WALTER LORENZ TOTAL TEMPOROMANDIBULAR JOINT REPLACEMENT SYSTEM
Applicant
Biomet Microfixation, Inc.
Product Code
LZD · Dental
Decision Date
Sep 21, 2005
Decision
APPR
Regulation
21 CFR 872.3940
Device Class
Class 3
Attributes
Therapeutic
Indications for Use
The Total Temporomandibular Joint(TMJ) Replacement System is indicated for reconstruction of the temporomandibular joint. The reconstruction is necessary due to one of the following diagnoses: 1. arthritic conditions: osteoarthritis, traumatic arthritis rheumatoid arthritis 2. ankylosis including but not limited to recurrent ankylosis with excessive heterotopic bone formation, 3. revision procedures where other treatments have failed (e.g. alloplastic reconstruction, autogenous grafts) 4. avascular necrosis 5. multiply operated joints 6. fracture 7. functional deformity 8. benign neoplasms 9. malignancy (e.g. post-tumor excision) 10. degenerated or resorbed joints with severe anatomic discrepancies 11. developmental abnormality
Device Story
Total TMJ Replacement System consists of mandibular condyle and glenoid fossa components; surgically implanted in jaw to reconstruct diseased/damaged TMJ. Mandibular component (Co-Cr-Mo alloy with Ti-6Al-4V plasma spray) and fossa component (UHMWPE) are fixed to bone via titanium screws. System includes trials and instruments. Used by surgeons in clinical settings to restore joint function; alleviate pain; improve maximal incisal opening. Clinical benefit demonstrated through prospective study showing significant reduction in pain/eating interference and increased jaw opening. Device requires surgeon training; post-market surveillance for long-term safety.
Clinical Evidence
Prospective, multi-center study of 224 cases (329 joints). Primary endpoints: VAS for jaw pain, VAS for eating interference, and Maximal Incisal Opening (MIO) at 3 years. Results showed statistically significant improvement (p<.0001) in all three endpoints from baseline to 3 years. 98.8% of unimputed cases met success criteria. Adverse events included infection, heterotopic bone formation, and component loosening/dislocation; 6.7% of cases required device removal.
Technological Characteristics
Two-component system: Mandibular condyle (Co-Cr-Mo per ASTM F 1537 or Ti-6Al-4V per ASTM F 136, with Ti-6Al-4V plasma spray per ASTM F 1580) and glenoid fossa (UHMWPE per ASTM F 648). Fixation via Ti-6Al-4V screws (ASTM F 136). Non-cemented design. Mechanical assembly.
Indications for Use
Indicated for skeletally mature patients requiring total TMJ reconstruction due to arthritis, ankylosis, avascular necrosis, benign/malignant neoplasms, functional deformity, revision of failed prior treatments, or fracture. Contraindicated in patients with active infection, insufficient bone quality/quantity, severe hyper-functional habits, or known material allergies.
Regulatory Classification
Identification
A total temporomandibular joint prosthesis is a device that is intended to be implanted in the human jaw to replace the mandibular condyle and augment the glenoid fossa to functionally reconstruct the temporomandibular joint.
Submission Summary (Full Text)
{1}
SUMMARY OF SAFETY AND EFFECTIVENESS DATA
I. General Information
Device Generic Name: Total temporomandibular joint implant
Device Trade Name: Total Temporomandibular Joint (TMJ) Replacement System
Applicant's Name: Walter Lorenz Surgical Incorporated
1520 Tradeport Drive
Jacksonville, Florida 32218-2480
Prenarket Approval Application (PMA) Number: P020016
Date of Panel Recommendation: August 22, 2002
Date of Notice of Approval to the Applicant: September 21, 2005
II. Indications for Use
The Total Temporomandibular Joint(TMJ) Replacement System is indicated for reconstruction of the temporomandibular joint. The reconstruction is necessary due to one of the following diagnoses:
1. arthritic conditions: osteoarthritis,
traumatic arthritis
rheumatoid arthritis
2. ankylosis including but not limited to recurrent ankylosis with excessive heterotopic bone formation,
3. revision procedures where other treatments have failed (e.g. alloplastic reconstruction, autogenous grafts)
4. avascular necrosis
5. multiply operated joints
6. fracture
7. functional deformity
8. benign neoplasms
9. malignancy (e.g. post-tumor excision)
10. degenerated or resorbed joints with severe anatomic discrepancies
11. developmental abnormality
{2}
12
# III. Device Description
The Total TMJ Replacement System is a two component system comprised of mandibular condyle and glenoid fossa components. Both components are available in multiple sizes as right and left side specific designs and are attached to bone by screws. The individual components are not for use in partial joint reconstruction. The Total Temporomandibular Joint (TMJ) Replacement System is implanted in the jaw to functionally reconstruct a diseased and/or damaged temporomandibular joint. Included in the system are trials, instruments and instrument cases.
## Materials:
Mandibular Component – Cobalt-Chromium-Molybdenum (Co-Cr-Mo) alloy per ASTM F 1537 with titanium alloy (Ti-6Al-4V) powder per ASTM F 1580) plasma spray coating or Titanium (Ti-6Al-4V) alloy per ASTM F 136 with titanium alloy (Ti-6Al-4V) powder per ASTM F 1580) plasma spray coating
Fossa Component – ArCom® ultra-high-molecular-weight (UHMWPE) per ASTM F 648
Screws – Titanium alloy (Ti-6Al-4V per ASTM F 136)
Trials – mandibular- aluminum fossa- Radel® plastic
Instruments – TMJ flat diamond rasp, TMJ diamond burs, TMJ double ended drill guide, retractors - stainless steel
Instrument Case – stainless steel, silicone, Radel® plastic
# IV. Contraindications
1. Active or chronic infection.
2. Patient conditions where there is insufficient quantity or quality of bone to support the components.
3. Systemic disease with increased susceptibility to infection.
4. Patients with extensive perforations in the mandibular fossa and/or bony deficiencies in the articular eminence or zygomatic arch that would severely compromise support for the artificial fossa component.
5. Partial TMJ joint reconstruction.
6. Known allergic reaction to any materials used in the components.
NOTE: Patients with known or suspected nickel sensitivity should not have Co-Cr-Mo devices implanted since this material contains nickel.
7. Patients with mental or neurological conditions who are unwilling or unable to follow postoperative care instructions.
8. Skeletally immature patients.
9. Patients with severe hyper-functional habits (e.g. clenching, grinding etc.)
10. Patients with a foreign body reaction due to previous implants.
{3}
3
13
# V. Warnings and Precautions
The warnings and precautions can be found in the Total TMJ Replacement System labeling.
# VI. Alternative Practices and Procedures
Alternative practices and procedures include autogenous or allogeneic bone grafting and implantation of other marketed devices for total TMJ reconstruction.
# VIII. Potential Adverse Effects
Adverse events that may occur following placement of the Total TMJ Replacement System are listed below. See Tables 7 and 8 for more detailed information on adverse events from the clinical trial.
- Removal of components(s) including, but not limited to the following:
- implant changes caused by loading and/or wear
- degenerative changes within the joint surfaces from disease or previous implants
- implant materials producing particles or corroding
- Loosening or displacement with or without removal of the implant
- Infection (systemic or superficial)
- Foreign body or allergic reaction to implant components
- Fossa wear through
- Facial swelling and/or pain
- Facial nerve dysfunction
- Excision of tissue
- Heterotopic bone formation
- Neuroma formation
- Ear problems
- Dislocation
# IX. Marketing History
Approval for marketing has been granted by Europe (EC-Certificate issued November 23, 2000). The system has been marketed in South Africa since January 2000. The medical device license for marketing from Canada was issued on January 14, 2004. The device has not been withdrawn from marketing for any reason relating to the safety and effectiveness of the device.
# X. Summary of Preclinical Studies
Shelf Life and Package Tests
{4}
All packages were gamma irradiated twice (2 X) for the purpose of package validation.
A. Bubble Emission Testing was performed: packages inflated to ~4 psi and packages were submerged for 2 minutes and observed for bubble leaks.
B. Burst Testing was performed: packages were slowly pressurized with air to the equivalent of submersion under 22.8 inches of water and held for 10 seconds, and then filled to burst.
C. Distribution Simulation: in accordance with ASTM D 4169, Distribution Cycle 13 (an appropriate cycle), Assurance Level I (the most rigorous level). Repeated 3 times, with real or mock devices.
D. Accelerated Aging was performed: at 55°C. At this temperature 37 days is equivalent to 1 year of real-time aging, and their 76 days exceeds the 2 year equivalent of 74 days by 2 additional days.
E. Real-time Aging was performed: on a significant number of fossa and mandibular packages. This was followed by Bubble and Burst testing, but no simulated shipping and handling was included.
F. Microbial Challenge test: performed on 47 Fossa and 47 Mandibular packages as baseline. Only 60 Fossa packages were tested after both accelerated aging and 3X Distribution Simulation
Four packages failed the Bubble Emission test after exposure to both 2 years of accelerated aging and 3X Distribution Simulation. All other test packages met the established acceptance criteria, including visual inspection. Package seals were strong and consistently met the acceptance criteria. A shelf life of one year was established for the Total TMJ Replacement System.
## Biomechanical Tests
The following biomechanical tests were conducted on the Total TMJ Replacement System. Test results were all determined to be sufficient for the intended use of the construct/component.
A. Fatigue Testing of Fossa and Mandibular Component Construct
B. Static Testing of the Mandibular Component
C. Fossa Screw Head Pull-Through Test
D. Compression Strength of Fossa Component Flange
E. 2.7mm Self-Tapping Screw Pull-Out Strength
## A. Fatigue Testing of Fossa and Mandibular Component Construct
Initial fatigue testing was performed on five joints with mandibular components minus the titanium plasma spray coating. No failures were seen after 10 million cycles at a maximum load of 145 lbs. at frequencies between 10 and 30 hertz. The same testing was repeated on four joints with mandibular components coated with titanium plasma spray. There were no failures after 10 million
14
{5}
cycles and no porous coating delamination was observed as expected from previous testing on orthopedic devices.
## B. Static Strength Testing of the Mandibular Component
A mandibular component was fixed to porcine bone using four 2.7mm diameter screws and a load was applied to ultimate failure. At 575.9 lbs. the component’s neck portion bent with neither fracture/pull-out of the bone screws or fracture of the component.
## C. Fossa Screw (2.0mm) Head Pull -Through Test
Twelve specimens were tested to determine the force required to pull the fossa screw head through the UHMWPE zygomatic arch flange of the fossa component. A standard static tensile test was performed using a cross-head travel rate of 0.05"/minute and the ultimate tensile loads were recorded. The mean tensile strength was 79.8 ± 2.5 lbs.
## D. Compression Strength of Fossa Component Flange
A fossa component was tested to establish the load required to collapse an unsupported fossa body and assure that failure in this fashion does not cause tearing or cracking of the UHMWPE junction between the body and flange of the fossa component. The fossa body collapsed against the flange at 83 lbs. without material failure at the body/flange junction.
## E. 2.7mm Self-Tapping Screw Pull-Out Strength
Five 2.7mm screws used to fixate mandibular components were tested for pull-out strength in fresh frozen bovine cortical bone. This substrate was chosen to mimic the clinical application. The mean pull-out strength was 373.2 ± 68.8 lbs.
## XI. Summary of the Clinical Studies
### A. Objective
The study was designed to obtain clinical data to support the safety and effectiveness of this device.
### B. Inclusion / Exclusion Criteria
Inclusion Criteria:
1. Patients requiring total joint reconstruction due to:
arthritis (osteo-, rheumatoid, traumatic)
ankylosis
avascular necrosis
benign neoplasms
multiple operated joints
malignancy
functional deformity
revisions
fracture
2. Patients who are skeletally mature.
{6}
3. Patients must have at least one of the following criteria for surgical TMJ treatment.
a. presence of considerable pain and/or limited function in the joint area.
b. clinical and imaging evidence consistent with anatomic joint pathology.
c. previous failure of non-surgical treatment/therapy or a failed implant.
d. high probability of patient improvement by surgical treatment.
4. Patients must be able to return for follow-up examinations.
5. Patients without serious compromising general medical conditions.
## Exclusion Criteria:
1. Patients with active infection.
2. Patient conditions where there is insufficient quantity or quality of bone to support the device.
3. Patients with perforations in the mandibular fossa and/or bony deficiencies in the articular eminence compromising support for the artificial fossa component.
4. Patients with mandibular and/or zygomatic arch screw holes compromising component fixation.
5. Patients requiring partial joint reconstruction or other TMJ procedures not listed as an indication.
6. Patients who are not skeletally mature.
7. Patients who are incapable or unwilling to follow postoperative care instructions.
8. Patients who are unable to return for follow-up examinations.
9. Patients with severe hyper-functional habits.
10. Patients on chronic steroid therapy.
## C. Patient Population and Demographics
A total of 224 cases (329 joints) with a mean patient age of 40 years (range 13-82 years) were enrolled into the study. There were 198 females (88%) and 26 males (12%) comprised of 105 (47%) bilateral cases and 119 (53%) unilateral cases. Of the 119 unilateral cases, 53 (45%) are the right side and 66 (55%) are left sides only. Demographic data are summarized in Table 1. Most cases had multiple diagnoses with osteoarthritis and ankylosis being the most common. See Table 2 for a complete listing of diagnoses.
The mean duration of symptoms prior to implantation with this device was 11 years (range 0.1-40 years) with the mean number of 4.8 (range 0-29) prior surgeries.
Patients were categorized according to the Wilkes Classification. There were 3 (1%) cases in Class I, 1 (1%) in Class II, 8 (4%) in Class III, 90 (40%) cases in Class IV, and 122 (54%) cases in Class V.
6
{7}
7
17
# TABLE 1
## Demographic Characteristics
| | Total Cases
n=224 |
| --- | --- |
| Age (years) | |
| Mean | 40.3 |
| Standard Deviation | ±10.6 |
| Range | 13-82 |
| Gender | |
| Female | 198 (88.3%) |
| Male | 26 (11.7%) |
| Side | |
| Unilateral | |
| Right | 53 (23.7%) |
| Left | 66 (29.5%) |
| Bilateral | 105 (46.9%) |
# TABLE 2
## Diagnosis
| | Total Cases
Right Side
n=158 | | Total Cases
Left Side
n=171 | |
| --- | --- | --- | --- | --- |
| | n | % | n | % |
| 1. Osteoarthritis | 93 | 28% | 107 | 30% |
| 2. Rheumatoid Arthritis | 9 | 3% | 12 | 3% |
| 3. Traumatic Arthritis | 60 | 18% | 64 | 18% |
| 4. Malignancy | 0 | 0% | 0 | 0% |
| 5. Benign Neoplasm | 1 | 0% | 1 | 0% |
| 6. Functional Deformity | 9 | 3% | 9 | 2% |
| 7. Revision: partial implant | 8 | 2% | 11 | 3% |
| 8. Revision: total implant | 45 | 14% | 49 | 14% |
| 9. Avascular Necrosis | 42 | 13% | 42 | 12% |
| 10. Ankylosis | 46 | 14% | 50 | 14% |
| 11. Fracture | 16 | 5% | 16 | 4% |
## D. Evaluation Schedule
Patients were evaluated preoperatively and postoperatively at 1 month, 3 months, 6 months, 1 year, 1.5 years, and 3 years. All data collected past the 3 years follow-up are included. The assessments carried out at each visit
{8}
labeled as Visit 1-Visit 11 are summarized in Table 3.
TABLE 3
Study Visits Schedule
| | Vs 1: Screen Base-line | Vs 2: Surgery | Vs 3: 1 month follow-up | Vs 4: 3 month follow-up | Vs 5: 6 month follow-up | Vs 6: 1 year follow-up | Vs 7: 1.5 years follow-up | Vs 8: 3 years follow-up | Vs 9: 4 years follow-up | Vs 10: 5 years follow-up | Vs 11: 6 years follow-up |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Inclusion/ Exclusion criteria met | X | | | | | | | | | | |
| Informed Consent | X | Or X | | | | | | | | | |
| Preoperative Record: Medical history clinical examination | X | | | | | | | | | | |
| Radiographic Assessment | X | | X | X | X | X | X | X | X | X | X |
| Jaw pain & function: Jaw pain intensity, Interference with eating, Maximal incisal opening, Occlusion, Anterior open bite, cross bite | X | | X | X | X | X | X | X | X | X | X |
| Operative Record | | X | | | | | | | | | |
| Patient satisfaction with surgery | | | X | X | X | X | X | X | X | X | X |
| Wound healing | | | X | X | X | X | X | X | X | X | X |
* immediate postoperative x-ray used for comparison only.
## E. Study Design
The study was a prospective, multi-center, single treatment study. It was designed to compare baseline clinical and radiographic assessments to assessments made postoperatively.
## F. Patient Accountability
Table 4 shows the number and percentage of cases with follow-up data at each of the visits. Compliance ranged from 91.0% at the 1 month follow-up visit to 72.4% at 3 years follow-up.
{9}
9
19
# TABLE 4
## Patient Accountability
| | Follow-Up Time Periods | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | 1mo | 3mos | 6mos | 1yr | 1.5yr | 3yrs | 4yrs | 5yrs | 6yrs |
| Theoretically Due (all cases) | 213 | 204 | 199 | 179 | 164 | 123 | 81 | 47 | 35 |
| Deaths | 1 | 1 | 1 | 1 | 1 | 2 | 2 | 3 | 3 |
| Permanent Removal of Total Joint | 0 | 0 | 1 | 2 | 2 | 2 | 2 | 1 | 0 |
| Have Follow-Up (all cases) | 193 | 181 | 177 | 150 | 129 | 85 | 48 | 20 | 14 |
| Percent Follow-Up | 91.0 | 89.2 | 89.8 | 85.5 | 80.1 | 72.4 | 64.2 | 51.1 | 48.6 |
| | | | | | | | | | |
| Have Follow-Up (all joints)
right side only | 44 | 46 | 45 | 42 | 39 | 26 | 15 | 4 | 4 |
| left side only | 56 | 50 | 52 | 42 | 29 | 25 | 14 | 7 | 5 |
| bilateral | 93 | 85 | 80 | 66 | 61 | 34 | 19 | 9 | 5 |
| Total # of joints | 286 | 266 | 257 | 216 | 190 | 119 | 67 | 29 | 19 |
## G. Efficacy and Safety Parameters
1. Primary efficacy endpoints include:
- Jaw pain intensity as measured on a 10 cm visual analogue scale (VAS) from preoperative assessment to assessment 3 years postoperative, adjusted for baseline at preoperative assessment,
- Interference with eating as measured on a 10 cm VAS from preoperative assessment to assessment 3 years postoperative, adjusted for baseline at preoperative assessment,
- Maximal incisal opening (MIO) measurement (in mm) from preoperative assessment to assessment 3 years postoperative, adjusted for baseline at preoperative assessment
### Patient and Study Success
a. Patient Success
A patient was determined to be a success if:
1. patient has not had a permanent total joint removal, and
2. patient meets two of the following three criteria:
- reduction of pain by 1 cm (VAS) from baseline to 3 years follow-up
- reduction of interference with eating by 1 cm (VAS) from baseline to 3 years follow-up
- increase in MIO of 10% from baseline to 3 years follow-up
{10}
b. Study Success
The study was deemed to be a success with 60% or more of the patients receiving the device having met the above Patient Success at 3 years follow-up.
In the cohort unimputed group, 84 of 85 (98.8%) cases are patient successes. In the cohort imputed group, 116 of 119 (97.5%) cases are patient successes. These patient success rates surpass the criteria for study success.
Analysis was performed on cases with 3 years follow-up postoperatively. These cases were defined as two groups. One is the cohort unimputed group comprised of 85 cases and the second group, cohort imputed, is comprised of 119 cases. The cohort imputed group used data points obtained at the follow-up visit closest to but not after the 3 years visit for analysis of the 34 cases missing data at the 3 years visit. The primary endpoints are summarized on the following table.
Table 5
Analysis of cases with 3 year follow-up
| Primary Efficacy Endpoints | Cohort Imputed Cases n=119 | Cohort Unimputed Cases n=85 |
| --- | --- | --- |
| | Difference between Vs 1 & Vs 8 ± SD | Difference between Vs 1 & Vs 8 ± SD |
| Jaw pain | 5.69 ± 2.33 cm | 6.03 ± 2.12 cm |
| Interference with eating | 5.42 ± 2.58 cm | 5.60 ± 2.32 cm |
| MIO | 10.69 ± 8.22 mm | 10.16 ± 8.72 mm |
These primary efficacy endpoints showed a significant improvement from baseline to 3 years postoperative. Multiple analyses (t-test and repeated measures) demonstrate that significant improvement is evidenced after implantation of the Total TMJ Replacement System, same patterned effect for the cohort imputed and unimputed groups.
Further t-test analysis shows that in both the total group (n = 224) and the cohort imputed group (n = 119), there was a statistical difference (p<.0001) in all three primary endpoints between baseline (Vs 1) and assessments at all time points from 1 month follow-up to 3 years follow-up.
Figures 1, 2, and 3 graphically display the three primary endpoints for the total study group and the two cohort groups from baseline to the 3 years visit.
{11}
# Comparison of Total Group (n=224) to 3 Year Cohort Unimputed and 3 Year Cohort Imputed
## Comparison Means per Visit
### Groups on Jaw Pain Intensity
Figure 1
Jaw Pain Intensity
| VAS 10 cm | baseline | 1 mo FUp | 3 mo FUp | 6 mo FUp | 1 yr FUp | 1.5 yr FUp | 3 yr FUp |
| --- | --- | --- | --- | --- | --- | --- | --- |
| total study group (224) | 8.5 | 4.6 | 3.7 | 3.4 | 3.1 | 3.4 | 2.8 |
| 3 yr cohort unimputed | 8.9 | 4.8 | 3.7 | 3.5 | 3.1 | 3.2 | 2.8 |
| 3 yr cohort imputed | 9 | 4.8 | 3.9 | 3.7 | 3.3 | 3.4 | 2.9 |
visits
11
21
{12}
# Comparison of Total Group (n=224) to 3 Year Cohort Unimputed and 3 Year Cohort Imputed
## Comparison Means per Visit
### Interference with Eating

Figure 2
Interference with Eating
{13}
# Comparison of Total Group (n=224) to 3 Year Cohort Unimputed and 3 Year Cohort Imputed
## Comparison Means per Visit
### Maximal Incisal Opening
Figure 3

2. Secondary efficacy endpoints include (Vs 1 comparison to Vs 3 – 8):
- Jaw pain intensity, interference with eating, and maximal incisal opening
- Patient satisfaction, with a focus on the comparison from postoperative baseline (Vs 5) to 3 years follow-up (Vs 8):
- in hindsight, whether the patient would choose to have this surgery;
- degree of satisfaction with surgery across time
The secondary efficacy endpoints demonstrated a gradual improvement over time in terms of jaw pain, interference with eating, and MIO. Table 6 lists the means and standard deviation for the three endpoints at visits Vs 1 - Vs 11.
13
23
{14}
24
# TABLE 6
## Secondary Efficacy Endpoints
Visits 1 - 11
| Visit (interval) | N | Jaw Pain
Mean ± SD | Interference with Eating
Mean ± SD | MIO
Mean ± SD |
| --- | --- | --- | --- | --- |
| Vs 1 (baseline) | 224 | 8.5 ± 2.3 | 8.5 ± 1.6 | 20.1 ± 10.0 |
| Vs 3 (1mo) | 193 | 4.6 ± 2.4 | 4.4 ± 2.3 | 24.9 ± 5.8 |
| Vs 4 (3 mos) | 181 | 3.7 ± 2.5 | 3.5 ± 2.4 | 28.5 ± 5.8 |
| Vs 5 (6 mos) | 177 | 3.4 ± 2.3 | 3.2 ± 2.4 | 29.4 ± 6.1 |
| Vs 6 (1 year) | 150 | 3.1 ± 2.4 | 3.0 ± 2.3 | 30.1 ± 5.8 |
| Vs 7 (1.5 yr) | 128 | 3.4 ± 2.3 | 3.2 ± 2.5 | 29.6 ± 6.1 |
| Vs 8 (3 yrs) | 85 | 2.8 ± 2.1 | 2.8 ± 2.0 | 29.3 ± 6.0 |
| Vs 9 (4 yrs) | 48 | 3.5 ± 2.4 | 3.4 ± 2.6 | 28.4 ± 6.6 |
| Vs 10 (5 yrs) | 20 | 4.0 ± 2.7 | 4.3 ± 2.3 | 28.9 ± 6.8 |
| Vs 11 (6 yrs) | 14 | 3.7 ± 2.1 | 3.2 ± 2.0 | 26.8 ± 5.9 |
Note: Visit 2 is the day of surgery.
Most patients were satisfied with their outcome as demonstrated with over 90% of cases reporting at least satisfied or better at every follow-up visit. Furthermore, over 90% of the cases in hindsight would choose to have this surgery at all time points. More specifically for Vs 3 – Vs 8, between 94–99% of the cases said yes to the question “In hindsight would you choose to have this surgery?”
## 3. Safety
a. Radiographic assessment (position of components, heterotopic bone formation, osseous erosion, fossa resorption) was performed at each follow-up visit.
The position of mandibular and fossa components and the mandibular and fossa screws were assessed by investigators in comparison to immediate postoperative radiographs. There were three mandibular components reported as having a change in position: one at Vs 4 and two at Vs 8. The case noted at Vs 4 also had a change of position of the mandibular screws and the joint was removed at 6 months postoperative. No change of position was reported for fossa screws.
Heterotopic bone formation was found in 15 joints, 8 rights and 7 left joints. There are no reports of osseous erosion or fossa resorption.
b. Adverse events
Adverse Events (AEs) were documented for all cases throughout the duration of the study. There have been no unanticipated device related
{15}
adverse events reported. Overall, 121 AEs were reported in 80 cases (35.7 %) of the 224 cases. Three cases (1.3 %) terminated the study due to their permanent total joint removal AEs. Table 7 summarizes AEs requiring device removal. Table 8 summarizes AEs not requiring device removal.
TABLE 7
Adverse Events Requiring Device Removal
| Device Removals | Cases (n=224) | | Joints (n=329) | |
| --- | --- | --- | --- | --- |
| | # | % | # | % |
| 1. Permanent removal of fossa component: | 5 | 2.2 % | 6 | 1.8 % |
| a. One due to aseptic necrosis | | | | |
| b. Two due to infection | | | | |
| c. One due to swelling | | | | |
| d. One due to heterotopic bone removal | | | | |
| 2. Removal (non-permanent)^{1} of mandibular component: | 5 | 2.2 % | 9 | 2.7 % |
| a. Two bilateral removals of heterotopic bone | | | | |
| b. One due to dislocation | | | | |
| c. Two due to reposition for malocclusion | | | | |
| 3. Permanent removal of mandibular component: | 1 | 0.4% | 1 | 0.3% |
| a. Larger component causing a dislocation removed and replaced with smaller component | | | | |
| 4. Permanent removal of total joint: | 4 | 1.8 % | 4 | 1.2 % |
| a. One unilateral patient requested removal due to pain and swelling after 6 months | | | | |
| b. Three removals due to infection | | | | |
| Permanent removal | 10 | 4.5 % | 11 | 3.3 % |
| Non-permanent removal | 5 | 2.2 % | 9 | 2.7 % |
| TOTAL | 15 | 6.7 % | 20 | 6.1% |
1 Mandibular components were taken out in the operating room for removal of heterotopic bone or re-positioning and then were placed back in the joint.
15
25
{16}
26
TABLE 8
Adverse Events Not Requiring Device Removal
| Adverse Events | Cases (n=224) | | Joints (n=329) | |
| --- | --- | --- | --- | --- |
| | # | % | # | % |
| Reflex Sympathetic Dystrophy (RSD) | 1 | 0.4 | 1 | 0.3 |
| Excision of tissue
(excluding neuroma and/or heterotopic bone) | 4 (10)* | 1.8 (4.5) | 6 | 1.8 |
| Heterotopic bone excision | 4 (9) | 1.8 (4.0) | 6 | 1.8 |
| Chronic severe masseter muscle spasms | 2 | 0.9 | 3 | 0.9 |
| Motor vehicle accident (MVA)
- increased pain regardless of facial impact | 14 | 6.3 | 22 | 6.7 |
| Facial trauma (excluding MVA) | 9 | 4.0 | 10 | 3.0 |
| Head trauma with no jaw involvement | 2 | 0.9 | 3 | 0.9 |
| Neuroma excision | 12 (13) | 5.4 (5.8) | 15 | 4.6 |
| Death (all unrelated) | 3 | 1.3 | 3 | 0.9 |
| Coronoidectomy | 16 (17) | 7.1 (7.6) | 25 | 7.6 |
| Unrelated disease diagnosis (multiple sclerosis, Multiple myeloma, meningitis) | 3 | 1.3 | 5 | 1.5 |
| Abscess (stitch/facial/intraoral) | 3 | 1.3 | 5 | 1.5 |
| Skin infection (not in area of prosthesis) | 1 | 0.4 | 2 | 0.6 |
| Dislocation (mandible) | 1 | 0.4 | 1 | 0.3 |
| Ear infection
(two with tympanic membrane perforation) | 5 | 2.2 | 8 | 2.4 |
| External ear canal problems:
1. Perforation
2. Granulation formation | 2 | 0.9 | 2 | 0.6 |
| Scalp alopecia from anesthesia tubing pressure | 1 | 0.4 | 2 | 0.6 |
| Muscle tenderness | 1 | 0.4 | 2 | 0.6 |
| Decreased range of motion | 1 | 0.4 | 1 | 0.3 |
| Allergy to resorbable sutures | 1 | 0.4 | 2 | 0.6 |
| Contralateral Subcondylar osteotomy for pre-existing disease | 1 | 0.4 | 1 | 0.3 |
| Patient reported episodic “floaters” in right eye | 1 | 0.4 | 2 | 0.6 |
| Dysesthesia of pre-auricular scar | 1 | 0.4 | 1 | 0.3 |
| Ankylosis | 2 | 0.9 | 3 | 0.9 |
| Facial numbness | 1 | 0.4 | 2 | 0.6 |
| Loose fossa screw | 1 | 0.4 | 2 | 0.6 |
| Fistula | 1 | 0.4 | 1 | 0.3 |
| Total Cases | 94 | 42.0% | 136 | 41.3% |
| Total Incidence | (107) | (47.8) | | |
* These numbers in parenthesis ( ) are the incidence.
H. Safety Analysis
1. Deaths
There have been three deaths reported in the study, none of which were device related.
{17}
27
# 2. Revisions/Removals
## a. Total joint removed
1. Case # 20
Bilateral patient first had the right fossa component removed (10 months postoperative) due to infection. A year later the right mandibular component was removed also due to infection. The right side (case #103) was re-implanted 7 months later. Case # 20 is now a left side only.
2. Case # 61
Unilateral patient first had her fossa removed 10 months postoperative and subsequent mandibular removal 6 months later due to infection. This case is lost to follow-up.
3. Case # 100
Unilateral (right side) patient had removal of prosthesis at 6 months postoperative due to chronic swelling and pain. This case is lost to follow-up.
4. Case#242
Bilateral patient had left prosthesis removed 2 months post-op due to chronic infection. The left side prosthesis was re-implanted and is now case #250. Case #242 is now right side only.
## b. Fossa component only revised/removed
1. Case # 1
Bilateral patient had removal of left fossa component due to aseptic necrosis at almost 2 years postoperative and 3.5 years after the removal had it replaced.
2. Case # 13
Fossa component was removed secondary to infection in the ear canal at 2.5 years postoperative and was replaced a month later.
3. Case # 19
Fossa component removed 4 years postoperative due to a late infection of the ear.
4. Case # 44
Bilateral fossa components replaced because they were damaged during surgery to remove heterotopic bone.
5. Case # 117
At 11 months postoperative the fossa was removed to see if this would decrease swelling. There were no signs of infection but heavy encapsulation was noted.
## c. Mandibular component only revised
1. Case # 183
{18}
This bilateral case was treated for an anterior dislocation by removing the right 50mm mandibular component and replacing it with a 45mm component.
## 3. Additional Safety Measurement
### a. Surgical Site (wound healing)
Most surgical wounds healed by 3 months postoperative with 100% (right side) and 98% (left side) healed. Redness and drainage accompanied with infection are documented as adverse events.
## XII. Conclusions Drawn from Studies
### Preclinical
The results of the pre-clinical studies demonstrate the Total Temporomandibular Joint Replacement System has adequate strength and durability for its intended use. The shelf life and package testing resulted in a shelf life of 1 year.
### Safety
The types of adverse events reported in the clinical study and the rate at which they occurred are not unexpected in this compromised patient population with many previous surgeries involving failed tissue grafts and/or failed implants which may leave behind material particulates.
### Efficacy
The clinical study showed that for patients with complete data at the 3 year follow-up (85 patients) the Total Temporomandibular Joint Replacement System provided statistically significant levels of reduced jaw pain, reduced interference with eating and increased maximal incisal opening. Similar trends, although not statistically significant were observed in the entire patient population. The cohort patients are representative of the target patient population.
## XIII. Panel Recommendation
At an advisory meeting held on August 22, 2002, the Dental Products Panel recommended that Walter Lorenz’s PMA for the Total TMJ Replacement System be approved subject to the following conditions:
1. The labeling should provide a clearer description of hyperfunctional habits such as clenching or bruxing and this information should be addressed in a different location in the labeling (i.e., moved from the contraindication section to the warnings or precaution section).
18
28
{19}
2. The Indications for Use section should state the indications for which the device has been well tested. For indications that were not evaluated in the clinical study, the Indications for Use section should state that the device has not been adequately evaluated for these indications.
3. Information regarding the potential for foreign body reaction should be included in patient labeling and physician's information.
4. The sponsor should remove all references to the use of a cemented fossa and state that the device will be marketed as a non-cemented device.
5. All surgeons implanting these devices should be required to receive didactic and hands-on training before they are able to use the device.
6. The following additional in vitro and in vivo testing should be performed:
a. The sponsor should perform wear testing that simulates the temporomandibular joint
b. The sponsor should test the fossa components for possible changes in fixation stability due to creep.
c. All explants should be retrieved and studied for wear, creep, and possible corrosion due to use of dissimilar metals.
d. For explant cases, the sponsor should perform histologic examination for wear particles.
7. The sponsor should submit data from the fatigue testing of the fossa with the post removed and the fossa made without a post to FDA.
8. The sponsor will seek full or partial data on all 180 cases, including retrieving VAS scores from patients at long distances and collecting full or partial post-market data. All 180 cases presently included in the study should be followed for 3 years for safety and effectiveness.
XIV. FDA Decision
CDRH concurred with the panel recommendations except for the engineering recommendation of wear testing that simulates the temporomandibular joint. This recommendation is not feasible given the lack of an adequate model for the loading of the temporomandibular joint. The labeling, training and other engineering recommendations have been completed.
A postapproval study will be conducted in order to collect additional long-term safety and effectiveness data. Three year follow-up data will be obtained on all subjects enrolled in the clinical study. Reports will be submitted to the PMA
29
{20}
annually. The labeling will be updated via a supplement, when the study is complete.
The applicant’s manufacturing faculties were inspected and found to be in compliance with the Quality System Regulations (21 CFR 820).
CDRH issued an approval order on September 21, 2005.
## XV. Approval Specification
- Directions for use: See the labeling.
- Hazards to Health from Use of the Device: See Indications, Contraindications, Warnings, Precautions and Adverse Events in the labeling.
- Postapproval Requirements and Restrictions: See approval order.
20
30
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.