The AMPLICOR™ HCV Test, v2.0 is an in vitro diagnostic, nucleic-acid amplification test for qualitative detection of HCV RNA in human serum or plasma from blood collected in EDTA. This test detects by reverse-transcribing target HCV RNA into complementary DNA (cDNA), amplifying cDNA by polymerase chain reaction (PCR), hybridizing amplified cDNA with an oligonucleotide probe that binds enzyme, and catalyzing conversion of substrate to a colored product that is recognized by a microwell plate reader. The AMPLICOR HCV Test, v2.0 is indicated for patients who have evidence of liver disease and antibody evidence of HCV infection, and who are suspected to be actively infected with HCV. Detection of HCV RNA indicates that the virus is replicating and therefore is evidence of active HCV infection.
Device Story
The AMPLICOR HCV Test, v2.0 is an in vitro diagnostic assay for qualitative detection of HCV RNA in human serum or EDTA plasma. The device processes specimens by lysing virus particles, reverse-transcribing HCV RNA into cDNA, and amplifying target sequences via PCR using thermostable rTth polymerase. An internal control (HCV IC) is included to monitor extraction and amplification efficiency. Contamination is controlled using AmpErase (uracil-N-glycosylase) and dUTP. Amplified products are hybridized to specific oligonucleotide probes on microwell plates and detected colorimetrically via an avidin-horseradish peroxidase conjugate and TMB substrate. Absorbance is measured at 450 nm using a microwell plate reader. The test is performed in clinical laboratories by trained personnel. Results provide direct evidence of active viral replication, aiding clinicians in diagnosing active HCV infection in patients with liver disease and antibody evidence of infection. The test helps differentiate active infection from inactive or cleared states, supporting clinical decision-making regarding further diagnostic procedures or antiviral therapy.
Clinical Evidence
Clinical study evaluated 846 patients with liver disease. Performance compared against anti-HCV testing (EIA/SIA), ALT levels, and liver histology. Results showed high concordance: 94% (serum) and 92% (plasma) positive in anti-HCV positive patients. In EIA-negative patients, 97% (serum) and 96% (plasma) were negative. Study included patients with and without histopathologic evidence of hepatitis. No clinical data for monitoring therapy or screening asymptomatic populations.
Technological Characteristics
In vitro nucleic-acid amplification test (PCR). Uses thermostable rTth DNA polymerase for simultaneous reverse transcription and amplification. Target: 5' UTR region of HCV genome (244 bp amplicon). Detection: Colorimetric microwell plate assay (450 nm). Reagents include guanidine thiocyanate lysis buffer, biotinylated primers, and AmpErase for carry-over prevention. Connectivity: Standalone microwell plate reader. Sterilization: Not applicable (reagent kits).
Indications for Use
Indicated for patients with evidence of liver disease and antibody evidence of HCV infection suspected of active HCV infection.
Regulatory Classification
Identification
A nucleic acid-based hepatitis C virus (HCV) ribonucleic acid (RNA) test is identified as an in vitro diagnostic device intended for prescription use as an aid in the diagnosis of HCV infection in specified populations, and/or as an aid in the management of HCV-infected patients including guiding the selection of genotype-specific treatment in individuals with chronic HCV infection. The test is intended for use with human serum or plasma. The test is not intended for use as a donor screening test for the presence of HCV antibodies in blood, blood products, or tissue donors.
Special Controls
*Classification.* Class II (special controls). The special controls for this device are:(1) For all nucleic acid-based HCV RNA tests, the labeling required under § 809.10(b) of this chapter must include:
(i) A prominent statement that the test is not intended for use as a donor screening test for the presence of HCV RNA from human cells, tissues, and cellular and tissue-based products.
(ii) A detailed explanation of the principles of operation and procedures for performing the assay.
(iii) A detailed explanation of the interpretation of results.
(iv) Limitations, which must be updated to reflect current clinical practice and disease presentation and management. These limitations must include, but are not limited to, statements that indicate:
(A) The specimen types for which the device has been cleared and that use of this test kit with specimen types other than those specifically cleared for this device may result in inaccurate test results.
(B) When applicable, that assay performance characteristics have not been established in populations of immunocompromised or immunosuppressed patients or, other populations where test performance may be affected.
(C) Test results are to be interpreted by qualified licensed healthcare professionals in conjunction with the individual's clinical presentation, history, and other laboratory results.
(2) For all nucleic acid-based HCV RNA tests, the design verification and validation must include:
(i) Detailed device description, including the device components, ancillary reagents required but not provided, and an explanation of the device methodology. Additional information appropriate to the technology must be included such as design of primers and probes, rationale for the selected gene targets, specifications for amplicon size, and degree of nucleic acid sequence conservation.
(ii) For devices with assay calibrators, the design and nature of all primary, secondary, and subsequent quantitation standards used for calibration as well as their traceability to a standardized reference material that FDA has determined is appropriate (
*e.g.,* a recognized consensus standard). In addition, analytical testing must be performed following the release of a new lot of the standard material that was used for device clearance or approval, or when there is a transition to a new calibration standard.(iii) Documentation and characterization (
*e.g.,* determination of the identity, supplier, purity, and stability) of all critical reagents (including nucleic acid sequences for primers and probes) and protocols for maintaining product integrity.(iv) Detailed documentation of analytical performance studies conducted as appropriate to the technology, specimen types tested, and intended use of the device, including, but not limited to, limit of detection (LoD), upper and lower limits of quantitation (ULoQ and LLoQ, respectively), linearity, precision, endogenous and exogenous interferences, cross reactivity, carryover, matrix equivalency, and sample and reagent stability. Samples selected for use in analytical studies or used to prepare samples for use in analytical studies must be from subjects with clinically relevant circulating genotypes in the United States. Cross-reactivity studies must include samples from HCV RNA negative subjects with other causes of liver disease, including autoimmune hepatitis, alcoholic liver disease, chronic hepatitis B virus, primary biliary cirrhosis, and nonalcoholic steatohepatitis, when applicable. The effect of each claimed nucleic-acid isolation and purification procedure on detection must be evaluated.
(v) Risk analysis and management strategies, such as Failure Modes Effects Analysis and/or Hazard Analysis and Critical Control Points summaries and their impact on test performance.
(vi) Final release criteria to be used for manufactured test lots with appropriate evidence that lots released at the extremes of the specifications will meet the claimed analytical and clinical performance characteristics as well as the stability claims.
(vii) Multisite reproducibility study that includes the testing of three independent production lots.
(viii) All stability protocols, including acceptance criteria.
(ix) Final release test results for each lot used in clinical studies.
(x) Analytical sensitivity and specificity of the test must be the same or better than that of other cleared or approved tests.
(xi) Lot-to-lot precision studies, as appropriate.
(3) For devices intended for the qualitative detection of HCV RNA, in addition to the special controls listed in paragraphs (b)(1) and (2) of this section, the design verification and validation must include detailed documentation of performance from a multisite clinical study. Performance must be analyzed relative to an FDA cleared or approved qualitative HCV RNA test, or a comparator that FDA has determined is appropriate. This study must be conducted using appropriate patient samples, with appropriate numbers of HCV positive and negative samples in applicable risk categories. Additional genotypes must be validated using appropriate numbers and types of samples. The samples may be a combination of fresh and repository samples, sourced from within and outside the United States, as appropriate. The study designs, including number of samples tested, must be sufficient to meet the following criteria:
(i) Clinical sensitivity of the test must have a lower bound of the 95 percent confidence interval of greater than or equal to 95 percent.
(ii) Clinical specificity of the test must have a lower bound of the 95 percent confidence interval of greater than or equal to 96 percent.
(4) For devices intended for the quantitative detection of HCV RNA, the following special controls, in addition to those listed in paragraphs (b)(1) and (2) of this section, apply:
(i) Labeling required under § 809.10(b) of this chapter must include a prominent statement that the test is not intended as a diagnostic test to confirm the presence of active HCV infection, when applicable.
(ii) Design verification and validation must include the following:
(A) Detailed documentation of the following analytical performance studies conducted as appropriate to the technology, specimen types tested, and intended use of the device, including but not limited to: LoD, ULoQ and LLoQ. LoD, LLoQ, and linearity studies must demonstrate acceptable device performance with all HCV genotypes detected by the device.
(B) Detailed documentation of clinical performance testing from either:
(
*1* ) A multisite clinical study with an appropriate number of clinical samples from chronically HCV infected patients in which the results are compared to an FDA-cleared or approved quantitative HCV RNA test, or a comparator that FDA has determined is appropriate. This study must include a sufficient number of HCV positive samples containing an analyte concentration near the LLoQ to describe performance at this level. Clinical samples must cover the full range of the device output and must be consistent with the distribution of these genotypes in the U.S. population. Clinical samples may be supplemented with diluted clinical samples for those viral load concentrations that are not sufficiently covered by natural clinical specimens, or(
*2* ) A clinical study with prospectively collected samples demonstrating clinical validity of the device.(C) Detailed documentation of a qualitative analysis near the lower end of the measuring range demonstrating acceptable performance when used as an aid in diagnosis.
(5) For devices intended for HCV RNA genotyping, in addition to the special controls listed in paragraphs (b)(1) and (2) of this section, design verification and validation must include the following:
(i) Detailed documentation of an analytical performance study demonstrating the LoD for all HCV genotypes detected by the device.
(ii) Detailed documentation, including results, of a multisite clinical study that assesses genotyping accuracy (
*i.e.,* the proportion of interpretable results that match with the reference method result) and the genotyping rate (*i.e.,* the proportion of results that were interpretable).(6) For any nucleic acid-based HCV RNA test intended for Point of Care (PoC) use, the following special controls, in addition to those listed in paragraphs (b)(1) and (2) of this section, apply:
(i) Clinical studies must be conducted at PoC sites.
(ii) Additional labeling must include a brief summary of the instructions for use that are appropriate for use in a PoC environment.
Reference Devices
COBAS AMPLICOR™ HCV Test, v2.0
Submission Summary (Full Text)
{1}
1
# Summary of Safety and Effectiveness Data
## 1. General Information
1.1. Device Generic Name: In vitro nucleic-acid amplification test for qualitatively detecting *Hepatitis C virus* (HCV) RNA in human serum or plasma from blood collected in EDTA
1.2 Device Trade Name: AMPLICOR® HCV Test, v2.0
1.3 Applicant’s Name and Address:
Roche Molecular Systems, Inc. (RMS)
4300 Hacienda Drive
Pleasanton, CA 94588
1.4 Premarket Approval Application (PMA) Number: P000010
1.5 Date of Panel Recommendation: July 28, 2000
1.6 Date of Notice of Approval: July 5, 2001
## 2. Indications for Use
The AMPLICOR™ HCV Test, v2.0 is an *in vitro* diagnostic, nucleic-acid amplification test for qualitative detection of HCV RNA in human serum or plasma from blood collected in EDTA. This test detects by reverse-transcribing target HCV RNA into complementary DNA (cDNA), amplifying cDNA by polymerase chain reaction (PCR), hybridizing amplified cDNA with an oligonucleotide probe that binds enzyme, and catalyzing conversion of substrate to a colored product that is recognized by a microwell plate reader. The AMPLICOR HCV Test, v2.0 is indicated for patients who have evidence of liver disease and antibody evidence of HCV infection, and who are suspected to be actively infected with HCV. Detection of HCV RNA indicates that the virus is replicating and therefore is evidence of active HCV infection.
## 3. Contraindications
None known.
## 4. Warnings and Precautions
The AMPLICOR HCV Test, v2.0 is for *in vitro* diagnostic use only. Refer to labeling for a listing of other warnings and precautions.
## 5. Device Description
### 5.1. Test Principles
The AMPLICOR HCV Test, v2.0 is based on five major processes:
{2}
- specimen preparation
- reverse transcription to generate cDNA from target HCV RNA and HCV Internal Control (HCV IC) RNA
- PCR amplification of target cDNAs by using HCV-specific primers
- hybridization of amplified cDNAs to target-specific oligonucleotide probes
- colorimetric detection of the probe-bound amplified cDNAs.
The AMPLICOR HCV Test, v2.0 permits simultaneous reverse transcription and PCR amplification of HCV and HCV IC target RNAs. The Master Mix reagent contains a primer pair that is specific for both HCV and HCV IC RNAs. Detection of amplified DNA is performed using target-specific oligonucleotide probes that permit independent identification of HCV amplicon and HCV IC amplicon.
## 5.1.1. Specimen Preparation
HCV RNA is isolated directly from serum or EDTA plasma by lysing virus particles with a chaotropic agent. HCV IC RNA, introduced into each specimen with Lysis Reagent, serves as an extraction and amplification control for each processed specimen. HCV and HCV IC RNAs are precipitated by using alcohol and then resuspended in Specimen Diluent.
## 5.1.2. Enzyme and Target Selection
**Enzyme**: Reactions are catalyzed by thermostable, recombinant *Thermus thermophilus* DNA Polymerase (rTth pol). In the presence of manganese ion (Mn²⁺) and the appropriate buffer, rTth pol has reverse transcriptase and DNA polymerase activities. This allows both reverse transcription and PCR amplification to occur in the same reaction mixture.
**Target selection**: Selection of primers and probe was critical for AMPLICOR HCV Test, v2.0 detection of all recognized genotypes (see Non-Clinical Performance section of the Package Insert). Accordingly, selection of the target RNA sequence was based on identifying an HCV-genome region that was maximally conserved among genotypes. HCV RNA sequences are most conserved in the 5' untranslated region (UTR). The AMPLICOR HCV Test, v2.0 uses primers KY78 and KY80 to amplify a 5' UTR sequence of 244 nucleotides. HCV RNA sequence corresponding to these primers and the capture probe are located in the most conserved 5' UTR domains.
## 5.1.3. Reverse Transcription
Processed specimens are added to the amplification mixture in reaction tubes, in which reverse transcription and PCR amplification occurs. The downstream or antisense primer (KY78) is biotinylated at its 5' end; the upstream or sense primer (KY80) is not biotinylated. The reaction mixture is heated in the thermal cycler to allow specific annealing of the downstream primer to target HCV and HCV IC RNAs. In the presence of Mn²⁺ and excess deoxynucleoside triphosphates (dNTPs), including
{3}
deoxyadenosine, deoxyguanosine, deoxycytidine and deoxyuridine (in place of thymidine) triphosphates, rTth pol extends the annealed primer to form cDNA.
## 5.1.4. Target Amplification
Following reverse transcription of target HCV and HCV IC RNAs, the reaction mixture is heated to denature RNA:cDNA hybrids and expose sequences that anneal with the primers. As the mixture cools, the upstream primer (KY80) anneals specifically to the cDNA strand representing each target RNA, rTth pol extends the primer, and a second DNA strand is synthesized. This completes the first cycle of PCR, yielding a double-stranded DNA copy of each RNA target region (HCV and HCV IC).
The reaction mixture is heated again to separate the double-stranded DNA and expose the primer-annealing sequences. As the mixture cools, primers KY78 and KY80 anneal to target DNA. The rTth pol enzyme, in the presence of Mn²⁺ and excess dNTPs, extends the annealed primers along the target templates to produce a 244 base pair double-stranded DNA "amplicon." The thermal cycler automatically repeats this process for 37 cycles, with each cycle intended to double the amount of amplicon DNA. Amplification occurs only in the region of the HCV genome between the primers; the entire genome is not amplified.
## 5.1.4.1. Amplification of HCV Internal Control (HCV IC) cDNA
In enzyme-based amplification processes such as PCR, inhibitors that may be present in the clinical specimen can reduce amplification efficiency. The HCV IC has been added to the AMPLICOR HCV Test, v2.0 to permit identification of processed specimens containing substances that may interfere with PCR amplification or specimens in which HCV RNA may have been lost during sample processing. The HCV IC is an RNA transcript with primer-annealing regions identical to those in the HCV genome, a randomized internal sequence of similar length and base composition as the HCV target sequence, and a unique probe-binding region that differentiates HCV IC amplicon from HCV amplicon. These features were selected to ensure equivalent amplification of HCV IC and HCV target RNAs. The HCV IC is introduced into each specimen with the Lysis Reagent and serves as an extraction and amplification control for each processed specimen.
## 5.1.4.2. Selective Amplification
Selective amplification of target nucleic acid from the clinical specimen is achieved in the AMPLICOR HCV Test, v2.0 by the use of AmpErase® (uracil-N-glycosylase) and deoxyuridine triphosphate (dUTP). AmpErase recognizes and catalyzes the destruction of DNA strands containing deoxyuridine, but not DNA containing thymidine. Deoxyuridine is not present in naturally occurring DNA, but is always present in amplicon from this assay due to the use of deoxyuridine triphosphate in place of thymidine triphosphate
{4}
as one of the dNTPs in the Master Mix reagent; therefore, only amplicon contains deoxyuridine.
Deoxyuridine renders contaminating amplicon susceptible to destruction by AmpErase prior to amplification of the target DNA. AmpErase, which is included in the Master Mix reagent, catalyzes cleavage of deoxyuridine-containing DNA at deoxyuridine residues by opening the deoxyribose chain at the C1-position. When heated in the first thermal cycling step at the alkaline pH of Master Mix, the amplicon DNA chain breaks at the position of the deoxyuridine, thereby rendering the DNA non-amplifiable. AmpErase is inactive at temperatures above 55°C (i.e., throughout the thermal cycling steps) and therefore does not destroy target amplicon. Following amplification, any residual AmpErase is denatured by the addition of Denaturation Solution, thereby preventing the degradation of any target amplicon. AmpErase in the AMPLICOR HCV Test, v2.0 has been demonstrated to inactivate at least 10³ copies of deoxyuridine-containing HCV amplicon per PCR.
## 5.1.5. Hybridization Reaction
Following PCR amplification, and after the addition of Denaturation Solution to the reaction tubes, the HCV amplicon and the HCV IC amplicon are chemically denatured to form single-stranded DNA. Aliquots of denatured amplicon are then transferred to separate wells of microwell plates (MWP) coated with either an HCV-specific (KY150) or HCV IC-specific (SK535) oligonucleotide probe. The biotin-labeled HCV and HCV IC amplicon are hybridized to the target-specific oligonucleotide probes bound to the wells of the MWP. This hybridization of amplicon to the target-specific probe increases the overall specificity of the test.
## 5.1.6. Detection Reaction
Following the hybridization reaction, the MWP is washed to remove unbound material and Avidin-Horseradish Peroxidase Conjugate is added to each well of the MWP. The Avidin-Horseradish Peroxidase Conjugate binds to the biotin-labeled amplicon which is hybridized to the target (HCV or HCV IC) specific oligonucleotide probes bound to the MWP. The MWP is washed again to remove unbound conjugate and a substrate solution containing hydrogen peroxide and 3,3',5,5'-tetramethylbenzidine (TMB) is added to each well. In the presence of hydrogen peroxide, the bound horseradish peroxidase catalyzes the oxidation of TMB to form a colored complex. The reaction is stopped by addition of a weak acid and the absorbance is measured at a wavelength of 450 nm (A₄₅₀) using a microwell plate reader.
4
{5}
# 5.2. Kit Configuration and Components
The AMPLICOR HCV Test, v2.0 consists of five separately packaged kits containing the controls and reagents required for specimen preparation, reverse transcription, and PCR amplification and detection of HCV and Internal Control target RNA (Table 1).
Table 1: AMPLICOR™ HCV Test, v2.0 Kits
| AMPLICOR HCV Specimen Preparation Kit, version 2.0 | 96 Tests | P/N: 21111086; US: 83126 |
| --- | --- | --- |
| AMPLICOR HCV Controls Kit, version 2.0 | 8 Sets | P/N: 21111175; US: 83131 |
| AMPLICOR HCV Amplification Kit, version 2.0 | 96 Tests | P/N: 21111094; US: 83127 |
| AMPLICOR HCV Detection Kit, version 2.0 | 96 Tests | P/N: 21118439; US: 833373 |
| AMPLICOR Internal Control Detection Kit | 96Tests | P/N: 20751952; US: 83068 |
The components of each kit are listed below (Table 2).
Table 2: Components of the AMPLICOR® HCV Test, v2.0
| KIT COMPONENT | FORMULATION |
| --- | --- |
| AMPLICOR HCV Specimen Preparation Kit, version 2.0 | |
| (1) HCV Lysis Reagent, v2.0
HCV LYS, v2.0 | Tris-HCl buffer containing 68% guanidine thiocyanate, 3% dithiothreitol and <1% glycogen, 8 x 6.9 mL |
| (2) HCV Specimen Diluent, v2.0
HCV DIL, v2.0 | Tris-HCl buffer containing, <0.005% poly rA RNA (synthetic), EDTA and 0.05% sodium azide, 8 x 4.8 mL |
| (3) HCV Internal Control, v2.0
HCV IC, v2.0 | A solution containing <0.001% non-infectious in vitro transcribed RNA (microbial) encoding HCV primer binding sequences and a unique probe binding region, <0.005% poly rA RNA (synthetic), EDTA, and 0.05% sodium azide, 8 x 0.1 mL |
| AMPLICOR HCV Controls Kit, version 2.0 | |
| (4) Negative Plasma (Human)
NHP | Human plasma non-reactive by FDA licensed tests for antibody to HCV, antibody to HIV-1 and HIV-2, HIV p24 antigen and HBsAg, and containing 0.1% ProClin® 300, 8 x 0.6 mL |
| (5) HCV (-) Control, v2.0
HCV (-) C, v2.0 | A solution containing <0.005% poly rA RNA (synthetic), EDTA and 0.05% sodium azide, 8 x 0.1 mL |
| (6) HCV (+) Control, v2.0
HCV (+) C, v2.0 | A solution containing <0.001% non-infectious in vitro transcribed RNA (microbial) encoding HCV sequences, <0.005% poly rA RNA (synthetic), EDTA and 0.05% sodium azide, 8 x 0.1 mL |
{6}
Table 2: (continued) Components of the AMPLICOR® HCV Test, v2.0
| AMPLICOR HCV Amplification Kit, version 2.0 | |
| --- | --- |
| (7) HCV Master Mix, v2.0
HCV MMX, v2.0 | A bicine buffered solution containing 16% DMSO, glycerol, < 0.01% rTth DNA Polymerase (rTth pol, microbial), potassium acetate, < 0.001% dATP, dCTP, dGTP, and dUTP, <0.005% KY78 and KY80 primers (one is biotinylated), < 0.01% AmpErase® (microbial) and 0.05% sodium azide, 8 x 0.7 mL |
| (8) HCV Manganese Solution, v2.0
HCV Mn²⁺, v2.0 | A solution containing <2% manganese, acetic acid, Amaranth dye, 0.05% sodium azide, 8 x 0.1 mL |
| KIT COMPONENT | FORMULATION |
| AMPLICOR HCV Detection Kit, version 2.0 | |
| (9) HCV Microwell plate, v2.0
HCV MWP, v2.0 | MWP coated with HCV-specific DNA Probe (KY150). Twelve, 8-well strips in one resealable pouch with desiccant |
| (10) Denaturation Solution
[1] DN | A solution containing 1.6% sodium hydroxide, EDTA and thymol blue, 1 x 12 ml |
| (11) HCV Hybridization Buffer
[2] HCV HYB | A sodium phosphate solution containing < 0.2% solubilizer and < 25% sodium thiocyanate, 1 x 20 mL |
| (12) Avidin-Horseradish Peroxidase Conjugate
[3] AV-HRP | A Tris-HCl buffered solution containing <0.001% avidin-horseradish peroxidase conjugate, bovine serum albumin (mammalian), Emulsit 25, 0.1% phenol, 1% ProClin® 150, 1 x 12 mL |
| (13) Substrate A
[4A] SUB A | A citrate solution containing 0.01% hydrogen peroxide, 0.1% ProClin 150, 1 x 12 mL |
| (14) Substrate B
[4B] SUB B | 0.1% 3,3',5,5'-Tetramethylbenzidine (TMB) in 40% dimethylformamide (DMF), 1 x 3 mL |
| (15) Stop Reagent
[5] STOP | Contains 4.9% Sulfuric acid, 1 x 12 mL |
{7}
Table 2 (continued). Components of the AMPLICOR™ HCV Test, v2.0
| KIT COMPONENT | FORMULATION |
| --- | --- |
| AMPLICOR Internal Control Detection Kit | |
| (17) Internal Control Microwell Plate
IC MWP | MWP coated with IC-specific DNA probe (SK 535), Twelve, 8-well strips in one resealable pouch with desiccant, 1 x 96 tests |
| (18) Avidin-Horseradish Peroxidase Conjugate
[3] AV-HRP | A Tris-HCl buffered solution containing <0.001% avidin-horseradish peroxidase conjugate, bovine serum albumin (mammalian), Emulsit 25, 0.1% phenol, 1% ProClin® 150, 1 x 12 mL |
| (19) Substrate A
[4A] SUB A | A citrate solution containing 0.01% hydrogen peroxide and 0.1% ProClin® 150, 1 x 12 mL |
| (20) Substrate B
[4B] SUB | Contains 0.1% 3,3',5,5'-tetramethylbenzidine (TMB) in 40% dimethylformamide (DMF), 1 x 3 mL |
| (21) Stop Reagent
[5] Stop | Contains 4.9% Sulfuric acid, 1 x 12 mL |
| (22) 10X-Wash Concentrate
10 X WB | A solution containing <2% phosphate buffer, <9% sodium chloride, EDTA, <2% detergent and 0.5% ProClin 300, 2 x 90 mL |
## 6. Alternative Practices and Procedures
### 6.1. Biochemical Methods
Alanine aminotransferase (ALT) is a marker of hepatocellular damage and its concentration in serum is often elevated during viral hepatitis. Prior to identification of HCV and subsequent development of tests for detecting anti-HCV, ALT was used as a surrogate marker for "parenterally transmitted non-A, non-B" hepatitis. As such, ALT was a non-specific test for HCV infection. More recently, and in the absence of a more specific or virologic test, changes in ALT concentration were used to assess effectiveness of antiviral therapy for chronic hepatitis C. While it is still used in assessing severity of chronic hepatitis C without therapy, ALT concentration poorly predicts the presence or absence of histopathologic changes in the liver.
{8}
8
## 6.2. Serologic or Virologic Methods
Currently, there are only a few methods for detecting evidence of HCV infection. Such evidence can be indirect (e.g., detection of an immunologic response) or direct (e.g., detection of the virus or its macromolecules).
## 6.2.1. Testing for Antibodies to HCV (anti-HCV)
Anti-HCV assays yield indirect evidence of HCV infection but cannot determine if the infection is active (i.e., HCV is actively replicating) or inactive. Furthermore, current antibody tests do not yield temporal information: results from a single specimen cannot be used to determine if an infection is acute or chronic.
**Enzyme immunoassay (EIA)** is the most widely used method for detecting evidence of HCV infection and is usually a first step in laboratory diagnosis of hepatitis C. Current EIAs (FDA-licensed as version 2 or 3) detect antibodies to antigens representing core (nucleocapsid) and certain non-structural HCV proteins. EIAs are readily performed but less than optimally specific. Therefore, package inserts for currently licensed EIAs advise supplemental testing to ensure that a “repeatedly reactive” result accurately indicates detection of antibodies to HCV.
**Strip immunoassay (SIA)** is the most widely used method for supplemental detection of antibody evidence of HCV infection and, currently, version 2 is licensed. CDC and its consultants² have also recommended using appropriate HCV RNA assays for supplemental evidence of infection (see below; Virologic Methods).
## 6.2.2. Virologic Methods
**Virus isolation (propagation)** cannot currently be used for clinical specimens: HCV is not known to replicate in cell culture. It is also not practical to inoculate susceptible animals (the only well-characterized model of HCV infection uses chimpanzees, *Pan troglodytes*).
**Nucleic-acid detection**: during recent years, analytically sensitive methods for detecting DNA or RNA were developed that have enabled practical detection or quantitation of the RNA genome inside HCV particles that circulate in the blood. These methods have been widely used for research applications, including detection of HCV RNA. Many are commercially available, often as generic “kits” that a laboratory can adapt for detection of particular analyte like HCV RNA.
All such methods are based on hybridization between reagent nucleic acid molecules and DNA or RNA “target” in the sample and subsequent production of an identifiable “signal.” These methods are frequently categorized as “nucleic-acid amplification tests” (NAT), terminology that is somewhat misleading because of wide variation in technological approaches and in analytical sensitivity. Typically, “target amplification” (polymerase-catalyzed production of nucleic acid representing the target, followed by generation of production-indicating signal) is analytically more sensitive than “signal
{9}
amplification" (enhanced production of signal that indicates direct hybridization with target). Examples of target amplification methods include PCR, nucleic acid sequence based amplification (NASBA), transcription mediated amplification (TMA), self-sustained sequence replication (3SR, a non-commercial predecessor to NASBA and TMA), strand displacement amplification (SDA), ligase chain reaction (LCR), and Qβ replicase (QβR). Examples of target amplification methods include branched chain DNA (bDNA) and hybrid capture (HC).
Assays for HCV RNA in the blood are intended to provide information about the viremic status of HCV-infected individuals, as demonstrated in peer-reviewed scientific literature. Other literature suggests that HCV viremia is a reliable marker of active HCV replication in the liver. Therefore, HCV RNA assays have been used in research for direct evidence of active HCV infection, in diagnostic or monitoring modes. The monitoring mode includes assessment of antiviral-therapy effects.
Several HCV RNA assays are commercially available but, other than the COBAS AMPLICOR™ HCV Test, v2.0 and AMPLICOR® HCV Test, v2.0, none have been FDA-approved for in vitro diagnostic use. Furthermore, none are FDA-approved for monitoring HCV infections (including monitoring of response to therapy for hepatitis C) or FDA-licensed for use in determining safety of blood, blood products, or tissue. Like the COBAS AMPLICOR HCV Test, v2.0 and the AMPLICOR® HCV Test, v2.0, other assays for HCV RNA often use reagent nucleic acids representing 5' UTR sequences that are highly conserved among HCV genotypes (to optimize sensitivity) but distinct from other known nucleotide sequences (in an effort to optimize specificity).
7. Marketing History
The AMPLICOR HCV Test, v2.0 has not been withdrawn from marketing in any country for reasons related to safety or effectiveness. It is currently available for commercial marketing in the following countries:
9
{10}
- Argentina
- Austria
- Australia
- Belarus
- Belgium
- Chile
- Cyprus
- Czech Republic
- Denmark
- Ecuador
- Egypt
- Finland
- France
- Germany
- Greece
- Hong Kong
- Hungary
- Iceland
- India
- Indonesia
- Ireland
- Israel
- Japan
- Korea
- Kuwait
- Lebanon
- Luxembourg
- Malaysia
- Malta
- Mexico
- Netherlands
- New Zealand
- Norway
- Oman
- Pakistan
- Paraguay
- Philippines
- Poland
- Qatar
- Russia
- Saudi Arabia
- Slovenia
- South Africa
- Spain
- Sweden
- Switzerland
- Syria
- Turkey
- United Arab Emirates
- United Kingdom
- Venezuela
## 8. Potential Adverse Effects of the Device on Health
### 8.1. Potential Adverse Effects for Patients
Possibilities for erroneous test results are inherent in all *in vitro* diagnostic products. An erroneous result can then lead to an adverse effect for the tested patient. Any current test for evidence of HCV infection, including those for detecting antibodies to HCV, can yield an inaccurate result, or a result that is inaccurately reported to the physician or inaccurately interpreted by the physician. The possibility of an erroneous AMPLICOR HCV Test, v2.0 result exists due to poor specimen quality (e.g., sample contamination or inadequate storage), potential malfunction of a test component, an equipment error, or an operator error. Because HCV RNA is the only current practical marker of active infection, an inaccurate result or interpretation can adversely affect a patient unless other information enables the physician to recognize the inaccuracy.
A False Positive result (i.e., a AMPLICOR HCV Test, v2.0 Positive result when HCV RNA was not present in the specimen) could lead to a misdiagnosis of active HCV infection. This misdiagnosis could lead to unnecessary diagnostic procedures (e.g., liver biopsy, which has associated risks and discomfort) or psychological trauma to the patient (e.g., concern about a chronic infection that can have severe disease consequences). While misdiagnosis could also lead to unnecessary treatment (current pharmacologic regimens for chronic hepatitis C are less than optimally efficacious and are associated with significant side effects), current practice recommendations include additional HCV RNA testing (quantitative and genotyping) for most patients before therapy begins.
{11}
A False Negative result (i.e., a AMPLICOR HCV Test, v2.0 Negative result when HCV RNA was present) could lead to a misdiagnosis of cleared, or inactive, HCV infection. This misdiagnosis could lead to unnecessary procedures in consideration of other diagnoses; a delay in, or failure to perform, appropriate additional diagnostic procedures (e.g., to determine the state of active HCV infection or severity of HCV-associated liver disease); or a delay in, or failure to provide, appropriate treatment.
Data submitted by RMS and in peer-reviewed scientific literature provided reasonable assurance of AMPLICOR HCV Test, v2.0 safety and effectiveness. That is, adverse effects could be severe but the potential for a false result appeared to be very low (and exceeded by the likelihood of benefit from an accurate result).
## 8.2. Potential Adverse Effects for Laboratory Personnel
There is also potential for adverse effects, from an AMPLICOR HCV Test, v2.0 component, on the health of laboratory personnel performing this assay. One component, Negative Plasma (Human), should be considered potentially infectious because it contains human-source materials. However, before the component is manufactured, these materials must be non-reactive for HBsAg and for antibodies to HCV, HIV-1, and HIV-2.
This potential risk occurs with many other in vitro diagnostic products. Appropriate warnings are included in AMPLICOR HCV Test, v2.0 vial and kit labels and in the Package Insert. Most laboratory personnel implement standard good laboratory practices and are protected from hepatitis B due to vaccination or past infection. Therefore, this risk is well known to laboratory personnel and its potential is considered to be minimal.
## 9. Summary of Non-Clinical Studies
RMS laboratories performed studies to assess analytical sensitivity and specificity of the AMPLICOR HCV Test, v2.0. Those studies that contributed to the determination of safety and effectiveness are summarized in the Non-Clinical Performance section of the AMPLICOR HCV Test, v2.0 Package Insert. Certain findings are also summarized below.
## 9.1. Limits of Detection (LODs)
LODs were determined by testing multiple specimens containing very low concentrations of the WHO International Standard for HCV genotype 1 RNA*. This WHO standard was diluted in HCV-negative serum and in HCV-negative EDTA plasma to concentrations of 100, 75, 60 and 50 IU/mL. Each concentration was tested with the AMPLICOR HCV Test, v2.0 in replicates ranging from 80 to 120. LODs were determined as the lowest HCV RNA concentration that resulted in ≥ 95% of results yielding A₄₅₀ ≥ 1.0, the cutoff for Positive results with the AMPLICOR HCV Test, v2.0. By this criterion, the LOD for HCV RNA in serum was 2.0 log₁₀ IU/mL, or 100 IU/mL. The LOD for HCV RNA in EDTA plasma was 1.7 log₁₀ IU/mL, or 50 IU/mL.
11
{12}
* The WHO International Standard for HCV genotype 1 RNA is available, as NIBSC code 96/790, from the National Institute for Biological Standards and Control, London, UK. IU designates International Units of this standard, which contains $10^{5}\mathrm{IU/mL}$ of genotype 1 HCV RNA as defined by consensus studies.
Here and in the AMPLICOR HCV Test, v2.0 Package Insert, $IU$ designates virion or subgenomic HCV RNA that has been quantified with reference to the WHO Standard. $IU / \mathrm{mL}$ affords a standardized approach to indicating HCV RNA concentration but it is not known if $IU / \mathrm{mL}$ accurately reflects the HCV RNA concentration of any particular specimen. Available data indicate that 1 IU corresponds to $>1$ HCV RNA molecule and that this number of molecules varies according to quantifying methods (which are imprecise with a single determination) and other variables:
| Source of data | Type of RNA | How quantified | Quantifier | Conversion factor |
| --- | --- | --- | --- | --- |
| Consensus | HCV (virion) | End-point dilution: qualitative assays | Copies | 1 IU ≈ 1.8 copies |
| | | Quantitative assays | Copies or genome equivalents | 1 IU ≈ 6.6 copies |
| Roche Molecular Systems (unpublished) | Subgenomic transcript of cloned HCV cDNA | UV spectroscopy | A260-molecules | 1 IU ≈ 2.7 A260 molecules (95% CI, 2.6-2.8) |
While RMS has demonstrated similar conversion factors for certain RNAs representing HCV genotypes 2-6, it is not known if quantitation with reference to the WHO Standard is affected by genotype, strain characteristics (including RNA structure and quasispecies), or concentration of HCV RNA. For example, IU-to-copy ratios for low concentrations of HCV RNA may be different than those for high concentrations of HCV RNA.
Specimens yielding a result in the Equivocal Zone were not repeat tested (per Interpretation of Results section of the Package Insert), so a final result was not obtained for these specimens. If repeat testing were done, it is reasonable to assume that frequencies of Positive results would increase, possibly resulting in lower LODs. By excluding Equivocal results, for example, the $95\%$ threshold of Positive results was reached at $1.7 \log_{10} \mathrm{IU} / \mathrm{mL}$ ( $50 \mathrm{IU} / \mathrm{mL}$ ) for serum.
# 9.2. Detection of HCV Genotypes
Quantified subgenomic RNAs, transcribed from cloned cDNAs, were used to approximate analytical sensitivity of the AMPLICOR HCV Test, v2.0 for five HCV genotypes: 1, 2, 3, 4, and 5. (Genotypes of seven viruses, including two representatives each of genotypes 1 and 2, were determined by using research methods that have not been evaluated or approved by FDA; however, genotyping by many such methods is generally recognized to be accurate whereas many subtyping techniques vary in accuracy.) Each RNA transcript, consisting of the $5^{\prime}$ -untranslated and core regions, was quantified by spectrophotometry $(\mathrm{A}_{260})$ and then diluted to three different concentrations in HCV Specimen Diluent, v2.0. Twenty-four replicates of each concentration were tested by the AMPLICOR HCV Test, v2.0. The number of Positive results at each concentration was determined. Positive result frequencies were similar for these subgenomic RNAs representing genotypes 1-5.
Another study evaluated the sensitivity of the AMPLICOR HCV Test, v2.0 under research laboratory conditions to determine if it would yield Positive results with 65 clinical specimens
{13}
containing HCV strains that represented the six recognized genotypes. The majority of these specimens were from patients in an antiviral efficacy study and were assumed to contain HCV RNA concentrations that were representative of those patients for whom the AMPLICOR HCV Test, v2.0 is indicated. All 65 specimens yielded Positive results.
## 9.3. Cross Contamination
Two studies were done under controlled laboratory conditions to assess the potential cross contamination rate of the AMPLICOR HCV Test, v2.0. In 22 runs of the AMPLICOR HCV Test, v2.0, multiple operators tested replicates of plasma specimens containing clinically-pertinent concentrations of HCV RNA (6.5 log₁₀ IU/mL) that alternated with replicates of an HCV-negative specimen. Combined results indicated a cross contamination rate of 1.2% (3/241).
The potential for cross contamination was also examined in two reproducibility studies performed under clinical laboratory conditions with lower concentrations of HCV RNA. In the first study, samples with [HCV RNA] ≤ 2.3 log₁₀ IU/mL were tested. Among 532 HCV-negative serum specimens, 531 yielded negative results, and 1 (0.2%) yielded an Equivocal result, which may or may not have indicated cross contamination. Among 537 HCV-negative plasma specimens, 3 (0.6%), yielded Equivocal results and 2 (0.4%) yielded Positive results, for a cross contamination rate between 0.4% and 0.9%. In the second study, with 153 specimens at [HCV RNA] ≤ 4.7 log₁₀ IU/mL, a cross contamination rate of 0.7% (1/153) was shown.
## 9.4. Cross-reactivity with or Interference from Other Microorganisms
The AMPLICOR HCV Test, v2.0 was evaluated by testing for potential cross-reactivity with, or interference from, certain pathogenic microorganisms and normal epidermal microflora that could be present in specimens. Twenty-four specimens that contained virus (adenoviruses, enteroviruses, herpesviruses, or retroviruses) and four that contained bacteria (Propionibacterium acnes, Staphylococcus aureus, or S. epidermidis) yielded Negative HCV and Positive HCV IC results with the AMPLICOR HCV Test, v2.0.
The assay was also evaluated for cross-reactivity by testing specimens that represented certain manifestations of infection with Hepatitis A virus (HAV) or Hepatitis B virus (HBV). Analytes known to be present in these specimens were HAV, IgM antibodies to HAV, HBsAg, or HBsAg with HBeAg. All tested serum and plasma specimens yielded Negative AMPLICOR HCV Test, v2.0 results.
13
{14}
14
## 9.5. Endogenous Substances
Serum specimens containing elevated concentrations of albumin, bilirubin, hemoglobin, triglycerides, and immunoglobulins were tested for interference with the AMPLICOR HCV Test, v2.0. These specimens were tested neat or after spiking with a near-LOD concentration of HCV RNA (100 IU/mL). Each specimen was tested in triplicate.
Specimens not spiked with HCV yielded Negative results, except for 2 of 11 specimens that contained elevated, or nearly elevated levels, of bilirubin. Negative AMPLICOR HCV Test, v2.0 results from the other 9 elevated-bilirubin specimens indicated lack of interference leading to False Positive results.
All HCV-spiked specimens yielded Positive results, indicating a lack of interference..
## 9.6. Therapeutic Drugs
Drugs for infectious diseases, or for several conditions associated with hepatitis C therapy, were evaluated for their potential to interfere with the AMPLICOR HCV Test, v2.0. Among the former were drugs for hepatitis C, hepatitis B, HIV, influenza A, or cytomegalovirus-associated syndromes. Each drug was spiked into two plasma concentrations, peak and 3-times peak (1X and 3X $C_{\mathrm{max}}$), that contained HCV at near-LOD (100 IU/ml) or was HCV RNA-negative. Specimens were tested in triplicate, with and without spiked drugs. Evaluated drugs did not yield False Positive or False Negative results. At each drug concentration, all HCV-containing (100 IU/ml) specimens yielded Positive results.
## 10. Summary of Clinical Studies
These studies are summarized in the Expected Values and Clinical Evaluation sections of the AMPLICOR HCV Test, v2.0 Package Insert. Certain findings are also summarized below.
## 10.1. Specimens from Characterized Patients
A prospective study was conducted at four US hepatology centers to evaluate clinical performance of the AMPLICOR HCV Test, v2.0 against results of testing for anti-HCV, and against a combination of anti-HCV results, serum ALT levels, and histologic findings in liver tissue. Studied patients were being investigated for evidence of liver disease, infection with HCV, or both. Patients were excluded if they had undergone liver transplantation or received antiviral therapy for hepatitis C within 6 months of study screening. While HCV RNA was not characterized via quantification or genotyping, the patients were assumed to represent US populations with active HCV infection and chronic liver disease; i.e., range of HCV RNA concentrations (majority likely > $10^{4}$ IU/mL) and HCV genotypes (majority of genotype 1 viruses, with genotypes 2 and 3 comprising most of the remainder).
## 10.2. Expected Values
**Specimens from patients with antibody evidence of HCV infection, not treated for hepatitis C:** During the clinical study, the vast majority of such specimens yielded AMPLICOR HCV Test, v2.0 absorbance values ($\mathrm{A}_{450} \geq 2.0$) much greater than those of the
{15}
Equivocal Zone (0.3 ≤ A₄₅₀ < 1.0): 92% of frozen or refrigerated serum and 93% of frozen EDTA plasma. Few specimens required repeat testing because of a Potentially Inhibited result (0.7% of frozen or 0% of refrigerated serum and frozen EDTA plasma), or an Equivocal Zone result (0.5% of frozen or 0% refrigerated serum and frozen EDTA plasma).
**Run failures**: Among 101 runs of the AMPLICOR HCV Test, v2.0 during the clinical study, 6 (6%) failed because Positive Controls were out of range (A₄₅₀ < 1.5) and 1 (1%) failed because Negative Controls were out of range (A₄₅₀ = 0.25). In addition, there were 2 failed runs due to protocol deviations (i.e., failure to follow procedures for specimen).
## 10.3. Comparison with Results of Anti-HCV Testing
Studied specimens were from 846 patients, including 153 who had received antiviral therapy for chronic hepatitis C that stopped > 6 months before collecting specimens. While these previously treated patients were included because those with active infections were considered to be virologically representative for determining assay performance, their data do not imply AMPLICOR HCV Test, v2.0 performance for monitoring HCV-associated disease or response to treatment. The 693 untreated patients included >200 who had negative anti-HCV EIA results and/or ALT concentration within reference range and no histopathologic evidence of hepatitis; they were studied for approximating test specificity, but their data do not imply performance for testing of individuals without liver disease or antibody evidence of HCV infections (see Warnings). In particular, Positive AMPLICOR HCV Test, v2.0 results were obtained for three patients who had liver disease other than hepatitis C and for whom there was no evidence, or insufficient evidence, to conclude that they were actively infected with HCV.
Analysis of data from specimens of serum and EDTA plasma demonstrated high concordance with results from anti-HCV testing:
- For specimens with EIA-repeat reactive (RR) and SIA-positive results of anti-HCV testing, the vast majority of AMPLICOR HCV Test, v2.0 results were Positive for both matrices: 94% for serum and 92% for EDTA plasma.
- For EIA-RR / SIA-indeterminate specimens, AMPLICOR HCV Test, v2.0 results were Positive for 66% of serum specimens and 50% of EDTA plasma specimens.
- For EIA-negative samples, the vast majority of AMPLICOR HCV Test, v2.0 results were Negative for both matrices (97% for serum and 96% for plasma).
## 10.4. Comparison with Anti-HCV, ALT, and Histologic findings
Test performance was further evaluated in the subset of patients from whom liver tissue had been collected in the past. AMPLICOR HCV Test, v2.0 results were compared with the combination of anti-HCV results, ALT concentrations, and histologic findings.
**Patients with elevated ALT levels and histopathologic evidence of hepatitis**: AMPLICOR HCV Test, v2.0 results were Positive for 99% of serum and 97% of EDTA plasma specimens that were anti-HCV EIA-RR and SIA-positive. AMPLICOR HCV Test, v2.0 results were also Positive for all 8 EIA-RR / SIA-indeterminate specimens. Negative
15
{16}
results were obtained with all 17 specimens from EIA-negative patients; their histopathologic characteristics were suggestive of autoimmune hepatitis, hepatitis B, or other forms of non-C hepatitis.
**Patients with normal ALT levels and histopathologic evidence of hepatitis:**
AMPLICOR HCV Test, v2.0 results were Positive for 95% of serum and 87% of EDTA plasma specimens among EIA-RR specimens (one of which was SIA-indeterminate; the others were SIA-positive). AMPLICOR HCV Test, v2.0 results were Negative for all 9 EIA-negative sera from patients with histologic features of hepatitis that were suggestive of diseases other than hepatitis C.
**Patients with elevated ALT levels and no histopathologic evidence of hepatitis:**
AMPLICOR HCV Test, v2.0 results were Positive for 5 of 7 serum specimens, and both EDTA plasma specimens, with EIA-RR / SIA-positive anti-HCV results. AMPLICOR HCV Test, v2.0 results were Negative for 97% of EIA-negative serum specimens.
**Patients with normal ALT levels and no histopathologic evidence of hepatitis:**
Positive AMPLICOR HCV Test, v2.0 results were obtained from (i) 2 of 4 serum specimen and the single EDTA plasma specimen with EIA-RR / SIA-positive results, and (ii) the single EIA-RR / SIA-indeterminate sample. AMPLICOR HCV Test, v2.0 results were Negative for 97% of serum specimens and for the one EDTA plasma specimen from EIA-negative patients. Histopathologic liver disease was evident in these cases (but without features of hepatitis).
## 10.5. Reproducibility
The AMPLICOR™ HCV Test, v2.0 Package Insert summarizes a reproducibility study that was performed in three clinical laboratories.
## 11. Conclusions Drawn from the Studies
## 11.1. Non-Clinical Studies
### Analytical Sensitivity
Regardless of the approach to analyzing submitted data, LODs with the WHO International Standard for HCV genotype 1 RNA were ≤2.0 log₁₀ IU/mL, an HCV RNA concentration that is much lower than those expected for most patients for whom the AMPLICOR HCV Test, v2.0 is indicated. (Similar data were obtained with other genotype 1 viruses in a reproducibility study that was performed in clinical laboratories: see Package Insert.)
Current technical limitations made it difficult to determine LODs for non-1 genotypes: in particular, there are no international reference materials for recognized HCV genotypes 2-6 and practical quantifying methods have approximately ±0.3 log₁₀ imprecision. While these limitations prevented LOD determinations from studies with RNAs representing non-1 HCV genotypes, the data were not inconsistent with those from the WHO Standard (i.e., the data were suggestive of LODs within 1 log₁₀ of those for genotype 1). These data also demonstrated that the AMPLICOR HCV Test, v2.0 was likely to detect HCV genotypes 2-
{17}
6 at much lower RNA concentrations that those expected for most patients for whom the assay is indicated.
These findings are evidence of the safety and effectiveness of the AMPLICOR HCV Test, v2.0 for its indicated use. In addition, data from an internationally recognized reference material, the WHO International Standard for HCV genotype 1 RNA, is appropriate and essential for enabling comparison among HCV RNA assays.
## Analytical Specificity
### Potential Cross-reactivity or Interference
Submitted data yielded no evidence of AMPLICOR HCV Test, v2.0 cross-reactivity with, or interference from, most of the analytes that were tested, including certain microorganisms, serum proteins, and pertinent therapeutic drugs.
The tested endogenous substances were shown not to interfere with the AMPLICOR HCV Test, v2.0, with the exception of certain specimens containing bilirubin, which might have yielded False Positive results. Among 11 specimens tested for analytical specificity in the presence of elevated concentrations of bilirubin and without spiked HCV, 2 did not yield Negative AMPLICOR HCV Test, v2.0 results. It could not be determined if the Positive and Equivocal results from these 2 specimens represented accurate detection of HCV RNA because anti-HCV results were not available. These concerns are noted in the procedural limitations section. Non-clinical studies of the AMPLICOR HCV Test, v2.0 did not include evaluation of potential interference with detection of low HCV RNA concentrations by other analytes that often circulate at elevated concentrations during viral hepatitis. While there was no apparent effect of ALT on frequencies of Positive results during the clinical study, it was assumed that most studied patients had >10⁴ IU/mL of HCV RNA; furthermore, most specimens were frozen (which reduces concentrations of these analytes) before AMPLICOR HCV Test, v2.0 testing. These concerns are noted in the procedural limitations section. While tested therapeutic drugs did not appear to interfere with the ability to detect HCV RNA, specimens were spiked with each drug and so the study did not provide data about drug metabolites that might affect assay performance. In addition, non-clinical studies did not evaluate potential effects of anti-bacterial or anti-fungal drugs that are frequently administered to patients. FDA did not recommend any pertinent post-approval studies.
### Cross Contamination
Cross contamination is a recognized problem for all laboratory assays, and particularly those based on target amplification occurred. Submitted data indicated infrequent (≤1.2%) cross-contamination rates during studies of the AMPLICOR HCV Test, v2.0.
## 11.2. Clinical Studies
### Study of patients with liver disease
The study of patients with liver disease demonstrated high concordance between AMPLICOR HCV Test, v2.0 results and those from anti-HCV testing. There was also high
17
{18}
concordance between AMPLICOR HCV Test, v2.0 results and the combination of serologic, biochemical and histologic findings. Such concordance was plausible because study data indicated that most of the studied patients were likely to have had chronic liver disease. Peer-reviewed scientific literature had previously demonstrated that (i) most HCV infections persist with detectable HCV viremia, particularly in individuals with elevated serum concentrations of ALT; and (ii) EIA-negative HCV infections with detectable RNA are very unusual, especially in a chronically infected population.
Limits of current technology and design of the study made it difficult to determine the true frequencies of false AMPLICOR HCV Test, v2.0 results. That is, while there may have been unrecognized False Positive and False Negative AMPLICOR HCV Test, v2.0 results, FDA decided that the study data were appropriate for contributing to the determination of safety and effectiveness. Most importantly, there is no reference assay with known accuracy for detecting HCV RNA.
Study design did not include determination of HCV genotypes or HCV RNA concentrations. While it could not be determined if these variables affected AMPLICOR HCV Test, v2.0 performance, FDA recognized the absence of reference assays for genotyping or quantifying HCV RNA and accepted the assumption that these variables were adequately represented among enrolled patients at the four US study sites. Furthermore, non-clinical studies demonstrated that any effects of genotype were unlikely at the high concentrations of HCV RNA (>10⁴ IU/mL) that most patients in this study were assumed to have had.
The data from this clinical study demonstrated that AMPLICOR HCV Test, v2.0 yielded Positive results for a very high proportion of patients with liver disease and antibody evidence of HCV infection. The assay also yielded Negative results for a very high proportion of patients who had liver disease without anti-HCV (these data do not imply performance for testing of individuals without liver disease or antibody evidence of HCV infections). Although the exact proportions of false results could not be determined, FDA decided that the data provided reasonable assurance of AMPLICOR HCV Test, v2.0 safety and effectiveness for diagnosis of active HCV infection among patients with evidence of liver disease and antibody evidence of HCV infection. Effectiveness of the AMPLICOR HCV Test, v2.0 was also supported by the low frequencies, during the clinical study, of failed runs and of Potentially Inhibited and Equivocal results that require repeat testing.
## Reproducibility
Performance of the AMPLICOR™ HCV Test, v2.0 assay was reproducible among studied variables that included reagent lots, study sites, and days, and sample matrices (See package insert).
## 11.3. Risk / Benefit Analysis
The presence of HCV viremia, as determined by detecting HCV RNA, is direct evidence of replication of the virus; i.e., active HCV infection. Thus, the most significant benefit is that AMPLICOR HCV Test, v2.0 results can provide evidence of active HCV infection for patients who have evidence of liver disease and antibody evidence of HCV infection.
18
{19}
(Antibody evidence of HCV infection is defined here as repeatedly reactive results from an anti-HCV EIA alone, or in combination with positive or indeterminate results from an anti-HCV SIA).
Available data indicate that the likelihood of benefit, to patients tested by the AMPLICOR HCV Test, v2.0, exceeds any known or potential adverse event or risk to patients or laboratory personnel.
The clinical and non-clinical laboratory studies described above provided reasonable assurance that the AMPLICOR HCV Test, v2.0 can be safely and effectively used for qualitative detection of HCV RNA in human serum or EDTA plasma. These studies also provided scientific evidence of diagnostic benefits to be gained from FDA approval for marketing of the AMPLICOR HCV Test, v2.0.
## 11.4. Safety
As a diagnostic test, the AMPLICOR™ HCV Test, v2.0 assay involves removal of blood from an individual for testing purposes. The test, therefore, presents no more safety hazards to and individual being tested than other tests where blood is removed.
## 12. Panel Recommendations
The Microbiology Devices Advisory Panel (MDAP) met on July 28, 2000 to consider safety and effectiveness of the AMPLICOR HCV Test, v2.0. MDAP recommended approval subject to the following conditions:
1) RMS should submit, and FDA should satisfactorily analyze, appropriate data for determining the assay's reproducibility.
2) The Package Insert should include revisions to the intended use, warnings, precautions and performance sections.
## 13. CDRH Decision
CDRH concurred with MDAP recommendations. RMS provided additional data, analyses, and peer-reviewed scientific literature to address assay reproducibility and other issues. Certain issues were rectified by appropriate data that RMS submitted in the Amendments, by FDA analyses of RMS data and peer-reviewed scientific literature, or by warnings and limitations in the Package Insert (Labeling restrictions). Other issues were rectified by deciding that the probability of unrecognized inaccurate performance or adverse effects was exceeded by public health importance of the first approval for US marketing of an HCV RNA assay. The Package Insert was revised by using the MDAP recommendations.
On October 6, 2000, the applicant's two manufacturing facilities were found to be in compliance with the Quality Systems Regulation (21 CFR 820).
Having determined that there was reasonable assurance of safety and effectiveness for the AMPLICOR HCV Test, v2.0, CDRH issued an approval order on July 3, 2001.
19
{20}
20
14. Approval Specifications
Directions for Use: see Package Insert and other Labeling
Conditions of Approval: CDRH Approval of this PMA is subject to full compliance with the conditions described in the approval order.
Postapproval Requirements and Precautions: See Approval Order
15. References
Alter MJ, Kruszon-Moran D, Nainan OV, McQuillan GM, Gao F, Moyer LA, Kaslow RA, Margolis HS. The prevalence of hepatitis C virus infection in the United States, 1988 through 1994. N Engl J Med 1999; 341:556-562.
Centers for Disease Control and Prevention. Recommendations for prevention and control of hepatitis C virus (HCV) infection and HCV-related chronic disease. MMWR Morb Mortal Wkly Rep 1998;47 (No. RR-19):1-54. Available at: http://www.cdc.gov/mmwr/PDF/rr/rr4719.pdf.
Gretch, DR. Diagnostic tests for hepatitis C. Hepatology 1997; 26 (Suppl 1): 43S-47S.
Schiff ER, de Medina M, Kahn RS. New perspectives in the diagnosis of hepatitis C. Semin Liver Dis 1999; 19 (Suppl1): 3-15.
Lok AS, Gunaratnam NT. Diagnosis of hepatitis C. Hepatology 1997; 26 (Suppl 1): 48S-56S.
National Institutes of Health Consensus Development Conference Panel statement: management of hepatitis C. Hepatology 1997; 26 (Suppl 1):2S-10S. Available at: http://odp.od.nih.gov/consensus/cons/105/105_intro.htm.
Myers TW, Gelfand DH. Reverse transcription and DNA amplification by a Thermus thermophilus DNA polymerase. Biochemistry 1991; 30:7661-7666.
Stuyver L, Rossau R, Wyseur A, et al. Typing hepatitis C virus isolates and characterization of new subtypes using a line probe assay. J Gen Virol 1993; 74:1093-1102.
Machida A, Ohnuma H, Tsuda F, et al. Two distinct subtypes of hepatitis C virus defined by antibodies directed to the putative core protein. Hepatology 1992;16:886-891.
Bukh J, Purcell RH, Miller RH. Importance of primer selection for the detection of hepatitis C virus RNA with the polymerase chain reaction assay. Proc Natl Acad Sci USA 1992; 89: 187-191.
Bukh J, Purcell RH, Miller RH. Sequence analysis of the 5' noncoding region of hepatitis C virus. Proc Natl Acad Sci USA 1992; 89:4942-4946.
Okamoto H, Okada S, Sugiyama Y, et al. The 5'-terminal sequence of the hepatitis C virus genome. Jpn J Exp Med 1990; 60:167-177.
Young KKY, Resnick R, Myers TW. Detection of hepatitis C virus RNA by a combined reverse transcriptase-polymerase chain reaction assay. J Clin Microbiol 1993; 31:882-886.
{21}
21
Longo MC, Berninger MS, Hartley JL. Use of uracil DNA glycosylase to control carry-over contamination in polymerase chain reactions. Gene 1990; 93:125-128.
Aoyagi K, Ohue C, Iida K, Kimura T, Tanaka E, Kiyosawa K, Yagi S. Development of a simple and highly sensitive enzyme immunoassay for hepatitis C virus core antigen. J Clin Microbiol 1999; 37:1802-1808.
Saldanha J, Lelie N, Heath A. Establishment of the First International Standard for Nucleic Acid Amplification Technology (NAT) Assays for HCV RNA. WHO Collaborative Study Group. Vox Sang 1999; 76:149-158.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.