← Product Code NIQ · P060008

# TAXUS LIBERTE PACLITAXEL-ELUTING CORONARY STENT SYSTEM (MONORAIL AND OVER-THE-WIRE) (P060008)

_Boston Scientific Corp · NIQ · Oct 10, 2008 · Cardiovascular · APWD_

**Canonical URL:** https://fda.innolitics.com/device/P060008

## Device Facts

- **Applicant:** Boston Scientific Corp
- **Product Code:** NIQ
- **Decision Date:** Oct 10, 2008
- **Decision:** APWD
- **Device Class:** Class 3
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic

## Intended Use

The TAXUS Liberté Paclitaxel-Eluting Coronary Stent System (Monorail and Over-the-Wire Systems) is indicated for improving luminal diameter for the treatment of de novo lesions in native coronary arteries ≥ 2.5 to ≤ 4.00 mm in diameter in lesions ≤ 28 mm in length.

## Device Story

Drug-eluting coronary stent system consisting of a 316L stainless steel Liberté stent coated with paclitaxel in a Translute (SIBS) polymer carrier. Delivered via Monorail or Over-the-Wire balloon catheter. Used in cardiac catheterization labs by interventional cardiologists to treat de novo coronary artery lesions. Stent is advanced over a guidewire to the lesion; balloon inflation deploys the stent, providing mechanical scaffolding while the polymer coating releases paclitaxel to inhibit neointimal hyperplasia. Output is a dilated, drug-eluting scaffold that improves luminal diameter and reduces restenosis. Clinical benefit includes reduced need for target vessel revascularization compared to bare metal stents.

## Clinical Evidence

Evidence includes the pivotal TAXUS ATLAS trial (n=871) comparing TAXUS Liberté to historical TAXUS Express controls (n=991). Primary endpoint (9-month TVR) met non-inferiority (8.0% vs 7.1%, p=0.4787). Secondary endpoints (binary restenosis, MLD, late loss) confirmed non-inferiority. Supporting data from TAXUS IV (randomized, n=1314) and TAXUS V (randomized, n=1156) demonstrated superiority over bare metal stents. Safety profile showed comparable MACE and stent thrombosis rates.

## Technological Characteristics

Stent: 316L stainless steel (ASTM F138-00). Coating: Paclitaxel (1μg/mm²) in Translute (SIBS) tri-block copolymer. Delivery: Monorail or Over-the-Wire balloon catheter. Sterilization: Ethylene oxide. MRI: MR Conditional (≤3 Tesla). Software: None.

## Regulatory Identification

Stent, coronary, drug-eluting -- a metal scaffold with a drug coating placed via a delivery catheter into the coronary artery or saphenous vein graft to maintain the lumen.  The drug coating is intended to inhibit restenosis.

## Predicate Devices

- Liberté Coronary Stent System ([P040016](/device/P040016.md))
- TAXUS Express² Paclitaxel-Eluting Coronary Stent ([P030025](/device/P030025.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a PMA](https://innolitics.com/services/regulatory/), [a 510(k)](https://innolitics.com/services/510ks/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

# Summary of Safety and Effectiveness Data

## I. GENERAL INFORMATION

|  Product Generic Name: | Drug-Eluting Coronary Stent System (NIQ)  |
| --- | --- |
|  Product Trade Name: | TAXUS® Liberté® Paclitaxel-Eluting Coronary Stent System (Monorail)  |
|   | TAXUS® Liberté® Paclitaxel-Eluting Coronary Stent System (Over-the-Wire)  |
|  Applicant’s Name and Address: | Boston Scientific Corporation
One Boston Scientific Place
Natick, MA 01760-1537
USA  |
|  Premarket Approval Application (PMA) Number: | P060008  |
|  Date of Panel: | None  |
|  Date of Notice of Approval to Applicant: | October 10, 2008  |

## II. INDICATIONS FOR USE

The TAXUS Liberté Paclitaxel-Eluting Coronary Stent System (Monorail and Over-the-Wire Systems) is indicated for improving luminal diameter for the treatment of de novo lesions in native coronary arteries $\geq 2.5$ to $\leq 4.00$ mm in diameter in lesions $\leq 28$ mm in length.

## III. CONTRAINDICATIONS

Use of the TAXUS Liberté Paclitaxel-Eluting Coronary Stent System is contraindicated in patients with:

- Known hypersensitivity to paclitaxel or structurally-related compounds.
- Known hypersensitivity to the polymer or its individual components (see details in Section V B2. Inactive Ingredients (Page 4).

Coronary Artery Stenting is contraindicated for use in:

- Patients who can not receive recommended anti-platelet and/or anticoagulant therapy.
- Patients judged to have a lesion that prevents complete inflation of an angioplasty balloon or proper placement of the stent or delivery device.

PMA P060008: FDA Summary of Safety and Effectiveness Data
page 1 9

{1}

PMA P060008: FDA Summary of Safety and Effectiveness Data
page 2

# IV. WARNING AND PRECAUTIONS

The warnings and precautions can be found in the TAXUS Liberté Paclitaxel-Eluting Coronary Stent System Directions for Use (DFU).

# V. PRODUCT DESCRIPTION

The TAXUS Liberté Paclitaxel-Eluting Coronary Stent System is a device / drug combination product comprised of two regulated components: a device (Liberté Coronary Stent System) and a drug product (a formulation of paclitaxel contained in a polymer coating). The characteristics of the TAXUS Liberté stent system are described in Table 1.

Table 1: TAXUS Liberté Stent System Product Description

|   | TAXUS Liberté
Monorail® Stent Delivery System | TAXUS Liberté
Over-the-Wire Stent Delivery System  |
| --- | --- | --- |
|  Available Stent
Lengths (mm) | 8, 12, 16, 20, 24, 28, 32  |   |
|  Available Stent
Diameters (mm) | 2.50, 2.75, 3.00, 3.50, 4.00  |   |
|  Stent Material | A 316L surgical grade stainless steel Liberté® stent  |   |
|  Drug Product | A conformal coating of a polymer carrier loaded with 1μg/mm² paclitaxel in a slow release (SR)* formulation applied to the stent with a maximum nominal drug content of 229 μg on the largest stent (4.00 x 32 mm).  |   |
|  Delivery System  |   |   |
|  Working Length | 140 cm | 135 cm  |
|  Delivery System Y-Adapter Ports | Single access port to inflation lumen.
Guidewire exit port is located
approximately 25 cm from tip. Designed
for guidewire ≤0.014" | Y-Connector (Side arm for access to balloon inflation/deflation lumen. Straight arm is continuous with shaft inner lumen).
Designed for guidewire ≤0.014"  |
|  Stent Delivery | A compliant balloon, nominally 0.3 mm longer than the stent, with two radiopaque markers.  |   |
|  Balloon Inflation
Pressure | Nominal Inflation Pressure: 9 ATM (Stent Diameter 2.50 mm)
Nominal Inflation Pressure: 8 ATM (Stent Diameters 2.75 – 4.00 mm)
Rated Burst Inflation Pressure: 18 ATM (stent Diameters 2.50 – 4.00 mm)  |   |
|  Guide Catheter Inner Diameter | ≥0.058" | ≥0.066"  |
|  Catheter Shaft Outer Diameter | 1.8F proximally and 2.7F distally:
• On all balloon lengths with diameters up to 3.0 mm
• On balloon lengths 8-20 mm with diameters of 3.5 mm
• On Balloon lengths 8-16 mm with diameters of 4.0 mm
2.0F proximally and 2.7F distally:
• On balloon lengths 24-32 mm with diameters of 3.5 mm
• On balloon lengths 20-32 mm with diameters of 4.0 mm | 3.2F proximally, 2.7F distally  |

*release rate is a function of weight/weight ratio of polymer and drug, and (SR) is the formulation that was studied clinically and is used in the marketed product

{2}

PMA P060008: FDA Summary of Safety and Effectiveness Data
page 3

# A. Device Component Description

The TAXUS Liberté Coronary Stent System consists of a balloon expandable Liberté stent, coated with paclitaxel in a slow-release (8.8% formulation) triblock copolymer system, and pre-mounted on either the Liberté Monorail™ or an Over-the-Wire (OTW) delivery system. TAXUS Liberté incorporates the identical bare Liberté stent component and a similar delivery system to that of Liberté Coronary Stent System (P040016, approved April 12, 2005), and the identical TAXUS technology as the TAXUS Express² Paclitaxel-Eluting Coronary Stent (P030025, approved March 4, 2004). The system is advanced over a guide wire through the coronary vasculature to deliver and dilate the stent at the target lesion location. Following stent deployment, the delivery balloon may be inflated with additional pressure in order to optimize the stent luminal diameter and strut apposition.

Liberté stents are manufactured from 316L stainless steel. The stent design consists of a dimensionally uniform pattern of radially expandable elements that share junctions with adjacent radially expandable elements. The TAXUS Liberté stent is available in 3 stent models each designed for specific diameters as follows:

- Small Vessel (SV): 2.50 mm
- Workhorse (WH): 2.75 - 3.50 mm
- Large Vessel (LV): 4.00 mm

# B. Drug Component Description

The drug component of the TAXUS Liberté Paclitaxel-Eluting Coronary Stent System consists of paclitaxel (the active ingredient) and Translute™ polymer carrier (the inactive ingredient).

# B1. Paclitaxel

The active pharmaceutical ingredient in the TAXUS Liberté stent is paclitaxel. It is a white powder, isolated from a spectrum of Taxus species and hybrids. The chemical name of paclitaxel is: Benzenepropanoic acid, β-(benzoylamino)-α-hydroxy-6, 12b-bis(acetyloxy)-12-(benzoyloxy)-2a,3,4,4a,5,6,9,10,11,12,12a, 12b-dodecahydro-4, 11-dihydroxy-4a,8,12,12-tetramethyl-5-oxo-7,11-methano-1H-cyclodeca{3,4}benz[1,2-b]oxet-9-yl ester, [2aR-[2aα, 4β, 4aβ, 6β, 9α (αr*, βs*) 11α, 12α, 12aα,12bα]].

Paclitaxel is a diterpenoid with a characteristic taxane skeleton of 20 carbon atoms, a molecular weight of 853.91 g/mol and a molecular formula of C₄₇H₅₁NO₁₄. It is highly lipophilic, insoluble in water, but freely soluble in methanol, ethanol, chloroform, ethyl acetate, and dimethyl sulfoxide.

The chemical structure of paclitaxel is shown in Figure 1. The nominal total loaded dose of paclitaxel per nominal stent length/diameter is shown in Table 2.

{3}

![img-0.jpeg](img-0.jpeg)
Figure 1: Chemical Structure of Paclitaxel

Table 2: Nominal Total Loaded Dose of Paclitaxel per Nominal Stent Length/Diameter

|   | Nominal Stent Length (mm)/Stent Model | 8 | 12 | 16 | 20 | 24 | 28 | 32  |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  Total loaded dose Paclitaxel /Stent (μg) | SV | 38 | 58 | 77 | 97 | 116 | 136 | 155  |
|   |  WH | 55 | 83 | 112 | 140 | 168 | 196 | 224  |
|   |  LV | 61 | 88 | 114 | 141 | 176 | 203 | 229  |

Small Vessel (SV): 2.50 mm; Workhorse (WH): 2.75-3.5 mm; Large Vessel (LV): 4.00 mm

# B2. Inactive Ingredients

The only inactive ingredient in the TAXUS Liberté stent is SIBS [poly(styrene-b-isobutylene-b-styrene)], a tri-block copolymer (trade name: Translute™) that is composed of styrene and isobutylene units built on 1,3-di(2-methoxy-2-propyl)-5-tert-butylbenzene. It is a hydrophobic elastomeric copolymer with a molecular weight (Mn-number average molecular weight) of 80,000 to 130,000 g/mol and a polydispersity index of 1.0 to 2.0. The polymer is mixed with the drug paclitaxel and then applied to the stents. There is no primer or topcoat layer. The drug/polymer coating is adhered to the entire surface (i.e., luminal and abluminal) of the stent. The structural formula for the polymer is shown in Figure 2 below.

![img-1.jpeg](img-1.jpeg)
Figure 2: The Chemical Structure of Translute™ Polymer Carrier

m = repeating units of styrene
n = repeating units of isobutylene

PMA P060008: FDA Summary of Safety and Effectiveness Data

{4}

# C. Mechanism of Action

The mechanism (or mechanisms) by which a TAXUS Liberté stent affects neointimal production as seen in clinical studies has not been fully established. Paclitaxel promotes the assembly of microtubules from tubulin dimers and stabilizes microtubules by preventing depolymerization. This stability results in the inhibition of the normal dynamic reorganization of the microtubule network that is essential for vital interphase and mitotic cellular functions.

# VI. ALTERNATIVE PRACTICES OR PROCEDURES

Treatment of patients with coronary artery disease may include exercise, diet, drug therapy, percutaneous coronary interventions (such as angioplasty, and placement of bare metal stents, coated stents, and other drug eluting stents), and coronary artery bypass surgery (CABG).

# VII. MARKETING HISTORY

The TAXUS Liberté Paclitaxel-Eluting Coronary Stent System is commercially available in the following countries:

|  • Albania | • Denmark | • Kuwait | • Portugal  |
| --- | --- | --- | --- |
|  • Algeria | • Dominican Rep | • Latvia | • Qatar  |
|  • Antigua/Barbuda | • Dutch Antilles | • Lebanon | • Romania  |
|  • Argentina | • Ecuador | • Lebanon | • Russia  |
|  • Armenia | • El Salvador | • Liechtenstein | • Saudi Arabia  |
|  • Aruba | • Estonia | • Lithuania | • Scotland  |
|  • Australia | • Finland | • Luxembourg | • Serbia/Montenegro  |
|  • Austria | • France | • Macau | • Singapore  |
|  • Bahamas | • Georgia | • Macedonia | • Slovakia  |
|  • Bahrain | • Germany | • Malaysia | • Slovenia  |
|  • Bangladesh | • Great Britain | • Malta | • South Africa  |
|  • Barbados | • Greece | • Martinique | • Spain  |
|  • Belarus | • Guatemala | • Mauritania | • Sri Lanka  |
|  • Belgium | • Guyana | • Mauritius | • Suriname  |
|  • Belize | • Haiti | • Mexico | • Sweden  |
|  • Bermuda | • Honduras | • Moldavia | • Switzerland  |
|  • Bolivia | • Hong Kong | • Myanmar | • Thailand  |
|  • Bosnia | • Hungary | • Nepal | • Trinidad/Tobago  |
|  • Brazil | • Iceland | • Netherlands | • Tunisia  |
|  • Brunei | • India | • New Zealand | • Turkey  |
|  • Bulgaria | • Indonesia | • Nicaragua | • United Arab Emirates  |
|  • Chile | • Iraq | • Norway | • Uruguay  |
|  • China | • Ireland | • Oman | • Venezuela  |
|  • Colombia | • Israel | • Pakistan | • Vietnam  |
|  • Costa Rica | • Italy | • Panama | • West Bank Gaza Strip  |
|  • Croatia | • Jamaica | • Paraguay | • Yemen  |
|  • Cyprus | • Jordan | • Peru |   |
|  • Czech Republic | • Kenya | • Philippines |   |
|  • Denmark | • Korea | • Poland |   |

PMA P060008: FDA Summary of Safety and Effectiveness Data

{5}

As of August 1, 2008, approximately 900,000 TAXUS Liberté stents have been distributed outside the U.S. No products have been withdrawn from the market in any country for any reason.

## VIII. SUMMARY OF NON-CLINICAL STUDIES

A series of non-clinical laboratory studies were performed – those related to the stent and the stent delivery system [i.e. the stent on either the Monorail (MR) or Over-The-Wire (OTW) stent delivery system (SDS)], the polymer substance [i.e., polyisobutylene styrene (SIBS)], the drug substance (i.e., paclitaxel) and the finished combination product (i.e., TAXUS Liberté Paclitaxel-Eluting Coronary Stent).

## A. Biocompatibility Studies

A series of biocompatibility tests and USP Physiochemical tests were conducted to demonstrate that the components of the TAXUS Liberté Paclitaxel-Eluting Stent System (Monorail and Over-The-Wire) are non-toxic. Tests were conducted on ethylene oxide-sterilized bare metal stents, stent delivery systems, polymer films, and polymer only coated stainless steel (SS) coupons. These test articles were processed in the same manner as the finished TAXUS Liberté product, except where polymers were present (i.e., films and coupons), the drug substance, paclitaxel, was not included in the polymer coating. With the exception of the inclusion of the drug substance, the surface treatment, coating processing, amount of coating per unit area, and sterilization processes were equivalent for both the stents and coupons utilized during testing. In all of these test systems, the materials were non-reactive and produced no greater response than then the negative control employed in each test system.

All biocompatibility testing was conducted in accordance with:

- FDA Guidance for Industry and Staff: Non-Clinical Tests and Recommended Labeling for Intravascular Stents and Associated Delivery Systems, January 13, 2005
- ISO 10993-1, Biological Evaluation of Medical Devices: Evaluation and Testing (1997)

The biocompatibility studies are summarized in Table 3.

PMA P060008: FDA Summary of Safety and Effectiveness Data
page 6 14

{6}

Table 3: Biocompatibility Test Summary

|  Test Name | Test Description | Test Article and Results  |
| --- | --- | --- |
|  Cytotoxicity | ISO 10993-5: In Vitro Cytotoxicity (L929 MEM Elution) | • Stent and delivery systems: Pass (non-cytotoxic)
• Polymer-only coated 316L SS coupon: Pass (non-cytotoxic)
• Polymer-only coated stent: Pass (non-cytotoxic)  |
|  Sensitization | ISO 10993-10: Sensitization (Kligman Maximization) | • Stent and delivery systems: Pass (non-sensitizing)
• Polymer-only coated 316L SS coupon: Pass (non-sensitizing)
• Polymer-only coated stent: Pass (non-sensitizing)  |
|  Intracutaneous Reactivity | ISO 10993-10:Irritation (Injection) | • Stent and delivery systems: Pass (non-irritant)
• Polymer-only coated 316L SS coupon: Pass (non-irritant)
• Polymer-only coated stent: Pass (non-irritant)  |
|  Acute Systemic Toxicity | ISO 10993-11: Systemic Toxicity (Acute) | • Stent and delivery systems: Pass (non-toxic)
• Polymer-only coated 316L SS coupon: Pass (non-toxic)  |
|  Pyrogenicity | LAL | • Stent and delivery systems: Pass (non-pyrogenic)  |
|   |  ISO 10993-11: Systemic Toxicity (Material-Mediated Rabbit) | • Polymer-only coated 316L SS coupon: Pass (non-pyrogenic)  |
|  Hemocompatibility | Direct Contact Hemolysis | • Stent and delivery systems: Pass (non-hemolytic)
• Polymer-only coated 316L SS coupon: Pass (non-hemolytic)
• Polymer-only coated stent: Pass (non-hemolytic)  |
|   |  Lee and White Coagulation | • Stent and delivery systems: Pass (no change in coagulation time)  |
|  Implantation | 14-days (Rabbit, Intramuscular) | • Stent and delivery systems: Pass (non-toxic)  |
|   |  30-days (Rabbit, Intramuscular) | • Stent and delivery systems: Pass (non-toxic)  |
|   |  14-days Repeat Dose Subchronic Toxicity (Mouse, Intravenous) | • Stent and delivery systems: Pass (non-toxic)  |
|   |  90-Day Chronic Toxicity (Mouse, Intraperitoneal) | • Stent and delivery systems: Pass (non-toxic)  |
|  Genotoxicity | Bacterial Reverse Mutation Assay (Ames Test) | • Polymer-only coated 316L SS coupon: Pass (non-mutagenic)  |
|   |  In Vitro Chromosomal Aberration (human blood lymphocytes) | • Polymer-only cast film: Pass (non-clastogenic)  |
|   |  In Vivo Mouse Micronucleus Test | • Polymer-only cast films: Pass (non-mutagenic)  |
|  Volatile/Metal Extracts | USP Physiochemical Extracts | • Stent and delivery systems: Pass  |

PMA P060008: FDA Summary of Safety and Effectiveness Data

{7}

# A. Biocompatibility Studies, continued

Since the sponsor did not conduct the traditional battery of ISO10993 testing on the finished TAXUS Liberté stent (i.e., containing the drug substance), subchronic toxicity, thrombogenicity, and implantation of the TAXUS Liberté stent, containing all components and processing, were evaluated in porcine, rabbit, and canine models of stent-mediated vascular injury. The significant animal studies are summarized separately in Section VIII H. Animal Studies.

Complement activation testing was not conducted on the TAXUS Liberté stent. However, there is extensive clinical experience with the previous generation TAXUS Express² stent, which has the same materials and manufacturing processes, with no clinical reports to suggest that anaphylactic shock is a concern. Given that the TAXUS Liberté stent does not introduce any new materials or manufacturing processes that would suggest a new source for chemical anaphylatoxins, complement activation testing was determined not to be necessary.

The genotoxicity, carcinogenicity, and reproductive toxicity of TAXUS Liberté stents have not been evaluated. However, the genotoxicity, carcinogenicity, and reproductive toxicity of paclitaxel have been investigated in bacterial and mammalian cells in vitro and in laboratory animals in vivo. Paclitaxel was not mutagenic when tested in two gene mutation assays.

Formal carcinogenicity testing on the final TAXUS Liberté stent was not conducted. Because some paclitaxel remains on the product for an extended period of time, and the carcinogenic potential of the SIBS polymer coating had not previously been investigated, an appropriate rationale was provided to demonstrate that the carcinogenic potential of the TAXUS Liberté stent was minimal, based on the types and quantities of starting materials (including any manufacturing additives).

There is no evidence to suggest that any chemical interactions which would form a new intermediate or molecular entity occur between paclitaxel or the polymer carrier used in the TAXUS Liberté stents.

Long term biocompatibility of the drug/polymer coating of the stent in humans is unknown.

PMA P060008: FDA Summary of Safety and Effectiveness Data
page 8

{8}

# B. In Vivo Pharmacokinetics

## B1. TAXUS Liberté Paclitaxel-Eluting Coronary Stent

Boston Scientific provided a letter from the drug substance manufacturer, authorizing access to a Drug Master File (DMF) in support of this application. The drug substance manufacturer produces a generic form of the drug Taxol®, a Bristol Myers Squibb drug product that is approved for injection of multiple oncologic indications. *In vivo* animal and *in vitro* pharmacology and toxicology studies, as well as *in vivo* and human pharmacokinetic studies, were conducted on Taxol to provide information about systemic, regional and local toxicity, dose-related toxicity, distribution profiles, end-organ disposition, drug metabolism and potential drug-drug interactions.

Given that the polymer coating and drug component of the TAXUS Liberté Paclitaxel-Eluting Coronary Stent System is identical to that of the TAXUS Express² Paclitaxel-Eluting Coronary Stent System (P030025), the evaluation of the TAXUS Express² Stent System is applicable. In the clinical studies, TAXUS I, II, and III, which evaluated the TAXUS Express² Stent System, no paclitaxel levels were detected after stent implantation using an analytical method with a lower limit of quantification (LLOQ) of 10 ng/ml. These findings were confirmed in preclinical studies using multiple stents with total loaded doses above the clinically available stent system and an assay with an LLOQ of 0.03 ng/ml. Hence, in the absence of systemically detectable levels, standard pharmacokinetic parameters were not established.

## B2. Drug Interactions

Paclitaxel is metabolized in the liver via CYP2C8 to 6-alpha-hydroxypaclitaxel and via CYP3A4 to 3'-p-hydroxypaclitaxel and 6-alpha, 3'-p-dihydroxypaclitaxel. Paclitaxel is a substrate of P-glycoprotein. Because metabolism appears to play an important role in the elimination of paclitaxel, agents that could compete with or inhibit the CYP2C8 and CYP3A4 isoenzymes may increase paclitaxel plasma levels. Potential drug interactions may occur with any drug that affects these isoenzymes.

Formal drug interaction studies have not been conducted with the TAXUS Liberté Stent. Consideration should be given to the potential for both systemic and local drug interactions in the vessel wall when deciding to place a TAXUS Liberté Stent in a patient who is taking a drug with known interactions to paclitaxel or when deciding to initiate therapy with such a drug in a patient that has recently received a TAXUS Liberté stent.

PMA P060008: FDA Summary of Safety and Effectiveness Data

{9}

# C. In Vitro Engineering Testing

In vitro engineering testing, in accordance with FDA “Guidance for Industry and Staff: Non-Clinical Tests and Recommended Labeling for Intravascular Stents and Associated Delivery Systems”, January 13, 2005, was conducted on the TAXUS Liberté stent and TAXUS Liberté stent delivery systems, as applicable.

Where appropriate, in vitro engineering testing was conducted on the uncoated, bare version of the Liberté stent mounted on either the MR or OTW delivery catheters, which were approved in P040016. Additional engineering testing was also performed on the TAXUS Liberté™ stent mounted on the delivery catheters. Some testing was not repeated since there was no change to the stent substrate, and where the effect of the coating did not impact test results; for the tests that were repeated, the results demonstrated adequate performance.

The in vitro engineering studies conducted are summarized in Table 4. “Pass” denotes that the test results met product specifications and/or the recommendation in the above-referenced guidance document.

Additional testing was conducted to support the integrity of the coating on the TAXUS Liberté stent as shown in Section VIII D. Coating Characterization Testing.

PMA P060008: FDA Summary of Safety and Effectiveness Data
page 10 18

{10}

Table 4: Stent and Delivery Catheter Engineering Testing

|  Test | Description of Test | Conclusion  |
| --- | --- | --- |
|  Material Characterization  |   |   |
|  Material Composition | Chemical analysis was conducted on the stainless steel ingot provided by the material supplier to confirm both chemical analysis and inclusion/impurity content as provided by ASTM F138-00 “Standard Specification for Wrought 18 Chromium-14 Nickel-2.5 Molybdenum Stainless Steel Bar and Wire for Surgical Implants (UNS S31673).” | Pass  |
|  Stent Corrosion Resistance | TAXUS Liberté stents were tested according to ASTM F2129-01 “Standard Test Method for Conducting Cyclic Potentiodynamic Measurements to Determine the Corrosion Susceptibility of Small Implant Devices” to demonstrate that the finished stents exhibit corrosion and repassivation characteristics comparable to or better than marketed 316L bare metal coronary stents. The results indicated that the corrosion resistance met product specification. Testing included assessment of Pitting (Potentiodynamic), Crevice, Fretting, Galvanic Corrosion. | Pass  |
|  Surface Contamination | TAXUS Liberté stents were examined via SEM at 500X and 2000X to detect evidence of surface contamination or impurities on the stent material not removed by cleaning processes. Results of SEM evaluation showed no evidence of contamination above the specified limits. | Pass  |
|  Stent Dimensional and Functional Attributes  |   |   |
|  Dimensional Verification | Testing was conducted to measure and optically inspect the stent to document that stent dimensional measurements do not deviate from product specifications. All products met specifications. | Pass  |
|  Percent Surface Area | Stent surface coverage as a function of stent diameter was calculated for the TAXUS Liberté stent. The percent surface area at each diameter is determined by dividing the total artery contact surface area of the coated stent by the surface area of the artery at any given nominal stent diameter. The highest percent surface area is 25.6% found for the 2.75 mm diameters stent, i.e. the smallest diameter WH stent. | Pass  |
|  Foreshortening | The length of the stents were measured prior to and after expansion to the largest nominal diameter. All stents met product specifications. | Pass  |
|  Recoil for Balloon Expandable Stents | Testing was conducted to quantify the amount of elastic recoil for the stent and correlate this parameter to the recommended sizing procedures. Results indicated that product specifications were met. | Pass  |
|  Stent Integrity | Testing was conducted to determine whether the deformation experienced by the stent undergoing expansion above the maximum rated diameter gives rise to stent or coating fractures. No stent exhibited any strut fracture or indications of coating integrity issues when visually examined at 32X following over-expansion. | Pass  |
|  Radial Stiffness and Radial Strength | Testing was conducted to determine the ability of the TAXUS Liberté stent to resist collapse under short term and long term loads. | Pass  |
|  Compression Resistance | Testing was conducted to determine the radial resistance of the TAXUS Liberté stent to external compression. | Pass  |
|  Mechanical Properties | Ultimate tensile strength, yield strength and elongation testing was performed on tubing (pre-processing) used to fabricate the stents as well as the stent component post-processing. Ultimate tensile strength, yield strength and elongation on pre-processed tubing met product specification. Testing on stent components determined that mechanical properties were not altered by processing. | Pass  |

PMA P060008: FDA Summary of Safety and Effectiveness Data

{11}

Table 4: Stent and Delivery Catheter Engineering Testing, continued

|  Test | Description of Test | Conclusion  |
| --- | --- | --- |
|  Stent Dimensional and Functional Attributes, continued  |   |   |
|  Magnetic Resonance Imaging (MRI) Safety and Compatibility | Through non-clinical testing, the TAXUS Liberté Stent has been shown to be MR Conditional (poses no known hazards under specified conditions). The conditions are as follows:
• Field strengths of 3 Tesla or less.
• A maximum whole body averaged specific absorption rate (SAR) of 2.0 W/kg or less for a total active MR scan time (with RF exposure) of 15 minutes or less.
• Maximum Spatial Field Gradient of 70 mT/cm or less.
• A rate of change of magnetic field (dB/dt) of 60 T/s or less.

The TAXUS Liberté Stent should not migrate in this MRI environment. MR imaging within these conditions may be performed immediately following the implantation of the stent. This stent has not been evaluated to determine if it is MR Conditional beyond these conditions.

Boston Scientific conducted tests using both single stents and overlapped stents. The maximum temperature rise was less than 2.0 degrees Celsius in all cases. The effect of heating in an MRI environment for stents with simulated fractures has been tested and found to be similar to single stents. In vivo, local SAR depends on MR field strength and may be different than the estimated whole body averaged SAR, due to body composition, stent position within the imaging field, and scanner used, thereby affecting the actual temperature rise.

MR imaging quality may be compromised if the area of interest is in exactly the same area or relatively close to the position of the stent. | Pass  |
|  Radiopacity | Testing was conducted on the bare metal stent as the addition of the coating did not add or detract from the radiopacity of the stent in clinical use. | Pass  |
|  Stent Conformability | Testing was conducted to determine the conformability (axial flexibility) of the stent in its expanded state by determining the pure bending moment of the stent. All diameters of stents were tested and met specifications. | Pass  |
|  Stent Delivery System Dimensional and Functional Attributes  |   |   |
|  Delivery, Deployment and Retraction | The delivery, deployment and retraction of the TAXUS Liberté Stent System was assessed by testing system track, crossing profile, stent deployment, guidewire movement at rated burst pressure and balloon withdrawal from a stent and into the guide catheter. Testing demonstrated that the TAXUS Liberté stent system could be delivered to the target location, deployed, and retracted, thus met required product specifications. | Pass  |
|  Balloon Rated Burst Pressure (RBP) | TAXUS Liberté stent systems were tested to failure to demonstrate that the stent system met rated burst pressure. All stent systems met specification and demonstrated with 95% confidence that at least 99.9% of balloons will not experience loss of integrity at or below the rated burst pressure. | Pass  |
|  Balloon Fatigue | TAXUS Liberté stent systems across the range of stent/balloon lengths and diameters were required to complete 20 pressurization cycles to Rated Burst Pressure (RBP). The results show statistically that, with 95% confidence, 90% of the catheters will not experience balloon, shaft, or proximal/distal seal loss of integrity at or below the maximum recommended rated balloon burst pressure. | Pass  |
|  Stent Diameter vs. Balloon Pressure (Compliance Chart) | Testing was performed to determine how the diameter of a deployed balloon varies with applied balloon pressures. The stent sizing results verify that the stent systems meet the labeled compliance values. | Pass  |

PMA P060008: FDA Summary of Safety and Effectiveness Data

{12}

Table 4: Stent and Delivery Catheter Engineering Testing, continued

|  Test | Description of Test | Conclusion  |
| --- | --- | --- |
|  Stent Delivery System Dimensional and Functional Attributes, continued  |   |   |
|  Catheter Bond Strength | Representative sizes of the TAXUS Liberté stent delivery system were tested to determine the balloon bond and full unit tensile strength of the delivery system. All stent systems exceeded the minimum specifications for full unit tensile and balloon bond. | Pass  |
|  Crossing Profile | Stent systems for each diameter balloon were tested to determine the non-deployed stent/balloon profile. All samples met the product specification. | Pass  |
|  Balloon Inflation and Deflation Time | TAXUS Liberté delivery systems across the range of balloon lengths and diameters were tested for inflation/deflation times, and all stent systems met specifications. | Pass  |
|  Stent Securement for Unsheathed Stents | Testing was conducted to assess the forces required to displace a crimped TAXUS Liberté stent from the delivery systems (1) directly from the balloon catheters, (2) after tracking through a simulated tortuous artery model and (3) after tracking through a simulated tortuous artery model and then through a simulated lesion. All stent systems met the stent securement specification. | Pass  |
|  Balloon Catheter Withdrawal Resistance | Testing was conducted to verify that the TAXUS Liberté stent system can be safely withdrawn back into the recommended guide catheter sizes both before and after stent deployment. All samples met the product specification. | Pass  |
|  Dye Flow | Testing was conducted to assess dye flow while the TAXUS Liberté Stent system is positioned in a guide catheter. Results demonstrated that dye flow met product specification. | Pass  |
|  Stent, System and Coating Durability Testing  |   |   |
|  Stress Analysis (FEA) | An in-depth analysis of the stent was conducted to ensure that the stent would not fail due to fatigue under implant conditions. The FEA evaluated the structural integrity of the stent and coating when subjected to the expected loading conditions generated in coronary arteries. The analysis took into account manufacturing, delivery, implantation and clinical loading over the implant life, and predicted that fatigue failures will not occur over 400 million cycles of loading. | Pass  |
|  Fatigue Analysis | Testing was conducted on the TAXUS Liberté stent and the stent coating to demonstrate that the stent and/or coating do not exhibit failure due to fatigue. All test samples met product specification. | Pass  |
|  Accelerated Durability Testing | Accelerated in vitro testing of approximately 10 years (400 million cycles) equivalent real time was conducted to ensure that the stents, when expanded to their largest intended diameters, will not show fatigue failure during simulated 10 year life span testing. The stents were dynamically cycled over simulated vessel conditions for 400 million cycles. Following cycling, stents were visually inspected using 40X optical microscopy. No signs of strut cracking or breaking were detected. Additionally, eight stents (four coated and four with coating removed after fatigue testing) were randomly analyzed using SEM. All tested stents were free from fatigue induced surface defects, and there was no evidence of coating fatigue or corrosion. The stent met the 10 year accelerated fatigue resistance requirement of the product specification. Overlapping stents were also evaluated in an accelerated in vitro test of approximately 2 year equivalent real time and met visual requirements for coating integrity and strut damage. | Pass  |
|  Coating Durability | The coating durability of the TAXUS Liberté stent coating was assessed via a series of acute and long term in vitro and in vivo tests performed on the coated stent and the SIBS polymer. The test results demonstrate that the paclitaxel/SIBS coating displays good durability and coating integrity that will be maintained throughout the lifetime of the coated stent implant. | Pass  |

PMA P060008: FDA Summary of Safety and Effectiveness Data

{13}

# D. Coating Characterization Testing

The coating characterization testing conducted on the TAXUS stent coating is summarized in Table 5.

Table 5: Coating Characterization Testing

|  Test | Description of Test  |
| --- | --- |
|  Material Characterization  |   |
|  Materials Analysis - Polymer | Polymer components were tested to ensure conformity to raw material specifications and incoming inspection procedures.  |
|  Chemical Analysis-Polymer | Assays were conducted to determine Mw, Mn, polydispersity, monomer content, presence/formation of oligomers and free monomers.  |
|  Chemical Analysis - Drug | Drug substance was tested to ensure conformity to incoming Certificate of Analysis.  |
|  Drug Content | Assay was conducted to quantitatively determine the total amount of the drug substance, paclitaxel, on the TAXUS Liberté stent.  |
|  Dose Density | Dose per unit area was calculated.  |
|  Drug Content Along Stent Length | Testing was conducted to characterize the uniform distribution of drug along the length of the TAXUS Liberté stent.  |
|  Coating Uniformity/Reproducibility | Testing was conducted to verify the reproducibility of coating uniformity from stent to stent and batch to batch.  |
|  Impurities/Degradation Products | Assays were conducted to quantitatively determine the type and amount of impurities and degradation products on the TAXUS Liberté stent.  |
|  In vitro Elution | Assay was developed to measure the in vitro release kinetics of paclitaxel off the TAXUS Liberté stent.  |
|  Particulates | Particulate levels were evaluated for the TAXUS Liberté stent system post tracking and deployment.  |

# E. Chemistry Manufacturing and Controls (CMC) Testing

Each batch of finished devices undergoes CMC testing. This testing is summarized in Table 6. Where applicable, the test methods follow International Conference on Harmonization (ICH) Guidelines. Information to support the stability of the TAXUS Liberté Stent is summarized separately in Section VIII F - Stability below.

Table 6: CMC Release Testing

|  Test | Description of Test  |
| --- | --- |
|  Material Analysis - Polymer | The polymer was tested to ensure conformity to specifications. The polymer met specifications prior to utilization in finished goods.  |
|  Drug Identity | Assay is conducted to verify the identity of the drug substance, paclitaxel, in the TAXUS Liberté stent.  |
|  Drug Content/Impurities | Assays are conducted to quantitatively verify the amount of drug and the type and amount of impurities on the TAXUS Liberté stent.  |
|  Drug Content Uniformity | Multiple stents are assayed to verify the uniformity of the drug content between individual stents is within specifications established for the TAXUS Liberté stent.  |
|  Residual Solvents | Assay is conducted to verify that residual levels of solvents used in the manufacturing process are below acceptable limits established for the TAXUS Liberté stent.  |
|  In vitro Drug Elution | The in vitro release profile of paclitaxel is measured to verify that the drug release is within the specifications established for the TAXUS Liberté stent.  |
|  Particulates | Particulate counts are measured to verify that they remain below acceptable levels established for the TAXUS Liberté stent.  |

PMA P060008: FDA Summary of Safety and Effectiveness Data

{14}

# F. Stability

Stability studies were conducted to establish a shelf life/expiration date for the TAXUS Liberté Paclitaxel-Eluting Coronary Stent System. Testing to establish package integrity and functional testing of the stent system were conducted on aged product. Testing evaluation included drug identity, assay, degradants, *in vitro* elution, particulates, sterility, drug content uniformity, residual solvents and endotoxin. Appropriate engineering tests were also performed on aged product to ensure that the TAXUS Liberté Stent System continues to meet specification throughout its shelf life. The data generated support a shelf life of 18 months.

In addition, the stability of the drug substance and inactive polymer has been independently verified.

# G. Sterilization

The TAXUS Liberté Paclitaxel-Eluting Coronary Stent System (Monorail and Over-The-Wire) is sterilized using ethylene oxide sterilization and has been validated per AAMI/ISO 1135:1994 “Medical Devices – Validation and Routine Control of Ethylene Oxide Sterilization.”

Results obtained from the sterilization studies show that the product satisfies a minimum Sterility Assurance Level (SAL) of $10^{-6}$.

The amount of bacterial endotoxin was verified to be within the specification limit for TAXUS Liberté stent delivery systems.

# H. Animal Studies

Detailed arterial histopathology and histomorphometry can not be obtained through human clinical trials, so a series of animal studies were conducted to evaluate safety, efficacy (proof of concept), and overall product performance.

The intravascular safety and biocompatibility of paclitaxel-eluting stents were evaluated in a series of animal studies in a porcine model of stent mediated vascular injury. Studies were conducted in accordance with §21 CFR 58 (Good Laboratory Practices) with the exception of study number S5458-123131. This study was not conducted in accordance to Good Laboratory Practices (Code of Federal Regulations 21:58). The data generated was collected under independent Quality Assurance oversight in strict accordance to the study protocol and site Standard Operating Procedures (SOPs) to ensure the integrity of the data and conclusions.

The results of these tests support the safety and biocompatibility of the TAXUS Liberté Stent. Summaries of these studies are included in Table 7.

PMA P060008: FDA Summary of Safety and Effectiveness Data
page 15 23

{15}

Table 7: Summary of Major Supportive Animal Studies

|  Study # | Stent Design | Type/# of Animals | # of Stents | Timepoints | Endpoints  |
| --- | --- | --- | --- | --- | --- |
|  RVF03-082
GLP: Yes | Test Article: TAXUS Liberté 3.0x16, 3.5x16
Control: Bare Liberté 3.0x16, 3.5x16 | Domestic Swine-2 overlap pairs/animal; 55 animals total
Vessels: LAD, LCX, RCA | Test: 69 pairs
Control: 38 pairs | 28, 90, 180, 360 days | • Histologic and histomorphometric evaluations
• Evaluation of degree of re-endothelialization by SEM
• Chronic Vascular Response  |
|  RVF03-115
GLP: Yes | Test Article: TAXUS Liberté 2.0x12, 2.5x20
Control: Bare Liberté 2.25x16, 2.5x16 | Domestic Swine - 2 stents/animal; 30 animals total
Vessels: LAD, LCX, RCA, OM
10 animals/time-point | Test: 30
Control: 30 | 30, 90, 180 days | • Histologic and histomorphometric evaluation
• Evaluation of degree of re-endothelialization by SEM
• Chronic Vascular Response  |
|  S5458-123131
GLP: No | Test Article: TAXUS Liberté MR & OTW models
Control: NA | Domestic Swine-4 animals total
Vessels: LAD, LCX, RCA | Test: 13 MR & 13 OTW
Control: NA | Acute – Implant and deployment procedure | • Acute evaluations included the performance assessment of the stent delivery system and stent deployment characteristics  |
|  BJAW-0171
GLP: Yes | Test Article: TAXUS Liberté 2.75x8
Control: TAXUS Express 2.75x8 | Rabbit- single test and control stent/animal; 110 animals total
Vessels: Right and Left Iliac
10 animals/time-point | Test: 110
Control: 110 | 2, 4, 10, 20, 30, 45, 60, 90, 135, 180, and 270 days | • Residual Stent drug content and tissue drug content comparison between the Liberté and Express stent platforms.  |

PMA P060008: FDA Summary of Safety and Effectiveness Data
page 16 24

{16}

# Stent Fractures:

Stent fractures were identified in the animal study results of evaluations to support the TAXUS Liberté Stent System. In order to better understand the significance of stent fractures in the animal studies, a root cause investigation with respect to stent fractures was conducted. In a total of 6 studies conducted in domestic swine (some of which were not used in support of this application), a total of 530 Liberté and TAXUS Liberté stents were examined by Faxitron analysis and 25 stents had fractures (4.72%). In an analysis of 607 additional Liberté, TAXUS Liberté and polymer coated Liberté stents from 4 studies conducted in Yucatan mini-swine (some of which were not used in support of this application), no fractures were identified. The overall stent fracture rate for the Liberté stent platform is 2.20% (25/1137), when combining all 10 studies. The majority of stent fractures (19/25) were identified between 180 and 580 days post-implant. In addition, 16/25 (64%) stent fractures were identified in stents in an overlapping configuration. Because these animal studies were designed to evaluate cardiac function and vascular safety and not in vivo device durability, it is unclear how the study results correlate to the clinical use of the device.

In order to better understand the clinical implications of stent fracture, an analysis was conducted of the incidence of stent fracture obtained from study angiograms evaluated by the core lab in the TAXUS ATLAS study. A total of 867 patients received a TAXUS Liberté stent during the index procedure. From this group, 863 angiographic films were able to be analyzed, with a total of 932 study stents implanted in these patients. No stent fractures were detected during or immediately following the index procedure. A sub-set of patients (n=543) were randomized to have follow-up angiography at 9-months post-procedure. From this subset, 454 patients returned for an angiogram and were evaluated for stent fracture. There were a total of 497 study stents implanted in these patients, and no stent fractures were detected. An analysis of angiographic data on 568 patients with 623 stents implanted was also conducted after 9 months follow-up. In this group there was one stent fracture identified; however, the core lab was unable to determine if the fracture occurred in the study stent or a commercially available stent which was used to treat in-stent restenosis in this patient. Based on this data available from the clinical study, a stent fracture rate of 0.18% per patient (1/568) or 0.16% (1/623) was observed.

Durability of the TAXUS Liberté Stent System was also evaluated as part of the in vitro engineering testing. Pulsatile fatigue testing was conducted on TAXUS Liberté stents in a straight configuration, both single and overlapping stents, to the equivalent of 10 years (400 million cycles). In addition, TAXUS Liberté stents were evaluated for pulsatile fatigue on a curve to the equivalent of two years (80 million cycles). No stent fractures were identified in any test samples.

As a condition of approval, additional testing to evaluate pulsatile fatigue on a curve to 400 million cycles will be conducted.

PMA P060008: FDA Summary of Safety and Effectiveness Data
page 17 25

{17}

PMA P060008: FDA Summary of Safety and Effectiveness Data
page 18

# IX. OVERVIEW OF CLINICAL STUDIES

The TAXUS Liberté clinical development program consists of a series of single-arm, historically-controlled, multicenter trials designed to assess the risk/benefit profile of the polymer-controlled, paclitaxel-eluting TAXUS Liberté stent. The specific goal of the TAXUS Liberté clinical trial program is to demonstrate that the TAXUS Liberté stent performs as well as the TAXUS Express stent to safely and significantly reduce the need for revascularization compared to bare metal stents within defined target lesions. The TAXUS Liberté clinical trial program was specifically designed to start with relatively simple lesions, and progress to increasingly more complex lesions, patient populations and procedures. This overview will focus on data generated with the pivotal TAXUS ATLAS trial comparing the TAXUS Liberté stent to a historical control population of TAXUS Express patients treated in the TAXUS IV and TAXUS V de novo clinical trials. A summary of the designs of these studies is presented in Table 8.

# A. TAXUS ATLAS

TAXUS ATLAS¹ is a multi-center, single-arm trial to evaluate the safety and efficacy of the 1 μg/mm² (loaded drug/stent surface area) slow-release (SR) formulation TAXUS Liberté stent in the treatment of de novo coronary lesions compared with the TAXUS Express Paclitaxel-Eluting Coronary Stent System (case-matched historic control data derived from the TAXUS IV and TAXUS V de novo studies). A total of 871 patients at 61 clinical sites were enrolled in this study. The primary endpoint for the study was the 9-month ischemia driven target vessel revascularization (TVR) rate. Secondary endpoints included 9-month clinical assessments for all patients and analysis of angiographic and intravascular ultrasound (IVUS) parameters in a subset of patients. After the procedure, patients were treated with aspirin indefinitely and with clopidogrel or ticlopidine for at least 6 months. Follow-up through 1 year is currently available, and yearly follow-up for clinical parameters through 5 years is ongoing.

The objective of TAXUS ATLAS was to demonstrate non-inferiority of clinical and angiographic outcomes for TAXUS Liberté when compared to TAXUS Express. Therefore, the treatment group is compared to a case-matched Control group derived from TAXUS IV and TAXUS V de novo. In order to case-match the Control group, all TAXUS IV and V patients randomized to the TAXUS Express group with (1) a reference vessel diameter (RVD) by visual estimate ≥2.5 mm and ≤4.0 mm, (2) a lesion length by visual estimate ≥10 mm and ≤28 mm, and (3) receiving 1 planned study stent were included. This resulted in inclusion of all 662 patients randomized into the TAXUS Express treatment arm of TAXUS IV and 329 out of 577 patients randomized into the TAXUS Express treatment arm of TAXUS V de novo.

¹ Turco MA, Ormiston JA, Popma JJ, et al. Polymer-based, paclitaxel-eluting TAXUS Liberté stent in de novo lesions: The pivotal TAXUS ATLAS trial. J Am Coll Cardiol. 2007;49(16):1676-1683.

{18}

# B. TAXUS IV

TAXUS IV² is a randomized, double-blind, controlled pivotal Phase III U.S. study of the safety and performance of the SR formulation TAXUS® Express™ Paclitaxel-Eluting Coronary Stent System in patients with low risk, de novo coronary artery lesions. A total of 1,326 patients at 73 U.S. sites were enrolled with patients randomized 1:1 to the TAXUS Express stent or the uncoated Express™ control stent. The primary endpoint for the study was the 9-month ischemia driven TVR rate. Secondary endpoints included 9-month clinical assessments for all patients and analysis of angiographic and IVUS parameters in a subset of patients. After the procedure, patients were treated with aspirin indefinitely and with clopidogrel or ticlopidine for at least 6 months. Follow-up through 4 years is currently available, and yearly follow-up for clinical parameters through 5 years is ongoing.

# C. TAXUS V

TAXUS V de novo³ is a randomized, double-blind, controlled, expansion study of the safety and performance of the SR formulation TAXUS® Express™ Paclitaxel-Eluting Coronary Stent in de novo lesions in small and large diameter vessels, as well as long lesions. TAXUS V de novo was designed to expand the data set beyond the standard-risk, de novo coronary artery lesions studied in the pivotal TAXUS IV trial. A total of 1172 patients at 66 U.S. sites were enrolled with patients randomized 1:1 to the TAXUS Express Stent System or the uncoated Express control stent. The primary end point was the incidence rate of ischemia-driven TVR through 9 months post-index procedure. Secondary end points included the cumulative major adverse cardiac event (MACE) rate at follow-up and detailed quantitative coronary analysis (QCA) and IVUS analysis in pre-specified subgroups at 9 months. After the procedure, patients were treated with aspirin indefinitely and with clopidogrel or ticlopidine for at least 6 months. Follow-up through 2 years is currently available, and yearly follow-up for clinical parameters through 5 years is ongoing.

PMA P060008: FDA Summary of Safety and Effectiveness Data
page 19
27

---

² Stone GW, Ellis SG, Cox DA, et al. One-year clinical results with the slow-release, polymer-based, paclitaxel-eluting TAXUS stent: the TAXUS-IV trial. *Circulation*. 2004;109(16):1942-1947.
Stone GW, Ellis SG, Cox DA, et al. A polymer-based, paclitaxel-eluting stent in patients with coronary artery disease. *N Engl J Med*. 2004;350(3):221-231.
³ Stone GW, Ellis SG, Cannon L, et al. Comparison of a polymer-based paclitaxel-eluting stent with a bare metal stent in patients with complex coronary artery disease: A randomized controlled trial. *JAMA*. 2005;294(10):1215-1223.

{19}

Table 8: Comparison of TAXUS Clinical Studies

|   | TAXUS ATLAS | TAXUS IV (Pivotal) | TAXUS V de novo (Expansion)  |
| --- | --- | --- | --- |
|  Study Type | Multi-center, single-arm registry study | Prospective, multicenter, randomized, double-blind | Prospective, multicenter, randomized, double-blind  |
|  Number of Patients (ITT) | Total: 871
TAXUS Liberté
Stent: 871
Blended TAXUS IV & V de novo
historical control: 991 | Total: 1314
TAXUS Express
Stent: 662
Uncoated control: 652 | Total: 1156
TAXUS Express
Stent: 577
Uncoated control: 579  |
|  Dose Release Formulation | Slow Release (SR) (1 μg/mm²)  |   |   |
|  Lesion Criteria: Vessel Diameter (by visual estimate) | ≥2.5 mm to ≤4.0 mm | ≥2.5 mm to ≤3.75 mm | ≥2.25 mm to ≤4.0 mm  |
|  Lesion Criteria: Lesion Length (by visual estimate) | ≥10 mm and ≤28 mm | ≥10 mm and ≤28mm | ≥10 mm and ≤46 mm  |
|  Product Used | Liberté™ stent on the Maverick™ Monorail Stent Delivery Balloon Catheter | Express™ Stent on the Maverick™ Monorail Stent Delivery Balloon Catheter | Express™ Stent on the Maverick™ Monorail Stent Delivery Balloon Catheter  |
|  Antiplatelet Therapy | Aspirin indefinitely and clopidogrel or ticlopidine for 6 months  |   |   |
|  Follow-Up | 30 days: clinical
4 months: clinical
9 months: clinical (all), QCA and IVUS (subset)
1 – 5 years: clinical  |   |   |

Abbreviations: ITT=intent-to-treat; IVUS=intravascular ultrasound; QCA=quantitative coronary angiography; SR=slow-release

PMA P060008: FDA Summary of Safety and Effectiveness Data
page 20

{20}

PMA P060008: FDA Summary of Safety and Effectiveness Data
page 21
29

# X. POTENTIAL ADVERSE EFFECTS OF THE PRODUCT ON HEALTH

## A. Observed Adverse Events

Observed adverse event experience comes from three clinical studies: TAXUS ATLAS, TAXUS IV and TAXUS V de novo. Principal adverse events for these trials are shown in Table 9. Stent apposition data for TAXUS ATLAS is presented in Table 10.

Table 9: TAXUS ATLAS, TAXUS IV, and TAXUS V de novo Major Adverse Cardiac Events (MACE) From Post-Procedure to Latest Follow-Up

|   | TAXUS ATLAS
to 1 Year* |   | TAXUS IV
to 4 Years** |   | TAXUS V de novo
to 2 Years†  |   |
| --- | --- | --- | --- | --- | --- | --- |
|   |  TAXUS Liberté | TAXUS Express Control | TAXUS Express | Uncoated Control | TAXUS Express | Uncoated Control  |
|  In-Hospital MACE | 2.4%
(21/871) | 2.6%
(26/991) | 2.4%
(16/662) | 2.1%
(14/652) | 4.0%
(23/577) | 3.1%
(18/579)  |
|  30-Day MACE, overall | 2.8%
(24/870) | 3.3%
(33/987) | 2.9%
(19/662) | 2.5%
(16/652) | 5.1%
(29/569) | 3.6%
(21/576)  |
|  9-Month MACE, overall | 11.0%
(95/862) | 10.5%
(102/974) | 8.5%
(56/662) | 15.0%
(98/652) | 15.0%
(84/560) | 21.2%
(120/567)  |
|  Cardiac Death | 0.8%
(7/862) | 0.9%
(9/974) | 1.4%
(9/662) | 1.1%
(7/652) | 0.5%
(3/560) | 0.9%
(5/567)  |
|  MI | 3.7%
(32/862) | 3.9%
(38/974) | 3.5%
(23/662) | 3.7%
(24/652) | 5.4%
(30/560) | 4.6%
(26/567)  |
|  Q-Wave MI | 0.7%
(6/862) | 0.6%
(6/974) | 0.8%
(5/662) | 0.3%
(2/652) | 0.5%
(3/560) | 0.2%
(1/567)  |
|  Non-Q-Wave MI | 3.0%
(26/862) | 3.3%
(32/974) | 2.7%
(18/662) | 3.4%
(22/652) | 4.8%
(27/560) | 4.4%
(25/567)  |
|  TVR, Overall | 8.0%
(69/862) | 7.1%
(69/974) | 4.7%
(31/662) | 12.0%
(78/652) | 12.1%
(68/560) | 17.3%
(98/567)  |
|  TLR, Overall | 5.7%
(49/862) | 4.5%
(44/974) | 3.0%
(20/662) | 11.3%
(74/652) | 8.6%
(48/560) | 15.7%
(89/567)  |
|  Non-TLR, Overall | 3.2%
(28/862) | 2.7%
(26/974) | 1.7%
(11/662) | 1.1%
(7/652) | 4.8%
(27/560) | 4.2%
(24/567)  |
|  1-Year MACE | 12.5%
(106/851) | 12.3%
(118/957) | 10.6%
(70/662) | 19.8%
(129/652) | 18.9%
(105/556) | 25.9%
(146/563)  |
|  2-Year MACE | NA | NA | 14.7%
(95/645) | 25.2%
(161/640) | 22.1%
(120/542) | 29.2%
(159/544)  |
|  3-Year MACE | NA | NA | 18.9%
(116/614) | 29.0%
(178/613) | NA | NA  |
|  4-Year MACE | NA | NA | 22.1%
(133/601) | 31.5%
(190/604) | NA | NA  |
|  Cardiac Death | NA | NA | 3.0%
(18/601) | 4.0%
(24/604) | NA | NA  |
|  MI | NA | NA | 7.2%
(43/601) | 7.1%
(43/604) | NA | NA  |
|  Q-Wave MI | NA | NA | 1.3%
(8/601) | 1.0%
(6/604) | NA | NA  |
|  Non-Q-Wave MI | NA | NA | 6.0%
(36/601) | 6.5%
(39/604) | NA | NA  |

{21}

Table 9: TAXUS ATLAS, TAXUS IV, and TAXUS V de novo Major Adverse Cardiac Events (MACE) From Post-Procedure to Latest Follow-Up

|  Trial Type | TAXUS ATLAS
to 1 Year* |   | TAXUS IV
to 4 Years** |   | TAXUS V de novo
to 2 Years†  |   |
| --- | --- | --- | --- | --- | --- | --- |
|   |  TAXUS Liberté | TAXUS Express Control | TAXUS Express | Uncoated Control | TAXUS Express | Uncoated Control  |
|  TVR, Overall | NA | NA | 16.0%
(96/601) | 26.0%
(157/604) | NA | NA  |
|  TLR, Overall | NA | NA | 7.8%
(47/601) | 20.2%
(122/604) | NA | NA  |
|  Non-TLR, Overall | NA | NA | 9.0%
(54/601) | 9.3%
(56/604) | NA | NA  |
|  4-Year Stent Thrombosis | NA | NA | 1.6%
(9/579) | 1.1%
(6/569) | NA | NA  |

* After 9 months, the TAXUS ATLAS study population was reduced to a pre-specified cohort (per protocol population), which consists of all patients who received a study stent at baseline.
** After 2 years the TAXUS IV study population was reduced to a pre-specified cohort, which consists of all patients who received a study stent at baseline (Safety Population). At 4 years, the safety population is comprised of 1290 (n=649 for TAXUS, n=641 for Control).
† After 1 year the TAXUS V de novo study population was reduced to a pre-specified cohort, which consists of all patients who received a study stent at baseline (Safety population).
NA= Not Applicable; variable and/or time point not calculated.

In TAXUS ATLAS, a pre-specified subset of patients underwent IVUS evaluation of the treated lesion immediately after treatment and as a part of a scheduled angiographic evaluation at 9 months. Table 10 presents incomplete apposition rates by treatment group for the IVUS subset (n=610), based on core lab identification of one or more struts not apposed to the vessel wall, with evidence of speckling indicative of blood flow. There were no statistically significant differences between treatment groups with respect to percent of patients with incomplete apposition post-procedure (P=0.7260). However, the rate of late incomplete apposition at 9-month follow-up was significantly lower in the TAXUS Liberté group than in the TAXUS Express control group (p=0.0461). Paired IVUS analysis for both post-procedure and 9 months was available for 285 patients. In this patient group, the rates were comparable between TAXUS ATLAS and Control with regard to resolved (present post-procedure, absent at 9 months), persistent (present post-procedure and at 9 months), or late-acquired (absent post-procedure, present at 9 months) incomplete apposition.

Table 10: Frequency of Incomplete Stent Apposition

|  Incomplete Apposition (IA) | TAXUS ATLAS (N=327) | Control (TAXUS IV & V) (N=283)  |
| --- | --- | --- |
|  Early (Post-Procedure) | 8.7% (22/254) | 7.3% (14/191)  |
|  Late (9-Month) | 4.3% (9/209) | 10.1% (14/139)  |
|  Paired Data  |   |   |
|  Resolved | 3.4% (6/177) | 2.8% (3/108)  |
|  Persistent | 2.3% (4/177) | 3.7% (4/108)  |
|  Late Acquired | 1.7% (3/177) | 5.6% (6/108)  |

Resolved = # patients with BL IA and without FU IA + # patients evaluable at baseline and follow-up.
Persistent = # patients with BL IA and with FU IA + # patients evaluable at baseline and follow-up.
Late Acquired = # patients without BL IA and with FU IA + # patients evaluable at baseline and follow-up.
Incomplete Apposition variables are from assessment by IVUS core laboratory

PMA P060008: FDA Summary of Safety and Effectiveness Data

{22}

# B. Potential Adverse Events

Potential adverse events (in alphabetical order) which may be associated with the use of a coronary stent in native coronary arteries include bit are not limited to:

- Abrupt stent closure
- Acute myocardial infarction
- Allergic reaction to anti-coagulant and/or antiplatelet therapy, contrast medium, or stent materials
- Angina
- Arrhythmias, including ventricular fibrillation (VF) and ventricular tachycardia (VT)
- Arteriovenous fistula
- Cardiac tamponade
- Cardiogenic shock/Pulmonary edema
- Coronary aneurysm
- Death
- Dissection
- Emboli, distal (air, tissue or thrombotic material or material from devices(s) used in the procedure)
- Heart failure
- Hematoma
- Hemorrhage, required transfusion
- Hypotension/Hypertension
- Infection, local or systemic
- Ischemia, myocardial
- Pain, at the access site
- Perforation or Rupture of coronary artery
- Pericardial effusion
- Pseudoaneurysm, femoral
- Renal Failure
- Respiratory Failure
- Restenosis of stented segment
- Stent embolization or migration
- Stent thrombosis/occlusion
- Stroke/cerebrovascular accident /TIA
- Total occlusion of coronary artery
- Vessel spasm
- Vessel trauma requiring surgical repair or reintervention

Potential adverse events not captured above, that may be unique to the paclitaxel drug coating:

- Allergic/immunologic reaction to drug (paclitaxel or structurally-related compounds) or the polymer stent coating (or its individual components)
- Alopecia
- Anemia
- Blood product transfusion
- Gastrointestinal symptoms

PMA P060008: FDA Summary of Safety and Effectiveness Data
page 23 31

{23}

- Hematologic dyscrasia (including leukopenia, neutropenia, thrombocytopenia)
- Hepatic enzyme changes
- Histologic changes in vessel wall, including inflammation, cellular damage or necrosis
- Myalgia/Arthralgia
- Peripheral neuropathy

There may be other potential adverse events that are unforeseen at this time.

## XI. SUMMARY OF CLINICAL STUDIES

### A. TAXUS ATLAS Clinical Trial

**Objective:** The primary objective of this study was to demonstrate non-inferiority of the TAXUS Liberté Stent as compared to the TAXUS Express Stent with respect to target vessel revascularization rate (TVR) 9 months post-index procedure.

**Design:** TAXUS ATLAS was a multi-center, single-arm trial in patients at 61 sites. Eligible patients were those presenting for stenting of *de novo* lesions of a single native coronary artery (RVD of 2.5 to 4.0mm) with a target lesion of 10 to 28 mm in length and stenosis &gt;50% in diameter (visual estimates) who are candidates for percutaneous coronary intervention (PCI) or coronary artery bypass graft surgery (CABG), and had documented angina pectoris or functional ischemia.

A total of 871 intent-to-treat (ITT) patients were enrolled and evaluable in this study. The Control group (991 total ITT patients) was comprised of case-matched, historic data derived from the TAXUS IV and TAXUS V *de novo* studies. Multiple stenting was allowed for bail-out only. After the procedure, patients who received the assigned study stent (protocol population) were treated with aspirin indefinitely and clopidogrel or ticlopidine for at least 6 months.

Follow-up included clinical assessments at 1, 4, and 9 months. In addition, patients agreed to annual telephone follow-up for clinical parameters through 5 years post-procedure. After the 9-month follow-up, the study population was reduced to a pre-specified cohort, which consists of all patients who received the assigned study stent at baseline (per protocol population). Follow-up through 1 year is currently available in 856/867 (98.7%) patients.

A subset of patients was pre-assigned to have angiographic (N=543) and IVUS (N=327) follow-up at 9 months. Angiographic assessments were performed for the area of the vessel within the stent margins (in-stent) and the areas immediately 5 mm proximal and distal from the stent margins (analysis segment).

PMA P060008: FDA Summary of Safety and Effectiveness Data
page 24 32

{24}

Demographics: Patients were well-matched for baseline demographics. QCA analysis of the baseline lesion characteristics showed well-matched RVD (mean $2.75 \pm 0.50$ mm versus $2.79 \pm 0.49$ mm, $p = 0.1274$) between TAXUS Liberté and the Control group. However, minimum lumen diameter (MLD) was smaller (mean $0.85 \pm 0.36$ mm versus $0.92 \pm 0.34$ mm, $p &lt; 0.0001$), % diameter stenosis was greater (mean $69.13 \pm 11.83\%$ versus $66.76 \pm 10.80\%$, $p &lt; 0.0001$), and lesion length was longer (mean $14.76 \pm 6.61$ mm versus $13.60 \pm 6.11$ mm, $p &lt; 0.0001$) for TAXUS Liberté compared to the Control group. In addition, QCA parameters assessing baseline lesion complexity (bend, tortuosity, calcification, and presence of branch vessel disease) were significantly higher for TAXUS Liberté, resulting in a significantly higher proportion of lesions with AHA/ACC Type B2 or C lesion complexity (75.5% for TAXUS Liberté versus 61.2% for Control, $p &lt; 0.0001$). In the presence of rigid matching criteria, these differences indicate a change in clinical practice patterns from the TAXUS Express Control (enrolled 2 to 3 years ago) to the TAXUS Liberté treatment group.

Despite the higher lesion complexity, stent placement in the TAXUS Liberté group was accomplished with shorter procedure times (47.8±25.5 minutes versus 53.0±49.5 minutes, $p = 0.0052$) and a lower incidence of geographic miss during the stent placement (5.6% versus 9.2%, $p = 0.0036$).

Results: The primary endpoint data (9 months) and latest available follow-up (1 year) results are presented below (Table 11, Table 12, Table 13, Table 14, Table 15, Table 16, Figure 3, and Figure 4).

Conclusion: Overall, the primary and all prespecified secondary endpoints of TAXUS ATLAS were met. In studied patients, the TAXUS Liberté stent was associated with a non-inferior rate of 9 month TVR (primary endpoint) as compared to TAXUS Express. Quantitative coronary angiography (QCA) and IVUS analyses confirmed similar results for both groups with non-inferiority for binary restenosis rate, minimum lumen diameter (MLD), percent diameter stenosis (%DS), late loss, and % in-stent net volume obstruction between the groups. The results were achieved with similar rates and results for edge restenosis and late loss at the proximal and distal edges. There were comparable MACE rates, rates of stent thrombosis, aneurysm and incomplete stent apposition between groups, demonstrating the similarity of the TAXUS Liberté Stent to the TAXUS Express Stent.

PMA P060008: FDA Summary of Safety and Effectiveness Data
page 25

{25}

Table 11: TAXUS ATLAS Clinical Results

|   | 9 months (ITT population) |   |   |  | 1 year (latest available follow-up) (per protocol population*)  |   |   |
| --- | --- | --- | --- | --- | --- | --- | --- |
|   |  TAXUS Liberté (N=871) | Control (N=991) | P-Value |   | TAXUS Liberté (N=867) | Control (N=978) | P-Value  |
|  EFFICACY |  |  |  |   |  |  |   |
|  TVR, Overall | 8.0% (69/862) | 7.1% (69/974) | 0.4787 |   | 9.2% (78/851) | 8.9% (85/957) | 0.8334  |
|  TLR, Overall | 5.7% (49/862) | 4.5% (44/974) | 0.2865 |   | 6.1% (52/851) | 5.5% (53/957) | 0.6035  |
|  TLR, PCI | 5.3% (46/862) | 3.9% (38/974) | 0.1472 |   | 5.9% (50/851) | 5.0% (48/957) | 0.4203  |
|  TLR, CABG | 0.3% (3/862) | 0.6% (6/974) | 0.5141 |   | 0.2% (2/851) | 0.5% (5/957) | 0.4579*  |
|  Non-TLR, Overall | 3.2% (28/862) | 2.7% (26/974) | 0.4911 |   | 4.2% (36/851) | 3.8% (36/957) | 0.6111  |
|  Non-TLR, PCI | 2.8% (24/862) | 2.1% (20/974) | 0.3596 |   | 3.5% (30/851) | 2.8% (27/957) | 0.3925  |
|  Non-TLR, CABG | 0.5% (4/862) | 0.6% (6/974) | 0.7578 |   | 0.8% (7/851) | 0.9% (9/957) | 0.7894  |
|  SAFETY |  |  |  |   |  |  |   |
|  Total Death | 1.2% (10/863) | 1.8% (18/977) | 0.2570 |   | 1.3% (11/854) | 2.3% (22/961) | 0.1110  |
|  Cardiac Death or MI | 4.2% (36/862) | 4.7% (46/974) | 0.6510 |   | 4.5% (38/851) | 4.7% (45/957) | 0.8102  |
|  Cardiac Death | 0.8% (7/862) | 0.9% (9/974) | 1.0000 |   | 0.8% (7/851) | 1.0% (10/957) | 0.6248  |
|  MI | 3.7% (32/862) | 3.9% (38/974) | 0.9030 |   | 4.0% (34/851) | 3.9% (37/957) | 0.8879  |
|  Q-wave MI | 0.7% (6/862) | 0.6% (6/974) | 1.0000 |   | 0.7% (6/851) | 0.6% (6/957) | 0.8383  |
|  Non-Q-wave MI | 3.0% (26/862) | 3.3% (32/974) | 0.7901 | 3.3% (28/851) | 3.2% (31/957) | 0.9515  |   |
|  Stent Thrombosis | 0.8% (7/858) | 0.7% (7/966) | 1.0000 |  | 0.9% (8/846) | 0.7% (7/947) | 0.6318  |

*After 9 months, the TAXUS ATLAS study population was reduced to a pre-specified cohort (per protocol population), which consists of all patients who received a study stent at baseline. Patients who did not receive a study stent were not followed beyond 9 months.
P-values are not adjusted for multiple comparisons.
This trial was not adequately powered to compare the rate of low frequency events to a bare metal control group, nor was it sized to determine the rate of low frequency events with a pre-specified precision.

PMA P060008: FDA Summary of Safety and Effectiveness Data
page 26

{26}

Table 12: TAXUS ATLAS Primary Endpoint

|  Per Protocol Population | TAXUS Express Control (N=980) | TAXUS Liberté (N=867) | Difference [Upper 1-Sided 95% CI] | p-Value | Δ  |
| --- | --- | --- | --- | --- | --- |
|  9-Month TVR | 7.01% (67/956) | 7.95% (68/855) | 0.94% [2.98%] | 0.0487^{a} | 3.0%  |
|  |   |   |   |   |   |
|  Intent-to-Treat Population | TAXUS Express Control (N=991) | TAXUS Liberté (N=871) | Difference [Upper 1-Sided 95% CI] | p-Value | Δ  |
|  9-Month TVR | 7.14% (69/967) | 8.03% (69/859) | 0.90% [2.94%] | 0.0454^{a} | 3.0%  |

a P-values represent unadjusted results from non-inferiority testing.

Table 13: TAXUS ATLAS Secondary Endpoints

|  Per Protocol Population | TAXUS Express Control (N=980) | TAXUS Liberté (N=867) | Bonferroni Adjusted Upper 1-Sided 95% CI^{b} | p-Value^{c} | Δ  |
| --- | --- | --- | --- | --- | --- |
|  In-Stent Percent Diameter Stenosis | 18.80±19.44 (486) (-23.34, 100.00) | 21.04±21.40 (448) (-21.01, 100.00) | 2.24 [5.35] | 0.0006** | 6.6%  |
|  In-Stent Binary Restenosis | 8.64% (42/486) | 11.38% (51/448) | 2.74% [7.32%] | 0.0354 | 6.3%  |
|  In-Stent MLD^{a} (mm) | 2.28±0.66 (486) (0.00, 4.08) | 2.19±0.71 (448) (0.00, 4.23) | -0.09 [-0.19] | 0.0316* | -0.17 mm  |
|  In-Stent Late Loss (mm) | 0.42±0.54 (484) (-0.85, 2.71) | 0.41±0.54 (446) (-0.77, 2.55) | -0.01 [0.07] | <0.0001* | 0.18 mm  |
|  % In-Stent Net Volume Obstruction | 12.26±13.73 (139) (-27.01, 53.96) | 13.92±11.30 (209) (-8.77, 50.96) | 1.66 [4.80] | 0.0021** | 5.7%  |

*Variances equal: Pooled t statistic
**Variances unequal: Satterthwaite’s approximate t statistic
a Lower 1-Sided 95% CI is reported for In-Stent MLD.
b Bonferroni Adjusted Upper 1-sided 95% CI calculated using a 1-sided 99% CI.
c P-values represent unadjusted results from non-inferiority testing.

Table 14: TAXUS ATLAS Procedural Results

|  Procedural Outcomes | TAXUS Express Control (N=991) | TAXUS Liberté (N=871) | P-Value  |
| --- | --- | --- | --- |
|  Procedure Time | 53.0±49.5 (991) | 47.8±25.5 (870) | 0.0052  |
|  Geographic Miss | 9.2% (91/985) | 5.6% (49/869) | 0.0036  |

P-values are not adjusted for multiple comparisons.

PMA P060008: FDA Summary of Safety and Effectiveness Data

{27}

Table 15: TAXUS ATLAS 9-Month Angiographic and IVUS Results

|  Angiographic Outcomes^{a} | TAXUS Express Control (N=704) | TAXUS Liberté (N=543) | P-Value  |
| --- | --- | --- | --- |
|  MLD (mm), In-stent |  |  |   |
|  Post-Procedure | 2.70±0.45 (484) | 2.60±0.46 (446) | 0.0006  |
|  9-Month | 2.28±0.66 (486) | 2.19±0.71 (448) | 0.0541  |
|  MLD (mm), Analysis Segment |  |  |   |
|  Post-Procedure | 2.28±0.50 (484) | 2.22±0.51 (447) | 0.0615  |
|  9-Month | 2.01±0.61 (486) | 1.97±0.67 (449) | 0.3132  |
|  % DS, In-stent |  |  |   |
|  Post-Procedure | 4.43±10.29 (484) | 7.30±8.87 (446) | <0.0001  |
|  9-Month | 18.80±19.44 (486) | 21.04±21.40 (448) | 0.0934  |
|  % DS, Analysis Segment |  |  |   |
|  Post-Procedure | 19.77±9.97 (484) | 21.30±9.47 (447) | 0.0169  |
|  9-Month | 28.47±17.24 (486) | 29.15±19.06 (449) | 0.5688  |
|  Late Loss, In-stent (mm) | 0.42±0.54 (484) | 0.41±0.54 (446) | 0.6872  |
|  Late Loss, Analysis Segment (mm) | 0.27±0.46 (484) | 0.25±0.50 (447) | 0.5889  |
|  Binary Restenosis |  |  |   |
|  In-stent restenosis | 8.6% (42/486) | 11.4% (51/448) | 0.1893  |
|  Analysis segment restenosis | 12.1% (59/486) | 14.3% (64/449) | 0.3836  |
|  |   |   |   |
|  IVUS Outcomes^{b} | TAXUS Express Control (N=283) | TAXUS Liberté (N=327) | P-Value  |
|  Neointimal Volume (mm^{3}) (9 months) | 21.6±25.0 (140) | 24.9±24.1 (209) | 0.2089  |
|  % Net Volume Obstruction (9 months) | 12.3±13.7 (139) | 13.9±11.3 (209) | 0.2197  |
|  Incomplete Apposition |  |  |   |
|  Late (9 months) | 10.1% (14/139) | 4.3% (9/209) | 0.0461  |
|  Late Acquired | 5.6% (6/108) | 1.7% (3/177) | 0.0871  |

a Includes all patients in the angiographic subset.
b Includes all patients in the IVUS subset.
P-values are not adjusted for multiple comparisons.
This trial was not adequately powered to compare the rate of low frequency events to a bare metal control group, nor was it sized to determine the rate of low frequency events with a pre-specified precision.

Table 16: TAXUS ATLAS Stent Thrombosis

|  Per Protocol Population | TAXUS Express Control (N=978) | TAXUS Liberté (N=867) | P-Value  |
| --- | --- | --- | --- |
|  Protocol Defined Stent Thrombosis^{a} |  |  |   |
|  Cumulative through 1 year | 0.7% (7/947) | 0.9% (8/846) | 0.6318  |
|  Acute ST (≤24 hrs) | 0.2% (2/978) | 0.0% (0/867) | 0.5015*  |
|  Subacute ST (>24 hrs and ≤30 days) | 0.3% (3/976) | 0.2% (2/865) | 1.0000*  |
|  Late ST (>30 days and ≤12 months) | 0.2% (2/972) | 0.7% (6/863) | 0.1583*  |

PMA P060008: FDA Summary of Safety and Effectiveness Data

{28}

Table 16: TAXUS ATLAS Stent Thrombosis (continued)

|  Per Protocol Population | TAXUS Express Control (N=978) | TAXUS Liberté (N=867) | P-Value  |
| --- | --- | --- | --- |
|  ARC Definite & Probable Stent Thrombosis^{b} |  |  |   |
|  Cumulative through 1 year | 0.8% (8/947) | 1.2% (10/846) | 0.4745  |
|  Acute ST (≤24 hrs) | 0.2% (2/978) | 0.0% (0/867) | 0.5015*  |
|  Subacute ST (>24 hrs and ≤30 days) | 0.3% (3/976) | 0.2% (2/865) | 1.0000*  |
|  Late ST (>30 days and ≤12 months) | 0.3% (3/972) | 0.9% (8/863) | 0.0868  |

To be included in the calculation of stent thrombosis rate for a given interval, a patient either had to have a stent thrombosis during the interval (e.g. 31-365 days inclusive) or they had to be stent thrombosis-free during the interval with last follow-up on or after the first day of the given interval (e.g. 31 days).

a Per protocol, stent thrombosis is defined as the occurrence of any of the following:

1. Clinical presentation of acute coronary syndrome with angiographic evidence of stent thrombosis:

a) Angiographic documentation of a complete occlusion (TIMI flow 0 or 1) of a previously successfully treated artery (TIMI flow 2 to 3 immediately after stent placement and diameter stenosis ≤30%) and/or
b) Angiographic documentation of a flow-limiting thrombus within or adjacent to a previously successfully treated lesion.

2. Acute MI of the distribution of the treated vessel.
3. Death within the first 30 days (without other obvious cause) is considered a surrogate for stent thrombosis when angiography is not available.

b Academic Research Consortium (ARC) stent thrombosis is defined as follows:

1. Definite ST is considered to have occurred after intracoronary stenting by either angiographic or pathologic confirmation of stent thrombosis.
2. Probable ST is considered to have occurred after intracoronary stenting in the following cases:

a) Any unexplained death within the first 30 days following stent implantation.
b) Irrespective of the time after the index procedure, any MI which is related to documented acute ischemia in the territory of the implanted stent without angiographic confirmation of St and in the absence of any other obvious cause.

After 9 months, the TAXUS ATLAS study population was reduced to a pre-specified cohort (per protocol population), which consists of all patients who received a study stent at baseline. Patients who did not receive a study stent were not followed beyond 9 months.

Numbers are % (Count/Sample Size).

*P-values are two-sided from Fisher's exact test; p-values without * are two-sided from the Chi-square test.

P-values are not adjusted for multiple comparisons.

This trial was not adequately powered to compare the rate of low frequency events to a bare metal control group, nor was it sized to determine the rate of low frequency events with a pre-specified precision.

PMA P060008: FDA Summary of Safety and Effectiveness Data
page 29

{29}

Figure 3: TAXUS ATLAS Freedom from TVR to 1 Year, Event-Free Survival ± 1.5 SE, Per Protocol Population, All Patients (N=1845)

![img-2.jpeg](img-2.jpeg)

|   | Event Rate | Event Free | P-value*  |
| --- | --- | --- | --- |
|  TAXUS Express Control | 8.9% | 91.1% | 0.8092  |
|  TAXUS Liberté | 9.2% | 90.8%  |   |

* Log-rank P-value. P-value is not adjusted for multiple comparisons.

PMA P060008: FDA Summary of Safety and Effectiveness Data
page 30
38

{30}

Figure 4: TAXUS ATLAS Cumulative Frequency Distribution of In-Stent Late Loss by QCA, Intent-to-Treat, All Angiographic Subset Patients (N=1247)

![img-3.jpeg](img-3.jpeg)

|   | TAXUS Express Control (N=704) | TAXUS Liberté (N=543)  |
| --- | --- | --- |
|  N | 484 | 446  |
|  Median | 0.33 | 0.29  |
|  Minimum | -0.85 | -0.77  |
|  Maximum | 2.71 | 2.55  |
|  Mean | 0.42 | 0.41  |
|  SD | 0.54 | 0.54  |
|  COV | 126.64% | 133.09%  |
|  Diff (95% CI) | -0.01 [-0.08, 0.06]  |   |
|  COV = coefficient of variation.
MLD = Minimum Lumen Diameter
Late Loss = Final MLD – 9-Month MLD  |   |   |

Diabetic Patients in TAXUS ATLAS: Patients with diabetes mellitus represent a high-risk group for adverse events following percutaneous coronary intervention. The TAXUS ATLAS clinical trial did not stratify for diabetic status, and this trial was not adequately powered to study safety and effectiveness of TAXUS® Liberté™ versus TAXUS® Express® in patients with diabetes. Diabetics were further defined as medically treated (all patients treated with oral medication and/or insulin) for diabetes mellitus.

The TAXUS ATLAS clinical trial was not designed to specifically support an approval for use in diabetic patients. The following table includes patient level data from the TAXUS ATLAS clinical trial in diabetic patients.

PMA P060008: FDA Summary of Safety and Effectiveness Data

{31}

Table 17: TAXUS ATLAS 1-year Clinical Results for Medically Treated Diabetic Patients

|  Per Protocol Population^{a} | TAXUS Express Control (N=241) | TAXUS Liberté (N=220) | P-Value  |
| --- | --- | --- | --- |
|  EFFICACY |  |  |   |
|  TVR, Overall | 12.9% (30/233) | 13.5% (29/215) | 0.8480  |
|  TLR, Overall | 8.2% (19/233) | 9.3% (20/215) | 0.6668  |
|  TLR, PCI | 7.7% (18/233) | 8.8% (19/215) | 0.6693  |
|  TLR, CABG | 0.4% (1/233) | 0.5% (1/215) | 1.0000*  |
|  TVR Remote, Overall | 5.2% (12/233) | 6.0% (13/215) | 0.6797  |
|  TVR Remote, PCI | 3.0% (7/233) | 4.7% (10/215) | 0.3621  |
|  TVR Remote, CABG | 2.1% (5/233) | 1.4% (3/215) | 0.7259*  |
|  SAFETY |  |  |   |
|  Total Death | 3.0% (7/235) | 2.3% (5/218) | 0.6500  |
|  Cardiac Death or MI | 4.3% (10/233) | 5.1% (11/215) | 0.6800  |
|  Cardiac Death | 1.7% (4/233) | 0.9% (2/215) | 0.6869*  |
|  MI | 3.0% (7/233) | 5.1% (11/215) | 0.2554  |
|  Q-Wave MI | 0.0% (0/233) | 0.9% (2/215) | 0.2298*  |
|  Non-Q-Wave MI | 3.0% (7/233) | 4.2% (9/215) | 0.5007  |
|  Stent Thrombosis^{b} | 0.4% (1/229) | 1.4% (3/213) | 0.3561*  |

a After 9 months, the TAXUS ATLAS study population was reduced to a pre-specified cohort (per protocol population), which consists of all patients who received a study stent at baseline. Patients who did not receive a study stent were not followed beyond 9 months.
b Per protocol stent thrombosis.
Numbers are % (Count/Sample Size).
*P-values are two-sided from Fisher exact test; p-values without * are two-sided from the chi-square test.
P-values are not adjusted for multiple comparisons.
This trial was not adequately powered to compare the rate of low frequency events to a control group, nor was it sized to determine the rate of low frequency events with a pre-specified precision.

PMA P060008: FDA Summary of Safety and Effectiveness Data
page 32 40

{32}

# B. Results of the TAXUS IV Pivotal Clinical Trial

**Primary Objective:** To demonstrate superiority of the TAXUS® Express™ Stent compared to a matched, uncoated control stent for reduction of the TVR rate at 9 months post-index procedure.

**Design:** This was a multi-center, prospective, randomized, double-blind study in patients at 73 U.S. sites. Eligible patients were those presenting for stenting of de novo lesions in a single native coronary artery (RVD 2.5 to 3.75 mm) with a target lesion 10 to 28 mm in length and stenosis ≥ 50% in diameter using visual estimates, and who were candidates for PCI or CABG, and had documented angina pectoris or functional ischemia.

A total of 1314 patients were enrolled and evaluable in this study: 662 in the TAXUS group and 652 in the Control group. Patients were randomized to receive either a TAXUS Express Stent or an uncoated Express™ coronary stent (bare metal control). Study randomization was sub-stratified for medically-treated diabetes, reference vessel diameter, and lesion length. Multiple stenting was allowed for bailout only. After the procedure, patients were treated with aspirin indefinitely and clopidogrel or ticlopidine for 6 months.

Follow-up included clinical assessments at 1, 4, and 9 months. In addition, patients agreed to annual telephone follow-up for clinical parameters through 5 years post-procedure. Follow-up through 2 years is available in 1238/1314 (94.2%) of patients. After the 2-year follow-up, the TAXUS IV study population was reduced to a pre-specified cohort, which consists of all patients who received a study stent at baseline (Safety Population). At 4 years, the safety population is comprised of 1290 (N=649 for TAXUS, N=641 for Control) and follow-up is available for 1230 patients (95.4%).

A subset of…

---

**Source:** [https://fda.innolitics.com/device/P060008](https://fda.innolitics.com/device/P060008)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a PMA](https://innolitics.com/services/regulatory/), [a 510(k)](https://innolitics.com/services/510ks/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
