THORATEC HEARTMATE II LEFT VENTRICULAR ASSIST SYSTEM

P060040S005 · ABBOTT MEDICAL · DSQ · Jan 20, 2010 · Cardiovascular

Device Facts

Record IDP060040S005
Device NameTHORATEC HEARTMATE II LEFT VENTRICULAR ASSIST SYSTEM
ApplicantABBOTT MEDICAL
Product CodeDSQ · Cardiovascular
Decision DateJan 20, 2010
DecisionAPPR
Device ClassClass 3
AttributesTherapeutic

Indications for Use

The HeartMate II LVAS is intended for use as a bridge to transplantation in cardiac transplant candidates at risk of imminent death from non-reversible left ventricular failure. The HeartMate II LVAS is also indicated for use in patients with New York Heart Association (NYHA) Class IIIB or IV end-stage left ventricular failure who have received optimal medical therapy for at least 45 of the last 60 days, and who are not candidates for cardiac transplantation. The HeartMate II LVAS is intended for use both inside and outside the hospital, or for transportation of Ventricular Assist Device (VAD) patients via ground ambulance, fixed-wing aircraft, or helicopter.

Device Story

HeartMate II LVAS is an implanted axial flow blood pump providing mechanical circulatory support. It consists of an internal pump connected via a percutaneous lead to an external microprocessor-based System Controller. The controller manages motor actuation, monitors system function, and serves as the primary user interface. Power is supplied by wearable rechargeable batteries or a stationary Power Module. Used in clinical and home settings, the device is operated by patients and monitored by physicians. It provides continuous blood flow to assist the failing left ventricle, improving patient survival, functional status, and quality of life. The device is specifically designed for smaller patients, including women and those with anatomic features precluding larger devices. Clinical output includes system status and alarms, which guide patient management and clinical decision-making regarding anticoagulation and device maintenance.

Clinical Evidence

Prospective, randomized, unblinded, non-inferiority pivotal study (n=200; 134 HeartMate II, 66 HeartMate XVE). Primary endpoint: composite of 2-year survival free of stroke (Modified Rankin > 3) or reoperation. HeartMate II achieved 46% success vs 11% for HeartMate XVE (p<0.0001). Secondary measures included functional status (6-minute walk, NYHA class) and quality of life (MLWHF, KCCQ), showing significant improvement in both groups. Adverse events were monitored; HeartMate II showed lower reoperation rates per patient-year compared to HeartMate XVE.

Technological Characteristics

Implanted axial flow blood pump; percutaneous lead; external microprocessor-based System Controller; wearable battery or Power Module power sources. Biocompatible materials; electrical safety and electromagnetic compatibility per international standards. Sterilization and packaging validated to international standards.

Indications for Use

Indicated for patients with NYHA Class IIIB or IV end-stage left ventricular failure, refractory to optimal medical therapy, who are not candidates for cardiac transplantation. Contraindicated for patients unable to tolerate anticoagulation therapy.

Regulatory Classification

Identification

Approved pma: P870072

Predicate Devices

Submission Summary (Full Text)

{0} # SUMMARY OF SAFETY AND EFFECTIVENESS DATA # I. GENERAL INFORMATION Device Generic Name: Ventricular bypass (assist) device (21CFR 870.3545) Device Trade Name: Thoratec HeartMate® II Left Ventricular Assist System (LVAS) Applicants Name and Address: Thoratec Corporation 6035 Stoneridge Drive Pleasanton, CA 94588 Date of Panel Recommendation: N/A Premarket Approval Application (PMA) Number: P060040/S005 Date of Notice of Approval to Applicant: January 20, 2010 The HeartMate II LVAS is currently indicated as a bridge to transplantation in cardiac transplant candidates at risk of imminent death from non-reversible left ventricular failure. The PMA for this device for that indication was approved on April 21, 2008. Thoratec Corporation submitted a PMA supplement to expand the Indications for Use of the HeartMate II LVAS to include patients who are not candidates for cardiac transplantation. This expanded indication for permanent support is called destination therapy (DT). The preclinical test results that apply to this device were presented in the original PMA application and are not repeated here. For information on the data that were used to support the original Indications for Use, the summary of safety and effectiveness data (SSED) from the previous application should be referenced. Written requests for copies can be obtained from the Dockets Management Branch (HFZ-305), Food and Drug Administration, 12420 Parklawn Drive, Rm. 1-23, Rockville, MD 20857, under Dockets 94M-0404 and 99M-1520, or via the Internet at http://www.fda.gov/cdrh/pmapage.html. PMA P060040/S5: FDA SUMMARY OF SAFETY AND EFFECTIVENESS DATA PAGE 1 6 {1} ## II. INDICATIONS FOR USE The HeartMate II LVAS is intended for use as a bridge to transplantation in cardiac transplant candidates at risk of imminent death from non-reversible left ventricular failure. The HeartMate II LVAS is also indicated for use in patients with New York Heart Association (NYHA) Class IIIB or IV end-stage left ventricular failure who have received optimal medical therapy for at least 45 of the last 60 days, and who are not candidates for cardiac transplantation. The HeartMate II LVAS is intended for use both inside and outside the hospital, or for transportation of Ventricular Assist Device (VAD) patients via ground ambulance, fixed-wing aircraft, or helicopter. ### III. CONTRAINDICATIONS The HeartMate II LVAS is contraindicated for patients who cannot tolerate anticoagulation therapy. ### IV. WARNINGS AND PRECAUTIONS See Warnings and Precautions in the final labeling (Instructions for Use) ### V. DEVICE DESCRIPTION The HeartMate II Left Ventricular Assist System (LVAS) consists of an implanted axial flow blood pump and external components as shown in Figure 1. The HeartMate II is smaller than the HeartMate XVE LVAS which is also approved for the DT indication. Because of its size, the HeartMate II LVAS can be used in the treatment of smaller sized non-cardiac transplant patients. These smaller sized patients include mostly women and men of small stature. It can also be used in patients with anatomic features that preclude use of the larger HeartMate XVE device. The HeartMate II LVAS is powered through the System Controller, a microprocessor-based unit that initiates motor actuation, monitors and reports on system function, and serves as the primary interface with the system. Electrical power is provided by either a pair of wearable, rechargeable batteries (Figure 1), or through connection to a dedicated power supply (Power Module) as shown in Figure 2. The electrical power to the implanted blood pump is delivered through a percutaneous lead that connects the blood pump to the external System Controller. PMA P060040/S5: FDA SUMMARY OF SAFETY AND EFFECTIVENESS DATA PAGE 2 1 {2} Figure 1 – HeartMate II LVAS, Implantable and External Components (Battery-powered Configuration) ![img-0.jpeg](img-0.jpeg) PMA P060040/S5: FDA SUMMARY OF SAFETY AND EFFECTIVENESS DATA PAGE 3 8 {3} Figure 2 – HeartMate II LVAS Configuration with Power Module and System Monitor ![img-1.jpeg](img-1.jpeg) # VI. ALTERNATIVE PRACTICES OR PROCEDURES The current standard of care for patients in end-stage heart failure includes three treatment modalities: pharmacologic therapy (including digoxin, ACE inhibitors/Angiotensin II receptor blockers, beta blockers, diuretics and inotropes), cardiac transplantation, and mechanical circulatory support devices. PMA P060040/S5: FDA SUMMARY OF SAFETY AND EFFECTIVENESS DATA PAGE 4 9 {4} # VII. MARKETING HISTORY The HeartMate II LVAS has met the conformity requirements of the European Union Active Implantable Medical Device Directive, as indicated by the CE Mark. Since November 7, 2005, the HeartMate II has been commercially distributed in the European Union, Switzerland, Iceland, Canada, Israel, Singapore, the Bahamas and South Africa. The HeartMate II LVAS was initially approved for marketing in the United States for the bridge to transplant indication on April 21, 2008. The HeartMate II LVAS has not been withdrawn from the market in any country. # VIII. POTENTIAL ADVERSE EVENTS Adverse events that may be associated with the use of the HeartMate II LVAS are listed below. Other than death, adverse events are listed in decreasing order of frequency observed in the clinical study. For additional information on adverse events that occurred in the destination therapy clinical study, please see Section X below. Death - Bleeding, perioperative or late Cardiac arrhythmia Local infection Respiratory failure Device malfunction Sepsis Right heart failure - Driveline or pocket infection Renal failure Stroke Neurologic dysfunction Psychiatric episode Thromboembolic event, peripheral - Hemolysis - Hepatic dysfunction Device thrombosis Myocardial infarction PMA P060040/S5: FDA SUMMARY OF SAFETY AND EFFECTIVENESS DATA PAGE 5 10 {5} # IX. SUMMARY OF PRE-CLINICAL STUDIES Thoratec conducted testing on the components and sub-systems of the HeartMate II LVAS. In vitro and in vivo system performance and characterization studies and long-term reliability studies demonstrated reasonable system safety of the HeartMate II LVAS. Pre-clinical testing demonstrated compliance with internationally recognized standards for electrical safety, electromagnetic compatibility, and biocompatibility. Packaging and sterilization processes were validated according to internationally recognized standards. Pre-clinical laboratory studies that were summarized in the Summary of Safety and Effectiveness for the original PMA (P060040) are equally applicable to the expanded Indications for Use. # X. SUMMARY OF CLINICAL STUDIES # Study Overview The objective of the study was to determine the safety and effectiveness of the HeartMate II LVAS as a destination therapy (DT) device in end-stage heart failure patients who were not candidates for cardiac transplantation. Effectiveness of the device was compared to the HeartMate XVE by evaluating a composite endpoint that included survival at two years free of stroke resulting in a Modified Rankin Score > 3 or reoperation to repair or replace the device. The safety of the HeartMate II was documented by the incidence of adverse events as well as device malfunctions and failures during LVAS support compared to the HeartMate XVE results. In addition, a number of secondary objectives were evaluated during the study, including separate evaluations of each component of the composite endpoint (2 year survival, stroke rates, and device reliability), functional status (6-minute walk, patient activity score, and NYHA classification), health status including quality of life (Minnesota Living with Heart Failure and Kansas City Cardiomyopathy Questionnaire), assessment of all adverse events, re-operations, re-hospitalizations, and neurocognitive assessments (memory, language, visual/spatial perception, processing speed and abstract/executive function). # Study Design The Destination Therapy (DT) pivotal study was a prospective, randomized, unblinded, non-inferiority evaluation of the HeartMate II LVAS in end-stage left ventricular failure patients who were not candidates for cardiac transplantation and were refractory to optimal medical therapy. The statistical analysis plan in the study protocol specified testing for superiority once non-inferiority was established. Patients were randomly assigned to treatment with the HeartMate XVE (control group) or to treatment with the HeartMate II. Two patients were randomized to the HeartMate II for every one patient randomized to the HeartMate XVE. The randomization was stratified by study center PMA P060040/S5: FDA SUMMARY OF SAFETY AND EFFECTIVENESS DATA PAGE 6 11 {6} and blocked to maintain the 2:1 ratio over time. Block sizes of 3, 6, or 9 patients were randomly selected to prevent manipulation of the treatment assignment. Two hundred patients were enrolled into the Primary Study Cohort (134 HeartMate II and 66 HeartMate XVE) at 38 investigational sites from March 2005 to May 2007. All 200 patients enrolled into the Primary Study Cohort were followed for at least two years. This study also enrolled patients into four additional study cohorts. Refer to Figure 3 for a summary of the cohorts and number of patients enrolled. - Small BSA Cohort: Patients who had a BSA of less than \(1.5\mathrm{m}^2\) and therefore, could not be randomized to the HeartMate XVE due to its size. - XVE Exchange Cohort: Destination Therapy patients who received the HeartMate II as a replacement for a failed HeartMate XVE originally implanted under commercial use. - Randomized Continued Access Protocol (CAP) Cohort: Upon completing enrollment in the Primary Study Cohort, patient enrollment continued under CAP, using the same study protocol as the Primary Study Cohort. - Anatomic Deviation Cohort: This cohort included patients that had a BSA \(>1.5\mathrm{m}^2\) but could not be randomized to the HeartMate XVE due to their body habitus or other anatomic considerations. Figure 3 – Total Number of Patients Enrolled in Each Cohort (as of December 15, 2008) ![img-2.jpeg](img-2.jpeg) PMA P060040/S5: FDA SUMMARY OF SAFETY AND EFFECTIVENESS DATA PAGE 7 12 {7} The study had two oversight committees: a Clinical Events Committee (CEC), which adjudicated all adverse events and deaths, and a Data and Safety Monitoring Board (DSMB), which reviewed the study data periodically to ensure that the study was safe to continue. The members of these committees were independent of Thoratec Corporation, the investigational sites, and the principal investigators. The primary endpoint of the study was a composite endpoint: two year survival free of stroke resulting in a Modified Rankin score > 3 or reoperation to repair or replace the device. Patients were considered a success if they achieved the composite endpoint and a failure if they did not. Patients who were urgently transplanted due to device failure were considered a failure. Patients who were electively transplanted after reversal of a pre-enrollment co-morbidity were followed and considered a success if they achieved two years of survival from the day of their VAD implant without experiencing a stroke resulting in a Modified Rankin score > 3. The HeartMate II was judged a success if the proportion of HeartMate II patients who achieved the composite endpoint was equal to or better than the HeartMate XVE comparison group. ## Primary Study Cohort Patient Population The HeartMate II was implanted as Destination Therapy in patients who were not candidates for cardiac transplantation. Patients were ineligible for transplant primarily due to age (28%), recent history of cancer (9%), obesity (7%), and substance abuse or insufficient social support (7%). The patients' ages ranged from 26 to 81 years, with a median of 64 years. The majority of patients were Caucasian males with ischemic heart disease. No significant differences were seen in age, BSA, body mass index (BMI), etiology or ethnicity between the HeartMate II and HeartMate XVE groups. Despite randomization, the HeartMate II arm included a significantly larger number of females than the HeartMate XVE arm. A gender analysis was performed and it was determined that there was no influence on the treatment effect (refer to the Gender Analysis Section at the end of this document). Patient demographics are presented in Table 1. Table 1 – Primary Study Cohort: Baseline Demographics | | HeartMate II (n=134) | HeartMate XVE (n=66) | P* | | --- | --- | --- | --- | | Age (years)** | 64 (26-79) | 65 (29-81) | 0.8125 | | Etiology | 66% Ischemic | 68% Ischemic | 0.7526 | | Gender | 81% Male 19% Female | 92% Male 8% Female | 0.0369 | | BSA (m2)** | 2.0 (1.6-2.8) | 2.0 (1.6-2.8) | 0.5438 | | BMI (kg/m2)** | 27.4 (18.0-43.4) | 27.9 (18.2-40.1) | 0.9913 | * Unpaired t-test or Fisher's exact test, as appropriate **Median and range PMA P060040/S5: FDA SUMMARY OF SAFETY AND EFFECTIVENESS DATA PAGE 8 13 {8} The baseline laboratory assessments, hemodynamic values and cardiovascular history did not reveal any statistically significant differences between the HeartMate II and HeartMate XVE group. Of note in the cardiovascular history is that 83% of the patients entered the study with implantable cardioverter defibrillators (ICD) and 16% of the patients had a history of stroke (refer to Table 2). Overall, 79% of the patients were on inotropes at baseline, 23% on intra-aortic balloon pump (IABP) and 8% on mechanical ventilation, thus indicating an end-stage heart failure patient population. The similarity in baseline characteristics indicates that the two treatment arms are comparable. Table 2 – Primary Study Cohort: Baseline Cardiovascular History | | HeartMate II (n=134) | | HeartMate XVE (n=66) | | | | --- | --- | --- | --- | --- | --- | | Cardiovascular | # Pts | % Pts | # Pts | % Pts | P* | | Arrhythmias | 115 | 86% | 55 | 83% | 0.6762 | | Ventricular Arrhythmias | 65 | 49% | 38 | 58% | 0.2337 | | Congenital Heart Disease | 1 | 1% | 1 | 2% | 0.5522 | | Coronary Artery Disease | 90 | 67% | 48 | 73% | 0.5159 | | Hypertension | 87 | 65% | 34 | 52% | 0.0903 | | Ischemic | 88 | 66% | 46 | 68% | 0.7526 | | Implantable Cardiac Defibrillator (ICD) | 111 | 83% | 52 | 79% | 0.5619 | | LV Aneurysm / Repair | 4 | 3% | 4 | 6% | 0.4433 | | Myocardial Infarction | 70 | 52% | 38 | 58% | 0.5468 | | Revascularization | 71 | 53% | 32 | 48% | 0.6550 | | Stroke | 21 | 16% | 11 | 17% | 0.8403 | | Valve Replacement / Repair | 14 | 10% | 7 | 11% | 1.0000 | | Valve Insufficiency | 103 | 77% | 46 | 70% | 0.3026 | | Ventricular Pacing | 105 | 78% | 45 | 68% | 0.1225 | | Biventricular Pacing | 85 | 63% | 39 | 59% | 0.6423 | *Fisher's exact test PMA P060040/S5: FDA SUMMARY OF SAFETY AND EFFECTIVENESS DATA PAGE 9 14 {9} # Primary Study Endpoint: The primary endpoint of this study was a composite endpoint: two year survival, free of stroke resulting in a Modified Rankin Score > 3 and free of reoperation to repair or replace the device. Of note, this endpoint combines the primary safety and effectiveness assessments for the device. Based on past experience with ventricular assist devices in this patient population, this type of endpoint does reflect the intended use of the device and represent a clinically meaningful outcome for patients, which are two critical factors in determining an appropriate primary endpoint. Complete follow-up was obtained for all 200 Primary Study Cohort patients. The results of the analysis demonstrated statistical superiority of the HeartMate II (p<0.0001). Forty-six percent (62/134) of the patients randomized into the HeartMate II cohort successfully achieved the composite endpoint, while only 11% (7/66) of patients randomized into the HeartMate XVE cohort achieved the composite endpoint. This analysis is displayed in Table 3. Table 3 – Primary Study Cohort (Intent To Treat): Final Analysis Results | Endpoint Analysis | HeartMate II (n=134) No. Pts (%) [95% CI] | HeartMate XVE (n=66) No. Pts (%) [95% CI] | P Value^{1} | | --- | --- | --- | --- | | Primary Composite Endpoint^{2} | 62 (46%) [38-55%] | 7 (11%) [3-18%] | 0.000000246 | | Components of Primary Composite Endpoint^{3} | | | | | 1) Stroke w/ Rankin score >3 | 15 (11%) [6-17%] | 8 (12%) [4-20%] | | | 2) Reoperation to repair or replace pump | 13 (10%) [5-15%] | 24 (36%) [25-48%] | | | 3) Death < 2 years | 44 (33%) [25-41%] | 27 (41%) [29-53%] | | | Total of Composite Events | 72 (54%) [45-62%] | 59 (89%) [82-97%] | | $^{1}$ Fisher exact test $^{2}$ Two-year survival free of stroke resulting in a Modified Rankin score > 3 or reoperation to repair or replace the device $^{3}$ Only the first event was counted (i.e., if a patient had a stroke then subsequently died, only the stroke would be counted) PMA P060040/S5: FDA SUMMARY OF SAFETY AND EFFECTIVENESS DATA PAGE 10 15 {10} # Composite Endpoint Component: Overall Survival (As Initially Treated) Subsequent to randomization, eight patients were not implanted and four patients were implanted with the alternate device. Therefore, 133 patients were initially implanted with the HeartMate II and fifty-nine (59) patients were initially implanted with the HeartMate XVE. Patients were followed to death or two years, whichever occurred first, regardless of whether they were transplanted, explanted, or had their device exchanged. Table 4 shows the survival at two years with a breakdown of the type of device present at two years if the device needed replacing. The protocol specified that the HeartMate XVE patients could have the HeartMate II implanted if their device needed replacing at the preference of the investigator. The HeartMate II patients could also crossover to the HeartMate XVE, but none did. Table 4 – Survival Status at 2 Years | Survival Status at 2 Years | Original Implanted Device | | | --- | --- | --- | | | HeartMate II (n=133) | HeartMate XVE (n=59) | | Ongoing on original implanted device | 50 (38%) | 0 (0%) | | Ongoing with replacement of original device with same type | 12 (9%) | 2 (3%) | | Ongoing with replacement of original device with alternate type | 0 (0%) | 14* (24%) | | Transplanted | 13 (10%) | 8 (19%) | | Explanted for recovery | 1 (1%) | 1 (2%) | | Total: | 76 (57%) | 25 (42%) | * Includes one patient who crossed-over to HeartMate II and was subsequently transplanted Figure 4 and Table 5 below present overall survival in an As Initially Treated Kaplan-Meier analysis, with no censoring of device cross-overs or transplants. As seen in Figure 4, the overall survival including device cross-overs and transplants favors the HeartMate II group when compared to the HeartMate XVE group. Competing outcomes (ongoing, transplant, death, etc.) over time for each of the two devices are shown in Figure 5 and Figure 6. Table 5 provides the details of the analysis, including the number of patients at each interval. Causes of death while on LVAS support are listed in Table 6. PMA P060040/S5: FDA SUMMARY OF SAFETY AND EFFECTIVENESS DATA PAGE 11 16 {11} Figure 4 – Primary Study Cohort (As Initially Treated): Kaplan-Meier Survival Including Device Cross-overs and Transplants ![img-3.jpeg](img-3.jpeg) Table 5 – Primary Study Cohort (As Initially Treated): Kaplan-Meier Survival Including Device Cross-overs and Transplants | HeartMate II | | | | | | | | | --- | --- | --- | --- | --- | --- | --- | --- | | | Time Interval (Months) | | | | | | | | | 0 - 1 | 1 - 3 | 3 - 6 | 6 - 9 | 9 - 12 | 12 - 18 | 18 - 24 | | Number of patients starting interval | 133 | 121 | 108 | 101 | 95 | 90 | 81 | | Number of patients who died this interval | 12 | 13 | 7 | 6 | 5 | 9 | 5 | | Number of cumulative patient deaths | 12 | 25 | 32 | 38 | 43 | 52 | 57 | | Number of patients censored in interval | 0 | 0 | 0 | 0 | 0 | 0 | 0 | | Number of cumulative censored patients | 0 | 0 | 0 | 0 | 0 | 0 | 0 | | Probability of surviving interval | 0.910 | 0.812 | 0.759 | 0.714 | 0.677 | 0.609 | 0.571 | | +/- 95% Confidence Limit at end of interval | 0.05 | 0.07 | 0.07 | 0.08 | 0.08 | 0.08 | 0.08 | | HeartMate XVE | | | | | | | | | | Time Interval (Months) | | | | | | | | | 0 - 1 | 1 - 3 | 3 - 6 | 6 - 9 | 9 - 12 | 12 - 18 | 18 - 24 | | Number of patients starting interval | 59 | 52 | 42 | 37 | 35 | 33 | 28 | | Number of patients who died this interval | 7 | 10 | 5 | 2 | 2 | 5 | 3 | | Number of cumulative patient deaths | 7 | 17 | 22 | 24 | 26 | 31 | 34 | | Number of patients censored in interval | 0 | 0 | 0 | 0 | 0 | 0 | 0 | | Number of cumulative censored patients | 0 | 0 | 0 | 0 | 0 | 0 | 0 | | Probability of surviving interval | 0.881 | 0.712 | 0.627 | 0.593 | 0.559 | 0.475 | 0.424 | | +/- 95% Confidence Limit at end of interval | 0.08 | 0.12 | 0.12 | 0.13 | 0.13 | 0.13 | 0.13 | PMA P060040/S5: FDA SUMMARY OF SAFETY AND EFFECTIVENESS DATA PAGE 12 17 {12} Figure 5 – Competing Outcomes for HeartMate II LVAS (As Treated) ![img-4.jpeg](img-4.jpeg) Figure 6 – Competing Outcomes for HeartMate XVE LVAS (As Treated) ![img-5.jpeg](img-5.jpeg) PMA P060040/S5: FDA SUMMARY OF SAFETY AND EFFECTIVENESS DATA PAGE 13 18 {13} Table 6 – Primary Study Cohort (As Treated): Causes of Death During LVAS Support | | HeartMate II (n=133) | | HeartMate XVE (n=59) | | | --- | --- | --- | --- | --- | | | # Pts | % Pts | # Pts | % Pts | | Bleeding | 4 | 3% | 1 | 2% | | Brain related: | | | | | | Air Embolism | 1 | 1% | 0 | 0% | | Anoxic Brain Injury | 0 | 0% | 1 | 2% | | Stroke | 13 | 10% | 11 | 19% | | Traumatic Brain Injury | 2 | 2% | 1 | 2% | | Cardiopulmonary: | | | | | | Cardiac Arrest | 4 | 3% | 1 | 2% | | Heart Failure | 1 | 1% | 1 | 2% | | Respiratory Failure | 4 | 3% | 2 | 3% | | Myocardial Infarction | 1 | 1% | 0 | 0% | | Right Heart Failure | 8 | 6% | 5 | 8% | | Device Malfunctions/Failure: | | | | | | Loss of Power | 5 | 4% | 0 | 0% | | Device Thrombosis | 2 | 2% | 0 | 0% | | Inflow Obstruction | 0 | 0% | 1 | 2% | | VAD Dysfunction/fail | 3 | 2% | 2 | 3% | | Infection | 5 | 4% | 6 | 10% | | Miscellaneous: | | | | | | Amyloidosis | 1 | 1% | 0 | 0% | | Cancer | 2 | 2% | 0 | 0% | | Ischemic Bowel | 0 | 0% | 1 | 2% | | Unknown | 5 | 4% | 0 | 0% | | Withdrawal of Support | 2 | 2% | 1 | 2% | | Multi-system Organ Failure | 3 | 2% | 4 | 7% | # Composite Endpoint Component: Stroke with Modified Rankin Score >3 (As Initially Treated Analysis) Figure 7 and Table 7 provide an As Initially Treated Kaplan-Meier analysis of survival free of stroke (Modified Rankin score > 3), with no censoring of patients at the time of any device cross-overs or transplants. Patients were followed to death or stroke, whichever occurred first, regardless of whether they were transplanted, explanted, or had their device exchanged. As can be seen in Figure 7, the overall stroke-free survival results favor the HeartMate II group when compared to the HeartMate XVE group. Patients initially treated with the HeartMate II have a two-year predicted survival free of stroke (Modified Rankin > 3) of 56% compared to 41% for the patients initially treated with the HeartMate XVE. Table 7 provides the details of the analysis including the number of patients at each time interval. PMA P060040/S5: FDA SUMMARY OF SAFETY AND EFFECTIVENESS DATA PAGE 14 19 {14} Figure 7 – Primary Study Cohort (As Initially Treated): Kaplan-Meier Survival Free of Stroke with Modified Rankin Score>3 ![img-6.jpeg](img-6.jpeg) Table 7 – Primary Study Cohort (As Initially Treated): Kaplan-Meier Survival Free of Stroke with Modified Rankin Score>3 | HeartMate II | | | | | | | | | --- | --- | --- | --- | --- | --- | --- | --- | | | Time Interval (Months) | | | | | | | | | 0 - 1 | 1 - 3 | 3 - 6 | 6 - 9 | 9 - 12 | 12 - 18 | 18 - 24 | | Number of patients starting interval | 133 | 119 | 107 | 99 | 93 | 88 | 78 | | Number of patients who had event this interval | 14 | 12 | 8 | 6 | 5 | 10 | 4 | | Number of cumulative patient events | 14 | 26 | 34 | 40 | 45 | 55 | 59 | | Number of patients censored in interval | 0 | 0 | 0 | 0 | 0 | 0 | 0 | | Number of cumulative censored patients | 0 | 0 | 0 | 0 | 0 | 0 | 0 | | Probability of surviving interval | 0.895 | 0.805 | 0.744 | 0.699 | 0.662 | 0.587 | 0.556 | | +/- 95% Confidence Limit at end of interval | 0.05 | 0.07 | 0.07 | 0.08 | 0.08 | 0.08 | 0.08 | | HeartMate XVE | | | | | | | | | | Time Interval (Months) | | | | | | | | | 0 - 1 | 1 - 3 | 3 - 6 | 6 - 9 | 9 - 12 | 12 - 18 | 18 - 24 | | Number of patients starting interval | 59 | 50 | 41 | 36 | 34 | 32 | 27 | | Number of patients who had event this interval | 9 | 9 | 5 | 2 | 2 | 5 | 3 | | Number of cumulative patient events | 9 | 18 | 23 | 25 | 27 | 32 | 35 | | Number of patients censored in interval | 0 | 0 | 0 | 0 | 0 | 0 | 0 | | Number of cumulative censored patients | 0 | 0 | 0 | 0 | 0 | 0 | 0 | | Probability of surviving interval | 0.848 | 0.695 | 0.610 | 0.576 | 0.542 | 0.458 | 0.407 | | +/- 95% Confidence Limit at end of interval | 0.09 | 0.12 | 0.12 | 0.13 | 0.13 | 0.13 | 0.13 | PMA P060040/S5: FDA SUMMARY OF SAFETY AND EFFECTIVENESS DATA PAGE 15 20 {15} # Composite Endpoint Component: Reoperation to Repair or Replace Pump (As Initially Treated Analysis) Figure 8 and Table 8 provide an As Initially Treated Kaplan-Meier analysis of survival free of reoperation to replace or repair the pump, with no censoring of patients at the time of any transplant. Patients were followed to death or pump replacement or repair, whichever occurred first, regardless of whether they were transplanted or explanted. As can be seen in Figure 8, the overall results for survival free of reoperation to repair or replace the pump favor the HeartMate II group when compared to the HeartMate XVE group. Patients initially treated with the HeartMate II have a two-year predicted survival free of pump replacement or repair of 49% compared to 15% for the patients initially treated with the HeartMate XVE. Table 8 provides the details of the analysis including the number of patients at each time interval. Figure 8 – Primary Study Cohort (As Initially Treated): Kaplan-Meier Survival Free of Reoperation to Repair or Replace the Device ![img-7.jpeg](img-7.jpeg) PMA P060040/S5: FDA SUMMARY OF SAFETY AND EFFECTIVENESS DATA PAGE 16 21 {16} Table 8 – Primary Study Cohort (As Initially Treated): Kaplan-Meier Survival Free of Reoperation to Repair or Replace the Device | HeartMate II | | | | | | | | | --- | --- | --- | --- | --- | --- | --- | --- | | | Time Interval (Months) | | | | | | | | | 0 - 1 | 1 - 3 | 3 - 6 | 6 - 9 | 9 - 12 | 12 - 18 | 18 - 24 | | Number of patients starting interval | 133 | 121 | 108 | 101 | 93 | 88 | 75 | | Number of patients with events this interval | 12 | 13 | 7 | 8 | 5 | 13 | 10 | | Number of cumulative patient events | 12 | 25 | 32 | 40 | 45 | 58 | 68 | | Number of patients censored in interval | 0 | 0 | 0 | 0 | 0 | 0 | 0 | | Number of cumulative censored patients | 0 | 0 | 0 | 0 | 0 | 0 | 0 | | Probability of surviving interval | 0.910 | 0.812 | 0.759 | 0.699 | 0.662 | 0.564 | 0.489 | | +/- 95% Confidence Limit at end of interval | 0.05 | 0.07 | 0.07 | 0.08 | 0.08 | 0.08 | 0.08 | | HeartMate XVE | | | | | | | | | | Time Interval (Months) | | | | | | | | | 0 - 1 | 1 - 3 | 3 - 6 | 6 - 9 | 9 - 12 | 12 - 18 | 18 - 24 | | Number of patients starting interval | 59 | 50 | 39 | 33 | 30 | 25 | 12 | | Number of patients with events this interval | 9 | 11 | 6 | 3 | 5 | 13 | 3 | | Number of cumulative patient events | 9 | 20 | 26 | 29 | 34 | 47 | 50 | | Number of patients censored in interval | 0 | 0 | 0 | 0 | 0 | 0 | 0 | | Number of cumulative censored patients | 0 | 0 | 0 | 0 | 0 | 0 | 0 | | Probability of surviving interval | 0.848 | 0.661 | 0.559 | 0.509 | 0.424 | 0.203 | 0.153 | | +/- 95% Confidence Limit at end of interval | 0.09 | 0.12 | 0.13 | 0.13 | 0.13 | 0.10 | 0.09 | # Safety: Adverse events The incidence of serious adverse events is presented in Table 9. Serious adverse events were defined as those that resulted in death, were life-threatening, resulted in permanent disability, required hospitalization, or prolonged a hospital stay. The study was not powered for a specific analysis of the adverse events. To take into account differences in patient support durations, adverse events were normalized to events per patient-year and analyzed using Poisson regression to obtain risk ratios with 95% confidence intervals. Table 12 presents the results of this As Treated analysis for serious adverse events. The rates of serious adverse events (as treated) over various time intervals are presented below. PMA P060040/S5: FDA SUMMARY OF SAFETY AND EFFECTIVENESS DATA PAGE 17 22 {17} Table 9 - Primary Study Cohort (As Treated): Serious Adverse Events | | HeartMate II (n=133) | | | HeartMate XVE (n=59) | | | | --- | --- | --- | --- | --- | --- | --- | | | # Pts | % Patients [95% Confidence interval] | # Events | # Pts | % Patients [95% Confidence interval] | # Events | | Bleeding | 102 | 77% [70 - 84%] | 278 | 41 | 69% [58 - 81%] | 70 | | Bleeding requiring surgery | 40 | 30% [22 - 38%] | 50 | 9 | 15% [6 - 24%] | 12 | | Stroke | 24 | 18% [12 - 25%] | 27 | 8 | 14% [5 - 22%] | 9 | | Peri-operative (≤POD2) | 3 | 2% [0 - 5%] | 3 | 1 | 2% [0 - 5%] | 1 | | Post-operative (>POD2) | 21 | 16% [10 - 22%] | 24 | 7 | 12% [4 - 20%] | 8 | | Peripheral TE | 14 | 11% [5 - 16%] | 21 | 6 | 10% [2 - 18%] | 6 | | Other Neurological* | 27 | 20% [13 - 27%] | 32 | 9 | 15% [6 - 24%] | 11 | | Psychological | 4 | 3% [0 - 6%] | 4 | 0 | 0% [0 - 0%] | 0 | | Local Infection | 40 | 30% [22 - 38%] | 60 | 19 | 32% [20 - 44%] | 30 | | Drive Line Infection | 39 | 29% [22 - 37%] | 75 | 14 | 24% [13 - 35%] | 22 | | Pocket Infection | 12 | 9% [4 - 14%] | 19 | 8 | 14% [5 - 22%] | 10 | | Sepsis | 48 | 36% [28 - 44%] | 80 | 26 | 44% [31 - 57%] | 45 | | Pump Housing | 1 | 1% [0 - 2%] | 1 | 2 | 3% [0 - 8%] | 2 | | Right Heart Failure | 31 | 23% [16 - 30%] | 34 | 19 | 32% [20 - 44%] | 22 | | Inotropes Only | 27 | 20% [13 - 27%] | 29 | 16 | 27% [16 - 38%] | 19 | | RVAD | 5 | 4% [1 - 7%] | 5 | 3 | 5% [0 - 11%] | 3 | | Renal Failure | 21 | 16% [10 - 22%] | 21 | 14 | 24% [13 - 35%] | 14 | | Hepatic Dysfunction | 3 | 2% [0 - 5%] | 3 | 0 | 0% [0 - 0%] | 0 | | Respiratory Failure | 47 | 35% [27 - 43%] | 61 | 24 | 41% [28 - 53%] | 33 | | Cardiac Arrhythmias | 62 | 47% [38 - 55%] | 110 | 21 | 36% [23 - 48%] | 30 | | Myocardial Infarction | 0 | 0% [0 - 0%] | 0 | 1 | 2% [0 - 5%] | 1 | *Includes transient ischemic attacks (TIA) and non-stroke neurological events PMA P060040/S5: FDA SUMMARY OF SAFETY AND EFFECTIVENESS DATA PAGE 18 23 {18} # Safety: Device Complications This section discusses three selected device complications that are inherent to mechanical circulatory support devices. These complications were device malfunctions, device thrombosis, and hemolysis. Ninety-three percent (93%, 25/27) of the HeartMate XVE malfunctions that resulted in adverse clinical effects were related to the implanted pump, primarily the result of wear-out of either the bearings or valve conduit (Table 16). In contrast, only 53% of the HeartMate II malfunctions with adverse clinical effects were related to the implanted pump. The balance of the malfunctions was primarily associated with the System Controller, an external component of the system that is designed for easy and rapid exchange in the event of a malfunction. The malfunctions of the HeartMate II pump were primarily related to the percutaneous lead that connects the implanted pump with the external System Controller. Repair procedures were developed that can extend the useful life of the external portion of the percutaneous lead; however, failures of the internal portion of the percutaneous lead require immediate pump replacement. The serious complications of device thrombosis and hemolysis each occurred in 4% (5 patients) of the HeartMate II patients. The HeartMate XVE patients did not experience any of these events. Table 10 lists serious adverse events that occurred because of the device while Table 11 illustrates the relative risk of those events. Three of the five (60%) hemolysis events were associated with device thrombosis. The other two hemolysis events resolved over time with no intervention. PMA P060040/S5: FDA SUMMARY OF SAFETY AND EFFECTIVENESS DATA PAGE 19 24 {19} Table 10 - Primary Study Cohort (As Treated): Device Related Serious Adverse Events | | HeartMate II (n=133) | | | HeartMate XVE (n=59) | | | | --- | --- | --- | --- | --- | --- | --- | | | # Pts | % Patients [95% Confidence interval] | # Events | # Pts | % Patients [95% Confidence interval] | # Events | | Device Thrombosis | 5 | 4% [1 - 7%] | 5 | 0 | 0% [0 - 0%] | 0 | | Hemolysis | 5 | 4% [1 - 7%] | 5 | 0 | 0% [0 - 0%] | 0 | | Confirmed Malfunctions | 30 | 23% [15 - 30%] | 39 | 12 | 20% [10 - 31%] | 13 | Table 11 - Primary Study Cohort (As Treated): Device Related Serious Adverse Events | | HeartMate II (n=133) 210 pt-years | HeartMate XVE (n=59) 41 pt-years | | | --- | --- | --- | --- | | Event | Events/pt-yr | Events/pt-yr | Risk Ratio | | Device Thrombosis | 0.02 | 0.00 | *** | | Hemolysis | 0.02 | 0.00 | *** | | Confirmed Malfunctions | 0.19 | 0.31 | 0.59 | *** Unable to calculate relative risk because of no occurrences in one group PMA P060040/S5: FDA SUMMARY OF SAFETY AND EFFECTIVENESS DATA PAGE 20 25 {20} Table 12 – Primary Study Cohort (As Treated): All Serious Adverse Events | | HM II (n=133) 210 pt-years | HM XVE (n=59) 41 pt-years | | | | --- | --- | --- | --- | --- | | Event | Events/pt-yr | Events/pt-yr | Risk Ratio | 95% Confidence Interval** | | Death | 0.31 | 0.92 | 0.34 | | | Bleeding | 1.32 | 1.69 | 0.78 | | | Bleeding requiring surgery | 0.24 | 0.29 | 0.81 | | | Stroke | 0.13 | 0.22 | 0.59 | | | Peri-operative (≤POD2) | 0.01 | 0.02 | 0.59 | | | Post-operative (>POD2) | 0.11 | 0.19 | 0.59 | | | Other Neurological* | 0.16 | 0.27 | 0.57 | | | Local Infection | 0.29 | 0.73 | 0.39 | | | Drive Line Infection | 0.36 | 0.53 | 0.67 | | | Pocket Infection | 0.09 | 0.24 | 0.37 | | | Pump Housing | 0.00 | 0.05 | 0.10 | | | Sepsis | 0.38 | 1.09 | 0.35 | | | Right Heart Failure | 0.16 | 0.53 | 0.30 | | | Inotropes Only | 0.14 | 0.46 | 0.30 | | | RVAD | 0.02 | 0.07 | 0.33 | | | Peripheral TE | 0.10 | 0.15 | 0.69 | | | Respiratory Failure | 0.29 | 0.80 | 0.36 | | | Cardiac Arrhythmias | 0.52 | 0.73 | 0.72 | | | Renal Failure | 0.10 | 0.34 | 0.30 | | | Hepatic Dysfunction | 0.01 | 0.00 | --- | | | Psychological | 0.02 | 0.00 | --- | | | Myocardial Infarction | 0.00 | 0.02 | --- | | | * Includes transient ischemic attacks (TIA) and non-stroke neurological events ** No adjustments made for multiplicity; no conclusions may be drawn regarding statistical significance *** Unable to calculate risk ratio because of no occurrence in one group | | | | 0.0 0.5 1.0 1.5 2.0 Favors HM II Favors HM XVE | PAGE 21 – THORATEC HEARTMATE® II LVAS 26 {21} Table 13 - Primary Study Cohort: Serious Adverse Event Rate per Patient-year by Time Interval (As Treated) | | Group | 0 - 30 days | 31 - 180 days | 181-365 days | 366-730 Days | | --- | --- | --- | --- | --- | --- | | Cumulative years support | HeartMate II | 10.30 | 43.10 | 44.4 | 70.0 | | | HeartMate XVE | 4.40 | 16.60 | 12.8 | 7.1 | | Adverse Event | | | | | | | Bleeding^{1} | HeartMate II | 12.33 | 0.86 | 1.01 | 0.81 | | | HeartMate XVE | 10.91 | 1.08 | 0.16 | 0.28 | | Stroke^{2} | HeartMate II | 0.49 | 0.09 | 0.16 | 0.13 | | | HeartMate XVE | 0.91 | 0.18 | 0.00 | 0.28 | | Other Neurological^{3} | HeartMate II | 0.78 | 0.23 | 0.00 | 0.16 | | | HeartMate XVE | 0.45 | 0.24 | 0.23 | 0.14 | | Local Infection | HeartMate II | 2.14 | 0.32 | 0.14 | 0.21 | | | HeartMate XVE | 3.18 | 0.66 | 0.23 | 0.28 | | Drive Line Infection | HeartMate II | 0.00 | 0.37 | 0.65 | 0.30 | | | HeartMate XVE | 0.23 | 0.60 | 0.63 | 0.42 | | Pocket Infection | HeartMate II | 0.39 | 0.14 | 0.09 | 0.06 | | | HeartMate XVE | 0.45 | 0.30 | 0.16 | 0.14 | | Pump Housing Infection | HeartMate II | 0.00 | 0.00 | 0.00 | 0.01 | | | HeartMate XVE | 0.00 | 0.12 | 0.00 | 0.00 | | Sepsis | HeartMate II | 1.65 | 0.37 | 0.29 | 0.40 | | | HeartMate XVE | 2.95 | 1.08 | 0.94 | 0.28 | | Right Heart Failure | HeartMate II | 2.33 | 0.07 | 0.02 | 0.04 | | | HeartMate XVE | 3.86 | 0.18 | 0.08 | 0.14 | | Peripheral TE | HeartMate II | 0.87 | 0.14 | 0.14 | 0.00 | | | HeartMate XVE | 0.91 | 0.00 | 0.16 | 0.00 | | Respiratory Failure | HeartMate II | 3.50 | 0.21 | 0.11 | 0.10 | | | HeartMate XVE | 1.82 | 0.48 | 0.23 | 0.00 | | Cardiac Arrhythmias | HeartMate II | 3.98 | 0.51 | 0.32 | 0.39 | | | HeartMate XVE | 4.09 | 0.48 | 0.16 | 0.14 | | Renal Failure | HeartMate II | 1.36 | 0.07 | 0.02 | 0.03 | | | HeartMate XVE | 2.27 | 0.18 | 0.08 | 0.00 | | Hepatic Dysfunction | HeartMate II | 0.19 | 0.00 | 0.00 | 0.00 | | | HeartMate XVE | 0.00 | 0.00 | 0.00 | 0.00 | | Device Thrombosis | HeartMate II | 0.10 | 0.02 | 0.02 | 0.03 | | | HeartMate XVE | 0.00 | 0.00 | 0.00 | 0.00 | | Hemolysis | HeartMate II | 0.19 | 0.00 | 0.02 | 0.03 | | | HeartMate XVE | 0.00 | 0.00 | 0.00 | 0.00 | | Psychological | HeartMate II | 0.19 | 0.02 | 0.00 | 0.01 | | | HeartMate XVE | 0.00 | 0.00 | 0.00 | 0.00 | | Myocardial Infarction | HeartMate II | 0.00 | 0.00 | 0.00 | 0.00 | | | HeartMate XVE | 0.23 | 0.00 | 0.00 | 0.00 | $^{1}$ HeartMate II: 80.52 events/pt-yr over 0-1 days; 6.78 events/pt-yr over 2-30 days HeartMate XVE: 100.00 events/pt-yr over 0-1 days; 3.92 events/pt-yr over 2-30 days $^{2}$ HeartMate II: 2.75 events/pt-yr over 0-1 days; 0.31 events/pt-yr over 2-30 days HeartMate XVE: 3.13 events/pt-yr over 0-1 days; 0.76 events/pt-yr over 2-30 days $^{3}$ Includes transient ischemic attacks (TIA) and non-stroke neurological events PAGE 22 - THORATEC HEARTMATE® II LVAS 27 {22} # Secondary Objectives: Secondary objectives were also assessed in this study, including the following: re-operations, clinical reliability, functional status, health status including quality of life, neurocognitive evaluation and post-explant follow-up. # Reoperations Reoperations included any return to the operating room for any reason following implant. Reoperations were device or patient related, such as driveline debridement or bleeding, and included routine operations, such as appendectomy and orthopedic procedures. As reflected in Table 14, the incidence of reoperations was similar between the groups. There was a higher percentage of patients in the HeartMate II group requiring reoperation. However, when normalized to events per patient-year, there is a lower rate of reoperations per patient-year in the HeartMate II group (risk ratio 0.53). As expected, the rate of reoperations is greatest during the first 30 days after LVAS implantation. Table 14 – Primary Study Cohort (As Treated): Reoperations | | # pts | # pts with reoperations | % pts with reoperations | # events | Reops/pt-yr | Risk Ratio | | --- | --- | --- | --- | --- | --- | --- | | HeartMate II | 133 | 106 | 80% | 325 | 1.55 | 0.53 | | HeartMate XVE | 59 | 43 | 73% | 120 | 2.91 | | Thirty-six (36) of the reoperations that occurred in the Primary Study Cohort were due to the need for pump replacements (21 HeartMate XVE and 15 HeartMate II). Of the 59 patients implanted with the HeartMate XVE, 20 patients received 21 pump exchanges. Of the 133 patients implanted with the HeartMate II, 14 patients received 15 pump exchanges. # Clinical Reliability Clinical reliability was evaluated by taking into account all HeartMate II study experience through January 20, 2009, regardless of study cohort. One hundred and one (101) reports of suspected malfunctions related to the implanted components of the HeartMate II LVAS were received from the 509 HeartMate II patients enrolled into the Destination Therapy clinical study. Eighty-five percent of those reports were related to the percutaneous lead. There were 28 reports of malfunctions that resulted in hemodynamic compromise, reoperations for pump replacement, pump explantation, or death, 27 of which were related to the percutaneous lead. As shown in Table 15, reliability of the current configuration of the percutaneous lead is improved compared to the overall reliability, as a result of design modifications to the external strain relief intended to reduce the most frequent failure modes. PMA P060040/S5: FDA SUMMARY OF SAFETY AND EFFECTIVENESS DATA PAGE 23 28 {23} Table 15 – HeartMate II Percutaneous Lead Reliability (All Cohorts) | Percutaneous Lead Configuration | Type of Malfunction | Reliability^{1} at: | | | | --- | --- | --- | --- | --- | | | | 1 yr | 2 yr | 3 yr | | All configurations^{2} | All malfunctions | 85% | 62% | 39% | | | Reoperation/Death | 96% | 87% | 75% | | Current Configuration^{3} | All malfunctions | 95% | 91% | 87% | | | Reoperation/Death | 97% | 95% | 92% | $^{1}$ Reliability estimates based on Weibull analysis $^{2}$ Both original external strain relief and current external strain relief design $^{3}$ External strain relief design at time of PMA approval The mean time to failure (defined as malfunctions resulting in hemodynamic compromise, reoperations for pump replacement, pump explantation, or death) was 1677 days (at the 80% confidence level). The 28 individual failures described above occurred between 36 and 1277 days of VAD support. The observed clinical reliability is less than the original reliability estimates based on in vitro testing because the in vitro test environment did not reflect all of the factors associated with the LVAD user environment. Twenty-five (25) clinical failures (defined as malfunctions or damage resulting in pump replacement, urgent transplant or death) occurred in the 59 HeartMate XVE patients in the Primary Study Cohort and 58 HeartMate XVE patients in the Randomized Continuous Access Protocol (CAP) cohort. In contrast to the HeartMate II in which failures were predominately related to one component, the percutaneous lead, there were several failure modes of the HeartMate XVE, as shown in Table 16 below. Table 16 – HeartMate XVE Failure Modes | Failure Mode | No. of Failures | | --- | --- | | Valve and/or conduit wear | 8 | | Bearing wear | 7 | | Low flow and/or pump stopped | 5 | | Other | 5 | | Total: | 25 | These 25 failures occurred over a range of 0 to 676 days. A Weibull analysis of the clinical reliability of the HeartMate XVE observed in this study is provided in Table 17. Table 17 – HeartMate XVE Clinical Reliability Calculations | Time (months) | Clinical Reliability* at 80% Confidence Level | | --- | --- | | 6 | 84% | | 12 | 69% | | 18 | 55% | *Weibull analysis parameters: beta=1.1515, eta=1023.9663, rho=0.9023 PMA P060040/S5: FDA SUMMARY OF SAFETY AND EFFECTIVENESS DATA PAGE 24 29 {24} Of the patients that were originally implanted with the HeartMate XVE, 18 were exchanged to the HeartMate II and three (3) patients were exchanged to another HeartMate XVE. Sixteen (16) of the 21 pump exchanges were due to: inflow or outflow valve malfunction (4), bearing wear (11), and infection (1). In the other 5 cases, pumps were exchanged in patients with clinical symptoms and diagnostic indicators (e.g. waveforms, vent filter analysis) suggestive of end of pump life or fluid ingress. Pumps were functional during explant analysis but fluid ingress was noted in three (3) of the five (5) and bearing wear was noted in the other two (2). All of the patients originally implanted with the HeartMate II that needed pump replacements were exchanged to another HeartMate II pump. The reasons for exchange included: suspected pump thrombus (2), suspected percutaneous lead wire breakage in external portion of driveline (7), percutaneous lead breakage at pump end (5), and outflow elbow disconnect (1). ## Functional Status, Quality of Life and Neurocognitive Measures The secondary objectives that were studied included functional evaluations based on NYHA class, six-minute walk, and Metabolic Equivalents scores (METs). Health Status including quality of life (QoL) was measured via the Minnesota Living with Heart Failure Questionnaire (MLWHF) and Kansas City Cardiomyopathy Questionnaire (KCCQ). A battery of neurocognitive evaluations was also performed. In summary, significant improvement was seen in both device groups in quality of life scores and in functional status over baseline and over time through 12 months, as can be seen in Figure 9 through Figure 13 below. After 12 months, there were too few HeartMate XVE patients to analyze. No significant difference in QoL was seen between the HeartMate II and HeartMate XVE groups. An additional measure of QoL was time spent out of hospital. Once implanted, HeartMate II patients spent 87% of their support time out of hospital compared to 69% for patients implanted with the HeartMate XVE (As Treated analysis). No decline in neurocognitive function was observed, and trends toward improvement over time were observed for some neurocognitive measures with both devices. The number of patients able to perform these tests of functional status and quality of life decreases at each time interval as patients expire, are transplanted, are weaned off the device, or are crossed over to the alternate device. Besides outcome-related reasons why testing could not be performed, reasons included: patient issue (e.g., patient too sick to perform the testing), management issue (e.g., scheduling or site staff oversight and testing not performed), patient refusal, or in some cases a reason was not provided. The tables following the graphs (Table 18 through Table 21) indicate the number and percentage of patients performing each test. In Figure 9 through Figure 13 the numbers of patients analyzed at each interval is shown above each bar. Error bars extending beyond the data bars in the Figures denote 95% confidence intervals or standard deviation, as noted in each Figure. PMA P060040/S5: FDA SUMMARY OF SAFETY AND EFFECTIVENESS DATA PAGE 25 30 {25} Figure 10 – Primary Study Cohort (As Treated): NYHA Class I or II over Time ![img-8.jpeg](img-8.jpeg) Table 18 - Primary Study Cohort (As Treated): NYHA Class I or II Compliance | | | Baseline | Month 1 | Month 3 | Month 6 | Month 12 | Month 18 | Month 24 | | --- | --- | --- | --- | --- | --- | --- | --- | --- | | # pts at interval | HeartMate II | 133 | 121 | 105 | 95 | 82 | 70 | 59 | | | HeartMate XVE | 59 | 52 | 39 | 32 | 19 | 5 | 2 | | # pts performing test | HeartMate II | 126 | 101 | 92 | 86 | 73 | 60 | 58 | | | HeartMate XVE | 57 | 47 | 38 | 30 | 18 | 5 | 2 | | % pt performing test | HeartMate II | 95% | 83% | 88% | 91% | 89% | 86% | 98% | | | HeartMate XVE | 97% | 90% | 97% | 94% | 95% | 100% | 100% | PMA P060040/S5: FDA SUMMARY OF SAFETY AND EFFECTIVENESS DATA PAGE 26 31 {26} Figure 11 – Primary Study Cohort (As Treated): Six Minute Walk – Meters Walked over Time ![img-9.jpeg](img-9.jpeg) Table 19 - Primary Study Cohort (As Treated): Six Minute Walk Test Compliance | | | Base-line | Month 1 | Month 3 | Month 6 | Month 12 | Month 18 | Month 24 | | --- | --- | --- | --- | --- | --- | --- | --- | --- | | # pts at interval | HeartMate II | 133 | 121 | 105 | 95 | 82 | 70 | 59 | | | HeartMate XVE | 59 | 52 | 39 | 32 | 19 | 5 | 2 | | # pts performing test | HeartMate II | 50 | 72 | 78 | 67 | 61 | 42 | 41 | | | HeartMate XVE | 19 | 30 | 29 | 21 | 12 | 3 | 2 | | % pt performing test | HeartMate II | 38% | 60% | 74% | 71% | 74% | 60% | 69% | | | HeartMate XVE | 32% | 59% | 74% | 66% | 63% | 60% | 100% | PMA P060040/S5: FDA SUMMARY OF SAFETY AND EFFECTIVENESS DATA PAGE 27 32 {27} Figure 12 – Primary Study Cohort (As Treated): MLWHF Scores over Time ![img-10.jpeg](img-10.jpeg) Table 20 - Primary Study Cohort (As Treated): MLWHF Testing Compliance | | | Baseline | Month 1 | Month 3 | Month 6 | Month 12 | Month 18 | Month 24 | | --- | --- | --- | --- | --- | --- | --- | --- | --- | | # pts at interval | HeartMate II | 133 | 121 | 105 | 95 | 82 | 70 | 59 | | | HeartMate XVE | 59 | 52 | 41 | 34 | 20 | 5 | 2 | | # pts performing test | HeartMate II | 116 | 96 | 89 | 86 | 75 | 61 | 48 | | | HeartMate XVE | 49 | 39 | 36 | 29 | 17 | 4 | 2 | | % pt performing test | HeartMate II | 87% | 79% | 85% | 91% | 91% | 87% | 81% | | | HeartMate XVE | 83% | 75% | 87% | 85% | 85% | 80% | 100% | PMA P060040/S5: FDA SUMMARY OF SAFETY AND EFFECTIVENESS DATA PAGE 28 33 {28} Figure 13 – Primary Study Cohort (As Treated): KCCQ Overall Summary Score over Time ![img-11.jpeg](img-11.jpeg) Figure 14 – Primary Study Cohort (As Treated): KCCQ Clinical Summary Score over Time ![img-12.jpeg](img-12.jpeg) PMA P060040/S5: FDA SUMMARY OF SAFETY AND EFFECTIVENESS DATA PAGE 29 34 {29} Table 21 - Primary Study Cohort (As Treated): KCCQ Testing Completion by Follow-up Intervals | | | Base-line | Month 1 | Month 3 | Month 6 | Month 12 | Month 18 | Month 24 | | --- | --- | --- | --- | --- | --- | --- | --- | --- | | # pts at interval | HeartMate II | 133 | 121 | 105 | 95 | 82 | 70 | 59 | | | HeartMate XVE | 59 | 52 | 40 | 33 | 19 | 5 | 2 | | # pts performing test | HeartMate II | 115 | 98 | 89 | 86 | 76 | 62 | 50 | | | HeartMate XVE | 47 | 39 | 36 | 29 | 16 | 4 | 2 | | % pt performing test | HeartMate II | 86% | 81% | 85% | 91% | 84% | 89% | 85% | | | HeartMate XVE | 80% | 75% | 90% | 88% | 90% | 80% | 100% | ## Gender Analysis An analysis was performed to determine if the treatment effect observed in the trial was influenced by the gender differences between treatment groups. In addition, Kaplan-Meier analyses were performed by gender to determine if the superior results obtained by patients supported with the HeartMate II were experienced by both males and females. In the Primary Study Cohort, despite randomizing patients into treatment arms, 16% of patients randomized into the HeartMate II cohort were female compared to 8% in the HeartMate XVE cohort. This difference in gender was statistically significant (P=0.0369). Logistic regression was used to determine that this gender difference did not influence the treatment effect seen in the trial. A Kaplan-Meier analysis of the study's primary composite endpoint, stratified by gender, was limited due to the small number of females enrolled in the trial. A post hoc analysis was conducted which included patients from other cohorts. Patients enrolled in the Primary Study Cohort were combined with patients randomized into the Continued Access Protocol (CAP) for the trial. This pooled cohort will be referred to as the Randomized Cohort. Sixteen percent (16%) of patients randomized into the HeartMate II group were female, compared to 10% in the HeartMate XVE Cohort. The difference in gender enrollment was no longer significant (P=0.2094). The Randomized Cohort provided 54 female patients to evaluate. An analysis of baseline demographics and etiology demonstrated that the groups remained comparable. A Kaplan-Meier analysis comparison of the Randomized Cohort males (222 HeartMate II vs. 111 HeartMate XVE) and females (41 HeartMate II vs. 13 HeartMate XVE), for the study's primary composite endpoint resulted in a significant survival advantage (p<0.0001) for HeartMate II patients irrespective of gender. This analysis provides evidence that gender differences have not influenced the outcome results observed in the trial and that the superior results of the HeartMate II are shared by both males and females. Adverse event rates between females who received a HeartMate II compared to females who received the HeartMate XVE were similar to overall study results for the Primary PMA P060040/S5: FDA SUMMARY OF SAFETY AND EFFECTIVENESS DATA PAGE 30 35 {30} Study Cohort, with differences in adverse event rates favoring the females implanted with the HeartMate II. The same result was also true for males implanted with the HeartMate II. A second post hoc analysis was performed to provide additional females by combining patients enrolled into the Small BSA Cohort (patients with BSA < 1.5m²) and the HeartMate II Anatomical Deviation Cohort (patients with BSA ≥ 1.5m², but who could not receive an HeartMate XVE due to body habitus or surgical issues) with the HeartMate II patients from the Randomized Cohort described above. This combined cohort included 286 males and 100 females, all supported with the HeartMate II. A Kaplan-Meier analysis of survival free of stroke (Rankin > 3) or reoperation to repair or replace the device resulted in no significant difference between males supported with the HeartMate II compared to females supported with the HeartMate II (p=0.2650). One noted observation is that males supported with the HeartMate II had better adverse event rates for local infection and peripheral thrombo-embolic events when compared to females supported with the HeartMate II (Table 22). However, the thrombo-embolic event rate was influenced by one female who had eight thrombo-embolic events in her lower extremities. In conclusion, the gender analysis shows that the benefits and risks of the HeartMate II are similar for males and females and that the observed treatment effect was not influenced by gender. Table 22 – Adverse Event Differences (As Treated), Males vs. Females (events/pt-year) | Adverse Event | Males (n=279) | Females (n=102) | | --- | --- | --- | | Local infection | 0.62 | 0.95 | | Peripheral thrombo-embolic events | 0.08 | 0.12 | PMA P060040/S5: FDA SUMMARY OF SAFETY AND EFFECTIVENESS DATA PAGE 31 36 {31} # XI. PANEL MEETING RECOMMENDATION In accordance with the provisions of section 515(c)(2) of the act as amended by the Safe Medical Devices Act of 1990, this PMA was not referred to the Circulatory System Devices Panel, an FDA advisory committee, for review and recommendation because the information in the PMA substantially duplicates information previously reviewed by this panel. # XII. CONCLUSIONS DRAWN FROM THE CLINICAL STUDY The sponsor conducted a multi-center randomized pivotal study comparing the safety and effectiveness of the HeartMate II LVAS to the HeartMate XVE for the destination therapy indication. The device is already approved as a bridge to transplant. The primary endpoint of the study was a composite of 2 year survival with freedom from debilitating stroke (Modified Rankin > 3) or reoperation to repair or replace the device. The composite endpoint analysis showed the HeartMate II to be superior to the control HeartMate XVE device. In addition, both intent to treat and per protocol analyses demonstrated a Kaplan Meier survival advantage with the HeartMate II compared to control. No safety or engineering problems were detected that suggested that the increased benefit seen with the HeartMate II device was accompanied by significantly increased risk compared to the HeartMate XVE control. Hence, a favorable risk-benefit profile has been established for the HeartMate II device. Approval is recommended with careful follow-up through a post-approval study. # XIII. CDRH DECISION CDRH issued an approval order on January 20, 2010. The final conditions of approval cited in the approval order are described below. You have agreed to implement a post-approval study that will enroll consecutive HeartMate II patients who give their consent for inclusion in the Interagency Registry for Mechanically Assisted Circulatory Support (INTERMACS) registry with an indication of Destination Therapy. The patients will be followed in the post-approval study until study outcome or 2 years, whichever occurs first. A detailed protocol, including but not limited to, patient characteristics at the time of implantation, incidence of adverse events (including definitions) while being supported by the device system, patient outcome(s), standardized anticoagulation protocol, quality of life assessment tools, functional status instruments, and the proposed reporting interval (e.g., 6 months) will be implemented. The study protocol will also collect data regarding sex, age, and race/ethnicity to determine if differences exist in the safety and effectiveness of the device in these populations as well as data regarding the relationship between bleeding, thrombosis, and anticoagulation in all patients. Furthermore, you have agreed to perform a second post-approval study to collect data regarding the relationship between bleeding, thrombosis, von Willebrand syndrome, and anticoagulation in LVAD patients. The post-approval study enrollment will begin immediately upon FDA approval of the HeartMate II® LVAS as a destination therapy device. Please note that FDA requests that you work PMA P060040/S5: FDA SUMMARY OF SAFETY AND EFFECTIVENESS DATA PAGE 32 37 {32} interactively with us to resolve the issues related to this second post-approval study as we believe some deficiencies from our approvable letter dated November 20, 2009 remain outstanding. The applicant's manufacturing facility was inspected and found to be in compliance with the device Quality System (QS) regulation (21 CFR 820). # XIV. APPROVAL SPECIFICATIONS Directions for Use: See Device Labeling (Instructions for Use) Hazards to Health from Use of the Device: See Indications, Contraindications, Warnings and Precautions, and Adverse Events in the labeling Post-approval Requirements and Restrictions: See Approval Order PMA P060040/S5: FDA SUMMARY OF SAFETY AND EFFECTIVENESS DATA PAGE 33 38
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...