P050051 · Abbott Laboratories, Inc. · SEI · Jun 1, 2006 · Microbiology
Device Facts
Record ID
P050051
Device Name
ABBOTT ARCHITECT AUSAB
Applicant
Abbott Laboratories, Inc.
Product Code
SEI · Microbiology
Decision Date
Jun 1, 2006
Decision
APRL
Regulation
21 CFR 866.3173
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The ARCHITECT AUSAB assay is a chemiluminescent microparticle immunoassay (CMIA) for the quantitative determination of antibody to hepatitis B surface antigen (anti-HBs) in human adult and pediatric serum and plasma (dipotassium EDTA, lithium heparin, and sodium heparin) and neonatal serum. It is intended for quantitative measurement of antibody response following hepatitis B virus (HBV) vaccination, determination of HBV immune status, and for the laboratory diagnosis of HBV disease associated with HBV infection when used in conjunction with other laboratory results and clinical information.
Device Story
ARCHITECT AUSAB is a two-step chemiluminescent microparticle immunoassay (CMIA) performed on the ARCHITECT i System. Input: human serum/plasma (adult, pediatric, neonatal). Process: sample incubated with recombinant HBsAg-coated paramagnetic microparticles; anti-HBs binds to particles. After washing, acridinium-labeled rHBsAg conjugate added. After second wash, pre-trigger and trigger solutions added; chemiluminescent reaction measured as relative light units (RLUs). Output: quantitative anti-HBs concentration determined via calibration curve. Used in clinical laboratories by technicians. Results aid physicians in assessing HBV vaccination response, immune status, and diagnosing HBV infection alongside other serological markers.
Clinical Evidence
Prospective multi-center study (n=2,389) evaluated performance in increased risk, symptomatic, vaccinated, and pediatric populations. Compared against a licensed anti-HBs assay. Overall positive percent agreement 98.6% (699/709); negative percent agreement 96.7% (1165/1205). Precision established via within-run, within-day, within-lab, and between-site studies. LoB 0.82 mIU/mL, LoD 1.21 mIU/mL, LoQ 3.00 mIU/mL.
Technological Characteristics
CMIA technology. Components: rHBsAg-coated paramagnetic microparticles (E. coli recombinant, subtypes ad/ay) in TRIS buffer; acridinium-labeled rHBsAg conjugate in MES buffer. Requires ARCHITECT i System analyzer, pre-trigger (hydrogen peroxide), trigger (sodium hydroxide), and wash buffer. Automated, networked system. Storage 2-8°C.
Indications for Use
Indicated for quantitative measurement of anti-HBs in human adult, pediatric, and neonatal serum and plasma. Used to assess HBV vaccination response, determine HBV immune status, and aid in laboratory diagnosis of HBV disease in conjunction with other clinical/laboratory data.
Regulatory Classification
Identification
A hepatitis B virus (HBV) antibody assay is identified as an in vitro diagnostic device intended for prescription use in the detection of antibodies to HBV in human serum, plasma, or other matrices, and as a device that aids in the diagnosis of HBV infection in persons with signs and symptoms of hepatitis and in persons at risk for hepatitis B infection. Results from assays may be qualitative or quantitative, such as quantitative anti-HBs. In addition, results from an anti-HBc IgM (IgM antibodies to core antigen) assay indicating the presence of anti-HBc IgM are indicative of recent HBV infection. Anti-HBs (antibodies to surface antigen) assay results may be used as an aid in the determination of susceptibility to HBV infection in individuals prior to or following HBV vaccination or when vaccination status is unknown. The assay is not intended for screening of blood, plasma, cells, or tissue donors. The assay is intended as an aid in diagnosis in conjunction with clinical findings and other diagnostic procedures.
Special Controls
*Classification.* Class II (special controls). The special controls for this device are:(1) The labeling required under § 809.10(b) of this chapter must include:
(i) A prominent statement that the assay is not intended for the screening of blood, plasma, cells, or tissue donors.
(ii) A detailed explanation of the principles of operation and procedures for performing the assay.
(iii) A detailed explanation of the interpretation of results.
(iv) Limitations, which must be updated to reflect current clinical practice and disease presentation and management. The limitations must include statements that indicate:
(A) When appropriate, performance characteristics of the assay have not been established in populations of immunocompromised or immunosuppressed patients or other special populations where assay performance may be affected.
(B) Detection of HBV antibodies to a single viral antigen indicates a present or past infection with hepatitis B virus, but does not differentiate between acute, chronic, or resolved infection.
(C) The specimen types for which the device has been cleared, and that use of the assay with specimen types other than those specifically cleared for this device may result in inaccurate assay results.
(D) Diagnosis of hepatitis B infection should not be established on the basis of a single assay result but should be determined by a licensed healthcare professional in conjunction with the clinical presentation, history, and other diagnostic procedures.
(E) A non-reactive assay result may occur early during acute infection, prior to development of a host antibody response to infection, or when analyte levels are below the limit of detection of the assay.
(F) Results obtained with this assay may not be used interchangeably with results obtained with a different manufacturer's assay.
(v) For devices intended for the quantitative detection of HBV antibodies (anti-HBs), in addition to the special controls listed in paragraphs (b)(1) and (2) of this section, labeling required under § 809.10(b) of this chapter must include:
(A) The assay calibrators' traceability to a standardized reference material that FDA has determined is appropriate (
*e.g.,* a recognized consensus standard) and the limit of blank (LoB), limit of detection (LoD), limit of quantitation (LoQ), linearity, and precision to define the analytical measuring interval.(B) Performance results of the analytical sensitivity study testing a standardized reference material that FDA has determined is appropriate (
*e.g.,* a recognized consensus standard).(2) Design verification and validation must include the following:
(i) Detailed device description, including all parts that make up the device, ancillary reagents required but not provided, an explanation of the device methodology, and design of the antigen(s) and capture antibody(ies) sequences, rationale for the selected epitope(s), degree of amino acid sequence conservation of the target, and the design and composition of all primary, secondary and subsequent standards used for calibration.
(ii) Documentation and characterization (
*e.g.,* supplier, determination of identity, and stability) of all critical reagents (including description of the antigen(s) and capture antibody(ies)), and protocols for maintaining product integrity throughout its labeled shelf life.(iii) Risk analysis and management strategies, such as Failure Modes Effects Analysis and/or Hazard Analysis and Critical Control Points summaries and their impact on assay performance.
(iv) Final release criteria to be used for manufactured assay lots with appropriate evidence that lots released at the extremes of the specifications will meet the identified analytical and clinical performance characteristics as well as stability.
(v) Stability studies for reagents must include documentation of an assessment of real-time stability for multiple reagent lots using the indicated specimen types and must use acceptance criteria that ensure that analytical and clinical performance characteristics are met when stability is assigned based on the extremes of the acceptance range.
(vi) All stability protocols, including acceptance criteria.
(vii) When applicable, analytical sensitivity of the assay that is the same or better than that of other cleared or approved assays.
(viii) Analytical performance studies and results for determining the limit of blank (LoB), limit of detection (LoD), cutoff, precision (reproducibility), including lot-to-lot and/or instrument-to-instrument precision, interference, cross reactivity, carryover, hook effect, seroconversion panel testing, matrix equivalency, specimen stability, reagent stability, and cross-genotype antibody detection sensitivity, when appropriate.
(ix) For devices intended for the detection of antibodies for which a standardized reference material (that FDA has determined is appropriate) is available, the analytical sensitivity study and results testing the standardized reference material. Detailed documentation of that study and its results must be provided, including the study protocol, study report, testing results, and all statistical analyses.
(x) For devices with associated software or instrumentation, documentation must include a detailed description of device software, including software applications and hardware-based devices that incorporate software. The detailed description must include documentation of verification, validation, and hazard analysis and risk assessment activities, including an assessment of the impact of threats and vulnerabilities on device functionality and end users/patients as part of cybersecurity review.
(xi) Detailed documentation of clinical performance testing from a clinical study with an appropriate number of HBV reactive and non-reactive samples in applicable risk categories and conducted in the appropriate settings by the intended users. Performance must be analyzed relative to an FDA cleared or approved HBV antibody assay or a comparator that FDA has determined is appropriate. Additional relevant patient groups must be validated as appropriate. The samples must include prospective (sequential) samples for each identified specimen type and, as appropriate, additional characterized clinical samples. Samples must be sourced from geographically diverse areas.
(3) For any HBV antibody assay intended for quantitative detection of anti-HBV antibodies, the following special controls, in addition to those special controls listed in paragraphs (b)(1) and (2) of this section, also apply:
(i) Detailed documentation of the metrological calibration traceability hierarchy to a standardized reference material that FDA has determined is appropriate.
(ii) Detailed documentation of the following analytical performance studies conducted, as appropriate to the technology, specimen types tested, and intended use of the device, including upper and lower limits of quantitation (UloQ and LloQ, respectively), linearity using clinical samples, and an accuracy study using the recognized international standard material.
Submission Summary (Full Text)
{0}
# SUMMARY OF SAFETY AND EFFECTIVENESS
## I. GENERAL INFORMATION:
Device Generic Name: Antibody to Hepatitis B Surface Antigen (anti-HBs)
Device Trade Name: ARCHITECT® AUSAB® Reagent Kit
ARCHITECT® AUSAB® Calibrators
ARCHITECT® AUSAB® Controls
Name and Address of Applicant: Abbott Laboratories
Abbott Diagnostics Division
100 Abbott Park Road
Abbott Park, IL 60064-3500
Date of Panel Recommendation: None
Premarket Approval Application (PMA) Number: P050051
Date of Notice of Approval to the Applicant: June 1, 2006
## II. INDICATIONS FOR USE:
ARCHITECT® AUSAB® Reagent Kit
The ARCHITECT AUSAB assay is a chemiluminescent microparticle immunoassay (CMIA) for the quantitative determination of antibody to hepatitis B surface antigen (anti-HBs) in human adult and pediatric serum and plasma (dipotassium EDTA, lithium heparin, and sodium heparin) and neonatal serum. It is intended for quantitative measurement of antibody response following hepatitis B virus (HBV) vaccination, determination of HBV immune status, and for the laboratory diagnosis of HBV disease associated with HBV infection when used in conjunction with other laboratory results and clinical information.
ARCHITECT® AUSAB® Calibrators
The ARCHITECT AUSAB Calibrators are used to calibrate the ARCHITECT i System when the system is used for the quantitative determination of antibody to hepatitis B surface antigen (anti-HBs) using the ARCHITECT AUSAB Reagent Kit. The performance of the ARCHITECT AUSAB Calibrators has not been established with any other anti-HBs assays.
{1}
SUMMARY OF SAFETY AND EFFECTIVENESS DATA
## ARCHITECT® AUSAB® Controls
The ARCHITECT AUSAB Controls are used for monitoring the performance of the ARCHITECT i System when used for the quantitative determination of antibody to hepatitis B surface antigen (anti-HBs) using the ARCHITECT AUSAB Reagent Kit. The performance of the ARCHITECT AUSAB Controls has not been established with any other anti-HBs assays.
### III. CONTRAINDICATIONS: None known.
### IV. WARNINGS AND PRECAUTIONS: For *in vitro* diagnostic use only.
Warnings and precautions for ARCHITECT AUSAB Reagent Kit, ARCHITECT AUSAB Calibrators, and ARCHITECT AUSAB Controls are stated in the respective product labeling.
### V. DEVICE DESCRIPTION:
#### Kit Configurations and Components
For detection of antibodies to hepatitis B virus, the ARCHITECT AUSAB Reagent Kit is composed of the following two components:
- ARCHITECT AUSAB Microparticles: 1 or 4 Bottle(s) (4.56 mL/16.80 mL) hepatitis B surface (*E. coli*, recombinant) antigen (subtypes *ad* and *ay*) coated microparticles in TRIS buffer with protein (bovine) stabilizers (76 µM). Minimum concentration: 0.125% solids. Preservatives: antimicrobial agents.
- ARCHITECT AUSAB Conjugate: 1 or 4 Bottle(s) (5.9 mL/26.3 mL) hepatitis B surface (*E. coli*, recombinant) antigen (subtypes *ad* and *ay*) acridinium-labeled conjugate in MES buffer with protein (112.5 g/L bovine serum and 102.7 g/L human plasma) stabilizers. Minimum concentration: 0.10 ug/mL. Preservatives: antimicrobial agents.
-Page 2
9
{2}
SUMMARY OF SAFETY AND EFFECTIVENESS DATA
In addition, the following components are required for the ARCHITECT AUSAB Reagent Kit:
- ARCHITECT i System is an analyzer designed to perform fully-automated immunoassay tests based on the use of CMIA detection technology.
- ARCHITECT AUSAB Calibrators, which consist of six calibrator levels (A through F) for the calibration of the instrument.
- ARCHITECT AUSAB Controls, which consist of a negative control and a positive control.
- ARCHITECT i Pre-Trigger Solution, which contains 1.32% (w/v) hydrogen peroxide.
- ARCHITECT i Trigger Solution, which contains 0.35N sodium hydroxide.
- ARCHITECT i Wash Buffer, which contains phosphate buffered saline solution with preservative.
The ARCHITECT AUSAB Calibrator Kit is composed of:
- ARCHITECT AUSAB Calibrator A: recalcified anti-HBs negative human plasma. Preservatives: sodium azide and ProClin® 950.
- ARCHITECT AUSAB Calibrators B – F: recalcified anti-HBs positive human plasma in recalcified anti-HBs negative human plasma. Preservatives: sodium azide and ProClin 950.
The ARCHITECT AUSAB Control Kit is composed of:
- ARCHITECT AUSAB Negative Control: recalcified anti-HBs negative human plasma. Preservatives: sodium azide and ProClin® 950.
- ARCHITECT AUSAB Positive Control: recalcified anti-HBs positive human plasma in recalcified anti-HBs negative human plasma. The positive control is blue and contains Acid Blue No. 9 dye. Preservatives: sodium azide and ProClin 950.
Page 3
10
{3}
SUMMARY OF SAFETY AND EFFECTIVENESS DATA
## Assay Principle and Format
The ARCHITECT AUSAB assay is a two-step immunoassay for the quantitative determination of anti-HBs in human serum and plasma using CMIA technology with flexible assay protocols, referred to as Chemiflex®.
In the first step, sample and recombinant hepatitis B virus surface antigen (rHBsAg) coated paramagnetic microparticles are combined. Anti-HBs present in the sample binds to the rHBsAg coated microparticles. After washing, acridinium-labeled rHBsAg conjugate is added in the second step. Following another wash cycle, pre-trigger and trigger solutions are added to the reaction mixture. The resulting chemiluminescent reaction is measured as relative light units (RLUs). A direct relationship exists between the amount of anti-HBs in the sample and the RLUs detected by the ARCHITECT i optical system. The concentration of anti-HBs in the sample is determined using an active ARCHITECT AUSAB calibration curve.
## VI. ALTERNATIVE PRACTICES AND PROCEDURES
Determining the presence of HBV in patients may be achieved by using a variety of commercially available, FDA-approved, serological tests. When these test results are used in combination with a physician’s assessment and other laboratory test results, HBV immune status can be identified.
## VII. MARKETING HISTORY
ARCHITECT AUSAB, List No. 1L82, has not been marketed in any other country.
## VIII. POTENTIAL ADVERSE EFFECTS OF THE DEVICE ON HEALTH
The ARCHITECT AUSAB assay together with Calibrators, and Quality Control Materials are for in vitro diagnostic use, thus there is no direct adverse effect on the patient.
Failure of the product to perform as intended, or errors in the use of the product, may lead to a false result. This assay is used as an aid in the diagnosis of individuals with acute or chronic HBV infection and in the determination of the vaccination status of HBV infected individuals in conjunction with other HBV serological markers for the laboratory diagnosis of HBV disease associated with HBV infection. This assay can also be used as an aid in the differential diagnosis in individuals displaying signs and symptoms of hepatitis in whom etiology is unknown.
A false nonreactive result does not exclude the possibility of exposure to HBV. A nonreactive result may be due to antibody levels below the detection limits of this assay. Since this assay is used in combination with other HBV assays, a nonreactive result cannot be considered a public health risk, as the individual
Page 4
11
{4}
SUMMARY OF SAFETY AND EFFECTIVENESS DATA
would be tested with other methodologies if signs and symptoms are indicative of HBV infection.
A false reactive result would not be considered a public health risk due to the fact that an individual would be tested with other hepatitis B virus marker assays to define the clinical status of the patient.
## IX. SUMMARY OF PRECLINICAL STUDIES
Nonclinical studies were performed at Abbott Laboratories to evaluate the performance characteristics of the ARCHITECT AUSAB assay. The studies are described below.
### Tube Type Interference
A study was conducted to evaluate which anticoagulants (blood collection tube types) are acceptable for use with the ARCHITECT AUSAB assay. Sample sets of human specimens were collected in the control tube type (glass serum) and the blood collection tube types selected for evaluation. Half of the sample sets were from vaccinee donors with an anti-HBs concentration within the analytical measurement range of the assay and half of the sample sets were from anti-HBs negative donors. The blood collection tubes for the anti-HBs negative sample sets were supplemented with anti-HBs positive stock from HBV recovered donors to a target concentration within the analytical measurement range of the assay. The control and test samples were tested.
On average, the tube types evaluated showed less than a 10% difference when compared to the control tube type (glass serum). The distribution of the percent differences per tube type is listed in the following table.
| Table 1
ARCHITECT AUSAB
Tube Type Interference
Distribution of Differences | | | |
| --- | --- | --- | --- |
| Evaluation Tube Type | < 10% | ≥ 10% to ≤ 20% | > 20% |
| Glass Serum Separator | 81.8%
(36/44) | 15.9%
(7/44) | 2.3%
(1/44) |
| Plastic Serum | 79.5%
(35/44) | 18.2%
(8/44) | 2.3%
(1/44) |
| Plastic Serum Separator | 79.5%
(35/44) | 20.5%
(9/44) | 0.0%
(0/44) |
| Plastic Lithium Heparin Plasma Separator | 86.0%
(37/43) | 14.0%
(6/43) | 0.0%
(0/43) |
| Plastic Sodium Heparin | 85.7%
(36/42) | 14.3%
(6/42) | 0.0%
(0/42) |
12
{5}
Page 6
13
# SUMMARY OF SAFETY AND EFFECTIVENESS DATA
| | Distribution of %Differences | | |
| --- | --- | --- | --- |
| Evaluation Tube Type | < 10% | ≥ 10% to ≤ 20% | > 20% |
| Plastic Dipotassium EDTA | 78.0%
(32/41) | 19.5%
(8/41) | 2.4%
(1/41) |
{6}
SUMMARY OF SAFETY AND EFFECTIVENESS DATA
The data support the use of the following blood collection tube types in the ARCHITECT AUSAB assay:
**Glass tubes**
- Serum
- Serum separator
**Plastic tubes**
- Serum
- Serum separator
- Dipotassium EDTA
- Lithium heparin plasma separator
- Sodium heparin
**Interferences – Bilirubin, Hemoglobin, Total Protein, and Triglycerides**
A study was conducted to evaluate the susceptibility of the ARCHITECT AUSAB assay to potentially interfering substances based on guidance from the Clinical Laboratory Standards Institute (CLSI) document EP7-A.
Bilirubin test samples were prepared by supplementing a high negative and a low positive sample with bilirubin (conjugated and unconjugated) at ≥ 20 mg/dL (targeted to 22 mg/dL). Hemoglobin test samples were prepared by supplementing a high negative and a low positive sample with hemolysate at ≥ 500 mg/dL (targeted to 550 mg/dL). A high protein test sample (≥ 12 g/dL [targeted to 13.2 g/dL]) was prepared by concentrating an HBV negative serum pool and supplementing with anti-HBs positive stock from HBV recovered donors to yield two test samples with different analyte levels (8.00 and 12.0 mIU/mL). Triglyceride test samples were prepared by supplementing a high negative and a low positive sample with Liposyn® III at ≥ 3000 mg/dL (targeted to 3300 mg/dL). Control samples were prepared for each test sample at each analyte level. The control and test samples were tested.
At the concentrations listed below, bilirubin, hemoglobin, total protein, and triglycerides showed less than 10% interference in the ARCHITECT AUSAB assay for high negative samples targeted to 8.0 mIU/mL (concentration range: 6.0 to 9.9 mIU/mL) and low positive samples targeted to 12.0 mIU/mL (concentration range: 10.0 to 14.0 mIU/mL):
- Bilirubin (≤ 20 mg/dL)
- Hemoglobin (≤ 500 mg/dL)
- Total Protein (≤ 12 g/dL)
- Triglycerides (≤ 3000 mg/dL)
14
{7}
SUMMARY OF SAFETY AND EFFECTIVENESS DATA
## Sample Stability of Serum and Plasma
A study was conducted to evaluate the sample storage temperatures and number of freeze/thaw cycles for each blood collection tube type acceptable for use with the ARCHITECT AUSAB assay. Sample sets of human specimens were collected in each of the blood collection tube types. The samples were tested at time point 0 and after being stored at 2 to 8°C for ≥ 7 days, at room temperature (study performed at 21 to 22°C) for ≥ 3 days, and after being subjected to three freeze/thaw cycles. Specimens that were stored at the room temperature condition and 2 to 8°C condition were tested from the blood collection tubes, as on the clot, cells, or gel represents worst-case condition. The specimens that were subjected to the freeze/thaw conditions were tested off the clot.
The data demonstrate that human serum (including serum collected in serum separator tubes) or plasma collected in dipotassium EDTA, lithium heparin plasma separator, or sodium heparin tubes may be used with the ARCHITECT AUSAB assay when:
- stored at 2 to 8°C for up to 7 days
- stored at room temperature (study performed at 21°C to 22°C) for up to 3 days
- subjected to up to 3 freeze/thaw cycles
## Sample On Board Stability
A study was conducted to evaluate samples when stored on the ARCHITECT i System (on board storage) and tested using the ARCHITECT AUSAB assay. A low positive sample pool was prepared at a target concentration of 12 mIU/mL (range: 10.0 mIU/mL to 14.0 mIU/mL). The sample pool was tested using one lot of reagents, one lot of calibrator, and one lot of controls on two instruments. Time point 0 consisted of testing the sample pool immediately after pipetting. Time point 1 consisted of testing the sample pool after being stored on board the instrument for longer than 3 hours.
The data support sample storage of up to 3 hours on board the ARCHITECT i System when tested with the ARCHITECT AUSAB assay.
Page 8
15
{8}
SUMMARY OF SAFETY AND EFFECTIVENESS DATA
## Within-Laboratory Precision
A 20-day precision study was conducted to evaluate the precision performance of the ARCHITECT AUSAB assay based on guidance from the CLSI document EP5-A2. Testing was performed using three ARCHITECT AUSAB reagent lots, three calibrator lots, and one control lot on two instruments. The ARCHITECT AUSAB Negative Control, Positive Control, and six panels were assayed in replicates of two at two separate times of day for 20 testing days.
The ARCHITECT AUSAB assay demonstrated acceptable precision. The results are summarized in Table 2.
Page 9
16
{9}
SUMMARY OF SAFETY AND EFFECTIVENESS DATA
Table 2
ARCHITECT AUSAB
Within-Laboratory Precision Study - Overall Precision
| Instrument | Sample | n | Grand Mean (mIU/ mL) | Within-Run | | Within-Day | | Within-Laboratory Precision (Total) | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | | SD | %CV | SD | %CV | SD | %CV |
| 1 | Negative Control | 240 | 0.15 | 0.135 | NA | 0.136 | NA | 0.150 | NA |
| | Positive Control | 240 | 14.99 | 0.499 | 3.3 | 0.572 | 3.8 | 0.811 | 5.4 |
| | Panel 1 | 240 | 7.55 | 0.339 | 4.5 | 0.339 | 4.5 | 0.467 | 6.2 |
| | Panel 2 | 240 | 11.32 | 0.471 | 4.2 | 0.479 | 4.2 | 0.666 | 5.9 |
| | Panel 3 | 240 | 49.71 | 1.407 | 2.8 | 1.733 | 3.5 | 2.656 | 5.3 |
| | Panel 4 | 240 | 98.24 | 2.845 | 2.9 | 3.497 | 3.6 | 5.040 | 5.1 |
| | Panel 5 | 240 | 497.89 | 13.496 | 2.7 | 17.405 | 3.5 | 26.504 | 5.3 |
| | Panel 6 | 240 | 837.29 | 21.876 | 2.6 | 25.947 | 3.1 | 45.386 | 5.4 |
| 2 | Negative Control | 240 | 0.16 | 0.180 | NA | 0.180 | NA | 0.180 | NA |
| | Positive Control | 240 | 15.70 | 0.496 | 3.2 | 0.669 | 4.3 | 0.907 | 5.8 |
| | Panel 1 | 240 | 8.18 | 0.507 | 6.2 | 0.544 | 6.7 | 0.667 | 8.2 |
| | Panel 2 | 240 | 12.25 | 0.564 | 4.6 | 0.573 | 4.7 | 0.776 | 6.3 |
| | Panel 3 | 240 | 51.63 | 1.388 | 2.7 | 1.735 | 3.4 | 2.557 | 5.0 |
| | Panel 4 | 240 | 101.84 | 2.457 | 2.4 | 3.229 | 3.2 | 4.871 | 4.8 |
| | Panel 5 | 240 | 511.13 | 12.171 | 2.4 | 16.383 | 3.2 | 24.991 | 4.9 |
| | Panel 6 | 240 | 852.69 | 24.829 | 2.9 | 33.928 | 4.0 | 51.692 | 6.1 |
NA = not applicable
Page 10
{10}
SUMMARY OF SAFETY AND EFFECTIVENESS DATA
## Analytical Specificity
A study was conducted to evaluate the ARCHITECT AUSAB assay for potential cross-reactivity with specimens from individuals with medical conditions unrelated to HBV infection. Specimens with various medical conditions were obtained and tested with the ARCHITECT AUSAB assay and a comparator anti-HBs assay. The final results for each of the specimens were compared between the two assays. Specimen results that were discordant between the two assays utilized supplemental testing results.
For the medical conditions evaluated, the ARCHITECT AUSAB assay demonstrated no more potential cross-reactivity than the comparator device. The data are summarized in Table 3.
Page 11
18
{11}
SUMMARY OF SAFETY AND EFFECTIVENESS DATA
Table 3
Reactivity of the ARCHITECT AUSAB Assay in Individuals with Medical Conditions Unrelated to HBV Infection
| | Comparator anti-HBs assay | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | | Negative | | | Positive | | |
| | | ARCHITECT | | | ARCHITECT | | |
| Category | No. of Specimens | NR^{a} | GZ^{a} | R^{a} | NR^{a} | GZ^{a} | R^{a} |
| Cytomegalovirus (anti-CMV positive) | 9 | 4 | 0 | 0 | 0 | 0 | 5 |
| Epstein-Barr Virus (anti-EBV positive) | 9 | 2 | 0 | 0 | 0 | 0 | 7 |
| Hepatitis A Virus (anti-HAV positive) | 10 | 9 | 0 | 0 | 0 | 0 | 1 |
| Hepatitis C Virus (anti-HCV positive) | 10 | 6 | 0 | 0 | 0 | 0 | 4 |
| Human immunodeficiency Virus (anti-HIV-1 positive) | 10 | 6 | 0 | 0 | 0 | 0 | 4 |
| Herpes Simplex Virus (anti-HSV positive) | 6 | 4 | 0 | 0 | 0 | 0 | 2 |
| Elevated Bilirubin | 9 | 9 | 0 | 0 | 0 | 0 | 0 |
| Elevated Protein | 8 | 5 | 0 | 0 | 0 | 0 | 3 |
| Human Anti-Mouse Antibodies (HAMA) positive | 10 | 10 | 0 | 0 | 0 | 0 | 0 |
| Influenza vaccine recipient | 10 | 4 | 0 | 0 | 0 | 0 | 6 |
| Multiparous Female | 10 | 10 | 0 | 0 | 0 | 0 | 0 |
| Non-viral liver disease | 8 | 6 | 0 | 0 | 0 | 0 | 2 |
| Rheumatoid factor positive | 6 | 5 | 0 | 0 | 0 | 0 | 1 |
| Rubella antibody positive | 10 | 7 | 0 | 0 | 0 | 0 | 3 |
| Syphilis | 10 | 5 | 0 | 0 | 0 | 0 | 5 |
| Toxoplasmosis IgG positive | 9 | 5 | 0 | 0 | 0 | 0 | 4 |
| Varicella Zoster Virus (VZV) positive | 7 | 4 | 0 | 0 | 1^{b} | 0 | 2 |
| Yeast infection | 9 | 6 | 0 | 0 | 0 | 0 | 3 |
| Total | 160 | 107 | 0 | 0 | 1 | 0 | 52 |
a NR = Nonreactive, GZ = Grayzone, R = Reactive
b The final interpretation of the VZV positive specimen was anti-HBs negative when tested using the supplemental AUSAB enzyme immunoassay.
19
{12}
SUMMARY OF SAFETY AND EFFECTIVENESS DATA
## WHO Standard Linearity
A study was conducted to evaluate the linearity performance of the ARCHITECT AUSAB assay calibrated with Abbott internal calibrators using dilutions of the World Health Organization (WHO) standard in anti-HBs negative plasma.
The linear regression slopes for the WHO Standard samples up to $250\mathrm{mIU/mL}$ and $800\mathrm{mIU/mL}$ were 1.05. The ARCHITECT AUSAB assay demonstrated acceptable WHO standard linearity.
## Dilution Linearity
A study was conducted to evaluate the dilution linearity performance of the ARCHITECT AUSAB assay by preparing 8 serial dilutions of 8 positive specimen pools. The dilutions were prepared with a negative specimen pool.
Each dilution series yielded a correlation coefficient of $\geq 0.90$. The ARCHITECT AUSAB assay demonstrated acceptable dilution linearity.
## Calibration Curve Storage
A study was conducted to evaluate the acceptability of an ARCHITECT AUSAB calibration curve stored on the ARCHITECT $i$ System for a minimum of 30 days. Testing was performed using three ARCHITECT AUSAB reagent lots, three calibrator lots, and one control lot on two instruments. Each reagent lot was matched with a different calibrator lot. A single calibration per reagent lot was performed on each instrument and the calibration curve generated was stored on each instrument for the duration of the study. The ARCHITECT AUSAB Negative Control and Positive Control were assayed in replicates of two, at two times per day, for a total of 20 time points across a minimum of 31 days. The last time point was performed at least 31 days after calibration.
The data support the storage of an ARCHITECT AUSAB calibration curve on the ARCHITECT $i$ System for a minimum of 30 days.
## Neonate Serum
A study was conducted to determine performance characteristics of neonatal samples when tested using the ARCHITECT AUSAB assay. Twenty-one neonate serum (cord blood) specimens were tested using the ARCHITECT AUSAB assay and three comparator anti-HBs assays. The results for each of the specimens were compared to the consensus result. Specimen results that were discordant between the assays utilized supplemental testing results.
One discordant specimen, 5 concordant reactive specimens, and 15 concordant nonreactive specimens were observed.
Page 13
20
{13}
SUMMARY OF SAFETY AND EFFECTIVENESS DATA
## Verification of Dilution
A study was conducted to evaluate the performance of specimens diluted by the ARCHITECT $i$ system versus specimens diluted manually. Positive specimens manually diluted 1:100 with negative serum, auto diluted 1:15 with on board ARCHITECT Wash Buffer, and manually diluted 1:100 with ARCHITECT Multi-Assay Manual Diluent were tested.
The mean % recovery of the manual dilutions with negative serum versus the auto dilutions was 115.17%. The mean % recovery of the manual dilutions with negative serum versus the manual dilutions with manual diluent was 99.07%. The ARCHITECT AUSAB demonstrated acceptable auto and manual dilution performance.
## Limit of Blank, Limit of Detection, and Limit of Quantitation
A study was conducted to evaluate the Limit of Blank (LoB), Limit of Detection (LoD), and the Limit of Quantitation (LoQ) of the ARCHITECT AUSAB assay based on guidance from the CLSI document EP17-A. Testing was performed using three ARCHITECT AUSAB reagent lots, one calibrator lot, and one control lot on four instruments. Samples were prepared at concentrations ranging from 0.0 mIU/mL to 3.5 mIU/mL. Five runs with all reagent lot/instrument combinations were performed across a minimum of two days. For each run, one replicate of each sample was tested from each of two sample cups for a total of six replicates per sample.
The ARCHITECT AUSAB assay demonstrated a LoB of 0.82 mIU/mL, a LoD of 1.21 mIU/mL, and a LoQ of 3.00 mIU/mL.
## High Dose Hook Effect
A study was conducted to characterize the performance of the ARCHITECT AUSAB assay when used to test specimens containing high levels of anti-HBs that have the potential to cause a high dose hook effect. Three unique stocks of recalcified anti-HBs positive human plasma with an ARCHITECT AUSAB concentration > 100,000 mIU/mL were each serially diluted with recalcified anti-HBs negative human plasma and tested on the ARCHITECT $i$ System.
The data demonstrate that the ARCHITECT AUSAB assay is not susceptible to interference from specimens with high levels of anti-HBs.
Page 14
21
{14}
SUMMARY OF SAFETY AND EFFECTIVENESS DATA
## ARCHITECT AUSAB Microbial Challenge Characterization
A Microbial Challenge Characterization (MCC) evaluation was performed according to USP requirements for Antimicrobial Effectiveness testing for the ARCHITECT AUSAB Reagents, Calibrators, and Controls, which consisted of an Antimicrobial Effectiveness Testing (AET) evaluation and a Microbial Interference Characterization (MIC) evaluation. The MCC evaluation integrated the results from both AET and MIC, which determined that the product is adequately protected.
## Reagent, Calibrator, and Control Developmental Stability
The developmental stability is an on-going study to establish the stability (shelf-life integrity) of the ARCHITECT AUSAB Reagents, Calibrators, and Controls at the intended storage condition of 2 to 8°C and during on board storage (for reagents only).
In addition, the developmental stability includes the in-use and freeze/thaw conditions. The in-use condition for the reagents, calibrators, and controls simulates customer use over time. The freeze/thaw condition for the reagents, calibrators, and controls supports the transport stability study.
Stability testing is performed on three lots of 100-test kit reagents, three lots of 500-test kit reagents, three lots of calibrators, and three lots of controls.
The developmental stability is scheduled to continue for a maximum of 15 months (with a minimum of 6 months). Six months of stability data has been obtained for the reagents, calibrators, and controls, which supports 6 months of expiration dating.
## Reagent Transport Stability
A study was conducted to support the stability of the ARCHITECT AUSAB Reagents following simulated transport stress conditions. One 100-test kit lot and one 500-test kit lot of the ARCHITECT AUSAB Reagents were tested after being subjected to simulated transport stress.
The data support the stability of the ARCHITECT AUSAB Reagents following transport at ambient temperatures.
Page 15
22
{15}
SUMMARY OF SAFETY AND EFFECTIVENESS DATA
## Calibrator and Control Transport Stability
A study was conducted to support the stability of the ARCHITECT AUSAB Calibrators and Controls following simulated transport stress conditions. One lot each of the ARCHITECT AUSAB Calibrators and Controls was tested after being subjected to simulated transport stress.
The data support the stability of the ARCHITECT AUSAB Calibrators and Controls following transport at ambient temperatures.
## X. SUMMARY OF CLINICAL STUDIES
A multi-center study was conducted to evaluate the efficacy of the ARCHITECT AUSAB assay for the quantitative detection of anti-HBs in human adult and pediatric serum and plasma and neonatal serum as measured by precision and method comparison.
## System Reproducibility
A study was conducted to validate the precision performance of the ARCHITECT AUSAB assay based on guidance from the CLSI document EP15-A2. Three lots of ARCHITECT AUSAB Reagents, Calibrators, and Controls were tested per site. The ARCHITECT AUSAB Negative Control and Positive Control, and a high negative panel member (Panel 1, targeted to 8.00 mIU/mL) and low positive panel member (Panel 2, targeted to 12.00 mIU/mL) were tested using two replicates of each of two aliquots at two separate times per day for five days. The data are summarized in Table 4.
Page 16 23
{16}
SUMMARY OF SAFETY AND EFFECTIVENESS DATA
Table 4
ARCHITECT AUSAB
System Reproducibility (5-Day Precision Study)
| Sample | n | Grand Mean (mIU/mL) | Within-Run | | Within-Day | | Within-Laboratory Precision (Total) | | Precision with Additional Component of Between-Site | | Precision with Additional Component of Between-Lot | | Precision with Additional Components of Site and Lot (Overall) | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| Negative Control | 360 | 0.09 | 0.108 | NA | 0.110 | NA | 0.123 | NA | 0.136 | NA | 0.162 | NA | 0.168 | NA |
| Positive Control | 360 | 14.78 | 0.502 | 3.4 | 0.542 | 3.7 | 0.739 | 5.0 | 0.995 | 6.7 | 0.878 | 5.9 | 1.015 | 6.9 |
| High Negative Panel | 360 | 7.20 | 0.334 | 4.6 | 0.370 | 5.1 | 0.434 | 6.0 | 0.814 | 11.3 | 0.570 | 7.9 | 0.846 | 11.7 |
| Low Positive Panel | 360 | 10.84 | 0.453 | 4.2 | 0.497 | 4.6 | 0.602 | 5.6 | 1.066 | 9.8 | 0.826 | 7.6 | 1.136 | 10.5 |
| NA = not applicable | | | | | | | | | | | | | | |
## Method Comparison
### Clinical Performance
A prospective multi-center study was conducted to evaluate the ability of the ARCHITECT AUSAB assay to detect anti-HBs antibodies in specimens from an intended use diagnostic population.
Of the 2,389 specimens tested and analyzed in the ARCHITECT AUSAB clinical study, 1,314 specimens were from individuals at increased risk of HBV infection due to lifestyle, behavior, occupation, disease state, or known exposure event; 704 specimens were from individuals exhibiting signs and symptoms of a hepatitis infection; 211 specimens were from vaccine recipients; 40 matched pre- and post-vaccination specimens were from 20 hepatitis B vaccine recipients; and 120 surplus specimens were from a pediatric population.
The 2,389 specimens were collected from specimen collection sites or were purchased from specimen vendors. Of the 2,389 specimens, 2,018 were from the
Page 17
24
{17}
SUMMARY OF SAFETY AND EFFECTIVENESS DATA
increased risk and signs and symptoms populations. The number and percent of specimens obtained from each specimen collection site/specimen vendor for the increased risk and signs and symptoms populations are listed in Table 5 below.
Table 5
Number and Percent of Specimens by Specimen Collection Sites/Specimen Vendors for Increased Risk and Signs and Symptoms Populations
| Collection Site/Vendor | Site/Location | Group (n) | Total (n) | Percent (%) |
| --- | --- | --- | --- | --- |
| Collection Site | | | | |
| Galveston, TX | 1 | 794 | 2018 | 39.35 |
| Dallas, TX | 2 | 117 | 2018 | 5.80 |
| Miami, FL | 3 | 89 | 2018 | 4.41 |
| St. Petersburg, FL | 4 | 85 | 2018 | 4.21 |
| Chicago, IL | 5 | 166 | 2018 | 8.23 |
| Denver, CO | 6 | 123 | 2018 | 6.10 |
| Collection Vendor | | | | |
| Vendor 1: High Point, NC | 1 | 185 | 2018 | 9.17 |
| Vendor 1: Colton, CA | 2 | 118 | 2018 | 5.85 |
| Vendor 1: Plymouth, MA | 3 | 341 | 2018 | 16.90 |
A demographic summary of the increased risk and signs and symptoms populations by race/ethnic group is provided in Table 6 below.
Table 6
Demographic Summary of Increased Risk and Signs and Symptoms Populations by Race/Ethnic Group
| Race/Ethnic Group | Group (n) | Total (n) | Percent (%) |
| --- | --- | --- | --- |
| African American | 577 | 2018 | 28.59 |
| American Indian/Alaska Native | 9 | 2018 | 0.45 |
| Asian | 40 | 2018 | 1.98 |
| Caucasian | 1067 | 2018 | 52.87 |
| Hispanic | 295 | 2018 | 14.62 |
| Other | 30 | 2018 | 1.49 |
Page 18
{18}
SUMMARY OF SAFETY AND EFFECTIVENESS DATA
Of the 2,018 increased risk and signs and symptoms subjects, 1,061 (52.58%) were female and 957 (47.42%) were male. The age was not reported for three subjects. Of the remaining 2,015 subjects, the mean age was 41 years (age range: 18 to 83 years).
Each specimen was tested using the ARCHITECT AUSAB assay at one of three clinical testing sites. Each of these specimens was also tested with a comparator anti-HBs assay method at ICON Laboratories. Specimens from the increased risk subgroup, and the signs and symptoms subgroup were also tested with FDA approved assays for HBsAg/HBsAg Confirmatory (where warranted), anti-HBc, and anti-HBc IgM to determine HBV classification. Post vaccine recipient specimens were also tested with an FDA approved anti-HBc assay. During the clinical study, all comparator and HBV classification testing was performed following manufacturers' instructions.
Testing was performed to determine the agreement of the investigational ARCHITECT AUSAB assay versus the comparator Anti-HBs assay. HBV Classification was determined based on the pattern of positive and negative results of four HBV serological markers: HBsAg, anti-HBc IgM, Total anti-HBc and anti-HBs. Nineteen unique reference marker patterns are represented. The classification being used is a modification of the National Centers of Infectious Diseases (CDC) interpretation of Viral Hepatitis B Panel testing.
Page 19
26
{19}
SUMMARY OF SAFETY AND EFFECTIVENESS DATA
# Results by Specimen Classification
Following testing by a comparator anti-HBs assay and the reference HBV assays, as necessary, HBV classification was determined for the 2,018 specimens from the increased risk and signs and symptoms subgroups. HBV classification was based on the reference marker patterns presented in Table 7 below.
Table 7
HBV Classification
| HBV Reference Markers | | | | HBV Classification |
| --- | --- | --- | --- | --- |
| HBsAg | Anti-HBc IgM | Total Anti-HBc | Anti-HBs | |
| + | - | - | - | Early Acute |
| + | + | + | - | Acute |
| + | + | + | I | Chronic |
| + | - | + | + | Chronic |
| + | - | + | - | Chronic |
| + | - | - | + | Chronic |
| + | - | + | I | Chronic* |
| + | + | + | + | Late Acute/Recovering* |
| - | + | + | + | Recovering Acute |
| - | + | - | + | Recovering Acute |
| - | + | + | - | Recovering Acute/Undetectable HBsAg |
| - | + | - | - | Possible Recovering Acute/Undetectable HBsAg |
| - | + | + | I | Early Recovery* |
| - | - | + | + | Immune Due to Natural Infection |
| - | - | + | I | Distantly Immune/Anti-HBs Unknown |
| - | - | + | - | Distantly Immune/Anti-HBs Not Detected |
| - | - | - | + | Immune Due to HBV Vaccination |
| - | - | - | I | Unknown |
| - | - | - | - | Susceptible |
I = indeterminate
*Three additional serological marker patterns were observed during the clinical evaluation.
Page 20
27
{20}
SUMMARY OF SAFETY AND EFFECTIVENESS DATA
## Comparison of Results
The ARCHITECT AUSAB assay results were compared to results from a comparator anti-HBs assay for each HBV classification for the increased risk and signs and symptoms subgroups. The comparison of ARCHITECT AUSAB results by HBV classification is presented in Table 8.
Page 21
28
{21}
Table 8
ARCHITECT AUSAB Results versus Comparator Anti-HBs Assay Results
Comparison of Results for Increased Risk and Signs and Symptoms Subgroups
by HBV Classification
| HBV Classification | Comparator anti-HBs Interpretation | | | | | | | | | | | | | | | | | | Total | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | Positive | | | | | | Indeterminate | | | | | | Negative | | | | | | | |
| | ARCHITECT AUSAB Interpretation | | | | | | ARCHITECT AUSAB Interpretation | | | | | | ARCHITECT AUSAB Interpretation | | | | | | | |
| | Reactive | | Grayzone | | Nonreactive | | Reactive | | Grayzone | | Nonreactive | | Reactive | | Grayzone | | Nonreactive | | | |
| | N | % | N | % | N | % | N | % | N | % | N | % | N | % | N | % | N | % | N | % |
| Early Acute | 0 | 0.00 | 0 | 0.00 | 0 | 0.00 | 0 | 0.00 | 0 | 0.00 | 0 | 0.00 | 0 | 0.00 | 0 | 0.00 | 2 | 0.10 | 2 | 0.10 |
| Acute | 0 | 0.00 | 0 | 0.00 | 0 | 0.00 | 0 | 0.00 | 0 | 0.00 | 0 | 0.00 | 0 | 0.00 | 0 | 0.00 | 5 | 0.25 | 5 | 0.25 |
| Late Acute/Recovering | 0 | 0.00 | 0 | 0.00 | 1 | 0.05 | 0 | 0.00 | 0 | 0.00 | 0 | 0.00 | 0 | 0.00 | 0 | 0.00 | 0 | 0.00 | 1 | 0.05 |
| Early Recovery | 0 | 0.00 | 0 | 0.00 | 0 | 0.00 | 2 | 0.10 | 0 | 0.00 | 1 | 0.05 | 0 | 0.00 | 0 | 0.00 | 0 | 0.00 | 3 | 0.15 |
| Recovering Acute | 4 | 0.20 | 0 | 0.00 | 0 | 0.00 | 0 | 0.00 | 0 | 0.00 | 0 | 0.00 | 0 | 0.00 | 0 | 0.00 | 0 | 0.00 | 4 | 0.20 |
| Chronic | 3 | 0.15 | 0 | 0.00 | 0 | 0.00 | 2 | 0.10 | 0 | 0.00 | 2 | 0.10 | 0 | 0.00 | 0 | 0.00 | 35 | 1.73 | 42 | 2.08 |
| Immune Due to Natural Infection | 192 | 9.51 | 1 | 0.05 | 0 | 0.00 | 0 | 0.00 | 0 | 0.00 | 0 | 0.00 | 0 | 0.00 | 0 | 0.00 | 0 | 0.00 | 193 | 9.56 |
| Distantly Immune/Anti-HBs Unknown | 0 | 0.00 | 0 | 0.00 | 0 | 0.00 | 21 | 1.04 | 6 | 0.30 | 4 | 0.20 | 0 | 0.00 | 0 | 0.00 | 0 | 0.00 | 31 | 1.54 |
| Distantly Immune/Anti-HBs Not Detected | 0 | 0.00 | 0 | 0.00 | 0 | 0.00 | 0 | 0.00 | 0 | 0.00 | 0 | 0.00 | 13 | 0.64 | 11 | 0.55 | 83 | 4.11 | 107 | 5.30 |
| Immune Due to HBV Vaccination | 500 | 24.78 | 7 | 0.35 | 1 | 0.05 | 0 | 0.00 | 0 | 0.00 | 0 | 0.00 | 0 | 0.00 | 0 | 0.00 | 0 | 0.00 | 508 | 25.17 |
| Unknown | 0 | 0.00 | 0 | 0.00 | 0 | 0.00 | 34 | 1.68 | 17 | 0.84 | 15 | 0.74 | 0 | 0.00 | 0 | 0.00 | 0 | 0.00 | 66 | 3.27 |
| Susceptible | 0 | 0.00 | 0 | 0.00 | 0 | 0.00 | 0 | 0.00 | 0 | 0.00 | 0 | 0.00 | 12 | 0.59 | 4 | 0.20 | 1040 | 51.54 | 1056 | 52.33 |
| Total | 699 | 34.64 | 8 | 0.40 | 2 | 0.10 | 59 | 2.92 | 23 | 1.14 | 22 | 1.09 | 25 | 1.24 | 15 | 0.74 | 1165 | 57.73 | 2018 | 100.00 |
Page 22
{22}
SUMMARY OF SAFETY AND EFFECTIVENESS DATA
## Percent Agreement
The negative percent agreement and positive percent agreement between the ARCHITECT AUSAB assay result and a comparator anti-HBs assay result, and their corresponding 95% confidence intervals were calculated for the increased risk and signs and symptoms subgroups. The negative percent agreement for the increased risk and signs and symptoms subgroups combined was 96.68% (1,165/1,205) with a 95% confidence interval of 95.51% to 97.62%. The positive percent agreement for the increased risk and signs and symptoms subgroups combined was 98.59% (699/709) with a 95% confidence interval of 97.42% to 99.32%.
The negative percent agreement and positive percent agreement results for the increased risk and signs and symptoms subgroups by HBV classification are presented in Table 9.
Page 23 30
{23}
Page 24
31
# SUMMARY OF SAFETY AND EFFECTIVENESS DATA
## Table 9
### ARCHITECT AUSAB Results versus Comparator Anti-HBs Assay Results Percent Agreement for Increased Risk and Signs and Symptoms Subgroups by HBV Classification
| HBV Classification | Positive Percent Agreement | 95% Exact Confidence Interval | Negative Percent Agreement | 95% Exact Confidence Interval |
| --- | --- | --- | --- | --- |
| Early Acute | NA (0/0) | NA | 100.00 (2/2) | 15.81-100.00 |
| Acute | NA (0/0) | NA | 100.00 (5/5) | 47.82-100.00 |
| Late Acute/Recovering | 0.00 (0/1) | 0.00-97.50 | NA (0/0) | NA |
| Early Recovery | NA (0/0) | NA | NA (0/0) | NA |
| Recovering Acute | 100.00 (4/4) | 39.76-100.00 | NA (0/0) | NA |
| Chronic | 100.00 (3/3) | 29.24-100.00 | 100.00 (35/35) | 90.00-100.00 |
| Immune Due to Natural Infection | 99.48 (192/193) | 97.15-99.99 | NA (0/0) | NA |
| Distantly Immune/Anti-HBs Unknown | NA (0/0) | NA | NA (0/0) | NA |
| Distantly Immune/Anti-HBs Not Detected | NA (0/0) | NA | 77.57 (83/107) | 68.49-85.07 |
| Immune Due to HBV Vaccination | 98.43 (500/508) | 96.92-99.32 | NA (0/0) | NA |
| Unknown | NA (0/0) | NA | NA (0/0) | NA |
| Susceptible | NA (0/0) | NA | 98.48 (1040/1056) | 97.55-99.13 |
| TOTAL | 98.59 (699/709) | 97.42-99.32 | 96.68 (1165/1205) | 95.51-97.62 |
{24}
SUMMARY OF SAFETY AND EFFECTIVENESS DATA
The negative percent agreement for the vaccine recipient subgroup was 84.78% (39/46) with a 95% confidence interval of 71.13% to 93.66%. The positive percent agreement for the vaccine recipient subgroup was 100.00% (146/146) with a 95% confidence interval of 97.51% to 100.00%. This is presented in Table 10 below.
Table 10
ARCHITECT AUSAB Results versus Comparator Anti-HBs Assay Results
Percent Agreement for Vaccine Recipients Subgroup
n=211
| | Comparator Anti-HBs Interpretation | | |
| --- | --- | --- | --- |
| ARCHITECT
AUSAB
Interpretation | Positive | Indeterminate | Negative |
| Reactive | 146
(A) | 9
(B) | 4
(C) |
| Grayzone | 0
(D) | 5
(E) | 3
(F) |
| Nonreactive | 0
(G) | 5
(H) | 39
(I) |
Negative Percent Agreement = (1 / (C+F+I)) x 100 = 84.78 %
Positive Percent Agreement = (A / (A+D+G)) x 100 = 100.00%
95% Confidence Interval for Negative Percent Agreement = (71.13 %, 93.66 %)
95% Confidence Interval for Positive Percent Agreement = (97.51 %, 100.00%)
Page 25 32
{25}
Page 26 33
# SUMMARY OF SAFETY AND EFFECTIVENESS DATA
The negative percent agreement for the pre- and post- vaccine recipients subgroup combined was 100.00% (22/22) with a 95% confidence interval of 84.56% to 100.00%. The positive percent agreement for the pre- and post- vaccine recipients subgroup combined was 100.00% (18/18) with a 95% confidence interval of 81.47% to 100.00%. This is presented in Table 11 below.
Table 11
ARCHITECT AUSAB Results versus Comparator Anti-HBs Assay Results
Percent Agreement for Pre- and Post- Vaccine Recipients Subgroup
| Vaccination Status | Positive Percent Agreement | 95% Exact Confidence Interval | Negative Percent Agreement | 95% Exact Confidence Interval |
| --- | --- | --- | --- | --- |
| Pre-Vaccination | NA (0/0) | NA | 100.00 (20/20) | 83.16-100.00 |
| Post-Vaccination | 100.00 (18/18) | 81.47-100.00 | 100.00 (2/2) | 15.81-100.00 |
| Combined | 100.00 (18/18) | 81.47-100.00 | 100.00 (22/22) | 84.56-100.00 |
{26}
SUMMARY OF SAFETY AND EFFECTIVENESS DATA
The negative percent agreement for the pediatric subgroup was: 92.68% (38/41) with a 95% confidence interval of 80.08% to 98.46%. The positive percent agreement for the pediatric subgroup was 93.44% (57/61) with a 95% confidence interval of 84.05% to 98.18%. This is presented in Table 12 below.
Table 12
ARCHITECT AUSAB Results versus Comparator Anti-HBs Assay Results
Percent Agreement for Pediatric Subgroup
n=120
| | Comparator Anti-HBs Interpretation | | |
| --- | --- | --- | --- |
| ARCHITECT
AUSAB
Interpretation | Positive | Indeterminate | Negative |
| Reactive | 57
(A) | 8
(B) | 1
(C) |
| Grayzone | 3
(D) | 4
(E) | 2
(F) |
| Nonreactive | 1
(G) | 6
(H) | 38
(I) |
Negative Percent Agreement = (I / (C+F+I)) x 100 = 92.68 %
Positive Percent Agreement = (A / (A+D+G)) x 100 = 93.44 %
95% Confidence Interval for Negative Percent Agreement = (80.08 %, 98.46 %)
95% Confidence Interval for Positive Percent Agreement = (84.05 %, 98.18%)
Page 27 34
{27}
SUMMARY OF SAFETY AND EFFECTIVENESS DATA
# Expected Results (Increased Risk Population and Pediatric Populations)
In addition to precision and method comparison, demographic summaries of the expected results from the increased risk and pediatric populations are provided.
# Increased Risk Population
Of the 2,389 specimens analyzed in the ARCHITECT AUSAB clinical investigation, a total of 1,314 specimens were from the individuals with increased risk of HBV infection. All subjects were at risk of HBV infection due to lifestyle, behavior, occupation, or known exposure event but were asymptomatic and reported no current signs or symptoms of hepatitis.
The 1,314 increased risk specimens were collected from specimen collection sites or were purchased from specimen vendors. The number and percent of specimens obtained from each collection site/vendor are listed in Table 13 below.
Table 13 Number and Percent of Specimens by Specimen Collection Sites/Specimen Vendors for Increased Risk Population
| Collection Site/Vendor | Site/Location | Group (n) | Total (n) | Percent (%) |
| --- | --- | --- | --- | --- |
| Collection Site | | | | |
| Galveston, TX | 1 | 743 | 1314 | 56.54 |
| Dallas, TX | 2 | 59 | 1314 | 4.49 |
| Miami, FL | 3 | 52 | 1314 | 3.96 |
| St. Petersburg, FL | 4 | 56 | 1314 | 4.26 |
| Chicago, IL | 5 | 8 | 1314 | 0.61 |
| Denver, CO | 6 | 36 | 1314 | 2.74 |
| Collection Vendor | | | | |
| Vendor 1: High Point, NC | 1 | 185 | 1314 | 14.08 |
| Vendor 1: Colton, CA | 2 | 76 | 1314 | 5.78 |
| Vendor 1: Plymouth, MA | 3 | 99 | 1314 | 7.53 |
{28}
SUMMARY OF SAFETY AND EFFECTIVENESS DATA
A demographic summary of the increased risk specimen population by race/ethnic group is provided in Table 14 below.
Table 14 Demographic Summary of Increased Risk Population by Race/Ethnic Group
| Race/Ethnic Group | Group (n) | Total (n) | Percent (%) |
| --- | --- | --- | --- |
| African American | 477 | 1314 | 36.30 |
| American Indian/Alaska Native | 6 | 1314 | 0.46 |
| Asian | 19 | 1314 | 1.45 |
| Caucasian | 625 | 1314 | 47.56 |
| Hispanic | 167 | 1314 | 12.71 |
| Other | 20 | 1314 | 1.52 |
Of the 1,314 subjects, 816 (62.10%) were female and 498 (37.90%) were male. The age was not reported for three subjects. Of the remaining 1,311 subjects, the mean age was 40 years (age range: 18 to 75 years).
The ARCHITECT AUSAB assay was reactive in 535 (40.72%) of the individuals in the increased risk population. The number and percent of ARCHITECT AUSAB reactive results observed at each collection location are presented in Table 15.
Page 29 36
{29}
SUMMARY OF SAFETY AND EFFECTIVENESS DATA
Table 15
ARCHITECT AUSAB Percent Reactive Results by Collection Location for the Increased Risk Population
| Collection Site/Vendor | Site/Location | Reactive (n) | Total (n) | Percent Reactive (%) |
| --- | --- | --- | --- | --- |
| Collection Site | | | | |
| Galveston, TX | 1 | 276 | 743 | 37.15 |
| Dallas, TX | 2 | 16 | 59 | 27.12 |
| Miami, FL | 3 | 33 | 52 | 63.46 |
| St. Petersburg, FL | 4 | 16 | 56 | 28.57 |
| Chicago, IL | 5 | 3 | 8 | 37.50 |
| Denver, CO | 6 | 24 | 36 | 66.67 |
| Collection Vendor | | | | |
| High Point, NC | 1 | 103 | 185 | 55.68 |
| Colton, CA | 2 | 35 | 76 | 46.05 |
| Plymouth, MA | 3 | 29 | 99 | 29.29 |
| Total | | 535 | 1314 | 40.72 |
Page 30 37
{30}
Page 31
38
# SUMMARY OF SAFETY AND EFFECTIVENESS DATA
The distribution of ARCHITECT AUSAB reactive, grayzone, and nonreactive results by age range and gender is presented in Table 16 below.
## Table 16
ARCHITECT AUSAB Results by Age Range and Gender for Individuals at Increased Risk of HBV Infection
| Age Group (Years)* | Gender | Reactive N (%) | Grayzone N (%) | Nonreactive N (%) | Total |
| --- | --- | --- | --- | --- | --- |
| 10 to 19 | Female | 9 (64.29%) | 1 (7.14%) | 4 (28.57%) | 14 |
| | Male | 7 (63.64%) | 0 (0.00%) | 4 (36.36%) | 11 |
| 20 to 29 | Female | 82 (44.57%) | 2 (1.09%) | 100 (54.35%) | 184 |
| | Male | 35 (36.08%) | 1 (1.03%) | 61 (62.89%) | 97 |
| 30 to 39 | Female | 78 (42.39%) | 1 (0.54%) | 105 (57.07%) | 184 |
| | Male | 33 (30.84%) | 0 (0.00%) | 74 (69.16%) | 107 |
| 40 to 49 | Female | 106 (42.23%) | 5 (1.99%) | 140 (55.78%) | 251 |
| | Male | 55 (34.59%) | 1 (0.63%) | 103 (64.78%) | 159 |
| 50 to 59 | Female | 68 (49.64%) | 3 (2.19%) | 66 (48.18%) | 137 |
| | Male | 30 (27.52%) | 4 (3.67%) | 75 (68.81%) | 109 |
| 60 to 69 | Female | 23 (65.71%) | 1 (2.86%) | 11 (31.43%) | 35 |
| | Male | 3 (25.00%) | 0 (0.00%) | 9 (75.00%) | 12 |
| 70 to 79 | Female | 3 (37.50%) | 1 (12.50%) | 4 (50.00%) | 8 |
| | Male | 2 (66.67%) | 0 (0.00%) | 1 (33.33%) | 3 |
| Total | | 534 (40.73%) | 20 (1.53%) | 757 (57.74%) | 1311 |
* Age was not reported for three subjects.
{31}
SUMMARY OF SAFETY AND EFFECTIVENESS DATA
# Pediatric Population
Of the 2,389 specimens tested and analyzed in the ARCHITECT AUSAB clinical investigation, 120 specimens were from a pediatric population. Specimens were obtained from ProMedDx, LLC, which collected the specimens from a collection site located in Fall River, MA. The specimens were obtained from children ages greater than 1 month to 18 years.
The distribution of ARCHITECT AUSAB reactive, grayzone, and nonreactive results by age range and gender is presented in Table 17 below.
Table 17 ARCHITECT AUSAB Results by Age Range and Gender for Pediatric Population
| Age Group (Years)^{a} | Gender | Reactive N (%) | Grayzone N (%) | Nonreactive N (%) | Total |
| --- | --- | --- | --- | --- | --- |
| Under 2 | Female | 8 (80.00%) | 1 (10.00%) | 1 (10.00%) | 10 |
| | Male | 12 (85.71%) | 0 (0.00%) | 2 (14.29%) | 14 |
| 2 to 12 | Female | 12 (54.55%) | 2 (9.09%) | 8 (36.36%) | 22 |
| | Male | 8 (21.05%) | 5 (13.16%) | 25 (65.79%) | 38 |
| 13 to 18 | Female | 18 (75.00%) | 1 (4.17%) | 5 (20.83%) | 24 |
| | Male | 8 (66.67%) | 0 (0.00%) | 4 (33.33%) | 12 |
| Total | | 66 (55.00%) | 9 (7.50%) | 45 (37.50%) | 120 |
a Children with ages under 2 are greater than 1 month old.
Page 32 39
{32}
SUMMARY OF SAFETY AND EFFECTIVENESS DATA
# XI. CONCLUSIONS DRAWN FROM THE STUDIES
Multi-centered clinical studies were conducted in the US to evaluate the ARCHITECT AUSAB anti-HBs assay. A method comparison was performed with a commercially available licensed assay to detect anti-HBs antibodies in specimens from an intended use diagnostic population.
Hepatitis B virus classification using the prospective population showed 19 unique reference marker patterns. The overall positive percent agreement between the ARCHITECT AUSAB assay and the reference assay was 98.6% (699/709) in the high risk, signs and symptoms, and vaccinated populations. The overall negative percent agreement between the ARCHITECT AUSAB assay and the reference assay was 96.7% (1165/1205) in the same population.
In the HBV vaccinated individuals, the positive agreement was 100% (146/146) and the negative percent agreement was 84.8% (39/46) with the comparison method.
The ability of the ARCHITECT AUSAB assay to detect the anti-HBS antibodies was demonstrated in pediatric and neonatal specimen testing.
Precision and reproducibility of the ARCHITECT AUSAB was established for within-run, within-day, within-lab, and between sites.
Tube Type Interference study results support the use of human serum and plasma (dipotassium EDTA, lithium heparin, and sodium heparin) and neonatal serum in the ARCHITECT AUSAB assay.
The results from both the non-clinical and clinical studies indicate that the ARCHITECT AUSAB assay can be used safely and effectively for the qualitative in vitro determination of anti-HBs antibodies in human serum and plasma. The assay may be used with other HBV serological markers to define the clinical status of patients known to be infected with HBV.
# RISK BENEFIT ANALYSIS
As a diagnostic test, the ARCHITECT AUSAB assay involves removal of blood from an individual for testing purposes. This test presents no more of a safety hazard to an individual than is presented to an individual who is having their blood drawn for any other diagnostic evaluation. The benefits to HBV-infected individuals tested by the assay outweigh any potential adverse event or risk to the patient or user due to assay malfunction or operator error.
The potential risks encountered with this in vitro diagnostic test are not unusual in the clinical laboratory setting. Appropriate warnings for these risks are contained
Page 33 40
{33}
SUMMARY OF SAFETY AND EFFECTIVENESS DATA
in the labeling and package inserts for the device. Standard good laboratory practices are considered sufficient to mitigate the risks to the end user.
## SAFETY
Based on the results of the preclinical and clinical laboratory studies, the ARCHITECT AUSAB assay, when used according to the provided directions and in conjunction with other serological and clinical information, should be safe and effective and pose minimal risk to the patient due to false test results.
## EFFECTIVENESS
The effectiveness of the ARCHITECT AUSAB has been demonstrated for use in determining if antibodies to the HBs antigen of the hepatitis B virus are present in an individual’s serum or plasma. A reasonable determination of effectiveness of the ARCHITECT AUSAB assay for aiding in the diagnosis of immunity and status of HBV infection in suspected individuals has been demonstrated.
## XII. PANEL RECOMMENDATIONS
In accordance with the provisions of section 515(c)(2) of the act as amended by the Safe Medical Devices Act of 1990, this PMA was not referred to the Microbiology Advisory Panel, an FDA advisory committee, for review and recommendation because the information in the PMA substantially duplicates information previously reviewed by this panel.
## XIII. CDRH DECISION
FDA issued an approval order on June 1, 2006.
The applicant’s manufacturing facility was inspected on 5/8/06 (N. Chicago), 5/16/06 (Abbott Park), & 5/19/04 (Puerto Rico) and found to be in compliance with the Quality Systems Regulation (21 CFR 820).
## XIV. APPROVAL SPECIFICATIONS
Directions for use: See the labeling.
Hazards to Health from Use of the Device: See Indications, Contraindications, Warnings, Precautions and Adverse Events in the labeling.
Postapproval Requirements and Restrictions: See approval order.
Page 34 41
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.