P010065 · E Med Future · MTV · Mar 14, 2003 · General Hospital
Device Facts
Record ID
P010065
Device Name
NEEDLE ZAP (R)
Applicant
E Med Future
Product Code
MTV · General Hospital
Decision Date
Mar 14, 2003
Decision
APRL
Regulation
21 CFR 880.6210
Device Class
Class 2
Indications for Use
The NeedleZap™ product is a portable needle destruction device that is intended for use by health care professionals to destroy previously used aluminum and stainless steel needles (gauges 16-30, ¾-2 inches in length) attached to syringes. The NeedleZap™ product is intended for use in health care facilities and treatment settings.
Device Story
Portable needle destruction device; battery-operated (two 2V D-size rechargeable batteries). User inserts used syringe needle vertically into top opening; rotates for 2-4 seconds. Device delivers 2.5A current between two angled copper electrodes to incinerate needle tip down to hub. Debris collected in internal magnetic stainless steel tray. Used in health care facilities by clinicians to disable sharps; reduces needlestick risk. Output is blunted/sealed needle tip. Device requires recharging via 6V charger (3-4 hour cycle). Safety features include flame-retardant ABS housing, rubber feet for stability, and 30A circuit breaker.
Clinical Evidence
Clinical studies involved 7 investigators (720 attempts) and hospital/nursing home settings (164 attempts). Primary endpoint: successful needle destruction/disabling. Results showed 100% destruction rate in study 1; average burn grade 2.65/3.0 in study 2. No adverse events reported. Bench testing confirmed no infectious aerosol generation (B. stearothermophilus), no hazardous toxic fumes, external temperatures <35°C, and compliance with UL 2601-1/EN 60601-1 electrical safety standards.
Technological Characteristics
Materials: ABS flame-retardant plastic housing, copper electrodes, magnetic stainless steel debris tray. Power: 2x 2V D-size rechargeable batteries (4.5V nominal). Dimensions: 3" x 4" x 2"; weight 1.4 lb. Principle: Electrical incineration/arcing. Standards: UL 2601-1, EN 60601-1. Connectivity: None (standalone). Sterilization: N/A (device is for destruction of sharps).
Indications for Use
Indicated for health care professionals to destroy used aluminum and stainless steel syringe needles (16-30 gauge, ¾-2 inches long) in health care facilities. Contraindicated for use near flammable liquids, gases, oxygen, operating rooms, or explosives, and for needles containing explosive or radioactive materials.
Regulatory Classification
Identification
A sharps needle destruction device is a prescription device that is intended to destroy needles or sharps used for medical purposes by incineration or mechanical means.
Special Controls
*Classification.* Class II (special controls). The special controls for this device are:(1) Performance testing must demonstrate the following during operation of the device:
(i) The device safely contains or ventilates aerosols or fumes from device operation.
(ii) Excessive heat or sparks are not generated that may injure users or patients.
(iii) Simulated use testing must demonstrate sharps and/or needles are completely destroyed using a range of types and sizes of sharps sufficient to represent actual use.
(iv) Simulated use testing must demonstrate that the device is physically stable on the surface for which it is intended to be mounted to ensure the risk of harm to the patient/user as a result of the device falling is minimized.
(2) Validation of cleaning and disinfection instructions must demonstrate that the device can be safely and effectively reprocessed after use per the recommended cleaning and disinfection protocol in the instructions for use.
(3) Analysis and/or testing must validate electromagnetic compatibility and electrical safety, including the safety of any battery used in the device, under conditions which are consistent with the intended environment of device use.
(4) Software verification, validation, and hazard analysis must be performed.
(5) Labeling must include:
(i) A clear description of the device and its technological features;
(ii) How the device is to be used, including validated cleaning and disinfection instructions;
(iii) Relevant precautions and warnings based on performance and in-use testing to ensure proper use of the device; and
(iv) Instructions to install device in adequately ventilated area and stable area.
Submission Summary (Full Text)
{0}
# SUMMARY OF SAFETY AND EFFECTIVENESS DATA
## I. GENERAL INFORMATION
Device Generic Name: Needle Destruction Device
Device Trade Name: NeedleZap™
Applicant Name and Address: E Med Future, Inc.
Robert Ochsendorf, President
794 Morrison Road
Columbus, Ohio 43230
PMA Application #: P010065
Date of Good Manufacturing Inspection: February 7 and 13, 2003
Date of Approval: March 14, 2003
## II. INDICATIONS FOR USE
The NeedleZap™ product is a portable needle destruction device that is intended for use by health care professionals to destroy previously used aluminum and stainless steel needles (gauges 16-30, ¾-2 inches in length) attached to syringes. The NeedleZap™ product is intended for use in health care facilities and treatment settings.
## III. CONTRAINDICATIONS
A. DO NOT USE THIS DEVICE NEAR FLAMMABLE LIQUIDS, GASES, OXYGEN, OPERATING ROOMS, OR EXPLOSIVES. NeedleZap™ can produce minor sparking which could create the risk of explosion or fire in this type of environment.
B. DO NOT DESTROY NEEDLES THAT CONTAIN EXPLOSIVE OR RADIOACTIVE MATERIALS.
## IV. WARNINGS AND PRECAUTIONS
A. DO NOT use two hands to operate the unit. It is designed for simple one-handed use to avoid your other hand becoming a target.
B. DO NOT use this product near a sink, washtub or other location where it could be immersed or subjected to high moisture.
C. DO NOT open the unit. Should your product require service, return it to the manufacturer.
{1}
D. DO NOT insert needle into the charger terminal (see Figure B below).
E. DO NOT operate while the unit is connected to charger. NeedleZap™ is designed as a portable device.
F. Should your device ever malfunction, maintain your level of caution and dispose of any remaining portion of the hypodermic needle according to proper disposal method regulations in your area.
G. As with any appliance or tool, keep out of reach of children.
V. DEVICE DESCRIPTION
The NeedleZap™ product is designed for one-hand use when destroying 16-30 gauge needles, ¾-2 inches in length. The needle is destroyed by placing it vertically into the opening on top of the NeedleZap™ and then is gently rotated for 2-4 seconds to destroy the tip down the entire length to the hub. The NeedleZap™ portable needle destruction device operates from a battery pack consisting of two (2) 2V D-size rechargeable batteries. When a needle is inserted in the opening, it delivers 2.5A of current between two angled copper electrodes. Inside the product is a magnetic stainless steel clean-out tray that collects the debris from the destroyed needles. NeedleZap™ can be used to destroy 50-100 needles before the batteries need to be recharged using the NeedleZap™ 6V Charger that comes with the unit. The recharging process takes three (3) to four (4) hours.
The product measures 3" x 4" x 2" and weighs 1.4 lb. The housing is made of ABS flame-retardant plastic, and four rubber feet secure the unit from moving while placed on a flat surface.
VI. ALTERNATIVE PRACTICES AND PROCEDURES
The proper disposal of sharps is subject to federal Occupational Safety and Health Administration ("OSHA") regulations. The final rule, "Occupational Exposure to Bloodborne Pathogens" at 29 C.F.R. § 1910.1030, states that sharps are to be placed in containers that are closeable, puncture-resistant, and leak-proof.
VII. MARKETING HISTORY
Since May, 2001, the applicant has marketed NeedleZap™ exclusively to veterinarians and to law enforcement personnel. The product is a tool designed to complement existing methods of sharps disposal, including as puncture-resistant sharps containers, and the applicant has not marketed this product for destruction of needles used in treatment settings and clinical laboratories. There have been no reported accidental needlesticks associated with the use of this product for law enforcement or veterinary use.
{2}
VIII. POTENTIAL ADVERSE EFFECTS OF THE DEVICE ON HEALTH
There are no anticipated adverse effects on patient health from the use of NeedleZap™. No adverse effects were reported during the investigator studies. Failure of the device to destroy the needle puts the user at no greater risk than would be encountered with no use of needle destruction device. Failure of the device to completely destroy the needle still results in a blunted and completely sealed needle tip.
IX. SUMMARY OF NON-CLINICAL LABORATORY STUDIES
Formation of Infectious Aerosols and Microbiological Testing
Testing was performed with 150 contaminated needles using a known inoculum of *B. stearothermophilus* to determine whether infectious aerosols were generated during the operation of NeedleZap™.
Results: There were no culturable microorganisms recovered from the collection plates placed around the device.
Conclusion: These results reveal that, under these test conditions, there is no evidence of bioaerosols formation after needle incineration by NeedleZap™. The labeling instructs to dispose of the needles and waste as biohazardous waste.
Test for Emissions of Toxic Fumes
Testing was performed using NIOSH test methods to determine if the destruction of 25 needles using NeedleZap™ produced toxic fumes.
Results: An independent laboratory report concluded that the vapors produced by the destruction of needles were not considered hazardous in all tests performed.
Conclusion: These levels do not represent a health concern to persons operating the device.
Generation of Heat
To determine the heat generation from NeedleZap™ when used in accordance with the labeling, testing was performed during the destruction of 20 gauge, 1¼ inch needles.
Results: The testing revealed external temperatures did not exceed 35°C and the internal temperature was 33°C during needle destruction.
Conclusion: No significant heating occurs after continuous use of NeedleZap™. These findings do not exceed the UL 2601-1 test standard.
{3}
# Generation of Noise
To determine the noise generation from NeedleZap™ when used in accordance with the labeling, three attributes were tested during the destruction of 16-30 gauge, ¾-2 inch needles: (1) OSHA Hearing Conservation 80 dB Threshold; (2) OSHA Noise Compliance 90 dB Threshold; and (3) IEC Noise Monitoring, Raw Data Measurement.
Results: The testing revealed a minimum change in noise and may not be normally detectable.
Conclusion: This device does not generate excessive noise.
# Formation of Sparks
The sparking observed inside NeedleZap™ is not uncommon when needles are destroyed. The fire risk has been further reduced by the use of two (2) D-size rechargeable batteries protected by a 30A circuit breaker. The labeling instructions contain the following contraindications for use of NeedleZap™
- DO NOT USE THIS DEVICE NEAR FLAMMABLE LIQUIDS, GASES, OXYGEN, OPERATING ROOMS, OR EXPLOSIVES. NeedleZap™ can produce minor sparking which could create the risk of explosion or fire in this type of environment.
- DO NOT DESTROY NEEDLES THAT CONTAIN EXPLOSIVE OR RADIOACTIVE MATERIALS.
In addition, there is a label with the following CAUTION near the needle entry point:
CAUTION
Do Not Operate
In the presence of flammable liquids,
vapors, operating rooms,
Or in the presence of elevated Oxygen
Levels
Some Sparking Will Occur
Results of the clinical studies did not report any problems associated with sparks.
Conclusion: NeedleZap™ should be contraindicated for use in any potentially explosive environment where flammable gases or liquids are used or stored.
# Completeness of Destruction of Needles and Simulated Use
Simulated use testing was completed with 25 units of NeedleZap™ used to destroy 16-30 gauge, ¾ -2 inch needles. The criteria for the testing were the following:
{4}
a. PASS – needle completely destroyed from the tip to the hub OR needle burned completely to the hub with a tiny rounded stub with the sharp disabled
b. FAIL – needle partially destroyed and the sharp intact
Results: All 2,000 needles tested met the passing test criteria; either the needles were completely destroyed or the sharps were disabled.
Conclusion: The NeedleZap™ device can successfully destroy 16-30 gauge ¾-2 inch needles.
## Stability
NeedleZap™ is noted to have 4 rubber feet and a low squat appearance. The applicant provided independent testing to demonstrate that the device remained stable and could withstand being dropped short distances.
Conclusion: NeedleZap™ is a stable device during normal use.
## Cleaning
A cleaning study of the exterior of NeedleZap™ was conducted that included 20 trials of two common hospital cleaning agents. Egg white material was applied to cover the top of device to present visible soil. The device then was cleaned with the recommended cleaning agents. The firm also provided detailed cleaning instructions using standard precautions.
Results: The cleaning agents were able to remove visible evidence of soil without visible degradation of the surface of the device.
Conclusion: Visible soil can be successfully removed from the exterior of NeedleZap™ with the cleaning agent and the steps provided in the labeling.
## Electrical Safety
An independent firm tested the electrical safety of NeedleZap™ and found it to be compliant with the applicable requirements of UL 2601-1 and EN 60601-1. The product is powered by two (2) rechargeable batteries connected in series and has a nominal 4.5V output. The risk of electrical shock to the user is limited because the device is to be used on battery.
To test for EMC compatibility, the applicant supplied independent testing with the charger connected to the device.
{5}
Results: NeedleZap™ is within the requirements for UL and conducted and radiated emissions testing. In addition, there were no reports of electrical shock during the clinical studies.
Conclusion: The results of testing NeedleZap™ during needle destruction demonstrate that the device does not represent an electrical hazard to the user or the clinical settings tested.
# X. SUMMARY OF INVESTIGATOR STUDIES
Objective: To demonstrate that NeedleZap™ can be used safely and effectively by health care professionals and in health care facilities to destroy previously used aluminum and stainless steel needles attached to syringes.
Study Design #1: A total of seven investigators, all health care professionals, participated in this study. Each investigator was provided a NeedleZap™ unit for destruction of their used syringe needles (gauges 21-30, 1-2 inches in length), and was instructed to read the labeling before using the product. For a period of one month, each investigator was asked to complete a log that would provide information on the use of the product.
Results #1: The investigators reported a total of 720 needle destruction attempts. In every case, the needle destruction rate was 100%. One investigator reported that a small "stub" remained on each unit. No adverse events were reported.
Study Design #2: Testing was conducted at a hospital and a nursing home. Each investigator was provided a NeedleZap™ unit for destruction of syringe needles (gauges 16-30, ¾-2 inches in length), and was instructed to read the labeling before using the product. Each investigator was asked to complete a log that would provide information on the use of the product. Needle destruction was evaluated on a "0" (needle unaffected) to "3" (needle disable) scale.
Results #2: The investigators reported a total of 164 needle destruction attempts. The average burn grade was 2.65. No adverse events were reported.
# XI. CONCLUSIONS DRAWN FROM THE STUDIES
The data from these studies provide reasonable assurance of safety and effectiveness of the product when used as intended and in accordance with the labeling.
Potential adverse effects have been assessed in the design of the product. No adverse events were reported during the clinical studies, or as a result of use for law enforcement or veterinary purposes.
{6}
XII. PANEL RECOMMENDATION
Based on the regulatory discretion provided in Section 515(c)(2) of the Federal Food, Drug, and Cosmetic Act, as amended by the Safe Medical Devices Act of 1990, this PMA was not referred to the General Hospital and Personal Uses Panel, an FDA advisory committee, for review and recommendation.
XIII. CDRH DECISION
Based on the data submitted, CDRH has determined that there is reasonable assurance that NeedleZap™ is safe and effective for its intended use. CDRH issued an approval order on March 14, 2003.
The applicant’s manufacturing facility was inspected on February 7 and 13, 2003 and found to be in compliance.
XIV. APPROVAL SPECIFICATIONS
Directions for Use: See the labeling.
Hazards to Health from Use of the Device: See Contraindications, Warnings and Precautions in the attached labeling.
Post-Approval Requirements and Restrictions: See approval order.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.