESTRO® Rechargeable System Device Procode: PIM Applicant’s Name and Address: EnteroMedics, Inc. 2800 Patton Road St. Paul, MN 55113 Date of Panel Recommendation: June 17, 2014 Premarket Approval Application (PMA) Number: P130019 Date of FDA Notice of Approval: January 14, 2015 ## II. INDICATIONS FOR USE The MAESTRO® Rechargeable System is indicated for use in weight reduction in patients aged 18 years through adulthood who have a Body Mass Index (BMI) of 40 to 45 kg/m², or a BMI of 35 to 39.9 kg/m² with one or more obesity related co-morbid conditions, and have failed at least one supervised weight management program within the past five years. ## III. CONTRAINDICATIONS The MAESTRO® Rechargeable System is contraindicated in individuals who: - Have cirrhosis of the liver, portal hypertension, esophageal varices or a clinically significant hiatal hernia; - Plan on needing magnetic resonance imaging (MRI); - Plan on needing shortwave, microwave, or therapeutic ultrasound diathermy; - Are at high risk for surgical complications; - Have a permanently implanted, electrical powered medical device, or gastrointestinal device or prosthesis (e.g., pacemakers, implanted defibrillators, or neurostimulators). PMA P130019: FDA Summary of Safety and Effectiveness Data {1} PMA P130019: FDA Summary of Safety and Effectiveness Data Page 2 # IV. WARNINGS AND PRECAUTIONS The warnings and precautions can be found in the MAESTRO® device labeling. # V. DEVICE DESCRIPTION The MAESTRO® Rechargeable System is comprised of implantable and external device components, which are listed in Table 1: Table 1: MAESTRO® Rechargeable system components | Implantable Component | Model numbers | | --- | --- | | Rechargeable Neuroregulator (RNR) | Model 2002 | | Posterior Lead-Marked | Model 2200P-47E | | Anterior Lead-Unmarked | Model 2200A-47E | | External Components | Model numbers | | --- | --- | | Mobile Charger (MC) | Model 2402 | | Patient Transmit Coil | Model 2403-60, Model 2403-60A | | Clinician Transmit Coil | Model 2403-300 | | Clinician Programmer (CP) | Model 2502 | | Programmer Cable | Model 1600 | | AC Recharger | Model 1620 | | Patient Transmit Coil Belt | Model 1660, Model 1660A | | Transmit Coil Foam Pad | Model 1670 | | Torque Wrench | Model 1680 | The MAESTRO System provides vagal blocking (VBLOC®) therapy by delivering intermittent, controllable electrical blocking signals to the abdominal anterior and posterior abdominal nerve trunks of the vagus nerve. ![img-0.jpeg](img-0.jpeg) Figure 1. MAESTRO® Rechargeable Neuroregulator and leads, with placement of the electrodes on the abdominal anterior and posterior nerve trunks. ![img-1.jpeg](img-1.jpeg) {2} The MAESTRO System is intended to promote weight loss by applying electrical pulse algorithms which block signals to the anterior and posterior trunks of the intra-abdominal vagus nerve. Other intended weight-reducing effects of electrical neural blockade include: - Reduced food intake by reducing gastric accommodation; - Promoting satiety by delaying food processing and gastric emptying; - Decreasing caloric intake. The System is designed to deliver square, biphasic, charge balanced, constant current to the abdominal nerve trunks (current range, 0-8 mA) at a pulse frequency of 5000 Hertz and a pulse width of 90 microseconds. The maximum charge density (8 mA) is estimated to be 5.3 μC/cm². ## Model 2002 Pulse Generator The Model 2002 Pulse Generator (RNR) is provided sterile and has a hermetic case enclosure with an integrated coil that acts as the telemetry and recharging antenna. The RNR is surgically implanted subcutaneously on the thoracic sidewall. The principal function of the RNR is to deliver current to the leads. It contains a Rechargeable 2.6 AH Li-ion battery (8 year battery life). When recharging, the mobile charger and transmit coil transfer power to the RNR and exchange information via a low power inductive telemetry link with a range of less than 5 cm. Power is transferred by emitting a continuous wave signal at 6.78 MHz, centered in the 6.765-6.795 MHz Industrial, Science and Medical (ISM) radio band. It is recharged transcutaneously using the transmit coil. The RNR is labeled as unsafe for magnetic resonance imaging (MRI). ## Model 2200-47E Leads (Anterior and Posterior) The Model 2200-47E Leads are provided sterile and are flexible and are approximately 47 cm in length. They contain bipolar platinum/iridium tip &amp; ring electrodes with an insulated lead body. The tip (i.e., nerve) electrode contains rigid (316L) stainless steel encased in silicone to provide structural support. Current is delivered to the nerve electrode via 90/10 platinum/iridium electrodes. A suture tongue anchors and stabilized the nerve electrode. The tip electrode measures lead impedance and delivers electrical pulses to the vagus nerve trunks. The ring electrode is sutured to the stomach, and is used for measuring lead impedance. The leads are placed on the anterior and posterior intra-abdominal nerve trunks. Unlike the helical or closed cuff designs used with other peripheral nerve stimulation electrodes, the MAESTRO leads are "C" shaped, and cradle rather than wrap around the abdominal vagus nerve trunks. Sutures anchor and stabilize lead placement. The leads are also labeled as unsafe for MRI. ## Model 2402 Mobile Charger The Model 2402 Mobile Charger (MC) is provided non-sterile and is powered by a rechargeable, lithium-ion battery which is recharged using the AC Recharger connected to a power outlet. The MC can be left plugged in to the AC Recharger when not used for recharging the RNR. When the MC battery is full, the AC recharger will stop charging. Patients are required to recharge the RNR battery as indicated by the display PMA P130019: FDA Summary of Safety and Effectiveness Data Page 3 {3} on the MC. The MC recharges the RNR when connected to the transmit coil positioned over the RNR. The MC uses the same connection to the AC recharger as the Patient Transmit Coil. Therefore, the Mobile Charger cannot be simultaneously connected to both the Patient Transmit Coil and the AC Recharger, thus insuring that the Transmit Coil cannot be connected when the MC is connected to an AC source of current. Patients are required to check the battery daily and recharge when needed. ## Transmit Coils There are three (3) different transmit coils available. These transmit coils are provided sterile. ## Model 2403-300 Clinician Transmit Coil The Model 24030300 Clinician Transmit Coil is for use in the operating room with the MAESTRO® Rechargeable System. The Clinician Transmit Coil is placed within a sterile sleeve for use in the operating room. ## Model 2430-60 Patient Transmit Coil The Model 2430-60 Patient Transmit Coil is for patient use with the MAESTRO® Rechargeable System and is five (5) inches in diameter with a 60cm long cable. The standard coil supplied in the implant kit has silicone as the skin contacting material (Model 2403-60). A second coil (Model 2403-60A) is available, which has an outer jacket of compressed foam and is otherwise identical to Model 2430-60. Physicians may request this coil if they decide it is better suited for a particular patient. Both transmit coils may be used with the optional Model 1670 Foam Pad to assist in achieving the desired spacing of 2-3cm between the transmit coil and the implanted neuroregulator to optimize recharging efficiency. ## Model 2430-60A Patient Transmit Coil The Model 2430-60A Patient Transmit Coil is identical to the Model 2430-60 device component, but has an optional foam pad to assist in achieving the desired spacing between the transmit coil and the implanted neuroregulator. ## Model 2501 Clinician Programmer The Model 2501 Clinician Programmer is a non-sterile, external, programmable, ambulatory microprocessor and controller with compatible, customized firmware. The commercially available laptop is used by the clinician to modify therapy parameters and treatment schedule. It transmits information to the Mobile Charger and uploads data from the Mobile Charger. ## Customized Software Software is provided with the Model 2501 Clinician Programmer/laptop computer and enables communication with the Mobile Charger and neuroregulator. The CP allows physicians to modify therapy parameters and delivery schedules and retrieve diagnostic information. Sample settings are set using the Clinician Programmer. PMA P130019: FDA Summary of Safety and Effectiveness Data Page 4 {4} PMA P130019: FDA Summary of Safety and Effectiveness Data Page 5 ## Therapy Algorithms The MAESTRO system delivers pulses of current to vagal nerve trunks at a high frequency, which maintains the nerve fibers in a refractory state and suppresses the natural impulses that are conveyed from the periphery (i.e., stomach) to the brain stem and higher centers of the central nervous system (CNS). ## Programmer Cable Description The Model 1600 Programmer Cable is non-sterile and is used to provide connectivity between the Model 2502 Clinician Programmer and the Model 2402 Mobile Charger. ## AC Recharger Description The Model 1620 AC Recharger is a non-sterile power supply used for charging the Mobile Charger battery. The AC Recharger is provided with an appropriate power cord to accommodate the electrical mains connection for the geographic location (Part Number P00899-001 for North America). The AC Recharger uses the same connection to the Mobile Charger as the Transmit Coil. Therefore, the Mobile Charger cannot be used to recharge the RNR when it is connected to the AC Recharger. ## Patient Transmit Coil Belt Description The Patient Transmit Coil Belts are non-sterile, external accessories that may be used to assist in holding the patient transmit coil over the location of the implanted RNR during charging. The Model 1660 Transmit Coil Belt is for use with the Model 2403-60 Patient Transmit Coil and the Model 1660A Transmit Coil Belt is for use with the Model 2403-60A Patient Transmit Coil. ## Transmit Coil Foam Pad Description The Model 1670 Transmit Coil Foam Pad is an optional non-sterile external accessory that may be used to assist in achieving the desired spacing between the patient's Transmit Coil and the implanted neuroregulator to optimize recharging efficiency. ## Torque Wrench Description The Model 1680 Torque Wrench is a sterile standard surgical wrench provided for use during the implant procedure to tighten set screws that secure the leads to the RNR as described in the implant procedure guide instructions for use. ## VI. ALTERNATIVE PRACTICES AND PROCEDURES There are several alternatives for the treatment of obesity. These alternative treatments include lifestyle interventions, such as behavioral therapy (e.g., diet and exercise), pharmacotherapy (e.g., phentermine based drugs), other medical devices (e.g., Lap Band®), and surgery (e.g., gastric bypass or sleeve gastrectomy). There are currently limited options for morbidly obese patients. Pharmacotherapy trials have generally demonstrated a $\geq 7 - 13\%$ total weight loss from initial body weight after subtracting for the placebo effect, but some patients have experienced adverse systemic side effects¹. Gastric bypass surgery is among the most effective means of obtaining durable and clinically significant weight loss, but can result in serious complications, including {5} perforation, hemorrhage, infections and bowel obstructions². The Lap Band® is also effective in providing long-term weight loss, but can produce adverse events of band prolapse and band erosion². Each alternative has its own advantages and disadvantages. A patient should fully discuss these alternatives with his/her physician to select the method that best meets expectations and lifestyle. ## VII. MARKETING HISTORY The MAESTRO® Rechargeable System received CE Mark for the treatment of obesity in March 2011 and has been listed on the Australian Register of Therapeutic Goods (ARTG) by the Therapeutic Goods Administration (TGA) for supply in Australia. The MAESTRO® Rechargeable System has not been marketed in the United States. The MAESTRO® Rechargeable System has not been withdrawn from any market. ## VIII. POTENTIAL ADVERSE EFFECTS OF THE DEVICE ON HEALTH Below is a list of the potential adverse effects (e.g., complications) associated with use of the device: - Allergic or immune responses to the implanted materials that could result in device rejections; - Pain at the neuroregulator and/or lead site, resulting from infection, skin irritation, wound dehiscence, erythema, erosion of the neurostimulator, seroma and hematoma; - Malfunction of the components of the MAESTRO Rechargeable System, including loss of therapy, lead migration, lead fracture, lead dislodgement, and bowel entanglement with the leads; - Device or therapy-related events, including heartburn/dyspepsia, abdominal pain, dysphagia (swallowing difficulty), belching, nausea and chest pain; - Damage to the vagal nerve trunks. For the specific adverse events that occurred in the clinical studies, please see Section X below. PMA P130019: FDA Summary of Safety and Effectiveness Data Page 6 {6} IX. SUMMARY OF PRECLINICAL STUDIES A. Laboratory Studies The MAESTRO® Rechargeable System and individual system components were evaluated using internationally accepted, and FDA recognized, standards for performance testing. General tests and conformance to specifications were performed for implantable components and accessories, including: - Visual inspection - Device marking requirements, per EN 45502-1: 1998, EN 45502-2-1:2003, and ISO 58413:2000 - Kit and component packaging, and shipping and temperature conditioning per ASTM D4169. Samples were preconditioned prior to bench testing, including: - Thermal Shock: The neuroregulator and leads were required to withstand a temperature/shock cycling, from -10°C to 55°C. - Environmental conditioning (shock, vibration, pressure, temperature): Temperature and humidity cycling per EN 45502-1: 1998. - Shipping simulation and temperature conditioning: Package free fall, loose load, and random vibrations after cycling between -10°C ± 3°C and 55°C ± 2°C per ASTM D4169-05 and ASTM D4169. - Shipping Conditioning per ASTM D4169-09. - Accelerated aging conditioning per ISO 11607. - Particulate release testing per EN 45502-1, EN 455020201 and ISO 14708-3:2008. In addition to evaluating the general testing, specific components and systems testing is summarized in the following tables. Model 220P-47 (Posterior) and Model 220A-47E (Anterior) Leads The MAESTRO leads were assessed for the expected loading and use conditions. The testing that was performed on the leads is summarized in Table 2. Bench testing of the leads and other device components on electromagnetic compatibility (EMC), radio-frequency (RF) wireless technology, and software are summarized Section IX. PMA P130019: FDA Summary of Safety and Effectiveness Data Page 7 {7} Table 2. Model 220P-47 (Posterior) and Model 220A-47E (Anterior) leads bench testing summary | Test | Purpose | Acceptance Criteria | Results | | --- | --- | --- | --- | | Visual Inspection | Legibility of markings. | Lead packaging must remain legible after exposure to cleaning solvents per IEC 60601-1:2012. | PASS | | | Test the dimensional specifications for the leads per ISO 5841-3:2000. | Must measure 18.50 ±” in length. | PASS | | | Test the dimensional specifications for the leads connectors per ISO 5841-3:2000. | Must measure 18.50 ±” in length. | PASS | | | Test particulate release from the leads per EN 45502-1: 1998, EN 45502-2-1:2003, and ISO 14708-3:2008. | Must produce particles counts lower than pre-specified criteria, in particles sized as 5, 100, and 1000 μm. | PASS | | Composite Tensile Load Testing Following 10 day immersion in 9.0 g/l saline solution at 37°C, leads must remain electrically, mechanically and functionally intact following tensile testing. | Test the tensile strength of lead to anchor per EN 45502-1:1998. | Withstand a pull test (lead to anchor) of 0.5 lbf. | PASS | | | Test the tensile strength of the suture wing and suture tongue peak removal force per EN 45502-1:1998. | Withstand a pull test of >1.1 N (0.25lb). | PASS | | | Test the composite tensile integrity per ISO 14708-3:2008. | Withstand a pull test of 5N (1.1 lbf) with lead elongation no greater than pre-specified criterion. | PASS | | Lead and Connector Flexure Testing | Test the ability of the lead body to withstand loading during long term use after conditioning per EN 45502-1. | Withstand flexure cycling (± 90 degrees) at a rate of 2 Hz for a minimum of 47,000 cycles, and maintain a DC resistance must be < 30 Ω. | PASS | | | Test the ability of the connector to withstand repeated flexure per EN 45502-1. | Withstand flexure cycling (± 45 degrees) at a rate of 2 Hz for a minimum of 82,000 cycles. DC resistances of the pin to | PASS | PMA P130019: FDA Summary of Safety and Effectiveness Data {8} | Test | Purpose | Acceptance Criteria | Results | | --- | --- | --- | --- | | | | distal electrode, the pin to distal electrode, and ring to ring electrode must meet the pre-specified criteria. | | | Fatigue Testing Leads must remain electrically and mechanically functionally intact following fatigue testing. | Test the ability of the lead body to withstand cyclical fatigue. | Withstand 400,000 cycles when flexed +/- 60 degrees at central lead body bend radius of 4.76 mm. | PASS | | | Test the ability of the suture wing and tongue to withstand tensile force. | Withstand minimum acceptance criterion of 0.25N. | PASS | | | Test the ability of the proximal connector to withstand cyclical fatigue. | Withstand a vertical load of 100 g oscillated at 45 degree angle for 82,000 cycles. | PASS | | | Test the ability of the lead tip to withstand withdrawal and insertion forces per ISO 5841-3:2000. | Go gauge can be fully inserted, with a maximum insertion and withdrawal with less force than the test limit. | PASS | | Corrosion Resistance | Test with cyclic polarization for Galvanic and pitting corrosion of the Pt-Ir electrodes per ASTM F2129-08. | No breakdown in potential, or signs of pitting or corrosion. | PASS | | | Test with chronopotentiometry for Faradaic corrosion of the electrodes lead, while manually flexing the lead at the proximal and distal junctions. | The leads must operate without corrosion for the stated lifetime under normal operation, and also exceeding maximum pulse amplitude and pulse width settings. | PASS | | Lead Cross Circuit DC Resistance | Test the resistance of the pin to distal electrode. | DC resistance must be less than 30 Ω. | PASS | | | Test the resistance of the ring to ring electrode | DC resistance must be less than 30 Ω. | PASS | | | Test the resistance of the pin to ring electrode | DC resistance must be less than 10 Ω. | PASS | | Dielectric Test | Test is to verify that the device has effective functional electrical insulation between Conductors per EN45502-2-1:2003. | After immersion in 9.0 g/L saline for a minimum of 10 days at 37°C, the leakage current between all conductors and the reference electrode must | PASS | PMA P130019: FDA Summary of Safety and Effectiveness Data {9} PMA P130019: FDA Summary of Safety and Effectiveness Data Page 10 | Test | Purpose | Acceptance Criteria | Results | | --- | --- | --- | --- | | | | be less than 2 mA during a voltage application. | | ## Model 2002 Rechargeable Neuroregulator Circuit assemblies for the of the Model 2002 Rechargeable Neuroregulator underwent electrical testing after accelerated life test exposure for over 1000 hours of operation at 120°C using MIL-STD 883 Revision G, Test Method Standard, Microcircuits as a guidance. Performance testing included evaluations of electrical safety, functionality (e.g., outputs), mechanical strength, and design validation. Battery testing is summarized in Section IX. Testing of the neuroregulator was performed per applicable regulations and standards including: - IEC 60601-1:2012 - IEC 60601-1-2: 2001 - EN 60601-1-2: 2007 - ANSI/AAMI/ISO 14708-3:2008 - EN45502-1:1998 - MIL-STD-202G:2002 The testing that was performed on the neuroregulator is summarized in Table 3. Table 3. Rechargeable Neuroregulator bench testing summary | Test | Purpose | Acceptance Criteria | Results | | --- | --- | --- | --- | | Visual Inspection | Ensure legibility of markings per EN 45502-1:1998 and ISO 5841-3:2000. | Neuroregulator packaging must remain legible after exposure to cleaning solvents per IEC 60601-1:2005. | PASS | | | Test protection of device due to mishandling during the implant procedure by visual inspection per EN 45502-1:1998 and EN 5841-3:2000. | Withstand 70 kPa for 1 hour and 150 kPa for 1 hour. | PASS | | | | Withstand half sine, 500g, 1 ms, 6 shocks, 3 axes. | PASS | | | | Withstand 3 cycles, 3 hours at -10°C, 3 hours at +55 °C. | PASS | | Mass Volume | Test to ensure the volume, mass and physical configuration meet design requirements per EN 45502-1:1998. | Components must comply with requirements for mass and volume, must not have sharp edges, and must meet the pre-specified criteria for internal atmosphere. | PASS | {10} | Test | Purpose | Acceptance Criteria | Results | | --- | --- | --- | --- | | Particulate Release Test | Test to ensure that final assembly process, and to the cleaning and packaging processes do not introduce unacceptable levels of particulates per EN 45502-1:1998. | Component must be removed from package, and meet maximum particulate content after immersion in 9.0 g/L saline. | PASS | | Dielectric Testing | Testing to verify that the device has effective functional electrical insulation between Conductors per EN45502-2-1:2003. | The leakage current between all conductors and the reference electrode must be less than 2 mA during the voltage application. | PASS | | Electrical Leakage Testing | Test for electrical neutrality per EN 45502-1: 1998. | Must not output > 1 μA net DC current when connected to the Model 220A/P Leads, Model 2402 Mobile Charger, Model 2403 Transmit Coil and nerve equivalent circuit. | PASS | | Vibration Testing | Test to validate that the neuroregulator remains electrically and mechanically functionally intact following fatigue testing. | Withstand 4-400 Hz, 0.7 (m/s2) 2 Hz, 30 minutes, 3 axes. | PASS | | | | Withstand 80,000 cycles when flexed +/- 60 degrees at central lead body bend radius of 4.76 mm. | PASS | | | | Withstand a vertical load of 100 g oscillated at 45 degree angle for 82,000 cycles. | PASS | | | | Withstand a displacement of 1 cm over 80,000 cycles. Extension cycling shall take place at a rate of 2 cycles per second. | PASS | | | | Withstand 80,000 flex (crush) cycles | PASS | | Connector Insertion Limits and | Test the connector retention after preconditioning in 9.0 g/L saline at room | Connector must withstand straight separating pulls of up | PASS | | inertion limits | to the connector retention after preconditioning in 9.0 g/L saline at room | transmitter must be inserted in the connector to the connector. | PASS | | | | transmitter must be inserted in the connector to the connector to the connector to the connector to the connector to the connector to the connector to the 10000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000 | PASS | PMA P130019: FDA Summary of Safety and Effectiveness Data {11} | Test | Purpose | Acceptance Criteria | Results | | --- | --- | --- | --- | | Dimensional Requirements | temperature per ISO 14708-1:2000. | to 10 N of force and stay connected to the neuroregulator. | | | | Test connector insertion. | The go gauge must be able to be fully inserted; The 2.7 mm gage pin must be inserted to the distance specified using a maximum of 9N of force. | PASS | | | Test connector withdrawal. | Disconnects with an insertion force of less than 14 N. | PASS | | | Test connector dimensions per ISO 5841-3:2000. | The dimensions of each lead barrel must be in compliance with figure 3 of ISO 5841-3:2000. | PASS | | Temperature Rise Test | Test temperature of outer surface when implanted and operated in normal condition per EN 45502-1: 1998. | Outer temperature does not exceed 2°C above normal body temperature. | PASS | | | Test comparing the heating effects of two different modes of the Neuroregulator; recharging of the neuroregulator and therapy delivery per EN 45502-1:1998. | Temperature increase measured during charging and therapy delivery is less than 2°C. | PASS | | Protection from Damage by External Defibrillators | Test to verify that the Neuroregulator can endure exposure to external defibrillation being applied near an implanted device per EN 45502-1. | Device must pass functional test after testing in conformance to Appendix C. | PASS | **Transmit Coils (TCs)** Tensile and flex testing, torsion testing, and dielectric testing depend on the connector and cable and the silicone overmold. Since the connector, cable, and silicone overmold are identical for the Model 2403-60 and 2403-60A Patient Transmit Coils (length, 60 cm) and Model 2403-300 Clinician Transmit Coil (length, 300 cm), performance testing for the Model 2403-60 is representative of testing of all TC Models. Results of bench testing are summarized in Table 4. PMA P130019: FDA Summary of Safety and Effectiveness Data {12} Table 4. Model 2403-60 Patient Transmit Coil testing summary | Test | Purpose | Acceptance Criteria | Results | | --- | --- | --- | --- | | Cable Performance | Test the electrical impedance, and resistance per IEC 60601-1:2005. | Cable must meet pre-specified criteria on the frequency of the coil assembly and acceptable impedance value, and withstand 60,000 flexure cycles. | PASS | | | Test mechanical performance per ANSI/AAMI/EC53:1995 (R2008). | Must withstand tensile, flexure, and torsion testing after 90,000 cycles and rotation from 0 to ±90 degrees. | PASS | | | Test dielectric withstand of cable with antenna and LEMO connector per ANSI/AAMI/EC53:1995 (R2008). | Withstand preconditioning in a humidified environment of 93% ±3% at 32°C, followed by immersion in 9.0 g/L saline, flex testing, and rotation cycles without cable failure. Leakage must not exceed 1 μA. | PASS | | Tensile Strength of Antenna and Connector | Test tensile strength of cable (LEMO) connector and antenna per ANSI/AAMI/EC53:1995 (R2008). | Antenna tensile strength must be equivalent or better than 32.8 lbf, and withstand torsion LEMO torsion flexion of 30,000 cycles without loss of conductivity. | PASS | | General Electrical Safety Testing | Test to verify that the MC meets international standards for electrical safety per 60601-1:2012. | Meets the requirements for dielectric withstand | PASS | Model 2402 Mobile Charger Testing for the Model 2402 Mobile Charger is summarized in Table 5. Table 5. Model 2402 Mobile Charger testing summary | Test | Purpose | Acceptance Criteria | Results | | --- | --- | --- | --- | | Sales Package High Humidity Storage Test | Test to verify that the sales package can endure high humidity storage conditions without deterioration. | Meets the requirements of EN 45502-1:1998 Section 10.2 | PASS | | Temperature Elevation of the MC Surface | Test to verify that the surface temperature does not exceed temperature | The MC, when provided with a representative load by the Model 2403-60 | PASS | PMA P130019: FDA Summary of Safety and Effectiveness Data {13} | Test | Purpose | Acceptance Criteria | Results | | --- | --- | --- | --- | | | limits per IEC 60601-1:2012. | transmit coil and MAESTRO Model 2002 Neuroregulator, must meet the pre-specified temperature criterion stated in IEC 60601-1. | | | Enclosure Rigidity Test | Test to verify that that the rigidity of the MC enclosure provides an adequate level of protection from internal damage from application of force, impact from misuse per IEC 60601:1990, Clause 21 and 60601-1:2005, Clause 15.3.2. | The MC enclosure, when subjected to the rigidity, must be sufficient to prevent an inward directed force from appreciably damaging the enclosure or reducing creepage distances and air clearance. | PASS | | Isolation between Clinician Programmer Port and Transmit Coil Connector | Test to verify that the electrical isolation between the Clinician Programmer port and the Transmit Coil connector per IEC 60601:2005. | Isolation is sufficient to prevent accidental coupling of mains terminal voltage to the transmit coil in the event of a malfunction of the Programmer's USB port. | PASS | | Immunity to Electrostatic Discharge | Test to verify that the electrical isolation between the Clinician Programmer port and the Transmit Coil connector per IEC 60601:2005. | Mobile Charger must sustain electrostatic discharges without any failures or changes in functionality. | PASS | | Electrical Safety Testing | Test to verify that the MC meets international standards as set forth in EN 45502-1:1998, ISO 6060101:1990, ISO 606010101:2005, ISO 6060102:2001, ISO 61000-4-2:2009, and ISO 8601:2004 | Meets the requirements for electrical isolation between the clinician programmer port and transmit coil connector, immunity from electrical discharge, and produce ≤1μA under dielectric strength testing. | PASS | Electromagnetic compatibility (EMC), electromagnetic interference (EMI), and radio-frequency (RF) wireless testing. The MAESTRO System was tested for EMC, EMI, and RF wireless performance. Results are summarized in Table 3. Testing was performed in accordance with applicable standards, including: PMA P130019: FDA Summary of Safety and Effectiveness Data {14} - IEC 60601-1-2 3rd Edition - IEC 61000-4-2 - IEC 61000-4-3 - IEC 61000-4-4 - IEC 61000-4-5 - IEC 61000-4-6 - IEC 61000-4-8 - IEC 61000-4-11 - EN 45502-1 - ISO 14708-3:2002 - EN45502-1:1997 - EN 300 330-2 V1.3.1 - EN 55011:2007 - FCC Part 18 Subpart C - FCC Part 15 Subpart C Table 6 summarizes the EMC, EMI, and wireless technology testing and results. Table 6. Testing for electromagnetic compatibility, electromagnetic interference and radio-frequency wireless technology | Test | Purpose | Acceptance Criteria | Results | | --- | --- | --- | --- | | Immunity Testing fields | Test for protection from static magnetic per ISO 14708-3. | Neuroregulator and leads operate normally when exposed to 1 mT and 50 mT, meeting A-line and B-line criteria. | PASS | | | Test for protection from magnetic fields in the range 10 Hz to 30 MHz per ISO 14708-3. | Neuroregulator and leads operate normally, with no loss of function when tested for performance criterion A of performance criterion B. | PASS | | | Test for protection from electromagnetic fields in the range of 30 MHz to 450 MHz, and electromagnetic fields in the range of 450 MHz to 3 GHz. | System operates normally, with no loss of function when tested for performance criterion A of performance criterion B. | PASS | | High Power Electrical Fields Exposure | Test to verify that the Neuroregulator can endure exposure to high powered electrical fields without harming the device per EN 45502-1. | Neuroregulator, Transmit Coil and Mobile charger must meet functional test criteria per Annex D after exposure. | PASS | PMA P130019: FDA Summary of Safety and Effectiveness Data {15} | Test | Purpose | Acceptance Criteria | Results | | --- | --- | --- | --- | | Protection from Electromagnetic Non-ionizing Radiation | Test for protection of the MAESTRO System to magnetic fields in the range of 30 Hz-30 MHz per 45502-2-1:1997 | Neuroregulator and leads operate with continuous therapy as intended, within pre-specified limits with no loss of function | PASS | | Protection from Electromagnetic Fields | Test for conducted and radiated emission testing per EN 60601-1-2:2007, EN 55011:2009 and FCC Part 15 Subpart B. | Neuroregulator and leads operate as intended, within pre-specified limits with no loss of function. | PASS | | Protection from Electromagnetic Fields | Test for immunity from electrostatic discharge, radiated fields, fast/transient bursts, surge transients, conducted disturbance, power frequency magnetic field, voltage dips/ interruptions/variations, and transients and surges per EN60601-1-2:2007, EN 61000-4-2:2009, EN 61000-4-3:2006, EN 61000-4-4:2004, EN 61000-4-5:2006, EN 61000-6:2009, EN 61000-8:2010, EN 61000-4-11:2004, ETSI EN 301-489-3, ISO 7636-2:2004, FCC Part 15 Subparts B and C, FCC Part 18 Subpart B. | Neuroregulator, transmit coil, anterior and posterior leads, mobile charger, programmer cable, clinician programmer and AC recharger operate as intended in the electromagnetic fields for the pre-specified frequency and power ranges, loads and burst conditions. | PASS | ## Software The software for the MAESTRO system components was determined to have a MAJOR level of concern. The software for the Model 2502 Clinician Programmer, Model 2402 Mobile Charger, Model 2002 Neuroregulator, and Operating Systems were evaluated through verification and validation testing to confirm that it met user needs and performed as intended. Software testing demonstrates conformance to the following standards and guidances: - ISO 13485:2003 - 90/385/EEC:1990 (Amended 2007/47/EC) - ISO/IEC 14971:2007 - EN 45502-1:1998 - IEC 60601-1:2005 (3rd Edition) PMA P130019: FDA Summary of Safety and Effectiveness Data {16} - IEC 60601-1-4:2005 (Amendment 1, 2nd Edition) - IEC 62304:2006 - IEEE Standards Collection for Software Engineering: 1999 - FDA Guidance, “Guidance for Industry, FDA Reviewers and Compliance on Off-The-Shelf Software Use in Medical Devices,” 1999 - FDA Guidance, “General Principles of Software Validation; Final Guidance for Industry and FDA Staff,” 2002 - FDA Guidance “Guidance for Industry – Cybersecurity for Networked Medical Devices Containing Off-The-Shelf (OTS) Software,” 2005 ## B. Animal Studies ### In Vivo Animal Studies EnteroMedics evaluated the safety of implantation and nerve blockade of the porcine abdominal vagus nerve trunk. The device was tested using a variety of device components. An earlier design of the neuroregulator (Radiofrequency 2, or RF2) was tested with a 100% platinum electrode Model 1200 electrode. The tested device was revised to include an RF2 neuroregulator for use with a Model 2200 platinum-iridium electrode. With iterations in device design, device components representing the test articles used in clinical study were tested, including testing of the Rechargeable neuroregulator in combination with the Model 2200 electrode. For the stimulation algorithm, pulse frequency was always maintained at 5000 Hz, and the duty cycle was always set to deliver 5 minutes of VBLOC therapy ON, followed by 5 minutes of VBLOC therapy OFF. Table 7. Summary of porcine studies conducted on VBLOC therapy | Animal Study ID | Model RNR + Leads | Study Duration | Therapy Algorithm: Current, Pulse Width, VBLOC | | --- | --- | --- | --- | | TR01 | #1000 RNR; #1200 leads | 1-3 weeks | 2-4mA, 100μS, 12 hrs./day | | TR02 | #1000 RNR; #1200 leads | 4 weeks | 2-6mA, 100μS, 12 hrs./day | | TR03 | #1000 RNR; #1200 leads | 8 weeks | 2-6mA, 100μS, 12 hrs./day | | TR04 | #1000 RNR; #1200 leads | 12 weeks | 4-6mA, 100μS, 12 hrs./day | | TR05 | #1000 RNR; #1200 leads | 1-3 weeks | 6mA, 100μS, 12 hrs./day | | TR06 | #1000 RNR; #1200 leads | 12 weeks | 6mA, 100μS, 12 or 24 hrs./day | | TR07 | #1002 RNR; #2200 leads | 9 days | 6mA, 90μS, 24 hrs./day | | TR08 | #1002 RNR; #2200 leads | 4-12 weeks | 6mA, 90μS, 24 hrs./day | | TR09 | #1002 RNR; #2200 leads | 12 weeks | 6mA, 90μS, 24 hrs./day | | TR10 | #1002 RNR; #2200 leads | 4 weeks | 8mA, 90μS, 24 hrs./day | | TR11 | #2002 RNR; #2200 leads | 4 weeks | 6mA, 90μS, 24 hrs./day | | TR12 | #2002 RNR; #2200 leads | 12 weeks | 8mA, 90μS, 24 hrs./day | PMA P130019: FDA Summary of Safety and Effectiveness Data Page 17 {17} # Analysis Device components that are designed to be internally implanted in patients include the neuroregulator and leads. For the animal studies, the neuroregulator and leads were exteriorized due to the anatomical limitations of using the porcine animal model. Exteriorization of these device components, and the natural growth of the animal subjects, resulted in chronic pulling forces that resulted in trauma to the nerve. Exteriorization of the neuroregulator and leads was reported to produce neural trauma which likely exacerbated the neurodegeneration observed in histological sections of the implanted nerves. Therefore, the data provided by EMI may not have been representative of the long term safety of device implantation in humans. Results supported device development, and confirmed that gastric contraction function was preserved with VBLOC therapy. # C. Additional Studies Additional performance studies are briefly described in Table 8. Table 8. Miscellaneous performance studies on the MAESTRO Rechargeable System | Test | Purpose | Acceptance Criteria | Results | | --- | --- | --- | --- | | Battery Testing | State of charge, voltage with relation to capacity, over-charging per IEC 62133 2nd Edition. | Batteries withstood an internal short circuit (nail test, n=14), external short circuit (n=8) and over-charging (n=3) without exhibiting adverse chemical/thermal reaction. | PASS | | Suture Needle Fit Test | Testing to verify that suture tabs on the Neuroregulator are adequately shaped to accommodate a range of suture needle sizes. | Each suture needle must pass freely through each of the 3 suture holes provided on the device. | PASS | | Sterilization of the Neuroregulator, Leads, and Torque Wrench | Testing to verify to ethylene oxide sterilization using the small batch, half cycle method, and acceptable bioburden, endotoxin, and ethylene oxide residual levels per ANSI/AAMI/ISO 11737-2:2009, ISO 11135:2014: Annex E, ANSI/AAMI/ISO 10993-7:2008, ISO 11737-1:2006, ANSI/AAMI/ISO 11737-1:2006, ISO 11138-2:2006, MTL-1005 rev 016, USP/EP/CFR Sterility Method | Sterilization must demonstrate per a sterility assurance of at least 10-6. Endotoxin levels must not have more than 20 Endotoxin Units (EU)/device for the summation of the blood-contacting devices, and meet test criteria for bioburden and colony forming units. Ethylene oxide residuals must not exceed test criteria. | PASS | PMA P130019: FDA Summary of Safety and Effectiveness Data {18} | Test | Purpose | Acceptance Criteria | Results | | --- | --- | --- | --- | | | Suitability (Bacteriostasis/Fungistasis) Te20).st, and U.S. Pharmaceutical Convention <85>. | | | | Packaging | Testing to ensure that the sterile integrity and physical properties of packaging meets shelf-life criteria per ASTM F1980-07:2011. | Packages tested met the acceptance criteria for seal strength and dye penetration to demonstrate a 3-year shelf life for the leads, 18-month shelf-life for the torque wrench and neuroregulator. Conditioning and test samples are used to evaluate each step in the packaging process. | PASS | | Biocompatibility | Testing of the Patient Transmit Coil as an externally patient contacting device with limited duration of exposure (≤ 24 hours) to ensure conformance to ISO-10993-1:2009, ISO-10993-2:2006, and ISO-10993-5:2009. | Must meet pre-specified criteria for cytotoxicity, irritation. and sensitization. | PASS | | | Testing of the of the neuroregulator and leads as permanent, tissue contacting, implanted device components to ensure conformance to ISO-10993-1:2009, ISO-10993-2:2006, ISO 10993-3:2003, ISO-10993-5:2009, ISO 10993-10:2010, 10993-11:2009, ISO 10993-18:2005, | Must meet pre-specified criteria for genotoxicity, mutagenicity, cytotoxicity, irritation, sensitization, implantation, acute systemic toxicity, and systemic toxicity, and demonstrate non-significant levels of risk with respect to leachables and extractables. | PASS | X. SUMMARY OF PRIMARY CLINICAL STUDIES Overview Three (3) studies (one (1) pilot and two (2) pivotal studies) were conducted. The primary effectiveness for each of the studies was to demonstrate a significantly greater percentage of weight loss at the 12 month time point, with a super-superiority margin of 10%. For the two (2) pivotal studies, which utilized a randomized, double blind approach, super-superiority was determined by comparing the percent excess weight loss (%EWL) at the 12 month time point in VBLOC treatment versus Control arms. For the pilot study, which utilized an open label approach, super-superiority was determined PMA P130019: FDA Summary of Safety and Effectiveness Data {19} by comparing the %EWL at baseline versus the %EWL at the 12 month time point. Although super-superiority was not demonstrated in any of the three (3) clinical trials, a statistically significant difference between the VBLOC therapy arm and the Sham control arm was observed. All three (3) studies met the safety endpoint, which was to show a serious adverse event (SAE) rate of ≤15%. ## EMPOWER Pivotal Study The first pivotal study to be conducted in the U.S. was entitled the EMPOWER Clinical Trial. The trial was initiated in 2007 and featured the MAESTRO RF2 system. Device components included the Model 1002 Neuroregulator, Model 2200 Leads, Model 104 Controller, Model 1403 Transmit Coil and accessories, Model 2500 Programmer Software, and Model P00062-000 Battery Charger. The transmit coil belt was worn "fanny pack" style to optimally align the coil to provide power to the RNR. EMPOWER was designed as a prospective, randomized (2:1), double-blind, controlled trial with evaluation of primary endpoints at 12 months. The intended patient population was those who have a BMI &gt;40 kg/m² to 45 kg/m², or ≥35 to 39.9 kg/m² with obesity related comorbidities. A total of 294 subjects were randomized to either VBLOC (192 subjects) or sham therapy (102 subjects) at 15 institutions. For the sham therapy, the Sham patients received the implantable device components as well as lead impedance and safety checks, but the therapy algorithm was set to deliver 0 mA of VBLOC therapy. There were 294 subjects implanted and randomized, including 192 VBLOC (treatment) and 102 Sham control subjects. At the 12 month time point, there were 22 SAEs (11.5%) in the VBLOC treatment arm, of which 22 (4.7%) were related to either implant/revision procedure, therapy, or the device. In the Sham control arm, there were 11 SAEs (10.8%) of which 8 (7.8%) were related to either implant/revision procedure, therapy, or the device. Table 9 presents longer term data on the most frequent types of adverse events through 48 months that were related to either implant/revision procedure, therapy, or the device. Table 9. EMPOWER Study: Most frequent (≥5%) implant/revision procedure, therapy, or device-related adverse events through 48 months. | Event Type | Treatment Group (n=192) | Sham Control (n=102) | | --- | --- | --- | | | Number (%) | Number (%) | | Pain, neuroregulator site | 56 (29.2%) | 31 (30.4%) | | Heartburn/dyspepsia | 46 (24.0%) | 17 (16.7%) | | Pain, other | 35 (18.2%) | 12 (11.8%) | | Other | 36 (18.8%) | 17 (16.7%) | | Pain, abdominal | 33 (17.2%) | 20 (19.6%) | | Nausea | 23 (12.0%) | 8 (7.8%) | | Skin reaction to coil/coil adhesion method | 21 (10.9%) | 13 (12.7%) | | Belching | 19 (9.9%) | 12 (11.8%) | | Constipation | 19 (9.9%) | 10 (9.8%) | PMA P130019: FDA Summary of Safety and Effectiveness Data Page 20 {20} | Event Type | Treatment Group (n=192) | Sham Control (n=102) | | --- | --- | --- | | | Number (%) | Number (%) | | Chest pain | 17 (8.9%) | 3 (2.9%) | | Wound redness | 13 (6.8%) | 4 (3.9%) | | Diarrhea | 9 (4.7%) | 8 (7.8%) | | Bloating, abdominal | 9 (4.7%) | 7 (6.9%) | | Incision pain/incision site | 8 (4.2%) | 8 (7.8%) | | Neuroregulator malfunction | 4 (2.1%) | 7 (6.9%) | The primary effectiveness endpoint was to demonstrate a significantly greater percentage of excess weight loss (% EWL, MetLife method, with a 10% supersuperiority margin) with subjects in the active VBLOC arm after 12 months of VBLOC Therapy when compared to sham control subjects. This objective was not met, since the mean %EWL in the VBLOC treatment and sham control arms was 12.1 ± 17.5 and 12.0 ± 20.8, respectively. EnteroMedics has suggested that the requirement for patients to wear the transit coil and controller for several hours per day in order ensure delivery of VBLOC therapy could have led to the lack of compliance with recharging instructions and thus contributed to the lack of effectiveness findings. The primary safety objective was to demonstrate that the long term (through 12 months), implant/revision procedure, device, and therapy related serious adverse event (SAE) rate is less than 15%. The safety objective was met, since the rate for these SAEs was 5.2% in the treatment arm and 2.9% in the control arm. There were no deaths or unanticipated adverse events (UAEs) observed. ## VBLOC DM2 Pilot Study VBLOC-DM2 was a pilot trial of the MAESTRO® Rechargeable System performed on obese subjects with Type 2 Diabetes. It was designed as an open label, single-arm, multi-center trial conducted on 28 subjects at five (5) centers outside the U.S. All subjects received all implantable components of the MAESTRO® Rechargeable System at the time of implantation. Mean weight loss of 24% EWL was reported at 12 months. There have been no deaths or unanticipated adverse events (UADEs); however, four (4) device or implant-related SAEs have been reported through 36 months. Table 10 lists AEs related to device or procedure that occurred in greater than 5% of Type 2 Diabetic subjects. PMA P130019: FDA Summary of Safety and Effectiveness Data {21} Table 10. VBLOC-DM2: Adverse Events Related to Device or Implant/Revision Procedure | AE Type | N patients (%) | N events | | --- | --- | --- | | Heartburn | 8 (28.6%) | 8 | | Constipation | 6 (21.4%) | 7 | | Other | 4 (14.3%) | 6 | | Pain (neuroregulator site) | 5 (17.9%) | 6 | | Nausea | 3 (10.7%) | 5 | | Pain | 3 (10.7%) | 4 | | Abdominal pain | 2 (7.1%) | 3 | | Abdominal cramps | 2 (7.1%) | 2 | | Chest pain | 2 (7.1%) | 2 | | Wound redness or irritation, neuroregulator site | 2 (7.1%) | 2 | | Wound redness or irritation, trocar site | 2 (7.1%) | 2 | # RECHARGE Pivotal Study The applicant performed a clinical study to establish a reasonable assurance of safety and effectiveness of abdominal vagus nerve blocking therapy with the Maestro® Rechargeable System for use in weight reduction in patients aged 18 years through adulthood who have a Body Mass Index (BMI) of 40 to $45\mathrm{kg} / \mathrm{m}^2$ , or a BMI of $\geq 35$ to $39.9\mathrm{kg} / \mathrm{m}^2$ with one or more obesity related co-morbid conditions, and have failed at least one supervised weight management program within the past five (5) years in the U.S. and Australia under Investigational Device Exemption (IDE) # G070025. Data from this clinical study were the basis for the PMA approval decision. A summary of the clinical study is presented below. Approval for the study was granted in March 2011. Subjects were implanted with the MAESTRO System between May 2011 and December 2011. # A. Study Design Patients were treated between May 2011 and January 2013. The database for this PMA reflected data collected through May 2013 and included 239 patients. There were 10 investigational sites (8 in the U.S. and 2 in Australia). This study was a prospective, randomized, double blind, parallel-group, multi-center trial to evaluate the safety and efficacy of the device in treating obesity with 12-month follow-up in 239 randomized patients. The intended patient population was those who had a BMI $40 - 45\mathrm{kg} / \mathrm{m}^2$ or $35 - 39.9\mathrm{kg} / \mathrm{m}^2$ with obesity related comorbidities. The subjects were randomized 2:1 to either VBLOC therapy or a Sham Control Group. Post-implant follow-up visits were scheduled at weeks 1-4, 6, 8 and 10, and then monthly from months 3-12. Longer term follow-up was scheduled out to 5 years post-implant. Subjects were blinded to the assigned treatment group and unblinding of subjects PMA P130019: FDA Summary of Safety and Effectiveness Data {22} began once all subjects had completed the month 12 evaluation. Subjects in the Control Group (sham) were implanted with a nonfunctional neuroregulator device and no leads were implanted. These subjects received a sham surgical procedure, which involved placing approximately the same number of incisions (3-5) and utilized general laparoscopic techniques. The nonfunctional neuroregulator operates in the same manner as the functional neuroregulator, but did deliver electrical stimulation. The neuroregulator's lead sockets were filled with medical grade silicone adhesive to ensure that no electrical current was delivered by the device. As with the active functional neuroregulator, the sham nonfunctional neuroregulator contains a battery and thus required recharging. The battery in the sham device becomes depleted and interacts with the programmer in the same fashion as the active device. All subjects remained blinded through at least the 12 month follow-up visit, after which the Sham subjects who chose to continue in the trial had the option of having the MAESTRO® Rechargeable System fully implanted, and receiving active therapy. A blinded Clinical Event Committee (CEC) reviewed and adjudicated all serious adverse events (SAEs). A blinded Data and Safety Monitoring Board (DSMB) reviewed the trial on an ongoing basis. The DSMB reviewed aggregated clinical laboratory and electrocardiographic data and concluded that there were no findings of clinical significance or concern. No changes in intra-cardiac conduction (PR interval, QRS duration), ventricular repolarization (QTcF interval), or ventricular arrhythmias were associated with either the treatment or control groups in the study. 1. Clinical Inclusion and Exclusion Criteria Enrollment in the ReCharge Study was limited to patients who met the following inclusion criteria: - Informed consent. - Body mass index (BMI) ≥ 40 kg/m² to 45 kg/m² or BMI ≥ 35 kg/m² to kg/m² with at least one (1) obesity related co-morbid condition. Co-morbid conditions may include one or more of the following and will be documented on the appropriate case report form: - Type 2 diabetes mellitus as defined in the 5th inclusion criteria (limited to 10% of randomized subjects) - Hypertension as defined by systolic pressure ≥140 mmHg and/or diastolic pressure ≥90 mmHg - Treated or untreated hypertension with systolic ≥140 mmHg and/or diastolic ≥90 mmHg - Treated hypertension with systolic &lt;140 mmHg and/or diastolic &lt;90 mmHg - Dyslipidemia as defined by total cholesterol ≥200 or LDL ≥130 PMA P130019: FDA Summary of Safety and Effectiveness Data Page 23 {23} ○ Treated or untreated dyslipidemia with total cholesterol ≥200 or LDL ≥130 ○ Treated dyslipidemia with total cholesterol &lt;200 or LDL &lt;130 ○ Sleep apnea syndrome (confirmed by overnight p02 studies) ○ Obesity-related cardiomyopathy - Females or males. Note: females of child-bearing potential must have a negative urine pregnancy test at Screen and also within 14 days of implant procedure followed by physician-approved contraceptive regimen for the duration of the study period. - 18-65 years of age inclusive. - Type 2 diabetes mellitus subjects with: ○ Glycosylated hemoglobin (HbA1c) 7.0-10% inclusive at screening visit (Undiagnosed subjects that are found to have a HbA1c 7-10% at screening must see their primary physician for diagnosis and medical treatment before continuing in trial), ○ Onset of Type 2 diabetes mellitus: 12 years or less since initial diagnosis, ○ Currently not using insulin therapy, GLP-1 receptor agonists (e.g., exenatide), for diabetes treatment and have not been on these treatments in the past 6 months, and ○ Serum creatinine within normal reference range: - No history of proliferative retinopathy - No history peripheral neuropathy - No history of autonomic neuropathy - No history of coronary artery disease, with or without angina pectoris - No history of peripheral vascular disease - Failure to respond to a supervised diet/exercise programs in which the subject was engaged within the last five (5) years. - Ability to complete all study visits and procedures. Patients were not permitted to enroll in the ReCharge Study if they met any of the following exclusion criteria: - Concurrent chronic pancreatic disease. - History of Crohn’s disease and/or ulcerative colitis. PMA P130019: FDA Summary of Safety and Effectiveness Data Page 24 {24} - History of bariatric surgery, fundoplication, gastric resection or major upper-abdominal surgery (acceptable surgeries include cholecystectomy, hysterectomy) - History of pulmonary embolism or blood coagulation disorders. - Clinically significant hiatal hernias (&gt;5cm) known from subject’s medical record or determined by barium swallow (upper GI x-ray) or upper endoscopy per PI discretion prior to implant. - Current cirrhosis, portal hypertension, and/or esophageal varices. - Intra-operative exclusion: hiatal hernia requiring surgical repair or extensive dissection at esophagogastric junction at time of surgery. - Treatment with prescription weight-loss drug therapy within the prior three (3) months and the use of prescription drug therapy or the use of over-the-counter weight loss preparations for the duration of the trial. - Smoking cessation within the prior six (6) months. - Known genetic cause of obesity (e.g., Prader-Willi Syndrome). - Weight loss of more than 10% of body weight in the previous 12 months. - Physician-prescribed pre-operative weight loss program prior to surgery. Note: Study subject may continue any personal eating plan they were on prior to study enrollment. - Current Type 1 diabetes mellitus (DM). - Current or recent history (within 12 months) of ongoing bulimia. - Current alterations in treatment for thyroid disorders (stable treatment regimen for prior three (3) months acceptable). - Current alterations in treatment for epilepsy (stable treatment regimen for prior six (6) months acceptable). - Current treatment for peptic ulcer disease (previous history acceptable). - Chronic treatment (more than four (4) weeks of daily use) with narcotic analgesic drug regimens (treatment with non-steroidal anti-inflammatory drugs is acceptable). PMA P130019: FDA Summary of Safety and Effectiveness Data Page 25 {25} - Current alterations in treatment regimens of anti-cholinergic drugs, including tricyclic antidepressants (stable treatment regimen for prior six (6) months is acceptable). - Current medical condition that, in the opinion of the investigator, would make subject unfit for surgery under general anesthesia or that would be exacerbated by intentional weight loss. Some examples include diagnosis of cancer, recent heart attack, recent stroke, or recent serious trauma. - Presence of permanently implanted electrical powered medical device or implanted gastrointestinal device or prosthesis (e.g., pacemakers, implanted defibrillators, neurostimulators etc.). - Planned or contemplated use of Magnetic Resonance Imaging (MRI) or oncologic radiation during the course of the trial. - Psychiatric disorders (including untreated severe depression, schizophrenia, substance abuse, bulimia nervosa, etc.) or limited intellectual functioning which would potentially compromise the participant's ability to fully comprehend and/or cooperate with the study protocol. Psychiatric disorders will be established by a review of subject's medical history. For depression, a Beck Depression Inventory (BDI) score $\geq 29$ will be considered to indicate severe depression. - Current, active member of an organized weight loss program (e.g., Weight Watchers, Take Off Pounds Sensibly (TOPS)). - Current participant in another weight loss study or other clinical trials. - Have a friend or family member who is currently participating or is planning to participate in this clinical trial. - Patient reported: - inability to walk for about 10 minutes without stopping - feeling of pain in chest when doing physical activity - feeling of pain in chest when not doing physical activity - Clinically significant cardiac rhythm disorder that requires either medical and/or surgical intervention (e.g., paroxysmal or chronic atrial fibrillation). 2. Follow-up Schedule Weight Management Counseling All subjects were required to participate in a weight management program, PMA P130019: FDA Summary of Safety and Effectiveness Data Page 26 {26} consisting of recommendations regarding diet, exercise, and behavior modification throughout the study. All subjects were taught the same basic information about weight loss and physical activity and were given the opportunity to practice related behavioral skills both during educational sessions and at home. Modifications to their current diet and exercise plan were taught by a trained adviser through seventeen (17) individual sessions during the first year along with the regularly scheduled trial visits. The subjects were required to complete a 7 day diet and exercise diary prior to the implant, weeks one (1) through four, and once per month during the first year of the study. Following the first year, group sessions were scheduled for the duration of the study. Subjects were required to cover the elements of the curriculum in a minimum 17 individual face-to-face sessions during the first 12 months after initiation in order to complete the year one behavioral weight loss instruction. ## Therapy Algorithms (Device Settings) The VBLOC treatment group neuroregulators were initially set to deliver current amplitude of 1 mA with a treatment schedule of 13 hours per day. The amplitude was increased to 3 mA at the week 1 visit, and increased by 1 mA each following week reaching 6 mA at week 4. The programming sessions and the systematic amplitude increases were performed for both VBLOC and Sham groups to maintain blinding. Subjects who could not tolerate 3 mA at week 1 or 1 mA incremental increases were increased at a slower rate and/or smaller increments. Other therapy parameters included a ramp-up time of 0 to 50 seconds, an ON time of 2 to 5 minutes, and an OFF time of 5 to 10 minutes. Therapy at 6 mA (or the maximal tolerated amplitude) and a 13 hour delivery schedule per day were then maintained for the remainder of the first 6 months. At month 6, the goal was for subjects to achieve a 15% EWL. Any subjects reporting unacceptable adverse events that were possibly related to therapy underwent modifications of the device parameters including a decrease in amplitude, an increase in the OFF Time, an increase or decrease ramp-up time, or an adjustment in the daily treatment schedule. Beyond the six (6) month visit, the therapy settings were left unchanged if the subject was losing weight and was not experiencing unacceptable adverse events. At the 6, 7, 8, 9, 10, and 11 month visits, the subjects had their % EWL from implant compared with the expected rate of 2.5% EWL per month. If the subject was either not losing weight at an expected rate or was experiencing unacceptable adverse events, the therapy settings were adjusted up or down. If a subject lost more than 2.5% EWL, no changes were made in the settings. Subjects that achieved the monthly %EWL but gained weight from the previous month had further adjustments in the therapy parameters. The maximum amplitude setting was 8.0 mA, and the maximum daily hours of VBLOC therapy was 18 hours. ## Scheduled visits Table 11 summarizes the scheduled visits and patient assessments. Adverse PMA P130019: FDA Summary of Safety and Effectiveness Data Page 27 {27} events and complications were recorded at all visits. Table 11. Schedule of trial events: Screening through 12 month follow-up | Screening [Enrollment] | Randomization/Implant/Initiation | Week 1 Visit 7 ±3 days after Implant | Follow-up Visits 2, 3, 4, 6, 8, 10, 12 weeks (±3 days); 4, 5, 6, 7, 8, 9, 10, 11, 12 months (±14 days) after randomization | | --- | --- | --- | --- | | • Informed consent • Inclusion/exclusion criteria assessments • Body weight • Body height • Vital signs* • Medication use assessment • Psychological assessment • Waist and hip circumferences • Clinical laboratory assessments • Subject Questionnaires • Physical exam • 7 day diet and activity diary • 12 lead ECG • Preoperative assessments (upper GI x-ray or upper endoscopy) • Device overview and training | • Body weight • Vital signs • Adverse event/medication use assessment • Randomized to treatment groups • Device implant (after all procedures above) | • Subject self-assessment (optional) • Body weight • Vital signs • Adverse event / medication use assessment • Device training • 7 day diet and activity diary • Blinding status • Weight management begins | • Subject self-assessment (optional) • Body weight • Vital signs* • Adverse event/ medication use assessment • Physical exam if needed • Clinical laboratory assessments (6 &12 months) • Waist and hip circumferences (12 months) • Weight management • Device interrogation • Current amplitude adjustments as indicated • Assess/maximize compliance with recharging • 12 lead ECG (4, 8, 12 months) • 7 day diet and activity diary • Blinding status (6 & 12 mo.) • Subject Questionnaires (3, 6 & 12 mo.) • Telephone contact with subject between visits (12 week- 6 months) | *Blood pressure collected in triplicate at screening, implant, and months 3, 6, 9, and 12 month visits. PMA P130019: FDA Summary of Safety and Effectiveness Data {28} PMA P130019: FDA Summary of Safety and Effectiveness Data Page 29 # 3. Clinical Endpoints With regards to safety, the primary safety endpoint of the ReCharge trial was to demonstrate that the 12-month serious adverse event (SAE) rate related to implant or revision procedures, device, or therapy was less than a performance goal of 15% among the subjects in the VBLOC group. With regards to effectiveness, there were two (2) co-primary effectiveness endpoints. The first co-primary effectiveness endpoint was percent excess weight loss (%EWL) at 12 months after randomization with ideal body weight calculated using the BMI method (i.e., the weight a subject would have with a BMI of 25 kg/m²). The objective of the analysis was to show that the mean %EWL in the VBLOC group is 10% greater than the %EWL in the control Sham group. The super-superiority margin of 10% was thought to represent a clinically meaningful difference between active VBLOC therapy and a Sham control. For each subject, the %EWL is the weight lost between baseline and 12 months divided by the subject's excess weight, where excess weight is the difference between baseline weight and ideal body weight. The analysis of %EWL was based on a t-test with a one-sided significance level of 2.5%. The primary analysis of this endpoint was based on the intent-to-treat group consisting of all randomized subjects. Note that not all subjects had a weight available at 12 months. For these subjects, their last available weight was used in place of the missing month 12 weight. The second co-primary effectiveness endpoint was based on responder rates with the following two requirements: (i) at least 55% of VBLOC subjects would achieve a %EWL of at least 20%; and (ii) at least 45% of VBLOC subjects would achieve a %EWL of at least 25%. With regard to success/failure criteria, the assessments of these objectives were based on observed rates rather than statistical hypothesis tests and according to the protocol both of these objectives should be met for trial success. # B. Accountability of PMA Cohort At the time of database lock, of 239 subjects enrolled in the PMA study, there were 162 subjects randomized to the VBLOC group and 77 subjects randomized to the Sham control group. However, only 233 subjects were implanted: Five (5) VBLOC subjects failed to get the fully-implanted device due to intra-operative exclusions and one (1) Sham control subject withdrew from the trial prior to the scheduled procedure time. Through the 12 month time point, the reasons given for subject withdrawal in the VBLOC treatment arm included three (3) intra-operative exclusions, 1 comorbid condition, and 1 surgeon discretion. "Subject decision" was the reason given for the subject who withdrew from the Sham control arm. Ninety-one percent (91%, n=147) of the randomized VBLOC and 86% (n=66) of the {29} randomized Sham subjects are available for analysis at the completion of the study, the 12 months post-operative visit (Figure 2). The combined number of VBLOC plus Sham subjects available at 12 months was eighty-nine percent $(89\%)$ . At 18 months, the follow-up rate was $72\%$ $(n = 117)$ in the VBLOC group and $55\%$ $(n = 42)$ in the Sham control group. ![img-2.jpeg](img-2.jpeg) Figure 2. ReCharge subject accountability # C. Study Population Demographics and Baseline Parameters Table 12 lists the baseline demographics of the Recharge study population of subjects enrolled in the U.S. During the June 17, 2014 Advisory Meeting, the Panel members commented that the majority of U.S. subjects were female and Caucasian. The trial included 239 randomized subjects (162 VBLOC and 77 Sham) at 10 investigational sites (8 in the U.S. and 2 in Australia). Of the 239 randomized subjects, 233 received an implanted device (157 VBLOC, 76 Sham). Among the randomized subjects, $84.9\%$ (141 VBLOC and 62 Sham) of the subjects were female, $92.9\%$ (149 VBLOC and 73 Sham) were Caucasian, the average age was 47 years (range: 18-65), average BMI at implant was $40.9\mathrm{kg} / \mathrm{m}^2$ (range: 34.4-48.4), and $6.3\%$ (9 VBLOC, 6 Sham Control) had type 2 diabetes mellitus. No significant differences were found between the VBLOC and sham groups for any of the recorded demographic and baseline variables. PMA P130019: FDA Summary of Safety and Effectiveness Data {30} Table 12. Baseline Demographics and Health Characteristics of Recharge Subjects | Characteristic | | VBLOC (n=162) | Sham (n=77) | P-value | | --- | --- | --- | --- | --- | | Gender | Female | 87.0% | 80.5% | 0.245 | | | Male | 13.0% | 19.5% | | | Age (years) at screening | | 47.1±10.3 | 46.6±9.4 | 0.693* | | Race | Caucasian | 92.0% | 94.8% | 0.592 | | | African American | 4.9% | 3.9% | 1.000 | | | Native American | 1.2% | 1.3% | 1.000 | | | Asian | 0.6% | 1 (1.3%) | 0.541 | | | Hawaiian/Pacific Islander | 0.6% | 0% | 1.000 | | Height (m) | | 1.7±0.1 | 1.7±0.1 | 0.112* | | BMI (kg/m², at implant) | | 40.9±2.8 | 40.9±3.1 | 0.969* | | Weight at implant (kg) | | 11216±11.8 | 123.0±11.3 | 0.117 | | Excess weight (kg) at implant | | 43.7±8.7 | 44.9±9.5 | 0.371 | | Waist circumference (cm) at screening | | 121.1±11.8 | 123.0±11.3 | 0.236 | | Type 2 diabetes mellitus | | 9 (5.6%) | 6 (7.8%) | 0.571 | | Hypertension | | 63 (38.9%) | 32 (41.6%) | 0.920 | | Dyslipidemia | | 91 (56.2%) | 46 (59.7%) | 0.884 | | Obstructive Sleep Apnea | | 33 (20.4%) | 23 (29.9%) | 0.267 | Note: Data are presented as mean ± SD for continuous variables. Data are presented as n (%) for categorical variables. P-values for continuous variables were calculated using a Student's t-test (no asterisk) or a Wilcoxon rank sum test (*) if the variable was not normally distributed based on the Shapiro-Wilk normality test. Categorical variables were compared using Fisher's exact test. ## D. Safety and Effectiveness Results ### 1. Safety Results The analysis of safety was based on the adverse event rates in the cohort of patients available for the 12 month evaluation. The key safety outcomes for this study are presented below. Adverse events are reported in Tables 13 thru 15. PMA P130019: FDA Summary of Safety and Effectiveness Data {31} # Serious Adverse Events in the Randomized Population Among the 162 subjects randomized to the VBLOC group (i.e., the intent-to-treat group), there were six (6) SAEs related to the device, implant/revision, or therapy. This yielded an observed SAE rate of $3.7\%$ (6/162, $95\%$ CI: $[1.4\%, 7.9\%]$ ) among the VBLOC subjects, which met the primary safety endpoint, because the upper bound of this confidence interval is less than $15\%$ . There were also nine (9) subjects who had SAEs related to the general surgical procedure. When these SAEs were counted as part of the primary safety endpoint, using an intent-to-treat analysis, the SAE rate was $8.6\%$ (14/162), with a $95\%$ CI of $[4.8\%, 14.1\%]$ , which also meets the performance goal of $15\%$ . The Sham group did not experience any adverse events related to the general surgical procedure or related to the device, implant/revision, or therapy. There were no deaths or unanticipated adverse events among subjects enrolled in the ReCharge Study. The Serious Adverse Events are summarized in Table 13. Table 13. Serious Adverse Events | Serious Adverse Events | VBLOC N=162 | | Sham Control N=77 | | | --- | --- | --- | --- | --- | | | N (%) Subjects | N Events | N (%) Subjects | N Events | | SAE related to device, implant/revision, or therapy | | | | | | Neuroregulator malfunction | 2 (1.2%) | 2 | 0 (0.0%) | 0 | | Atelectasis | 1 (0.6%) | 1 | 0 (0.0%) | 0 | | Gallbladder disease | 1 (0.6%) | 1 | 0 (0.0%) | 0 | | Emesis/vomiting | 1 (0.6%) | 1 | 0 (0.0%) | 0 | | Pain, neuroregulator site | 1 (0.6%) | 1 | 0 (0.0%) | 0 | | SAEs related to general surgical procedure | | | | | | Nausea | 6 (3.7%) | 6 | 0 (0.0%) | 0 | | Cirrhosis* | 1 (0.6%) | 1 | 0 (0.0%) | 0 | | Generalized ileus | 1 (0.6%) | 1 | 0 (0.0%) | 0 | | Intra-operative oozing | 1 (0.6%) | 1 | 0 (0.0%) | 0 | | SAEs related to pre-existing condition or not related | | | | | | Allergic reaction | 1 (0.6%) | 1 | 0 (0.0%) | 0 | | Chest pain | 1 (0.6%) | 1 | 0 (0.0%) | 0 | | Colitis | 1 (0.6%) | 1 | 0 (0.0%) | 0 | | Gallbladder disease | 1 (0.6%) | 1 | 0 (0.0%) | 0 | | Gastritis | 0 (0.0%) | 0 | 1 (1.3%) | 1 | | Infection, other | 1 (0.6%) | 1 | 1 (1.3%) | 1 | | Osteoarthritis | 1 (0.6%) | 1 | 0 (0.0%) | 0 | | Pain, abdominal | 1 (0.6%) | 1 | 0 (0.0%) | 0 | | Pain, other | 2 (1.2%) | 2 | 0 (0.0%) | 0 | | Palpitations | 1 (0.6%) | 1 | 0 (0.0%) | 0 | | Pericarditis | 1 (0.6%) | 1 | 0 (0.0%) | 0 | | Breast cancer | 0 (0.0%) | 0 | 1 (1.3%) | 1 | PMA P130019: FDA Summary of Safety and Effectiveness Data {32} | Serious Adverse Events | VBLOC N=162 | | Sham Control N=77 | | | --- | --- | --- | --- | --- | | | N (%) Subjects | N Events | N (%) Subjects | N Events | | Worsening back pain | 0 (0.0%) | 0 | 1 (1.3%) | 1 | *Subject not implanted. ## Adverse Events in the Implanted Population Five (5) subjects in the VBLOC arm and one (1) subject in the Sham control arm were enrolled in the ReCharge study, but did not receive device implantations. Using an analysis set that includes only those subjects who received the device implant, the most common adverse events (AEs) in the VBLOC Treatment group and those subjects randomized to the Sham Control group were pain and gastrointestinal (GI) symptoms, including heartburn, dysphagia, belching and nausea. AEs that occurred in ≥ 5% of VBLOC and/or Sham control subjects through 12 and 18 months are summarized in Tables 14 and 15. Table 14. Most common device, implant/revision procedure, or therapy related AEs through 12 months* | Adverse Event Type | VBLOC (n=157) | | Sham (n=76) | | | --- | --- | --- | --- | --- | | | Subjects | Events | Subjects | Events | | Neuroregulator site pain | 60 (38%) | 72 | 32 (42%) | 35 | | Other pain | 38 (24%) | 43 | 0 (0%) | 0 | | Heartburn/dyspepsia | 38 (24%) | 42 | 3 (4%) | 3 | | Other | 34 (22%) | 43 | 7 (9%) | 10 | | Abdominal pain | 20 (13%) | 26 | 2 (3%) | 2 | | Dysphagia | 13 (8%) | 13 | 0 (0%) | 0 | | Eructation/belching | 13 (8%) | 13 | 0 (0%) | 0 | | Nausea | 7 (4%) | 8 | 1 (1%) | 1 | | Chest pain | 9 (6%) | 9 | 2 (3%) | 2 | *Based on subjects with implanted device PMA P130019: FDA Summary of Safety and Effectiveness Data {33} Table 15. Most common device, implant/revision procedure, or therapy related AEs through 18 months* | Adverse Event Type | VBLOC (n=157) | | Sham (n=76) | | | --- | --- | --- | --- | --- | | | Subjects | Events | Subjects | Events | | Neuroregulator site pain | 60 (38%) | 75 | 32 (42%) | 36 | | Heartburn/dyspepsia | 41 (26%) | 45 | 3 (4%) | 3 | | Other pain | 41 (25%) | 49 | 0 (0%) | 0 | | Other | 37 (24%) | 46 | 7 (9%) | 10 | | Abdominal pain | 22 (14%) | 30 | 2 (3%) | 2 | | Eructation/belching | 14 (9%) | 14 | 0 (0%) | 0 | | Chest pain | 13 (8%) | 13 | 2 (3%) | 2 | | Dysphagia | 13 (8%) | 13 | 0 (0%) | 0 | | Nausea | 8 (5%) | 11 | 1 (1%) | 1 | | Abdominal cramps | 8 (5%) | 8 | 0 (0%) | 0 | *Based on subjects with implanted device ## Surgical removals of the device Through 12 months, there were 13 surgical explants of the device for the subjects who withdrew from the trial, including eight (8) in the Sham control group and five (5) in the VBLOC treatment group. Reasons given for explants in the VBLOC arm included subject decision (n=2), pain when doing physical activity (n=1), pain at the neuroregulator site (n=1), and heartburn (n=1). In the Sham control arm, reasons given for device explants included subject decision (n=4), shoulder pain which required an MRI (n=1), worsening irritable bowel symptoms (n=1), need for mastectomy (n=1), and pain at the neuromodulator implant site (n=1). Between the 12- and 18-month time points, there were 16 additional explants, including nine (9) in the Sham control group and 14 in the VBLOC treatment group. Twelve (12) explants in the VBLOC arm and seven (7) in the Sham arm were attributed to subject decision. Other reasons for device explant in the VBLOC arm included right arm pain (n=1), pain at the neuroregulator site (n=1), and upper quadrant pain (n=2). Other reasons for device explant in the Sham arm included the need for MRI for neck and back pain (n=1) and pain at the neuroregulator site (n=1). ## Surgical revisions Using per protocol analysis (i.e., those subjects who completed the study), eight (8) subjects (8/167, 4.8%) in the randomized VBLOC arm had nine (9) surgical revisions performed through 12 months: four (4) for device malfunction (2.4%), three (3) for pain at the neuroregulator site (1.8%), and two (2) for neuroregulator tilt (12.0%). There were no surgical revisions in the Sham control group. PMA P130019: FDA Summary of Safety and Effectiveness Data {34} # Device explants between 12 and 18 months Sixteen (16) patients in the VBLOC group had their device removed between 12 and 18 months after implant. Twelve (12) explants were for subject decision, one (1) for right arm pain due to thoracic outlet syndrome, one (1) for pain at the neuroregulator site, and two (2) for upper quadrant pain. Nine (9) patients in the Sham control group had their device removed between 12 and 18 months after implant. Seven (7) explants were for subject decision, one (1) for need for MRI for neck and back pain, and one (1) due to pain at the neuroregulator site. # 2. Effectiveness Results The analysis of effectiveness was based on the 233 evaluable at the 12 month time point. Key effectiveness outcomes are presented in Figure 3. ![img-3.jpeg](img-3.jpeg) Figure 3. Percent Excess Weight loss (%EWL) in VBLOC and Control Arms through 18 months. The dashed red line indicates the 12 month time point of the primary analysis of %EWL in the ReCharge trial. The first co-primary effectiveness endpoint was percent excess weight loss (%EWL) at 12 months after randomization, with ideal body weight calculated using the BMI method (i.e., the weight a subject would have with a BMI of $25\mathrm{kg} / \mathrm{m}^2$ ). The goal of the analysis of this co-primary endpoint was to show that the mean %EWL in the VBLOC group is at least $10\%$ greater than the %EWL in the Sham control group (i.e., there was a pre-specified super-superiority margin of $10\%$ ). Using the intent-to-treat study population (i.e., all subjects who enrolled in the study), the average %EWL at 12 months was $24.4\%$ ( $\mathrm{SD} = 23.6\%$ ) in the VBLOC group and $15.9\%$ PMA P130019: FDA Summary of Safety and Effectiveness Data {35} $(SD = 17.7\%)$ in the Sham control group, resulting in an average difference between the VBLOC and sham control groups of $8.5\%$ (95% CI: [3.1%, 13.9%]). While these results would support a conclusion that average $\% \mathrm{EWL}$ is higher in the VBLOC group than in the Sham control group, the pre-specified superiority margin of $10\%$ was not achieved, because the lower bound of the confidence interval is less than $10\%$ . At 18 months, the observed mean $\% \mathrm{EWL}$ in the VBLOC group was $25.2\%$ (95% CI: [20.6, 29.8]) and $11.7\%$ (95% CI: [5.4, 18.0]) in the Sham control group, resulting in a treatment difference of $13.5\%$ (95% CI: [5.7, 21.3]). Figure 3 shows the $\% \mathrm{EWL}$ through 18 months. Analysis of the 18-month data from the ReCharge study suggests maintenance of the weight loss at 12 months. The second co-primary effectiveness endpoint had two (2) requirements: (i) at least $55\%$ of VBLOC subjects would achieve a $\% \mathrm{EWL}$ of at least $20\%$ ; and (ii) at least $45\%$ of VBLOC subjects would achieve a $\% \mathrm{EWL}$ of at least $25\%$ . The assessments of these objectives were based on observed rates. According to the protocol both of these objectives should be met for trial success. This co-primary objective was also not achieved: (i) $52.5\%$ ( $&lt; 55\%$ ) of VBLOC subjects had a $\% \mathrm{EWL}$ of at least $20\%$ ; and (ii) $38.3\%$ ( $&lt; 45\%$ ) of VBLOC subjects had a $\% \mathrm{EWL}$ of at least $25\%$ . The Sham control group had $32.5\%$ ( $n = 25$ ) of subjects with $20\%$ ( $n = 18$ ) or greater EWL and $23.4\%$ of subjects with $25\%$ or greater EWL. Although the study did not include a pre-determined endpoint for factors associated with health improvements, data were collected on the 12 month change in parameters such as cholesterol, triglycerides, blood pressure, fasting glucose, and HbA1c. Results suggested that there were small improvements of various parameters in both the VBLOC group and the Sham control group from baseline to 12 months, but the change in the VBLOC group was not statistically significantly different from the change in the Sham control group (Table 16). Table 16. Summary of factors associated with co-morbidities at screening, month 12, and changes from screening to month 12. | Parameter | Study Visit | VBLOC Mean ± SD (N Subjects) | Sham Mean ± SD (N Subjects) | Difference (VBLOC – Sham) Mean ± SD | | --- | --- | --- | --- | --- | | Systolic Blood Pressure | Screening | 127.9 ± 12.5 (162) | 129.9 ± 12.8 (77) | | | | Month 12 | 121.9 ± 11.8 (147) | 125.5 ± 15.7 (66) | | | | Change | -5.5 ± 14.2 (147) | -4.0 ± 13.5 (66) | -1.5 ± 14.0 | PMA P130019: FDA Summary of Safety and Effectiveness Data {36} | Parameter | Study Visit | VBLOC Mean ± SD (N Subjects) | Sham Mean ± SD (N Subjects) | Difference…
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...