RayOne EMV Toric

P060011S039 · Rayner Intraocular Lenses , Ltd. · HQL · Oct 9, 2025 · Ophthalmic

Device Facts

Record IDP060011S039
Device NameRayOne EMV Toric
ApplicantRayner Intraocular Lenses , Ltd.
Product CodeHQL · Ophthalmic
Decision DateOct 9, 2025
DecisionAPPR
Regulation21 CFR 886.3600
Device ClassClass 3
AttributesTherapeutic

Indications for Use

The RayOne EMV Toric (Model RAO210T) consists of the EMV Toric IOL (210T) preloaded within the RayOne injection system (RAO). The EMV Toric IOL is intended for primary implantation in the capsular bag of the eye for the visual correction of aphakia, in adult patients in whom a cataractous lens has been removed by phacoemulsification and providing reduction of residual refractive astigmatism in adult patients with greater than or equal to 1.00 diopter (D) of corneal astigmatism The RayOne injection system is used to fold and assist in inserting the IOL into the eye.

Device Story

RayOne EMV Toric is a single-piece, aspheric, toric IOL manufactured from Rayacryl (hydroxyethyl methacrylate/methyl methacrylate copolymer). The device is preloaded in a RayOne injection system for delivery through a micro-incision into the capsular bag during cataract surgery. The IOL features a 360° posterior enhanced square edge and closed C-loop haptics. The optic design includes positive spherical aberration controlled by pupil size to provide smooth transitions. The device is used by ophthalmologists in surgical settings to replace the crystalline lens, correct aphakia, and reduce astigmatism. The injector mechanically folds the lens for insertion. Clinical outcomes are assessed by surgeons via visual acuity and refractive measurements; the device provides improved uncorrected distance vision and reduced residual astigmatism compared to monofocal IOLs.

Clinical Evidence

Prospective, multi-center, randomized, active-controlled clinical study (IDE #G230072) of 274 subjects (118 Toric, 120 Monofocal). Primary endpoints at 6 months (Visit 4) met: mean residual manifest cylinder difference of -0.43 D (p<0.0001) vs. control; 93.1% of eyes had <10° axis misalignment; 99.0% rotational stability (≤5° rotation). Safety endpoints met: adverse event rates were not statistically higher than ISO 11979-7 SPE rates. 100% of Toric subjects achieved BCDVA of 0.30 logMAR or better.

Technological Characteristics

Material: Rayacryl (HEMA/MMA copolymer). Design: Aspheric, toric, single-piece, 6.0 mm optic, 12.5 mm overall diameter, closed C-loop haptics. Energy: Passive optical device. Connectivity: None. Sterilization: Moist heat. Software: None. Standards: ISO 11979-2 (optical), ISO 11979-3 (mechanical), ISO 11979-5 (biocompatibility).

Indications for Use

Indicated for primary implantation in the capsular bag of adult patients (22+ years) following phacoemulsification for visual correction of aphakia and reduction of residual refractive astigmatism (≥1.00 D corneal astigmatism).

Regulatory Classification

Identification

An intraocular lens is a device made of materials such as glass or plastic intended to be implanted to replace the natural lens of an eye.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0} # SUMMARY OF SAFETY AND EFFECTIVENESS DATA (SSED) ## I. GENERAL INFORMATION | Device Generic Name: | Intraocular Lens | | --- | --- | | Device Trade Name: | RayOne EMV Toric | | Model Number: | RAO210T | | Device Procode: | HQL, MJP | | Applicant's Name and Address: | Rayner Intraocular Lenses Limited The Ridley Innovation Centre 10 Dominion Way, Worthing, West Sussex BN14 8AQ, United Kingdom | Date(s) of Panel Recommendation: None Premarket Approval Application (PMA) Number: P060011/S039 Date of FDA Notice of Approval: TBD The C-flex™ intraocular lens (IOL; model 570C) is the parent device, approved in P060011 on May 3, 2007. RayOne EMV Toric is a modification to the approved parent device by introducing a toric surface to the lens to provide a reduction of residual refractive astigmatism in adult patients with greater than or equal to 1.00 diopters (D) of corneal astigmatism, compared to a monofocal IOL. The C-flex™ was approved with the following Indications for Use (IFU) statement: The Rayner C-flex™ intraocular lens is indicated for primary implantation for the visual correction of aphakia in adults in whom a cataractous lens has been removed by phacoemulsification. The lens is intended to be placed in the capsular bag. The SSED supporting the original indication is available on the CDRH website. ## II. INDICATIONS FOR USE The RayOne EMV Toric (Model RAO210T) consists of the EMV Toric IOL (210T) preloaded within the RayOne injection system (RAO). The EMV Toric IOL is intended for primary implantation in the capsular bag of the eye for the visual correction of aphakia, in adult patients in whom a cataractous ---PMA P060011/S039: FDA Summary of Safety and Effectiveness Data 1 of 18 Version: 24th September 2025 {1} lens has been removed by phacoemulsification and providing reduction of residual refractive astigmatism in adult patients with greater than or equal to 1.00 diopter (D) of corneal astigmatism The RayOne injection system is used to fold and assist in inserting the IOL into the eye. ### **III. CONTRAINDICATIONS** There are no known contraindications. ### **IV. WARNINGS AND PRECAUTIONS** The warnings and precautions can be found in the RayOne EMV Toric labeling. ---PMA P060011/S039: FDA Summary of Safety and Effectiveness Data 2 of 18 Version: 24th September 2025 {2} ## V. DEVICE DESCRIPTION Rayner intraocular lenses (IOLs) are single piece optical devices, manufactured from Rayacryl (hydroxyethyl methacrylate/methyl methacrylate copolymer with UV blocker). These devices are designed to be surgically implanted in the human eye as a replacement for the crystalline lens and are intended for placement in the capsular bag following phacoemulsification. Rayner IOLs are intended to provide adjustment to the dioptic power of the eye. Toric models are intended to provide adjustment to the astigmatism of the eye. The RayOne EMV Toric (RAO210T) has an aspheric posterior surface designed to control the spherical aberration of the IOL; this lens is designed to have positive spherical aberration and is intended to add to the positive spherical aberration of the cornea. This addition to positive spherical aberration is controlled with pupil size and smooth transitions to negative spherical aberration towards the edge of the optic to prevent excessive positive spherical aberration reducing image quality at larger pupil sizes. In addition, RayOne EMV Toric (Model RAO210T) is intended to provide adjustment to the astigmatism of the eye. The RayOne injection system is a Fully Preloaded system that delivers a 6 mm optic diameter lens into the eye through a micro incision. The RayOne injector is designed to mechanically fold a lens and insert it into the eye during routine cataract surgery. The physical characteristics of RayOne EMV Toric are shown in TABLE 1. TABLE 1 RAYONE EMV TORIC PHYSICAL CHARACTERISTICS | Characteristic | RayOne EMV Toric | | --- | --- | | Lens image | Front view of IOL and side view of IOLAll dimensions in mm![img-0.jpeg](img-0.jpeg) | | Injector image | ![img-1.jpeg](img-1.jpeg) | | Lens material | Rayacryl | PMA P060011/S039: FDA Summary of Safety and Effectiveness Data 3 of 18 Version: 24th September 2025 {3} | Characteristic | RayOne EMV Toric | | --- | --- | | Injector material | Polypropylene & TPE | | Design features | Aspheric, Toric, Monofocal capsular bag fixated | | Injection system | Fully preloaded | | Overall diameter (mm) | 12.5 ± 0.2 | | Optic diameter (mm) | 6.0 ± 0.1 | | Limb width (mm) | 0.38 ± 0.1 | | Limb thickness (mm) | 0.41 ± 0.15 | | Haptic style | Closed C-loop | | Square edge | 360° posterior Enhanced Square Edge | | Sphere power range (increments) | +7.75 to 23.875 D | | Spherical equivalence (increments) | +10 to 25 D (0.5 D) | | Cylinder power range (increments) | +1.5 to 4.5 D (0.75 D) | | Index of refraction | 1.46 | The effective corneal powers for each of the test lens plane cylindrical powers of the test IOLs are shown in **TABLE 2**. TABLE 2 CYLINDER POWER AND CORNEAL ASTIGMATISM CORRECTION RANGE | IOL Type | IOL Power at IOL Plane (D) | Cyl Power on Corneal Plane (D) | Specified Range (D) Lower | Specified Range (D) Upper | | --- | --- | --- | --- | --- | | T3 | 1.5 | 1.03 | 1.03 | 1.53 | | T4 | 2.25 | 1.54 | 1.54 | 2.04 | | T5 | 3 | 2.05 | 2.05 | 2.56 | | T6 | 3.75 | 2.57 | 2.57 | 3.07 | | T7 | 4.5 | 3.08 | 3.08 | 3.59 | ## VI. ALTERNATIVE PRACTICES AND PROCEDURES There are several other alternatives for restoring functional vision of the aphakic eye. Non-surgical options include special cataract glasses or contact lenses. Surgical options such as monofocal, multifocal, simultaneous vision or accommodative IOLs are also available. Each alternative has its own advantages and disadvantages. A patient should fully discuss these alternatives with his/her physician to select the method that best meets expectations and lifestyle. PMA P060011/S039: FDA Summary of Safety and Effectiveness Data 4 of 18 Version: 24th September 2025 {4} ## VII. MARKETING HISTORY The RayOne EMV Toric has not been marketed in the United States. This product has been available in the European Union since receiving CE Mark in September 2022. Since launch, there have been no recalls of RayOne EMV Toric (Model RAO210T). Additionally, there are no countries where the device has been removed from the market for any reason(s) related to the safety or effectiveness of the device. ## VIII. POTENTIAL ADVERSE EFFECTS OF THE DEVICE ON HEALTH Below is a list of the potential adverse effects (e.g., complications) associated with the use of the device: - Secondary glaucoma - IOL replacement or extraction - Precipitates - Reduced vision - Vitreous herniation - Excessive intraoperative vitreous loss - IOL decentration - Secondary membrane - Expulsive hemorrhage - IOL dislocation and subluxation - Retrolenticular membrane - Corneal edema - Endophthalmitis and panophthalmitis - Retinal detachment - Corneal dystrophy - Hemorrhage - Iris atrophy - Pupillary block - Cystoid macular edema - Severe ametropia and aniseikonia - Iridocyclitis and hyalitis - Deviation from target refraction - Fibrin reaction. Secondary surgical interventions include, but are not limited to, lens repositioning, lens replacement, vitreous aspiration or iridectomy for pupillary block, wound leak repair, and retinal detachment repair. PMA P060011/S039: FDA Summary of Safety and Effectiveness Data 5 of 18 Version: 24th September 2025 {5} For the specific adverse events that occurred in the clinical studies, please see Section X below. ## IX. SUMMARY OF NON-CLINICAL STUDIES The RayOne EMV Toric (Model 210T) is composed of identical materials to the currently approved family of hydrophilic Rayacryl intraocular lenses (P060011). Non-clinical studies were performed on the material parent device (P060011) instead of the final proposed device. These tests are as follows: physicochemical and biocompatibility testing, sterilization, packaging, shelf-life, and transport stability testing. ### Physicochemical Testing The results of physicochemical testing are shown in **TABLE 3**. PMA P060011/S039: FDA | Test | Purpose | Acceptance Criteria | Results | | --- | --- | --- | --- | | ISO 11979-5 Exhaustive Extraction | Soxhlet extraction to recover polymerization residuals, impurities, and additives, quantitative analysis of extracts | No prespecified acceptance criteria. The results shall be evaluated to assess the risk for potentially harmful effects due to extractable components. | Performed | | ISO 11979-5 Leachable study | The IOL material was extracted at 35°C for 72 h with water and hexane, respectively. The extracts were analysed for potential leachables. | No prespecified acceptable criteria The results shall be evaluated to assess the risk for potentially harmful effects due to leachable components. | Performed | | ISO 11979-5 Test for Hydrolytic Stability | Test to verify material does not degrade by hydrolysis | Hydrolytic stability of IOLs by investigation of dioptic power, recording of transmission spectra, light microscopic inspection and scanning electron microscopic (SEM) inspection. The exposure medium shall be qualitatively and quantitatively analysed for any chemical entities at the end of the exposure period. | Pass | | ISO 11979-5 Photostability test | Exposure to UV radiation in a saline solution using a radiation source (Xenon arc lamp) in order to simulate an in vivo exposure time of 20 years to evaluate the photostability. | The investigated lens material stable against ultraviolet/visible light exposure. No significant change shall be detected between the physical appearance, UV/Vis spectra, dioptic power and image quality of the test | Pass | ---PMA P060011/S039: FDA Summary of Safety and Effectiveness Data 6 of 18 Version: 24th September 2025 {6} | Test | Purpose | Acceptance Criteria | Results | | --- | --- | --- | --- | | | | material exposed to UV radiation and controls receiving no radiation. The exposure medium shall be qualitatively and quantitatively analysed for any chemical entities after irradiation and compared to non-irradiated controls. | | | ISO 11979-5 Nd-YAG laser exposure test | To confirm IOL sample's stability when exposed to Nd:YAG lasers, and leakage of cytotoxic compounds. | No significant difference in physical appearance, spectral transmittance, or dioptic power after the exposure. No cytotoxic substances after laser exposure were detected. The exposure medium shall be qualitatively and quantitatively analysed for any chemical entities after laser exposure. | Pass | | ISO 11979-5 Evaluation of insoluble inorganics | To assess the IOL material for the presence of residual insoluble inorganics on and in the lens arising from manufacturing materials and process aids. | No prespecified acceptance criteria. Quantitative determination of the inorganic ions presents in the test material. | Performed | ### Biocompatibility Testing The RayOne EMV Toric (Model RAO210T) is made of the same material that was used with previously approved hydrophilic Rayacryl intraocular lenses (P060011). Biocompatibility testing (**TABLE 4 BIOCOMPATIBILITY TESTING**) was performed to support P060011 in accordance with all relevant ISO Standards (ISO 11979-5 and ISO 10993-1), as well as the United States Food and Drug Administration. Use of International Standards ISO 10993-1, "Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process". All tests to evaluate the biocompatibility were conducted in accordance with provisions of 21 CFR 58, Good Laboratory Practice (GLP) for Nonclinical Laboratory Studies. TABLE 4 BIOCOMPATIBILITY TESTING OF THE EMV TORIC (MODEL RAO210T) | Test | Purpose | Acceptance Criteria | Result | | --- | --- | --- | --- | | **Intraocular Lens*** | | | | | Cytotoxicity - MEM elution (ISO 10993-5, ISO 11979-5) | To evaluate the potential for cellular toxicity. | Non-cytotoxic | Pass | PMA P060011/S039: FDA Summary of Safety and Effectiveness Data 7 of 18 Version: 24th September 2025 {7} | Test | Purpose | Acceptance Criteria | Result | | --- | --- | --- | --- | | Cytotoxicity - Inhibition of Cell Growth Test (ISO 10993-5) | To evaluate the potential for cellular toxicity. | Non-cytotoxic | Pass | | Cytotoxicity – Agarose Overlay (ISO 10993-5, ISO 11979-5) | To evaluate the potential for cellular toxicity. | Non-cytotoxic | Pass | | Sensitization - Guinea Pig Maximization (ISO 10993-10, ISO 11979-5) | To assess dermal contact sensitization in guinea pig at 24 and 48 hours after patch removal. | Non-sensitizer | Pass | | Sensitization - Magnusson-Kligman Maximization Test (ISO 10993-11) | To investigate delayed contact hypersensitivity of the IOL in guinea pigs. | Non-sensitizer | Pass | | Ocular Irritation Study in Rabbits (ISO 10993-10) | To evaluate the potential for ocular reactions at 1, 24, 48 and 72 hours after the single exposure in rabbits. | Non-irritant | Pass | | Acute Toxicity (Limit-Test) (ISO 10993-11) | To evaluate the potential for acute toxicity in rats. | Non-toxic | Pass | | Genotoxicity - Mouse Peripheral Blood Micronucleus Study (ISO 10993-3, ISO 11979-5) | To evaluate the potential to produce cytogenetic damage, resulting in micronuclei formation. | No micronuclei formation | Pass | | Genotoxicity - In vitro chromosome aberration test (ISO 10993-3, ISO 11979-5) | To determine genotoxic potential in Chinese hamster ovary. | Non-genotoxic | Pass | | Genotoxicity - Mouse Lymphoma Assay (ISO 10993-3, ISO 11979-5) | To detect potential gene mutations or chromosomal damage. | Non-genotoxic | Pass | | Genotoxicity – Mutagenic activity (ISO 10993-3, ISO 11979-5) | To evaluate mutagenic activity. | Non-mutagenic | Pass | Note: In vivo 'Test for Local Effects' (non-ocular) and 'Ocular Implantation' test is deemed unnecessary for RayOne Family of hydrophilic IOL devices, considering that they are manufactured from a raw material with extensive historical clinical use. (Based on the recommendations of ISO 10993-1; and as per ISO 11979-5:2020, the risks arising from the use of the material are deemed acceptable based on the information from previous clinical use and other relevant literature.) Biocompatibility testing (TABLE 5) was performed on the patient-contacting components of the RayOne injection system. PMA P060011/S039: FDA Summary of Safety and Effectiveness Data 8 of 18 Version: 24th September 2025 {8} TABLE 5 BIOCOMPATIBILITY TESTING OF RAYONE INJECTION SYSTEM | Test | Purpose | Acceptance Criteria | Result | | --- | --- | --- | --- | | Cytotoxicity - Growth Inhibition Test in L929 Mouse Fibroblasts (Elution Test) (ISO 10993-5) | To evaluate the potential for cytotoxic effects using an in vitro mammalian cell culture test. | Non-cytotoxic | Pass | | Systemic Toxicity – Acute Systemic Toxicity (in Swiss Albino Mice) (ISO 10993-11) | To evaluate toxic characteristics. | Non-toxic | Pass | | Sensitization - Guinea Pig Maximization Sensitization Test (ISO 10993-10) | To evaluate the potential for sensitization. | Non-sensitizer | Pass | | Sensitization - Test for delayed-type hypersensitivity (Guinea Pig Maximisation Test) (ISO 10993-10) | To evaluate the potential for sensitization. | Non-sensitizer | Pass | | Intracutaneous Reactivity Testing in Rabbits (ISO 10993-10) | To evaluate the potential for irritation. | Non-irritant | Pass | | Intraocular Irritation Study in Rabbits (ISO 10993-10) | To evaluate the potential for irritation. | Non-irritant | Pass | ### Optical and Mechanical Testing The optical properties of the Model 210T were tested per “ISO 11979-2: Ophthalmic implants – Intraocular lenses: Optical properties and test methods and ANSI Z80.30: Toric intraocular lenses.” The mechanical properties were tested per “ISO 11979-3: Ophthalmic implants – Intraocular lenses: mechanical properties and test methods.” In cases where the mechanical properties were related to the Rayacryl® material and overall IOL design (e.g. haptic design, optic body diameter), the testing is being leveraged from the parent device the Model 600C for the Model 210T, since both IOL models are manufactured from the same Rayacryl® material and are identical in non-optical design features (i.e. only differ by anterior and posterior optical surface design). Testing was performed on lenses representing the low, the medium and the high diopter ranges: (spherical equivalent power (SE) = +10.0 D with cylinder = 0.75 D; SE = +20.0 D with cylinder = 0.75 D; SE = +25.0 D with cylinder = 0.75; SE = + 10.0 with cylinder 4.50; SE = + 20.0 with cylinder 4.50 D and SE = + 25.0 D with cylinder = 4.50 D.). The results of the testing demonstrate that the Model 210T meets the ISO and ANSI standards for optical and mechanical properties. The results of the testing demonstrate that the Model 210T meets all of the ISO and ANSI standards for optical and mechanical properties, as summarized in TABLE 6 and TABLE 7. PMA P060011/S039: FDA Summary of Safety and Effectiveness Data 9 of 18 Version: 24th September 2025 {9} TABLE 6 SUMMARY OF OPTICAL TESTING FOR RAYONE EMV TORIC – ISO 11979-2 & ANSI Z80.30 | Test | Purpose | Acceptance Criteria | Results | | --- | --- | --- | --- | | BS EN ISO 11979-2 Clauses4.2.2 - Dioptric power for toric IOL (TIOL) – Average Power (SE) 4.2.2 - Dioptric power for toric IOL (TIOL) – Cylinder Power | To assess conformance to optical power tolerances (post injection*) | +10.0 ≤ SE ≤ +15.0 D: ± 0.3 D +15.0 < SE ≤ +25.0 D: ± 0.4 D SE < 25 D and 0 < C ≤ 2,5 :±0,3 ±0,4 D SE < 25 D and 2,5 < C ≤ 4,5: ±0,4 D | Passed | | BS EN ISO 11979-2 Clause 4.3.3 - Determination of imaging quality - Toric IOL (TIOL) | To assess conformance to image quality tolerances (post injection*) | MTF (modulation transfer function) measurement >0.43 100% Pass | Passed | | BS EN ISO 11979-2 Clause 4.2.2 – Axis marks for toric IOL (TIOL) | To determine axis marks align with the meridian of lowest dioptric power. | ≤ 5° | Passed | | BS EN ISO 11979-2 Spectral transmittance | To characterize the spectral transmission in the range 300 nm to 1100 nm, and the 10% cutoff wavelength | ≥380nm | Passed | TABLE 7 SUMMARY OF MECHANICAL TESTING FOR RAYONE EMV TORIC - ISO 11979-3 | Mechanical Test | Purpose | Acceptance Criteria | Results | | --- | --- | --- | --- | | BS EN ISO 11979-3:2012 Clause 4.4 Compression force | To characterize the mechanical force exerted by the IOL haptics under compression initially | BS EN ISO 11979-3:2012: Measure and Report | Passed | | BS EN ISO 11979-3:2012 Clause 4.5 Axial displacement in compression | To characterize the axial displacement of the IOL under compression | BS EN ISO 11979-3:2012: Measure and Report | Passed | PMA P060011/S039: FDA Summary of Safety and Effectiveness Data 10 of 18 Version: 24th September 2025 {10} | Mechanical Test | Purpose | Acceptance Criteria | Results | | --- | --- | --- | --- | | BS EN ISO 11979-3:2012 Clause 4.6 Optic Decentration | To assess optic decentration under compression | $$(\bar{X} + 2\sigma) < 10\%$$ Clear Optic Diameter | Passed | | BS EN ISO 11979-3:2012 Clause 4.7 Optic Tilt | To assess optic tilt under compression | $$(\bar{X} + 2\sigma) < 5^\circ$$ | Passed | | BS EN ISO 11979-3:2012 Clause 4.8 Angle of Contact | To characterise angle of contact of haptics under compression | Measure and Report | Passed | | BS EN ISO 11979-3:2012 Clause 4.9 Compression Force Decay | To characterize the mechanical force exerted by the IOL haptics under compression after 24 hours | Measure and Report | Passed | | BS EN ISO 11979-3:2012 Clause 4.10 Dynamic Fatigue Durability | To assess the capability of the haptics to cyclic compressive displacement | Withstand 250,000 cycles without damage | Passed | | BS EN ISO 11979-3:2012 Clause 4.11 Surgical Manipulation | To assess the force required to separate the haptic from the IOL body | Force >0.25N | Passed | | BS EN ISO 11979-3:2012 Clauses 4.12 and 5(a) Surface and Bulk Homogeneity * IOL injections of the pre-loaded IOL through RayOne injector performed with OphteisBio 3.0%, a sodium hyaluronate OVD. | To ensure the IOL is free of surface and bulk defects | The IOL shall be essentially free from defects | Passed | | BS EN ISO 11979-3:2012 Clause 5 (b) Overall Diameter and Sagitta | To assess conformance to dimensional tolerances | Overall Diameter ±0.20 mm Sagitta ± 0.35 mm | Passed | *Each 210T is optically tested and inspected for Surface and Bulk Homogeneity prior to packaging and loading into the delivery system as part of standard manufacturing quality control. These measurements confirm compliance with ISO 11979-2 requirements for Dioptric Power and Imaging Quality, and compliance with ISO 11979-3 requirements for Surface and bulk Homogeneity Pre-Injection. Post-injection testing was performed to confirm compliance with ISO 11979-2 and ISO 11979-3 requirements, to verify the integrity of the delivery system and ensure clinical performance. PMA P060011/S039: FDA Summary of Safety and Effectiveness Data 11 of 18 Version: 24th September 2025 {11} ## Injector Validation Injector validation was conducted as part of the mechanical and optical testing of the EMV Toric, including the visual inspection of the IOL and injector pre-and post-injection. The test results summarized above reflect the compatibility of the RayOne injector and EMV Toric IOL. The testing demonstrated that the EMV Toric IOL can be successfully delivered across the entire range of IOLs using OphteisBio 3.0% viscoelastics. ## Sterilization and Shelf-Life Evaluation The RayOne EMV Toric is moist heat sterilized. Sterilisation evaluation, shelf-life and transport stability testing, and bacterial endotoxin testing were performed to support the RayOne EMV Toric. The results are summarized in TABLE 8. TABLE 8 SHELF-LIFE TESTING | Test | Purpose | Acceptance Criteria | Results | | --- | --- | --- | --- | | Moist heat sterilisation validation (Overkill method) (ISO 17665-1, ISO 17665-2 & ISO 17665-3) | Determination of process parameters for routine sterilisation of product, and demonstration that these process parameters can adequately achieve a Sterility Assurance Level (SAL) of 1.0E-6, based upon inactivation of control biological indicators (BIs, G. stearothermophilus) in process challenge devices (PCDs). Validated process parameters will be used for parametric release upon confirmation that Independent Recorder and controller meet batch release requirements. | Successful performance of Installation Qualification (IQ), Operational Qualification (OQ) and both Microbiological and Physical Performance Qualification (PQ) studies. The Microbiological PQ study requires total inactivation of BIs upon exposure to a half-cycle (e.g., half of the routine full-cycle exposure time) that meets the designated parameters. The Physical PQ study requires that routine full-cycle parameters can be achieved with minimum and maximum sterilisation loads. Together these PQ results will demonstrate the routine sterilisation process can achieve an SAL ≤ 1.0E-6. | Passed | PMA P060011/S039: FDA Summary of Safety and Effectiveness Data 12 of 18 Version: 10th September 2025 {12} | Test | Purpose | Acceptance Criteria | Results | | --- | --- | --- | --- | | Transport stability (ISO 11979-6, ISO 11607-1, ASTM F1886, ASTM D4991, ASTM F1929, ASTM F88, & ASTM F1140) | Confirm that both the inner tray and outer pouch sterile barrier packaging can maintain device sterility throughout anticipated transport conditions. Testing includes visual inspection, whole package integrity (dye immersion – weight change, validated alternative method) and seal integrity assessments (seal peel and dye penetration testing for inner tray; burst test and dye penetration for outer pouch) after transport conditioning. | Meets sterile barrier packaging visual inspection, whole package integrity & seal integrity requirements. For visual inspection, no channels, voids, punctures, or breaches observed, and labels are legible. For dye immersion, - weight change, there is no reduction in weight of the inner tray after it is held under vacuum for 30 minutes in vacuum chamber. For seal peel strength, average peel strength exceeds minimum force. For burst test, internal pressure exceeds minimum burst pressure before package failure. For dye penetration, no penetration of dye into seal area is observed. | Passed | | Shelf-life (ISO 11979-6, ISO 11607-1, ASTM F1886, ASTM D4991, ASTM F1929, ASTM F88, & ASTM F1140) | Confirm that both the inner tray and outer pouch sterile barrier packaging can maintain device sterility throughout claimed shelf life. Testing includes visual inspection, whole package integrity (dye immersion– weight change, validated alternative method) and seal integrity assessments (seal peel and dye penetration testing for inner tray; burst test and dye penetration for outer pouch) after aging (both accelerated aging and real-time aging conditions). | Meets sterile barrier packaging visual inspection, whole package integrity & seal integrity requirements. For visual inspection, no channels, voids, punctures, or breaches observed, and labels are legible. For dye immersion, - weight change, there is no reduction in weight of the inner tray after it is held under vacuum for 30 minutes in vacuum chamber. For seal peel strength, average peel strength exceeds minimum force. For burst test, internal pressure exceeds minimum burst pressure before package failure. For dye penetration, no penetration of dye into seal area is observed. | Passed | | Bacterial Endotoxin Testing (ANSI/AAMI ST72, USP<85>, FDA Guidance on Endotoxin Testing Recommendations for Single-Use Intraocular Ophthalmic Devices) | Confirm that Endotoxin present on product is below the permanent intraocular device limit to confirm product is non-pyrogenic. | ≤ 0.2 Endotoxin Units (EU)/device | Passed | | Bioburden Estimation Testing (ISO 11737-1) | Confirm that naturally occurring Bioburden is controlled on product and in manufacturing environment. | Less than the Specification Limit of < 100 Colony Forming Units (CFU)/device | Passed | PMA P060011/S039: FDA Summary of Safety and Effectiveness Data 13 of 18 Version: 24th September 2025 {13} ## **X. SUMMARY OF PRIMARY CLINICAL STUDY** The applicant performed a clinical study to establish a reasonable assurance of safety and effectiveness of subjects undergoing cataract extraction and IOL implantation with the RayOne EMV Toric, Model RAO210T for primary implantation in the capsular bag of the eye in adult subjects for visual correction of aphakia and corneal astigmatism following removal of a cataractous lens for improved uncorrected distance vision in the US under IDE # G230072. Data from this clinical study were the basis for the PMA approval decision. A summary of the clinical study is presented below. ### **A. Study Design** Subjects were treated between August 10, 2023, and June 19, 2024. The database for this Panel Track Supplement reflected data collected through October 23, 2024, and included 274 subjects. There were 9 investigational sites. This study was a prospective, multi-center, randomized, active controlled, masked (assessor and subject) clinical study for unilateral implantation of low cylinder (1.50 D) toric IOL. Subjects who met all protocol-specified eligibility criteria were to be randomized to receive either the Rayner RAO210T Toric IOL (1.50 D low cylinder) or the Rayner RAO600C aspheric monofocal IOL in the study eye according to a 1:1 ratio if the study eye's estimated IOL cylinder power using Barrett Toric Calculator was 1.50 D. If a subject had significant cataract in both eyes, it was recommended that cataract surgery be performed in one eye before the subject enrolled in the study. Once a subject was enrolled, it was also recommended that the fellow eye not undergo cataract surgery (except for a YAG capsulotomy) throughout the duration of the study. At screening, if both eyes qualified for the study, the eye to undergo cataract surgery and IOL implantation was the eye with worse pre-operative best corrected distance visual acuity (BCDVA). If pre-operative BCDVA was the same for each eye, the right eye was to be the study eye. Subjects completed 6 study visits between the pre-operative visit to completion of Visit 4 (120 to 180 days post-operative). Nine (9) U.S. sites were to be encouraged to enroll a minimum of 20 subjects in the study. No site could enroll more than 25% of the subjects enrolled in the study. The control group was treated with Rayner RAO600C aspheric monofocal IOL.. Study sample size was based on the ANSI Standard for Toric IOLs, Z80.30, as well as ISO 11979-7 requirements for a modification of a parent lens, which specify a minimum of 100 evaluable implanted subjects. #### **1. Clinical Inclusion and Exclusion Criteria** ---PMA P060011/S039: FDA Summary of Safety and Effectiveness Data 14 of 18 Version: 24th September 2025 {14} Enrollment in the RayOne EMV Toric study was limited to patients who met the following inclusion criteria. All ocular criteria must have been met in the eye receiving the study lens (only one eye was to be enrolled): 1) Male or female, 22 years or older at the pre-operative visit who have cataract with best corrected distance visual acuity of 0.30 logMAR (20/40) or worse in at least one eye with or without a glare source present who are eligible for phacoemulsification cataract surgery 2) Subjects who are projected to have best corrected distance visual acuity 0.20 logMAR (20/30) or better after IOL implantation by potential acuity meter (PAM) or Investigator estimation 3) Clear intraocular media other than cataract 4) Contact lens wearers must demonstrate stability of biometry 5) Have the capability to understand and sign an IRB approved informed consent form and privacy authorization in accordance with local regulations 6) Female subjects must be 1-year postmenopausal, surgically sterilized, or, if of childbearing potential, have a negative urine pregnancy test at the Pre-operative Visit. Women of childbearing potential must use an acceptable form of contraception throughout the study. Acceptable methods included at least one of the following: intrauterine (intrauterine device), hormonal (oral, injection, patch, implant, ring), barrier with spermicide (condom, diaphragm), or abstinence. 7) Have Investigator selected IOL spherical equivalent power between +10.0 D to +25.0 D in 0.5 D steps and IOL cylinder power of +1.50 D 8) Have pre-existing corneal astigmatism of 1.00 D to 1.50 D as determined by keratometry 9) Dilated pupil size 5.5 mm or greater to allow visualization of the toric IOL axis markings post-operatively Patients were not permitted to enroll in the RayOne EMV Toric study if they met any of the following exclusion criteria: The ocular exclusion criteria applied to both eyes, unless otherwise noted as only the planned operative eye: 1) Previous intraocular, corneal, or retinal detachment surgery, including corneal transplant, LASIK / LASEK / PRK, SMILE, astigmatic keratotomy and limbal relaxing incisions in the planned operative eye. 2) Diagnosed degenerative visual disorders (e.g., macular degeneration, retinal detachment, proliferative diabetic retinopathy, or other retinal disorders) that PMA P060011/S039: FDA Summary of Safety and Effectiveness Data 15 of 18 Version: 24th September 2025 {15} are predicted to cause future acuity losses to a level of 0.20 logMAR (20/30) or worse 3) Significant anterior segment pathology that might increase intraoperative risk or compromise IOL stability (e.g., pseudoexfoliation syndrome, any iris pathology) 4) Subjects with conditions associated with increased risk of zonular rupture (that may affect post-operative centration or tilt of IOL) in the planned operative eye 5) Potentially occludable angle or ciliary body tumor, or other pathology that might increase risk to subject safety, based on gonioscopic observation 6) Subjects reasonably expected to require secondary ocular surgical intervention or laser treatment (other than YAG capsulotomy) 7) Subjects with clinically significant corneal pathology, potentially affecting corneal topography 8) Subjects with traumatic cataract in the planned operative eye 9) Participating in a concurrent drug or device clinical trial or who have participated in a drug or device trial within 30 days of the pre-operative visit 10) Subjects with any other serious ocular pathology (e.g., glaucoma, severe dry eye, history of intraocular inflammation, history of retinal surgery or retinal laser procedure) or underlying systemic medical condition (e.g., uncontrolled diabetes) or circumstance that, based on the Investigator's judgment, poses a concern for the subjects' safety or could confound the results of the study (History of cataract surgery with PC-IOL in one eye was allowed.) 11) Use of medications known to interfere with visual performance, pupil dilation, or iris structure within 30 days of the pre-operative visit, at the discretion of the Investigator 12) Pregnant or nursing females 13) Irregular astigmatism in the planned operative eye ## 2. Follow-up Schedule All patients were scheduled to return for follow-up post-operative examinations at: 1. (Form 1) Visit 1 - Postoperative days 1-2 2. (Form 2) Visit 2 - Postoperative days 7-14 3. (Form 3) Visit 3 - Postoperative days 30-60 4. (Form 4) Visit 4 - Postoperative days 120-180 Preoperatively, subjects were evaluated to obtain demographic information and a medical history and to establish a baseline for their ocular condition, PMA P060011/S039: FDA Summary of Safety and Effectiveness Data 16 of 18 Version: 24th September 2025 {16} including manifest refraction, uncorrected and best-corrected photopic distance visual acuity using ETDRS chart at 4m, biometry, IOL power calculation, dilated pupil size, and anterior and posterior segment evaluations. In addition, prior to the operative procedure, the intended axis was marked with a corneal marker. Postoperatively, the objective parameters measured during the study included the following for the study eye: - Residual manifest cylinder (determined by manifest refraction) – Visits 1, 2, 3 and 4 - IOL axis orientation (retro-illuminated slit lamp images with dilated pupils acquired in Toric group subjects; images were sent to independent reading center for evaluation) – at the conclusion of surgery at the Operative visit and at Visits 1, 2, 3 and 4 - Uncorrected distance visual acuity (UCDVA) (photopic ETDRS, 4m) – Visits 1, 2, 3 and 4 - Best-corrected distance visual acuity (BCDVA) (photopic ETDRS, 4m) – Visits 2, 3 and 4 - Keratometry – Visits 3 and 4 - Lens stability (decentration and tilt) - Visits 2, 3 and 4 In addition, slit lamp biomicroscopy for ocular health, intraocular pressure using applanation tonometry, posterior capsule opacification assessment, PC-IOL observations and fundus examination were performed at all post-operative visits. The Schedule of Events can be found in TABLE 9 below. TABLE 9 SCHEDULE OF EVENTS | Procedure | Both Eyes | Study Eye^{1} | | | | | | --- | --- | --- | --- | --- | --- | --- | | | Visit 00 Pre-Op | Visit 0 Operative | Visit 1 Post-Op | Visit 2 Post-Op | Visit 3 Post-Op | Visit 4 Post-Op | | | Day -90 - 0 | Day 0 | Day 1-2 | Day 7-14 | Day 30-60 | Day 120- 180 | | Informed Consent/HIPAA | X | | | | | | | Inclusion/Exclusion Criteria Review | X | X^{2} | | | | | | Demographics | X | | | | | | | Ocular and Significant Non-ocular Medical History | X* | X | | | | | | Urine Pregnancy Test (if applicable) | X | | | | | | | Potential Visual Acuity | X* | | | | | | | Corneal Topography | X* | | | | | | | Axial length/ Anterior chamber depth (biometry)/ Target refraction/ IOL power calc | X* | | | | | | PMA P060011/S039: FDA Summary of Safety and Effectiveness Data 17 of 18 Version: 24th September 2025 {17} | Procedure | Both Eyes | Study Eye^{1} | | | | | | --- | --- | --- | --- | --- | --- | --- | | | Visit 00 Pre-Op | Visit 0 Operative | Visit 1 Post-Op | Visit 2 Post-Op | Visit 3 Post-Op | Visit 4 Post-Op | | | Day -90 - 0 | Day 0 | Day 1-2 | Day 7-14 | Day 30-60 | Day 120- 180 | | (Barrett Toric and any non-toric Monofocal IOL formula) | | | | | | | | Keratometry | X* | | | | X | X | | Gonioscopy | X* | | | | | | | Dilated Pupil size | X* | | | | | | | Randomization (for subjects with estimated IOL cyl power of 1.50 D with Barrett Toric) | X | | | | | | | Operative Procedures^{3} | | X | | | | | | Manifest Refraction^{4} | X | | | X | X | X | | UCDVA - Monocular (ETDRS) – 4 m^{4} | X | | X | X | X | X | | BCDVA - Monocular (ETDRS) – 4 m^{4} | X | | | X | X | X | | Intraocular Pressure^{4} | X* | | X | X | X | X | | Slit lamp Biomicroscopy^{4} | X* | | X | X | X | X | | Posterior Capsule Opacification (PCO) Assessment^{4} (**dilated**) | | | X | X | X | X | | IOL Observations^{4} (**dilated**) | | | X | X | X | X | | Lens Stability - decentration and tilt (**dilated**) | | | | X | X | X | | IOL Axis Orientation - retroilluminated slit lamp photo^{5} (**dilated**) | | X | X | X | X | X | | Dilated Fundus Exam^{[4]} | X* | | X | X | X | X | | Adverse Events^{4} | X | X | X | X | X | X | | Device Deficiencies^{4} | | X | X | X | X | X | | Concomitant Medications^{4} | X | X | X | X | X | X | | Exit from Study | | | | | | X | 1. If both eyes qualify, the study eye will be the eye with the worse preoperative BCDVA. If BCDVA of both eyes is the same, the study eye will be the right eye 2. Review of inclusion/exclusion criteria before surgery 3. Surgical incision will be standardized to the temporal horizontal meridian and approximately 2.4 mm incision size to minimize variations in changes in corneal astigmatism between toric and control subjects; incision location and size information should be noted in Form 0. The intended axis will be marked preoperatively with two marks, 180 degrees apart in the peripheral cornea near the limbus, using a corneal marker such as a pre-inked RoboMarker (Surgilum, Wilmington, NC) or similar device. 4. Additionally, to be completed at any post-operative Unscheduled Visit 5. Subjects implanted with RAO210T toric IOL only. Images will be evaluated by an independent Reading center to assess IOL axis orientation from IOL axis marks correlated with intended axis and anatomic landmarks at Day 0 and all subsequent visits. *Assessments performed as standard of care within the -90 – 0 pre-operative visit screening window may be used as qualifying assessments prior to date of informed consent. [ ] Optional at Unscheduled Visit, based on Investigator's discretion PMA P060011/S039: FDA Summary of Safety and Effectiveness Data 18 of 18 Version: 24th September 2025 {18} Adverse events and complications were recorded at all visits. The key timepoints are shown below in the tables summarizing safety and effectiveness. 3. Clinical Endpoints With regards to safety: Co-primary safety endpoints - Rates of IOL adverse events through 120 to 180 days post-operatively (Visit 4) for eyes implanted with Toric IOLs compared to ISO Safety and Performance Endpoint (SPE) rates as defined in ISO 11979-7. Per the performance criteria specified in Annex E of ISO 11979-7, the null and alternative hypotheses for each IOL adverse event were structured as follows: $$H_{0s}: p_{ts} \leq p_0$$ $$H_{1s}: p_{ts} > p_0$$ Where $p_{ts}$ denoted the true proportion of Toric test group eyes at Visit 4 suffering from the adverse event, and $p_0$ represented the corresponding SPE rate specified in ISO 11979-7. If none of the null hypotheses were rejected at a one-sided 0.05 significance level using exact binomial test, then it would be concluded that the Toric IOL successfully met this endpoint according to the ISO 11979-7 performance criteria. - Rates of all other adverse events not included in IOL adverse event list from ISO 11979-7 through 120 to 180 days post-operatively (Visit 4). These rates were planned to be descriptively summarized. - Rates of secondary surgical interventions for IOL repositioning due to IOL misalignment through 120 to 180 days post-operatively (Visit 4). These rates were planned to be descriptively summarized. - Rate of BCDVA 0.30 logMAR or better at 120 to 180 days post-operatively (Visit 4) in eyes implanted with Toric IOLs compared to the ISO SPE rate specified in ISO 11979-7. For the proportion of Toric test group eyes with 0.30 logMAR or better at Visit 4, the null and alternative hypotheses were as follows: $$H_{0s}: p_{ts} \geq p_0$$ $$H_{1s}: p_{ts} < p_0$$ PMA P060011/S039: FDA Summary of Safety and Effectiveness Data 19 of 18 Version: 24th September 2025 {19} Where $p_{ts}$ denoted the true proportion of Toric test group eyes achieving a BCDVA of 0.30 logMAR or better at Visit 4 and $p_0$ was the corresponding SPE rate specified in ISO 11979-7. If the null hypothesis was not rejected at a one-sided 0.05 significance level using exact binomial test for both the Safety and Best Case Sets (defined in Section X.A.4), then it would be concluded that the Toric IOL successfully met this endpoint according to the ISO 11979-7 performance criteria. With regards to effectiveness: Co-primary effectiveness endpoints - Mean magnitude of residual manifest cylinder (as measured by manifest refraction) at 120 to 180 days post-operatively (Visit 4). The primary effectiveness hypotheses were: $$H_{0e}: \mu_{te} - \mu_{ce} \geq 0$$ $$H_{1e}: \mu_{te} - \mu_{ce} < 0$$ Where $\mu_{te}$ and $\mu_{ce}$ denoted the population mean magnitude of residual manifest cylinder at Visit 4 for the Toric test group and Monofocal control group, respectively. If the null hypothesis was rejected at a one-sided 0.025 significance level, then it would be concluded that the 1.50 D Toric IOL is statistically successful in this endpoint. - Percentage of Toric IOLs with axis misalignment at Visit 4 (as determined by photographic method) less than 10 degrees and 20 degrees Per ANSI Z80.30, the co-primary endpoint for axis misalignment would be considered successful if IOL axis misalignment was less than 10 degrees for 90% of eyes and less than 20 degrees for 95% of eyes from the Toric test group. - Stability of Toric IOL axis orientation, expressed as percentage of Toric IOLs that rotate $\leq 5$ degrees postoperatively between 30 to 60 days (Visit 3) and 120 to 180 days (Visit 4). Per ANSI Z80.30, the co-primary endpoint for axis stability would be considered successful at Visit 4 if at least 90% of the Toric IOL eyes rotated less than or equal to five degrees between Visits 3 and 4. The following secondary effectiveness endpoints were planned to be descriptively summarized: - Residual manifest cylinder by subgroups of 0.25 D preoperative keratometric cylinder at 120 to 180 days post-operatively (Visit 4) - Percent reduction in absolute cylinder (calculated as reduction of magnitude of residual manifest cylinder relative to preoperative keratometric cylinder) PMA P060011/S039: FDA Summary of Safety and Effectiveness Data 20 of 18 Version: 24th September 2025 {20} at 120 to 180 days post-operatively (Visit 4) Additional effectiveness variables were: - Percentage of eyes achieving UCDVA at 4 m at 120 to 180 days post-operatively (Visit 4): ≤ 0.30 logMAR, ≤ 0.20 logMAR, ≤ 0.00 logMAR - Percentage of eyes achieving BCDVA at 4 m at 120 to 180 days post-operatively (Visit 4): ≤ 0.30 logMAR, ≤ 0.20 logMAR, ≤ 0.00 logMAR - Percentage of eyes achieving accuracy of cylinder (to target) at 120 to 180 days post-operatively (Visit 4) within: ±0.25 D, ±0.50 D, ±0.75 D - Lens axis misalignment (with toric IOLs) (compared to intended) at 120 to 180 days post-operatively (Visit 4): Absolute value of misalignment, Signed value of misalignment, Two-sided tolerance interval around mean of the signed value of misalignment, Percentage of eyes with lens axis misalignment: < 10 degrees, < 20 degrees, > 30 degrees - Reduction in cylinder power at 120 to 180 days post-operatively (Visit 4), defined as the difference between pre-operative magnitude of keratometric cylinder and magnitude of manifest cylinder at Visit 4 - Reduction in cylinder power by subgroups of 0.25 D preoperative keratometric cylinder at 120 to 180 days post-operatively (Visit 4) - Percentage of eyes with reduction of cylinder, at 120 to 180 days post-operatively (Visit 4), within 0.50 D and within 1.00 D of intended - Descriptive statistics concerning the distribution of surgically induced astigmatism (at Visit 4) - Scatterplots and regression analyses of the change in magnitude of corneal astigmatism (at Visit 4) from preop as a function of the incision location - Scatterplots and regression analyses of change in corneal cylinder axis (at Visit 4) from preop, as a function of preoperative corneal cylinder magnitude - Double-angle plots and vector analyses (intended refractive correction, surgically induced refractive correction, error vector, correction ratio, error ratio) at Visit 4, stratified by preoperative corneal cylinder magnitude #### 4. Analysis Sets - All Enrolled Set – All subjects who signed informed consent. This set would be used for disposition summaries. - Intent-to-Treat (ITT) Set – All enrolled subjects who were randomized. All ITT analyses would be done according to randomized lens assignment. - Safety Set (SAF) – All ITT subjects who underwent surgery. Summaries and analyses based on the Safety Set would be analyzed according to IOL actually implanted. If no IOL was successfully implanted, but the IOL touched the eye, the subject would be analyzed by the attempted lens. If a subject underwent surgery yet no IOL touched the eye, the subject would be analyzed in a separate Not Treated group. If it was confirmed that a subject did not undergo surgery, then the subject would be excluded from all safety PMA P060011/S039: FDA Summary of Safety and Effectiveness Data 21 of 18 Version: 24th September 2025 {21} analyses. - Per Protocol (PP) Set – All SAF subjects who were successfully implanted with a study lens and who had no major protocol deviations. - Best Case Set (BCS) – All PP subjects who also met the following criteria: o No clinically significant pre-operative ocular pathology o No macular degeneration detected at any time o No previous surgery for the correction of refractive errors, which would have resulted in exclusion of the subject from the study ### B. Accountability of PMA Cohort At the time of database lock, of 274 patients enrolled (consented) in the PMA study, 85.0% (233) patients were available for analysis at the completion of the study, the 6-month post-operative visit (Day 120-180, Visit 4). Refer to TABLE 10 for accountability at each scheduled visit and TABLE 11 for the number and percentage of subjects included in the study analysis sets. In all tables, 'N' refers to the total number of subjects (Toric or Monofocal) in the specified analysis set (e.g., ITT or Safety, as indicated). TABLE 10 ACCOUNTABILITY OF SUBJECTS AT EACH FORM VISIT (ITT SET) | Randomized Lens | Accountability | Operative n (%) | Visit 1 (Day 1-2) n (%) | Visit 2 (Day 7-14) n (%) | Visit 3 (Day 30-60) n (%) | Visit 4 (Day 120-180) n (%) | | --- | --- | --- | --- | --- | --- | --- | | Toric IOL (N=118) | Available for Analysis (in Window) | 117 (99.2%) | 117 (99.2%) | 116 (98.3%) | 115 (97.5%) | 117 (99.2%) | | | Visit Out of Window | 0 | 0 | 0 | 1 (0.8%) | 0 | | | Missed Visit | 0 | 0 | 1 (0.8%) | 1 (0.8%) | 0 | | | Discontinued | 1 (0.8%) | 1 (0.8%) | 1 (0.8%) | 1 (0.8%) | 1 (0.8%) | | | Lost to Follow-up | 0 | 0 | 0 | 0 | 0 | | | % Accountability | 117 (100%) | 117 (100%) | 116 (99.1%) | 115 (98.3%) | 117 (100%) | | Monofocal IOL (N=120) | Available for Analysis (in Window) | 117 (97.5%) | 115 (95.8%) | 115 (95.8%) | 113 (94.2%) | 116 (96.7%) | | | Visit Out of Window | 0 | 1 (0.8%) | 1 (0.8%) | 3 (2.5%) | 0 | | | Missed Visit | 0 | 0 | 0 | 0 | 0 | | | Discontinued | 3 (2.5%) | 4 (3.3%) | 4 (3.3%) | 4 (3.3%) | 4 (3.3%) | | | Lost to Follow-up | 0 | 0 | 0 | 0 | 0 | | | % Accountability | 117 (100%) | 115 (99.1%) | 115 (99.1%) | 113 (97.4%) | 116 (100%) | PMA P060011/S039: FDA Summary of Safety and Effectiveness Data 22 of 18 Version: 24th September 2025 {22} N = total eyes in the analysis set n = total number of eyes in the specified row category at that visit Total number of subjects (Toric or Monofocal) in the ITT Set (N=118 Toric; N=120 Monofocal) TABLE 11 ANALYSIS SETS | Population (N=238 Randomized) | n (%) | | --- | --- | | Intent-to-Treat Set[1] | 238 | | Toric IOL | 118 (49.6%) | | Monofocal IOL | 120 (50.4%) | | Safety Set[2] | 234 (98.3%) | | Toric IOL | 117 (50.0%) | | Monofocal IOL | 116 (49.6%) | | Not Treated | 1 (0.4%) | | Per Protocol Set[3] | 226 (95.0%) | | Toric IOL | 111 (49.1%) | | Monofocal IOL | 115 (50.9%) | | Best Case Set[4] | 222 (93.3%) | | Toric IOL | 111 (50.0%) | | Monofocal IOL | 111 (50.0%) | [1]All randomized subjects (eyes). [2]All ITT subjects who undergo surgery. [3]All Safety Set subjects who are implanted with the randomized study lens and who have no major protocol deviations. [4]All Per Protocol Set subjects meeting the conditions specified in the SAP. All subjects with Major Protocol Deviations are excluded from the PP Set. Per Protocol Subjects were excluded from the BCS due to significant pre-operative ocular pathology in the study eye. ### C. Study Population Demographics and Baseline Parameters The demographics of the study population are typical for an IOL study performed in the US. The proportions enrolled were consistent with the sex, age, racial and ethnic prevalence of disease in the US. TABLE 12 shows the distribution of demographics of all ITT subjects in the study. Subjects ranged in age from 48 to 85 years in the Toric group (Mean: 67.7 ± 7.4 years) and 44 to 85 years in the Monofocal group (Mean: 69.4 ± 7.0 years). Females comprised 56% (66/118) in the Toric group and 58% (70/120) of the subjects enrolled in the Monofocal group. The study eye was the right eye in a majority of subjects in both groups. TABLE 13 shows the key baseline parameters for the Toric and Control groups. The distribution of axial length, pre-operative keratometry cylinder, and the grade of cataract are similar between randomized groups. PMA P060011/S039: FDA Summary of Safety and Effectiveness Data 23 of 18 Version: 24th September 2025 {23} TABLE 12 DEMOGRAPHICS (ITT SET) | | | Randomized Lens | | | --- | --- | --- | --- | | Demographic Parameter | | Toric IOL (N=118) | Monofocal IOL (N=120) | | Age (years) | N | 118 | 120 | | | Mean ± SD | 67.7 ± 7.4 | 69.4 ± 7.0 | | | Median | 68 | 70 | | | Min, Max | 48, 85 | 44, 85 | | | | | | | Age Group, n (%) | < 65 years | 36 (30.5%) | 23 (19.2%) | | | ≥ 65 years | 82 (69.5%) | 97 (80.8%) | | | | | | | Sex, n (%) | Male | 52 (44.1%) | 50 (41.7%) | | | Female | 66 (55.9%) | 70 (58.3%) | | | | | | | Study Eye, n (%) | OD | 62 (52.5%) | 73 (60.8%) | | | OS | 56 (47.5%) | 47 (39.2%) | Total number of subjects (Toric or Monofocal) in the ITT Set (N=118 Toric; N=120 Monofocal) TABLE 13 BASELINE PARAMETERS (ITT SET) | | | Randomized Lens | | | --- | --- | --- | --- | | Parameter | Statistic | Toric IOL (N=118) | Monofocal IOL (N=120) | | Axial Length (mm) | N | 118 | 120 | | | Mean ± SD | 24.2 ± 1.1 | 24.1 ± 1.1 | | | Median | 24.1 | 24.1 | | | Min, Max | 21.9, 27.5 | 21.8, 26.7 | | | | | | | Keratometric Cylinder (D)[1] | N | 118 | 120 | | | Mean ± SD | 1.21 ± 0.15 | 1.25 ± 0.14 | | | Median | 1.18 | 1.23 | | | Min, Max | 1.00, 1.50 | 1.02, 1.50 | | | | | | | LOCS III: Nuclear Opalescence, n (%) | N | 117 | 120 | | | NO1 | 11 (9.4%) | 7 (5.8%) | | | NO2 | 31 (26.5%) | 38 (31.7%) | | | NO3 | 44 (37.6%) | 43 (35.8%) | | | NO4 | 6 (5.1%) | 7 (5.8%) | | | NO5 | 6 (5.1%) | 4 (3.3%) | | | NO6 | 1 (0.9%) | 0 | PMA P060011/S039: FDA Summary of Safety and Effectiveness Data 24 of 18 Version: 24th September 2025 {24} | | | Randomized Lens | | | --- | --- | --- | --- | | Parameter | Statistic | Toric IOL (N=118) | Monofocal IOL (N=120) | | | N/A | 18 (15.4%) | 21 (17.5%) | | | | | | | LOCS III: Cortical Opacity, n (%) | N | 117 | 120 | | | C1 | 28 (24.1%) | 29 (24.4%) | | | C2 | 20 (17.2%) | 34 (28.6%) | | | C3 | 18 (15.5%) | 18 (15.1%) | | | C4 | 4 (3.4%) | 3 (2.5%) | | | C5 | 2 (1.7%) | 1 (0.8%) | | | N/A | 44 (37.9%) | 34 (28.6%) | | | | | | | LOCS III: Posterior Subcapsular, n (%) | N | 117 | 120 | | | P1 | 27 (23.3%) | 27 (22.9%) | | | P2 | 12 (10.3%) | 13 (11.0%) | | | P3 | 5 (4.3%) | 4 (3.4%) | | | P4 | 3 (2.6%) | 4 (3.4%) | | | P5 | 0 | 0 | | | N/A | 69 (59.5%) | 70 (59.3%) | [1]One subject was randomized (Toric IOL) but not treated due to a Screen Failure upon re-screening due to a preop keratometric cylinder of 0.61 D. The subject's initial preop value of 1.14 D is used for all keratometric cylinder tables and analyses. Total number of subjects (Toric or Monofocal) in the ITT Set (N=118 Toric; N=120 Monofocal) ### D. Safety and Effectiveness Results #### 1. Safety Results The analysis of safety was based on the Safety set cohort (117 subjects implanted with Toric IOL and 116 subjects implanted with Monofocal IOL) of 233 subjects available for the 6-month (Visit 4) evaluation (Day 120-180). The key safety outcomes for this study are presented below in TABLE 14 and TABLE 18. The results of adverse events analyses based on the consensus definitions as set forth by American Academy of Ophthalmology's Task Force (Masket et al. Ophthalmology 2017) are shown in TABLE 15. All ocular adverse events (serious and non-serious combined) are reported in TABLE 16. Secondary Surgical Interventions (SSIs) due to IOL misalignment are reported in TABLE 17. PMA P060011/S039: FDA Summary of Safety and Effectiveness Data 25 of 18 Version: 24th September 2025 {25} ### **Adverse effects that occurred in the PMA clinical study:** The first co-primary safety endpoint was to evaluate rates of cumulative and persistent adverse events in first operative eyes at Visit 4 in comparison to ISO 11979-7 Safety and Performance Endpoints grid (SPE rates), and the results are summarized in **TABLE 14**. Of the AEs listed in the ISO grid, the two cumulative AEs observed in this study were cystoid macular edema (2/117 subjects, 1.7%) and secondary surgical interventions (SSIs) (2/117 subjects, 1.7%). There were no persistent AEs observed in the trial among those listed in the ISO grid. At Visit 4, no rates of the adverse events in the Toric group were identified to be statistically higher than the ISO grid at one-sided 0.05 significance level, and this endpoint was successfully met according to the ISO performance criteria. TABLE 14 RATES OF IOL ADVERSE EVENTS COMPARED TO ISO 11979-7 SAFETY AND PERFORMANCE ENDPOINTS (SPE) RATES (SAFETY, TORIC IOLS) | Adverse Event | n/N (%) | SPE Rate (%) | P-value^{[1]} | 95% Confidence Interval^{[2]} | | --- | --- | --- | --- | --- | | **Cumulative** | | | | | | Cystoid macular edema | 2/117 (1.7%) | 3.0 | 0.8691 | (0.208%, 6.039%) | | Hypopyon | 0/117 (0.0%) | 0.3 | >.9999 | (0%, 3.104%) | | Endophthalmitis* | 0/117 (0.0%) | 0.1 | >.9999 | (0%, 3.104%) | | Lens dislocation from posterior chamber | 0/117 (0.0%) | 0.1 | >.9999 | (0%, 3.104%) | | Mechanical pupillary block | 0/117 (0.0%) | 0.1 | >.9999 | (0%, 3.104%) | | Retinal detachment | 0/117 (0.0%) | 0.3 | >.9999 | (0%, 3.104%) | | Secondary Surgical Intervention^{[3]} | 2/117 (1.7%) | 0.8 | 0.2406 | (0.208%, 6.039%) | | **Persistent** | | | | | | Cornea stroma edema | 0/117 (0.0%) | 0.3 | >.9999 | (0%, 3.104%) | | Cystoid macular edema | 0/117 (0.0%) | 0.5 | >.9999 | (0%, 3.104%) | | Iritis | 0/117 (0.0%) | 0.3 | >.9999 | (0%, 3.104%) | | Increased IOP | 0/117 (0.0%) | 0.4 | >.9999 | (0%, 3.104%) | $^{[1]}$One-sided Exact Binomial Test comparing the proportion of Toric eyes to the SPE Rate, without multiplicity adjustment. $^{[2]}$Two-sided 95% Exact (Clopper-Pearson) confidence interval for the Toric IOL AE Rate, without multiplicity adjustment. $^{[3]}$Excludes Wound Burp SSI. Posterior capsulotomies were excluded as SSIs. * Endophthalmitis is defined as inflammatory reaction (sterile or infectious) involving the vitreous body. n=number of subjects that experienced the event; N = total number of subjects in the Safety Set implanted with the Toric IOL. The second co-primary safety endpoint was to evaluate the rates of all other adverse events not included in the IOL adverse event list from ISO 11979-7 through Visit 4. The results of adverse events analyses based on the consensus PMA P060011/S039: FDA Summary of Safety and Effectiveness Data 26 of 18 Version: 24th September 2025 {26} definitions as set forth by American Academy of Ophthalmology’s Task Force (Masket et al. Ophthalmology 2017) are shown in **TABLE 15**. All ocular adverse events (serious and non-serious combined) are reported in **TABLE 16**. The most common adverse events reported (out of an N=117) were posterior capsular opacification (n=14 subjects, 12.0%), posterior vitreous detachment (n=7 subjects, 6.06%) and corneal edema (n=7 subjects, 6.0%). No unexpected safety concerns were identified with the RayOne EMV Toric. TABLE 15 RATES OF ADVERSE EVENTS BASED ON A MODIFIED VERSION OF THE AAO TASK FORCE CONSENSUS (MASKET ET AL, 2017), SAFETY SET | Adverse Event | Toric IOL (N=117) | | | Monofocal IOL (N=116) | | | | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | | | Subjects with at least one Masket et al Study Eye AE | | | 5 (4.3%) | | | 4 (3.4%) | | | | n (%) | Two-sided 95% CI^{[1]} | E^{[2]} | n (%) | Two-sided 95% CI^{[1]} | E^{[2]} | | | | Chronic anterior uveitis | 0 | [0.0%, 3.1%] | 0 | 0 | [0.0%, 3.1%] | 0 | | | | Clinically significant cystoid macular edema: Macular edema diagnosed by exam and testing (e.g., OCT, FA) resulting in BCDVA of 0.30 logMAR or worse at ≥ Visit 3 | 1 (0.9%) | [0.0%, 4.7%] | 1 | 0 | [0.0%, 3.1%] | 0 | | | | Visually significant corneal edema | 0 | [0.0%, 3.1%] | 0 | 0 | [0.0%, 3.1%] | 0 | | | | Endophthalmitis | 0 | [0.0%, 3.1%] | 0 | 0 | [0.0%, 3.1%] | 0 | | | | Mechanical pupillary block | 0 | [0.0%, 3.1%] | 0 | 0 | [0.0%, 3.1%] | 0 | | | | Increased IOP: Elevation of IOP by ≥10 mmHg above baseline to a minimum of 25 mmHg | 3 (2.6%) | [0.5%, 7.3%] | 3 | 4 (3.4%) | [0.9%, 8.6%] | 4 | | | | Rhegmatogenous RD | 0 | [0.0%, 3.1%] | 0 | 0 | [0.0%, 3.1%] | 0 | | | | Toxic anterior segment syndrome | 0 | [0.0%, 3.1%] | 0 | 0 | [0.0%, 3.1%] | 0 | | | | Secondary IOL Intervention | 1 (0.9%) | [0.0%, 4.7%] | 1 | 0 | [0.0%, 3.1%] | 0 | | | | Exchange (the investigational device is replaced with the same lens model) | 0 | [0.0%, 3.1%] | 0 | 0 | [0.0%, 3.1%] | 0 | | | | Removal (the investigational device is removed and replaced with a non-investigational lens or no lens is implanted) | 0 | [0.0%, 3.1%] | 0 | 0 | [0.0%, 3.1%] | 0 | | | | IOL repositioning (existing IOL is surgically moved to another location or rotated) | 1 (0.9%) | [0.0%, 4.7%] | 1 | 0 | [0.0%, 3.1%] | 0 | | | $^{[1]}$ Two-sided 95% Exact (Clopper-Pearson) confidence interval for the AE Rate, without multiplicity adjustment. $^{[2]}$ E = Number of events. n=number of subjects that experienced the event; N = total number of subjects in the Safety Set (N=117 Toric; N=116 Monofocal) PMA P060011/S039: FDA Summary of Safety and Effectiveness Data 27 of 18 Version: 24th September 2025 {27} TABLE 16 OCULAR ADVERSE EVENTS (SERIOUS AND NON-SERIOUS COMBINED), SAFETY SET | Adverse Event | Toric IOL (N=117) | | | Monofocal IOL (N=116) | | | | --- | --- | --- | --- | --- | --- | --- | | | n (%) | Two-sided 95% CI^{[1]} | E^{[2]} | n (%) | Two-sided 95% CI^{[1]} | E^{[2]} | | Subjects with at least one Study Eye AE | 49 (41.9%) | [32.8%, 51.4%] | 75 | 40 (34.5%) | [25.9%, 43.9%] | 49 | | Posterior capsular opacification requiring YAG capsulotomy | 14 (12.0%) | [6.7%, 19.3%] | 14 | 4 (3.4%) | [0.9%, 8.6%] | 4 | | Posterior vitreous detachment | 7 (6.0%) | [2.4%, 11.9%] | 7 | 11 (9.5%) | [4.8%, 16.3%] | 11 | | Corneal edema^{[3]} | 7 (6.0%) | [2.4%, 11.9%] | 7 | 7 (6.0%) | [2.5%, 12.0%] | 7 | | Cornea stroma edema^{[4]} | 6 (5.1%) | [1.9%, 10.8%] | 6 | 5 (4.3%) | [1.4%, 9.8%] | 5 | | Increased IOP | 4 (3.4%) | [0.9%, 8.5%] | 4 | 7 (6.0%) | [2.5%, 12.0%] | 7 | | Anterior chamber cells greater than Grade 2 at Visit 1 | 5 (4.3%) | [1.4%, 9.7%] | 5 | 1 (0.9%) | [0.0%, 4.7%] | 1 | | Corneal abrasion | 3 (2.6%) | [0.5%, 7.3%] | 3 | 0 | [0%, 3.1%] | 0 | | Cystoid macular edema | 2 (1.7%) | [0.2%, 6.0%] | 2 | 1 (0.9%) | [0.0%, 4.7%] | 1 | | Epithelial defect | 2 (1.7%) | [0.2%, 6.0%] | 2 | 1 (0.9%) | [0.0%, 4.7%] | 1 | | Epiretinal membrane | 2 (1.7%) | [0.2%, 6.0%] | 2 | 0 | [0%, 3.1%] | 0 | | Choroidal nevus | 1 (0.9%) | [0.0%, 4.7%] | 1 | 1 (0.9%) | [0.0%, 4.7%] | 1 | | Corneal dellen | 1 (0.9%) | [0.0%, 4.7%] | 1 | 1 (0.9%) | [0.0%, 4.7%] | 1 | | Dermatochalasis | 1 (0.9%) | [0.0%, 4.7%] | 1 | 1 (0.9%) | [0.0%, 4.7%] | 1 | | Subconjunctival hemorrhage | 1 (0.9%) | [0.0%, 4.7%] | 1 | 1 (0.9%) | [0.0%, 4.7%] | 1 | | Worsening dry eye disease | 1 (0.9%) | [0.0%, 4.7%] | 1 | 1 (0.9%) | [0.0%, 4.7%] | 1 | | Conjunctival injection | 0 | [0%, 3.1%] | 0 | 2 (1.7%) | [0.2%, 6.1%] | 2 | | Central capsular wrinkle | 1 (0.9%) | [0.0%, 4.7%] | 1 | 0 | [0%, 3.1%] | 0 | | Chemosis | 1 (0.9%) | [0.0%, 4.7%] | 1 | 0 | [0%, 3.1%] | 0 | | Double vision secondary to neurological adjustment following cataract surgery | 1 (0.9%) | [0.0%, 4.7%] | 1 | 0 | [0%, 3.1%] | 0 | | Epithelial basement membrane dystrophy | 1 (0.9%) | [0.0%, 4.7%] | 1 | 0 | [0%, 3.1%] | 0 | | IOL decentration/malposition resulting in secondary surgical intervention | 1 (0.9%) | [0.0%, 4.7%] | 1 | 0 | [0%, 3.1%] | 0 | PMA P060011/S039: FDA Summary of Safety and Effectiveness Data 28 of 18 Version: 24th September 2025 {28} | | Toric IOL (N=117) | | | Monofocal IOL (N=116) | | | | --- | --- | --- | --- | --- | --- | --- | | Adverse Event | n (%) | Two-sided 95% CI[1] | E[2] | n (%) | Two-sided 95% CI[1] | E[2] | | IOL repositioning (existing IOL is surgically moved to another location or rotated) | 1 (0.9%) | [0.0%, 4.7%] | 1 | 0 | [0%, 3.1%] | 0 | | Intraretinal hemorrhage | 1 (0.9%) | [0.0%, 4.7%] | 1 | 0 | [0%, 3.1%] | 0 | | Iris pigment epithelium loss | 1 (0.9%) | [0.0%, 4.7%] | 1 | 0 | [0%, 3.1%] | 0 | | Photophobia | 1 (0.9%) | [0.0%, 4.7%] | 1 | 0 | [0%, 3.1%] | 0 | | Progression or onset diabetic retinopathy | 1 (0.9%) | [0.0%, 4.7%] | 1 | 0 | [0%, 3.1%] | 0 | | Retained lens material | 1 (0.9%) | [0.0%, 4.7%] | 1 | 0 | [0%, 3.1%] | 0 | | Retinal pigment epithelial mottling | 1 (0.9%) | [0.0%, 4.7%] | 1 | 0 | [0%, 3.1%] | 0 | | Superficial punctate keratitis | 1 (0.9%) | [0.0%, 4.7%] | 1 | 0 | [0%, 3.1%] | 0 | | Surgical removal of retained lens fragment | 1 (0.9%) | [0.0%, 4.7%] | 1 | 0 | [0%, 3.1%] | 0 | | Trace punctate epithelial erosions | 1 (0.9%) | [0.0%, 4.7%] | 1 | 0 | [0%, 3.1%] | 0 | | Visually significant corneal edema resulting in BCDVA of 0.3 or worse | 1 (0.9%) | [0.0%, 4.7%] | 1 | 0 | [0%, 3.1%] | 0 | | Worsening punctate epithelial erosion | 1 (0.9%) | [0.0%, 4.7%] | 1 | 0 | [0%, 3.1%] | 0 | | Wound burp (aqueous tap, or paracentesis) | 1 (0.9%) | [0.0%, 4.7%] | 1 | 0 | [0%, 3.1%] | 0 | | Dry eye disease | 0 | [0%, 3.1%] | 0 | 1 (0.9%) | [0.0%, 4.7%] | 1 | | Iritis | 0 | [0%, 3.1%] | 0 | 1 (0.9%) | [0.0%, 4.7%] | 1 | | Keratoconjunctivitis Sicca, not specified as Sjogrens | 0 | [0%, 3.1%] | 0 | 1 (0.9%) | [0.0%, 4.7%] | 1 | | Retinal pigment epithelial changes | 0 | [0%, 3.1%] | 0 | 1 (0.9%) | [0.0%, 4.7%] | 1 | [1]Two-sided 95% Exact (Clopper-Pearson) confidence interval for the AE Rate, without multiplicity adjustment. [2]E = Number of events. [3] Corneal edema of grade 2+ or more reported after IOL surgery in the early post-operative period (prior to one week). [4] Edema of corneal stroma of any degree at any location regardless of effect on BCDVA at Visit 2 (Day 7-14) or later. n=number of subjects that experienced the event; N = total number of subjects in the Safety Set (N=117 Toric; N=116 Monofocal) The third co-primary safety endpoint was to evaluate the rates of secondary surgical interventions for IOL repositioning due to IOL misalignment through Visit 4 (TABLE 17). One out of 117 (0.9%) toric subjects underwent IOL PMA P060011/S039: FDA Summary of Safety and Effectiveness Data 29 of 18 Version: 24th September 2025 {29} repositioning surgery within the first few weeks. The need for repositioning was not attributed to the device itself, but to the initial placement. TABLE 17 RATE OF SECONDARY SURGICAL INTERVENTION (SSI) DUE TO IOL MISALIGNMENT (SAFETY, TORIC IOLS) | | n/N (%) | 95% Confidence Interval^{[1]} | | --- | --- | --- | | Secondary Surgical Intervention for IOL Repositioning due to IOL Misalignment | 1/117 (0.9%) | (0.022%, 4.670%) | $^{[1]}$Two-sided 95% Exact (Clopper-Pearson) confidence interval for the Toric IOL AE Rate, without multiplicity adjustment. n=number of subjects that experienced the event; N = total number of subjects in the Safety Set implanted with the Toric IOL. The fourth co-primary safety endpoint was to evaluate the rate of BCDVA of 0.30 logMAR or better at Visit 4 compared to the ISO SPE rates as described in ISO 11979-7, and the results are summarized in **TABLE 18**. All Toric subjects achieved BCDVA of 0.30 logMAR (20/40) or Month 6 post-operatively based the Safety set (117/117, 100%, 95% CI [96.9%, 100%]) and the Best-Case set (111/111, 100%, 95% CI [96.7%, 100%]). The rate of BCDVA of 0.30 logMAR or better at Visit 4 for the test Toric group was not identified to be statistically significantly lower than the corresponding ISO SPE rate based on both the Safety (92.5%) and Best Case sets (96.7%), therefore this endpoint was successfully met per the ISO performance criteria. Therefore, all co-primary safety endpoints were met. TABLE 18 RATE OF BCDVA OF 0.30 LOGMAR OR BETTER AT VISIT 4 (SAFETY AND BEST-CASE SETS) | Analysis Population | Implanted Lens | n/N (%) | SPE Rate (%) | P-value^{[1]} | 95% Confidence Interval^{[2]} | | --- | --- | --- | --- | --- | --- | | Safety Set | Toric IOL | 117/117 (100.0%) | 92.5 | >.9999 | (96.90%, 100%) | | | Monofocal IOL | 116/116 (100.0%) | 92.5 | >.9999 | (96.87%, 100%) | | | | | | | | | Best Case Set | Toric IOL | 111/111 (100.0%) | 96.7 | >.9999 | (96.73%, 100%) | | | Monofocal IOL | 111/111 (100.0%) | 96.7 | >.9999 | (96.73%, 100%) | $^{[1]}$One-sided Exact Binomial Test comparing the proportion of Toric eyes to the SPE Rate, without multiplicity adjustment. $^{[2]}$Two-sided 95% Exact (Clopper-Pearson) confidence interval for the Toric IOL AE Rate, without multiplicity adjustment. PMA P060011/S039: FDA Summary of Safety and Effectiveness Data 30 of 18 Version: 24th September 2025 {30} Total number of subjects (Toric or Monofocal) in the Safety Set (N=117 Toric; N=116 Monofocal) ## 2. Effectiveness Results The analysis of effectiveness was based on the 238 evaluable patients at the 6-month time point (Day 120-180). Key effectiveness outcomes are presented in **TABLE 19**, **TABLE 20** and **TABLE 21**. The primary analysis was based on the Intent to Treat Analysis Set (ITT) with missing data being handled using multiple imputation (MI) method. Visual acuity was measured in this study with the Early Treatment of Diabetic Retinopathy Study (ETDRS) charts using the M&S Clinical Trial Suite system (M&S Technologies, Niles, IL) at 4 meters. All three co-primary effectiveness endpoints were met according to the ANSI Z80.30 performance criteria: the adjusted mean difference in magnitude of residual manifest cylinder between the randomized groups (Test-Control) was -0.430 D (95% CI [-0.544, -0.316]), which was statistically ( $P < 0.0001$ ) and clinically significant, meeting the minimum clinically important difference (MCID) of 0.40 D recommended by FDA (**TABLE 19**); the percentages of Toric IOL axis misalignment (from pre-operatively marked intended axis) under 10 degrees and 20 degrees exceeded the minimum percentages of 90% and 95%, respectively, (93.1%, 95% CI [86.1%, 96.8%] less than 10 degrees and 99.99%, 95% CI [96.9%, 100%] less than 20 degrees) per the primary imputation analysis (**TABLE 20**) and the percentage of Toric IOLs that rotated $\leq$ five degrees between Visit 3 (Day 30-60) and Visit 4 (Day 120-180) was 99.0% (95% CI [93.5%, 99.9%]) and exceeded the pre-specified success criteria of 90% for rotational stability using the primary imputation strategy (**TABLE 21**). The results of the secondary effectiveness endpoints are as follows: at Visit 4, the magnitude of residual manifest cylinder was stratified by preoperative keratometric cylinder for the Toric group compared with the Monofocal group, with between-group differences of -0.381 D $\pm$ 0.078 D in the 1.00 to <1.25 D subgroup and -0.478 D $\pm$ 0.088 D in the 1.25 to 1.50 D subgroup. Percent reduction in absolute cylinder at Visit 4 was 68.6% in the Toric group versus 36.1% in the Monofocal group (**TABLE 22** and **TABLE 23**). In addition to the co-primary and secondary effectiveness endpoints, several additional effectiveness analyses were performed to further characterize visual and refractive outcomes. These included uncorrected and best-corrected distance visual acuity (**TABLE 24** and **TABLE 25**), accuracy of cylinder to target (**TABLE 26**), lens axis misalignment (**TABLE 27**), mean reduction in cylinder power overall and by preoperative keratometric cylinder subgroup (**TABLE 28** and **TABLE 29**), and the percentage of eyes with ---PMA P060011/S039: FDA Summary of Safety and Effectiveness Data 31 of 18 Version: 24th September 2025 {31} reduction in cylinder within 0.50 D and 1.00 D of intended (TABLE 30). These analyses were evaluated descriptively using summary statistics. TABLE 19 RESIDUAL MANIFEST CYLINDER (D) AT VISIT 4 – PRIMARY IMPUTATION STRATEGY (ITT SET) | Statistic | Result | | --- | --- | | **Observed data** | | | Toric IOL | 117/118 | | Mean ± SD | 0.368 ± 0.334 | | Median | 0.25 | | Min, Max | 0.00, 1.25 | | Q1, Q3 | 0.00, 0.50 | | | | | Monofocal (Control) IOL | 116/120 | | Mean ± SD | 0.780 ± 0.532 | | Median | 0.75 | | Min, Max | 0.00, 2.50 | | Q1, Q3 | 0.38, 1.25 | | | | | Difference in Means ± SE | -0.413 ± 0.0582 | | **With Imputation** | | | Number of Imputations^{[1]} | 50 | | LS Mean Difference ± SE | -0.430 ± 0.058^{[2]} | | 95% Confidence Interval | [-0.544, -0.316] | | t Value | -7.39 | | P-value | <.0001 | Results are presented in absolute cylinder (D); more negative differences are more favorable for the Toric lens. One toric IOL subject underwent an SSI for IOL repositioning; Per ANSI Z80.30, the last Residual Manifest cylinder (1.25 D) prior to the SSI was carried forward to subsequent Visits 3 and 4. [1]Missing values imputed with the Markov chain Monte Carlo (MCMC) method. Following imputation, the primary hypothesis was tested using an Analysis of Covariance (ANCOVA) model with randomized treatment as factor and preoperative keratometric cylinder as covariate. [2]Meets the minimum clinically significant difference of -0.40 D TABLE 20 ABSOLUTE IOL AXIS MISALIGNMENT (DEGREES) AT VISIT 4 – PRIMARY IMPUTATION STRATEGY (ITT SET) | | < 10 Degrees Misalignment at V4 | < 20 Degrees Misalignment at V4 | | --- | --- | --- | | Absolute Misalignment (degrees), at Visit 4 | | | | % (n/N), observed | 93.9% (92/98) | 100% (98/98) | | %, with imputation^{[1]} | 93.1% | 99.99% | | 95% Confidence Interval ^{[2]}, with imputation ^{[1,3]} | [86.1%, 96.8%] | [96.9%, 100%] | PMA P060011/S039: FDA Summary of Safety and Effectiveness Data 32 of 18 Version: 24th September 2025 {32} IOL Axis Misalignment is the difference in intended axis of orientation, marked pre-operatively on the day of surgery (Day 0), and the IOL axis at the postoperative visit. One subject underwent an SSI for IOL repositioning; Per ANSI Z80.30, the last IOL Misalignment (9.04°) prior to the SSI was carried forward to subsequent Visits 3 and 4. The IOL axis misalignment endpoint is successful if IOL axis misalignment is less than 10 degrees for 90% of eyes and less than 20 degrees for 95% of eyes from the Toric test group in the fifty imputed datasets. [1] Missing values were handled through multiple imputation (50 sets) using with the Markov chain Monte Carlo (MCMC) method. [2] 95% Exact (Clopper-Pearson) confidence interval (CI), without multiplicity adjustment. [3] All Missing data were imputed as <20 degrees misalignment in 49/50 imputations. The 95% CI presented for the percent of eyes with <20 degrees misalignment is based on a single imputation with 118/118 successes. Total number of subjects (Toric or Monofocal) in the ITT Set (N=118 Toric; N=120 Monofocal) TABLE 21 STABILITY OF TORIC IOL AXIS ORIENTATION (DEGREES) AT VISIT 4 - PRIMARY IMPUTATION STRATEGY (ITT SET) | | Absolute IOL Axis Rotation ≤ 5 degrees between Visit 3 and Visit 4 | | --- | --- | | Stability of Toric IOL Axis Orientation (degrees) at Visit 4 | | | % (n/N), observed | 99.0% (100/101) | | %, with imputation [1] | 99.0% | | 95% Confidence Interval [2], with imputation [1] | [93.5%, 99.9%] | Absolute IOL Axis Rotation is the absolute difference in the actual IOL axis at the postoperative visits The IOL axis stability endpoint is successful if at least 90% of all toric lenses in the fifty imputed datasets rotate less than or equal to five degrees between Visits 3 and 4. [1] Missing values were handled through multiple imputation (50 sets) using with the Markov chain Monte Carlo (MCMC) method. [2] 95% Exact (Clopper-Pearson) confidence interval (CI), without multiplicity adjustment. Total number of subjects (Toric or Monofocal) in the ITT Set (N=118 Toric; N=120 Monofocal) TABLE 22 RESIDUAL MANIFEST CYLINDER (D) AT VISIT 4 STRATIFIED BY PREOPERATIVE KERATOMETRIC CYLINDER (ITT SET) | | Residual Manifest Cylinder (D) at Visit 4 | | | | --- | --- | --- | --- | | Pre-operative Keratometric cylinder | Toric IOL Mean ± SD | Monofocal IOL Mean ± SD | Difference Difference in means (± SE) | | 1.00 to < 1.25 D | | | | | n | 75 | 62 | | | | 0.413 ± 0.333 | 0.794 ± 0.535 | -0.381 (± 0.078) | | 1.25 to 1.50 D | | | | | n | 42 | 54 | | | | 0.286 ± 0.325 | 0.764 ± 0.533 | -0.478 (± 0.088) | PMA P060011/S039: FDA Summary of Safety and Effectiveness Data 33 of 18 Version: 24th September 2025 {33} Abbreviations: D=Diopter; SD=Standard deviation; SE=Standard Error Results are presented in absolute cylinder (D); more negative differences are more favorable for the Toric lens. For one subject who underwent an SSI for IOL repositioning, the last Residual Manifest cylinder (1.25 D) prior to the SSI was carried forward to subsequent Visits 3 and 4, per ANSI Z80.30. Total number of subjects (Toric or Monofocal) in the ITT Set (N=118 Toric; N=120 Monofocal) TABLE 23 PERCENT REDUCTION IN ABSOLUTE CYLINDER (ITT SET) | Percent Reduction in cylinder at Visit 4 | Toric IOL Mean ± SD | Monofocal IOL Mean ± SD | | --- | --- | --- | | n | 117 | 116 | | Mean ± SD | 68.59% ± 29.17 % | 36.11% ± 45.69 % | | Difference | | | | Difference in means (± SE) | 32.48% (± 5.03) % | | Abbreviations: SD=Standard deviation; SE=Standard Error Percent Reduction in absolute cylinder is calculated for each eye as (1 - |residual manifest cylinder|/(preoperative keratometric cylinder)) x 100%. For one subject who underwent an SSI for IOL repositioning, the last Residual Manifest cylinder (1.25 D) prior to the SSI was carried forward to subsequent Visits 3 and 4, per ANSI Z80.30. Total number of subjects (Toric or Monofocal) in the ITT Set (N=118 Toric; N=120 Monofocal); subjects implanted with the Toric IOL (N=117) and Monofocal IOL (N=116). TABLE 2423 UNCORRECTED DISTANCE VISUAL ACUITY (4M) AT VISIT 4 (ALL TORIC IOLS, ITT SET) | UCDVA (logMAR) Outcomes | UCDVA (Snellen Equivalent) | Visit 4 (Day 120-180) n (%) | | --- | --- | --- | | N (eyes) | | 117 | | -0.20 or better | 20/12 | 1 (0.9%) | | -0.10 or better | 20/15 | 10 (8.5%) | | 0.00 or better | 20/20 | 44 (37.6%) | | 0.10 or better | 20/25 | 83 (70.9%) | | 0.20 or better | 20/32 | 102 (87.2%) | | 0.30 or better | 20/40 | 112 (95.7%) | | Count Fingers | Worse than 20/800 | 0 | PMA P060011/S039: FDA Summary of Safety and Effectiveness Data 34 of 18 Version: 24th September 2025 {34} | UCDVA (logMAR) Outcomes | UCDVA (Snellen Equivalent) | Visit 4 (Day 120-180) n (%) | | --- | --- | --- | | n | | 117 | | Mean ± SD | | 0.061 ± 0.130 | | Median | | 0.02 | | Min, Max | | -0.20, 0.52 | | Q1, Q3 | | -0.02, 0.14 | | Not Done | | 0 | Total number of subjects (Toric or Monofocal) in the ITT Set (N=118 Toric; N=120 Monofocal) TABLE 2524 BEST CORRECTED DISTANCE VISUAL ACUITY (4M) AT VISIT 4 (ALL IOLS, ITT SET) | Randomized Lens | BCDVA (logMAR) Outcomes | UCDVA (Snellen Equivalent) | Visit 4 (Day 120-180) n (%) | | --- | --- | --- | --- | | Toric IOL | N (eyes) | | 117 | | -0.20 or better | 20/12 | 7 (6.0%) | | -0.10 or better | 20/15 | 43 (36.8%) | | 0.00 or better | 20/20 | 90 (76.9%) | | 0.10 or better | 20/25 | 114 (97.4%) | | 0.20 or better | 20/32 | 116 (99.1%) | | 0.30 or better | 20/40 | 117 (100%) | | Mean ± SD | -0.055 ± 0.084 | | Median | -0.06 | | Min, Max | -0.20, 0.24 | | Q1, Q3 | -0.10, 0.00 | | Not Done | 0 | | Monofocal IOL | N (eyes) | | 116 | PMA P060011/S039: FDA Summary of Safety and Effectiveness Data 35 of 18 Version: 24th September 2025 {35} | Randomized Lens | BCDVA (logMAR) Outcomes | UCDVA (Snellen Equivalent) | Visit 4 (Day 120-180) n (%) | | --- | --- | --- | --- | | | -0.20 or better | 20/12 | 3 (2.6%) | | | -0.10 or better | 20/15 | 34 (29.3%) | | | 0.00 or better | 20/20 | 83 (71.6%) | | | 0.10 or better | 20/25 | 114 (98.3%) | | | 0.20 or better | 20/32 | 116 (100%) | | | 0.30 or better | 20/40 | 116 (100%) | | | Mean ± SD | | -0.039 ± 0.081 | | | Median | | -0.04 | | | Min, Max | | -0.20, 0.20 | | | Q1, Q3 | | -0.10, 0.02 | | | Not Done | | 0 | Categorical outcomes include count and percentage (%). Total number of subjects (Toric or Monofocal) in the ITT Set (N=118 Toric; N=120 Monofocal) TABLE 2625 ACCURACY OF CYLINDER TO TARGET (ITT SET) | Randomized Lens | Accuracy of Cylinder to Target | Visit 4 (Day 120-180) | | --- | --- | --- | | Toric IOL | ±0.25 D | 66/117 (56.4%) | | ±0.50 D | 100/117 (85.5%) | | ±0.75 D | 110/117 (94.0%) | | >0.75 D | 7/117 (6.0%) | | Monofocal IOL | ±0.25 D | 32/116 (27.6%) | PMA P060011/S039: FDA Summary of Safety and Effectiveness Data 36 of 18 Version: 24th September 2025 {36} | Randomized Lens | A…
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...