← Product Code [MOS](/productcode/MOS) · K994345

# MAGNA 5000 PHASED ARRAY CTL SPINE COIL (K994345)

_Usa Instruments, Inc. · MOS · Feb 14, 2000 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K994345

## Device Facts

- **Applicant:** Usa Instruments, Inc.
- **Product Code:** [MOS](/productcode/MOS.md)
- **Decision Date:** Feb 14, 2000
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1000
- **Device Class:** Class 2
- **Review Panel:** Radiology

## Indications for Use

The Magna 5000 Phased Array CTL Spine Coil is a receive-only phased array RF coil, used for obtaining diagnostic images of the spine region (cervical, thoracic and lumbar anatomy) in Magnetic Resonance Imaging Systems. The indications for use are the same as for standard MR Imaging. The Magna 5000 Phased Array CTL Spine Coil is designed for use with the Signa HFO/i 0.7T MRI scanner manufactured by GE Medical Systems.

## Device Story

Magna 5000 Phased Array CTL Spine Coil is a multi-element, receive-only RF coil for MRI systems. Device conforms to spine contours; features removable anterior section for cervical imaging and three bridges with coil elements for thoracic/lumbar imaging. Operates as passive accessory to GE Signa HFO/i 0.7T scanner; receives hydrogen NMR signals. Decoupling achieved via RF chokes and switching diodes to isolate elements during RF transmission, preventing RF burns and resonant loops. Used in clinical imaging environments by trained MRI technologists/radiologists. Output is raw signal data processed by the host MRI system into diagnostic images for physician interpretation. Benefits include enhanced signal-to-noise ratio and image quality for spine anatomy assessment.

## Clinical Evidence

No clinical data. Bench testing only.

## Technological Characteristics

Multi-element receive-only phased array coil. Materials: ABS/PVC plastic alloy, polyurethane, glass fiber reinforced polyester (flame retardant fiberglass), Naughahyde. Decoupling: RF chokes with switching diodes. Non-conductive housing. Standalone accessory for GE Signa HFO/i 0.7T MRI scanner.

## Regulatory Identification

A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).

## Special Controls

*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

## Predicate Devices

- Premier 7000 Phased Array C/T/L Spine Coil ([K980157](/device/K980157.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K994345
Page 1 of 2

## SUMMARY OF SAFETY AND EFFECTIVENESS

| 1. Device Name :                  | Magnetic Resonance Imaging Accessory                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
|-----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 2. Proprietary Name :             | Magna 5000 Phased Array CTL Spine Coil                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| 3. Classification :               | Class II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| 4. Establishment Registration #:  | 1529041                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| 5. Manufacture Facility Location: | USA Instruments, Inc<br>1515 Danner Drive<br>Aurora, Ohio 44202, USA<br>Telephone: 330-562-1000; Fax: 330-562-1422.                                                                                                                                                                                                                                                                                                                                                                                                                           |
| 6. Performance Standard:          | No applicable performance standards have been issued<br>under Section 514 of the Food, Drug and Cosmetic Act.                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| 7. Intended Use:                  | The Magna 5000 Phased Array CTL Spine Coil is a<br>receive-only phased array RF coil, used for obtaining<br>diagnostic images of the spine region (cervical, thoracic<br>and lumbar anatomy) in Magnetic Resonance Imaging<br>Systems. The indications for use are the same as for<br>standard MR Imaging. The Magna 5000 Phased Array<br>CTL Spine Coil is designed for use with the Signa HFO/i<br>0.7T MRI scanner manufactured by GE Medical Systems.                                                                                     |
| 8. Device Description:            | The Magna 5000 Phased Array CTL Spine Coil is a<br>multi-element phased array receive-only coil. The coil is<br>shaped to conform to the contours of the spine and has a<br>removable anterior section for imaging the anterior<br>cervical region. The coil also has three bridges consisting<br>of coil elements for enhanced imaging of the thoracic and<br>lumbar spine. The elements and associated circuitry are<br>enclosed in a housing made of plastic materials, which are<br>fire rated and have high impact and tensile strength. |

please turn over

.

{1}------------------------------------------------

K994345
Page 2 of 2

.

## 9. Safety and Effectiveness

| Magna 5000 Phased Array CTL Spine Coil<br>Product Features                                                                                                                                                         | Comparison to predicate device or other<br>510(k) cleared product                                               |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------|
| Intended Use: Imaging of the spine.                                                                                                                                                                                | -Similar to the Premier 7000 Phased Array<br>C/T/L Spine Coil manufacture by USA<br>Instruments, Inc. (K980157) |
| Indications for Use: Identical to routine MRI<br>imaging                                                                                                                                                           | -Similar to the Premier 7000 Phased Array<br>C/T/L Spine Coil manufacture by USA<br>Instruments, Inc. (K980157) |
| Coil Material: ABS/PVC Plastic alloy,<br>Polyurethane Plastic, Glass fiber reinforced<br>polyester (Flame retardant Fiberglass),<br>Naughahyde (fabric material)                                                   | -Similar to the Premier 7000 Phased Array<br>C/T/L Spine Coil manufacture by USA<br>Instruments, Inc. (K980157) |
| Coil Design: Receive-only phased array design                                                                                                                                                                      | -Similar to the Premier 7000 Phased Array<br>C/T/L Spine Coil manufacture by USA<br>Instruments, Inc. (K980157) |
| Decoupling: RF Chokes with Switching<br>Diodes                                                                                                                                                                     | -Similar to the Premier 7000 Phased Array<br>C/T/L Spine Coil manufacture by USA<br>Instruments, Inc. (K980157) |
| Prevention of RF Burns: Does not transmit<br>RF Power, Decoupling isolates the coil<br>elements from RF fields during RF<br>transmission, Coil elements and circuitry are<br>enclosed in a non-conductive housing. | -Similar to the Premier 7000 Phased Array<br>C/T/L Spine Coil manufacture by USA<br>Instruments, Inc. (K980157) |
| Radio Frequency Absorption: Coil is a<br>receive only coil and does not transmit RF<br>power                                                                                                                       | -Similar to the Premier 7000 Phased Array<br>C/T/L Spine Coil manufacture by USA<br>Instruments, Inc. (K980157) |
| Formation of Resonant Loops: Decoupling<br>isolates coil elements from RF fields during RF<br>transmission. Length of cable and stiffness<br>does not permit looping                                               | -Similar to the Premier 7000 Phased Array<br>C/T/L Spine Coil manufacture by USA<br>Instruments, Inc. (K980157) |

:

.

.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of three human profiles facing right, stacked on top of each other, with flowing lines suggesting movement or connection. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular fashion around the emblem.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

FEB 1 4 2000

Rony Thomas Manager, Regulatory Affairs USA Instruments, Inc. 1515 Danner Drive Aurora. Ohio 44202

Re:

Magna 5000 Phased Array CTL Spine Coil Dated: December 22, 1999 Received: December 23, 1999 Regulatory class: II 21 CFR 892.1000/Procode: 90 MOS

Dear Mr. Thomas:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been redassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

K994345

If vour device is classified (see above) into either class II (Special Controls) or dass III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Capt. Daniel G. Schultz, M.D.

Acting Director, Division of Reproductive. Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

K994345 510(k) Number (if known):

Device Name: Magna 5000 Phased Array CTL Spine Coil

Indications for Use: The Magna 5000 Phased Array CTL Spine Coil is designed to provide Magnetic Resonance Images of the spine anatomy. The Magna 5000 Phased Array CTL Spine Coil is designed for use with the GE Signa HFO/i 0.7T scanner.

> Anatomic Regions: Spine Nuclei Excited: Hydrogen

The indications for use are the same as for standard imaging:

The GE Signa HFO/i 0.7T system is indicated for use as an NMR device that produces images that: (1) correspond to the distribution of protons exhibiting NMR signal, (2) depend upon NMR parameters (proton density, spin lattice relaxation time T1, spin-spin relaxation time T2) and (3) display the soft tissue structure of the head and whole body. When interpreted by a trained physician, these images yield information that can be useful in the determination of a diagnosis.

(PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use
(Per 21 CFR 801.109) OR Over-The-Counter Use
(Optional Format 1-2-96)

(Division Sign-Off)
Division of Reproductive, Abdominal, ENT,
and Radiological Devices
510(k) Number K994345

---

**Source:** [https://fda.innolitics.com/device/K994345](https://fda.innolitics.com/device/K994345)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
