← Product Code [MOS](/productcode/MOS) · K994436

# SURGI-VISION ESOPHAGEAL STYLET COIL (K994436)

_Surgi-Vision, Inc. · MOS · Mar 14, 2000 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K994436

## Device Facts

- **Applicant:** Surgi-Vision, Inc.
- **Product Code:** [MOS](/productcode/MOS.md)
- **Decision Date:** Mar 14, 2000
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1000
- **Device Class:** Class 2
- **Review Panel:** Radiology

## Indications for Use

The Surgi-Vision Esophageal Stylet Coil is recommended for highresolution Magnet Resonance Imaging of the human esophagus. The single use, disposable Stylet Coil has been designed to be inserted in the esophagus of the patient during MRI scans in order to obtain improved image quality in the anatomical regions surrounding the esophagus. The unique coil shape facilitates orientation of the coil to the anatomy. This product is to be used with a 1.5T MRI machine.

## Device Story

Esophageal stylet coil; accessory to 1.5T MRI systems. Device consists of inter-cavitary probe with radiofrequency receiving coil, electronic matching circuit, and coaxial cable. Inserted into patient esophagus to improve image quality of surrounding anatomical regions; signals transmitted via cable to standard surface coil connector of GE MRI systems. Single-use, disposable design. Used in clinical MRI settings; operated by trained personnel. Benefits include enhanced resolution of aortic regions visualized from esophagus. Safety testing confirms no excess RF heating and no increased susceptibility to peripheral nerve stimulation under clinical conditions.

## Clinical Evidence

Bench testing only. Evaluated RF heating in clinical-representative phantoms (no significant temperature increase vs. control); assessed current leakage from MRI pulsed gradient fields (no increased nerve stimulation risk); verified imaging performance (enhanced resolution of aortic regions).

## Technological Characteristics

Inter-cavitary probe with radiofrequency receiving coil, electronic matching circuit, and coaxial cable. Fully sealed in insulating layer. Designed for 1.5T MRI systems. Passive device; no software or active energy source.

## Regulatory Identification

A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).

## Special Controls

*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

## Predicate Devices

- Surgi-Vision Endo-Esophageal MRI Coil ([K992193](/device/K992193.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

MAR 1 4 2000

# 510(k) Summary

# Surqi-Vision Esophageal Stylet Coil

Common/Classification Name: Accessory to Magnetic Resonance Diagnostic Device, 21 CFR 892.1000

> Surgi-Vision, Inc. 9250 Rumsey Road, Suite 100 Columbia, MD 21045

Contact: Nancy E. Taylor, Prepared: December 30, 1999

# LEGALLY MARKETED PREDICATE DEVICES A.

The Surgi-Vision Esophageal Stylet Coil is substantially equivalent to the Surgi-Vision Endo-Esophageal MRI Coil, which was cleared for marketing on September 3, 1999, in premarket notification K992193.

#### B. DEVICE DESCRIPTION

The Surgi-Vision Esophageal Stylet Coil is a specialty coil for use in MRI imaging of the anatomical regions surrounding the esophagus. The signals picked up by the coil are conducted through a small coaxial cable to a connection with the standard surface coil connector for GE MRI systems. The coil and cable are completely sealed inside the insulating layer.

### C. INTENDED USE

The Surgi-Vision Esophageal Stylet Coil is recommended for highresolution Magnet Resonance Imaging of the human esophagus. The single use, disposable Stylet Coil has been designed to be inserted in the esophagus of the patient during MRI scans in order to obtain improved image quality in the anatomical regions surrounding the esophagus. The unique coil shape facilitates orientation of the coil to the anatomy. This product is to be used with a 1.5T MRI machine.

#### D. SUBSTANTIAL EQUIVALENCE SUMMARY

The Surgi-Vision Esophageal Stylet Coil has identical indications for use as the legally marketed predicate device.

Surgi-Vision, Inc. Esophageal Stylet Coil December 30, 1999

Page 1-1

{1}------------------------------------------------

The Surgi-Vision Esophageal Stylet Coil has the same technological characteristics as the predicate device. Both proposed and predicate devices have an electronic matching circuit, a connecting coaxial cable, and an inter-cavitary probe with a radiofrequency receiving coil. However, there are differences in construction and design that make it necessary to provide performance data to assure substantial equivalence. Such performance data are available and do demonstrate substantial equivalence.

# E. TECHNOLOGICAL CHARACTERISTICS

See Section D. above.

## F. TESTING

Surgi-Vision carried out testing and/or analysis of the Surgi-Vision Esophageal Stylet Coil that addressed the following issues:

- (1) Possibility of excess RF heating;
- Possibility of increased susceptibility of patients to peripheral nerve (2) stimulation: and
- (3) Imaging performance.

The results of the heating experiments demonstrate that there is no excess heating when SVESC is positioned in a phantom that is representative of clinical conditions. The change in temperature observed during use of the SVESC is not significantly different then that observed without the coil. The experiments done to determine current leakage by the MRI pulsed gradient field demonstrate that there is no possibility of increased susceptibility of patients to nerve stimulation. - Finally, the imaging performance demonstrates enhanced resolution of aortic regions visualized from the esophaqus. The results of the testing demonstrate that there are no safety problems for imaging of a patient using the Surgi-Vision Esophageal Stylet Coil if the instructions for use are followed and if the whole-body average SAR is no more than the present recommended limit of 4 W/kg.

## G. CONCLUSIONS

This pre-market submission has demonstrated Substantial Equivalence as defined and understood in the Federal Food Drug and Cosmetic Act and

Surgi-Vision, Inc. Esophageal Stylet Coil December 30, 1999

{2}------------------------------------------------

various guidance documents issued by the Center for Devices and Radiological Health.

Surgi-Vision, Inc. Esophageal Stylet Coil
December 30, 1999

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the department's name around the perimeter. In the center of the seal is a stylized image of an eagle with three wing segments.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAR 1 4 2000

Nancy E. Taylor CEO/President Surgi-Vision, Inc. 9250 Rumsey Rd., Suite 100 Columbia, MD 21045

Re:

K994436 Esophageal Sylet Coil Dated: December 30, 1999 Received: December 30, 1999 Regulatory class: II 21 CFR 892.1000/Procode: 90 MOS

Dear Ms. Taylor:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce pror to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act Include requirements for annual registration, listing of devices, good manufacturing and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class III (Premarket Aproval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Reculation (QS) for Medical Devices: General requiration (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements conceming your device in the Federal Register. Please note: this response to your premarket notfication submission you might have under sections 531 through 542 of the Act for devices under Radiation Control provisions, or other Federal laws or regulations.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4591. Additionally, for questions and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entititled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-5597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

David C. Schuller, M.D.

Daniel G. Schultz, M.D. Captain, USPHS Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure(s)

{4}------------------------------------------------

# STATEMENT OF INDICATIONS FOR USE

510(K) Number (if known):

Device name: Surgi-Vision Esophageal Stylet Coil

Indications for Use:

The Surgi-Vision Esophageal Stylet Coil is recommended for highresolution Magnetic Resonance Imaging of the human esophagus. The single use, disposable coil has been designed to be inserted in the esophagus of the patient during MRI scans in order to obtain improved image quality in the anatomical regions surrounding the esophagus. The unique coil shape facilitates orientation of the coil to the anatomy. This product is to be used with a 1.5T MRI machine.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use
(Per 21 CFR 801.109)
OR
Over-the-Counter Use

(Division Sign-Off)
Division Reproductive, Abdominal, ENT,
and Radiological Devices
J9944310

510(k) Number

Surgi-Vision, Inc. Esophageal Stylet Coil December 30, 1999

---

**Source:** [https://fda.innolitics.com/device/K994436](https://fda.innolitics.com/device/K994436)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
