HER2 CISH PHARMDX KIT

P100024 · Dako Denmark A/S · NYQ · Nov 30, 2011 · Pathology

Device Facts

Record IDP100024
Device NameHER2 CISH PHARMDX KIT
ApplicantDako Denmark A/S
Product CodeNYQ · Pathology
Decision DateNov 30, 2011
DecisionAPWD
Regulation21 CFR 864.1890
Device ClassClass 2

Indications for Use

HER2 CISH pharmDx™ Kit is intended for dual-color chromogenic visualization of signals achieved with directly labeled in situ hybridization probes targeting the HER2 gene and centromeric region of chromosome 17. The Kit is designed to quantitatively determine HER2 gene status in formalin-fixed, paraffin-embedded breast cancer tissue specimens. Red and blue chromogenic signals are generated on the same tissue section for evaluation under bright field microscopy. The CISH procedure is automated using Dako Autostainer instruments. HER2 CISH pharmDx™ Kit is indicated as an aid in the assessment of patients for whom Herceptin™ (trastuzumab) treatment is being considered. Results from the HER2 CISH pharmDx™ Kit are intended for use as an adjunct to the clinicopathologic information currently used for estimating prognosis in stage II, node-positive breast cancer patients. This kit is for in vitro diagnostic (IVD) use only.

Device Story

The HER2 CISH pharmDx™ Kit is an in vitro diagnostic assay for identifying and quantifying HER2/neu gene amplification in FFPE breast cancer tissue. The device uses Texas-Red labeled DNA probes for the HER2 gene and fluorescein-labeled PNA probes for the centromeric region of chromosome 17 (CEN-17). The procedure is automated on Dako Autostainer instruments. Following hybridization, fluorescent signals are converted to red (HER2) and blue (CEN-17) chromogenic signals via an immunohistochemical reaction using anti-Texas Red/alkaline phosphatase and anti-FITC/horseradish peroxidase conjugates. Pathologists evaluate the stained tissue using bright field microscopy to enumerate copy numbers and calculate the HER2/CEN-17 ratio. A ratio ≥2.0 indicates gene amplification. This output assists clinicians in determining patient eligibility for Herceptin therapy and provides prognostic information for stage II, node-positive breast cancer patients.

Clinical Evidence

Clinical validation involved a comparative study of 365 invasive breast cancer specimens. The study demonstrated 97.7% overall agreement (Kappa 0.90) between HER2 CISH and PathVysion FISH, with 90.9% positive agreement and 98.7% negative agreement. A separate comparison between HER2 CISH and HER2 FISH pharmDx™ showed 98.3% overall agreement (Kappa 0.92). Reproducibility studies showed site-to-site/day-to-day CVs ≤16% for amplified and ≤12% for non-amplified specimens. The CISH success rate was 96.7%.

Technological Characteristics

Dual-color chromogenic in situ hybridization (CISH) assay. Uses Texas-Red labeled DNA probes (HER2) and fluorescein-labeled PNA probes (CEN-17). Employs anti-Texas Red/alkaline phosphatase and anti-FITC/horseradish peroxidase conjugates for signal visualization. Form factor: reagent kit for use with automated Dako Autostainer instruments. Analysis via bright field microscopy. Storage: 2-8 °C.

Indications for Use

Indicated for patients with invasive breast cancer to determine HER2 gene status as an aid in assessing eligibility for Herceptin (trastuzumab) therapy and as an adjunct for estimating prognosis in stage II, node-positive breast cancer patients.

Regulatory Classification

Identification

This device is intended to detect her2 gene amplification in formalin-fixed, paraffin-embedded breast carcinoma tissue sections using chromogenic in situ hybridization and brightfield microscopy. Indicated as an aid in the assessment of patients for whom herceptin. (trastuzumab) treatment is being considered. Interpretation of test results must be made within the context of the patients clinical history by a qualified pathologist.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{1} Summary of Safety and Effectiveness Data (SSED) # SUMMARY OF SAFETY AND EFFECTIVENESS DATA (SSED) ## I. GENERAL INFORMATION Device Generic Name: In vitro diagnostic test kit for HER2 gene amplification in formalin-fixed, paraffin-embedded (FFPE) tissue sections using Chromogenic In Situ Hybridization (CISH) Device Trade Name: HER2 CISH pharmDx™ Kit Applicant’s Name and Address: Dako Denmark A/S Produktionsvej 42 DK-2600 Glostrup Denmark Date(s) of Panel Recommendation: None Premarket Approval Application (PMA) Number: P100024 Date of FDA Notice of Approval: November 30, 2011 Expedited: Not applicable. ## II. INDICATIONS FOR USE HER2 CISH pharmDx™ Kit is intended for dual-color chromogenic visualization of signals achieved with directly labeled in situ hybridization probes targeting the HER2 gene and centromeric region of chromosome 17. The Kit is designed to quantitatively determine HER2 gene status in formalin-fixed, paraffin-embedded breast cancer tissue specimens. Red and blue chromogenic signals are generated on the same tissue section for evaluation under bright field microscopy. The CISH procedure is automated using Dako Autostainer instruments. HER2 CISH pharmDx™ Kit is indicated as an aid in the assessment of patients for whom Herceptin™ (trastuzumab) treatment is being considered. Results from the HER2 CISH pharmDx™ Kit are intended for use as an adjunct to the clinicopathologic information currently used for estimating prognosis in stage II, node-positive breast cancer patients. This kit is for in vitro diagnostic (IVD) use only. ## III. CONTRAINDICATIONS The contraindications can be found in the HER2 CISH pharmDx™ kit labeling. PMA P100024: FDA Summary of Safety and Effectiveness Data page 1 of 20 5 {2} Summary of Safety and Effectiveness Data (SSED) ## IV. WARNINGS AND PRECAUTIONS The warnings and precautions can be found in HER2 CISH pharmDx™ Kit labeling. ## V. DEVICE DESCRIPTION HER2 CISH pharmDx™ Kit is a dual-color chromogenic in situ hybridization (CISH) assay used for slide-based, bright field microscope identification and quantification of HER2/neu gene amplification in formalin-fixed, paraffin-embedded (FFPE) tissue sections. HER2 CISH pharmDx™ Kit contains all specific reagents required to complete a CISH procedure. In the procedure, the targets (HER2 and CEN-17) are labeled by in situ hybridization using fluorescent-labeled DNA and PNA probes. Before signal evaluation the fluorescent signals are converted to chromogenic signals in an immunohistochemical reaction that includes blocking of endogenous peroxidase activity, incubation with CISH Antibody Mix and subsequently visualization of the signals by incubation with chromogens. The HER2/CEN-17 Probe Mix consists of a mixture of Texas-Red labeled DNA cosmid clones that cover 218 kb of the chromosomal region that includes the HER2 gene plus a mixture of fluorescein-labeled (PNA) probes targeted towards the centromeric region of chromosome 17. The probes are pre-mixed in hybridization buffer for ease of use. Unlabeled PNA probes are also included to suppress sequences contained within the target loci that are common to other chromosomes. The CISH Antibody Mix includes anti-Texas Red conjugated with alkaline phosphatase (AP) that convert the Texas Red-labeled signals (HER2) to red chromogenic signals and anti-FITC conjugated with horseradish peroxidase (HRP) that convert the FITC-labeled signals (CEN-17) to blue chromogenic signals. Upon specific hybridization at the two targets and subsequent conversion to chromogenic signals a red signal is seen at each HER2 gene and a blue signal is seen at the chromosome 17 centromere. In each cell, the copy numbers of HER2 and CEN-17 are enumerated. The presence of amplified HER2 is determined by the ratio of the average copy number of HER2 and CEN-17. Gene amplification is defined as a HER2/CEN-17 ratio ≥2.0, while HER2/CEN-17 < 2.0 indicates non-amplification. ## VI. ALTERNATIVE PRACTICES AND PROCEDURES There are several alternative methods for the detection of HER2 gene amplification. Devices utilizing fluorescent in situ hybridization (FISH) or chromogenic in situ hybridization (CISH) methodologies for gene amplification determination in human breast cancer tissue specimens are commercially available. This is in addition to immunohistochemistry (IHC), another alternative procedure for detection of gene product overexpression in human breast. A physician should consider these alternatives and select the appropriate method to be used. PMA P100024: FDA Summary of Safety and Effectiveness Data page 2 of 20 {3} Summary of Safety and Effectiveness Data (SSED) ## VII. MARKETING HISTORY HER2 CISH pharmDx™ Kit has been available in the following European Union countries - Switzerland, Norway, Lichtenstein and Iceland since June 2010 and in Canada since September 2010. This device has not been withdrawn for any reason related to safety and effectiveness. ## VIII. POTENTIAL ADVERSE EFFECTS OF THE DEVICE ON HEALTH HER2 CISH pharmDx™ Kit is intended for in vitro diagnostic use only. As with any in vitro diagnostic test, there are potential risks associated with incorrect result interpretation. A false positive test result would likely assign patients to receive a more aggressive therapy, possibly exposing the patient to serious side effects and, in rare cases, death. Alternatively, a false negative test result may exclude a patient who might benefit from therapy, potentially resulting in a poor clinical outcome. ## IX. SUMMARY OF NON-CLINICAL STUDIES Analytical performance testing was performed to evaluate the safety and effectiveness of HER2 CISH pharmDx™ Kit. The validation included analytical sensitivity, analytical specificity, locus specificity, robustness, reproducibility, repeatability and stability testing. ### A. Laboratory Studies #### 1. Analytical Specificity and Sensitivity ##### a. Hybridization Efficiency To verify that the HER2 CISH pharmDx Kit specifically detect their target sequences, the HER2 CISH pharmDx probe mix was used to stain human metaphase spreads. The HER2 DNA probes in the HER2/CEN-17 Probe Mix have been end-sequenced and mapped to confirm total coverage of 218 kb including the HER2 gene. The CEN-17 PNA probes in the HER2/CEN-17 Probe Mix have been tested individually and in combination to confirm their specific hybridization to the centromeric region of chromosome 17. A total of 250 metaphase spreads were evaluated for specific hybridization of the HER2 DNA and CEN-17 PNA probe mixes. This study was done according to the instructions provided in the labeling. To ensure that there was no cross-hybridization to chromosomes other than chromosome 17, studies were performed on metaphase spreads according to standard Dako QC procedures. In all 250 cases the hybridization was specific for chromosome 17. No cross-hybridization to loci on other chromosomes was observed in any of the 250 cases. PMA P100024: FDA Summary of Safety and Effectiveness Data page 3 of 20 {4} Summary of Safety and Effectiveness Data (SSED) ## b. Analytical Sensitivity (quantitative) The analytical sensitivity for HER2 CISH measures the assay’s ability to detect the target substances (HER2/CEN-17). The analytical sensitivity of HER2 CISH pharmDx™ Kit probes was determined using 18 normal human breast tissue specimens. All specimens showed a HER2/CEN-17 ratio within the preset acceptance criteria for normal breast tissue (0.97-1.09) and with consistent results between FISH (previously approved in P040005) and CISH. Assay sensitivity showed that the assay performs according to specifications (within the preset acceptance criteria) with an estimated mean ratio of HER2/CEN-17 of 1.03 (S.D. = 0.04) (normal breast tissue has an estimated HER2/CEN-17 ratio between 0.8-1.4) and there is consistency between HER2 CISH and HER2 FISH measurements. ## c. Analytical Specificity Studies were performed to assess the analytical specificity for HER2 CISH and to measure the assay’s ability to solely identify target substances HER2/CEN-17 without interference from other substances. Stained slides were evaluated for presence of signals in the absence of probes or antibodies with complete FISH and CISH staining used as reference. Analytical specificity was tested on 18 normal human breast tissue specimens. Assay specificity showed that there is no non-specific binding of CISH Antibody Mix or Red and Blue Chromogen that result in signals visible in CISH upon omission of Probe Mix and Antibody Mix. ## 2. Assay Robustness Studies ### a. Tissue Thickness The effect of section thickness on the performance of HER2 CISH pharmDx was studied. A total of 10 serial sections of breast carcinoma (two specimens representing ± HER2 gene amplification on each slide) with different thickness (3-7 μm) were tested. Repeatability of HER2/CEN-17 ratio evaluated on section of different thickness showed a coefficient of variance of 3% for the non-amplified specimen and 6% for the amplified specimen, within the preset acceptance criteria (see Table 1). This is within the interval observed for sections of equal thickness. Based on the studies, the recommended thickness for sections is 4 - 6 μm. Table 1: Section thickness PMA P100024: FDA Summary of Safety and Effectiveness Data page 4 of 20 {5} Summary of Safety and Effectiveness Data (SSED) | Duplicate measurements with varying slide thickness | Ratio (HER2/CEN-17) Non-amplified samples | Ratio (HER2/CEN-17) Amplified samples | | --- | --- | --- | | 3 μm | 1.06 | 9.81 | | 3 μm | 1.00 | 10.28 | | 4 μm | 1.03 | 9.57 | | 4 μm | 1.06 | 10.33 | | 5 μm | 1.00 | 11.38 | | 5 μm | 1.06 | 9.82 | | 6 μm | 1.00 | 9.47 | | 6 μm | 1.03 | 10.33 | | 7 μm | 1.09 | 9.35 | | 7 μm | 1.09 | 10.16 | | %C.V. | 3 | 6 | ## b. Incubation Time and Temperature Robustness of HER2 CISH pharmDx™ Kit assay was tested by varying time and temperature for incubation with CISH Antibody Mix, Red Chromogen Solution, Blue Chromogen Solution and counterstaining. No significant difference in results was observed at the following experimental conditions (unless otherwise indicated): - Antibody Mix incubation time: 25, 27, 30, 33 and 35 minutes. 25 minutes incubation with Antibody Mix resulted in a loss of red signal intensity while no significant differences were observed at all other time points. Recommended incubation time is 30 minutes. - Red Chromogen Solution incubation time for 8, 9, 10, 11 and 12 minutes. Recommended incubation time is 10 minutes. - Blue Chromogen Solution incubation time for 8, 9, 10, 11 and 12 minutes. Recommended incubation time is 10 minutes. - Counterstaining incubation time for 4, 5 and 6 minutes and concentration at 1:4, 1:5 and 1:6 dilutions with a recommended incubation time of 5 minutes at a 1:5 dilution. The robustness of the steps including pre-treatment, denaturation/hybridization and post hybridization have been tested previously as part of the HER2 FISH pharmDx™ Kit assay development. No significant difference in results was observed at the following experimental conditions (unless otherwise indicated): - Pre-treatment at 7, 10 and 13 minutes combined with each of the temperatures 89, 92 and 95-99 °C. Recommendation is 10 minutes at ≥ 95 °C. PMA P100024: FDA Summary of Safety and Effectiveness Data page 5 of 20 {6} PMA P100024: FDA Summary of Safety and Effectiveness Data page 6 of 20 # Summary of Safety and Effectiveness Data (SSED) - Pepsin incubation times of 2, 5, 10, 15 and 18 minutes at room temperature. Recommendation is 5-15 minutes (depending on tissue fixation). - Denaturation temperatures of 72, 77, 82, 87 and 92 °C were tested. Recommendation is 5 minutes at 82 °C. - Hybridization time of 17 hours combined with each of the temperatures 40, 45 and 50 °C and hybridization times of 10, 12 and 14 hours at a temperature of 45 °C. Recommendation is 14-20 hours at 45 °C. - The stringent wash was tested for 10 minutes at 60, 65 and 70 °C and for 5, 10 and 15 minutes at 65 °C. Stringent wash for 10 minutes at 70 °C, and stringent wash for 15 minutes at 65 °C resulted in loss of signals, whereas no significant difference in results was observed at the other time/temperature combinations. Recommendation is 10 minutes at 65 °C. - Dilution of Stringent Wash Buffer was tested: 1:10, 1:15, 1:20, 1:30 and 1:40. The 1:40 dilution of Stringent Wash Buffer resulted in loss of signals, whereas no significant difference in signal intensity was observed at the other dilutions. Recommendation is 1:20 dilution. ## c. Stability Testing Results of real-time stability studies indicate that HER2 CISH pharmDx™ Kit is stable for 10 months when stored per instructions in the package insert, i.e., 2-8 °C in the dark. The stability studies included transport simulation (48 hours, 22 °C), 20 minutes working stability of the Red Chromogen Solution (at room temperature) and 8 days working stability of the Blue Chromogen Solution when stored at 2-8 °C in the dark and on-board stability for 30 hours (Ready to Use reagents). The performance of the kits was measured on tissue specimens according to the device labeling. ## d. Repeatability Repeatability of the HER2/CEN-17 ratio was investigated using 9 different FFPE breast carcinoma specimens with different HER2 gene status. Consecutive sections of each specimen were tested three times under the same conditions (in the same run) and the HER2/CEN-17 ratio was determined. Repeatability of HER2/CEN-17 ratio showed a coefficient of variance between 1% and 7% (within the preset acceptance criteria) with a tendency for the higher values found for the amplified specimens (see Table 2). Repeatability on consecutive sections of breast cancer specimens with different thickness (3-7 μm) was tested with HER2 CISH pharmDx™ Kit. The coefficient of variance of the HER2/CEN-17 ratio in this study was found to be 3-6%, i.e. in the same range as for tissue of equal thickness. {7} Summary of Safety and Effectiveness Data (SSED) Table 2: Repeatability (intra-run) | Specimen ID | Ratio (HER2/CEN-17) Repeat 1 | Ratio (HER2/CEN-17) Repeat 2 | Ratio (HER2/CEN-17) Repeat 3 | %CV | | --- | --- | --- | --- | --- | | 57210 | 3.12 | 3.33 | 3.23 | 3.0 | | 57316 | 1.02 | 1.03 | 1.04 | 1.0 | | 57342 | 1.00 | 1.09 | 1.05 | 4.0 | | 54573 | 1.15 | 1.10 | 1.18 | 4.0 | | 57240 | 1.03 | 1.05 | 1.03 | 1.0 | | 57264 | 1.06 | 1.05 | 1.13 | 4.0 | | 57304 | 7.00 | 6.77 | 6.50 | 4.0 | | 57314 | 8.50 | 8.21 | 8.33 | 2.0 | | 57308 | 7.36 | 8.09 | 7.15 | 7.0 | e. Reproducibility Reproducibility – lot-to-lot Lot-to-lot reproducibility demonstrates the degree of agreement between evaluations of HER2/CEN-17 ratio carried out using three different production lots of HER2 CISH pharmDx™ Kit. Reproducibility was tested on nine different breast carcinoma specimens with different HER2 gene status. HER2 gene status for each specimen remained unchanged from lot-to-lot. The coefficient of variation of the HER2/CEN-17 ratio was found to be between 1-7% for non-amplified specimens and 7-12% for amplified specimens (see Table 3) within the preset acceptance criteria. Table 3: Reproducibility (lot-to-lot) | Specimen ID | Ratio (HER2/CEN-17) Lot 1 (Lot A) | Ratio (HER2/CEN-17) Lot 2 (Lot B) | Ratio (HER2/CEN-17) Lot 3 (Lot C) | %CV | | --- | --- | --- | --- | --- | | 57308 | 7.36 | 7.48 | 8.33 | 7.0 | | 57304 | 5.82 | 6.56 | 6.95 | 9.0 | | 54573 | 1.13 | 1.23 | 1.22 | 5.0 | | 57314 | 8.19 | 8.32 | 9.3 | 7.0 | | 57564 | 1.15 | 1.02 | 1.05 | 6.0 | | 57240 | 1.05 | 1.06 | 1.08 | 1.0 | | 57210 | 3.55 | 3.79 | 3.0 | 1.2 | | 57342 | 1.06 | 1.15 | 1.0 | 7.0 | | 57316 | 1.09 | 1.02 | 1.0 | 5.0 | 3. Non-clinical Studies – External f. Site-to-Site and Day-to-Day Reproducibility The reproducibility study was a three-site, blinded, randomized study, using formalin-fixed, paraffin-embedded (FFPE) human breast cancer specimens with different levels of HER2 gene amplification. Nine specimens were analyzed and scored on three non-consecutive days at three study sites. These nine specimens included three non-amplified specimens, three HER2 IHC 2+ specimens, as determined by HercepTest™ according to the manufacturer’s recommendations, and three amplified specimens. PMA P100024: FDA Summary of Safety and Effectiveness Data page 7 of 20 {8} Summary of Safety and Effectiveness Data (SSED) # Results for site-to-site and day-to-day reproducibility Data for the HER2/CEN-17 ratios obtained from the reproducibility study are tabulated in Table 4. Table 4: Site-to-Site and Day-to-Day Reproducibility - HER2/CEN-17 ratios using automated CISH. | Specimen | Type | Day | HER2/CEN-17 Ratio | | | | --- | --- | --- | --- | --- | --- | | | | | US Site 1 | US Site 2 | Dako Internal | | 54573 | Non-amplified | 1 | 1.21 | 1.05 | 1.15 | | | | 2 | 1.05 | 1.16 | 1.17 | | | | 3 | 1.26 | 1.17 | 1.06 | | 57264 | Non-amplified | 1 | 1.20 | 1.08 | 1.09 | | | | 2 | 1.23 | 1.31 | 0.98 | | | | 3 | 1.13 | 1.06 | 0.97 | | 57316 | Non-amplified | 1 | 1.05 | 1.14 | 1.10 | | | | 2 | 0.97 | 1.15 | 1.04 | | | | 3 | 1.11 | 1.31 | 1.07 | | 57210 | IHC 2+ | 1 | 2.59 | 2.33 | 3.15 | | | | 2 | 2.63 | 2.29 | 2.94 | | | | 3 | 2.46 | 3.52 | 3.03 | | 57240 | IHC 2+ | 1 | 0.98 | 1.09 | 1.06 | | | | 2 | 1.26 | 1.16 | 1.03 | | | | 3 | 1.08 | 1.17 | 1.00 | | 57342 | IHC 2+ | 1 | 1.28 | 1.21 | 1.06 | | | | 2 | 1.19 | 1.13 | 1.00 | | | | 3 | 0.90 | 1.12 | 1.03 | | 57304 | Amplified | 1 | 8.22 | 6.94 | 7.02 | | | | 2 | 6.37 | 7.12 | 6.62 | | | | 3 | 5.89 | 6.00 | 6.11 | | 57308 | Amplified | 1 | 9.08 | 5.16 | 7.72 | | | | 2 | 8.50 | 7.60 | 7.41 | | | | 3 | 8.55 | 6.48 | 6.76 | | 57314 | Amplified | 1 | 7.64 | 6.33 | 8.26 | | | | 2 | 6.68 | 7.12 | 8.92 | | | | 3 | 8.24 | 5.96 | 8.32 | The HER2/CEN-17 ratios were analyzed using a variance component model where estimated CVs were determined from log-transformed HER2/CEN-17 ratios. This resulted in CV estimates just below 10% for non-amplified specimens, at 12% for IHC 2+ specimens and close to 15% for the amplified specimens (Table 5). The measurements made on PMA P100024: FDA Summary of Safety and Effectiveness Data {9} # Summary of Safety and Effectiveness Data (SSED) different sites and different days seem to differ only slightly more than putative measurements made the same day at the same site (the residual error). Table 5: Variance Component %CV estimates | | %CV estimate | | | | --- | --- | --- | --- | | | Non-amplified | IHC 2+ | Amplified | | Site-to-site | 9.0% | 12.2% | 15.8% | | Day-to-day | 8.4% | 12.2% | 13.6% | | Residual error | 8.4% | 12.2% | 13.1% | ## g. Observer-to-Observer Reproducibility The objective of the study was to evaluate observer-to-observer variation on the HER2/CEN-17 ratio when specimens stained by HER2 CISH pharmDx™ Kit was read by three different observers: - Observer 1 – US certified pathologist (original observer) - Observer 2 – Additional US certified pathologist - Observer 3 – Internal observer Dako Denmark A/S The observer-to-observer study was performed as a blinded, randomized, observer study using formalin-fixed, paraffin-embedded (FFPE) human breast cancer specimens that were a subset of the FFPE breast cancer specimens stained in the method comparison study. The 100 specimens (one specimen per slide) included in the observer-to-observer study were selected using the following rules: 1) Out of the 295 consecutive specimens with successful staining, 50 specimens were randomly selected (47 non-amplified, 3 amplified). 2) Out of the 57 additional specimens (with IHC HER2 2+) with successful staining, 50 specimens were randomly selected (39 non-amplified, 11 amplified). Furthermore, out of the 100 specimens there are 5 with a ratio within the borderline range (1.8-2.2) based on Observer 1 results. Table 6 – Observer-to-Observer Reproducibility Results | | Observer 2 Status | | Total | | | --- | --- | --- | --- | --- | | | | Non-amplified | | Amplified | | Observer 1 Status | Non-amplified | 85 | 1 | 86 | | | Amplified | 7 | 7 | 14 | | Total | | 92 | 8 | 100 | PMA P100024: FDA Summary of Safety and Effectiveness Data page 9 of 20 13 {10} Summary of Safety and Effectiveness Data (SSED) | | Observer 3 Status | | Total | | | --- | --- | --- | --- | --- | | | | Non-amplified | | Amplified | | Observer 1 Status | Non-amplified | 84 | 2 | 86 | | | Amplified | 2 | 12 | 14 | | Total | | 86 | 14 | 100 | | | Observer 3 Status | | Total | | | --- | --- | --- | --- | --- | | | | Non-amplified | | Amplified | | Observer 2 Status | Non-amplified | 86 | 6 | 92 | | | Amplified | 0 | 8 | 8 | | Total | | 86 | 14 | 100 | Each of the 100 selected specimens originally evaluated in the Method Comparison study was evaluated by two additional observers for a total of three observations per slide. The %CV for the HER2/CEN-17 ratio was calculated for each specimen across all three observers. The mean %CV for the HER2/CEN-17 ratio for all 100 specimens is 12.4% while for the 87 non-amplified samples it was 10.9% and for the 13 amplified samples, 22.6%. ## B. Animal Studies None. ## C. Additional Studies None. ## X. SUMMARY OF PRIMARY CLINICAL STUDIES The safety and effectiveness of HER2 CISH pharmDx™ Kit has been evaluated in the Dako sponsored clinical study: Comparative study of HER2 CISH pharmDx™ Kit with the PathVysion® HER-2 DNA Probe Kit and HER2 FISH pharmDx™ Kit. This study provided comparative data between HER2 CISH pharmDx™ Kit and PathVysion HER-2 DNA Probe Kit, an FDA approved, and commercially available fluorescence In-situ hybridization (FISH) method (P980024). Data from this clinical study was the basis of this PMA approval decision. In order to support a prognostic claim a method comparison study between HER2 CISH pharmDx™ Kit and HER2 FISH pharmDx™ Kit (P040005) was performed. A summary of the clinical study is presented below. ### A. Study Design The design of the study was comparative, where HER2 gene status achieved by the HER2 CISH pharmDx™ Kit was compared to HER2 gene status obtained by the PathVysion® HER-2 DNA Probe Kit (FISH) and to the HER2 FISH pharmDx™ Kit PMA P100024: FDA Summary of Safety and Effectiveness Data {11} Summary of Safety and Effectiveness Data (SSED) on formalin-fixed and paraffin-embedded (FFPE) histological sections from breast cancer specimens. Only complete tissue sections were stained and evaluated in the study. 1. Clinical Inclusion and Exclusion Criteria Enrollment in the Comparative study of HER2 CISH pharmDx™ Kit with the PathVysion® HER-2 DNA Probe Kit and HER2 FISH pharmDx™ Kit study was limited to specimens that met the following inclusion criteria: - A confirmed pathology diagnosis of invasive breast cancer - Adequate tissue specimen was available (a minimum of 7 slides were needed) - The sample was residual - The HercepTest™ IHC score, as per the manufacturer's Package Insert, was known for the specimen - Fixation history of the specimen was known Specimens were not permitted to enroll in the Comparative study of HER2 CISH pharmDx™ Kit with the PathVysion® HER-2 DNA Probe Kit and HER2 FISH pharmDx™ Kit study if they met any of the following exclusion criteria: - Inadequate or no existing tissue specimen for the entire study available - Missing tissue specimen - Tissue specimen was from a core or needle biopsy - A specimen could be excluded from some or all of the analysis if either CISH or FISH fails in both of the two attempts - Tissue sections from specimen that do not contain invasive tumor - Specimen was not fixed in neutral buffered formalin 2. Follow-up Schedule Not applicable. 3. Clinical Endpoints The primary objective of the study was to investigate the concordance between the HER2 gene status obtained using HER2 CISH pharmDx™ Kit, and the HER2 gene status obtained using PathVysion® HER-2 DNA Probe Kit (FISH) on serial sections of the same breast cancer specimens. The secondary objective of the study was to investigate the concordance between the HER2 gene status obtained using Dako HER2 CISH pharmDx™ Kit, and the HER2 gene status obtained using Dako HER2 FISH pharmDx™ Kit on serial sections of the same breast cancer specimens. PMA P100024: FDA Summary of Safety and Effectiveness Data page 11 of 20 15 {12} # Summary of Safety and Effectiveness Data (SSED) The tertiary objective was to compare slide evaluation times with the HER2 CISH pharmDx™ Kit and the PathVysion® HER-2 DNA Probe Kit (FISH). ## B. Description of Samples in the PMA Cohort Three hundred sixty-five invasive breast cancer specimens selected in chronological order were included in the study. For the first 304 specimens the incoming specimens were selected regardless of the HercepTest™ score. For the next 61 specimens only those with a HercepTest™ IHC 2+ score were selected in order to enrich the study population with IHC 2+ cases. From these 365 specimens, consecutive sections from 350 breast cancer specimens were successfully analyzed by both HER2 CISH pharmDx™ Kit and PathVysion HER-2 DNA Probe Kit (FISH). ## C. Study Population Demographics and Baseline Parameters The demographics of the study population are typical for a HER2 testing study performed in the US at a reference laboratory. A total of 365 FFPE breast cancer specimens from different patients were enrolled in this comparative study. The specimens were selected chronologically as they arrived at the US reference laboratory for HER2 status determination. As the specimens were consecutive specimens coming in to central lab from hospitals across the United States it is expected that the specimens in the study represent characteristics of the US breast cancer population. ## D. Safety and Effectiveness Results ### 1. Method Comparison Study - HER2 CISH pharmDx™ Kit and PathVysion HER-2 DNA Probe Kit Test results were obtained from 365 breast cancer specimens selected in chronological order. For the first 304 specimens the incoming specimens were selected regardless of the HercepTest™ score. For the next 61 specimens only those with a HercepTest™ IHC 2+ score were selected in order to enrich the study population with IHC 2+ cases. Table 7: Numbers of valid and missing test results for the comparison between HER2 CISH pharmDx™ Kit and PathVysion HER-2 DNA Probe Kit (FISH) assays. | Type | | | HercepTest | CISH | PathVysion | | --- | --- | --- | --- | --- | --- | | Consecutive | N | Valid | 304 | 295 | 300 | | | | Missing | 0 | 9 | 4 | | IHC2+ | N | Valid | 61 | 57 | 61 | | | | Missing | 0 | 4 | 0 | PMA P100024: FDA Summary of Safety and Effectiveness Data page 12 of 20 16 {13} # Summary of Safety and Effectiveness Data (SSED) The numbers of valid and missing test results are presented in Table 7. For HER2 CISH pharmDx™ assay 13 results were not available and for PathVysion HER-2 DNA Probe Kit (FISH) 4 specimens were without a final test result. # Cross Tabulations of HER2 Status for All Valid Specimens For the comparison of HER2 status obtained by the HER2 CISH pharmDx™ Kit and the PathVysion HER-2 DNA Probe Kit (FISH) a total of 350 valid specimens were eligible For HER2 CISH staining, two (2) out of four (4) of the missing test results were due to 2 test failures with regard to staining quality. The last two missing CISH result were due to lack of patient sample to perform repeat test (see Table 8). Table 8: Case processing summary for HER2 CISH pharmDx™ Kit versus PathVysion FISH HER2 gene status – all cases. | | Cases | | | | | | | --- | --- | --- | --- | --- | --- | --- | | | Valid | | Missing | | Total | | | | N | Percent | N | Percent | N | Percent | | CISH status * PathVysion FISH status | 350 | 95.9% | 15 | 4.1% | 365 | 100.0% | For these 350 specimens, the HercepTest™ IHC scores distribution of amplified and non-amplified specimens as determined by HER2 CISH pharmDx™ Kit is shown in Table 9. Table 9: Cross tabulation of HercepTest™ IHC score and CISH HER2 gene status | | HercepTest IHC score | | | | Total | | | --- | --- | --- | --- | --- | --- | --- | | | | 0 | 1+ | 2+ | | 3+ | | CISH status | non-amplified | 98 | 108 | 95 | 5 | 306 | | | amplified | 0 | 1 | 17 | 26 | 44 | | Total | | 98 | 109 | 112 | 31 | 350 | By cross tabulation of the HER2 status an overall agreement at 97.7% and a Kappa value at 0.90 were found (Table 10). Positive agreement was 90.9% with lower and upper 95% confidence intervals based on the binomial distribution at 79.8% and 96.9%, respectively. Negative agreement was 98.7% with lower and upper 95% confidence intervals at 96.9% and 99.6%, respectively. McNemar’s test for a systematic bias between HER2 CISH pharmDx™ Kit and PathVysion HER-2 DNA Probe Kit (FISH) assays revealed a non-significant two-tailed p value at 1.00, showing that no bias between the assays was present. PMA P100024: FDA Summary of Safety and Effectiveness Data page 13 of 20 {14} Summary of Safety and Effectiveness Data (SSED) Table 10: Cross tabulation of CISH HER2 gene status versus PathVysion FISH HER2 gene status. | | PathVysion FISH status | | Total | | | --- | --- | --- | --- | --- | | | | non-amplified | | amplified | | CISH status | non-amplified | 302 | 4 | 306 | | | amplified | 4 | 40 | 44 | | Total | | 306 | 44 | 350 | Overall agreement: $(342/350 \times 100) = 97.7\%$ (CI95: $95.7\%$; $98.9\%$) Positive agreement: $(40/44 \times 100) = 90.9\%$ (CI95: $79.8\%$; $96.9\%$) Negative agreement: $(302/306 \times 100) = 98.7\%$ (CI95: $96.9\%$; $99.6\%$) ![img-0.jpeg](img-0.jpeg) Figure 1. Plot of HER2/CEN-17 ratios for Dako HER2 CISH pharmDx™ Kit and PathVysion HER-2 DNA Probe Kit (FISH) Linear regression of the logarithmically transformed HER2/CEN-17 ratios revealed a correlation coefficient at 0.90 and a slope of 0.71 From the cross tabulation of specimens analyzed with HER2 CISH pharmDx™ Kit and PathVysion HER-2 DNA Probe Kit (FISH), it is observed that eight specimens do not get the same HER2 gene status. Below, in Table 11, a short data summary is provided with the HercepTest™ IHC scores, HER2/CEN-17 ratios, and HER2 status for these specimens. These eight specimens represent HercepTest™ IHC scores of $2+$ or $3+$. Five of the specimens (1001, 1098, 1401, 1411, 2056) have CISH HER2/CEN-17 ratios close to or within the borderline area of 1.8 to 2.2 in which the obtained results should be interpreted with caution. Thorough analysis of the discrepant cases have lead to the inclusion of several precautions in the instruction for use of the product (see safety conclusion). PMA P100024: FDA Summary of Safety and Effectiveness Data {15} Summary of Safety and Effectiveness Data (SSED) Table 11: Specimens that disagree in CISH versus PathVysion FISH HER2 gene status. | | Specimen | HercepTest IHC score | CISH HER2/CEN-17 ratio | CISH status | PathVysion HER2/CEN-17 ratio | PathVysion FISH status | | --- | --- | --- | --- | --- | --- | --- | | 1 | DAKO 09-1001 | 2+ | 1.72 | non-amplified | 2.27 | amplified | | 2 | DAKO 09-1004 | 3+ | 1.21 | non-amplified | 4.02 | amplified | | 3 | DAKO 09-1020 | 3+ | 6.13 | amplified | 1.00 | non-amplified | | 4 | DAKO 09-1098 | 3+ | 1.74 | non-amplified | 3.10 | amplified | | 5 | DAKO 09-1401 | 2+ | 2.07 | amplified | 1.90 | non-amplified | | 6 | DAKO 09-1411 | 2+ | 2.09 | amplified | 1.41 | non-amplified | | 7 | DAKO 09-1460 | 2+ | 1.38 | non-amplified | 2.27 | amplified | | 8 | DAKO 09-2056 | 2+ | 2.44 | amplified | 1.88 | non-amplified | ## Cross Tabulations of HER2 Status for IHC 2+ Specimens Only In the current study a total of 112 specimens with a HercepTest™ IHC 2+ score were eligible for HER2 status comparison by the HER2 CISH pharmDx™ Kit and PathVysion HER-2 DNA Probe Kit (FISH). Cross tabulation of the HER2 status revealed an overall agreement for HER2 status at 95.5% (Table 12) and a Kappa value at 0.82. Positive agreement was 87.5% with lower and upper 95% confidence intervals based on the binomial distribution at 65.6% and 97.3%, respectively. Negative agreement was 96.9% with lower and upper 95% confidence intervals at 91.9% and 99.1%, respectively. The two-tailed confidence intervals are based on a binomial distribution. McNemar’s test for a systematic bias between CISH and PathVysion assays revealed a non-significant two-tailed p-value at 1.00, showing that no bias was present. Table 12: Cross tabulation of CISH HER2 gene status versus PathVysion FISH HER2 gene status – IHC 2+ specimens only. | | PathVysion FISH status | | Total | | --- | --- | --- | --- | | | non-amplified | amplified | | | CISH status | non-amplified | 93 | 2 | | | amplified | 3 | 14 | | Total | | 96 | 16 | Overall agreement: (107/112 x 100) = 95.5% (CI95: 90.5%; 98.3%) Positive agreement: (14/16 x 100) = 87.5% (CI95: 65.6%, 97.3%) Negative agreement: (93/96 x 100) = 96.9% (CI95: 91.9%, 99.1%) ## HER2 CISH Success Rate Of the 365 specimens enrolled, the HER2 CISH pharmDx™ test was not performed for one specimen. For a total of 364 initial CISH tests performed, a HER2/CEN-17 ratio could be obtained in 352 tests. Therefore, the CISH success rate was (352/364 x 100) %, i.e. 96.7%. PMA P100024: FDA Summary of Safety and Effectiveness Data page 15 of 20 {16} Summary of Safety and Effectiveness Data (SSED) # Conclusion on HER2 CISH pharmDx™ Kit and PathVysion HER-2 DNA Probe Kit (FISH) Comparison Study The clinical utility of HER2 CISH pharmDx™ Kit has been investigated using 365 invasive breast cancer specimens that were analyzed by HER2 CISH pharmDx™ Kit and PathVysion HER-2 DNA Probe Kit (FISH). Specimens were enrolled consecutively as they arrived at the US reference laboratory for HER2 status analysis and the study population was enriched by HercepTest™ IHC 2+ specimens to challenge testing close to the cut-off value. Consecutive sections from 350 breast cancer specimens were successfully analyzed by both HER2 CISH pharmDx™ Kit and PathVysion HER-2 DNA Probe Kit (FISH). The overall agreement between HER2 status obtained by HER2 CISH pharmDx™ Kit and PathVysion HER-2 DNA Probe Kit (FISH) for 350 breast cancer specimens was 97.7% (CI95 limits 95.7%; 98.9%), with a Kappa of 0.90 indicating almost perfect agreement. Positive agreement was 90.9% (CI95 limits 79.8%; 96.9%), whereas, negative agreement was 98.7% (CI95 limits 96.9%; 99.6%). There was no systematic bias between the two tests. The acceptance criteria set prior to the clinical study were based on the positive and negative HER2 status agreements including the lower bound confidence intervals for positive and negative agreements previously obtained in a comparison between HER2 FISH pharmDx™ Kit and PathVysion HER-2 DNA Probe Kit (FISH) assays. The agreements and confidence limits obtained are summarized together with the acceptance criteria in the table below. Table 13: Acceptance criteria defined prior to study initiation and the specific agreements obtained for the comparison between HER2 CISH pharmDx™ Kit and PathVysion HER-2 DNA Probe Kit (FISH). | | Positive agreement | | Negative agreement | | | --- | --- | --- | --- | --- | | | Positive Agreement | Lower 95% CI limit | Negative agreement | Lower 95% CI limit | | Acceptance criteria | 86% | 77% | 97% | 94% | | CISH versus PathVysion | 90.9% | 79.8% | 98.7% | 96.9% | The positive and negative agreements including the lower 95% confidence intervals for the comparison between HER2 CISH pharmDx™ Kit and PathVysion HER-2 DNA Probe Kit (FISH) met the acceptance criteria defined prior to the initiation of the study. These study results support the predictive claim for HER2 CISH pharmDx™ Kit indicated as an aid in the assessment of patients for whom Herceptin™ is being considered. Of the 364 specimens for which a CISH assay was initiated, a HER2/CEN-17 ratio could be obtained in 352 tests resulting in a CISH success rate of 96.7%. PMA P100024: FDA Summary of Safety and Effectiveness Data {17} Summary of Safety and Effectiveness Data (SSED) 2. Method Comparison Study - HER2 CISH pharmDx™ Kit and Dako HER2 FISH pharmDx™ (Prognostic value of HER2 CISH pharmDx™ Kit) In order to link the prognostic value of HER2 FISH pharmDx™ Kit to the HER2 CISH pharmDx™ Kit the concordance between test results obtained with these assays was investigated. The prognostic value of HER2 FISH pharmDx™ Kit was previously shown based on the data in the DBCG (Danish Breast Cancer Cooperative Group) study 89D that analyzed HER2 status using Dako HER2 FISH pharmDx™ Kit. In the DBCG 89D study test results from 649 patient specimens were available for multivariate analysis, 417 had a HER2/CEN-17 ratio below 2.0 (normal or non-amplified HER2 gene status) and 232 had a HER2/CEN-17 ratio above or equal to 2.0 (amplified HER2 gene status). The results showed that patients with HER2 amplification had more severe prognosis, as they were elder, had higher malignancy grade, greater tumor size, more tumor positive lymph nodes, more abnormal TOP2A status and higher HER2 IHC score. HER2 amplification had no significant prognostic value when tested in all patients (N=649) using the multivariate Cox proportional hazards model. However, HER2 amplification had an independent prognostic value for both recurrence free survival (RFS): HR=1.42 (1.08-1.85), p=0.011, and overall survival (OS): HR=1.40 (1.07-1.85), p=0.015, when testing in the sub-group of node-positive patients (n=423). Therefore, it was concluded that HER2 amplification had an independent prognostic value for both RFS and OS. The concordance between HER2 FISH pharmDx™ Kit and HER2 CISH pharmDx™ Kit assays was investigated in a clinical study comprising 365 invasive breast cancer specimens collected in chronological order. The first 304 specimens were selected regardless of the HercepTest™ score, and the next 61 specimens were selected based on a HercepTest™ IHC 2+ score in order to enrich the study population with IHC 2+ cases. A total of 348 specimens with a valid test result for both assays were available for the concordance analysis between HER2 CISH pharmDx™ Kit and the HER2 FISH pharmDx™ Kit. The cross tabulation of the HER2 status for the two Dako assays is shown in Table 14. PMA P100024: FDA Summary of Safety and Effectiveness Data {18} Summary of Safety and Effectiveness Data (SSED) Table 14. Cross tabulation HER2 status obtained using the HER2 CISH pharmDx™ Kit and the HER2 FISH pharmDx™ Kit. | DAKO CISH vs. FISH | FISH status | | Total | | | --- | --- | --- | --- | --- | | | | non-amplified | | amplified | | CISH status | non-amplified | 301 | 3 | 304 | | | amplified | 3 | 41 | 44 | | Total | | 304 | 44 | 348 | Overall agreement: (342/348 x 100) = 98.3% (CI95: 96.5%; 99.3%) Positive agreement: (41/44 x 100) = 93.2% (CI95: 82.9%; 98.0%) Negative agreement: (301/304 x 100) = 99.0% (CI95: 97.4%; 99.7%) The cross tabulation presented in Table 14 showed an overall agreement between the two assays of 98.3% (CI95: 96.5%; 99.3%). The positive and negative agreement were 93.2% (CI95: 82.9%; 98.0%) and 99.0% (CI95: 97.4%; 99.7%) respectively. Likewise, a Kappa coefficient of 0.92 underlined the high agreement between HER2 CISH pharmDx™ Kit and HER2 FISH pharmDx™ Kit assays. ## Conclusion on HER2 CISH pharmDx™ Kit and HER2 FISH pharmDx™ Kit Comparison A high overall agreement between the two Dako assays of 98.3% was obtained in this study, with positive and negative agreements at 93.2% and 99.0%, respectively. The acceptance criteria set prior to study initiation were identical to the criteria for the HER2 CISH pharmDx™ Kit and PathVysion HER-2 DNA Probe Kit (FISH) comparison. In Table 15 these acceptance criteria are listed together with the agreements obtained for the HER2 status comparison between HER2 CISH pharmDx™ Kit and HER2 FISH pharmDx™ Kit. Table 15. Acceptance criteria defined prior to study initiation for the HER2 status comparison between HER2 CISH pharmDx™ Kit and HER2 FISH pharmDx™ Kit. | | Positive agreement | Positive agreement lower 95% CI limit | Negative agreement | Negative agreement lower 95% CI limit | | --- | --- | --- | --- | --- | | Acceptance criteria | 86% | 77% | 97% | 94% | | CISH versus FISH | 93.2% | 82.9% | 99.0% | 97.4% | ## Conclusion on Studies to Demonstrate Prognostic Claim for HER2 CISH pharmDx™ Kit A large number of studies have confirmed the prognostic values of HER2 status in breast cancer, both with respect to overexpression of the protein and amplification of the gene. One of these studies is the DBCG 89D study where HER2 amplification measured by the HER2 FISH pharmDx™ Kit showed to have an independent prognostic value for both RFS and OS when testing in node positive breast cancer patients. In order to transfer the prognostic claim from the HER2 FISH pharmDx™ Kit (PMA P040005) to the HER2 CISH pharmDx™ Kit a concordance study demonstrating overall agreement at 98.3% between the two PMA P100024: FDA Summary of Safety and Effectiveness Data {19} # Summary of Safety and Effectiveness Data (SSED) pharmDx™ Kits has been performed. Acceptance criteria based on positive and negative agreements as well as the lower 95% confidence limits for these were met. Based on the result from the DBCG 89D study and the comparison between HER2 CISH pharmDx™ Kit and HER2 FISH pharmDx™ Kit it is justified that the prognostic claim of HER2 FISH pharmDx™ Kit is transferred to HER2 CISH pharmDx™ Kit. ## XI. PANEL MEETING RECOMMENDATION AND FDA’S POST-PANEL ACTION In accordance with the provisions of section 515(c)(2) of the act as amended by the Safe Medical Devices Act of 1990, this PMA was not referred to the Immunology Devices Panel, an FDA advisory committee, for review and recommendation because the information in the PMA substantially duplicates information previously reviewed by this panel. ## XII. CONCLUSIONS FROM PRECLINICAL AND CLINICAL STUDIES ### A. Safety Conclusions The adverse effects of this device are based on data collected in clinical study conducted to support PMA approval as described above. As a diagnostic test, HER2 CISH pharmDx™ assay involves testing on formalin-fixed, paraffin-embedded human breast cancer tissue sections. These tissue sections are routinely removed for breast cancer diagnosis. The test, therefore, presents no additional safety hazard to the patient being tested if the following precautions, based on the preclinical and clinical studies, are followed as outlined in the instructions for use: - If a specimen includes cluster of blue signals (amplification of CEN-17) that makes it difficult to count the red HER2 signals and thereby interpret the staining, it is recommended to note the score and the staining pattern for reference. The user then needs to refer to other test methods (e.g. FISH or IHC) to make a final conclusion on HER2 status. - Due to potential heterogeneity of the HER2 signal distribution, it is important to perform a thorough scanning of the complete CISH stained specimen to evaluate signal distribution before selecting the area for signal enumeration. - The appearance of “Giant cells” within the tumor area should be noted and interpreted with caution since these cells can potentially affect the status of the specimen. ### B. Effectiveness Conclusions Results of the non-clinical and clinical studies using HER2 CISH pharmDx™ Kit on formalin-fixed, paraffin-embedded human breast cancer tissue sections with varying levels of gene amplification demonstrated: PMA P100024: FDA Summary of Safety and Effectiveness Data page 19 of 20 23 {20} Summary of Safety and Effectiveness Data (SSED) HER2 status determined in the comparative clinical study using HER2 CISH pharmDx™ Kit showed an overall agreement to that obtained by PathVysion HER-2 DNA Probe Kit (FISH) at 98% (95% CI: 96%-99%), with positive and negative agreements at 91% and 99%, respectively. Likewise, determination of HER2 status when using HER2 CISH pharmDx™ Kit and HER2 FISH pharmDx™ Kit showed an overall agreement at 98.3% (95% CI: 96.5% - 99.3%) with positive and negative agreements at 93% and 99%, respectively. Based on the presentation of data in the DBCG 89D study using HER2 FISH pharmDx™ Kit the prognostic claim from the HER2 FISH pharmDx™ Kit can be transferred to the HER2 CISH pharmDx™ Kit. Analysis of the data from the clinical reproducibility study using variance component models revealed that the HER2 CISH pharmDx™ Kit demonstrated day-to-day, site-to-site reproducibility ≤ 12% CV for non-amplified (normal) and HER2 IHC 2+ (HercepTest™) specimens and ≤ 16% CV for amplified specimens. Analysis of data from the observer-to-observer reproducibility study showed a CV=12.4% for all specimens. This could be further divided into a CV=10.9% for non-amplified and 22.6% for amplified specimens defined by the consensus status. The success rate for HER2 CISH pharmDx™ Kit was determined to be 96.7% in the comparative clinical study. ## C. Overall Conclusions The data in this application support the reasonable assurance of safety and effectiveness of this device when used in accordance with the intended use. ## XIII. CDRH DECISION CDRH issued an approvable letter on November 30, 2011 The final conditions of approval cited in the approval order are described in the approval order. The applicant’s manufacturing facility was inspected and found to be in compliance with the device Quality Systems (QS) regulation (21 CFR 820) on August 11, 2011. ## XIV. APPROVAL SPECIFICATIONS Directions for use: See device labeling. Hazards to Health from Use of the Device: See Indications, Contraindications, Warnings, Precautions, and Adverse Events in the device labeling. Post-approval Requirements and Restrictions: See approval order. PMA P100024: FDA Summary of Safety and Effectiveness Data page 20 of 20
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...