← Product Code [LGW](/productcode/LGW) · P030017S275

# PrecisionTM and Spectra WaveWriterTM Spinal Cord Stimulation (SCS) Systems (P030017S275)

_Boston Scientific Corp · LGW · Aug 11, 2017 · Neurology · APPR_

**Canonical URL:** https://fda.innolitics.com/device/P030017S275

## Device Facts

- **Applicant:** Boston Scientific Corp
- **Product Code:** [LGW](/productcode/LGW.md)
- **Decision Date:** Aug 11, 2017
- **Decision:** APPR
- **Device Class:** Class 3
- **Review Panel:** Neurology
- **Attributes:** Therapeutic, Real-World Evidence

## Real-World Evidence

| Submission | Device | Sponsor | RWD Sources | RWE Use Summary | Key Tags |
| --- | --- | --- | --- | --- | --- |
| P030017S275 · Aug 11, 2017 | PrecisionTM and Spectra WaveWriterTM Spinal Cord Stimulation (SCS) Systems | Boston Scientific Corp | Published clinical literature; Retrospective clinical studies; Prospective clinical studies; Randomized controlled trials (RCTs) | The FDA conducted a systematic review of 44 publications (3,413 patients) for safety and 22 publications (633 patients) for effectiveness to support the expanded indication. The data were used to demonstrate the safety profile and clinical effectiveness of SCS therapy in the target patient populations. | Systematic literature review; Retrospective analysis; Chronic intractable pain; Complex Regional Pain Syndrome (CRPS); Failed Back Surgery Syndrome (FBSS) |

### Clinical Evidence

| Study Design | Population | Comparator | Key Endpoints |
| --- | --- | --- | --- |
| Systematic Literature Review (Safety); Systematic review of published clinical studies; Study Period: Last 20 years | Patients with chronic intractable pain of the trunk and/or limbs (including FBSS, CRPS, radiculopathy, etc.); Sample Size: 3413 implanted patients | Not applicable for this study | Incidence of adverse events and device-related complications |
| Systematic Literature Review (Effectiveness); Systematic review of published clinical studies (including P, R, and RCT designs); Follow-up/Duration: At least 6 months | Patients with CRPS Type I/II, radicular pain syndrome, FBSS, epidural fibrosis, etc.; Sample Size: 633 implanted patients | Not applicable for this study | Reduction in VAS score, 50% reduction in pain, or clinically validated patient-reported outcomes |

## Indications for Use

The Boston Scientific Spinal Cord Stimulator Systems are indicated as an aid in the management of chronic intractable pain of the trunk and/or limbs including unilateral or bilateral pain associated with the following: failed back surgery syndrome, Complex Regional Pain Syndrome (CRPS) Types I and II, intractable low back pain and leg pain. Associated conditions and etiologies may be radicular pain syndrome, radiculopathies resulting in pain secondary to failed back syndrome or herniated disc, epidural fibrosis, degenerative disc disease (herniated disc pain refractory to conservative and surgical interventions), arachnoiditis, multiple back surgeries.

## Device Story

Totally implanted spinal cord stimulation (SCS) system; delivers electrical pulses to dorsal column to manage chronic intractable pain. Components: Implanted Pulse Generator (IPG) (rechargeable/non-rechargeable), percutaneous/surgical leads, lead extensions, splitters, anchors. External components: External Trial Stimulator (ETS), Clinician Programmer (CP), Remote Control (RC), charger, programming wand. Physician implants leads in epidural space; connects to IPG. Clinician uses CP to program stimulation parameters; patient uses RC to adjust therapy (on/off). System provides paresthesia or sub-paresthesia stimulation to modulate pain signals. Benefits: pain relief, improved function. Risks: lead migration, infection, hardware malfunction, surgical complications.

## Clinical Evidence

Systematic literature review of 44 publications (3413 patients) for safety and 22 publications (633 patients) for effectiveness. Primary endpoints: pain reduction (VAS) and 50% pain relief. Results: 23.9% required additional intervention; 12% lead migration; 3.5% infection. Effectiveness: CRPS Type I success 51% (pooled); CRPS Type II success 88% (pooled); radicular/back pain conditions success 37-77%. Evidence supports safety/effectiveness based on 30 years of clinical experience with similar SCS systems.

## Technological Characteristics

Totally implanted SCS system. Components: IPG, percutaneous/surgical leads, extensions, trial stimulators. Energy: Battery-powered (rechargeable/non-rechargeable). Connectivity: Telemetry between RC/CP and IPG. Sterilization: Standard medical device sterilization. Software: Programmable stimulation parameters. Materials: Standard implantable-grade materials (leads, IPG housing).

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a PMA](https://innolitics.com/services/regulatory/), [a 510(k)](https://innolitics.com/services/510ks/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

# SUMMARY OF SAFETY AND EFFECTIVENESS DATA (SSED)

## I. GENERAL INFORMATION

Device Generic Name: Stimulator, Spinal-Cord, Totally Implanted For Pain Relief

Device Trade Name: Precision™ Spinal Cord Stimulator System
Precision Spectra™ Spinal Cord Stimulator System
Precision Novi™ Spinal Cord Stimulator System
Precision™ Montage™ MRI Spinal Cord Stimulator System
Precision™ Montage™ Spinal Cord Stimulator System
Spectra WaveWriter™ Spinal Cord Stimulator System

Collectively referred to as the Precision™ and Spectra
WaveWriter™ Spinal Cord Stimulation (SCS) Systems

Device Procode: LGW

Applicant's Name and Address: Boston Scientific Neuromodulation Corporation
24242 Rye Canyon Loop
Valencia, CA 91355

Date(s) of Panel Recommendation: None

Premarket Approval Application (PMA) Number: P030017/S275

Date of FDA Notice of Approval: August 11, 2017

The original PMA P030017 was approved on April 27, 2004 and is indicated as an aid in the management of chronic intractable pain of the trunk and/or limbs, including unilateral or bilateral pain associated with the following: failed back surgery syndrome, intractable low back pain and leg pain. The SSED to support the indication is available on the CDRH website and is incorporated by reference here. The current supplement was submitted to expand the indication for the Boston Scientific Spinal Cord Stimulator Systems.

## II. INDICATIONS FOR USE

The Boston Scientific Spinal Cord Stimulator Systems are indicated as an aid in the management of chronic intractable pain of the trunk and/or limbs including unilateral or bilateral pain associated with the following: failed back surgery syndrome, Complex Regional Pain Syndrome (CRPS) Types I and II, intractable low back pain and leg pain.

Associated conditions and etiologies may be

- radicular pain syndrome,

PMA P030017-S275: FDA Summary of Safety and Effectiveness Data

Page 1

{1}

- radiculopathies resulting in pain secondary to failed back syndrome or herniated disc,
- epidural fibrosis,
- degenerative disc disease (herniated disc pain refractory to conservative and surgical interventions),
- arachnoiditis,
- multiple back surgeries.

### III. CONTRAINDICATIONS

The following patients are contraindicated from being treated with Boston Scientific Spinal Cord Stimulator Systems:

- Poor surgical candidates;
- Unable to operate the SCS system;
- Failed trial stimulation by failing to receive effective pain relief;
- Pregnant.

### IV. WARNINGS AND PRECAUTIONS

The warnings and precautions can be found in the Boston Scientific Spinal Cord Stimulator Systems labeling.

### V. DEVICE DESCRIPTION

Boston Scientific Spinal Cord Stimulator (SCS) Systems are totally implanted devices that deliver electrical stimulation to the dorsal column of the spinal cord for the treatment of chronic intractable pain of the trunk and/or limbs. Figures 1 and 2 display main system components and the typical location of implanted components.

![img-0.jpeg](img-0.jpeg)

PMA P030017-S275: FDA Summary of Safety and Effectiveness Data

Page 2

{2}

**Figure 1:** The Boston Scientific Spinal Cord Stimulation System

![img-1.jpeg](img-1.jpeg)

**Figure 2:** Typical location of implanted pulse generator and percutaneous leads.

#### **A. Implanted Components**

The implanted components of the Boston Scientific SCS System include the following:

- **Implanted Pulse Generator (IPG):** Generates programmable electrical pulses that are conducted to the spinal cord via leads. Includes rechargeable or non-rechargeable batteries.
  - IPG models include Precision, Precision Spectra, Precision Novi, Spectra Wavewriter, Precision Montage and Precision Montage MRI
- **Percutaneous and Surgical Leads:** Available in various lengths and configurations, the leads are connected to the IPG to deliver stimulation to the spinal cord.
  - Percutaneous lead models include Linear, Linear ST, Linear 3-4, Linear 3-6, Infinion, Infinion CX, Avista MRI that is used specifically with Precision Montage MRI System
  - Surgical lead models include CoverEdge that is used with the Precision Spectra and Spectra Wavewriter Systems, and Artisan
- **Lead Extension:** Lead Extensions are designed to provide additional length to connect the leads to the stimulator. Lead extensions come in lengths of 25cm, 35cm, and 55cm.

PMA P030017-S275: FDA Summary of Safety and Effectiveness Data

Page 3

{3}

- Lead Splitters: Optional component used to connect multiple leads to the IPG.
- Implantable Adaptors: Adaptors are provided to connect other manufacturer's leads to Boston Scientific IPGs.
- Suture Sleeves and Anchors: Used to anchor the lead to the supraspinous ligament or deep fascia.

### B. External Components

The external components of the Boston Scientific SCS System include the following:

- External Trial Stimulator (ETS): The ETS is intended to provide trial stimulation with implanted leads before permanent placement of the IPG. It provides the identical stimulation capabilities as the IPG.
  - ETS models are the Precision ETS used with the Precision System, and the Precision Spectra ETS used with all other Boston Scientific SCS Systems
- Clinician Programmer (CP): The CP is used by the clinician to program the IPG and ETS, and thus prescribe stimulation therapy for the patient.
- Remote Control (RC): The Remote Control is a hand-held, battery operated unit that uses telemetry to communicate with the IPG and ETS. It allows the patient to control the stimulation therapy prescribed by the clinician (e.g., turn SCS system on and off).
  - RC models are the Precision RC used with the Precision System, and the FreeLink RC used with all other Boston Scientific SCS Systems
- Programming Wand: The programming wand is used with some systems to allow the CP to communicate wirelessly with the IPG and ETS.
- Charger: The Charger is used with all rechargeable IPGs to transcutaneously charge the IPG battery.

### C. Accessories

Accessories provided with the Boston Scientific SCS Systems include the following:

- Torque Wrench: Used to tighten the set screws that lock the lead into the IPG.
- Stylets: Used to maneuver the lead through the epidural space to the desired implant location.

PMA P030017-S275: FDA Summary of Safety and Effectiveness Data

Page 4

{4}

- IPG Template: guides the physician to create the correct sizing of the subcutaneous pocket.
- Insertion Needle: used during implant procedure to introduce the percutaneous lead into the epidural space.
- Lead Blank: optionally used during implant procedure to clear a path for the introduction of the lead into the epidural space.
- Tunneling Tool: Used to create a subcutaneous tunnel from the IPG site to the lead implant location.
- IPG Connector Plug/Port Plug: Provided to seal the port(s) of the IPG that are not in use.
- OR (Operating Room) Cable and Extension: Used to connect the lead to the ETS during intraoperative testing and trial phase.
- External Adaptors: For connecting other manufacturer's SCS leads to the Boston Scientific external stimulators during in-office evaluation.

# VI. ALTERNATIVE PRACTICES AND PROCEDURES

There are several other alternatives for the treatment of chronic intractable pain of the trunk and/or limbs. Patients are typically treated on a treatment continuum with less invasive therapies prescribed first. Established non-surgical treatment options include, but are not limited to: oral medications (including Non-Steroidal Anti-inflammatory Drugs and opioids), massage therapy, physical/occupational/exercise therapy, psychological therapies (e.g., behavior modification, hypnosis), Transcutaneous Electrical Nerve Stimulation (TENS), acupuncture, sympathetic nerve blocks, epidural blocks, intrathecal blocks, and facet joint blocks. The surgical treatment options for these patients include sympathectomy, implantable intrathecal drug delivery systems, partially implanted SCS systems (power source is external) and commercially available fully implantable SCS systems. Each alternative has its own advantages and disadvantages. A patient should fully discuss these alternatives with his/her physician to select the method that best meets expectations and lifestyle.

# VII. MARKETING HISTORY

Currently, the Boston Scientific Spinal Cord Stimulator System for the treatment of chronic pain of the trunk and limbs is approved for commercial distribution in the United States, European Community (EC) countries, Canada, Australia/New Zealand, Brazil, Kuwait, Israel, Japan, Argentina, Chile, Colombia, Costa Rica, Ecuador, El Salvador, Guatemala, Mexico, Panama, Peru, Russia, South Africa, Turkey, and United Arab Emirates.

PMA P030017-S275: FDA Summary of Safety and Effectiveness Data

Page 5

{5}

The device has not been withdrawn from marketing for any reason related to its safety or effectiveness.

### VIII. POTENTIAL ADVERSE EFFECTS OF THE DEVICE ON HEALTH

Below is a list of the potential adverse effects (e.g., complications) associated with the use of SCS Systems. The Boston Scientific SCS systems are similar to other legally-marketed SCS systems in intended use, target patient population, technology, device design and output characteristics. Therefore, the following list of potential adverse effects have been identified from peer-reviewed published literature that describes studies of all legally-marketed SCS systems.

The adverse effects include: (1) those associated with any surgical procedure, (2) those associated with the SCS system placement procedures, and (3) those associated with having an implanted SCS system to treat pain, including the Boston Scientific SCS System. In addition to the risks listed below, there is the risk that the SCS therapy may not be effective in relieving symptoms, or may cause worsening of symptoms. Additional intervention may be required to correct some of the adverse effects.

- Risks associated with any surgical procedure: abscess; cellulitis; excessive fibrotic tissue; wound dehiscence; wound, local or systemic infection; wound necrosis; edema; inflammation; foreign body reaction; hematoma; seroma; thrombosis; ischemia; embolism; thromboembolism; hemorrhage; thrombophlebitis; adverse reactions to anesthesia; hypertension; pulmonary complications; organ, nerve or muscular damage; gastrointestinal or genitourinary compromise; seizure, convulsion, or changes to mental status; inability to resume activities of daily living; and death.
- Risks associated with SCS system placement procedures: temporary pain at the implant site, infection, cerebrospinal fluid (CSF) leakage, CSF fistula, epidural hemorrhage, bacterial meningitis, seroma, hematoma, and paralysis. Patient use of anticoagulation therapies may increase the risk of procedure-related complications such as hematomas, which could produce paralysis.
- Risks associated with the use of a SCS system: lead migration; IPG migration; allergic response or tissue reaction to the implanted system material; hematoma or seroma at the implant site; skin erosion at the implant site; persistent pain at the IPG, extension, or lead site; radicular chest wall stimulation; disturbed urination; dysesthesia; decubitus; premature battery depletion; loss of pain relief over time; and uncomfortable stimulation or ineffective pain control caused by random failure of the system components or battery, changes in electrode position, loose electrical connections, lead or extension insulation breaches or fractures.

### IX. SUMMARY OF NONCLINICAL STUDIES

PMA P030017-S275: FDA Summary of Safety and Effectiveness Data

Page 6

{6}

Pre-clinical studies (bench and animal) previously submitted to FDA in the Original PMA application (P030017) and supplements continue to support the safety of the commercially available Boston Scientific SCS System for treatment of chronic intractable pain of the trunk and/or limbs. No additional preclinical studies were required to evaluate the safety of Boston Scientific SCS therapy for the treatment of the new patient populations. The previously approved supplements which support the device and its components are listed below in Table 1.

**Table 1. Summary of System/Device Components and Their Respective Approval References**

|  System/Device Component | Approval Reference  |
| --- | --- |
|  Precision™ Spinal Cord Stimulator System Includes IPG, Linear Leads, Lead Extensions, External Trial Stimulator, Remote Control, Charger, Base Station, Clinician Programmer | P030017  |
|  Artisan Leads | P030017/S008  |
|  Linear ST Leads | P030017/S020  |
|  Connector M1 | P030017/S025  |
|  Linear 3-4 and Linear 3-6 Leads | P030017/S100  |
|  Infinion Leads | P030017/S119  |
|  Precision Spectra™ Spinal Cord Stimulator System Includes IPG, FreeLink Remote Control, External Trial Stimulator, Programming Wand, Clinician Programmer | P030017/S134  |
|  CoverEdge Leads | P030017/S152  |
|  Infinion CX Leads | P030017/S191  |
|  M8 Adapter | P030017/S202  |
|  S8 Adapter | P030017/S210  |
|  Precision Novi™ Spinal Cord Stimulator System | P030017/S217  |
|  Precision™ Montage™ MRI Spinal Cord Stimulator System Includes Avista MRI leads | P030017/S235  |
|  Precision™ Montage™ Spinal Cord Stimulator System | P030017/S235  |
|  Spectra WaveWriter™ Spinal Cord Stimulator System | P030017/S271  |

## **X. SUMMARY OF PRIMARY CLINICAL STUDIES**

The Boston Scientific SCS Systems are approved as an aid in the management of chronic intractable pain of the trunk and/or limbs, including unilateral or bilateral pain associated with the following: failed back surgery syndrome, intractable low back pain and leg pain.

### **A. Study Design**

The safety and effectiveness of the Boston Scientific SCS System was based on a systematic review of published clinical studies of the safety and effectiveness of commercially available, fully implantable SCS systems in treating chronic intractable pain of the trunk and/or limb, including unilateral or bilateral pain associated with Complex Regional Pain Syndrome Types I and II, radicular pain syndrome,

PMA P030017-S275: FDA Summary of Safety and Effectiveness Data

Page 7

{7}

radiculopathies resulting in pain secondary to failed back syndrome or herniated disc, epidural fibrosis, degenerative disc disease (herniated disc pain refractory to conservative and surgical interventions), arachnoiditis, and multiple back surgeries. The Boston Scientific SCS System is similar in design, technology, performance, and intended use to the SCS systems used to treat the patient populations in the submitted published clinical studies. Therefore, data obtained from these published studies were used to establish a reasonable assurance of safety and effectiveness of the Boston Scientific SCS Systems for use in these specific patient populations, and were the basis for the PMA approval decision.

A total of 44 publications that reported data on 40 study populations (some publications reported data on the same patient population at different time points), representing a total of 3413 patients were identified for inclusion in the safety analysis. A total of 22 papers (19 studies) representing a total of 633 implanted patients were included in the effectiveness analysis.

The Boston Scientific SCS System is similar to the SCS systems reported in the published literature in intended use, target patient population, device design, and output characteristics. Based on these similarities the primary objective of the literature search was to provide clinical evidence of the effectiveness of SCS Systems when they are used to treat chronic intractable pain of the trunk and/or limb, including unilateral or bilateral pain associated with Complex Regional Pain Syndrome Types I and II, radicular pain syndrome, radiculopathies resulting in pain secondary to failed back syndrome or herniated disc, epidural fibrosis, degenerative disc disease (herniated disc pain refractory to conservative and surgical interventions), arachnoiditis, and multiple back surgeries.

Effectiveness was demonstrated by patient-reported outcome measures of pain, and using one or more of the following:

1. A reduction of pain as demonstrated by a clinically significant reduction in the Visual Analog Scale (VAS) score;
2. A 50% reduction in pain using a validated patient-reported assessment of pain relief (e.g., a 3 or 4 point scale), in at least 30% of patients included in the study;
3. A clinically significant difference in pain reduction as measured by a clinically validated patient-reported outcome (e.g., VAS) when compared to a control group.

Safety of the Boston Scientific SCS System was established using literature articles by examining the incidence of complications of the SCS systems used in each study. The articles describe studies of SCS Systems when used to treat chronic intractable pain of the trunk and/or limb, including unilateral or bilateral pain associated with Failed Back Surgery Syndrome (FBSS), intractable low back and leg pain, Complex Regional Pain Syndrome Types I and II, radicular pain syndrome, radiculopathies resulting in pain secondary to failed back syndrome or herniated disc, epidural

PMA P030017-S275: FDA Summary of Safety and Effectiveness Data

Page 8

{8}

fibrosis, degenerative disc disease (herniated disc pain refractory to conservative and surgical interventions), arachnoiditis, and multiple back surgeries. Literature articles that identified patients as having FBSS, with no further delineation of their pain condition or etiology, were included in the safety analysis because we believe it is reasonable to assume that the majority of these patients received surgery for one of the conditions of interest (e.g., herniated disc) or developed one or more of the conditions or etiologies following the surgery (e.g., arachnoiditis, radicular pain syndrome).

## B. Literature Search Strategy

The strategy for the SCS literature search was chosen to ensure inclusion of specific populations of SCS subjects into the safety and effectiveness analysis. These populations were patients who had pain from: multiple back operations, post-laminectomy, post-discectomy or other disc surgery, epidural fibrosis, arachnoiditis, herniated disc, degenerative disc disease, radiculopathy or other radicular syndromes, and complex regional pain syndrome (CRPS), previously known as reflex sympathetic dystrophy (RSD) and causalgia.

The literature search strategy consisted of the following steps:

1. Search of PubMed database for indexed articles using search terms for SCS systems, treatment of trunk and/or limb pain, and treatment of the specific pain conditions and etiologies of interest (e.g., CRPS Type I and II, arachnoiditis) (Pubmed database available through National Center for Biotechnology Information [NCBI] at the National Library of Medicine [NLB] located at the National Institute of Health [NIH]).
2. The following exclusion criteria were applied to the search results:

- Angina, peripheral nerve stimulation, peripheral vascular disease, peripheral artery disease, spinal cord injury, SCS to induce movement, cervical SCS.
- All data in the study were for single contact leads.
- Case reports with 3 or fewer subjects.
- Trial SCS data only.
- SCS used only with other treatments under study; e.g. intrathecal drug pump.
- Article in language other than English.

3. Determination of studies appropriate for the safety summary (total of 44 publications and 40 subject populations), which required that the article be published within the last 20 years, and that it include a report of adverse event or device complication data.
4. Determination of studies appropriate for the effectiveness summary (total of 22 publications and 19 subject populations), which required the following:

- A defined population that included one or more of the conditions or etiologies of interest (i.e., adequate inclusion/exclusion criteria).

PMA P030017-S275: FDA Summary of Safety and Effectiveness Data

Page 9

{9}

- Effectiveness outcomes assessed at least 6 months after implantation.
- Specific pain endpoints (e.g., VAS)
- Analysis of effectiveness outcomes for specific patient population having one or more of the conditions or etiologies of interest.

# C. Safety and Effectiveness Results

# 1. Safety Results

Table 2 provides a summary of Adverse Events (AEs) reported in 44 publications that reported data on 40 study populations (4 publications reported on the same subjects at later time points), and that included a total of 3413 implanted patients, were analyzed and summarized. Some studies reported the number of AEs whereas others reported the number of patients in which the AEs occurred. To combine data from these studies, each event was included as an AE count. For example, if one patient was reported as having three additional interventions for two infections, this was counted as five total AE counts, three toward Additional Intervention Required and two toward Infection. The incidence of each event is reported as a percentage of occurrences in total implanted patients, i.e. Incidence = (Number of AEs / Number of implanted patients) x 100.

Table 2. Summary of SCS Risks Identified in Clinical Studies

|  Clinical Harm | Counts | Incidence = (Count/3413 implanted patients)*100  |
| --- | --- | --- |
|  Accidental Injuries, Secondary to Stimulation | 4 | 0.12%  |
|  Abnormal Healing | 4 | 0.12%  |
|  Adhesions | 4 | 0.12%  |
|  Additional Intervention Required | 816 | 23.91%  |
|  Allergic Reaction | 3 | 0.09%  |
|  Cerebrospinal Fluid (CSF) Leak | 8 | 0.23%  |
|  Death (none described as related to SCS) | 4 | 0.12%  |
|  Dehiscence | 6 | 0.18%  |
|  Discomfort | 51 | 1.49%  |
|  Erosion | 6 | 0.18%  |
|  Hematoma | 21 | 0.62%  |
|  Inadequate Stimulation | 103 | 3.02%  |
|  Infection | 119 | 3.49%  |
|  Inflammation | 5 | 0.15%  |
|  Muscle Spasm, Stimulation Related | 4 | 0.12%  |

PMA P030017-S275: FDA Summary of Safety and Effectiveness Data

Page 10

{10}

|  Clinical Harm | Counts | Incidence = (Count/3413 implanted patients)*100  |
| --- | --- | --- |
|  Musculoskeletal Stiffness or Spasm, Not Stimulation Related | 2 | 0.06%  |
|  Nerve Injury | 23 | 0.67%  |
|  Neurostimulation Effects on Other Systems, e.g. dizziness, urinary effects | 8 | 0.23%  |
|  Overstimulation of Tissue | 2 | 0.06%  |
|  Pain | 276 | 8.09%  |
|  Paralysis | 13 | 0.38%  |
|  Seroma | 14 | 0.41%  |
|  Tissue Damage | 1 | 0.03%  |
|  Undesired Sensations, Target or Non-Target Stimulation Area | 127 | 3.72%  |

Additional interventions were the most commonly occurring AEs at a rate of 23.91%. This category combines surgical revisions to correct lead migration, IPG discomfort, and battery depletion, as well as device explants to treat infection and replace fractured leads or malfunctioning equipment.

Among the other reported AEs, pain (8%), undesired sensations (4%), infection (3.5%), inadequate stimulation (3%), and discomfort (1.5%) were the five most common clinical harms. All other harms were seen at less than 1%. In the population of chronic pain patients, incidence of pain is commonly reported, and these counts include device-related pain, postoperative pain, and pain in other areas. Most of these events were temporary and related to the procedure itself and were resolved. Subjects with inadequate stimulation either had incomplete paresthesia coverage or experienced other problems with the device.

Table 3 provides a summary of device-related or hardware complications. A total of 759 device-related complications were reported in the safety data population of 3413 patients, for a rate of 22.24%. The incidence of each event is reported as a percentage of occurrences in total implanted patients, i.e. Incidence = (Number of device complications/Number of implanted patients) x 100.

**Table 3. Summary of Device-Related Complications**

|  Complications | Counts | Incidence = (Count/3413 implanted patients)*100  |
| --- | --- | --- |
|  Lead Migration | 413 | 12.1%  |
|  Lead Breakage or Malfunction | 121 | 3.55%  |

PMA P030017-S275: FDA Summary of Safety and Effectiveness Data

Page 11

{11}

|  Complications | Counts | Incidence = (Count/3413 implanted patients)*100  |
| --- | --- | --- |
|  IPG Malfunction | 83 | 2.43%  |
|  Disconnection of Lead or Extension | 51 | 1.49%  |
|  IPG Migration or Discomfort | 45 | 1.32%  |
|  Loss of Paresthesia | 33 | 0.97%  |
|  Battery depletion within 1st year of implant | 12 | 0.35%  |
|  Charger Dysfunction | 1 | 0.03%  |
|  *Total* | *759* | *22.24%*  |

Among the reported device-related complications, lead migration had the most common incidence of 12% followed by lead breakage or malfunction at about 4%. Lead migrations and malfunctions typically required lead repositioning or replacement.

## 2. Effectiveness Results

Table 4 provides a summary of effectiveness data reported in 15 publications and obtained from 13 studies of patients diagnosed with CRPS Type I, CRPS Type II, or CRPS Not-Otherwise-Specified (NOS). The study design (SD) varied among the 13 studies with five prospective (P) studies, one randomized controlled trial (RCT), and seven retrospective (R) studies.

We indicate both the number of patients enrolled in each study and the number of patients permanently implanted with an SCS System. For the RCT and the prospective studies, the principal reason the number of patients implanted is different than the number of patients enrolled is because some patients did not experience at least 50% pain relief during trial stimulation with an external trial stimulator (and therefore, they are not implanted with a permanent system). For the retrospective studies, the principal reason the number of enrolled vs. implanted patients are different is due to the heterogenous nature of the patient populations included in the study.

We refer to the time between when patients were implanted, and the time when device effectiveness was assessed, as follow-up (FU) time. For most studies, FU time varied among patients and therefore, the mean number of FU months when averaged for all evaluated patients is indicated in Table 4. For the few studies that evaluated patients at specific timepoints following implantation, the FU range is indicated and the success data at each reported timepoint is provided. Success percentages are computed by dividing the number of patients meeting one or more of our definitions of device effectiveness by the total number of patients that were evaluated at least 6 months after implantation.

PMA P030017-S275: FDA Summary of Safety and Effectiveness Data

Page 12

{12}

**Table 4. Summary of Effectiveness Results in Clinical Studies:

|  Author | Year | # Implanted (# Enrolled) | SD | FU | Success % (# of patients)  |
| --- | --- | --- | --- | --- | --- |
|  Geurts | 2013 | 84 CRPS I (84) | P | 132 | 40.5% (34/84)  |
|  Kemler | 1999 | 18 CRPS I (23) | R | 32 | 72.2% (13/18)  |
|  Kemler | 2000 2004 2008 | 24 CRPS I (54) | RCT | 6 - 60 | 6 mo: 58.3% (14/24) 2 yrs: 62.5% (15/24) 5 yrs: 29.2% (7/24) Significantly reduced VAS vs control ($p<0.001$), except @ 5 years ($p=0.06$)  |
|  Kumar | 1996 | 9 CRPS II (30) | R | 87 | 66% (6/9)  |
|  Kumar | 1997 | 12 CRPS I (12) | R | 41 | 100% (12/12)  |
|  Kumar | 2006 | 32 CRPS I and II (452) | R | 97.6 (median) | 71.9% (23/32)  |
|  Lame | 2009 | 32 CRPS I (32) | P | 9 | 38% (12/32)  |
|  Moriyama | 2012 | 14 CRPS NOS (55) | P | 6 | 83.3% (10/12)  |
|  Oakley | 1999 | 19 CRPS NOS (19) | P | 7.9 | 80% (8/10)  |
|  Robaina | 1989 | 6 CRPS I (11) | P | 27 | 90.9% (5/6)  |
|  Sanchez- Ledesma | 1989 | 11 CRPS II 8 CRPS I (49) | R | 60 | 100% (19/19)  |

PMA P030017-S275: FDA Summary of Safety and Effectiveness Data

Page 13

{13}

|  Sears | 2011 | 18 CRPS NOS (35) | R | 52.8 | 55.6% (10/18)  |
| --- | --- | --- | --- | --- | --- |
|  Verdolin | 2007 | 4 CRPS I 6 CRPS II (10) | R | 6 | 100% (10/10)  |

Collectively, the data in Table 4 were obtained from a total of 297 implanted patients diagnosed with CRPS Type I, CRPS Type II, or CRPS-NOS. The range of success rates observed across all 13 studies and FU times is 29.2% to 100%, with twelve studies observing success rates greater than 30% and ten studies observing success rates greater than 50%.

Patients specifically diagnosed with CRPS Type I represent 188 of the implanted patients. The range of success rates observed across all FU times and all studies of CRPS Type I patients is 29.2% to 100%. When the success data for CRPS Type I patients is pooled across all studies, the collective success rate is 51% (95 / 188).

Patients specifically diagnosed with CRPS Type II represent 26 of the implanted patients. The range of success rates observed across all FU times and all studies of CRPS Type II patients is 66% to 100%. When the success data for CRPS Type II patients is pooled across all studies, the collective success rate is 88% (23 / 26).

Table 5 provides a summary of effectiveness data reported in 7 publications and obtained from 6 studies of patients diagnosed with radicular pain syndrome, radiculopathies resulting in pain secondary to failed back syndrome or herniated disc, epidural fibrosis, degenerative disc disease (herniated disc pain refractory to conservative and surgical interventions), arachnoiditis, and multiple back surgeries. The study design (SD) varied among the studies with two randomized controlled trials (RCT) and four retrospective studies (R).

The definitions for most columns are the same as those for Table 4. The number of surgeries refers to the mean number of prior back surgeries that occurred prior to treatment with an SCS device, when averaged over all patients. When the article did not specify the mean number of back surgeries for the patient population, we included the author’s description of the population’s surgical history when it was provided.

PMA P030017-S275: FDA Summary of Safety and Effectiveness Data

Page 14

{14}

**Table 5. Summary of Effectiveness Results in Clinical Studies:

|  Author | Year | Condition / Etiology | # Implant (# Enroll) | Number of Surgeries | SD | FU | Success % (# of Patients)  |
| --- | --- | --- | --- | --- | --- | --- | --- |
|  De La Porte | 1983 | Arachnoiditis | 38 (94) | 3.5 | R | 35.8 | 12 months: 68% (21/31)  |
|  Fiume | 1995 | Epidural Fibrosis | 36 (55) | 1.8 | R | 55 | 56% (20/36)  |
|  Kumar | 2007 2008 | Radicular pain after surgery for a herniated disc | 50 (100) | More than one surgery for 28 subjects implanted | RCT | 24 | 6 months: 48% (24/50) 24 months: 40% (17/42)  |
|  North | 2005 | Radicular pain after spine surgery | 31 (60) | 2.5 | RCT | 34.8 | 52% (15/29)  |
|  Probst | 1990 | Radicular Pain due to Epidural and Intradural Fibrosis after surgery for lumbar disc herniation | 92 (112) | 85% of patients had history of multiple operations | R | 54 | 6-12 months: 77% (71/92) 13-24 months: 69% (63/92)  |
|  Siegfried | 1982 | Arachnoid and Epidural Fibrosis | 89 (191) | Not Reported | R | 48 | 37%  |

Collectively, the data in Table 5 were obtained from a total of 336 implanted patients diagnosed with radicular pain syndrome, radiculopathies resulting in pain secondary to failed back syndrome or herniated disc, epidural fibrosis, degenerative disc disease (herniated disc pain refractory to conservative and surgical interventions), arachnoiditis, and multiple back surgeries. The range of success rates observed across all 6 studies and FU times is 37% to 77%, with all studies observing success rates greater than 30% and 4 of the 6 studies observing success rates greater than 50%.

### 3. Pediatric Extrapolation

PMA P030017-S275: FDA Summary of Safety and Effectiveness Data

Page 15

{15}

In this premarket application, existing clinical data was not leveraged to support approval of a pediatric patient population.

#### **D. Financial Disclosure**

A clinical study was not performed and thus, the Financial Disclosure by Clinical Investigators regulation (21 CFR 54) is not applicable to this PMA.

### **XI. PANEL MEETING RECOMMENDATION AND FDA'S POST-PANEL ACTION**

In accordance with the provisions of section 515(c)(3) of the act as amended by the Safe Medical Devices Act of 1990, this PMA was not referred to the Neurological Devices Panel, an FDA advisory committee, for review and recommendation because the information in the PMA substantially duplicates information previously reviewed by this panel.

### **XII. CONCLUSIONS DRAWN FROM PRECLINICAL AND CLINICAL STUDIES**

#### **A. Effectiveness Conclusions**

The evaluation of effectiveness was conducted using published clinical studies relevant to Boston Scientific Spinal Cord Stimulation (SCS) System features and indications. A total of 22 publications and 19 subject populations (some articles reported on the same subjects at a later time point) diagnosed with the conditions and etiologies of interest were reviewed. Collectively, these studies report data for 633 implanted patients: 188 diagnosed with CRPS Type I, 26 diagnosed with CRPS Type II, 83 diagnosed with CRPS-NOS, and 336 diagnosed with one or more of the associated conditions or etiologies. SCS treatment was demonstrated to be effective in all 19 subject populations described in these articles.

Patients specifically diagnosed with CRPS Type I represent 188 of the implanted patients. The range of success rates observed across all FU times and all studies of CRPS Type I patients is 29.2% to 100%. When the success data for CRPS Type I patients is pooled across all studies, the collective success rate is 51% (95 / 188).

Patients specifically diagnosed with CRPS Type II represent 26 of the implanted patients. The range of success rates observed across all FU times and all studies of CRPS Type II patients is 66% to 100%. When the success data for CRPS Type II patients is pooled across all studies, the collective success rate is 88% (23 / 26).

Patients specifically diagnosed with radicular pain syndrome, radiculopathies resulting in pain secondary to failed back syndrome or herniated disc, epidural fibrosis, degenerative disc disease (herniated disc pain refractory to conservative and surgical interventions), arachnoiditis, and multiple back surgeries represent 336 of the implanted patients. The range of success rates observed across all 6 studies and FU

PMA P030017-S275: FDA Summary of Safety and Effectiveness Data

Page 16

{16}

times is 37% to 77%, with all 6 studies observing success rates greater than 30%, and 4 of the 6 studies observing success rates greater than 50%.

The results of the review support the effectiveness of SCS therapy in treating patients who suffer from chronic, intractable pain of the trunk and/or limbs, including unilateral or bilateral pain associated with failed back surgery syndrome, Complex Regional Pain Syndrome Types I and II, intractable leg and back pain, radicular pain syndrome, radiculopathies resulting in pain secondary to failed back syndrome or herniated disc, epidural fibrosis, degenerative disc disease (herniated disc pain refractory to conservative and surgical interventions), arachnoiditis, and multiple back surgeries.

### **B. Safety Conclusions**

The clinical evidence provided to support safety of the Boston Scientific Neuromodulation Spinal Cord Stimulator Systems includes a systematic literature review of clinical research conducted on this and other comparable devices. Adverse Events (AEs) reported in 44 publications that reported data on 40 study populations (4 publications reported on the same subjects at later time points), and that included a total of 3413 implanted patients, were analyzed and summarized.

Additional interventions were the most commonly occurring AEs at a rate of 23.91%. This category combines surgical revisions to correct lead migration, IPG discomfort, and battery depletion, as well as device explants to treat infection and replace fractured leads or malfunctioning equipment. Among the other reported AEs, pain (8%), undesired sensations (4%), infection (3.5%), inadequate stimulation (3%), and discomfort (1.5%) were the five most common clinical harms. All other harms were seen at less than 1%. In the population of chronic pain patients, incidence of pain is commonly reported, and these counts include device-related pain, postoperative pain, and pain in other areas.

A total of 759 device-related complications were reported in the safety data population of 3413 patients, for a rate of 22.24%. Among the reported device-related complications, lead migration had the most common incidence of 12% followed by lead breakage or malfunction at about 4%. Lead migrations and malfunctions typically required lead repositioning or replacement.

The results of the systematic review support the safety of SCS therapy in treating patients who suffer from chronic, intractable pain of the trunk and/or limbs. The Boston Scientific Neuromodulation system is similar to SCS systems approved by the FDA.

### **C. Benefit-Risk Determination**

Chronic pain is very debilitating. Patients have difficulty performing activities of daily living.

PMA P030017-S275: FDA Summary of Safety and Effectiveness Data

Page 17

{17}

Depending upon the condition or etiology, the chronic pain may last for years. Chronic pain by definition has been present for at least 3 months. Treatment options for this population of patients are of great importance. Patients are willing to accept the probable risk given the probable benefit as evidenced by 30 years of SCS use.

The probable benefits of the device are based on data collected in a systematic literature review. A total of 22 publications and 19 subject populations (some articles reported on the same subjects at a later time point) diagnosed with the conditions and etiologies of interest were reviewed. Collectively, these studies report data for 633 implanted patients. SCS treatment was demonstrated to be effective in all 19 subject populations described in these articles.

The improvement in pain ranged from 29.2%-100% for CRPS patients, and 37-77% for those with back and leg pain due to surgery associated conditions and etiologies across the studies.

Follow-up ranged from 6 months to 11 years. At 11 years, Guerts et al (2013) reported a decrease in pain due to CRPS of at least 50% in 40.5% of patients studied (34/84 patients). At a mean of 4.5 years, Fiume et al (1995) found a reduction of at least 50% in pain due to epidural fibrosis in 56% of patients (20/36 patients).

The most common adverse event (AE), at 23.9% was the need for an additional intervention (surgical revisions to correct lead migration, IPG discomfort, battery depletion, infection, fractured leads). Other reported adverse events included pain (8%), unpleasant sensation (4%), infection (3.5%), inadequate stimulation (3%), and discomfort (1.5%). Infection is a common cause for explantation of the device.

As described above, the risks were evaluated in a systematic literature review. This analysis evaluated AEs reported in 44 publications that reported data on 40 study populations (4 publications reported on the same subjects at later time points), and that included a total of 3413 implanted patients. The safety information provided was consistent with the well-known safety profile of SCS systems.

Additional factors to be considered in determining probable risks and benefits for the Boston Scientific SCS System included the open label design of the clinical studies during FDA review. In some of the studies reviewed, open label studies may cause an overestimation of the treatment effect in investigator and subject ratings. Also, open label studies do not assess the magnitude of the placebo response, regression to the mean, the effect of changes in medications or other treatments to alleviate pain or changes in the underlying severity of the pain disorder. SCS is an option for patients who do not have adequate pain relief from medications and/or other treatments for pain.

This submission did not include specific information on patient perspectives for this device.

PMA P030017-S275: FDA Summary of Safety and Effectiveness Data

Page 18

{18}

In conclusion, given the available information above, the data support that the probable benefits outweigh the probable risks for the Boston Scientific SCS System when used as an aid in the management of chronic intractable pain of the trunk and/or limbs in patients, including unilateral or bilateral pain associated with the following: failed back surgery syndrome, Complex Regional Pain Syndrome (CRPS) Types I and II, intractable low back pain and leg pain. Associated conditions and etiologies may be radicular pain syndrome, radiculopathies resulting in pain secondary to failed back syndrome or herniated disc, epidural fibrosis, degenerative disc disease (herniated disc pain refractory to conservative and surgical interventions), arachnoiditis, and multiple back surgeries.

#### **D. Overall Conclusions**

The data in this application support the reasonable assurance of safety and effectiveness of this device when used in accordance with the indications for use. The results from the clinical evaluation support reasonable assurance of the safety and efficacy of the Boston Scientific SCS System, as well its long-term performance, when used in a manner consistent with its labeling and intended use. The evidence supporting the safety and effectiveness of the Boston Scientific SCS System is based on a foundation of 30 years of clinical research and experience as documented in the literature with fully implantable SCS systems and the similarities of the Boston Scientific system to market released implantable SCS systems. The analyses also support a clinical benefit to risk determination that is favorable.

### **XIII. CDRH DECISION**

CDRH issued an approval order on August 11, 2017.

The applicant's manufacturing facilities have been inspected and found to be in compliance with the device Quality System (QS) regulation (21 CFR 820).

### **XIV. APPROVAL SPECIFICATIONS**

Directions for use: See device labeling.

Hazards to Health from Use of the Device: See Indications, Contraindications, Warnings, Precautions, and Adverse Events in the device labeling.

Post-approval Requirements and Restrictions: See approval order.

### **XV. REFERENCES**

#### **A. References Used in Safety Analysis**

NOTE: Publications marked with a '*' were used in both the Safety and Effectiveness analysis

PMA P030017-S275: FDA Summary of Safety and Effectiveness Data

Page 19

{19}

Air E. L., Toczyl G. R., & Mandybur G. T. (2012). Electrophysiologic monitoring for placement of laminectomy leads for spinal cord stimulation under general anesthesia. *Neuromodulation*, 15(6):573-579.

Baldeschi G. C. & De Carolis G. (2014). The Italian experience with octopolar perc-paddle leads. *Neuromodulation*, 17(4):349-353.

Colombo, E. V., Mandelli, C., Mortini, P., Messina, G., De Marco, N., Donati, R., & Dones, I. (2015). Epidural spinal cord stimulation for neuropathic pain: a neurosurgical multicentric Italian data collection and analysis. *Acta Neurochir (Wien)*, 157(4):711-720.

Connor D. E. Jr, Cangiano-Heath A., Brown B., Vidrine R., Battley T. 3rd, Nanda A., & Guthikonda B. (2012). The utility of bone cement to prevent lead migration with minimally invasive placement of spinal cord stimulator laminectomy leads. *Neurosurgery*, 71(1):157-63.

De Andres J., Perotti L., Villanueva-Perez V. L., Asensio-Samper J. M., & Fabregat-Cid G. (2013). Role of lumbosacral retrograde neuromodulation in the treatment of painful disorders. *Pain Physician*, 16(2):143-153.

de Vos C. C., Bom M. J., Vanneste S., Lenders M. W., & De Ridder D. (2014). Burst spinal cord stimulation evaluated in patients with failed back surgery syndrome and painful diabetic neuropathy. *Neuromodulation*, 17(2):152-159.

de Vos C. C., Dijkstra C., Lenders M. W., & Holsheimer J. (2012). Spinal cord stimulation with hybrid lead relieves pain in low back and legs. *Neuromodulation*, 15(2):118-123.

Deer T., Skaribas I., McJunkin T., Nelson C., Salmon J., Darnule A., Fernando Gomezese, O. (2016). Results from the partnership for advancement in neuromodulation registry: A 24-Month follow-up. *Neuromodulation*, 19(2):179-187.

Forouzanfar T., Kemler M. A., Weber W. E., Kessels A. G., & van Kleef M. (2004). Spinal cord stimulation in complex regional pain syndrome: cervical and lumbar devices are comparably effective. *Br J Anaesth*, 92(3):348-353.

\*Geurts J. W., Smits H., Kemler M. A., Brunner F., Kessels A. G., & van Kleef M. (2013). Spinal cord stimulation for complex regional pain syndrome type I: A prospective cohort study with long-term follow-up. *Neuromodulation*, 16(6):523-529.

Gopal H., Fitzgerald J., & McCrory C. (2016). Spinal cord stimulation for FBSS and CRPS: A review of 80 cases with on-table trial of stimulation. *J Back Musculoskelet Rehabil*, 29(1):7-13.

Hamm-Faber T. E., Aukes H. A., de Loos F., & Gultuna I. (2012). Subcutaneous stimulation as an additional therapy to spinal cord stimulation for the treatment of lower limb pain and/or back pain: A feasibility study. *Neuromodulation*, 15(2):108-117.

PMA P030017-S275: FDA Summary of Safety and Effectiveness Data

Page 20

{20}

Harke H., Gretenkort P., Ladleif H. U., & Rahman S. (2005). Spinal cord stimulation in sympathetically maintained complex regional pain syndrome type I with severe disability. A prospective clinical study. Eur J Pain, 9(4):363-373.

Hayek S. M., Veizi E., & Hanes M. (2015). Treatment-limiting complications of percutaneous spinal cord stimulator implants: A review of eight years of experience from an academic center database. Neuromodulation, 18(7):603-608.

Kapural L., Yu C., Doust M. W., Gliner B. E., Vallejo R., Sitzman B. T., & Burgher A. H. (2015). Novel 10-kHz high-frequency therapy (HF10 Therapy) is superior to traditional low-frequency spinal cord stimulation for the treatment of chronic back and leg pain: The SENZA-RCT randomized controlled trial. Anesthesiology, 123(4):851-860.

*Kemler M. A., Barendse G. A., Van Kleef M., de Vet H. C., Rijks C. P., Furnee C. A., & Van Den Wildenberg F.A. (2000). Spinal cord stimulation in patients with chronic reflex sympathetic dystrophy. N Engl J Med, 343(9):618-624.

*Kemler M. A., Barendse G. A. M., Van Kleef M., Van Den Wildenberg F. A., & Weber W. J. (1999). Electrical spinal cord stimulation in reflex sympathetic dystrophy: retrospective analysis of 23 patients. J Neurosurg, 90(1 Suppl):79-83.

*Kemler M. A., de Vet H. C., Barendse G. A., Van Den Wildenberg F. A., & Van Kleef M. (2004). The effect of spinal cord stimulation in patients with chronic reflex sympathetic dystrophy: two years' follow-up of the randomized controlled trial. Ann Neurol, 55(1):13-18.

*Kemler M. A., de Vet H. C., Barendse G. A., Van Den Wildenberg F. A., & Van Kleef M. (2008). Effect of spinal cord stimulation for chronic complex regional pain syndrome type I: Five-year final follow-up of patients in a randomized controlled trial. J Neurosurg, 108(2):292-298.

Kinfe T. M., Quack F., Wille C., Schu S. & Vesper J. (2014). Paddle versus cylindrical leads for percutaneous implantation in spinal cord stimulation for failed bck surgery syndrome: A single-center trial. J Neurol Surg A Cent Eur Neurosurg, 75(6):467-473.

Kinfe T. M., Schu S., Quack F. J., Wille C., & Vesper J. (2012). Percutaneous implanted paddle lead for spinal cord stimulation: Technical considerations and long-term follow-up. Neuromodulation, 15(4):402-407.

*Kumar K., Hunter G., & Demeria D. (2006). Spinal cord stimulation in treatment of chronic benign pain: challenges in treatment planning and present status, a 22-year experience. Neurosurgery, 58(3):481-496.

Kumar K., Rizv S., & Bnurs S. B. (2011). Spinal cord stimulation is effective in management of complex regional pain syndrome I: fact or fiction. Neurosurgery, 69(3):566-578.

PMA P030017-S275: FDA Summary of Safety and Effectiveness Data

Page 21

{21}

\*Kumar K., Taylor R. S., Jacques L., Eldabe S., Meglio M., Molet J., & North R. B. (2007). Spinal cord stimulation versus conventional medical management for neuropathic pain: a multicentre randomised controlled trial in patients with failed back surgery syndrome. *Pain*, 132(1-2):179-188.

\*Kumar K., Taylor R. S., Jacques L., Eldabe S., Meglio M., Molet J., & North, R. B. (2008). The effects of spinal cord stimulation in neuropathic pain are sustained: a 24-month follow-up of the prospective randomized controlled multicenter trial of the effectiveness of spinal cord stimulation. *Neurosurgery*, 63(4):762-770.

Logé D., Vanneste S., Vancamp T., & Rijckaert D. (2013). Long-term outcomes of spinal cord stimulation with percutaneously introduced paddle leads in the treatment of failed back surgery syndrome and lumboischialgia. *Neuromodulation*, 16(6):537-545.

Mammis A. & Mogilner A. Y. (2012). The use of intraoperative electrophysiology for the placement of spinal cord stimulator paddle leads under general anesthesia. *Neurosurgery*, 70(2 Suppl Operative):230-236.

Matias C. M., Amit A., Lempka S. F., Ozinga J. G. 4th, Nagel S. J., & Lobel D. A. (2014). Long-term outcomes after replacement of percutaneous leads with paddle leads in a retrospective cohort of patients with spinal cord stimulation systems. *Neurosurgery*, 75(4):430-436.

McAuley J., Farah N., van Groningen R., & Green C. (2013). A questionnaire-based study on patients' experiences with rechargeable implanted programmable generators for spinal cord stimulation to treat chronic lumbar spondylosis pain. *Neuromodulation*, 16(2):142-146.

McRoberts W. P., Wu P., Bentley I. (2012). Effect of a novel fixation method for spinal cord stimulators. *Neuromodulation*, 16(5):449-453.

Mekhail N. A., Mathews M., Nageeb F., Guirguis M., Mekhail M. N., & Cheng J. (2011). Retrospective review of 707 cases of spinal cord stimulation: Indications and complications. *Pain Practice*, 11(2):148-153.

\*North R. B., Kidd D. H., Farrokhi F., & Piantadosi S. A. (2005). Spinal cord stimulation versus repeated lumbosacral spine surgery for chronic pain: a randomized, controlled trial. *Neurosurgery*, 56(1):98-106.

North R. B., Kidd D. H., Olin J., Sieraki J. M., Farrokhi F., Petrucci L., & Cutchis P. N. (2005a). Spinal cord stimulation for axial low back pain. *Spine*, 30(12):1412-1418.

Oakley J. C., Krames E. S., Prager J. P., Stamatos J., Foster A. M., Weiner R., & Henderson J. (2007). A new spinal cord stimulation system effectively relieves chronic, intractable pain: a multicenter prospective clinical study. *Neuromodulation*, 10(3):262-278.

Ohnmeiss D. D., & Rashbaum R. F. (2001). Patient satisfaction with spinal cord stimulation for predominant complaints of chronic, intractable low back pain. *Spine J*, 1(5):358-363.

PMA P030017-S275: FDA Summary of Safety and Effectiveness Data

Page 22

{22}

Pahapill P. A. (2014). Surgical paddle-lead placement for screening trials of spinal cord stimulation. *Neuromodulation*, 17(4):346-348.

Rigoard P., Delmotte A., D'Houtaud S., Misbert L., Diallo B., Roy-Moreau A., Bataille B. (2012). Back pain: a real target for spinal cord stimulation? *Neurosurgery*, 70(3):574-584.

Schu S., Slotty P. J., Bara G., von Knop M., Edgar D., & Vesper J. (2014). A prospective, randomised, double-blind, placebo-controlled study to examine the effectiveness of burst spinal cord stimulation patterns for the treatment of failed back surgery syndrome. *Neuromodulation*, 17(5):443-450.

Schultz D. M., Webster L., Kosek P., Dar U., Tan Y., & Sun M. (2012). Sensor-driven position-adaptive spinal cord stimulation for chronic pain. *Pain Physician*, 15(1):1-12.

Son B. C., Kim D. R., Lee S. W., & Chough C. K. (2013). Factors associated with the success of trial spinal cord stimulation in patients with chronic pain from failed back surgery syndrome. *J Korean Neurosurg Soc*, 54(6):501-506.

Spincemaille G. H., Beersen N., Dekkers M. A., & Theuvenet P. J. (2004). Neuropathic limb pain and spinal cord stimulation: Results of the Dutch prospective study. *Neuromodulation*, 7(3):184-192.

Van Buyten, J. P., Fowo S., Spincemaille G. H., Tronnier V., Beute G., Pallares J. J., & Lazorthes, Y. (2008). The restore rechargeable, implantable neurostimulator: handling and clinical results of a multicenter study. *Clin J Pain*, 24(4):325-334.

Vonhogen L.H., Vancamp T., Vanneste S., Pollet W., Dirksen R., Bakker P., de Ridder D. (2011). Percutaneously implanted plates in failed back surgery syndrome (FBSS). *Neuromodulation*, 14(4):319-324.

Wolter T., & Winkelmuller M. (2012). Continuous versus intermittent spinal cord stimulation: an analysis of factors influencing clinical efficacy. *Neuromodulation*, 15(1):13-19.

## **B. References Used Only in Effectiveness Analysis**

De La Porte C. & Siegfried J. (1983). Lumbosacral Spinal Fibrosis (Spinal Arachnoiditis): Its Diagnosis and Treatment by Spinal Cord Stimulation. *Spine*, 8(6):593-603.

Fiume D., Sherkat S., Callovini G. M., Parziale G., & Gazzeri G. (1995). Treatment of the failed back surgery syndrome due to lumbo-sacral epidural fibrosis. *Acta Neurochir Suppl*, 64:116-118.

Kumar K., Toth R. K., & Nath C. (1996). Spinal cord stimulation for chronic pain in peripheral neuropathy. *Surgical Neurology*, 46(4):363-369.

Kumar K., Nath R. K., & Toth C. (1997). Spinal cord stimulation is effective in the management of reflex sympathetic dystrophy. *Neurosurgery*, 40(3):503-509.

PMA P030017-S275: FDA Summary of Safety and Effectiveness Data

Page 23

{23}

Lame I. E., Peters M. L., Patijn J., Kessels A. G., Geurts J., & van Kleef M. (2009). Can the outcome of spinal cord stimulation in chronic complex regional pain syndrome type I patients be predicted by catastrophizing thoughts? *Anesth Analg*, 109(2):592-599.

Moriyama K., Murakawa K., Uno T., Oseto K., Kawanishi M., Saito Y., Taira T., & Yamauchi M. (2012). A Prospective, Open-Label, Multicenter Study to Assess the Efficacy of Spinal Cord Stimulation and Identify Patients Who Would Benefit. *Neuromodulation*, 15(1):7-11.

Oakley J. C., Weiner R. L. (1999). Spinal cord stimulation for complex regional pain syndrome: A prospective study of 19 patients at two centers. *Neuromodulation*, 2(1): 47-50.

Probst C. (1990). Spinal cord stimulation in 112 patients with epi-/intradural fibrosis following operation for lumbar disc herniation. *Acta Neurochir (Wien)*, 107(3-4):147-151.

Robaina F. J., Rodriguez J. L., De Vera J. A., & Martin M. A. (1989). Transcutaneous electrical nerve stimulation and spinal cord stimulation for pain relief in reflex sympathetic dystrophy. *Stereotact Funct Neurosurg*, 52(1):53-62.

Sanchez-Ledesma M. J., Garcia-March G., Diaz-Cascajo P., Gomez-Moreta J., & Broseta J. (1989). Spinal cord stimulation in deafferentation pain. *Stereotact Funct Neurosurg*, 53(1):40-45.

Sears N. C., Machado A. G., Nagel S. J., Deogaonkar M., Stanton-Hicks M., Rezai A. R., & Henderson J. M. (2011). Long-Term Outcomes of Spinal Cord Stimulation With Paddle Leads in the Treatment of Complex Regional Pain Syndrome and Failed Back Surgery Syndrome. *Neuromodulation*, 14(4):312-318.

Siegfried J. & Lazorthes Y. (1982). Long-term follow-up of DCS for chronic pain syndrome after multiple lumbar operations. *Appl Neurophysiol*, 45:201-204.

Verdolin M.H., Stedje-Larsen E.T., & Hickey A.H. (2007). Ten consecutive cases of complex regional pain syndrome of less than 12 months duration in active duty United States military personnel treated with spinal cord stimulation. *Anesth Analg*, 104(6):1557-1560.

PMA P030017-S275: FDA Summary of Safety and Effectiveness Data

Page 24

---

**Source:** [https://fda.innolitics.com/device/P030017S275](https://fda.innolitics.com/device/P030017S275)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a PMA](https://innolitics.com/services/regulatory/), [a 510(k)](https://innolitics.com/services/510ks/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).