THE DISINTEGRATOR INSULIN NEEDLE DESTRUCTION DEVICE
P010040 · Redhawk Medical Products & Services, LLC · MTV · Mar 15, 2002 · General Hospital
Device Facts
Record ID
P010040
Device Name
THE DISINTEGRATOR INSULIN NEEDLE DESTRUCTION DEVICE
Applicant
Redhawk Medical Products & Services, LLC
Product Code
MTV · General Hospital
Decision Date
Mar 15, 2002
Decision
APRL
Regulation
21 CFR 880.6210
Device Class
Class 2
Indications for Use
The Disintegrator™ is an insulin needle destruction unit for home use. It is a sharps needle destruction device that is designed to incinerate 27-30 gauge insulin needles, 5/16 to ½ inch in length, attached to insulin pens and disposable insulin syringes of 1/3cc to 1cc volume sizes.
Device Story
The Disintegrator™ is a portable, battery-powered device for home destruction of insulin needles. It accepts 27-30 gauge needles attached to pens or syringes. Operation involves inserting the needle into a port and pressing a button, which triggers a 9.6V electrical arc to melt and disintegrate the needle tip. Melted metal residue is collected in an internal sealed chamber. The device includes a needle straightener and low-battery indicators. It is intended for use by patients in close proximity to the injection site. By destroying the needle, the device reduces sharps waste and potential for accidental injury. It is powered by a rechargeable Ni-Cad battery or AC wall transformer.
Clinical Evidence
Prospective 30-day home-use study with 18 patients (14 completed) using 905 needles. Primary endpoints: patient adherence to instructions, needle destruction to <1/16 inch stub, and adverse events. Results: 93.3% successful destruction rate; failures attributed to battery charging or user error. No adverse events reported. Bench testing included mechanical stability, structural integrity (UL-2601), noise, heat, spark generation, electrical safety, EMI (EN-55011), toxic fume analysis (ICP/Mass Spec), and infectious aerosol testing (Bacillus subtilis).
Technological Characteristics
Portable unit (6"x4"x2", 2 lbs). Exterior: flame-retardant PC/ABS resin. Power: 9.6V Ni-Cad rechargeable battery or AC transformer. Principle: electrical arc incineration. Features: needle straightener, internal residue collection chamber, low-battery indicator. Electrical safety per UL 2601-1. EMI shielding via conductive paint for clinical versions.
Indications for Use
Indicated for home use by diabetic patients requiring insulin injections to destroy 27-30 gauge insulin needles (5/16 to 1/2 inch length) attached to insulin pens or 1/3cc-1cc disposable syringes.
Regulatory Classification
Identification
A sharps needle destruction device is a prescription device that is intended to destroy needles or sharps used for medical purposes by incineration or mechanical means.
Special Controls
*Classification.* Class II (special controls). The special controls for this device are:(1) Performance testing must demonstrate the following during operation of the device:
(i) The device safely contains or ventilates aerosols or fumes from device operation.
(ii) Excessive heat or sparks are not generated that may injure users or patients.
(iii) Simulated use testing must demonstrate sharps and/or needles are completely destroyed using a range of types and sizes of sharps sufficient to represent actual use.
(iv) Simulated use testing must demonstrate that the device is physically stable on the surface for which it is intended to be mounted to ensure the risk of harm to the patient/user as a result of the device falling is minimized.
(2) Validation of cleaning and disinfection instructions must demonstrate that the device can be safely and effectively reprocessed after use per the recommended cleaning and disinfection protocol in the instructions for use.
(3) Analysis and/or testing must validate electromagnetic compatibility and electrical safety, including the safety of any battery used in the device, under conditions which are consistent with the intended environment of device use.
(4) Software verification, validation, and hazard analysis must be performed.
(5) Labeling must include:
(i) A clear description of the device and its technological features;
(ii) How the device is to be used, including validated cleaning and disinfection instructions;
(iii) Relevant precautions and warnings based on performance and in-use testing to ensure proper use of the device; and
(iv) Instructions to install device in adequately ventilated area and stable area.
Submission Summary (Full Text)
{0}
# SUMMARY OF SAFETY AND EFFECTIVENESS DATA
## I. GENERAL INFORMATION
Device Generic Name: Sharps Needle Destruction Device for Insulin Needles
Device Trade Name: The Disintegrator™ Insulin Needle Destruction Unit
Applicant’s Name and Address:
Mr. Joey Adkins
Safeguard Medical Devices, Inc.
403 Ken Mar Industrial Parkway
Unit # 475
Broadview Heights., Ohio 44147
Premarket Approval (PMA) Application number: P010040
Date of Notice of Approval to the Applicant: March 15, 2002
## II. INDICATIONS FOR USE
The Disintegrator™ is an insulin needle destruction unit for home use. It is a sharps needle destruction device that is designed to incinerate 27-30 gauge insulin needles, 5/16 to ½ inch in length, attached to insulin pens and disposable insulin syringes of 1/3cc to 1cc volume sizes.
## III. CONTRAINDICATIONS
There are no known contraindications for the use of this device.
## IV. WARNINGS AND PRECAUTIONS
The warnings and precautions can be found in the attached instruction sheet and device labeling for the Disintegrator™.
## V. DEVICE DESCRIPTION
The Disintegrator™ is a home use, portable sharps needle destruction device for diabetics. It uses electrical current to incinerate 27-30 gauge insulin hypodermic needles 5/16 to 1/2 inch in length attached to insulin pens and disposable insulin syringes of 1/3 cc to 1 cc volume in size. The Disintegrator uses a rechargeable 9.6-volt battery to deliver an electrical current to produce an arc of electricity directed at the tip of the needle. This arc of electricity melts the tip of the needle and disintegrates the needle. Any balls of melted metal that fall off during the process are collected into a chamber where the ash or ball is sealed and does not
{1}
need to be emptied. The remaining stub of the needle on the hub is less than 1/16th inch in length. The device also includes a small molded port at the top, which is a needle straightener, for use in straightening needles that are bent prior to destruction.
The Disintegrator™ uses a large button as the switch to turn on the device and activate the circuit that supplies the charge. There is a low battery charge light that is yellow, which is found at the top of the device. Before first use, the Disintegrator battery should be charged for 12 hours. A fully charged battery is indicated by a green light located next to the yellow light. When the green light is lit this device will disintegrate 30-40 needles before needing to recharge. The Disintegrator™ unit can be powered from a battery and/or AC current by a wall transformer. The battery can be recharged by the wall transformer within 45 to 90 minutes. The Disintegrator™ can be utilized when the battery is recharging. The device exterior surface is off-white PC/ ABS (flame-retardant) Resin with a blue needle port. The tear shaped Disintegrator™ measures 6"X4"X 2" and weighs 2 pounds. The Disintegrator™ is to be used in closed proximity to the site of needle use.
## VI. ALTERNATIVE PRACTICES AND PROCEDURES
Alternative disposal methods of insulin syringes and pen needles include placement in a sealed puncture proof container prior to discarding in the trash or transporting to a sharps collection center; or use of another commercially available needle destruction device.
## VII. MARKETING HISTORY
The Disintegrator™ Insulin Needle Destruction Unit was introduced into Canada in June of 2001. Approximately 150 units are currently in use with no report of any customer complaints. The Disintegrator™ Insulin Needle Destruction Unit has not been withdrawn from marketing for any reason relating to the safety and effectiveness of the device.
## VII. POTENTIAL ADVERSE EFFECTS OF THE DEVICE ON HEALTH
There are no anticipated adverse effects on patient health from the use of The Disintegrator™. No adverse effects were reported during the clinical studies. Failure of the device to destroy the needle subjects the user to no greater risk than would be encountered with no use of a destruction device. Failure of the device to fully destroy the needle still results in a blunted and completely sealed needle tip.
{2}
IX. SUMMARY OF NON-CLINICAL STUDIES
Mechanical / Positional Stability
The device is supported with raised legs along the outer perimeter of the bottom, resulting in extreme stability during use. Tests were performed to determine whether the unit could be made to tip over by activating the “button” from various angles.
Results: The device remained stable during all tests.
Conclusion: The results of stability testing indicate that The Disintegrator™ is not prone to tipping during use.
Structural Integrity
Independent laboratory tests were performed to test the structural strength of The Disintegrator™ according to standards outlined in UL-2601.
Results: The device passed all tests.
Conclusion: The housing strength of the device is acceptable.
Ergonomic Considerations
The Disintegrator™ is designed with the needle port, needle straightening aid, and actuator button all on centerline. The indicators are equidistant from center and are visible from a wide viewing angle of approximately 120 degrees. In-house tests were conducted with both left and right-handed individuals to determine comparative ease of use.
Results: The device was found equally usable by both left and right handed individuals.
The prospective clinical study protocols required representation of both left-handed and right handed patients.
Results: There were no reports of difficulty in use attributable to the patient being left or right handed.
Conclusion: The Disintegrator™ may be used with equal efficiency and effectiveness by either left or right handed individuals.
{3}
# Generation of Noise
Measurement of noise during the destruction cycle was taken at a distance of 18 inches from the device.
Results: The increase in noise level during destruction was less than 3 dB, the minimum change normally detectable by the human ear.
Conclusion: The device poses no potential of excess noise generation.
# Generation of Heat
The temperature of the case at the port entry was monitored during the destruction of 10 needles over a 10 minute period (much more rapid than anticipated for a single insulin user).
Result: A total temperature rise of less than 5 degrees F was measured.
Independent laboratory tests were performed to determine the temperature rise on various internal components during both normal and intentional misuse, as well as in specific "failure" modes, according to safety standards outlined in UL-2601.
Results: All tested components had a temperature rise within the limits of the UL-2601 standard.
Conclusion: Neither external nor internal temperature rise during the needle destruction process poses a threat to the user.
# Formation of Sparks
During the in-house destruction of over 4,000 needles to date, no notice was made of any sparks generated outside the unit. Although reports of visible sparks inside the unit were recorded by some of the Study Subjects involved in the Clinical Trials, there were no reports of visible external sparks.
Results: There were no visible sparks generated outside the unit during destruction.
Conclusion: As no sparks were visible outside the unit, it is concluded that no direct user hazard exists from sparks. However, it is known that the device uses an internal low-powered electric arc to dispose of the metal needle body. For this reason, the unit should not be used near flammable gases or liquids. The Disintegrator™ is labeled accordingly.
{4}
# Electrical Shock Hazard and Electromagnetic Compatibility
The Disintegrator™ is powered with a rechargeable 9.6 volt Ni-Cad battery incapable of shock hazard. The device is designed to prevent operation in the event that the case or battery / ash door has been opened. The AC adapter selected for recharging the battery is an off-the-shelf sealed unit listed under both UL and CSA standards.
An independent laboratory study was conducted to evaluate the electrical safety of The Disintegrator™ according to UL 2601-1 (1997 edition).
Results: The unit passed all applicable requirements for electrical safety.
Additional laboratory testing of EMC levels was performed to EN-55011 standards. Tests were performed on a “Disintegrator” unit that contained no added shielding for EMI, as manufactured for the consumer market. Tests were also performed on a unit with shielding added in the form of EMI conductive paint, as might be the case for “Disintegrator” units intended for U.S. clinical sales.
Results: The unshielded unit showed short peaks of some radiated signals slightly above the limits of EN-55011. Modifications were made to the unit in the form of EMI conductive paint on the internal walls. This unit showed reductions in all radiated emissions to well below the limits of EN-55011.
Conclusions: Electrical safety and EMI test results are acceptable for the device as intended for use by consumers. EMI test results show that units intended for use in U.S. clinical settings would need to be manufactured utilizing EMI shielding materials.
# Emission of Toxic Fumes
The Disintegrator™ was tested during operation to identify and qualify airborne contaminates potentially generated by destruction of ten 27 gauge needles 1/2 inch in length. The test was designed to analyze the components of the needles based on their potential to cause deleterious health effects. These components were determined to be chromium, nickel, manganese and iron. An airtight chamber with a sampling pump was constructed to collect and filter air samples during operation of the unit. Three trials were performed with 10 syringes, each equipped with a ½ inch by 27 gauge needle, were tested over a 10 minute period per trial and the mass of each of these elements found in the filtered air was measured using a Perkin Elmer Sciex, Elan 5000, ICP / Mass Spec system. The concentration of the materials found in the sampled air were determined, by extrapolation, to be 256 times the concentration level which would have been found in 1 cubic meter of air.
{5}
Results: The levels found in the concentrated air were found to be less than 0.002 (0.2%) of the safe limits established by the EPA. If the samples had in fact been taken from a 1 cubic meter enclosure, the concentration levels would have been below detection limits.
Conclusions: The concentrations of these metals or their oxides that are emitted from use of The Disintegrator™ represent an unusual use of the device but are within safe limits.
## Formation of Infectious Aerosols
A Disintegrator™ device was sent to U.S. Micro-Solutions, Inc. of Greensburg, PA for testing to determine the potential for the formation of infectious aerosols generated from use of the unit. An Anderson N-6 Single Stage Air Sampler was placed next to the Disintegrator™ inside a Class II Biological Safety Cabinet, with a TSA plate placed into the sampler and 4 TSA plates placed around the unit. A total of 20 sterile insulin needles were destroyed and the plates incubated at 35 degrees (C) for 48 hours. This process was then repeated using 20 insulin needles contaminated with a 1 to 5 x 10⁵ suspension of *Bacillus subtilis*. A TSA plate inoculated with the suspension of *Bacillus subtilis* bacteria was used as the Positive Control. The air inside the biosafety cabinet was sampled with the N-6 Sampler, before destruction of any needles, and used as the Negative Control.
Results: All trypticase soy agar (TSA) plates had no growth of *Bacillus subtilis* except for the Positive Control plate.
Conclusion: The Disintegrator™ does not appear to generate an infectious aerosol during its use.
## Ability to Completely Destroy the Needle
Simulated laboratory tests were performed on more than 4,000 insulin needles, ranging from 27 to 30 gauge and lengths of 5/16 and ½ inch long, attached to syringes ranging in dosage from 3/10cc to 1cc. Insulin pens were also similarly tested. The study evaluated adequate destruction if the remaining "stub" was less than 1/16 inch long, whether or not terminated by a small BB shaped ball.
Results: All needle sizes and gauges were adequately destroyed.
The prospective clinical studies also evaluated actual needle destruction by insulin users in the home. In these studies the destruction criteria remained the same as in the laboratory setting.
Results: In the clinical studies the worst case rate of successful needle destruction was at least 93.3%.
{6}
Conclusion: The Disintegrator™ successfully demonstrated needle destruction for 27-30 gauge insulin needles in 5/16 and ½ inch lengths. Additional labeling instructions clarifying the battery charging requirements were added for the second clinical study, and are now added to all devices produced. This should help to maintain a high percentage of successful needle destruction.
## X. SUMMARY OF CLINICAL STUDIES
The objective of the clinical studies was to determine the effectiveness of The Disintegrator when used in the home by a variety of patients who inject insulin, following only the directions supplied by the manufacturer.
## Study Design
The assessment of safe use and the effectiveness of needle destruction was based on:
1) Patients' ability to properly follow written instructions provided with the device;
2) Destruction of the needle body to a "stub" of 1/16 inch or less in length;
3) The number of adverse device events; and
4) User comments regarding the use of the device.
During the clinical studies a total of 18 patients were enrolled for a 30-day study period. Patient selection criteria were based on a mixture of patients across a wide age range, both left and right handed, who inject insulin a minimum of 2 times per day. In addition, at least one patient was selected who suffered from some degree of diabetes induced visual impairment. Fourteen patients provided documentation demonstrating completion of all tests. The patients used an adjusted total of 905 needles. Included in the total number of needles are 169 "test syringes" representing 4 different syringe manufacturers and 11 different gauge and size combinations. These "test syringes" were supplied to the patients in their test kits. The study was approved by an IRB associated with the Cleveland Clinic where the patients were obtained. The tests took place in the patient's home.
## Results
Of the 905 needles tested, there were 42 reports of unsuccessful needle destruction. Of these failures, the study subjects documented 12 as attributable to their failure to recharge the battery, and another 7 were reported as due to improper use of the device inserts. The remaining 23 failures are attributed to imperfect mechanism operation, which could occur due to the manufacturing process used for the clinical trial production units. The firm has addressed the
{7}
issue of improper battery charging by revisions in the device labeling. The issue of mechanical imperfections is self-resolving in that all future production of mechanical components will utilize final injection molding with much tighter tolerances. There were no adverse events reported during the clinical study.
## Conclusion
The device safely and successfully destroys 27-30 gauge insulin needles when used as directed in the instructions for use.
## XI. SUMMARY CONCLUSIONS DRAWN FROM THE STUDIES
The pre-clinical and clinical testing provides reasonable assurance of the safety and effectiveness of the Disintegrator™ Insulin Needle Destruction Unit device when used in accordance with the manufacturers' instructions for use.
## XII. PANEL RECOMMENDATION
In accordance with the provisions of section 515(c)(2) of the act as amended by the Safe Medical Devices Act of 1990, this PMA was not referred to the General Hospital and Personal Uses Devices Advisory Panel, an FDA advisory committee, for review and recommendation because the information in the PMA substantially duplicates information previously reviewed by this panel.
## XIII. CDRH DECISION
The applicant's manufacturing facility was inspected on January 18, 2002 and was found to be in compliance with the Quality Systems Regulations. FDA issued an approval order on March 15, 2002.
## XIV. APPROVAL SPECIFICATIONS
Directions for use: See the attached labeling.
Hazards to Health from Use of the Device: See Contraindications, Warnings, Precautions and Adverse Reactions in the attached labeling.
Post-Approval Requirements and Restrictions: See approval order.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.